Professional Documents
Culture Documents
HA5 User Manual
HA5 User Manual
HA5 User Manual
ENGLISH
User manual
HA5
Automated 5-part Hematology Analyzer
Operator’s Manual
HA5 software version: 1.0.241.0
Manual Issue date: November-2017
Manual Revision: 1.0
User manual
Biosystems
4
Table of contents
1. Introduction...............................................................................10
1.1. Who should read this manual..............................................................10
1.2. Symbols used in the manual.................................................................10
1.3. Intended Use........................................................................................10
1.4. Disclaimer............................................................................................10
1.5. Symbols on the outside of the analyzer................................................11
1.6. Technical contact..................................................................................11
1.7. Warranty..............................................................................................12
2. Measurement technology............................................................13
2.1. Laser based cell counting.....................................................................13
2.2. Impedance based cell counting.............................................................13
2.3. Interpreting scattergrams.....................................................................14
2.4. Interpreting histograms........................................................................14
2.5. Hematology parameters.......................................................................15
2.6. 5-part differentiation of White Blood Cells..........................................16
2.7. The measurement process.....................................................................16
2.8. Supported blood collection vials..........................................................17
3. Package contents.........................................................................17
3.1. Accessories...........................................................................................18
3.2. Identifying parts of the analyzer...........................................................19
4. Installing the analyzer.................................................................20
4.1. User serviceable parts...........................................................................20
4.2. Connecting peripheral devices.............................................................21
4.2.1. Keyboard (optional)...................................................................................... 21
4.2.2. Mouse (optional).......................................................................................... 21
4.2.3. Bar code scanner (optional)........................................................................... 21
4.2.4. USB Wi-Fi Dongle (optional)....................................................................... 21
4.2.5. Network (“Ethernet”)................................................................................... 21
4.2.6. Reagent pack................................................................................................. 22
4.2.7. Reagent bottles............................................................................................. 22
4.3. Analyzer packaging material................................................................23
4.3.1. Power............................................................................................................ 23
4.3.2. External power supply................................................................................... 23
5
User manual
Biosystems
6
7.2.4.5. Profiles .................................................................................................. 45
7.2.4.6. LIS......................................................................................................... 45
7.3. Setting up reagents...............................................................................46
7.3.1. Reagent: Pack................................................................................................ 46
7.3.2. Reagent: Bulk............................................................................................... 47
7.3.3. Network Settings.......................................................................................... 49
7.3.3.1. WiFi settings........................................................................................... 50
7.3.3.2. Wired network Settings............................................................................ 50
8. Daily operation...........................................................................51
8.1. Power on..............................................................................................51
8.2. Running samples.................................................................................52
8.2.1. Introducing a Sample.................................................................................... 52
8.3. Results..................................................................................................53
8.3.1. Short (Quick) result...................................................................................... 53
8.3.2. Full result (Full screen).................................................................................. 53
8.3.3. Flags............................................................................................................. 54
8.3.4. Technical flags............................................................................................... 55
8.4. Reports.................................................................................................59
8.4.1. PDF.............................................................................................................. 59
8.4.2. Printer (Thermal).......................................................................................... 59
8.4.3. Printer (PCL5).............................................................................................. 60
8.5. Power off..............................................................................................61
9. Calibration..................................................................................62
9.1. Target values........................................................................................62
9.2. Automatic calibration..........................................................................62
9.3. Manual calibration...............................................................................63
10. QC – Quality Control.................................................................64
10.1. Expected values....................................................................................64
10.2. Assay sheets..........................................................................................65
10.3. QC diagrams........................................................................................67
10.4. QC reports...........................................................................................67
11. Database functions......................................................................69
11.1. Browsing records..................................................................................69
11.2. Database context menu........................................................................69
11.3. Selection...............................................................................................70
11.4. Moving to specific date in database......................................................70
11.5. Managing data.....................................................................................71
11.6. Sending e-mail.....................................................................................71
7
User manual
12. Maintenance...............................................................................72
12.1. Cleaning the analyzer...........................................................................72
12.2. Daily maintenance...............................................................................72
12.3. Emergency procedures.........................................................................72
12.4. Reagent pack replacement....................................................................72
12.5. Software updates..................................................................................72
12.6. Management menu...............................................................................73
12.6.1. Maintenance................................................................................................. 73
12.6.2. Prime & Drain.............................................................................................. 73
12.6.3. Clean............................................................................................................ 74
12.6.4. Service.......................................................................................................... 74
12.7. Supervision..........................................................................................75
12.7.1. History......................................................................................................... 75
12.7.2. History details............................................................................................... 75
12.7.3. About............................................................................................................ 75
12.7.4. Status bar...................................................................................................... 77
13. Technical Specifications..............................................................78
14. Reagent consumption.................................................................79
15. Performance Data.......................................................................79
16. Error messages............................................................................81
16.1. Mechanical Errors................................................................................82
16.2. Operation related errors.......................................................................82
16.3. Electronic errors...................................................................................84
16.4. System (software) errors.......................................................................84
16.5. Troubleshooting...................................................................................86
16.5.1. Measurement problems................................................................................. 86
16.6. Electrical problems...............................................................................88
16.7. Mechanical problems...........................................................................88
16.8. Hydraulic problems..............................................................................89
16.9. Network problems................................................................................89
16.10. Printer problems..................................................................................89
16.11. Certificates, standards..........................................................................90
16.12. Declaration of Conformity...................................................................90
17. Appendix A – Signs of improper sample handling......................91
17.1. Pre-analytical factors............................................................................91
17.1.1. Distinct layers............................................................................................... 91
Biosystems
8
17.1.2. Large visible bubbles..................................................................................... 91
17.1.3. Pink plasma (hemolysis)................................................................................ 91
17.1.4. Clots, gel-like substance inside vial................................................................ 92
17.2. Post-analytical factors..........................................................................92
17.2.1. Elevated PLT, decreased RBC, RDWsd/cv increased..................................... 92
17.2.2. Elevated PLT, no PLT result, noise flag(s)...................................................... 92
17.2.3. Elevated MON, lowered NEU: . .................................................................. 93
9
User manual
1. Introduction
HA5 is a small, 60 tests/hour laser scatter and impedance based hematology analy-
zer for small to medium sized laboratory use, performing 5-population analysis of
human whole blood samples.
