Professional Documents
Culture Documents
Manual Técnico PENTRA 60 (2020 - 08 - 20 02 - 50 - 44 UTC)
Manual Técnico PENTRA 60 (2020 - 08 - 20 02 - 50 - 44 UTC)
Manual Técnico PENTRA 60 (2020 - 08 - 20 02 - 50 - 44 UTC)
P/n: RAA016CEN
Contents
1 - Introduction
1. Revision ..................................................................................................................2
5. Specifications ........................................................................................................8
Technical Manual
RAA045AEN 0-1
Contents
Technical Manual
0-2 RAA045AEN
Contents
2. Boards ....................................................................................................................4
4. Connections ...........................................................................................................9
5. Synoptic ...............................................................................................................25
Technical Manual
RAA045AEN 0-3
Contents
Technical Manual
0-4 RAA045AEN
Contents
5 - Software
1. Versions .................................................................................................................2
Technical Manual
RAA045AEN 0-5
Contents
1. Printer .....................................................................................................................2
2. Transmission .........................................................................................................2
3. Calibration .............................................................................................................2
5. Reagents ................................................................................................................3
6. Temperature ..........................................................................................................3
Technical Manual
0-6 RAA045AEN
Contents
7 - Maintenance
1. Maintenance ..........................................................................................................4
2. Procedures ............................................................................................................7
RAS235C: Installation
Technical Manual
RAA045AEN 0-7
Contents
8 - Exploded views
1. List ..........................................................................................................................2
2. Valves .....................................................................................................................6
14. Chambers for instrument NON equipped with Takasago valves .................18
Technical Manual
0-8 RAA045AEN
Contents
Technical Manual
RAA045AEN 0-9
Contents
Technical Manual
0 - 10 RAA045AEN
Introduction
1 - Introduction
1. Revision ..................................................................................................................2
5. Specifications ........................................................................................................8
5.1. Technical specifications ....................................................................................................... 8
5.2. Physical specifications ....................................................................................................... 11
Technical Manual
RAA016CEN 1-1
Introduction
Revision
1. Revision
This document applies to the latest software listed and higher versions.
When a subsequent software version changes the information in this document, a new electronic issue
(CD-ROM and/or online help) is released and supplied by HORIBA Medical.
To update a paper document, please contact your local HORIBA Medical representative.
Technical Manual
1-2 RAA016CEN
Introduction
Legal information
2. Legal information
This instrument responds to the Standards and Directives named in the Declaration of Conformity.
Latest version of the CE Declaration of conformity for this instrument is available on www.horiba-
abx.com
The Information in this manual is distributed on an "As Is" basis, without warranty. While every
precaution has been taken in the preparation of this manual, HORIBA Medical will not assume any
liability to any persons or entities with respect to loss or damage, caused or alleged to be caused directly
or indirectly by not following the instructions contained in this manual, or by using the computer software
and hardware products described herein in a manner inconsistent with our product labelling.
To alert the operator of potentially hazardous conditions, one of the headings which are described below
is provided wherever necessary throughout this text.
Emphasizes information that must be followed to avoid hazard to either the operator or the
environment, or both.
Emphasizes information that must be followed to avoid possible damage to the instrument
or erroneous test results.
Emphasizes information that can be helpful to the operator before, during or after a specific
operational function.
2.4. Graphics
All graphics including screens and printouts, photographs are for illustration purposes only and are not
contractual.
Technical Manual
RAA016CEN 1-3
Introduction
Legal information
2.5. Trademarks
Product names mentioned within this publication may be trademarks or registered trademarks of their
respective owners.
All rights reserved. No part of this book may be reproduced or transmitted in any form or by any means,
electronic, mechanical, photocopying, recording, or otherwise, without the prior written permission of
HORIBA Medical.
Technical Manual
1-4 RAA016CEN
Introduction
Operational conditions
3. Operational conditions
3.1. Environment
■ The operation of the Pentra 60 should be restricted to indoor location use only. Instrument is
operational at an altitude of maximum 3000 meters (9840 feet).
■ The Pentra 60 is designed for safety from voltages surges according to INSTALLATION CATEGORY II
and POLLUTION DEGREE 2 (IEC 61010-1).
■ Please contact your local HORIBA Medical representative for information regarding operation
locations, when it does not comply with the recommended specifications.
3.2. Location
The power switch and input voltage supply connection should always be accessible. When
positioning the system for operational use, leave the required amount of space for easy
accessibility to these items.
3.3. Grounding
Proper grounding is required when installing the system. Check the wall outlet ground (Earth) for proper
grounding to the facilities electrical ground. If you are unsure of the outlet grounding, contact your
facilities engineer to verify the proper outlet ground.
Technical Manual
RAA016CEN 1-5
Introduction
Operational conditions
■ The Pentra 60 must operate between temperatures of 16°C to 34°C (61°F to 93°F).
■ Maximum relative humidity 80% for temperature up to 31°C (88°F) without condensation.
■ If it is stored at a temperature less than 10°C (50°F), the instrument should stand for 1 hour at the
correct room temperature before use.
■ The Pentra 60 has been designed to produce less than the accepted level of electromagnetic
interference in order to operate in conformity with its destination, allowing the correct operation of
other instruments also in conformity with their destination.
■ In case of suspected electromagnetic noise, check that the instrument has not been placed in the
proximity of electromagnetic fields or short wave emissions, (i. e. Radar, X-rays, Scanners, Cell
phones, etc...).
■ Grounding is required. Check that the earth wall-plug is correctly connected to the laboratory
grounding system. If there is no such system a ground stake should be used.
■ Use only main supply cable delivered with the Pentra 60.
■ Main supply voltage fluctuations must not exceed +/-10% of the nominal voltage.
Technical Manual
1-6 RAA016CEN
Introduction
Environmental protection
4. Environmental protection
It should be disposed of, in accordance with local legislation, and should be treated as being
contaminated with blood. The appropriate biological precautions should be taken.
If any doubt, please contact your HORIBA Medical representative service department.
Condition for storage and transportation: Temperature from -20°C to +50°C (-4°F to 122°F).
Technical Manual
RAA016CEN 1-7
Introduction
Specifications
5. Specifications
5.1.2. Parameters
CBC mode
Parameters Descriptions
WBC White blood cells
RBC Red blood cells
HGB Hemoglobin concentration
HCT Hematocrit
MCV Mean corpuscular volume
MCH Mean corpuscular Hemoglobin
MCHC Mean corpuscular Hemoglobin concentration
RDW Red distribution width
PLT Platelets
PDW* Platelets distribution width
PCT* Plateletcrit
MPV Mean platelet volume
* PDW and PCT have not been established as indications for this product, in the United
States. The use of PCT and PDW should be restricted to research and Investigational
measurements only.
Technical Manual
1-8 RAA016CEN
Introduction
Specifications
* PDW and PCT have not been established as indications for this product, in the United
States. The use of PCT and PDW should be restricted to research and Investigational
measurements only.
5.1.3. Units
Technical Manual
RAA016CEN 1-9
Introduction
Specifications
5.1.6. Reagents
■ ABX Diluent (10 or 20 l)
■ ABX Cleaner (1 l, integrated)
■ ABX Eosinofix (1 l, integrated)
■ ABX Basolyse 2 (1 l, integrated)
■ ABX Alphalyse or ABX Lysebio ( 0,4 l, integrated)
The CD-ROM RAX055 delivered with your instrument provides Reagents, Controls and
Calibrators leaflets/msds. Latest versions of these documents are available on www.horiba-
abx.com/documentation.
Technical Manual
1 - 10 RAA016CEN
Introduction
Specifications
Technical Manual
RAA016CEN 1 - 11
Introduction
Specifications
Technical Manual
1 - 12 RAA016CEN
Hydraulic & pneumatic principles
Technical Manual
RAA016CEN 2-1
Hydraulic & pneumatic principles
Hydropneumatic connections
1. Hydropneumatic connections
Read this table as follows: LV3_2«S» means connection to port 2 of valve 3 with a «Sleeving».
Technical Manual
2-2 RAA016CEN
Hydraulic & pneumatic principles
Hydropneumatic connections
Technical Manual
RAA016CEN 2-3
Hydraulic & pneumatic principles
Hydropneumatic connections
Technical Manual
2-4 RAA016CEN
Hydraulic & pneumatic principles
Hydropneumatic connections
Read this table as follows: LV3_2«S» means connection to port 2 of valve 3 with a «Sleeving».
Technical Manual
RAA016CEN 2-5
Hydraulic & pneumatic principles
Hydropneumatic connections
Technical Manual
2-6 RAA016CEN
Hydraulic & pneumatic principles
Hydropneumatic connections
Technical Manual
RAA016CEN 2-7
Hydraulic & pneumatic principles
Function of valves
2. Function of valves
Technical Manual
2-8 RAA016CEN
Hydraulic & pneumatic principles
Pneumatic diagrams
3. Pneumatic diagrams
Technical Manual
RAA016CEN 2-9
Hydraulic & pneumatic principles
Pneumatic diagrams
Technical Manual
2 - 10 RAA016CEN
Electric & electronic principles
2. Boards ....................................................................................................................4
2.1. XAA423B, Optical Preampli board ....................................................................................... 4
2.2. XAA425C, Main board .......................................................................................................... 5
2.3. XAA427B, Motor relay board ................................................................................................ 6
2.4. XAA428B, Carriage board .................................................................................................... 7
2.5. XAA429A, LED board ........................................................................................................... 7
4. Connections ...........................................................................................................9
4.1. List ........................................................................................................................................ 9
4.2. XBA193A, Coax BASO-O.D LMNE-LMNE ........................................................................... 9
4.3. XBA387A, Reagent heating coil harness ............................................................................ 10
4.4. XBA389A, Photometer HGB ............................................................................................... 11
4.5. XBA390A, Motor for syringe blocks ................................................................................... 11
4.6. XBA391B, Carriage motor .................................................................................................. 11
4.7. XBA392A, Sampling needle motor ..................................................................................... 11
4.8. XBA393A, Main fan 24V ..................................................................................................... 12
4.9. XBA394A, Carriage sensor ................................................................................................. 13
4.10. XBA396A, Sampling needle sensor .................................................................................. 14
4.11. XBA398B, RBC/WBC coaxial cable ................................................................................. 15
4.12. XBA399C, Ground flowcell fitting ..................................................................................... 16
4.13. XBA402A, Barcode reader harness .................................................................................. 17
4.14. XBA458A, Syringe switch cable Lg=300 mm ................................................................... 18
4.15. XBA460A, Syringe switch cable Lg=500 mm ................................................................... 18
4.16. XBA462A, Motor cable Lg=300 mm ................................................................................. 19
4.17. XBA480A, Optical preampli board harness ...................................................................... 20
4.18. XBA481A, LED board harness .......................................................................................... 20
4.19. XBA482A, Door security switch harness .......................................................................... 21
4.20. XBA620A, Draining detection sensor ............................................................................... 21
4.21. XBA645A, Takasago valve extension harness ................................................................. 22
4.22. XBA646A, Takasago valve wired ...................................................................................... 23
4.23. XDA605A, Diluent tank assembly ..................................................................................... 24
4.24. XDA812A, T° room fan + harness ..................................................................................... 24
5. Synoptic ...............................................................................................................25
Technical Manual
RAA016CEN 3-1
Electric & electronic principles
Main board test points
Test
Designation Potentiometer Target Tolerance Comment
point
TP1 HGB Blank adjust. R248 4.7V +/-0.02 V
TP2 PLT Threshold adjust. R159 300 mV +/-5 mV
TP3 LMNE CIS Threshold adjust. R160 650 mV +/-5 mV
TP4 LMNE OD Threshold adjust. R161 350 mV +/-5 mV
TP5 Draining syringe motor R149 4.8V +/-0.05V
TP6 Counting syringe motor R150 4.8V +/-0.05V
TP7 Reagent syringe motor R151 4.0 V +/-0.05V
TP8 LMNE syringe motor R152 3.0 V +/-0.05V
TP9 Unused
TP10 Horizontal carriage motor R154 3.0 V +/-0.05V
TP11 Sampling syringe motor R155 2.0 v +/-0.05V
TP12 Needle carriage motor R156 4.5 V +/-0.05V
TP13 RBC Threshold adjust. R158 300 mV +/-5 mV
TP14 BASO Threshold adjust. R157 300 mV +/-5 mV
TP15 BASO Comparator
TP16 RBC Comparator
TP17 PLT Comparator
TP18 LMNE CIS Comparator
TP19 LMNE OD Comparator
TP20 RBC line height adjust. R133 Factory Adjusted
TP21 BASO gain adjust. R134 Factory Adjusted
TP22 PLT line height adjust. R135 Factory Adjusted
TP23 LMNE Resistive gain adjust. R136 Factory Adjusted
TP24 LMNE Optical gain adjust. R148 Factory Adjusted
TP25 RBC line width adjust.
TP26 RBC line reject adjust. R142 10 µs +/- 0.5µs Factory Adjusted
TP27 RBC line pulse adjust. Factory Adjusted
TP28 BASO line width adjust.
