Professional Documents
Culture Documents
XR115
XR115
XR115
5694013-1EN
Revision 4.0
US English
© 2020 General Electric Company
All rights reserved.
Legal Notices
GE EXCLUSIVE USE ONLY
Restricted To General Electric Healthcare (GEHC) This Material is the property of GEHC and contains
information that is restricted to use by GEHC Manufacturing, Engineering and Service personnel. Any
reproduction in whole or in part is strictly prohibited. Neither this material nor any information
contained herein shall be disclosed to anyone outside of GEHC Manufacturing, Engineering and Service
personnel. If you are not a GE Healthcare Employee, you must exit this material now.
Trademarks
Brivo XR115 TM is a trademark of GE Healthcare. All other products and their name brands are
trademarks of their respective holders.
Copyrights
All material copyright © 2020 by General Electric Company, Inc. All rights reserved. The material
presented and contained herein may not be reproduced in any form or manner, without the written
permission of General Electric Company, Inc.
DANGER
Danger is used to identify conditions or actions for which a specific hazard is known to
exist which will cause severe personal injury, death, or substantial property damage if the
instructions are ignored.
WARNING
Warning is used to identify conditions or actions for which a specific hazard is known to
exist which may cause severe personal injury, death, or substantial property damage if the
instructions are ignored.
CAUTION
Caution is used to identify conditions or actions for which a potential hazard may exist
which will or can cause minor personal injury or property damage if the instructions are
ignored.
NOTE
A Note provides additional information that may emphasize certain information regarding
special tools or techniques, items to check before proceeding or factors to consider about a
concept or task.
WARNING
Safe equipment performance and operation requires that personnel servicing the
equipment be fully trained on servicing medical X-ray apparatus. General Electric
Company, GE Healthcare, and its associates, maintain a world-wide organization that
provides skilled X-ray service.
WARNING
X-RAY EQUIPMENT IS DANGEROUS TO BOTH PATIENT AND OPERATOR UNLESS
MEASURES OF PROTECTION IS STRICTLY OBSERVED.
Though this equipment is built to the highest standards of electrical and mechanical
safety, the useful x-ray beam becomes a source of danger in the hands of the
unauthorized or unqualified operator. Excessive exposure to x-radiation causes damage to
human tissue.
Therefore, adequate precautions must be taken to prevent unauthorized or unqualified
persons from operating this equipment or exposing themselves or others to its radiation.
It is important that everyone having anything to do with x-radiation be fully acquainted
with the local regulation recommendations. For India refer to Atomic Energy Regulatory
Board (AERB), India regulations and take adequate steps to insure protection against
injury.
DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent, have notation
“damage in shipment” written on all copies of the freight or express bill before delivery is accepted or
“signed for” by a General Electric representative or a hospital receiving agent. Whether noted or
concealed, damage MUST be reported to the carrier immediately and local GE Healthcare.
CERTIFIED ELECTRICAL CONTRACTOR STATEMENT
All electrical installations that are preliminary to positioning of the equipment at the site prepared for
the equipment should be performed by licensed electrical contractors. In addition, electrical feeds into
the Power Distribution Unit should be performed by licensed electrical contractors. Other connections
between pieces of electrical equipment, calibrations and testing should be performed by qualified GE
Healthcare personnel. The products involved (and the accompanying electrical installations) are highly
sophisticated, and special engineering competence is required. In performing all electrical work on these
products, GE will use its own specially trained field engineers. All of GE’s electrical work on these
products will comply with the requirements of the applicable electrical codes. The purchaser of GE
equipment should only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party
service companies with equivalent training, or licensed electricians) to perform electrical servicing on
the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.
Revision History
Revision Date Reason for change
1.0 24-Jul-2020 Initial draft
2.0 10-Sep-2020 Chapter 4, added Section 4.7 Electrical
Safety Tests Using FLUKE
ESA612 on page 40
3.0 17-Dec-2019 Chapter 3, Section 3.1, removed Shock
Watch term and images.
4.0 25-Sep-2020 Chapter 4, modified the Maximum Val-
ue from ≤ 500 to ≤ 2000 for leakage
current test in Section 4.7.2 Equip-
ment Leakage Current
Test on page 44
Chapter 4 Maintenance................................................................................................................. 27
4.1 Before You Begin............................................................................................................................................................... 27
4.2 Planned Maintenance Schedule ................................................................................................................................27
4.3 Service Tools....................................................................................................................................................................... 27
4.4 System Performance ......................................................................................................................................................28
4.5 Planned Maintenance Procedures............................................................................................................................28
4.5.1 Before You Begin - Safety Precautions ......................................................................................................28
4.5.2 PM Procedures...................................................................................................................................................... 28
4.6 Calibration........................................................................................................................................................................... 30
4.6.1 For Control Board Firmware Version 1.10................................................................................................31
4.6.2 For Control Board Firmware Version 1.20................................................................................................36
4.7 Electrical Safety Tests Using FLUKE ESA612........................................................................................................40
4.7.1 Protective Earth Resistance Test .................................................................................................................42
4.7.2 Equipment Leakage Current Test.................................................................................................................44
2. Turn OFF the system by pressing the Power OFF button in console.
3. Turn OFF the Rocker switch on the bottom of the system and remove the Power Cable from the
mains.
4. Remove the mains cable connected to the system and Remove the Mains Fuses ( on page 2). 1
5. Place the mains Cable and fuses in a LOTO container and apply your personal lock and tag.
Networks built into many integrated circuits provide some protection, but in many cases, the discharge
contains enough power to alter device parameters or melt silicon junctions.
Important ESD considerations when working on a mobile system
NOTE 1
Static Discharge Risk to Components
The output of the inverter is sent to the primary of the HV transformer. The High frequency AC is
then sent to a voltage rectifier and voltage doubler circuit.
• HV Rectifier and Voltage Doubler
The High frequency AC signal is rectified and kVp is applied across the X-ray tube.
• Control board
The control board establishes CAN (Controller Area Network) communication with the console. It
2
sends PWM (Pulse Width Modulated) signals to drive the inverter and filament drive circuit. These
System Overview
drive signals are adjusted according to the kVp and mAs selection made on the console. The kVp
and mAs feedback signals are received by the control board, based on this feedback the drive
signals can be adjusted by the control board, if needed. The calibration database resides in the
control board.
• Collimator
The collimator receives 24 VDC from the Power board to energize the 100 Watt collimator lamp.
• Console
The console receives 5 VDC from the PDU. When the console ON button is pressed a trigger is sent
to a relay in the PDU which in turn triggers the Contactor and provides 230 AC power to the
monoblock. Additionally, the console establishes CAN communication with the control board in the
monoblock.
System Overview
System Overview
Max current 16 Amps
Output Power 4 kW
System frequency 200 kHz
Boot time <10 secs
Reset time <5 secs
Duty cycle time 1:30
Technique selection 2-point technique (kV, mAs)
kV range 40 to 100 kV
mA range 10 to 100 mA
mAs range 0.1 to 250 mAs
Exposure time range 10 ms to 5 sec@230 VAC
kV step Jump step of 2 kV for 40-70 kV, then step of 5 till 100
mAs step Renard scale
Tube type Split Architecture, Stationary anode
Tube focal spot 1.8 mm
Collimator roatation +90 to -90 degrees
Collimator lamp lux 160 lux @ 1m SID
Cassette holder Up to 6 cassettes
Physical Specifications
Table 2-2 Physical Specifications
Symbol Description 2
This symbol on the equipment indicates General Warning Sign.
System Overview
Hand crushing hazard. This symbol indicates that serious injury to the hand may occur.
Symbol Description
The Mains ON LED glows green when the Mains wall socket and system rocker switch is switched ON.
The POWER ON button is used to turn ON the system.
The fault LED glows RED when there is an error in the system.
Pressing the Reset button will bring back the technique parameters to its previously set values.
Symbol Description
This symbol indicates the equipment is suitable for alternating current only.
Symbol Description
Protective earth (ground) identifies any terminal which is intended for connection of an external pro-
tective conductor to protect against electrical shock in case of a fault.
2
Table 2-7 Collimator
System Overview
Symbol Description
Collimator lamp button for indicating radiation field by using light.
Symbol Description
Used to indicate that the X-ray tube head is emitting X-rays.
Take adequate precautions to prevent the possibility of any persons carelessly, unwisely, or un-
knowingly exposing themselves or others to radiation.
System Overview
Manufacturer Column
System Overview
When not in the transport position, do Column
not drive the mobile system up on an
incline greater than 5 degrees
System Overview
Caution: X-rays Tube Head As-
sembly
Label Location
X-Ray Tube
System Overview
Table 2-14 Collimator Knob Control Label (x2)
System Overview
Table 2-16 Caution-Hot Surface Label (x2)
LH Bracket RH Bracket
Label Location
One logo on the Tube Head and two logos on either side of the Cabinet.