HA5 is easy to use, uses low amounts of reagent for determining 26 hematology
parameters.
HA5’s small footprint allows operating in space-constrained environments.
Symbol Meaning
WARNING
Risk of personal injury
BIOHAZARD
Risk of biological infection, contamination
LASER RADIATION
Risk of exposure to laser radiation
1.4. Disclaimer
The manufacturer reserves the right to:
• modify the contents of this manual without prior notice,
• change technology applied within the analyzer without prior notice,
• change technical specifications without prior notice.
Biosystems
10
The manufacturer does not warrant this manual to be 100% free of unintentional
errors.
Please note that this Manual may be revised without prior notification. The Ma-
nufacturer reserves the right to change the specifications of the product and/or the
contents of this manual as deemed necessary, without prior notice.
Pictures included in this manual may differ from the actual delivered product.
Performance and reliability are never influenced by minor visual differences between
this manual and the actual product.
BIOHAZARD
Risk of biological infection, contamination
CE mark
Address of manufacturer
Serial number
11
User manual
1.7. Warranty
Your HA5 comes with two years manufacturer’s warranty against workmanship
defects. Warranty claims must be made through your local representative.
Your supplier will only be able to repair and warrant operation of the analyzer as per
its specifications if all service actions are performed by qualified service personnel,
and the Warranty Label is intact.
Warranty is void if any of the below can be identified, or inoperability, malfunction
can be traced back to any of the events listed below:
• broken warranty seal;
• intentional damage and/or modification to the analyzer;
• improper use, use against instructions in this manual, against intended use;
• damage from intentional activities overriding built-in safety logic;
• natural disaster;
• uncertified power supply, or unapproved peripheral has been connected to
the analyzer;
• disassembly, repair attempt performed by unauthorized staff;
• damage rooted back to unreported shipping and/or installation induced
activities.
The manufacturer does not recommend and support modifying the operating
software of this hematology analyzer. The analyzer is able to log and track modi-
fications to the operating software.
• In the case, when the End User required or performed installation of any
software product not controlled by the manufacturer and would afterwards
experience erroneous or unreliable operation of the analyzer, or would expe-
rience erroneous results, then the manufacturer cannot accept any complaints
about operation, reliability or performance.
Exclusions: The below listed items are not covered by warranty unless they were
found damaged upon arrival to the End User. In such cases, proper complaint must
be filed along with a complaint registered with the forwarder/Distributor company:
• Batteries, external power cable
• Plastic cover elements (including transparent display cover), external reagent
tubes,
• Labels, printed material
• Internal piercing needle
2. Measurement technology
HA5 uses flow cytometry and volumetric impedance method to provide 26 hema-
tology parameters from 25µl of whole human blood samples. The measurement
cycle time is 60 seconds yielding 60 tests/hour throughput.
Biosystems
12
2.1. Laser based cell counting
Flow cytometry is a technology that simultaneously measures and then analyzes
multiple physical characteristics of single cells, as they flow in a fluid stream through
a beam of laser light. The properties measured include a particle’s relative size and
relative granularity.
These characteristics are determined using an optical-to-electronic coupling system
that records how the cell scatters incident laser light in different angles. Low angle
(few degrees) scattered light is proportional to the cell’s volume as it is scattered
from the surface of cell membrane; while high angle (few 10 degrees) scattered
light is proportional to the cell’s internal complexity as it is scattered from the
intra-cellular granules and nucleus.
13
User manual
Biosystems
14
2.5. Hematology parameters
15
User manual
Biosystems
16
wavelength) laser beam. The detector tube and amplifier collects scattered data
about each individual passing cell during the 15 seconds of measurement.
The second 1.2µl processed blood sample is moved into the mix chamber using
diluent, creating a dilution of approximately 1:1000. This mix dilution is then
forced through a 70 µm aperture with the help of a sheathing fluid, creating an
overall dilution of 1:20000 The electrical signal of each individual cell is measured
(through stainless steel and platinum electrodes and is collected for 8 seconds) for
RBC and PLT counts.
As a final step, the system empties the measuring elements and is ready to take
the consecutive sample.
WARNING
Vials with rubber stoppers (caps) are designed for a limited
number of penetrations with a needle. Using a closed vial more
than 5 times poses a risk of damaging the rubber cap and causing
liquid damage in the analyzer.
3. Package contents
HA5 comes in a double cardboard packaging. The outer box provides protection
against shipping damage to the analyzer’s internal box.
17
User manual
The internal box contains the analyzer and its accessories inside protective foams.
The analyzer comes with a protective film covering the scratch resistant polycar-
bonate cover elements.
Upon arrival, carefully check the contents of the package and look for visible
damages, even on the outer packaging box. File any obvious damage through the
shipping company to be eligible for compensation and support from your local
dealer and from the manufacturer.