TP29 BASO line reject adjust. R139 10 µs +/- 0.5µs Factory Adjusted
TP30 BASO line pulse adjust. Factory Adjusted
TP31 PLT line width adjust.
TP32 PLT line reject adjust. R144 10 µs +/- 0.5µs Factory Adjusted
TP33 PLT line pulse adjust. Factory Adjusted
TP34 LMNE Res. line width adjust.
TP35 LMNE Res. line reject adjust. R145 15 µs +/- 0.5µs Factory Adjusted
TP36 LMNE Res. line pulse adjust Factory Adjusted
TP37 LMNE Opt. line width adjust.
TP38 LMNE Opt. line reject adjust. R147 5 µs +/- 0.5µs Factory Adjusted
TP39 LMNE Opt. line pulse adjust. Factory Adjusted
TP40 5 V Power supply 5V +/- 0.15 V Non adjustable
Technical Manual
3-2 RAA016CEN
Electric & electronic principles
Main board test points
Test
Designation Potentiometer Target Tolerance Comment
point
TP41 12 V Power supply 12 V +/- 0.5 V Non adjustable
TP42 24 V Power supply 24 V +/- 1.4 V Non adjustable
TP43 -12 V Power supply -12 V +/- 0.5 V Non adjustable
TP44 GROUND
TP45 GROUND
TP46 GROUND
TP47 GROUND
TP48 LMNE Draining sensor R286 4.5 V +/- 0.3 V On air
<1 V With water
TP49 Unused
Reagent heating system CTN
TP50
voltage
Thermostated room heating
TP51
sytem CTN voltage
TP52 Draining sensor R287 4.5 V +/- 0.3 V On air
<1 V With water
TP53 LMNE Lamp R11 6V
TP54 GROUND
TP55 GROUND
TP56 GROUND
TP57 GROUND
TP58 GROUND
TP59 GROUND
LMNE Res. line reject 2
TP60 R409 50 µs +/- 2 µs Factory Adjusted
adjust.
LMNE Res. line reject 2
TP61 R410 250 µs +/- 5 µs Factory Adjusted
adjust.
Technical Manual
RAA016CEN 3-3
Electric & electronic principles
Boards
2. Boards
XBA480A (+ 12V)
to J36 on Main board
Technical Manual
3-4 RAA016CEN
Electric & electronic principles
Boards
DAD113A to
J2 on motor
relay board
DAD112A to
J5 on carriage
board + TAKASAGO
valve assy
Not used
DAD107A to
valves #1 - 6
DAD109A to
valves #7 - 11
DAD075A to
valves #20 - 26
DAD109A to
valves #27 - 31
DAD107A to
valves #12 - 16
Not used
Heater assy
XBA387A
Not used Room heater Not used Power supply Start cycle Not used Barcode
(XDA812A) switch (XBA482A) reader output
(XBA402A)
Door security LMNE transfer DAD110A to Keyboard & Display assy J2 on optical
switch (XBA482A) sensor M9 (XBA620A) (XAA444A) preampli board
(XBA480A)
Technical Manual
RAA016CEN 3-5
Electric & electronic principles
Boards
XBA391B, B3 XBA394A, M3
Horizontal carriage motor Horizontal carriage home sensor
■ Interconnection board where all the different motors and end of run sensor are connected
■ This board is connected to the main board with a flat cable
Technical Manual
3-6 RAA016CEN
Electric & electronic principles
Boards
DAD112A to J1
XBA390A, B2
on Main board
Sample syringe motor
XBA392A, B1
Vertical carriage motor
XBA458A, SW2
Sample syringe home switch
XBA396A, M1
Vertical carriage
home sensor
■ Interconnection board where all the carriage electrical functions are connected
■ This board is connected to the main board with one flat cable
XBA421B to J9
on Main board
■ This board supports LEDs on the front panel. They give instrument state:
- Red: instrument is busy
- Green: instrument is ready to sample
Technical Manual
RAA016CEN 3-7
Electric & electronic principles
Flat cable list
Length / number of
Reference From To
points
DAD075A 350 mm / 14 pts J18 on main board J1 on 7 valves assy
DAD078A 405 mm / 10 pts J4 on main board RS232 output
DAD107A 480 mm / 14 pts J16 on main board J2 on 6 valves assy
J15 on main board J1 on 5 valves assy
DAD109A 560 mm / 14 pts J17 on main board J1 on 5 valves assy
J14 on main board J1 on 5 valves assy (Non Takasago)
DAD136A J1 on 5 valves assy (Takasago)
DAD110A 750 mm / 40 pts J9 on main board J1 on Keyboard & LCD board
DAD111A 300 mm / 26 pts J2 on main board Printer output
DAD112A 930 mm / 26 pts J1 on main board J5 on carriage board
DAD113A 470 mm / 50 pts J20 on main board J2 on motor relay board
DAD136A 80 mm / 14 pts J14 on main board DAD109A and XBA645A (Takasago only)
Technical Manual
3-8 RAA016CEN
Electric & electronic principles
Connections
4. Connections
4.1. List
Reference From To
XBA193A J22 on main board LMNE flowcell
J25 on main board J1 on Optical preampli board
XBA387A J31 on main board Heating coil harness
XBA389A J26 on main board HGB photometer
XBA390A J4 on carriage board Sampling syringe motor (B2)
J11 on motor relay board Reagent syringe motor (B6)
XBA391B J14 on motor relay board Horizontal carriage motor (B3)
XBA392A J3 on carriage board Vertical carriage motor (B1)
XBA393A J35 on main board Fan
XBA394A J1 on motor relay board Horizontal carriage motor sensor (M3)
XBA396A J1 on carriage board Vertical carriage motor sensor (M1)
XBA398B J23 on main board RBC/PLT counting head
J24 on main board WBC/BASO counting head
XBA399C XBA193A on LMNE flowcell LMNE flowcell
XBA402A J5 on main board Barcode reader output
XBA421B J9 on main board J1 on LEDS board
XBA458A J2 on carriage board Sampling syringe motor home (SW2)
J7 on motor relay board Count syringe motor home (SW4)
J8 on motor relay board Drain syringe motor home (SW7)
XBA460A J5 on motor relay board LMNE syringe motor home (SW5)
J6 on motor relay board Reagent syringe motor home (SW6)
XBA462A J10 on motor relay board LMNE syringe motor (B5)
J12 on motor relay board Count syringe motor (B4)
J13 on motor relay board Drain syringe motor (B7)
XBA480A J36 on main board J2 on Optical preampli board
XBA482A J27 on main board Door security switch
J34 on main board Start cycle switch (Pentra MS60 only)
XBA620A J29 on main board Transfer sensor (M9)
J30 on main board Draining sensor (M8)
DAD136A (connected between J14 on main
XBA645A TAKASAGO valve assy XBA645A
board and DAD109A)
XBA646A XBA645A TAKASAGO valve assy
XDA605A J4 on motor relay board Diluent reservoir
XDA812A J33 on main board Chambers heating harness
No diagram available
Technical Manual
RAA016CEN 3-9
Electric & electronic principles
Connections
Technical Manual
3 - 10 RAA016CEN
Electric & electronic principles
Connections
No diagram available
No diagram available
No diagram available
No diagram available
Technical Manual
RAA016CEN 3 - 11
Electric & electronic principles
Connections
Technical Manual
3 - 12 RAA016CEN
Electric & electronic principles
Connections
Technical Manual
RAA016CEN 3 - 13
Electric & electronic principles
Connections
Technical Manual
3 - 14 RAA016CEN
Electric & electronic principles
Connections
Technical Manual
RAA016CEN 3 - 15
Electric & electronic principles
Connections
Technical Manual
3 - 16 RAA016CEN
Electric & electronic principles
Connections
Technical Manual
RAA016CEN 3 - 17
Electric & electronic principles
Connections
No diagram available
Technical Manual
3 - 18 RAA016CEN
Electric & electronic principles
Connections
Technical Manual
RAA016CEN 3 - 19
Electric & electronic principles
Connections
No diagram available
Technical Manual
3 - 20 RAA016CEN
Electric & electronic principles
Connections
No diagram available
Technical Manual
RAA016CEN 3 - 21
Electric & electronic principles
Connections
Technical Manual
3 - 22 RAA016CEN
Electric & electronic principles
Connections
Technical Manual
RAA016CEN 3 - 23
Electric & electronic principles
Connections
No diagram available
Technical Manual
3 - 24 RAA016CEN
Electric & electronic principles
Synoptic
5. Synoptic
Technical Manual
RAA016CEN 3 - 25
Electric & electronic principles
Synoptic
Technical Manual
3 - 26 RAA016CEN
Analysis cycle technology
Technical Manual
RAA016CEN 4-1
Analysis cycle technology
Analysis cycle description
Dilution:
Blood Volume: 10 µl
ABX Basolyse 2 Volume: 2000 µl
Dilution Rate: 1/200
Technical Manual
4-2 RAA016CEN
Analysis cycle technology
Analysis cycle description
Dilution:
Blood Volume: 25 µl
ABX Eosinofix Volume: 1000 µl
ABX Diluent Volume: 1000 µl
Final Dilution Rate: 1/80
Technical Manual
RAA016CEN 4-3
Analysis cycle technology
Analysis cycle description
Dilution:
Initial Blood Volume: 10 µl
Reagent Volume (RBC chamber): 2500 µl
Final Dilution Rate: 1/10000
HGB:
■ Distribution of 0.4 ml of ABX Diluent then
distribution of 0.36 ml of ABX Lyse into DIL1
chamber while bubbling.
■ WBC\BASO chamber rinsing: Draining and
filling with 1ml of ABX Cleaner and 1.5ml of
ABX Diluent.
■ RBC\PLT counting and HGB measurement.
Technical Manual
4-4 RAA016CEN
Analysis cycle technology
Analysis cycle description
Dilution:
Blood Volume: 10 µl
ABX Basolyse 2 Volume: 2000 µl
Dilution Rate: 1/200
Technical Manual
RAA016CEN 4-5
Analysis cycle technology
Analysis cycle description
Dilution:
Blood Volume: 25 µl
ABX Eosinofix Volume: 1000 µl
ABX Diluent Volume: 1000 µl
Final Dilution Rate: 1/80
Technical Manual
4-6 RAA016CEN
Analysis cycle technology
Analysis cycle description
Dilution
Initial Blood Volume: 10 µL
Reagent Volume (RBC chamber): 2500 µL
Final Dilution Rate: 1/10000
HGB:
■ Distribution of 0.4 ml of ABX Diluent then
distribution of 0.36 ml of ABX Lyse into DIL1
chamber while bubbling.
■ WBC\BASO chamber rinsing: Draining and
filling with 1ml of ABX Cleaner and 1.5ml of
ABX Diluent.
■ RBC\PLT counting and HGB measurement.
Technical Manual
RAA016CEN 4-7
Analysis cycle technology
Measuring principles
2. Measuring principles
diluent
In CBC mode, 30µl of whole blood is
Air
aspirated then delivered with reagents into
chambers as follows: 7 µl discarded
3 µl discarded
Diluent
10 µl RBC/PLT/HGB
3 µl discarded
Technical Manual
4-8 RAA016CEN
Analysis cycle technology
Measuring principles
2.2. RBC/PLT
Technical Manual
RAA016CEN 4-9
Analysis cycle technology
Measuring principles
■ Results:
Number of cells counted per volume unit x
calibration coefficient
Histograms
Technical Manual
4 - 10 RAA016CEN
Analysis cycle technology
Measuring principles
2.3.1. Principles:
■ Alphalyse:
This reagent breaks down the RBC cell membrane and releases the Hemoglobin within the cell. The
hemoglobin, released by the lysing reagent, combines with the Potassium cyanide from the lysing
reagent to form a chromogenous cyanmethemoglobin compound. This compound is then measured
through the optical part of the first dilution chamber by way of spectrophotometry at a wavelength of
550nm.
■ Lysebio:
Reagent for erythrocyte lysis and cyanide-free determination of hemoglobin. By addition of agent of
lysis, hemoglobin is released. All the heme iron is oxidized and stabilized. Oxidation resulting
complexes are quantified by spectrophotometry at a wave length of 550nm.
Technical Manual
RAA016CEN 4 - 11
Analysis cycle technology
Measuring principles
2.3.3. Result
Final HGB result represents: Absorbance value obtained x coefficient of calibration.
■ The height of the impulse generated by the passage of a cell through the micro-aperture is directly
proportional to the volume of the analyzed RBC.
■ The hematocrit is measured as a function of the numeric integration of the MCV.
■ The study of the RBC distribution detects erythrocyte anomalies linked to anisocytosis.
■A Red Cell Distribution Width (RDW) will enable you to follow the evolution of the width of the curve in
relation to the cell number and average volume.
■ With:
- K = system constant
- SD = Determined standard deviation according to statistical studies on cell distribution.
