Professional Documents
Culture Documents
Stereotaxy - Operator Manual - UM - 5262169-8-1EN - 3
Stereotaxy - Operator Manual - UM - 5262169-8-1EN - 3
Operator manual
5262169-8-1EN
Revision 3
© 2008-2017 by General Electric Company
All Rights Reserved.
Senographe Essential Stereotaxy 5262169-8-1EN
IMPORTANT...X-RAY PROTECTION
IMPORTANT...X-RAY PROTECTION
CAUTION
If not properly used, x-ray equipment may cause injury. Accordingly, it is your obligation to
confirm that the instructions herein contained are thoroughly read and understood by everyone
who will use the equipment before you attempt to place this equipment in operation. The General
Electric Company, Healthcare Technologies, will be glad to assist and cooperate in placing this
equipment in use.
Although this apparatus incorporates a high degree of certain protections against x-radiation
other than the useful beam, no feasible design of equipment can provide complete protection
from all potential injury. Nor can any feasible design force the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves or others to
radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
knowledgeable about the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910
Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the International
Commission on Radiation Protection. It is your obligation and responsibility to take adequate
steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Healthcare
Technologies, its agents, and representatives have no responsibility for injury or damage, which
may result from improper use of the equipment. Various protective materials and devices are
available. It is urged that such materials or devices be used in accordance your site’s clinical
practice.
IMPORTANT...X-RAY PROTECTION
Table of Contents
Table of Contents
IMPORTANT...X-RAY PROTECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Table of Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Table of Contents
Table of Contents
Table of Contents
Publication presentation
1 Applicability
This manual is provided for operators using a Senographe Essential Acquisition System for Stereotaxy.
The manual has been written to describe the use of the system in its most complete configuration. If any
of the options described in this manual are not included in your system, skip the corresponding chapter
or section.
See your GE Representative for the options available with the Senographe Essential Acquisition
System.
Physical description
chapters.
Procedural description
chapters.
Functional description
chapters
Chapters are arranged in an order which is intended to suit the learning requirements of a new Operator.
This means that similar types of chapter are not necessarily grouped together.
Publication presentation
Publication presentation
Left 3 and 4 1
1 6
1 3
Middle 2 and 5 2
2 5 2
Right 1 and 6 3
(on the side
of the
trackball)
3-2 Typographical
• In text describing the Operator interface, italic and bold characters are used to differentiate between
body text and on-screen or hardware captions and labels, entered text, etc.:
- Italics are used for references to hardware items and text which appears on-screen (e.g., menu
titles, button labels, on-screen messages, etc.).
- Bold face is used when specific instructions are given for the entry of text or selection of a menu
option, etc. If the operation or text entry includes the use of a keyboard or mouse button, its name
is shown between angle brackets, to distinguish it from entered text, e.g., 123<Enter> would be
written for an instruction to type 123 followed by the Enter or Return key.
• In other text:
- Italics are used when a specific concept is introduced.
- Italic and bold characters may also be used for local emphasis.
- In electronic versions of the document, cross-references appear in blue italics.
Publication presentation
DANGER
Indicates an imminently hazardous situation that, if not avoided, will result in death or serious
injury.
WARNING
Indicates a potentially hazardous situation that, if not avoided, could result in death or serious
injury.
CAUTION
Indicates a potentially hazardous situation that, if not avoided, may result in minor or moderate
injury.
! Notice:
Used for instructions to the Operator to prevent damage to property.
Note:
Used to draw attention to information that is important for the Operator to know.
Publication presentation
4 Revision History
This table is intentionally left in English.
Publication presentation
1 Standards compliance
This product complies with the regulatory requirements of the following:
• If the 0459 label is affixed to the product, it complies with Council Directive 93/42/EEC concerning
medical devices.
European registered place of business:
GE MEDICAL SYSTEMS SCS
283, RUE DE LA MINIERE
78530 BUC France
• The first year of CE marking of this product is 2009.
• Code of Federal Regulations Title 21, Subchapter J - Radiological health.
• Underwriters' Laboratories, Inc. (UL), an independent testing laboratory.
• Canadian Standards Association (CSA).
• International Electrotechnical Commission (IEC), international standards organization, when
applicable, in particular IEC 60601-2-45:2001 compliance notice.
• USA/HHS: United States Federal law restricts this device to use by or on the order of a physician.
• The original document was written in English.
Use within the patient environment
All components of the Senographe Essential Stereotaxy system are designed to be suitable for use
within the patient environment, and are compliant with the relevant standards.
2 Installation requirements
Packing materials
The materials used to pack our equipment are recyclable. They must be collected and processed in
accordance with the regulations in force for the country where the machines or accessories are
unpacked.
Use in the presence of flammable gases
The Stereotactic Positioner must not be used in the presence of flammable gases.
Electromagnetic Compatibility (EMC)
This equipment complies with the IEC 60601-1-2 Edition 2 EMC standard for medical devices.
This equipment generates, uses, and can radiate radio frequency energy. The equipment may cause
radio frequency interference to other medical and non-medical devices and radio communications.
To provide reasonable protection against such interference, this product complies with emissions
limits for a Group 1, Class A Medical Devices Directive as stated in EN60601-1-2. However, there is
no guarantee that interference will not occur in a particular installation.
If this equipment is found to cause interference (which may be determined by turning the equipment
on and off), the user (or qualified service personnel) should attempt to correct the problem by one or
more of the following measure(s):
- Reorient or relocate the affected device(s).
- Increase the separation between the equipment and the affected device.
- Power the equipment from a source different from that of the affected device.
- Consult the point of purchase or service representative for further suggestions.
The manufacturer is not responsible for any interference caused by using other than recommended
interconnect cables or by unauthorized changes or modifications to this equipment. Unauthorized
changes or modifications could void the users' authority to operate the equipment.
To comply with the regulations on electromagnetic interference for a Class A FCC Device, all
interconnect cables to peripheral devices must be shielded and properly grounded. Use of cables not
properly shielded and grounded may result in the equipment causing radio frequency interference in
violation of the FCC regulations.
Do not use devices which intentionally transmit RF Signals (Cellular Phones, Transceivers, or Radio
Controlled Products) in the vicinity of this equipment as it may cause performance outside the
published specifications. Keep the power to these types of devices turned off when near this
equipment.
The medical staff in charge of this equipment is required to instruct technicians, patients, and other
people who may be around this equipment to comply fully with the above requirement.
3 Usage requirements
• The equipment includes a function IMAGE ANNOTATIONS, which allows the physician to store
information on the patient.
• Note that the European Directive regarding “the protection of the people with regard of data
management on their private life and to the free circulation of these data” requests the users of
computerized files (radiologists, physicians) not to store data related to:
- race,
- philosophical opinions,
- religious opinions,
- political opinions,
- etc.
4 Operator profile
The use of the Senographe Essential Stereotaxy positioning system requires specific technical and
medical knowledge and skills. As a minimum, operators must have acquired knowledge and skill in the
areas of:
• radiation protection,
• safety procedures,
• patient safety.
through approved training and professional education.
5 Training program
Before using the Senographe Essential Stereotaxy system to work with patients, it is mandatory that all
users have received adequate training. Contact your GE Representative for information on suitable
training programs.
For systems subject to Mammography Quality Standards Act (MQSA) rules, the Radiologist, Interpreting
Physician, and the Medical Physicist are each required to obtain eight hours of new-modality training for
full-field digital mammography.
Additional information can be found on the Food and Drug Administration’s (FDA's) website:
http://www.fda.gov
in the Mammography pages.
7 Meaning of symbols
Ground (earth)
Dangerous voltage
Type B equipment
This symbol indicates that waste electrical and electronic equipment must not be disposed
of as unsorted municipal waste and must be collected separately. Please contact an
authorized representative of the manufacturer for information concerning the
decommissioning of your equipment.
Name and address of the manufacturer
Date of manufacture
This symbol on the console indicates that you must have read and understood the
operator manual before taking any action.
1 Intended Use
Senographe Stereotaxy is an optional accessory for the Senographe full field digital system. It is
designed to allow the accurate location of lesions in the breast in three dimensions, using information
extracted from stereotactic pairs of two-dimensional images. It is intended to provide guidance for
interventional purposes (such as biopsy, pre-surgical localization or FNA).
Senographe Stereotaxy includes the Senographe Stereotactic Positioner with adaptors for vertical and/
or lateral approach biopsies, and the Stereotaxy software application.
2 Overview
It is assumed that the reader of this manual is familiar with the Senographe Essential Acquisition System
Operator manual.
• The Stereotaxy application is entered automatically when the Stereotactic Positioner is installed. It is
used to perform a stereotactic exam, to locate breast masses or micro-calcifications in three
dimensions. When it has been located, information of coordination is sent to the Stereotactic
Positioner, where it is used to accurately position a suitable needle to perform cytology/histology or
hook wire placement.
• Standard X-ray images are made with the X-ray beam
perpendicular to the image receptor; only 2-dimensional
information (X and Y coordinates) is available. In
Stereotaxy, the Senographe tube arm can be angled
independently of the image receptor, allowing the
calculation of 3-dimensional information (X, Y, Z) from
images taken at ±15° with respect to a line perpendicular
to the plane of the image receptor. This information is
then sent to the Stereotactic Positioner.
• The Stereotactic Positioner is a self-contained unit which
is mounted on the Senographe image receptor, in place
of the Bucky. It includes a compression paddle
(independent of the main compression system), used to
immobilize the breast during the procedure. The paddle
used for vertical approach procedures has an aperture to
allow the passage of needles; a paddle without aperture
is used for lateral approach procedures. The unit
includes a needle holder which accepts interchangeable
needle guides for accurate positioning of various types
of needles. Provision is made for the use of spring-
loaded and vacuum assisted biopsy devices.
• When an area of interest has been identified, and
located by the software, its 3-dimensional coordinates
are sent to the positioner, finally the needle holder is manually operated or motorized driven to the
exact required position which depends on the needle and device to be used . The needle is then
introduced through the needle guides, in the breast.
3 Principles
3-1 General
Stereotaxy is a process which allows you to determine the location (X, Y and Z coordinates) of an object
of interest (such as a suspected lesion), using a stereo pair of X-ray images. It is then possible to
accurately perform a puncture at this location.
0°
0°
+15° -15°
A B
• A single image (such as the scout image) taken with the tube arm at the 0° position, perpendicular to
the plane of the image receptor, can only show the X and Y coordinates of the projection of an object
on the image plane. See illustration A.
• If an image of the same object is taken with the tube arm at a different angle with respect to the
image receptor, the object appears to shift along a line called epipolar line. The shifted distance
along the epipolar line depends on the tube arm angle, and is proportional to the height (Z) of the
object above the image plane. See illustration B.
• Stereotaxy uses a stereo pair of images, taken with the tube arm angled to +15° and -15° with
respect to a line perpendicular to the image receptor.
• By measuring the position of the object on each of the two images of the stereo pair, its position in all
three dimensions (X, Y, and Z) can be computed.
• A scout view and an associated stereo pair may be referred to as a stereo triplet.
a a
As shown in the drawing, the lateral arm (b) simulates a needle having a length of (a) when
introduced vertically.
• It is the responsibility of the physician to decide which access is most suitable for each procedure,
depending on the position of the lesion and the biopsy device that will be used.
4 Compatibility
• The Stereotaxy option has been designed to be compatible with the needles and biopsy devices
listed below.
• Where appropriate, installation instructions for biopsy devices listed here are given in this manual.
• For the full operating instructions of biopsy devices, you must refer to the manufacturer’s instructions
for use. Contact the biopsy device manufacturer for ordering information.
Needle or Needle or Needle guide manufacturer Needle and needle guide dimensions
device device
manufacturer
BARD Vacora BARD (**) For compatible needle guide , - Needle: 10G.
(**) please refer to manufacturer - Plastic needle guides for metallic
user documentation of needle coaxial cannula (4 mm).
or device. - Plastic needle guides for plastic
coaxial cannula (4.5 mm).
- Other needles/gauges may be available.
(*) Perouse medical is not a distributor. For distribution informations, contact your GE Sales and
Service representative.
(**) See Trademark information on page 208.
Stereotaxy components
1 Overview
This chapter gives an introduction to components used in the Senographe Essential Stereotaxy system.
• The standard configuration of Senographe Stereotaxy includes:
- The Stereotactic Positioner, shown in section 2 Stereotactic Positioner on page 28, and fully
described in Chapter 5 The Stereotactic Positioner.
- Accessories provided with the Stereotactic Positioner, shown in section 3 Accessories supplied
with the Stereotactic Positioner on page 29. They include the aperture paddle, metal needle guide
adaptors for vertical and lateral approach, the calibration needle for vertical approach and a
phantom used for calibration and verification of accuracy, in vertical approach (FNA and Hook
wire needles). Some other service tools are also included in this kit.
- The Vertical Holder kit; components of the kit are shown in section 4 Parts contained in the
Vertical Holder Kit on page 31. This kit, supplied in a carrying case, provides parts required for
mounting interventional tools (needles, biopsy devices, etc.) to be used in vertical approach.
- The Lateral Holder kit; components of the kit are shown in section 5 Parts contained in the Lateral
Holder Kit on page 33. This kit, supplied in a carrying case, contains components required for
mounting interventional tools (needles, biopsy devices, etc.) to be used in lateral approach. It
includes the non-aperture paddle, and a phantom used for calibrating lateral approach tools and
verifying accuracy in vertical and lateral approach.
• Typical needle guides are shown in section 6 Needle guides on page 38. According to the biopsy
device in use, and the type of intervention, needle guides may be provided by GE or by the
manufacturer of the biopsy device.
Stereotaxy components
2 Stereotactic Positioner
Part name Code Comment Illustration
Stereotactic Positioner SP The Stereotactic Positioner is a
self-contained unit which is
mounted on the Senographe
image receptor, in place of the
Bucky. A compression paddle
fitted to the positioner is used,
independently of the main
compression system, to
immobilize the breast during
stereotactic procedures. The
paddle used for vertical approach
procedures has an aperture to
allow the passage of needles; a
paddle without aperture is used for
lateral approach procedures. The
unit includes a needle holder
which accepts interchangeable
needle guides for accurate
positioning of various types of
needle
Stereotaxy components
Stereotaxy components
Stereotaxy components
Stereotaxy components
Stereotaxy components
Stereotaxy components
Stereotaxy components
spring clip A
Needle guide holder for L14 Fits to the guide holder support
lateral approach: 6 mm L13 or L15 to hold a 6 mm needle
dia guide for use in lateral approach.
Needle guide holder for Reusable, to be disinfected before use. Supplied in packs of 5
lateral approach: 6 mm axis.
Stereotaxy components
Stereotaxy components
Stereotaxy components
6 Needle guides
Note:
Needle guides may also be referred to as bushings. The term needle guide is used throughout this
manual for consistency.
• Needle guides may be provided by GE or by biopsy device manufacturers, and exist in many sizes
and forms for different uses. The information provided here is indicative only.
Part name Code Comment Illustration
Plastic needle guide NGP To be inserted in
metal needle guide adaptors
(upper and lower) to position the
needle.
Plastic needle guides: Supplied sterilized, for one-time use, in packs of 10 (5 pairs).
Diameters (mm): 0.9, 1.0, 1.2, 1.4, 1.8, 2.25.
Plastic needle guide with HWG To be inserted in metal needle
slot, for use with hook guides adptators (upper and
wires. lower) to position the needle for a
hook wire localization exam.The
needle passes through the holes
in the needle guides, and the slot
in the front of the guides allow the
wire to be disengaged after Plastic needle guide
Metal needle guides: Reusable, to be sterilized before use. Supplied in packs of 4 (2 pairs).
Diameters (mm): 0.9, 1.0, 1.2, 1.4, 1.8, 2.25, 2.55*, 2.9*, 3.2* (* = without
retaining catch).
Metal needle guides for NG Fits in upper and lower SP needle
VAD. holder :
- Available for 8G,11G and 14G
Mammotome needles (non-
circular needles).
- Available for 9 G and 12 G
circular needles.
Metal needle guides. - Reusable, to be sterilized before use. Supplied in packs of 5.
Available for 8G, 11G And 14G Mammotome needles.
