Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 10

Opeoluwa Matthews

Opeoluwa Matthews
Advanced Lab Individual Experiment

The Frank-Hertz Experiment

Thursday, March 3, 2011

The Frank-Hertz Experiment: Demonstrating the Quantization of Kinetic Energy


Transfer in Electron-Atomic Collisions

Abstract:
The purpose of this experiment was to show that the transfers of energy to atomic electrons of
mercury through collisions occurs for discreet energies and are consistent with the quantum
mechanically allowable energy transitions for mercury. This was demonstrated by accelerating electrons
through increasing potentials in a Frank-Hertz tube and observing the corresponding currents at the end
of the tube.

Introduction:
This experiment replicates the Frank-Hertz experiment conducted by German physicists James
Frank and Gustav Ludwig Hertz in 1914. It succeeded the Bohr-Heisenberg Model of the atom
introduced in 1913, which sought to explain the spectral emission lines of the hydrogen atom. Bohr
proposed that the change in energy levels were related to the frequency of the radiations by ΔE=hv. It
was eventually shown that that absorption of photons occurred for only discrete frequencies of light and
thus discrete energies hv.

The Frank-Hertz experiment tests the generalization of Bohr’s findings. It tests whether or not
kinetic energy transfers, like photons, occurred also in discrete energies. To test this, Frank and Hertz
bombarded gases with electrons in a chamber and tested whether the electrons lost their kinetic energy
at discrete energies. Their findings won them the Nobel Prize in Physics in 1925.

Experimental Procedure:
Experimental Setup:

The apparatuses required for the Frank-Hertz Experiment include Frank-Hertz tube containing
mercury, a copper filament, a heating grill, a variable-potential Anode and a variable-potential Collector.
These can be seen in Figure 1 below.

1
Opeoluwa Matthews

Figure 1. Schematic of the Frank-Hertz Experiment.

As can be seen in the schematic above, the mercury atoms are vaporized within the Frank-Hertz tube.
Electrons are then emitted by thermionic emission from the filament. The Collector at the end of the
tube, connected to a programmable DC power supply, accelerates the electrons through the Frank-Hertz
tube at varying voltages. The electrons are then able to collide with the mercury atoms and move
towards the Collector.

If the electrons are at the right energies to transit the mercury atoms to a higher energy level, the
collision will be inelastic. But if the electrons are not at the discreet energy levels required, the collision
is simply elastic and the electrons retain their energy.

Note that after the collision of the electrons with a mercury atom, the Collector could still accelerate
them to higher energies. For this reason, there is a Grid Anode at the end of the Frank-Hertz tube that
provides a retarding potential to ensure that those electrons that have had inelastic collisions but have
still been accelerated after the collision will be drawn to the Grid Anode and not make it to the
Collector.

2
Opeoluwa Matthews

At the Collector is a picoammeter to measure the current at the end of the tube. The current indicates
how many electrons are losing their kinetic energies to inelastic collisions.

Conducting the Experiment:

The Frank-Hertz tube was placed at a certain temperature, typically between 150 and 200 oC. The Grid
Anode was placed at a retarding potential, typically between 1.5 and 2 volts. Then the Collector’s
accelerating voltage was varied from around 2 volts to 30 volts at increments of .1 volts, with the
current at the collector recorded at each increment. Then, a graph of current against voltage was
plotted to show the collision patterns (if any) of the electrons.

Results and Interpretation:


Below will be a series of graphs detailing what experiments were performed. The error bars will
be in black.

Figure 2 below is a graph of current at the Collector against voltage taken with Annode Grid
retarding voltage of 1.5v.

1_5v__150-155
-9
4 10

-9
3.5 10

-9
3 10

-9
2.5 10
current(nA)

-9
2 10

-9
1.5 10

-9
1 10

-10
5 10

0
0 5 10 15 20 25

volts(v)

3
Opeoluwa Matthews

Figure 2. Graph of current at collector versus voltage. Results were recorded for .1 volt increments of the
Collector’s potential ranging from 2 volts to 23 volts. The Annode Grid was set at a retarding potential of 1.5volts.
The chamber was at a temperature between 150 and 155 oC.

In the above graph, the average distance between the peaks was found to be 4.96±.07v.

In Figure 3. Below, measurements are recorded for Annode Grid retarding voltage of 1.5v, like figure 1,
but the temperature of the chamber is raised.

1_5v__170-180
-9
1.2 10

-9
1 10

-10
8 10

-10
Current(nA)

6 10

-10
4 10

-10
2 10

-10
-2 10
0 5 10 15 20 25 30

Volts(v)

Figure 3. Graph of current at collector versus voltage. Results were recorded for .1 volt increments of the
Collector’s potential ranging from 2 volts to 23 volts. The Annode Grid was set at a retarding potential of 1.5volts.
The chamber was at a temperature between 170 and 180 oC.

4
Opeoluwa Matthews

Here, the average distance between peaks was found to be 5.07 ±.8 volts.

Comparing Figure 2 and 3, it is interesting to see how raising the temperature reduces the overall
current at the Collector. Figure 4 is Figure 2 and 3 plotted on the same axis to demonstrate this more
clearly.

1_5v__170-180
1_5v__150-155

1_5v__170-180
-9
4 10

-9
3 10

-9
2 10
Current(nA)

-9
1 10

-9
-1 10
0 5 10 15 20 25 30

Volts(v)

Figure 4. Graph of current versus voltage for 1.5v Annode Grid, 150-155 oC chamber temperature plotted
on the same axis with graph of current versus voltage for 1.5v Annode Grid and 170-180 oC.

