General Specifications: Model RC400G Residual Chlorine Analyzer
General Specifications: Model RC400G Residual Chlorine Analyzer
n Overview
Chlorine treatment in water treatment systems, such
as municipal water supplies, serves to destroy micro-
organisms, fungi, aquatic plants and shellfish that live
in water, and thus to yield appropriate quality for the
application. However, adding too much chlorine cause
negative effects; in municipal water supplies, it may
give the water an unpleasant chlorine odor; in indus-
try, excessive chlorine may corrode plumbing; and in
waste water this may release dangerous substances
into streams and rivers. For this reason, it has be-
come very important in recent years to use a residual
chlorine analyzer when chlorine is added to water, in
order to measure the concentration of chlorine, and to
monitor and control this concentration.
The RC400G residual chlorine analyzer employs a po-
larographic system using rotating gold alloy electrodes
to provide continuous online measurement of residual
chlorine concentration. It is said that total chlorine
(residual chlorine) is the sum of free chlorine and • Uses reliable sliding contacts in contact elements.
combined chlorine. By appropriate use of reagents, • Easily maintained (easy-to-clean) cell construction.
RC400G can have three kinds of measurement possi- • Available for special version (for free chlorine meas-
bilities, namely free chlorine measurement, total chlo- urement) designed to cope with very high concentra-
rine (residual chlorine) measurement, and free chlorine tions of combined chlorine (see other literature).
measurent in case of existence of combined chlorine. Sampling equipment (conditioning equipment)
If the raw water is particularly dirty, the concentration • Adds a 2-cylinder sand filter for particularly dirty
of combined chlorine may become extremely high applications to the product line.
immediately after chlorine is added. To compensate
for this, we also offer a free chlorine measurement
system featuring improved separation performance n Standard Specifications
over the standard system (Combined chlorine insensi- 1. Residual chlorine analyzer
tive version). Measurement object: Free chlorine or residual chlorine
The RC400G is an intelligent residual chlorine ana- (total chlorine) in water
lyzer making extensive use of microprocessors incor- Measurement method: Rotating electrode type
porating know-how based on Yokogawa’s many years polarographic system
of experience in residual chlorine analysis. Measurement range: 0 to 10 mg/l (mg/l used as
concentration units per JIS K0101 and
municipal water test methods)
n Features Output range: Can be freely set within ranges with
Intelligent microprocessor-based converter spans of 1 mg/l or greater. (Range set when
• Simple, one-touch calibration. shipment. No.1: 0 to 5 mg/l No.2: 0 to 10 mg/l.
• Automatic sensor self-diagnostics (zero point, slope, Allows selection between 2 ranges (selected
response) during calibration. with contact input signal).
• Change of output range for 1mg/l span or greater, and 2-line-segment output available: Any desired
remote range switching are available. point between 0% and 100% of measurement
• Line-segment function available for output signal. span can be set as the 50% point of the output
• Simple verification of applied voltage/current charac- range.
teristics (Plateau characteristic) between electrodes. Output signal: 4 to 20 mA (load resistance, 550 Ω
Allows detailed knowledge of sensor characteristics Max.) or 1 to 5 V DC (output resistance, 300 Ω
for the target application to be obtained. Max.), isolated
• A wealth of display functions and contact output functions. Display range: -1.00 to 12.00 mg/l
• Easy-to-use operating panel. Display method: Digital display (4-digit LED)
Upgraded detector based on extensive experience Output contacts: Voltage-free (“dry”) contacts
• Improvements to rotating electrodes and glass bead Contact rating: Max. 250 V AC, Max. 2 A,
cleaning have been added for even more effective Max. 125 VA (resistive load)
automatic continuous cleaning. Max. 220 V DC, Max. 2 A, Max. 60 W
• Waterjet cleaning can be applied for the purpose of (resistive load)
particularly dirty applications.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition May 27, 2014-03
3
Paint colors Tank capacity: 100 L
Mounting stand: Munsell 0.6GY3.1/2.0 Tank materials: Rigid PVC
Other parts: Munsell 0.6GY3.1/2.0 and 2.5Y8.4/1.2 Cart materials:
Finish: Baked polyurethane resin coating Frame; Steel pipe (SPG30A)
Bracket; Steel plate (SPCC)
Operating conditions Pump (Seal-less Pump):
Ambient temperature: -5 to 55 °C Discharge flow rate; 14 to 35 L/min
(However, measure to prevent freezing are Wetted part material; Polypropylene resin, Hastel-
required if the water sample or reagent loy, ceramics, fluorinated rubber
appears to freeze.) Power supply; 100 V AC, 50/60 Hz
Ambient humidity: 5 to 95% RH (non-condensing) Power consumption; Approx. 100 VA
Storage temperature: -30 to 70 °C Power cord; PVC sheathed cable, 5 m
Installation: Indoors (A separate rainproof cover is Weight; Approx. 40 kg
required for outdoor installation. Avoid direct sunlight.)