- Reusable, to be sterilized before use. Available for 9G and 12 G circular
needles.
1 Introduction
This chapter provides information on the Stereotactic Positioner and the Stereo Compression Paddle.
• The main positioner components are described in section 2 The Stereotactic Positioner on page 39.
• Installation and removal of the stereo compression paddle is described in section 3 The stereo
compression paddle on page 41.
Needle holder
Compression knob
Stereo compression
paddle Movement control knobs
(X, Y, Z)
Table support
Breast support
• Display panel. This panel displays positioning information. When not in use, the panel is stored in
its holder on top of the unit, as shown. In use, it may be placed on the side of the unit, held in
place by its magnetic back.
• Stereo compression paddle and compression knobs. The compression paddle is used to
immobilize the breast during the exposure and puncture procedures. An aperture in the paddle
allows the needle to penetrate the breast in vertical approach. A non-aperture paddle is provided
for use in lateral approach. Refer to section 3 The stereo compression paddle on page 41 for
information on paddle installation and removal.
Compression is applied manually, using one of the two compression knobs (one on each side of
the unit).
• Breast support. The breast support forms part of the Stereotactic Positioner assembly. The
assembly is mounted on the Image Detector in the same way as other breast supports (Bucky
and Magnification Stands). Four crosses are engraved on the breast support to indicate the limits
of the biopsy window.
• Movement control knobs. Three knobs are provided to allow manual positioning of the needle
holder on the three axes, X, Y, and Z. These knobs are duplicated on each side of the SP.
• Movement control buttons. Two sets of three buttons (one set on each side of the unit) are
provided to control motorized movement of the needle holder: Go to Park P , Go to Target
, and Next Target . A light next to each Go to Target button becomes green to
indicate that the target position has been reached, or flashes green to indicate that the needle
holder is in the Park position.
• Rotation/Angulation buttons. Two buttons (one on each side of the
unit) are provided to enable switching between rotation and
angulation modes when positioning the gantry for stereotactic
exposures. A green light next to each button is lit when the
angulation mode is in use. Refer to Chapter 10 Preparation for
examinations, section 2 Arm Positioning on page 94.
CAUTION
1 2 3 4
• Reverse the procedure to install the plastic paddle. On completion, remember to lock the paddle in
position by turning the two locking levers to the horizontal position.
WARNING
The rear of the Stereotactic Positioner assembly is very heavy. Take great care when installing or
removing the assembly. It is important to hold it using both metal rods, to ensure the best
possible grip. Especially, do not try to hold the assembly using the compression knobs, the
carbon cover, or any attached parts (e.g., the lateral approach tool if it is mounted).
If the Stereotactic Positioner assembly falls on the detector, or is allowed to slide at an angle on
the detector, it can cause irreparable damage.
When not installed, the assembly should be placed on its special table stand. If it is placed on a
flat surface without support, its own weight may cause it to tip over and cause damage.
Note:
Installation of the Stereotactic Positioner is easiest when the Arm angle is 0°, and the operator is
directly facing the Image Detector. Adjust the Arm elevation as required for comfort and ease of
installation.
• To install the Stereotactic Positioner on the Senographe Arm:
1. Remove Senographe accessories:
- Face shield
- Compression paddle
- Bucky or magnification platform
2. Move the compression paddle holder on the Senographe arm to the
maximum height, using the footpedal and manual controls.
Note:
If the compression paddle holder on the gantry is not in its upper
position, angulation will not be allowed.
3. Hold the Stereotactic Positioner and breast support assembly by the
metal rods as shown. Carefully align the assembly with the rails, and
slide it onto the rails as shown. Press on the front of the breast
support until motor noise is heard, indicating the start of the locking
phase. You should hear a click (as for Bucky installation). The display
panel then displays Communcation not established, followed by
Ready.
After Stereotactic Positioner locking has been completed, the red line on the rear cover must not be
visible. If it is visible, the breast support has not been inserted correctly. Remove it and insert it
correctly, sliding it along the rails until motorized locking begins.
4. Attach the Display Unit to the positioner housing so that it is easily visible, held in position by its
magnetic base.
5. Use the manual control knobs to move the stereo compression paddle to its highest position.
6. Press and hold the Go to Park button P to move the needle holder out of the field of view. The
display panel should now display position coordinates for the upper needle holder support.
! Notice:
The knobs at the sides of the Senographe compression paddle holder (used for manual
compression in standard mammography) must not be operated once the Stereotactic Positioner is
in place.
• The Senographe can now be positioned for the projection required as described in Chapter 10
Preparation for examinations, section 2 Arm Positioning on page 94.
WARNING
The rear of the Stereotactic Positioner assembly is very heavy. Take great care when installing or
removing the assembly. It is important to hold it using both metal rods to ensure the best possible
grip, as described above (section 1). Especially, do not try to hold the assembly using the
compression knobs, the carbon cover, or any attached parts (e.g., the lateral approach tool if it is
mounted).
Take great care, and be prepared to take the weight when removing the assembly. If it falls on
the detector, or is allowed to slide at an angle on the detector, it can cause irreparable damage.
After removal, the assembly should be placed on its special table stand. If it is placed on a flat
surface without support, its own weight may cause it to tip over and cause damage.
To remove the Stereotactic Positioner assembly, press the two breast
support unlocking buttons (located underneath the Image Detector)
simultaneously. It is only necessary to press them briefly to start the
unlocking phase. Do not pull the assembly; unlocking takes place
automatically. Motor noise is heard during the entire unlocking phase.
When the unlocking movement stops (after about three seconds),
remove the assembly by manually sliding it along the rails.
Breast support
unlocking buttons
1 Introduction
This chapter provides a step-by-step guide to stereotaxy examination procedures.
Section 3 Pre-examination check on page 45 refers to procedures required to make needle path tests
and accuracy tests for vertical and lateral approaches using a stereotaxy phantom.
Section 5 Patient examination on page 46 describes the normal procedures required for use with a
patient.
2 Quick reference
Section 3 Pre-examination check on page 45
Section 4 Cleaning, disinfection, and sterilization on page 45
Section 5 Patient examination on page 46
5-1 Preparation on page 46
5-2 Stereotaxy exposures on page 47
5-3 Define puncture on page 48
5-4 Send the puncture set and move the Stereotatic Positioner to the target position on page 49
5-5 Puncture on page 50
5-6 Secondary targets (if required) on page 50
5-7 Completion on page 51
3 Pre-examination check
• The following tests are described in Chapter 16 Planned Maintenance section 3 QC Tests for the
Radiologic Technologist or Medical Physicist on page 161.
- QC Test 1 - Needle Path Block Test on page 163.
- QC Test 2 - Localization Accuracy - Vertical Approach on page 165.
- QC Test 5 - Localization Accuracy - Lateral Approach on page 176.
For information on the frequency of the above listed tests, refer to the description of each test.
If no stereotactic exams are carried out, the tests should be carried out at least once a month to
ensure continued accuracy:
• Record the test results as instructed in Chapter 16 Planned Maintenance section 4 Test Result
Record sheets on page 185.
• Refer to local rules and regulations for the required frequency of geometrical accuracy tests using the
phantoms.
5 Patient examination
5-1 Preparation
1. Switch on the Senographe and install the Stereotactic Positioner as described in Chapter 6
Installation of the Stereotactic Positioner, section 1 Installing the Stereotactic Positioner on page 43.
After pushing the Go to Park button P as instructed in Chapter 6 Installation of the Stereotactic
Positioner, the display panel should display position coordinates for the upper needle holder support.
2. For biopsy devices in vertical approach, refer to Chapter 8 Installation for Vertical Approach:
- If required, install the vertical adaptor or other parts as described.
3. For biopsy devices in lateral approach, refer to Chapter 9 Installation for Lateral Approach:
- If required, install the lateral adaptor or other parts as described.
- Depending on the position of the targeted lesion, the puncture should be performed from the side
(right or left) that allows the shortest route. Check the Accessibility:
- Any lesion located at the Right part of the Scout (0°) image should be accessed from the Right
Any lesion located at the Left part of the Scout (0°) image should be accessed from the Left.
- Any lesion located within 2 cm of the center of the Scout (0°) image can be accessed from
either Left or Right, if the distance between the lesion and the entrance point in the breast is
less than the useful needle length.
4. Prepare the biopsy device to be used, with its mounting bracket if required.
5. Ensure that the needle or biopsy device to be used is specified in the Needle Panel Refer to
Chapter 11 Needle Setup for more information.
6. Open the Worklist window and verify that the selection for the type of exam is Stereo.
7. Select the patient’s name/ID in the Patients list on the Worklist by clicking on it.
If the Patient has not yet been defined, select New Patient... in the Worklist, and complete the new
Medical Procedure Card as required, before starting image acquisition. Refer to Chapter 12
Stereotaxy Image Acquisition section 1 Starting the Stereotaxy application on page 125.
8. Click the Start Exam button. The Stereotactic Viewer appears, with the Viewer area at the right of the
screen remaining black until an exposure is taken.
9. Move the compression paddle on the Stereotatic Positioner to its highest position.
10. Press the rotation buttons to set the gantry at the selected incidence.
11. Press and hold the Go to Park button P to move the needle holder out of the field of view.
Image acquisition is allowed only when the display shows parking.
12. Position the patient as described in Chapter 10 Preparation for examinations section 3-3 Positioning
the Patient on page 99. Ensure that the area of interest is in the field of view and compress the breast
to immobilize it. Note that the system does not recognize the presence of a breast unless the
compressed thickness is less than10 cm.
13. Press the Rotation/Angulation button on the Stereotactic Positioner to switch from rotation mode to
angulation mode. The gantry readout now displays the stereo angle, not the gantry angle. Refer to
Chapter 10 Preparation for examinations, section 2 Arm Positioning on page 94.
1. Click (left mouse button) on the Stereotaxy selection tab to open the Stereotaxy Panel
2. Click the New Primary button to define a new puncture set, then select the puncture marker and
drag it over the exact area of interest on each of the zoomed images. Use the Show/Hide Epipolar
buttons to view the epipolar line if necessary. Refer to Chapter 13 Stereotaxy Punctures section 1-
4 Placing the Primary Puncture Marker on page 136.
Note:
You must accurately select the same point seen on the two images. Inaccuracies may lead to
deviations when the needle is being inserted in the breast.
CAUTION
Before starting a puncture, make sure that the Z value of the target is greater than the specified
minimum accessible value for the biopsy device in use. If not, the breast must be repositioned for
access to the lesion.
3. Click the Check on scout button to display the scout image in the top half of the screen. This allows
you to check that the correct puncture site has been selected.
See Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on page 127 for more
information on this feature.
4. Click the Back to Pair button to return to Stereo Mode.
See Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on page 127 for more
information on this feature.
5. If you wish to define Secondary puncture sites, click the Secondary button. See
Chapter 13 Stereotaxy Punctures section 1-5 Placing the Secondary Puncture 5.5
Markers on page 138 for more information on this feature.
To change the distance of the secondary markers from the primary site, click on the Secondary
up/down arrows to adjust the value displayed in the offset window (the value can be
changed by increments of 0.5 mm to the maximum of 20.0 mm).
5-4 Send the puncture set and move the Stereotatic Positioner to the target position
1. Click the Needle & Puncture button to open the Needle & Puncture panel. Note that this panel opens
only if the puncture set is within the puncture volume. Puncture volume is the volume in which the
Stereotactic Positioner can physically reach the defined puncture target position.(see Chapter 13
Stereotaxy Punctures for more information).
2. Select a pre-programmed needle or biopsy device, or define the needle if it does not already exist.
3. Click the Send button.
If this button is inactive, the cause is given in the status field below the button. Device-target
incompatibility may be caused by
- an unsuitable needle length,
- a risk of collision with the breast support,
- a puncture target outside the puncture volume, or
- because a correction to the Z target is needed to reach the lesion.
If a correction is required, click the Correction button to acknowledge and accept any required
changes (the correction is made automatically; if you make the correction manually you risk
missing the target). Now check at the bottom of the panel to see if there are any active warnings
present. Click Send if you still want to perform the puncture. Refer to Chapter 13 Stereotaxy
Punctures, section 2-2 Correction on page 145 for more information.
4. Angle the tube arm to either of the parking positions as required, (or +/-33°) to ease insertion of the
needle.
5. Press and hold the Go To Target button (or manually adjust the X, Y and Z knobs) on the
Stereotactic Positioner until the target position is reached; the light will now be green; the display
indicates TARGET REACHED and 0-0-0.
See Chapter 13 Stereotaxy Punctures section 4 Performing the puncture on page 151 for more
information.
Note:
Motorized movement is allowed only when the needle path is closed.
6. When the target has been reached, press the Go To Target button again to display the
programmed needle. Check that the correct needle or biopsy device has been chosen.
7. Install the required needle guides.
For the Lateral Arm, see Chapter 9 Installation for Lateral Approach, section 4-2-2 Install needle
guide holders and needle guides on page 81.
5-5 Puncture
1. If using a biopsy device, prepare the device and the adaptator (if needed).
- For a biopsy device in vertical approach, see Chapter 8 Installation for Vertical Approach.
- For a biopsy device in lateral approach, see Chapter 9 Installation for Lateral Approach.
2. Position the patient.
3. Perform the stereo triplet.
4. Mark the target, select the needle/device that will be used.
5. Clean/disinfect the breast.
6. Move the needle holder to target position.
7. Insert a fine needle through the sterile needle guides to check correct location to perform anesthesia
and inject anesthesic.
8. When the anesthesic is injected, remove the needle in use.
9. Move the needle holder back to make an incision, if necessary.
10. Move the needle holder back to target position.
11. Place the biopsy device in the cocked position and insert the needle through the sterile needle
guides.
Note:
Performing an incision and injecting anesthesic depends on the device in use so, the application
of the previous procedures is the decision of the performing physician.
12. Take pre-fire images (if CB or VAD). Take needle tip check images (if hook wire or FNA).
CAUTION
Pre-fire stereo images allow the needle position to be checked in the X and Y axes.
To check the Z position of the needle, it is recommended that you define a new target at the tip of
the needle, and compare the Z value with the Z value of the lesion.
13. Confirm needle placement. You can compare the pre-fire and the stereo pair by using the Pairs
Comparison feature (see Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on
page 127).
14. Proceed to the sample collection as described in the Biopsy Device User Manual (if CB or VAD); get
the cellls sample (suction or capillarity if FNA).
15. A stereo pair can be taken after the sample collection, to check if more samples are needed, in case
of CB or VAD. A stereo pair could be performed to check the wire tip position in case of hook wire
exam.
16. Proceed with secondary targets if defined (Chapter 5-6 Secondary targets (if required)).
17. Remove the needle / device when all the necessary samples are obtained.
18. Marker clip insertion : for core biopsy, it might be useful to perform a marker clip insertion to mark the
targeted area for additional procedures. Follow the instructions provided by the clip manufacturer for
correct insertion and placement.
2. Press the Go To Target button again to move the needle holder over the selected target (only
if the needle path was closed between sampling; otherwise move the holder manually).
3. Proceed with the remaining secondary targets.
5-7 Completion
1. Before releasing compression:
You may want to make a final image at 0° to check that affected tissue has been removed. To do this,
close the needle path and drive the needle holder to parking position.
2. Specimen imaging:
If another mammography system is available, leave the patient compressed and use the other
system to image the specimen. If no other system is available, release the patient from compression
before imaging specimens.
3. Slowly relase patient compression and proceed with appropriate patient after-care.
4. Print or archive the image(s) for permanent documentation.
Refer to Chapter 14 Image Printing and Review.
5. Specimen imaging after target completion:
- If the samples image (s) are performed on the Stereotatic Positioner, the compression paddle
holder must be moved to a height smaller than 10 cm to allow exposures.
- Samples image(s) can also be taken using the magnification stand.
6. Set the arm angulation to 0.0° (scout position), removal of the Stereotactic Positioner is allowed if
the compression paddle holder is moved to a height over 10 cm.