The lowering of the overall current is expected when the temperature is raised because the mercury
atoms have more kinetic energy with higher temperatures, thus allowing for more collisions of the
electrons at lower electron energies.

5
Opeoluwa Matthews

Figure 5 below is a graph of current versus voltage for 1.7v Annod Grid and 150-155 oC chamber
temperature.

1_7v__150-155
-9
3.5 10

-9
3 10

-9
2.5 10

-9
current(nA)

2 10

-9
1.5 10

-9
1 10

-10
5 10

0
0 5 10 15 20 25

volts(v)

Figure 5. Graph of current at collector versus voltage. Results were recorded for .1 volt increments of the
Collector’s potential ranging from 2 volts to 23 volts. The Annode Grid was set at a retarding potential of 1.7volts.
The chamber was at a temperature between 150 and 155 oC.

Here, the average distance between peaks was found to be 5.19 ± .08v. Again, comparing Figure 5 with
Annode Grid at 1.7v with Figure 1 with Annode Grid at 1.5v, both at the same chamber temperature, we
can see how a change in the Annode Grid voltage can affect the current v voltage curve. Figure 6 below
plots the graphs of both above delineated configurations on the same axis.

6
Opeoluwa Matthews

1_7v__150-155
1_5v__150-155

1_7v__150-155
-9
4 10

-9
3.5 10

-9
3 10

-9
2.5 10
current(nA)

-9
2 10

-9
1.5 10

-9
1 10

-10
5 10

0
0 5 10 15 20 25

volts(v)

Figure 6. Graph of current versus voltage for 1.5v Annode Grid, 150-155 oC chamber temperature plotted
on the same axis with graph of current versus voltage for 1.7v Annode Grid and 150-155 oC.

As expected, we see that increasing the Annode Grid voltage lowers the overall current at the Collector.
This is because a higher retarding voltage at the Annode Grid means less electrons have the energy to
“make it” to the Collector.

In Figure 7 below, measurements are taken along the inside of a dip to detect smaller energy transitions.
Hence measurements are taken from 16 to 17volts in .01volt increments for 1.5v Grid voltage and 170-
180 oC chamber temperature. Hence it is similar to zooming into Figure 3 between 16 and 17 volts.

7
Opeoluwa Matthews

zoomed in 16-17v 1_5v__170-180


-10
3.5 10

-10
3 10

-10
2.5 10
Current(nA)

-10
2 10

-10
1.5 10

-10
1 10
15.8 16 16.2 16.4 16.6 16.8 17 17.2

Volts(v)

Figure 7. Graph of Collector current versus voltage taken between 16 and 17 volts in .01volt increments.
Annode Grid retarding voltage is 1.5 volts and chamber temperature is 170-180 oC.

As can be seen, there is also oscillation going on in the small interval between 16 and 17 volts. The
average distance between peaks was found to be 0.20 ± .01v.

Figure 8 below, like Figure 7, takes measurements at .01volt increments but between 19.5 and 20.5
volts.

8
Opeoluwa Matthews

zoomed in 19.5-20.5v 1_5v__170-180


-10
1.8 10

-10
1.6 10

-10
1.4 10

-10
Current(nA)

1.2 10

-10
1 10

-11
8 10

-11
6 10

-11
4 10
19.4 19.6 19.8 20 20.2 20.4 20.6

Volts(v)

Figure 8. Graph of Collector current versus voltage taken between 16 and 17 volts in .01volt increments.
Annode Grid retarding voltage is 1.5 volts and chamber temperature is 170-180 oC.

Here, since the peaks are all but prominent and the errors are so large that the peaks could be slid up
and down to produce a linear curve, it seems unreasonable to calculate the distance between peaks.

Error Analysis:

9
Opeoluwa Matthews

Current vs Voltage Graphs:

Current:

All the error expected in this area is related to the equipment. For experiments with 0.1 volt increments,
the error due to the machine was measured by taking 2 volt increments within the range of the
measurements taken and calculating 10 times each what the value of the current is. The same was done
for the .01v-increment measurements, except the errors were measured at .1volt increments, instead.
The standard deviation was found and each value was then mapped to the error regime that
corresponds to the mark of the last regime and the standard deviation calculated for its maximum value.

Of course, a computer algorithm, in Python, was used to make this mapping as it would be impossible
within the time frame of the experiment to do it manually. A python algorithm was also used to “clean
up” the data as copied from the logger pro interface, which would again be impossible if done manually.

Voltage:

All the error expected is related to the fact that when a command from the computer is sent to the
variable-potential collector to change to a certain potential, the value is always off by a small amount.
That amount was found to be typically a constant 0.16% of the command that was sent. That value was
used as the error.

Conclusion:
The average distance between peaks for the normal 0.1v-increment sweeps was found to be 5.07
±0.07eV (as it is electrons we are dealing with). This value is 3.5% off the expected value of the transition
of Mercury from the 6s6s 1S0 ground state to the 6s6p 3P1 excited state, which is 4.9eV.

By quantum mechanics, the above stated transition is by far the most probable transition given the
accelerating potentials used. Hence, at higher accelerating potentials, the peaks simply correspond to
when electrons have integer multiples of the energy for transition of Mercury from the 6s6s 1S0 ground
state to the 6s6p 3P1 excited state. So, what happens in the chamber is they are able to have multiple
inelastic collisions until they have no more kinetic energy.

For the smaller voltage increments, an oscillation was spotted, with an average peak width of 0.20 ± .
01v. Nonetheless, after significant searches, a transition with energy as low as 0.20eV was not found.
Nonetheless, the clear sign of periodicity in the current-voltage graphs show there is a good chance that
such a transition can be found.

10

You might also like