Professional Documents
Culture Documents
Click On Above Icons To Reach The Pages - : Uptodate? Uptodate?
Click On Above Icons To Reach The Pages - : Uptodate? Uptodate?
Click On Above Icons To Reach The Pages - : Uptodate? Uptodate?
Up to date ? Up to date ?
Outside GE Inside GE
LOGIQ E9 LOGIQ S8 Invenia Abus 2.0 LOGIQ E10 LOGIQ S7 LOGIQ P7 & P9
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
7. Emergency Oxygen
8. Suction Line
Height *:
9. LOGIQ E10
10. Dedicated Power Outlet - Circuit Breaker
protected and easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of LOGIQ E10
from wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and Equipment
16. Examination Table – 1930 x 610 mm (76 x
24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
126 kg
278.0 lbs.
* Dimensions given with floating keyboard stowed for transport and the Main Monitor down. Min. Floor Plan Need UPS ?
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 70 –106 kPa A dedicated AC power outlet must be within reach of the unit without
Site Power Outlets extension cords
3070 BTU/hour + 300 Btu for each person. 264 VAC, 50 Hz -> 6A, 60 Hz -> 5A
Heat Dissipation Without lights, other equipment etc. 220 VAC, 50 Hz -> 5.5A , 60 Hz -> 6A
Inrush current 110 VAC, 50 Hz -> 9A, 60 Hz -> 11A
Storage Temperature -10 to 50 C (-14 to 122 F) , (These values are estimations) 90 VAC, 50 Hz -> 13A, 60 Hz -> 12A
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 30% - 80 % rH, non-condensing , Pressure : 70 –106 kPa possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality.
Acoustic Noise output Less than 48 dB(A)
3 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
135 kg
298.0 lbs.
85 kg
187.4 Ibs
5 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer - Dual Level
Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. LOGIQ S7
10. Dedicated Power Outlet - Circuit Breaker
protected and easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of LOGIQ E9
from wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and Equipment
16. Examination Table – 1930 x 610 mm (76 x
24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
90 kg
198.4 Ibs
6 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Floor Plan 4.3m x 5.2m (14ft x 17ft)
Please Note: The picture may vary according to order configuration.
60 kg
132.3 Ibs
Need UPS ?
External Inputs and Outputs : Electrical Specifications
Environmental Requirements 100 - 120 VAC ±10% 5.0 ... 4.1 A 50, 60 Hz (±2%)
Voltage 220 - 240 VAC ±10% 2.2 ... 2.0 A 50, 60 Hz (±2%)
Operating Temperature 10 to 35 C (50 to 96F)
Operating Humidity 30 to 80 % rH, non-condensing , Pressure : 700 –1060 hPa Power Consumption 500VA
680 BTU/hour +300 Btu for each person. A dedicated AC power outlet must be within reach of the unit without
Heat Dissipation Without lights, other equipment etc. Site Power Outlets extension cords
Inrush current is not a factor to consider due to the inrush current
Storage Temperature -10 to 50 C (-14 to 122 F) , Inrush current limiting properties of the power supplies.
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing , Pressure : 700 –1060 hPa possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise output Less than 50 dB(A)
7 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight :
2.0 Please Note: The picture may vary according to order configuration.
103 kg
271.0 lbs.
Power Consumption Max 2080 W Heat Dissipation 3800 BTU/hour, + 300 Btu for each person.
A dedicated AC power outlet must be within reach of the unit without
Site Power Outlets extension cords Storage Temperature -10 to 60 C (14 to 140 F)
Requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any Storage Humidity 10—85 % rH, non-condensing
Power outage may occur other equipment operating on the same circuit. Combination lighting system (dim/bright) is recommended. Keep in mind that
lighting controls and dimmers can be a source of EMI which could degrade
Lighting image quality.
8 Please Note: For more information please refer to the product’s service & user manuals.
Scan Station Room Layout
The Invenia ABUS Scan Station can be positioned at the head of the table (Position
A), or at an angle (Position B). Position A is the preferred layout because it allows
the hinge on the Display Arm to function smoothly. Position B is acceptable, but
significantly reduces the Display Arm range of motion.
Please Note: For more information please refer to the product’s service & user manuals.
Cardiovascular
Site Preparation
7. Emergency Oxygen
8. Suction Line
Height :
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
Width 15. Storage for Linens and
128 kg
Equipment
283.0 lbs. 58.5 cm / 23 inch
16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
11 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm (76
x 24 inches)
75 kg 17. Lavatory and Dressing Room
165.0 lbs. 18. Door – at least 762 mm (30 inches)
External Inputs and Outputs : + Echopac Floor Plan Min. Floor Plan Need UPS ?
Electrical Specifications
Environmental Requirements
100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
Operating Temperature 10 to 35 C (50 to 95 F) Voltage At 230 VAC -> up to 3A , at 100 VAC—> up to 7A
Operating Humidity 30 to 85 % rH, non-condensing / 700— 1060 hPa Power Consumption 500 VA
2000 BTU/hour +300 Btu for each person. Dedicated AC power outlet must be within reach of the unit without extension
Heat Dissipation Without on-board peripherals Site Power Outlets cords
100 V— 4.5A
Storage Temperature -20 to 60 C (-4 to 140 F) Inrush current 240 V—2.3 A
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 70 % rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any other
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise Level Less than 55 dB(A)
12 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
External Inputs and Outputs : + Echopac Floor Plan Min. Floor Plan Need UPS ?
