Professional Documents
Culture Documents
Wavelength Accuracy JRNL Aug2017
Wavelength Accuracy JRNL Aug2017
Abstract—Wavelength aspects of UV/VIS spectropho- The same sensitivity towards wavelength accuracy
tometers are discussed in this paper. The wavelength re- remains in the higher-level time-dependent kinetic or
quirements of different pharmacopeias, ASTM standards concentration-dependent quantification methods.
and applications are outlined. This article is organized as follows: In section II,
Wavelength calibration and adjustment1 with mercury
the wavelength-accuracy requirements for UV/VIS spec-
argon, deuterium, and xenon lamps as well as different
certified reference materials (CRMs), which are liquid trophotometers stated in pharmacopeias and for high-
solutions or glass containing holmium-, didymium- or demanding applications are presented. In these docu-
other rare earth oxides, are analyzed with the focus ments, calibration standards are recommended. In section
on their wavelength-range coverage, accuracy limits and III, these standards, ranging from gas-discharge lamps
usability. to rare earth solutions and rare earth ion doped glass,
Three different pharmacopeia-compliant scanning and are discussed and compared. Based on measurements
array spectrophotometers are tested with respect to wave- on three different instruments, section IV is dedicated
length accuracy and repeatability. The role of resolution,
to an analysis of theoretical and practical aspects re-
the number of data points on the wavelength axis, software
features and calibration procedures are investigated and
lated to wavelength accuracy and calibration procedures.
recommendations based on practical experiences are given. The section contains valuable information for users of
UV/VIS spectroscopy to better understand the instru-
ments’ behavior and performance. The article ends with
a short conclusion.
I. I NTRODUCTION
II. R EGULATORY AND APPLICATION REQUIREMENTS
TABLE I TABLE II
WAVELENGTH REQUIREMENTS OF DIFFERENT P HARMACOPEIA P HARMACOPOEIA & ASTM APPLICATIONS WITH THIGH
AND ONE ASTM STANDARD TOLERANCES ON WAVELENGTH ACCURACY & REPRODUCIBILITY
when considering that the spectrophotometers used to B. Instrument-internal tungsten-deuterium and xenon
determine certified reference substances are themselves lamps - TgDe / Xe
calibrated with HgAr lamps (see, e.g., [11]).
The majority of spectrophotometers relies on tungsten-
Conceptually simple, wavelength adjustment with deuterium or xenon lamps as a light source. The emission
HgAr lamps exhibits some practical complications: i) spectra of these lamps also feature sharp lines that can be
The light from the external lamp needs to be inserted into used for wavelength calibration. Using the internal lamp
the beam path of the spectrophotometer. In the design for calibration requires no additional equipment, no test
of some (mostly scanning) spectrophotometers this is substances, and no time-consuming calibration setup and
solved by special slots to insert pen-shaped HgAr lamps installation by the user. Because of these advantages, it
(e.g. LOT Oriel, Pen-Ray line sources). A convenient so- has become an established procedure.
lution for array-based instruments are HgAr lamps which However, the disadvantage lies in the fact that only
send the light through an optical fiber to an adapter that a few lines have accurately known spectral positions
directly fits into the standard 1-cm cuvette holder (e.g. and can actually be used as reference lines. In case
Ocean Optics, HG-1 Mercury Calibration Source). ii) of the deuterium lamp there are two peaks at 486.0 nm
The intensity differences among the lines of HgAr lamps, and 656.1 nm which are recommended as references in
particularly between the very intense lines of mercury the pharmacopeias, and in case of Xe lamps various
and the much weaker lines of argon, are typically larger single peaks are used, but none is clearly established
than the dynamic range of the spectrophotometers. This in regulations. Effectively, this results in a two- or one-
requires separate measurements with adopted illumina- point calibration or adjustment, which cannot compete in
tion times for the differently-intense spectral regions. performance with the previously mentioned HgAr lamp.
iii) Instrument-specific behavior that has an influence
on the wavelength-adjustment, i.e. wavelength-dependent
detection sensitivity, resolution-dependent shapes of the C. Holmium oxide solution
recorded emission lines and geometrical aspects. These
Holmium oxide (4%) in perchloric acid (10%), sealed
complications all dictate the use of special algorithms
in 1 cm optical path length cuvettes, is typically mea-
outside of the instrument’s application toolbox to fully
sured against air as blank (no cuvette at all) for cal-
exploit the HgAr lamp’s advantages for wavelength
ibration. Intensive work has been done on this stan-
adjustment.
dard by Travis et al. [10]. They report on tests that
Typically, the wavelength adjustment with an HgAr were performed on 17 instruments in 14 countries with
lamp requires the following successive steps: i) Selection spectral bandwidths of 0.1 nm, 1.0 nm and 3.0 nm and a
of reference lines and definition of their positions. ii) wavelength sampling interval of 0.1 nm. Peak positions
Recording the lamp spectrum, usually involving different were determined to the nearest 0.01 nm by use of peak-
illumination times for the differently-intense spectral re- location algorithms. All instruments were adjusted with
gions. iii) Spectral smoothing and interpolation. iv) Peak respect to atomic spectral lines prior to the measure-
finding of the chosen reference peaks, thereby connecting ments. Figure 1 shows the spectrum of an holmium oxide
the wavelength λ with an instrument parameter x, such solution in the top panel.