Packing list:
• Analyzer outer packaging box
• Accessories box
• Power supply; Power cord
• Reagent connector
• Operator’s manual
• Shipping foams (top, bottom)
• Analyzer packaging bag
• HA5 analyzer
3.1. Accessories
external power supply
(SYS1443-6512-T3)
Biosystems
18
3.2. Identifying parts of the analyzer
1
2
3
6
7
9
10
19
User manual
1. Carefully remove HA5 from the shipping carton. Look for signs of damage,
such as cracks on the outer covers or loose/missing screws. If you find such
signs, please file a complaint to the shipping company to be eligible for com-
pensation and support from your local dealer and from the manufacturer.
Make sure you can find all the accessories listed in the packing list.
2. Prior to powering on HA5, allow the analyzer to reach room temperature
to avoid dew condensation. Sudden temperature changes may cause dew to
condensate on colder internal structures and can lead to damage of electronic
components.
3. Place the analyzer on a desktop bench and find a nearby grounded power outlet.
Please avoid power extension cords, use direct connection to the power outlet.
Always use the power supply packaged along with the analyzer.
4. Connect the power supply to the socket on the rear plate of the analyzer.
5. Connect the power cord to the power outlet.
6. Remove the protective foil from the analyzer’s front.
Attention
If you experience any error, like smoke, immediately disconnect the
power cord from the power outlet.
Use a fire extinguisher if necessary.
Biosystems
20
4.2. Connecting peripheral devices
HA5 offers 2 USB sockets for connecting external peripheral devices. To expand
USB connectivity options, you can connect any standard USB HUBs.
21
User manual
Diluent3 Diff
Lyse 3 Diff
WASTE
Each tube is equipped with a label indicating which bottle to connect it to. The
caps match the standard HA5 reagent bottles.
Biosystems
22
Attention
For best results and performance, place the reagents to the same
level (on the same desk) as the analyzer.
4.3.1. Power
23
User manual
4.3.4. Power on
The single push of the START button will power on the electronics and HA5 will
display the welcome screen and the main menu automatically.
When HA5 is turning on, the screen remains dark, only backlight illumination
can be seen at the edges of the screen. This is normal.
To completely and immediately power down HA5, remove the power cord plug
from the mains outlet or the power connector from the back of the analyzer.
Biosystems
24
4.5. REAGENT PRECAUTION
The reagents are considered chemical products.
INTENDED USE
Diluent 5 Diff is a buffered, stabilized and micro-filtered electrolyte solution for
automated dilution of human blood samples, quantitative and qualitative deter-
mination of erythrocytes (RBC), leukocytes (WBC) and the leukocyte subpo-
pulations, and the measurement of thrombocytes (PLT) concentration on HA5
haematology analyzers.
The diluent should only be used with BioSystems reagents. Erroneous results may
be obtained if the reagent is applied with third party reagents.
Please refer to the instrument’s User Manual for further information.
INGREDIENTS
Sodium chloride....................................< 1.6%
Buffers...................................................< 1.1%
Stabilizers...............................................< 1.0%
Preservatives...........................................< 0.7%
In ion-free water
25
User manual
HANDLING PRECAUTIONS
• Diluent 5 Diff is environmental friendly, azide-free reagent; does not contain
harmful ingredients.
• Avoid contact with eyes, skin and clothing.
• In case of eye or skin contact flush eyes with copious amounts of water for
several minutes or wash skin area with water.
• Keep the reagent container closed when not in use.
• Wear laboratory gloves when handling the reagent.
• All body fluid samples should be considered potentially infectious materials.
Treat all blood and other potentially infectious materials with appropriate
precautions. Use gloves, masks and gowns if blood exposure is anticipated.
• Please refer to the SDS associated with the reagent.
• Use Good Laboratory Practices (GLP) when handling the reagents.
REAGENT PREPARATION
This reagent is ready for use and can be applied straight from the container; no
special reagent preparation is necessary. Leave the reagent to reach the room
temperature.
REAGENT REPLACEMENT
• Person installing the reagent must be a trained laboratory professional.
• Remove the inlet cap from the reagent container to be replaced. Cut the sealing
foil on the new reagent using a sharp object to allow the tubes go through.
Connect the reagent inlet to the new reagent container.
• Be sure that the colour on each tube, reagent container label and connector
in the back of the instrument match.
• Avoid any dust or microbial contamination of the tubing and reagents. Do
not pour and mix the remains of a reagent from a container into other one.
• Prime thoroughly the new reagent and measure backgrounds according to the
instrument’s User Manual.
• When installing a new lot of reagent, verify calibration by running control ma-
terial; recalibrate the instrument if necessary as specified in your User Manual.
Biosystems
26
• Dispose of waste product, unused product and contaminated packaging in
compliance with local regulations.
INTENDED USE
Lyse 5 Diff lysing reagent is a stabilized and micro-filtered lysing agent for stro-
matolysis of erythrocytes (RBC), for quantitative determination of leukocytes
(WBC), leukocyte five-part differentiation (LYM, MID, NEU, EOS, BAS) and
hemoglobin (HGB) concentration measurement in human blood on HA5 hae-
matology analyzers.
The lysing reagent should only be used with BioSystems reagents. Erroneous results
may be obtained if the reagent is applied with third party reagents.
Please refer to the instrument’s User Manual for further information.
INGREDIENTS
Surfactants.............................................< 0.5%
Buffers...................................................< 1.8%
Stabilizers...............................................< 1.1%
Preservatives...........................................< 1.0%
In ion-free water
HANDLING PRECAUTIONS
• Lyse 5 Diff lysing reagent is environmental friendly, azide-free reagent; does
not contain harmful ingredients.