- MCV = Mean Corpuscular Volume of erythrocytes
■ MCV (Mean Cell Volume) is calculated directly from the RBC histogram.
■ MCH (Mean Corpuscular Hemoglobin) is calculated from the Hgb value and the RBC number.
The mean hemoglobin weight in each RBC is given by the formula:
■ MCHC (Mean Corpuscular Hemoglobin Concentration) is calculated according to the HGB and HCT
values. Mean HGB concentration in the total volume of RBC is given by the formula:
The MPV (Mean Platelet Volume) is directly derived from the analysis of the platelet distribution curve.
Technical Manual
4 - 12 RAA016CEN
Analysis cycle technology
Measuring principles
■ All
the WBC are counted between the
electrical threshold <0> threshold <BA3>.
The basophils are located from threshold
<BA2> to threshold <BA3>.
■ Results
- WBC: Number of cells per volume x coefficient of calibration.
- BASO: Number of cells per volume x coefficient of calibration in percentage regarding the total
number of leukocytes (BASO + WBC nuclei)
Technical Manual
RAA016CEN 4 - 13
Analysis cycle technology
Measuring principles
Technical Manual
4 - 14 RAA016CEN
Analysis cycle technology
Measuring principles
■ Incubation duration: 12 s
■ Initial blood volume: 25 µl (CBC or CBC/DIFF)
■ Vol. ABX Eosinofix: 1000 µl
■ Vol. ABX Diluent: 1000 µl
■ Final dilution rate: 1/80
■ Results:
From these measurements, a matrix is drawn up with volumes on the X-axis and optical transmission
on the Y-axis. The study of the matrix image allows the clear differentiation of 4 out of 5 leukocyte
populations. As a matter of fact, the basophil population is very small compared to the others.
■ LYMPHOCYTES: The lymphocyte is a very small round shaped cell with condensed cytoplasm and
large nucleus. This cell is normally positioned in the lower part of the Y-axis, as well as the lower part
of the X-axis because of its small size. The far left side of the lymphocyte area (LL) should normally be
empty. Any detection of cells in the LL area indicates Small lymphocytes, Platelet aggregates, NRBCs
Technical Manual
RAA016CEN 4 - 15
Analysis cycle technology
Measuring principles
(Nucleated Red Blood Cells), and/or improperly adjusted flowcell alignment. Background noise may
also be detected in this area if the interference is important.
■ MONOCYTES: The monocyte is a very large irregular shaped cell with large convoluted nuclei. The
nucleus contains folds and sometimes vacuoles. The cytoplasm is also large with non-granular
intracellular material. This cell does not scatter or absorbs a large amount of light when passing
through the flowcell. It is positioned in the lower part of the Y-axis. Because the monocyte is a large
cell, it is positionned on the right side of the X-axis.
■ NEUTROPHILS: The neutrophil is larger in size than the lymphocyte. It contains granular material in its
cytoplasm along with a segmented nucleus. Due to these cellular features, more light will pass through
neutrophils in the flowcell. As a result, neutrophils go higher on the Y-axis and spread along the X-axis
according to their maturity. Hyper-segmentation and increased granules place this population higher
along the Y-axis.
■ EOSINOPHILS: The eosinophil is somewhat like the neutrophil. It contains granular material and
segmented nuclei within the cytoplasm. The granular material is colorized with a reagent before
passing through the light beam in the flowcell. Due to the colorization action of the reagent, the
eosinophils are placed in the highest part of the Y-axis. Hyper-segmentation and increased granules
place this population in the top right area of the matrix.
■ Additional parameters: LIC (Large Immature Cells) and ALY (Atypical Lymphocytes) complete the panel
available on the matrix.
Immature granulocytic cells are detected by their larger volumes and by the increased granules, which
allow more light to pass through the cells and increase the intensity of scattered light. Therefore, cells
such as metamyelocytes will be found to the right of the neutrophils and almost at the same level.
Myelocytes and promyelocytes will be found on the far right of the matrix in the saturation position. The
metamyelocytes, myelocytes, and promyelocytes will all be classed as (LIC) and there given results will
be included in the neutrophil value. The Blast cells will be generally located to the right of the monocyte
population and, as such, will increase the (LIC) count. Small blast cells will be found between the
normal lymphocyte and mococyte populations (ALY).
Platelet aggregates and debris from RBC cell fragments are found in the background noise area, at the
lower left corner of the matrix. Most of the cell population thresholds are fixed and give the normal
limits for the normal leukocyte morphologies. Changes in the morphology of a specific population will
be indicated on the matrix by a shift in the corresponding population.
A Blast alarm is generated from increased counts within the (LIC) area; this is correlated with the Blast
detection on the Basophil histogram.
Large lymphocytes are usually detected in the (ALY) area, where reactive lymphoid forms, stimulated
lymphocytes, and plasmocytes are also found as well.
Technical Manual
4 - 16 RAA016CEN
Analysis cycle technology
Raw counts
3. Raw counts
The raw counts are printed at the bottom of the Report printout.
■ WBC1 / WBC2: Number of WBCs counted in WBC/BASO chamber during the first and the second
period of count.
■ RBC1 / RBC2: Number of RBCs counted in RBC/PLT chamber during the first and the second period
of count.
■ PLT1 / PLT2: Number of PLTs counted in RBC/PLT chamber during the first and the second period of
count.
■ LYM / MON / NEU / EOS: Number of cells located in their respective population areas.
■ NO / LL / RM / LN / FLN / FMN / FNE / RN : Number of cells located in their respective flag areas.
Technical Manual
RAA016CEN 4 - 17
Analysis cycle technology
Raw counts
C127
EOS
NO
FNE
RN
LN NEU
N
FM
FLN
MON RM
LL LYM
ALY
■ WBC: WBCs used for for the LMNE and BASO balance adjustments.
■ CO%: Correlation percentage: % of cells passing through the flowcell in the correct time (transfer time,
between the aperture and the optical measurement).
Technical Manual
4 - 18 RAA016CEN
Software
5 - Software
1. Versions .................................................................................................................2
Technical Manual
RAA016CEN 5-1
Software
Versions
1. Versions
Technical Manual
5-2 RAA016CEN
Software
USER menu overview
1 - START CYCLE
2 - CHANGE COEFF.
3 - PRINT COEFF.
3 - CALIBRATION 4 - REPEATIBILITY
2 - DAILY WORKLOAD
3 - PRIME
Technical Manual
RAA016CEN 5-3
Software
MAINTENANCE menu overview
MAINTENANCE
1 - AUTOCONTROL
1 - BACK FLUSH
2 - CLEANING
1 - CHAMBERS
2 - RINSE 2 - CYTOMETER
1 - SAMPLING NEEDLE
2 - CARRIAGE
3 - SAMPLING SYR.
4 - MECHANICAL SYS. 1 - INITIALIZATION 4 - DRAINING SYR.
5 - COUNTING SYR.
6 - CYTOMETER SYR.
2 - CHECK MOTORS 7 - DILUTION SYR.
3 - CHECK VALVES 1 - 1 to 11
2 - 12 to 16
3 - 17 to 19
5 - TECHNICIAN 4 - 20 to 26
4 - MAINTENANCE 5 - 27 to 31
CARRIAGE POS.
Technical Manual
5-4 RAA016CEN
Software
SETUP menu overview
1 -CBC
6 - SETUP
1 - NORMAL RANGES 2 - DIFF
4 - THRESHOLDS
2 - UNITS
1 - RS232 CONFIG.
1 - NUMERICAL RESULT
2 - SENDING CONFIG. 2 - FLAGS & PATHOLOGIES
3 - HISTOGRAMS & THRESHOLDS
4 - PATIENT FILE
3 - LAB.LIMITS 5 - RAW DATA
3 - ABX FORMAT
1 - CALIBRATION
4 - RS232
2 - IDENTIFICATION MODE
3 - AUTOCLEAN FREQUENCY
3 - PRINTER CONFIG.
5 - PRINTER
4 - CHANGE PASSWORD
4 - PRINT LATEST RES.
5 - LANGUAGES
7 - CYCLES #
8 - CONFIGURATION
9 - WBC BALANCE
Technical Manual
RAA016CEN 5-5
Software
TECHNICIAN menu overview
2 - MEASUREMENT
1 - ADJUSTMENT
2 - REFERENCE
1 - AUTOCONTROL
1 - REAGENT HEATING SYSTEM
3 - HEATING SYSTEMS 2 - AMBIANCE BOX
2 - CLEANING
1 - ADJUSTMENT
2 - REFERENCE
3 - HYDRAULIC SYS.
4 - BUBBLING
4 - MECHANICAL SYS.
1 - DRAINING
5 - TECHNICIAN 5 - SENSOR CHECK 2 - LMNE TRANSFER
1 - COUNTING
6 - VACUUM CHECK 2 - DRAINING
7 - BURN-IN
8 - USER MODE
1 - CYCLE #
2 - MAINTENANCE CARRIAGE POSITION
9 - OTHERS 3 - PARK SYRINGES POSITION
4 - SPECIAL VALUES
Technical Manual
5-6 RAA016CEN
Alarm & error list
1. Printer .....................................................................................................................2
2. Transmission .........................................................................................................2
3. Calibration .............................................................................................................2
5. Reagents ................................................................................................................3
6. Temperature ..........................................................................................................3
Technical Manual
RAA016CEN 6-1
Alarm & error list
Printer
1. Printer
2. Transmission
3. Calibration
Technical Manual
6-2 RAA016CEN
Alarm & error list
Technical Operations
4. Technical Operations
5. Reagents
6. Temperature
Technical Manual
RAA016CEN 6-3
Alarm & error list
Temperature
Technical Manual
6-4 RAA016CEN
Maintenance
7 - Maintenance
RAS235C: Installation
1. Installation kit XEA791D .......................................................................... RAS235C-2
2. Installation ............................................................................................... RAS235C-3
Technical Manual
RAA016CEN 7-1
Maintenance
Technical Manual
7-2 RAA016CEN
Maintenance
Technical Manual
RAA016CEN 7-3
Maintenance
Maintenance
1. Maintenance
Technical Manual
7-4 RAA016CEN
Maintenance
Maintenance
The Pentra 60 maintenance is scheduled on different maintenance kits. The maintenance schedule
depends on the number of cycles per day, follow the chart table to know the kit part number to use:
XEA485AS
Part number Qty Designation
FAA053A 1 O’ring 1.4x1.25 Fluocarbon (sampling needle)
FAA065A 1 O’ring 6.30x2.40 Silicon ( Reagent syringe )
XDA621A 2 O’ring 30.80x3.60 + washer ( counting + draining syringe)
XDA622A 4 O’ring 15.54x2.62 + washer ( Reagent syringe )
XEA019A 1 Silicon grease
XEA486BS
Part number Qty Designation
DAJ007A 1 Optical bench lamp
FAA040A 1 O’ring 12.1x2.7 5DIFF syringe
FAA046A 2 O’ring 2.75x1.6 Coaxial cable
FAA064A 2 O’ring 14.2x1.52 Sampling syringe
Technical Manual
RAA016CEN 7-5
Maintenance
Maintenance
XEA486BS
Part number Qty Designation
FAA066A 4 O’ring 13.1x1.6 Draining chambers
FAA067A 4 O’ring 2.4x1.9 5DIFF syringe
GBG275A 4 Counting head 0.5 joint
XEA663A 1 LMNE flowcell coaxial kit
XEA286AS 1 Waste chamber washer + joint
JAJ007A 1,5ML Polyeth. Tape
GBL0073 2 Jaws springs
KAL017A 4 Pin 3x12
FAA049A 2 O’ring 2.2x1 silic. 70 SH
XEA581BS
Part number Qty Designation
GBC030A 4 Reagent syringe piston
GBC031A 1 Lyse reagent syringe piston
GBG091A Carriage,needle guide P60
XDA619BS 1 P60 OT equiped needle
GBG260A 2 Piston D=31 Spacer
GBG274A 2 Piston D=31
KAA010A 6 CHC M3x25 screws
Technical Manual
7-6 RAA016CEN
Maintenance
Procedures
2. Procedures
Maintenance and adjustments required for Pentra 60 are divided into procedures according
to the concerned assemblies. This should make any update easier as all interventions can be
done with the corresponding procedure on its own.
Each procedure has to be read entirely before starting intervention.
Technical Manual
RAA016CEN 7-7
Maintenance
Procedures
Technical Manual
7-8 RAA016CEN
RAS235C
Installation
RAS235C: Installation
■ Concerns
■ Installation of Pentra 60
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
Technical Manual
RAS235C - 2 RAA016CEN
Installation
2. Installation
To operate the instrument correctly, install it in a place fulfilling the following conditions:
■ Place where the instrument and reagents are not thermally affected by direct sunlight and intense
radiant heat.
■ Place where it is not exposed to water or vapor.
■ Place where it is flat and free from vibration or shock.
■ Place where an independent power receptacle can be used.
■ Use a receptacle different from the one used by a device that easily generate noise such as a
centrifuge, etc...
