Download as pdf or txt
Download as pdf or txt
You are on page 1of 11

METTLER TOLEDO GmbH, AUGUST 2017 1

Regulatory and Practical Aspects of Wavelength


Accuracy in UV/VIS Spectrophotometers
Mario Crevatin and Heiner Saßmannshausen
Mettler-Toledo GmbH

Abstract—Wavelength aspects of UV/VIS spectropho- The same sensitivity towards wavelength accuracy
tometers are discussed in this paper. The wavelength re- remains in the higher-level time-dependent kinetic or
quirements of different pharmacopeias, ASTM standards concentration-dependent quantification methods.
and applications are outlined. This article is organized as follows: In section II,
Wavelength calibration and adjustment1 with mercury
the wavelength-accuracy requirements for UV/VIS spec-
argon, deuterium, and xenon lamps as well as different
certified reference materials (CRMs), which are liquid trophotometers stated in pharmacopeias and for high-
solutions or glass containing holmium-, didymium- or demanding applications are presented. In these docu-
other rare earth oxides, are analyzed with the focus ments, calibration standards are recommended. In section
on their wavelength-range coverage, accuracy limits and III, these standards, ranging from gas-discharge lamps
usability. to rare earth solutions and rare earth ion doped glass,
Three different pharmacopeia-compliant scanning and are discussed and compared. Based on measurements
array spectrophotometers are tested with respect to wave- on three different instruments, section IV is dedicated
length accuracy and repeatability. The role of resolution,
to an analysis of theoretical and practical aspects re-
the number of data points on the wavelength axis, software
features and calibration procedures are investigated and
lated to wavelength accuracy and calibration procedures.
recommendations based on practical experiences are given. The section contains valuable information for users of
UV/VIS spectroscopy to better understand the instru-
ments’ behavior and performance. The article ends with
a short conclusion.

I. I NTRODUCTION
II. R EGULATORY AND APPLICATION REQUIREMENTS

W AVELENGTH accuracy and repeatability are key


performance aspects of UV/VIS instruments. An
accurate calibration of the wavelength axis is crucial for
To understand the requirements on wavelength aspects
we first need to understand the needs of regulatory
standards and applications.
every application relying on a spectroscopic technique
and is the basis for its use as a reliable analytical tool.
For example, the typical UV/VIS measurement methods A. Pharmaceutical and ASTM requirements
of fixed-wavelength or scanning deliver two types of Pharmaceutical regulations, so called pharmacopeias
results both sensitive to the wavelength accuracy of the [2]–[5], list quantitative requirements of UV/VIS spec-
instrument: Absorbance or transmittance values at pro- trophotometers. These requirements are separated into
grammed wavelengths and wavelengths values of spec- the four independent categories of wavelength accuracy
tral peaks or valleys. If there is an error in the wavelength and repeatability; photometric accuracy, repeatability and
x-axis, both results will deviate from the “true values”. In linearity; resolution; and straylight. The first require-
the first case, the measurement actually takes place at a ments concern wavelength accuracy, and the perfor-
wavelength different from the programmed wavelength. mance tests for the latter requirements are carefully
In the second case, the position of the spectral peak is designed to eliminate influences from it. These phar-
wrongly translated into a wavelength value. macopeia wavelength accuracy requirements are sum-
marized in Table I. Most countries and laboratories are
1
Calibration and Adjustment [1]: In this document we use cali- referring to and testing in line with the European or
bration for the evaluation of the deviation, in particular wavelength
US Pharmacopeia (USP). Among the pharmacopeias, the
deviations, of values compared to target values from a reference. An
adjustment is the process to reduce these deviations to the minimum US version has the most detailed outline on the test
possible and store the correction for future calibrations. parameters to be used.
2 METTLER TOLEDO GmbH, AUGUST 2017

TABLE I TABLE II
WAVELENGTH REQUIREMENTS OF DIFFERENT P HARMACOPEIA P HARMACOPOEIA & ASTM APPLICATIONS WITH THIGH
AND ONE ASTM STANDARD TOLERANCES ON WAVELENGTH ACCURACY & REPRODUCIBILITY

Pharmacopeia Range Accuracy Repeatability Application Range Accuracy Rep.


