Maintenance Schedule DH76

You might also like

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 1

Daily/weekly/monthly/as needed maintenance plan

Operator: month: Inst. Model : DH76 S/N:

DATE 1 2 3 4 5 6 7 8 9 1 1 1 1 1 1 1 1 1 1 2 2 2 2 2 2 2 2 2 2 3 31
0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6 7 8 9 0

DAILY
MAINTENANC
E

Check reagents

Check waste
container

Check
Backgrounds

Perform clean
WBC

Perform clean
RBC

Perform clean
Flow chamber

Perform clean
diff. path

Perform clean
sample probe

Perform
cleanser soak

Weekly
Maintenance

WBC Channel
Cleanser soak

RBC Channel
Cleanser soak

Check
WBC/RBC/DIFF
. chambers
area for any
calcification
and clean as
needed

Clean sample *Remove dust, wipe with 10%


probe outside sodium hypochlorite.Then wipe
surface with D. Water.

Clean input & Remove dust, wipe with 10%

output rack sodium hypochlorite.Then wipe


area with D. Water.

Monthly
Maintenance

Clean and *Remove dust, wipe with 10% sodium hypochlorite.


Decontaminate
Then wipe with D. Water.
analyser outer
surfaces.

As needed
Maintenance

Fluidics NEEDED AFTER MAINTENANCE OR CLEANING PROCEDURES DONE ON FLUIDICS SYSTEM INSIDE ANALYZER
initialization

Clean fluidics NEEDED WHEN BACKGROUNDS RESULTS OUT OF RANGE.

Empty fluidics NEEDED WHEN: 1-ANALYZER WILL NOT USED FOR 3- 10 DAYS 2- INCORRECT REAGENTS USED.

*Backgrounds Reference ranges: •WBC ≤ 0.2 •RBC ≤ 0.02 •HGB ≤ 1 •PLT ≤ 10