■ Provide a space of at least 10 cm at the back of the instrument for arranging the power cable and
pipes.
Technical Manual
RAA016CEN RAS235C - 3
Installation
Risk of erroneous results if diluent container is installed further than 80cm (31.5 in.) bellow
the instrument.
■ Diluent input tubing: cristal 3x6 / 2 meters (80 in.) maximum.
■ Waste output tubing: cristal 4x6 / 2 meters (80 in.) maximum.
Never pour reagent from one container to another. Particles present in the bottom of the old
container can contaminate the new reagent, causing unacceptable background counts,
especially for PLTs.
Technical Manual
RAS235C - 4 RAA016CEN
Installation
Power supply
cable
2.4. Start Up
■ Before analyzing samples, visually inspect reagent lines and pumps for air bubbles. Repeat
priming if air bubbles are still present. Call the HORIBA Medical representative service
department if priming does not eliminate air bubbles.
■ Never initiate two Lyse prime cycles back-to-back. This causes excessive foaming in the
waste chamber. Run a blank cycle between each Lyse prime cycle.
Technical Manual
RAA016CEN RAS235C - 5
Installation
Technical Manual
RAS235C - 6 RAA016CEN
RAS236C
6 month maintenance
■ Concerns
■ O’ring replacement:
- Rinsing block
- Reagent syringe
- Draining & counting syringes
■ Axis lubricating
■ Instrument cleaning
■ Required tools
■ Hexagonal keys
■ Dynamometric screw driver A302, A301, A300
■ Cutting pliers
■ Flat screw driver
■ Philips screw driver
■ Torx keys
■ Required products
■ Minoclair
■ Ethanol or essence C
■ Intervention time
■2 hours
■ Frequency
XEA485AS
Part number Qty Designation
FAA053A 1 O’ring 1.4x1.25 Fluocarbon (sampling needle)
FAA065A 1 O’ring 6.30x2.40 Silicon ( Reagent syringe )
XDA621A 2 O’ring 30.80x3.60 + washer ( counting + draining syringe)
XDA622A 4 O’ring 15.54x2.62 + washer ( Reagent syringe )
XEA019A 1 Silicon grease
Technical Manual
RAS236C - 2 RAA016CEN
6 month maintenance
2. Rinsing block
(1x FAA053A)
Use a dynamometric screw driver to tighten the Rinsing block, tightening torques:
120 mN.m (17 Ozf.in)
Technical Manual
RAA016CEN RAS236C - 3
6 month maintenance
3. Reagent syringe
Technical Manual
RAS236C - 4 RAA016CEN
6 month maintenance
Technical Manual
RAA016CEN RAS236C - 5
6 month maintenance
1 x XDA621A
■ Drain the syringe by pushing the piston nut up and down several times.
■ Open the syringe:
- 4x CHC M4x16 screws on old model.
- Rotation of the syringe body on new model.
Molded syringe (new model) Machined syringe (old model)
Open
Close
The chamber body of the new syringe is directly screwed on the motor block.
When the body is reinstalled on the motor block, it has to be rotated until the mark on the
body is located between the two marks of the motor block
Technical Manual
RAS236C - 6 RAA016CEN
6 month maintenance
■ Replace the O’ring by the new one from the 6 month maintenance kit.
■ Use one drop silicon grease (LAM004A) in between 2 fingers to lubricate O’ring.
■ Apply a thin film of grease on the piston.
■ Make sure that the washer (GBG148A) is present on old model then reassemble in reverse order.
1_Motor assembly
2_O’ring (FAA017A)
4_Washer (GBG148A)
5_Syringe body
■ Clean the endless screw with ethanol or essence C (Do not grease the endless screw).
1 x XDA621A
Technical Manual
RAA016CEN RAS236C - 7
6 month maintenance
5. Axis lubricating
Technical Manual
RAS236C - 8 RAA016CEN
6 month maintenance
6. Instrument cleaning
■ Dilute the 12°cl bleach to 1 part of bleach for 4 of deionize water (1/5).
■ Instrument environment must be cleaned.
■ No sponge, nor cloth must be used. Only absorbant paper, thrown after use, in contamination bins,
have to be used. For small or sensitive assemblies use accurate drier papers.
■ All assemblies suspected to have been in contact with biohazardous material must be disinfected with
diluted bleach (the stainless steel must be bleached below 30°Celsius).
■ Blood stains or salt marks must be cleaned with spray detergent first.
■ Concerned assemblies:
- All the thermostated compartment
- Outer surfaces of the instrument (perpex, covers, reagent locations....)
- Waste connector plug
- Liquid valve push button
- Assemblies close to the needle
- Tube holder assembly
- Overflow trays
■ Reinstall all the assemblies and set up the instrument back to its initial configuration.
■ Before reassembling the covers, start the instrument and check that there is no leak.
■ Reassemble the covers then perform the «RAS252C: Check up after intervention» procedure to check
instrument accuracy and eventually re-calibrate it.
Technical Manual
RAA016CEN RAS236C - 9
6 month maintenance
Technical Manual
RAS236C - 10 RAA016CEN
RAS237C
Yearly maintenance
■ Concerns
■ Required tools
■ Hexagonal & tork keys
■ Dynamometric screwdrivers A300, A301, A302
■ Cutting pliers
■ Flat & cruciform screwdriver
■ Required products
■ Minoclair
■ Distilled water
■ RBC/PLT latex
■ Intervention time
■2 h
■ Frequency
Technical Manual
RAS237C - 2 RAA016CEN
Yearly maintenance
2 x FAA064A
Technical Manual
RAA016CEN RAS237C - 3
Yearly maintenance
1 x FAA040A + 4 x FAA067A
■ Switch off the instrument.
■ Open the left side panel and the mother
board door (lock it).
■ Gentlydisconnect the tubes from the 5DIFF
syringe and unscrew the 2 fixation screws.
FAA040A
■ Open the syringe:
- 2xFX M3x12
- 4xCHC M3x12 screws .
■ Replace all the O’rings (Use one drop of sil-
FAA067A
icon grease between 2 fingers to lubricate
O’rings).
■ Reassemble in reverse order.
Technical Manual
RAS237C - 4 RAA016CEN
Yearly maintenance
1 x DAJ007A
Please refer to "RAS254C, 3. Lamp replacement, page 5"
Technical Manual
RAA016CEN RAS237C - 5
Yearly maintenance
5. LMNE flowcell
1 x XEA663A
■ Switch off the instrument.
■ Open the cover.
■ Open the left side panel and the mother
board door (lock it).
■ Open the optical bench cover.
Technical Manual
RAS237C - 6 RAA016CEN
Yearly maintenance
■ Slowly
turn the flowcell (take care to the
washers).
■ Unscrew the ground screw using a small
philips screwdriver.
■ Put the new LMNE coaxial instead.
■ Reassemble following previous steps back-
ward (except optical bench cover).
■ Enter in: Service\Hydraulics systems\Rinse menu and click the «Rinse cytometer» button to get rid
of the air bubbles stuck to the inner optical surfaces.
■ Check that there is no leak and that flowcell contains no or just a very few air bubbles.
■ Put back the optical bench cover.
■ Check that the displayed values are within the range in the following table.
■ If you are out of range or to get a better adjustment follow the RAS240C: LMNE flowcell adjustment
procedure.
Technical Manual
RAA016CEN RAS237C - 7
Yearly maintenance
1 x XEA286AS
■ Switch on the instrument.
■ Open the pneumatic access door (Right side
of the instrument).
■ RunMenu: Maintenance\Hydraulics sys-
tems\Drain chamber\Diluent reservoir
■ When cycle is finished disconnect the drain
tube under the reservoir.
■ Unscrew the 2 fixation screws under the di-
luent reservoir (few turns to release diluent
reservoir support).
Technical Manual
RAS237C - 8 RAA016CEN
Yearly maintenance
4 x FAA066A
Technical Manual
RAA016CEN RAS237C - 9
Yearly maintenance
8. Counting heads
2 x FAA046A
■ Run Menu: Maintenance\Hydraulics sys-
tems\Drain chamber\All chambers to drain
the chambers.
■ Switch off the instrument.
■ Open the pneumatic access door (right side
of the instrument).
■ Put a piece of absorbant paper or cloth un-
der the chambers assembly.
■ Record the tube positions before disman-
tling the chambers assembly.
■ Disconnect chamber’s tubes except for the
wastes.
■ Unscrew the 3 screws from the chambers
assembly.
Technical Manual
RAS237C - 10 RAA016CEN
Yearly maintenance
4 x GBG275A
Technical Manual
RAA016CEN RAS237C - 11
Yearly maintenance
■ Do not introduce any sharp instrument inside so as to avoid damaging the inside of the
chamber and the aperture.
■ Do not manipulate the aperture using hard instruments. Clean the aperture with a piece of
soft paper or preferably in between 2 fingers.
■ It is recommended to reconnect the tubes on the counting head before reassembling the
electrode and the chamber in order to avoid applying constraint on the chamber.
■ Do not apply too much pressure on the electrode fixation screws as it can break the
aperture.
■ Tightening torque = 400 m.Nm (56.8 Ozf.in).
9. Additional maintenance
■ Before reassembling the covers, start the instrument and check that there is no leak.
■ Reassemble the covers then perform the «RAS252C: Check up after intervention» procedure
Technical Manual
RAS237C - 12 RAA016CEN
RAS238C
■ Concerns
■ Required tools
■ Hexagonal keys
■ Dynamometric screw driver A302, A301, A300
■ Flat screw driver
■ Required products
■ Ethanol or essence C
■ Intervention time
■2 hours
■ Frequency
1. Kit XEA581BS
Technical Manual
RAS238C - 2 RAA016CEN
Every 2 years maintenance
2. Sampling parts
(1 x XDA619BS + 1 x GBG091A)
Technical Manual
RAA016CEN RAS238C - 3
Every 2 years maintenance
3. Reagents syringe
Technical Manual
RAS238C - 4 RAA016CEN
Every 2 years maintenance
Technical Manual
RAA016CEN RAS238C - 5
Every 2 years maintenance
Open
Close
The chamber body of the syringe is directly screwed on the motor block.
When the body is reinstalled on the motor block, it has to be rotated until the mark on the
body is located between the two marks of the motor block
Technical Manual
RAS238C - 6 RAA016CEN
Every 2 years maintenance
2
■ Prepare the piston: on the new molded old piston
syringe, the piston (1) is shorter.
■ For new molded syringe, install only the
piston GBG274A (1).
■ For old machined syringe, use the piston 1
GBG274A (1) + the cross piece GBG260A
(2) and use the CHC M3x20 screws from the
kit.
■ Use one drop silicon grease (LAM004A) in between 2 fingers to lubricate O’ring.
■ Apply a thin film of grease on the piston.
■ Make sure that the washer (GBG148A) is present on old model then reassemble in reverse order.
1_Motor assembly
2_O’ring (FAA017A)
4_Washer (GBG148A)
5_Syringe body
■ Clean the endless screw with ethanol or essence C (Do not grease the endless screw).
■ Open the left side panel and the mother board door (lock it) to access to the syringe.
■ Follow the same procedure than for Draining syringe.
Technical Manual
RAA016CEN RAS238C - 7
Every 2 years maintenance
5. Additional maintenance
■ The «Every 2 years maintenance» includes the «Yearly maintenance» and the «6 month maintenance».
■ Therefore, perform the RAS236C: 6 month maintenance and the RAS237C: Yearly maintenance
procedures.
■ Before reassembling the covers, start the instrument and check that there is no leak.
■ Reassemble the covers then perform the «RAS252C: Check up after intervention» procedure to check
instrument accuracy and eventually re-calibrate it.
Technical Manual
RAS238C - 8 RAA016CEN
RAS239C
■ Concerns
■ Required tools
■ Hexagonal keys
■ Clamps
■ Flat screw driver
■ Torx keys
■ Required products
■ Fungicidal, bacterial, virus killing detergent spray,
non corrosive for metals, non plastic altering.
■ Minoclair
■ 9° bleach
■ Deionized water
■ Distilled water
■ Absorbant paper
■ Intervention time
■ 1h30min.
■ Frequency
■ On request
1. Preparation
■ When cycle is finished, switch the instrument off and remove the power supply cable.
■ Open the instrument covers.
■ Spray the bactericidal cleaner on all biohazardous areas (assemblies in contact with the biohazardous
material such as instrument cover, tube holder, keyboard...) and wait for 10 minutes.
2. Manual decontamination
■ Dilute the 9°Cl bleach to 1 part of bleach for 4 of deionize water (1/5).
■ Instrument environment must be cleaned and decontaminated.
■ No sponge, nor cloth must be used, only absorbant paper thrown after use in contamination bins, have
to be used. For small or sensitive assemblies use accurate drier papers.
■ All assemblies suspected to have been in contact with biohazardous material must be disinfected with
diluted bleach (the stainless steel must be bleached below 30°C).
■ Blood stains or salt marks must be cleaned with spray detergent first.
