Professional Documents
Culture Documents
Basic Hemodynamic Monitoring Notebook
Basic Hemodynamic Monitoring Notebook
Inside:
• Module Outline
• Lesson Objectives
• Lesson Summary
• Lesson Resource Files
• Lesson Practice Pearls
2
Module Outline
Module 4 - Basic Hemodynamic Monitoring
Introduction
In this lesson we look at the hemodynamic monitoring system itself which offers the clinician a relatively simple way of
measuring the dynamic conditions of the cardiovascular system at the bedside. We learn how to set up a hemodynamic
pressure monitoring system and explore the fact that with advances in electronics and computer systems, there have also
been advances in the monitoring of cardiovascular hemodynamics.
Blood Pressure
• Monitoring blood pressure is the most basic way to monitor a patient’s hemodynamic status.
• Can be done noninvasively using a sphygmomanometer and stethoscope or invasively using an arterial catheter and
electronic monitor
• If system is set up correctly, these BP values are highly reliable and valid:
• When measured noninvasively via a cuff, BP values depend on the detection of pulsations within the artery to
determine the systolic and diastolic values.
• When measured directly via a catheter placed inside the artery, BP values are taken directly from pressures detected
inside the vessel itself.
• Note: Different standards exist for using the fluid filled pressure monitoring system to measure nonhemodynamic
pressures.
Strip Recorder
• Allows user to document the waveform on paper for further analysis
• Especially important when waveform has significant movement from baseline
• Most can record in single channel (one waveform per strip) or dual channel (2 waveforms per strip) - When analyzing
hemodynamic waves, select dual channel when available.
Introduction
Here we explore how to ensure that the hemodynamic values you get are accurate and that your monitoring system is
accurate.
Transducer Placement
• Using a leveling device, place transducer at a level horizontal with the phlebostatic axis.
• Any type of level fine, electric, laser or even simple carpenter’s level
• Never eyeball or estimate. Can result in large variation in your numbers.
• Transducer placed by positioning the air-reference stopcock at the same horizontal level as the phlebostatic axis.
• Some institutions place transducers directly on bed or on patient’s chest at phlebostatic axis.
• Many institutions mount transducer in a manifold on an IV pole at the bedside (instead of directly on bed or
patient’s chest.)
• In IV pole approach, important to relevel transducer whenever the head of the bed position is changed.
Zero Referencing
• After leveling transducer, we zero the transducer. This is done to ensure that pressures being measured are result of
patient’s hemodynamic system and not of atmosphere or fluid.
• Zeroing basically tells transducer to ignore the effects of:
• Weight of the atmosphere
• Weight of the fluid in the system
• 5 basic steps for zeroing a transducer:
• Level transducer.
• Turn stopcock nearest transducer off to patient and open to the capped stopcock port.
• Remove cap to stopcock port, opening it to air.
• Activate zero function key on monitoring device.
• When monitor indicates that system is properly zeroed, replace cap to stopcock port. Then turn stopcock back so
that it’s off to the cap and open to the patient.
Scale for pressure Set pressure monitoring scale Check monitor to assure
being measured to appropriate level for appropriate level has been
incorrect. pressure being measured. set.
Clotted catheter. Keep flush bag inflated to Check flush bag to assure it
300mmHg is inflated properly.
Aspirate catheter if possible.
Follow hospital policy
regarding clearing of clotted
catheters. DO NOT
ATTEMPT TO FLUSH
FORWARD.
Dampened Air bubbles in tubing / Flush system with gravity Flush bubbles from system.
waveform (over stopcocks. only. Check tubing and stop
damped) cocks carefully for air bubbles
prior to connecting.
Readout values Transducer level too ALWAYS check level with Confirm level at phlebostatic
too high or too high (readouts too leveling device. axis.
low. low).
Transducer level too ALWAYS check level with Confirm level at phlebostatic
low (readouts too leveling device. axis.
high).
Ringing / fling Excess tubing length Use non compliant tubing. Change tubing, shorten
or whip in or use of compliant Make sure tubing is less than length.
catheter. (under tubing. 4 feet from the catheter
damped) connection to the transducer.
Excessive stop cocks Keep inline stop cocks to a Remove excess stop cocks.
in system minimum.
Bleed back into Loose connections in NOTE: If blood gets into the Check & tighten all
tubing system. transducer, the transducer will connections.
need to be changed.
High Risk for Breaks in sterile ALWAYS maintain sterile If infection is suspected,
infection from technique technique. Keep stop cock contact MD to review
invasive lines. ports sterile and always antibiotic coverage, When
maintain sterility of stopcock discontinuing catheter,
ports. culture catheter tip.
Practice Pearls
Transducer Placement
Remember, the phlebostatic axis is on the patient’s chest. Every time you raise or lower
the head of the patient’s bed, the transducer will need to be releveled.
Zero Referencing
When removing the cap from the stopcock, care must be taken not to contaminate the
cap. You can do so by placing it on a piece of sterile gauze while zeroing, or you can
replace the cap with a new sterile one.
Lesson 2
Arterial Pressure Monitoring
Introduction
In this lesson we explore arterial pressure monitoring, one of the most common forms of invasive pressure monitoring and a
common invasive procedure among the critically ill. We learn about commonly used arterial catheter insertion sites and look
at different arterial catheter types and the insertion techniques. We also explore the complications you may encounter with
an arterial line. We analyze and discuss some different arterial waveforms and review arterial line blood sampling removal.
We also briefly discuss the difference between direct and indirect blood pressure monitoring.
Arterial Catheter
• Catheter inserted into an artery • Arterial sites used
Allen’s Test
• Assesses the adequacy of blood flow and perfusion through ulnar artery
• Performed by compressing both the radial and ulnar arteries at wrist
• Use nondominant hand
• 2 Steps
Pressure Monitoring System Setup
• Needs to be done prior to catheter insertion
• Includes setting up transducer system, flush solution and pressure bag should be set up at the bedside.
• Process of set-up
• Disadvantages
• Complications can result from prolonged use
• Thrombus formation possible
• Cannulation of artery requires smaller catheter which can cause waveforms that “over-shoot” giving inaccurate/overly
large results
• Potential injury to adjacent nerves with hematoma formation
• Disadvantages:
• Requires patient remain supine • Cannot be used if intra-aortic balloon pump is to be
• Difficult to immobilize if patient agitated or restless used
Introduction
Here we look at the infrequent, but possible complications that can occur with arterial pressure monitoring.
Patient Safety
• Patient safety issues related to arterial pressure monitoring:
• The chance for accidental administration of medication through arterial line
• Blood loss due to stopcock turned wrong way
• After dicrotic arch, the fall in waveform pressure is smooth and progressive (called the reflection wave) during diastole until
the next systolic upstroke.
Atrial Fibrillation
• Often associated with variation in amplitude of the arterial systolic waveform
• The longer the R to R interval, the more time the ventricle has to fill, and therefore the larger stroke volume it ejects.
