Professional Documents
Culture Documents
Service Manual Universal Radiographic System X PLUS LP
Service Manual Universal Radiographic System X PLUS LP
SM-1028R13
Service Manual
Universal Radiographic System
X PLUS LP
Universal Radiographic System X PLUS LP
Service Manual
REVISION HISTORY
This Document is the English original version, edited and supplied by the manufacturer.
The Revision state of this Document is indicated in the code number shown at the bottom of this page.
ADVISORY SYMBOLS
The following advisory symbols will be used throughout this manual. Their
application and meaning are described below.
SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SAFETY SIMBOLS
The following safety symbols will be used in the equipment. Their meaning are
described below.
Ionizing radiation.
Type B equipment.
Dangerous voltage.
Ground.
SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
TABLE OF CONTENTS
Section Page
1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
2 UNPACKING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
4.1 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
SM-1028R13 i
Universal Radiographic System X PLUS LP
Service Manual
Section Page
5 ADJUSTMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
ii SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Section Page
6 TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
7 DISASSEMBLE / REASSEMBLE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
8 MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
SM-1028R13 iii
Universal Radiographic System X PLUS LP
Service Manual
iv SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SECTION 1 INTRODUCTION
1.1 TOOLS
SM-1028R13 1
Universal Radiographic System X PLUS LP
Service Manual
• Column Data
2 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Illustration 1-1
dimensions
310¦25
12.2”¦1
45o 45o
(104.3”)
2650
Max Travel = 1265 (49.8”)
2240
(88”)
(27.5”)
700
(16.7”)
425
(49.2”)
1250
1680
(66”)
SM-1028R13 3
Universal Radiographic System X PLUS LP
Service Manual
Illustration 1-2
Room Layout -- Installation at Wall and Floor
CONTROL BOX
GENERATOR
AND
(112.5”)
2860
1300
(51”)
1980 2170
(78”) (85”)
4200 (165”)
4 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Illustration 1-3
Room Layout -- Installation at Ceiling and Floor
4000 (157.4”)
CONTROL BOX
GENERATOR
AND
(137.7”)
3500
1300
(50”)
1980 2170
(78”) (85”)
4200 (165”)
SM-1028R13 5
Universal Radiographic System X PLUS LP
Service Manual
Illustration 1-4
Drill Template
(27.1”)
690
COMPACT GENERATOR
6 Wall Drilling Holes
(6.7”)
170
(23.6”)
CONTROL BOX
600
Lateral Cable Access
65 462 65
2.5” 18.1” 2.5”
37 52
1.4” 2”
R9.5
8.3”
213
1.9” 1.1” .8”
50 28 21
Anchoring Holes
6 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Illustration 1-5
Anchoring bolts
719
(28”)
1500
(59”)
1500
(59”)
301 204
(11.8”) (8”)
719
(28”)
1500
(59”)
1500
(59”)
301 204
(11.8”) (8”)
SM-1028R13 7
Universal Radiographic System X PLUS LP
Service Manual
1. Install Tube-collimator Support and Control Panel (column laid on the pallet).
2. Install Column Upper Support in back top of the column, mark and drill
the anchoring holes install Wall Support, anchor the Unit to the wall and floor.
Wall Support
Delta Support
8 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
A8127-XX Column Harness J1 - URS PLUS Interface BD A8124-07 (Control Box) Interface Cable
(Interface Cable)
A8136--XX) Column Harness J3 - URS PLUS Interface BD A8124-07 (Control Box) Supply Cable
(Supply Cable)
A8143-01 Column Harness J4 - URS PLUS Interface BD A8124-07 (Control Box) Detec. Mot cable
Emergency Cable
J1 J3 TS1 J4
TS3 TS1
SM-1028R13 9
Universal Radiographic System X PLUS LP
Service Manual
OTHER:
Emergency Stop wires located at the Back of the Receptor are bridged (3/6 and 4/5 to allow motion during installation),
remove bridge and connect wires in Terminal Strip TS4 (at the Back of the Receptor).
Special Receptor Signals: For Receptor Signals to be connected to Image System or to Control Box refer to Schematics Section.
10 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SECTION 2 UNPACKING
1. Place the shipping pallet near its final site in the room and remove all its
laterals. Do not discard any packing material such as envelopes, boxes,
bags until all parts are accounted for as listed on the packing list.
2. When the equipment is unpacked, verify that all items on the customer
order are present, and the hardware and internal wiring is firmly
connected.
3. Check the part numbers / serial numbers of each component with its
identification labels, and inspect all pieces for visible damage. If any
damaged parts are found, repair or order replacements to prevent
unnecessary delay in installation.
SM-1028R13 11
Universal Radiographic System X PLUS LP
Service Manual
1. Open the Generator shipping box, then move away the Interconnection
Cables, Cabinet Cover and other furnished parts. Do not discard any
packing material such as envelopes, boxes, bags until all parts are
accounted for as listed on the packing list.
3. Remove the Generator Cabinet from the shipping pallet and place it near
to its site in the room.
4. When the equipment is unpacked, verify that all items on the customer
order are present, and the hardware and internal wiring is firmly
connected.
5. Check the part numbers / serial numbers of each component with its
identification labels, and inspect all pieces for visible damage. If any
damaged parts are found, repair or order replacements to prevent
unnecessary delay in installation.
12 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SECTION 3 INSTALLATION
1. With the Column laid on the Pallet, install the Support of the
Tube-Collimator Assembly / Control Panel to the Upper Carriage of the
Swivel Arm using the four Safety Screws installed in the carriage.
All the Control Panel wires are factory connected except the SID Cable
(A8128--02), connect both connectors marked J2.
Safety Screws
Upper Carriage of the Swivel Arm
SM-1028R13 13
Universal Radiographic System X PLUS LP
Service Manual
2. Mark and drill the Column anchoring holes at floor and wall. (Refer to
Illustration 1-4).
3. Install the Wall Support in the wall with the screws, washers and nuts
provided.
4. With the Column laid on the Pallet, install the Column Upper Support in
the top of the Column with the screws, washers and nuts provided.
Column Base
5. Place the Column standing-up and position it at its final site in the room.
It is important to place the Wall Support above the Column Upper
Support. While one person is holding up the Column, the other one
should level it vertically on both lateral sides and on the front.
6. Anchor the Column Upper Support to the Wall Support with the bolts and
washers and nuts supplied.
7. Anchor the Column Base to the floor. Perform the leveling adjustments
on Column Base and on the Column Upper Support. Make sure that the
Column is correctly leveled and tighten firmly the Column.
8. Place the Control Box on its final position of the room, keep in mind:
G the length of the Column Harness (approx. 1.5 m -- 59”) that goes
from the Column to the Control Box,
14 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
9. Level the Control Box using the Adjustable Leveling Legs. Keep the Base
at the maximum distance from the floor.
Seismic areas and other conditions require to secure firmly the Control
Box to the floor through the mounting holes at its base. In this case, place
the four spacers (provided) under the base and secure them to the floor.
Keep the four Leveling Legs at the same height than the spacers.
Cable Access
10. Usually the Generator Cabinet is installed on the Control Box, in case it
is not done this way, refer to the Generator Installation Chapter in the
Generator Service Manual and follow with step 16.
11. Assemble the supplied Back Cover for the Generator Cabinet to the
upper side of the Control Box using two M6x20 screws (supplied).
Back Cover
Fixing Screws
Control Box
12. Remove the four Adjustable Leveling Legs from the base of the
Generator Cabinet.
SM-1028R13 15
Universal Radiographic System X PLUS LP
Service Manual
13. Hang the Generator Cabinet on the Back Cover and secure it to the
Control Box using four M6x20 screws (supplied). At least two people are
required for this operation. Fix the two screws that join the rear side of
the Cabinet to the Back Cover.
Generator Cabinet
Back Cover
Hook
Fixing Screws
Control Box
14. If the HV Transformer is shipped out of the Generator Cabinet, install the
HV Transformer inside the Cabinet (upper area) and secure it with the
respective anchors or plates, then connect the following cables from the
Power Module to the corresponding terminals on the HV Transformer:
G Connector J1.
16 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
15. Leave a working area around the Generator Cabinet until its final
installation.
ON / OFF Button
16. Install the lateral access cover of the Control Box. Perform the
corresponding connections from the Column (Main Harness) in the
Control Box as per schematics 54302080 and IM-XXX (refer to these
schematics at the end of this manual).
Main Harness
SM-1028R13 17
Universal Radiographic System X PLUS LP
Service Manual
GENERATOR CABINET
AND CONTROL BOX
-- Stator / Thermostat
COLLIMATOR
4. Cables to be connected to the Col. Lamp
Generator Cabinet:
-- Power Line
-- Door Interlock Switches
-- Warning Room Lamps
-- another cables
J2
TS4
GND
18 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
17. Inverters located inside the Control Box control the Motor parameters.
These Inverters are factory pre-configured.
18. Connect the Ground cables (GND) from the Column to the Main Ground
Stud (GND) of the Control Box.
19. Verify that the wire “ ” in Transformer TR1 is set according to the Power
Line.
Note . The Transformer is factory set at 230 V, if the Local Power Line is
220V, connect wire “ ” to the 208 V input in Transformer TR1.
20. Connect the Power Line cables to the Terminal Strip TS1 of the Control
Box, considering Line, Neutral and GND.
21. Install the X-ray Tube in the Support of the Tube-Collimator Assembly
using the Collimator Adaptation Ring and its four Safety Screws (Allen).
SM-1028R13 19
Universal Radiographic System X PLUS LP
Service Manual
22. Before installing the Collimator, unscrew equally the four Centering
Adjustment and Safety Screws (Allen) to allow the Collimator installation
in the Collimator Adaptation Ring. Adjust the Collimator Blades to their
widest setting and carefully install the Collimator centering it with the
X-ray Tube window.
Collimator Cover
Collimator Cable
23. Remove the X-ray Tube and Collimator covers and connect the Stator
/ Thermostat wires to the X-ray Tube and the Collimator Lamp cable
(A3388--02) from the Upper Harness of the Column, accordingly to their
identification. The Collimator Lamp is powered through its connection on
the Control Box (refer to schematic 54302080), do not power the
Collimator Lamp through the Generator Cabinet.
25. Install the Front Housing (cover of Control Panel and Tube-Collimator
Assembly) by snapping it first and then with the two screws at the back
of the cover.
20 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Prepare the High Voltage terminals that will be installed in the X-ray Tube
receptacles. Apply Silicone Paste over the entire surface of the Plug
including the Pins.
Carefully connect cables to their related receptors of the Tube and fix
their nuts tightly.
27. Fasten the Stator and HV cables to Upper Carriage of the Swivel Arm
with the clamp at its lateral side. Give the cable length enough to enable
movements of the of Swivel Arm and fasten them again to the Cable
Support at the upper lateral side of the Column together with the Column
Harness.
Detector Harness
28. Turn the Unit ON with the Switch located at the Control Box door.
29. Remove now the Safety Locking Rod from the Central Carriage.
30. Position the Swivel Arm at a comfortable position for installing the
Receptor. It is recommended to place the Swivel Arm in horizontal
position close to the floor.
SM-1028R13 21
Universal Radiographic System X PLUS LP
Service Manual
31. Turn the Unit OFF with the Switch located at the Control Box door.
32. Route the Receptor Cables through the Swivel Arm as indicated below:
a. Take all the Receptor Cables and identify the cable end to be
connected to the Receptor Assembly and the cable end to be
connected to the Control Box. Tie the cables to get a single
harness to ease the routing thorugh the Swivel Arm.
b. Check that all connectors in both cable ends pass through the
cable access or cable hoses (Vacuflex), if not, remove connectors.
