General Model RC400G

Specifications Residual Chlorine Analyzer


GS 12F4A1-E

n Overview
Chlorine treatment in water treatment systems, such
as municipal water supplies, serves to destroy micro-
organisms, fungi, aquatic plants and shellfish that live
in water, and thus to yield appropriate quality for the
application. However, adding too much chlorine cause
negative effects; in municipal water supplies, it may
give the water an unpleasant chlorine odor; in indus-
try, excessive chlorine may corrode plumbing; and in
waste water this may release dangerous substances
into streams and rivers. For this reason, it has be-
come very important in recent years to use a residual
chlorine analyzer when chlorine is added to water, in
order to measure the concentration of chlorine, and to
monitor and control this concentration.
The RC400G residual chlorine analyzer employs a po-
larographic system using rotating gold alloy electrodes
to provide continuous online measurement of residual
chlorine concentration. It is said that total chlorine
(residual chlorine) is the sum of free chlorine and • Uses reliable sliding contacts in contact elements.
combined chlorine. By appropriate use of reagents, • Easily maintained (easy-to-clean) cell construction.
RC400G can have three kinds of measurement possi- • Available for special version (for free chlorine meas-
bilities, namely free chlorine measurement, total chlo- urement) designed to cope with very high concentra-
rine (residual chlorine) measurement, and free chlorine tions of combined chlorine (see other literature).
measurent in case of existence of combined chlorine. Sampling equipment (conditioning equipment)
If the raw water is particularly dirty, the concentration • Adds a 2-cylinder sand filter for particularly dirty
of combined chlorine may become extremely high applications to the product line.
immediately after chlorine is added. To compensate
for this, we also offer a free chlorine measurement
system featuring improved separation performance n Standard Specifications
over the standard system (Combined chlorine insensi- 1. Residual chlorine analyzer
tive version). Measurement object: Free chlorine or residual chlorine
The RC400G is an intelligent residual chlorine ana- (total chlorine) in water
lyzer making extensive use of microprocessors incor- Measurement method: Rotating electrode type
porating know-how based on Yokogawa’s many years polarographic system
of experience in residual chlorine analysis. Measurement range: 0 to 10 mg/l (mg/l used as
concentration units per JIS K0101 and
municipal water test methods)
n Features Output range: Can be freely set within ranges with
Intelligent microprocessor-based converter spans of 1 mg/l or greater. (Range set when
• Simple, one-touch calibration. shipment. No.1: 0 to 5 mg/l No.2: 0 to 10 mg/l.
• Automatic sensor self-diagnostics (zero point, slope, Allows selection between 2 ranges (selected
response) during calibration. with contact input signal).
• Change of output range for 1mg/l span or greater, and 2-line-segment output available: Any desired
remote range switching are available. point between 0% and 100% of measurement
• Line-segment function available for output signal. span can be set as the 50% point of the output
• Simple verification of applied voltage/current charac- range.
teristics (Plateau characteristic) between electrodes. Output signal: 4 to 20 mA (load resistance, 550 Ω
Allows detailed knowledge of sensor characteristics Max.) or 1 to 5 V DC (output resistance, 300 Ω
for the target application to be obtained. Max.), isolated
• A wealth of display functions and contact output functions. Display range: -1.00 to 12.00 mg/l
• Easy-to-use operating panel. Display method: Digital display (4-digit LED)
Upgraded detector based on extensive experience Output contacts: Voltage-free (“dry”) contacts
• Improvements to rotating electrodes and glass bead Contact rating: Max. 250 V AC, Max. 2 A,
cleaning have been added for even more effective Max. 125 VA (resistive load)
automatic continuous cleaning. Max. 220 V DC, Max. 2 A, Max. 60 W
• Waterjet cleaning can be applied for the purpose of (resistive load)
particularly dirty applications.