• Therefore, peak systolic pressure is higher in the beat terminating the longer R to R interval.
Introduction
Here we explore the fact that noninvasive (cuff) blood pressure and direct arterial pressure are not the same.
Introduction
Here we look at both open and closed systems for drawing blood for lab tests.
Closed System
• Highly recommended because:
• Reduces likelihood of contamination or introduction of bacteria into system
• Contributes to blood conservation since no need to discard aspirated solution/blood
• Procedures for using closed system
Here we look at the timing and steps of removing the arterial catheter.
Arterial Catheter Removal
• Done when it is no longer required for continuous arterial pressure monitoring or frequent blood gas samples
• Steps:
• Remove dressing
• Aspiration of blood
• Apply pressure
• Pressure dressing
Catheter
Placement Advantages Disadvantages
Radial Artery • The artery is easy to locate • Prolonged use may lead to
and cannulated. complications such as vascular
• The artery is accessible occlusion.
during most types of • Thrombus formation may occur
surgery. occluding the catheter.
• Immobilization of the site is • Cannulation of the artery
fairly easy and comfortable requires a smaller catheter
for the patient. which can contribute to “over-
shoot” of the waveform giving
an “over estimation” of the
systolic pressure.
• Potential injury to the adjacent
nerves with hematoma
formation.
Femoral Artery • Pressure line tends to • Increased risk of bleeding
remain patent longer with (larger artery) and
fewer complications. retroperitoneal bleeding.
• Waveform tends to be more • Requires the patient to remain
reflective of the actual supine to prevent bending the
aortic waveform, i.e., less catheter.
distortion. • Difficult to immobilize if patient
• Femoral artery is larger is agitated or restless.
accommodating a larger • Cannot use if an intra-aortic
catheter. balloon pump is to be used.
25
Many clinicians will compare the noninvasive (cuff) blood pressure with the direct
arterial pressure assuming both are equal. A common practice is to state the arterial line
pressure correlates with the cuff pressure. Of interest is which set of numbers to accept
if they do not correlate.
27
Practice Pearls
• Complications
Introduction
In this lesson we learn about various different central venous catheter types, insertion techniques, and sites. We also
explore how to manage CVP lines, analyze CVP waveforms, and handle some of the more common complications
encountered with CVPs. We also briefly discuss the use of water manometers to obtain a CVP reading and compare that
with values obtained using the pressure transducer.
• Because CVP was discovered to be unreliable for use in measuring left heart function, pulmonary artery catheter was
introduced 1970.
• Inflation of small balloon on tip of catheter allowed for indirect assessment of left heart pressures and function.
• Ultimately came to measure not just pulmonary artery pressure but CVP too
Central Venous
• Term “central venous” generally considered to be within the thoracic cavity
• CV pressure often used interchangeably with right atrial pressure (RAP) since the difference between the two is generally
insignificant clinically
• Indications:
• Fluid resuscitation
• Hemodynamic monitoring
• Medication administration
• Inability to access peripheral veins
• Dialysis
• Plasmapheresis
• Transvenous pacemaker placement
• Contraindications:
• Infection
• Trauma
• Venous thrombosis at the selected site
• A multilumen catheter
• Can have 2, 3, or 4 ports; most commonly 3 (called a “triple lumen” catheter)
• Allows for simultaneous medication/fluid infusions as well as one port to be used for continuous CVP monitoring
• Each lumen must be flushed and capped prior to insertion.
• Use distal port to guide catheter into blood vessel.
Introduction
Here we discuss insertion points for a central venous catheter.
Accessing a Vein
• Done by performing one or the other of these:
• Cutdown
• An incision directly over the target vein
• Used only if/when the vein is undetectable
• Percutaneous insertion
• Into the vessel, introducing a needle attached to a syringe
• Blood is aspirated into the syringe.
• Can use either of two access systems or insertion techniques: -
• Seldinger technique
• Catheter-Over-Needle technique
Introduction
Here we look at the management of the central venous catheter once it has been inserted.
Catheter Removal
• Proper technique crucial since during this procedure, patient at a very high risk for developing air embolus.
• Patient should remain in bed for at least 30 minutes following the removal of a central venous catheter from the internal
jugular vein or the subclavian vein.
• Removal of a femoral vein catheter involves a different procedure and requires placing the patient at bed rest for a longer
period time.
• Steps of removal
Introduction
Here we look at how to analyze the waveform and obtain values.
CVP/RAP Waveform
• Components of a CVP/RAP waveform:
• Waveform • V Wave
• A Wave • X Descent
• C Wave • Y Descent
Introduction
Here we explore some of the complications associated with insertion of central venous catheters and with central venous
pressure monitoring.
Bleeding
• Is usually occult (not visible) Instead blood is in either the tissue or a body cavity.
• The reason that a chest x-ray following insertion is so important.
Vascular Erosion
• Rare but can occur any time after insertion of the catheter
• Patients at increased risk if hypertonic solutions being infused or if endothelial lining is friable.
• More common with Internal Jugular (IJ) or subclavian insertions
Arrhythmias
• Occur when catheter tip is introduced into the right atrium or right ventricle.
• A central venous catheter inserted into subclavian vein or IJ may spontaneously advance into the right ventricle. The
arrhythmia occurs from the tip of the catheter touching the endocardial wall.
Nosocomial Infection
• All central venous catheters increase patient’s risk of developing a nosocomial infection and severe sepsis.
• Possible reasons for infection:
• Poor technique during catheter insertion
• Introduction/migration of the skin flora into the insertion site
• Contaminated tubing and pressure transducer/flush system.
• Use strict sterile technique when changing dressings, replacing tubing and flush solutions, or accessing the line for a blood
sample.
Thromboembolic Complications
• Occur with kink in catheter or when fluid flow through system is poor
• A blood clot can form at tip of the catheter. This blood clot could become dislodged during catheter repositioning or with
flushing.
• Most thrombi clinically silent
• Risk Factors for thromboembolic complications
• Hypercoagulability associated with:
• Myocardial infarction • Antithrombin
• Fever • Cancer
• Polycythemia • Estrogen therapy
• Vessel wall trauma (during insertion)
Introduction
Here we explore the taking of pressure measurements via different methods-pressure transducer and water manometer.
Water Manometer
• Before the advent of electronic pressure monitoring systems, water manometer systems were used to take pressure
measurements.
• Today, rarely used in critical care but sometimes used outside ICU for patients experiencing fluid shifts.
• Measurements are taken in centimeters of water pressure (cm H20).
Page 1 of 3
Practice Alerts
The goal of the practice alerts is to help nurses and other health care practitioners carry their bold
voices to the bedside to directly impact patient care. Practice alerts are succinct, dynamic directives
from AACN that are supported by authoritative evidence to ensure excellence in practice and a safe and
humane work environment. Practice Alerts are short directives designed for easy reference.
