Professional Documents
Culture Documents
Abx (P60)
Abx (P60)
Abx (P60)
Technical Manual
P/n: RAA016B
ABX DIAGNOSTICS
B.P. 7290
Rue du caducée
Parc Euromédecine
34184 MONTPELLIER Cedex 04 - FRANCE
• REVISIONS
• CONTENTS
1. SPECIFICATIONS
2. HYDRAULIC & PNEUMATIC PRINCIPLES
3. ELECTRIC & ELECTRONIC PRINCIPLES
4. ANALYSIS CYCLE TECHNOLOGY
5. SOFTWARE VERSION
6. OUTPUT FORMAT
7. ALARMS & ERROR LIST
8. MAINTENANCE PROCEDURES
9. TRAINING SLIDES
10. BLOWUPS VIEWS
11. SPARE PART LIST
Technical Manual RAA016 B
SPECIFICATIONS
page 3/8
Technical Manual RAA016 B
3. Connections .......................................................................................................... 12
3.1. Chamber Heating Connector XBA386A ................................................................................ 12
3.2. Reagent Heating Coil Connector XBA387A .......................................................................... 13
3.3. Horizontal Carriage Motor XBA391A ..................................................................................... 14
3.4. Upper Fan XBA393A ........................................................................................................... 15
3.5. IR sensor XBA394AS ........................................................................................................... 16
3.6. IR sensor XBA395AS ........................................................................................................... 17
3.7. IR sensor XBA396AS ........................................................................................................... 18
3.8. IR sensor XBA397AS ........................................................................................................... 19
3.9. RBC / WBC coaxe XBA398A ............................................................................................... 20
3.10. LMNE flowcell coaxe XEA663A (XBA399B+XBA490A) ....................................................... 21
3.11. Barcode reader cable XBA402AS ...................................................................................... 22
3.12. Diluent level sensor XDA605AS ......................................................................................... 23
page 4/8
Technical Manual RAA016 B
5-SOFTWARE VERSIONS
1. Software Overview: User Menu Overview .......................................................... 2
6-OUTPUT FORMAT
page 5/8
Technical Manual RAA016 B
1. Printer ................................................................................................................. 2
2. Transmission ...................................................................................................... 2
3. Calibration .......................................................................................................... 2
5. Reagents ............................................................................................................ 3
6. Temperature ........................................................................................................ 3
8-MAINTENANCE PROCEDURES
1. Maintenance ...................................................................................................... 2
1.1. Introduction ........................................................................................................................ 2
1.2. Daily customer maintenance .............................................................................................. 2
1.3. Weekly customer maintenance .......................................................................................... 2
1.4. Maintenance table ............................................................................................................. 3
3. Procedures ......................................................................................................... 8
3.1. Procedure chart tables ....................................................................................................... 9
3.2. Required tools and products ........................................................................................... 10
page 6/8
Technical Manual RAA016 B
9-TRAINING SLIDES
10-BLOWUPS VIEWS
page 7/8
Technical Manual RAA016 B
page 8/8
SPECIFICATIONS
1. TECHNICAL SPECIFICATIONS
• The ABX PENTRA 60 is a fully automated hematology analyzer used for in vitro diagnostic
testing of whole blood specimens.
• The ABX PENTRA 60 is able to operate either in CBC mode (Cell Blood Count:12 parameters)
or in CBC + 5DIFF mode (5 population Differential count: 26 parameters).
1.1. Parameters
Plt Platelets
PDW * Platelet Distribution Width
MPV Mean Platelet Volume
Pct * Plateletcrit
* PDW and PCT have not been established as indications for this product,in the Uni-
ted States. The use of PCT and PDW should be restricted to research and
Investigational measurements only.
page 2/20
1. Specifications RAA016B
Plt Platelets
PDW * Platelet Distribution Width
MPV Mean Platelet Volume
Pct * Plateletcrit
* PCT, PDW, ALY and LIC have not been established as indications for this product,in
the United States. The use of PCT, PDW, ALY and LIC should be restricted to research
and Investigational measurements only.
page 3/20
1. Specifications RAA016B
1.1.3. Units
UNITS
WBC RBC HGB HCT PLT MCV
STD 103/mm3 106/mm3 g/dl % 103/mm3 µm3
SI 109/l 1012/l g/l l/l 109/l fl
mmol/l 109/l 1012/l mmol/l l/l 109/l fl
page 4/20
1. Specifications RAA016B
1.4. Reagents
• ABX DILUENT (20 litres),
• ABX CLEANER (1 litre, Integrated),
• ABX EOSINOFIX (1litre, Integrated),
• ABX BASOLYSE II (1 litre, Integrated),
• ABX ALPHALYSE or LYSEBIO (0.4 litre, Integrated).
page 5/20
1. Specifications RAA016B
2. PHYSICAL SPECIFICATIONS
• Maximum relative humidity 80% for temperatures up to 31°C (88°F) decreasing linearly to
50% relative humidity at 40°C (104°F).
page 6/20
1. Specifications RAA016B
page 7/20
1. Specifications RAA016B
page 8/20
1. Specifications RAA016B
• Three levels of ABX MINOTROL material (Lot No: JX108) were run in duplicate once daily
for a prolonged period (26th July 2002 to 30 August 2002) on all parameters.
The results were used to quantify within run precision, and Total Precision in accordance
with the NCCLS EP 5-A Guidelines.
page 9/20
1. Specifications RAA016B
Precision claims*
page 10/20
1. Specifications RAA016B
3.3. Linearity*
• Linearity range: The Manufacturer’s tested linearity zone of the instrument using linearity
kits and/or human blood.
• Linearity limits: Maximum and minimum values within instrument returns no dilution alarm.
• Visible range: Range values given by the instrument. These values (above linearity limits)
are given as an indication. They are given associated with a «D» flag. This Visible range is
outside Manufacturer’s range.
• Linearity kits
Linearity was tested using available «Low Range» and «Full Range» Linearity Test kits. The
Test kits were analyzed and data was computed according to the Manufacturer’s instruc-
tions.
• Human Blood
Linearity was also obtained on human blood, using a minimum of 5 dilution points. The
re-sults of this study are as followed:
3.4. Carryover
• The ABX PENTRA 60 carry-over effects were evaluated by assaying a sample with high cell
concentrations three consecutive times (i1-3), followed immediately by testing a diluted sample
consecutively 3 times (j1-3).
(j1-j3)
Carryover = X 100
(i3-j3)
Carry-over % is then:
This is the method as described in Guidelines for the Evaluation of blood cell analyzers
including those used for differential leukocyte and reticulocyte counting and cell marker
applications. ISLH, 14 January, 1994.
page 11/20
1. Specifications RAA016B
• Carry-over Conclusion:
Results provided are extremely satisfactory. In order to provide for eventual possibilities
within the laboratory environment the following claims shall be made :
IMPORTANT
Expected values will vary with sample population and/or geographical
location. It is highly recommended that each Laboratory establish its
own Normal ranges based upon the local population!
*Bibliography:
AIDE MEMOIRE D’HEMATOLOGIE
Prof : C.SULTAN / M. GOUAULT- HELMANN / M. IMBERT
Service Central d’Hématologie de l’Hopital Henri Mondor
Faculté de médecine de Créteil (Paris XII)
page 12/20
1. Specifications RAA016B
3.6. Accuracy*
• The data shows good correlation between results achieved on the ABX PENTRA 60 versus
the reference system, which can be resumed as follows:
page 13/20
1. Specifications RAA016B
Results displayed
NOTE
Use the instrument diluent to dilute the sample if a ---- D flag occurs
on WBC or Hct.
page 14/20
1. Specifications RAA016B
4. REAGENTS SPECIFICATIONS
NOTE
The ABX Diagnostics reagents specified for this instrument may have
followed one or both of the approval methods below:
1 - have been registered by the A.F.S.S.A.P.S. «Agence Française de
Sécurité Sanitaire des Produits de Santé» according to the procedure
relative to laboratory reagents used for biological analyses.
2 - or approved in accordance with the European Directive 98/79/CE
(Annex III) for in-vitro medical devices.
Please refer to the packaging of each reagent concerned to establish
the approval method.
WARNING
When disposing of waste, protective clothing must be worn (lab coat,
gloves, eye protection, etc…). Follow your local and /or national
guidelines for biohazard waste disposal.
• If required, waste can be neutralized before being discarded. Follow your laboratory’s
protocol when neutralizing and disposing of waste.
• Dispose of the waste container according to the local or national regulatory requirements.
page 15/20
1. Specifications RAA016B
5. LIMITATIONS
WARNING
Whilst every effort is taken by ABX Diagnostics to investigate and
indicate all known interference’s, it is by no means possible to guarantee
that all interference’s have been identified. At all times, results should
be validated and communicated only once all information relating to
the patient has been assessed and taken into account.
5.1. Maintenance
• In Chapter 5. Maintenance & Troubleshooting, specific maintenance procedures are listed.
The maintenance procedures identified are mandatory for proper use and operation of the
ABX PENTRA 60. Failure to execute any of these recommended procedures may result in poor
reliability of the system.
IMPORTANT
Failure to execute any of these recommended procedures may result
in poor reliability of the system.
page 16/20
1. Specifications RAA016B
Unlysed Red Cells - In some rare instances, the erythrocytes in the blood sample may not
be completely lysed. These non-lysed red blood cells may be detected on the WBC histogram
with an L1 alarm or as an elevated baseline on the side (leading edge) of the lymphocytes
population. Non-lysed erythrocytes will cause a falsely elevated WBC count.
Multiple myeloma - The precipitation of proteins in multiple myeloma patients may give
high WBC counts.
Leukemia - A very low WBC count may result from this disease because of possible increased
fragility of the leukocytes leading to destruction of some of these cells during counting.
These white cell fragments will also interfere with the white cell differential parameters.
These white cell fragments will also interfere with the white cell partial differential parameters:
LYM% + #, MON% + #, GRAN% + #. A suspiciously low WBC count may also be seen in
patients with lymphocytic leukemias due to the presence of abnormally small lymphocytes
which may not be counted by the instrument.
Chemotherapy - Cytotoxic and immunosuppressive drugs may increase the fragility of the
leukocytes which may cause low WBC counts.
Agglutinated erythrocytes - May cause a low incorrect RBC count. Blood samples
containing the agglutinated red blood cells may be suspected by elevated MCH and MCHC
values and shown by examination of the stained blood film.
Cold agglutinins - IgM immunoglobulins which are high in cold agglutinin disease may
cause lower RBC and PLT counts and increase MCV.
page 17/20
1. Specifications RAA016B
Hgb (Hemoglobin):
Turbidity of the blood sample - Any number of physiological and/or therapeutic factors
may produce high incorrect HGB results. To obtain accurate hemoglobin results when
increased turbidity of the blood sample occurs, determine the cause of the turbidity and
follow the appropriate method below:
High WBC: An extremely high WBC will cause excessive light scatter. In these cases use
reference (manual) methods.The diluted sample should be centrifuged, and the supernatant
fluid measured with a spectrophotometer.
High lipid concentration: A high concentration of lipids in the blood sample will give the
plasma a “milky” appearance. This condition can occur with hyperlipidemia, hyperproteinemia
(as in gammapathies) and hyperbilirubinemia. Accurate hemoglobin determinations can be
achieved by using reference (manual) methods and a plasma blank.
Increased turbidity may also be seen in cases where the red blood cells are resistant to
lysing. This condition will cause an incorrect high HGB result, but may be detected by
observing the abnormal MCH, MCHC values, and the increased baseline on the leading
edge of the WBC histogram. Erroneous hemoglobin results will cause the results of the
MCH and MCHC to be incorrect as well.
Fetal bloods - The mixing of fetal and maternal bloods may produce a high inaccurate HGB
value.
Hct (Hematocrit)
Red blood cells agglutination - May produce an inaccurate HCT and MCV values. Red
blood cell agglutination may be detected by observing abnormal MCH and MCHC values,
as well as by examination of the stained blood film In such cases, manual methods may be
required to obtain an accurate HCT value.
Excessive numbers of large platelets and/or the presence of an excessively high WBC
count may interfere with the accurate determination of the MCV value. In such cases, careful
examination of the stained blood film may reveal the error.
page 18/20
1. Specifications RAA016B
Nutritional deficiency or blood transfusion - May cause high RDW results due to iron
and/or cobalamin and /or folate deficiency.
Plt (Platelets)
Very small erythrocytes (microcytes), erythrocyte fragments (schizocytes) and WBC frag-
ments may interfere with the proper counting of platelets and cause elevated PLT counts.