EU & UK, UV range 200 − 400 nma ±1 nm - Cyanocobalamin peaks 278 & 361 mm ±1 nm -
EU & UK, VIS range 400a− 800 nm ±3 nm - USP peak 550 nm ±2 nm -
EU & UK single wavelengthb ±2 nm - Aspartame peaks 247 & 252 nm ±1 nm -
US, UV range 200 − 400 nm ±1 nm 0.5 nm stdc USP peaks 258 & 264 nm ±1 nm -
US, VIS range 400 − 700 nm ±2 nm 0.5 nm stdc Thiophene in Benzene 589 nm - ≤1.0 nm
CN, UV range - ±1 nm - ASTM D1685-00
CN, VIS range 500 nm region ±2 nm - Monoethylene Glycol 220.6 nm ±0.5 nm ≤0.3 nm
JP, optical filters 200 − 800 nm ±0.5 nm ±0.2 nmd ASTM E2193-02
JP, Hg or De peaks 200 − 800 nm ±0.3 nm ±0.2 nmd Petroleum Products 220 − 400 nm ±0.2 nm ≤0.2 nm
ASTM E925-09 [6] 200 − 700 nm ±1.0 nm ±0.1 nm ASTM D2008-12
a No exact transition wavelength between UV and VIS defined in the EU
pharmacopeia.
b Accuracy criteria if only a single wavelength is required to be measured. TABLE III
c Standard deviation (calculated from six measurements) not to be exceeded. N UMBER OF PEAKS RECOMMENDED FOR
No requirement for array instruments. WAVELENGTH - CALIBRATION STANDARDS IN PHARMACOPEIAS
d When the measurement is repeated three times, each value obtained needs
to be within this tolerance around the mean. Pharmacopeia EU US CN JP
calibration source & UK
Mercury argon lamp 10 8 11 4 optional
B. Application requirements
Deuterium lamp 1 2 2 2 optional
The requirements on the wavelength accuracy coming Holmium oxide solution 4 14 11 not quantified
from applications are illustrated by six examples pre- Holmium oxide glass - 11 optional 9 -
sented in Table II. These examples comprise two mea- Rare earth & Didymium - optional - -
surements which are part of the USP, and three examples
selected from 12 available ASTM applications2 . They are
selected among many because they impose high demands line sources. Next, we will describe the most relevant
on wavelength accuracy for UV/VIS spectrophotometers. certified reference materials (CRMs). These CRM stan-
It should be noted that many applications have require- dards are typically delivered in the form of 1 cm optical
ments on the photometric accuracy, which, particularly path length cuvettes from different vendors.
for applications with narrow peaks and valleys, translates
into a tolerance for wavelength accuracy. Nevertheless, A. Mercury argon reference lamps - HgAr
most applications include broad spectra and are therefore In gas discharge lamps, atoms in the gas phase are
not particularly sensitive to wavelength deviations (e.g. electronically excited using an electric discharge and
ASTM D4046, where the absorbance is to be measured then decay upon emission of light at characteristic
at 452 nm ± 5 nm) or include a quantification method wavelengths. Consequently, the intensity spectra of such
with multiple standards. For these reasons, many appli- lamps consist of narrow lines, the wavelength positions
cations are insensitive to wavelength deviations. of which trace back to the internal energy structure
of the corresponding gas-phase atoms. In contrast to
III. WAVELENGTH - ADJUSTMENT STANDARDS AND broad absorption bands of solutions, those positions are
WAVELENGTH - ACCURACY TEST METHODS stable, precise and traceable back to the metric SI unit.
Those are the main reasons why these lamps are used as
In this section, the technical limits and uncertain-
a primary wavelength-adjustment standard for UV/VIS
ties as well as practical aspects of the most common
spectrophotometers.
wavelength-accuracy test and calibration methods are in-
HgAr lamps in particular, containing a mixture of
vestigated. The focus lies on the standards recommended
mercury vapor and argon gas, are used because of several
in the different pharmacopeias, as presented in Table III.
distinct advantages. They are widely available, easy to
The numbers in this table denote the stated number of
use and durable. Their line spectrum is well understood
test wavelengths for each method. First, we will discuss
and features many narrow lines spanning the full UV/VIS
the use of mercury argon, deuterium and xenon atomic
part of the electromagnetic spectrum. 19 positions are
2
ASTM C25, D1685, D2008, D2269, D3231, D4046, D4590, known to very high precision, with uncertainties being
E533, E819, E1615, E1784, E2193 on the order of 10−4 nm [7]. Their importance is revealed
METTLER TOLEDO GmbH, AUGUST 2017 3

when considering that the spectrophotometers used to B. Instrument-internal tungsten-deuterium and xenon
determine certified reference substances are themselves lamps - TgDe / Xe
calibrated with HgAr lamps (see, e.g., [11]).
The majority of spectrophotometers relies on tungsten-
Conceptually simple, wavelength adjustment with deuterium or xenon lamps as a light source. The emission
HgAr lamps exhibits some practical complications: i) spectra of these lamps also feature sharp lines that can be
The light from the external lamp needs to be inserted into used for wavelength calibration. Using the internal lamp
the beam path of the spectrophotometer. In the design for calibration requires no additional equipment, no test
of some (mostly scanning) spectrophotometers this is substances, and no time-consuming calibration setup and
solved by special slots to insert pen-shaped HgAr lamps installation by the user. Because of these advantages, it
(e.g. LOT Oriel, Pen-Ray line sources). A convenient so- has become an established procedure.
lution for array-based instruments are HgAr lamps which However, the disadvantage lies in the fact that only
send the light through an optical fiber to an adapter that a few lines have accurately known spectral positions
directly fits into the standard 1-cm cuvette holder (e.g. and can actually be used as reference lines. In case
Ocean Optics, HG-1 Mercury Calibration Source). ii) of the deuterium lamp there are two peaks at 486.0 nm
The intensity differences among the lines of HgAr lamps, and 656.1 nm which are recommended as references in
particularly between the very intense lines of mercury the pharmacopeias, and in case of Xe lamps various
and the much weaker lines of argon, are typically larger single peaks are used, but none is clearly established
than the dynamic range of the spectrophotometers. This in regulations. Effectively, this results in a two- or one-
requires separate measurements with adopted illumina- point calibration or adjustment, which cannot compete in
tion times for the differently-intense spectral regions. performance with the previously mentioned HgAr lamp.
iii) Instrument-specific behavior that has an influence
on the wavelength-adjustment, i.e. wavelength-dependent
detection sensitivity, resolution-dependent shapes of the C. Holmium oxide solution
recorded emission lines and geometrical aspects. These
Holmium oxide (4%) in perchloric acid (10%), sealed
complications all dictate the use of special algorithms
in 1 cm optical path length cuvettes, is typically mea-
outside of the instrument’s application toolbox to fully
sured against air as blank (no cuvette at all) for cal-
exploit the HgAr lamp’s advantages for wavelength
ibration. Intensive work has been done on this stan-
adjustment.
dard by Travis et al. [10]. They report on tests that
Typically, the wavelength adjustment with an HgAr were performed on 17 instruments in 14 countries with
lamp requires the following successive steps: i) Selection spectral bandwidths of 0.1 nm, 1.0 nm and 3.0 nm and a
of reference lines and definition of their positions. ii) wavelength sampling interval of 0.1 nm. Peak positions
Recording the lamp spectrum, usually involving different were determined to the nearest 0.01 nm by use of peak-
illumination times for the differently-intense spectral re- location algorithms. All instruments were adjusted with
gions. iii) Spectral smoothing and interpolation. iv) Peak respect to atomic spectral lines prior to the measure-
finding of the chosen reference peaks, thereby connecting ments. Figure 1 shows the spectrum of an holmium oxide
the wavelength λ with an instrument parameter x, such solution in the top panel.
as pixel number (in case of array instruments) or a In the following, the paper of Travis et al. [10] is
generalized position of a dispersive element (for scan- summarized:
ning instruments). v) Fitting of a parametrized formula a) Temperature sensitivity: The positions of ab-
λ(x), which can be a simple polynomial of the form sorption maxima of holmium oxide solutions have shown
λ(x) = a0 + a1 x + a2 x2 + a3 x3 , in order to obtain this to be very temperature insensitive. Peak shifts < 0.05 nm
relationship not only on the positions of the reference have been reported in the range between 25◦ C − 45◦ C.
lines but over the whole spectrum. The determination of Nevertheless it is recommended to use holmium oxide
parameters with physical meaning have been suggested solution between 20◦ C and 30◦ C.
instead of polynomial coefficients [8]. b) Concentration sensitivity: Changes in the con-
Such HgAr-lamp based adjustments can achieve a centration between 10 − 60 g/l result in peak shifts of
wavelength accuracy of 0.1 nm over the whole spectrum less than 0.05 nm.
on instruments with 0.2 nm data intervals, using sophis- Proper concentration adherence is in the responsibility
ticated calibration algorithms (but are limited by the in- of the certified reference material supplier. It is recom-
strument’s actual technical capabilities, see section IV-C) mended that concentration variations shall be within 20%
[8], [9]. compared to the nominal value of 40 g/l.
4 METTLER TOLEDO GmbH, AUGUST 2017