System Overview
2.7 Packing Symbol
Table 2-19 Packaging Labels
Installation
WARNING
All packages should be closely examined at time of delivery. If damage is apparent write
“Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is
accepted or “signed for” by a GE representative or Hospital receiving agent. Whether
noted or concealed, damage MUST be reported to the carrier immediately and local GE
Healthcare office.
3.2 Safety
Refer the safety instruction provided in the section Safety of this Manual.
CAUTION
Take necessary safety precautions while lifting or moving heavy objects during
installation.
Site Preparation
Installing Environment The unit is designed for indoor usage only.
• It should not be subject to direct sunlight for expanded duration.
• Place it away from sources of liquid ingress.
• If the X-Ray unit is stored below 10°C, time must be allowed for X-Ray unit to reach room
temperature (keeping in room temperature) before connecting it to the mains voltage.
Operating conditions:
Temperature: +10 °C to +40 °C
3
Max temp gradient of +10°C /hour (Operating)
Installation
Humudity: 20% to 80% (RH) non-condensing, 30%
Gradient/Hour (Operating)
Pressure: Operating : 700 to 1060 Hpa.
Non-operating: 500 to 1060 Hpa.
Electrical Outlets & Ensure that there is wall socket available near the installation location. The power
Requirements requirements are as mentioned in the User manual.
• The mains outlet should have a good Earth connection. Grounding of the system must be
checked before connecting the System.
• The mains cord shipped with the unit should not be extended further using extension
boxes or wires. In case extra wiring is required, ensure that this is done by a qualified
electrician.
• The mains outlet should be capable of supplying 230V ac and 16A of current. It should
have Fuse protection or provided with a circuit breaker of 16A (230V) or more.
• It is recommended to have an ELCB (Earth Leakage Circuit Breaker) or RCCB (Residual
Current Circuit Breaker) for protection against earth leakage.
Dimentional Drawing
Installation
3.4 Installation Procedure
Follow these steps for Installation:
1. Unpack the system, see Unpack system.
2. Power On the system
3. Perform Initial testing
4. Fill the Installation Checklist
Installation
5. Remove 4 screws on the ramp as shown in Figure 3-5 Shrink wrap on page 20 and position it on
the floor as shown in Figure 3-6 Aluminium foil on page 21.
Figure 3-5 Shrink wrap
Installation
6. Remove the shrink wrap inside the packing materials from the system and cut the aluminum foil as
shown in Figure 3-7 Cut foil on page 21.
Figure 3-7 Cut foil
7. Unscrew the fasteners and take out the metal clamp on the front side of the system and remove 12
screws on L –clamp as shown in Figure 3-8 Unscrew fasteners on page 22.
Installation
8. Release the plunger and raise the swivel arm to access the front metal brackets.
9. Remove the 4 M10x25mm bolts and take out the metal clamp in the front of the positioner as
shown in Figure 3-9 Remove bolts on page 22 and Figure 3-10 Remove clamp on page 23.
Figure 3-9 Remove bolts
Installation
CAUTION
Ensure without raising the swivel arm upwards, should not access the front metal
brackets. This may cause damage to the plunger in the parking slot.
10. Remove the 2 rear blocks by tilting the system on either sides as shown in Figure 3-11 Remove
blocks on page 23.
Figure 3-11 Remove blocks
11. Tilt the system slightly up towards the front and take out blocks below the base of the positioner as
shown in on page 24.
Installation
12. Release the brake and unload the system from the pallet by pulling it through the ramp as shown in
on page 24.
13. Visually inspect the system and components for any damage that might have occurred during
shipping.
• If no damage is observed, continue with the installation.
• If damage is found, contact GE Service or the installation specialist.
4. Turn the system ON by pressing the power ON button, on the console. The unit performs a few
diagnostic self-tests to ensure that the system is functioning correctly. (This process takes a few
seconds).
5. When the console display lights up with default technique values (KV and mAs) the unit is ready for
operation.
Installation
2. Check for the functioning of the parking brake by engaging the brake and checking for non-
movement of the system.
3. Check the Arm balancing in all positions i.e arm in horizontal position and in maximum extended
position.
4. Check for the functioning of the parking lock by moving the arm to parked position.
5. Check for free movement of the cassette holder assembly and its locking with magnetic catch.
Collimator
1. Open collimator shutters using the two knobs provided in collimator.
2. Press LAMP KEY to turn on the light field.
3. Adjust the light field on the area to be irradiated, using the 2 knobs provided in collimator.
4. The light field adjusted will be the X-ray field while taking X-ray. There is an automatic timer of 30
seconds for collimator light. This will switch off the light after 30 seconds.
5. If the light field adjustment is still required, press the LAMP KEY again. Press LAMP KEY to switch
off the collimator lamp, if it is on.
Exposure
1. When re-starting after long (more than 3 months) storage, take the unit through a Tube seasoning
procedure to ensure the tube head operates at its optimum. Following procedure explains the
seasoning process. Check the date on test report along with the system to confirm this. If the date
of installation is less than 3 months from date of testing, skip this step.
a. Set the kV and mAs to the minimum value (40 kV, 0.1 mAS).
b. Close the Collimator shutters.
c. Give 5 exposures.
d. Increase kV in steps of 10 kV and give 5 exposures.
2. Set the kV and mAs parameters to required values (Refer User Manual for Usage details).
3. Press and hold Exposure Handswitch to level I as shown in Figure 3-12 Exposure Hand
Switch on page 26 (second image).
Installation
4. This indicates system is getting prepared to take X-ray by preheating filament.
5. When the generator is ready to take X-ray, PREP/READY LED will turn to Green and buzzer will give
an audible single beep sound.
6. Press the Exposure Handswitch to level II as shown in Figure 3-12 Exposure Hand
Switch on page 26 (third image) to take X-Ray and hold it pressed till the buzzer stops. Buzzer and X-
RAY ON LED Green turns to Yellow and will be continuously ON throughout X-ray on time.
7. Feedback kV and mAs values will be displayed on the console display after exposures for few
seconds.
Consumables
1 Insulation Tape (if required for Trouble shooting)
2 Cable tie 4-inch, 5.9 inch & 11 inch & Cable tie mount 0.75-inch x 0.75 inch (if required)
Maintenance
Never use cleaners or solvents of any kind if you are uncertain of the nature of the
cleaning agent.
Ensure that no liquid seeps into the unit, so as to avoid short-circuiting or corroding the
electrical and electromechanical parts.
Personal caution should be used when removing any accumulating foreign Material/debris
with live system components.
4.5.2 PM Procedures
PM procedures are divided into the following groups:
1. Visual Inspection on page 28
2. Functional Checks on page 29
3. Cleaning on page 29
4. Calibration on page 30
Table 4-1 Visual Inspection
Maintenance
Inspect any damage to the column assembly, column cover, park-
ing lock & friction lock assembly.
Cables and Connection Inspect system cable for damage and connections for tightness. 5 minutes
Inspect all Monoblock cables connected inside the power board
for tightness.
Inspect EMI filter box and PDU connection cables are properly
tightness.
Inspect PDU to Monoblock power cables for damage and connec-
tion for tightness.
Perform the KVp and mAs accuracy test by using standard measurement tool. The accuracy parameters are mentioned 4
below, if any drift observed during the measurement then perform full system calibration.
Maintenance
Accuracy:
KVp - 40 to 100 +/- 10% of selected value
mAs - +/-10% +0.2mAs of selected value
4.6 Calibration
Calibration Procedure
To check the Control board version:
1. Turn on the system.
2. Press and hold the Reset button for 1 second and then release it to view the diagnostic
parameters.
3. Press the KVp increase button till the KVp display shows the letter CoN in the KVp.
4. Note the version of the control board.
5. Press the Reset button again to return to application mode.
Maintenance
RELATED LINKS
4.6.1 For Control Board Firmware Version 1.10 on page 31
4.6.2 For Control Board Firmware Version 1.20 on page 36
CAUTION
Make sure both the shutters are closed before starting the calibration.
3. Execute the steps given in Removal of Monoblock Skin Guard on page 69.
4. Execute the steps given in Removal of Input Panel Cup on page 69.
5. Execute the steps given in Removal of Tube Head Cover on page 71.
6. Execute the steps given in Removal of Console Unit.
7. Remove the console from the cabinet.
8. Remove the exposure handswitch J4 from the console.
9. Connect the handswitch and 'console to exposure handswitch cable' as shown in Figure
4-2 on page 32.
4
Figure 4-2
Maintenance
10. Connect the 'console to exposure handswitch cable' (4pin) connector of control board as shown in
Figure 4-3 on page 33.
Figure 4-3
Maintenance
11. Remove the tubehead to console communication cable J5 from control board as shown in Figure
4-4 on page 33.
Figure 4-4
12. Remove the FRC cable from the control board as shown in Figure 4-5 on page 34.
Figure 4-5
Maintenance
13. Connect the FRC cable from Calibration test jig shown in Figure 4-6 on page 34 to control board as
shown in Figure 4-7 on page 34.