Agglutinated erythrocytes - May trap platelets, causing an erroneously low platelet count.
The presence of agglutinated erythrocytes may be detected by observation of abnormal
MCH and MCHC values and by careful examination of the stained blood film.
Giant platelets in excessive numbers - may cause a low inaccurate platelet count as
these large platelets may exceed the upper threshold for the platelet parameter and are not
counted.
Hemolysis - Hemolysed specimens contain red cell stroma which may increase platelet
counts.
A.C.D. blood - Blood anticoagulated with acid-citrate-dextrose may contain clumped platelet
which could decrease the platelet count.
Elevated triglycerides and/or cholesterol: may interfere with correct platelet counting.
Platelet agglutination - Clumped platelets may cause a decreased platelet count and/or a
high WBC count. The specimen should be recollected in sodium citrate anticoagulant to
ensure the anticoagulated character depending on agglutination and reanalyzed only for the
platelet count. The final PLT result must be corrected for the sodium citrate dilution effect.
However, these platelet clumps do trigger flags L1, LL and LL1.
Very small erythrocytes (microcytes), erythrocytic fragments (Schizocytes) and white blood
cell fragments may interfere with the proper counting and sizing of Platelets.
Agglutinated erythrocytes - May trap Platelets, causing an incorrect MPV result. The
presence of agglutinated erythrocytes may be detected by observation of abnormal MCH
and MCHC values and by careful examination of the stained blood film.
IMPORTANT
Blood samples collected in EDTA will not maintain a stable Mean
Platelet Volume. Platelets collected in EDTA swell depending on the
time post-collection and storage temperature.
page 19/20
1. Specifications RAA016B
page 20/20
HYDRAULIC&PNEUMATIC
PRINCIPLES
1. PENTRA 60 HYDRAULIC
Liquid circuits, hydropneumatic connections, as well as the transmission tubes
used, are described in the following chart tables.
PART
DESIGNATION DIAMETER LENGTH QUANTITY
NUMBER
Y Connector EAB021A 3 1
Y Connector EAB026A 2.5 1
T410 Connector EAB033A 1.6 1
T220 Connector EAB035A 2.3 8
0.040" TYGON tubing EAE005A 1.02 10 1
0.040" TYGON tubing EAE005A 1.02 95 1
0.040" TYGON tubing EAE005A 1.02 110 1
0.040" TYGON tubing EAE005A 1.02 130 1
0.040" TYGON tubing EAE005A 1.02 175 1
0.040" TYGON tubing EAE005A 1.02 205 2
0.040" TYGON tubing EAE005A 1.02 240 1
0.051" TYGON tubing EAE006A 1.30 20 1
0.051" TYGON tubing EAE006A 1.30 140 1
0.051" TYGON tubing EAE006A 1.30 320 1
0.060" TYGON tubing EAE007A 1.52 15 1
0.060" TYGON tubing EAE007A 1.52 20 1
0.060" TYGON tubing EAE007A 1.52 50 4
0.060" TYGON tubing EAE007A 1.52 60 1
0.060" TYGON tubing EAE007A 1.52 70 4
0.060" TYGON tubing EAE007A 1.52 80 1
0.060" TYGON tubing EAE007A 1.52 90 1
0.060" TYGON tubing EAE007A 1.52 100 1
0.060" TYGON tubing EAE007A 1.52 120 6
0.060" TYGON tubing EAE007A 1.52 130 1
0.060" TYGON tubing EAE007A 1.52 160 1
0.060" TYGON tubing EAE007A 1.52 200 4
0.060" TYGON tubing EAE007A 1.52 220 1
0.060" TYGON tubing EAE007A 1.52 230 1
0.060" TYGON tubing EAE007A 1.52 250 2
0.060" TYGON tubing EAE007A 1.52 280 1
0.060" TYGON tubing EAE007A 1.52 290 1
0.060" TYGON tubing EAE007A 1.52 300 1
0.060" TYGON tubing EAE007A 1.52 310 1
0.060" TYGON tubing EAE007A 1.52 330 1
0.060" TYGON tubing EAE007A 1.52 335 1
0.060" TYGON tubing EAE007A 1.52 350 1
0.060" TYGON tubing EAE007A 1.52 380 1
page 2/8
2. Hydraulic & Pneumatic principles RAA016B
PART
DESIGNATION DIAMETER LENGTH QUANTITY
NUMBER
0.060" TYGON tubing EAE007A 1.52 400 1
0.060" TYGON tubing EAE007A 1.52 430 1
0.060" TYGON tubing EAE007A 1.52 480 1
0.060" TYGON tubing EAE007A 1.52 500 2
0.060" TYGON tubing EAE007A 1.52 640 1
0.060" TYGON tubing EAE007A 1.52 820 1
0.081" TYGON tubing EAE008A 2.05 20 9
0.081" TYGON tubing EAE008A 2.05 30 2
0.081" TYGON tubing EAE008A 2.05 40 1
0.081" TYGON tubing EAE008A 2.05 55 1
0.081" TYGON tubing EAE008A 2.05 60 1
0.081" TYGON tubing EAE008A 2.05 65 2
0.081" TYGON tubing EAE008A 2.05 70 2
0.081" TYGON tubing EAE008A 2.05 100 1
0.081" TYGON tubing EAE008A 2.05 100 1
0.081" TYGON tubing EAE008A 2.05 120 5
0.081" TYGON tubing EAE008A 2.05 125 1
0.081" TYGON tubing EAE008A 2.05 140 1
0.081" TYGON tubing EAE008A 2.05 140 1
0.081" TYGON tubing EAE008A 2.05 170 1
0.081" TYGON tubing EAE008A 2.05 280 1
0.081" TYGON tubing EAE008A 2.05 300 1
0.081" TYGON tubing EAE008A 2.05 350 1
0.081" TYGON tubing EAE008A 2.05 400 3
0.081" TYGON tubing EAE008A 2.05 420 1
0.081" TYGON tubing EAE008A 2.05 480 1
0.045" TYGON tubing EAE033A 1.14 35 2
0.100" TYGON tubing EAE034A 2.54 20 1
0.100" TYGON tubing EAE034A 2.54 130 1
0.100" TYGON tubing EAE034A 2.54 150 1
0.100" TYGON tubing EAE034A 2.54 340 1
0.0075" TYGON tubing EAE047A 0.19 4 1
0.073" TYGON tubing EAE049A 1.85 11 1
Sleevings GAL098A 25
Tube shielding XBA403A 1
page 3/8
2. Hydraulic & Pneumatic principles RAA016B
page 4/8
2. Hydraulic & Pneumatic principles RAA016B
ABX
Abx Eosinofix Bottle 2.05 400 EV8_1
EOSINOFIX
EV8_3 2.05 120 Diluter Syringe_3
EV8_2 1.52 500 S Reagent Heater_3
Reagent Heater_4 S 1.52 120 LMNE chamber_3
ABX
Abx Basolyse II Bottle 2.05 400 EV11_1
BASOLYSE II
EV11_3 2.05 120 Diluter Syringe_5
EV11_2 1.52 500 S Reagent Heater_10
Reagent Heater_9 S 1.52 60 S Reagent Heater_11
BASO/WBC
Reagent Heater_12 S 1.52 120
chamber_2
ABX
Abx Alphalyse Bottle 1.52 290 EV6_1
ALPHALYSE
EV6_3 1.52 130 Diluter Syringe_1
EV6_2 1.52 480 DIL1/HGB chamber_2
LMNE
LMNE chamber_4 1.30 20 M1_2 photocell
COUNTING
M1_1 photocell 1.30 320 EV4_1
EV4_2 1.02 130 T2_2
T2_3 1.02 240 Injector Syringe_5
Injector Syringe_3 1.02 95 EV5_1
EV5_2 2.05 420 T6_2
T2_1 S 1.85 11 LMNE flowcell_6
0.19 4 LMNE flowcell_6
LMNE flowcell_Output 1.52 20 E1_1 Anode fitting
E1_2 Anode fitting 1.52 70 Isolator_1
Isolator_2 1.52 80 E2_1 Ground fitting
E2_2 Ground fitting 1.52 335 EV24_2
EV24_1 1.52 120 T4_2
T4_1 1.52 160 S Reagent Heater_5
Reagent Heater_6 S 1.52 120 LMNE chamber_2
page 5/8
2. Hydraulic & Pneumatic principles RAA016B
WBC/RBC BASO/WBC
1.52 200 EV23_3
COUNTING chamber_3
EV23_1 1.52 200 RBC chamber_3
RBC chamber_2 1.52 430 EV14_2
EV14_1 1.52 70 Counting Syringe_2
EV23_2 1.52 310 Counting Syringe_3
page 6/8
2. Hydraulic & Pneumatic principles RAA016B
Valves 1 to 11:
On the left side,
Horizontal block,
On the top of the reagent and injection
syringes.
Valves 12 to 16:
On the left side,
Vertical block,
Beside the counting syringe.
Valves 17 to 19:
On the right side,
Horizontal block,
On the sample carriage.
Valves 20 to 26:
On the right side,
Vertical block,
Beside the draining syringe.
Valves 27 to 31:
On the right side,
Horizontal block,
On top of the blending and count
chambers.