Chapter 8
Operator manual
V3 V2 V4
Needle guides in Needle guide in Needle guide in Adaptor and Needle guides in
SP needle holder SP needle holder SP needle holder needle guide in SP needle holder
Page no. 54
upper and lower lower support lower support SP needle holder upper and lower
supports lower support supports
SenoRx EnCor
Hand held biopsy devices holder
Puncture to be Suros Atec BARD Vacora Mammotome ST Mammotome SenoRx EnCor BARD Magnum
made manually biopsy device biopsy device biopsy device Biopsys biopsy device biopsy device
No biopsy device biopsy device
Revision 3
V6
V5
V7
V2 V3
Refer to Chapter 4 Stereotaxy components (section 4 Parts contained in the Vertical Holder Kit on
page 31) for an overview and illustrations of parts contained in the kit.
Note:
It is strongly recommended that all kit components are returned to the carrying case after use.
Parts contained in the kit are listed in the following table:
Part Reference Quantity
Biopsy device support V1 1
Mammotome Biopsys holder V2 1
Mammotome ST holder V3 1
BARD Magnum guide V4 1
Securing rods V5 2
Securing plate V6 1
Knurled securing screw V7 1
Note:
Different biopsy devices require different parts from the kit. Refer to the installation instructions in
this chapter.
Securing rods V5
Securing plate V6
WARNING
The use of Plexiglas or other material to build up the breast for a vertical approach procedure
may cause:
- Image quality degradation, so the lesion may therefore be missed.
- Risk of needle damage or breast perforation because the system is not aware of the thickness
of the breast as compared to the thickness of the supporting material.
It is the user’s responsability to ensure that no injury to the patient or damage to the equipment
results from the procedure.
Follow the instructions supplied for the biopsy device to be used:
• Mammotome ST
Refer to section 3-1 Fitting the Mammotome ST or Mammotome Biopsys biopsy device (vertical
approach) on page 59.
• Mammotome Biopsys
Refer to section 3-1 Fitting the Mammotome ST or Mammotome Biopsys biopsy device (vertical
approach) on page 59.
• SenoRx EnCor
Refer to section 3-2 Fitting the SenoRx EnCor biopsy device (vertical approach) on page 61.
• BARD Magnum
Refer to section 3-3 Fitting the BARD Magnum biopsy device (vertical approach) on page 62.
• BARD Vacora
Refer to section 3-4 Using the BARD Vacora biopsy device (vertical approach) on page 63.
• Suros ATEC
Refer to section 3-5 Using the Suros ATEC biopsy device (vertical approach) on page 64.
Mammotome ST or Biopsys
Mammotome Biopsys V2
holder
Mammotome ST holder V3
CAUTION
SenoRx EnCor
3-2-2 Fit the SenoRx EnCor device to the biopsy device support
Note:
A sterile disposable needle guide is provided
with the SenoRx EnCor device.
1. Fit the SenoRx EnCor biopsy device to the
adaptor supplied. Follow the instructions
provided with the adaptor.
CAUTION
BARD Magnum
CAUTION
BARD Vacora
CAUTION
Suros ATEC
CAUTION
WARNING
The Use of Plexiglas or other material to build up the breast for a lateral approach procedure may
cause:
- Image quality degradation so the lesion may therefore be missed.
- Risk of needle damage or unnecessary skin perforation, because the system is not aware of the
presence of the supporting material.
It is the responsibility of the user to ensure that no injury to the patient or damage to the
equipment results from the procedure.
Chapter 9
Install the lateral approach tool support
Operator manual
See section 3 Installing the lateral needle Needle guide holder L14, L16, or L18, with
guide holder on page 71. support L13 or L15 as required
Fit the biopsy device, with holder if Biopsy device holder BARD Magnum holder
Overview of installation for lateral approach
Page no. 66
SenRx EnCor BARD Vacora rail
holder holder and
adaptor
Needle guides Needle guides Needle guides Needle guides Needle guides Needle guides Needle guides
Suros Atec Mammotome ST Mammotome SenoRx EnCor BARD Magnum BARD Vacora Puncture to be
biopsy device biopsy device Biopsys biopsy device biopsy device biopsy device made manually
biopsy device No biopsy device
Revision 3
L3
L3 L11 L6 L19 L13
L15
L8
L1
L9
L7
L1
L17
L13
Note:
It is strongly recommended that all kit components are returned to the carrying case after use
Parts contained in the kit are listed in the following table:.
Part Ref. Qnty Part Ref. Qnty
Lateral approach tool support L1 1 Biopsy device support L11 1
Securing rod L2 1 Biopsy device support screw L12 1
Securing plate L3 1 Guide holder L13 1
Lateral adaptor bar L4 1 Needle guide 6 mm dia. L14 5
(delivered in a separate envelope)
Positioning cursor L5 1 Guide holder L15 1
Non-aperture paddle L6 1 Needle guide 8 mm dia. L16 5
(delivered in a separate envelope)
Rigidity bar supports L7 2 BARD Magnum holder L17 1
(not used for Essential)
Key (hex wrench) L8 1 Guide Holder (BARD Magnum) L18 1
Rigidity bar support screws L9 4 Pin Phantom L19 1
(in plastic bag)
Rigidity bar L10 1 Calibration needle L20 1
Note:
Different biopsy devices require different parts from the kit. Refer to the installation instructions in
this chapter.
The lateral approach tool support and adaptor bar are provided to position and guide biopsy
devices used in lateral approach. Biopsy devices must be hand-held when in use.
• Install the lateral approach tool support and adaptor bar as described below.
Parts required from the lateral holder kit are shown here:
Lateral approach L1 Securing rod L2
tool
support
Non-aperture L6
compression
paddle.
1. If present, remove the compression paddle with aperture used for vertical approach. Refer to
Chapter 5 The Stereotactic Positioner.
2. Make sure that the needle path is closed, then press the Go to Park button to move the needle
holder out of the field of view.
3. Clip the lateral approach tool support L1 over the
Stereotactic Positioner head.
CAUTION
With the bar on the side to be used, make sure that the “OK” engraved mark is facing the
patient, otherwise the targeted lesion won’t be reached even if the needle holder is in the target
position.
Correct Not correct
Metal guide
for securing
rod
WARNING
L13 must never be used with the lateral arm mounted on the right side of the breast support.
The protruding second ball may impact the patient’s chest wall.
4. Slide the selected lateral needle guide holder or support (L13, L15, or L18) onto the lateral adaptor
bar L4.The engraved arrow must point towards the Stereotactic Positioner (note that when the
approach is from the left, the arrow is found at the rear of the holder).
5. Rotate the adaptor bar towards the rear, to be sure that the holder and the bar are far enough from
the 0 position to prevent them from appearing on images.
6. Tighten the screw of the rigidity bar L10 to lock the assembly in position; be sure that the assembly
can NOT rotate freely.
Ensure that the guide or holder and the adaptor bar are outside the field of view.
Suros ATEC
WARNING
When the Suros ATEC is used with a needle length of 90 mm or 120 mm, the target area must be
at least 23 mm above the breast support. For target areas below this height, use the140 mm
needle, to allow the biopsy device to be positioned to the side of the breast support.
• First install the lateral approach adaptor and needle guide holder as described in sections 2 Installing
the lateral approach tool support on page 68 and 3 Installing the lateral needle guide holder on
page 71.
• The guide holder support used must be L15.
• Install the biopsy device as described in the instructions which follow.
Additional parts required from the lateral holder kit are shown here:
4-1-1 Fit the Suros ATEC device to the biopsy device support
Fit the Suros ATEC biopsy device to the adaptor supplied. Follow the instructions provided with the
adaptor.
1. Remove the lateral lock spring clip A (see L11
in the table in Chapter 4 Stereotaxy
components, Section 5 Parts contained in the
Lateral Holder Kit on page 33); it is not required
for use with the Suros ATEC device.
4-1-2 Installation of guides and needle guides for the Suros ATEC in lateral approach
1. Install one of the five metal needle guide
holders L16, by sliding the holder into the guide
holder support L15.
2. Secure the holder in position by tightening the
knurled screw on the support.
WARNING
The metal needle guide holders L14 and L16 are supplied non-sterile. They MUST be
disinfected before use. Refer to Chapter 15 System Hygiene,.
3. Install the needle guide:
Select the appropriate size of sterile needle
guide supplied by GE, and introduce it into the
needle guide holder L16.
WARNING
Biopsy devices should never be used without either disposable needle guides supplied by the
biopsy device manufacturer or sterilizable needle guides supplied by GE.
Always make sure that the size of the needle guide corresponds to the diameter of the needle to
be used.
WARNING
When a biopsy device is installed on the lateral approach adaptor bar, the use of the rigidity bar
L10 is mandatory. To avoid unexpected rotation of the bar while the needle is in the patient
breast, the securing screw of the rigidity bar L10 must not be released.
CAUTION
The lateral approach tool support and adaptor bar are provided to position and guide biopsy
devices used in lateral approach. They must not be considered as a firm support: biopsy devices
should be supervised and hand-held when in use.
The target position should be reached before installing the biopsy device on the lateral arm.
! Notice:
To avoid damage to the equipment, the Z positioning knob must NOT be turned after a biopsy
device has been installed on the lateral approach adaptor bar. If it is necessary to move the Z axis,
the biopsy device must first be removed, or it must be held so as to prevent its weight from being
applied to the adaptor assembly during movement.
3. Release the securing screw of the biopsy
device support L11. Fit the support, together
with the attached Suros ATEC assembly, on to
the lateral adaptor bar L4, so that the needle of
the biopsy device passes through the needle
guide. Note that the biopsy device must be held
perfectly parallel to the bar for ease of
mounting.
Mammotome ST or Biopsys
WARNING
The Mammotome ST and Biopsys biopsy devices cannot be used for biopsies on patients with
target areas less than 13 mm above the breast support.
• First install the lateral approach adaptor and needle guide holder as described in sections 2 Installing
the lateral approach tool support on page 68 and 3 Installing the lateral needle guide holder on
page 71.
• The guide holder support used must be L15.
• Install the biopsy device as described in the instructions which follow.
Parts which may be required from the lateral holder kit are shown here:
WARNING
The metal needle guide holders L14 and L16 are supplied non-sterile. They MUST be
disinfected before use. Refer to Chapter 15 System Hygiene.
3. Install the needle guide:
Introduce a disposable needle guide supplied
by the manufacturer into the needle guide
holder L14, or a GE sterilizable metal needle
guide into the needle guide holder L16. The
size of needle guide must be chosen according
to the needle to be used.
WARNING
Biopsy devices should never be used without either disposable needle guides supplied by the
manufacturer or sterilizable needle guides supplied by GE.
Always make sure that the size of the needle guide corresponds to the diameter of the needle to
be used.
WARNING
When a biopsy device is installed on the lateral approach adaptor bar, the use of the rigidity bar
L10 is mandatory. To avoid unexpected rotation of the bar while the needle is in the patient
breast, the securing screw of the rigidity bar L10 must not be released.
CAUTION
The lateral approach tool support and adaptor bar are provided to position and guide biopsy
devices used in lateral approach. They must not be considered as a firm support: biopsy devices
should be supervised and hand-held when in use.
The target position should be reached before installing the biopsy device on the lateral arm.
! Notice:
To avoid damage to the equipment, the Z positioning knob should not be turned after a biopsy
device has been installed on the lateral approach adaptor bar. If it is necessary to move the Z axis,
the biopsy device must be removed, or it must be held so as to prevent its weight from being
applied to the adaptator assembly during movement.
Note:
The position used should be verified after installation, using the pin phantom and the biopsy device
in use. Refer to Chapter 16 Planned Maintenance, section QC Test 6 - Accuracy check before
using new CB or VAD needles in lateral approach on page 180. The required position of cursor L5
is obtained if the tip of the pin phantom is in the middle of the notch (center of B) at the end of the
procedure.
SenoRx EnCor
WARNING
The SenoRx EnCor biopsy device cannot be used for biopsies on patients with target areas less
than 15 mm above the breast support.
• First install the lateral approach adaptor and needle guide holder as described in sections 2 Installing
the lateral approach tool support on page 68 and 3 Installing the lateral needle guide holder on
page 71.
• The guide holder support used must be L15.
• Install the biopsy device as described in the instructions which follow.
Parts which may be required from the lateral holder kit are shown here:
4-3-1 Fit the SenoRx EnCor device to the biopsy device support
1. Fit the SenoRx EnCor biopsy device to the
EnCor adaptor supplied. Follow the instructions
provided with the adaptor.
WARNING
The metal needle guide holders L14 are supplied non-sterile. They MUST be disinfected before
use. Refer to Chapter 15 System Hygiene.
3. Install the needle guide:
Select the appropriate size of sterile needle
guide supplied by the manufacturer, and
introduce it into the needle guide holder L14.
WARNING
Biopsy devices should never be used without either disposable needle guides supplied by the
manufacturer or sterilizable needle guides supplied by GE.
Always make sure that the size of the needle guide corresponds to the diameter of the needle to
be used.
WARNING
When a biopsy device is installed on the lateral approach adaptor bar, the use of the rigidity bar
L10 is mandatory. To avoid unexpected rotation of the bar while the needle is in the patient
breast, the securing screw of the rigidity bar L10 must NOT be released.
CAUTION
The lateral approach tool support and adaptor bar are provided to position and guide biopsy
devices used in lateral approach. They must not be considered as a firm support: biopsy devices
should be supervised and hand-held when in use.
The target position should be reached before installing the biopsy device on the lateral arm.
! Notice:
To avoid damage to the equipment, the Z positioning knob must NOT be turned after a biopsy
device has been installed on the lateral approach adaptor bar. If it is necessary to move the Z axis,
the biopsy device must first be removed, or it must be hand-held so as to prevent its weight from
being applied to the adaptor assembly during movement.
3. Release the securing screw of the SenoRx
EnCor assembly. Fit the assembly on to the
lateral adaptor bar L4, so that the needle of the
biopsy device passes through the needle
guide. Note that the device must be held
perfectly parallel to the bar for ease of
mounting.
4. The biopsy device support L11 must be
positioned on the adaptor bar L4 so that the
position cursor L5 contacts the holder L15.
Determine the L5 ruler stop position according
to the needle parts used: the fired needle
length, the sample notch and the throw. When
the correct position has been attained, tighten
the securing screw of the holder L11 securely;
ensure that the holder is now unable to slide on
the bar.
Note:
The position used should be verified after installation, using the pin phantom and the biopsy device
in use. Refer to Chapter 16 Planned Maintenance, section QC Test 6 - Accuracy check before
using new CB or VAD needles in lateral approach on page 180. The required position of cursor L5
is obtained if the tip of the pin phantom is in the middle of the notch (center of B) at the end of the
procedure.
BARD Magnum
WARNING
The BARD Magnum biopsy device cannot be used for biopsies on patients with target areas less
than 8 mm above the breast support.
• First install the lateral approach adaptor and needle guide as described in sections 2 Installing the
lateral approach tool support on page 68 and 3 Installing the lateral needle guide holder on page 71.
• The needle guide holder used must be L18.
• Install the biopsy device as described below.
Additional parts required from the lateral holder kit are shown here:
WARNING
Metal needle guides are supplied non-sterile. They MUST be sterilized before use. Refer to
Chapter 15 System Hygiene.
WARNING
Biopsy devices should never be used without either disposable needle guides or sterilizable
needle guides.
Always make sure that the size of the needle guide corresponds to the diameter of the needle to
be used.
WARNING
When a biopsy device is installed on the lateral approach adaptor bar, the use of the rigidity bar
L10 is mandatory. To avoid unexpected rotation of the bar while the needle is in the patient
breast, the securing screw of the rigidity bar L10 must not be released.
CAUTION
The lateral approach tool support and adaptor bar are provided to position and guide biopsy
devices used in lateral approach. They must not be considered as a firm support: biopsy devices
should be supervised and hand-held when in use.
The target position should be reached before installing the biopsy device on the lateral arm.
! Notice:
To avoid damage to equipment, the Z positioning knob must NOT be turned after a biopsy device
has been installed on the lateral approach Biopsy adaptor. If it is necessary to move the Z axis, the
biopsy device must first be removed, or it must be hand-held so as to prevent its weight from being
applied to the adaptor assembly during movement.
3. Position the cursor L5 on adaptor bar L4, using the scale engraved on the bar.
Determine the L5 ruler stop position according to the needle parts used: the fired needle length, the
sample notch, the throw and coax.