D. Holmium oxide glass


0.8 Holmium oxide solution
0.6
Solid holmium oxide glass filters (fitting in stan-
0.4 dard 1-cm-cuvette holders) have been used since 1961
0.2 as wavelength-calibration references [13]. A measured
0 spectrum is shown in the middle panel of Figure 1. These
filters provide an alternative to the holmium oxide solu-
Absorbance

Holmium oxide glass


1.5
tion, but only provide 11 wavelengths in the same spec-
1
tral range. In a recent publication [14], several problems
0.5
of these filters are outlined, among which are the ab-
0 sorption cut-off effect of the glass itself for wavelengths
Didymium oxide glass below 300 nm and the necessity of regular recertification
2
due to their hygroscopic nature. Practical experiments
1 with such a filter on a Mettler-Toledo UV7 spectropho-
tometer have shown that the lowest wavelength peak at
0
200 300 400 500 600 700 800 900 241.7 nm can still be just resolved on a huge absorption
Wavelength [nm] background from the glass. Towards longer wavelengths,
the peaks at 536.5 nm and 638.0 nm are very broad
Fig. 1. Spectra of different certified reference materials measured and not recommended as reliable calibration standards.
on a Metter-Toledo UV7 spectrophotometer. Upper panel: Holmium
Consequently, 8 reference peaks remain in the range
oxide solution; middle panel: holmium oxide glass; lower panel:
didymium-oxide glass. Sharp peaks are highlighted with stars ∗ and from 279.4 nm to 460 nm. The peak positions are found
optional peaks with larger uncertainties in their peak position are in the vicinity but not at the same wavelengths as those
denoted with ◦. of a holmium oxide solution [15]. Regarding this type of
holmium oxide glass filter, the possibility of differences
between glass batches is mentioned in the USP. The
c) Contamination sensitivity: Baseline absorbance uncertainty of the certified wavelength positions lies in
values at 230 nm greater than 0.3 A (<50%T) indicate the range of ±0.11 nm [16] to ±0.20 nm [15].
problems with contamination of the solution or of the
cuvette surface. E. Rare earth & didymium glass filters
It is therefore crucial to prevent finger prints or other Rare earth and didymium glass filters expand
contaminations on the cuvette surfaces. the wavelength range towards the lower and upper
d) Resolution sensitivity: Based on a mathematical wavelength range [17]. USP refers to the didymium
model, the positions of 14 absorption peaks are provided filters (solution and glass) as an option to expand
for 0.1 nm steps of the spectral bandwidths (SBW) the wavelength range up to 865 nm if the application
between 0.1 nm up to 3.0 nm SBW (and 1 nm SBW steps requires it. The lower panel of Figure 1 shows the
up to 10 nm SBW). The calculated values have 95% spectrum. Measurements with a didymium glass filter
confidence interval of 0.05 nm. Because significant shifts on a Mettler-Toledo UV7 spectrophotometer have
of the peak positions can occur, it is crucial to know the shown that some peaks (at 573 nm, 741 nm, 807 nm,
instruments spectral bandwidth (resolution, slit-width) to and 879 nm) only have a partially resolved substructure,
apply the correct reference values from the holmium or lie at a shoulder (431 nm). Practically, that restricts
certificate. This aspect is discussed in more detail in the applicability at longer wavelength up to 749 nm.
section IV-B. Rare earth or didymium glass filters may be useful for
e) Accuracy limits: The 95% confidence interval calibration for applications in the low-UV range or the
for final peaks for 13 instruments is smaller than ± 0.05 VIS to NIR range, respectively.
nm. The prediction intervals P 95 for measuring holmium
peaks on a single instrument on the basis of a single
spectrum are between ±(0.16 − 0.26) nm for a SBW of F. Holmium-didymium oxide solutions
1 nm. For a dense and broad wavelength-range coverage,
Starna Scientific Ltd. and Hellma GmbH & Co. KG state mixtures of holmium oxide and didymium oxide in solu-
on their holmium oxide solution certificates uncertainties tion are available as CRMs [18], [19]. The recommended
(k = 2, 95%) of ±0.11 nm [11] and ±0.20 nm [12], reference peaks range from 241.12 nm to 794.16 nm and
respectively. include 7 peaks of holmium and 7 peaks of didymium.
METTLER TOLEDO GmbH, AUGUST 2017 5

