Professional Documents
Culture Documents
1 Photostability Testing Shedding Light On A Not Well Understood Guideline
1 Photostability Testing Shedding Light On A Not Well Understood Guideline
net/publication/312466067
CITATIONS READS
3 6,986
1 author:
Allen Zielnik
Ametek Inc.
18 PUBLICATIONS 63 CITATIONS
SEE PROFILE
Some of the authors of this publication are also working on these related projects:
ICH Q1B & Q5C, VICH GL-05 Pharmaceutical Photostability Testing View project
All content following this page was uploaded by Allen Zielnik on 08 September 2017.
• Three months of continuous As photon energy is inversely variation between sources, however. The
exposure to visible light without proportional to wavelength, the lower source for visible light is “cool white”
protective packaging, as a 90-day UV cut-on and higher low wavelength fluorescent as specified in ISO 10977,
supply could be transferred to a non- spectral irradiance of D65 compared to although window light does contain
protective package. This exposure ID65 (Figure 1) can be significant in terms visible wavelengths not found in cool
was estimated to be at 500 lux (about of photolytic effects. Although either white lamps.
normal office lighting) for 24h x 100 SPD is permitted in Q1B for confirmatory These two fluorescent lamps may be
days = 1.2 x 106 lux•h. studies, the results may be dissimilar. Note used in combination or in sequential
• During this same time, indirect that the key point is that both UV and visible exposures. Depending on the
sunlight filtered through window exposures are conducted simultaneously manufacturer, however, there may be an
glass was estimated at 200 W•h•m² under one full spectrum source. emission gap between the two sources
(~320-400nm), roughly the equivalent
of 1-2 days of windowsill exposure
where the glass attenuates the lower
wavelength UV.
• Combine Option 1 and Option 2. Forced degradation, on the other technique as direct sunlight exposure can
Use an Option 1 source for the UV hand, attempts to force the API or FPP occur in the real world.
confirmatory minimum and Option 2 to degrade by any means necessary.
cool white fluorescent for the visible. For photostability, performing testing to Other Considerations for Testing
5-10X the confirmatory minimums often Option 1 sources generally provide more
Another possibility is to use appropriate serves this purpose. Sometimes more specimen heating effects than Option
UV cut-on and cut-off filters between the extreme measures are required. This 2, especially at higher irradiances. With
Option 1 source and the specimens to may include performing the more severe both options, a “dark control” specimen
isolate the UV and visible exposure bands. D65 rather than the low UV-filtered ID65 (e.g., wrapped in aluminum foil) shielded
While this may be possible for small source tests, or using other photolysis from light is recommended during
individual specimens, it is not practical sources. The main purpose is to achieve exposures. The guideline specifies “room
for the larger area of a typical exposure some degradation, then to validate the temperature” testing, but in practice
chamber. However, the use of sharp cut ability of the stability-indicating assays, most chambers provide some specimen
on or bandpass filters may be useful such as HPLC, to detect the change. heating. The confounding effects of
for determining the actinic wavelengths Due to minimum detection levels temperature (and, in some instances,
responsible for photodegradation. or interferences, a sufficient level of humidity) can then be evaluated
degradation products need be created to relative to the light-exposed specimens.
Irradiance Levels, Exposure Time and validate the assay method (5-15% API loss Sometimes the higher near-infrared
Dose is often used). However, care must be wavelengths present in xenon sources
Another aspect of Q1B is that the taken when degradation is forced using may cause “greenhouse effect” heating
irradiance levels are not specified, only wavelengths more damaging, such as of sealed specimen containers, so some
the radiant energy dose. However, high UV-C germicidal lamps, as this may force air exchange should be permitted.
irradiance levels can alter the degradation degradation mechanisms that would not The higher chamber cooling airflow
response of the product if reciprocity occur under normal conditions. requirements of Option 1 sources
is not obeyed. Fluorescent lamps can An important aspect of validating may also be problematic when testing
be operated over a relatively narrow the assay method is the concept of powders. Powders should be maintained
power and irradiance range, although mass balance, where a decrease in at a controlled depth during exposure,
the number of lamps can be increased the API must be accompanied by a covered with appropriate UV transmitting
or the distance to specimen decreased to reasonable accounting for the mass quartz, for example. This will allow
provide higher irradiance. of the degradation products. Loss of uniform sampling of a degraded
Xenon arc lamps can be operated API without detecting degradants is of surface relative to the bulk material;
over a wider power range, but the limited value. If using Option 2 or Option some commercial powder testing
infrared heating effects become 1/ID65 sources for confirmatory testing, apparatus are available for this
disproportionately larger. In general, switching to D65 is usually a reasonable purpose (Figure 2).
Option 1 xenon sources are higher output
and result in relatively short exposure
times, often as short as 3-9h for the
minimum confirmatory UV requirement,
or 7.5-22h for the visible exposure, by
varying the irradiance.