Figure 4-6
Figure 4-7
Maintenance
1. Check if calibration jig is in standby mode (Default kV and mAs values as 40 kV, 0.1 mAs will be
displayed after boot).
2. Press the RESET key of calibration jig. The kV display will show as as "Uin" and the mAs display will
show the input AC voltage value.
3. Press the RESET key again after 2 seconds. The calibration jig shows 40 kV and 0.1 mAs. After 1
second press mAs up key followed by lamp key in quick succession within a time frame of 6
seconds. Immediately after this step, display shows "40" in kV and "10.0" in mAs.
4. Press the Exposure Handswitch to complete PREP and observe whether PREP/READY LED
(indication) glows Green. This will ensure that the system is now in calibration mode.
5. Press Exposure Handswitch to complete X-ray ON indicated by Yellow with beep.
6. Wait for 20 seconds.
7. Press the Exposure Handswitch half-way down (PREP command) and observe whether PREP/
READY LED (indication) glows Green. Then press the Exposure Handswitch full way down (X Ray
ON command) indicated by yellow and beep.
8. Repeat Step 6 and Step 7 above till display shows 40 kV, 0.1 mAs. When 40 kV and 0.1 mAs is seen
then calibration is complete. In case If there is any fault during calibration, turn off the unit and
repeat the above steps again.
Setting up the system after Calibration
Follow the below procedure after Calibration:
1. Switch OFF the unit.
2. Connect the tubehead to console communication cable from control board as shown in Figure
4-2 on page 32.
3. Connect the FRC cable available in the unit to the control board as shown in Figure 4-3 on page 33. 4
4. Remove the 4-way connector from control board as shown in Figure 4-4 on page 33.
Maintenance
5. Assemble the 'console to exposure handswitch cable' to the J2 connector of the console.
6. Assemble exposure handswitch to trolley.
7. Reassemble the console.
Figure 4-9
Maintenance
7. Connect the FRC cable from Calibration test jig as shown in Figure 4-10 on page 37 to control
board as shown in Figure 4-11 on page 37.
Figure 4-10
Figure 4-11
8. Remove the tubehead to console communication cable (6 pin) J5 and cable from exposure hand
switch to console (4 pin) J4 from console as showin in Figure 4-12 on page 38.
Figure 4-12
Maintenance
9. Switch ON the mains power.
10. Press the Power ON switch on the console.
11. E34 CAN error is displayed on console with continuous beep.
Calibration jig
Perform the following serirs of steps:
Figure 4-13
1. Make sure that the jig is in standby mode (Default kV and mAs values as 40 kV, 0.1 mAs will be
displayed after boot).
2. Press RESET key. The kV display will show as as "Uin" and the mAs display will show the input AC
Voltage value.
3. Press the RESET key again after 2 seconds. The calibration jig shall show 40 kV and 0.1 mAs. After 1
second Press mAs up key followed by LAMP key in quick succession within a time frame of 6
seconds. Immediately after this step, display should show "40" in kV and "10.0" in mAs.
4. Within a span of 10 seconds, put back the tubehead to console communication cable (6 way) to the
J5 connector of the console and short all 4 pins of the connector J4 of console board using the
shorting link (Refer Figure 4-14 on page 39 and Figure 4-15 on page 39) provided with the
calibration jig. The calibration starts.
Figure 4-14
Maintenance
Figure 4-15
• Remove the shorting link shown in Figure 4-9 on page 36 and put back the cable from exposure
hand switch to console (4 way) to the J4 of console as shown inFigure 4-10 on page 36.
• Reassemble the console back to the cabinet.
• Remove the FRC cable of the calibration jig from control board as shown in Figure 4-6.
• Connect the FRC cable available in the unit to the control board as shown in Figure 4-4.
• Execute the steps in Fixing tube head covers.
• Execute the steps in Fixing Input panel Cup.
• Execute the steps in Fixing of monoblock Skin Guard.
4
• Execute the steps in Initial Checks and Power ON.
Maintenance
• Execute the steps in Giving Exposure.
Test Standard Selections ANSI/AAMI ES-1, IEC62353, IEC60601-1, and AN/NZS 3551
Modes 2-Wire
Test Current >200mA ac
Range 0.000 Ω to 2.000 Ω
Accuracy (2% of reading + 0.015 Ω)
Maintenance
Modes* AC+DC (True-rms)
AC Only
DC Only
* For tests that do not use MAP voltage, AC+DC, AC ONLY, and DC ONLY modes are available for all leakages. MAP
voltages are available only in True-rms (shown as AC+DC)
Patient Load Selection AAMI ES1-1993, IEC 60601
Crest factor ≤3
Ranges 0.0 µA to 199.9 µA
200 µA to 1999 µA
2.00 mA to 10.00 mA
Condition
None
WARNING
Follow the LOTO procedure as per the procedure written in corresponding product document.
Service Procedure
Under the system periodic maintenances (PM) requirements, the FE must perform the Protective Earth
Resistance Test and Equipment Leakage Current Test, to ensure the electrical safety of the system. The
test procedures are described in this section.
The test procedure below contains the specifics tool setup / to use with the Fluke ESA612. Refer to
device’s user manual for setup details if instead using an equivalent testing device. 4
• 4.7.1 Protective Earth Resistance Test on page 42
Maintenance
• 4.7.2 Equipment Leakage Current Test on page 44
Protective Earth Resistance Test
NOTE
Before you start the test, ensure the system is in OFF position and the system input power
plug is connected to test socket on the side of Fluke ESA612 meter.
It is recommended to use Fluke ESA612 to set up the test conditions. The nominal AC Mains should be
220VAC to match the standard system plug. Refer to Table 1 for earth resistance test points of Brivo
XR115 systems 220VAC AC Mains (site-specific input source).
Maintenance
3. Connect one end of a test lead to the V/ Ω /A jack.
4. If you use an accessories probe, connect it to the other end of the test lead and put the probe tip
into the ∅/Null jack. If you use an alligator clip accessory, connect it to the other end of the test
lead, put the null post adapter in the ∅/Null jack, and clamp the alligator clip to the null post
adapter.
NOTE
The ∅/Null jack does not accept the test leads supplied with the Product.
CAUTION
To avoid electric shock, remove the null post adapter from the Null jack after a test
lead zero is performed. The Null jack becomes potentially hazardous during some of
the other test conditions.
5. Push ZERO LEADS button until the display shows 0.000 Ω.
6. Now connect one end of the test lead between the V/Ω/A jack of the test meter and other end to an
exposed conductive surface of the DUT ensuring that the input power plug of the DUT remains in
the test socket on the side on the Fluke ESA612. Refer to Table 4-17 Conductive surface test points
for BRIVO XR115 on page 44 for the list of conductive surfaces (accessible dead metal parts) of the
BRIVO XR115 system.
NOTE
When a current of 200mA from the current source with a no-load voltage not exceeding
24V at a frequency of 50/60Hz is applied between conductor terminals for 5 to 10
seconds, the voltage drops between the terminals. The resistance determined using the
current and voltage drop should not be greater than 0.3 Ohm (300 mΩ).
The resistance value will be displayed on the Fluke. Record the test results in Table 4-18 Results of
the Protective Earth Resistance Test for BRIVO XR115 on page 44.
7. Repeat table above steps 4-7 for each test point (conductive) mentioned in Table 4-17 Conductive
surface test points for BRIVO XR115 on page 44/ and record the test results for each test point in
Table 4-17 Conductive surface test points for BRIVO XR115 on page 44.
NOTE
Manufacturer recommendation is to perform Null test before testing each test point.
Table 4-17 Conductive surface test points for BRIVO XR115 4
Maintenance
Test Point Test Point Description
1 Input Panel Cup screws
2 Collimator handle
3 Monoblock cover screws
4 Swivel Arm screws
5 Foot Pedal screws
Maintenance
2. Press F2 button to select Instrument to select the instrument setup.
Figure 4-17 F2 button
Maintenance
4. Press up or down to scroll through the standard selections.
Figure 4-19 Up and Down buttons
5. When the desired standard is displayed, press F1 button Standard to confirm the standard.
Maintenance
6. Press F4 button to select Done to complete the standard setup.
7. Push µA to access the leakage current main menu.
8. Push the Enclosure/chassis/Direct Equipment button as per the regulatory requirement on the
Fluke.
Figure 4-21 Equipment selections
13. Repeat above steps with test conditions 2 and 3 as per Table 4-19 Test Conditions for 220VAC
AC on page 48.
Table 4-19 Test Conditions for 220VAC AC
Maintenance
1 Input Panel Cup screws
2 Collimator handle
3 Monoblock cover screws
4 Swivel Arm screws
5 Foot Pedal screws
NOTE
Refer to 4.7.3 Appendix 1 on page 50 for the locations of the test points in the system.