■ Concerned assemblies:
- Outer surfaces of the instrument (perpex, covers, reagent locations....)
- Waste connector plug
- Liquid valve push button
- Assemblies close to the needle
- Overflow trays
■ Reinstall all the assemblies and set up the instrument back to its initial configuration.
Technical Manual
RAS239C - 2 RAA016CEN
Decontamination & rinse
■ Prepair 1 bottle containing 1/2 litre of bleach diluted to 1 part of bleach for 9 parts of deionize water
(1/10).
■ Prepair 1 bottle containing 1/2 litre of distilled water.
■ Replace the reagent bottles by the diluted bleach bottle.
■ Switch the instrument on.
■ Run 6 consecutive «All reagents» prime cycles (Menu: Reagents\Prime\All Reagents).
■ Fill a sample tube with diluted bleach to 1 part of bleach for 4 of deionize water (1/5)
■ Run 15 burn-in cycles (Menu: Technician\User mode\Burn-in).
■ Leave the instrument operating until it stops.
■ Remove the reagent straws from the bottle containing the diluted bleach.
■ Wrap the straws in absorbant paper.
■ Run 2 consecutive «All reagents» prime cycles (Menu: Reagents\Prime\All Reagents): the bleach is
drained.
■ Replace the diluted bleach bottle by the distilled water bottle and re-plunge the straws in distilled
water.
■ Run 6 consecutive «All reagents» prime cycles (Menu: Reagents\Prime\All Reagents).
■ Remove the reagent straws from the distilled water (Wrap the straws in absorbant paper).
■ Run 2 consecutive «All reagents» prime cycles (Menu: Reagents\Prime\All Reagents): the distilled
water is drained.
■ Run a STAND BY cycle.
■ Re-install the reagent bottles and the straws.
■ Switch off the instrument.
■ Close the instrument cover.
Technical Manual
RAA016CEN RAS239C - 3
Decontamination & rinse
Technical Manual
RAS239C - 4 RAA016CEN
RAS240C
■ Concerns
■ LMNE Flowcell
■ Required tools
■ Hexagonal keys
■ Flatscrew driver
■ Torx keys
■ Required products
■ RBC/PLT Latex (LAD002DS)
■ Fresh blood sample
■ Calibrator
■ Intervention time
■1 h
■ Frequency
■ On request
This adjustment procedure must be carried out when LMNE flowcell is moved or replaced, or when
LMNE tube is changed.
This adjustment must be checked in case of matrix quality trouble.
Corresponds to the voltage on the LMNE amplifier board, between TP1 and TP2.
This voltage is defined by the light (from the lamp) received on the cell of the LMNE amplifier board.
Target value: 6,00 Volts +/- 0.50, adjustable with R11 on LMNE amplifier board.
Technical Manual
RAS240C - 2 RAA016CEN
LMNE flowcell adjustment
Time between the resistive detection of the cell in the micro aperture (1) and the absorbance detection
of the cell in the optical flow (2). This time is defined by the height of the optical flowcell, adjustable
with gauges.
Gauges
This volumetric adjustment defines the size of the cells counted in the micro aperture of the LMNE
flowcell.
Target value: 53 +/-2, adjustable with R136, located on the mother board.
LMNE amplifier
LMNE ABS board cell
Technical Manual
RAA016CEN RAS240C - 3
LMNE flowcell adjustment
2. Requirements
Technical Manual
RAS240C - 4 RAA016CEN
LMNE flowcell adjustment
Never dismantle the emission gun or even unlock it. The adjustment is factory made and
cannot be performed in the field.
■ Run a Rinse cytometer cycle to get rid of air bubbles stuck to the inner optical surfaces:
■ Run Menu: Maintenance\Hydraulics systems\Rinse\Cytometer.
■ Check that the flowcell contains no or just a very few air bubbles.
■ Check that the displayed values are within the range in the following table.
■ If you are out of range or to get a better adjustment follow the procedure.
Technical Manual
RAA016CEN RAS240C - 5
LMNE flowcell adjustment
■ Run a Rinse cytometer cycle to get rid of air bubbles stuck to the inner optical surfaces:
■ Run Menu: Maintenance\Hydraulics systems\Rinse\Cytometer.
■ Check that the flowcell contains no or just a very few air bubbles.
Technical Manual
RAS240C - 6 RAA016CEN
LMNE flowcell adjustment
■ First,
turn the lateral screw (A) to obtain a A
stable transfer on the LMNE adjustment
screen.
■ Run several cycles on normal human fresh blood and check if the CO% (See Section 4, 3. Raw counts,
page 4-17) may be increased by slightly turning the axial toothed wheel (B).
Technical Manual
RAA016CEN RAS240C - 7
LMNE flowcell adjustment
5. Final adjustment
■ Check the position of the populations on the matrix: the populations must be correctly
placed in their respective zone (box), without flags.
■ Check that the LYA and GCI values are as lower as possible (average maximum 1,3%).
■ Check that no alarm is triggered.
Resistivity axis
1- Resistive gain adjustment:
- The resistive gain is set by R136 on the mother board.
- According to the resistive gain adjustment, populations of cells will move along the resistive axis in
different proportions according to the population:
- If the resistive gain is increased, cell populations will move to the right of the matrix.
- If the resistive gain is decreased, cell populations will move to the left of the matrix.
■A good adjustment is obtained when the Lymphocyte population is correctly centered in the LYM box.
Technical Manual
RAS240C - 8 RAA016CEN
LMNE flowcell adjustment
■ Run this procedure in case of Transfer time out of range only (target value: between 150 and 250).
■ In case of replacement of an optical flowcell, a gauge set may be necessary to adjust the transfer time.
Before adjusting the height of the flowcell, make sure the other parameters are correct (lamp
voltage, resistive gain, LMNE Abs gain).
After intervention,check with normal human fresh blood that you have a correct matrix.
Technical Manual
RAA016CEN RAS240C - 9
LMNE flowcell adjustment
Technical Manual
RAS240C - 10 RAA016CEN
RAS241C
Chambers adjustment
■ Concerns
■ Required tools
■ Hexagonal keys
■ Gauge set MAJ004A
■ Required products
■ None
■ Intervention time
■ 15 mn
■ Frequency
■ On request
■ Itis mandatory to perform the «RAS242C: Needle adjustment» procedure after this
adjustment.
■ This adjustment procedure must be carried out when the chamber assembly is moved or
replaced.
1. Alignment check
■ Switch off the instrument and disconnect the power supply cable.
■ Open the right access door.
Technical Manual
RAS241C - 2 RAA016CEN
Chambers adjustment
2. Alignment adjustment
■ If
you need to get a better position, free the
chamber support by loosening the 6
screws.
■ Install
the gauge over the Rinse chamber
then move the carriage over the gauge.
■ Gently
push down on the needle until it
comes just in contact with the gauge.
■ Tighten gently the left upper screw.
Technical Manual
RAA016CEN RAS241C - 3
Chambers adjustment
Technical Manual
RAS241C - 4 RAA016CEN
RAS242C
Needle adjustment
■ Concerns
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■ 15 mn
■ Frequency
■ On request
1. Introduction
Check that the chamber assy is parallel to the carriage motion before checking or making an
adjustment of the Needle position. Follow RAS241C: Chambers adjustment procedure of this
manual.
Needle Home: Check (1): Allows to check Check (2): Checks the
Adjustment to get the best needle cleaning position deepness of the needle
needle extremity cleaning in front of LMNE-3 inlet
Carriage Home:
Allows adjustment of the
carriage sampling position
Needle position:
needle deepness
up & down adjustment
Technical Manual
RAS242C - 2 RAA016CEN
Needle adjustment
Technical Manual
RAA016CEN RAS242C - 3
Needle adjustment
To check the right position of the needle, you must see inlet as in «A» and not as in «B» (where
circles are not concentric).
A B
■ Needle Up/Down adjustment is the same for all the chambers and cannot be modified from
one chamber to another (adjustment by the «NEEDLE POSITION» value)
■ Needle Left/Right position in the chambers can be adjusted separately for each chamber
(except the Rinse chamber).
Technical Manual
RAS242C - 4 RAA016CEN
Needle adjustment
.
os
g ep s.
arria e po
c iag
ASO carr
/B NE
BC LM
(2 )W (3)
s. s.
e po po
iag iag
e 2
carr rr
1 ca
DIL RB
C
(1) (4) 4
3
1
Technical Manual
RAA016CEN RAS242C - 5
Needle adjustment
Technical Manual
RAS242C - 6 RAA016CEN
RAS243C
■ Concerns
■ Required tools
■ Smallflat screw driver
■ Hexagonal keys
■ Voltmeter
■ Required products
■ RBC/PLT Latex (LAD002DS)
■ Intervention time
■ 30min
■ Frequency
■ On request
■ The main board is located on the left hand side of the instrument.
■ Unscrew the 4 CHC screws and remove the left panel to access to the main board.
Technical Manual
RAS243C - 2 RAA016CEN
Main board adjustment
■ The power supply test points are located at the bottom of the main board and are not adjustable.
Technical Manual
RAA016CEN RAS243C - 3
Main board adjustment
■ The motor current test points are located on the left upper corner of the main board.
■ Check and if necessary adjust the motor currents according to the following table (ground on TP57)
Technical Manual
RAS243C - 4 RAA016CEN
Main board adjustment
4. Threshold adjustment
■ The threshold test points are located at the top of the main board.
■ Check and if necessary adjust the threshold voltages according to the following table (ground on TP57)
Technical Manual
RAA016CEN RAS243C - 5
Main board adjustment
■ Adjust the voltage between TP52 and ground to 4.5Volts +/- 0.3 using R287.
■ Press the Esc key.
■ The utility then fills the sensor with the diluent and the following message is displayed:
■ Check that the voltage between TP52 and ground is less than 1 Volt.
■ The test point is located in the left lower corner of the main board.
Technical Manual
RAS243C - 6 RAA016CEN
Main board adjustment
■ Adjust the voltage between TP48 and ground to 4.5Volts +/- 0.3 using R286.
■ Press the Esc key.
■ The utility then fills the sensor with the diluent and the following message is displayed:
■ Check that the voltage between TP48 and ground is less than 1 Volt.
■ The test point is located in the left lower corner of the main board.
Technical Manual
RAA016CEN RAS243C - 7
Main board adjustment
This adjustment is made with the right side door closed and under stabilised thermic
conditions (5 min at least after startup).
Technical Manual
RAS243C - 8 RAA016CEN
Main board adjustment
■A special cycle allows the dilution of the RBC/PLT LATEX. A special count programme carries out the
calculations of the mean volume in the predefined zones and displays them every 700 ms. The
duration of the measurement cycle is 21 seconds.
■ The adjustments are made using R133 potentiometer for the RBC and R135 for the PLT.
PLT RBC
Technical Manual
RAA016CEN RAS243C - 9
Main board adjustment
■ A special cycle allows the dilution of the RBC/PLT LATEX. A special count program carries out the
calculations of the mean volume in the predefined zones and displays them every 700 ms. The
duration of the measurement cycle is 21 seconds
WBC
Technical Manual
RAS243C - 10 RAA016CEN
RAS244C
Temperature adjustment
■ Concerns
■ Required tools
■ Thermometer
■ Required products
■ None
■ Intervention time
■ 30 mn
■ Frequency
■ On request
■ Concerns the control of the Heating coil temperature (reagent heating system).
This control must be made with the right side door closed and under stabilised thermic
conditions (5min at least after startup)
Make sure you have plunged the thermometer probe deeply into the liquid.
■ Check is validated if you get during control at least 10 times the right temperature.
■ Ifan adjustment is required, when heating block has been replaced for example, perform
the Heating coil temperature adjustment procedure.
Technical Manual
RAS244C - 2 RAA016CEN
Temperature adjustment
The Heating coil CTN value is located on the CTN wire, close to the main board, connected
on J31.
Be sure to read the right value, as the sticker may be up or down.
Technical Manual
RAA016CEN RAS244C - 3
Temperature adjustment
■ Make a Heating coil temperature check (See “1.1. Heating coil temperature check, page 2”).
■ If required change the «Reference temperature» value and check temperature again, until you get an
acceptable temperature.
Technical Manual
RAS244C - 4 RAA016CEN
Temperature adjustment
2. Thermostatic compartment
This control must be made with the right side door closed and under stabilised thermic
conditions (5min at least after startup)
■ Click the «Run» button and check that temperature is aroud 35°C +/-1°C every time you hear a «bip».
■ Check is validated if you get during control at least 10 times the right temperature.
■ If an adjustment is required, perform the Heating coil temperature adjustment procedure.
Technical Manual
RAA016CEN RAS244C - 5
Temperature adjustment
The Thermostatic compartment CTN value is located on the CTN wire, close to the main
board, connected on J33.
Be sure to read the right value, as the sticker may be up or down.
Technical Manual
RAS244C - 6 RAA016CEN
Temperature adjustment
■ If
you do not get the right temperature,
adjust it by the mean of the «Reference
temperature».