Utilities 3. Reagent sets (for one year)
Power supply: 100/110/220 V AC ±10%, 50/60 Hz (1) For free chlorine
Power consumption: Potassium bromide extra pure:
RC400G-1¨; Approx. 65 VA 64 containers, 500 g each
RC400G-2¨; Approx. 125 VA Anhydrous sodium acetate extra pure:
RC400G-3¨; Approx. 210 VA 16 containers, 500 g each
Cleaning water (required with sand filter system): Acetic acid extra pure: 7 containers, 500 ml each
Quality; clean water (2) For residual chlorine (Total chlorine)
Pressure; 100 to 500 kPa Potassium iodide extra pure:
Flow; 10 to 12 l/min 8 containers, 500 g each
Consumption; Anhydrous sodium acetate extra pure:
Approx. 130 l/day (1-cylinder sand filter type) 3 containers, 500 g each
Approx. 470 l/day (2-cylinder sand filter type) Acetic acid extra pure:
(With standard cleaning sequence [fac- 17 containers, 500 ml each
tory settings])
(3) For high concentration of residual chlorine
Air purge (using instrument air):
(Total chlorine)
Supply pressure; 140 kPa
Potassium iodide extra pure:
Air consumption; Approx. 5 l/min
16 containers, 500 g each
Weight Anhydrous sodium acetate extra pure:
For pure water: Approx. 65 kg 3 containers, 500 g each
For raw water (1-cylinder): Approx. 70 kg Acetic acid extra pure:
For raw water (2-cylinder): Approx. 75 kg 17 containers, 500 ml each
Optional specifications (4) For combined chlorine insensitive version (Free chlorine)
Pump for air purge (if instrument air cannot be used) Potassium bromide extra pure:
Install to enable intake of clean air. 64 containers, 500 g each
Intake/discharge: 5 l/min, at 50 Hz (Intake/discharge Anhydrous sodium acetate extra pure:
outlets: Open) 87 containers, 500 g each
Maximum pressure: 80 kPa Acetic acid extra pure:
Power consumption: Approx. 23 VA 4 containers, 500 ml each
Weight: Approx. 2 kg (Note) These reagents sets are not imported from Japan be-
Regulatory Compliance: cause of both safety and transportation cost issues.
EMC Regulatory Arrangement in Australia and New Purchase them directly in your local reagent supplier.
Zealand (RCM) EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard Class A n Characteristics
한국 전자파적합성 기준 (Percent display is computed with respect to whichever
of output range 1 or output range 2 has the highest up-
2. Reagent Tank Model RC401G per range value)
(1) Reagent tank (RC401G-A) Repeatability: 2%
Capacity: 100 L Linearity: ±3%
Materials: Rigid PVC Drift: Zero drift; ±1%/month Max.
Weight: Approx. 15 kg Span drift; -5%/month Max.
