Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/312466067

Photostability testing: Shedding light on a not well understood guideline

Article · January 2013

CITATIONS READS

3 6,986

1 author:

Allen Zielnik
Ametek Inc.
18 PUBLICATIONS   63 CITATIONS   

SEE PROFILE

Some of the authors of this publication are also working on these related projects:

ICH Q1B & Q5C, VICH GL-05 Pharmaceutical Photostability Testing View project

All content following this page was uploaded by Allen Zielnik on 08 September 2017.

The user has requested enhancement of the downloaded file.


CLINICAL RESEARCH

Photostability Testing: Shedding Light


on a Not Well Understood Guideline
The subject of pharmaceutical However, the Q1B document is not but only for the visible region of
photostability testing often raises without its critics. At a high level it the spectrum, and is expressed as
many questions for several reasons. primarily pertains to the photostability of lumens/m², or more commonly, lux.
Predominantly governed by ICH active pharmaceutical ingredients (API) • Spectral irradiance is irradiance
Guideline Q1B for small molecules, or and manufactured final pharmaceutical expressed at a specific wavelength
ICH Q5C for biotechnology products, products (FPP) and does not address or integrated over a range of
the standards are not particularly well the photostability of a product under in- wavelengths, (e.g., 320-400nm),
written and often cause confusion. use conditions. It also does not address usually expressed as W/m²•nm
Q1B, in particular, provides options specific photochemical characteristics where the wavelength(s) is (are)
to use two non-equivalent light or quantitatively determining rates specified.
sources and additional variants. The of photochemical degradation. At a • Spectral Power Distribution (SPD)
standard also mixes radiometric and practical level the vagueness of the refers to a continuous plot of
photometric light measurements language, and the mixing of scientific and the wavelengths over a specified
for the ultraviolet and visible light lay terms - which are either unfamiliar range versus the irradiance at each
dosages. Following one of these to the pharmaceutical laboratory or not point, and is useful for visualising
options, which specifies the spectral used in photochemistry - to describe or comparing the source spectral
power distribution of the light source, the radiation output can lead to output.
will require different exposures in misunderstanding. The implication is • Dose is the measured irradiance or
order to conform to the guideline that the various photolysis source options illuminance over time. For example,
minimum requirements. This article are technically or effectively equivalent, the Q1B 200 W•h/m² UV (320-
will attempt to explain the guideline which is not the case. 400nm) minimum dose would be
options available, and provide some achieved with a spectral irradiance of
practical guidance. 20 W/m²•320-400nm for 10 hours.
Following Newton’s inverse-square
ICH Q1B guideline “Photostabililty law, this holds true for point sources
Testing of New Drug Substances and or for specimens greater than five
Products”1 was introduced in 1997 times distant than the diameter of
and has provided necessary guidance the source.
to pharmaceutical applicants about ICH Q1B Exposure Dose
the regulatory requirements for The guideline sets required confirmatory
photostability testing. It has provided testing minimum doses for the UV and
a valuable framework for harmonising visible light exposures. Note that these
global laboratory practice, and provides Terminology of Photochemistry are dose minimums and not exposure
a sequential approach when identifying To better understand photostability endpoints, so “overexposure” is
protective packaging requirements for testing a brief guide to photochemistry technically not possible. In fact, as the
photosensitive pharmaceutical products. terminology is useful: majority of photolability issues can be
It also contains a forced degradation • The photolysis source provides the resolved with appropriate packaging, a
provision to help validate stability flux (optical power). common practice is to expose to 5-10X
indicating assays. Prior to the introduction the ICH confirmatory minimums and first
of the guideline, photostability protocols º Radiant flux is a radiometric determine if there is a potential stability
varied considerably with regard to the unit of the total rate of emitted problem. If photolability is found, further
types of photolysis sources and their radiation from the source and is studies may be implemented to identify
spectral characteristics, specimen expressed in watts. the amount and rate of photodegradation.
presentation, exposure duration, etc. º Luminous flux is similar, except The developers of the guideline assumed
Exposures often vary by several orders that it is a photometric unit of that light exposure could occur anywhere
of magnitude, making comparative visible radiation only, as viewed by during manufacturing, packaging,
studies difficult. The adoption of Q1B set the spectral sensitivity of the human warehousing and distribution, or in a
minimum requirements for test protocols eye, and expressed in lumens. pharmacy or home setting. Light sources
and duration, and comprises one element • Irradiance is the total rate of energy would most commonly be various types of
of a larger set of environmental stability (radiant flux, all wavelengths) fluorescent lamps, indirect sunlight filtered
testing guidelines, and has been adopted incident upon a given surface area through window glass, or a combination.
by the majority of pharmaceutical and commonly expressed in W/m². The product should be expected to
producing country regulatory bodies. • Illuminance is analogous to irradiance withstand a minimum of:

50 INTERNATIONAL PHARMACEUTICAL INDUSTRY Spring / Summer 2013 Volume 5 Issue 2


CLINICAL RESEARCH

• Three months of continuous As photon energy is inversely variation between sources, however. The
exposure to visible light without proportional to wavelength, the lower source for visible light is “cool white”
protective packaging, as a 90-day UV cut-on and higher low wavelength fluorescent as specified in ISO 10977,
supply could be transferred to a non- spectral irradiance of D65 compared to although window light does contain
protective package. This exposure ID65 (Figure 1) can be significant in terms visible wavelengths not found in cool
was estimated to be at 500 lux (about of photolytic effects. Although either white lamps.
normal office lighting) for 24h x 100 SPD is permitted in Q1B for confirmatory These two fluorescent lamps may be
days = 1.2 x 106 lux•h. studies, the results may be dissimilar. Note used in combination or in sequential
• During this same time, indirect that the key point is that both UV and visible exposures. Depending on the
sunlight filtered through window exposures are conducted simultaneously manufacturer, however, there may be an
glass was estimated at 200 W•h•m² under one full spectrum source. emission gap between the two sources
(~320-400nm), roughly the equivalent
of 1-2 days of windowsill exposure
where the glass attenuates the lower
wavelength UV.

In order to provide these conditions,


while providing testing flexibility, the
guideline provides for two laboratory
testing options using different photolysis
sources.

Option 1 and Option 2 Sources for


Confirmatory Requirements
Option 1 sources may be xenon or metal
halide arc discharge lamps that are
appropriately filtered, or “full spectrum”
fluorescent lamps providing both UV-A/
UV-B and visible energy to provide a Figure 1. Comparison of xenon arc SPD for D65 and ID65 requirements.
D65 (outdoor daylight) or ID65 (behind
glass indoor daylight) SPD. In practice, Option 2 radiation sources attempt between 390 and 430nm. If this is the
however, fluorescent lamps of this type to mimic indoor (fluorescent) lighting case it is important to verify that the API
are difficult to source and the UV content conditions of a windowed room. This or drug substance does not absorb in this
ratio in specific wavelength bands can is provided by two separate fluorescent region to avoid an incorrect test result.
vary. Likewise, metal halide lamps sources. The one for UVA emission is to A dichotomy exists in meeting the
providing the proper spectra are also have a spectral distribution from 320- specified SPD and the exposure
limited. Therefore, optically filtered long 400nm with a maximum energy emission durations required to attain the minimum
arc xenon lamps are predominantly used; of 250-370nm, with a “significant” confirmatory requirements for UV and
they have a high-fidelity spectral match portion of the UV in both the bands visible. For an Option 1 xenon source,
to D65 and ID65 sunlight and are the 320-260nm and 360-400nm. The term the 200 W•h•m² (~320-400nm) dose
“gold standard” for simulating sunlight. “significant” may lead to considerable corresponds to about 0.45 million lux•h
of visible radiation. Therefore exposing
to 1.2 million lux•h would result in
exceeding the UV minimum by a factor
of 2.5 to 3 times by using D65 or ID65
sources respectively.
The “overexposure problem” can be
avoided in one of two ways in Option 1:

• Exposing two sets of specimens


simultaneously, removing the first set
when the UV minimum is met, and
the second when the visible exposure
minimum is met. Of course both
• specimens receive the full spectrum,
but for different dose minimums.
ICH does not require that the UV
and visible exposures be separated;
in fact, using Option 1 they can’t be
separated.