Yokogawa Electric Corporation GS 12F4A1-E


2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan ©Copyright Jun. 1994
Tel.: 81-422-52-5617 Fax.: 81-422-52-6792 1st Edition Jun. 1994
6th Edition Apr. 2013
2
Input contacts: Voltage-free (“dry”) contact Converter functions:
ON resistance: 200 Ω Max. Display functions:
OFF resistance: 100 kΩ Min. Data: Concentration, temperature, applied volt-
Open circuit voltage: 10 V age, diffusion current, output signal %,
Closed circuit current: 100 mA pump flow, zero point, slope
Automatic zero calibration: Zero calibration using acti Status indicators: Measurement, maintenance, hold,
vated charcoal filter (optional) calibration in progress, cleaning, fail
Contact inputs: Operating status : Rotating electrode, metering
Remote range switching: Switching between 2 preset pump, air pump, solenoid valve ON/OFF
ranges (range 1 and range 2) Diagnostic functions: Concentration over range,
Contact open: Range 1 measurement temperature error, no liquid in
Contact closed: Range 2 flow cell, converter error, temperature
Contact outputs: compensation over range, setting upper limit
FAIL contact: Activate when concentration over exceeded, zero point error, slope error,
range, measurement temperature response error
error, no liquid in flow cell converter Maintenance functions (MAINT mode):
error, temperature compensation over One-touch calibration, flow cell and sand filter
range, setting upper limit exceeded system cleaning, pump flow measurement,
Range switching contact one-touch plateau characteristic acquisition,
MAINT contact: Activate when analyzer is in failure error information
maintenance mode Setup functions:
Main power PROGRAM 1 mode
Main power ON Output Range 1 and Range 2 settings, 2-line-
OFF
--- Not operating Operating segment output setting, applied voltage
FAIL contact Open Closed Open
setting, response time and stability
parameter settings, contact output during FAIL
MAINT contact Closed Open Closed
condition (ON/OFF), indication hold during
Sample conditions: MAINT (ON/OFF), remote range switching
Temperature: 0 to 50 °C ON/OFF
pH: 3 to 9 pH PROGRAM 2 mode
Flow: 1 to 4 l/min, for pure water or Cleaning, auto zero calibration sequence setup
municipal water application. Automatic cleaning functions: Electrode and flow cell
5 to 10 l/min, for raw water or sea cleaning methods
water application. For pure water: Glass-bead cleaning
10 to 20 l/min, for secondary sewage For raw/sea water, secondary sewage treatment:
treatment application. Glass-bead cleaning+water jet cleaning;
Pressure: 20 to 500 kPa Water jet cleaning flow: 2 to 3 l/min
Electrodes:
Indicator electrode: Rotating gold alloy electrode Sand filter system:
Counter electrode: Platinum electrode Filter sand: Filter sand for municipal water (high-
(built-in Pt1000 Ω RTD), the com- speed filtering sand, approx. 550 ml per tube)
bined chlorine insensitive version Filtering flow: Approx. 500 ml/min (per cylinder)
uses a silver chloride electrode (built- Backwash water flow: 8 to 9 l/min
in Pt1000 Ω RTD). (per cylinder)
Reagents: Standard cleaning sequence (factory settings)
Composition: The following reagents are dissolved 1-cylinder 2-cylinder
Setting range
in pure water to give 100 l of solution. system system
(Note) Pure water should be equal to or exceed level of Sand filter back- 0.1 to 24 h, 0.1 h steps
2 hours 30 min
tap water and should not contain oxidizing and wash interval
reducing substances such as chlorine. Sand filter back- 0.1 to 25 min, 0.1 min
1 min 1 min
For residual For free chlo- For combined wash time steps
chlorine rine measure- chlorine Flow cell jet 0.1 to 24 h, 0.1 h steps
measurement ment measurement 2 hours 1 hour
cleaning interval
Maximum Up to Over 6 up Up to Up to Flow cell jet 0.1 to 25 min, 0.1 min
concentration 6 mg/l to 10 mg/l 10 mg/l 10 mg/l 1 min 1 min
cleaning time steps
Potassium iodide, 0.1 to 25 min, 0.1 min
500 g 1000 g --- --- Waiting time 5 min 3 min
extra pure steps
Potassium bro- Jet cleaning time +
--- --- 4000 g 4000 g Output hold time 6 min 4 min
mide, extra pure waiting time
Anhydrous sodium In a 1-cylinder system, sand filter backwash and flow
acetate, extra pure 150 g 1000 g 5400 g cell jet cleaning are performed at the same time.
Wetted part materials
Glacial acetic acid, Measurement tank: Acrylic resin
1000 ml 1000 ml 200 ml
extra pure
Pump: Fluorinated rubber (Viton), hardened PVC,
Consumption: 1.5 ml/min ±10% SUS316
Constant flow pump: Tubing/piping: Polyethylene, hardened PVC
Sample flow: 50 ml/min ±10%
Stand materials: Carbon steel or stainless steel
Reagent flow: 1.5 ml/min ±10%