as pixel number (in case of array instruments) or a In the following, the paper of Travis et al. [10] is
generalized position of a dispersive element (for scan- summarized:
ning instruments). v) Fitting of a parametrized formula a) Temperature sensitivity: The positions of ab-
λ(x), which can be a simple polynomial of the form sorption maxima of holmium oxide solutions have shown
λ(x) = a0 + a1 x + a2 x2 + a3 x3 , in order to obtain this to be very temperature insensitive. Peak shifts < 0.05 nm
relationship not only on the positions of the reference have been reported in the range between 25◦ C − 45◦ C.
lines but over the whole spectrum. The determination of Nevertheless it is recommended to use holmium oxide
parameters with physical meaning have been suggested solution between 20◦ C and 30◦ C.
instead of polynomial coefficients [8]. b) Concentration sensitivity: Changes in the con-
Such HgAr-lamp based adjustments can achieve a centration between 10 − 60 g/l result in peak shifts of
wavelength accuracy of 0.1 nm over the whole spectrum less than 0.05 nm.
on instruments with 0.2 nm data intervals, using sophis- Proper concentration adherence is in the responsibility
ticated calibration algorithms (but are limited by the in- of the certified reference material supplier. It is recom-
strument’s actual technical capabilities, see section IV-C) mended that concentration variations shall be within 20%
[8], [9]. compared to the nominal value of 40 g/l.
4 METTLER TOLEDO GmbH, AUGUST 2017
tempering process over a wide temperature range in pro- the TE-260 with 1 nm or 2 nm and the A-8453 with
duction to compensate the internal stress due to mount- 1.5 nm, the holmium oxide solution reference values
ing. After tempering, tests on a MT-UV7 instrument given in Table V are used for calibration. Using the
show no measurable temperature dependence of the 14 correct reference values, taking into account even sub-
peak positions of an holmium oxide solution in the range tle physical effects, minimizes systematic, instrument-
of 10◦ C − 40◦ C [25]. These results confirm the tempera- independent deviations.
ture insensitivity of the holmium oxide solution standard The described resolution dependency of peak positions
and the stability of the array technology. However, small occurs generally with asymmetric lines, and care has to
shifts due to large thermal differences were observed in be taken when comparing application results obtained on
measurements taken with instruments immediately after instruments with different SBWs.
significant temperature changes (e.g., unloading from
cold/hot car and transfer into the laboratory), without 0.8
SBW = 0.1 - 3 nm (1 nm center)
giving the equipment the time to equilibrate. SBW = 1.5 nm
0.6 SBW = 2 nm
For best practices, we recommend using the internal
[nm]
start-up calibration for scanning systems and to always
0.4
peak, SBW = 1 nm
ensure temperature equilibrium with the environment
before wavelength calibration. 0.2
B. Resolution dependency 0
The wavelength positions and photometric values of
absorption peaks are important for many different appli- -0.2
TABLE V
H OLMIUM OXIDE SOLUTION ABSORPTION PEAKS , MEASURED AT DIFFERENT SPECTRAL BANDWIDTHS (SBW S ) AND USED AS
REFERENCES WAVELENGTHS FOR CALIBRATION , e.g. 1.0 NM FOR THE MT-UV7 & TE-260, 1.5 NM FOR MT-UV5NANO & A-8453 AND
2 NM FOR THE MT-UV5/UV5B IO SPECTROPHOTOMETERS . DATA TAKEN FROM R EF. [10], [11].
SBW
1.0 nm 241.12 249.89 278.13 287.22 333.48 345.38 361.25 385.61 416.25 451.45 467.82 485.23 536.56 640.50
1.5 nm 241.13 249.98 278.11 287.40 333.49 345.38 361.18 385.70 416.42 451.33 467.86 485.26 536.71 640.62
2.0 nm 241.12 250.03 278.10 287.52 333.47 345.42 361.12 385.80 416.57 451.32 467.90 485.25 536.86 640.79
0.2
Reference 333.48 nm
b) Wavelength repeatability: Whereas for the abso-
1 nm, Agilent A-8453
0.2 nm, Mettler Toledo UV7 lute wavelength accuracy a possible error in the order of
0.1 nm, ThermoFisher TE-260
0.15
λI /2 can never be ruled out, the situation is different for
the corresponding repeatability measurement. Repeata-
Absorbance
R EFERENCES [22] Starna Scientific Ltd., Didymium glass filter for use as a wave-
length accuracy reference, Sample Certificate Number: 42476,
Set Serial No: 15356, 2013.
