Not Required: Verification of Equipment (Weekly)
Not Required: Verification of Equipment (Weekly)
) UCDS(W)_QC-002511_05-07-2021_v01_00
ULTRASONIC Form Rev. 5
CALIBRATION DATA
Date 7/5/2021
SHEET
Page 1 of 3
1. EQUIPMENT
1.1 Instrument
Model No. Manufacturer Serial number Calibration Date
Mainframe OMNISCAN-MX2 Olympus OMNI2-101098 1/26/2021
UT module OMNI-M2-PA32128PR Olympus QC-002511 1/26/2021
1.2 Probes and Wedges
TECHNIQUE T1-T3 Model No. Manufacturer Serial number
Probe 2.25DM7X4PM-19X12-A17 Olympus U4107
Wedge SA17-DN55L0-IHC Olympus ABWX2786
TECHNIQUE T5 Model No. Manufacturer Serial number
Probe 2.25DM7X4PM-19X12-A17 NOT REQUIRED
Olympus
Wedge SA17-DN55L0-IHC Olympus
Tolerance: The sensivity and signal-to-noise ratio shall be within ± 6 dB of the base measurement.
Tolerance: The pulse duration shall not be greater than 1.5 times the base measurement.
Tolerance: The deviation for the squint angle shall be within ± 1 dB
Doc. No. UCDS(W)_QC-002511_05-07-2021_v01_00
ULTRASONIC Rev. 5
CALIBRATION DATA
Date 7/5/2021
SHEET
Page 3 of 3
Tolerance: The sensivity and signal-to-noise ratio shall be within ± 6 dB of the base measurement.
Tolerance: The pulse duration shall not be greater than 1.5 times the base measurement.
Tolerance: The deviation for the squint angle shall be within ± 1 dB