Recognizing that practice is dynamic, the Practice Alerts will be reviewed and updated as needed, and
can be accessed by going to: http://www.aacn.org/aacn/practicealert.nsf/vwdoc/pa2
Following is the current Practice Alert for Preventing Catheter Related Bloodstream Infection, but be
sure to check the link for the most up-to-date information.
41
Page 2 of 3
Expected Practice:
; Cleanse hands with waterless cleaning solution or, if visibly soiled, with soap and water before
and after patient contact.
; Disinfect clean skin utilizing friction with an appropriate antiseptic (preferably 2% chlorhexadine)
before catheter insertion and during site care.
; Utilize full barrier precautions when inserting central venous access devices.
; Educate all staff inserting and caring for intravascular catheters, assess competency of same at
regular intervals, advocate adherence to standards of care.
; Replace peripheral IV sites in the adult patient population at least every 96 hours but no more
frequently than every 72 hours. Leave peripheral venous catheters in children until IV therapy
is completed, unless complications (e.g., phlebitis and infiltration) occur.
; Replace IV tubing at least every 96 hours but no more frequently than every 72 hours.
; When adherence to aseptic technique during intravascular catheter insertion cannot be
ensured (i.e. prehospital, code situation), replace the catheter soon as possible, but within 48
hours.
Supporting Evidence:
A substantial proportion of hospital-acquired infections result from cross-contamination from the
hands of healthcare workers. Alcohol-based hand rub, compared with traditional handwashing
with unmedicated soap and water or medicated hand antiseptic agents, may offer better results
because it requires less time, acts faster, and is less likely to irritate skin. Thus, the CDC
recommends the use of alcohol-based hand rubs between patient contacts as an adjunct to
traditional handwashing alone.
Chlorhexidine gluconate solutions utilized for vascular catheter site care reduce catheter related
bloodstream infections and catheter colonization more effectively than povidone-iodine solutions.
Moreover, 80% of resident and transient skin flora are found in the first five epidermal layers of
the skin. There is clinical evidence to support the efficacy of applying antiseptics with sufficient
friction to assure that the solution reaches into the cracks and fissures of the skin. There is no
evidence that supports use of traditional concentric prepping technique. Although a 2%
chlorhexidine-based preparation is preferred, tincture of iodine, an iodophor, or 70% alcohol can
be used. Allow any solution used to dry before the catheter is inserted.
Compared with peripheral venous catheters, CVCs carry a substantially greater risk for infection;
therefore, the level of barrier precautions needed to prevent infection during insertion of CVCs
should be more stringent. Maximal sterile barrier precautions (e.g., cap, mask, sterile gown,
sterile gloves, and full body sterile drapes) during the insertion of CVCs substantially reduce the
incidence of CRBSI compared with standard precautions (e.g., sterile gloves and small
drapes)1,3,2,10 There are some studies that demonstrate infection rates are lower when the
42
Page 3 of 3
subclavian site is used. Selection of central line insertion site, however, is based on patient risk
factors.
Healthcare workers who insert and care for intravascular devices should receive formalized
education and training in indications for intravascular catheterization, proper placement,
maintenance, and infection control. Educational programs focusing on central venous catheter
insertion and care have lead to a substantial decrease in cost, morbidity, and mortality
attributable to central venous catheterization. Ongoing education and reinforcement of
appropriate technique serve as a reminder of current best practices, and studies demonstrate
that consistent reinforcement of aseptic technique leads to decreased CRBSI.
Studies of peripheral intravenous catheters show that there is not a substantial difference in
phleblitis rates between catheters left in place 72 hours and those left in place 96 hours. There is
no evidence to support that routine replacement of central venous catheters is more effective in
decreasing blood stream infections than replacing central venous catheters as needed.
Studies show that IV tubing containing crystalloids can be replaced every 72 – 96 hours. If
monitoring using a transducer system, replace the transducer, tubing, flush device and flush
solution every 96 hours.
Page 1 of 1
Practice Pearls
Bleeding
Always assess the abdominal girth as well as the softness of the abdomen. If
retroperitoneal bleeding is present, the abdomen will become very taught and the girth
will increase. Patients often will complain of severe back pain.
Nosocomial Infection
Most facilities have a central line “bundle” which consists of mandatory steps/processes
to be used during insertion of a central line. The purpose is to prevent a central line
infection. Ask about your hospital’s “central line bundle.”
Thromboembolic Complications
Avoid forceful flushing with a syringe which may cause the thrombus to become
dislodged.
Lesson 4
Pulmonary Artery Catheters
• Catheter Types
• Insertion
• Waveform Analysis
• Clinical Applications
51
Introduction
In this lesson we focus on the pulmonary artery catheter (PA catheter) and discuss types, insertion responsibilities, and
indications for use. For patients with a PA catheter, interpreting waveforms and documenting values, troubleshooting the
system, and assessing for complications are priorities.
• Fluid
• Perioperative
• Use in preventing/treating Shock
Introduction
Here we explore different types of PA catheters.
Introduction
Here we look at insertion methods for a PA catheter.
Relative Contraindications
• No known absolute ones but relative ones are:
• Presence of fever, mechanical tricuspid valve, and anticoagulated state
• Patients with Wolff, Parkinson, White syndrome, and Ebstein malformation
• Have risk for tachydysrhythmias
• Closely monitor the ECG in patients with left bundle branch block.
• Have increased risk of complete heart block
• In case that heart block develops, emergent pacemaker insertion may be necessary.
Equipment Preparation
• Nurse responsibility to prepare equipment including the pressure tubing system, transducer, and monitoring system
• To ensure accuracy, setting the scales for pressure tracing is important.
• Procedure for setting the scales
PA Catheter Insertion
• Gather and prepare supplies and equipment (steps):
• Infection prevention
• Gather supplies
• Connect
• Calibrate fiberoptics
• Check balloon
• Also vital to determine patient’s baseline cardiovascular, peripheral vascular and neurovascular status.
• Advance Catheter
Introduction
Here we look at ways to get accurate readings, manage the PA catheter, and troubleshoot it when it’s not accurate or
functioning properly.
Patient Positioning
• Helps get accurate PA pressure measurements
• Position patient in supine position with the head of bed elevated from 0 to 60 degrees.
• Ensure air/fluid interface is level with phlebostatic axis.
• Reposition transducer.
• When patient in a side-lying or prone position, phlebostatic axis is not at the fourth ICS, midclavicular.
• For patient in right lateral position, reference point is intersection of the fourth intercostal space and the
midsternum.
• For patient in left lateral position, reference point is intersection of the fourth intercostal space and the left
parasternal border.
Troubleshooting
• Important for maintaining catheter patency, ensuring that data from PA catheter is accurate, and preventing occurrence of
catheter-related and patient-related complications.
• Possible reasons why waveform may not be displaying on the monitor:
• Catheter has kink.
• Stopcock not correctly positioned.
• An incorrect scale size was chosen.
