General Imaging
Site Preparation

LOGIQ E9 LOGIQ S8 Invenia Abus 2.0 LOGIQ E10 LOGIQ S7 LOGIQ P7 & P9
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.

1. Secretaries or Doctors Desk


2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer - Dual Level
Lighting (bright and dim)
1300 mm / 51 inch

7. Emergency Oxygen
8. Suction Line
Height *:

9. LOGIQ E10
10. Dedicated Power Outlet - Circuit Breaker
protected and easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of LOGIQ E10
from wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and Equipment
16. Examination Table – 1930 x 610 mm (76 x
24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
126 kg
278.0 lbs.

* Dimensions given with floating keyboard stowed for transport and the Main Monitor down. Min. Floor Plan Need UPS ?

External Inputs and Outputs : Electrical Specifications


100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Environmental Requirements Voltage At 230 VAC -> up to 8A , at 100 VAC—> up to 10A

Operating Temperature 10 to 35 C (50 to 96F) Power Consumption 1 KVA

Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 70 –106 kPa A dedicated AC power outlet must be within reach of the unit without
Site Power Outlets extension cords
3070 BTU/hour + 300 Btu for each person. 264 VAC, 50 Hz -> 6A, 60 Hz -> 5A
Heat Dissipation Without lights, other equipment etc. 220 VAC, 50 Hz -> 5.5A , 60 Hz -> 6A
Inrush current 110 VAC, 50 Hz -> 9A, 60 Hz -> 11A
Storage Temperature -10 to 50 C (-14 to 122 F) , (These values are estimations) 90 VAC, 50 Hz -> 13A, 60 Hz -> 12A
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 30% - 80 % rH, non-condensing , Pressure : 70 –106 kPa possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality.
Acoustic Noise output Less than 48 dB(A)
3 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.

1. Secretaries or Doctors Desk


2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer - Dual Level
Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. LOGIQ E9
10. Dedicated Power Outlet - Circuit Breaker
protected and easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of LOGIQ E9
from wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and Equipment
16. Examination Table – 1930 x 610 mm (76 x
24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

135 kg
298.0 lbs.

Min. Floor Plan Need UPS ?

External Inputs and Outputs : Electrical Specifications


100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Voltage At 230 VAC -> up to 8A , at 100 VAC—> up to 10A
Environmental Requirements Power Consumption 1100 W
Operating Temperature 10 to 35 C (50 to 96F) A dedicated AC power outlet must be within reach of the unit without
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 70 –106 kPa Site Power Outlets extension cords
264 VAC, 50 Hz -> 6A, 60 Hz -> 5A
4712 BTU/hour + 300 Btu for each person.
Heat Dissipation Without lights, other equipment etc. 220 VAC, 50 Hz -> 5.5A , 60 Hz -> 6A
110 VAC, 50 Hz -> 9A, 60 Hz -> 11A
Storage Temperature -10 to 50 C (-14 to 122 F) , Inrush current 90 VAC, 50 Hz -> 13A, 60 Hz -> 12A
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 30% - 80 % rH, non-condensing , Pressure : 70 –106 kPa possible loss of critical care equipment, make sure you DO NOT have any
Power outage may occur other equipment operating on the same circuit.
Combination lighting system (dim/bright) is recommended. Keep in
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise output Less than 48 dB(A)
4 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.

1. Secretaries or Doctors Desk


2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer - Dual Level
Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. LOGIQ S8
10. Dedicated Power Outlet - Circuit Breaker
protected and easily accessible
11. Network Interface
520 12. 457 mm (18 inches) distance of LOGIQ S8
from wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and Equipment
16. Examination Table – 1930 x 610 mm (76
x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

85 kg
187.4 Ibs

Min. Floor Plan Need UPS ?