Test Results
Table 4-21 Results of the leakage current test under normal conditions for BRIVO XR115
Test Point (SEE APPENDIX Units Maximum Value Test Condition Measured Result
FOR TEST POINT LOCATIONS) Value (Pass/Fail)
Input Panel Cup Screws µA ≤ 2000 Condition 1
Condition 2
Condition 3
Collimator Handle µA ≤ 2000 Condition 1
Condition 2
Condition 3
Monoblock cover Screws µA ≤ 2000 Condition 1
Condition 2
Condition 3
Swivel Arm Screws µA ≤ 2000 Condition 1
Condition 2
Condition 3
Foot Pedal Screws µA ≤ 2000 Condition 1
Condition 2
Condition 3
In SFC, a single fault means of protection against electrical hazard is defective or faulty. To create Single
Fault Condition, refer to Table 4-22 SFC Test conditions for 220VAC AC Mains on page 49:
Table 4-22 SFC Test conditions for 220VAC AC Mains
To perform the Equipment Leakage Current Test under SFC, follow the steps listed below:
1. Set up test condition 1 listed in Table 4-22 SFC Test conditions for 220VAC AC Mains on page 49,
4
follow step 2 below.
Maintenance
2. Press the Polarity button to select Reversed on the Fluke ESA612. Refer to the user manual if
using an equivalent testing device.
3. Now connect one end of the test lead between the V/ Ω /A jack of the test meter and another end
to exposed conductive surface of the DUT ensuring that the input power plug of the DUT remains in
the test socket on the side on the Fluke ESA612.
4. The leakage value will be displayed. Record the test results in Table 4-23 Results of the leakage
current test under SFC conditions for BRIVO XR115 on page 49.
5. Set up test condition 2 listed in Table 4-22 SFC Test conditions for 220VAC AC Mains on page 49,
Press the Earth button to select Open on the Fluke ESA612. Refer to the user manual if using an
equivalent testing device.
6. Connect the test lead between the V/ Ω /A jack of the test meter and an exposed conductive
surface of the DUT ensuring that the input power plug of the DUT remains in the test socket on the
side on the Fluke ESA612. Record the test results in Table 4-23 Results of the leakage current test
under SFC conditions for BRIVO XR115 on page 49.
7. Repeat the above steps for each leakage current test point mentioned in Table 4-20 Test Points for
BRIVO XR115 on page 48.
Expected Test Results
Table 4-23 Results of the leakage current test under SFC conditions for BRIVO XR115
Test Point (SEE APPENDIX FOR Units Maximum Test Condition Measured Result
TEST POINT LOCATIONS) Value Value (Pass/Fail)
Input Panel Cup Screws µA ≤ 2000 Condition 1
Condition 2
Collimator Handle µA ≤ 2000 Condition 1
Condition 2
Monoblock Cover Screws µA ≤ 2000 Condition 1
Condition 2
Swivel Arm Screws µA ≤ 2000 Condition 1
Condition 2
Foot Pedal Screws µA ≤ 2000 Condition 1
Condition 2
4.7.3 Appendix 1
Test Point Location for Brivo XR115
This section contains the illustrations of the test points listed in Table 4-18 Results of the Protective
Earth Resistance Test for BRIVO XR115 on page 44.
Figure 4-22 Input Panel Cup Screws
Maintenance
Figure 4-23 Collimator Handle
Maintenance
Figure 4-26 Foot Pedal Screws
Condition
None
1. Move the swivel arm assembly to the top most or maximum position, this reduces the spring
tension for easier adjustment and provides access to the tie-rod nut.
2. Once the swivel arm assembly is moved to the top position, you can get access to the tie-rod nut at
the bottom of the base assembly rear end.
3. Use an open-end spanner of size 19 mm to tighten the tie-rod nut. Tighten the tie-rod nut with 2 or
3 rotations and then check for arm balancing as shown in Figure 5-1 on page 53.
4. Position the arm as shown in Figure 5-2 on page 53 and check if it balances.
5. Repeat the step of tightening the tie-rod nut until the arm get balanced.
Figure 5-1
Adjustment Procedures
Figure 5-2
Figure 5-3
Adjustment Procedures
NOTE
Access to the Tie-rod nut is possible only when the Swivel arm is in its top most position.
Condition
None
1. Remove the brake cover by loosening the fasteners using 2 mm Allen key.
2. Using a 14 mm open end spanner loosen the hexagonal nut as shown in Figure 5-4 on page 55.
Figure 5-4
Adjustment Procedures
3. Set the gap between the brake pad surface and the wheel surface to 2 mm.
4. To adjust the gap loosen or tighten the brake pad, once the hexagonal nut is loosened.
5. Once the gaps of 2 mm set, hold the brake pad with hand or position a 2 mm spacer between the
wheel surface and brake pad and tighten the hexagonal nut as shown in Figure 5-5 on page 55.
Figure 5-5
6. Apply Loctite 243 to the brake pad threading and tighten the hexagonal nut.
NOTE
Ensure that the gap of 2 mm is set in the brake pedal/cam assembly in disengaged
position and the bearing of the brake follower assembly is mating with the cam home
position.
NOTE
Ensure that the gap of 2 mm is set on both sides of the brake, as this is a synchronized
braking system.
Adjustment Procedures
Standard Tool Kit 1 - - -
Condition
None
1. Place a cassette on the object plane/table top and set the SID to 40"/ (1000mm).
2. Center the collimator tool on the cassette.
3. Collimate the beam such that it aligns with the outline of the collimator tool.
4. Set the parameters to 40kV/ 5mAs and make an exposure.
5. Develop the film and, measure the shift of collimator test tool edge and the darker x-ray edge.
6. If there is a shift in alignment, adjust the screw provided in the collimator Figure 5-6 on page 57
and repeat the above steps.
Figure 5-6
Adjustment Procedures
Functional Test
Test Setup
Connect the Ground Impedance test meter (Hioki/Equivalent) between any accessible dead metal parts
of the system as mentioned below, which is protectively earthed to PE terminal of the system. When a
current of 25A from the current source with a no-load voltage of 6v is passed between the conductor
terminals. The voltage drops between the terminal and the resistance determined from the current and
voltage drop should not be greater than 0.1 Ohm.
Test Procedure
1. Turn off the system and Remove the Mains Cable from the power outlet.
2. Connect the Ground Impedance test meter between Accessible dead metal parts of the system to
P.E (Protective Earth).
3. PE terminal to Monoblock head cover.
4. PE terminal to collimator knob/handle.
5. PE terminal to System metal storage box bolt.
6. PE terminal to Metal part of Front castor wheel.
Functional Test
Acceptance Creiteria
1. Increment of kV should be displayed on the console.
2. No increment should happen beyond 100 kV.
3. Decrement of kV should be displayed on the console.
4. No decrement should happen beyond 40 kV.
5. Increment of mAs should be in steps as per Renard scale. Each step should be displayed on the
console.
6. No increment should happen beyond 250 mAs.
7. Decrement of mAs should be in steps as per Renard scale. Each step should be displayed on the
console.
8. No decrement should happen beyond 0.1 mAs.
3. After exposure, the system should go to standby mode indicated by 'X-ray On' indicator going OFF.
Functional Test
Acceptance Criteria
1. Shift in radiation field ((X, X', Y & Y') Tolerance should be 2% of SID.
2. Difference in radiation and optical field (X+X' & Y+Y') -Tolerance should be 3% of SID.
3. Difference between sums of lengths and widths of optical and radiation fields ( X+X'+Y+Y') Tolerance
should be 4% of SID.
Arm movement found tight/hard Frictional lock is engaged Disengage the fricitional locak
Castors not moving properly Castor fasteners are loose. Tighten the castor fasteners.
Swivel arm parking lock not function- Wedge inside the column cover as- Remove the column cover and check
ing sembly is loose the parking pin actions.
Replace the parking pin if defective.
Important
Whenever during troubleshooting if PCB or HV tank is replaced then calibration to be done
mandatorily.
Calibration will ensure no drift to accuracy values mentioned.
Spare Parts
Sl No. Item
1 Monoblock assembly
2 Collimator
3 Conduit
4 Mains Input Cable
5 Rocker Switch
6 Front Castor
7 Inderxing Plunger (parking pin)
8 Hand Switch
9 Brake Shaft Assembly (inside the cover)
10 Rear Wheel
Console Unit
Figure 9-3 9
Spare Parts
9
Collimator
Spare Parts
Figure 9-5
Fixing Procedure
Fix the Input Panel 4 nos of M4x10 CSK screws using Star Screwdriver as shown in Figure
10-1 on page 69. Ensure the cut out is positioned to accommodate the cable routing.
Figure 10-2
10
Replacement Procedure
Figure 10-3
Figure 10-4
10
Replacement Procedure
Fixing of Monoblock Skin Guard
1. Fix the Skin Guards as shown in Figure 10-4 on page 71.
2. Fix the Skin Guard's M5x12 Socket Head CSK 4no's using 3mm bit / allen key as shown in Figure
10-3 on page 70.
