Professional Documents
Culture Documents
Olympus CV180 User Manual-CAP4 OPERATION
Olympus CV180 User Manual-CAP4 OPERATION
Chapter 4 Operation
This chapter explains the work flow of endoscopic observation using the video
system center. For information on how to use the functions that are not
explained in this chapter, refer to the reference pages.
1. Connect the endoscope to the video system center and the light source.
Section 4.2, “Connection of an endoscope” on page 46
7. Perform examination.
Chapter 5, “Functions” on page 62
Changes iris mode,
Image enhancement,
Freezes image,
Zooms image,
Recording and printing of the images,
NBI observations, etc.
9. Disconnect the endoscope from the video system center and the light source.
Section 4.9, “Termination of the operation” on page 60
11. Reprocess and store the instrument and ancillary equipment as appropriate after use.
Chapter 7, “Care, Storage and Disposal” on page 155
For details on endoscope and light source, see the respective instruction manuals.
Figure 4.1
• Make sure that the video plug and its electrical contacts are
completely dry before connecting the plug to the video
system center (see Figure 4.2). Wet equipment could cause
the image to flicker or disappear.
“UP” mark
Video plug
Electrical contacts
Figure 4.2
1. Ensure that this instrument and all connected devices are turned OFF.
3. Push the video plug into the video connector socket of the instrument all the
way until it clicks, holding this instrument with a hand so that it will not move.
Confirm that the “UP” mark points upwards (see Figure 4.3).
“UP” mark
Video connector socket
Video plug
Figure 4.3
1. Ensure that this instrument and all connected devices are turned OFF.
3. Push the video plug of the scope cable EXERA II into the video connector
socket of the instrument all the way until it clicks, holding this instrument
with a hand so that it will not move. Confirm that the “UP” mark points
upwards (see Figure 4.3).
4. Connect the scope side connector of the scope cable EXERA II to the
endoscope, referring to the instruction manual of the endoscope.
1. Ensure that this instrument and all connected devices are turned OFF.
3. Push the video plug of the camera head into the video connector socket of
the instrument all the way until it clicks, holding this instrument with a hand
so that it will not move. Confirm that the “UP” mark points upwards (see
Figure 4.3).
4. Connect the video adapter and camera head to the eyepiece section of the
fiberscope, referring to the instruction manuals for the video adapter and
camera head.
1. Ensure that this instrument and all connected devices are turned OFF.
2. Connect the light guide cable to the light source, referring to the instruction
manual for the light source.
3. Push the video plug of the camera head into the video connector socket of
the instrument all the way until it clicks, holding this instrument with a hand
so that it will not move. Confirm that the “UP” mark points upwards (see
Figure 4.3).
4. Attach the light guide cable, video adapter and camera head to the
rigidscope, referring to the instruction manuals for the light guide cable,
video adapter and camera head.
Summary of connection
This diagram gives an overview of all connections. For further details of the
connection, refer to the instruction manuals of the ancillary equipment being
used.
Light source
CV-180
1
2
VISERA series
2 3
Video adapter 1
3
Camera head
Fiberscopes
3 1
2
Light guide cable
4
Video adapters, camera heads, etc.
Rigidscope
The numbers show the order of connection.
Figure 4.4
2. Turn the instrument ON by pressing the power switch (see Figure 4.5). The
indicator above the power switch lights up.
Power indicator
Power switch
Figure 4.5
ID:
Name:
Sex: Age:
D.O.B.
12/12/2005
12:12:12
Physician:
Comment:
Figure 4.6
1. Press the “Shift” and “F2” keys together. The “User preset” menu appears
on the monitor (see Figure 4.7).
F2
Shift
Figure 4.7
Figure 4.8
2. Click the desired user name in the user name dialog box. The selected user
name is highlighted.
3. Check if the correct user name is selected, then click “Select”. The
endoscopic image appears on the monitor and the user preset data is
loaded to the video system center.
