DETECTORS
Nitrogen/Phophorus Detector - NPD
Theory of Operation
l capnut
NPD heater —.
block:
HPO baad tt fats bo
ionization area of the detector body)
Electrical leads provide current to the NPD—
bbead to maintain the hydrogen-air plasma
Rrea where
ionization
takes place
NPDIet Detector viewport
JB. — colector electrode
Positively charged
Electrical lead
toampliier
Inside the NPD detector body, an electrically heated thermionic bead (NPD bead) is positioned between
the jt orifice andthe collector electrode. The bead is coated with an alkali metal which promotes the ionization
of compounds that contain nitrogen or phosphorus, Hydrogen and air lows ereate a hydrogen plasma around,
thehot NPD bead. When molecules containing nitrogen or phosphorus enter the plasma from the column and
jetori
they undergo a catalytic surface chemistry reaction, producing thermionic electrons, Therresulting
ions are attracted toa positively charged collector electrode, then amplified and output to the data system. The
hydrogen to ar ratio is too lean to sustain aflame, therefore minimizing hydrocarbon ionization and contributing,
tothe NPD detector’s selectivity.
Detector viewport
Hydrogen
‘Sample-laden carier gas flowing plasma, Positively charged collector
from the column to the jet Stainless steo! jet electrode doubles as the
‘exhaust vent
io
Thermionic NPO bead
N
a
Compressed airfoeds
the nyerogen fame
4291421, 538 PM
DETECTORS
Nitrogen/Phophorus Detector - NPD
Expected Performance
‘The following chromatograms are from an SRI GC equipped with NPD, DELCD and FID detectors
Sincethe NPD isnot the only detector, helium carrier gas was used instead of nitrogen. The first chromatogram,
showsa separation of 200ppm Organophosphorus pesticide standard, Mix 8270. The second chromatogram,
shows both the NPD and DELCD responses to a mixture of 100ppm Mix 8270 and 100ppm Organochilorine
pesticide standard, Mix 8081. Other than the sample and length of time, the analytical parameters were the
same for both runs. The NPD has a much smaller response to the Organochlorine standard (Mix 8081).
Since the DELCD is selective to chlorinated molecules, its response to Mix 8081 supplements the NPD
response for better identification. and quantification.
Sample: ipl 200ppm |OsUeiS \HLUL[v?r sss ealnnny
COrganophophorus standard 7
rmx8270 zm
Method: dirctinjection 0-37 minutes
Column: 60-meter MXT-VOL
Cari: helium at Om inin
NPD gain: HIGH
NPD temperature: 250°C
NPD bead current: -370
Injector temperature: 150°C
Temperature program:
Initial Hold Ramp Final
1000 600 1500 2e0C
280°C 20.00 0.00 280°C
Sample: 1yL_100ppm
‘Organophophorus pesticide
standard mix 8270 +
100ppm Organochlorine
pesticide standard mix 8081
Method: direct injection
Injector temperature: 150°C
‘Column: 60-meterMXT-VOL,
‘Cari: helium at 1OmLfmin
NPD gain: HIGH
NPD temperature: 250°C
NPO bead current:-370
DELCO gain: LOW
DELCO reactor: 260
DELCO temp: 300°C
a
@ peLcp
22