TABLE IV A. Optical setup


DATA FROM CERTIFICATES OF WAVELENGTH - CALIBRATION
STANDARDS AND PRACTICAL LIMITS
There are two main categories of UV/VIS spectropho-
tometers that can be differentiated by their basic optical
Calibration source Rangea #WLb Uncertainty design:
Mercury argon lamp 253 − 922 nm 19 10−4 nmc a) Scanning: The original and most common type
Deuterium lamp ∗486.0, 656.1 nm 2 - are scanning instruments. Such systems scan the wave-
Holmium solution ∗241 − 641 nm ∗14 0.11 / 0.20 nmd length over the desired range by moving an optical
Holmium oxide glass ∗280 − 460 nm 11, ∗8 0.11 / 0.20 nme element inside the instrument. More specifically, a light
Rare earth solution 201 − 253 nm 5 -
beam of the selected wavelength passes through the
Didymium solution 731 − 864 nm 5 0.20 / 0.40 nmf
Didymium glass ∗473 − 749 nm 11, ∗6 0.11 / 0.20 nmg
sample and is measured by a detector. Sequentially,
Ho-didymium solution 241 − 794 nm 14h/ 7i 0.20 / 0.40 nmi and in tunable increments, the wavelength positions are
a Wavelength range according data sheets, or, if marked with an ∗, practical scanned in a pre-defined interval, typically taking a
recommended limits.
b Number of wavelengths total and practical (indicated by ∗).
scan time of seconds up to minutes. More time-efficient
c Value taken from [7]. measurements are available in “fixed wavelength” modes
d Certificates of Starna Scientific Ltd. [11] and Hellma Analytics [12].
e Certificate of Starna Scientific Ltd. [16].
where the absorbance is determined only at a single
f Certificates of Starna Scientific Ltd. [11] and Hellma Analytics [20]. wavelength. Many pharmaceutical test procedures are
g Certificates of Hellma Analytics [21] and Starna Scientific Ltd. [22].
written according to the specific behavior of scanning
h Recommended by Starna Scientific Ltd., private communication.
i Certificate of Hellma Analytics [23]. instruments.
Due to the mechanical moving parts of scanning
instruments they are prone to misalignments and, as a
Because the spectrum of the mixture is spectrally dense, result, a loss of wavelength accuracy or repeatability
it is recommended to use Ho-didymium mixed solutions over time. To counteract these drifts, the tested TE-
only for instruments with a high resolution better than 260 instrument and, as reported in [24], a Perkin Elmer
1 nm. instrument perform a wavelength calibration with the
peaks of the internal light source as reference at start-up
In Table IV, the findings of this section are sum-
(TE-260: xenon lamp peaks at 484.45 nm and 541.90 nm
marized. The table contains information from different
or Perkin Elmer deuterium lamp peak at 656.1 nm).
sources combined with recommendations based on our
This process takes around half a minute for two Xe
own practical experiences (indicated with a star).
wavelengths on the TE-260.
b) Array: Array based instruments are capable of
recording the full spectral range in a single measure-
IV. P RACTICAL I NSTRUMENT A SPECTS ment. These instruments do not have mechanical moving
components but comprise a spectrograph with an array
In this section, wavelength aspects related to the spec- of sensors dedicated to the different wavelengths. To
troscopic instrument used, the optical setup and software obtain the full spectrum, white light of all wavelengths is
features are outlined. We have performed measurements passed through the sample and later spectrally analysed.
on a Thermo Fisher Scientific Evolution 260 Bio (TE- The measurement time only depends on the required
260) scanning instrument and two array based systems, averages (i.e. the signal-to-noise ratio), and typically
Agilent 8453 (A-8453) and Mettler Toledo UV7 (MT- takes between sub seconds and a few seconds. Because
UV7) to demonstrate practical wavelength related issues array systems have no moving parts they are generally
of UV/VIS spectrophotometers. This chapter is not in- less prone to the wavelength stability problems outlined
tended to compare the performance or features of these for scanning instruments. This has been recognized by
particular instruments, but to study instrument-dependent the USP, which requires only a test of wavelength
aspects related to wavelength accuracy and their impact accuracy at a single wavelength and no repeatability
on applications. The Agilent and Thermo Fisher Scien- test for array instruments [4]. Regardless, both tested
tific instruments were not updated to the latest software array instruments provide means for calibration with
version, nor were they adjusted by a service technician the holmium oxide solution and with the internal light
prior to our measurements. For this reason, no judgment sources.
related to a specific brand is intended and future software However, wavelength-accuracy changes can also occur
releases may outdate any brand-specific statements. Our with array spectrographs, primarily because of mechani-
intent is to provide general, but practical considerations cal stress released by temperature changes. To minimize
and information when using these or similar instruments. this effect, spectrographs often undergo an extensive
6 METTLER TOLEDO GmbH, AUGUST 2017

tempering process over a wide temperature range in pro- the TE-260 with 1 nm or 2 nm and the A-8453 with
duction to compensate the internal stress due to mount- 1.5 nm, the holmium oxide solution reference values
ing. After tempering, tests on a MT-UV7 instrument given in Table V are used for calibration. Using the
show no measurable temperature dependence of the 14 correct reference values, taking into account even sub-
peak positions of an holmium oxide solution in the range tle physical effects, minimizes systematic, instrument-
of 10◦ C − 40◦ C [25]. These results confirm the tempera- independent deviations.
ture insensitivity of the holmium oxide solution standard The described resolution dependency of peak positions
and the stability of the array technology. However, small occurs generally with asymmetric lines, and care has to
shifts due to large thermal differences were observed in be taken when comparing application results obtained on
measurements taken with instruments immediately after instruments with different SBWs.
significant temperature changes (e.g., unloading from
cold/hot car and transfer into the laboratory), without 0.8
SBW = 0.1 - 3 nm (1 nm center)
giving the equipment the time to equilibrate. SBW = 1.5 nm
0.6 SBW = 2 nm
For best practices, we recommend using the internal

[nm]
start-up calibration for scanning systems and to always
0.4

peak, SBW = 1 nm
ensure temperature equilibrium with the environment
before wavelength calibration. 0.2