• Avoid contact with eyes, skin and clothing.
• In case of eye or skin contact flush eyes with copious amounts of water for
several minutes or wash skin area with water.
• Keep the reagent container closed when not in use.
• Wear laboratory gloves when handling the reagent.
• All body fluid samples should be considered potentially infectious materials.
Treat all blood and other potentially infectious materials with appropriate
precautions. Use gloves, masks and gowns if blood exposure is anticipated.
• Please refer to the SDS associated with the reagent.
27
User manual
REAGENT PREPARATION
This reagent is ready for use and can be applied straight from the container; no
special reagent preparation is necessary. Leave the reagent to reach the room
temperature.
REAGENT REPLACEMENT
• Person installing the reagent must be a trained laboratory professional.
• Remove the inlet cap from the reagent container to be replaced. Cut the sealing
foil on the new reagent using a sharp object to allow the tubes go through.
Connect the reagent inlet to the new reagent container.
• Be sure that the colour on each tube, reagent container label and connector
in the back of the instrument match.
• Avoid any dust or microbial contamination of the tubing and reagents. Do
not pour and mix the remains of a reagent from a container into other one.
• Prime thoroughly the new reagent and measure backgrounds according to the
instrument’s User Manual.
• When installing a new lot of reagent, verify calibration by running control ma-
terial; recalibrate the instrument if necessary as specified in your User Manual.
Biosystems
28
analyzers’ capillaries, tubing and chambers, removing blood component precipitates
and lipoprotein deposits on HA5 haematology analyzers.
The cleaning reagent should only be used with BioSystems reagents. Erroneous
results may be obtained if the reagent is applied with third party reagents.
Please refer to the instrument’s User Manual for further information.
INGREDIENTS
Surfactants.............................................< 1.1%
Buffers...................................................< 1.0%
Stabilizers...............................................< 0.9%
Preservatives...........................................< 0.8%
In ion-free water
HANDLING PRECAUTIONS
• Cleaning Solution 5 Diff reagent is environmental friendly, azide-free reagent;
does not contain harmful ingredients.
• Avoid contact with eyes, skin and clothing.
• In case of eye or skin contact flush eyes with copious amounts of water for
several minutes or wash skin area with water.
• Keep the reagent container closed when not in use.
• Wear laboratory gloves when handling the reagent.
• All body fluid samples should be considered potentially infectious materials.
Treat all blood and other potentially infectious materials with appropriate
precautions. Use gloves, masks and gowns if blood exposure is anticipated.
• Please refer to the SDS associated with the reagent.
• Use Good Laboratory Practices (GLP) when handling the reagents.
REAGENT PREPARATION
This reagent is ready for use and can be applied straight from the container; no
special reagent preparation is necessary. Leave the reagent to reach the room
temperature.
REAGENT REPLACEMENT
• Person installing the reagent must be a trained laboratory professional.
• Remove the inlet cap from the reagent container to be replaced. Cut the sealing
foil on the new reagent using a sharp object to allow the tubes go through.
Connect the reagent inlet to the new reagent container.
• Be sure that the colour on each tube, reagent container label and connector
in the back of the instrument match.
29
User manual
INTENDED USE
Reagent Pack 5 Diff is a set of reagents intended to be used on HA5 haematology
analyzers.
Reagent Pack 5 Diff reagents should not be replaced with third party reagents.
Erroneous results may be obtained if the third party reagent is applied.
Please refer to the instrument’s User Manual for further information.
COMPONENTS
60126 60128
Diluent 5 Diff 3400 mL 5000 mL
Lyse 5 Diff 600 mL 1000 mL
Cleaning Solution 5Diff 300 mL 300 mL
Biosystems
30
HANDLING PRECAUTIONS
• Reagents inside Reagent Pack 5 Diff are environmental friendly, azide-free
reagent; do not contain harmful ingredients.
• Avoid contact with eyes, skin and clothing.
• In case of eye or skin contact flush eyes with copious amounts of water for
several minutes or wash skin area with water.
• Keep the reagent container closed when not in use.
• Wear laboratory gloves when handling the reagent.
• All body fluid samples should be considered potentially infectious materials.
Treat all blood and other potentially infectious materials with appropriate
precautions. Use gloves, masks and gowns if blood exposure is anticipated.
• Please refer to the SDS associated with the reagent.
• Use Good Laboratory Practices (GLP) when handling the reagents.
REAGENT PREPARATION
This reagent pack is ready for use and can be applied straight from the container;
no special reagent preparation is necessary. Leave the reagent to reach the room
temperature.
REAGENT REPLACEMENT
• Person installing the reagent must be a trained laboratory professional.
• Remove the inlet cap from the reagent container to be replaced. Cut the sealing
foil on the new reagent using a sharp object to allow the tubes go through.
Connect the reagent inlet to the new reagent container.
• Be sure that the colour on each tube, reagent container label and connector
in the back of the instrument match.
• Avoid any dust or microbial contamination of the tubing and reagents. Do
not pour and mix the remains of a reagent from a container into other one.
• Prime thoroughly the new reagent and measure backgrounds according to the
instrument’s User Manual.
• When installing a new lot of reagent, verify calibration by running control ma-
terial; recalibrate the instrument if necessary as specified in your User Manual.
31
User manual
5. Blood samples
Proper handling of a hematology sample is of key importance for accurate,
dependable and reproducible results. Blood is a living tissue composed cellular
structures suspended in liquid plasma. The cellular structures are composed of cell
membranes with liquid and intracellular structures inside. Hematology analyzers
are designed to detect, calculate and classify these cellular bodies and their content.