3. Fix the Skin Guard's Rubber Cap 4 nos from the front as shown in Figure 10-2 on page 70.
Removal Procedure
1. Position the Swivel Arm such that the Tube Head is at a height suitable for easy cover removal.
2. Cut the Cable Ties on the top of Tube head covers as shown in Figure 10-5 on page 72.
Figure 10-5
10
Replacement Procedure
3. Remove the Monoblock skin Guards (See 10.2 Monoblock Skin Guard on page 69).
4. Remove the Input Panel Cup (See 10.1 Input Panel Cup on page 69).
5. Remove 2 No’s of M3x6 socket head screws from the strain relief using 2.5mm allen key as shown in
Figure 10-6 on page 72) Remove three screws and Disconnect the Power Cable as shown in Figure
10-7 on page 73).
Figure 10-6
Figure 10-7
10
Replacement Procedure
6. Remove 4 No's M3X10 button head using 2mm allen key from terminal block as shown in Figure
10-8 on page 73.
Figure 10-8
7. Remove the M4x10 Standoff's (4 nos) using 7mm nut driver as shown in Figure 10-9 on page 74.
Figure 10-9
10
Replacement Procedure
8. Remove wire to wire connector 4 No's M3X10 button head using 2mm allen key and disconnect the
8-way connector as in Figure 10-10 on page 74 and Figure 10-11 on page 75.
Figure 10-10
Figure 10-11
10
Replacement Procedure
9. Remove 4 nos M4x20 socket head screw from bottom Tube Head cover using 3 mm allen key as
shown in Figure 10-12 on page 75 and remove the bottom Cover.
Figure 10-12
10. Turn over the Tube head and remove 4 No’s of M4x10 CSK screws from the top cover using star
screw driver as shown in Figure 10-13 on page 76.
Figure 10-13
10
Replacement Procedure
11. After removing the screws lift the upper cover just enough to remove the Flat cable connection to
the Console.
12. Remove the Console board connector by pressing the locking tab and pull the cable connector (10-
way FRC cable) connected to J9 connector of console board as shown in Figure 10-14 on page 77
and take out the cover.
13. Remove the 4nos of M4x10 screws of remote console board inside the Tubehead cover as shown in
Figure 10-14 on page 77.
Figure 10-14
10
Replacement Procedure
Fixing Procedure
1. Before fixing top cover connect the console board using M4x10 screws and cable connector (10way
FRC cable) by pressing the connector in to the J9 connector as shown in Figure 10-14 on page 77.
2. Fix the top cover to the tube head assembly using 4 nos of M4x10 CSK screws using star screw
driver as shown in .
3. Take the bottom cover and fix in to the tube head assembly using 4 No’s of M4x20 socket head
screws as shown in Figure 10-13 on page 76 using 3 mm Allen key.
4. Fix the M4x10 Standoff's (4 nos) using 7 mm nut driver as shown in Figure 10-9 on page 74.
5. Place the PP with Indications (L N E) on it and fix the barrier strip with 4 No's M3X10 button head
2mm allen key from as shown in Figure 10-8 on page 73.
6. Fix the power cable on the barrier strip using star screw driver as shown in Figure 10-7 on page 73.
7. Fix the clamp 2 No’s of M3x6 socket head screws using 2.5mm allen key as shown in Figure
10-6 on page 72.
8. Execute the steps as mentioned in Fixing Input panel on page 69.
9. Connect the 8-way connector and fix the wire to wire connector 4 No's M3X10 button head using
2mm allen key as shown in Figure 10-10 on page 74, Figure 10-11 on page 75.
WARNING
After mounted the Tubehead cover in monoblock, paste the identification label and Rating
label on parallel to english label pasted depends on the country specific language. This
label is available as part of FRU box. Refer Appendix B on page 142 for language vs
applicable language.
Finalization
1. Perform the following Functional Tests 10
• Exposure Test on page 61
Replacement Procedure
10.4 Monoblock Assembly
Monoblock Assembly Replacement Procedure
1. Power Off the system and remove the Power cable from socket.
2. Raise the Arm and position the tube head facing ground and keep the arm parallel to ground.
3. Tie the horizontal arm to base assembly with a Plastic Belt as shown in Figure 10-15 on page 78.
4. Place a dead weight (10 to 20 Kg) on the edge of the Base assembly to balance the counterweight.
Figure 10-15
CAUTION
DROP HAZARD: The weight of the tube head assembly is about 25Kg. Take necessary 10
precaution to handle weight of the assembly. Make sure the Base assembly is
adequately compensated with dead weight to avoid toppling of the positioner
Replacement Procedure
assembly. It is recommended to lift the monoblock by two people for the safety
reason.
5. Cut the Cable Ties on the top of ID covers as shown in Figure 10-16 on page 79.
Figure 10-16
6. Remove the Input Panel Cup by removing 4 nos of M4x10 CSK screws using Star Screwdriver as in
Figure 10-17 on page 79.
Figure 10-17
7. Disconnect power cable from the Tube Head as shown in Figure 10-18 on page 80.
Figure 10-18
10
Replacement Procedure
8. Remove wire to wire connector 4 No's M3X10 button head using 2 mm allen key and disconnect the
8-way connector as in Figure 10-19 on page 80.
Figure 10-19
9. Remove the Nut cap, hexagonal Nut and Friction washer connected to the tube head bracket as
shown in Figure 10-20 on page 80.
Figure 10-20
10. Remove the Tube head assembly by sliding it out from the arm by holding the bracket as shown in
Figure 10-21 on page 81.
Figure 10-21
10
Replacement Procedure
Fixing the Monoblock Assembly
1. Position the New Tube head to the horizontal arm and position the whole in the arm to the bolt
projecting on the end. See Figure 10-21 on page 81.
2. Place the friction washer and tighten the hexagon nut to the bolt followed by the nut cap. See
Figure 10-20 on page 80.
3. Connect the 8-way connector and its cap. See Figure 10-19 on page 80.
4. Connect the Power Cable and clamp. Check for the L, N, E marking on the Cable and the connecting
point and match the connection. See Figure 10-18 on page 80.
5. Fix the input Panel Cup with 4 screws. Ensure the cut out is positioned to accommodate the cable
routing.
6. Install the cable ties to replace those removed in step 2. See Figure 10-16 on page 79.
7. Remove the dead weights from the base assembly and remove the belt tied to arrest the base and
arm.
CAUTION
After mounted monoblock in Arm, paste the identification label and Rating label on
parallel to English label pasted depends on the country specific language. This label is
available as part of FRU box. Refer Appendix B on page 142 for language vs applicable
language.
Finalization
Perform the following Functional Tests:
Replacement Procedure
• Exposure Test, see 7.8 Exposure Test on page 61
• Congruency of Radiation and Optical Filed Test, see 7.9 Congruency of Radiation and Optical Field
Test on page 62
• Complete the Regulatory test as recommended by local agency.
Figure 10-23
10
Replacement Procedure
Figure 10-24
Figure 10-25
10
Replacement Procedure
Fixing of Collimator
1. Assemble the lead cup as shwon in Figure 10-25 on page 84.
Figure 10-26
2. Take the collimator assembly shown in Figure 10-23 on page 83 and connect the bulb connector to
the power board as shown in Figure 10-22 on page 82.
3. Place the collimator assembly on the tank assembly and assemble the screws as shown in Figure
10-27 on page 85 and Figure 10-28 on page 85 and connect the ground cable.
Figure 10-27
10
Replacement Procedure
Figure 10-28
4. Fix the hot surface label on the side of collimator bulb box and label is available in FRU box.
WARNING
After mounted the collimator in monoblock, paste the hot surface label on the collimator
depends on the country specific language. This caution label is available as part of FRU
box. Refer Appendix B on page 142 for language vs applicable language.
Finalization
Perform the following Functional tests:
• 7.5 Ground Impedance Test on page 60
• 7.8 Exposure Test on page 61
10
10.6 Electrolytic Capacitor
Replacement Procedure
Removal Procedure
1. Execute the procedure in Monoblock Cover removal.
2. Measure the voltage across the capacitor with a DMM and ensure that the electrolytic capacitor is
fully discharged beofre removing it.
3. Disconnet the two DC Bus cables from the capacitor terminals (Blue='+'&Black='-') as showin in
Figure 10-29 on page 86 using flat screw driver.
Figure 10-29
4. Remove 2 clamp of the bracket arm using 3 mm allen key as shwon in Figure 10-30 on page 87.
Figure 10-30
10
Replacement Procedure
Fixing the Capacitor
1. Connect the DC Bus cables (Blue='+'&Black='-') from power board using flat screw driver as shwon
in Figure 10-29 on page 86.