page 7/8
2. Hydraulic & Pneumatic principles RAA016B
The following table shows the functions of each valve in the Pentra 60:
VALVE
VALVE SUBJECT VALVE FUNCTIONS
NUMBER
1 DIL LMNE Directing optical chamber / LMNE chamber
2 DIL LMNE Directing Internal / External sleeving
3 DIL LMNE Control Entry / Exit syringe
4 LMNE Transfer control LMNE tray /Injection syringe
5 LMNE Transfer control LMNE tray /Injection syringe
6 HEMOLYSE Control Entry / Exit syringe
7 CLEANER Control Entry / Exit syringe
8 EOSINOFIX Control Entry / Exit syringe
9 DIL Control Entry / Exit syringe
10 DIL Directing Circuit chamber / Needle circuit
11 BASOLYSE II Control Entry / Exit syringe
Directing WBC/BASO chamber / Needle
12 CLEANER
circuit
13 COUNTING SYRINGE Set in the atmosphere
14 COUNTING SYRINGE RBC/PLT count
15 TANK Set in the atmosphere
16 COUNTING SYRINGE Draining
17 NEEDLE Flushing out of rinsing unit
18 NEEDLE Commutation Diluent interior/exterior Needle
19 NEEDLE Reserved
20 DRAINING SYRINGE Set in the atmosphere
21 DIL Directing head for WBC count / Valve 22
22 DIL Directing diluent chamber dil.1
23 COUNTING SYRINGE WBC/RBC selection count
24 LMNE Exit from the optical chamber
25 DILUENT TANK Entry of the diluent
26 FLUSHING SYRINGE Flushing of the syringe
27 CHAMBER GROUP Drain rinse chamber
28 CHAMBER GROUP Drain first dilution chamber
29 CHAMBER GROUP Drain LMNE chamber
30 CHAMBER GROUP Drain RBC/PLT chamber
31 CHAMBER GROUP Drain WBC/BASO chamber
2. PNEUMATIC DIAGRAMS
page 8/8
ELECTRIC&ELECTRONIC
PRINCIPLES
3. Connections ............................................................................ 12
3.1. Chamber Heating Connector XBA386A ............................................ 12
3.2. Reagent Heating Coil Connector XBA387A ...................................... 13
3.3. Horizontal Carriage Motor XBA391A ................................................. 14
3.4. Upper Fan XBA393A ....................................................................... 15
3.5. IR sensor XBA394AS ....................................................................... 16
3.6. IR sensor XBA395AS ....................................................................... 17
3.7. IR sensor XBA396AS ....................................................................... 18
3.8. IR sensor XBA397AS ....................................................................... 19
3.9. RBC / WBC coaxe XBA398A ........................................................... 20
3.10. LMNE flowcell coaxe XEA663A (XBA399B+XBA490A) ................... 21
3.11. Barcode reader cable XBA402AS .................................................. 22
3.12. Diluent level sensor XDA605AS ..................................................... 23
page 2/24
3. Electric & Electronic principles RAA016B
TEST POTENTI-
DESIGNATION TARGET TOLERANCE REMARQUE
POINT OMETRE
TP1 HB Gain adjustment R248 4.7V +/- 0.05V
TP2 PLT Threshold adjustment R159 300mV +/- 5mV
TP3 LMNE CIS Threshold adjustment R160 650mV +/- 5mV
TP4 LMNE OD Threshold adjustment R161 350mV +/- 5mV
TP5 Draining syringe motor R149 4.8V +/- 0.05V
TP6 Counting syringe motor R150 4.8V +/- 0.05V
TP7 Dilutor syringe motor R151 4V +/- 0.05V
TP8 Injection syringe motor R152 3V +/- 0.05V
TP9 UNUSED
TP10 Horizontal carriage motor R154 3V +/- 0.05V
TP11 Sample syringe motor R155 2V +/- 0.05V
TP12 Needle carriage motor R156 4.5V +/- 0.05V
TP13 RBC Threshold adjustment R158 300mV +/- 5mV
TP14 BASO Threshold adjustment R157 300mV +/- 5mV
TP15 BASO Comparator
TP16 RBC Comparator
TP17 PLT Comparator
TP18 LMNE CIS Comparator
TP19 LMNE OD Comparator
TP20 RBC line height adjsutment R133 XX V +/- 50mV Factory Adjusted
TP21 BASO gain adjustment R134 XX V +/- 50mV Factory Adjusted
TP22 PLT line height adjustment R135 XX V +/- 50mV Factory Adjusted
TP23 LMNE Resistive gain adjustment R136 XX V +/- 50mV Factory Adjusted
TP24 LMNE Optical gain adjustment R148 XX V +/- 50mV Factory Adjusted
TP25 RBC line width adjustment
TP26 RBC line reject adjustment R142 10µs +/- 0.5µs Factory Adjusted
TP27 RBC line pulse adjustment
TP28 BASO line width adjustment
TP29 BASO line reject adjustment R139 10µs +/- 0.5µs Factory Adjusted
TP30 BASO line pulse adjustment
TP31 PLT line width adjustment
TP32 PLT line reject adjustment R144 10µs +/- 0.5µs Factory Adjusted
TP33 PLT line pulse adjustment
TP34 LMNE Resistive line width adjustment
TP35 LMNE Resistive line reject adjustment R145 15µs +/- 0.5µs Factory Adjusted
TP36 LMNE Resistive line pulse adjustment
TP37 LMNE Optical line width adjustment
TP38 LMNE Optical line reject adjustment R147 5µs +/- 0.5µs Factory Adjusted
TP39 LMNE Optical line pulse adjustment
page 3/24
3. Electric & Electronic principles RAA016B
page 4/24
3. Electric & Electronic principles RAA016B
POTENTIO- TEST
DESIGNATION TARGET TOLERANCE REMARQUE
METRE POINT
R133 RBC line height adjsutment TP20 Factory Adjusted
R134 BASO gain adjustment TP21 Factory Adjusted
R135 PLT line height adjustment TP22 Factory Adjusted
R136 LMNE Resistive gain adjustment TP23 Factory Adjusted
R139 BASO line reject adjustment TP29 Factory Adjusted
R142 RBC line reject adjustment TP26 Factory Adjusted
R144 PLT line reject adjustment TP32 Factory Adjusted
R145 LMNE Resistive line reject adjustment TP35 Factory Adjusted
R147 LMNE Optical line reject adjustment TP38 Factory Adjusted
R148 LMNE Optical gain adjustment TP24 Factory Adjusted
R149 Draining syringe motor TP5 4.8V +/- 0.05V
R150 Counting syringe motor TP6 4.8V +/- 0.05V
R151 Dilutor syringe motor TP7 4V +/- 0.05V
R152 Injection syringe motor TP8 3V +/- 0.05V
R154 Horizontal carriage motor TP10 3V +/- 0.05V
R155 Sample syringe motor TP11 2V +/- 0.05V
R156 Needle carriage motor TP12 4.5V +/- 0.05V
R157 BASO Threshold adjustment TP14 300mV +/- 5mV
R158 RBC Threshold adjustment TP13 300mV +/- 5mV
R159 PLT Threshold adjustment TP2 300mV +/- 5mV
R160 LMNE CIS Threshold adjustment TP3 650mV +/- 5mV
R161 LMNE OD Threshold adjustment TP4 350mV +/- 5mV
R248 HB Gain adjustment TP1 4.7mV +/- 0.05V
page 5/24
3. Electric & Electronic principles RAA016B
DO
NO
TM
IFY
OD
D
O IFY
OTM
N
DO
NOTE If you have initialization’s problem with 68HC11 Microcontrolers, shut E20
jumper to reset Pentra 60 (instrument must be switched on).
page 6/24
3. Electric & Electronic principles RAA016B
2. Pentra 60 Boards
page 7/24
3. Electric & Electronic principles RAA016B
Includes Keyboard and LCD screen (Allows to have one flat connector for all connections)
LED board is connected to this board
page 8/24
3. Electric & Electronic principles RAA016B
This an interconnection board where all the different motor and end of run sensor are connected
This board is connected to the main board with one flat cable
page 9/24
3. Electric & Electronic principles RAA016B
CARRIAGE BOARD:
This an interconnection board where all the carriage electrical functions are connected
This board is connected to the main board with one flat cable
page 10/24
3. Electric & Electronic principles RAA016B
LED BOARD:
This board support LEDs on the front panel. They give instrument state:
Red: instrument is busy
Green: instrument is ready to sample
page 11/24
3. Electric & Electronic principles RAA016B
3. Connections
page 12/24
3. Electric & Electronic principles RAA016B
page 13/24
3. Electric & Electronic principles RAA016B
page 14/24
3. Electric & Electronic principles RAA016B
page 15/24
3. Electric & Electronic principles RAA016B
page 16/24
3. Electric & Electronic principles RAA016B
page 17/24
3. Electric & Electronic principles RAA016B
page 18/24
3. Electric & Electronic principles RAA016B
page 19/24
3. Electric & Electronic principles RAA016B
page 20/24
3. Electric & Electronic principles RAA016B
page 21/24
3. Electric & Electronic principles RAA016B
page 22/24
3. Electric & Electronic principles RAA016B
page 23/24
3. Electric & Electronic principles RAA016B
4. Pentra 60 Synoptic
See diagram next page.
page 24/24
ANALYSIS CYCLE TECHNOLOGY
Dilution
Blood Volume 10µl
BASOLYSE II Volume 2000µl
Dilution Rate 1/200
Dilution
Blood Volume 25µl
Eosinofix Volume 1000 µl
Diluent Volume 1000µl
Final Dilution Rate 1/80
page 2/12
4. Analysis cycle technology RAA016B
Dilution
Initial Blood Volume 10 µl
Reagent Volume (RBC chamber) 2500 µl
Final Dilution Rate 1/10000
page 3/12
4. Analysis cycle technology RAA016B
2. Measuring principles
Mixing chamber
page 4/12
4. Analysis cycle technology RAA016B
page 5/12
4. Analysis cycle technology RAA016B
The hemoglobin freed by the lysis of the red blood cells combines with the
potassium cyanide to form the chromogenous cyanmethemoglobin compound.
The compound is then measured by spectrophotometry, through the optical
part of the first dilution chamber, with a wave length of 550 nm.
Dilution Measurement
Volume of blood 10 µl Method Photometry
Volume ABX DILUENT 1700 µl Wavelength 550 nm
Volume ABX ALPHALYSE 400 µl
Complement ABX DILUENT 400 µl
Final dilution rate 1/250
Reaction temperature 35°C
Result
Final HGB result represents: Absorbance value obtained x Coefficient of
calibration
The height of the impulse generated by the passage of a cell through the micro-
aperture is directly proportional to the volume of the analyzed RBC.
The hematocrit is measured as a function of the numeric integration of the MCV.
K SD
RDW =
MCV
page 6/12
4. Analysis cycle technology RAA016B
• MCV (Mean Cell Volume) is calculated directly from the RBC histogram.
• MCH (Mean Cell Hemoglobin) is calculated from the HGB value and the RBC
number.
The mean hemoglobin weight in each RBC is given by the formula:
HGB
MCH (pg) = x 10
RBC
HGB
MCHC (g/dL) = x 100
HCT
The MPV (Mean Platelet Volume) is directly derived from the analysis of the
platelet distribution curve.
page 7/12
4. Analysis cycle technology RAA016B
S1 S2
PDW
The WBC count is carried out two times by two different sensors:
• In the BASO count chamber at the same time as the BASOS count.
• In the optical chamber during the acquisition of the LMNE matrix (available
from the V1.1 software version)
The reference count is the one obtained in the WBC and BASOS count chamber.
The basophils are located from threshold <2> to the end. 100 % of the leukocytes
are represented by the total number of nucleic particles plus the basophils from
the threshold <1> to <3>.
Percentage of basophils is calculated from the number of particles existing
from the threshold <2> to <3>.
WBC BASO
page 8/12
4. Analysis cycle technology RAA016B
Results
WBC: Number of cells per volume x Coefficient of calibration.
BASO: Number of cells per volume x Coefficient of calibration in percentage
regarding the number of counted cells (BASO + WBC nuclei)
Technical characteristics of the WBC count during the acquisition of the matrix
Dilution Measurement
Volume of blood 25 µl Method Impedance with hydrofocus
Volume Eosinofix 1000 µl Ruby diameter 60 µm
Volume diluent 1000 µl Diameter of the flow 42 µm
Rate of final dilution 1/80 Injection duration 12 seconds
Reaction temperature 35 ° C Volume injected 72 µl
Incubation duration 12 s.
page 9/12
4. Analysis cycle technology RAA016B
Results
From these measurements, a matrix is drawn up with volumes on the X-axis
and optical transmission on the Y-axis
The study of the matrix image permits the clear differentiation of 4 out of 5
leukocytes populations. As a matter of fact, the basophil population is very
small compared to the other 5 in a small blood sample.
NEUTROPHILS EOSINOPHILS
ABSORBANCE
LMNE
LIC
VOLUME
page 10/12
4. Analysis cycle technology RAA016B
MONOCYTES: The monocytes, being cells with large kidney shaped nuclei
and a large non-granular cytoplasm, will neither be scattered nor absorb a large
amount of light. They will therefore be positioned in the lower part of the optical
axis but clearly to the right of the volume axis. Certain large monocytes can be
found on the right side of the matrix in the lower LIC (Large Immature Cells)
zone. The immature granulocytic cells are detected by their larger volumes and
by the presence of granules which increase the intensity of the scattered light.
Therefore, cells such as metamyelocytes will be found clearly to the right of the
neutrophils and nearly at the same level. Myelocytes and promyelocytes will
be found in saturation position on the far right of the matrix. These last three
populations will be counted as LIC (Large Immature Cells) and their given results
are included in the neutrophil value. The blast cells will be found generally to
the right of the monocytes, and, as such, will increase the LIC count. Small
blasts will be found between the normal lymphocytes and monocytes. Platelets
and debris from erythrocyte lysis represent the background noise population
located in the lower left area of the matrix. Most of the population partition
thresholds are fixed and give the limits of the morphological normality of
leukocytes. Changes in the morphology of a population will be expressed on
the matrix by a shifting of the corresponding population.
page 11/12
4. Analysis cycle technology RAA016B
DILUENT 1700µl
Hemoglobin Measurement 10µl LYSE 400µl 1/250 35°C
DILUENT (Compl.) 400µl
page 12/12
SOFTWARE VERSIONS
page 2/4
5. Software versions RAA016B
page 3/4
5. Software versions RAA016B
3. Software Release
Software Pentra 60
Release Serial Number Note
First release,
V0.82 60 002 to 60 010 Not on field
page 4/4
OUTPUT FORMAT
See ABX Diagnostics output format for haematology analyzers P/n: RAA024
ALARMS & ERROR LIST
1. Printer ................................................................................. 2
2. Transmission ...................................................................... 2
3. Calibration .......................................................................... 2
5. Reagents ............................................................................ 3
6. Temperature ........................................................................ 3
7. Alarms & Error list RAA016B
1. Printer
DISPLAYED MESSAGES CAUSES USER ACTIONS
2. Transmission
3. Calibration
Perform an autoconcentrated
Coefficients out of the
Coefficients outside the limits cleaning
ranges given section 3
Run another calibrator vial
page 2/3
7. Alarms & Error list RAA016B
4. Technical Operations
DISPLAYED MESSAGES CAUSES USER ACTIONS
Blocked motor
Emergency stop, Run an
Incorrect Drains control the motor operations : Menu
autocontrol
Thermal door opened "MAINTENANCE / MECHANICAL
SYSTEMS / CHECK MOTORS"
.....not reaching home Blocked motor
Thermal door opened open during a cycle close the door and rerun the cycle
5. Reagents
6. Temperature
page 3/3
MAINTENANCE
1. Maintenance ...................................................................... 2
1.1. Introduction .................................................................................. 2
1.2. Daily customer maintenance ........................................................ 2
1.3. Weekly customer maintenance .................................................... 2
1.4. Maintenance table ....................................................................... 3
3. Procedures ......................................................................... 8
3.1. Procedure chart tables ................................................................. 9
3.2. Required tools and products ....................................................... 10
8. Maintenance RAA016B
1. Maintenance
1.1. Introduction
WARNING
Customer maintenance has to be carried out according to the recommended
frequency chart table and after having attended an ABX approved customer trai-
ning course.
