Gap Analysis ISO 15223-1 (2016 Vs 2021)

EN ISO 15223-1:2016 EN ISO 15223-1:2021 CHANGES COMMENT


Scope Scope
This document identifies requirements This document identifies requirements for
for symbols used in medical device symbols used in medical device labelling that
labelling that convey information on the convey information on the safe and effective
safe and effective use of medical use of medical devices. It also lists symbols
devices. It also lists symbols that satisfy that satisfy the requirements of this
the requirements of this document. document.
This document is applicable to symbols
used in a broad spectrum of medical This document is applicable to symbols used Same
devices, which are marketed globally in a broad spectrum of medical devices,
and therefore need to meet different which are marketed globally and therefore
regulatory requirements. need to meet different regulatory
These symbols may be used on the requirements.
medical device itself, on its packaging or
in the associated These symbols are marked on the medical
documentation. device itself, placed on its packaging or
The requirements of this placed in the associated accompanying
document are not intended to apply to information. The requirements of this
symbols specified in document are not intended to apply to
other standards. symbols specified in other standards.
Normative References Normative References
ISO 70001), Graphical symbols for use ISO 7000 (database), Graphical symbols for
on equipment — Registered symbols use on equipment — Registered symbols
ISO 7001:2019, Graphical symbols — Public According to new
ISO 8601, Data elements and information symbols references, ……
interchange formats — Information ISO 8601-1:2019, Date and time —
interchange — Representation of Representations for information interchange
dates and times — Part 1: Basic rules ISO 8601-2:2019, Date
and time — Representations
for information interchange — Part 2:
Extensions
Gap Analysis ISO 15223-1 (2016 Vs 2021)
ISO 15223-2, Medical devices — ISO 14971:2019, Medical devices -- Application of
Symbols to be used with medical risk management to medical devices
device labels, labelling, and ISO 13485:2016, Medical devices – Quality
information management systems – Requirements for
to be supplied — Part 2: Symbol regulatory purposes
development, selection and validation ISO 15223-2:2010, Medical devices — Symbols to
be used with medical device labels, labelling, and
information to be supplied — Part 2: Symbol
development, selection and validation
ISO 20417 1, Medical Devices - Information to
be supplied by the manufacturer
IEC 60417 (database), Graphical symbols for use
on equipment
IEC 80416-1:2008, Basic principles for graphical
symbols for use on equipment — Part 1: Creation
of
graphical symbols for registration
Terms and definitions Terms and definitions New terms and
definitions will examine
3.1 to 3.6 3.1 to 3.22 and if we need a
change, they will
use in procedure.

ISO 15223-1:2016 Table 1 — Symbols to


ISO 15223-1:2021 Table 1 — Symbols to convey
convey information essential for
proper use information essential for proper use
5.1.1 5.1.1 Same
5.1.2 5.1.2 Same
5.1.3 5.1.3 Same
5.1.4 5.1.4 Same
5.1.5 5.1.5 Same
5.1.6 5.1.6 Same
Gap Analysis ISO 15223-1 (2016 Vs 2021)
5.1.7 5.1.7 Same
5.1.8 - Importer New Symbol
New Symbol
5.1.9 - Distributor
N/A
5.1.10 – Model Number New Symbol
New Symbol
5.1.11 - Country of manufacturer
5.2.1 5.2.1 Same
5.2.2 5.2.2 Same
5.2.3 5.2.3 Same
5.2.4 5.2.4 Same
5.2.5 5.2.5 Same
5.2.6 5.2.6 Same
5.2.7 5.2.7 Same
5.2.8 5.2.8 Same
5.2.9 5.2.9 Same
5.2.10 - Sterilized using vaporized
New Symbol
hydrogen peroxide
5.2.11 - Single sterile barrier New Symbol
system
N/A
5.2.12 - Double sterile barrier New Symbol
system
5.2.13 - Single sterile barrier system
New Symbol
with protective packaging inside
5.2.14 - Single sterile barrier
New Symbol
system
with protective packaging outside
5.3.1 5.3.1 Same
5.3.2 5.3.2 Same
5.3.3 5.3.3 Same
Gap Analysis ISO 15223-1 (2016 Vs 2021)
5.3.4 5.3.4 Same
5.3.5 5.3.5 Same
5.3.6 5.3.6 Same
5.3.7 5.3.7 Same
5.3.8 5.3.8 Same
5.3.9 5.3.9 Same
5.4.1 5.4.1 Same
5.4.2 5.4.2 Same
5.4.3 5.4.3 Same
5.4.4 5.4.4 Same
5.4.5 5.4.5 Same
5.4.6 – Contains human blood or plasma
New Symbol
derivatives

5.4.7 - Contains a medicinal New Symbol


substance
5.4.8 - Contains biological material of
New Symbol
animal origin
N/A 5.4.9 - Contains biological material of
New Symbol
human origin

5.4.10 - Contains hazardous substances New Symbol

5.4.11 - Contains nano materials New Symbol

5.4.12 - Single patient multiple use New Symbol

N/A Our products are


5.5 IVD-specific 5.5 IVD-specific
medical
device.
5.6.1 5.6.1 Same
Gap Analysis ISO 15223-1 (2016 Vs 2021)
5.6.2 5.6.2 Same
5.6.3 5.6.3 Same
5.6.4 5.6.4 Same
5.6.5 5.6.5 Same
5.6.6 5.6.6 Same
5.7.1 5.7.1 Same

5.7.2 - Patient name New Symbol

5.7.3 - Patient identification New Symbol

5.7.4 - Patient information website New Symbol

5.7.5 - Health care center or doctor New Symbol


N/A
5.7.6 - Date New Symbol

5.7.7 - Medical device New Symbol


5.7.8 - Translation
New Symbol

5.7.9 - Repackaging New Symbol

5.7.10 - Unique Device Identifier New Symbol