2. Fix the capacitor in capacitor clamp using 3 mm allen key as shown in Figure 10-30 on page 87.
3. Execute the steps Fixing Procedure on page 77.
4. Execute the steps in Fixing Procedure on page 69.
5. Execute the steps in Fixing of Monoblock Skin Guard on page 71.
Finalization
Peform the following Functional tests:
• 7.8 Exposure Test on page 61
Removal Procedure
1. Perform LOTO.
2. Remove 2 no's of M4*10 button head screws from the outer venilation cover using 2.5 Allen key as
shown in Figure 10-31 on page 88 and remove the outer ventilation cover.
Figure 10-31
10
Replacement Procedure
3. Next remove 2 no's of M4*10 button head screws using 2.5 Allen key as shown in Figure
10-32 on page 88 and remove the inner ventilation cover.
Figure 10-32
4. Now remove the bulb carefully from connector as shown in Figure 10-33 on page 89.
Figure 10-33
10
Replacement Procedure
Fixing the Bulb
1. Fix the collimator bulb carefully to the connector as shown in Figure 10-33 on page 89.
2. Fix the inner ventilation cover with 2 nos of M4*10 button head screws as shown in Figure
10-32 on page 88 using 2.5 Allen key.
3. Fix the outer ventilation cover with 2 nos of M4*10 button head screws as shown in Figure
10-31 on page 88 using 2.5 Allen key.
Finalization
Perform the following Functional tests:
• 7.9 Congruency of Radiation and Optical Field Test on page 62
2. Remove the hex screw with a 4 mm Allen key and remove the dowel pins on the bottom of the
cassette holder by pushing it inside the cassette holder on both sides, see Figure
10-35 on page 90.
Figure 10-35
10
Replacement Procedure
3. Remove the cassette holder from the system.
4. Remove 4 nos of 4 mm hex screws with an Allen Key and Remove the black PVC sheets as in Figure
10-36 on page 90.
Figure 10-36
Fixing Procedure
1. Fix the Black PVC Sheets on the system with 4 mm hex screws (2 each) as in Figure
10-36 on page 90.
2. Insert the cassette holder into the system and secure the dowel pins into the bottom of the
cassette holder and fix it with 4mm hex screw on both sides as shown in Figure 10-35 on page 90.
3. Fix the stoppers on both sides of the cassette holder as shown in Figure 10-34 on page 89.
Figure 10-37
10
Replacement Procedure
3. Remove the Filter Unit output cable and ground Cables connected to the filter unit box as shown in
Figure 10-38 on page 91 and Figure 10-39 on page 92.
Figure 10-38
Figure 10-39
10
Replacement Procedure
4. Remove the black metal part of the filter unit and take out the whole filter unit assembly as shown
in Figure 10-40 on page 92.
Figure 10-40
Replacement Procedure
Removal Procedure
1. Execute procedure in Removal Procedure on page 89.
2. Remove 3 Heaxagon Nut with Integral Locak Washer as shown in Figure 10-41 on page 93.
Figure 10-41
3. Remove power cable and cable connected to console from PDU and ground wire to remove the PDU.
Fixing the PDU
1. Connect the Monoblock Power Cable line (L=brown) and Neutral (N=blue) to the PDB as shown in
Figure 10-42 on page 93. Connect the cable between PDU to Monoblock Assembly.
Figure 10-42
2. Mount the PDU to the positioner Assembly base using 3 Hexagon screws with Intergral Lock
Washer at the given locations as shown in Figure 10-41 on page 93.
Replacement Procedure
Perform the following Functional Tests:
• 7.5 Ground Impedance Test on page 60
• 7.8 Exposure Test on page 61
3. Remove the Supply Cable, communication Cable and Hand switch Cable connected to the Console
unit.
4. Remove 4 nuts from the corners of the console to Remove the console assembly as shown in Figure
10-44 on page 95.
Figure 10-44
10
Replacement Procedure
Console Unit Assembly Procedure
1. Place the Console board assembly and secure it with 4 nuts on the corners as shown in Figure
10-44 on page 95.
2. Connect the Communication Cable, supply cable and Hand switch Cables to the console board.
3. Fix the Metal storage box with 4 screws as shown in Figure 10-43 on page 94.
CAUTION
For other than English language follow the step 4 to 8, refer Appendix B on page 142
for language applicable countries.
4. Clean the console panel with dry cloth or cotton material to remove dust particle as shown in Figure
10-45 on page 95.
Figure 10-45
5. Remove applicable language label as mentioned in B.1 Appendix B on page 142 from the sheet
provided in FRU box as shown in Figure 10-46 on page 96
Figure 10-46
10
Replacement Procedure
6. Ensure the full label stick on top of the panel above the white line shown in Figure
10-47 on page 96 and should align the words on top the buttons as shown in Figure
10-48 on page 97.
Figure 10-47
Figure 10-48
10
Replacement Procedure
7. Press hard and rub on top of the label for 5 times with equal pressure to ensure the label properly
stick on the console as shown in Figure 10-49 on page 97.
Figure 10-49
8. Remove the white label support strip as shown in Figure 10-50 on page 98 and final view as
shown in Figure 10-51 on page 98.
Figure 10-50
10
Replacement Procedure
Figure 10-51
Finalization
Perform the following Functional Tests:
• 7.7 KV/mAs Increment/ Decrement on page 61
• 7.8 Exposure Test on page 61
Figure 10-52
10
Replacement Procedure
Fixing Procedure
Connect the Hand switch to the rear side of the system and tighten the screws of the connectors.
Finalization
Perform the following Functionla Tests:
• Exposure Tests (7.8 Exposure Test on page 61)
5. Remove all cable ties and remove the Cables together with the conduit.
Monoblock to Console and PDU Cable Assembly Procedure 10
1. Route the cables through the conduit.
Replacement Procedure
2. Insert the conduit into the positioner arm from the monoblock side and remove it out through the
opening at the positioner end.
3. Remove the Column cover by unscrewing 4 screws and keep the screws and cover in the kit trolley
as shown in Figure 10-53 on page 100.
Figure 10-53
4. Insert the conduit into the positioner column and remove it from the cabinet end.
5. Connect The Monoblock Power Cable and Communication Cables- refer procedures in Monoblock
cover replacement fixing and PDU fixing for cable terminations.
6. Secure the Cables with Cable ties.
7. Assemble the Cassette Holder, Column Cover and Input panel cup.
Finalization
Perform the following Functional Tests:
• 7.2 Tube Position Checks on page 59
• 7.5 Ground Impedance Test on page 60
• 7.8 Exposure Test on page 61
Replacement Procedure
• 7.8 Exposure Test on page 61
6. Remove the 10-way cable from the console board Figure 10-55 on page 102.
Figure 10-55
10
Replacement Procedure
Assembly Procedure
1. Connect the 10-way cable from the control board Figure 10-55 on page 102.
2. Connect the 10-way cable from the control board Figure 10-54 on page 101.
3. Install the tube head covers.
4. Install the input panel cup.
5. Install the monoblock skin guard.
Figure 10-56
10
Replacement Procedure
Fixing Procedure
1. Install the new plunger and tight it with 19 mm spanner by turning clockwise.
2. Park the monoblock and Check whether the parking lock latches and de-latches properly. If not
loosen the plunger slightly and recheck the function.
3. Remove the Plunger and apply lock-tight in the plunger threads and re fix the plunger in the proper
position.
Finalization
Perform the following Functional Tests:
• Parking Lock Test (7.1 Parking Lock Test on page 59)
Figure 10-57
10
Replacement Procedure
2. Remove the Handle, Metal Cover, and Friction Washer.
Fixing Procedure
1. Place the Friction Washer and Metal cover to the friction plate.
2. Fix the handle and tighten it by turning clockwise.
3. Adjust the tightness to enable required Friction to support the arm balance.
Finalization
Perform the following Functional Test:
• Swivel Arm Position Check (7.3 Swivel Arm Position Checks on page 59)
10.19 Conduit
Removal Procedure
See Removal of Monoblock Bracket on page 130.
Fixing Procedure
See Replacing Monoblock Bracket on page 130.
Finalization
Perform the following Functional Tests:
• Tube Position Check (7.2 Tube Position Checks on page 59)
• Exposure Test (7.8 Exposure Test on page 61)
Replacement Procedure
floor as shown in Figure 10-58 on page 105.
Figure 10-58
4. Remove the Screw, Washer, Covering Plate, Wheel, and Inner Covering Plate.
Fixing Procedure
1. Assemble the wheel with wheel covers on both sides and fix it with 6 mm Allen screw as shown in 10
Figure 10-59 on page 105.
Replacement Procedure
2. Engage the brake and remove the packing.
Finalization
Perform the following Functional Tests:
• Movement and Brake Tests (7.4 Movement and Brake Test on page 60)
Replacement Procedure
1. Open the cassette tray.
2. Remove the hex screws with 2 nos of 2.5 mm Allen keys on both sides as shown in Figure
10-61 on page 107.
Figure 10-61
Fixing Procedure
1. Place the magnetic catches and secure it with 2 nos of 2.5 mm hex screws as shown in Figure
10-61 on page 107.