The system warranty may be affected if damage occurs after a non trained technician
intervenes or if replaced spare parts and consumables do not come from an ABX
approved origin.
CAUTION
Never use solvent or abrasive cleaning material to clean the system.
page 2/10
8. Maintenance RAA016B
Sampling syringe X X X X X X
5DIFF syringe X X X X X X
Draining chamber X X X X X X
XEA486BS
Optical bench lamp X X X X X X Procedure
RAS237
LMNE flowcell coaxial X X X X X X
Counting heads X X X X X X
page 3/10
8. Maintenance RAA016B
2. Maintenance kits
Follow the RAS236 procedure for maintenance with the XEA485AS kit. The kit
is including:
Follow the RAS237 procedure for maintenance with the XEA486BS kit. The kit
is including:
page 4/10
8. Maintenance RAA016B
Follow the RAS238 procedure for maintenance with the XEA581BS kit. The kit
is including:
PISTONS KIT
Part Number Qty
XEA581BS
The fitting kit contains the most common pneumatic and hydraulic parts:
FITTING KIT
Part Number Qty
XEA311A
EAA005A Straight connector Ø1.6mm 1/8 10
EAA006A Straight connector 10
EAA009A Bent connector I=1.5 10/32 10
EAB002A "L" Connector 10
EAB003A Connector Ø4mm H19.5 20
EAB005A "T" Connector Ø1.6mm T10-6 30
EAB006A "T" Connector Ø2.3mm 30
EAB009A Straight connector Ø1.6mm 30
EAB010A Straight connector Ø2.3mm 30
EAA013A Bent connector 1/8 10
EAA014A Straight connector 1/8 10
EAB026A "Y" Connector Ø2.5mm 3
EAB021A "Y" Connector Ø3mm 3
EAB035A "T" Connector Ø2.3mm T220-6 10
EAB014A Straight connector "Y" Ø1.5mm/1.5mm 10
page 5/10
8. Maintenance RAA016B
The screws kit contains the most common screws used in Pentra 60:
SCREWS KIT
Part Number Qty
XEA293AS
KAH001A HU M3 Nut 20
KAH002A HU M4 Nut 20
KAH003A HU M5 Nut 20
page 6/10
8. Maintenance RAA016B
page 7/10
8. Maintenance RAA016B
3. Procedures
CONCERNING PARTS:
- Hydraulic maintenance and adjustments.
- Pneumatic maintenance and adjustments.
- Electrical maintenance and adjustments.
- Power supply maintenance and adjustments.
- Electronic maintenance and adjustments.
- Printer maintenance and setup.
NOTE Each procedure has to be read entirely before beginning the intervention:
RED Procedures are Adjustment procedure
BLEU Procedures are Maintenance procedure
GREEN Procedures are Dismantling procedure
WARNING
When cleaning instruments, disposable gloves should be worn.
page 8/10
8. Maintenance RAA016B
EVERY 2 YEARS MAINTENANCE Sampling probe: Probe & Guide Replacement - Reagents syringe:
RAS238
KIT Pistons Replacement - Vacuum/Waste pump: Piston Replacement
LMNE FLOWCELL
RAS240 Optical Bench Check and Adjustment
ADJUSTMENT
RAS247 FRONT PANEL DISMANTLING Front panel Dismantling - Front panel Mounting
RAS248 HEATER ASSY REPLACEMENT Heater Assy Dismantling - Heater Assy Mounting
POWER SUPPLY
RAS250 Power supply Dismantling & Replacement
REPLACEMENT
START CYCLE MICROSWITCH Start cycle microswitch Dismantling - Start cycle microswitch
RAS251
REPLACEMENT Mounting
CHECK UP AFTER
RAS252 Instrument check up: Repeatability - Calibration
INTERVENTION
OPTICAL BENCH DISMANTLING Optical bench Dismantling - Optical bench Replacement - Emission
RAS254
& REPLCEMENT gun location Control - Optical bench lamp alignement Control
BARCODE READER
RAS255 Instrument configuration for Barcode Reader
CONFIGURATION
page 9/10
8. Maintenance RAA016B
T O O L S P R O D U C T S
THERMOMETER
TORX KEYS
page 10/10
INSTALLATION
• CONCERNS
1 - Working conditions
2 - Unpacking
3 - General description
4 - Installation
5 - LX300+ Printer connection
6 - Reagent connections
7 - Priming and startup
• REQUIRED TOOLS
Hexagonal keys
Pentra 60
• REQUIRED PRODUCTS
Pentra 60 Reagents:
Bottles,
Waste container.
• INTERVENTION TIME
60 minutes
• FREQUENCY
RAS235B
INSTALLATION
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
INSTALLATION KIT
Part Number Qty
XEA791A
WORKING CONDITIONS
1 - Environment
The ABX Pentra 60 should be operated in an indoor location only. Operation at an altitude over
2000 meters (6000 Feet) is not recommended. Instrument is designed to be safe for transient
voltages according to INSTALLATION CATEGORY II and POLLUTION DEGREE 2.
Please ask your ABX representative service center for any information about operating loca-
tion when it does not comply with the recommended specifications.
2 - Location
The ABX Pentra 60 should be placed on a clean and level table or work station. Please note
that the ABX Pentra 60, printer and reagents weigh approximately 40 kilograms (88 lbs). Avoid
exposure to sunlight. Proper ventilation requires that a space of at least 20 cm (8 inches) must
be left behind the apparatus.
INSTALLATION
3 - Grounding
Proper grounding is required. Check that the wall ground (earth) plug is correctly connected to
the laboratory grounding electricity installation. If there is no ground then use a ground stake.
Current electricity Standards must be applied.
The ABX PENTRA 60 must function between 16 to 34°C (61 to 93°F). Maximum relative humidity
80% for temperatures up to 31°C (88°F) decreasing linearly to 50% relative humidity at 40°C
(104°F). If it is kept at a temperature less than 10°C (50°F), the instrument should be allowed to
sit for an hour at the correct room temperature before use.
The ABX Pentra 60 has been designed to produce less than the required level of electromagnetic
interferences in order to operate in conformity to its destination. The electromagnetic interferences
by the ABX Pentra 60 are limited to a level allowing the correct operation of other instruments in
conformity with their destination.
In case of problems, check that the instrument is not placed in the proximity of electromagnetic
fields, or short wave emissions (radars, X-rays, scanner, etc...).
6 - Environment protection
When old the instruments and their accessories must be collected by a laboratory specialized
in elimination and recycling of this kind of equipment according to the legislation.
p.3/11
INSTALLATION
GENERAL DESCRIPTION
1. Cover:
Before any attempt to remove the cover, open the
pneumatic access door.
2. Pneumatic access door:
allows the operator to access hydraulic parts for mainte-
nance operations. It is mandatory to keep the door
locked during the measuring cycles as it ensures the
heating of the dilutions !
3. Sampling probe:
This needle is used for the 53µl whole blood sampling.
4. Control panel:
This panel allows the operator to communicate with the
instrument :
- To access the different cycles
- To identify the patients
- To setup the instrument, etc...
5. Door to reagents bottles:
Pentra 60
Turn the locker as shown on the Diag. 2 to open the pneumatic protection door.
Diag.3
Diag.2
1. Sampling carriage:
• Ensuring the needle positioning for the different sampling stages and the distribution.
• Support the sampling syringe and the distribution of the blood
2. Sampling syringe:
• Take the specimen and distribute the portions into the different chambers.
• Take the sample from the first dilution and distribute into the RBC/PLT chamber.
3. Count assembly:
• Receives the different rinsing and dilutions.
• Ensure the temperature control of the dilutions.
• Ensure the counts for WBC, BASOS, RBC, PLT, Hct, Hb.
p.4/11
RAS235B
INSTALLATION
Diag.4&5
4. Draining syringe:
• Draining of the different chambers.
• Bubbling of the mixtures.
• Transfer, by vacuum, the LMNE specimen of the mixing chamber towards the injector of the
optical chamber.
5. Diluent tank:
• To hold the necessary diluent for an analysis cycle.
• To avoid the risk of the «degassing» of the diluent as it is being aspirated by the syringes.
This tank is vacuum filled by the counting syringe.
6. Counting syringe:
• Ensure the vacuum for the WBC and the BASOS counts.
• Ensure the vacuum for the RBC and the PLT counts.
• Ensure the vacuum for the filling of the internal tank with diluent.
9. Optical bench:
• Ensure the support and the adjustment of the optical chamber
• Ensure the support and the adjustment of the lamp
• Ensure the support and the adjustment of the optical and electronic elements.
p.5/11
INSTALLATION
2 - Rear panel
Internal left
1 -hand
Wasteside,
output
2 - Diluent input
Mother board
3 - Serial number label
4 - Barcode reader connector
5 - Printer connector
6 - RS 232 C Output
7 - Power supply cable
Pentra 60
Diag.6
INSTALLATION
- Turn the locker (as shown on the Diag.2) to open the pneumatic protection door.
Diag.7
p.6/11
RAS235B
1 - Reagent connections
INSTALLATION
- Install the straw deep in the container.
IMPORTANT
Diluent Tube:
Use 3x6 cristal tube for diluent connection. Tube length must be under 2 meters
Waste Tube:
Use 4x6 cristal tube for waste connection. Tube length must be under 2 meters
Diag.8
p.7/11
INSTALLATION
3 - Bottle connection
- Open reagent bottles and replace the caps by the «bottle stoppper and reagent straw» assembly.
Diag.9 Diag.10
IMPORTANT
After the priming of one or several reagents, close the pneumatic access door (Right side
of the instrument), switch off the PENTRA 60, then re-start it
Diag.11
p.8/11
RAS235B
INSTALLATION
Diag.12&13
Open the ink ribbon access door at the top of the printer and install the ribbon as shown in Diag
14 & 15:
- Guide the ribbon between the print head and ribbon guide.
- Slide the printer head from side to side to make sure it moves smoothly.
Diag.15
Diag.14
p.9/11
INSTALLATION
- Install the paper supports (Diag.16) for single sheets paper use only.
Diag.16
- The printer is connected to the ABX Pentra 60 with the cable delivered with the instrument.
Lock the connector in place by tightening the 2 screws on each end of the connector to the ABX
Pentra 60. Attach the other end of the cable to the printer and lock the printer connector in place
Pentra 60
Diag.17
- For continuous paper, introduce it in the slot at the back of the printer and use the sprocket
covers to load paper, or feeding the paper frontward when using single sheets (see printer user
manual).
Diag.19
Diag.18
p.10/11
RAS235B
Instrument priming
DILUENT
0% CHANGE
EOSINOFIX 1 - LEVEL/CHANGE
11% CHANGE
BASOLYSE II
23% CHANGE
CLEANER
20% CHANGE
HGB LYSE
52% CHANGE
INSTALLATION
REAGENTS\LEVEL/CHANGE 16:05 «CHANGE» key in reverse video.
DILUENT
100% CHANGE
EOSINOFIX 1 - LEVEL/CHANGE - Press : a prime cycle is automatically run.
99% CHANGE
BASOLYSE II
98% CHANGE - The reagent level is updated to «100%»
CLEANER
98% CHANGE
ALPHALYSE
98% CHANGE
Never pour reagent from one container to another. Particles present in the bottom of the old
container can contaminate the new reagent, causing unacceptable background counts, especially
for PLTs.
IMPORTANT
Before analyzing samples, visually inspect reagent lines and pumps for air bubbles. Repeat
priming if air bubbles are still present. Call the ABX representative service department if
priming does not eliminate air bubbles.
Never initiate two Lyse prime cycles back-to-back. This causes excessive foaming in the
waste chamber. Run a blank cycle between each Lyse prime cycle.