• Flush bag is empty or there is less than 300 mmHg of pressure on the pressure bag.
• Catheter is clotted.
• Cable may be fractured and need replacing. (rare)
Other Issues
• Other possible reasons for an inaccurate or absent PA reading on the monitor:
• Artifact • Blood back-up in line
• Inability to flush catheter • Ruptured balloon
• Continuous wedge waveform (a medical emergency)
Introduction
Here we explore the process of interpreting waveforms.
Introduction
Here we explore some of the complications that can arise with PA catheters either during insertion or during maintenance.
Introduction
Here we look at how monitoring the waveforms not only helps ensure that the catheter is still properly placed, but also aid in
the diagnosis of cardiac and pulmonary diseases. We look at the implication of both PA pressures and elevations in certain
wave types can mean to for diagnosing a patient.
PAD vs PAOP
• For blood flow through lungs to occur, mean PA pressure must always be higher than left atrial pressure.
• PAD pressure should be higher than left atrial pressure (PAOP) and if it isn’t, it usually means either a very low pulmonary
blood flow state or the waveform has been misinterpreted.
• Reduce possible pulmonary complications by using the PAD to depict PAOP.
• Conditions that make PAD not equal to PAOP:
• Pulmonary diseases that increase the PVR • Rapid heart rates that decrease filling time causing
• Mitral valve diseases PAD to elevate with no change in PAOP
As the critical care nurse, you will need to gather all the supplies and equipment for the
insertion of the PA catheter. Below is a sample supply checklist.
Fluoroscope
Emergency equipment
Temporary pacing equipment
Indelible marker
Transducer holder
62
Page 1 of 5
Practice Alerts
The goal of the practice alerts is to help nurses and other health care practitioners carry their bold
voices to the bedside to directly impact patient care. Practice alerts are succinct, dynamic directives
from AACN that are supported by authoritative evidence to ensure excellence in practice and a safe and
humane work environment. Practice Alerts are short directives designed for easy reference.
Recognizing that practice is dynamic, the Practice Alerts will be reviewed and updated as needed, and
can be accessed by going to: http://www.aacn.org/aacn/practicealert.nsf/vwdoc/pa2
Following is the current Practice Alert for Pulmonary Artery Pressure Measurement, but be sure to
check the link for the most up-to-date information.
63
Page 2 of 5
Expected Practice:
; Verify the accuracy of the transducer-patient interface by performing a square waveform test at
the beginning of each shift.
; Position the patient supine prior to PAP/RAP (CVP) measurements. Head of the bed (HOB)
elevation can be at any angle from 0° (flat) to 60°.
; Level the transducer air-fluid interface to the phlebostatic axis (4th ICS/½ AP diameter of the
chest) with the patient in a supine position prior to PAP/RAP measurements.
; Obtain PAP/RAP measurements from a graphic (analog) tracing at end-expiration.
; Use a simultaneous ECG tracing to assist with proper PAP/RAP waveform identification.
; PA catheters can be safely withdrawn and removed by competent registered nurses.
Supporting Evidence:
The square waveform test, or dynamic response test, determines the ability of the transducer
system to correctly reflect pressures in the pulmonary artery.1-5 This test can identify system
problems, such as air bubbles in the tubing, excessive tubing length, loose fitting equipment,
and/or patient problems, such as catheter patency. Any of these problems may affect accuracy of
PAP/RAP measurements and should be corrected prior to pressure measurement. Experts
recommend the following situations as appropriate to perform the square waveform test: on the
initial system setup, at least once each shift, after opening the catheter system (e.g. for rezeroing,
drawing blood, or changing tubing), and whenever the PAP waveform appears to be damped or
distorted. 1-4,6
Consider the following changes in PA pressures as clinically significant (i.e., not reflective of the
normal variability in PA pressures): UPAS > 4-7 mm Hg; UPAEDP > 4-7 mm Hg; UPAWP > 4
mm HG.7,8
Studies in a variety of patient populations have found that PAP/RAP measurements are accurate
when the head of the bed is elevated to any angle between 0o and 60o, as long as the patient is in
the supine position.9-11 Two studies have also shown that PAP/RAP readings are accurate with
the patient in a lateral position if the angle of rotation is exactly 30° or 90° with the head of the bed
flat, and the location of the transducer air-fluid interface changed to the appropriate external
landmarks for lateral positioning (30o lateral: ½ distance from surface of bed to the left sternal
border; 90o right lateral: 4th ICS/midsternum; 90o left parasternal border).1,12-14 When utilizing a
30o side lying angle a method of ensuring an accurate angle is needed and should be readily
available to the bedside practitioner.13
Leveling the transducer air-fluid interface to the left atrium corrects for changes in hydrostatic
pressure in vessels above and below the heart.1,3 In the supine position, the external landmark for
the left atrium is the phlebostatic axis (4th ICS/½ AP diameter of the chest).15,16 Studies have
found that improper positioning of the air-fluid interface can lead to significantly different PAP/RAP
reading.17,18 Once the correct location for the phlebostatic axis is identified, a mark should be
64
Page 3 of 5
placed on the chest wall and a laser pointer level or a carpenter’s level should be used to properly
level the transducer air-fluid interface anytime the patient is repositioned.1,2.,18 Changes in patient
position, even slight HOB m or n, require releveling of the transducer air-fluid interface before
obtaining PAP/RAP measurements.
Changes in intrathoracic pressure during respiration significantly alter hemodynamic pressures.
Obtaining accurate PAP/RAP measurements requires reading pressure waveforms during end
expiration only.1-4,9,19 Digital readouts on pressure monitors reflect pressures obtained throughout
respiration and are significantly different from end expiratory pressures, requiring pressure to be
read from graphic waveform tracings. 6,20,21
Levels of evidence supporting validation of PAP/RAP waveform measurement with simultaneous
ECG tracings include clinical literature, expert opinion and sound theoretical principles of
hemodynamic measurement. 1-4
Studies and surveys show that after education and clinical monitoring to assess competency,
registered nurses can safely withdraw and/or remove PA catheters.22,23,24 Before incorporating
withdrawing and/or removing PA catheters into nursing practice, verify that it is within your state’s
scope of practice for registered nurses.
Page 4 of 5
REFERENCES:
1. Keckeisen M, Chulay M, Gawlinski A, eds. Pulmonary and artery pressure monitoring. In
AACN’s Protocols for Practice: Hemodynamic Monitoring Series. Aliso Viejo, Calif: AACN;
1998.
2. Quaal S. Quality assurance in hemodynamic monitoring. AACN Clin Issues. 1993;4:197-206.
3. Daily E, Schroeder J. Techniques in Bedside Hemodynamic Monitoring. St Louis, Mo: Mosby-
Year Book; 1994.
4. Quaal S. Ask the experts. Crit Care Nurse. 1995;10:92-93.
5. Gardner R. Direct blood pressure measurement: Dynamic response requirements.
Anesthesiology. 1981;54:227-236.