2. Check for the proper closing of cassette holder Assembly.
Figure 10-62
10
Replacement Procedure
2. Remove 4 hex screws connected to the brake shaft assembly from inside (marked in Figure
10-63 on page 108) and takeout the brake shat assembly.
Figure 10-63
Fixing Procedure
1. Position the brake shaft assembly and Secure it with 4 hex screws from inside as marked in the
Figure 10-63 on page 108.
2. For brake alignment and checking follow procedure in Section 5.1.2 on page 54 in Chapter 5
Adjustment procedures.
3. Fix the brake Cover with 2 hex screws with a 2 mm Allen key as shown in Figure 10-62 on page 108.
Finalization
Perform the following Functional Test:
• Movement and Brake Test (7.4 Movement and Brake Test on page 60)
Replacement Procedure
1. Perform LOTO.
2. Remove the fuse holder shown in Figure 10-64 on page 109 and Figure 10-65 on page 109.
Figure 10-64
Figure 10-65
Replacement Procedure
Allen key as shown in Figure 10-66 on page 110.
Figure 10-66
7. Remove 2 nos M4x6 socket head screw shown in Figure 10-67 on page 110 using 3mm Allen key.
Figure 10-67
8. Remove clamp of the bracket arm using 3 mm Allen key as shown in Figure 10-68 on page 111.
Figure 10-68
10
Replacement Procedure
9. Take off the top ID cover as shown in Figure 10-69 on page 111 and Figure 10-70 on page 112.
Figure 10-69
Figure 10-70
10
Replacement Procedure
Figure 10-71
Fixing of Goniometer
1. Keep the goniometer shown in Figure 10-71 on page 112 face down as shown in Figure
10-70 on page 112.
2. Place the top ID cover as shown in Figure 10-69 on page 111.
3. Fix the clamp shown in Figure 10-67 on page 110 with 2 nos of M4x6 socket cap head screw and
tighten it by applying 2.5Nm torque using 3mm torque bit.
4. Fix 2 nos M4x8 socket head screw and M4 washers- 2 plain 1 spring washer each using 3mm allen
key as shown in Figure 10-66 on page 110.
5. Execute the steps in Fixing Procedure on page 77.
Replacement Procedure
9. Execute the steps in Giving Exposure.
7. Cut the cable tie 2 nos from cover mounting strips as shown in on page 113 and on page 113.
8. Cut the cable tie 2 nos from input panel as shown in on page 114 & on page 114.
10
Replacement Procedure
9. Disconnect the communication cable from control board as shown in on page 114.
10
Replacement Procedure
11. Disconnect the console cable from the control board as shown in on page 115.
12. Disconnect the receptacle from the collimator to the power board as shown in on page 116.
10
Replacement Procedure
13. Remove 6 nos M4*8 button head screws using 2.5 Allen key as shown in on page 116.
14. Remove 2 nos of M4x6 socket head screws mounted on tank using 3mm allen key as shown
on page 117.
10
Replacement Procedure
15. Remove the Input cable from J3 connector of Power board using flat screw driver as shown in
on page 117 and also remove ground cable (M4x6 socket head) using 3 mm allen key as shown in
on page 117.
10
Replacement Procedure
10
Replacement Procedure
18. Remove 1 no of M4x30 socket head screw from PFC inductor using 3 mm allen key as shown in
on page 119.
19. Remove 2 nos M3x25 button head screws from the fan fixed to the clamp using 2 mm allen key as
shown in on page 120.
10
Replacement Procedure
20. Remove the heat sink cover M3x6 button head screw using 2 mm allen key as shown in
on page 120.
21. Remove 8 nos of M4x6 socket head screws from power board using 3mm allen key shown in
on page 121
10
Replacement Procedure
22. Remove 4 nos M4x8 socket head screws from heat sink using 3 mm allen key shown in
on page 121.
23. Remove the Inverter cables (Blue&black) from J5 connector on the sealing board using star screw
driver as shown in on page 122.
10
Replacement Procedure
24. Remove the filament cable connector from the J6 of the power board which is connected to the J3
connector of sealing board as shown in on page 122.
25. If the board contains FRC cable, remove the cable from the J5 connector of the power board which
is connected to the J2 connector of the sealing board as shown in on page 123.
10
Replacement Procedure
26. If the board contains microfit cable, remove the cable from the J5 connector of the power board
which is connected to the J2 connector of the sealing board as shown in on page 123
27. Remove the collimator bulb power cable connector (2 way connector) by pressing the locking tab of
the J4 connector in the power board as shown in on page 124.
10
Replacement Procedure
Fixing the Board Set
1. Ensure that PP sheet is in original position.
2. Place the Power board on the tank assembly. Connect the collimator lamp connector to the J4
connector of the power board as shown in on page 116.
3. Fix the Inverter cables (Blue&black) to J5 connector on the sealing board using star screw driver as
shown in on page 122.
4. Fix the filament cable connector from the J6 of the power board to the J3 connector of sealing
board as shown in on page 122.
5. If the board contains FRC cable, fix the feedback cable from the J5 connector of the power board to
the J2 connector of the sealing board as shown in on page 123.
6. If the board contains microfit cable, fix the feedback cable from the J5 connector of the power board
to the J2 connector of the sealing board as shown in on page 123.
7. Fix all 8 Nos of M4*6 socket head screws of the power board to the tank assembly as shown in
on page 121 using 3 mm Allen key.
8. Fix the resonant inductor PP (plastic) sheet cover by fixing M4*30 socket head screw driver as
shown in on page 119.
9. Fix the 4 NOS M4*10 hex-head heat sink fixing screws to the tank assembly using 3 mm allen key as
shown in on page 121.
10. Fix the fan by fixing 2nos of M3*25 button head screws as shown in on page 120.
11. Fix one no M3*6 button head screws using 2 mm Allen key as shown in on page 120.
12. Fix 6 nos M4x8 socket head screws from the Bottom ID mounting strips 4 nos using 3mm allen key
as shown in on page 118, on page 118 and on page 119.
13. Fix the 4 Cover-mounting strips to the tank by fixing M4*8 button head screws using 2.5 Allen key as
shown in on page 116.
14. Connect the cable to the capacitor using flat screw driver as shown in on page 113.
15. Execute the steps in Execute the steps in ''Fixing of tube head to Remote console Communication
Cable''.
16. Execute the steps in Fixing Procedure on page 77.
10
Replacement Procedure
17. Execute the steps in Fixing Procedure on page 69.
18. Execute the steps in Fixing of Monoblock Skin Guard on page 71.
19. Execute the steps in Initial Checks & Power ON.
20. Execute the steps in Giving Exposure.
CAUTION
DROP HAZARD: Make sure the Base assembly is adequately compensated with dead
weight to avoid toppling of the positioner assembly before dismounting HV tank.
8. Remove M8 plain washer, M8 delrin washer, Arm bush, Compression spring, Washer, Delrin washer
as shown in Figure 10-73 on page 125, Figure 10-74 on page 126, Figure 10-75 on page 126, Figure
10-76 on page 126, Figure 10-77 on page 126, Figure 10-78 on page 126 and Figure
10-79 on page 126.
10
Table 10-2
Replacement Procedure
Figure 10-74 Figure 10-75
9. Remove 2 nos M4 nuts and M4 washers of tank mounting bracket-RH using as shown in Figure
10-80 on page 127, Figure 10-81 on page 127, Figure 10-82 on page 127, and Figure
10-83 on page 127. 10
Table 10-3
Replacement Procedure
Figure 10-80 Figure 10-81
10. Remove 2 No's of M4×10 HSHC, 2 No's of M4 washers using 3mm allen bit as shown in Figure
10-84 on page 128 and Figure 10-85 on page 128.
Table 10-4
Replacement Procedure
11. Repeat the same for tank mounting bracket-LH as shown in Figure 10-86 on page 128 and Figure
10-87 on page 128.
Table 10-5
WARNING
During HV tank replacement does not use old Tube head covers instead use new covers shipped
along with HV tank FRU.
1. Fix one side of Tank mounting bracket-1/Tank mounting bracket LH to the Tank assembly with 2
No's of M4×10 HSHC, 2 No's of M4 washers using 3mm torque bit and with a torque of 4.5 Nm. Once
torque is applied, draw a torque mark from screw head to screw fixing surface using marker pen.
2. Fix other side of Tank mounting bracket-1/Tank mounting bracket LH to the Tank assembly with 2
No's of M4 washers & 2 No's of M4 Nuts as shown in Figure 10-86 on page 128 and Figure
10-87 on page 128 using 7 mm socket by applying a torque of 2.5 Nm. Once torque is applied, draw
a torque mark from nut to nut fixing surface using marker pen.
3. Repeat the above steps to fix Tank mounting bracket-2/Tank mounting bracket RH OEM1 to Tank
assembly as shown in Figure 10-80 on page 127, Figure 10-81 on page 127, Figure
10-82 on page 127, and Figure 10-83 on page 127. 10
4. Take 1 no each of Delrin washer, compression spring, tank bracket arm bush, washer, M8 nyloc nut,
Replacement Procedure
Friction pad, M8 plain washer as shown in Figure 10-88 on page 129.