- Run a STARTUP cycle and follows the ABX Pentra 60 User Manual's instructions.
p.11/11
6 MONTH MAINTENANCE
• CONCERNS
Instrument maintenance step by step
• REQUIRED TOOLS
Hexagonal keys
Dynamometric screw driver A302, A301, A300
Cutting pliers
Flat screw driver
Philips screw driver
Torx keys
Pentra 60
• REQUIRED PRODUCTS
Minoclair
• INTERVENTION TIME
2h
• FREQUENCY
See Maintenance table
RAS236B
6 MONTH MAINTENANCE
SAMPLING PROBE
REAGENTS SYRINGE
DRAINING & COUNTING SYRINGES
INSTRUMENT CLEANING
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Sampling needle
FAA065A O’ring 6.30x2.40 Silicon 1
Reagent syringe
XDA621A O’ring 30.80x3.80 + wedge 2
Vacuum/Waste pump
XDA622A O’ring 15.54x2.62 + wedge 4
Reagent syringe
XEA019A Silicon grease 1
NOTE
p. 2/10
RAS236B
SAMPLING PROBE
6 MONTH MAINTENANCE
carriage (Diag.1)
Diag.1
Diag.2
ü Take out the Probe and the guide at the same time (lift the Probe locker to free the
Probe) (Diag.2 & 3)
ü Put the Probe back to its support
Diag.4
Diag.3
TORQUE
Rinsing block screws:
120mN.m (17 Ozf.in)
ü Open rinsing block assembly and replace the old O’ring joint by the new one in the 6
month maintenance kit (use one drop of silicon grease between 2 fingers for O’ring) (Diag.4)
ü Reassemble following previous steps backward
ü When startup is done check that there is no leak
p. 3/10
6 MONTH MAINTENANCE
REAGENTS SYRINGE
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.5
p. 4/10
RAS236B
TORQUES
9x CHC M3x12:
400mN.m (56.8 Ozf.in)
2x FX M3x12:
6 MONTH MAINTENANCE
400mN.m (56.8 Ozf.in)
Diag.6
WARNING
A washer is adapted to each O’ring joint tore diameter, so that it is mandatory to keep
O’ring joint and its washer together.
Lyse piston O’ring had no washer.
p. 5/10
6 MONTH MAINTENANCE
Diag.7
Diag.8 Diag.9
p. 6/10
RAS236B
6 MONTH MAINTENANCE
Diag.11
Diag.10
ü Unscrew the 4 CHC M4x16 syringe screws (just few turns to free silent block) (Diag.10)
ü Remove the syringe
ü Drain the syringe pushing the piston nut up and down several times
ü According to the type of syringe fitted, unscrew the 4x CHC M4x16 screws (Diag.11) or
unscrew the syringe body and open it (Diag.12).
Diag.12 Diag.13
ü Replace the O’ring (Diag.14) by the new one from the 6 month maintenance kit (use one drop
of silicon grease in between 2 fingers to lubricate O’ring and piston).
ü Reassemble in reverse order according to diagram 14 (Molded syringe body Diag.12 has to
be rotated until the mark on the body is in between the two marks of the motor Diag.15).
O’ring
Anti-extrusion ring
O’ring thickness
wedge
Diag.14
Diag.15
p. 7/10
6 MONTH MAINTENANCE
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.17
Diag.16
Diag.19
Diag.18
p. 8/10
RAS236B
ü Replace the old O’ring joint and washer by new one from the 6 month maintenance kit (use
one drop of silicon grease in between 2 fingers to lubricate the O’ring) (Diag.22).
6 MONTH MAINTENANCE
Diag.20 Diag.21
ü Reassemble in reverse order according to diagram 22 (Molded syringe body Diag.20 has to
be rotated until the mark on the body is in between the two marks of the motor Diag.23).
O’ring
Anti-extrusion ring
O’ring thickness
wedge
Diag.22
Diag.23
p. 9/10
6 MONTH MAINTENANCE
INSTRUMENT CLEANING
ü Dilute the 12°cl bleach to 1 part of bleach for 4 of deionize water (1/5).
ü No sponge, nor cloth must be used. Only absorbant paper, thrown after use, in contamination
bins, have to be used. For small or sensitive assemblies use accurate drier papers.
ü All assemblies suspected to have been in contact with biohazardous material must be
disinfected with diluted bleach (the stainless steel must be bleached below 30°Celsius).
ü Blood stains or salt marks must be cleaned with spray detergent first.
ü Concerned assemblies:
ü Reinstall all the assemblies and setup the instrument back to its initial configuration.
ü Follow procedure RAS252 to check instrument accuracy and eventually re-calibrate it.
p. 10/10
YEARLY MAINTENANCE
• CONCERNS
• REQUIRED TOOLS
Hexagonal keys Philips screw driver
Dynamometric screw driver A302, A301, A300 Torx keys
Clamps
Scalpel
Cutting pliers
Flat screw driver
Pentra 60
• REQUIRED PRODUCTS
Empty sample tubes
Silicone grease
Soft tissue
Liquid soap
Distilled water
Micropipette tip
RBC/PLT Latex
• INTERVENTION TIME
2h
• FREQUENCY
The yearly maintenance frequencies vary with the instrument output.
According to the number of cycle, 3 categories are created :
0 - 30 cycles/day -> 1 overall maintenances/year
30 - 120 cycles/day -> 2 overall maintenances/year
> 120 cycles/day -> 3 overall maintenances/year
RAS237B
YEARLY MAINTENANCE
SAMPLING SYRINGE
5DIFF SYRINGE
OPTICAL BENCH LAMP
LMNE FLOWCELL COAXIAL
DILUENT RESERVOIR JOINT
DRAINING CHAMBER
COUNTING HEADS
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
p. 2/12
RAS237B
SAMPLING SYRINGE
YEARLY MAINTENANCE
Diag.1 Diag.2
Diag.3 Diag.4
p. 3/12
YEARLY MAINTENANCE
5DIFF SYRINGE
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.5
Diag.6
ü Remove the bottom plate (2x FX M3x12, 4x CHC M3x12), the piston axles and the
O’ring joints (Diag.7)
Diag.7 Diag.8
p. 4/12
RAS237B
ü Replace all O’ring joints by new one in the yearly maintenance kit (use one drop of
silicon grease between 2 fingers for each O’ring and piston except the Diameter 12 piston
of the 5DIFF syringe) (Diag.8)
ü Reassemble following previous steps backward
ü When startup is done check that there is no leak
TORQUES
YEARLY MAINTENANCE
4x CHC M3x12: 400mN.m (56.8 Ozf.in)
p. 5/12
YEARLY MAINTENANCE
Diag.9
Diag.10
WARNING
Do not touch lamp bulbe with the fingers, use a piece of cloth or paper to manipulate
the lamp.
p. 6/12
RAS237B
WARNING
YEARLY MAINTENANCE
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.11
ü Disconnect the «T» connector (few diluent drops will leak) (Diag.12&13)
Diag.12
Diag.13
p. 7/12
YEARLY MAINTENANCE
ü Disonnect ground fitting and put the new LMNE coaxial instead
ü Reassemble following previous steps backward (except optical bench cover)
ü Run Menu: Maintenance\Hydraulics systems\Rinse\Cytometer to get rid of the air
bubbles stuck to the inner optical surfaces
ü Check that there is no leak and that flowcell contains no or just a very few air bubbles
ü Put back the optical bench cover
Diag.14
MEASUREMENT\LMNE ADJUSTMENT
LMNE LAMP
6.00
4.00 8.00
TRANSFERT 1TIME
- LEVEL/CGE
198
100 300
RESISTIVE CHANNEL
51
5 70
ABSORBANCE CHANNEL
180
10 240
Diag.15
ü Check that the displayed values are within target values range:
ü If you are out of range perform a RAS240 LMNE Flowcell adjustment procedure
p. 8/12
RAS237B
YEARLY MAINTENANCE
Diag.16
NOTE
Reservoir body must be turned to have
inferior support backward and tubes in the
upper part in frontward
TORQUE
4x CHC M3x20:
120mN.m (17 Ozf.in)
Diag.17
p. 9/12
YEARLY MAINTENANCE
DRAINING CHAMBER
Diag.19
Diag.18
Diag.20
p. 10/12
RAS237B
COUNTING HEADS
YEARLY MAINTENANCE
ü Disconnect the chamber tubes (not the wastes)
ü Dismantle the Counting chamber assembly, 3 screws (Diag.21)
Diag.21
Diag.22 Diag.23
ü Use a previously cut micropipette tip to replace the electrode O’ring joint as shown in
diagram 25
ü Reassemble electrode
ü Repeat with the other electrode (Diag.23)
ü When startup is done check that there is no leak
p. 11/12
YEARLY MAINTENANCE
Diag.24
TORQUE
2x CHC M3x12:
120mN.m (17 Ozf.in) Diag.25
CAUTION
Do not manipulate the aperture using hard instruments. Clean the aperture with a
piece of soft paper or preferably, in between 2 fingers.
CAUTION
Do not apply too much pressure on the electrode fixation screws, as it can
break the aperture (tightening torque = 120mN.m / 17 Ozf.in).
It is recommended to reconnect the tubes on the counting head before
reassembling the "electrode/chamber/counting head" assy in order to avoid
applying constraint on the chamber.
p. 12/12
EVERY 2 YEARS MAINTENANCE
• CONCERNS
Instrument maintenance step by step
• REQUIRED TOOLS
Hexagonal keys Philips screw driver
Dynamometric screw driver A302, A301, A300 Torx keys
Clamps
Scalpel
Cutting pliers
Flat screw driver
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
2 hours
• FREQUENCY
0 to 120 cycles per day: Every 2 years
>120 cycles per day: Every year
RAS238B
EVERY 2 YEARS MAINTENANCE
SAMPLING PROBE
REAGENT SYRINGE
COUNTING/DRAINING SYRINGE, PISTON
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
PISTONS KIT
Part Number Qty
XEA581BS
TIGHTENING
REP SCREWS
TORQUES
(DIAG n°) DESIGNATION
mN.m (Ozf.in)
p. 2/11
RAS238B
SAMPLING PROBE
Diag.1 Diag.2
ü Take out the Probe and the rinsing block at the same time (lift the Probe locker to free
the Probe) (Diag.2&3)
ü Open guide assembly and replace the probe rinsing block and the probe (Diag.4)
Diag.4
Diag.3
IMPORTANT
Use a new FAA053A O’ring joint when performing 6 Month maintenance at the same
time
p. 3/11
EVERY 2 YEARS MAINTENANCE
REAGENTS SYRINGE
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.5
WARNING
A washer is adapted to each O’ring joint tore diameter, so that it is mandatory to keep
O’ring joint and its washer together
p. 4/11
RAS238B
Diag.6
ü Replace Pistons by new ones in the every 2 years maintenance kit (use one drop of
silicon grease between 2 fingers for each piston) (Diag.6)
IMPORTANT
Be sure to use the new washers corresponding to the new O’ring joints, when
performing 6 Month maintenance at the same time
TORQUES
9x CHC M3x12: 400mN.m (99.4 Ozf.in)
p. 5/11
EVERY 2 YEARS MAINTENANCE
COUNTING/DRAINING SYRINGE
Diag.7
Diag.8 Diag.9
p. 6/11
RAS238B
Diag.10
ü Unscrew the 4 CHC M4x16 syringe screws (just few turns to free silent block) (Diag.10)
ü Remove the syringe
ü Drain the syringe pushing the piston nut up and down several times
ü According to the type of syringe fitted, remove the syringe guiding plate (4 torx screws)
(Diag.11)
ü According to the type of syringe fitted, unscrew the 4x CHC M4x16 screws (Diag.12) or
unscrew the syringe body and open it (Diag.13).
Diag.12 Diag.13
p. 7/11
EVERY 2 YEARS MAINTENANCE
Diag.14
Pentra 60
2
IMPORTANT
Take care to keep syringe nut in place on the motor screw
1
ü Piston replacement:
Old model of syringe: Use parts 1 + 2
New model of syringe: Use part 1
ü Replace Piston by new one in the every 2 years maintenance kit (use a bit of silicon
grease)
WARNING
Be sure to use the new washer corresponding to the new O’ring joint, when performing
6 Months maintenance at the same time
ü Reassemble in reverse order according to diagram 15 (Molded syringe body has to be rotated
until the mark on the body is in between the two marks of the motor Diag.16).