6. Ahrens T, Penick J, Tucker M. Frequency requirements for zeroing transducers in
hemodynamic monitoring. Am J Crit Care. 1995;4:466-471.
7. Moser D, Woo M. Normal fluctuation in pulmonary artery pressure and cardiac output in
patients with severe left ventricular dysfunction abstract. Am J Crit Care. 1996;5:236.
8. Nemens EJ, Woods SL. Normal fluctuations in pulmonary artery and pulmonary capillary
wedge pressure in acutely ill patients. Heart Lung. 1982;11:393-398.
9. Dobbin K, Wallace S, Ahlberg J, et al. Pulmonary artery pressure measurement in patients
with elevated pressures: effect of backrest elevation and method of measurement. Am J Crit
Care. 1992;1:61-69.
10. Wilson A, Bermingham-Mitchell K, Wells N, et al. Effect of back position on hemodynamic and
right ventricular measurements in critically ill adults. Am J Crit Care. 1996;5:264-270.
11. Woods S, Mansfield L. Effect of body position upon pulmonary artery and pulmonary capillary
wedge pressures in noncritically ill patients. Heart Lung. 1976;5:83-90.
12. VanEtta D, Gibbons E, Woods S. Estimation of left atrial location in supine and 30° lateral
position abstract. Am J Crit Care. 1993;2:264.
13. Bridges EJ, Woods SL, Brengelmann GL, et al. Effect of the 30 degree lateral recumbent
position on pulmonary artery and pulmonary artery wedge pressures in critically ill adult
cardiac surgery patients. Am J Crit Care. 2000;9:262-275.
14. Kennedy GT, Bryant A, Crawford MK. The effects of lateral body positioning on
measurements of pulmonary artery and pulmonary wedge pressures. Heart Lung.
1984;13:155-158.
15. Paolella L, Dortman G, Cronan J, et al. Topographic location of the left atrium by computed
tomography: reducing pulmonary artery catheter calibration errors. Crit Care Med.
1988;16:1154-1156.
16. Courtois M, Fattal P, Kovacs S, Tiefenbrunn A, Ludbrook P. Anatomically and physiologically
based reference level for measurement of intracardiac pressures. Circulation. 1995;92:1994-
2000.
17. Kee L, Simonson J, Stotts N, Skov P, Schiller N. Echocardiographic determination of valid
zero reference levels in supine and lateral positions. Am J Crit Care. 1993;2:72-80.
18. Bartz B, Maroun C, Underhill S. Differences in midanteroposterior level and midaxillary level of
patients with a range of chest configurations. Heart Lung. 1988;17:309.
19. Ahrens T. The effects of mechanical ventilation on hemodynamic waveforms. Crit Care Clin
North Am. 1991;3:629-639.
20. Ahrens T, Schallom L. Comparison of pulmonary atery and central venous pressure waveform
measurements via digital and graphic measurement methods. Heart Lung. 2001;30:26-38.
21. Lipp-Ziff E, Kawanishi D. A technique for improving the accuracy of the pulmonary artery
diastolic pressure as an estimate of left ventricular end-diastolic pressure. Heart Lung.
1991;20:107-115.
22. Wadas TM. Pulmonary artery catheter removal. Crit Care Nurse. 1994;14;62-72.
66
Page 5 of 5
23. Roundtree WD. Removal of pulmonary artery catheters by registered nurses: a study in safety
and complications.
Focu s Crit Care. 1999;18:313-318.
24. Zevola, DR, Maier B. Improving the care of cardiothoracic surgery patients through advanced
nursing skills. Crit Care Nurse. 1999;19:34-44.
67
Practice Pearls
Relative Contraindications
When the decision to insert a PA catheter has been made, the physician will need to
speak to the patient and family regarding the reason for insertion, overview of the
procedure and possible complications. A signed informed consent is the ideal. In most
instances the insertion is emergent and the preparation of the patient and family may be
inadequate. It is important to take the time after the patient has stabilized to discuss the
information with the family.
68
• Methods Of Calculating CO
• Clinical Application Of CO
• Non-Invasive CO Monitoring
71
Page of 9
Cardiac Output
• Cardiac Output (CO) one the two most widely utilized pieces of information from the PA catheter
• Understanding CO critical to assessing if cardiac function adequate
• CO is the amount of blood ejected by the ventricle each minute.
• Equals HR x SV
• Is measured in L/min.
• Normal CO: 4 - 8 L/min.
Cardiac Index
• Cardiac Index (CI) is a better measure of cardiac function than CO and is the value that should be used to trend cardiac
function.
• CI is CO adjusted to individual body size
• You divide CO by the individual’s BSA (which you get from the Dubois’ Surface Area chart on the cardiac monitor.)
• CI is displayed on most monitors either continuously or at the time that CO is obtained.
• This value only accurate if person’s height and weight were entered into the database to ensure accuracy of this
value.
• Normal value for CI: 2.5 -4.3 L/min/m2
Stroke Volume
• Stroke Volume (SV) one of the two most widely utilized pieces of information from the PA catheter
• Understanding SV critical to assessing if cardiac function adequate
• Afterload
• Vascular resistance; the resistance to ventricular emptying during systole
• The pressure/resistance that the ventricles must overcome in order for them to be able to open the aortic and
pulmonic valves and pump blood into the systemic and pulmonary vasculature.
• Contractility
• The strength of the myocardial contraction • The degree of myocardial fiber shortening
Preload
• Clinical indicators for right and left ventricular preload • Reasons for reduced preload
• Influences on preload • Reasons for increased preload
Afterload
• Influences on afterload • Reasons for reduced afterload • Reasons for increased afterload
Calculating Afterload
• Afterload cannot be measured directly; it must be calculated using values from the PA catheter.
Contractility
• Affected by many factors including:
• Preload and afterload • Functioning of myocardium
• Vasodilators, vasoconstrictors, and intropes • Myocardial oxygenation
• Electrolyte levels
Thermodilution Method
• At bedside, CO measurements obtained through PA catheter via one of these methods:
• Intermittent Bolus Thermodilution (TDCO)
• Continuous CO (C CO)
• Steps:
• TDCO injection
• Thermistor detects temperature change
• CO curve interpreted
TDCO Technique
• Xx
• A closed injectate delivery system is most often used for reasons of reducing multiple entries into the system and
controlling infections.
• Iced injectate verses room temperature injectate
• A 10+ degree gradient between temperature of injectate and of patient’s blood
• Cardiac output computer
Proper Technique
• Use of proper technique and observing of CO curve will bolster accuracy.
• End-expiration • Rapid injection • Interpretation • Abnormal curves
• Catheters with continuous CCO capability have a 10 cm thermal filament that emits a pulsed, low heat energy signal in a
30 to 60 second sequence.
• Proper placement: The thermal filament section of catheter is located in RV
• Bedside computer constructs thermodilution curves and measures CO automatically.