Swivel Arm
Detector Cables
Leveling Screws
Detector Cables
Cable Access
22 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
d. Remove the cover of the cable hose (Vacuflex) located close the
Central Carriage in the Swivel Arm. A wire is provided from factory
as a cable guide to ease the routing of the Receptor Cables
through the Swivel Arm. The other end of this wire is at the cable
access of the lower side.
Cable Access
e. Attach all the Receptor Cables to the cable guide. Route the
Receptor Cables through the cable access at the lower side of the
Swivel Arm towards the central cable access at the Swivel Arm.
Swivel Arm
Cable Access
Detector Cables
SM-1028R13 23
Universal Radiographic System X PLUS LP
Service Manual
Detector Harness
34. Install now the Receptor Assembly in the Delta Support of the Swivel Arm
with its fixing screws. Once the Receptor Assembly is installed, it is
recommended to seal the Delta support screws with Loctite 243
adhesive.
EMERGENCY SWITCH
TS4
J9
FIXING SCREWS
HANDLE
DELTA SUPPORT
J2
b. Emergency Stop wires: they are bridged (3/6 and 4/5) to allow
motion during installation. Remove the bridge and connect the
Emergency Stop wires in the Terminal Strip TS4.
24 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
37. Turn the Unit ON with the Switch located at the Control Box door.
39. Turn the Unit OFF with the Switch located at the Control Box door.
41. Install the Base and Top Covers of the Columm with the supplied screws.
42. Route, to the upper cable access of the Generator Cabinet, the cables
from the Column to be connected in the Generator (HV Cables, Stator
/ Thermostat, if applicable AEC and Bucky signal cables).
SM-1028R13 25
Universal Radiographic System X PLUS LP
Service Manual
Levels
26 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Illustration 4-1
Control Panel Buttons
CONTROL PANEL
8 12 10
1 4
2 5
3 6
Enter Calibration Mode Enter Configuration Mode
9 11
ENTER IN SERVICE MODE: Turn On the Unit and press and hold (9) while “Vxx” is
being displayed on the Angle display until “CAL” appears, release all buttons and wait
five seconds until the real angle degree appears on Angle display.
SM-1028R13 27
Universal Radiographic System X PLUS LP
Service Manual
4.1 CONFIGURATION
The following is a guiding table for the configuration of the Unit. Write down the
values configured in “SET VALUE” column. Usually C04 and C05 are not to be
modified as they correspond to Standard Radiographic values.
C02 C21 Safety Height (This value corresponds to the minimum height 15 cm (6”)
from floor to Receptor with the Arm in vertical position).
position)
C22 Thorax SID (This value corresponds to the SID to be applied 180 cm (70.8)
when using the automatic Thorax position).
C03 C31 Minimum Height in horizontal position. 45 cm (17.7”)
C32 Maximum Height in horizontal position. 168 cm (66”)
C33 Half way Height. 105 cm (41.3”)
C04 C41 Minimum angle CCW (--30o) --30o
C42 Vertical Angle (0o) 0o
C43 Horizontal Angle (90o) 90o
C44 Maximum Angle CW (+120o) +120o
C05 C51 Minimum SID (100) 100 cm (40”)
C52 Maximum SID (180) 180 cm (70.8”)
C06 ((* onlyy for C61 Minimum Rotation Angle of Receptor CCW (--45o) --45o
U it with
Units ith
motorized C62 0o Detent of Receptor Rotation (Perpendicular to X-Ray 0o
rotation Beam)
Receptor) C63 Maximum Rotation Angle of Receptor CW (+45o) +45o
28 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
1. Turn On the Unit and press and hold (9) while “Vxx” is being displayed
on the Angle display until “CAL” appears, release all buttons and wait five
seconds until the real angle degree appears on Angle display, then go to
step 2.
1. Turn On the Unit and press and hold (9) while “Vxx” is being displayed
on the Angle display until “CAL” appears, release all buttons and wait five
seconds until the real angle degree appears on Angle display, then go to
step 2.
SM-1028R13 29
Universal Radiographic System X PLUS LP
Service Manual
The standard height of the room for normal functioning of the Unit corresponds
to a configuration value of 350 cm (138”).
30 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
This value provides information to the Unit that avoids collision of the Receptor
on the floor when going to vertical position. Depending on the height of the
Receptor, this value is measured in Horizontal position of the Receptor from the
Receptor Panel to the end of the Receptor in its maximum height. In this case
the Standard value is 15 cm (6”).
1. Measure with metric tape the height of the Receptor, that is, from the
Receptor Panel to the base of the Receptor (Refer to Illustration 4-2.
Write the data obtained in the table).
XX CM
MEASURE FROM RECEPTOR SUPPORT
TO END OF RECEPTOR
SM-1028R13 31
Universal Radiographic System X PLUS LP
Service Manual
This value provides information to the Unit that avoids collision of the Receptor
when going to vertical position. Depending on the width of the Receptor, this
value is measured in Horizontal position of the Receptor from the Receptor half
to the left of the Receptor in its maximum point. Standard value is 40 cm (15.7”).
1. Measure with metric tape the width of the Receptor, that is, from the
center axis of the Receptor to the left end (if C17) or to the right end (if
C18), refer to Illustration 4-3). Write the data obtained in the table.
2. Once C17 appears on Angle Display (follow steps 1, 2, 3, 4 of General
Configuration Procedure in case you are not coming from previous step),
press 1 or 3 to set the desired value.
3. Press 2 to save the data. Angle Display shows “OK”.
4. Press 5 to exit submenu and press 4 to go to C18 or Press 1, 2 and 3 at
the same time to exit.
This value provides a information to the Unit that avoids collision of the
Receptor when going to vertical position. Depending on the width of the
Receptor, this value is measured in Horizontal position of the Receptor from the
Receptor half to the right of the Receptor in its maximum point. Standard value
is 40 cm.
Illustration 4-3
Configuration of C17/C18 Receptor width -- center to left distance and center to right distance
40 cm 40 cm (15.7”)
C17 C18
32 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
This value provides information to the Unit of a third SID detent point. In case
it is not needed, set 190 cm (74.8”).
This value provides information to the Unit of the minimum height allowable of
the Arm in vertical position. In this case the standard value is 15 cm (6”)
measured from the Receptor end to the floor.
This value provides information to the Unit of the SID to be applied when using
the automatic Thorax positioning. In this case the standard value is 180 cm
(70.8”). If the operator requires to modify this values proceed in the following
way:
SM-1028R13 33
Universal Radiographic System X PLUS LP
Service Manual
This value provides information to the Unit of the minimum height allowable of
the Arm in horizontal position. In this case the standard value is 45 cm (17.7”)
measured from the arm central axis to the floor.
This value provides information to the Unit of the maximum height allowable of
the Arm in horizontal position. In this case the standard value is 168 cm (66”)
measured from the Arm Central Axis to the floor.
This value provides information to the Unit of the middle height of the Arm in
horizontal position. In this case the standard value is 105 cm (41.3”) measured
from the Arm Central Axis to the floor.
3. Press 5 to exit submenu and press 4 three times to go to C60 and press
5 again to enter C6.1 or Press 1, 2 and 3 at the same time to exit.
34 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
These values do not need to be modified as they are part of the standard
radiographic exams. In case these need to be modified, follow same procedure
as before.
This value provides information to the Unit of the Receptor angle when turned
counterclockwise. In this case the standard value is --45o.
This value provides information to the Unit of the Receptor angle when turned
to 0o. Then rotational movement will stop at this position. The standard value
is 0o .
This value provides information to the Unit of the Receptor angle when turned
clockwise. In this case the standard value is +45o.
SM-1028R13 35
Universal Radiographic System X PLUS LP
Service Manual
The following is a set of tips that may help to solve/understand some common
errors / problems found during the Configuration process.
• The maximun height value (C32) is less than the Undertable Height
(C14) + Receptor arm lengh (90 cm -- 35.4”).
• The maximun room height value (C12) is less than the tube arm lengh
(110 cm) + Receptor arm lengh (90 cm -- 35.4”) + the Undertable Height
(C14).
• The Undertable Height (C14) is less than the Receptor height (C16) + the
Safety Distance C21).
• The maximun height value (C32) is less than the Undertable Height
(C14).
• The maximun room height value (C12) is less than the tube arm lengh
(110 cm) + the Undertable Height (C14).
• The Undertable Height (C14) is less than Receptor arm lengh (90 cm --
35.4”) + the Receptor height (C16) + the Safety Distance C21).
• The maximun height value (C32) is less than the Thorax Height (C13).
• The maximun room height value (C12) is less than tube arm lengh (110
cm -- 43.3”) + the Thorax Height (C13).
36 SM-1028R13
C12: Max. height of he room
SM-1028R13
(350 cm -- 138”) Illustration 4-4
C21:
motor (152 cm -- 60”)
C17 C18
Detector Arm Tube Arm
-- 17.7”)
C33
(105 cm --41.3”)
C16
37
Universal Radiographic System X PLUS LP
Service Manual
4.5 CALIBRATION
The following is a guiding table for the calibration of the Unit. Write down the
calibrated values in the “SET VALUE” column.
P04 --(*
( only
y for Units with P41 Minimum Angle CCW --45o (configured in C61)
motorized
t i dR Receptor)
t )
Arm in vertical position. Measure P42 Vertical Angle 0O (configured in C62)
the angles with an Inclinometer.
P43 Maximum Angle CW +450 (configured in C63)
P21 (--30o)
P22 (0o)
P41 (--45o)
P24(+120o)
P43 (+45o)
P12
P13
P31 (100 cm--40”)
P11
P32 (180 cm--70.8”)
38 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
1. Turn On the Unit and press and hold (9) while “Vxx” is being displayed
on the Angle display until “CAL” appears, release all buttons and wait five
seconds until the real angle degree appears on Angle display, then go to
step 2.
3. Press 4 to go to P01.
4. Press 5 to enter into the Sub--level (it goes directly to the last selected
8 10 Sub--level).
1 4
2 5 5. Press 4 or 6 to go to PXX, then PXX appears on Angle display and it is
3 6 followed by the Configurated Value.
9 11
6. Press 1+3 to allow movement of the arm. Now the operation movement
buttons are enabled.
7. Position the Swivel Arm at the calibration point (measure with metric tape
or use an Inclinometer).
SM-1028R13 39
Universal Radiographic System X PLUS LP
Service Manual
1. Turn On the Unit and press and hold (9) while “Vxx” is being displayed
on the Angle display until “CAL” appears, release all buttons and wait five
seconds until the real angle degree appears on Angle display, then go to
step 2.
3. Press 5 to enter into P11 Sub--level (it goes directly to the last selected
Sub--level).
40 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SM-1028R13 41
Universal Radiographic System X PLUS LP
Service Manual
42 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
2. Position the Receptor Arm in configured position (0o). Measure with the
Inclinometer on the Receptor surface.
SM-1028R13 43
Universal Radiographic System X PLUS LP
Service Manual
44 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SECTION 5 ADJUSTMENTS
The following special Tools have been used for adjustment of the X-ray System
(these tools are not included with the System):
SM-1028R13 45
Universal Radiographic System X PLUS LP
Service Manual
Illustration 5-1
Checkings overview
3 Check / Correct alignment of the light field with the X-ray field
CENTERING ADJUSTMENT
SCREWS
46 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
A B C
CENTERING ADJUSTMENT
SCREWS
DEVIATION
6 SID Indicator Test
SM-1028R13 47
Universal Radiographic System X PLUS LP
Service Manual
Levels
4. Place the Swivel Arm in vertical position and Place SID at 1 meter.
48 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Illustration 5-2
Vertical Position of Swivel Arm with Alignment Tools
5. Turn on the Collimator light and center the Collimator in relation to the
Detector Table-Top. Horizontal and transversal position of the light axes
projected by the Collimator Lamp must be in line with the axes or Field
Sizes marked on the Detector Table-Top.
If needed, modify Collimator position by carefully unscrewing and
screwing the four Centering Adjustment and Safety Screws (Allen) of
Collimator.