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition May 27, 2014-03
3
Paint colors Tank capacity: 100 L
Mounting stand: Munsell 0.6GY3.1/2.0 Tank materials: Rigid PVC
Other parts: Munsell 0.6GY3.1/2.0 and 2.5Y8.4/1.2 Cart materials:
Finish: Baked polyurethane resin coating Frame; Steel pipe (SPG30A)
Bracket; Steel plate (SPCC)
Operating conditions Pump (Seal-less Pump):
Ambient temperature: -5 to 55 °C Discharge flow rate; 14 to 35 L/min
(However, measure to prevent freezing are Wetted part material; Polypropylene resin, Hastel-
required if the water sample or reagent loy, ceramics, fluorinated rubber
appears to freeze.) Power supply; 100 V AC, 50/60 Hz
Ambient humidity: 5 to 95% RH (non-condensing) Power consumption; Approx. 100 VA
Storage temperature: -30 to 70 °C Power cord; PVC sheathed cable, 5 m
Installation: Indoors (A separate rainproof cover is Weight; Approx. 40 kg
required for outdoor installation. Avoid direct sunlight.)
Utilities 3. Reagent sets (for one year)
Power supply: 100/110/220 V AC ±10%, 50/60 Hz (1) For free chlorine
Power consumption: Potassium bromide extra pure:
RC400G-1¨; Approx. 65 VA 64 containers, 500 g each
RC400G-2¨; Approx. 125 VA Anhydrous sodium acetate extra pure:
RC400G-3¨; Approx. 210 VA 16 containers, 500 g each
Cleaning water (required with sand filter system): Acetic acid extra pure: 7 containers, 500 ml each
Quality; clean water (2) For residual chlorine (Total chlorine)
Pressure; 100 to 500 kPa Potassium iodide extra pure:
Flow; 10 to 12 l/min 8 containers, 500 g each
Consumption; Anhydrous sodium acetate extra pure:
Approx. 130 l/day (1-cylinder sand filter type) 3 containers, 500 g each
Approx. 470 l/day (2-cylinder sand filter type) Acetic acid extra pure:
(With standard cleaning sequence [fac- 17 containers, 500 ml each
tory settings])
(3) For high concentration of residual chlorine
Air purge (using instrument air):
(Total chlorine)
Supply pressure; 140 kPa
Potassium iodide extra pure:
Air consumption; Approx. 5 l/min
16 containers, 500 g each
Weight Anhydrous sodium acetate extra pure:
For pure water: Approx. 65 kg 3 containers, 500 g each
For raw water (1-cylinder): Approx. 70 kg Acetic acid extra pure:
For raw water (2-cylinder): Approx. 75 kg 17 containers, 500 ml each
Optional specifications (4) For combined chlorine insensitive version (Free chlorine)
Pump for air purge (if instrument air cannot be used) Potassium bromide extra pure:
Install to enable intake of clean air. 64 containers, 500 g each
Intake/discharge: 5 l/min, at 50 Hz (Intake/discharge Anhydrous sodium acetate extra pure:
outlets: Open) 87 containers, 500 g each
Maximum pressure: 80 kPa Acetic acid extra pure:
Power consumption: Approx. 23 VA 4 containers, 500 ml each
Weight: Approx. 2 kg (Note) These reagents sets are not imported from Japan be-
Regulatory Compliance: cause of both safety and transportation cost issues.
EMC Regulatory Arrangement in Australia and New Purchase them directly in your local reagent supplier.
Zealand (RCM) EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard Class A n Characteristics
한국 전자파적합성 기준 (Percent display is computed with respect to whichever
of output range 1 or output range 2 has the highest up-
2. Reagent Tank Model RC401G per range value)
(1) Reagent tank (RC401G-A) Repeatability: 2%
Capacity: 100 L Linearity: ±3%
Materials: Rigid PVC Drift: Zero drift; ±1%/month Max.
Weight: Approx. 15 kg Span drift; -5%/month Max.
(2) Reagent tank with stand (RC401G-B) Response time: Displayed as 90% response time
Incorporates the reagent tank in item (1) with a stand For pure water; Approx. 3 min
and a needle valve. For raw water; Approx. 4 min
Stand materials: carbon steel (Within 3 min from standard liquid inlet)
Stand color: Munsell 2.5Y8.4/1.2 Temperature compensation error (water tempera-
Stand finish: Baked polyurethane resin coating ture): ±1% Max. (Temperature compensa-
Weight: Approx. 25 kg tion range: 0 to 40 °C)
(3) Reagent mixing tank with cart (RC401G-C) Effect of ambient temperature: ±0.5%/10 °C
A reagent mixing tank mounted on a cart. Effect of power variation: ±0.5%/10% of rated voltage
With manual mixer, and pump for transferring reagents. Effect of combined chlorine: 5% or less, of concentra-
Note: This tank is for reagent mixing, and is not a substi- tion of combined chlorine for combined
tute for the reagent tank. chlorine insensitive version

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Feb.28, 2014-02
4
n Model and Suffix Codes
(1) Residual chlorine analyzer [Style: S3]