B. Resolution dependency 0
The wavelength positions and photometric values of
absorption peaks are important for many different appli- -0.2

cations in UV/VIS spectroscopy. Using analyte-specific


-0.4
peak positions is beneficial for several reasons: The 200 300 400 500 600 700
relevance of spectral peaks is easily understood; they are peak
at SBW = 1 nm [nm]
easily determined numerically; and, the first derivative of
the signal being zero, peak absorbance values are stable Fig. 2. Sensitivity of 14 commonly used Ho reference peak positions
towards the spectral resolution (in the form of a spectral bandwidth)
with respect to small wavelength fluctuations between
of the measurement. The horizontal axis shows the wavelength
the measurements. positon of the peaks and the vertical axis shows the wavelength shift
However, in the case of asymmetric lines and band- of the peak when measured with SBW between 0.1 and 3 nm. The
shapes, which often occur in solution spectroscopy, the shift is given with respect to the measurement at a SBW of 1 nm.
The green and blue marks highlight the shifts at SBW of 1.5 nm and
peak positions depend on the instrument’s spectral res- 2.0 nm, respectively. Data taken from Ref. [10].
olution, which can lead to unintended peak shifts after
adjustment procedures.
As indicated in section III-C, Travis et al. [10] have
determined 14 peak positions of the holmium oxide
solution as a function of the spectral bandwidth (SBW) C. Technical limitations to peak determination
of the measurement. A part of their results is summarized This section is dedicated to the fundamental accuracy
in Fig. 2, where the range of peak-position differences and repeatability limits that are a direct result of a
compared to a SBW of 1 nm, measured with SBWs discrete measured spectrum. In a discrete spectrum,
between 0.1 and 3 nm, are presented. The positions of the absorbance values are measured in a given interval
SBWs of 1 nm, 1.5 nm and 2 nm are commonly available along the wavelength axis. In scanning instruments, the
on certificates and are highlighted by black bars, green wavelength interval λI can be tuned in certain limits,
dots and blue crosses, respectively, as shown in Fig. 2. whereas in array instruments it is given by the number of
Shifts up to 0.3 nm occur between measurements at single sensors per nanometer. In both cases, interpolation
SBW of 1 nm and 2 nm. Systematic wavelength shifts can be used to decrease the wavelength interval.
on the order of a third of the SBW change are clearly To simulate an application, and to demonstrate
measurable. Consequently, the reference peak values wavelength-interval-related accuracy issues, a holmium
need to be adopted to the instruments specific SBW in oxide solution was measured around the absorption
the wavelength adjustment and calibration processes. peak at 333.48 nm on the three tested instruments. In
For the Mettler-Toledo UV/VIS spectrophotometers Fig. 3, the spectra obtained with the three instruments
UV7, UV5Nano and UV5/UV5Bio, with spectral res- are depicted. The solid vertical line shows the reference
olutions around 1 nm, 1.5 nm and 1.7 nm, respectively, wavelength at 333.48 nm. The wavelength interval λI
METTLER TOLEDO GmbH, AUGUST 2017 7

TABLE V
H OLMIUM OXIDE SOLUTION ABSORPTION PEAKS , MEASURED AT DIFFERENT SPECTRAL BANDWIDTHS (SBW S ) AND USED AS
REFERENCES WAVELENGTHS FOR CALIBRATION , e.g. 1.0 NM FOR THE MT-UV7 & TE-260, 1.5 NM FOR MT-UV5NANO & A-8453 AND
2 NM FOR THE MT-UV5/UV5B IO SPECTROPHOTOMETERS . DATA TAKEN FROM R EF. [10], [11].

SBW
1.0 nm 241.12 249.89 278.13 287.22 333.48 345.38 361.25 385.61 416.25 451.45 467.82 485.23 536.56 640.50
1.5 nm 241.13 249.98 278.11 287.40 333.49 345.38 361.18 385.70 416.42 451.33 467.86 485.26 536.71 640.62
2.0 nm 241.12 250.03 278.10 287.52 333.47 345.42 361.12 385.80 416.57 451.32 467.90 485.25 536.86 640.79

0.2
Reference 333.48 nm
b) Wavelength repeatability: Whereas for the abso-
1 nm, Agilent A-8453
0.2 nm, Mettler Toledo UV7 lute wavelength accuracy a possible error in the order of
0.1 nm, ThermoFisher TE-260
0.15
λI /2 can never be ruled out, the situation is different for
the corresponding repeatability measurement. Repeata-
Absorbance

bility is the standard deviation of several measurements


0.1 of the peak position. In the calculation of the standard
deviation, the average λ̄ of the measurements is used
instead of the known reference value. With n being the
0.05
number of measurements at a specific wavelength and
∆λi = λ̄ − λi being the wavelength deviation of a spe-
0 cific measurement from the average of all measurements,
330 331 332 333 334 335 336
Wavelength [nm]
the repeatability is calculated as
v
u n
Fig. 3. Holmium peak at 333.48nm measured on three instruments u 1 X
with different data intervals of 1 nm, 0.2 nm and 0.1 nm. The mea- λrepeatability = t (∆λi )2 .
n − 1 i=1
surements were taken against air as blank and only the wavelength
positions and not the photometric absorbance values can be compared
between the measurements. Independent of the data interval λI , the value of
λrepeatability can be small, even zero, if the instru-
ments measure with high photometric accuracy the same
for the TE-260 scanning instrument was set to 0.1 nm, absorption values repeatedly at the same wavelength.
whereas the A-8453 and MT-UV7 array instruments This can be seen also in Table VI, where both array
have a fixed wavelength interval of 1 nm and 0.2 nm, instruments have obtained the same peak value in ten
respectively. successive measurements, leading to a standard deviation
a) Wavelength accuracy: The measurement was of zero.
repeated ten times on each instrument. For each mea- Photometric noise is always present to some degree,
surement, the peak positions λi determined by the in- and it is visible in the TE-260 spectrum, but to a lesser
struments’ own algorithms were extracted. The average extent also in the MT-UV7 and the A-8453 spectra.
values λ̄ = n1 · i λi are shown in Table VI. The
P
Due to the photometric fluctuations from measurement
wavelength accuracy is calculated as the difference be- to measurement, the peak position can toggle between
tween the average measured peak position λ̄ and a given two, or even more, wavelength points. This effect is
reference value λref according to enhanced in case of broad peaks, with many wavelength
λaccuracy = λ̄ − λref . points having similar photometric values. In case that
measured peak positions toggle evenly between two
For the two array instruments A-8453 and MT-UV7, the neighboring wavelength points, the resulting effect of the
reference wavelength is located almost in the middle wavelength interval λI on the wavelength repeatability
between two data points in our example (see Fig. 3). This can be calculated as
is the reason why the observed deviations λaccuracy of
n λI
r
the measured peak positions from the reference value λrep. toggling = · ,
correspond closely to λacc. limit = λI /2, which is the n−1 2
maximal wavelength-accuracy error resulting from the where ∆λi = λI /2 was inserted in the general
finite wavelength interval λI . The TE-260 reports the equation of λrepeatability .
peak position at 333.60 nm independent of the data For the three instruments discussed here, we can
interval. calculate this repeatability λrep. toggling on the basis of
8 METTLER TOLEDO GmbH, AUGUST 2017