52 INTERNATIONAL PHARMACEUTICAL INDUSTRY Spring / Summer 2013 Volume 5 Issue 2


CLINICAL RESEARCH

• Combine Option 1 and Option 2. Forced degradation, on the other technique as direct sunlight exposure can
Use an Option 1 source for the UV hand, attempts to force the API or FPP occur in the real world.
confirmatory minimum and Option 2 to degrade by any means necessary.
cool white fluorescent for the visible. For photostability, performing testing to Other Considerations for Testing
5-10X the confirmatory minimums often Option 1 sources generally provide more
Another possibility is to use appropriate serves this purpose. Sometimes more specimen heating effects than Option
UV cut-on and cut-off filters between the extreme measures are required. This 2, especially at higher irradiances. With
Option 1 source and the specimens to may include performing the more severe both options, a “dark control” specimen
isolate the UV and visible exposure bands. D65 rather than the low UV-filtered ID65 (e.g., wrapped in aluminum foil) shielded
While this may be possible for small source tests, or using other photolysis from light is recommended during
individual specimens, it is not practical sources. The main purpose is to achieve exposures. The guideline specifies “room
for the larger area of a typical exposure some degradation, then to validate the temperature” testing, but in practice
chamber. However, the use of sharp cut ability of the stability-indicating assays, most chambers provide some specimen
on or bandpass filters may be useful such as HPLC, to detect the change. heating. The confounding effects of
for determining the actinic wavelengths Due to minimum detection levels temperature (and, in some instances,
responsible for photodegradation. or interferences, a sufficient level of humidity) can then be evaluated
degradation products need be created to relative to the light-exposed specimens.
Irradiance Levels, Exposure Time and validate the assay method (5-15% API loss Sometimes the higher near-infrared
Dose is often used). However, care must be wavelengths present in xenon sources
Another aspect of Q1B is that the taken when degradation is forced using may cause “greenhouse effect” heating
irradiance levels are not specified, only wavelengths more damaging, such as of sealed specimen containers, so some
the radiant energy dose. However, high UV-C germicidal lamps, as this may force air exchange should be permitted.
irradiance levels can alter the degradation degradation mechanisms that would not The higher chamber cooling airflow
response of the product if reciprocity occur under normal conditions. requirements of Option 1 sources
is not obeyed. Fluorescent lamps can An important aspect of validating may also be problematic when testing
be operated over a relatively narrow the assay method is the concept of powders. Powders should be maintained
power and irradiance range, although mass balance, where a decrease in at a controlled depth during exposure,
the number of lamps can be increased the API must be accompanied by a covered with appropriate UV transmitting
or the distance to specimen decreased to reasonable accounting for the mass quartz, for example. This will allow
provide higher irradiance. of the degradation products. Loss of uniform sampling of a degraded
Xenon arc lamps can be operated API without detecting degradants is of surface relative to the bulk material;
over a wider power range, but the limited value. If using Option 2 or Option some commercial powder testing
infrared heating effects become 1/ID65 sources for confirmatory testing, apparatus are available for this
disproportionately larger. In general, switching to D65 is usually a reasonable purpose (Figure 2).
Option 1 xenon sources are higher output
and result in relatively short exposure
times, often as short as 3-9h for the
minimum confirmatory UV requirement,
or 7.5-22h for the visible exposure, by
varying the irradiance.

Forced Degradation Requirement


While the minimum confirmatory
requirements for UV and visible
exposure may seem confusing, the
forced degradation requirement is more
straightforward, though less defined.
The Q1B guideline has a sequenced
structure with a flowchart and qualified
exit points. This leads the researcher
through testing requirements first with
the API and then, if needed, to the
API with excipients through to product
with packaging. The guideline also
specifies retest requirements, such as a
change in synthesis or plant. The result
helps determine any protective packaging
requirements to prevent photolability
resulting in toxicity, loss of efficacy or physical
changes such as dissolution or friability. Figure 2. Example of powders and tablets under Option 1 exposure.