Model Suffix code Option code Description


RC400G --------------------------------- -------------------- Residual chlorine analyzer
Application -1 -------------------- For pure water, for municipal water (with glass-bead cleaning)
(cleaning unit) (*1)
For raw water (1-cylinder) (with glass-beads cleaning, jet cleaning, and one
-2 --------------------
cylinder sand filter unit) *6
For raw water (2-cylinder) (with glass-beads cleaning, jet cleaning, and two
-3 --------------------
cylinder sand filter unit) *6
Output signal 5 -------------------- 1 to 5 V DC
6 -------------------- 4 to 20 mA DC
Power supply 1 -------------------- 200 V AC, 50 Hz
2 -------------------- 200 V AC, 60 Hz
3 -------------------- 220 V AC, 50 Hz
4 -------------------- 220 V AC, 60 Hz
5 -------------------- 100 V AC, 50 Hz
6 -------------------- 100 V AC, 60 Hz
7 -------------------- 110 V AC, 50 Hz
8 -------------------- 110 V AC, 60 Hz
Measurement object (*2) F -------------------- Free chlorine
T -------------------- Residual chlorine (total chlorine)
C -------------------- Combined chlorine insensitive version (free chlorine) (*3)
- -N -------------------- Always -N
- *A -------------------- Always A
Optional specifications /AP1 With air purge pump (200 V AC, 50/60 Hz) *7
/AP3 With air purge pump (220 V AC, 50/60 Hz) *7
/AP5 With air purge pump (100 V AC, 50/60 Hz) *7
/AP7 With air purge pump (110 V AC, 50/60 Hz) *7
/PPM Units: ppm
/NR Without reagent set for start-up (*4)
/SCT Stainless tag plate
/S Stainless stand (*5)
/AZC With automatic zero calibration
/ARS With arrestor (power and signal lines)

*1: “Application” indicates general guidelines. Select the cleaning equipment appropriate for the contamination in the sample.
*2: Measurement object selection is indicated in the following.
*3: Less affected by combined chlorine in free chlorine measurement.
*4: When ordering the RC400G, select /NR and get reagent sets for start-up from local reagent supplier.
These reagent sets can not be exported from Japan due to both safety and transportation issues.
*5: Stanchion, base and bracket are stainless steel.
*6: Please contact Yokogawa regarding the adequate number of the cylinder.
*7: Installation of air purge is always required for all aplication. If instrument air cannot be used, select the air purge pump.

Measurement object
Application : Can be measured
Free chlorine Total chlorine x: Cannot be measured
Water purification: raw water  (*1) 
*1: If ammoniacal contaminants are present in large quantity,
Water purification: mixed water, sedi- high concentrations of combined chlorine may remain
 (*1) 
mentation water if sampling is performed soon after chlorine injection in
Water purification: mains water   prechlorination treatment, or in intermediate treatment
without prechlorination treatment. In this case, combined
Sea water x x (*3) chlorine type is recommended
Factory wastewater, treated effluent x x *2: Oxidizing or reducing agents other than chlorine may be
present. If so, chlorine concentration measurement may
Factory cooling water (industrial water) x  (*2) not be possible.
Factory drinking water   *3: There are Residual chlorine analyzer for sewage
secondary treatment or sea water. Please ask
Sewage secondary treatment x x (*3) Yokogawa.

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition May 27, 2014-03
5
Accessories
1. Residual chlorine meter (RC400G) 2. Reagent tank (RC401G-A, RC401G-B)
Name Quantity Remark
Name Q’ty Remark
Polishing powder 1 bottle For polishing electrode
Couplings 2 sets
Lubricating oil 1 For metering pump drive
Outside diam-
Fuses 4 each 1 A and 3 A (for spare) Polyethylene tubing 5m eter 6 mm, Inside
Glass beads 1 (including for spare) (2 bags) diameter 4 mm.

Valve sheet 4 For metering pump (for spare) Note: These are not included in reagent mix-
ing tank (RC401G-C.)
1 For sample pump (for spare)
Bellophram
1 For reagent pump (for spare)
Special tool 1 For pump valve replacement (*) These reagents sets are not imported
Tool 1 For Bellophram replacement from Japan because of both safety and
transportation cost issues.
Allen wrenches 1 set 1.5, 2, 2.5, 4, 5, 6 (mm) Purchase them directly in your local
reagent supplier.
Reagent set (*) For startup
(2) Reagent tank
Model Suffix code Option code Description
RC401G ----------------------- -------------------- Reagent tank
Type -A -------------------- Reagent tank only
-B -------------------- Reagent tank with stand
-C -------------------- Reagent mixing tank with mobile stand
- -HM -------------------- With manual mixer
- *A -------------------- Style A