As the cells are actually alive in the meaning of active chemical interaction with
their environment, keeping them close to their ideal environmental conditions is
important for correct analysis. Blood samples intended to be analyzed for hema-
tology purposes are sensitive to temperature, anticoagulant quantity and quality,
and mechanical effects.
Follow the guidelines below for optimal sample handling.
Biosystems
32
• Samples can be stored for up to 6 hours at 15 ºC - 30 ºC or up to 10 hours
after collection when refrigerated at 2 ºC - 8 °C.
• If samples need to be transported after collection, provide a cooled environ-
ment (2 ºC - 8 °C).
• Hematology blood samples must not freeze. Avoid direct contact of sample
containers and cooling agent (cold surface, ice pack, etc.). The frozen water
will damage the structures and cellular membranes and will produce false and/
or erroneous results.
33
User manual
6. Menu system
1
2
Biosystems
34
Individual Individual reagent containers, usage
6.3. Gestures
HA5’s touch screen user interface allows operating the analyzer with light taps
and swipe actions listed below.
All actions (touch and release) are acknowledged by a small circular symbol appea-
ring at the point where the user tapped the screen, and an audible “tick” is head
if sound was enabled.
Using the above basic actions, HA5 can interpret various combined actions, so-
called gestures.
6.4. Tap
Touching (tapping) an active area (icon, data input field) will activate the co-
rresponding function, or open a sub-menu. For data input fields, an on-screen
keyboard is displayed at the top or bottom quarter of the screen depending on the
position of the data input field.
35
User manual
6.5. Swipe
HA5 is able to display data that spans over several screens (pages). Such pages are
indicated with small circles in the center near the top half of the screen.
To switch (scroll) between such screens, you need to perform a sweeping gesture
by lightly touching the screen at any point of the screen showing data, and quickly
sweep your finger in the direction of the intended scrolling action (up, down, left,
right). The screen will be scrolled in the requested direction.
Biosystems
36
6.7. Data input
The on-screen keyboard allows changing symbols entered by changing the keys.
The keyboard can be changed by swiping left or right to reveal capital letters, small
letters and symbols.
SPACE adds a space at the cursor.
CLEAR deletes the entire field.
BACKSPACE erases the character left of the cursor.
37
User manual
Actual keyboard layout depends on the language of the user interface. Always the
standard keyboard layout is used to interpret keystrokes.
Symbols ---, -- and – buttons will have decreasing effect on first, second and the
third digit, respectively.
Symbols +, ++ and +++ buttons will have increasing effect on first, second and the
third digit, respectively.
[Options] Settings
Biosystems
38
[System] System administration utilities
39
User manual
Biosystems
40
7. Setup
HA5 comes preprogrammed with the operating software, and has most its options
set to standard values, values that suits most laboratories.
However, you can customize these settings to make HA5 fit into your daily routine
even better. Here is a brief list of options that you are recommended to adjust or
verify to make the use of the analyzer simple and convenient.
7.2.1. Appearance
[Page1]
You can disable the visual feedback of
touchscreen gestures.
You can enable/disable audible touch
feedback.
You can change the background image.
To use Custom wallpaper, use “Custom”
setting, and load your wallpaper by acces-
sing local menu and tapping the “Load
Wallpaper” icon.
41
User manual
HA5 will confirm the new wallpaper file, or will display an error message if the
image is not of the right size, format, or it cannot be found in the root folder of
the USB memory device.
7.2.2. Languages
[Page2]
Language of the user interface can be
changes. Changes have an immediate
effect.
7.2.3. Units
[Page3]
It also allows changing measurement units
for cell count (WBC, RBC, PLT),
HGB, HCT/PCT.
Biosystems
42
7.2.4. Functionality
This menu allows setting operation related parameters.
43
User manual
WARNING
Forcing an incorrect language to a printer may result in scrambled
printouts, and empty pages, even one single symbol per page!
7.2.4.4. Customization
[Page4]
Defining Comment 1-3 values will chan-
ge header of comment fields on printouts
and user interface as well.
QC charts include only 11 parameters
(WBC, LYM%, MON%, NEU%, EO%,
BAS%, RBC, HGB, MCV, PLT, MPV)
if Short QC parameters mode is selected.
Used Reagent volume: you can define the
volume (in Liters) of the external contai-
ner collecting waste liquids. When the free
volume drops below 0.3L, there will be a
Full container symbol in the upper right
corner of the screen.
If you set 0 as volume, HA5 will not
display a warning at all.
Biosystems
44
7.2.4.5. Profiles
[Page5]
Minimum and maximum values of nor-
mal ranges can be set for each measured
parameters. Measured values will be
verified against these limits.
HA5 offers 8 slots (Human, Male, Fema-
le, Profile 4-8). Swap between profiles by
swiping up/down.
All limits can be changed. You can define
custom names for Profile 4-8.
To backup/restore Profiles tap the “Load
Profiles” or “Save Profiles” icon by acces-
sing local menu.
7.2.4.6. LIS
[Page6]
HA5 is capable of transmitting reports
using HL7 v2.5 protocol to a host com-
puter.
This page allows setting HL7 server IP
address along with the communication
port.
Automatic transmission of results can be
enabled.
By enabling repeatation of Sample ID as
Patient ID the transmitting report will
contain standard HL7 PID field which
includes the sample identification string.
Keeping HL7 connection open after
transmission can be enabled.
HL7 message format is depends on
selected version. For more information
requests for HL7 transmission protocol
document.
This page also allows setting up Biosystems Access parameters. For details, please
contact your service personnel.