Illustration 5-3
Collimator Screws
6. Position the Collimator Test Tool (RMI model 161B) on the Detector
Table-Top.
SM-1028R13 49
Universal Radiographic System X PLUS LP
Service Manual
8. Adjust the Light Field of the Collimator Lamp to the rectangle drawn
inside the Collimator Test Tool.
9. Place centered the Beam Alignment Test Tool (RMI model 162A) on the
Collimator Test Tool.
Illustration 5-4
Alignment Test Tools
DETECTOR TABLE-TOP
a. Check that the X-ray Field falls just within the image of the inner
rectangle of the Collimator Test Tool.
If an edge of the X-ray Field falls out of the inner rectangle means
a misalignment of the Light Field respect to the X-ray Field. The
maximum misalignment allowed is 2 % of SID (for SID 1m = 2 cm
tolerance).
Refer to Section 5.2.1 for alignment of Light Field with X-ray Field.
50 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
-- If the image of the two balls overlap (A) the X-ray Beam is
perpendicular to within 0.5o.
-- If the image of the top ball (larger shadow) intercepts the first
circle (B), the X-ray Beam is about 1.5o away from the
perpendicular.
-- If the image of the top ball (larger shadow) intercepts the second
circle (C), the X-ray Beam is about 3o away from the
perpendicular.
A B C
c. Check that the X-ray Beam is properly centered with the Image
Receptor. The maximum misalignment allowed is 2 % of SID (for
SID 1m = 2 cm tolerance).
SM-1028R13 51
Universal Radiographic System X PLUS LP
Service Manual
3. Position the Collimator Test Tool (RMI model 161B) on the Detector
Table-Top.
4. Turn on Collimator light and center the Collimator Test Tool with the light
axes projected by the Collimator Lamp.
5. Adjust the Light Field of the Collimator Lamp to the rectangle drawn
inside the Collimator Test Tool.
52 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
6. Check on the acquired Digital Image the adjustment required to meet the
Light Field with the X-ray Field. Identify the deviation on the axes imaged
on the image.
SM-1028R13 53
Universal Radiographic System X PLUS LP
Service Manual
7. Do not remove the Collimator Test Tool from its original position and
adjust the Light Field by moving the Collimator Unit and/or the Collimator
Lamp.
Illustration 5-5
Collimator Screws
54 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
In case that perpendicularity is out of tolerance (top ball is out of first circle),
adjust perpendicularity as follows:
3. Position the Collimator Test Tool (RMI model 161B) on the Detector
Table-Top.
4. Turn on Collimator light and center the Collimator Test Tool with the light
axes projected by the Collimator Lamp.
5. Place centered the Beam Alignment Test Tool (RMI model 162A) on the
Collimator Test Tool and observe if a shadow of the Beam Alignment Test
Tool is projected in equal proportion around it.
8. Repeat exposure and procedure until the result is satisfactory (top ball
must be inside of the first circle).
SM-1028R13 55
Universal Radiographic System X PLUS LP
Service Manual
Illustration 5-6
Adjustments in the Tube-Collimator Assembly
LOCK TOOTH
SET SCREWS
56 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
The error between centers of the X-ray Field and the Image Receptor should
not be greater than 2% of the SID. (for SID 1m = 2 cm tolerance).
DEVIATION
1. Print the acquired Digital Image. Check on the acquired Image the
correction required for centering the X-ray Field with the Image Receptor.
Adjustments on the Tube-Collimator Assembly will be performed as the
same way described for Perpendicularity correction, so only re-adjust it
if really is necessary.
3. Now place the Arm in Thorax position, acquire an image and repeat the
procedure from step 1. Due to the weight of the Detector, the weight of
the Tube and also due to the Arm torsion, sometimes the Centering of
the X-Ray field with the Receptor Assembly in Undertable and Thorax
position will have to be left at a middle position between the center in
Undertable position and the center in Thorax position. If this is the case,
the Collimator Laser lamp position can be slightly modified with its
regulating Screw.
SM-1028R13 57
Universal Radiographic System X PLUS LP
Service Manual
Note . Before starting with the SID Indicator Test, the Alignment of X-Ray
Beam Test and the Alignment of Light Field with X-Ray Field Test
should be performed.
Illustration 5-7
SID Test Stand Tool
3. Place the SID Test Stand Tool on the Detector Table-Top. (Refer to
Illustration 5-7).
58 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
5. Adjust the Light Field to cover the window (circle) on the top SID Stand
Test Tool. (Refer to Illustration 5-8).
6. Place the Small Tool included in the SID Stand Test Tool with two straight
narrow wires crossed on the center of circle at top SID Stand Test Tool.
Place a small signal on the circle in order to mark the position.
Illustration 5-8
Light Field over SID Test Stand Tool
a. Measure the height h, it is the height of the SID Test Stand Tool
plus the distance from the Digital Detector to the Detector
Table-Top.
b. Measure the radius of window (d/2) on the SID Test Stand Tool.
Illustration 5-9
SID Calculation
FOCAL SPOT
X-RAY TUBE
COLLIMATOR
(x) TO CALCULATE
SID ?
WINDOW CIRCLE
d
SID TEST STAND TOOL
h
DIGITAL DETECTOR
SM-1028R13 59
Universal Radiographic System X PLUS LP
Service Manual
Illustration 5-10
Radius in Acquired Image
SIGNAL
D/2
PROCESSED FILM
9. With the measurements obtained on the step-8., calculate the “X” value,
where:
(d/2) x h
SID X=
d/2 (D/2) - (d/2)
D/2
11. The difference between the indicated SID and the calculated SID may
not exceed 1.8 % (rejection limit) of indicated SID. Therefore the
deviation of SID calculated should be not higher than 18 mm for SID at
1 meter.
12. If the SID value calculated is out acceptance limits, it will be necessary
to perform the configuration and calibration of the Unit for the Minimum
and maximum SID.
60 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Note . Before starting with the Field Size Indicator Test, the Alignment of
X-Ray Beam Test, the Alignment of Light Field with X-Ray Field
Test and the SID Indicator Test should be performed.
4. Turn on the Collimator Light and center the Collimator in relation to the
Detector Table-Top. Horizontal and transversal position of the light axes
projected by the Collimator Lamp must be in line with the axes or Field
Size marked on the Detector Table-Top.
SM-1028R13 61
Universal Radiographic System X PLUS LP
Service Manual
6. Acquire the image on the Control station and check the following
measurements known (refer to Illustration 5-11):
Illustration 5-11
X-ray Field Size
IMAGE 35x43 cm
PRINTED IMAGE
8. The difference between the indicated Field Size and the obtained Field
Size may not exceed 1.5 % (rejection limit) of the SID in either direction.
Therefore the deviation in any direction should not be higher than 15 mm
(rejection limit) for SID at 1 meter.
62 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Note . Before starting with the Collimator Lamp Brightness Test, the SID
Indicator Test should be performed.
4. Turn on the Collimator Light and center the Collimator in relation to the
Detector Table-Top. Horizontal and transversal position of the light axes
projected by the Collimator Lamp must be in line with the axes or Field
Size marked on the Detector Table-Top.
5. Place the Light Meter on the Detector Table-Top with sensor directed
toward the Light Source.
6. Place the Light Meter in the center of one quadrant of the light field (refer
to Illustration 5-12).
Illustration 5-12
Light Meter on Detector Table-Top
4th QUADRANT
3rd QUADRANT
LIGHT FIELD
LIGHT METER
SM-1028R13 63
Universal Radiographic System X PLUS LP
Service Manual
7. With Collimator Lamp OFF, measure and keep record of the ambient light
level.
8. Do not move the Light Meter. With the Collimator Lamp ON, measure and
keep record of light level of that quadrant.
11. With the measurements obtained on the step-10., the results of Light
Field Intensity in all quadrants should be higher of 170 lux
(16 foot-candles) and among all the quadrants the intensity light field
shall be not differ more than 10%.
12. The average illumination at a distance of 1 meter from the focal spot shall
not be less than 160 lux (15 foot-candles). Rejection limit is 170 lux
(16 foot-candles).
b. If the light level in all quadrants is less than 170 lux, it is necessary
to check the following:
G Check that the Collimator Lamp, the Mirror and the Mylar
window are not dirty or discolored.
64 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
3. Open the Collimator blades to set a Field Size of 25 x 25 cm for SID 1m.
4. Turn ON the Collimator Light and center the Collimator in relation to the
Detector Table-Top. Horizontal and transversal position of the light axes
projected by the Collimator Lamp must be in line with the axes or Field
Size marked on the Detector Table-Top.
5. Place the Light Meter on the Detector Table-Top with sensor directed
toward Light Source.
6. Place the Light Meter in the center of one quadrant of the light field.
(Refer to Illustration 5-13).
7. Minimize the room lighting. With the Collimator Lamp OFF, measure and
record the ambient light level.
Illustration 5-13
Light Contrast Calculation
EDGE OF A QUADRANT
LIGHT SENSOR ON THE LEFT EDGE LIGHT SENSOR ON THE RIGHT EDGE
BRH POINT
(LIGHT DROPS 25%)
SM-1028R13 65
Universal Radiographic System X PLUS LP
Service Manual
9. Measure the illumination at a point 3 mm from the edge of the field toward
the center of the field (Light sensor on the left of edge). (Refer to
Illustration 5-13). Record this as I1.
10. Measure the illumination at a point 3 mm from the edge of the field away
from the center of the field (Light sensor on the right of edge). (Refer to
Illustration 5-13). Record this as I2.
11. Correct the values of I1 and I2 by subtracting from each value the ambient
light level measured in step-7. Now divide the corrected value of I1 by
I2. This ratio should be 4 or more.
12. Repeat the process from step-8. for all quadrants of Light Field.
13. If the deviation of Light Field Contrast Ratio is out acceptance limits, it
is necessary check the following:
G Check that the Collimator Lamp, the Mirror and the Mylar window
are not dirty or discolored.
G The ambient light level shall be low as it affects the accuracy of the
measurements.
66 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SECTION 6 TROUBLESHOOTING
2. Press at the same time “1” and “4” to enable Packing Mode, this mode
disables the Software limits of movements or the Gauge limits. The
following message appears on the display “Caution movement without
software restriction, place the equipment and switch it off”.
4. Position the Swivel Arm in horizontal position and at the correct height
in order to install the Safety Locking Rod.
5. Install the Safety Locking Rod and Turn the System Off.
SM-1028R13 67
Universal Radiographic System X PLUS LP
Service Manual
The EEPROM memory (U27) contains all the configuration and calibration Data
of the Unit. In case the data need to be reset due to E01 or any other reason,
perform the following steps;
1. Turn on the Unit and press and hold 9 while “Vxx” is being displayed
on the Angle display until “CAL” appears, release all buttons and wait five
seconds until the real angle degree appears on the Angle display.
2. Press at the same time 1+3+4+6, the display shows the message “INI
EEP”. The memory has been reset.
In case the Unit movements are suddenly stopped, there are two ways to
disable the protection Gauges:
SAFETY LOCK PLUS ANY MOVEMENT BUTTON: In case the Unit remains
+ Any movement continuously blocked (button lights up), it can be moved by pressing and holding
Button
the “Safety Lock” button and then pressing the movement button required.
1. Unscrew the four screws of the PCB cover located at the back of the
column.
3. Insert the Black Jumper in both Pins. Now the Gauges are disabled. Do
not forget to remove the jumper after relocating the arm position or “Error
07” will appear on display after powering On the Unit.
Illustration 6-2
Gauges Disabling Jumper.
BLACK JUMPER IN PCB TAKE OUT BLACK JUMPER INSERT JUMPER IN BOTH PINS
68 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
The Control Box includes three Motor Inverters factory configured. All units are
factory tested and configured according to customer electrical requirements
(50/60 Hz).