Figure 10-88
Replacement Procedure
Perform the following Functional Tests:
• 7.5 Ground Impedance Test on page 60
• 7.8 Exposure Test on page 61
• 7.9 Congruency of Radiation and Optical Field Test on page 62
Complete the regulatory test as recommended by local agency.
3. Remove two nos of M5x20 socket button head screw using 3 mm bit as shown in Figure
10-91 on page 130.
Figure 10-91
Finalization
Perform the follwing Functional Test: 10
• Swivel Arm Position Check (7.3 Swivel Arm Position Checks on page 59)
Replacement Procedure
10.29 Replacing Filter Box Plate
Removal of Filter box plate
1. Execute the steps in "Power off".
2. Remove the top cover of Filter unit by removing 3 No's of M3×6 HSHC screw along with M3 plain
washer as shown in Figure 10-92 on page 131.
Figure 10-92
3. Remove the input panel from the filter unit by removing 2 nos of M4x8 socket button cap head
screw fully threaded as shown in Figure 10-93 on page 132.
Figure 10-93
10
Replacement Procedure
4. Remove the receptacle from J1 connector as shown in Figure 10-94 on page 132.
Figure 10-94
5. Remove the filter box plate assembly from EMI filter box.
6. Remove the yellow / green wire from the center hole by removing M4x6 HSHC screw using 3 mm
allen key as shown in Figure 10-95 on page 133.
Figure 10-95
10
Replacement Procedure
Fixing of Filter Plate Box
1. Fix the 2 way receptacle to J1 connector as shown in Figure 10-94 on page 132.
2. Fix the lug of yellow/green wire to the center hole which is along the side of the Filter box with M4×6
HSHC using 3mm Allen key as shown in Figure 10-95 on page 133.
3. Insert the input panel to the filter box and fix it with 2 No's of M4×8 Socket Button Cap head Screw.
4. Fully threaded using 2.5 mm torque bit by applying 1.5 Nm torque.
5. Fix the Top cover to Filter unit using 3 No's of M3×8 HSHC screw along with M3 plain washer using
2.5 mm torque bit by applying 1 Nm torque as shown in Figure 10-93 on page 132.
6. Execute the steps in Initial Checks & Power ON.
7. Execute the steps in Giving Exposure.
Figure 10-96
10
Replacement Procedure
3. Remove the fastener in top cover with a 3mm Allen key as shown in Figure 10-97 on page 134.
Figure 10-97
4. Dismount the top cover separately as shown Figure 10-98 on page 135.
Figure 10-98
10
Replacement Procedure
5. Remove the wedge with 21mm spanner by rotating anticlockwise as shown in Figure
10-99 on page 135.
Figure 10-99
SL No kV mAs Exp-Time(mSec) mA
1 40 0.1 10 10
2 40 0.5 10 50
3 40 1 10 100
4 40 1.5 15 100
5 40 2 20 100
6 40 2.5 25 100
7 40 3 30 100
8 40 3.5 35 100
9 40 4 40 100
10 40 5 50 100
11 40 6.5 65 100
12 40 8 80 100
13 40 10 100 100
14 40 12.5 156.25 80
15 40 16 228.5 70
16 40 20 320 62.5
17 40 25 435 57.5
18 40 32 581.8 55
19 40 40 800 50
20 40 50 1000 50
21 40 63 1260 50
22 40 80 1600 50
23 40 100 2000 50
24 40 125 2631 47.5
25 40 160 3368 47.5
26 40 200 4210 47.5
27 40 250 5263 47.5
28 50 0.1 10 10
29 50 0.5 10 50
SL No kV mAs Exp-Time(mSec) mA
30 50 1 12.5 80
31 50 1.5 19.23 80
A
32 50 2 25 80
33 50 2.5 32 80
34 50 3 37.5 80
35 50 3.5 44.8 80
36 50 4 50 80
37 50 5 62.5 80
38 50 6.5 83.33 80
39 50 8 100 80
40 50 10 142.86 70
41 50 12.5 223.2 56
42 50 16 320 50
43 50 20 400 50
44 50 25 568.2 44
45 50 32 761.9 42
46 50 40 1000 40
47 50 50 1250 40
48 50 63 1575 40
49 50 80 2000 40
50 50 100 2500 40
51 50 125 3289.5 38
52 50 160 4210 38
53 50 200 5263 38
54 60 0.1 10 10
55 60 0.5 10 50
56 60 1 15 66.7
57 60 1.5 22.5 66.7
58 60 2 30 66.7
59 60 2.5 38.46 66.7
60 60 3 45 66.7
61 60 3.5 53.84 66.7
62 60 4 60.01 66.7
63 60 5 75.01 66.7
64 60 6.5 100 65
65 60 8 137.15 58.3
SL No kV mAs Exp-Time(mSec) mA
66 60 10 200 50
67 60 12.5 288.5 43.3
A
68 60 16 400 40
69 60 20 545.5 36.7
70 60 25 681.8 36.7
71 60 32 914.3 35
72 60 40 1201.2 33.3
73 60 50 1501.5 33.3
74 60 63 1890 33.3
75 60 80 2526 31.7
76 60 100 3003 31.7
77 60 125 3157.9 31.7
78 60 160 5052 31.7
79 70 0.1 10 10
80 70 0.5 10 50
81 70 1 17.5 57.1
82 70 1.5 26.2 57.1
83 70 2 35 57.1
84 70 2.5 43.75 57.1
85 70 3 52.5 57.1
86 70 3.5 61.25 57.1
87 70 4 70 57.1
88 70 6.5 133.82 48.6
89 70 8 186.7 42.9
90 70 10 269.23 37.1
91 70 12.5 364.58 34.3
92 70 16 486.95 32.9
93 70 20 636.36 31.4
94 70 25 833.33 30
95 70 32 1120 28.8
96 70 40 1400 28.6
97 70 50 1750 28.6
98 70 63 2321 27.1
99 70 80 2947.4 27.1
100 70 100 3684.2 27.1
101 70 125 4605 27.1
SL No kV mAs Exp-Time(mSec) mA
102 80 0.1 10 10
103 80 0.5 10 50
A
104 80 1 20 50
105 80 1.5 30 50
106 80 2 40 50
107 80 2.5 50 50
108 80 3 60 50
109 80 3.5 70 50
110 80 4 80 50
111 80 4.5 90 50
112 80 5 100 50
113 80 6.5 167.74 38.8
114 80 8 228.57 35
115 80 10 320 31.3
116 80 12.5 434.8 28.8
117 80 16 581.8 27.5
118 80 20 762 26.3
119 80 25 952.4 26.3
120 80 32 1280 25
121 80 40 1600 25
122 80 50 2000 25
123 80 63 2652.6 23.8
124 80 80 3368.4 23.8
125 80 100 4210.5 23.8
126 80 125 5263.2 23.8
127 90 0.1 10 10
128 90 0.5 11.25 44.4
129 90 1 22.5 44.4
130 90 1.5 34.61 44.4
131 90 2 45 44.4
132 90 2.5 57.69 44.4
133 90 3 69.23 44.4
134 90 3.5 80.76 44.4
135 90 4 92.3 44.4
136 90 5 132.35 37.8
137 90 6.5 201.72 32.2
SL No kV mAs Exp-Time(mSec) mA
138 90 8 276.92 28.9
139 90 10 375 26.7
A
140 90 12.5 489.13 25.6
141 90 16 654.54 24.4
142 90 20 857.14 23.3
143 90 25 1125 22.2
144 90 32 1440 22.2
145 90 40 1800 22.2
146 90 50 2368.4 21.1
147 90 63 2984.2 21.1
148 90 80 3789.5 21.1
149 90 100 4736.8 21.1
150 100 0.1 10 10
151 100 0.5 12.5 40
152 100 1 25 40
153 100 1.5 37.5 40
154 100 2 50 40
155 100 2.5 62.5 40
156 100 3 75 40
157 100 3.5 87.5 40
158 100 4 100 40
159 100 5 156.25 32
160 100 6.5 240.74 27
161 100 8 320 25
162 100 10 434.8 23
163 100 12.5 568.2 22
164 100 16 761.9 21
165 100 20 952.38 21
166 100 25 1250 20
167 100 32 1600 20
168 100 40 2000 20
169 100 50 2631.6 19
170 100 63 3315.8 19
171 100 80 4210.5 19
172 100 100 5263.2 19
NOTE
Above details are at input voltage 230V AC.
Max exposure time changes based on the input voltage
Exposure Timing is per IEC60601_2_54 cl203.6.4.3.104.5: Accuracy of LOADING TIME A
Exposure Current is per IEC60601_2_54 cl203.6.4.3.104.4: Accuracy of X-RAY TUBE CURRENT