20.5559.19 - 124UD-20 - FGM 160 (With Dual Path) Ultrasonic Sensor Flowrig Calibration Certificate - Redacted

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

72.120.201.

C - FGM 160/160 Dual Path Ultrasonic Sensor


Flowrig Calibration Certificate

Fluenta Flare Gas Meter FGM 160/160 Dual Path


Ultrasonic Sensor Flowrig Calibration Certificate
Project:

Client:

Calibration
Error of the ultrasonic flare gas meter FGM 160 is determined by comparision
Method:
with 4-path ultrasonic flow meter with assured accuracy of 0.5%. Calibration is
carried out with air under atmospheric pressure at an ambient temperature
and humidity. Average volume flow for period of 3 min. is used for error
calculation.

Date of 29/07/2020
Calibration:

Traceability: Calibration results and raw data logs are stored on Fluenta server and is
availble upon customer's request

Parameters: FGM flow: Meter under test indicated actual volume flow in m3/h
Ref. flow: Reference Meter indicated actual volume flow in m3/h
%Error: Error calculated from reported volume flow in %
Press: Meter OUTLET static pressure in bar absolute
Temp: Meter OUTLET temperature, degrees Celsius

Performed by: Eng. Jacek Reglinski


29/07/2020
Gdańsk, Fluenta AS
Fluenta AS
Haraldsgate 90
N-5501 Haugesund
Norway
Phone: +47 21 02 19 27
info@fluenta.com

This certificate is issued under the provision that no liability is accepted and that the applicant gives warranty for each responsibility
against third parites

72.120.201 Page 1 of 2
72.120.201.C - FGM 160/160 Dual Path Ultrasonic Sensor
Flowrig Calibration Certificate

Fluenta Flare Gas Meter FGM 160/160 Dual Path


Ultrasonic Sensor Flowrig Calibration Certificate
Sensor S/N: 124U/D-20
Model : FGM 160
Serial Number: 2013-1478
Job: 20.5559.19
Data File: 124UD20.zip
Test gas : AIR Standard density= 1.225 kg/m3
at standard conditions of 15 C deg., and 1015.325 hPa
Qmax/Qmin: 5300 [m3/h] / 280 [m3/h]

Results before calibration

Pt. FGM flow Ref. flow %Error Press Temp


1 292.3995 295.3154 -0.9874 1.0016 23.5334
2 1158.2890 1175.5020 -1.4644 1.0012 23.5610
3 2807.5790 2884.7700 -2.6758 0.9994 23.6364
4 3840.4030 4009.7160 -4.2226 0.9976 23.7230
5 4964.4630 5125.6550 -3.1448 0.9953 23.8185

Verification results after calibration

Pt. FGM flow Ref. flow %Error Press Temp


1 3967.3920 3994.9390 -0.6895 0.9976 23.9183
2 5085.0830 5124.0970 -0.7614 0.9952 24.0026

Error is determined by:

Fluenta AS
Haraldsgate 90
N-5501 Haugesund
Norway
Phone: +47 21 02 19 27
info@fluenta.com

This certificate is issued under the provision that no liability is accepted and that the applicant gives warranty for each responsibility against
third parites

72.120.201 Page 2 of 2