O’ring
Anti-extrusion ring
O’ring thickness
wedge
Diag.15
Diag.16
ü Reassemble following previous steps backward
ü When startup and prime are done check for that there is no leak
p. 8/11
RAS238B
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.17
Diag.19
Diag.18
ü Disconnect grounding wire from the frame (Diag.19)
p. 9/11
EVERY 2 YEARS MAINTENANCE
ü Put a piece of aborbant paper under the syringe and disconnect all the tubes from the
syringe (Diag.20)
Pentra 60
Diag.20 Diag.21
Diag.23
Diag.22
p. 10/11
RAS238B
IMPORTANT
Take care to keep syringe nut in place on the motor screw 1
ü Piston replacement:
Old model of syringe: Use parts 1 + 2
New model of syringe: Use part 1
ü Replace Piston by new one in the every 2 years maintenance kit (use a bit of silicon
grease)
WARNING
Be sure to use the new washer corresponding to the new O’ring joint, when performing
6 Month maintenance at the same time
ü Reassemble in reverse order according to diagram 25 (Molded syringe body has to be rotated
until the mark on the body is in between the two marks of the motor Diag.26).
O’ring
Anti-extrusion ring
O’ring thickness
wedge
Diag.25
Diag.26
ü Reassemble following previous steps backward
ü When startup and prime are done check for that there is no leak
p. 11/11
DECONTAMINATION & RINSE
• CONCERNS
Instrument decontamination before maintenance operation in the following
cases:
- Instrument removed from a biohazardous area
- Maintenance intervention on contaminated assemblies
• REQUIRED TOOLS
Hexagonal keys
Clamps
Flat screw driver
Torx keys
Pentra 60
• REQUIRED PRODUCTS
Fungicidal, bactericidal, virus killing detergent spray, non corrosive for metals,
Non plastic altering.
Bleach solution 12°Cl
Deionize water
Protection gloves
Absorbant paper
Distilled water
• INTERVENTION TIME
1h35min
• FREQUENCY
On request
RAS239B
DECONTAMINATION & RINSE
DECONTAMINATION
1 - Preparation (20min)
ü Switch on the instrument
ü Open the pneumatic access door (Right side of the instrument)
ü Run Menu: Maintenance\Hydraulics systems\Concentrated cleaning (Diag.1)
WARNING
Disposable gloves and lab coat must be worn by the operator
Local or national regulations must be applied in at all the times
1-BACKFLUSH
1- BACKFLUSH
2-RINSE
2- RINSE 3-DRAIN CHAMBERS
3- DRAIN CHAMBERS 4-CONCENTRATED CLEANING
4- CONCENTRATED CLEANING
Diag.1 Diag.2
POUR2-HYDRAULICS
3 ml OF MINOCLAIR INTO CHAMBERS
SYSTEMS
PRESS ESC KEY TO CONTINUE
3-MECHANICAL SYSTEMS
Diag.3
p. 2/4
RAS239B
• Concerned assemblies
ü Outer surfaces of the instrument (perpex, covers, LCD, reagent locations....)
ü Keyboards
ü Waste connector plug
ü Liquid valve push button
ü Assemblies close to the needle
ü Tube holder assembly
ü Overflow trays
ü Reinstall all the assemblies and setup the instrument back to its initial configuration
DILUENT
1- DILUENT 100% CHANGE
2- CLEANER EOSINOFIX 1 - LEVEL/CHANGE
3- EOSINOFIX 99% CHANGE
4- BASOLYSE II BASOLYSE II
5- ALPHALYSE 98% CHANGE
6- ALL REAGENTS CLEANER
7- UNPRIME ALL 98% CHANGE
ALPHALYSE
98% CHANGE
Diag.4 Diag.5
ü Fill a sample tube with diluted bleach to 1 part of bleach for 4 of deionize water (1/5)
ü Run 15 cycles Burn-in Menu: Technician\User mode\Burn-in leave the instrument
operating until it stops
p. 3/4
DECONTAMINATION & RINSE
DECONTAMINATION
WARNING
When rinse is finished check that LMNE chamber is empty
p. 4/4
LMNE FLOWCELL ADJUSTMENT
• CONCERNS
Instrument adjustment:
LMNE Flowcell
• REQUIRED TOOLS
Hexagonal keys
Dynamometric screw driver A302, A301, A300
Flat screw driver
Torx keys
Pentra 60
• REQUIRED PRODUCTS
Latex RBC\PLT
Fresh blood samples
Calibrator
• INTERVENTION TIME
1h
• FREQUENCY
On request
RAS240B
LMNE FLOWCELL ADJUSTMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p. 2/6
RAS240B
WARNING
Never dismantle Emission gun or even unlock it. The adjustment is factory made and
cannot be performed in the field.
1 - Preparation
ü Open the cover
ü Open the left side panel and the main board door (block it)
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.1
p. 3/6
LMNE FLOWCELL ADJUSTMENT
IMPORTANT
Mix RBC/PLT LATEX thoroughly and for few minutes !
MEASUREMENT\LMNE ADJUSTMENT
LMNE LAMP
6.00
4.00 8.00
TRANSFERT 1TIME
- LEVEL/CGE
Pentra 60
198
100 300
RESISTIVE CHANNEL
51
5 70
ABSORBANCE CHANNEL
180
10 240
Diag.2
Diag.3
p. 4/6
RAS240B
NOTE Test points on Optical bench board, near R11 potentiometer, can be
used to check 6V with a voltmeter.
R11
Diag.4
Diag.5 Diag.6
p. 5/6
LMNE FLOWCELL ADJUSTMENT
R148
R136
Diag.7
ü Put back the Optical Bench cover
After intervention check, with normal fresh blood, that you have a cor-
NOTE rect matrix.
p. 6/6
CHAMBERS ADJUSTMENT
• CONCERNS
Instrument adjustment:
Chambers assembly adjustment
• REQUIRED TOOLS
Hexagonal keys
Dynamometric screw driver A302, A301, A300
Flat screw driver
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
15min
• FREQUENCY
RAS241B
CHAMBERS ADJUSTMENT
CHAMBERS ADJUSTMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p. 2/5
RAS241B
CHAMBERS ADJUSTMENT
ü This adjustment procedure must be carried out when chamber assembly is moved or
replaced.
1 - Alignment Check
ü Open the cover and the right access door
CHAMBERS ADJUSTMENT
ü Switch on the instrument
ü Menu: Maintenance\Technician\Others\Maintenance carriage position
ü Switch off the instrument
ü Locate probe as shown in diagrams 1 & 2
Diag.1 Diag.2
ü Push the carriage backward (without lifting the probe), check that there is the same
height between probe and first and last tops of chamber (Diag.3 & 4)
Diag.3 Diag.4
p. 3/5
CHAMBERS ADJUSTMENT
2 - Alignement Adjustment
ü Free the chamber support by loosening the 4 screws (Diag.5)
Diag.5
Pentra 60
Diag.6
Diag.7
p. 4/5
RAS241B
ü Push the carriage backward (without lifting the probe), check that there is the same
height between probe and last top of chamber (Diag.8 & 9)
CHAMBERS ADJUSTMENT
Diag.8 Diag.9
ü If necessary adjust position moving up and down chamber
ü Perform an alignment check as described at the begenning of this procedure
ü When adjustment is correct, tighten screws and perform a Probe Adjustment procedure
WARNING
It is mandatory to perform the Probe adjustment procedure after this adjustment
p. 5/5
PROBE ADJUSTMENT
• CONCERNS
Instrument adjustment:
Probe adjustment
• REQUIRED TOOLS
None
Pentra 60
• REQUIRED PRODUCTS
None
• INTERVENTION TIME
15 min
• FREQUENCY
On request
RAS242B
PROBE ADJUSTMENT
INTRODUCTION
NEEDLE HOME CHECK & ADJUSTMENT
ALL CARRIAGE POSITIONS CHECK & ADJUSTMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p. 2/7
RAS242B
INTRODUCTION
IMPORTANT
Check that the chamber assy is parallel to the carriage motion before checking or
making an adjustment of the Probe position. Follow CHAMBERS ADJUSTMENT
procedure of this manual
PROBE ADJUSTMENT
ü Probe adjustments can be accessed by the menu «Maintenance\
Technician\Dilution» as explained below (Diag.1):
ü Needle Home: Adjustment to get ü Check (2): Allows to ü Check (1): Allows to
the best probe extremity cleaning check that probe is aligned check probe cleaning
with LMNE-3 inlet position
IMPORTANT
Running Check functions check the Probe positions before any adjustment.
p. 3/7
PROBE ADJUSTMENT
IMPORTANT
When the probe is in «NEEDLE HOME» position it is also in cleaning position.
Check and adjust the NEEDLE HOME position to achieve the best cleaning of the
probe extrimity
Diag.2 Diag.3
p. 4/7
RAS242B
IMPORTANT
To check the right position of the probe you must see inlet as in Diagram 4 and not
as in diagram 5 (where circles are not concentric)
PROBE ADJUSTMENT
Diag.4 Diag.5
Diag.6
ü Press ESC key to move to the next chamber, and continue until the end of the cycle
ü When the cycle is ended check that no tube is disconnected
IMPORTANT
Probe Up/Down adjustment is the same for all the chambers and cannot be modified
from one chamber to another (adjustment by the «NEEDLE POSITION» value)
Probe Left/Right position in the chambers can be adjusted separately for each
chamber (except the Rinse chamber)
p. 5/7
PROBE ADJUSTMENT
Left/Right adjustment:
Diag.7
ü If you want to modify only one of the position, remind to increase or decrease the next
motion with the same number of steps, not to loose the other adjustments.
Pentra 60
S.
E PO
IAG S.
RR
CA
E PO
O
BA
S IAG
B/ A RR
G
(2) EC
) L MN
(3
.