■ Make a Thermostated compartment temperature check as described at the begining of the procedure.
■ If required change the «Reference temperature» value and check temperature again, until you get an
acceptable temperature.
Technical Manual
RAA016CEN RAS244C - 7
Temperature adjustment
Technical Manual
RAS244C - 8 RAA016CEN
RAS245C
Vacuum adjustment
■ Concerns
■ Required tools
■ Barflex
■ Required products
■ None
■ Intervention time
■ 15 mn
■ Frequency
■ On request
No adjustment available
■ When installed in altitude, some instruments can trigger some «Transfer Sensor Timeout»
alarms. This is due to the decreased atmospheric pressure.
■ In these conditions, the draining vacuum can lower down to about – 200mb/–210mb
instead of –260mb (20% less): this results in a too long transfer time of the solution
between LMNE chamber to T1-T2 segment before the flowcell.
■ In such cases, we recommend to replace some 1.02 tubing (P/n: EAE005AS) by 1.3 tubing
(P/n: EAE006AS) in the following locations:
1- from valve #4 / port 2 to «T2/2»: 1.02x130 replaced by 1.3x130.
2- from «T2/3» to bottom right of 5Diff syringe: 1.02x240 replaced by 1.3x240.
3- from top right of 5Diff syringe to valve #5 / port 1: 1.02x95 replaced by 1.3x95.
Technical Manual
RAS245C - 2 RAA016CEN
Vacuum adjustment
1 2
Technical Manual
RAA016CEN RAS245C - 3
Vacuum adjustment
■ Menu: Maintenance\Technician\Vacuum
Check\Counting\Check.
■ Check vacuum on the Barflex while the fol-
lowing screen is displayed, vacuum must be
at 220mb.
■ Disconnect the Barflex and connect back
the tube on the counting syringe.
■ Close the right side door.
■ Menu: Maintenance\Technician\Vacuum
Check\Counting\Check.
■ Check vacuum on the Barflex, vacuum
must be at 220mb.
■ Press Esc key and change the step value. If
vacuum is too low increase step value.
■ Repeat vacuum check and step value adjustment until you get the correct vacuum.
■ Disconnect the Barflex and connect back the tube on the counting syringe.
■ Close the right side door.
Technical Manual
RAS245C - 4 RAA016CEN
RAS247C
Covers dismantling
■ Concerns
■ Required tools
■ Hexagonal keys
■ Torx keys
■ Required products
■ None
■ Intervention time
■ 10 mn
■ Frequency
■ On request
1. Dismantling
Technical Manual
RAS247C - 2 RAA016CEN
Covers dismantling
B
A
■ Make sure the needle is lift and push the
carriage backward.
■ Loosen the K30x16 torx screw (B).
■ Unscrew the 2 CHC M4x6 upper screws (A).
Technical Manual
RAA016CEN RAS247C - 3
Covers dismantling
2. Mounting
■ Be careful not to clamp one of the reagent’s tube behind the reagent compartment.
Technical Manual
RAS247C - 4 RAA016CEN
RAS248C
■ Concerns
■ Required tools
■ Hexagonal keys
■ Dynamometric screw driver A302, A301, A300
■ Flat screw driver
■ Torx keys
■ Required products
■ None
■ Intervention time
■1 h.
■ Frequency
■ On request
Technical Manual
RAS248C - 2 RAA016CEN
Heater assy replacement
- «LMNE2»
(LMNE2 pass through the third hole,
between LMNE and RBC chamber).
- BASO2 (connector).
(BASO2 pass through the hole in the
chamber plate, on the right side of WBC/
BASO chamber.
Technical Manual
RAA016CEN RAS248C - 3
Heater assy replacement
Technical Manual
RAS248C - 4 RAA016CEN
Heater assy replacement
■ Put the new heater assy on the instrument frame and connect following tubes:
From To
Heater assy rep n°1 Valve 22 Inlet 2
Heater assy rep n° 3 Valve 8 Inlet 2
Heater assy rep n° 5 T4
Heater assy rep n° 7 Valve 22 Inlet 1
Heater assy rep n° 10 Valve 11 Inlet 2
■ Lift
the heater assy, make the tubes go
through the chamber’s support:
E
- Dil1 &DIL 2 pass through the first hole (A). D
C
B
- LMNE3 & LMNE4 pass through the A
second hole (B).
Technical Manual
RAA016CEN RAS248C - 5
Heater assy replacement
■ Put back the chamber assy and connect the following tubes on chambers:
- «DIL1»
- «DIL2»
- «LMNE4» (on transfer sensor)
- «LMNE3»
- «LMNE2»
- BASO2 (connector).
■ Install the CTN on its support.
■ Connect heating system wiring on the mother board in J33.
■ Put back the Reagents syringe and the LMNE syringe and screw silent blocks (check that no tube is
disconnected).
■ Put back the covers (see RAS247C: Covers dismantling).
■ Check chambers position (see RAS241C: Chambers adjustment).
■ When startup and prime are done check that there is no leak.
Technical Manual
RAS248C - 6 RAA016CEN
RAS249C
Bubbling adjustment
■ Concerns
■ Bubbling adjustment.
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■ 15 mn
■ Frequency
■ On request
■ Bubbling is factory adjusted and normally does not required further adjustment.
■ If an adjustment is necessary follow this procedure:
2. Bubbling values
Technical Manual
RAS249C - 2 RAA016CEN
RAS250C
■ Concerns
■ Required tools
■ Hexagonal keys
■ Voltmeter
■ Required products
■ None
■ Intervention time
■ 30 mn
■ Frequency
■ On request
Technical Manual
RAS250C - 2 RAA016CEN
Power supply replacement
Be careful to the lamp power supply cable when you remove the power supply.
Technical Manual
RAA016CEN RAS250C - 3
Power supply replacement
Technical Manual
RAS250C - 4 RAA016CEN
RAS251C
■ Concerns
■ Required tools
■ Hexagonal keys
■ Torx key
■ Required products
■ None
■ Intervention time
■ 30 mn
■ Frequency
■ On request
1. Introduction
■ The start cycle microswitch has to be changed or dismantle only when required
■ To access to the start cycle microswitch you need to remove the Front panel (see RAS247C: Covers
dismantling).
■ Open the pneumatic access door (right side of the instrument).
Technical Manual
RAS251C - 2 RAA016CEN
Start cycle microswitch replacement
■ Install
the fixing plate (A) previously
dismantled on this switch.
■ Put back the start cycle microswitch in the
2 slots (nearly full backward).
■ Tighten the 2 microswitch screws.
■ Check with the front panel in its location, that the Start cycle key is operating normally. If not move
forward or backward the microswitch until you get a correct location.
■ You must ear a «click» when key is pressed, and one when key is released.
■ Reinstall fan.
■ Reinstall covers.
Technical Manual
RAA016CEN RAS251C - 3
Start cycle microswitch replacement
Technical Manual
RAS251C - 4 RAA016CEN
RAS252C
■ Concerns
■ Required tools
■ None
■ Required products
■ Fresh and normal blood samples
■ Calibration blood samples
■ Intervention time
■ 45 mn
■ Frequency
■ On request
1. Preliminary
2. Repeatability
■ Based on 20 consecutives analyses without alarm from one fresh and normal blood sample.
■ Run 20 consecutives analyses.
■ Control to have variation coefficients within the following acceptable limits.
2
Xi X – Xi SD
X = ------------ CV = ------- 100
n SD = ------------------------------
n–1 X
X: Mean
Xi: Measure value
n: Measure number
SD: Standard deviation
■ With one fresh and normal blood sample check that there is no alarm in CBC+5DIFF mode.
Technical Manual
RAS252C - 2 RAA016CEN
Check up after intervention
3. Calibration
■ If necessary, change Lot number, Expiration date or Target values according to the lot you are going
to use.
■ Prepare the calibrator according to the specific instructions detailed in the calibrator package insert
(temperature, mixing, etc...).
■ Open the vial and position the sampling needle deeply inside the bottle
■ Press the sampling bar located behind the needle
■ When the cycle LED stops flashing, remove the vial and replace the cap on the calibrator.
■ Risk of erroneous results if the calibrator is not continously mixed between each analysis.
Continue mixing the calibrator between each analysis.
■ The calibration can be performed on 3 to 11 analyses. In order to obtain the best possible
calibration, it is recommended to run at least 5 calibrator samplings.
■ When the analysis cycle ends, the first result is displayed on the result chart table.
■ Run another calibrator sample when Green LED turns on.
Technical Manual
RAA016CEN RAS252C - 3
Check up after intervention
■ Check that the calibration coefficients remain within the following ranges:
Technical Manual
RAS252C - 4 RAA016CEN
RAS253C
■ Concerns
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■ 15 mn
■ Frequency
■ On request
1. Introduction
3. Chambers
Technical Manual
RAS253C - 2 RAA016CEN
Step by step analysis cycle
4.1.3. Second dilution for the RBC and the PLT (for instrument NON
equipped with Takasago valves):
■ Transfer of the needle from the Rinse chamber to the Dil1/HGB chamber (sampling needle motor,
carriage motor).
■ Sample of 42.5 µl of the first dilution (sampling syringe).
■ Rinsing of needle exterior (in the Dil1/HGB chamber) with 0.4 ml of ABX Diluent (reagents syringe).
■ Transfer of the needle from the Dil1/HGB chamber to the RBC chamber (sampling needle motor,
carriage motor).
■ Distribution of the 42.5 µl of first dilution into a flux of 2 ml of ABX Diluent (sampling needle motor,
carriage motor).
Technical Manual
RAA016CEN RAS253C - 3
Step by step analysis cycle
■ Distribution of 0.5 ml of ABX Diluent from the interior of the needle (sampling needle motor, carriage
motor).
■ In the case of this RBC/PLT dilution, it is the entire blood sample that is distributed (and not
a portion as in the preceding dilution). The distribution in a flux is kept in order to benefit
from the homogenisation aspect of the procedure.
■ In order to attain the hemoglobin measurement we add 0.4 ml of lysing agent into the first
dilution chamber immediately following the blood sample for the RBC/PLT dilution.
4.1.4. Second dilution for the RBC and the PLT (for instruments equipped
with Takasago valves):
■ Valves 32 and 33 commute (ON).
■ Valves 9 & 10 commute (ON).
■ Aspiration of a part of the first dilution through EV32 & 33 by the mean of the Diluent reagent syringe
(the lenght beetwen EV32 and EV33 gives the exact volume for the RBC sample).
■ EV 32 and 33 commute (OFF).
■ The sample (42 µL) is pushed into RBC chamber with a flux of 2.2 mL of ABX Diluent by the mean of
the Diluent reagent syringe through valves 9 & 10.
■ EV 9 and 10 commute (OFF).
■ Bubbling by the mean of the Draining syringe through valve 30 (ON).
■ LMNE: Loading of the vacuum injection circuit of the optical bench (draining syringe, control of valves
4 and 5).
■ Start flux in the optical chamber (approximately 2 seconds).
■ Injection, acquisition of the matrix and WBC count for 12 seconds.
■ Simultaneously: WBC and BASO counts: Rinsing of the counting head (diluter pushed into position,
control of valves 9, 10, 21, 23 and 13).
■ Start the vacuum and first part of count (duration of 6 seconds, control of valve 23).
■ Rinsing of counting head and adjustment of the vacuum count.
■ Second part of count (duration of 6 seconds).
■ After this first series of measurements:
- Measurement of the hemoglobin.
- Rinsing of the WBC/BASO chamber with a mixture of 1 ml of ABX Cleaner and 1.5 ml of ABX Diluent.
- Rinsing of the WBC/BASO counting head.
■ Rinsing of the RBC/PLT counting head. Start the vacuum for the RBC/PLT counting head.
■ First series of counts (duration of 5 seconds, control valve 14).
■ Rinsing of the RBC/PLT counting head and adjustment of the vacuum.
■ Second series of counts (duration of 5 seconds).
Technical Manual
RAS253C - 4 RAA016CEN
Step by step analysis cycle
■ At the end of the RBC/PLT count, the RBC count vacuum is «finished» and used for the filling up of the
diluent tank (control valves 15 and 25).
■ In order to improve the filling speed, the syringe is initialised for a few seconds then drawn again to
complete the filling.
■A float located in the tank controls the level of liquid.
■ For reasons of maximising the mechanical resources, the rinsing of the RBC/PLT chamber is
«normally» programmed for the beginning of the succeeding cycle (which can be started during the
RBC and PLT count).
■ If no other cycle is started during RBC/PLT count or during the filling up of the tank, a special rinsing
sequence is initiated. The sample signal light changes to red and the instrument is «jammed» for a
few seconds.
■ This sequence involves draining of the RBC chamber followed by a refilling with the diluent.
Technical Manual
RAA016CEN RAS253C - 5
Step by step analysis cycle
Technical Manual
RAS253C - 6 RAA016CEN
RAS254C
■ Concerns
■ Required tools
■ Hexagonal keys
■ Flat screw driver
■ Required products
■ None
■ Intervention time
■ 30 min.