Hemodynamic Instability
• Presents as either a high blood flow state or a low blood flow state
• Body’s efforts to compensate work only for a while.
• Low CO states occur due to hypovolemia or left ventricular dysfunction.
• Early warning of decreasing CO is a drop in the SV/SVI. CO will still appear normal even if this is dropped, so
paying attention to this can help you anticipate and ward off a low CO or low-flow state.
Low-flow State
• To treat a low CO state you must first determine whether it’s hypovolemia or a problem of LV dysfunction. This is done by a
combination of clinical and hemodynamic assessments:
• Patient’s physical assessment
• Patient’s history
• Use of intracardiac pressures such as RAP, PAP and PAOP
Elevated CO Values
• In healthy people, CO values elevate in response to increased oxygen demand.
• In critically ill patients, an elevated CO is always an indicator of some other problem.
• Elevated CO values can mean systemic inflammation, anaphylaxis, or neurogenic-mediated vasodilation, all of
which result in a decreased SVR (afterload.)
• The increase in CO might be minimal or marked. The key to remember is that the CO elevation is a sign of a problem
rather than the problem.
• Contraindications:
• Coarctation of the aorta • Esophageal pathology • Coagulopathy • IABP
Thoracic Bioimpedance
• Determines SV using a current that flows from an outer electrode to an inner sensor
• Thoracic bioimpedance: the relationship between impedance and SV
• The resistance of current flow (impedance) across the chest is inversely related to the thoracic fluid.
• A useful method for trend analysis, but not accurate enough for diagnostic interpretation
• Major application has been outside the critical care setting.
• Management of acutely ill patients in the outpatient setting may be the most important contribution of this technology.
Gastric Tonometry
• Provides CO information based on adequacy of gastric mucosal perfusion.
• Gastric very sensitive to altered mucosal perfusion.
• Carbon dioxide partial pressure PCO2 measured in the stomach or intestine.
Practice Pearls
Thermodilution Method
The specific gravity of D5W is used in the formula to derive CO by the TDCO method.
The use of saline can result in a 2% decrease in the TDCO measurement. Saline
should be substituted if the patient’s medical condition warrants it.
TDCO Technique
Iced injectate may be preferable to room temperature injectate in patients with poor
forward blood flow, such as those with right ventricular failure, tricuspid stenosis or
tricuspid insufficiency. These conditions make it difficult to get accurate cardiac output
measurements, but the larger signal-to-noise ratio obtained with iced injectate may
improve the estimates obtained with the thermodilution method. Hypothermic patients
(cardiac surgery, trauma) may also require iced injectate to achieve greater accuracy
and precision of measurement.
The computation constant for the size and type of catheter used can be found in the
package insert of the PA catheter.
Lesson 6
Oxygenation And Transport
• Clinical Applications
84
- Describe the use of SVO2 and ScVO2 monitoring in the critically ill
patient.
85
Page of 7
Oxygen Delivery
• Goal of hemodynamic monitoring: Evaluate whether oxygen delivery to tissues is enough to meet their metabolic demands.
If oxygen delivery not adequate, we must determine steps for enhancing this delivery.
• Adequate Oxygen Delivery
• Inadequate Oxygen Delivery
Oxygen Demand
• Oxygen demand: Amount of oxygen cells require to meet their metabolic process
• Oxygen consumption: Amount of oxygen cells actually use
• More recently, central venous oxygen saturation has been reintroduced as a less invasive alternative to mixed
venous oxygen saturation.
Changes in SvO2
• Can be due to changes in oxygen delivery or to changes in oxygen consumption
Fiberoptic Technology
• Used in both fiberoptic pulmonary artery catheter and fiberoptic central venous triple lumen catheter
• How the technology works:
• Fiberoptic network travels length of catheter.
• An optical module transmits light down length of the catheter. Hemoglobin absorbs a certain amount of light
relative to oxygen saturation.
• The light is then reflected back to optical module where it is converted into an electrical signal and transmitted to
the monitor.
Insertion Guidelines
• Insertion guidelines for SvO2 and ScvO2 monitoring similar.
• In Vitro calibration • Signal strength • Transport
• In Vivo calibration considerations
DO2/VO2 Imbalances
• Oxygen delivery can fall with any of these occurrences:
• Decreased hemoglobin • Leftward shift of • Loss of autoregulation
• Decreased CO Oxyhemoglobin Saturation
• Decreased SaO2 Curve
•Increased SvO2
• Indicates that less oxygen is being extracted
Introduction
Here we look at the impact of the components of oxygen delivery.
Normal Patient
• How to calculate oxygen delivery for a patient:
• Given:
- CO = 5.0 l/min - SaO2 = 98% - Hgb = 15 g/dl - SvO2 = 75%
• Note, this is a normal patient since the components of oxygen delivery are within normal limits since
• All of the “Given” values are within normal range
• All of the calculated values are within normal range:
VO2 is 238 ml O2/min DO2 is 1014 mL O2/min O2 Extraction is 23%
General ICU factors that increase oxygen demand and the percent of increase include:
1. Dressing change 10%
2. Bed bath 23%
3. Each position change 31%
4. Increased work of breathing 40%
5. Shivering 50-100%
6. Visitor 22%
7. Chest X-ray 25%
8. ET suctioning 27-70%
9. Getting out of bed 39%
10. Weight on a sling scale 36%
11. For every 1 degree centigrade the body temperature is elevated, there is a 10-13%
increase in demand
12. Inotropic infusions 6-20%
94
Calculate the total percent increase in oxygen demand for this patient:
Consider that you are caring for an abdominal surgery patient. He is intubated and on a
ventilator. You are preparing to bathe the patient at 0400. The following sequence of
events occurs.
1. Bath is given and the patient is turned twice during the bath.
2. Three sets of dressings are changed.
3. Bed linens are changed and the patient is turned three times during the process.
4. The patient required suctioning after being turned and repositioned.
5. The patient’s body temperature is 38 degrees centigrade.
6. At 5:30 am radiology is at the bedside to obtain the portable upright chest x-ray. In
the process of positioning the plate behind the patient’s back, his position is
changed a total of three times. Upon completion of the x-ray, the plate is removed
and the patient is repositioned.
Event Percent Change
Bed Bath 23%
Position Change (x2) 31%
Dressing Change (3 sets) 10%
ET Suctioning 27-70%
Body Temp. For every 1 degree centigrade the body temperature is
elevated, there is a 10-13% increase in demand.