External Inputs and Outputs :
Electrical Specifications
Environmental Requirements 100 - 120 VAC ±10% 9.0 ... 7.0 A 50/60 Hz (±2%)
Voltage 220 - 240 VAC ±10% 4.0 ... 3.7 A 50/60 Hz (±2%)
Operating Temperature 10 to 35 C (50 to 96F)
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa Power Consumption 900VA
680 BTU/hour + 300 Btu for each person. A dedicated AC power outlet must be within reach of the unit without
Heat Dissipation Without lights, other equipment etc. Site Power Outlets extension cords
Inrush current is not a factor to consider due to the inrush current
Storage Temperature -10 to 50 C (-14 to 122 F) , Inrush current limiting properties of the power supplies.
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise output Less than 50 dB(A)

5 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer - Dual Level
Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. LOGIQ S7
10. Dedicated Power Outlet - Circuit Breaker
protected and easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of LOGIQ E9
from wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and Equipment
16. Examination Table – 1930 x 610 mm (76 x
24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

90 kg
198.4 Ibs

Min. Floor Plan Need UPS ?


External Inputs and Outputs :
Electrical Specifications
Environmental Requirements 100 - 120 VAC ±10% 9.0 ... 7.0 A 50/60 Hz (±2%)
Voltage 220 - 240 VAC ±10% 4.0 ... 3.7 A 50/60 Hz (±2%)
Operating Temperature 10 to 35 C (50 to 96F)
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa Power Consumption 900VA
680 BTU/hour +300 Btu for each person. A dedicated AC power outlet must be within reach of the unit without
Heat Dissipation Without lights, other equipment etc. Site Power Outlets extension cords
Inrush current is not a factor to consider due to the inrush current
Storage Temperature -10 to 50 C (-14 to 122 F) ,
Inrush current limiting properties of the power supplies.
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa
possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise output Less than 50 dB(A)

6 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Floor Plan 4.3m x 5.2m (14ft x 17ft)
Please Note: The picture may vary according to order configuration.

60 kg
132.3 Ibs

Need UPS ?
External Inputs and Outputs : Electrical Specifications
Environmental Requirements 100 - 120 VAC ±10% 5.0 ... 4.1 A 50, 60 Hz (±2%)
Voltage 220 - 240 VAC ±10% 2.2 ... 2.0 A 50, 60 Hz (±2%)
Operating Temperature 10 to 35 C (50 to 96F)
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa Power Consumption 500VA
680 BTU/hour +300 Btu for each person. A dedicated AC power outlet must be within reach of the unit without
Heat Dissipation Without lights, other equipment etc. Site Power Outlets extension cords
Inrush current is not a factor to consider due to the inrush current
Storage Temperature -10 to 50 C (-14 to 122 F) , Inrush current limiting properties of the power supplies.
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise output Less than 50 dB(A)

7 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight :
2.0 Please Note: The picture may vary according to order configuration.

103 kg
271.0 lbs.

Site Survey Planner : Please Note : Please check next


To help the customer with planning activities, fill out the Site Survey Planner. page for the floor plane.
You can download this from MyWorkShop, DOC1864913.
Need UPS ?
Electrical Specifications Environmental Requirements
Operating Temperature 5 to 35 C (41 to 95 F)
Voltage 100 –240 VAC +- %10 / 10 A / 50, 60 Hz ( +-2%)
Operating Humidity 30 to 75 % rH, non-condensing

Power Consumption Max 2080 W Heat Dissipation 3800 BTU/hour, + 300 Btu for each person.
A dedicated AC power outlet must be within reach of the unit without
Site Power Outlets extension cords Storage Temperature -10 to 60 C (14 to 140 F)
Requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any Storage Humidity 10—85 % rH, non-condensing
Power outage may occur other equipment operating on the same circuit. Combination lighting system (dim/bright) is recommended. Keep in mind that
lighting controls and dimmers can be a source of EMI which could degrade
Lighting image quality.

8 Please Note: For more information please refer to the product’s service & user manuals.
Scan Station Room Layout

The Invenia ABUS Scan Station can be positioned at the head of the table (Position
A), or at an angle (Position B). Position A is the preferred layout because it allows
the hinge on the Display Arm to function smoothly. Position B is acceptable, but
significantly reduces the Display Arm range of motion.