■ Frequency
■ On request
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it.
C
■ Disconnect tubes from LMNE syringe: D
- Inlet 1 (A)
- Inlet 5 (B)
■ Disconnect tubes from valves:
- 1 Inlet 1 (C)
- 4 Inlet 2 (D)
A
B
Technical Manual
RAS254C - 2 RAA16CEN
Optical bench dismantling & replacement
Do not tighten silent blocks too hard, check that optical bench cannot be lifted.
Technical Manual
RAA016CEN RAS254C - 3
Optical bench dismantling & replacement
2. Control
■ Be careful, if you use a 3mm allen key to control gap, not to damage the flowcell or the
emission gun.
■ This adjustment is factory made and must not be changed.
Technical Manual
RAS254C - 4 RAA16CEN
Optical bench dismantling & replacement
3. Lamp replacement
■ Switch off the instrument and wait for the lamp to cool down for at least 10 minutes.
As for halogen/quartz lamp, do not touch the bulb with your fingers. This will reduce
significantly the shelf life of the lamp. In case of finger contact, clean the bulb with a solution
of 90% alcohol and a soft paper.
Technical Manual
RAA016CEN RAS254C - 5
Optical bench dismantling & replacement
■ Make sure the flowcell is full of liquid, if not run a «Rinse Diluent» cycle.
■ Make sure that there is no bubble inside the flowcell.
■ Make sure of the cleanliness of the flowcell optical window and objectives.
Technical Manual
RAS254C - 6 RAA16CEN
Optical bench dismantling & replacement
■ If
the projection of the filament is not correct, as shown on previous picture, adjust the lamp support
position: (a small screw, 1.2mm allen key, allows adjutment of lamp).
Technical Manual
RAA016CEN RAS254C - 7
Optical bench dismantling & replacement
Technical Manual
RAS254C - 8 RAA16CEN
RAS255C
■ Concerns
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■ 15 mn
■ Frequency
■ On request
■ Check that the barcode reader is working properly by means of the following reading test.
with C.D.
1224488
reading12244881
without C.D.
121314151617
reading 121314151617
with C.D.
reading 12345ABCDE
without C.D.
reading 12345ABCDEW
12345abcde QWERTY
12345ABCDE azerty
37859 123456/$
reading 37859 reading 123456/$
Technical Manual
RAS255C - 2 RAA016CEN
External barcode reader configuration
■ An audible beep occurs after each reading. Read all the labels from the top to the bottom.
■ When the last label is read, check once more on the test labels.
For the code I2 of 5, to avoid a bad interpretation, it is mandatory to hold the reader in order
to read the entire code.
■ If the reading test has failed for the DATALOGIC reader, proceed to the barcode setup as follows:
Technical Manual
RAA016CEN RAS255C - 3
External barcode reader configuration
Technical Manual
RAS255C - 4 RAA016CEN
RAS304C
■ Concerns
■ Required tools
■ None
■ Required products
■ Fresh and normal blood samples (5 different
samples at least)
■ Intervention time
■ 15 min
■ Frequency
■ On request
1. Calibration
Technical Manual
RAS304C - 2 RAA016CEN
LMNE balance adjustment
If the variation between the previous and the new coefficients exceeds 20%, the calibration
fails.
LMNE Balance coefficient range: 0,5 to 1,5
Technical Manual
RAA016CEN RAS304C - 3
LMNE balance adjustment
2. Forced calibration
Technical Manual
RAS304C - 4 RAA016CEN
RAS680A
■ Concerns
■ Required tools
■ Hexagonal keys
■ Cruciform screw driver
■ Required products
■ None
■ Intervention time
■ 15 mn
■ Frequency
■ On request
1. Purpose
Because of Takasago valve, it could be more difficult to access to the chambers or to the heater
assembly.
Follow the first part (“2.Chamber access, page 2”) of this procedure to remove the Takasago valves for
maintenance on those assemblies.
Follow the second part of this procedure (“3.Valve assembly replacement, page 3”) for the Takasago
valve assembly replacement or for one valve dismantling only.
The second part of the procedure has to be used in case of failure of the Takasago valve
assembly only.
2. Chamber access
2.1. Disassembly
■ Turn off the instrument and disconnect the power supply cable.
Technical Manual
RAS680A - 2 RAA016CEN
Takasago valve assembly dismantling
2.2. Reassembly
■ Install
the valve plate on the crosspieces
using the 2 CHC screws.
■ Turn off the instrument and disconnect the power supply cable.
Technical Manual
RAA016CEN RAS680A - 3
Takasago valve assembly dismantling
■ Reassemble in reverse order. Take care to the tubing position (See “2.2. Reassembly, page 3”).
Technical Manual
RAS680A - 4 RAA016CEN
Takasago valve assembly dismantling
Technical Manual
RAA016CEN RAS680A - 5
Takasago valve assembly dismantling
Technical Manual
RAS680A - 6 RAA016CEN
Exploded views
8 - Exploded views
1. List ...........................................................................................................................2
2. Valves ......................................................................................................................6
3. Takasago valve assembly .......................................................................................7
4. 5 Diff syringe ...........................................................................................................8
5. Sampling syringe .....................................................................................................9
6. Reagents syringe ..................................................................................................10
7. Waste syringe ........................................................................................................11
8. Counting syringe ...................................................................................................12
9. Diluent tank assy ...................................................................................................13
10. Rinse block & needle ..........................................................................................14
11. Carriage assy ......................................................................................................15
12. Carriage peripherals ............................................................................................16
13. Optical bench ......................................................................................................17
14. Chambers for instrument NON equipped with Takasago valves ........................18
15. Chambers for instrument equipped with Takasago valves .................................19
16. Chamber parts ....................................................................................................20
17. Thermostated room ............................................................................................21
18. Electronic boards ................................................................................................22
19. Straws & tubes ....................................................................................................23
20. Reagent connection assembly ............................................................................24
21. Miscellaneous .....................................................................................................24
22. Power supply ......................................................................................................25
23. Covers .................................................................................................................26
Technical Manual
RAA016CEN 8-1
Exploded views
List
1. List
N° in
Drawing Reference Designation
drawing
“2.Valves, p 6” 1 XDB123A VALVE, LIQ 11 VALVE ASSY P60 (1 to 11)
“2.Valves, p 6” 2 XDB124A VALVE, LIQ 5 VALVE ASSY P60 (12 to 16)
“2.Valves, p 6” 3 XDB125A VALVE, LIQ 2 VALVE ASSY P60 (17 to 18)
“2.Valves, p 6” 4 XDB126A VALVE, LIQ 7 VALVE ASSY P60 (20 to 26
“2.Valves, p 6” 5 XDB127A VALVE, LIQ 5 VALVE ASSY P60 (27 to 31)
“2.Valves, p 6” 6 XDB135A 2W VALVE WITHOUT COIL
“2.Valves, p 6” 7 XDB136A 3W VALVE WITHOUT COIL
“3.Takasago valve assembly, p 7” 1 XBA685A VALVE, TAKASAGO + MANIFOLD ASSY
“3.Takasago valve assembly, p 7” 2 XBA646AS TAKASAGO VALVE
“3.Takasago valve assembly, p 7” 3 EZZ003A TAKASAGO VALVE MANIFOLD
“3.Takasago valve assembly, p 7” 4 XEA960AS KIT, TAKASAGO KIT O'RING + SCREW
“3.Takasago valve assembly, p 7” 5 XEA959AS TAKASAGO WIRING KIT
“4.5 Diff syringe, p 8” 1 GBG037A SYRINGE, 5DIFF BLOC BODY P60/80
“4.5 Diff syringe, p 8” 2 FAA040A O'RING, 5DIFF SYRINGE D=12,1
“4.5 Diff syringe, p 8” 3 FAA067A O'RING, 5DIFF SYRINGE D=2,4
“4.5 Diff syringe, p 8” 4 GBG042A SYRINGE, CROSSPIECE P60/SPS/P80
“4.5 Diff syringe, p 8” 5 GBG040A SYRINGE, 5DIFF PISTON P60/P80
“4.5 Diff syringe, p 8” 6 XDA616AS NEEDLE, 190µL SYR. P60/SPS/P80
“4.5 Diff syringe, p 8” 7 XDA591AS SYRINGE, 5 DIFF SYRINGE ASSY P60/P80
“4.5 Diff syringe, p 8” 8 XDB034AS SYRINGE, 5DIFF BLOCK+MOTOR ASSY
“4.5 Diff syringe, p 8” 9 XDA826AS MOTOR, BLOCK 5DIFF SYR. P60/P80
“4.5 Diff syringe, p 8” 10 XBA548AS MOTOR, FOR 5 DIFF SYRINGE
“4.5 Diff syringe, p 8” 11 GBG038A SYRINGE, 5DIFF BOTTOM PLATE
“4.5 Diff syringe, p 8” 12 XBA462A CABLE, SYR. MOTORS L=300 P60/P80
“5.Sampling syringe, p 9” 1 GBG044A SYRINGE, SAMP. SYR. BODY P60/P80
“5.Sampling syringe, p 9” 2 FAA064A O’RING, SAMPLING SYR. P60/P80
“5.Sampling syringe, p 9” 3 GBG048A SYRINGE, SAMPLING CROSSPIECE
“5.Sampling syringe, p 9” 4 XDA617AS NEEDLE, 100 µL SAMPL. SYR. P60/P80
“5.Sampling syringe, p 9” 5 XDA593AS SYRINGE, SAMPLING SYRINGE ASSY
“5.Sampling syringe, p 9” 6 XDA834AS SYRINGE, SAMPL BLOCK+MOTOR ASSY
“5.Sampling syringe, p 9” 7 XDA824AS MOTOR, BLOCK SAMPL. SYR.P60/P80
“5.Sampling syringe, p 9” 8 XDA547AS MOTOR, FOR SAMPLING SYRINGE
“5.Sampling syringe, p 9” 9 GBG045A SYRINGE,SAMPLING BOTTOM PLATE
“5.Sampling syringe, p 9” 10 XBA462A CABLE, MOTORS L=300 P60/P80
“6.Reagents syringe, p 10” 1 XDB001AS SYRINGE,REAGENT ASSY P60/P80
“6.Reagents syringe, p 10” 2 GBG033A SYRINGE, REAG BLOCK BODY P60/P80
“6.Reagents syringe, p 10” 3 XDA622A O’RING, REAGENT SYRINGE + WASHER