Note:
The position used should be verified after installation, using the pin phantom and the biopsy device
in use. Refer to Chapter 16 Planned Maintenance, section QC Test 6 - Accuracy check before
using new CB or VAD needles in lateral approach on page 180. The required position of cursor L5
is obtained if the tip of the pin phantom is in the middle of the notch (center of B) at the end of the
procedure.
4. Slide the needle guide holder L18 onto the bar
L4, until it contacts cursor L5.
5. Slide the BARD Magnum holder L17 over the
bar L4, and secure it in position, using the
upper screw of the holder.
BARD Vacora
WARNING
When the BARD Vacora is used with a needle length of 118 mm, the target area must be at least
15 mm above the breast support. When it is used with a needle length of 140 mm, the target area
must be at least 13 mm above the breast support. Lower targets cannot be accessed using the
BARD Vacora device in lateral approach.
• Install the lateral approach adaptor and needle guide holder as described in sections 2 Installing the
lateral approach tool support on page 68 and 3 Installing the lateral needle guide holder on page 71.
• The guide holder support used must be L15 when entering the breast from the right side of the
needle positioner.
Note:
A second ball is provided on the holder L13, to allow the bar to be positioned at a greater angle
when using the BARD Vacora with the lateral arm mounted on the left side of the breast support
(entering the breast from the left side of the needle positioner).
CAUTION
L13 must never be used with the lateral arm mounted on the right side of the breast support. The
protruding second ball may impact the patient’s chest wall.
• Install the biopsy device as described below.
4-5-1 Install the needle guide holder and needle guide
1. Fit the needle guide holder supplied with the
Vacora biopsy device onto the guide holder
support L15 or L13. Secure it in place by
tightening the lower screw of L15 or L13.
2. Introduce the sterile needle guide (4 mm for
metallic coaxial cannula, 4.5 mm for plastic
coaxial cannula) through the opening on the
needle guide holder.
WARNING
Biopsy devices should never be used without either disposable needle guides supplied by the
biopsy device manufacturer or sterilizable needle guides supplied by GE.
Always make sure that the size of the needle guide corresponds to the diameter of the needle to
be used.
WARNING
When a biopsy device is installed on the lateral approach adaptor bar, the use of the rigidity bar
L10 is mandatory. To avoid unexpected rotation of the bar while the needle is in the patient
breast, the securing screw of the rigidity bar L10 must NOT be released.
CAUTION
The lateral approach tool support and adaptor bar are provided to position and guide biopsy
devices used in lateral approach. They must not be considered as a firm support: biopsy devices
should be supervised and hand-held when in use.
The target position should be reached before installing the biopsy device on the lateral arm.
! Notice:
To avoid damage to the equipment, the Z positioning knob must NOT be turned after a biopsy
device has been installed on the lateral approach adaptor bar. If it is necessary to move the Z axis,
the biopsy device must first be removed, or it must be supported so as to prevent its weight from
being applied to the adaptor assembly during movement.
3. The cursor L5 must be positioned on the adaptor bar L4, using the scale engraved on the bar.
Determine the L5 ruler stop position according to the needle parts used: the fired needle length, the
sample notch, the throw and coax (if a spacer is used, this should be considered also).
Note:
The position used should be verified after installation, using the pin phantom and the biopsy device
in use. Refer to Chapter 16 Planned Maintenance, section QC Test 6 - Accuracy check before
using new CB or VAD needles in lateral approach on page 180. The required position of cursor L5
is obtained if the tip of the pin phantom is in the middle of the notch (center of B) at the end of the
procedure.
4. Slide the cursor L5 to the chosen position on
the adaptor bar L4. Lock it in position by
tightening its screw.
5. Move the needle guide holder L15 or L13 up to
the cursor L5. Lock it in position by tightening
its screw.
1 Preparation
1-1 Preparing the Senographe
The Senographe is switched on and used in the usual way as described in the Senographe Essential
Acquisition System Operator manual.
• To view or change the current setting press the Set-up button , then select in succession
MEDICAL, STEREO and LIGHT.
• The factory default setting for light duration in stereotaxy is 5 minutes. It can be changed in steps of
1 minute between the minimum of 1 minute and the maximum of 10 minutes.
2 Arm Positioning
2-1 Rotation and Angulation Modes
• The term rotation is used to describe the rotation of the
complete Arm assembly, with the tube head arm
remaining at its current angle with respect to the image
receptor. The illustration shows rotation of the arm from
CC to MLO. Rotation
1 4 5 5 4
1
2 2
3 3
Arm rotation
control buttons
• When the Stereotactic Positioner is installed:
- Buttons shown with reference 4 in the illustration are used to control rotation or angulation,
according to the current mode of operation. Note that in angulation mode only one speed of
movement is available; additional pressure does not change the speed of movement.
- Buttons shown with reference 5 are used for up/down (lift) movement.
- Buttons with references 1, 2, and 3 (used for preset movements in routine mammography) are not
operational.
CAUTION
When the patient is in position, take great care when angulating to avoid any possibility of hitting
the patient with the arm assembly.
Note:
While compression is applied to the breast, all rotation and lift movements are inhibited. Only
angulation movement is possible.
2-4-1 Angulation movement for stereotaxy
• When in angulation mode, only the buttons shown with reference 4 (in the illustration in section 2-
2 Movement control buttons on page 95) are operational. They are used to move the tube head arm
between the five permitted positions as shown in the diagram below.
1. Press the top or bottom of the button to cause angular movement of the tube head arm in the
direction indicated by the arrow on the button.
2. Continue to press the button until angulation stops automatically, at the next permitted position.
For example, if the angulation is now 0°, press and hold the top of button 4 on the left hand side of
the arm, to move the tube head arm clockwise, until it stops automatically at the -15° position.
Press and hold the same button again to move the arm to the -Park position.
3. If the button is released during movement, the movement stops immediately.
4. The +Park and -Park positions are preset at ±33°, but the user can choose to stop the gantry at
any position.
Note:
With the Stereotactic Positioner installed, exposure is only
possible with the tube arm at a nominal angulation of +15°,
-15°, or 0°, as shown on the gantry readout (note that the
numeric display may show a slightly different value, for
example -15.3° instead of -15.0°, as shown here).
In all other cases, exposure is inhibited and a warning
message appears on the Senographe X-ray Console.
- Within five seconds, press the 0° button. The interlock is temporarily removed.
WARNING
- If metal needle guides (reusable) are in use, ensure that they have been sterilized before use.
- The internal diameter of needle guides must always be one size larger than the external
diameter of the needle in use; e.g., a needle of 0.8 mm should be used with needle guides of
0.9 mm.
CAUTION
The patient must be comfortably seated, or otherwise well supported (e.g., on a stretcher or a
suitable table), to ensure that she remains motionless, and that there is no danger of falling if she
is unwell during the examination. She should be supervised at all times.
CAUTION
While the images are being processed on the workstation, it is recommended that an operator
should remain with the patient to ensure that she does not move. If the patient moves between
the +15° and - 15° exposures, the scout and stereo images should be repeated (see Chapter 13
Stereotaxy Punctures section 1 Defining punctures on page 135).
3-3-3 The seated patient
Use a suitable examination chair, with adjustable height and locking wheels.
To adjust the chair (functions of the chair vary according to the model):
• Set chair height and depth of back of chair so that the patient is sitting comfortably with her back
straight and firmly supported.
• Set the height of the foot support so that the patient’s feet are resting comfortably.
• Set the height of the arm rests to support the patient’s arms comfortably.
Note:
1. The patient should be seated in a comfortable position, in which she can remain motionless for
the duration of the exam. This condition is essential for the accuracy of the puncture.
2. Be sure that the patient’s arms or hands do not rest on any of the Stereotactic Positioner
controls, such as the X, Y, Z knobs or the compression knobs.
3-3-4 Recumbent Patient Position
The patient should be positioned in a lateral decubitus position. She should be secured on the stretcher
and have good back support. Bending the knees with a pillow between them often makes the patient
more comfortable. The arms should be positioned so that the patient is not lying on them or restricting
the blood flow by being bent abnormally. The patient’s hand(s) may be directed to hold onto a handle on
the system if convenient. The breast can then be positioned into a LM/ML position or a CC or FB
position, making sure to check tube rotation/angulation prior to the start of the procedure. If the selected
projection for the exam is CC, FB or LM, the patient lies on the unaffected breast side, however if the
selected projection is ML, the patient lies on the affected breast side and the opposite breast side needs
to be positioned so that it is not included in the FOV. Refer to the Senographe Essential Acquisition
System Operator manual for information on selection of view names for recumbent patients.
3-3-5 Positioning the Breast
• Position the breast under the compression paddle so that the area to be examined is in the center of
the aperture in the paddle, or in the area defined by the four engraved crosses on the breast support
plate if using a non-aperture paddle.
CAUTION
For maximum accuracy, it is essential that the area of interest is as close to the center of the
compression paddle as possible.
3-3-6 Compression
• Compress the breast by using the manual compression knobs on the Stereotactic Positioner.
Compression must be firm to prevent movement of the breast, but not so firm as to cause discomfort
or to cause the breast to protrude through the paddle aperture.
WARNING
The Senographe Essential Acquisition System was designed for patient breast thicknesses of
less than 100 mm when under compression. The system is not recommended for use with patient
breast thickness exceeding 100 mm under compression.
• The system assumes that compression is applied when the distance between the Stereotactic
Positioner paddle and the breast support is reduced to less than 100 mm. It assumes that
compression has been removed, and therefore require the acquisition process to be restarted, when
the distance is increased to more than 105 mm.
• The system monitors breast thickness. When the operator defines targets, the system will give a
warning if the breast thickness has changed during the exam. See Chapter 13 Stereotaxy Punctures.
Needle Setup
1 Needle parameters
• This chapter describes Needle Setup procedures. These
procedures require the entry of biopsy device and needle
Tube head
dimensions.
• It is important to obtain and enter these measurements correctly,
Risk of collision
so that the system can accurately position the needle, and can
take all dimensions into account when calculating permissible
movements and access to the target area.
Biopsy device
• Use of incorrect needle and biopsy device dimension can cause:
- Risk of collision between the angulating tube head and the
biopsy device.
- Risk of skin tear.
- Risk of collision with the breast support.
- Risk of missing the target (needle too long or too short).
• For information on the Needle Management function for creation and modification of needle
parameters, refer to section 2 Needle Setup on page 104.
• For information on measurements required for needle management, refer to section 3 Needle and
biopsy device dimensions on page 108.
• For specific information on the measurement of biopsy devices for use in vertical approach, refer to
section 4 Biopsy device measurements on page 114.
WARNING
All needles and biopsy devices to be used should be programmed in the system before a patient
procedure is started.
All needle measurements (needle and biopsy device parameters) must be entered in millimeters
in the Needle Panel.
Needle Setup
2 Needle Setup
2-1 The Needle Panel
The needle to be used for a puncture is selected from a list of available needles.
To manage the needle list, i.e., to declare and specify a new needle, modify the specification of an
existing needle, or delete one that is no longer used, open the Needle Panel and use the Needle
Management function.
To open the Needle Panel window:
• Click the Needle or Needle & Puncture button in the Stereotaxy Viewer Needle & Puncture
Control Panel (see Chapter 12 Stereotaxy Image Acquisition section
2 Stereotaxy viewer on page 126). If a valid puncture target is currently
selected, the legend on the button is Needle & Puncture, and a puncture control panel is displayed
below the needle description.
Or:
• Click the Tools icon, and then select Needle Setup from the drop down menu.
The Needle Panel is displayed:
Needle Panel
Type FNA
Fna72
Name FNA72
Core92
Hook107
Needle (A) 110 mm
VacuumDevice
Needle guide (n) 4.0 mm
Modify Delete
Create needle
Quit
Note:
Buttons for OK and Cancel are displayed only after a function (Create needle, Modify, or Delete)
has been selected.
Needle Setup
Needle Setup
Needle Setup
Tip (C): 20 mm
Quit
Needle Setup
Needle Setup
3-4 Core Biopsy or Vacuum Assisted Device without holder in vertical approach
Measure and enter the following dimensions when creating or modifying either a core biopsy device
(CB), or a vacuum assisted device without holder (VAD without holder) for use in vertical approach.
Refer to section 4 Biopsy device measurements on page 114 for more information:
• Needle post-fire (A):
Post-fire length from the tip of the needle to the
bottom of the body of the biopsy device (see
drawing). The needle length must be measured in a
D
post-fire position (i.e. with the biopsy device not
cocked)
• Notch (B):
Length of the collection chamber.
n
• Tip (C):
Distance from the end of the collection chamber to
the tip of the needle.
• Needle guide (n):
Thickness of the needle guide.
A
• Device (D):
Length of the biopsy device body. Measure from the
top of the biopsy device to the bottom of the body. B
This length is taken into account when calculating the
clearance available for angulation: you must include
any protruding buttons. If tubes are present, include C
an allowance for the vertical part of the tubes (to the
point where their flexibility allows them to bend).
Needle Setup
Needle Setup
Note:
When using a vacuum assisted biopsy device with vertical holder, the bottom of the body of the
biopsy device may be above the top of the needle holder support (Mammotome ST or Biopsys), or
may be below it (SenoRx EnCor). In the latter case, the measured distance e is negative, and the
value E may be negative. The two conditions are shown here:
D
+
+
e
E
0 3 mm
3 mm 0 e
E
-
-
B A
C B
C
The bottom of the biopsy device is above the The bottom of the biopsy device is below the
holder; e is positive. holder; e is negative.
E = measured distance e + 3 mm E = 3 mm - |measured distance e|
Needle Setup
WARNING
Regardless of the type of needle used, do not enter the
actual needle length for Needle (A)
Instead, use the dimension shown in the diagram, measured
as described below.
A
XX= 0.0
+ 0.0mm
mm
Y = 0.0 mm
Z = 0.0 mm
Needle Setup
8. Open the Needle Panel again, select the previous lateral arm and click on Modify. Change the value
in field Length (A) to be equal to A + Z, where A is the current value (e.g., 107.2 mm) and ∆Z is the
value obtained in step 7.
Example: (A) = 107.2 mm, ∆Z = +1.0 mm
New value (A) = 108.2 mm
Needle Panel
Needle description
Needle Management
OK Cancel
Modify Delete
Create needle
Puncture
Quit Send
Needle Setup
Needle Setup
Quit
Needle Setup
A
B
C
Notch (B)
Tip (C)
Before any stereotactic procedure, the required dimensions must have been carefully measured and
entered in the Needle Description for each needle and each configuration to be used.
Note:
The BARD Magnum biopsy device has two possible throw lengths, 15 mm and 22 mm.
It may also be used with a coaxial cannula.
The throw length in use affects dimensions for needle length (A) and notch length (B).
Be sure to measure A and B after firing with the throw length to be used (15 mm or 22 mm).
Using the cannula reduces the value of (A) and increases the value of (D).
If the cannula is to be used, be sure to measure A and D with the cannula in place.
Needle Setup
Tip (C): 20 mm
Quit
Needle Setup
A
B
C
D D Device (D)
Notch (B)
B B
B Tip (C)
C C C
Before any stereotactic procedure, the required dimensions must have been carefully measured and
entered in the Needle Description for each needle and each configuration to be used.
Note:
The BARD Vacora biopsy device is normally used with a coaxial cannula which enables insertion and
removal of the biopsy device several times to collect samples.
It may also be used with a spacer to reduce the effective length of the notch.
The presence of a cannula reduces the value of (A) and increases the value of (D).
The presence of a spacer reduces the value of (A) and (B), and increases the value of (D).
Any cannula and spacer to be used must be in position on the device when making the measurements.
Needle Setup
A
B
C
Device (D)
Needle guide
(n)
Needle post-fire
(A)
Notch (B)
Tip (C)
Before any stereotactic procedure, the required dimensions must have been carefully measured and
entered in the Needle Description for each needle and each configuration to be used.
Needle Setup
Needle list
Type VAD with holder
Fna72
Name VADwith2
VADwith1
VADwithout1
Hook107 Needle post-fire (A) 132.0 mm
Create needle
OK Cancel
Quit
Needle Setup
D
+
E
0
-
A
B
C
Device (D)
Notch (B)
Tip (C)
Before any stereotactic procedure, the required dimensions must have been carefully measured and
entered in the Needle Description for each needle and each configuration to be used.