54 INTERNATIONAL PHARMACEUTICAL INDUSTRY Spring / Summer 2013 Volume 5 Issue 2


CLINICAL RESEARCH

The spectral transmittance of any


specimen container is an important factor,
especially for the UV exposures. Many
common glass vials and some plastic foils
do not transmit low wavelength UV.
Liquid specimens should usually be tested
with as large an exposed surface area
and short path length as is reasonable,
like powders, to account for the Beer-
Lambert Law. FPPs should be tested in
their final dosage form as well as in final
packaging. Specimen containers should
be positioned so as to avoid shading or
reflective effects, and special care should
be taken regarding container closures
shielding the specimens.

Commercial Photostability Chambers


A number of testing chambers are
commercially available. These vary
Figure 3. A commercial Option 1 photostability chamber with auxiliary cooling system for thermally
in photolysis source, optional optical labile specimens.
filtering, thermal and humidity control,
irradiance control and geometry. small molecules and Q5C3 for biotechnology its Q5C corollary, provide a reasonable
Regardless of design, the user should products such as peptides, proteins, structure for testing to ensure that potential
be aware of spatial variations in spectral monoclonal antibodies, etc., is the lack of a pharmaceutical photostability issues are
irradiation (and temperature) and confirmatory minimum requirement. Many suitably managed.
verify through mapping studies; some of these products are light-sensitive, such as
chambers have considerable variability protein disulfide bond linkages, which can References
near the edges and corners. This may lead to loss of conformation and bioactivity.
be accomplished with a radiometer or In essence, only the forced degradation 1. International Commission on
spectroradiometer with a positionable Harmonization, ICH Guideline Q1B,
requirement, and subsequent stability assay
sensing head. Stability Testing: Photostability Testing
verification, is required.
A calibrated luxmeter may be used of New Drug Substances and Products,
for the visible range, but neither it nor Nov. 1995.
a UV filter radiometer is suitable to 2. Baertschi, S., Alsante, Tønnesen,
obtain an absolute measurement of the H. A critical assessment of the ICH
guideline on photostability testing of
irradiance, or to compare irradiance
new drug substances and products
between sources. Various sources, while
(Q1B): Recommendation for revision,
appearing continuous in output, may J.Pharm.Sci, Vol. 99(7), Jul 1 2010.
be driven by square or sine waves, or 3. International Commission on
rapidly pulsed; unless calibrated for Harmonization, ICH Guideline Q5C
the specific light source, erroneous Stability Testing of Biotechnological/
readings may result. One should always Biological Products
consult the source manufacturer for
proper measurement and calibration
techniques. Many chambers have an Allen Zielnik is
integral measurement or control system a Senior Consul-
to provide the proper irradiance. The tant within Atlas
use of chemical actinometers, such as Material Test-
quinine hydrochloride, is problematic2 Conclusion ing Technology’s
The ICH Q1B guideline is the principal Global Consulting
and should probably be avoided,
document governing photostability testing. It Group. He has
especially with Option 1 sources.
Some photostability exposure apparatus is, however, an imperfect one and can result been with Atlas for 19 years. Prior to
in some confusion, especially regarding the Atlas his career spanned 19 years with
are equipped with active cooling to mitigate
choice of Options 1 or 2 light sources and analytical instrument companies, prin-
temperatures, a consideration when
their equivalence. Further, the measurement cipally in pharmaceutical chromatogra-
testing thermally labile compounds such as
phy and spectroscopy. He is a frequent
biotechnology products (Figure 3). of irradiance and timing of exposures is often
conference speaker and has authored
a source of confusion. Hopefully these, and
over 100 publications.
A Note on ICH Q5C others, will be addressed in future revisions.
Email: al.zielnik@ametek.com.
The primary difference between ICH Q1B for In the meanwhile, the Q1B guideline, and

56 INTERNATIONAL PHARMACEUTICAL INDUSTRY Spring / Summer 2013 Volume 5 Issue 2

View publication stats

You might also like