[1] JCGM 200:2008, International vocabulary of metrology Basic
[23] Hellma Analytics, Calibration certificate, Liquid calibration
and general concepts and associated terms (VIM), JCGM, 2008.
standard, Type UV45, Serial number: 0015, 2017.
[2] European Pharmacopoeia, 2.2.25 Absorption Spectrophotometry, [24] Steve Upstone, Validating UV/VISible Spectrophotometers,
Ultraviolet and Visible, Rev. 8.0. Technical Note, PerkinElmer, Inc., Seer Green, UK.
[3] British Pharmacopoeia, Appendix II B. Ultraviolet and Visible [25] K.-D. Anders, Mettler-Toledo UV/VIS Spectrometer Ambient
Absorption Spectrophotometry, (Updated Ph. Eur. 9.0, effective Temperature Dependency Evaluation, Technical Note, asens AG,
01/01/2017, Ph.Eur.mmethod 2.2.25), 2017. 2015.
[4] U.S. Pharmacopeia National Formulary USP 40 NF 35, < 857 > [26] Agilent Technologies, Inc., Wavelength accuracy and repro-
Ultraviolet-Visible Spectroscopy, Official from August 1, 2017, ducibility of the Agilent UV-Vis diode array using holmium oxide,
The United States Pharmacopeial Convention, 2017. Technical Note, Publication number: 5991-3446EN, 2013.
[5] JP XVII, The Japanese Pharmacopeia, 17’th edition, English [27] Agilent Technologies, Inc., Agilent 8453 UV-visible Spec-
Version, Official from April 1, 2016, The Ministry of Health, troscopy System, Service Manual, Edition 02/00.
Labour and Welfare of Japan, 2016.
[6] E925-09, Standard Practice for Monitoring the Calibration of
Ultraviolet-Visible Spectrophotometers whose Spectral Bandwith
does not Exceed 2 nm, ASTM International, reapproved 2014.
[7] C. J. Sansonetti, M. L. Salit, and J. Reader, Wavelengths of
spectral lines in mercury pencil lamps, APPLIED OPTICS, 35,
74-77, 1996.
[8] K. Liu and F. Yu, Accurate wavelength calibration method using
system parameters for grating spectrometers, Optical Engineer-
ing, 51, 013603, 2013.
[9] Y. C. Sun, C. Huang, G. Xia, S. Q. Jin, and H. B. Lu, Accurate
wavelength calibration method for compact CCD spectrometer,
Journal of the Optical Society of America A, 34, 498-505, 2017.
[10] J. C. Travis, J. C. Acosta, G. Andor, J. Bastie, P. Blatner, C.
Chunnilall, S. C. Crosson, D. L. Duewer, E. A. Early, F. Hengst-
berger, C. S. Kim, L. Liedquist, F. Manoocheri, F. Mercader, A.
Mito, L. A. G. Monard, S. Nevas, M. Nilsson, M. Noel, A. C.
Rodriquez, A. Ruiz, A. Schirmacher, M. V. Smith, G. Valencia,
N. van Tonder, and J. Zwinkels, Intrinsic Wavelength Standard
Absorption Bands in Holmium Oxide Solution for UV/VISible
Molecular Absorption Spectrophotometry, J. Phys. Chem. Ref.
Data Vol. 34, No. 1, 2005.
[11] Starna Scientific Ltd., Certified Reference Materials for UV and
Visible Spectrophotometry, RH-HL, Holmium solution, Certifi-
cate Number: 47944, Set Serial No: 19781, 2014.
[12] Hellma Analytics, Calibration certificate, Liquid calibration
standard, Type 667-UV5, Holmium solution, Serial number:
0597, 2016.
[13] H. J. Keegan, J. C. Schleter, V. R. Weidner, Ultraviolet Wave-
length Standard for Spectrophotometry, J. Opt. Soc. Am. 51,
1470, 1961.
[14] Ch. Burgess, J. Hammond, Exploding the myths: an update on
UV-visible spectrometry, Spectroscopy Europe, Vol. 21 No. 1,
2009.
[15] D. W. Allen, Holmium Oxide Glass Wavelength Standards,
Journal of Research of the National Institute of Standards and
Technology, Vol. 112, No., 2007.
[16] Starna Scientific Ltd., Holmium glass filter for use as a wave-
length accuracy reference, Sample Certificate Number: 42504,
Set Serial No: 18088, 27, 2013.
[17] Starna Scientific Ltd., A User Guide: Why do we need Reference
Materials?, Starna Certified Reference Materials for UV, Visible
and NIR Spectrophotometry Guide.
[18] Starna Scientific Ltd., private communication.
[19] Hellma Analytics, private communication.
[20] Hellma Analytics, Calibration certificate, Liquid calibration
standard, Type 667-UV25, Didymium solution, Serial number:
0019, 2017.
[21] Hellma Analytics, Calibration certificate, Didymium Glass Fil-
ter, Type 666-F7, Serial number: E1875, 2017.