1. Maximum arm reach is ~1 m, or 39”


2. ~0.76 m, or 30” clear space recommended
Example layout in the room 3. ~0.61 m, or 24” clear space recommended

Position A – Desired Footprint

Position B– Desired Footprint

Please Note: For more information please refer to the product’s service & user manuals.
Cardiovascular
Site Preparation

VIVID E80/E90/E95 VIVID S60/S70 VIVID T8 VIVID IQ


Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
141.0 cm / 55.5 inch

7. Emergency Oxygen
8. Suction Line
Height :

9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
Width 15. Storage for Linens and
128 kg
Equipment
283.0 lbs. 58.5 cm / 23 inch
16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

+ Echopac Floor Plan Min. Floor Plan Need UPS ?


External Inputs and Outputs :
Electrical Specifications
100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Environmental Requirements Voltage At 230 VAC -> up to 3A , at 100 VAC—> up to 7A
Operating Temperature 10 to 35 C (50 to 95 F)
Operating Humidity 30 to 85 % rH, non-condensing
Power Consumption 700 W, 770 VA, including all on board peripherals.
2390 BTU/hour + 300 Btu for each person. A dedicated AC power outlet must be within reach of the unit without
Heat Dissipation Without on-board peripherals Site Power Outlets extension cords
264 VAC, 50 Hz -> 13A, 60 Hz -> 12A
Storage Temperature -20 to 60 C (-4 to 140 F) 220 VAC, 50 Hz -> 9A , 60 Hz -> 11A
110 VAC, 50 Hz - > 5.5A, 60 Hz -> 6 A
Storage Humidity 30% - 95 % rH, non-condensing Inrush current 90 VAC, 50 Hz -> 6A, 60 Hz -> 5A
Combination lighting system (dim/bright) is recommended. Keep in Requires a dedicated single branch circuit. To avoid circuit overload and
mind that lighting controls and dimmers can be a source of EMI which possible loss of critical care equipment, make sure you DO NOT have any
Lighting could degrade image quality. Power outage may occur other equipment operating on the same circuit.

11 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm (76
x 24 inches)
75 kg 17. Lavatory and Dressing Room
165.0 lbs. 18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : + Echopac Floor Plan Min. Floor Plan Need UPS ?

Electrical Specifications
Environmental Requirements
100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Operating Temperature 10 to 35 C (50 to 95 F) Voltage At 230 VAC -> up to 3A , at 100 VAC—> up to 7A

Operating Humidity 30 to 85 % rH, non-condensing / 700— 1060 hPa Power Consumption 500 VA
2000 BTU/hour +300 Btu for each person. Dedicated AC power outlet must be within reach of the unit without extension
Heat Dissipation Without on-board peripherals Site Power Outlets cords
100 V— 4.5A
Storage Temperature -20 to 60 C (-4 to 140 F) Inrush current 240 V—2.3 A
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 70 % rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any other
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise Level Less than 55 dB(A)

12 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.

1. Secretaries or Doctors Desk


2. File Cabinet
3. Film Viewer
65 kg 4. Counter Top
144.0 lbs. 5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : + Echopac Floor Plan Min. Floor Plan Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 3 to 40 C ( 38 to 104 F)
Operating Humidity 30 to 80 % rH, non-condensing / 700— 1060 hPa Voltage 100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
3800 BTU/hour + 300 Btu for each person. Power Consumption 400 VA
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C ( 23 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality.
Acoustic Noise Level Less than 39 dB(A)

13 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
VIVID IQ ; 2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
5 kg
11 Ibs. 6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
Cart ; 10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
45 kg Equipment
99 lbs. 16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : + Echopac Floor Plan Min. Floor Plan Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 3 to 40 C ( 38 to 104 F)
Voltage 100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Operating Humidity 30 to 80 % rH, non-condensing / 700— 1060 hPa
3800 BTU/hour + 300 Btu for each person. Power Consumption VIVID IQ : Max 150 VA - Cart : 750 VA
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C ( 23 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise Level VIVID IQ : 47 dB(A) VIVID IQ Cart : 43 dB(A)

14 Please Note: For more information please refer to the product’s service & user manuals.
Suggested floor plan for Echopac & Ultrasound system

1. Workstation parts 4. 3x mains power outlets


2. UPS 5. Hot and Cold water
3. Ethernet network wall outlet 6. Dedicated mains power outlet
7. Ethernet network wall outlet
Minimal floor plan suggestion ;

Scale: Each square equals one square foot (app. 31 x 31 cm)

