RBC Apperture Problem Dymind

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

Dymind Global Customer Service Center QA-F00XX

0xB1000506-RBC aperture voltage abnormal

Use Background

When analyzer gives a false alarm - RBC aperture voltage abnormal.

Mechanism: The RBC aperture voltage is out of the allowable range [1.65, 2.25] V, which can be
checked in Self-test > Other self-test > RBC aperture voltage.

Declaration

20/8/2020, V1.0, For reference only.

Reference Standard

Model Possibility Solution

Check the diluent box, the syringe, the valve


connected to syringe. Replace LV2, LV5, LV6,
isolation chamber, LV11 respectively to
locate it.

1. There’s no diluent in the RBC


chamber or unstable liquid level.
DH31/DH33/DH
36/DH36VET

2. Some bubbles filled in the tube Check the tube, connectors and valves

near RBC aperture. connected to RBC aperture. Replace LV1, LV7,


the tube connected, and the connectors

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

respectively to locate it.

Replace another box diluent.

3. Diluent get contaminated or


expired.

Replace the RBC aperture or chamber.

4. The RBC aperture damaged.

Replace the RBC chamber.

5. The electrode of RBC chamber


damaged.

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

Replace the main board together with the SD,


6. The main board damaged. keep the software from main board and SD
card the same, otherwise it will not work.

Check the diluent box, the syringe, the valve


connected to syringe. Replace LV1, LV5, LV7,
isolation chamber, LV13 to locate it.

1. There’s no diluent in the RBC


chamber or unstable liquid level.

DF50/DF51/DF5
2/DF50VET
Check the tube, connectors and valves
connected to RBC aperture. Replace LV9,
LV18, the tube connected, and the
connectors respectively to locate it.

2. Some bubbles filled in the tube


near RBC aperture.

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

Replace another box diluent.

3. Diluent get contaminated or


expired.

Replace the RBC aperture or chamber.

4. The RBC aperture damaged.

Replace the RBC chamber.

5. The electrode of RBC chamber


damaged.

Replace the main board together with the SD,


6. The main board damaged. keep the software from main board and SD
card the same, otherwise it will not work.

Check the diluent box, the syringe, the valve


1. There’s no diluent in the RBC
DH73/DH76 connected to syringe. Replace LV4, LV7, LV8,
chamber or unstable liquid level.
isolation chamber, LV16, LV26 to locate it.

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

Replace LV6, LV24, LV28, the tube connected,


and the connectors respectively to locate it.

2. Some bubbles filled in the tube


near WBC aperture.

3. Diluent get contaminated or Replace another box diluent.

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

expired.

Replace the RBC aperture or chamber.

4. The RBC aperture damaged.

Replace the RBC chamber.

5. The electrode of RBC chamber


damaged.

Replace the main board, and keep the


6. The main board damaged. software of the machine (host) the same
with IPU, otherwise it will not work.

Relative spare part code

Model Part Name P/N Picture

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

Three-way valve (L) - SMC 23.99.0002A

Two-way valve (L) - SMC 23.99.0001A

Three-way valve (S) - SMC 23.99.0004A


DH31/DH33/DH36/DH36VET,

DF50/DF51/DF52/DF50VET,

DH73/DH76

Two-way valve (S) - SMC 23.99.0003A

RBC aperture 24.11.0002A

T420-1 Connector 53.05.0003A

DH31/DH33/DH36/DH36VET,
SD card 34.01.0030A
DF50/DF51/DF52/DF50VET

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

RBC bath assembly(with


20.01.0144A
Ruby)

Isolation Chamber with


Strainer(Upper port 68.02.0022A
blocked)

DH31/DH33/DH36/DH36VET Mainboard 22.01.0023A

DF50/DF51/DF52/DF50VET Mainboard 22.01.0021C

Mainboard 22.01.0001C

Isolation Chamber with


DH73/DH76 60.01.0012C
Strainer

RBC bath assembly(with


20.01.0008A
Ruby)

Dymind Global Customer Service Center Roy


Dymind Global Customer Service Center QA-F00XX

Y230-1 Connector 53.05.0011A

If you have any problem with this file, please contact our engineer ASAP, thank you!