Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Tn

SIL
Part Number: RAN352H

ABX Micros ES60 / ABX Micros ESV60 /


ABX Micros Care ST / SCIL Vet abc+
Diagnostic Kit
TEMP-0220_rev16

Date: 29/5/17
ECR: 6134

Ref: 1300032689
Diagnostic Kit SIL: RAN352H

Objective This tool allows you to enter the MAC address of the instrument and the test of all hardware and
peripherals.
It should be particularly used in case of SBC9312 board replacement. This evolution is related to new
availability of MAC addresses.

Intervention
Intervention priority For information - As needed - Mandatory on next visit - Immediate and mandatory
information
Tracking number SIL-RAN352H Intervention estimated duration: 15 minutes

User process impact YES NO Potential patient result impact YES NO

User training needed YES NO User documentation update YES NO

on instrument serial #

ABX Micros ES 60:ESOH11239


Applied in factory ABX Micros ESV 60:11201
since instrument serial #
Micros Care ST:CSTH11206
SCIL Vet abc+: EVOH11201

Note history Note index Reason for change


A to B Bug correction and software translation in English

B to C Addition of ABX Micros Care ST instrument

C to D Service information letter updated with ABX Micros Care ST

D to E New availability of MAC addresses

E to F New kit to order and new MAC addresses

F to G New kit to order and MAC addresses correction

G to H New kit to order and MAC addresses correction

Software Note index Previous New Mini. requirements

version A to B

New parts Parts designation Previous New Reason for change


Diagnostic kit ABX Micros
ES60/ABX Micros ESV60/ XEC027FS Kit to order: 1300032700 Correction of MAC addresses
SCIL Vet abc+

New Adjustments Previous New Reason for change

adjustments

New Documentation Previous New Reason for change

documents

2/8
Diagnostic Kit SIL: RAN352H

Additionnal Designation Part number Link

requirements External keyboard CBK053A (AZERTY keyboard)


(USB) CBK054A (QWERTY keyboard)

Part details Component


availa-
kit or Assy Designation Component designation Qty ble for
p/n
order
Diagnostic kit ABX
Micros ES60/ABX
1300032700 1207951099 USB key 4 Go, loaded 1 No
Micros ESV60/SCIL
Vet abc+

1201891028 RS232 cable 1 Yes

1300032689 Service information letter 1 No

1. Subject
The purpose of this kit is, in case of SBC9312 board replacement, to set up back the MAC address and test all
hardware and peripherals features.
It is delivered on a USB key, 4 Go, pre loaded with DAGNOSTIC software.
The modules are tested ones after the others. The course of the test follows the order of the below procedure.

An external keyboard (USB) is necessary to run this procedure.

2. Evolution and Corrected Bug


■ MAC addresses are unique.
■ MAC addresses start by the company ID 70-B3-D5.
■ MAC addresses are included between AB-80-00 and AB-8F-FF.

■ This diagnostic tool can now be used on the Micros Care ST instrument.
■ The software is available in English. You can choose between English or French languages.

■ Test RTC: you can now enter a date equal or above 2011.

3. Procedure
■ Connect the keyboard and the power supply cable.
■ Insert the USB key in the front USB port then turn ON the instrument.

3/8
Diagnostic Kit SIL: RAN352H

3.1. Touchscreen Test

■A cross appears on the screen. Press the center of this cross to move to the next position.
Press the cross for each position.
The message Selectionner la langue/Please select the language appears.
■ Selectthe language.
The message Touchscreen calibrated, you can launch the tests appears.

If the touchscreen test fails, you are not allowed to correctly perform the rest of the procedure.

■ Click 2- SAV.

3.2. Printer and Barcode Reader

■ The messages Have you got a printer: Internal? External? and Have you got a barcode reader: Yes No appear.
■ Select your printer.
■ Declare or not your barcode reader.
■ Click Next.

3.3. Serial Number/MAC Address

■ Enter
your instrument number and the day. The serial number is located on a label under the instrument.
Example: Micros n: 7 06 ESO H 00008*
JOUR: 10 (actual day)

The MAC address appears.


Example 00:50:C2:58:93:F1
■ Click Next.