TABLE VI by parametrisation. Even with these improvements, some


R ESULTS OF MEASUREMENTS OF THE HOLMIUM OXIDE desired peaks were still omitted or picked wrongly by the
ABSORPTION PEAK AT 333.48 NM ON THREE DIFFERENT
SPECTROPHOTOMETERS : WAVELENGTH ACCURACY AND
peak-finding algorithm. The complication for automated
REPEATABILITY peak finding originates from the difficult distinction
between peaks which result from photometric noise or
A-8453 TE-260 MT-UV7
those that result solely from the absorption of the sample.
Data interval λI 1 nma 0.1 nm 0.2 nm
A pragmatic, but powerful solution to this problem, is
Accuracy (n = 10)
λ̄ 333.00 nma 333.60 nmb 333.40 nm
the definition of specific search areas in the application
λaccuracy −0.49 nmc +0.12 nm −0.09 nm method. This can be particularly helpful for scanning
λacc. limit ±0.50 nm ±0.05 nm ±0.10 nm instruments to reduce the overall measurement time.
Repeatability (n = 10) To conclude, the analysis of the tests has shown that
λrepeatability 0 nmd 0.02 nmb 0 nmd there are differences regarding wavelength accuracy and
λrep. toggling 0.53 nmf 0.05 nmf 0.11 nmf repeatability between different UV/VIS spectrophotome-
λrep. s.o. 0.32 nme 0.03 nme 0.06 nme
ters, but also between different measuring modes, peak-
a Peak/valley finding without further mathematical functions.
b Use of interpolation. finding and data-interval parametrisation on the same
c Reference value of 333.49 nm for 1.5 nm resolution [26]. instrument. These differences were traced back primarily
d All obtained wavelength values are equal.
e One single data point deviates from the others
by λI .
to the instrument-specific wavelength data intervals and
fThe peak positions of multiple measurements are evenly distributed if additional interpolation was performed on the data.
over two neighboring data points.
In order to cover the vast majority of applications, it
is sufficient to obtain a wavelength data interval of
the instrument-specific λI values (see results listed in ≤ 0.2 nm, which translates into a maximal technically-
Table VI). The number of measurements n does not have limited wavelength error of λacc. limit = ±0.10 nm. That
a significant impact on this value, that asymptotically accuracy value exceeds the pharmacopeia requirements
approaches λI /2 with increasing measurements. and is of a similar magnitude as the uncertainty of
In another scenario, where the peak is found only once reference materials.
out of n measurements on a neighboring wavelength
D. Application software features
data point, λrep. s.o. is smaller than λrep. toggling but still
surprisingly large. It can be expressed as λrep. s.o. = √λnI In practice during above outlined measurements, it
(where “s.o.” abbreviates “single outlier”). Both repeata- was found that for all applications, particularly for
bility limits, λrep. toggling and λrep. s.o. , are given in typical applications with broad spectral peaks, further
Table VI for the three tested instruments. measurement aspects are as important as the performance
c) Interpolation: In the peak-position determina- specifications. The authors’ suggestions for more time-
tion, some instruments additionally interpolate the ab- efficient working procedures are listed next:
• The possibility of application method programming,
sorbance in between the wavelength points. Thereby, the
above described artifacts can be reduced. The A-8453 e.g., to allow for a pre-defined number of wave-
and TE-260 instruments use interpolation algorithms in length positions to be measured, to obtain only the
the calibration procedure3 for best performance results. desired peak positions.
• In order to prevent error-prone manual data anal-
In normal applications, the A-8453 does not use the same
interpolation function by default, but it can be activated ysis, modern UV/VIS instruments should provide
by the user. The TE-260 provides the interpolation-based mathematical functions capable of calculating de-
peak-finding procedure automatically in all measurement rived application-specific values on the basis of the
modes. The MT-UV7 does not provide an interpolation measured data (e.g., the wavelength deviation from
beyond the wavelength data interval of 0.2 nm. a target wavelength, differences or ratios between
d) Peakfinding sensitivity: During the measurement absorbances measured at pre-defined wavelength
of the holmium oxide solution on the tested instruments positions or peaks, or statistics over multiple mea-
it was revealed that the instruments did not find the surements).
• A simple way of exporting the spectral data (in
same number of peaks for each scan. Generally the user
is interested in an application-specific, clearly defined case the mathematical functions do not deliver the
number of peaks. Therefore, it is beneficial that the peak required results) as well as derived quantities.
• Acceptance limits to validate the calculated results
finding algorithms of the instruments can be improved
according to boundaries set by company or regula-
3
For the A-8453 instrument [27]: ”99-point spline function is used” tory standards.
METTLER TOLEDO GmbH, AUGUST 2017 9