SID INVERTER
The Table 6-2 includes the factory configuration data. Due to the weight of the
Receptor, sometimes a fine adjustment can be performed to improve the
accuracy of the stop position, as an example; F8 is the parameter for “Slow
Speed”, F26 is the parameter for “Positioning Speed” and F27 is the parameter
for “Precision Speed”.
The following steps show how to modify a parameter in an Inverter, the process
of configuration is the same for the three of them. (Refer to the Inverter Manual
for extended information).
1. Turn On the Unit with the On/Off Switch-button of the Control Box Door.
Wait 30 seconds until the Inverters are ready. “045” appears on display,
it corresponds to standard speed. (if “045” does not appear, set it with
up or down arrow keys).
SM-1028R13 69
Universal Radiographic System X PLUS LP
Service Manual
Table 6-2
Inverters Factory Configuration Table
F_01 1 0.3 1
F_03 0 0 0
F_04 0 0 0
F_05 2 2 2
F_06 50 50 50
F_07 0 0 0
F_09 6 6 6
F_10 1 1 1
F_11 0 0 0
F_12 10 10 10
F_13 10 10 10
F_14 0 0 0
F_17 10 10 10
F_18 100 50 70
F_19 2 2 2
F_20 6 6 6
F_21 1 1 1
F_22 0 0 0
F_23 1 1 1
F_24 0 0 0
F_25 0 0 0
F_28 1 1 1
70 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Besides the Light beams installed in the Arm that slow down the speed and
stops motion in case the light beam is cut, this unit includes two Safety Devices
that will react to any inadverted bump into the Arm Assembly (Anticrushing
System) or into the Detector (Detector Bumper).
These safety Devices are designed to prevent high risk situations with patients,
that is, in the extremely rare case of collision into the Arm or the Detector, it will
automatically stop or reverse (in low speed and during 1.5 seconds in rotational
movement and during 3 seconds in up or down movement), the following is an
explanation of when the unit stops or reverses.
Detector Bumper: When the Arm of the URS--LP is moving down or when it
rotates, it will always reverse the movement if the Detector Bumper is activated
to avoid something kept between the detector and the floor.
3
2
2 1
When the Arm of the URS--LP is moving up and the Detector Bumper is
activated, it will make the reverse movement or stop depending on the following
situations:
3 2
• In the rest of the angle ranges (from 1o to 120o) and in horizontal position,
the system never reverses automatically, it stops.
SM-1028R13 71
Universal Radiographic System X PLUS LP
Service Manual
Anticrushing System: When the Arm of the URS--LP is moving down or when
it rotates, it will always reverse the movement if the Anticrushing System is
activated to avoid something kept between the detector and the floor.
1
2 3
3
2
When the Arm of the URS--LP is moving up and the Anticrushing System is
activated, it will make the reverse movement or stop depending on the following
situations:
2 3
• In the rest of the angle ranges (from 51o to 120o) and in horizontal
position, the system just stops.
510
1200
72 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
In the back face of the main carriage there are 4 bearings that must be properly
tighted against the linear guide.
a. First we must check if the gap between the bearings and the guide
is correct; the gap between any of the 4 bearings and the guide
must be less than 0.3 mm.
SM-1028R13 73
Universal Radiographic System X PLUS LP
Service Manual
c. The next step is to check if the screws are properly tighten. Each
screw must be tighted with 15 Nm torque. To check this value we
should use a dynamometric wrench. We must use an open wrench
to hold the eccentrirc nut, avoiding any movement of it:
74 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Error codes indicate the potential cause of a system failure. They appear on the
Control Panel Display at Column and will enable the service personnel to
identify and fix the problem.
Note . In case of bump or collision of any type, please position the Arm
in Horizontal position and medium height, then turn OFF the Unit,
wait two minutes and turn ON to allow the system performing a self
test that will apply correct values.
Table 6-1
Error Codes
Turn the System OFF / ON. If the error remains, reset the
“EEPROM ERROR”. Memory error. Failure in
E01 EEprom (U27) as explained in this Troubleshooting Section.
Integrated Circuit.
If error remains, replace PCB A8123--XX.
“MOVEMENT NOT REQUESTED”. Motors start Turn OFF/ON and recalibrate. If error remains replace
E03
motion without software request. (virtual error) EEprom.
Room maximum height too short. Configure again C12 and C16 to a higher value.
SM-1028R13 75
Universal Radiographic System X PLUS LP
Service Manual
“HEIGHT POTENTIOMETER ERROR”. Height Turn the System OFF / ON. If the error remains, Calibrate
E14
potentiometer out of calibration range. (P01), If error remains, replace Height potentiometer.
“ROTATION INCLINOMETER ERROR”. Turn the System OFF / ON. If the error remains, Calibrate
E15
Rotation Inclinometer out of calibration range. (P02). If error remains, replace Inclinometer.
“SID POTENTIOMETER ERROR”. SID Turn the System OFF / ON. If the error remains, Calibrate
E16
potentiometer out of calibration range. (P03). If error remains, replace SID potentiometer.
“RAM ERROR”. Memory error. Failure in Turn the System OFF / ON. If the error remains, turn it OFF
RAM ERR
Integrated Circuit. and Replace the PCB A-8123--XX.
* E07, is new error as from V6R1.1 (Receptor with Servomotor ) and V5R1.1
(Receptor without Servomotor).
* * * E14, E15, E16 are new errors as from V6R2.5 (Receptor with Servomotor)
and V5R2.5 (Receptor without Servomotor).
Description: Steel Cable that moves the Carriage Assembly of the system is
either broken or loose or
The Safety Device (”Parachute”) has been activated. This device prevents the
carriage from falling to the ground in case the steel cable breaks or is loose.
76 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Actions:
Remove the upper back cover of the Column and check that the Steel Cable
holding the Carriage Assembly on the back of the column is not broken, or
loose or winded in opposite direction.
If the steel cable is broken. Perform the Steel Cable replacement procedure as
explained in Job Card 1.10 in this Manual.
Verify that the carriage assembly is in its proper position, that no bearing
is out of place or broken as the carriage may have shifted it from its proper
position, if so, tighten or fix the bearing. Verify also the proper rolling of
the cable arround the pulley.
Then, at the same time, press the “Safety Lock button” and the “up
movement” button to rewind the cable in the motors pulley.
Turn the Unit OFF, wait two minutes, remove the sawhorses in the
meantime, and then turn it ON again, the system should be fully
operational.
If E06 still appears on screen, turn the Unit OFF, remove the Tube-Collimator
cover and remove the screws attached to the A8123-XX (in order to have
access to the PCB components) and check the electrical signal of the switch
SW7 in the following way:
Measure Ohms between Pin 8 and 9 in J1 connector at Control Box, the circuit
is normally open.
SM-1028R13 77
Universal Radiographic System X PLUS LP
Service Manual
In case of replacement of any of the three motors of the Unit (Height, SID or
Rotation), please check that the Connection Diagram corresponds to the
following:
U2 V2 W2
YY
U5 V5 W5
U1 V1 W1
L1 L2 L3
78 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
REPLACEMENT
ADJUSTMENTS
SOFTWARE UPDATING
SM-1028R13 79
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
Note . If possible, insert the Safety Locking rod in the Carriage to block
the carriage, in case it is not possible two persons will be needed
to support the carriage when releasing the Steel Cable.
Motor
and
Pulley
Rope or similar
80 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
9. (with the Motor Assembly on a working surface). Disassembly the Top Cover of the Defective
Motor.
Potentiometer Cable
10. Release the Potentiometer Support Screws and release the Tooth Belt from the Pot Gear.
11. Disassembly (1) one allen Screw at the back of the Pulley and remove it (with its shaft) from the
defective motor. (Use a Plastic Head hammer to bump out the shaft).
12. Unscrew the Potentiometer Assembly Screws from the defective Motor and take the
Potentiometer Assembly out (cable and connector included).
13. Reinstall the Potentiometer Assembly in the new motor (leave it loose).
14. Insert the Shaft/Pulley in the new motor (do not forget the Potentiometer Tooth Belt).
15. Reinstall the Top Cover of the Defective Motor in the new motor.
16. Install the new Motor in the Column and connect the Motor and potentiometer Assembly
connectors.
17. Wire the Steel cable around the spiral pulley in clockwise direction and install the Steel Cable
Screw. (it may be needed to power on the Unit and move up the carriage to release tension in
the Steel Cable).
18. Carefully remove the rope and reinstall the Upper Column Support with the Wall Support.
19. Power on the Unit and place the Carriage at middle travel. Then set the Potentiometer shaft in
its middle way, that is, five turns. Another way is to connect a multimeter (volts) in the
potentiometer contacts and move the potentiometer shaft to 2.5 Volts.
20. Place the tooth belt over the Gear and place the Mounting Plate to its final position. The belt
should be straightened by hand without any pressure at all. The idea is to have the belt
straightened enough so the teeth will stay in place but a little bit loose so the pot will not be pulled
and consequently damaged. It would be better to leave it a little bit loose than tighten. Normally
if you place some normal manual (finger) pressure in the middle of the belt ,it should bend around
2 to 3 mm. For fine positioning use the Tensioner Plate Screws, do not forget to tighten them when
in place.
21. Calibrate maximum, half and minimum height, for that, follow procedures in this manual.
When calibrating, the potentiometer readout (it appears during the calibration process after
pressing store button, the displays shows “OK” and then the potentiometer readout) for
minimum height should not be lower than 20 (typ. 23) and for maximum height should not be
higher than 80 (typ 78). If everything is well adjusted the value for half height should be very close
to 50. If the values are not within the limits, repeat step 3 and turn the potentiometer shaft to
increase or decrease all these values.
23. Reinstall the upper back cover.
SM-1028R13 81
Universal Radiographic System X PLUS LP
Service Manual
____________________________________________________________
Note . In case the Height Motor has to be replaced on the floor, perform
the following steps. At least three persons are needed to lay the
Unit on the floor.
1. Turn the Arm in Vertical position and lower the carriage as much as possible to keep the gravity
center close to the floor.
2. Power off the Unit.
3. Remove the upper back cover of the Column.
4. Unscrew the Floor anchoring screws and then the Upper Column Support.
5. Carefully, lay down the Unit. Rest the upper part on a piece of wood and the base on the floor.
Carriage
Upper Back Cover
6. Follow from Step 5 of previous procedure. When the new motor is assembled and the Unit is in
vertical position, do not forget to to install the base anchoring screws.
82 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
TOOLS : Standard Service Tool Kit. Two working benches or similar are advisable.
PROCEDURE
1. A minimum of two persons and a complete working day is needed to perform this task. Also two
working benches or similar support should be taken by the field engineers to comfortably perform
the replacement.
2. Position the machine to middle height and insert the Safety Locking Rod.
3. Rotate to Thorax position (if possible) this greatly facilitates the job.
4. Dismount the Tube and the Detector.
5. Dismount the cover for the rotation gear.
6. Place the working benches below the arm.
7. Turn off the machine.
8. Unscrew the arm from the carriage. First the 3 screws located at the lower right side, then the
rest of fixing screws at the back of the arm with the exception of the corner screws that will be
unscrewed the last ones to avoid the arm to bend.
Rotation Gear
SM-1028R13 83
Universal Radiographic System X PLUS LP
Service Manual
84 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Power on the Unit and position the Arm so that the carriage is at the middle height of the total
travel (typically 105 cm from the centre of the carriage to the floor).
2. Power Unit down.
Tooth Belt
Potentiometer Gear
Potentiometer Cable
SM-1028R13 85
Universal Radiographic System X PLUS LP
Service Manual
10. Mount the new potentiometer in the Tensioner Plate. Insert first the round washer, then the
locking washer. Position the potentiometer more or less where the other one was mounted before
and tighten the screws.
11. Mount the Mounting Plate in the Column, insert the gear in the potentiometer shaft and tighten
the two Grub screws.