(3) Spare parts


Replacement intervals vary depending on the application.
Recommended
Name Part No. Description Q'ty
replacement interval
Indicator electrode
K9334JP Gold alloy electrode (*2) 1 Yearly
(Rotating electrode)
Counter electrode K9332MJ Reference electrode 1 (*1)
Counter electrode
K9332MK Silver chloride electrode 1 (*1)
(for combined chlorine)
Glass beads (2 bags/Q’ty) K9332ZJ For cleaning indicator electrode 1 Yearly
Polishing powder (Alumina) K9088PE For polishing indicator electrode 1 -
Brush K9332JX Part for electrode mechanism 1 2 years
Slip ring K9332JZ Part for electrode mechanism 1 2 years
Drive belt L9804UK Part for electrode mechanism 1 3 years
Driven shaft assembly K9334JV Part for electrode mechanism (*2) 1 3 years
O-ring Y9115XB Part for electrode mechanism 1 3 years
Motor assembly (100 V) K9334JY Part for electrode mechanism 1 3 years
Motor assembly (110 V) K9334VQ Part for electrode mechanism 1 3 years
Motor assembly (200 V) K9334VR Part for electrode mechanism 1 3 years
Motor assembly (220 V) K9334VS Part for electrode mechanism 1 3 years
Gear head (100 V / 110 V) K9332JP Part for electrode mechanism 1 3 years
Gear head (200 V / 220 V) K9334VA Part for electrode mechanism 1 3 years
Fuse (1 A) A1109EF For electric circuit 1 Yearly (*3)
Fuse (3 A) A1113EF For power supply 1 Yearly (*3)
Sand K9720FZ For sand filter unit (1 liter) 1 Yearly
(*1) At the time of damage. (*2) P/N changed by style S3.
(*3) Fuse may be used more than one year, we recommend periodical replace for planning maintenance. Continued next page.

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Jun.05, 2013-00
6

Recommended
Name Part No. Description Q'ty
replacement interval
Air pump (100 V) K9087XA For air purgepump 1 3 years
Air pump (110 V) K9087XF For air purgepump 1 3 years
Air pump (200 V) K9087XG For air purgepump 1 3 years
Air pump (220 V) K9087XH For air purgepump 1 3 years
Lubricating oil (engine oil) K9041RA For metering pump drive 1 1 month (*4)
Bellophram L9819AA For reagent pump 1 6 months
Bellophram L9819AB For sample water pump 1 6 months
Valve sheet K9041HC For metering pump 1 6 months
Activated charcoal filter L9862AY For automatic zero calibration 1 Yearly
Filter K9332NN For head tank or sand filter 1 Yearly
(*4) Recommended maintenance interval for lubricating the metering pump drive assembly.

n External Dimensions
(1) RC400G-1 Residual chlorine analyzer for pure water and municipal water
unit : mm
Standard solution inlet Air purge inlet
Rc1/4(female) Rc1/4(female) Maintenance Maintenance
space space
Reagent inlet
4-ø15 holes Approx. 200 530 Approx. 200
Rc1/4 (female)
for anchor bolt
100

Stand
600
Ball valve (V3)
350

Approx. 500

Strainer (S1) Maintenance


Front
space
100

20 490 20

Cross section A-A

Air pump (*3)


Converter (AP)
(CON)

Detector
(CELL)

Power terminal
Head tank
box (*1) (HD)

Metering
Meas. / Std. solution pump Drain
1450

switching valve VP40 pipe


(PU)
(V2)
Activated Ball valve Pressure reducing
charcoal filter (V11,12) valve (*2) (PR)
(*2) Ball valve
(F11) (*2) (V1) Ball valve (*2)
3-way solenoid (V4)
A A valve (SV6) Measuring
sample inlet Tap water inlet
(*2)
VP16 pipe VP16 pipe
(*2)
120
93
80

530 550 50 85 60 40
150 150

(*1) Option / ARS applies to model with arrestors


(*2) Option / AZC applies to model with auto zero calibration
(*3) Option / AP applies to model with air purge pump

F01E.ai

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Jun.06, 2013-00
7
(2) RC400G-2 Residual chlorine analyzer for raw water (1-cylinder)
unit : mm
Standard solution inlet Air purge inlet
Rc1/4 (female) Rc1/4 (female)

Reagent inlet Maintenance Maintenance


Rc1/4 (female) 4-ø15 holes space space
for anchor bolt
Approx. 200 530 Approx. 200

Ball valve (V6) 100


350 Stand

600
Strainer (S1)

Approx. 500
100

Maintenance
Front
20 490 20 space

Cross section A-A

Air pump (*3)


Converter (AP)
(CON)

Detector Sand filter


(CELL) Solenoid (F1)
valve
(SV1,2)

Ball valve Ball valve


Power terminal (V1)
(V2)
box (*1)

Metering
Meas. / Std. solution pump
1450

switching valve (PU) Drain VP40 pipe


(V5)
Ball valve Pressure reducing
Activated (V11,12) valve (PR)
charcoal filter
(*2) Ball valve
(F11) (*2) Ball valve
(V3)
3-way solenoid (V4)
A A valve (SV6) Measuring
sample inlet Tap water inlet
(*2) VP16 pipe
VP16 pipe
120
93
80