E. Wavelength adjustment & calibration TABLE VII


WAVELENGTH CALIBRATION AND ADJUSTMENT FEATURES
Among other tests such as stray light, resolution, and
photometric accuracy, all instruments provide special Wavelength test A-8453 TE-260 MT-UV7
software features for the user to execute wavelength Accuracy
calibration and repeatability tests. In this short section, Holmium #wavelengths 14 6 14
oxide solution time [min] 1:45a 8:20 0:10
the possibilities offered by the three tested instruments
Internal lamp lamp Tg Xe Xe
are presented (see Table VII). On the Agilent instru-
wavelength(s) 486.0 nm 542 nm 260.6 nm
ment, these tests are part of a defined performance-test 656.1 nm 528.6 nm
sequence. For the other instruments, the tests are freely time [min] 0:15 0:40 0:05
configurable for the user. As outlined in Table VII, the External lamp HgAr − optionalc −
instruments provide the possibility to execute perfor- Repeatabilityb
mance tests with holmium oxide solution or solely with Holmium #wavelengths 14 1 14
the internal light source. For both array instruments the oxide solution 536.6 nm
time [min] 1:45a 3:15 0:25
holmium oxide solution tests are executed for all 14
Internal lamp lamp Xe Xe
wavelength as stated by USP in Table III. All instruments
wavelength(s) − 542 nm 260.6 nm
do provide a performance test report for traceability. As time [min] 6:10 0:25
expected, the array instruments perform these perfor- Adjustment
mance tests faster than the scanning instruments. External lamp HgAr − optionalc serviced
As stated in section III-A on mercury argon lamps, time [min] 0:10
special software features are required to perform the Internal lamp time [min] 0:20 5:10 −
a
wavelength adjustment and calibration procedures, which Time for the holmium oxide wavelength accuracy and repeatability part of
an extended performance test.
are always tailored to a specific reference lamp or b All instrument perform 10 measurements and calculate the standard devi-

material. On the A-8453 and TE-260 instruments, the ation.


c Optionally available as Thermo Scientific EvolutionTM Mercury Lamp
wavelength adjustment is accessible for the customer. Accessories.
d Executed by a service technician and HgAr lamp tool.
There, selected peaks in the spectrum of the internal
lamp are used as reference. For the UV7 instrument,
wavelength adjustment is part of the instrument’s service.
It is performed on the basis of 13 reference lines of an In line with many regulatory standards, the holmium
external HgAr lamp. oxide solution, certified and sealed in 1 cm optical
path length cuvettes, is the first choice for instrument
calibration. It has advantages in the user friendliness,
V. C ONCLUSION robustness, reproducibility and temperature insensitivity.
Typical applications in UV/VIS spectroscopy for 14 peaks in the 241 − 641 nm range can be certified
liquids and solids have broad peaks. Pharmacopoeia with uncertainties of 0.11 nm or 0.20 nm, depending
standards therefore require a wavelength accuracy of on the manufacturer. The cuvettes shall be clean and
±1 nm in the 200 − 400 nm UV range and ±2 nm used between 20◦ C and 30◦ C. Other substances like
in the 400 − 800 nm VIS range. For the wavelength didymium or rare earth oxides can be used to expand
repeatability a standard deviation value of 0.5 nm of the wavelength range from 201 nm to 864 nm with larger
at least six replicates is needed. Specific countries like uncertainties up to 0.40 nm. A mixture of didymium with
Japan, or a few ASTM standards, as outline in this holmium oxide solution is available as well.
paper, have higher requirements. The differences in the wavelength-range coverage
and uncertainties among the wavelength adjustment
To reach these specifications, wavelength adjustment and calibration procedures are large. It is therefore
using mercury argon lamps is recommended. These important that the instrument suppliers clearly state
lamps provide more than a dozen sharp peaks with uncer- which procedures are offered and over which wavelength
tainties in the order of 10−4 nm. Elaborated adjustment range the specifications can be guaranteed with these
algorithms allow for wavelength-accuracy values down procedures.
to 0.1 nm. The instrument internal tungsten deuterium
or xenon lamp peaks provide means for a simple one to The resolution of an instrument can have an impact on
two point calibration or adjustment. In case of scanning measured peak positions. Therefore, the resolution needs
instruments, the internal lamps provide good means for to be considered in wavelength calibration procedures
a start-up performance check. and when comparing application results of different
10 METTLER TOLEDO GmbH, AUGUST 2017

instruments. For the calibration, the correct resolution ACKNOWLEDGMENT


specific reference values have to be taken from the The authors would like to thank John P. Hammond
reference material certificates. If values of the wrong - Technical & Marketing Manager, Starna Scientific
spectral bandwidth are applied, systematic errors of Ltd, Hainault, UK and Birgit Kehl - Compliance Man-
0.3 nm and more are possible. All tested instruments ager Calibration Laboratory, Hellma GmbH & Co. KG,
have these values incorporated in their automated test Müllheim, Germany for their valuable inputs. Further-
sequences for 14 or 6 reference values of the holmium more, we would like to thank Rujuta Bhandarkar, In-
oxide solution. Preferably, wavelength repeatability is ternational Market Support Group, Mettler-Toledo India,
tested at all reference wavelengths. As a compromise for her contribution on the ASTM applications.
for a faster measuring process on scanning instruments,
a subset of the wavelengths might be applied, which
should preferably cover a broad wavelength range.
To reduce the effects of wavelength sampling to the
same level as the uncertainty of certified reference ma-
terials, an instrument shall provide a data point on the
wavelength x-axis every 0.2 nm. This will prevent instru-
ment specific wavelength errors greater than ±0.1 nm
and reproducibility errors greater than 0.11 nm. Inter-
polation can reduce these technical errors further, but
it needs to be available in both the calibration and
application modes.
For fast and application specific wavelength mea-
surements, the instrument should provide comprehensive
software support. The software should enable the user
to scan at application specific peak locations, to use
mathematical result calculations, and to check accep-
tance criteria.
The three tested instruments, among many others,
fulfill the wavelength-accuracy requirements imposed
by the different pharmacopeias. In the measurements
presented here, those instruments performed even better
than specified with wavelength accuracies below 0.5 nm.
With further technical improvements it can be envi-
sioned, that the instrumental accuracy will soon approach
the uncertainties of commonly used and accepted refer-
ence substances, such as the holmium oxide solution. In
the future, it would be desirable that a broader wave-
length range could be addressed with a single certified
reference material containing a combination of rare earth
oxides in solution. Such a combination would need to be
intensely tested and should be incorporated in different
regulatory standards. It would also be desirable to define
common reference peaks of xenon, which can then be
used in efficient calibration procedures of modern xenon
lamp based spectrophotometers.
METTLER TOLEDO GmbH, AUGUST 2017 11