12. Tighten the two Potentiometer Mounting Plate screws (M17) and place a Tie wrap to route the
Potentiometer cable, connect it.
13. Power on the Unit and place the Carriage at middle travel. Then set the Potentiometer shaft in
its middle way, that is, five turns. Another way is to connect a multimeter (volts) in the
potentiometer contacts and move the potentiometer shaft to 2.5 Volts.
14. Place the tooth belt over the Gear and place the Mounting Plate to its final position. The belt
should be straightened by hand without any pressure at all. The idea is to have the belt
straightened enough so the teeth will stay in place but a little bit loose so the pot will not be pulled
and consequently damaged. It would be better to leave it a little bit loose than tighten. Normally
if you place some normal manual (finger) pressure in the middle of the belt ,it should bend around
2 to 3 mm. For fine positioning use the Tensioner Plate Screws, do not forget to tighten them when
in place.
15. Calibrate maximum, half and minimum height, for that, follow procedures in this manual.
When calibrating, the potentiometer readout (it appears during the calibration process after
pressing store button, the displays shows “OK” and then the potentiometer readout) for
minimum height should not be lower than 20 (typ. 23) and for maximum height should not be
higher than 80 (typ 78). If everything is well adjusted the value for half height should be very close
to 50. If the values are not within the limits, repeat step 3 and turn the potentiometer shaft to
increase or decrease all these values.
16. Reinstall the upper back cover.
86 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Place the system so that the tube carriage is at the middle of the total travel of the SID.
2. Power system down.
3. Remove the Arm Back Cover.
SID Potentiometer Standard Version SID Potentiometers for Systems with Automatic Collimator
Chain Gear
Pot Gear
SM-1028R13 87
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Place the system in Thorax position.
2. Power system down.
3. Remove the Detector Back Cover.
Terminal Block
Motor Gear
4. Release the Pot Support screws to separate the Pot Gear from the Motor Gear.
5. Remove the Electrical connections from the Terminal block.
6. Unscrew the Motor Plate support screws (2).
7. Unscrew the Motor Support screws (4).
8. Carefully take out the motor. The Detector may turn, just hold it.
9. Install the new motor. Be careful with the Motor Gear and the Pot Gear do not apply force, they
should be just touching, enough for the movement to be transmitted to the Pot. Tighten the Pot
Support screws.
10. Connect the Motor cables to the Terminal Block and Ground.
11. Turn On the System and position the Detector at --45o.
12. Turn the potentiometer clockwise to the maximum and then turn 1/8 counterclockwise.
13. Check with a level the correct readout of the Detector angle.
14. Turn the System Off and Arm Back cover.
15. Start up the system and recalibrate the Motor Angle.
88 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Place the system in Thorax position.
2. Power system down.
3. Remove the Detector Back Cover.
Motor Gear
4. Release the Pot Support screws to separate the Pot Gear from the Motor Gear.
5. Unsolder the wires that feed the broken potentiometer.
6. Replace the broken Pot by the new one and sold the wires. Be careful with the Motor Gear and
the Pot Gear do not apply force, they should be just touching, enough for the movement to be
transmitted to the Pot. Tighten the Pot Support screws.
7. Turn On the System and position the Detector at --45o.
8. Turn the potentiometer clockwise to the maximum and then turn 1/8 counterclockwise.
9. Check with a level the correct readout of the Detector angle.
10. Turn the System Off and Arm Back cover.
11. Start up the system and recalibrate the Motor Angle.
SM-1028R13 89
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Place the system in Thorax position.
2. Power system down.
3. Remove the X-ray Tube Cover (screws at the back of the cover and pull).
PCB Display
Control PCB
4. Remove the PCB Support Screws and disconnect all the connectors at the Control PCB.
5. Replace the Control PCB or the Display PCB and connect all connectors.
6. Reinstall the cover.
7. Perform Configuration and Calibration processes (Refer to this manual).
90 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Place the system in Thorax position.
2. Power system down.
3. Remove the PCB Cover at the back of the column (a small square cover with four screws).
PCB Anticrushing
Connectors
4. Disconnect all the connectors at the PCB and remove the PCB Support Nuts.
5. Replace the Anticrushing PCB and connect all connectors.
6. Reinstall the cover and power on the Unit.
SM-1028R13 91
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Power system down.
2. Open the Control Box.
3. Disconnect all the connectors at the PCB Interface X Plus and remove the PCB Support Nuts.
4. Replace the PCB Interface X Plus and connect all connectors.
92 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
Pin
Safety Washers
SM-1028R13 93
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Please follow Job Card 1.1 (Replacement of the Height Motor). It includes the disassembly of
both the Pulley and the Tooth Belt.
94 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Turn the System Off and remove the Tube-Collimator cover.
2. Carefully disconnect the High Voltage cables from the Tube, for that, screw out first the Grub
Screw located at the Ring Nut, then unscrew Ring Nut and pull out the Termination Plug of the
High Voltage cable.
3. Remove the X-Ray Tube cover and disconnect the Stator / Thermostat cables.
4. Remove the Collimator back cover and disconnect the Collimator Lamp Cable.
5. Screw out the four Centering Adjustment Screws of the Collimator and remove the Collimator.
6. Unscrew the four allen screws of the Collimator Adaptation Ring and remove the X-Ray Tube.
7. Replace the X-Ray Tube and reinstall the assembly in reverse order.
Ring Nut
Grub Screw
H V Cable
Terminal of H V Cable
Collimator Cover
Collimator Cable
SM-1028R13 95
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Turn the System Off and remove the Tube-Collimator cover.
2. Remove the Collimator back cover and disconnect the Collimator Lamp Cable. (The Automatic
Collimator also includes a cable connection for the potentiometer signal, disconnect it as well).
3. Screw out the four Centering Adjustment Screws of the Collimator and remove it.
4. Replace the Collimator and reinstall the assembly in reverse order.
PROCEDURE
1. Turn the System Off and remove the Tube-Collimator cover.
2. Carefully disconnect the High Voltage cables from the Tube, for that, screw out first the Grub
Screw located at the Ring Nut, then unscrew by hand the Ring Nut and pull out the Termination
Plug of the High Voltage cable.
3. Now disconnect in the same way the High Voltage Cables from the Generator.
4. Assemble the mounting accessories of each Termination Plug following the Cable manufacturer
instructions.
5. Before installing the cables in the Generator Tank and in the X-Ray Tube, prepare the High
Voltage terminals that will be installed in the X--ray Tube(s) receptacles; Apply Silicone Paste
over the entire surface of the Plug including the Pins and the Tube receptacle, then place 1cm
of oil in the HV Cable receptacle in the Generator.
6. Carefully connect the Anode and Cathode cables into the respective X--rayTube receptacles
(Apply Silicone Paste on the X-Ray Tube Receptacle). Make sure that all connections are made
correctly, that is, keep correct Anode and Cathode orientation. Thoroughly tighten the Ring Nut
by hand, once finished, tighten the Grub Screw located at the Ring Nut.
7. Put approximately 1cm (0.5”) of HVOil in the HV Transformer receptacles.
8. Carefully connect the Anode and Cathode cables from the X--ray Tube(s) into the respective HV
Transformer receptacles. Make sure that all connections are made correctly, that is, keep
correct Anode and Cathode orientation. Thoroughly tighten the Ring Nut by hand, once finished,
tighten the Grub Screw located at the Ring Nut.
9. Replace the X-Ray Tube and reinstall the assembly in reverse order.
96 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Turn the System Off and remove the Collimator Back cover.
2. Remove the lamp protection heatsink.
3. Carefully remove the faulty lamp.
4. Replace the lamp with an identical lamp.
5. Make sure that the lamp pins are completely inserted in the lamp-holder.
6. Check on Light Field / X-Ray field correspondence.
7. If necessary remove the lamp, rotate it 180o axially and re-insert.
SM-1028R13 97
Universal Radiographic System X PLUS LP
Service Manual
SUBASSEMBLY : ARM
PROCEDURE
1. Turn on the system and move the Unit to Thorax position.
2. Turn the System Off and remove the Rotation Gear cover.
3. The Rotation Switches appear accessible. Remove then and replace.
4. The Height switches are located inside the carriage. A square blank cover at the back of the
carriage provides the access to the Height switches.
5. Remove the screws of the Blank cover and replace the switches.
SW5-SW6
98 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SUBASSEMBLY : CARRIAGE
PROCEDURE
1. Turn the System Off and remove the small square cover located at the back of the carriage.
2. This black cover provides the access to the Height limit switches. Remove and replace.
Carriage
SM-1028R13 99
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Turn the Unit OFF.
2. Remove the front and back covers of the Receptor Assembly.
3. Disconnect the Ion Chamber Cable.
4. Unscrew the four Plate Support s (two screws each) and lift carefully the defective Ion Chamber.
5. Place the new Ion Chamber at the same position on the Receptor Assembly.
6. Route and connect the Ion Chamber cables.
7. Fix it with the Four Plate Supports.
8. Reinstall the cover front and back covers of the Receptor Assembly.
Ion Chamber
100 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Power on the Unit and position the Arm so that the carriage is at the middle height of the total
travel (typically 105 cm from the centre of the carriage to the floor).
2. Power Unit down.
Tooth Belt
Potentiometer Gear
Potentiometer Cable
SM-1028R13 101
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Place the system so that the tube carriage is at the middle of the total travel of the SID.
2. Power system down.
3. Remove the Arm Back Cover.
SID Potentiometer Standard Version SID Potentiometers for Systems with Automatic Collimator
Chain Gear
Pot Gear
4. Set the Pot (both pots if system with autocollimation) in the middle position, that is five turns. One
can also power up the machine and adjust the voltage at the pot’s shaft 2.5 Volts.
5. Place the pot support so that the pot’s gear are in touch with the chain gear, do not apply force,
they should be just touching, enough for the movement to be transmitted to the Pot (make sure
that you keep the pot at its middle position).
6. Tighten the screws fixing the support.
7. Reinstall the Arm Back cover.
8. Start up the system and recalibrate the SID.
102 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
1. Remove the central back Cover of the Arm.
2. Turn On the System and position the Arm in horizontal position (check with a level).
3. Loose the Inclinometer fixing screws.
4. Turn the Inclinometer position in order to obtain 0o. The photo shows its correct position.
5. Tighten the Inclinometer screws.
6. Turn the System Off and assemble the Cover.
SM-1028R13 103
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
Note . This section only applies to systems where the X-ray Tube or the
Collimator has to be replaced by a new different X-ray Tube or
Collimator (different weight).
1. Turn the unit ON. Place the Swivel Arm in horizontal position and open SID at maximum meters.
2. Press the “Rotation” button and check that the Swivel Arm is balanced, that is, the Swivel Arm
should not tilt in any direction without force applied at any point.
If the Swivel Arm is correctly balanced, check the next section. If not, perform the following steps:
a. Remove the end cover of the Arm.
b. Install or remove the Counterweight Plates required to balance the Swivel Arm, so that the
Tube-Collimator Assembly end weights the same than the Detector Assembly end.
Re-install the Top Cover.
COUNTERWEIGHT PLATES
104 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
PROCEDURE
Note . This section only applies to systems where the speed of the
Receptor has to be modified (sometimes the speed has to be
reduced as the Receptor does not stop at 0 o or 90 o due to the force
of inertia).
TP3
TP4
R13
R16
GND PWR
SM-1028R13 105
Universal Radiographic System X PLUS LP
Service Manual
TOOLS : Special Serial Cable RS232. PC (Laptop) with RS232 serial port. Program “Flip V2.4.4” (to load
software on to the board). URS LP Control Board A8123--04.
PROCEDURE
1. First of all, you have to install the software for microcontrollers ATMEL serial programming in the
PC (flip 2.4.4). Save the dll file “AtRs232no.dll” on C:\Program files\Atmel\bin.
