Specifications for Environment test chamber with temperature and humidity control

(Note: It is mandatory for the bidders to provide the compliance statement (comply/not comply)
for the below points with document proof as required)
1. Stainless steel interior (chambers) and exterior
a) Internal chamber volume : 250 l or larger
b) Number of grids – 1; with a provision to attached 5 additional grids in future
c) Door – stainless steel – with provision of condensation prevention mechanism such as
heating
2. Temperature
a) Temperature resolution: lower than 0.5 °C
b) Temperature range, when humidity is not being controlled: -40 °C to 180 °C
c) Temperature range, when humidity is being controlled: 10 °C to 95 °C
d) Spatial variation of temperature inside chamber: maximum of ± 2 °C
e) Minimum two sensors for temperature measurement
f) 3 point calibration
3. Humidity
a) Humidification and dehumidification – both should be controlled
b) Humidity range : 10-98 % RH
c) RH set point control: 1 %
d) Condensate collection mechanism
e) Compressor-refrigeration systems should be environment friendly and with chlorine-free
refrigerant
4. Control of humidity/temperature
a) Given that temperature and humidity have to be controlled simultaneously, provide the
overall chart of achievable envelope such that
 10% RH should be possible at 45 °C or lower
 95% RH should be possible at 15 °C or lower
 at 20 °C, 40% RH or lower should be possible
b) Microprocessor based control
 programmable profiles for both humidity and temperature
 10 ramps or higher number
c) 3 point calibration
d) Storage of data for 6 months or more
5. Ventilation with air convection
6. Computer interfacing for collection of humidity/temperature data, as well as programming
7. Safety provisions
8. Conformation to standards
a) ASTM304 for interior of the chamber
b) DIN 12880:2007-05 for temperature variation
c) DIN 12880 for temperature safety
9. Documentation/technical support
a) Chart of achievable envelope in temperature-humidity axes
b) Standard factory calibration certificate
c) Installation, commisioning and training to be provided
d) Operation manual
e) All the supporting documents, such as necessary catalogues and product data sheet
f) Standrad warranty and maintenance contract for 3 years from installation

Point wise technical compliance statement

Quoting for optional accessories is mandatory.


Note: If the compliance statement (comply/Not comply) is not furnished for the evaluation.