R EFERENCES [22] Starna Scientific Ltd., Didymium glass filter for use as a wave-
length accuracy reference, Sample Certificate Number: 42476,
Set Serial No: 15356, 2013.
[1] JCGM 200:2008, International vocabulary of metrology Basic
[23] Hellma Analytics, Calibration certificate, Liquid calibration
and general concepts and associated terms (VIM), JCGM, 2008.
standard, Type UV45, Serial number: 0015, 2017.
[2] European Pharmacopoeia, 2.2.25 Absorption Spectrophotometry, [24] Steve Upstone, Validating UV/VISible Spectrophotometers,
Ultraviolet and Visible, Rev. 8.0. Technical Note, PerkinElmer, Inc., Seer Green, UK.
[3] British Pharmacopoeia, Appendix II B. Ultraviolet and Visible [25] K.-D. Anders, Mettler-Toledo UV/VIS Spectrometer Ambient
Absorption Spectrophotometry, (Updated Ph. Eur. 9.0, effective Temperature Dependency Evaluation, Technical Note, asens AG,
01/01/2017, Ph.Eur.mmethod 2.2.25), 2017. 2015.
[4] U.S. Pharmacopeia National Formulary USP 40 NF 35, < 857 > [26] Agilent Technologies, Inc., Wavelength accuracy and repro-
Ultraviolet-Visible Spectroscopy, Official from August 1, 2017, ducibility of the Agilent UV-Vis diode array using holmium oxide,
The United States Pharmacopeial Convention, 2017. Technical Note, Publication number: 5991-3446EN, 2013.
[5] JP XVII, The Japanese Pharmacopeia, 17’th edition, English [27] Agilent Technologies, Inc., Agilent 8453 UV-visible Spec-
Version, Official from April 1, 2016, The Ministry of Health, troscopy System, Service Manual, Edition 02/00.
Labour and Welfare of Japan, 2016.
[6] E925-09, Standard Practice for Monitoring the Calibration of
Ultraviolet-Visible Spectrophotometers whose Spectral Bandwith
does not Exceed 2 nm, ASTM International, reapproved 2014.
[7] C. J. Sansonetti, M. L. Salit, and J. Reader, Wavelengths of
spectral lines in mercury pencil lamps, APPLIED OPTICS, 35,
74-77, 1996.
[8] K. Liu and F. Yu, Accurate wavelength calibration method using
system parameters for grating spectrometers, Optical Engineer-
ing, 51, 013603, 2013.
[9] Y. C. Sun, C. Huang, G. Xia, S. Q. Jin, and H. B. Lu, Accurate
wavelength calibration method for compact CCD spectrometer,
Journal of the Optical Society of America A, 34, 498-505, 2017.
[10] J. C. Travis, J. C. Acosta, G. Andor, J. Bastie, P. Blatner, C.
Chunnilall, S. C. Crosson, D. L. Duewer, E. A. Early, F. Hengst-
berger, C. S. Kim, L. Liedquist, F. Manoocheri, F. Mercader, A.
Mito, L. A. G. Monard, S. Nevas, M. Nilsson, M. Noel, A. C.
Rodriquez, A. Ruiz, A. Schirmacher, M. V. Smith, G. Valencia,
N. van Tonder, and J. Zwinkels, Intrinsic Wavelength Standard
Absorption Bands in Holmium Oxide Solution for UV/VISible
Molecular Absorption Spectrophotometry, J. Phys. Chem. Ref.
Data Vol. 34, No. 1, 2005.
[11] Starna Scientific Ltd., Certified Reference Materials for UV and
Visible Spectrophotometry, RH-HL, Holmium solution, Certifi-
cate Number: 47944, Set Serial No: 19781, 2014.
[12] Hellma Analytics, Calibration certificate, Liquid calibration
standard, Type 667-UV5, Holmium solution, Serial number:
0597, 2016.
[13] H. J. Keegan, J. C. Schleter, V. R. Weidner, Ultraviolet Wave-
length Standard for Spectrophotometry, J. Opt. Soc. Am. 51,
1470, 1961.
[14] Ch. Burgess, J. Hammond, Exploding the myths: an update on
UV-visible spectrometry, Spectroscopy Europe, Vol. 21 No. 1,
2009.
[15] D. W. Allen, Holmium Oxide Glass Wavelength Standards,
Journal of Research of the National Institute of Standards and
Technology, Vol. 112, No., 2007.
[16] Starna Scientific Ltd., Holmium glass filter for use as a wave-
length accuracy reference, Sample Certificate Number: 42504,
Set Serial No: 18088, 27, 2013.
[17] Starna Scientific Ltd., A User Guide: Why do we need Reference
Materials?, Starna Certified Reference Materials for UV, Visible
and NIR Spectrophotometry Guide.
[18] Starna Scientific Ltd., private communication.
[19] Hellma Analytics, private communication.
[20] Hellma Analytics, Calibration certificate, Liquid calibration
standard, Type 667-UV25, Didymium solution, Serial number:
0019, 2017.
[21] Hellma Analytics, Calibration certificate, Didymium Glass Fil-
ter, Type 666-F7, Serial number: E1875, 2017.

You might also like