2. Connect the cable to the PC serial port and to the J10 control board connector. To program the
microcontroller set the W1 and W2 switches on “B” position and the supply connected to J6.
a. Start--> Programs--> flip
b. Press Device --> Select and mark T89C51RD2.
106 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
a. Press Setting --> Communications --> RS232. Select the serial port, the baud rate ( 38400)
and press Connect.
b. Make sure that Level is 0, if not, select Level 0 and press SET.
c. Now press File--> Load HEX and select the new control board software and press Open.
d. Select the “SXXXX.hex” file for Motorized detector systems or “SXXXX.hex” for
non--motorized systems.
e. Press Run.
4. After programming the microcontroller return W1 and W2 to “A” position.
SM-1028R13 107
Universal Radiographic System X PLUS LP
Service Manual
108 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SECTION 8 MAINTENANCE
Heavy duty installations (more than125 patients per day) require a regular six
(6) month maintenance.
GENERAL CLEANING
EXTERNAL SURFACES Clean external Covers and surfaces frequently, particularly if corroding chemicals are
present, with a cloth moistened in warm water with mild soap. Rinse wipe with a cloth
moistened in clean water.
Do not use cleaners or solvents of any kind.Also check painted surfaces for scratching
and touch up as required.
INTERNAL CLEANING DANGER: NEVER ATTEMPT TO CLEAN INTERNALLY THE UNIT. THE UNIT
MUST BE ALWAYS KEPT SHUT UNLESS A SPECIFIC RENEWAL PART
PROCEDURE IS TO BE UNDERTAKEN BY HIGHLY QUALIFIED MAINTENANCE
PERSONNEL.
SM-1028R13 109
Universal Radiographic System X PLUS LP
Service Manual
ELECTRICAL CHECKS
ELECTRICAL CABLES Check that all electrical connections are firm and secure and that all cable clamps and
AND CONNECTIONS strain reliefs are in place. Also check that connectors do not have any exposed wires
and check cable sheaths for wear and fraying.
MECHANICAL CHECKS
MOVEMENT COMPONENTS Check for correct operation of the controls, detents, limits and movements in all
possible positions.
STEEL CABLE Verify that the steel cable is in good condition and has a uniform appearance. Replace
the Cable if any steel wire is damaged. Refer to section 8.3.1 for greasing.
BRAKES / LOCKS Check correct operation of the Brakes / Locks. Replace if necessary.
RAILS: There are three rails along the Column to support the Swivel Arm
vertical movement: left side, right side and back side.
Clean the tracks of the rails before greasing them. Use some sandpaper on the
track surface if necessary.
110 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
Once the track surfaces are cleaned, apply a thin film of grease using a brush
or a piece of paper along the two tracks of the three different rails. The grease
must be applied only on the tracks where the bearings run. It is
recommended to stick some paper tape to the column next to the tracks to
prevent dirt in the column.
Do not apply any grease on the side of the rails where the screws are placed.
After greasing the tracks, clean this surface of the rail and remove the paper
tape.
STEEL CABLE AND PULLEY: Remove the upper back cover of the column and
place the swivel arm in the lower position to have the maximum length of the
wire rope extended out of the pulley.
Using a piece of paper or cloth, grease the extended wire rope placed on the
back of the column allowing grease to saturate the cable. Using a brush, apply
the grease to the rest of the cable in the pulley.
PULLEY
STEEL CABLE
BACK RAIL
SIDE RAIL
SM-1028R13 111
Universal Radiographic System X PLUS LP
Service Manual
112 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
A3 HANDLE 1 2 N 54404106
A6 OVERLAY 1 2 N 55801411
A0
A1
A4 A5
A2 A3
A6
A7
SM-1028R13 113
Universal Radiographic System X PLUS LP
Service Manual
B3 INCLINOMETER 1 2 N 50501016
B4 PHOTO-SENSOR 2 2 N 53412101
B0
B2 B10 B5 B3 B1 B4
SW1-SW2
B6
B8
B9
B7
114 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
C0 CARRIAGE A8103-- 01
C5 SPRING 4 2 N 10372-01
C0
C2
C3 x8
C6
C4 x4
C1
C5
C1
SW3-SW4
SW5-SW6
SM-1028R13 115
Universal Radiographic System X PLUS LP
Service Manual
D4 PULLEY 1 2 N 10367--02
D0
D4
D2
D1
D5
D3
116 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
SM-1028R13 117
Universal Radiographic System X PLUS LP
Service Manual
E0
E2 E12 E3
E10 x2 E6 x 2 E14
E4 E5 E4 E15
118 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
F1 TABLE-TOP 1 2 N 10443-01
F4 BUMPER 1 2 N 55102008
F0
F1 F2
F3
F6
F5
F4
SM-1028R13 119
Universal Radiographic System X PLUS LP
Service Manual
120 SM-1028R13
Universal Radiographic System X PLUS LP
Service Manual
• Anti-Crushing . . . . . . . . . . . . . . . . . . . . . . . . . . A3531--01
SM-1028R13 121
Universal Radiographic System X PLUS LP
Service Manual
122 SM-1028R13
A B C D E
COLLIMATOR
+24 VAC LAMP
0 VAC LAMP 2
4 F3, T2A
(A8130-05/15)
GND 1
7
208V 4 Emergency
8V FAN SW1 Stop Button Power Cable
LF1 TS1 (A3153-XX)
F1,T10A
0V 2 1 230 VAC
3 1 L1
0 VAC/Neutral from VAC LINE
4 2 L2/N
19V F7, 400mA
5 4 F2,T10A/N GND
GND
2
GND Stud
0V 0V
WARNING GND Stud GND Cable of the Column GND Cable to GND Stud of the
1
4
THE EMERGENCY STOP BUTTON SHOULD BE Generator Cabinet 4
INSTALLED IN ACCESSIBLE LOCATION FOR USE IN AN
EMERGENCY. IT SHOULD BE CONNECTED AS SHOWN.
TS2 J11 J1
Emitter 1 Cable (A8140-01) J9
Photocel +V (brown) +5 VDC
Emitter 1 0V (blue) 1 1 +5 VDC 2
2 1 2 +5 VDC 1
11 2 3 GND 4
(A8123-04) 16
Safety Device TS4 Safety Cable (A8142-01)
UNLOCK
GND
1
2
SAFETY1 J5
DET CW IN
J7 1
CAN L
SID ARM CAN H for BUS CAN
2 GND DET CCW IN
SW2 4 1 3
2
SID 180 J2 SID Cable (A8128-02) 3
3
4 J10
SID 180
SW1 2 4 SID 180 RTN 5 TxD
3 6 1 BUCKY HANDLE OVERLAY
SID 100 SID 100 GND for RS232
4
1 2 SID 100 RTN 7 2 RxD (55801369) 4
1 8 3
Detector Assembly
SAFETY Cable
J9 (A8142-02) TS9
+5 VDC
4 1
4 4
SAFETY SIGNAL R1
12 2 ROT. POT
GND
5 3
DET ROT
6 4
Shield
TS3 TS3
X-Plus Unit
+24V UNR
1 1
GND
2 2
Interface
Panel Interface
Power Gnd 5
DET ROT Cable Power Panel Detector Motor Cable
(A8155-01) +Vcc 4 (A8149-01)
J5 J4 Earth 3 J6
Shield
DET CW OUT MOTOR - black
3
7 1 1 -Motor 2 1 7 3
Ready
8
9
Detector Assembly
URS Plus LP
+12V/12mA OUT 10
Interface
-12V/12mA OUT 11
Board
(A8124-07) GND 12
WARNING :
Servoamplifier
2 2
CCW ROT
Board is not used. That is 2
CW SW
to enable the X Plus 3
CCW SW
Note 1.- Configurate switches as IxR mode:
normal operation without 4
BRAKE SW1-OFF, SW2-ON, SW3-ON, SW4-OFF, SW5-OFF, and SW6-OFF
5
no interface interruption. 6
GND (see documentation of the Servoamplifier)
1 1
UNLOCK
8 cap x 100 nF
LOW SPEED
UNDERTABLE
HEIGHT UP Rotate Relay
HEIGHT DWN Sid Relay
