Chlorine Dioxide

DR 900 Analytical Procedure AP-126/900

SPECIFICATIONS
Method: Spectrophotometric (DR 900)
Range: 0.04 to 5.00 mg/L as ClO2
Testing Time: 1 minute

DESCRIPTION
The following test method provides a direct means
to measure the concentration of Chlorine Dioxide
present in your system.
Chlorine dioxide reacts with DPD (N, N-diethyl-
p-phenylenediamine) to the extent of one-fifth of
its total available chlorine content corresponding
to reduction of chlorine dioxide to chlorite to form
a pink color. The color intensity is proportional
to the ClO2 in the sample. Chlorine interference is
eliminated by adding glycine, which converts free
chlorine to chloroaminoacetic acid, but has no effect Nalco DR 900 Colorimeter
on chlorine dioxide at the test pH.
Accessories
ORDERING INFORMATION Description Part No.
Order all replacement parts and reagents by their Sample cells, round, 10-20-25 mL,
part numbers. To be sure this product is compatible with cap, 6/pkg 500-P2555.88
with your treatment program, contact your local Sample cells, round, 10 mL,
Nalco representative. Items marked with an “*” with cap, 6/pkg 500-P1323.88
are included with the kit or set. To place your Beaker, glass, 50 mL 500-P0119.88
order, please contact your local customer service DPD-F Chlorine, Powder Pillows,
department. 10 mL, 100/pkg 460-S0460A.87
Glycine reagent, 29 mL 460-S0310.70
REPLACEMENT PARTS AND REAGENTS Data transfer adapter 500-P0892A.88

Required Parts
Description Part No. SAFETY PRECAUTIONS
Nalco DR 900 Colorimeter 400-P0900.88 Refer to Material Safety Data Sheet (MSDS) for
Reagent Set, Chlorine Dioxide 460-P3122.88 specific reagent personal safety and disposal
information.

SUPPORT
If you have any questions regarding this procedure,
please contact your Nalco representative. In North
America you can also contact the Equipment Help
Desk at 630-305-CHEM.
INTERFERENCES

Interfering Substance Interference Levels and Treatments


Acidity Greater than 150 mg/L CaCO3. May not develop full color or color may fade instantly. Neutral-
ize to pH 6–7 with 1 N Sodium Hydroxide (460-S0279.71). Determine amount to be added
on separate sample aliquot, then add the same aount to the sample being tested. Correct for
volume addition (see Correcting for Volume Additions in General Information Section.)
Alkalinity Greater than 250 mg/L CaCO3. Color may not develop fully or may fade instantly. Neutral-
ize to pH 6–7 with 1 N Sulfuric Acid (460-FTGR0350.71). Determine amount to be added on
separate sample aliquot, then add the same amount to the sample being tested. Correct for
the volume addition (see Correcting for Volume Additions in General Information Section.)
Bromine, Br2 Interferes at all levels.
Chlorine, Cl2 May interfere at levels greater than 6 mg/L. Additional glycine may be able to compensate for
this interference.
Chloramines, organic May interfere.
Flocculating agents High levels of most flocculating agents can be tolerated. This tolerance is decreased if chlo-
rine is present. See the information about metals in this table. In the presence of 0.6 mg/L Cl2,
Al(SO4)3 (<500 mg/L) and FeCl2 (<200 mg/L) may be tolerated.
Hardness No effect at less than 1,000 mg/L as CaCO3.
Iodine, I2 Interferes at all levels.
Manganese, oxidized Oxidized manganese interferes at all levels. Oxidized chromium interferes at levels
(Mn4+, Mn7+) or greater than 2 mg/L. To remove the interferences:
Chromium, oxidized (Cr6+) 1. Adjust sample pH to 6–7.
2. Add 3 drops Potassium Iodide (30 g/L) (001-H01252.88) to a 25-mL sample.
3. Mix and wait one minute.
4. Add 3 drops Sodium Arsenite* (5 g/L) (001-H01253.88) and mix.
5. Analyze 10 mL of the treated sample as described in the procedure.
Subtract the result of this test from the original analysis to obtain the correct chlorine dioxide
concentration.
Metals Various metals may interfere by combining with the glycine needed to remove the chlorine
interference. Metal interference is limited except when chlorine is present. In the presence of
0.6 mg/L Cl2, both copper (>10 mg/L) and nickel (>50 mg/L) interfere. Other metals may also
interfere, depending on their ability to prevent glycine from reacting with any Cl2 in the sample.
It may be necessary to add more glycine to overcome this interference.
Monochloramine Causes a gradual drift to higher readings. When read within 1 minute after reagent addition,
3 mg/L monochloramine causes less than a 0.1 mg/L ClO2 increase in the reading.
Ozone Interferes at levels greater than 1.5 mg/L.
Peroxides May interfere.
Extreme sample pH Adjust to pH 6–7. (See Interferences in General Information Section.)
Highly buffered samples Adjust to pH 6–7. (See Interferences in General Information Section.)
*Samples treated with sodium arsenite for interferences will be hazardous waste as regulated by Federal RCRA for arsenic (D004).
*See Form ADV-652 for further information on proper disposal of these materials
CHLORINE DIOXIDE DR 900 PROCEDURE