Minimal floor plan,


2.5 m x 3 m (8 by 10 foot)

1. Door – at least 762 mm (30 inches) 9. Ultrasound system


2. Film Viewer 10. External Peripherals
3. Counter Top, Sink with hot and cold water and 11. Dedicated Power Outlet -
Supplies Storage Circuit Breaker protected and easily accessible
4. Linen Supply 12. Network Interface
5. Probes/Supplies 13. 457 mm (18 inches) distance
6. Examination Table – 1930 x610 mm (76 x 24 inches) of Ultrasound system from wall or objects
7. Footswitch 14. GE Cabinet for Software and
8. Stool Manuals

16
Women's Health
Site Preparation

VOLUSON E10/E8/E6 VOLUSON S10 VOLUSON S6/S8 VOLUSON P6/P8


Physical Dimensions and Weight (mm): Recommended Ultrasound Room Layout
Please Note: The picture may vary according to order configuration.

1. door – at least 76 cm (2.5 foot /


30 inches)
2. film viewer
3. counter top, sink with hot and
cold water, supplies storage
4. linen supply
5. probes/supplies
6. examination table
7. footswitch
8. stool
9. ultrasound system
10. external peripherals
11. dedicated power outlet - circuit
breaker protected and easily
accessible
12. network interface
13. distance from wall or objects
14. cabinet for software and
manuals
150 kg
330.0 lbs.

External Inputs and Outputs : Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 10 to 30 C (50 to 86 F)
Voltage 100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Operating Humidity 30 to 80% rH, non-condensing
2730 BTU/hour +300 Btu for each person. Power Consumption 800 VA, including all on board peripherals.
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -10 to 50 C (14 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity < 90% rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise Output max. 60 dB(A)

18 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.3m x 5.2m (14ft x 17ft)
Please Note: The picture may vary according to order configuration.

90 kg
200.0 lbs.

External Inputs and Outputs : Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 18 to 30 C (64 to 86 F) 100 –120 VAC +- %10 / 9.0….7.0 A / 50, 60 Hz (2%)
Voltage 220 - 240 VAC +-%10 / 4.0….3.7 A / 50, 60 Hz (2%)
Operating Humidity 30 to 80% rH, non-condensing
820 BTU/hour + 300 Btu for each person. Power Consumption 900 VA, including all on board peripherals.
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -10 to 50 C (14 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 to 90% rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality.
Acoustic Noise Output max. 55 dB(A)

19 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.3m x 5.2m (14ft x 17ft)
Please Note: The picture may vary according to order configuration.

90 kg
200.0 lbs.

External Inputs and Outputs : Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 18 to 30 C (64 to 86 F) 100 –120 VAC +- %10 / 9.0….7.0 A / 50, 60 Hz (2%)
Voltage 220 - 240 VAC +-%10 / 4.0….3.7 A / 50, 60 Hz (2%)
Operating Humidity 30 to 80% rH, non-condensing
S6:680 BTU/Hour , S8:780 BTU/hour +300 Btu for each person. Power Consumption 900 VA, including all on board peripherals.
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -10 to 50 C (14 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 to 90% rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality.
Acoustic Noise Output max. 55 dB(A)
20 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.3m x 5.2m (14ft x 17ft)
Please Note: The picture may vary according to order configuration.

69 kg
200.0 lbs.

High Version

68 kg
150 Ibs.
Low Version

External Inputs and Outputs : Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 18 to 30 C (64 to 86 F) 100 –120 VAC +- %10 / 9.0….7.0 A / 50, 60 Hz (2%)
Voltage 220 - 240 VAC +-%10 / 4.0….3.7 A / 50, 60 Hz (2%)
Operating Humidity 30 to 80% rH, non-condensing
1200 BTU/hour +300 Btu for each person.
Heat Dissipation Without on-board peripherals
Power Consumption 900 VA, including all on board peripherals.
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -10 to 50 C (14 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 to 90% rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise Output max. 55 dB(A)

21 Please Note: For more information please refer to the product’s service & user manuals.
Primary Care
Site Preparation

LOGIQ F6 & F8 LOGIQ V3 & V5 Versana Premier Versana Essential LOGIQ V1 & V2
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
65 kg
144 Ibs
16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : Min. Floor Plan Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 3 to 40 C (38 to 104 F)
Voltage 100-240 VAC ±10% 50/60 Hz
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa
3800 BTU/hour + 300 Btu for each person. Power Consumption Max 400 VA
Heat Dissipation Without lights, other equipment etc.
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C (-23 to 122 F) , Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which Acoustic Noise output 47 dB(A)
Lighting could degrade image quality.