530 550 50

(*1) Option / ARS applies to model with arrestors


(*2) Option /AZC applies to model with auto zero calibration
(*3) Option / AP applies to model with air purge pump
F05E.ai

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
8
(3) RC400G-3 Residual chlorine analyzer for raw water (2-cylinder)

unit : mm
Standard solution inlet Air purge inlet
Rc1/4 (female) Rc1/4 (female)
Maintenance Maintenance
Reagent inlet space space
Rc1/4 (female) 4-ø15 holes
Approx. 200 530 Approx. 200
for anchor bolt

100
Stand

600
Ball valve (V8)
350

Approx. 500
Strainer (S1)
Maintenance
100

Front
space
20 490 20

Cross section A-A

Air pump (*3)


Converter (AP)
(CON)

Detector Sand filter Sand filter


(CELL) (F1) (F2)
Solenoid
valve (SV1) Solenoid
valve (SV3, 4)
Ball valve Ball valve
(V1, 5) Solenoid
(V2)
valve (SV1, 2, 5)
Power terminal
box (*1)
Metering
pump Drain
1450

Meas. / Std. solution VP40 pipe


(PU)
switching valve (V7)
Ball valve Pressure reducing
Activated
(V11,12) valve (PR)
charcoal filter
(F11) (*2) (*2)
Ball valve
3-way solenoid Measuring (V4)
A A valve (SV6) sample inlet
VP16 pipe Tap water inlet
(*2)
VP16 pipe
120
93
80

530 550 50 85 60 40
150 150

(*1) Option /ARS applies to model with arrestors


(*2) Option /AZC applies to model with auto zero calibration
(*3) Option / AP applies to model with air purge pump
F09E.ai

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
9
Reagent tanks
l RC401G-A Reagent tank l RC401G-B Reagent tank

Unit: mm

Reagent tank

415 450

Reagent output
415 (Rc1/4)
Needle
valve
Manual mixer Approx.1600
Reagent input

Level gauge
Approx.
Reagent Approx. 700
output Approx.
(Rc1/4) 955
805
225
20 410 (20) 250 4-ø13 holes
100
450 450 (100) F0204.ai

l RC401G-C Reagent preparation tank

Approx.1280

770 460 F0205.ai

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
10
n Piping Diagrams
(1) RC400G-1 Residual chlorine analyzer for pure water and municipal water

Tubing materials
ø6 x ø4
polyethylene tube Converter

ø22 x ø15 Head tank CON


HD
braided wire reinforced
soft PVC tube
ø33 x ø25
braided wire reinforced
soft PVC tube

VP16 pipe
VP40 pipe CELL Measuring
tank
* option

PU Metering
pump

Ball valve
V2

COM

V1
Measuring sample (VP16)
Standard solution (Rc1/4)
V3 Strainer
Reagent (Rc1/4) S1

Air (Rc1/4) AP

Vat
*Air pump
Drain (VP40)
F02E.ai

RC400G-1 /AZC with auto zero calibration


Tubing materials
ø6 x ø4
polyethylene tube Converter
ø22 x ø15 Head tank CON
HD
braided wire reinforced
soft PVC tube
ø33 x ø25
braided wire reinforced
soft PVC tube

VP16 pipe
V12
VP40 pipe CELL Measuring
tank
* option SV6 NO
3-way solenoid
valve
COM
NC Activated
charcoal
filter PU Metering
F11 pump
Ball
valve
Ball valve V11 V2
V1
Measuring sample (VP16) COM
Pressure reducing valve
V13
Calibration solution (VP16) PR
Standard solution (Rc1/4)
V3 Strainer
Reagent (Rc1/4) S1

Air (Rc1/4) AP

*Air pump Vat

Drain (VP40)
F04E.ai

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
11
(2) RC400G-2 Residual chlorine analyzer for raw water (1-cylinder)
Tubing materials
ø6 x ø4
polyethylene tube
Converter
Sand filter
ø8 x ø6 CON
polyethylene tube F1

ø22 x ø15
braided wire reinforced
soft PVC tube
ø33 x ø25
braided wire reinforced
soft PVC tube
CELL Measuring
V1 tank
VP16 pipe
VP40 pipe
V2
* option

PU Metering
pump
Ball valve
V3 V5
COM
Measuring sample (VP16) SV1 SV2
Pressure reducing valve Solenoid
V4 valve
Tap water (VP16)
Ball valve PR
Standard solution (Rc1/4)
V6 Strainer
Reagent (Rc1/4) S1

Air (Rc1/4) AP

*Air pump
Vat
Drain (VP40)
F06E.ai

RC400G-2 /AZC with auto zero calibration


Tubing materials

ø6 x ø4
polyethylene tube Converter
Sand filter CON
ø8 x ø6 F1
polyethylene tube
ø22 x ø15
braided wire reinforced soft PVC tube