Chest X-Ray 25%
Getting Out of Bed 39%
CO = 5.0 l/min
SaO2 = 98%
Hgb = 15 g/dl
SvO2 = 75%
Oxygen Consumption
VO2 = CO x Hgb x 13.8 (SaO2 - SvO2)
5 x 15 x 13.8 x (.98-.75) = 238 ml O2/min
Oxygen Delivery
DaO2 = CO x Hgb x 13.8 x SaO2
45 x 15 x 13.8 x .98 = 1014ml O2/min
O2 Extraction
VO2 / DaO2 = % of O2 extracted
238 / 1014 = 23%
96
Practice Pearls
Oxygen Demand
Measuring oxygen consumption in the critically ill patient may be difficult unless the
patient is on a ventilator. In this case a “metabolic assessment" cart is used. In clinical
practice, we can use indirect methods to assess the amount of oxygen that is extracted
at the cellular level. We assume that oxygen extracted is actually consumed. Therefore,
we often use the terms interchangeably.
Oxygen Delivery
It is important to remember the formula for cardiac output as well as the determinants of
stroke volume.
CO – Cardiac Output
CO – HR x SV
CaOa – Arterial Oxygen Content
CaO2 – (Hgb x 1.38 x SaO2) + (PaO2 x 0.0031)
98
Insertion Guidelines
If you do not have time to perform the “in vitro” calibration prior to insertion, you can
perform a calibration once the catheter is inserted. The catheter will still provide
necessary information to care for the patient.
The optical signal may be impaired if the tip of the catheter is against the wall of the vein
or artery, or if a small piece of fibrin covers the tip of the catheter. Kinking of the
catheter can interrupt the light emission as well.
99
Practice Exercise
If it is greater than 30%, the metabolic demands of the tissues are not being met
indicating the tissues are trying to maximize oxygen extraction. On the other hand, if the
O2ER is less than 22%, the metabolic demands are being met or the patient is unable to
extract oxygen as in the case of severe sepsis.
Lesson 7
Pharmacological Management Of
Hemodynamics
Included in this Lesson:
• Managing Preload
• Managing Afterload
• Managing Contractility
102
Preload
• The force that stretches ventricles during diastole • Is measured on the right side by RAP/CVP and on the
• The amount of blood that fills ventricles during diastole left by the PAOP.
Increasing Preload
• Low PAOP (PAOP <8 mmHg) may be due to:
• Dehydration • Excessive diuresis • Third-spacing • Bleeding
• Preload increased by increasing circulating blood volume.
• Done by administering crystalloid or colloid
Crystalloids
• Crystalloids commonly used to enhance preload:
• Normal saline (NS) • Lactated ringers (LR)
• Hypotonic solutions (D5W, 0.45 NS) are not for use in rapid fluid resuscitation.
• Possible contraindications of LR
• Possible complications of normal saline in large volumes
• Monitor electrolytes closely in patients receiving aggressive fluid resuscitation.
Pulmonary Dysfunction
• A fluid challenge in patients with severe pulmonary dysfunction often results in increased CO and left ventricular stroke
work index without a significant increase in PAOP
• Indicates that further fluid administration may be beneficial.
• A PAOP of 15 -18 mmHg generally indicates adequate preload and warrants consideration for an afterload reduction with a
vasodilator (for purposes of improving SVI and CI.)
Preload Reducers
• Diuretics (Furosemide) • Nitroglycerin • Morphine sulfate
• Milrinone (a phosphodiesterase inhibitor, Type III) • Nitroprusside (Nipride)
Afterload
• The resistance ventricle must overcome to eject blood in the forward direction.
• Two determinants of afterload:
• Volume and mass of blood ejected from ventricle • Compliance of vascular space into which the blood is ejected.
• Right ventricular afterload measured by PVR; left ventricular afterload measured by SVR.
• Pulmonary hypertension reflects in an elevated PVR.
• Cardiac diseases that may cause elevated PVR.
• Pulmonary diseases that may cause an elevated PVR.
Reducing PVR
• Developing an effective therapy for pulmonary hypertension is difficult since pulmonary vasculature reacts less to neural
and pharmacologic stimulation than systemic vasculature does.
• Ways to manage pulmonary hypertension (lower PVR):
• First ensure adequate oxygenation
• If PVR not improving and patient shows signs of right-sided failure, a more aggressive therapy is warranted.
SVR Indications
• High SVR (>1200 dynes/sec/cm5) may reflect physiologic stress or left ventricular failure
• Medications that decrease SVR: nitroprusside, hydralazine, captopril, and to a lesser extent nitroglycerin
• Low SVR (<800 dynes/sec/cm5) may reflect septic shock or excessive administration of afterload reduction medications.
• Medications that increase SVR: dopamine, epinephrine, norepinephrine, phenylephrine, and vasopressin
Increasing SVR
• Drugs most commonly used to increase SVR:
• Dopamine • Norepinephrine • Vasopressin
• Epinephrine • Phenylephrine
Contractility
• The ability to shorten and develop tension within the myocardial cells
• Not possible to directly measure; instead use these indirect measures of contractility:
• SVI • RVSWI • LVSWI
• Patients with poor contractility may also have decreased CI, elevated filling pressures (RAP and PAOP,) and a low SvO2.
• Blood pressure not a reliable indicator of low contractility.
• Conditions generally present with low contractility:
• Myocardial ischemia or infarction • Cardiomyopathy
• Congestive heart failure • Early stages of septic shock
IV Drip Rate:
Drug Concentration:
Amount of drug in solution (g, mg, mcg)
Amount of solution (mL)
Example: Calculate the intravenous infusion rate in mL/h for a 70-kg patient
requiring dobutamine, 5 mcg/kg /min, using a dobutamine drip of
500 mg in 250 mL D5W.
Example: Calculate the IV infusion rate in mL/h for a 70-kg patient requiring a
procainamide, maintenance drip at 2 mg/min.
Maintenance Dose to be infused: 2 mg/min
Procainamide Concentration: 2000 mg/500 mL = 4 mg/mL
Calculation: 2mg/min x 60 min/h ÷ 4 mg/mL = 30 mL/h
Answer: Setting the infusion pump to 30 mL/h will deliver procainamide at 2 mg/min.
Calculation of mg/h
Examples: Theophylline, versed, diazepam, morphine, diltiazem
To calculate dose with infusion in progress:
_________mg/mL x _______mL/h = ______mg/h
concentration pump setting dosage
Example: Calculate the IV infusion rate in mL/h for a 70-kg patient requiring
morphine 5mg/h.
Concentration: 20 mg/100 mL = 0.2 mg/mL.
Calculation: 5 mg/h ÷ 0.2 mg/mL = 25 mL/h
Answer: Setting the infusion pump to 5 mL/h will result in final infustion rate
of 5 mg/h.
Conversion Factors:
1 mg = 1000mcg 1m = 100 cm
1 kg = 2.2 lb 1/150 gr = 0.4 mg
1 gr = 60 mg 1 tsp = 5 ml
1/100 gr = 0.6 mg 1 Tbsp = 15 ml
1 inch = 2.5 cm 1 oz = 30 ml
1 cm = 10 mm 1 mm Hg = 1.36 cm H20
F C
89.6 32
91.4 33
93.2 34
95. 35
96.8 36
98.6 37
100.4 38
102.2 39
104.2 40
105.8 41
Practice Pearls
Increasing SVR
Epinephrine should not be used as the initial vasopressor for management of
hypotension in the septic patient. It will improve the BP, but decrease mesenteric blood
flow and increase the production of lactate.
