Liquick Cor-GGT

DIAGNOSTIC KIT
FOR DETERMINATION OF
γ-GLUTAMYLTRANSFERASE
ACTIVITY

Kit name Cat. No GGT activity remains stable in specimen up to 2 days at 15-25ºC or
Liquick Cor-GGT mini 1-296 1 week at 2-8ºC or 1 month at -25ºC. Freezing of sample causes
Liquick Cor-GGT 30 1-226 a loss of enzyme activity. Frozen specimens should be thawed and
Liquick Cor-GGT 60 1-224 kept at room temperature for 18 to 24 hours before measurement to
achieve full enzyme reactivation.
INTRODUCTION Nevertheless it is recommended to perform the assay with freshly
γ-Glutamyltransferase (GGT, GGTP) is a membrane localized collected samples!
enzyme that catalyzes the transfer of glutamyl groups from
glutathione to amino acids or peptides. Large GGT amounts are PROCEDURE
present in secretory organs: kidney, liver, bile duct, pancreas. These reagents may be used both for manual assay (Sample Start and
Although the GGT activity is highest in renal tissue, serum GGT is Reagent Start method) and in several automatic analysers.
generally elevated as a result of liver disease. Since alcohol induces Applications for them are available on request.
GGT production, measurement of GGT activity is used for
monitoring of abstinence in withdrawal treatment. Manual procedure
wavelength 405 nm
METHOD PRINCIPLE temperature 25ºC / 30°C / 37°C
Kinetic method with L-γ-glutamyl-3-carboxy-4-nitroanilide. cuvette 1 cm

γ-GT Sample Start method


L-γ-glutamyl-3-carboxy-4-nitroanilide + glycylglycine Pipette into the cuvette:
L-γ-glutamyloglycylglycine + 5-amino-2-nitrobenzoate working reagent 1000 µl
Bring up to the temperature of determination. Then add:
sample 100 µl
The rate of 5-amino-2-nitrobenzoate creation measured colorimetrically
Mix and incubate at adequate temperature. After about 1 min. read
is directly proportional to γ-glutamyltransferase activity.
the absorbance against air or water. Repeat the reading after exactly
1, 2 and 3 minutes. Calculate the mean absorbance change per minute
REAGENTS
(∆A/min.).
Package
Liquick Cor- Liquick Cor- Liquick Cor-
Calculation
GGT mini GGT 30 GGT 60
1-GGT 2 x 24 ml 5 x 24 ml 5 x 48 ml GGT activity [U/l] = ∆A/min. x 1158
2-GGT 1 x 12 ml 1 x 30 ml 1 x 60 ml 1 U/l = 0.0167 µkat/l

Reagent Start method


The reagents when stored at 2-8°C are stable up to expiry date
The determination can be also performed with use of separate 1-GGT
printed on the package. The reagents are stable for 12 weeks on
and 2-GGT reagents.
board the analyser at 2-10°C. Protect from light and avoid
Pipette into the cuvette:
contamination!
1-GGT 1000 µl
Bring up to the temperature of determination. Then add:
Working reagent preparation and stability
Assay can be performed with use of separate 1-GGT and 2-GGT sample 100 µl
reagents or with use of working reagent. For working reagent Mix well, incubate for 1 min. Then add:
preparation mix gently 4 parts of 1-GGT with 1 part of 2-GGT. 2-GGT 250 µl
Avoid foaming. Mix well; perform measurement as described for Sample Start
Stability of working reagent in darkness: 4 weeks at 2-8°C method.
5 days at 15-25°C
Protect from light and avoid contamination! Calculation
GGT activity [U/l] = ∆A/min. x 1421
Concentrations in the test 1 U/l = 0.0167 µkat/l
Tris (pH 8.25) 100 mmol/l
glycylglycine 100 mmol/l REFERENCE VALUES 6
L-γ-glutamyl-3-carboxy-4-nitroanilide 4 mmol/l serum / plasma 37ºC
female 7 – 32 U/l (0.117 – 0.533 µkat/l)
Warnings and notes male 11 – 49 U/l (0.183 – 0.817 µkat/l)
 Product for in vitro diagnostic use only. It is recommended for each laboratory to establish its own reference
ranges for local population.
ADDITIONAL EQUIPMENT
 automatic analyzer or photometer able to read at 405 nm; QUALITY CONTROL
 thermostat at 25ºC, 30ºC or 37ºC; For internal quality control it is recommended to use the CORMAY
 general laboratory equipment; SERUM HN (Cat. No 5-172) and CORMAY SERUM HP (Cat. No
5-173) with each batch of samples.
SPECIMEN For the calibration of automatic analysers systems the CORMAY
Serum, EDTA plasma free from haemolysis. MULTICALIBRATOR LEVEL 1 (Cat. No 5-174; 5-176) or
Do not use citrate, oxalate and fluoride as anticoagulants because of LEVEL 2 (Cat. No 5-175; 5-177) is recommended.
GGT activity inhibition! The calibration curve should be prepared every 12 weeks, with
Heparin causes turbidity in the reaction mixture! change of reagent lot number or as required e.g. quality control
findings outside the specified range.

Liquick Cor-GGT page 1


PERFORMANCE CHARACTERISTICS LITERATURE
These metrological characteristics have been obtained using the 1. Szasz G., Weimann G., Suhler F., Wahlefrld A.W., Persijn J.P. :
automatic analyser Biolis 24i Premium. Results may vary if Z. Klin. Chem. Klin. Biochem. 12, 228 (1974).
a different instrument or a manual procedure is used. 2. Persijn J.P., van der Silk W.: J. Clin. Chem. Clin. Biochem. 14,
421-427 (1976).
 Sensitivity: 5.1 U/l (0.085 µkat/l). 3. Burtis C.A., Ashwood E.R., ed. Tietz Textbook of Clinical
Chemistry, 2nd ed. Philadelphia, PA: WB Saunders, 850-1,
 Linearity: up to 500 U/l (8.33 µkat/l). (1994).
4. Tietz N.W., ed. Clinical Guide to Laboratory Tests, 3rd ed.
 Specificity / Interferences Philadelphia, PA: WB Saunders, 286 (1995).
Haemoglobin up to 0.16 g/dl, ascorbate up to 62 mg/l, bilirubin 5. Kaplan L.A., Pesce A.J.: Clinical Chemistry. Theory, analysis
up to 20 mg/dl and triglycerides up to 1000 mg/dl do not and correlation 3rd Ed., the C.V. Mosby Company, St. Louis
interfere with the test. 1996, p.1072.
6. Alan H.B. Wu: Tietz Clinical Guide to Laboratory Tests, 4th ed.
 Precision WB Saunders, 470, (2006).
Repeatability (run to run) Mean SD CV
n = 20 [U/l] [U/l] [%]
level 1 29.06 1.44 4.97 Date of issue: 04. 2013.
level 2 134.05 1.86 1.39

Reproducibility (day to day) Mean SD CV


n = 80 [U/l] [U/l] [%]
level 1 29.05 2.14 7.36
level 2 132.33 3.50 2.64

 Method comparison
A comparison between GGT values determined at Biolis 24i
Premium (y) and at ADVIA 1650 (x) using 50 samples gave
following results:
y = 0.9791 x - 1.322 U/l;
R = 0.9990 (R – correlation coefficient)

WASTE MANAGEMENT
Please refer to local legal requirements.

MANUFACTURER

PZ CORMAY S.A.
22 Wiosenna Street,
05-092 Łomianki, POLAND
tel.: +48 (0) 22 751 79 10