Needle dimensions: In the needle dimension fields for (A), (B), and (C), enter the measured dimensions
(in mm) for the needle in use with the biopsy device.
Device dimensions: In the biopsy device dimension fields (D) and (E), enter the measured dimensions
(in mm) of the biopsy device (E = e + 3 mm).
Needle Setup
D
+
E
0
-
A
B
Device (D)
C
Needle post-fire
(A)
Notch (B)
Tip (C)
Before any stereotactic procedure, the required dimensions must have been carefully measured and
entered in the Needle Description for each needle and each configuration to be used.
Needle dimensions: In the needle dimension fields for (A), (B), and (C), enter the measured dimensions
(in mm) for the needle in use with the biopsy device.
Device dimensions: In the biopsy device dimension fields (D) and (E), enter the measured dimensions
(in mm) of the biopsy device (E = e + 3 mm).
Needle Setup
D
Device (D) +
E
0
-
A
B
C
Needle post-fire
(A)
Notch (B)
Tip (C)
Before any stereotactic procedure, the required dimensions must have been carefully measured and
entered in the Needle Description for each needle and each configuration to be used.
Needle dimensions: In the needle dimension fields for (A), (B), and (C), enter the measured dimensions
(in mm) for the needle in use with the biopsy device.
Device dimensions: In the biopsy device dimension fields (D) and (E), enter the measured dimensions
(in mm) of the biopsy device (E = e + 3 mm).
Needle Setup
Stereotaxy panel button , in the Viewer Command Window. Refer to section 2-4 Control
Panels on page 130.
Note:
The stereotaxy viewer is available only for acquisition, not for review. Refer to Chapter 14 Image
Printing and Review, section 2 Image Review on page 158.
2 Stereotaxy viewer
2-1 Overview
The Stereotactic Viewer, shown below, includes two main areas:
• The Image Window or Viewing Area, in which images are displayed.
• A Command window at the left of the Viewer window. It contains:
- The Mosaic Image Browser (top).
- Control Panels (center);
- Middle mouse button function panel (bottom).
Command window
Mosaic Image
Browser
• Stereo mode
When the stereo images are used for placing
puncture markers, the images are shown at
two different resolutions.
The complete left and right images are shown
side by side in the upper half of the viewing Complete -15° image Complete +15° image
area, in “fit-to-view” format.
Portions of the same images are shown side
by side as high-resolution visualizers in the
lower half of the viewing area. The zoom
mode used is 1 : 1 (as used for Scout Mode).
The high-resolution visualizers are scrollable
to allow regions of interest to be shown. High resolution High resolution
-15° image +15° image
Primary Secondary
Annotation Level
Coordinates (mm)
Graphics & Meas.
Secondary
Check on Scout
Pairs Comparison
WARNING
Measurements shown in the Annotation control panel are valid only for images taken with the X-
ray beam perpendicular to the detector; they are not valid in angulation. When angulation is in
use, annotations may be used as relative values for comparison, but not as absolute
measurements.
Note:
Annotations added to the images are local only; they are not sent with the images to review
workstations or archiving devices.
2-4-3 The Stereotaxy Control panel
The Stereotaxy Control panel is used to define primary and secondary punctures, and to
access the Puncture window and Needle Management window. Refer to Chapter 13
Stereotaxy Punctures section 1 Defining punctures on page 135 for more information.
CAUTION
While the images are being analyzed on the workstation, the patient MUST remain motionless. It
is recommended that the patient should be supervised at all times.
In particular, patient motion between successive exposures (scout image, and left and right
images of the stereo pair) may result in clearly perceptible differences in position of the same
anatomical feature on the different images.
8. If more exposures are required (e.g., check exposures), set the tube arm to the required angulation
(0°, -15°, or 15°) and make the exposure as for the previous images.
Stereotaxy Punctures
1 Defining punctures
1-1 Overview
After acquiring one or more scout images and stereo pairs, you can define one or more puncture targets
using the Stereotaxy Viewer, as described in the following sections.
You can define several puncture sets (each consisting of a primary puncture and optional secondary
punctures) within the same exam.
You can at any time during the procedure add or delete graphic and text annotations, or make
adjustments to the contrast and brightness settings and invert function (to better see the
microcalcifications), using the Viewer command window. The magnifying glass, scroll, and contrast/
brightness functions of the Multi-Function mouse button are also available at all times.
Note:
Annotations added on the Acquisition Workstation are local, and are lost during image transfer to
a Review Workstation.
Stereotaxy Punctures
CAUTION
CAUTION
Check coincidence between Reference Markers and lead marker shadows on each scout image.
If necessary, correct the Reference Marker positions on the final scout image used for the
procedure.
Stereotaxy Punctures
• Define the puncture position. On each of the two images of the stereo pair, use the trackball or
mouse to select the Puncture Marker (a cross) and drag it until it is centered exactly over the
structure of interest (suspected lesion). Use the lower views for final positioning because of their
greater accuracy. If the position defined is within the puncture volume (accessible by the Stereotactic
Positioner), the legend OK or OK? is displayed beside the puncture number. If not, no text is
displayed beside the number. See section 1-8 Puncture set marker status on page 141.
Note:
If the system detects that the Stereo triplet used to define the puncture set may not correspond to
the position of the targeted area, it warns the operator with a pop-up message (this is due to a
thickness change, for instance) .
See Chapter 17 Operator messages Table 3 Patient position status change messages displayed
on the AWS on page 199. The message suggests that a new triplet must be acquired for the next
target creation.
- If the operator chooses to acquire a new triplet, the New Primary button becomes disabled
until a new triplet is acquired. The primary markers already set will show OK?
- If the operator chooses to continue the procedure when the system has detected a change in
breast compression, the legend of accessible puncture set is OK? The New Primary button is
still enabled. See section 1-8 Puncture set marker status on page 141.
CAUTION
The user must accurately select the same point on the two images. Inaccuracies may lead to
deviations when the needle is inserted in the lesion.
• The computed X, Y, and Z coordinates of the selected puncture target are displayed continuously in
the Coordinates (mm) window of the control panel.
Coordinates (mm)
X= X:
- 2.3
- 2.3
Y=26.37
y: 26.37
Z=23.28
Y z: 23.28
Z
X
Epipolar line
0
• The epipolar line display shows the location of the puncture target in relation to the
Show Epipolar
breast support. It should display the same value as the Z value in the Coordinates
window.
• By default, the epipolar line is hidden on the view. To show the epipolar line associated with the
marker, click the Show Epipolar button on the control panel.
Stereotaxy Punctures
The graduations along the epipolar line are in half centimeters of depth in the Z direction. The 0 indicates
the bottom, i.e., the breast support end (Z = 0).
When you move a marker up or down away from the epipolar line (in the Y direction) on one of the two
views, the corresponding marker on the other view will follow this movement as soon as you release the
mouse button. Both markers are displayed with the same Y coordinates, since for any structure of
interest (suspected lesion) the Y coordinates are the same on the left and right view.
Note:
The Y coordinate may vary from one image to another if the patient has moved between
exposures, or if it relates to two separate areas within the breast.
When you move a marker left or right along the epipolar line on one of the views, the corresponding
marker on the other view does not move. You will note, however, that the epipolar line moves relative to
the marker, to show the newly computed height (Z) that corresponds to the shift in the X direction
Show/hide markers and epipolar line
To view the image without the markers or the epipolar line:
• Click the Hide Marker or Hide Epipolar button on the control panel.
Hide Marker
The button legend changes to Show Marker or Show Epipolar.
• A second click on the button brings the markers, or the epipolar line, back into
Show Epipolar
view.
Delete markers
To delete any of the primary puncture markers you have defined previously:
• Select (highlight) the number of the puncture in the Primary puncture selection window by clicking on
it.
• Click the Delete Primary button.
Delete Primary
Note that deletion is not possible if the puncture information has already been sent to
the Stereotactic Positioner (status shown as S).
CAUTION
Be careful to superimpose the inverted T reference markers precisely before defining the
coordinates to the stereo positioner. The accuracy of the biopsy targets depends on the correct
positioning of these reference markers over the inverted Ts.
You can define a set of secondary markers, to perform sampling around the B
target area defined by the Primary Puncture Marker, at the same height (Z) as the
primary, but with an offset in X and Y.
The purpose is to provide a sample at a given offset around the primary. A C
Y
CAUTION D
Stereotaxy Punctures
• To add secondary markers, click the Secondary button. The secondary markers are
displayed around the primary marker on all displayed images. 5.5
• To cancel the definition of secondary markers, click again on the Secondary button.
Secondary
The secondary markers disappear from the display.
• The value of the X and Y offsets is shown in mm in the offset window above the
Secondary button. Minimum offset is 0.5 mm and maximum offset is 20.0 mm. To change the value
(by 0.5 mm steps), click the up or down arrow at the side of the window.
Note:
If the system detects that the Stereo triplet used to define the puncture set may not correspond to
the current position of the target area, it warns the operator with a pop-up message (for instance
due to a tickness change).
See Chapter 17 Operator messages Table 3 Patient position status change messages displayed
on the AWS on page 199. The message suggests that a new triplet be acquired for the next target
creation.
If the system has detected a change in breast compression, the legend displayed beside the
accessible secondary points is OK?. See section 1-8 Puncture set marker status on page 141.
Stereotaxy Punctures
Stereotaxy Punctures
Stereotaxy Punctures
Modify Delete
Create needle
Puncture
Quit Send
Warning(s)
4. Select the needle to be used for the puncture by clicking on its name in the list of predefined needles
shown in the Needle list.
The description of the selected needle is displayed in the Needle description window. A status field at
the bottom of the Needle & Puncture panel indicates whether or not the puncture is authorized, and a
message giving additional information may be displayed in the Warning(s) field.
WARNING
Always verify that the needle selected is indeed the needle that is in use.
Note:
It is highly recommended that the operator configures all needle characteristics (length, notch
width for a biopsy needle, etc.) BEFORE starting the exam (from the Tool menu of the Main
Browser or during the accuracy test procedure).
However, the operator can still create a new needle or modify an existing one during the
procedure.
Stereotaxy Punctures
Needle list
Modify Delete
Create needle
Puncture
Quit Send
Warning(s)
CAUTION
It is the responsibility of the operator to decide whether the thickness variation is acceptable for
the definition of the stereotactic puncture(s) to be performed, or whether the acquisition of new
images is required.
Stereotaxy Punctures
CAUTION
When using a Core biopsy or Vacuum Assisted Device VAD in vertical approach,
there can be a risk of collision between the tube head when angling the tube arm
to perform the pre-fire and post-fire images. The operator must take great care to avoid this risk.
The risk of collision with the tube head is especially great when trying to puncture a target with a
high Z. The system recognizes this situation and warns the operator by displaying the following
message in the Warning(s) panel: Angulation motion limited. Angulation from -15° to 0° and from
+15° to 0° is forbidden.
Needle Panel
Puncture
Quit Send
Warning(s)
Stereotaxy Punctures
• The two conditions described above combine to limit punctures with a Mammotome in vertical
approach to puncture targets in a range of Z coordinates. The limits of these ranges are:
Zmin Zmax
Mammotome Biopsys 11G 13 mm 43mm
Mammotome ST 11G 13 mm 67mm
• When an interlock is required, it is activated as soon as the needle path is opened. The interlock is
then released as soon as the needle path is closed.
2-2 Correction
The combination of a puncture target position too close to the breast Before After
support (in the “danger zone”) and the use of a core biopsy needle correction correction
or a VAD in vertical approach is not accepted by the system,
because of the risk of the tip of the needle hitting the breast support.
In this case the Send button is inactive.
The “puncture target position” corresponds to the center of the notch Puncture
of the core biopsy needle. If the puncture can be performed with the target
structure to be punctured still located within the lower part of the
notch, the Needle Status display shows Corrective action possible
and the Correction button is active.
In this case, clicking the Correction button will move the puncture
target position upwards, so that the puncture is accepted by the
system. Click on Undo to remove the correction.
It is the operator’s responsibility to decide whether the nature of the
Breast support Danger zone
structure to be punctured, and the width of the notch of the core
biopsy needle, allow a satisfactory puncture under these conditions.
CAUTION
1. When corrective action is accepted by clicking the Correction button, clicking Send performs
the corrective action automatically. Do not also make the correction manually (e.g., by moving the
positioner Z knob), or you will miss the target.
2. Corrective action is not proposed if the target is too low (i.e., if the target defined is lower than
the tip of the needle plus the danger zone height).
Stereotaxy Punctures
Send
2. The puncture data is transferred to the Stereotactic Positioner, and the Puncture panel closes.
3. The needle holder can now be positioned ready for the puncture as described in section
3 Positioning the Needle Holder on page 147.
Stereotaxy Punctures
Needle holder
• The Stereotactic Positioner is equipped with a needle path block. Its purpose is to prevent any
motorized movement of the Stereotactic Positioner when a needle is being inserted or already in the
breast or phantom.
The needle path block is controlled by the small lever next to the Z manual adjustment knob. When
this is pushed to the right, a plastic disk between the two needle holder supports blocks the needle
path.
• For patient safety, the motorized movement of the Stereotactic Positioner can only be used if the
needle path is blocked. In this position it is not possible to insert a needle in the holder.
• To open the needle path for the insertion of a needle, push the lever to the left.
• When the needle path is open, all motorized movement of the positioner is inhibited.
Note:
The needle path is always blocked when using the lateral arm. Only manual movement is allowed
when the lateral arm is in use.
WARNING
The needle path block does not prevent manual movement of the positioner using the X, Y, Z
knobs.
It is the responsibility of the operator to ensure that the positioner manual controls are not moved
after the needle is inserted.
WARNING
Needle path block test:
The needle path block must be tested on each day that stereotactic exams are performed, before
the first exam of the day.
If the needle path block is broken, the Stereotactic Positioner must NOT be used.
Stereotaxy Punctures
Next target
3-2-1 Go To Park P
The Go To Park button is used to move the needle holder out of the field of view, to obtain the best view
of the area of interest.
1. Press and hold the Go To Park button.
The needle holder starts moving to the park position, i.e., centered left-right (X), maximum
displacement towards the rear (Y), and centered height (Z).
2. Hold the button down until the needle holder has reached this position and stops, then release the
button.
When the needle holder is at the park position, the light beside the control buttons on the Stereotactic
Positioner will blink green.
During the operation, the display unit shows the distances X,Y,
and Z between the current position and the park position. parking X=+01.3
Z=+012.3 Y=-08.5
3-2-2 Go To Target
The Go To Target button is used to move the needle holder directly over the target.
1. Press and hold the Go To Target button.
The needle holder starts moving to the target position. The distances X, Y, and Z between the current
position of the needle holder and the target position are displayed continuously on the Display Unit.
2. Hold the button down until the needle holder has reached the target position and stops, then release
the button.
3. During the operation, the display unit continues to show the distances
X,Y, and Z between the current position and the destination position, P3 X=+00.0
until the three coordinates become zero at the destination position. Z=+002.3 Y=-00.0
Stereotaxy Punctures
4. When the needle holder is at the destination position, the light beside
the Go To Target button illuminates green, and the display unit shows TARGET REACHED
the message TARGET REACHED.
5. To confirm that the correct needle is in use:
Press the Go To Target button again when the needle holder is at the
destination position. FNA: N1
NL = 110.0
The display unit now shows the programmed needle characteristics:
- Needle type:
Display Declared needle type
FNA Fine Needle Aspiration FNA
CORE Core Biopsy CB
HOOK Pre-surgical hook Hook
VAD Vacuum Assisted Device with vertical holder VAD with holder
VADN Vacuum Assisted Device without vertical holder VAD without holder
LAT Lateral Approach Tool LAT
- Needle name
- Needle dimension.
The example shows the display for an FNA needle named N1, having a length of 110 mm.
Note:
If the needle holder is moved/located to a position where the tip of the needle would enter the
“danger zone” (i.e., it would reach a position close to the breast support plate, where it may touch
the plate), the light next to the Go To Target button illuminates red, warning you that there is a risk
of the needle hitting the plate.