23 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm: Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.

V3 with 15inch display 1. Secretaries or Doctors Desk


V5 with 15/17/19inch display 2. File Cabinet
V5 Expert with 15/17/19inch display 3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
48 kg
106 Ibs 17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : Min. Floor Plan Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 3 to 40 C (38 to 104 F)
Voltage 100-240 VAC ±10% 50/60 Hz
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa
3800 BTU/hour + 300 Btu for each person. Power Consumption Max 300 VA
Heat Dissipation Without lights, other equipment etc.
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C (-23 to 122 F) , Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality.
Acoustic Noise output 42 dB(A)

24 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
36 kg
79 Ibs (76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : Min. Floor Plan Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 3 to 40 C (38 to 104 F)
Voltage 100-240 VAC ±10% 50/60 Hz
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa
3800 BTU/hour + 300 Btu for each person. Power Consumption Max 300 VA
Heat Dissipation Without lights, other equipment etc.
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C (-23 to 122 F) , Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting Acoustic Noise output 47 dB(A)
could degrade image quality.

25 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
“Flexible LCD Arm vertical lift: 150mm & Flexible Neck vertical lift: 160mm” 1. Secretaries or Doctors Desk
2. File Cabinet
fixed Neck and fixed LCD Arm :
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
70 kg (76 x 24 inches)
154.3 Ibs 17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : Min. Floor Plan Need UPS ?

Environmental Requirements Electrical Specifications


3 to 40 C without battery (37.4 to 104 F)
Operating Temperature 10—30 C with battery (50-86 F ) Voltage 100-240 VAC ±10% 50/60 Hz
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa
3800 BTU/hour + 300 Btu for each person. Power Consumption Max 450 VA
Heat Dissipation Without lights, other equipment etc.
A dedicated AC power outlet must be within reach of the unit without
Site Power Outlets extension cords
Storage Temperature -5 to 50 C (-23 to 122 F) ,
Requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa
Power outage may occur other equipment operating on the same circuit.
Combination lighting system (dim/bright) is recommended. Keep in
mind that lighting controls and dimmers can be a source of EMI which Acoustic Noise output 43 dB(A)
Lighting could degrade image quality.

26 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.

LOGIQ V1 & V2 ; 1. Secretaries or Doctors Desk


2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
6 kg 8. Suction Line
13.23 Ibs. 9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
V Cart ; 11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
40 kg
(76 x 24 inches)
88.2 lbs. 17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : Min. Floor Plan Need UPS ?

Environmental Requirements
Electrical Specifications
Operating Temperature 10 to 40 C ( 38 to 104 F)
Operating Humidity 30 to 80 % rH, non-condensing / 700— 1060 hPa
Voltage 100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
3800 BTU/hour + 300 Btu for each person. Power Consumption LOGIQ V1 & V2 : Max 200 VA - Cart : VA
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C ( 23 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in
Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which Acoustic Noise Level 54 dB(A)
Lighting could degrade image quality.

27 Please Note: For more information please refer to the product’s service & user manuals.
Point of Care
Site Preparation

VENUE VENUE 50 LOGIQ e


Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
63 kg (76 x 24 inches)
139 Ibs 17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)

External Inputs and Outputs : Min. Floor Plan Need UPS ?