(2) Reagent tank with stand (RC401G-B) Response time: Displayed as 90% response time
Incorporates the reagent tank in item (1) with a stand For pure water; Approx. 3 min
and a needle valve. For raw water; Approx. 4 min
Stand materials: carbon steel (Within 3 min from standard liquid inlet)
Stand color: Munsell 2.5Y8.4/1.2 Temperature compensation error (water tempera-
Stand finish: Baked polyurethane resin coating ture): ±1% Max. (Temperature compensa-
Weight: Approx. 25 kg tion range: 0 to 40 °C)
(3) Reagent mixing tank with cart (RC401G-C) Effect of ambient temperature: ±0.5%/10 °C
A reagent mixing tank mounted on a cart. Effect of power variation: ±0.5%/10% of rated voltage
With manual mixer, and pump for transferring reagents. Effect of combined chlorine: 5% or less, of concentra-
Note: This tank is for reagent mixing, and is not a substi- tion of combined chlorine for combined
tute for the reagent tank. chlorine insensitive version
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Feb.28, 2014-02
4
n Model and Suffix Codes
(1) Residual chlorine analyzer [Style: S3]
*1: “Application” indicates general guidelines. Select the cleaning equipment appropriate for the contamination in the sample.
*2: Measurement object selection is indicated in the following.
*3: Less affected by combined chlorine in free chlorine measurement.
*4: When ordering the RC400G, select /NR and get reagent sets for start-up from local reagent supplier.
These reagent sets can not be exported from Japan due to both safety and transportation issues.
*5: Stanchion, base and bracket are stainless steel.
*6: Please contact Yokogawa regarding the adequate number of the cylinder.
*7: Installation of air purge is always required for all aplication. If instrument air cannot be used, select the air purge pump.
Measurement object
Application : Can be measured
Free chlorine Total chlorine x: Cannot be measured
Water purification: raw water (*1)
*1: If ammoniacal contaminants are present in large quantity,
Water purification: mixed water, sedi- high concentrations of combined chlorine may remain
(*1)
mentation water if sampling is performed soon after chlorine injection in
Water purification: mains water prechlorination treatment, or in intermediate treatment
without prechlorination treatment. In this case, combined
Sea water x x (*3) chlorine type is recommended
Factory wastewater, treated effluent x x *2: Oxidizing or reducing agents other than chlorine may be
present. If so, chlorine concentration measurement may
Factory cooling water (industrial water) x (*2) not be possible.
Factory drinking water *3: There are Residual chlorine analyzer for sewage
secondary treatment or sea water. Please ask
Sewage secondary treatment x x (*3) Yokogawa.
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition May 27, 2014-03
5
Accessories
1. Residual chlorine meter (RC400G) 2. Reagent tank (RC401G-A, RC401G-B)
Name Quantity Remark
Name Q’ty Remark
Polishing powder 1 bottle For polishing electrode
Couplings 2 sets
Lubricating oil 1 For metering pump drive
Outside diam-
Fuses 4 each 1 A and 3 A (for spare) Polyethylene tubing 5m eter 6 mm, Inside
Glass beads 1 (including for spare) (2 bags) diameter 4 mm.
Valve sheet 4 For metering pump (for spare) Note: These are not included in reagent mix-
ing tank (RC401G-C.)
1 For sample pump (for spare)
Bellophram
1 For reagent pump (for spare)
Special tool 1 For pump valve replacement (*) These reagents sets are not imported
Tool 1 For Bellophram replacement from Japan because of both safety and
transportation cost issues.
Allen wrenches 1 set 1.5, 2, 2.5, 4, 5, 6 (mm) Purchase them directly in your local
reagent supplier.