Stereotaxy Punctures
the four positions A, B, C, D, shown in section 1-5 Placing the Secondary Puncture Markers on
page 138).
To cycle through the targets, press the button repeatedly. After the last target, the display
automatically wraps around, back to the first puncture target.
An asterisk (*) is displayed in front of the puncture target reference
when the target (primary or secondary) has been reached. *P3 X=00.0
Z=000.0 Y=00.0
3-3 Manual Positioning
Manual positioning of the needle holder of the Stereotactic Positioner can be used in the following cases:
• To perform fine tuning of the X, Y and Z coordinates to make slight changes from the values
computed on the workstation. For instance, when placing of a hook wire for pre-surgical localization.
Final adjustment of the Z coordinate is done manually.
Note:
If the needle holder is moved/located to a position where the tip of the needle would enter the
“danger zone” (i.e., it would reach a position close to the breast support plate, where it may touch
the plate), the light next to the Go To Target button illuminates red, warning you that there is a risk
of the needle hitting the plate.
• When using the lateral approach arm accessory (only manual positioning is possible, in this case).
• To correct the position of the biopsy device after viewing pre-fire or post-fire images, when the biopsy
device is still in place.
Stereotaxy Punctures
Needle holder
• Move the tube arm to the right hand or left hand side Park position (see Chapter 10 Preparation for
examinations section 2 Arm Positioning on page 94).
This frees the space above the needle holder, to facilitate the installation of needle guides and
needles.
Note:
At the Park positions, exposure is inhibited; a warning message appears on the Senographe
Control Console read-out if exposure is attempted.
Stereotaxy Punctures
CAUTION
Whatever the type of puncture performed, check exposures must be made with the needle in
place in the breast.
CAUTION
Angulating to the opposite side when the biopsy device is in place is not possible if a risk of
collision between the biopsy device and the tube head covers has been detected. In that case,
only a check in the current angulation is enabled. It is the responsibility of the user to determine
whether this check is sufficient to accept the puncture or not.
Under some circumstances, it may be possible to override the interlock, after temporarily removing the
biopsy device to avoid the danger of collision. Refer to Chapter 10 Preparation for examinations, section
2-5 Angulation interlock on page 97.
4-3-1 Performing Check Exposures
As soon as the puncture data is transferred to the Stereotactic Positioner by pressing the Send button,
the Senographe workstation returns to the Stereotaxy application screen, ready for further exposures,
such as pre-fire and post-fire check exposures during core biopsy, or exposures to verify correct
positioning of a pre-surgical localization hook.
To perform check exposures:
• Reposition the tube arm as required (angulation +15°, -15°, or 0°).
• Make the check exposures as described in Chapter 12 Stereotaxy Image Acquisition, section
3 Making Stereotaxy exposures on page 132.
Stereotaxy Punctures
CAUTION
If there is a perceptible difference between the needle position and the puncture target (for
example, due to patient motion), it is the responsibility of the operator to either accept the needle
position or to define a new puncture set.
4-3-2 Pairs Comparison Function
The display mode Pairs Comparison shows two different stereo pairs simultaneously, allowing
comparison of stereo pair images captured at different instants of the procedure, e.g., pre-fire and post-
fire images.
1. Select the stereo pair to be used as the main stereo pair for reference (e.g., the pre-fire or post-fire
images), by clicking on it in the Mosaic Image Browser. The selected pair is displayed in Stereo
mode.
2. Click the Pairs Comparison button. The legend on the button changes to Back to Pair.
3. Select the stereo pair to be used as the second stereo pair for comparison, by clicking on it in the
Mosaic Image Browser.
4. The selected pair is displayed in the upper half of the screen, with the main stereo pair shown in the
lower half.
See Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on page 127 for more
information.
Stereotaxy Punctures
5 Additional punctures
5-1 Performing secondary punctures
If secondary punctures are required, proceed as follows for each puncture in turn:
1. After completion of the previous puncture procedure, remove the needle and close the needle path in
preparation for the next puncture (move the needle path block lever to the right, as described in
section 3-1 Needle Path block on page 147).
2. Select the next secondary puncture, using the Next Target button on the Positioner.
3. Move the needle holder to the secondary puncture position using the Go to Target button, or
manually with X, Y and Z knobs.
4. Proceed exactly as described for the primary puncture.
Stereotaxy Punctures
WARNING
When pre-surgical localization is performed, the operator must ensure that after removal of the
needle:
1. The compression paddle is released slowly.
2. The localization wire emerging from the skin is moved carefully, so as not to move the wire
hook from the center of the lesion.
Stereotaxy Punctures
1 Printing Images
1-1 Introduction
Print management is the same as for the screening/diagnostic application, described in the Senographe
Essential Acquisition System Operator manual.
The chapter on Printing in the Senographe Essential Acquisition System Operator manual is applicable
for the Stereotaxy Application, except for the section describing Print Annotation Models.
Note:
Printing of stereotaxy images on 10" x 12" film is not recommended. The effective image size is
too small to be useful.
WARNING
All measurement calculations for graphical
annotations (length, surface area, etc.) are
made in a reference plane which is 2 cm
above the breast contact surface. These
values therefore do not correspond to the
true size of an object unless it is situated in
the reference plane.
It is not possible to determine in the 2D
projection image if a pathological feature is
situated in the reference plane, so these
measurement tools must not be used to
measure the size of pathological features.
Furthermore, it is not possible to duplicate
breast position from one exam to another,
so they must not be used to compare the
size of pathological features over time.
Films printed using the stereo model contain puncture information (e.g., targets, angulation, etc.).
2 Image Review
To review stereo images stored on the workstation image disk, return to the Browser screen.
• In the Browser screen, select the desired stereo images. First select the patient, then the exam
(study), by clicking on them.
Note:
Stereo images can be recognized in the Browser images list because the angulation (0°, +15°, -
15°) is shown in the Angulation field.
• Click the Review exam button.
• When the Review exam button is clicked after selection of a series created by the Stereotaxy
application, the Stereotaxy Viewer is launched automatically. The Viewer window opens and displays
the first selected image.
• In the Viewer Control Panel, the same functions that are available during stereo image acquisition
(refer to Chapter 12 Stereotaxy Image Acquisition, section 2-4 Control Panels on page 130) are also
available in review mode. However, the stereo puncture data defined during the procedure are
recorded as part of the exam and cannot be modified or deleted during image review.
• The following needle and target conditions are displayed when an exam is opened in review mode:
- If the status of a primary target was S (sent) at acquisition, its status appears as S in review
mode.
- If the status of a primary target was OK, OK?, or blank, at acquisition, its status appears as “--” in
review mode.
- The Needle button is enabled only if the status is S.
If the Needle button is clicked and the Needle Panel opens, the needle used for sending the
target is selected by default, and the user is not able to change this selection.
- The needle parameters shown are those of the needle at the time the target was sent.
- Whatever the status of secondary targets in acquisition mode, no status is displayed for them in
review mode (displayed as blank).
• The Mosaic Image Browser is available during stereo image review. Refer to Chapter 12 Stereotaxy
Image Acquisition, section 2-2 Mosaic Image Browser on page 127 for more information.
• The Function Panel is available during stereo image review. Refer to the Viewer chapter in the
Senographe Essential Acquisition System Operator manual for more information.
System Hygiene
WARNING
Adequate cleaning, disinfection, and sterilization is necessary to prevent disease transmission.
Be sure to keep all equipment in a clean condition.
Thoroughly clean and disinfect or sterilize all equipment surfaces that contact the patient, and all
equipment surfaces likely to become soiled during use.
The tables below indicate typical requirements, but they do not cover all situations, and should be
used only as a guide.
System Hygiene
3 Hygiene practices
Part category Example Conditions Required hygiene practice
No contact Display panel on the N/A. Cleaning :
Stereotactic Positioner Follow the instructions given in the
System Hygiene chapter in the
Senographe Essential Acquisition
System Operator Manual.
Contact Stereotactic Positioner No risk of cross Disinfection - low level:
body, including: contamination. Follow the instructions given in the
- Needle path block lever System Hygiene chapter in the
- Rotation/Angulation button Senographe Essential Acquisition
- Movement control buttons System Operator Manual.
- Manual control knobs (X,
Risk of cross Disinfection - high level:
Y, Z)
contamination Follow the instructions given in the
Compression paddles
System Hygiene chapter in the
Breast support
Senographe Essential Acquisition
All parts of the Lateral and
System Operator Manual.
Vertical Holder kits
Acute contact Metal needle guides (also N/A Sterilization:
known as metal bushings). Follow the instructions given in the
Operator Manual Sterilization of metal
components used in Stereotaxy
procedures .
Planned Maintenance
1 Introduction
This chapter describes maintenance procedures intended to maintain the performance of the equipment
at the highest level, including quality control tests for the Senographe Stereotaxy.
The chapter includes the following sections:
• Section 2 Field Engineer Maintenance on page 161
• Section 3 QC Tests for the Radiologic Technologist or Medical Physicist on page 161
• Section 4 Test Result Record sheets on page 185
Planned Maintenance
Frequency:
On each day that stereotactic exams are performed, before the first exam of the day.
If no stereotactic exams are carried out, the test should be carried out at least once a month to ensure
continued safe use.
Objective:
To ensure that motorized movement of the needle positioner is inhibited when the needle path is open
(unblocked).
CAUTION
The purpose of the needle path block is to inhibit any motorized movement of the positioner when
a needle is installed. The needle path block does not prevent manual movement of the positioner
using the X, Y, and Z knobs.
Equipment: None
Action Limit:
- The positioner must not make any motorized movement when the needle path is open.
- If it is possible to insert a needle with the needle path closed, or if motorized movement of the needle
positioner is possible when the needle path is open, the needle path block has malfunctioned, and it
is recommended that you contact your GE Service Engineer.
3-1 Procedure
1. Manually move the needle positioner so that the needle path is approximately centered in the field of
view.
2. Check that it is not possible to insert a needle when the needle path is closed (needle path block
lever to the right).
3. Open the needle path by moving the needle path block lever to the left.
4. Activate the Go To Park button on the positioner
5. Record completion of the test in section QC Test Chart 1- NeedlePath Block Test on page 186.
Frequency:
On each day that stereotactic exams using the vertical approach are to be performed, before the first
exam of the day. If no stereotactic exams are carried out, the test should be carried out at least once a
month to ensure continued accuracy. See Chapter 7 Procedure Steering Guide section 3 Pre-
examination check on page 45.
Objective:
To ensure localization accuracy in the vertical approach
Equipment:
Pin phantom L19 and calibration needle 120mm.
Action Limit:
The magnitude of the offset for any axis (X, Y, or Z) must not exceed 1 mm.
Record completion of the test in section QC Test Chart 2 - Stereotactic
Localization Accuracy: Vertical Approach on page 187.
Pin phantom
3-2 Procedure
The procedure used for the test with the phantom is similar to a normal stereotactic patient procedure.
3-2-1 Preparation
1. Prepare the Senographe and install the Stereotactic Positioner as described in Chapter 6 Installation
of the Stereotactic Positioner, section 1 Installing the Stereotactic Positioner on page 43.
2. Open the Worklist window on the Senographe AWS and
verify that the selection for Type of Exam is Stereo. Browser
3. On the Worklist, select the “Phantom” Patient or create it. Type of Exam Stereo Start Exam
4. If a “Phantom” Patient has not yet been defined for use
with the phantom accuracy tests, select New Patient... in
the Worklist, and complete the new Medical Procedure Card as required, with “Phantom” as the
Patient Name, before starting image acquisition. Refer to Chapter 12 Stereotaxy Image Acquisition
section 1 Starting the Stereotaxy application on page 125 for details.
5. Click Start Exam. The Stereotactic Viewer appears, with the Viewer area at the right of the screen
remaining black until an exposure is taken. Refer to Chapter 12 Stereotaxy Image Acquisition section
2 Stereotaxy viewer on page 126.
6. Position the Pin Phantom under the compression paddle so that the tip of the phantom rod faces the
opening in the paddle, and is approximately in the middle of the volume (Z = 30 mm).
7. Use the compression knobs on either side of the Stereotactic Positioner to apply just enough
compression to hold the phantom in place.
Note:
- When the stereotaxy compression paddle leaves the park position (i.e., it is less than 10 cm from
the breast support), the up/down and rotation movements of the Senographe Positioner Arm are
inhibited.
- After the phantom has been placed in position and compressed, it must not be moved until the
test has been completed.
8. Press the Rotation/Angulation button on the Stereotactic Positioner to switch from rotation mode to
angulation mode. The gantry readout now displays the stereo angle, not the gantry angle. Refer to
Chapter 10 Preparation for examinations, section 2 Arm Positioning on page 94.
9. Press the Go to Park button P on the Stereotactic Positioner to move it to the parking position.
3-2-2 Exposure
1. Select Manual exposure mode (Mo-Mo, 24kV, 4mAs) on the X-ray Console. Select laterality.
2. Make the scout exposure. The scout image appears in the Viewer area.
3. The scout image shows the inverted T reference marks, the projection of the phantom rod, and the
projection of the compression paddle aperture.
4. Verify that the projection of the tip of the rod appears inside the projection of the compression paddle
aperture.
5. Adjust the window width and level if necessary for best viewing level.
6. Move the cursor onto the center of one of the two Reference Markers . Press and hold the left
mouse button until the cursor icon becomes a pencil, then drag the Reference Marker to place it
exactly on top of one of the images of the inverted T reference marks, as described in Chapter 13
Stereotaxy Punctures section 1 Defining punctures on page 135. Repeat the procedure to drag the
other Reference Marker and place it on top of the image of the other inverted T reference mark.
Use the invert function (Chapter 12 Stereotaxy Image Acquisition, section 2-4 Control Panels on
page 130) to confirm the positioning.
7. Make the two stereo exposures, with the tube angled to +15°, then to -15°.
See Chapter 10 Preparation for examinations section 2 Arm Positioning on page 94 for a description
of tube arm movements.
See Chapter 12 Stereotaxy Image Acquisition section 3 Making Stereotaxy exposures on page 132
for an explanation of the exposure sequence.
The stereo images are displayed in the monitor Viewer area in Stereo mode (the images in lower half
of the viewing area are zoomed images of the top half; see Chapter 12 Stereotaxy Image Acquisition
section 2 Stereotaxy viewer on page 126).
Use the right mouse button to click on the area of interest on each of the top images, to center the
lower zoomed images on the area of interest.
3-2-3 Define target
1. Click (left mouse button) on the Stereotaxy selection button to open the Stereotaxy Panel.
2. Click the New Primary button to define a new puncture set, then select the puncture
New Primary
marker and drag it over the tip of the rod on each of the zoomed images. Refer to
Chapter 13 Stereotaxy Punctures section 1-4 Placing the Primary Puncture Marker
on page 136.
Note:
You must accurately select the same point seen on the two images. Inaccuracies may lead to
deviations when the needle is being inserted in the phantom.
3. Click the Check on Scout button to display the scout image in the top half of the screen. This allows
you to check that the correct puncture site has been selected.
See Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on page 127 for more
information on this feature.
Planned Maintenance QC Test 3 - Accuracy check before using new CB or VAD needles in vertical approach
Frequency:
Whenever a new CB or VAD needle is defined or a new type of needle is to be used.
Note:
For needles used without a biopsy device (FNA or Hook), use the procedure described in section
QC Test 4 - Accuracy check before using new FNA/Hook-wire needles in vertical approach on
page 172.
Objective:
Confirm needle definition whenever a new gauge or length needle is
introduced for an existing device.
Equipment:
Pin phantom L19, biopsy device and the new CB or VAD needle to be used.
Action Limit:
The magnitude of the offsets for the axes (X, Y, and Z) is recorded during
the check. The target must be inside the central part of the needle notch.
Record completion of the test in section QC Test Chart 3 - Accuracy check Pin phantom
before using new CB or VAD needles in vertical approach on page 188.
3-3 Procedure
The procedure used for the test with the phantom is similar to a normal stereotactic patient procedure.
3-3-1 Preparation
1. Prepare the Senographe and install the Stereotactic Positioner as described in Chapter 6 Installation
of the Stereotactic Positioner, section 1 Installing the Stereotactic Positioner on page 43.