S. P OS
PO GE
IAG
E
RIA
2
ARR C AR
C
1 BC
DIL (4)
R
(1) 4
3
1
Diag.8
p. 6/7
RAS242B
ü The right location for the probe is when you see it through the inlet as in the following
diagram (Diag.9):
Diag.9
PROBE ADJUSTMENT
ü When the cycle is ended check that no tube is diconnected
Up/Down adjustment:
Diag.10
Diag.11
p. 7/7
MAIN BOARD ADJUSTMENT
• CONCERNS
Instrument adjustment:
Main board adjustment
• REQUIRED TOOLS
Small flat screw driver
Hexagonal keys
Voltmeter
Pentra 60
• REQUIRED PRODUCTS
None
• INTERVENTION TIME
30min
• FREQUENCY
On request
RAS243B
MAIN BOARD ADJUSTMENT
HB BLANK ADJUSTMENT
APERTURE CURRENT CHECK
RBC/PLT GAIN ADJUSTMENT
WBC/BASO GAIN ADJUSTMENT
DRAIN SENSOR ADJUSTMENT
LMNE SENSOR ADJUSTMENT
MOTOR CURRENT ADJUSTMENT
THRESHOLDS ADJUSTMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p. 2/10
RAS243B
Diag.1
NOTE This adjustment is made with the right side door closed and under
stabilised thermic conditions (5min at least after startup)
Diag.2
p. 3/10
MAIN BOARD ADJUSTMENT
Diag.3
Pentra 60
MAINTENANCE\TECHNICIAN\MEASUREMENT 6:43
Diag.4
WARNING
Do not damage connectors while measuring
IMPORTANT
Use an old coaxe that you have previously cut to control voltages with the voltmeter
ü Check Aperture currents in J22, J23 and J24, aperture current measured must be
around 60V (not ajustable)
ü Press ESC to exit
ü Reconnect LMNE CIS, RBC and WBC coaxes (Diag.3)
p. 4/10
RAS243B
RBC
77
PRESS ESCAPE TO SAMPLE THE LATEX 56 96
PLT
111
80 135
Diag.5 Diag.6
IMPORTANT
Mix RBC/PLT Latex thoroughly before sampling
Diag.7
p. 5/10
MAIN BOARD ADJUSTMENT
WBC/BASO
PRESS ESCAPE TO SAMPLE THE LATEX 101
57 156
Diag.8 Diag.9
Pentra 60
IMPORTANT
Mix Latex thoroughly before sampling
Diag.10
p. 6/10
RAS243B
Diag.11 Diag.12
Diag.13
p. 7/10
MAIN BOARD ADJUSTMENT
Diag.14 Diag.15
Pentra 60
Diag.16
p. 8/10
RAS243B
ü Potentiometers and Test points are located on Main board as shown on the next diagram:
Draining syringe
Counting syringe
Reagents syringe
Optical Bench
Injector syringe
Horizontal carriage
Sample syringe
Probe carriage
Up/Down
Diag.17
p. 9/10
MAIN BOARD ADJUSTMENT
THRESHOLDS ADJUSTMENT
1 - Thresholds Adjustment
ü Open the cover and the left access door to main board
ü Check and adjust, when necessary, all following voltages:
ü Potentiometers and Test points are located on Main board as shown on the next diagram:
Pentra 60
Diag.18
PLT
LMNE OD LMNE CIS RBC BASO
p. 10/10
THERMIC ADJUSTMENT
• CONCERNS
Instrument adjustment:
Concerns the enclosed thermic and the preheating of the reagents
• REQUIRED TOOLS
Thermometer
Pentra 60
• REQUIRED PRODUCTS
None
• INTERVENTION TIME
30min
• FREQUENCY
On request
RAS244B
THERMIC ADJUSTMENT
THERMIC CHECKS
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p. 2/7
RAS244B
THERMIC CHECKS
THERMIC ADJUSTMENT
TECHNICIAN\HEATING SYSTEMS 15:38
Diag.1
IMPORTANT
This control must be made with the right side door closed and under stabilised
thermic conditions (5min at least after startup)
Diag.3
IMPORTANT
Make sure to have plunged the thermometer probe in the liquid
p. 3/7
THERMIC ADJUSTMENT
REFERENCE 15:08
Diag.4
ü Every time you hear one beep read the Thermometer value: it must be 35°C +/-1°C
To validate check you must get, during control, 10 times the right
NOTE temperature
If an adjustment is required, when heating block is replaced, follow this
procedure
Pentra 60
p. 4/7
RAS244B
ADJUSTMENT 15:05
THERMIC ADJUSTMENT
Diag.5
The reagent heating system CTN value is located on the CTN wire near
NOTE the main board (left connector on J31) (Diag.6)
Be sure you read the right value, sticker could be up or down
Diag.6
ü Make a reagents preheating temperature check (at the begining of this procedure)
IMPORTANT
This control must be made with the right side door closed and under stabilised
thermic conditions (5min at least after startup)
ü If you don’t get the right temperature (35°C +/-1°C), adjust it by the mean of the
Reference temperature (Diag.7)
REFERENCE 15:08
Diag.7
p. 5/7
THERMIC ADJUSTMENT
Diag.8 Diag.9
IMPORTANT
Pentra 60
This control must be made with the right side door closed and under stabilised
thermic conditions (5min at least after startup)
Diag.10
REFERENCE 15:08
Diag.11
ü Every time you hear one beep read the Thermometer value: it must be 35°C +/-1°C
To validate check you must get, during control, 10 times the right
NOTE temperature
If an adjustment is required, when heating block is replaced, follow this
procedure
p. 6/7
RAS244B
ADJUSTMENT 15:05
THERMIC ADJUSTMENT
Diag.12
ü Check and eventualy change the CTN value, thermostated compartment CTN value
is located on the CTN wire near the main board (right one connected on J33) (Diag.13)
Diag.13
IMPORTANT
This control must be made with the right side door closed and under stabilised
thermic conditions (5min at least after startup)
ü If you don’t get the right temperature (35°C +/-1°C), adjust it by the mean of the
Reference temperature (Diag.14)
REFERENCE 15:08
Diag.14
p. 7/7
VACUUM ADJUSTMENT
• CONCERNS
Instrument adjustment:
Concerns counting and draining syringe vacuum control and adjustment
• REQUIRED TOOLS
Barflex
Pentra 60
• REQUIRED PRODUCTS
None
• INTERVENTION TIME
15min
• FREQUENCY
On request
RAS245B
VACUUM ADJSUTMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
ü When installed in altitude, some intrument can exhibit some «Transfer Sensor Timeout»
alarms. This is due to the decreased atmospheric pressure. In these conditions, the draining
vacuum can lower down to about 200mb/210mb instead of 260mb (20% less): this results in
Pentra 60
a too long transfer time of the solution between LMNE chamber to T1-T2 segment before the
flowcell.
IMPORTANT
The tubing length must remain identical
1
2
p. 2/5
RAS245B
VACUUM ADJUSTMENT
Diag.1
1 - COUNTING
2 - DRAINING CHECK VACUUM
Diag.2 Diag.3
ü Check vacuum on the Barflex, vacuum must be around 260mb and must be stable
ü Disconnect the Barflex and connect back the tube on the counting syringe
p. 3/5
VACUUM ADJSUTMENT
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.4
VACUUM 850
1 - COUNTING
2 - DRAINING
CHECK
Diag.5 Diag.6
ü Check vacuum on the Barflex while the following screen is displayed, vacuum must
be around 220mb
VACUUM 850
CHECK
Diag.7
NOTE To get a better adjustment (difference of altitude for example) follow this
procedure
ü Disconnect the Barflex and connect back the tube on the counting syringe
ü Close the right side door
p. 4/5
RAS245B
VACUUM ADJUSTMENT
Diag.8
VACUUM 50
CHECK
Diag.9
ü Repeat vacuum check and step value adjustment until you get the correct vacuum
ü Disconnect the Barflex and connect back the tube
ü Close the right side door
p. 5/5
FRONT PANEL DISMANTLING
• CONCERNS
Instrument dismantling:
Front panel
• REQUIRED TOOLS
Hexagonal keys
Dynamometric screw driver A302, A301, A300
Flat screw driver
Torx keys
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
30min
• FREQUENCY
On request
None
RAS247B
FRONT PANEL DISMANTLING
FRONT PANEL
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p. 2/5
RAS247B
FRONT PANEL
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
WARNING
Fully lift sampling probe before pushing backward sampling carriage
ü Push the carriage back and unscrew the 2 Torx M3x16 screws (several turns) (Diag.1&2)
Diag.1 Diag.2
Diag.3
p. 3/5
FRONT PANEL DISMANTLING
Diag.4
Pentra 60
Diag.5
ü Disconnect the flat connector on the LCD & Keyboard board (Diag.6)
Diag.6
p. 5/5
HEATER ASSY REPLACEMENT
• CONCERNS
Instrument dismantling:
Replacement of the heater assy
• REQUIRED TOOLS
Hexagonal keys
Dynamometric screw driver A302, A301, A300
Flat screw driver
Torx keys
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
1h
• FREQUENCY
On request
RAS248B
HEATER ASSY REPLACEMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p.2/8
RAS248B
1 - Introduction
ü Heater assy has to be changed only when required
Diag.1
p.3/8
HEATER ASSY REPLACEMENT
Diag.2
Pentra 60
- BASO 2
- RBC 1
- LMNE 2 & 3
- DIL 1
ü The heater assy is free, let it lay on the frame and by the front aperture disconnect
following tubes:
Diag.3
Diag.4
p.4/8
RAS248B
ü Disconnect heating system wiring on the mother board plugged in J31: (Diag.5)
Diag.5
HEATER ASSY REPLACEMENT
ü Take out the cable through the instrument
ü Take out the heater assy and drain it on absorbant paper
Diag.6
p.5/8
HEATER ASSY REPLACEMENT
NOTE You can put the former and new heater beside and connect following
tubes one to one
REP. ON
TO
HEATER
ü Put new heater, with previous tubing connected, on instrument frame (Diag.8)
ü Connect following tubes: (Diag.9)
REP. ON
TO
HEATER
1 Valve 22 Inlet 2
3 Valve 8 Inlet 2
5 T4
7 Valve 22 Inlet 1
10 Valve 11 Inlet 2
Diag.8 Diag.9
p.6/8
RAS248B
ü Dismantle chamber assy to get the tubes through the chamber’s support (Diag.10)
ü Lift the heater assy, make the 5 tubes go through the chamber’s support steel sheet
(Diag.11)
Diag.11
Diag.12
p.7/8
HEATER ASSY REPLACEMENT
ü Put back the chamber assy and connect the following tubes on chambers:
- BASO 2
- RBC 1
- LMNE 2 & 3
- DIL 1
Diag.13
ü Push back the Reagents syringe and the LMNE syringe and screw silent blocks (check
that no tube is disconnected)
ü Put back the Front panel, follow RAS247 Front panel dismantling of this manual
ü Check chambers position, follow RAS241 Chambers Adjustment of this manual
ü When startup and prime are done check that there is no leak
p.8/8
BUBBLING ADJUSTMENT
• CONCERNS
Instrument adjustment:
Bubbling
• REQUIRED TOOLS
None
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
15min
• FREQUENCY
On request
RAS249B
BUBBLING ADJUSTMENT
BUBBLING ADJUSTMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p.2/3
RAS249B
BUBBLING ADJUSTMENT
1 - Introduction
ü Bubbling is factory adjusted and normally does not required further adjustments
ü If an adjustment is necessary follow this procedure
BUBBLING ADJUSTMENT
2 - Bubbling Adjustment
ü Enter Menu: Maintenance\Technician\Bubbling\ (Diag.1)
ü Enter new step value for each bubbling
MAINTENANCE\TECHNICIAN\BUBBLING 16:05
1-AUTOCONTROL CYCLE
FIRST DILUTION 00
LMNE 300
WBC/BASO 300
RBC/PLT 300
ALPHALYSE 400
Diag.1
p.3/3
POWER SUPPLY REPLACEMENT
• CONCERNS
Instrument dismantling:
Power supply
• REQUIRED TOOLS
Hexagonal keys
Dynamometric screw driver A302, A301, A300
Flat screw driver
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
30min
• FREQUENCY
On request
RAS250B
POWER SUPPLY REPLACEMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p.2/5
RAS250B
Diag.1
ü With a flat screwdriver disconnect, on the mother board, the supply connector in J37
(Diag.2)
Diag.2
ü Open the left side panel and the main board door (block it)
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
p.3/5
POWER SUPPLY REPLACEMENT
Diag.3
ü Unscrew the 2 CHC M3x6 back screws of the power supply (Diag.4)
Pentra 60
Diag.4
ü Unscrew the 2 CHC M3x6 front screws of the power supply (Diag.5)
Diag.5
p.4/5
RAS250B
TEST
DESIGNATION TARGET
POINT
TP40 5V Power supply +5V
TP41 12V Power supply +12V
TP42 24V Power supply +24V
TP43 -12V Power supply -12V
Diag.6
p.5/5
START CYCLE MICROSWITCH REPLACEMENT
• CONCERNS
Instrument dismantling:
Start cycle microswitch replacement
• REQUIRED TOOLS
Hexagonal keys
Torx keys
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
30min
• FREQUENCY
On request
RAS251B
START CYLE MICROSWITCH REPLACEMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p.2/4
RAS251B
1 - Introduction
ü Start cycle microswitch has to be changed or dismantle only when required
ü To access start cycle microswitch you need to remove the Front panel, follow RAS 247
A Front panel dismantling of this manual
Diag.2
Diag.1
ü Release the Start cycle microswitch loosening the 2 screws (Diag.2)
Diag.3
p.3/4
START CYLE MICROSWITCH REPLACEMENT
IMPORTANT
Microswitch is connected Normally Open
ü Put back the Start cycle microswitch in the 2 slots (nearly full backward) (Diag.4)
ü Tighten the 2 microswitch screws
Pentra 60
Diag.4
ü Check with the front panel in its location, that the Start cycle key is operating normally.
If not move forward or backward the microswitch until you get a correct location
NOTE You must ear a «click» when key is pressed, and one when
key is released
ü Reinstall fan
WARNING
Use the washer to connect ground wiring to the fan
ü Follow RAS247 Front panel dismantling of this manual to reassemble front panel
p.4/4
CHECK UP AFTER INTERVENTION
• CONCERNS
Instrument maintenance:
Check up after intervention
• REQUIRED TOOLS
Fresh and normal blood samples
Calibration blood samples
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
1h
• FREQUENCY
On request
RAS252B
CHECK UP AFTER INTERVENTION
p.2/5
RAS252B
1 - Preliminary
WARNING
This procedure must be performed on a clean instrument
If instrument is suspected to be not perfectly clean, perform an Autoconcentrated
cleaning
WARNING
Disposal gloves and white coat must be worn by the operator
Local or national regulations must be applied in all the operations
2 - Repeatability
ü Based on 20 consecutives analyses without alarm from one fresh and normal blood
sample
ü Run 20 consecutives analyses
ü Control to have variation coefficients within the following acceptable limits (Diag.1):
TEST
DESIGNATION TARGET
POINT
TP40 5V Power supply +5V
TP41 12V Power supply +12V
TP42 24V Power supply +24V
TP43 -12V Power supply -12V
Diag.1
X : Mean X=
∑ Xi
Xi : Measure value n
n : Measure number
SD : Standard deviation
SD =
∑ ( X − Xi )²
n −1
SD
CV (%) = x100
X
ü With one fresh and normal blood sample check that there is no alarm in CBC+5DIFF
mode
p.3/5
CHECK UP AFTER INTERVENTION
3 - Calibration
Autocalibration:
ü Enter Menu: Calibration\Autocalibration\ (Diag.2)
Diag.2 Diag.3
Pentra 60
Diag.4
IMPORTANT
Change Lot number, Expiration date or Target values if necessary
Run Calibration:
ü Prepare the calibrator according to the specific instructions (temperature, mixing, etc...)