45
User manual
Biosystems
46
To add reagent packs to the system:
HA5 offers entering reagent pack data
using the onboard camera. Biosystems
provides reagent packs with QR codes.
To activate QR code reading sequence:
• tap the local menu icon
• tap the small camera Icon.
The screen will change, and you will see
a live image of what the camera “sees”.
The camera is located in the center of
the letter “o” in the HA5 text at the top
right corner of the display.
Hold the QR code in front of the analy-
zer, making sure that the code is not
curved or bent. The QR code should be
located in the center of the live image.
Use the ON/OFF button with low
ambient light and the screen will pro-
vide sufficient backlight for scanning
a QR code.
HA5 will acknowledge the QR code
with an audible tone. The data will be
entered as a new reagent pack. If you
scanned a code that already exists, Icon
will show a message.
To load codes from a USB flash drive,
use the “Load” icon.
47
User manual
Biosystems
48
7.3.3. Network Settings
You can make the most of HA5’s ser-
vices if you let it connect to a network
with access to the Internet so that it can
create and send reports in e-mails, trans-
mit data to your laboratory information
system (LIS) and update software when
necessary.
Network availability is displayed in the
top row of the screen.
When WiFi is available (a USB WiFi
adapter is connected) then a little signal
strength icon appears in the top row.
When Wired network is availabe, a
network plug icon is displayed.
Tapping on this icon brings up the list
of available wireless networks.
Tap on a network to access its details.
Tapping on the X in the row of a net-
work will remove it from the list of
networks.
49
User manual
Biosystems
50
8. Daily operation
Your daily routine will mostly consist
of powering on the analyzer, running
samples with unique sample ID’s and
creating reports.
It is good laboratory practice to run QC
material on a daily basis to ensure and
allow controlling the performance of the
analyzer on a day-to-day basis.
You will most probably create reports of
the samples run and occasionally you will
need to find earlier results of individual
patients.
All these tasks are outlined below to give
you an easy to follow guide and make
laboratory work routine and fun to do.
8.1. Power on
HA5 will mostly be waiting in stand-by mode.
Push the START button to wake up the analyzer,
the status ring on the top will change color.
The user interface will be displayed shortly, and the
analyzer is ready for work.
The fluidic system will always require initialization,
a programmed startup. This is indicated by a yellow
analyzer in measurement mode.
Tap the yellow icon to start the initializa-
tion, and will display FastBlank (system
check) results. HA5 will accept Fast
Blank values if the values are below the
following thresholds:
WBC < 0.2 103/µL
RBC < 0.05 106/µL
PLT < 40 103/µL
51
User manual
Avoid putting more than one barcode label onto the blood
collection tube to avoid the tube getting stuck in the sampling
mechanics.
HA5’s sampling mechanics is equipped with a cap detector.
Cap detector operation may be impaired by additional labels put
on the sample vial, and HA5 may refuse to run such tubes.
Tapping the analyzer symbol in the center will start the measurement.
HA5 will lower the sample vial into the sampling mechanics and will take a sample
with its built-in cap piercing needle.
Biosystems
52
When the sample has been taken, the vial will be returned to the
Operator.
Results will be available in a minute.
A new available result will be indicated by a small icon in the lower
right corner of the screen. Tapping this icon will take you to the
Results screen and the latest data will be displayed on the screen.
8.3. Results
HA5 can display results in two ways:
a short version of results during measurement in the lower half of the screen;
a full (detailed) view accessible in “Results” (database)
53
User manual
8.3.3. Flags
The screen below intentionally contains error flags.
Biosystems
54
A small red flag indicates that the system detected a condition that requires atten-
tion. Flags next to result values indicate that the numerical result is beyond actual
minimum or maximum accepted values (normal range). Flags with yellow text
indicate technical flags, messages related to the evaluation process of the sample.
Most of these flags require the sample to be rerun.
55
User manual
Biosystems
56
Flag name Type Flag meaning Letter
RBC Slice errors System RBC/PLT measurement s
encountered an internal
homogeneity error. Rerun the
sample. If the problem persists,
contact Service.
Unintelligible WBC The internally prepared W
histogram sample did not meet proper
measurement criteria; sample
flow in the flow cell was not
homogenous.
Rerun the sample.
If persists, run WBC
Debubbling. If it cannot be
resolved with repeated WBC
Debubbling, contact service, as
temperature stability or liquid
flow might have been influenced.
Unstable HGB WBC Improper internal H
homogenization of sample.
Rerun the sample. If
maintenance related cleaning
actions have been omitted,
please do them now. Run WBC
Debubbling
Unstable HGB WBC The HGB measurement signaled h
baseline an error. Probably some bubbles
got stuck in the flow cell. Run
WBC Debubbling.
Unstable Voltage WBC Internal liquid flow was A
disturbed, or temporary block
of the liquid flow was detected.
Rerun the sample.
Unstable Voltage RBC Internal liquid flow was a
disturbed, or temporary block
of the liquid flow was detected.
Rerun the sample.
WBC Noise WBC The flow cell detected an N
abnormal number of disturbing
signals. This may be the result
of air bubbles blocking the flow
cell. Run a Prime All function.
57
User manual
8.4. Reports
8.4.1. PDF
HA5 is able to send reports ready to be printed via e-mail or is able to save reports
to USB HD. Reports include all entered patient data, along with reported para-
meters, scattergram, histograms and additional notes, comments created using the
analyzer during or after analysis.
Biosystems
58
8.4.3. Printer (PCL5)
HA5 is able to send report to connected printer which is capable to interpret PCL5
printer language. When HA5 detects PCL5 compatible printer is connected proper
symbol will be shown on status bar.