ø33 x ø25
braided wire reinforced soft PVC tube
CELL Measuring
VP16 pipe V1 tank
V12
VP40 pipe
SV6 NO 3-way solenoid V2
* option valve
COM Metering
NC Activated PU pump
charcoal
filter Ball
V3
F11 valve
V5
Measuring sample (VP16) SV1 COM
Solenoid V11
Pressure reducing valve
valve
V4 SV2
Tap water (VP16)
PR
Ball valve
Standard solution (Rc1/4)
V6 Strainer
Reagent (Rc1/4) S1

Air (Rc1/4) AP

*Air pump
Vat
Drain (VP40)
F08E.ai

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
12
(3) RC400G-3 Residual chlorine analyzer for raw water (2-cylinder)

Tubing materials
ø6 x ø4 Converter
polyethylene tube
Sand filter CON
F1 F2
ø8 x ø6 Sand filter
polyethylene tube
ø22 x ø15
braided wire reinforced Solenoid valve
soft PVC tube SV3 SV4
ø33 x ø25
braided wire reinforced CELL Measuring
soft PVC tube V1 V5 tank

VP16 pipe
V2
VP40 pipe
* option PU Metering
pump
V4 V3
Ball valve V7
Measuring sample (VP16) COM SV2
Pressure reducing valve SV1 SV5
V6
Tap water (VP16) PR
Standard solution (Rc1/4)
Strainer
Reagent (Rc1/4) S1
V8
Air (Rc1/4) AP

*Air pump

Vat

Drain (VP40)
F10E.ai

RC400G-3 /AZC with auto zero calibration

Tubing materials
ø6 x ø4 Converter
polyethylene tube
CON
ø8 x ø6 F1 Sand filter F2
polyethylene tube Sand filter
ø22 x ø15
braided wire reinforced Solenoid valve
SV3 SV4
soft PVC tube
ø33 x ø25
braided wire reinforced CELL Measuring
soft PVC tube V1 V5 tank
V12
Metering
VP16 pipe 3-way solenoid pump
valve SV6 COM V2
VP40 pipe
NO
* option NC PU

V4 V3 F11
Ball valve Activated V7
charcoal
Measuring sample (VP16) SV1 V11 filter COM SV2
Pressure reducing valve Ball SV5
V6 valve
Tap water (VP16) PR
Standard solution (Rc1/4)
V8 Strainer
Reagent (Rc1/4) S1
Air (Rc1/4) AP

*Air pump

Vat
Drain (VP40)
F12E.ai

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
13
n Wiring Diagram
RC400G-1 /ARS /AZC
Converter Dedicated cable Detector
(CON) K9332WA (CELL)
Bk
F1 ME ME
Fail contact output W
F2 RE RE
Bw
M1 T1 T1
Maintenance contact output Gn
M2 T2 T2
(Wired by customer)
G Terminal
O1 mounting
Range switching output Rd
O2 C1 C1 screw
Bl
R1 C2 C2
Remote range switching input
R2

Power terminal box (*2) Pump


Dedicated cable (PU)
- Analog output - K9332WB
+ P1 P1
+ Bk
P2 P2
S W
G G G
AR2x2 -
+
G G
Solenoid valve
L2 L2 (option)
L1 L1 (SV6) (*1)
Dedicated cable
G K9460CN
V1 1
L2 Bk
L1 V2 2
W
G
CB1
G L2 Air pump
2 (option)
L1
(AP) (*3)
1 AR1 Dedicated cable
K9332WJ
A1 A1
Bk
* Case ground terminal must be A2 A2
grounded. W
* Ground

If this is not possible connect


to power source ground wire. (Note)
AR1,2 : Arrestors Dotted lines : external wiring F16.EPS

CB1 : Circuit breaker Use cable of 6 to 12 mm OD, however when connecting via power
Power Analog output terminal box option use cable of 9 to 11 mm OD for both power and
analog output cables.
(*1) Option /AZC applies to auto zero calibration
(*2) Option /ARS applies to version with arrestor
(*3) Option /AP applies to version with air purge pump

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
14
RC400G-2 /ARS /AZC
Converter Dedicated cable Detector
(CON) K9332WA (CELL)
Bk
F1 ME ME
Fail contact output W
F2 RE RE
Bw
M1 T1 T1
Maintenance contact output Gn
M2 T2 T2
(Wired by customer)
G Terminal
O1 mounting
Range switching output Rd
O2 C1 C1 screw
Bl
R1 C2 C2
Remote range switching input
R2