Reagent set (*) For startup
(2) Reagent tank
Model Suffix code Option code Description
RC401G ----------------------- -------------------- Reagent tank
Type -A -------------------- Reagent tank only
-B -------------------- Reagent tank with stand
-C -------------------- Reagent mixing tank with mobile stand
- -HM -------------------- With manual mixer
- *A -------------------- Style A
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Jun.05, 2013-00
6
Recommended
Name Part No. Description Q'ty
replacement interval
Air pump (100 V) K9087XA For air purgepump 1 3 years
Air pump (110 V) K9087XF For air purgepump 1 3 years
Air pump (200 V) K9087XG For air purgepump 1 3 years
Air pump (220 V) K9087XH For air purgepump 1 3 years
Lubricating oil (engine oil) K9041RA For metering pump drive 1 1 month (*4)
Bellophram L9819AA For reagent pump 1 6 months
Bellophram L9819AB For sample water pump 1 6 months
Valve sheet K9041HC For metering pump 1 6 months
Activated charcoal filter L9862AY For automatic zero calibration 1 Yearly
Filter K9332NN For head tank or sand filter 1 Yearly
(*4) Recommended maintenance interval for lubricating the metering pump drive assembly.
n External Dimensions
(1) RC400G-1 Residual chlorine analyzer for pure water and municipal water
unit : mm
Standard solution inlet Air purge inlet
Rc1/4(female) Rc1/4(female) Maintenance Maintenance
space space
Reagent inlet
4-ø15 holes Approx. 200 530 Approx. 200
Rc1/4 (female)
for anchor bolt
100
Stand
600
Ball valve (V3)
350
Approx. 500
20 490 20
Detector
(CELL)
Power terminal
Head tank
box (*1) (HD)
Metering
Meas. / Std. solution pump Drain
1450
530 550 50 85 60 40
150 150
F01E.ai
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Jun.06, 2013-00
7
(2) RC400G-2 Residual chlorine analyzer for raw water (1-cylinder)
unit : mm
Standard solution inlet Air purge inlet
Rc1/4 (female) Rc1/4 (female)
600
Strainer (S1)
Approx. 500
100
Maintenance
Front
20 490 20 space
Metering
Meas. / Std. solution pump
1450
530 550 50
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
8
(3) RC400G-3 Residual chlorine analyzer for raw water (2-cylinder)
unit : mm
Standard solution inlet Air purge inlet
Rc1/4 (female) Rc1/4 (female)
Maintenance Maintenance
Reagent inlet space space
Rc1/4 (female) 4-ø15 holes
Approx. 200 530 Approx. 200
for anchor bolt
100
Stand
600
Ball valve (V8)
350
Approx. 500
Strainer (S1)
Maintenance
100
Front
space
20 490 20
530 550 50 85 60 40
150 150
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
9
Reagent tanks
l RC401G-A Reagent tank l RC401G-B Reagent tank
Unit: mm
Reagent tank
415 450
Reagent output
415 (Rc1/4)
Needle
valve
Manual mixer Approx.1600
Reagent input
Level gauge
Approx.
Reagent Approx. 700
output Approx.
(Rc1/4) 955
805
225
20 410 (20) 250 4-ø13 holes
100
450 450 (100) F0204.ai
Approx.1280
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
10
n Piping Diagrams
(1) RC400G-1 Residual chlorine analyzer for pure water and municipal water
Tubing materials
ø6 x ø4
polyethylene tube Converter
VP16 pipe
VP40 pipe CELL Measuring
tank
* option
PU Metering
pump
Ball valve
V2
COM
V1
Measuring sample (VP16)
Standard solution (Rc1/4)
V3 Strainer
Reagent (Rc1/4) S1
Air (Rc1/4) AP
Vat
*Air pump
Drain (VP40)
F02E.ai
VP16 pipe
V12
VP40 pipe CELL Measuring
tank