INCREASE SID Height Relay
DECREASE SID Thorax Key
R7
ROTATE CW Rotate Ccw Key
4K7 U4 TLP621
ROTATE CCW Rotate Cw Key
8 7 7 10
A4 C4 THORAX Decrease Sid Key
8 K4 E4 9 Increase Sid Key
J8 6 5 5 12
A3 C3 Height Dwn Key
6 K3 E3 11 Height Up Key
GND 4 3 3 14
1 A2 C2 Undertable Key
HEIGHT UP 4 13 +5V
2 HEIGHT DWN 2 1 K2 E2 R10
3 1 A1 C1 16
2 INCREASE SID 2 15 1 470 2
4 DECREASE SID K1 E1 9 1 8
5 ROTATE CW R6 8 2 7 U16
6 ROTATE CCW 4K7 U3 TLP621 7 3 6 74HCT244
7 UNDERTABLE 8 7 6 4 5 D0
8 7 A4 C4 10 2 1A1 1Y1 18
THORAX 8 9 2 4 16 D1
9 LOW SPEED 6 5 K4 E4 3 R9 1A2 1Y2 D2
10 5 A3 C3 12 6 1A3 1Y3 14
UNLOCK 6 11 4 470 8 12 D3
11 DET CW IN 4 3 K3 E3 5 4 5 1A4 1Y4 D4
12 3 A2 C2 14 11 2A1 2Y1 9
DET CCW IN 4 13 3 6 13 7 D5
13 GND 2 1 K2 E2 R8 2 7 2A2 2Y2 D6
14 1 A1 C1 16 15 2A3 2Y3 5
2 15 4K7 1 8 17 3 D7
15 K1 E1 2A4 2Y4
16 R2 C20
C17 C15 C22 19 CS9\
4K7 U2 TLP621 2G
1G 1
8 7 7 10 +5V C18 C16 C19 C21
A4 C4 R5
8 K4 E4 9
6 5 5 12 1 470
A3 C3 9 1 8
6 K3 E3 11
4 3 3 14 8 2 7 8 cap x 100 nF U14
A2 C2 7 3 6 74HCT244
4 K2 E2 13
2 1 1 16 6 4 5 2 18 D0
A1 C1 2 1A1 1Y1 D1
2 K1 E1 15 4 1A2 1Y2 16
3 R4 6 14 D2
R1 4 470 1A3 1Y3 D3
8 1A4 1Y4 12
4K7 U1 TLP621 5 4 5 11 9 D4
8 7 3 6 2A1 2Y1 D5
7 A4 C4 10 13 2A2 2Y2 7
8 9 R3 2 7 15 5 D6
6 5 K4 E4 4K7 1 8 2A3 2Y3 D7
3 5 A3 C3 12 17 2A4 2Y4 3 3
6 K3 E3 11
4 3 3 14 C6 C4 C9 C11 19
A2 C2 2G CS10\
J9 4 K2 E2 13 1G 1
2 1 1 16 C7 C5 C8 C10
IN1 A1 C1
1 2 K1 E1 15
IN2
2 IN3 +5V
3 U15 8 cap x 100 nF
4 R25 ULN2803 U17
5 DET ROT 22 74HCT573
6 OUT1 2 1 D0
7 18 Q1 I1 1 19 Q1 D1 2
OUT2 4 3 17 2 18 3 D1
8 OUT3 6 5 Q2 I2 Q2 D2 D2
9 16 Q3 I3 3 17 Q3 D3 4
OUT4 8 7 15 4 16 5 D3
10 OUT5 Q4 I4 Q4 D4 D4
11 14 Q5 I5 5 PAL1 15 Q5 D5 6
R53, 22 13 6 14 7 D5
12 SAFETY 2 Q6 I6 PAL2 Q6 D6 D6
IN4 12 7 13 8
13 IN5 Q7 I7 Q7 D7 D7
14 11 Q8 I8 8 12 Q8 D8 9
15 GND
16 LIMIT\ C 11 12 13 CS11\
R24 1
TP1 2K 9 OC U33F
+5V
+24 VDC DET 74HC04
PHOTOCEL 1
C2 D2
100nF 1N4448 R23 C40
2K 100nF
R17 2 PHOTOCEL 2
2K 1 DET ROT
4 3 R22 C39 4
U13A 2K 100nF
R16 D1 LMC6482
LIMIT
1M 2K 1N4448 C1
33nF C38
100nF NAME DATE SHEET / OF
DRAWING F. García 07/07/04 A8123-04
1/4
REVISED F. García 09/09/04 Bb1 B1 REV
PHOTOCEL 2
+5V
PHOTOCEL 1
J4
8 7 R35 TP4
+V2r 16 A(0...12)
4 3 +24 VDC 2K49 R36 R37 ROT
PHOTOCEL 2 15 2 1 2K 9K09 +24 VDC
0V 14 6 5
+V2e 13 2.5 V
+5V R31 R32 U28
0V 12 R40 3K4 2K 74HC138 2,13
GND 11 1K D17 D16 C58 A12 D13 K11
+V1r 10 15 Y0 A 1
1N4448 LM336M2.5 100nF 14 2 A13 +5V C31 1N4148 DECREASE SID
PHOTOCEL 1 9 E\ Y1 B A14
C55 6 13 3 100nF 6
0V 8 CS2\ Y2 C
1 100nF R33 2 7 12 1
+V1e 7 CS3\ Y3 +5V
2K 1 5 11
28
GND 6 CS4\ Y4
3 U21B 10 6
ROT POS 5 CS5\ Y5 G1 2,13
U21A R38 LMC6482 9 4 A15 2
+5 VDC CS6\
VCC
4 +5V R30 D18 LMC6482 2K R39 Y6 G2A CLK/I D14 K12
GND 3 7 Y7 G2B 5 PAL1 13 I I/Q 23
2K 1N4448 C56 R34 9K09 C59 16 21 1N4148 INCREASE SID
SID POS 2 PAL2 I I/Q
33nF 49K9 33nF 3 20 6
+5 VDC 1 U30 I I/Q
4 I I/Q 19
+5V TP6 74HC138 5 18
SID A12 I I/Q
CS8\ 15 Y0 A 1 PHOTO EN\ 6 I I/Q 17
C64 D21 14 2 A13 7 25 2,13
CS9\ Y1 B A14 BROKEN\ I I/Q
100nF 1N4448 13 3 9 26 D9 K7
CS10\ Y2 C I I/Q
U8C 12 10 24 1N4148 ROTATE CCW
CS11\ Y3 I I/Q
R43 6 74HC00 U8B 11 11 27 6
GND
CS12\ Y4 I I/Q
2K 7 9 74HC00 10 6 A15 12
Y5 G1 I
5 CS1\ 8 4 9 Y6 G2A 4
U22B 10 6 7 5
Y7 G2B
14
R44 D22 LMC6482 5 U7 2,13
1M 2K 1N4448 C63 PAL22V10 D10 K8
33nF 1N4148 ROTATE CW
D[7..0] 6
+5V TP5
HEIGHT
C61 D19 U19 U10
RES\ 2,13
100nF 1N4448 74HCT573 ULN2803
J3 +5V D0 2 D1 Q1 19 A0 D5 K3
R41 2 D1 3 18 A1 D0 2 19 1 18 1N4148 HEIGHT UP
2K D2 D2 Q2 A2 D1 D1 Q1 I1 Q1
+5 VDC 1 4 17 3 18 2 17 6
1 D3 D3 Q3 A3 D2 D2 Q2 I2 Q2
HEIGHT POS 2 3 5 D4 Q4 16 4 D3 Q3 17 3 I3 Q3 16
2 U22A +5V D4 6 15 A4 D3 5 16 4 15 2
GND 3 R42 D20 LMC6482 D5 D5 Q5 A5 D4 D4 Q4 I4 Q4
7 D6 Q6 14 6 D5 Q5 15 5 I5 Q5 14
1M 2K 1N4448 C60 D6 8 13 A6 D5 7 14 6 13 2,13
D7 D7 Q7 A7 D6 D6 Q6 I6 Q6 D6 K4
33nF 9 D8 Q8 12 8 D7 Q7 13 7 I7 Q7 12
C44 U24 +5V D7 9 12 8 11 1N4148 HEIGHT DWN
100nF T89C51RD2 D8 Q8 I8 Q8
38 11 6
R18 C GND
1 OC 11 C
10K 1 1
VDD D0 OC 9
29 37 2
+5V EA/VP P0.0 3 D1 U9 2,13
P0.1 36
+5V 4 7 35 4 D2 74HCT573 D11 K9
PFI RESET P0.2 ALE
34 5 D3 1N4148 SP1 SID
C72 P0.3 6 D4
1 8 4 33 6
U25 100nF MR RESET RESET P0.4 7 D5 W2
32
GND
AD7812 P0.5 8 D6 W1
20 2 VCC PFO 5 P0.6 31 A B 2 1 CS2\
C73 30 9 D7 ALE
100nF P0.7 A B U33A 2,13
Vdd
3 19 U32 27 74HC04 D12 K10
3
+5V
J7
RD\
1
WR\ 2 GND
3 +5 VDC
R47 +24 VDC
SAFETY 1 4 SAFETY 1
1K K12
5 SID 180
R46 1,14 7,8
6 SID 180 RTN
22K1 K11
7 SID 100
Q1 1,14 7,8
VLCD ON\ 8 SID 100 RTN
BC857
+5V
1 J1 1
+24V
D23 1 +5 VDC
LIMIT 2 LIMIT
DISPLAY BZV55C2V4 C76
MOV START\ 3 MOV START\
J13 +5V 100nF R49
2R21 4 GND
VSS 5 +5 VDC
1 VDD 6
2 7 GND
3 BROKEN 8 BROKEN
A0 U34 K7
4 9 GND
6
WR\ MC34063AD 1,14 7,8
5 RD\ L1 10 -30º
1 8
Vpower
6 D0 470uH SW Collector DRIVE K8 11 -30º RTN
7 D1 1,14 7,8 12 +120º
8 D2 13 +120º RTN
9 2 SW Emitter Isense 7 14
D3 K3
10 D4 C77 1,14 7,8 15 GND
GND
11 D5 15uF 16 UPPER LOCK
12 3 Timing Comp Input 5 17 UPPER LOCK RTN
D6 16V + K4
13 D7 D24 1,14 7,8 18 LOWER LOCK
14 CS\ C75 19 LOWER LOCK RTN
10BQ060
4
15 RST\ 220pF 20
16 VEE
17 SEL1
18 J2
VLED
19 VLSS
20 1 MOTOR INCREASE SID
2 MOTOR DECREASE SID
R50
3 MOTOR SID COMMON
2 12R4 2
4 MOTOR ROTATE CW
5 MOTOR ROTATE CCW
RES\ 6 MOTOR ROTATE COMMON
7 MOTOR HEIGHT DWN
E\ 8 MOTOR HEIGHT UP
9 MOTOR HEIGHT COMMON
10 MOTOR SID SP1
D[0..7] K9
11 MOTOR SID RST
7,8 1,14
12 MOTOR ROTATE SP1
13 MOTOR ROTATE RST
A(0...12) C36 +5V 7,8 1,14
14 MOTOR HEIGHT SP1
100 nF K10
15 MOTOR HEIGHT RST
K5
16 SID RELAY
U20 7,8 1,14
17 ROTATE RELAY
J12 CY6264-70
28
18 HEIGHT RELAY
7,8 1,14
A0 D0 K6 19 RELAY COMMON
10 11
VCC
1 CS4\ A1 A0 D0 D1 20 INVERTERS OFF
9 12 K1
2 CS5\ A2 A1 D1 D2
8 13 7,8 1,14
3 CS6\ A3 A2 D2 D3
4 MODE 7 A3 D3 15
D0 A4 6 16 D4 7,8 1,14
5 D1 A5 A4 D4 D5 K2
6 5 A5 D5 17
D2 A6 4 18 D6
7 A7 A6 D6 D7 SID RELAY
D3 3 19
8 A8 A7 D7 ROTATE RELAY
D4 25 C46 +5V
9 A9 A8 HEIGHT RELAY
D5 24 100 nF
10 D6 A10 A9 C78 +5V
11 21 A10 INVERTERS OFF
D7 A11 23 U26 100 nF
12 A12 A11 82C900 U35
2
22
13 A12
7
3
+5V PCA82C250 3
14
3
CS\ 20 +5V
VCC
VCC
15 CS1 D0 TRANS-CAN
11 5
VCC
16 RD\ D1 P0 Vref T1
22 OE 18 P1 J5
WR\ 27 26 D2 12 1 4 6
GND
4
D4 13 2 1 7 100pF CAN0001
D5 P4 TxDCA TXD CANH 2 CAN H
16
GND
P5 3 GND
14
D6 14 28 8 R51 C80
D7 P6 RxDCB RS 121 100pF V3
15 P7
27 CAN0001
3
CS1\ CS\ TxDCB
9
2
CS12\ ALE CTRL0
20 +5V C45
ALE CTRL1 100 nF
OUT0 6
RD\ 19 CTRL3
OUT1 23
4
5 6 3 R26
RES\ CAN RESET 10
U33C WR\ 10 4 3 X2 1
74HC04 CTRL2 XTAL1 24MHz
26 MODE0 XTAL2 5
J11 +5V C49 +5V
100 nF 25 24
2
MODE1 RDY
GND
GND
1 U29A
2 +5V C50 74HCT123
16
3 47pF
4
21
R27 14
8
VCC
5 10K Cext
4 6 Q 4 4
7 UNLOCK 15 Rext/Cext
2 13
from KEYPAD
8 LOW SPEED B Q
3
GND
9 UNDERTABLE CLR
10 HEIGHT UP 1 A
11 HEIGHT DWN NAME DATE SHEET / OF
12 INCREASE SID DRAWING F. García 07/07/04 A8123-04