ü Open the vial and position the sampling needle deeply inside the bottle
ü Press the sampling bar located behind the needle
ü When the cycle LED stops flashing, remove the vial and replace the cap on the calibrator
ü When the analysis cycle ends, the first result is displayed on the result chart table
ü Run the second calibrator sample
Diag.5
p.4/5
RAS252B
Calibration passed:
ü If the statistic figures are within the acceptable limits:
- Coefficient of variation is within the limits setup as described section and
- The percentage difference between the target and the mean value is less than 20
Diag.6 Diag.7
CALIBRATION\AUTOCALIBRATION 16:05
WBC RBC HGB HCT PLT
1 9.5 4.92 13.1 43.3 172
2 9.3 4.90 13.0 42.7 171
3 9.5 4.82 13.0 42.2 175
ENTER TO PRINT RESULTS
ESC TO CONTINUE
STAT WBC RBC HGB HCT PLT
COEFF 127 225 41.0 251 225
NEW 136 209 41.9 222 328 H
MEAN 9.4 4.88 13.0 42.7 173
CV% 1.31 1.07 0.32 1.27 1.39
Diag.8
ü Check that the calibration coefficients remain within the following ranges (Diag.9):
Calibration
Standard value minimum maximum
coefficient
Diag.9
p.5/5
STEP BY STEP ANALYSIS CYCLE
• CONCERNS
Instrument control:
Analysis cycle
• REQUIRED TOOLS
Fresh blood samples
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
15min
• FREQUENCY
On request
RAS253B
STEP BY STEP ANALYSIS CYCLE
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p.2/8
RAS253B
1 - Introduction
ü The main cycle consists of four principal phases:
1 - Preparation Of The Sample Phase: Dilutions
2 - Count And Measurement Of The WBC Group
Diag.2
Diag.1
Draining
Syringe
Diag.3 Diag.4
p.3/8
STEP BY STEP ANALYSIS CYCLE
ü Transfer of the needle toward the rinsing chamber (Sampling needle motor, Carriage
motor)
ü Distribution of 3 µl of blood during the rinsing of the needle (Sampling syringe, Reagents
syringe)
ü Transfer of the needle from rinse chamber to the first dilution chamber (Sampling needle
motor, Carriage motor)
syringe)
ü Transfer of the needle from the first dilution chamber to the WBC and BASO count
chamber (Sampling needle motor, Carriage motor)
ü Transfer of the needle from the WBC and BASO count chamber to the LMNE mixing
chamber (Sampling needle motor, Carriage motor)
ü Transfer of the needle from the LMNE mixing chamber to the rinse chamber (Sampling
needle motor, Carriage motor)
ü Distribution of the 5 µl of residual blood during the double rinsing (interior and exterior) of
the needle (Sampling syringe, Reagents syringe)
p.4/8
RAS253B
ü Transfer of the needle from the first dilution chamber to the RBC and PLT count
chamber (Sampling needle motor, Carriage motor)
ü Distribution of the 42.5 µl of first dilution into a flux of 2 ml of diluent (Sampling needle
motor, Carriage motor)
ü Distribution of 0.5 ml of diluent from the interior of the needle (Sampling needle motor,
Carriage motor)
ü In the case of this RBC/PLT dilution, it is the entire blood sample that
NOTE is distributed (and not a portion as in the preceding dilution). The distri-
bution in a flux is kept in order to benefit from the homogenisation as-
pect of the procedure
p.5/8
STEP BY STEP ANALYSIS CYCLE
ü Start the vacuum and first part of count (duration of 6 seconds, control of valve 23)
p.6/8
RAS253B
4 - Filling Of Tank
ü At the end of the RBC/PLT count, the RBC count vacuum is «finished» and used for the
filling up of the diluent tank (control valves 15 and 25)
ü In order to improve the filling speed, the syringe is initialised for a few seconds then drawn
again to complete the filling
p.7/8
STEP BY STEP ANALYSIS CYCLE
ü If no other cycle is started during RBC/PLT count or during the filling up of the tank, a
special rinsing sequence is initiated. The sample signal light changes to red and the instru-
ment is «jammed» for a few seconds
ü This sequence involves draining of the RBC chamber followed by a refilling with the diluent
Pentra 60
p.8/8
OPTICAL BENCH REPLACEMENT
• CONCERNS
Instrument dismantling:
Dismantling, Replacement and Control of Optical Bench
• REQUIRED TOOLS
Hexagonal keys
Dynamometric screw driver A302, A301, A300
Flat screw driver
Torx keys
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
30min
• FREQUENCY
On request
RAS254B
OPTICAL BENCH REPLACEMENT
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p.2/6
RAS254B
WARNING
A flat cable is connected at the back of the mother board, becareful when you open the
door that you don’t disconnect or damage it
Diag.1
Diag.2
p.3/6
OPTICAL BENCH REPLACEMENT
Diag.3
Pentra 60
NOTE Do not tighten silent blocks too hard, check that optical bench Cannot be
lifted
p.4/6
RAS254B
ü Check that gap between emission gun and flowcell is around 3mm (Diag.5)
WARNING
Becareful, if you use a 3mm allen key to control gap, that you don’t damage flowcell or
emission gun
Diag.5
WARNING
This adjustment is factory made and must not be changed
p.5/6
OPTICAL BENCH REPLACEMENT
Diag.6
ü To control lamp put a piece of white paper between reception gun and flowcell (Diag.7)
Pentra 60
Diag.7
Diag.9
Diag.8
WARNING
A small screw (1.2mm allen key) allows adjutment of lamp (Diag.9)
This adjustment is factory made and must not be changed
p.6/6
BARCODE READER CONFIGURATION
• CONCERNS
Instrument adjustment:
Barcode Reader test and configuration
• REQUIRED TOOLS
None
Pentra 60
• REQUIRED PRODUCTS
• INTERVENTION TIME
15min
• FREQUENCY
On request
RAS255B
BARCODE READER CONFIGURATION
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p.2/6
RAS255B
1 - Test
ü Check that barcode reader is working properly by means of the reading test (Diag.1)
1234567 TEST
EAN 8 2/5
9076543210
1234567
CODABAR
EAN 13
543210
123456789012
C128
EAN 13
0123456789
123456789012
Diag.1
ü If the reading test is correct, the procedure is completed
p.3/6
BARCODE READER CONFIGURATION
ü When the last label is read, check once more on the test labels (Diag.1)
WARNING
For code I2 of 5, to avoid a bad interpretation, it is mandatory to hold the reader in
order to read the entire code
ü If the reading test has failed for the DATALOGIC reader, proceed to the barcode setup as
follows:
Pentra 60
$+ AD111AD*0316 EB3EHRFFFF
AZ0AA3 EB3EHA45FF $-
p.4/6
RAS255B
ü Make sure that the BARCODE reader is properly fitted in the DIN plug of the RS adaptor
wiring.
ü Check the barcode reader works properly by means of the reading test on this page. If the
reading test is correct, the procedure is completed. If not, continue below.
ü Switch on the instrument and within the 15 first seconds, read from the top to the bottom the
barcode labels located on page 4 of this procedure. The audible beep occurs after each reading.
Read all the labels from top to bottom and from left to right.
ü When the last label is read, the signal beeps 5 times in order to indicate the end of the setup.
Check once more on the test labels.
WARNING
For code I2 of 5, to avoid a bad interpretation, it is mandatory to hold the reader in
order to read the entire code
C39
TEST
2/5
9076543210
CODABAR
543210
C 128
98765
p.5/6
BARCODE READER CONFIGURATION
p.6/6
LMNE BALANCE ADJUSTMENT
• CONCERNS
Instrument adjustment:
Adjustment of the LMNE Balance.
• REQUIRED TOOLS
None
Pentra 60
• REQUIRED PRODUCTS
None
• INTERVENTION TIME
15 min
• FREQUENCY
On request
RAS304B
LMNE BALANCE ADJUSTMENT
INTRODUCTION
AUTO-CALIBRATION
FORCED CALIBRATION
WARNING
Disposal gloves and white coat must be worn by the operator.
Local or national regulations must be applied in all the operations.
Pentra 60
p.2/5
RAS304B
INTRODUCTION
AUTO-CALIBRATION
Diag.1
ü Prepare the sample according to the specific instructions (temperature, mixing, etc...),
insert the tube and close the tube holder.
IMPORTANT
Risk of erroneous results if the calibrator is not continously mixed between each
analysis. Continue mixing the calibrator between each analysis.
ü When the analysis cycle ends, the first result is displayed on the result chart table.
ü When the Green LED turns on, run the same specimen again.
IMPORTANT
At least 5 results are necessary to calibrate the LMNE Balance.
p.3/5
LMNE BALANCE ADJUSTMENT
ü The results selected to be involved in the statistical calculation have a dot in the square
column. To discard a result from the statistical calculation, select the line and press the
DOT key (Diag.2):
Diag.2
ü When ready to calibrate press ENTER to apply new calibration coefficient (Diag.3):
Diag.3
Diag.4
Diag.5
p.4/5
RAS304B
FORCED CALIBRATION
WBC 131
HCT 218
RBC 251
HGB 41
PLT 232
RDW 0.50
MPV 1.00
WBC LMNE 0.95
Diag.7
ü With Up and Down arrows select the WBC LMNE calibration coefficient and change it.
p.5/5
TRAINING SLIDES
EXPLODED VIEWS
1. 11 Valves assy ............................................................................ 1
2. 5 Valves assy (12-16) ................................................................... 2
3. 2 Valves assy .............................................................................. 3
4. 7 Valves assy .............................................................................. 4
5. 5 Valves assy (27-31) ................................................................... 5
6. Optical bench syringe assy ......................................................... 6
7. Reagents syringe assy ................................................................ 7
8. Sampling syringe assy ................................................................. 8
9. Syringe motorization carriage assy ............................................... 9
10. Syringe motor assy ...................................................................... 10
11. Syringe motor body assy ............................................................. 11
12. Sampling motor assy ................................................................... 12
13. Optical syringe motor assy .......................................................... 13
14. Reagents motor assy ................................................................... 14
15. Sampling syringe assy ................................................................. 15
16. Vacuum syringe piston assy ........................................................ 16
17. Vacuum syringe assy ................................................................... 17
18. Short vacuum syringe assy .......................................................... 18
19. Long vacuum syringe ................................................................... 19
20. Long vacuum syringe assy .......................................................... 20
21. Short vacuum syringe assy .......................................................... 21
22. Needle carriage motor assy ......................................................... 22
23. Horizontal sampling carriage notched belt ................................... 23
24. Notched belt ................................................................................ 24
25. Needle carriage assy ................................................................... 25
26. Sampling horizontal carriage ........................................................ 26
27. Sampling horizontal carriage assy ................................................ 27
28. Sampling carriage assy ............................................................... 28
29. Carriage vertical belt .................................................................... 29
30. Carriage rinsing block assy .......................................................... 30
31. Home carriage cell ....................................................................... 31
32. Needle assy ................................................................................ 32
33. Carriage motor assy ..................................................................... 33
34. Sampling needle assy ................................................................. 34
35. Reagents heating coils assy ........................................................ 35
36. Thermostated room assy ............................................................. 36
37. Thermostated room plate assy .................................................... 37
38. Thermostated room fan assy ....................................................... 38
39. Frame assy .................................................................................. 39
40. Diluent chamber assy .................................................................. 40
41. Diluent chamber .......................................................................... 41
42. RETIC flowcell assy ..................................................................... 42
43. Optical bench assy ..................................................................... 43
44. Optical bench lamp ..................................................................... 44
45. Preamplifier support .................................................................... 45
46. Cover pentra 60 assy .................................................................. 46
47. LED board pentra 60 assy ........................................................... 47
48. Pentra 60 chamber assy .............................................................. 48
49. Pentra 60 chamber assy .............................................................. 49
50. WBC/HGB/BASO Chamber assy ................................................. 50
SPARE PART LIST
REF DESIGNATION PACKAGING CATEGORY