2. Open the Worklist window on the Senographe AWS and
verify that the selection for Type of Exam is Stereo. Browser
3. On the Worklist, select the “Phantom” Patient or create it. Type of Exam Stereo Start Exam
If a “Phantom” Patient has not yet been defined for use with
the phantom accuracy tests, select New Patient... in the
Worklist, and complete the new Medical Procedure Card as required, with “Phantom” as the Patient
Name, before starting image acquisition. Refer to Chapter 12 Stereotaxy Image Acquisition section
1 Starting the Stereotaxy application on page 125 for details.
4. Click Start Exam. The Stereotactic Viewer appears, with the Viewer area at the right of the screen
remaining black until an exposure is taken. Refer to Chapter 12 Stereotaxy Image Acquisition
section 2 Stereotaxy viewer on page 126.
5. Position the Pin Phantom under the compression paddle so that the tip of the phantom rod faces the
opening in the paddle and is approximately in the middle of the volume (Z = 30 mm).
6. Use the compression knobs on either side of the Stereotactic Positioner to apply just enough
compression to hold the phantom in place.
Note:
- When the stereotaxy compression paddle leaves the park position (i.e., it is less than 10 cm from
the breast support), the up/down and rotation movements of the Senographe Positioner Arm are
inhibited.
Planned Maintenance QC Test 3 - Accuracy check before using new CB or VAD needles in vertical approach
- After the phantom has been placed in position and compressed, it must not be moved until the
test has been completed.
7. Press the Rotation/Angulation button on the Stereotactic Positioner to switch from rotation mode to
angulation mode. The gantry readout now displays the stereo angle, not the gantry angle. Refer to
Chapter 10 Preparation for examinations, section 2 Arm Positioning on page 94.
8. Press the Go to Park button P on the Stereotactic Positioner to move it to the parking position.
3-3-2 Exposure
1. Select Manual exposure mode (Mo-Mo, 24kV, 4mAs) on the X-ray Console, select laterality.
2. Make the scout exposure. The scout image appears in the Viewer area.
The scout image shows the inverted T reference marks, the projection of the phantom rod, and the
projection of the compression paddle aperture.
3. Adjust the window width and level if necessary for best viewing level.
4. Move the cursor onto the center of one of the two Reference Markers . Press and hold the left
mouse button until the cursor icon becomes a pencil, then drag the Reference Marker to place it
exactly on top of one of the images of the inverted T reference marks, as described in Chapter 13
Stereotaxy Punctures section 1 Defining punctures on page 135. Repeat the procedure to drag the
other Reference Marker and place it on top of the image of the other inverted T reference mark.
Use the invert function (Chapter 12 Stereotaxy Image Acquisition, section 2-4 Control Panels on
page 130) to confirm the positioning.
5. Make the two stereo exposures, with the tube angled to +15°, then to -15°.
See Chapter 10 Preparation for examinations section 2 Arm Positioning on page 94 for a description
of tube arm movements.
See Chapter 12 Stereotaxy Image Acquisition section 3 Making Stereotaxy exposures on page 132
for an explanation of the exposure sequence.
6. The stereo images are displayed in the monitor Viewer area in Stereo mode (the images in lower half
of the viewing area are zoomed images of the top half; see Chapter 12 Stereotaxy Image Acquisition
section 2 Stereotaxy viewer on page 126).
7. Use the right mouse button to click on the area of interest on each of the top images, to center the
lower zoomed images on the area of interest.
Planned Maintenance QC Test 3 - Accuracy check before using new CB or VAD needles in vertical approach
1. Click (left mouse button) on the Stereotaxy selection button to open the Stereotaxy Panel.
2. Click the New Primary button to define a new puncture set, then select the puncture
New Primary
marker and drag it over the tip of the rod on each of the zoomed images. Refer to
Chapter 13 Stereotaxy Punctures section 1-4 Placing the Primary Puncture Marker
on page 136.
Note:
You must accurately select the same point seen on the two images. Inaccuracies may lead to
deviations when the needle is being inserted in the phantom.
3. Click the Check on Scout button to display the scout image in the top half of the screen. This allows
you to check that the correct puncture site has been selected.
See Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on page 127 for more
information on this feature.
3-3-4 Prepare needle
1. Click the Needle & Puncture button to open the Needle & Puncture panel. Note that this panel opens
only if the puncture status is OK.
2. Select the new pre-programmed needle or program it as described in Chapter 11 Needle Setup
section 2 Needle Setup on page 104.
3. Click the Send button.
4. Angle the tube arm to either of the parking positions, to ease insertion of the needle. Refer to
Chapter 10 Preparation for examinations, section 2-4-1 Angulation movement for stereotaxy on
page 96.
5. With the needle path block closed, press and hold the Go to Target button on the Stereotactic
Positioner until the target position is reached; the light will now be green.
See Chapter 13 Stereotaxy Punctures section 4 Performing the puncture on page 151 for more
information.
Note:
When the needle path is open, all motorized movements are inhibited. Manual movement is
allowed with the needle path open.
3-3-5 Position the needle
1. Press the Go to Target button again to display the programmed needle. Check that the
correct needle has been chosen.
2. Install the appropriate needle guides, and insert the needle (with the biopsy device in its pre-fire
state) through the needle guides
3. The tip of the needle must be a few millimeters above the target.
4. Acquire a pre-fire stereo pair to verify the positioning of the puncture tool.
5. Turn the Y knob slightly so as to allow the device to be fired without impacting the tip of the pin. Make
all movements very carefully, and ensure that the notch remains facing the tip of the pin.
6. Fire the device.
7. Reverse the movement of the Y knob to return to the previous position (Y = 0).
8. Check that the target is inside the notch, at a position you would accept for a puncture (usually
around the center of the notch).
Planned Maintenance QC Test 3 - Accuracy check before using new CB or VAD needles in vertical approach
9. Perform a post-fire check by acquiring a pair of stereo views. The target must be inside the notch at
an acceptable position.
3-3-6 Repeat for other test points
• Decompress the Pin Phantom.
• Reposition it on the breast support, so as to have the tip of the rod in a different position in the
puncture volume. Turn it to use a different value of Z (do not use Z = 10, as the target would then be
too low). Press Go to Parking till parking position is reached, and move to a different location
(changing X and Y), under the paddle aperture area, away from the center.
• Carry out the procedures described above (sections 3-3-3 Define target on page 170, 3-3-4 Prepare
needle on page 170 and 3-3-5 Position the needle on page 170) for at least two other positions of the
target.
Note:
Depending on the device /needle length (post fire, notch or tip length) different warnings may
appear in the Needle Panel.
Planned Maintenance QC Test 4 - Accuracy check before using new FNA/Hook-wire needles in vertical approach
Frequency:
Whenever a new FNA/Hook-wire needle is defined or a new type of FNA/Hook-wire needle is to be used.
Note:
For needles used with a biopsy device (CB or VAD), use the procedure described in section QC
Test 3 - Accuracy check before using new CB or VAD needles in vertical approach on page 168.
Objective:
Confirm needle definition whenever a new gauge or length needle is introduced.
Equipment:
Vertical approach phantom and the new FNA needle.
Note:
The needle used must be long enough to touch
the bottom of the deepest hole of the phantom.
A needle length of at least 94-95 mm is
recommended when the phantom is
compressed to 50 mm.
• The phantom is a block measuring 50 x 100 x
72 mm (approximately 2 x 4 x 3 inches).
• Five holes of 1 mm diameter and different depths
are provided for use in accuracy tests. They are 5
positioned as shown in the diagram. The z value 3
4
for those five holes is as follows: 2
Reference 1 2 3 4 5 1
Z value 45 20 30 40 10
(mm)
Action Limit:
The magnitude of the offset for any axis (X, Y, or Z)
must not exceed 1 mm.
Record completion of the test in section QC Test
Chart 4 - Accuracy check before using new FNA needles in vertical approach on page 189.
3-4 Procedure
The procedure used for the test with the phantom is similar to a normal stereotactic patient procedure.
3-4-1 Preparation
1. Prepare the Senographe and install the Stereotactic Positioner as described in Chapter 6 Installation
of the Stereotactic Positioner, section 1 Installing the Stereotactic Positioner on page 43.
2. Open the Worklist window on the Senographe AWS and
verify that the selection for Type of Exam is Stereo. Browser
3. On the Worklist, select the “Phantom” Patient or create it. Type of Exam Stereo Start Exam
If a “Phantom” Patient has not yet been defined for use
with the phantom accuracy tests, select New Patient... in
Planned Maintenance QC Test 4 - Accuracy check before using new FNA/Hook-wire needles in vertical approach
the Worklist, and complete the new Medical Procedure Card as required, with “Phantom” as the
Patient Name, before starting image acquisition. Refer to Chapter 12 Stereotaxy Image Acquisition
section 1 Starting the Stereotaxy application on page 125 for details.
4. Click Start Exam. The Stereotactic Viewer appears, with the Viewer area at the right of the screen
remaining black until an exposure is taken. Refer to Chapter 12 Stereotaxy Image Acquisition section
2 Stereotaxy viewer on page 126.
5. Position the vertical approach phantom under the compression paddle so that the holes in the
phantom face the opening in the paddle.
6. Place 2cm or more of Plexiglas filtration at the rear of the phantom to prevent any possibility of ghost
images in the clinical images acquired immediately after QC tests.
7. Use the compression knobs on either side of the Stereotactic Positioner to apply just enough
compression to hold the phantom in place.
Note:
- When the stereotaxy compression paddle leaves the park position (i.e., it is less than 10 cm from
the breast support), the up/down and rotation movements of the Senographe Positioner Arm are
inhibited.
- After the phantom has been placed in position and compressed, it must not be moved until the
test has been completed.
8. Press the Rotation/Angulation button on the Stereotactic Positioner to switch from rotation mode to
angulation mode. The gantry readout now displays the stereo angle, not the gantry angle. Refer to
Chapter 10 Preparation for examinations, section 2 Arm Positioning on page 94.
9. Press the Go to Park button P on the Stereotactic Positioner to move it to the parking position.
3-4-2 Exposure
1. Select Manual exposure mode (Rh-Rh, 29 kV, 64 mAs) on the X-
ray Console.
2. Make the scout exposure. The scout image appears in the Viewer
area.
The scout image shows the inverted T reference marks, the
projection of the lead plugs contained in the phantom, and the
projection of the compression paddle aperture.
Verify that the projections of all lead plugs appear inside the
projection of the compression paddle aperture.
3. Adjust the window width and level if necessary for best viewing
Scout image
level.
4. Move the cursor onto the center of one of the two Reference Markers . Press and hold the left
mouse button until the cursor icon becomes a pencil, then drag the Reference Marker to place it
exactly on top of one of the images of the inverted T reference marks, as described in Chapter 13
Stereotaxy Punctures section 1 Defining punctures on page 135. Repeat the procedure to drag the
other Reference Marker and place it on top of the image of the other inverted T reference mark.
Use the invert function (Chapter 12 Stereotaxy Image Acquisition, section 2-4 Control Panels on
page 130) to confirm the positioning.
5. Make the two stereo exposures, with the tube angled to +15°, then to -15°.
See Chapter 10 Preparation for examinations section 2 Arm Positioning on page 94 for a description
of tube arm movements.
Planned Maintenance QC Test 4 - Accuracy check before using new FNA/Hook-wire needles in vertical approach
See Chapter 12 Stereotaxy Image Acquisition section 3 Making Stereotaxy exposures on page 132
for an explanation of the exposure sequence.
-15° +15°
Inverted T Reference
Markers
Lead plugs
6. The stereo images are displayed in the monitor Viewer area in Stereo mode (the images in lower half
of the viewing area are zoomed images of the top half; see Chapter 12 Stereotaxy Image Acquisition
section 2 Stereotaxy viewer on page 126).
7. Use the right mouse button to click on the area of interest on each of the top images, to center the
lower zoomed images on the area of interest.
3-4-3 Define target
1. Click (left mouse button) on the Stereotaxy selection button to open the Stereotaxy Panel.
2. Click the New Primary button to define a new puncture set, then select the puncture
marker and drag it over the same lead plug on each of the zoomed images. Refer to New Primary
Chapter 13 Stereotaxy Punctures section 1-4 Placing the Primary Puncture Marker
on page 136.
Note:
You must accurately select the same point seen on the two images. Inaccuracies may lead to
deviations when the needle is being inserted in the phantom.
3. Click the Check on Scout button to display the scout image in the top half of the screen. This allows
you to check that the correct puncture site has been selected.
See Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on page 127 for more
information on this feature.
3-4-4 Prepare needle
1. Click the Needle & Puncture button to open the Needle & Puncture panel. Note that this panel opens
only if the puncture status is OK.
2. Select a pre-programmed needle or program a new needle as described in Chapter 11 Needle Setup
section 2 Needle Setup on page 104.
Note:
- Only FNA needles should be used with the vertical approach phantom.
- The needle used must be long enough to touch the bottom of the deepest hole of the vertical
approach phantom. A needle length of at least 94-95 mm is recommended when the phantom is
compressed to 50 mm.
3. Click the Send button.
Planned Maintenance QC Test 4 - Accuracy check before using new FNA/Hook-wire needles in vertical approach
4. Angle the tube arm to either of the parking positions, to ease insertion of the needle. Refer to
Chapter 10 Preparation for examinations, section 2-4-1 Angulation movement for stereotaxy on
page 96.
5. With the needle block closed, press and hold the Go to Target button on the Stereotactic
Positioner until the target position is reached; the light will now be green.
See Chapter 13 Stereotaxy Punctures section 4 Performing the puncture on page 151 for more
information.
Note:
When the needle path is open, all motorized movements are inhibited. Manual movement is
allowed with the needle path open.
3-4-5 Position the needle
1. Press the Go to Target button again to display the programmed needle. Check that the
correct needle has been chosen.
2. Install the appropriate needle guides, and insert the needle into the
needle guides. Needle hub
3. The needle should move easily into the hole in the phantom above
the selected target, and the tip of the needle should just contact the Needle guide
lead plug with the base of the needle hub resting on the needle
guide. Phantom
- If not, adjust the X and Y positions of the needle manually until it
is centered in the hole, and adjust the Z position until the needle Lead plug
tip just contacts the lead plug with the base of the hub resting on
the needle guide.
- Read the resulting offsets between the computed target position and the actual needle position
on the Display Unit.
• If the magnitude of any of these offsets exceeds 1.0 mm, repeat the test.
If it is not possible to reduce the magnitude of all offsets below 1.0 mm, call your Field Engineer.
3-4-6 Repeat for all test points
• Carry out the procedures described above (sections 3-4-3 Define target on page 174, 3-4-4 Prepare
needle on page 174 and 3-4-5 Position the needle on page 175) for each of the remaining holes in
the phantom.
• To be sure that the Stereotactic Positioner is operating correctly, the check should be made for at
least three of the holes in the phantom, using the same reference exposure (section 3-4-2 Exposure
on page 173).
Note:
Do not move the phantom at any time during the test. The procedure must be repeated if there has
been any movement of the phantom.
Frequency:
On each day that stereotactic exams using the lateral approach are to be performed, before the first
exam of the day. If no stereotactic exams are carried out, the test should be carried out at least once a
month to ensure continued accuracy. See Chapter 7 Procedure Steering Guide section 3 Pre-
examination check on page 45.
Objective:
To ensure localization accuracy in the lateral approach
Equipment:
Pin phantom L19 and calibration needle L20.
Action Limit:
The magnitude of the offset for the X or Y or Z axis must not exceed ±1.0 mm with the calibration needle.
Record completion of the test in section QC Test Chart 5 - Localization Accuracy: Lateral Approach on
page 190.
3-5 Procedure
Note:
Before beginning the procedure, the Lateral Approach Tool must have
been declared as described in Chapter 11 Needle Setup section 2-
2 To create a new needle (with or without a biopsy device) on
page 105.
The procedure used for the test with the phantom is similar to a normal
stereotactic patient procedure using the lateral arm.
Pin phantom
3-5-1 Preparation
1. Prepare the Senographe and install the Stereotactic Positioner as described in Chapter 6 Installation
of the Stereotactic Positioner, section 1 Installing the Stereotactic Positioner on page 43.