8
1 1
+5V
C90
100 nF
16
+5V +5V +5V
C92 14 C91 6 C37 6
VCC
R20 1 uF Cext R19 1 uF Cext R56 1 uF Cext
2M23 4 5M23 12 5M23 12
Q Q Q
15 Rext/Cext 7 Rext/Cext 7 Rext/Cext
3 CLR Q 13 11 CLR Q 5 11 CLR Q 5
1 2 2 3 4 10 5 6 10
GND
GND
GND
B B B
1 A 9 A 9 A
+5V R57 R57 R57
C83 10K C84 10K C85 10K C86
100nF 1 uF 1 uF 1 uF
8
U36
PAL22V10
28
18
VCC
I/Q
17 I/Q
13 I
Sid Relay 16 I I/Q 27
2
Height Up Key 3 I I/Q 26 2
Height Relay I
2 R57
Rotate Relay CLK/I 10K
8 9
INV OFF
14
R61
10K C87
0.1 uF
C89
1 uF
TP9
+24 VDC +24V
TP8 +5V
+5V D15
1N4007
R52
3
R28 R29 10K 3
1K U38 1
LM7805 1w
C79
VOUT VIN 220nF J6
TP7 TP3 TP2 C54
+5V GND GND GND 100uF GND C53
100uF 1 +24 VDC
16V
16V 2 +9 VDC
C3,C57,C62
100nF 3 GND
8
6
7 6
5 7
LMC6482 5
4 U13B,U21B,U22B U13B
LMC6482
16
20
74HCT573 74HCT244
4 14 4
12
U8D
13 74HC00 U31,U33 U28,U30
74HC04 74HC138
7
7
10
K5
7,8 1,14
K4
7,8 1,14
J1 J2
MOTOR INCREASE SID
1 MOTOR DECREASE SID 1
2 MOTOR SID COMMON 2
3 MOTOR ROTATE CW 3
4 4 4 4
MOTOR ROTATE CCW
5 MOTOR ROTATE COMMON 5
6 MOTOR HEIGHT DWN 6
7 H-DWN 7
MOTOR HEIGHT UP
8 H-UP 8
MOTOR HEIGHT COMMON
9 MOTOR SP1 SID 9
10 MOTOR RST SID 10
11 MOTOR SP1 ROTATE 11
12 MOTOR RST ROTATE 12
13 MOTOR SP1 HEIGHT 13
14 MOTOR RST HEIGHT 14
15 SID RELAY 15
16 ROTATE RELAY 16
17 HEIGHT RELAY 17
18 RELAY COMMON 18
19 INVERTERS OFF INVERTER GND 19
20 20
+24 VDC
K3
J3 8
1 9
+24 VDC
1 +9 VDC D10
2 GND 1N4148
3 16 12
5
3 3
TP1 TP6
TS1 F1 F2 +32 VDC +24 VDC
6A 3A
24 VAC LAMP D1
1 FB1002L R7 R6 D15 R23
0 VAC LAMP 1K 6.8 1N4007 1K
2 5w
24 VAC - +
3 VIN VOUT
0 VAC1 C2 C3 R1
4
ADJ
2200uF 2200uF 2K7 R21
+24 VDC-2 50V 50V 2w 243
5 TP2 U2 D16
INVERTERS OFF F3 +9 V LM317 1N4007 C18
6 2A C16 100 nF
8 VAC D4 Q2 100 nF
7 B250C1500 R10 MTD3055E R22 C19
0 VAC2 1K R8 R12 D 500 10uF
8 2M 100 50V
GND - + G C17 C20
9 S 10uF 10uF
+32 VDC C4 C5 R9 50V 50V
10 2200uF 2200uF 2K C6 R20 TP7
19 VAC 25V 25V 10 uF 4K12 GND
11 19 VAC
0 VAC3
12 0 VAC3
2 2
D5
DB104 D19 K6 1
1N4148 6 K6
- + 4
16 8
R11 1 D20
17
10K K7 1N4148
2 U7
C7 ULN2803 16
1 uF R34
TP8 9 33K2
D14 +24 VDC-2 Q4
1N4007 +24 VDC-2 2N4401
R29
R5 100
6.8 U1 R45 C26
5w LM317 1K 9 K6 47uF
13 35V
VIN VOUT 11
ADJ
R19
243 D13
1N4007
C15
100 nF
R18 C14
4K32 10uF
1 50V 1
0 VAC3
+24 VDC +24 VDC-2
19 VAC
F4 1 1 K1 D6 Q1
4A R36 R36 1N4148 2N4403
TS3 NO 10K 10K R4
COM 1K
+24V UNR NC 2 3 R39 R38 16
1 K10 100 10K
4
U5 4
0V D2 1 2 3 4 3 ULN2803
2 FB1002L R38 18
6 K9 10K U5 C32
- + 4 1 2 1 1 uF 9
ULN2803
8
R2 C31
2K7 TP5 10 nF 9 R3
C8 C9 C10 C11 2w GND PWR +24 VDC 2K74
2200uF 2200uF 2200uF 2200uF
50V 50V 50V 50V 1
R36
10K
3
4
TS2 K1 R14 R17 C1
NC 4K12 4K12 R39 2.2 uF
17
CW ROT COM 100
1 U5
B
NO TP3 2 D23 4
ULN2803
CCW ROT CW 1N4148
2 W1
A
CW
CW SW R13 9
3 15
D8 2K +24 VDC U5
CCW SW BZX55C10V C12 4
4 K2 NC 100 nF ULN2803
1
BRAKE COM R36
5
B
NO TP4 10K D24 9
GND CCW CCW 1N4148
6 W2
A
R16 9
3 D3 2K 3
J4 8 7 R39 R38 13
100 10K U5
DET CW 8 7 6 5 6 ULN2803 +24 VDC
1 R38 12
DET CCW 6 K8 10K U5 C34
2 4 7 8 7 1 uF 9 2 3
ULN2803
GND 11 K9 8 R37 R37
3 13 C33 3K3 3K3
CW LIMIT 9 K7 11 10 nF 9
4 13 1 4
CCW LIMIT 11 K8 9
5 13
ENABLE 9
6 R15 16 1
2 +12 VDC 5 47.5 U6 2
7 R30 TLP621-4
ROT SPEED C27 10K R27 14 15 2
8 100 nF 10K U3 K10 D25
-12 VDC U4B 6 3 5 ULN2803 1N4148
9
8
18 R24
U3 3K32
C28 ULN2803 1
100 nF
C21
9 100 uF
35V
1 1
+24 VDC-2 Q3
2N4403 W3
A B
1 1 K4 2,13 D11 J5
R42 R42 H-UP 1N4148
10K 10K
1 H BRAKE
6 R44 D28
H-UP 2 GND
2 3 R41 R40 15 1K 1N4007
8 100 10K 1
U7
R37 2 1 3 4 4 ULN2803 R42 R43
3K3 R40 18 10K 3K32
10K U7
2 1 1 C35 9
7 ULN2803 4
220 uF
C37 50V
7 10 10 nF 9
U6 16 D26 3
TLP621-4 U7 1N4148 R41 C39
8 9 3 ULN2803 100 100 uF
35V
4
9
14
U7
3 +24 VDC-2 5 ULN2803
3
9
1 1 K5 2,13 D12
R42 R42 H-DWN 1N4148 D27
10K 10K 1N4148
H-DWN 6
5 7 R41 R40 13
6 100 10K U7
R37 7 8 8 7 6 1
ULN2803
3K3 R40 12 R42
10K U7 10K
5 6 7 C36 9
5 ULN2803
220 uF 8
C38 50V
5 12 10 nF 9
U6 11
TLP621-4 U7
6 11 8 ULN2803
2 2
1 1
U1 +5V C1,C2,C3
ICM7228AIBI 100nF
D0 12 18 KEY PAD
D1 ID0 D8 U1,U2,U3
11 ID1 D7 21
D2 13 22 ICM7228AIBI 19
D3 ID2 D6
14 ID3 D5 17 S5
D4 10 20 +5V
ID4(SHUT) D4 V+
D5 6 23 S7 J1
D6 ID5(DEC) D3
1 5 ID6(HEXA) D2 16 1
D7 7 15 S8
ID7(DATA) D1 1
12
GND 2
9
7
3
4
9 MODE DP 4 S6 3
8 27
C5
C4
C3
C2
C1
WRITE SEGA 28 4
SEGB 3 S1 5
SEGC 1 1 R1 6
2 2 S9 LOW SPEED
SEGE R2 UNLOCK 7
SEGG 26 11 R3 8
24 10 D1 S3 HEIGHT UP
SEGF R4 ANGLE UNDERTABLE 9
SEGD 25 8 R5 10
5 UNITS S2 INCREASE SID
R6 HEIGHT DWN 11
6 R7 12
S4 ROTATE CW
DECREASE SID 13
THORAX 14
S10 15
ROTATE CCW
16
U2 S6 S7 S5 S4
ICM7228AIBI UNDERTABLE UNLOCK LOW SPEED THORAX
D0 12 18
D1 ID0 D8
11 ID1 D7 21
D2 13 22
D3 ID2 D6
14 ID3 D5 17
D4 10 20
D5 ID4(SHUT) D4
6 ID5(DEC) D3 23
D6 5 16
D7 ID6(HEXA) D2
7 ID7(DATA) D1 15
12
9
7
3
4
2 9 4 2
MODE DP
8 27
C5
C4
C3
C2
C1
WRITE SEGA
SEGB 3
SEGC 1 1 R1
SEGE 2 2 R2
SEGG 26 11 R3
24 10 D2
SEGF R4 ANGLE
SEGD 25 8 R5
5 TENS
R6
6 R7
U3 S10 S3 S2 S1 S9 S8
ICM7228AIBI ROT CCW ROT CW DECREASE SID INCREASE SID HEIGHT DWN HEIGHT UP
D0 12 18
D1 ID0 D8
11 ID1 D7 21
D2 13 22
D3 ID2 D6
14 ID3 D5 17
D4 10 20
D5 ID4(SHUT) D4
3 6 ID5(DEC) D3 23 3
D6 5 16
D7 ID6(HEXA) D2
7 ID7(DATA) D1 15
12
9
7
3
4
9 MODE DP 4
8 27
C5
C4
C3
C2
C1
WRITE SEGA
SEGB 3
SEGC 1 1 R1
J2 SEGE 2 2 R2
SEGG 26 11 R3
CS5 24 10 D3
1 CS4 SEGF R4 ANGLE
2 SEGD 25 8 R5
MODE 5 HUNDRED
3 CS6 R6
4 6 R7
D1
5 D0
6 D3
7 D2
8 D5
9 D4
10 D7
11 D6 +5V
12
13
14
15
16
4 4
TP2
D D
C4
470nF
+2.5V
+2.5V
C5 +2.5V
470nF U5 C10
100nF
43 P0.0 AVDD 5
R5 R3 44
C14 C11 240 270K P0.1 C18 C19 +2.5V C7 U2
45 34
16
100 nF 100 nF P0.2 DVDD 100nF 100nF 1uF MAX232CSE -2.5V
46 P0.3 DVDD 20
Galga1 Galga3 49 48 2 C1+ 1
V+
P0.4 DVDD -2.5V -2.5V R6 V+ C6
50 P0.5
J1 R4 51 16 1k
P0.6 P3.0 RX
U4A 10K U1B 52 17 W1 6 3 1uF
white P0.7 P3.1 TX V- C1-
7
1
8
-2.5V 6 18 4
1 1 - P3.2 C2+
black 2 2 3 . 7 1 P1.0 P3.3 19 C9 C8
red 6 5 2 22 CALIBRACION 1uF
green 3 3 + P1.1 P3.4 1uF
4 4 2 3 P1.2 P3.5 23 W2 C2- 5
AD620 LMC6482 +2.5V 4 24
P1.3 P3.6 R10 100 -2.5V -2.5V
11 P1.4 P3.7 25 14 11
12 A B
5
+2.5V P1.5
13 P1.6 ALE 42 13 12
Galga2 Galga4 -2.5V 14 41 +2.5V -2.5V J3 -2.5V
EARTH STUB -2.5V P1.7 PSEN#
EA# 40 1 7 10 TX
-2.5V 28 R9 100
R7 P2.0 2
29 33 8 9
GND
P2.1 XTAL2 3 RX
10K 30 32
C TP3 P2.2 XTAL1 4 C
31 P2.3
R8 36 27 C12 11.059Mhz C13 CON4
P2.4 SDATA/MOSI
15
100 37 26 XT1
P2.5 SLOCK
33p
33p
\MOV_START 38 P2.6
39 +2.5V
LIMIT P2.7
9 10 -2.5V -2.5V -2.5V
DAC0 DAC1 -2.5V -2.5V
C15 6 3 15 RESET VREF 8
33p 7
V1 V1 CREF
35 DGND
MMQA5V6T1 21 6 C17 C16
-2.5V DGND AGND
47 DGND
2-5 100nF 100nF
ADuC831
\MOV_START
-2.5V
LIMIT
+2.5V +2.5V
+2.5V W3
+2.5V
U3
B B
8 ST24C16
R2
1k U1A 1 5
C1 A0 VCC SDA
2 - 2 6
100 nF U1 A1 SCL 1
. 1 3 A2 2
8
LMC6482 3 +
C3
WC 7 1 3
J2
6 100nF
-2.5V LMC6482 GND C2 C20 C21 C22 V1 4
7
5 MMQA5V6T1
R1 4 10uF 10uF 10uF 10uF
1k 2-5
TP1
4
-2.5V
A
90045-03 A