Environmental Requirements Electrical Specifications


Operating Temperature 10 to 40 C (50 to 104 F) Voltage 100-240 VAC ±10% 50/60 Hz
Operating Humidity 30 to 85 % rH, non-condensing , Pressure : 700 –1060 hPa 240 VAC the current may be up to 2.08 A
2000 BTU/hour + 300 Btu for each person. Current 100 VAC the current may be up to 5 A
Heat Dissipation Without lights, other equipment etc. Power Consumption Max 500 VA

Storage Temperature -20 to 60 C (-4 to 122 F) , A dedicated AC power outlet must be within reach of the unit without
Site Power Outlets extension cords
Storage Humidity 10 % - 70 % rH, non-condensing , Pressure : 700 –1060 hPa Requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise output Less than 55 dB(A)

29 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
VENUE 50 ;
1. Secretaries or Doctors Desk
2. File Cabinet 200.0 lbs.
3. Film Viewer
4. Counter Top
4 kg 5. Counter Top and Sink with hot and cold
8.8 lbs. water
6. Overhead Lights Dimmer -
Docking Cart : Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
28.5 kg
62.8 lbs.
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
Docking Station : 15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
4.6 kg 17. Lavatory and Dressing Room
10.1 lbs. 18. Door – at least 762 mm (30 inches)

Min. Floor Plan Need UPS ?


External Inputs and Outputs :
Environmental Requirements Electrical Specifications
Operating Temperature 3 to 40 C (37 to 104 F) Voltage Venue 50 : 100-240 VAC ±10% 50/60 Hz , 1.6A Max
Docking Cart: 100-240 VAC ±10% 50/60 Hz , 2.7A Max
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa
Venue 50 : Max 180 VA
3800 BTU/hour + 300 Btu for each person. Power Consumption Docking Cart : Max 380 VA
Heat Dissipation Without lights, other equipment etc.
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C (23 to 122 F) , Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa
possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise output Less than 55 dB(A)

30 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.

LOGIQ e ;

5.2 kg
11.5 lbs.

Docking Cart :

53 kg
116.8 lbs.

Min. Floor Plan Need UPS ?


External Inputs and Outputs :
Environmental Requirements Electrical Specifications
Voltage LOGIQ e R7 : 100-240 VAC ±10% 50/60 Hz
Operating Temperature 10 to 40 C (50 to 104 F) Docking Cart: 100-240 VAC ±10% 50/60 Hz
Operating Humidity 30 to 75 % rH, non-condensing , Pressure : 700 –1060 hPa LOGIQ e R7 : Max 180 VA
3800 BTU/hour + 300 Btu for each person. Power Consumption Docking Cart : More than or equal to 750 VA
Heat Dissipation Without lights, other equipment etc.
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C (23 to 122 F) , Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa possible loss of critical care equipment, make sure you DO NOT have any
Power outage may occur other equipment operating on the same circuit.
Combination lighting system (dim/bright) is recommended. Keep in
mind that lighting controls and dimmers can be a source of EMI which Less than 55 dB(A)
Lighting could degrade image quality. Acoustic Noise output

31 Please Note: For more information please refer to the product’s service & user manuals.
UPS (Uninterruptible Power Supply)
UPS (Uninterruptible Power Supply) for 100-130V AC countries
Manufacturer : Tripp Lite Model : SMART1200XLHG
Tripp Lite’s SMART1200XLHG medical-grade UPS system provides dependable battery backup and
is fully compliant with UL 60601-1 regulations when plugged into a hospital-grade outlet, making it
ideal for use in patient-care vicinities. The 1kVA/750W capacity provides up to 28 minutes of
runtime for a half load or 10 minutes for a full load, so you can work through short power failures
and perform an orderly system shutdown in a prolonged blackout.

PART NO :H48691PD

UPS (Uninterruptible Power Supply) for 220-240V AC countries


Manufacturer : Tripp Lite Model : SMX1200XLHG
Tripp Lite’s SMX1200XLHG medical-grade UPS system provides dependable battery backup and
is fully compliant with UL 60601-1 regulations when plugged into a hospital-grade outlet, making
it ideal for use in patient-care vicinities. The 1kVA/750W capacity provides up to 27.5 minutes of
runtime for a half load, or 11 minutes for a full load so you can work through short power failures
and perform an orderly system shutdown in a prolonged blackout.

PART NO :H48691PE
PLEASE NOTE : Before ordering any UPS, check and compare UPS Capacity and Ultrasound
Power consumption If you need different model or more information please visit the link :
https://www.tripplite.com

PLEASE CONTACT : if you want to order UPS through Ge Healthcare, please contact your sales representative.