* option SV6 NO
3-way solenoid
valve
COM
NC Activated
charcoal
filter PU Metering
F11 pump
Ball
valve
Ball valve V11 V2
V1
Measuring sample (VP16) COM
Pressure reducing valve
V13
Calibration solution (VP16) PR
Standard solution (Rc1/4)
V3 Strainer
Reagent (Rc1/4) S1
Air (Rc1/4) AP
Drain (VP40)
F04E.ai
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
11
(2) RC400G-2 Residual chlorine analyzer for raw water (1-cylinder)
Tubing materials
ø6 x ø4
polyethylene tube
Converter
Sand filter
ø8 x ø6 CON
polyethylene tube F1
ø22 x ø15
braided wire reinforced
soft PVC tube
ø33 x ø25
braided wire reinforced
soft PVC tube
CELL Measuring
V1 tank
VP16 pipe
VP40 pipe
V2
* option
PU Metering
pump
Ball valve
V3 V5
COM
Measuring sample (VP16) SV1 SV2
Pressure reducing valve Solenoid
V4 valve
Tap water (VP16)
Ball valve PR
Standard solution (Rc1/4)
V6 Strainer
Reagent (Rc1/4) S1
Air (Rc1/4) AP
*Air pump
Vat
Drain (VP40)
F06E.ai
ø6 x ø4
polyethylene tube Converter
Sand filter CON
ø8 x ø6 F1
polyethylene tube
ø22 x ø15
braided wire reinforced soft PVC tube
ø33 x ø25
braided wire reinforced soft PVC tube
CELL Measuring
VP16 pipe V1 tank
V12
VP40 pipe
SV6 NO 3-way solenoid V2
* option valve
COM Metering
NC Activated PU pump
charcoal
filter Ball
V3
F11 valve
V5
Measuring sample (VP16) SV1 COM
Solenoid V11
Pressure reducing valve
valve
V4 SV2
Tap water (VP16)
PR
Ball valve
Standard solution (Rc1/4)
V6 Strainer
Reagent (Rc1/4) S1
Air (Rc1/4) AP
*Air pump
Vat
Drain (VP40)
F08E.ai
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
12
(3) RC400G-3 Residual chlorine analyzer for raw water (2-cylinder)
Tubing materials
ø6 x ø4 Converter
polyethylene tube
Sand filter CON
F1 F2
ø8 x ø6 Sand filter
polyethylene tube
ø22 x ø15
braided wire reinforced Solenoid valve
soft PVC tube SV3 SV4
ø33 x ø25
braided wire reinforced CELL Measuring
soft PVC tube V1 V5 tank
VP16 pipe
V2
VP40 pipe
* option PU Metering
pump
V4 V3
Ball valve V7
Measuring sample (VP16) COM SV2
Pressure reducing valve SV1 SV5
V6
Tap water (VP16) PR
Standard solution (Rc1/4)
Strainer
Reagent (Rc1/4) S1
V8
Air (Rc1/4) AP
*Air pump
Vat
Drain (VP40)
F10E.ai
Tubing materials
ø6 x ø4 Converter
polyethylene tube
CON
ø8 x ø6 F1 Sand filter F2
polyethylene tube Sand filter
ø22 x ø15
braided wire reinforced Solenoid valve
SV3 SV4
soft PVC tube
ø33 x ø25
braided wire reinforced CELL Measuring
soft PVC tube V1 V5 tank
V12
Metering
VP16 pipe 3-way solenoid pump
valve SV6 COM V2
VP40 pipe
NO
* option NC PU
V4 V3 F11
Ball valve Activated V7
charcoal
Measuring sample (VP16) SV1 V11 filter COM SV2
Pressure reducing valve Ball SV5
V6 valve
Tap water (VP16) PR
Standard solution (Rc1/4)
V8 Strainer
Reagent (Rc1/4) S1
Air (Rc1/4) AP
*Air pump
Vat
Drain (VP40)
F12E.ai
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
13
n Wiring Diagram
RC400G-1 /ARS /AZC
Converter Dedicated cable Detector
(CON) K9332WA (CELL)
Bk
F1 ME ME
Fail contact output W
F2 RE RE
Bw
M1 T1 T1
Maintenance contact output Gn
M2 T2 T2
(Wired by customer)
G Terminal
O1 mounting
Range switching output Rd
O2 C1 C1 screw
Bl
R1 C2 C2
Remote range switching input
R2
CB1 : Circuit breaker Use cable of 6 to 12 mm OD, however when connecting via power
Power Analog output terminal box option use cable of 9 to 11 mm OD for both power and
analog output cables.