Many vasopressor agents can cause an extravasation if there is an infiltration of the IV
catheter and the drug leaks into the surrounding tissue. There are several mechanisms
for treating an extravastation from phentolamine administration to application of
nitroglycerine paste. Check your pharmacy policies to determine the procedure for
managing and extravasation.
116
The Essentials of
Hemodynamic Monitoring Glossary
Air embolus Air entering the venous or arterial system inadvertently through an
arterial or venous line. (GE Healthcare - Medcyclopaedia, 2008)
Allen's test (for occlusion of ulnar or radial arteries) The patient makes a tight
fist so as to express the blood from the skin of the palm and fingers;
the examiner digitally compresses either the radial or the ulnar
artery. When the patient unclenches the fist, if blood fails to return
to the palm and fingers, there is indicated obstruction to blood flow
in the artery that has not been compressed. (Dorland's Illustrated
Medical Dictionary, 30th ed., 2003)
Arterial blood The pressure in the larger arteries, arteries being the blood vessels
pressure which take blood away from the heart (Wikipedia, 2008)
Cardiac output The volume of blood ejected from the left side of the heart in one
minute (Merriam Webster's Medical Dictionary, 2005)
118
Central Venous The venous pressure of the right atrium of the heart obtained by
Pressure (CVP) inserting a catheter into the median cubital vein and advancing it to
the right atrium through the superior vena cava (Merriam Webster's
Medical Dictionary, 2005)
Congestive heart Heart failure in which the heart is unable to maintain adequate
failure (CHF) circulation of blood in the tissues of the body or to pump out the
venous blood returned to it by the venous circulation (Merriam
Webster's Medical Dictionary, 2005)
Deadender cap A terminal cap that does not allow air into the line or blood to leak
out of the port.
119
Fast flush The pressure in the arterial line is rapidly increased to 300 mmHg by
flushing the system with the continuous flow mechanism and the
resulting waveform was analyzed to determine the response of the
system; ideally, one large and one small oscillation should occur,
after which the waveform should be returned to the baseline (Critical
Care. 2006;10(2):R43. ©2006 BioMed Central, Ltd.)
Hydrodynamic Related to a branch of physics that deals with the motion of fluids
and the forces acting on solid bodies immersed in fluids and in
motion relative to them (Merriam Webster's Medical Dictionary,
2005)
Hypercoaguable The state of being more readily coagualated than normal (Dorland's
state Illustrated Medical Dictionary, 30th ed., 2003)
Ischemic Deficient supply of blood to a body part (as the heart or brain) that
is due to obstruction of the inflow of arterial blood (as by the
narrowing of arteries by spasm or disease) (Merriam Webster's
Medical Dictionary, 2005)
Korotkoff sounds Arterial sounds heard through a stethoscope applied to the brachial
artery distal to the cuff of a sphygmomanometer that change with
varying cuff pressure and that are used to determine systolic and
diastolic blood pressure. Korotkoff, Nikolai Sergeievich (1874-
1920), Russian physician. Korotkoff introduced the auscultation
method of determining blood pressure in 1905. (Merriam Webster's
Medical Dictionary, 2005)
121
Low compliance Stiff tubing that is used in hemodynamic monitoring. The rigidity of
tubing the tubing will not “absorb the wave” of pressure and create
inaccurate readings. With stiff tubing, the “wave” travels all the way
to the transducer so that accurate values are obtained.
Mean arterial The average pressure within an artery over a complete cycle of one
pressure (MAP) heartbeat; in the brachial artery, MAP = diastolic pressure +
((Systolic Pressure - Diastolic Pressure)/ 3) (Dorland's Illustrated
Medical Dictionary, 30th ed., 2003)
Mixed venous The oxygen saturation of venous blood as it returns to the heart as
oxygen measured at the vena cava. SvO2 represents the difference between
saturation oxygen delivery and oxygen consumption. SvO2 = DO2 - VO2.
(SvO2)
Multi-lumen A type of central venous catheter that may have two, three or four
catheter ports. The most common configuration is the triple lumen (3
lumens) catheter. This catheter allows for simultaneous medication/
fluid infusions as well as one port to be used for continuous CVP
monitoring.
Oxygen delivery The amount of oxygen delivered to the tissues in one minutes. DO2
(DO2) = Hgb x 1.38 x SaO2 x CO x 10. Normal DO2 = 1000 mL/min.
Oxygen demand The amount of oxygen the cells require to meet their metabolic
process.
122
Phlebostatic axis An anatomic landmark located near the midaxillary line and fourth
intercostal space. (Wikipedia, 2008)
Pneumothorax A condition in which air or other gas is present in the pleural cavity
and which occurs spontaneously as a result of disease or injury of
lung tissue, rupture of air-filled pulmonary cysts, or puncture of the
chest wall or is induced as a therapeutic measure to collapse the
lung (Merriam Webster's Medical Dictionary, 2005)
Preload The stretched condition of the heart muscle at the end of diastole
just before contraction (Merriam Webster's Medical Dictionary,
2005)
Septic shock A life-threatening severe form of sepsis that usually results from the
presence of bacteria and their toxins in the bloodstream and is
characterized especially by persistent hypotension with reduced
blood flow to organs and tissues and often organ dysfunction
(Merriam Webster's Medical Dictionary, 2005)
Square wave Activation of the fast flush device on the transducer for 1 to 2
test seconds, which causes the monitor wave form peak out at maximum
pressure. The waveform will “bounce” back from the peaked level
and oscillate back and forth before resuming the readout of the
pressure waveform. The evaluation of this “bounce back” reflects
the dynamic response of the system. An optimally damped system
will drop straight to the bottom when the flush device is released, it
will then bounce no more than twice before returning to the
waveform. (aka: dynamic response testing)
124
Strip recorder A feature of the hemodynamic monitor that allows the user to
document the waveform on paper for further analysis
Stroke Volume The volume of blood pumped from a ventricle of the heart in one
beat (Merriam Webster's Medical Dictionary, 2005)
Stroke Volume The stroke volume per heartbeat corrected for body surface area;
Index usually expressed in mL per beat per square meter. (Dorland's
Illustrated Medical Dictionary, 30th ed., 2003)
Systole The contraction of the heart by which the blood is forced onward and
the circulation kept up (Merriam Webster's Medical Dictionary,
2005)
Transducer A device that is actuated by power from one system and supplies
power usually in another form to a second system (Merriam
Webster's Medical Dictionary, 2005)
Venous Oxygen The amount of oxygen remaining after the blood has passed through
Transport the tissues. This represents the oxygen reserve for the body.
Venous Oxygen Transport = Hgb x 1.38 x SvO2 x CO x 10. Normal
Venous Oxygen Transport = 750 mL/min.