3.4. RTC EEPROM Test

■ Enter the date and time:


Date: 10 31 2006 (Month, day, year)
Heure: 10 36 52 (hours, minutes, secondes)
■ Click 1- RTC.
■ Click Next.

4/8
Diagnostic Kit SIL: RAN352H

3.5. Screen Test

■ Click 1- Start the screen test.


■A black screen appears, check that there is no default (missing pixel, for example).
■ The screen Have you seen some default on the black screen? appears.
- Click 4- Start the test again to test the black screen.
- Click 2- Yes if you have seen some defaults.
The message Error ’’black test’’ failed appears.
Click Accept so that the diagnostic software goes to next test.
- Click 3- No if the black screen is correct.
■A red screen appears, then a green one and a blue one. Follow the same procedure for those screens.

Zone 1 Zone 3

Zone 2

■ For each color, check the three diffrent zones as follow:


- Zone 1: Progressive range from dark to clear.
- Zone 2: Range vertical dark/clear hatches.
- Zone 3: Diagonal dark/clear hatches starting with a dark triangle in the left upper corner.
■ The following message appears Have you seen some default on the test screen?.
- Click 4- Start the test again if you want to test the screen again.
- Click 2- Yes if you have seen some defaults.
The message Error ’’screen test’’ failed appears.
Click Accept so that the diagnostic software goes to next test.
- Click 3- No if the screen is correct.
■ Click 4- Start the test again to test the screen again, or click Next.
■ Click 1- Start the luminosity test.
■ The message Have you seen a luminosity change? appears.
- Click 4- Start again to run the luminosity test again.
- Click 3- No if you have not seen a luminosity change.
The message Error: luminosity test failed is displayed.
Click Accept so that the diagnostic software goes to next test.
- Click 2- Yes if you have seen a luminosity change.

5/8
Diagnostic Kit SIL: RAN352H

3.6. Buzzer Test

■ Click Launch the buzzer test.


■ The following message appears Did you hear the 3 different frequencies?.
- Click 4- Start again if you want to test the buzzer again.
- Click 3- No if you did not correctly heard the buzzer.
The message Error: Buzzer test invalid appears.
Click Accept so that the diagnostic software goes to next test.
- Click 2- Yes if you heard correctly the buzzer.

3.7. Barcode Test

■ Click 1- Start the barcode reader test.


■ Waitfor the scanner initialization, then pass the following barcodes in front of the barcode reader, at a distance from
approximately 10 to 15 cm.
■ Check that all the barcodes are correctly read.

CODE 39

Interleaved 2 of 5

CODARBAR

CODE 128

ISTB128

■ Click Next.

6/8
Diagnostic Kit SIL: RAN352H

3.8. RS232 Ports Test

■ Plug a serial cable (DAC028A) between serial port 1 and serial port 2 at the rear of the instrument.
■ Click 1- Start the serial port test.
■ Click Next when it becomes green.

3.9. Memory Test

■ Select Disk on chip in the User choice column.


■ Click 1- Memory test.
■ Click Next when it becomes green.

3.10. RTC and EEPROM Test

■ Click Stop.
■ Turn the instrument OFF.
■ Unplug the serial cable (DAC028A) from the serial port 1 & 2 at the rear of the instrument (leave the USB key).
■ Turn the instrument ON.
The first screen displays the date and time of the instrument when it has been turned off and the date and time after
boot.
■ Click Next.
In order to test the ethernet connexion, the instrument should be connected to a PC by the mean of an ethernet cross
cable on the RJ42 port at the rear of the instrument.

■ Enter the IP address of the test computer and then click 2- IP fixe.
To get the IP address, the instrument must be connected to a PC and be turned ON.
- Click Start/Run on the computer.
- Enter cmd and click OK.
- Enter ipconfig and press Enter to display the IP address.
■ Click Next.
■ Click 3- Bypass if the instrument is not connected.

3.11. Printer Test

■ Click 1- Print results.


■ Check the correct printout of the results.
ok: test done and passed
nok: test done, but not passed
n/a: test for factory only or not
■ Click Turn off the ABX Micros ES60/ABX Micros ESV60.
■ Remove the USB key.

7/8
Diagnostic Kit SIL: RAN352H

8/8