“6.Reagents syringe, p 10” 4 GBC030A SYRINGE, REAGENT PISTON P60/P80
“6.Reagents syringe, p 10” 5 GBC031A SYRINGE, LYSE PISTON MIC/P60/P80
“6.Reagents syringe, p 10” 6 FAA065A O’RING, REAGENT SYRINGE D=6.3+
“6.Reagents syringe, p 10” 7 XDB005A SYRINGE, REAGENT+MOTOR ASSY P80
Technical Manual
8-2 RAA016CEN
Exploded views
List
N° in
Drawing Reference Designation
drawing
“6.Reagents syringe, p 10” 8 XDB004AS MOTOR, BLOCK FOR REAG. SYR. P80
“6.Reagents syringe, p 10” 9 XBA390A MOTOR, SYRINGE EQUIPPED P80
“6.Reagents syringe, p 10” 10 GBG034A SYRINGE,5DIFF BLOCK BOTTOM P60
“6.Reagents syringe, p 10” 11 GBG035A SYRINGE,REAGENT HOLDER PLATE
“7.Waste syringe, p 11” 1 XDA827BS SYRINGE, WASTE CPTE P60/C+/P80
“7.Waste syringe, p 11” 2 CAE010A SWITCH, MICROSWITCH XCC5-81
“7.Waste syringe, p 11” 3 FAA017A O'RING, TANK MIN/AG+WASTE MIC
“7.Waste syringe, p 11” 4 GBG219A O'RING, ANTI EXTRUSION RING
“7.Waste syringe, p 11” 5 GBG272A SYRINGE,VACCUM PUMPBODY P60/80
“7.Waste syringe, p 11” 6 FAL009A SILENT BLOCK, FOR SUB ASSEMBLIES
“7.Waste syringe, p 11” 7 XBA458A CABLE,SYR.SWITCH L=300 P60/P80
“7.Waste syringe, p 11” 8 XBA462A CABLE,MOTORS L=300 P60/P80
“7.Waste syringe, p 11” 9 (3+4) XDA621A O’RING, VAC/WASTE PUMP + WASHER
“8.Counting syringe, p 12” 1 XDA828BS SYRINGE, VACUUM CPTE P60/C+/P80
“8.Counting syringe, p 12” 2 CAE010A SWITCH, MICROSWITCH XCC5-81
“8.Counting syringe, p 12” 3 FAA017A O'RING, TANK MIN/AG+WASTE MIC
“8.Counting syringe, p 12” 4 GBG219A O'RING, ANTI EXTRUSION RING
“8.Counting syringe, p 12” 5 GBG273A SYRINGE,VACCUM PUMPBODY P60/80
“8.Counting syringe, p 12” 6 FAL009A SILENT BLOCK, FOR SUB ASSEMBLIES
“8.Counting syringe, p 12” 7 XBA458A CABLE,SYR.SWITCH L=300 P60/P80
“8.Counting syringe, p 12” 8 XBA462A CABLE,MOTORS L=300 P60/P80
“8.Counting syringe, p 12” 9 (3+4) XDA621A O’RING, VAC/WASTE PUMP + WASHER
“9.Diluent tank assy, p 13” 1 XDA626AS CHAMBER, DIL.TANK COVER P60/P80
“9.Diluent tank assy, p 13” 2 GAL094A CHAMBER, WAST.P120/DIL TANK P60
“9.Diluent tank assy, p 13” 3 XEA286AS KIT, O'RING + WASHER P60/P80/P120
“9.Diluent tank assy, p 13” 4 XDA605A CHAMBER, DIL. TANK ASSY P60/P80
“10.Rinse block & needle, p 14” 1 XDA619BS NEEDLE, SAMPLING NEEDLE
“10.Rinse block & needle, p 14” 2 GBG091A CARRIAGE, NEEDLE GUIDE P60
“10.Rinse block & needle, p 14” 3 FAA053A O'RING, SAMPL. NEEDLE
“10.Rinse block & needle, p 14” 4 GBG090A CARRIAGE, RINSING BLOCK P60
“11.Carriage assy, p 15” 1 XDA599AS KIT, CARRIAGE ASSY P60 OT
“11.Carriage assy, p 15” 2 FBR011A BELT, SYRINGE BLOCK 5DIFF/P120
“11.Carriage assy, p 15” 3 XBA396A SENSOR, SAMP NEEDLE MOVE P60/80
“11.Carriage assy, p 15” 4 XBA392A MOTOR, SAMPLING NEEDLE PENTRA60
“11.Carriage assy, p 15” 5 XDA618AS CARRIAGE, NEEDLE CARRIAGE P60
“11.Carriage assy, p 15” 6 DBE014A CABLE, BUSHING D=3,2 BLACK
“11.Carriage assy, p 15” 7 GBG081A NEEDLE, BLOCK SHAFT
“11.Carriage assy, p 15” 8 GBG093A MOTOR, PULLEY P60
“11.Carriage assy, p 15” 9 FBR012A BELT, CARRIAGE L=544 P60
“12.Carriage peripherals, p 16” 1 XBA391B MOTOR, CARRIAGE P60 ACT5D OV&CP
“12.Carriage peripherals, p 16” 2 GBG078A CARRIAGE, HORIZONTAL SHAFT D7.9
“12.Carriage peripherals, p 16” 3 GBG077A CARRIAGE, HORIZONTAL SHAFT D=8
“12.Carriage peripherals, p 16” 4 XBA394A SENSOR, CARRIAGE MOVE PENTRA60
“12.Carriage peripherals, p 16” 5 GBG093A MOTOR, PULLEY P60
Technical Manual
RAA016CEN 8-3
Exploded views
List
N° in
Drawing Reference Designation
drawing
“13.Optical bench, p 17” 1 DAJ007A OPTICAL, LAMP FOR BENCH P60/P80
“13.Optical bench, p 17” 2 FAL010A AMORTISSEUR HEYMANWELL IBM6047
“13.Optical bench, p 17” 3 XDA601CS CHAMBER, LMNE FLOWCELL P60/P80
“13.Optical bench, p 17” 4 XBA399C FITTING, GROUND FLOWCELL P60/80
“13.Optical bench, p 17” 5 XBA490A CABLE, LMNE FLOWCEL CATHOD WIRE
“13.Optical bench, p 17” 6 XBA403A TUBING, SHIELD REPARTITOR P60
“13.Optical bench, p 17” 7 XEA616BS KIT, FLOWCELL INJECTOR TUBE P60
“13.Optical bench, p 17” 8 XBA193A CABLE, COAX BASO-O.D LMNE-LMNE
“13.Optical bench, p 17” 9 XEA663A CABLE, COAX LMNE FLOWCEL P60/80
“13.Optical bench, p 17” 10 XDA600AS OPTICAL, BENCH COMPLETE P60
“14.Chambers for instrument NON
1 XDB035C 5 CHAMBERS BLOCK ASSY EQUIPED
equipped with Takasago valves, p 18”
“14.Chambers for instrument NON
2 XDB036BS 4 CHAMBERS BLOCK ASSY EQUIPED
equipped with Takasago valves, p 18”
“14.Chambers for instrument NON
3 XDB019BS WBC/BASO CHAMBER EQUIPED
equipped with Takasago valves, p 18”
“14.Chambers for instrument NON
4 XBA620A SENSOR, CHAMBERS DRAINING P60
equipped with Takasago valves, p 18”
“15.Chambers for instrument equipped
1 XDB021E 5 CHAMBERS BLOCK ASSY EQUIPED
with Takasago valves, p 19”
“15.Chambers for instrument equipped
2 XDB017ES 4 CHAMBERS BLOCK ASSY EQUIPED
with Takasago valves, p 19”
“15.Chambers for instrument equipped
3 XDB019BS WBC/BASO CHAMBER EQUIPED
with Takasago valves, p 19”
“15.Chambers for instrument equipped
4 XDB250AS 4 CHAMBERS BLOCK ASSY + COUNTING HEAD
with Takasago valves, p 19”
“15.Chambers for instrument equipped
5 XBA620A SENSOR, CHAMBERS DRAINING P60
with Takasago valves, p 19”
“16.Chamber parts, p 20” 1 XDB021E 5 CHAMBERS BLOCK ASSY
“16.Chamber parts, p 20” 2 GBG326A WBC/BASO CHAMBER NAKED
“16.Chamber parts, p 20” 3 GBG275A FLAT JOINT, COUNTING HEAD
“16.Chamber parts, p 20” 4 FAA066A O’ RING, DRAINING CHAMBER
“16.Chamber parts, p 20” 5 FAK001A APERTURE 50µ
“16.Chamber parts, p 20” 6 GBG003A DRAINING, 1/4 TURN
“16.Chamber parts, p 20” 7 GBG007A CHAMBER, DRAINING CHICANE
“16.Chamber parts, p 20” 8 GBG157A CHAMBER, COUNTING HEAD
“16.Chamber parts, p 20” 9 XBA398B COAXIAL RBC & WBC PENTRA 60 CE
“16.Chamber parts, p 20” 10 FAK003A APERTURE 80µ
HGB PHOTOMETER (HGB PHOTOCELL ALONE:
“16.Chamber parts, p 20” 11 + 11* XBA389A
CAX005A)
“17.Thermostated room, p 21” 1 XDB032BS HEATER, BLOCK COMPLETE P60/ACT
“17.Thermostated room, p 21” 2 XCA166A CHAMBER, ISOLATOR (LONG)
“17.Thermostated room, p 21” 3 GBC015A CLIP, MIX CH. HOLDER MIC/P120
“17.Thermostated room, p 21” 4 GBG013A CUP, OVERFLOW T° ROOM
“17.Thermostated room, p 21” 5 GBG205A PROTECTION, FOR VALVES T°ROOM
“17.Thermostated room, p 21” 6 CAE006A SWITCH, MICROSWITCH XC5-81-82
Technical Manual
8-4 RAA016CEN
Exploded views
List
N° in
Drawing Reference Designation
drawing
“17.Thermostated room, p 21” 7 XDA812A KIT, T° ROOM FAN + HARNESS P60
“17.Thermostated room, p 21” 8 XEA906AS KIT, INSULATING PADS P60/ACT
“18.Electronic boards, p 22” 1 XAA425C PCB, MAIN BOARD PENTRA60 C+
“18.Electronic boards, p 22” 2 XAA423BS PCB, PREAMP+MECA ASSY PENTRA60
“18.Electronic boards, p 22” 3 XAA427BS PCB, LINK BOARD PENTRA60
“18.Electronic boards, p 22” 4 XAA429AS PCB, LED BOARD FOR COVER P60/80
“19.Straws & tubes, p 23” 1 GBG145A REAGENT, STRAW D=20 P60/P80
“19.Straws & tubes, p 23” 2 GBG155A REAGENT, BOTTLE STOPPER D=25
“19.Straws & tubes, p 23” 3 GBG245A REAGENT, STRAW D=28 P60/P80/ACT
“19.Straws & tubes, p 23” 4 GAK302A REAGENT, BOTTLE STOPPER D=40
“19.Straws & tubes, p 23” 5 XDB119A TUBING, DILUENT ASPIRATION P60
“19.Straws & tubes, p 23” 6 XDA693A TUBING, WASTE P60/CRP200
“19.Straws & tubes, p 23” 7 EAC018A FITTING, LUER MALE/FEMALE I=2,5
“19.Straws & tubes, p 23” 8 XEA018A REAGENT, STRAW DILUENT L=360
“20.Reagent connection assembly, p 24” 1 XDA623AS REAGENT, BLOCK EQUIPED P60/P80
“20.Reagent connection assembly, p 24” 2 EAC008A FITTING, ANTI ROTATION WASHER
“20.Reagent connection assembly, p 24” 3 EAC010A FITTING, LUER FEMALE I=3MM
“21.Miscellaneous, p 24” 1 XBA393A FAN, MAIN FAN 24V P60/P80/P400
“21.Miscellaneous, p 24” 2 XBA509B PCB, BARCODE HAND READER
“22.Power supply, p 25” 1 DBN004C PCB, POWER SUPPLY BLOCK P60
“23.Covers, p 26” 1 XDA644A COVER, LEFT SIDE EQUIPPED P60
“23.Covers, p 26” 2 GBG130A COVER, TOP COVER P60
“23.Covers, p 26” 3 XDA624BS DOOR, FOR T° ROOM EQUIPPED P60
“23.Covers, p 26” 4 XDA628BS COVER, FRONT COVER EQUIPPED P60
“23.Covers, p 26” 5 HAW008A STICKER, REAGENTS NAMES P60
“23.Covers, p 26” 6 HAX0312 STICKER, REAGENT DOOR P60
“23.Covers, p 26” 7 HAX0310 STICKER, SCREEN & KEYBOARD P60
“23.Covers, p 26” 8 FZZ024A DOOR, PUSH PUSH P60
“23.Covers, p 26” 9 XD629AS COVER,REAGENT DOOR EQUIPED P60
Technical Manual
RAA016CEN 8-5
Exploded views
Valves
2. Valves
1 5
6
7
Technical Manual
8-6 RAA016CEN
Exploded views
Takasago valve assembly
3
4
Technical Manual
RAA016CEN 8-7
Exploded views
5 Diff syringe
4. 5 Diff syringe
7 1
8
2
3 x4
4 x2
12
10 5
11
6 x2
Technical Manual
8-8 RAA016CEN
Exploded views
Sampling syringe
5. Sampling syringe
2 x2
8
4
10
Technical Manual
RAA016CEN 8-9
Exploded views
Reagents syringe
6. Reagents syringe
2
11
3
x4
6
4
8 x4
7 5
10
Technical Manual
8 - 10 RAA016CEN
Exploded views
Waste syringe
7. Waste syringe
2 9
3
4
6 x4
Technical Manual
RAA016CEN 8 - 11
Exploded views
Counting syringe
8. Counting syringe
2 9
3
4
6 x4
Technical Manual
8 - 12 RAA016CEN
Exploded views
Diluent tank assy
Technical Manual
RAA016CEN 8 - 13
Exploded views
Rinse block & needle
1 2 3 4
Technical Manual
8 - 14 RAA016CEN
Exploded views
Carriage assy
1
2
6
9
7
8
Technical Manual
RAA016CEN 8 - 15
Exploded views
Carriage peripherals
4 5
Technical Manual
8 - 16 RAA016CEN
Exploded views
Optical bench
10
2 x4
7
3
4
8
6
5
Technical Manual
RAA016CEN 8 - 17
Exploded views
Chambers for instrument NON equipped with Takasago valves
1 2
Technical Manual
8 - 18 RAA016CEN
Exploded views
Chambers for instrument equipped with Takasago valves
1 2
Technical Manual
RAA016CEN 8 - 19
Exploded views
Chamber parts
9 x2
2
3 x4
11
8 x2
10
4
x4
7
x4
1 6 x4 11*
Technical Manual
8 - 20 RAA016CEN
Exploded views
Thermostated room
1
8 3
5 4
Technical Manual
RAA016CEN 8 - 21
Exploded views
Electronic boards
4
1
Technical Manual
8 - 22 RAA016CEN
Exploded views
Straws & tubes
3
2
5
6
Technical Manual
RAA016CEN 8 - 23
Exploded views
Reagent connection assembly
2
x2
3
1
x2
21. Miscellaneous
1
2
Technical Manual
8 - 24 RAA016CEN
Exploded views
Power supply
Technical Manual
RAA016CEN 8 - 25
Exploded views
Covers
23. Covers
3 1
5 4
9 8
Technical Manual
8 - 26 RAA016CEN