(*1) Option /AZC applies to auto zero calibration
(*2) Option /ARS applies to version with arrestor
(*3) Option /AP applies to version with air purge pump
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
14
RC400G-2 /ARS /AZC
Converter Dedicated cable Detector
(CON) K9332WA (CELL)
Bk
F1 ME ME
Fail contact output W
F2 RE RE
Bw
M1 T1 T1
Maintenance contact output Gn
M2 T2 T2
(Wired by customer)
G Terminal
O1 mounting
Range switching output Rd
O2 C1 C1 screw
Bl
R1 C2 C2
Remote range switching input
R2
grounded. V1 1
Bk
If this is not possible connect to V2 2
power source ground wire. W
AR1,2 : Arrestors
CB1 : Circuit breaker Air pump (option)
Power Analog output (AP) (*3)
Dedicated cable
K9332WJ
A1 A1
Bk
A2 A2
W
(Note)
Dotted lines : external wiring
Use cable of 6 to 12 mm OD, however when connecting via power
terminal box option use cable of 9 to 11 mm OD for both power and
analog output cables.
(*1) Option /AZC applies to auto zero calibration
(*2) Option /ARS applies to version with arrestor
(*3) Option /AP applies to version with air purge pump F17.EPS
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
15
RC400G-3 /ARS /AZC
Converter Dedicated cable Detector
(CON) K9332WA (CELL)
(Wired by customer) Bk
F1 ME ME
Fail contact output W
F2 RE RE
Bw
M1 T1 T1
Maintenance contact output M2 T2
Gn
T2
G Terminal
O1 mounting
Range switching output Rd
O2 C1 C1 screw
Bl
R1 C2 C2
Remote range switching input
R2
G
1 K9332WH
B1 1
2 Bk
B2 2
W
Power Analog
output B3
3 Solenoid valve
4 Dedicated cable (SV5)
B4
K9332WG
* Case ground terminal must be 1
Bk
grounded. 5 2
S1 W
If this is not possible connect to
6
power source ground wire. S2
Solenoid valve
AR1,2 : Arrestors
Dedicated cable (SV3)
CB1 : Circuit breaker
7 K9332WE
S3 1
8 Bk
S4 2
W
Solenoid valve
Dedicated cable (SV4)
K9332WF
Air pump (option) 1
Bk
Dedicated cable (AP) (*3) 2
W
K9332WJ
A1 A1
Bk
A2 A2
W
(Note)
Dotted lines : external wiring
Use cable of 6 to 12 mm OD, however when connecting via
power terminal box option use cable of 9 to 11 mm OD for both
power and analog output cables.
(*1) Option /AZC applies to auto zero calibration
(*2) Option /ARS applies to version with arrestor
(*3) Option /AP applies to version with air purge pump F18.EPS
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
16
Enquiry Specifications Sheet for RC400G Residual Chlorine Analyzer
Please place checkmarks () in the pertinent boxes and filling in the blanks.
1. General information
Name of your company:
Name of inquirer: Dept. or sect.: (telephone: )
Name of plant:
Measurement location:
Purpose of use: Indication Record Alarm Control
Power supply: V AC Hz
2. Process conditions
(1) Solution temperature: to , normally [°C]
(2) Solution pressure: to , normally [kPa]
(3) Ammoniacal contaminants?: Yes No
(4) Slurry or soiling components?: Yes No
(5) Name of measured solution:
(6) Composition of measured solution:
(7) Other remarks:
3. Installation environment
(1) Ambient temperature: to [°C]
(2) Installation location: Indoors
(3) Other remarks:
4. Specification requirements
(1) Measurement object: Free chlorine
Residual (total) chlorine
Combined (free) chlorine
(2) Measuring range: to mg/L
(3) Transmission output: 4 to 20 mA DC 1 to 5 V DC
(4) Selection of component instruments: Residual chlorine analyzer
Reagent tank
Accessories
(5) Options: With air purge pump Unit: ppm Stainless tagplate Stainless stand
With arrestor With automatic zero calibration
All Rights Reserved. Copyright © 1994, Yokogawa Electric Corporation GS 12F4A1-E 6th Edition Apr.01, 2013-00
Subject to change without notice.