59
User manual
Note:
HA5 supports a limited set of pr-
inter languages. The printer must
support one of the following lan-
guages:
PLC5
ESC/POS
Printers not supporting the above
printer languages will trigger a printer symbol with a question mark on the screen,
and will print scrambled reports, or a lot of empty pages.
HA5 can be forced to use ESC/POS language for unknown printers. Forcing a
specific language for a printer may result in empty pages or missing, unrecognized
characters.
Shutdown will prepare the system for a longer period of inactivity. The process
takes about a minute, after which the system will automatically power off, leaving
HA5 ready to be disconnected from the mains if necessary.
Biosystems
60
9. Calibration
HA5 as any laboratory analyzer can be calibrated. For this purpose, HA5 offers
two ways to perform this task. HA5 arrives to your laboratory factory calibrated.
Calibration is necessary if laboratory standards require it, or when maintenance
actions require you to do so. You may need to calibrate upon installing new reagents.
61
User manual
Insert the vial with the calibrator material into HA5’s sample opening and press
START.
HA5 will mix and run the sample and will calculate the mean values of the mea-
sured parameters. The mean value will be used to calculate the calibration factors
using the formula below
Biosystems
62
These expected values represent the standard to which your analyzer’s reported
values are compared.
It is recommended to use control material Control Blood 5 diff, developed spe-
cifically for HA5.
63
User manual
Biosystems
64
Assay sheet data can also be entered
manually.
The list of QC materials can be seen
with a manual entry point in the top
row.
Banks can be selected as targets for
measurement by ticking the box in
front of the specific row.
Banks can be deleted by ticking the box
in front of the specific row and going to
the local menu and selecting “DROP”.
Tapping the first (top) row will enter
the screen where manual definition of
parameters is possible.
65
User manual
10.3. QC diagrams
HA5 can represent QC data in two
views. QC data are only shown if there
actual QC measurements stored in the
analyzer’s memory.
One view lists data in a graph where va-
lues of parameters are displayed against
measurements in sequence (time).
You can initiate a measurement by ta-
pping the analyzer Icon on the screen.
New QC measurements will be displa-
yed on the chart.
10.4. QC reports
HA5 can print Levey-Jennings diagram via thermal and PCL5 printer as well. To
print diagram tap “Print” button via accessing local menu.
Biosystems
66
67
User manual
Results Human
Date Record Details
QC
Date Record Details
Calibration
Date Record Details
Fast Blank
Date Record Details
Swiping the screen in the left or right direction will switch pages displayed.
Swiping the screen up or down will scroll the view up or down: you can scroll
between dates, groups of records and individual records.
Biosystems
68
11.3. Selection
By tapping “Same Day” you can select records measured on the day of the actual
result.
By tapping “Same SID” you can select records matching same SampleID of the
actual result.
Tapping “Selection Start” will allow multi-selection of results. You will be taken
back to the database table view. You can select the END of the selection by ticking
the box in front of any sample. All samples between the “first” (where selection
started) and the “last” (the one you ticked) will be selected.
The number of selected records will be shown under the list of records. By (un)
ticking the box by the (e.g.) 16 of 273 will deselect ALL records.
69
User manual
print results via connected thermal or PCL5 printer
export results in TSF format to USB HD (Tab Separated File – ideal for external
processing)
e-mail results in PDF format
save reports in PDF format to USB HD
save reports in raw data format to USB HD
send results to LIS
drop selected records
CV% calculates CV data of selected results.
Biosystems
70
12. Maintenance
71
User manual
12.6.1. Maintenance
Maintenance menu allows running
processes related to maintaining reliable
operation:
Priming, draining reagents
Cleaning procedures
General
Biosystems
72
12.6.3. Clean
It is possible to run various cleaning
actions that help keeping the system in
good working conditions:
RBC cleaning
WBC debubbling
12.6.4. Service
This menu is password protected and it
is only available for and should be used
by authorized service personnel.
73
User manual
12.7. Supervision
12.7.1. History
History will list all user activity in a log,
grouped into activity types. You can
review Information, User action, Errors.
12.7.3. About
The About screen will display instrument specific information
• SW version,
Biosystems
74
• Product ID,
• Manufacturer data,
• Service contact to your local service
personnel.
• You can restart the setting wizard
• You can initiate Remote Manage-
ment (refer to your service person-
nel)
Page 3 allows accessing Software upgrade and sending or saving diagnostic files
containing information for the technical support.
75
User manual
1 2 3 4 5 6
The status bar contains symbols and shortcut to main menu. Instrument statuses
are including following informations (from left to right):
home shortcut to return to Main Menu
progress symbol when fluidics action is in progress (animated)
printer symbol when printer is connected
3 versions: PCL5, ESC/POS, Unknown
wireless network symbol when wireless USB dongle is connected
Ethernet connector symbol when wired network cable is connected
Available measurement count with messages:
Sample count based on actual reagent level
Expired
Biosystems
76
Inactive
Reagent sensor indicators, when a reagent sensor is disabled.
External waste container full symbol
77
User manual
Biosystems
78
Parameter CV% Conditions
WBC 3 2.0 < WBC < 20.0
HGB 2 100 < HGB < 240
RBC 2 3.0 < RBC < 6.0
MCV 2 70 < MCV < 100
PLT 5 100 < PLT < 800
MPV 2 5 < MPV < 15
Accuracy (when compared to Abbott Cell-Dyn 3700)
Parameter Carryover: %
WBC <1%
RBC <1%
HGB <1%
PLT <1%
Display range
79
User manual
Values measured below or beyond the limits listed above will be marked with
< or > symbols and the actual limit itself. E.g. PLT 1200 will be displayed as: “PLT
1000<” accompanied by a linearity range flag.