Power terminal box (*2) Pump


Dedicated cable (PU)
Analog output
- K9332WB
- P1 P1
+ Bk
+ P2 P2
S G W
G G
AR2x2 -
+
Solenoid valve
G Dedicated cable (SV1)
G
L2 K9332WC
L2 B1 1
Bk
L1 L1 B2 2
G W
L2
Solenoid valve
L1 Dedicated cable (SV2)
G K9332WD
CB1 J1 1
G L2 Bk
J2 2
2 L1 W
1
AR1 Solenoid valve
(SV6) (*1)
Dedicated cable
* Case ground terminal must be K9460CN
* Ground

grounded. V1 1
Bk
If this is not possible connect to V2 2
power source ground wire. W
AR1,2 : Arrestors
CB1 : Circuit breaker Air pump (option)
Power Analog output (AP) (*3)
Dedicated cable
K9332WJ
A1 A1
Bk
A2 A2
W

(Note)
Dotted lines : external wiring
Use cable of 6 to 12 mm OD, however when connecting via power
terminal box option use cable of 9 to 11 mm OD for both power and
analog output cables.
(*1) Option /AZC applies to auto zero calibration
(*2) Option /ARS applies to version with arrestor
(*3) Option /AP applies to version with air purge pump F17.EPS

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
15
RC400G-3 /ARS /AZC
Converter Dedicated cable Detector
(CON) K9332WA (CELL)
(Wired by customer) Bk
F1 ME ME
Fail contact output W
F2 RE RE
Bw
M1 T1 T1
Maintenance contact output M2 T2
Gn
T2
G Terminal
O1 mounting
Range switching output Rd
O2 C1 C1 screw
Bl
R1 C2 C2
Remote range switching input
R2

Power terminal box (*2) Analog Pump


Dedicated cable (PU)
-
output
- K9332WB
+ + P1 P1
Bk
P2 P2
S G W Solenoid valve
G G
AR2x2 - (SV2)
Dedicated cable
+ K9332WL
J1 1
G Bk
G J2 2
L2 L2 W
L1 L1
G Solenoid valve
L2 (SV6) (*)
L1 Dedicated cable (option)
G K9460CN
CB1 V1 1
G L2 Bk
V2 2
2 L1 W
1
AR1
Dedicated cable Terminal box Solenoid valve
K9332WK Dedicated cable (SV1)
* Ground

G
1 K9332WH
B1 1
2 Bk
B2 2
W
Power Analog
output B3
3 Solenoid valve
4 Dedicated cable (SV5)
B4
K9332WG
* Case ground terminal must be 1
Bk
grounded. 5 2
S1 W
If this is not possible connect to
6
power source ground wire. S2
Solenoid valve
AR1,2 : Arrestors
Dedicated cable (SV3)
CB1 : Circuit breaker
7 K9332WE
S3 1
8 Bk
S4 2
W

Solenoid valve
Dedicated cable (SV4)
K9332WF
Air pump (option) 1
Bk
Dedicated cable (AP) (*3) 2
W
K9332WJ
A1 A1
Bk
A2 A2
W

(Note)
Dotted lines : external wiring
Use cable of 6 to 12 mm OD, however when connecting via
power terminal box option use cable of 9 to 11 mm OD for both
power and analog output cables.
(*1) Option /AZC applies to auto zero calibration
(*2) Option /ARS applies to version with arrestor
(*3) Option /AP applies to version with air purge pump F18.EPS

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
16
Enquiry Specifications Sheet for RC400G Residual Chlorine Analyzer
Please place checkmarks () in the pertinent boxes and filling in the blanks.

1. General information
Name of your company:
Name of inquirer: Dept. or sect.: (telephone: )
Name of plant:
Measurement location:
Purpose of use:  Indication  Record  Alarm  Control
Power supply: V AC Hz

2. Process conditions
(1) Solution temperature: to , normally [°C]
(2) Solution pressure: to , normally [kPa]
(3) Ammoniacal contaminants?:  Yes  No
(4) Slurry or soiling components?:  Yes  No
(5) Name of measured solution:
(6) Composition of measured solution:
(7) Other remarks:

3. Installation environment
(1) Ambient temperature: to [°C]
(2) Installation location:  Indoors
(3) Other remarks:

4. Specification requirements
(1) Measurement object:  Free chlorine
 Residual (total) chlorine
 Combined (free) chlorine
(2) Measuring range:  to mg/L 
(3) Transmission output:  4 to 20 mA DC  1 to 5 V DC
(4) Selection of component instruments:  Residual chlorine analyzer
 Reagent tank
 Accessories
(5) Options:  With air purge pump  Unit: ppm  Stainless tagplate  Stainless stand
 With arrestor  With automatic zero calibration

5. Cleaning methods:  Glass-bead cleaning


 Glass-bead cleaning, jet cleaning, and one sand filter unit
 Glass-bead cleaning, jet cleaning, and two sand filter units
6. Others :

All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
Subject to change without notice.