Start Zero

1 2 3 4
Start Program Fill a sample cell with 10 mL Place the blank into the Press: ZERO
of sample (the blank). cell holder. Tightly cover
76 Chlor Diox DPD The cursor will move to the
the sample cell with the
right, then the display will
Note: Samples must be instrument cap.
show:
analyzed immediately and
cannot be preserved for Note: For best results, run 0.00 mg/L ClO2
later analysis. a reagent blank using
deionized water as the
sample. Subtract the blank Note: If Reagent Blank
Note: Wipe off any liquid value from the sample Correction is on, the
or fingerprints before reading to obtain the final display may flash “limit”.
inserting the sample cell result. See Reagent Blank See Section 1 of the
into the instrument. Correction in Section 1 of the DR 900 Procedures
DR 900 Procedures Manual. Manual.

Read

5 6 7 8
Add four drops of Glycine Add the contents of one Allow 30 seconds for Press: READ
Reagent to the sample cell. DPD Free Chlorine Powder undissolved powder to settle.
The cursor will move to
Swirl to mix. Pillow to the sample cell (the Place the prepared sample
the right, then the result in
prepared sample). Cap the into the cell holder. Tightly
mg/L chlorine dioxide will be
cell and swirl to mix. cover the sample cell with
displayed.
the instrument cap.
Note: A pink color will Note: Wipe off any liquid Note: If the sample
develop if free chlorine or fingerprints before temporarily turns yellow
dioxide is present. inserting the sample cell after reagent addition, or
into the instrument. the display flashes “limit”,
Note: Perform step 9 it is due to high chlorine
within one minute of dioxide levels. Dilute a
reagent aeddition. fresh sample with chlorine
dioxide-free water and
repeat the test. A slight
loss of chlorine dioxide
may occur during dilution.
Multiply the result by the
dilution factor.
Nalco, an Ecolab Company
North America: Headquarters – 1601 West Diehl Road • Naperville, Illinois 60563 • USA
Nalco Champion – 7705 Highway 90-A • Sugar Land, Texas 77487 • USA
Europe: Richtistrasse 7 • 8304 Wallisellen • Switzerland
Asia Pacific: 2 International Business Park • #02-20 The Strategy Tower 2 • Singapore 609930
Latin America: Av. das Nações Unidas 17.891 • 6° Andar 04795-100 • São Paulo • SP • Brazil
www.nalco.com

Ecolab, NALCO and the logos are trademarks of Ecolab USA Inc.
©2013 Ecolab USA Inc. All Rights Reserved 8/13

You might also like