The user preset data used in the last operation before turning
the video system center OFF comes up initially after turning
the instrument ON.
• Before observation.
• After exchange of the light source.
• When any abnormality can be seen on the color of the image even if
white balance adjustment has been completed.
1. Confirm the lighting status of the “Wh/B OK” (Wh/B = White balance)
indicator on the front panel (see Figure 4.9 and Table 4.1).
Wh/B OK indicator
Wh/B button
Figure 4.9
Table 4.1
1. Insert the endoscope's distal end into the white cap and hold the white cap
and endoscope stable to avoid wash-out of the monitor image (see Figure
4.10).
White cap
Figure 4.10
2. Maintaining the stable condition in step 1., press the “Wh/B” button until a
short beep is generated. The result of the adjustment will be displayed on
the monitor for a few seconds (see Table 4.2).
Result Message
Completed successfully “White balance complete!”
“Check NBI white balance.”
(When using NBI compatible endoscope.)
Failed “White balance incomplete! Perform again!”
Table 4.2
3. Confirm that the “Wh/B OK” indicator is ON (see Table 4.1). If it is OFF, go
back to step 1. again.
1. Hold the endoscope stable to avoid wash-out of the monitor image, and
enlarge the image to the full monitor, monitoring a white object such as a
piece of gauze in such a way that it does not contact the endoscope.
2. Maintaining the stable condition in step 1., press the “Wh/B” button until a
short beep is generated. The result of white balance adjustment will be
displayed on the monitor for a few seconds (see Table 4.3).
Result Message
Completed successfully “White balance complete!”
“Check NBI white balance.”
(When using NBI compatible endoscope.)
Failed “White balance incomplete! Perform again!”
Table 4.3
3. Confirm that the “Wh/B OK” indicator is ON (see Table 4.1). If it is OFF, go
back to step 1. again.
1. Adjust the white balance for normal light observation according to “For
normal light observation” on page 53.
2. Switch to the NBI mode by operating the mode button of the light source.
Refer to the instruction manual of the light source.
3. Confirm that “NBI” appears on the monitor, and that the “NBI” indicator on
the front panel of the video system center turns from green to white.
5. Maintaining the stable condition in step 4., press the “Wh/B” button until a
short beep is generated. The result of the adjustment will be displayed on
the monitor for a few seconds (see Table 4.4).
Result Message
Completed successfully “NBI White Balance Complete!”
Failed “NBI White Balance Incomplete!”
Table 4.4
7. Switch to the normal mode by operating the mode button of the light source.
Refer to the instruction manual of the light source.
8. Confirm that “NBI” disappears on the monitor, and that the “NBI” indicator on
the front panel of the video system center turns from green to white.
This section explains how to enter patient data immediately before the
examination. To enter a list of several patient data in advance, see Section 5.6,
“Pre-entry of patient data” on page 136.
1. Press the “F1” key to change the monitor display full-patient-data display.
2. Press the “EXAM END” key to clear the previous patient data.
ID:
Name:
Sex: Age:
Cursor D.O.B.
12/12/2005
12:12:12
Physician:
Comment:
Figure 4.11
3. Enter data.
Table 4.5
ABC123
Mike Johnson
M 51
03/03/1954
12/12/2005
12:12:12
John Smith
Cardiac end of the stomach
An example of displaying
Figure 4.12
4. When modifying data, press the arrow key to move the cursor to the input
position and edit the data.
5. When deleting all patient data displayed on the monitor, press the “EXAM
END” key on the keyboard (see Figure 4.13).
EXAM END
Figure 4.13
Table 4.6
1. Press the “EXAM END” key (see Figure 4.14) to execute the following
process;
• clearing patient data on the monitor;
• completing printing of un-printed images;
• closing the digital file system.
EXAM END
Figure 4.14
Video plug
Videoscope cable
A) When the scope side connector B) When both connectors are
is not connected. not connected.
Figure 4.15
Locking lever
Video plug
Figure 4.16