13 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
VIVID IQ ; 2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
5 kg
11 Ibs. 6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
Cart ; 10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
45 kg Equipment
99 lbs. 16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
External Inputs and Outputs : + Echopac Floor Plan Min. Floor Plan Need UPS ?
14 Please Note: For more information please refer to the product’s service & user manuals.
Suggested floor plan for Echopac & Ultrasound system
16
Women's Health
Site Preparation
18 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.3m x 5.2m (14ft x 17ft)
Please Note: The picture may vary according to order configuration.
90 kg
200.0 lbs.
19 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.3m x 5.2m (14ft x 17ft)
Please Note: The picture may vary according to order configuration.
90 kg
200.0 lbs.
69 kg
200.0 lbs.
High Version
68 kg
150 Ibs.
Low Version
21 Please Note: For more information please refer to the product’s service & user manuals.
Primary Care
Site Preparation
LOGIQ F6 & F8 LOGIQ V3 & V5 Versana Premier Versana Essential LOGIQ V1 & V2
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
65 kg
144 Ibs
16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
23 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm: Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
24 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
1. Secretaries or Doctors Desk
2. File Cabinet
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
36 kg
79 Ibs (76 x 24 inches)
17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
25 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
“Flexible LCD Arm vertical lift: 150mm & Flexible Neck vertical lift: 160mm” 1. Secretaries or Doctors Desk
2. File Cabinet
fixed Neck and fixed LCD Arm :
3. Film Viewer
4. Counter Top
5. Counter Top and Sink with hot and cold
water
6. Overhead Lights Dimmer -
Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
70 kg (76 x 24 inches)
154.3 Ibs 17. Lavatory and Dressing Room
18. Door – at least 762 mm (30 inches)
26 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight (mm) : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
Environmental Requirements
Electrical Specifications
Operating Temperature 10 to 40 C ( 38 to 104 F)
Operating Humidity 30 to 80 % rH, non-condensing / 700— 1060 hPa
Voltage 100 –240 VAC +- %10 / 50, 60 Hz ( +-2%)
3800 BTU/hour + 300 Btu for each person. Power Consumption LOGIQ V1 & V2 : Max 200 VA - Cart : VA
Heat Dissipation Without on-board peripherals
A dedicated AC power outlet must be within reach of the unit without
Storage Temperature -5 to 50 C ( 23 to 122 F) Site Power Outlets extension cords
Requires a dedicated single branch circuit. To avoid circuit overload and
Storage Humidity 10 % - 90 % rH, non-condensing possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in
Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which Acoustic Noise Level 54 dB(A)
Lighting could degrade image quality.
27 Please Note: For more information please refer to the product’s service & user manuals.
Point of Care
Site Preparation
Storage Temperature -20 to 60 C (-4 to 122 F) , A dedicated AC power outlet must be within reach of the unit without
Site Power Outlets extension cords
Storage Humidity 10 % - 70 % rH, non-condensing , Pressure : 700 –1060 hPa Requires a dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you DO NOT have any
Combination lighting system (dim/bright) is recommended. Keep in Power outage may occur other equipment operating on the same circuit.
mind that lighting controls and dimmers can be a source of EMI which
Lighting could degrade image quality. Acoustic Noise output Less than 55 dB(A)
29 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
VENUE 50 ;
1. Secretaries or Doctors Desk
2. File Cabinet 200.0 lbs.
3. Film Viewer
4. Counter Top
4 kg 5. Counter Top and Sink with hot and cold
8.8 lbs. water
6. Overhead Lights Dimmer -
Docking Cart : Dual Level Lighting (bright and dim)
7. Emergency Oxygen
8. Suction Line
9. Ultrasound system
10. Dedicated Power Outlet -
Circuit Breaker protected and
easily accessible
11. Network Interface
28.5 kg
62.8 lbs.
12. 457 mm (18 inches) distance of
Ultrasound system from
wall or objects
13. Stool
14. Footswitch
Docking Station : 15. Storage for Linens and
Equipment
16. Examination Table – 1930 x 610 mm
(76 x 24 inches)
4.6 kg 17. Lavatory and Dressing Room
10.1 lbs. 18. Door – at least 762 mm (30 inches)
30 Please Note: For more information please refer to the product’s service & user manuals.
Physical Dimensions and Weight : Recommended Ultrasound Room Layout 4.27 x 5.18 m (14 x 17 foot)
Please Note: The picture may vary according to order configuration.
LOGIQ e ;
5.2 kg
11.5 lbs.
Docking Cart :
53 kg
116.8 lbs.
31 Please Note: For more information please refer to the product’s service & user manuals.
UPS (Uninterruptible Power Supply)
UPS (Uninterruptible Power Supply) for 100-130V AC countries
Manufacturer : Tripp Lite Model : SMART1200XLHG
Tripp Lite’s SMART1200XLHG medical-grade UPS system provides dependable battery backup and
is fully compliant with UL 60601-1 regulations when plugged into a hospital-grade outlet, making it
ideal for use in patient-care vicinities. The 1kVA/750W capacity provides up to 28 minutes of
runtime for a half load or 10 minutes for a full load, so you can work through short power failures
and perform an orderly system shutdown in a prolonged blackout.
PART NO :H48691PD
PART NO :H48691PE
PLEASE NOTE : Before ordering any UPS, check and compare UPS Capacity and Ultrasound
Power consumption If you need different model or more information please visit the link :
https://www.tripplite.com
PLEASE CONTACT : if you want to order UPS through Ge Healthcare, please contact your sales representative.
Thank you !
Channel Partners
Ge Healthcare