Professional Documents
Culture Documents
Special Cleaning Processes
Special Cleaning Processes
Special Cleaning Processes
69
Figure 5.1 Illustration of the drag method: (a) position a small wipe over the optical
surface, and (b) using a solvent, squeeze 2–4 drops directly on the wipe. (c) Slowly
pull the wipe across the optic while maintaining surface contact, and (d) continue
pulling the wipe—the solvent will evaporate behind the edge of the wipe. As the wipe
nears the edge, continue pulling the wipe up and away from the surface.
Figure 5.2 After the strip coating has dried, slowly pull the coating up and away
using a pull tab.
Pull tabs (supplied by the manufacturer) are then applied on the hard
coating with light pressure (see Fig. 5.2). Consult the manufacturer’s
instructions to see if cellophane or masking or Kapton® tape can be used
as a pull tab. After applying the tab or tape, rub across its surface to ensure
that it sticks evenly. Remove the coating by slowly but firmly pulling the
tape up at an angle. The strip coating should pull off. Observe and check
the optical surface as it is being pulled; if any strip-coating residue is left on
the optical surface, apply the liquid again and repeat the process.
clear coating may not be visible, so it is best to assume that the optic is
not coated. Handle the plastic optic with extra care. The use of hand
tools should be avoided during the cleaning process (see Fig. 5.3).
Remove any dust or fibers using a soft brush or compressed air.
A microfiber wipe, a soft clean-room wipe, or a swab can be used.
Fold the wipe to the size required, and use a corner of the fold to get
into small areas. Observe the result: if fingerprints or oils are still
visible, a cleaning liquid may be needed.
Before using any liquids, however, the optic should be removed
from the housing or device. Always check the manufacturer’s manual
before proceeding, as the optic may require special assembly, disassem-
bly, or testing before and after the process. The plastic surface should
be accessible enough to reduce the risk of damage while cleaning (see
Fig. 5.4). Make sure that no liquid can leak into the device.
Figure 5.4 Use small pointed-tip applicators to reach enclosed plastic optics.
surface to remove stains, oils, and glue. It can be used when other
cleaning methods have failed or as part of the final cleaning of either
an optical surface or a ground edge/side. Magnesium oxide (MgO)
talc powder can be substituted for chalk.
Prepare the workstation with the tools needed; note that the
station can get wet during this process. Place a large wipe on the
worktable, centered under the light source. Add a 1- to 2-mm-thick
pad as a cushion, and place the optic on the pad. Transfer
about 10–20 g of chalk from its large container into a smaller one.
Cleaning requires about 1–2 g; sprinkle this amount in a small,
clean, flat dish.
The chosen swab or wipe should be free of particles. A wipe is
used to clean a large optic (>50 mm), whereas a swab is used for
smaller optics. However, a swab can be used to clean a localized area
on a large optic.
Dampen a folded wipe by either dipping its corner into a dish of
soap and distilled water or spraying the solution on the wipe using a
spray bottle. Take the dampened corner of the wipe and touch the
outer edge of the chalk to collect a small amount. Place the corner
on the area of the optical surface to be cleaned. Use a slight
downward pressure to scrub the surface in a small circular pattern
(see Figs. 5.5 and 5.6). Do not allow the chalk to dry during the
process—clean excess chalk off of the surface, and then inspect
using the light. Repeat the process as needed, but dampen a new
corner each time to prevent cross-contamination. Finish cleaning
by using a swab dampened with a solvent, and then inspect the
optical surface using the light source. The drag method may be
useful after this procedure. Be sure to remove any remaining chalk
particles from the edges.
Figure 5.5 Apply light pressure when using a damped wipe and chalk.
Figure 5.6 Apply light pressure while using an applicator to clean, using multidirec-
tional movements, such as circles and short back-and-forth strokes.
Figure 5.7 CO2 nozzle and control applicator with a pressure gauge (image
courtesy of Applied Surface Technologies).
Figure 5.8 Optical surface before the CO2 cleaning process (image courtesy of
Applied Surface Technologies).
Figure 5.9 Optical surface after the CO2 cleaning process (image courtesy of
Applied Surface Technologies).
Figure 5.10 UC Berkeley physicist Charles Townes cleaning one of the large
mirrors of the Infrared Spatial Interferometer (image courtesy of Cristina Ryan, 2008).
and the snow should flow away from them. For safety reasons, the
area should have good airflow. Do not let the CO2 and nitrogen
accumulate in the room or under the flow bench.
Figure 5.11 Cleaning the surface of a cell phone screen using a carbon-tip wipe.
Figure 5.12 Cleaning the eye lens of an eyepiece with a small brush before using
the carbon tip.
surface and slowly moved in a spiral motion from the center out to
the edge (Fig. 5.13). During the final rotation, lift the tip up and
away from the surface.
High-rotation (>1000 rpm) spin cleaners were originally developed for the
semiconductor wafer process. These cleaners use spray cleaning, rinsing
liquids, and drying with gases under high pressure.
Figure 5.13 Cleaning the eye lens of an eyepiece using a carbon tip.
Figure 5.14 A motorized spin cleaner, with foot pedal and digital speed controller
(image courtesy of Headway Technologies).