Note (1): display precision is 1 decimal digit for diff absolute counts
Note (2): If WBC is below 0.1, the diff values will NOT be displayed.
0.1 < WBC < 1.0 the diff values get an * mark
Biosystems
80
16.1. Mechanical Errors
Error message Reason for error Remedy
Shear Valve error. An error Mechanical Error Rerun operation. If
occurred while testing persists, contact Service.
Shear Valve backlash.
An error occurred while Pump not operating, Rerun operation. If
attempting to use pump. pump jammed. persists, contact Service.
Sampling Rotor error. Mechanical error. Rerun operation. If
An error occurred persists, contact Service.
while attempting to use Check whether closed
Sampling Rotor. mode sampling opening
is obstructed.
Sampling Rotor error. Mechanical error, Rerun operation. If
An error occurred sensor induced. persists, contact Service.
while attempting to
use Sampling Rotor
(Sampling Rotor invalid
position).
Valve error. An error Valve controller Rerun operation. If
occurred while attempting Error persists, contact Service.
to set valve.
Shear valve recovery The shear valve had Confirm, and run
problems moving, initialization.
but the system
resolved the error
81
User manual
Biosystems
82
16.3. Electronic errors
Error message Reason for error Remedy
Camera error. An error The camera could Power off and on again.
occurred while activating not be detected. If the problem persists,
camera. use electronic files to load
data.
Contact service to
schedule a visit.
Touch screen does not Faulty touch screen Connect an external
react to gestures mouse for temporary
operation.
Contact Service
83
User manual
Biosystems
84
Error message Reason for error Remedy
System error. An A key system element Contact Tech support
error occurred while is missing or is for guidance.
attempting to initiate malfunctioning.
system (unspecified
error).
Reagent detection error. A key system element Contact Tech support
An error occurred while is missing or is for guidance.
attempting to detect malfunctioning.
Lyse.
Reagent detection error. A key system element Contact Tech support
An error occurred while is missing or is for guidance.
attempting to detect malfunctioning.
System Solution.
Reagent system error. A key system element Contact Tech support
An error occurred while is missing or is for guidance.
attempting to activate malfunctioning.
reagent (expired).
System error. An A key system element Contact Tech support
error occurred while is missing or is for guidance.
attempting to modify malfunctioning.
reagent’s tracking level.
System error. An A key system element Contact Tech support
error occurred while is missing or is for guidance.
attempting to query malfunctioning.
active reagents (inactive
or expired reagents).
Reagent detection error. A key system element Contact Tech support
An error occurred while is missing or is for guidance.
attempting to read malfunctioning.
reagent sensor data.
Firmware Error A key system element Contact Tech support
is missing or is for guidance.
malfunctioning.
Remote Management IP address of Access Contact Service for
error Server is incorrect, proper setup
or a Firewall blocks
connection
16.5. Troubleshooting
85
User manual
Biosystems
86
16.6. Electrical problems
Symptom Possible cause Remedy
Instrument cannot be Power is not connected Connect power
powered on source, verify
connections
Power voltage is too high, Replace / check power
or too low supply (is LED on?)
Fuse is blown inside Contact Service
analyzer
Indicator on top/front Power voltage is too high/ Replace power supply
of the analyzer glows low
in RED or BLUE color
BEFORE power ON
Instrument powers Power system fluctuations Try to start the
down during operation analyzer again. Use a
UPS.
Contact Service
Software error message Corrupted SW system. Contact Service
87
User manual
Biosystems
88
Symptom Possible cause Remedy
Printer connected, but no Printer is not Test the printer and
printer Icon in top row of supported, USB cable USB cable with a PC.
screen is wrong
Printer is OFF Restart HA5.Make
sure to use the same
grounded electrical
outlet for HA5 and
peripherals. Test USB
sockets with an external
Keyboard or with a
USB stick (and try to
save diagnostic files to
the USB)
89
User manual
Biosystems
90
17.1.4. Clots, gel-like substance inside vial
When using vials without pre-filled anticoagulant, there is a risk of adding unsa-
tisfactory amount of anticoagulant, or adding it late. In such cases an irreversible
clotting procedure will start very fast, or cannot be stopped in time, which will
bind cells in a fibrin net making hematology analysis impossible. Depending on
the efficacy of the coagulation process, either all cells will be grouped into a dark
red chunk, or a considerable ratio of the cells will be bound inside the fibrin net
taking the shape and form of a dark red gel-like substance. The plasma will have
no significant number of cells suspended anymore, and the chunks if aspirated by
the analyzer will very likely cause a technical problem by clogging delicate tubing
systems inside the analyzer.
Such samples must be rejected with an indication of sample induced technical
problem.
Normal PLT/RBC histogram Hemolysed PLT/RBC histogram
Such samples should be rested, analyzed for possible hemolysis, and rerun for
verification if possible.
91
User manual
Normal PLT/RBC histogram PLT/RBC histogram with micro-
bubble interference
Biosystems
92
Remedy:
Leave the sample rest for 30 minutes and rerun it to obtain much better scatter
and WBC differential results.
Errata
This manual is under continuous revision and improvement. The sections listed
below are known to describe functions focusing on convenience of use that may
not yet be available in your analyzer.
Missing descriptions or functions do not impair the operation of the analyzer.
References:
• 98/79 EN Directive
• pubmed.org
• eurlex.eu
93