Part Number: RAN352H
Date: 29/5/17
ECR: 6134
Ref: 1300032689
Diagnostic Kit SIL: RAN352H
Objective This tool allows you to enter the MAC address of the instrument and the test of all hardware and
peripherals.
It should be particularly used in case of SBC9312 board replacement. This evolution is related to new
availability of MAC addresses.
Intervention
Intervention priority For information - As needed - Mandatory on next visit - Immediate and mandatory
information
Tracking number SIL-RAN352H Intervention estimated duration: 15 minutes
on instrument serial #
version A to B
adjustments
documents
2/8
Diagnostic Kit SIL: RAN352H
1. Subject
The purpose of this kit is, in case of SBC9312 board replacement, to set up back the MAC address and test all
hardware and peripherals features.
It is delivered on a USB key, 4 Go, pre loaded with DAGNOSTIC software.
The modules are tested ones after the others. The course of the test follows the order of the below procedure.
■ This diagnostic tool can now be used on the Micros Care ST instrument.
■ The software is available in English. You can choose between English or French languages.
■ Test RTC: you can now enter a date equal or above 2011.
3. Procedure
■ Connect the keyboard and the power supply cable.
■ Insert the USB key in the front USB port then turn ON the instrument.
3/8
Diagnostic Kit SIL: RAN352H
■A cross appears on the screen. Press the center of this cross to move to the next position.
Press the cross for each position.
The message Selectionner la langue/Please select the language appears.
■ Selectthe language.
The message Touchscreen calibrated, you can launch the tests appears.
If the touchscreen test fails, you are not allowed to correctly perform the rest of the procedure.
■ Click 2- SAV.
■ The messages Have you got a printer: Internal? External? and Have you got a barcode reader: Yes No appear.
■ Select your printer.
■ Declare or not your barcode reader.
■ Click Next.
■ Enter
your instrument number and the day. The serial number is located on a label under the instrument.
Example: Micros n: 7 06 ESO H 00008*
JOUR: 10 (actual day)
4/8
Diagnostic Kit SIL: RAN352H
Zone 1 Zone 3
Zone 2
5/8
Diagnostic Kit SIL: RAN352H
CODE 39
Interleaved 2 of 5
CODARBAR
CODE 128
ISTB128
■ Click Next.
6/8
Diagnostic Kit SIL: RAN352H
■ Plug a serial cable (DAC028A) between serial port 1 and serial port 2 at the rear of the instrument.
■ Click 1- Start the serial port test.
■ Click Next when it becomes green.
■ Click Stop.
■ Turn the instrument OFF.
■ Unplug the serial cable (DAC028A) from the serial port 1 & 2 at the rear of the instrument (leave the USB key).
■ Turn the instrument ON.
The first screen displays the date and time of the instrument when it has been turned off and the date and time after
boot.
■ Click Next.
In order to test the ethernet connexion, the instrument should be connected to a PC by the mean of an ethernet cross
cable on the RJ42 port at the rear of the instrument.
■ Enter the IP address of the test computer and then click 2- IP fixe.
To get the IP address, the instrument must be connected to a PC and be turned ON.
- Click Start/Run on the computer.
- Enter cmd and click OK.
- Enter ipconfig and press Enter to display the IP address.
■ Click Next.
■ Click 3- Bypass if the instrument is not connected.
7/8
Diagnostic Kit SIL: RAN352H
8/8