2. Install the lateral approach tool support and adaptor bar as described in Chapter 9 Installation for
Lateral Approach.
3. Open the Worklist window on the Senographe AWS and
verify that the selection for Type of Exam is Stereo. Browser
4. On the Worklist, select the “Phantom” patient or create it. Type of Exam Stereo Start Exam
If no “Phantom” patient has been defined for use with the
phantom accuracy test, select New Patient... in the
Worklist, and complete the new Medical Procedure Card as required, with “Phantom” as Patient
Name, before starting image acquisition. Refer to Chapter 12 Stereotaxy Image Acquisition Section
1 Starting the Stereotaxy application on page 125 for details.
5. Click the Start Exam button. The Stereotactic Viewer appears, with the Viewer area at the right of the
screen remaining black until an exposure is taken.
6. Position the phantom L19 under the compression paddle:
- With the tip of the pin within the area of the biopsy window as defined by the four engraved
crosses on the breast support.
- With the pin perpendicular to the chest wall, pointing toward the chest wall side.
The upper face of the phantom indicates the Z position of the target (10, 20, 30, or 40 mm).
7. Use the compression knobs on either side of the Stereotactic Positioner to apply just enough
compression to hold the phantom in place.
Note:
- When the stereotaxy compression paddle leaves the park (high) position, the up/down and
rotation movements of the Senographe Positioner Arm are inhibited.
- After the phantom has been placed in position and compressed, it must not be moved until the
test has been completed
8. Press the Rotation/Angulation button on the Stereotactic Positioner to switch from rotation mode to
angulation mode. The gantry readout now displays the stereo angle, not the gantry angle.
9. If stereo positioner is not in parking position , turn the X, Y, and Z knobs until the parking position is
reached (green light will blink), to move the needle holder out of the field of view.
3-5-2 Exposure
1. Select the manual exposure mode with the following parameters: Mo-Mo, 24kV, 4mAs. This type of
phantom does not give a good image in AOP mode.
2. Make the scout exposure. The scout image appears in the Viewer area.
- Verify that the tip of the Phantom is centered within the four engraved crosses.
- Verify that both inverted T reference marks are visible and not hidden by the phantom. If not,
reposition the phantom and make the scout exposure again.
3. Adjust the window width and level if necessary for best viewing level.
4. Move the cursor onto the center of one of the two Reference Markers . Press and hold the left
mouse button until the cursor icon becomes a pencil, then drag the Reference Marker to place it
exactly on top of one of the images of the inverted T reference marks, as described in Chapter 13
Stereotaxy Punctures section 1 Defining punctures on page 135. Repeat the procedure to drag the
other Reference Marker and place it on top of the image of the other inverted T reference mark.
Use the invert function (Chapter 12 Stereotaxy Image Acquisition, section 2-4 Control Panels on
page 130) to confirm the positioning.
5. Make the two stereo exposures, with the tube angled to +15°, then to -15°.
See Chapter 10 Preparation for examinations section 2 Arm Positioning on page 94 for a description
of tube arm movements.
See Chapter 12 Stereotaxy Image Acquisition section 3 Making Stereotaxy exposures on page 132
for an explanation of the exposure sequence.
6. The stereo images are displayed in the monitor Viewer area in Stereo mode (the images in lower half
of the viewing area are zoomed images of the top half; see Chapter 12 Stereotaxy Image Acquisition
section 2 Stereotaxy viewer on page 126.
7. Use the right mouse button to click on the area of interest on each of the top images, to center the
lower zoomed images on the area of interest.
3-5-3 Define puncture
1. Click (left mouse button) on the Stereotaxy selection button to open the Stereotaxy Panel.
2. Click the New Primary button to define a new puncture set, then select the puncture marker and drag
it over the exact area of interest on each of the zoomed images. Refer to Chapter 13 Stereotaxy
Punctures section 1-4 Placing the Primary Puncture Marker on page 136.
Note:
You must accurately select the same point seen on the two images. Inaccuracies may lead to
deviations when the needle is being moved to the phantom position.
3. Click the Check on Scout button to display the scout image in the top half of the screen. This allows
you to check that the correct puncture site has been selected.
See Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on page 127.
3-5-4 Install and position the needle
1. Click the Needle & Puncture button to open the Needle & Puncture panel. Note that this panel opens
only if the puncture status is OK
2. Select the pre-programmed lateral approach tool.
3. Click the Send button.
4. Angle the tube arm to either of the parking positions, to ease insertion of the needle.
5. Go to target manually by turning the X, Y, Z knobs on the
Stereotactic Positioner until the following position is reached:
X= 0.0 mm
Y= 0.0 mm
Z= 0.0 mm
This movement must be made manually because motorized
movement is not allowed with the lateral approach tool installed. X = 0.0 mm
Y = 0.0 mm
Z = 0.0 mm
6. Slide the cursor L5 up to the stop on the lateral adaptor bar L4 (position 0), and slide the needle
guide holder L18 onto the bar until it touches the cursor. Lock the cursor and holder in position by
tightening their securing screws.
7. Place the calibration needle L20 through the holes of the needle guide holder L18 (no needle guide is
required).
3-5-5 Test results
1. Check that the tip of the phantom pin touches the tip of the calibration
needle.
2. If not, use the X and Y knobs to move the needle until contact is
established between the two tips.
3. To be correct, the X, Y coordinates must be equal to or less than
±1.0 mm
• If X or Y are not within the values specified above, repeat the test. If the problem persists, call your
Field Engineer.
• If the tip of the phantom does not touch the tip of the calibration needle because of displacement on
the Z axis, refer to Chapter 11 Needle Setup and perform the procedures described in section 3-
6 Lateral Approach adaptor on page 112. Repeat the test. If the problem persists, call your Field
Engineer.
Planned Maintenance QC Test 6 - Accuracy check before using new CB or VAD needles in lateral approach
Frequency:
Whenever a new CB or VAD needle is defined or a new type of needle is to be used.
Note:
For needles used without a biopsy device (FNA or Hook), use the procedure described in section
QC Test 4 - Accuracy check before using new FNA/Hook-wire needles in vertical approach on
page 172.
If no stereotactic exams are carried out, the test should be carried out at least once a month to ensure
continued accuracy. See Chapter 7 Procedure Steering Guide section 3 Pre-examination check on
page 45.
Objective:
To ensure localization accuracy in the lateral approach.
Equipment:
Pin phantom L19, biopsy device and the new CB or VAD needle to be used.
Action Limit:
With any biopsy device, the tip of the pin of the test phantom must be at the center of the notch of the
biopsy needle.
Record completion of the test in section QC Test Chart 6 - Accuracy check before using new CB or VAD
needles in lateral approach on page 191.
3-6 Procedure
Note:
Before beginning the procedure, the Lateral Approach Tool must have
been declared as described in Chapter 11 Needle Setup section 2-
2 To create a new needle (with or without a biopsy device) on
page 105.
The procedure used for the test with the phantom is similar to a normal
stereotactic patient procedure using the lateral arm.
Pin phantom
CAUTION
It is NOT recommended to use the firing mechanism of a biopsy device when not introduced in a
breast or a soft phantom; it could result in injury to the operator.
This procedure tests the accuracy of the system when using the lateral arm with a VAD or similar biopsy
device. The principle is to introduce the needle to a position offset by approximately 10 mm in the Y
direction from the tip of the Phantom pin, and to carefully move the needle toward the 0, 0, 0 position,
using the Y knob only.
3-6-1 Preparation
1. Prepare the Senographe and install the Stereotactic Positioner as described in Chapter 6 Installation
of the Stereotactic Positioner, section 1 Installing the Stereotactic Positioner on page 43.
2. Install the lateral approach tool support and adaptor bar as described in Chapter 9 Installation for
Lateral Approach.
Planned Maintenance QC Test 6 - Accuracy check before using new CB or VAD needles in lateral approach
3. If required, install the mounting bracket onto the biopsy device as described in Chapter 9 Installation
for Lateral Approach section 4 Fitting the lateral approach biopsy device on page 73
Planned Maintenance QC Test 6 - Accuracy check before using new CB or VAD needles in lateral approach
Planned Maintenance QC Test 6 - Accuracy check before using new CB or VAD needles in lateral approach
5. Make the two stereo exposures, with the tube angled to +15°, then to -15°.
See Chapter 10 Preparation for examinations section 2 Arm Positioning on page 94 for a description
of tube arm movements.
See Chapter 12 Stereotaxy Image Acquisition section 3 Making Stereotaxy exposures on page 132
for an explanation of the exposure sequence.
6. The stereo images are displayed in the monitor Viewer area in Stereo mode (the images in lower half
of the viewing area are zoomed images of the top half; see Chapter 12 Stereotaxy Image Acquisition
section 2 Stereotaxy viewer on page 126).
7. Use the right mouse button to click on the area of interest on each of the top images, to center the
lower zoomed images on the area of interest.
3-6-3 Define puncture
1. Click (left mouse button) on the Stereotaxy selection button to open the Stereotaxy Panel.
2. Click the New Primary button to define a new puncture set, then select the puncture marker and drag
it over the exact area of interest on each of the zoomed images. Refer to Chapter 13 Stereotaxy
Punctures section 1-4 Placing the Primary Puncture Marker on page 136.
Note:
You must accurately select the same point seen on the two images. Inaccuracies may lead to
deviations when the needle is being moved to the phantom position.
3. Click the Check on Scout button to display the scout image in the top half of the screen. This allows
you to check that the correct puncture site has been selected.
See Chapter 12 Stereotaxy Image Acquisition section 2-3 Viewing Area on page 127.
3-6-4 Install and position the needle
1. Click the Needle & Puncture button to open the Needle & Puncture panel. Note that this panel opens
only if the puncture status is OK
2. Select the pre-programmed lateral approach tool.
3. Click the Send button.
4. Angle the tube arm to either of the parking positions, to ease insertion of the needle.
5. Go to target manually by turning the X, Y, Z knobs on the
Stereotactic Positioner until the following position is reached:
X= 0.0 mm
Y= -10.0 mm
Z= 0.0 mm
This movement must be made manually because motorized
movement is not allowed with the lateral arm installed.
X = 0.0 mm
Y = -10.0 mm
Z = 0.0 mm
6. Install the appropriate needle guide holders and needle guides as described in Chapter 9 Installation
for Lateral Approach.
7. With the biopsy device in a post-fire condition (the biopsy device has already been fired), install it as
described in Chapter 9 Installation for Lateral Approach.
8. Turn the Y knob carefully to reach the 0, 0, 0 position.
Planned Maintenance QC Test 6 - Accuracy check before using new CB or VAD needles in lateral approach
• If the tip of the phantom is not within the notch because of displacement on the Z axis, refer to
Chapter 11 Needle Setup and perform the procedures described in section 3-6 Lateral Approach
adaptor on page 112. Repeat the test. If the problem persists, call your Field Engineer.
3-6-6 Repeat for other test points
• Decompress the pin phantom.
• Reposition it on the breast support, so as to have the tip of the rod in a different position in the
puncture volume. Turn it to use a different value of Z, and move it to a different location (changing X
and Y). Press Go to Parking till the parking position is reached.
• Carry out the procedures described above (sections 3-6-3 Define puncture on page 183, and 3-6-
4 Install and position the needle on page 183) for at least two other positions of the target.
Test results
Facility: Room:
Year Month
Date 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Initials
Needle Path Block Test
Year Month
Date 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Initials
Needle Path Block Test
Remarks:
Observations:
As read from the Stereotactic Positioner Display Unit, x is the offset between the final needle position
and the computed target position in the X direction, y is the offset in the Y direction, and z is the offset
in the Z direction.
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Planned Maintenance QC Test Chart 3 - Accuracy check before using new CB or VAD needles in vertical approach
Observations:
As read from the Stereotactic Positioner Display Unit, x is the offset between the final needle position
and the computed target position in the X direction, y is the offset in the Y direction, and z is the offset
in the Z direction.
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Planned Maintenance QC Test Chart 4 - Accuracy check before using new FNA needles in vertical approach
QC Test Chart 4 - Accuracy check before using new FNA needles in vertical
approach
Observations:
As read from the Stereotactic Positioner Display Unit, x is the offset between the final needle position
and the computed target position in the X direction, y is the offset in the Y direction, and z is the offset
in the Z direction.
Date
Vertical approach phantom hole number Vertical approach phantom hole number
Offset 1 2 3 4 5 1 2 3 4 5
Δx (mm)
Δy (mm)
Δz (mm)
Date
Vertical approach phantom hole number Vertical approach phantom hole number
Offset 1 2 3 4 5 1 2 3 4 5
Δx (mm)
Δy (mm)
Δz (mm)
Date
Vertical approach phantom hole number Vertical approach phantom hole number
Offset 1 2 3 4 5 1 2 3 4 5
Δx (mm)
Δy (mm)
Δz (mm)
Use this figure to identify the plugs when testing Stereotactic
Localization Accuracy with an FNA needle. The view is from
the top of the test phantom.
5
4
2
1
Observations:
As read from the Stereotactic Positioner Display Unit, x is the offset between the final needle position
and the computed target position in the X direction, y is the offset in the Y direction, and z is the offset
in the Z direction.
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Planned Maintenance QC Test Chart 6 - Accuracy check before using new CB or VAD needles in lateral approach
QC Test Chart 6 - Accuracy check before using new CB or VAD needles in lateral
approach
Observations:
As read from the Stereotactic Positioner Display Unit, x is the offset between the final needle position
and the computed target position in the X direction, y is the offset in the Y direction, and z is the offset
in the Z direction.
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Date
Pin Phantom Position Pin Phantom Position Pin Phantom Position
Offset 1 2 3 1 2 3 1 2 3
Δx (mm)
Δy (mm)
Δz (mm)
Planned Maintenance QC Test Chart 6 - Accuracy check before using new CB or VAD needles in lateral approach
Operator messages
1 Introduction
Any operational anomalies found by the system during Senographe operation are signalled by Error
Messages. These may be displayed on the X-ray Console, in a pop-up window on the AWS monitor, or
on the Gantry Readout. They may be simple information messages requiring no action, or they may
indicate error or fault conditions requiring action from the operator and/or GE Service engineers.If you
need to call your Field Service representative about an Operator message, you should quote the entire
message or messages, complete with any identification characters.
• The following tables list error messages relating to stereotactic operation, with explanations and
recommended actions.
- Table 1 Operator Messages displayed on the X-ray Console on page 194
- Table 2 Operator Messages displayed on the AWS monitor on page 196
- Table 3 Patient position status change messages displayed on the AWS on page 199
- Table 4 Operator Messages Displayed on the Readout Panel at the foot of the Gantry on
page 201
• Only messages relating to stereotactic operation are listed here. If you see an error message which
is not listed here, and for more information about error messages, refer to the Error Messages
chapter in the Senographe Essential Acquisition System Operator manual.
2 Operator messages
• Messages displayed on the X-ray Console monitor begin with a three-character identification code,
followed by the message itself.
The first letter of each code indicates the type of message, for example:
- Exx: Error messages indicating that a non-recoverable error has been detected. The system must
be restarted by switching OFF/ON from the X-ray Console.
- Ixx: Information or warning messages indicating operational errors due to temporary fault
conditions or attempts to use non-permitted configurations. It is usually possible to continue use
of the system without restarting.
- Sxx: System exposure messages indicating conditions in which an exposure may not be possible
or has not been correctly terminated.
• Four main types of messages are displayed on the AWS monitor;
- Information messages relating to power on/off events.
- Information messages relating to boot/reset events.
- Information messages relating to disk or memory problems.
- Information messages relating to the stereotaxy interface.
• Four main types of message are displayed on the Readout panel at the foot of the Gantry:
- Information messages with no prefix: Displayed to acknowledge operator actions such as the
selection of a new field of view.
- Advisory messages with a prefix Axx: Displayed to warn the operator when a requested operation
is not possible.
- Halt messages with a prefix Hxx: Displayed to warn the operator that a non-recoverable error has
been detected. The system must be restarted by switching ON/OFF from the X-ray Console.
- Warning messages with a prefix Wxx: Displayed to warn the operator when an error has been
detected that does not prevent continued operation.
Operator messages
Operator messages
Operator messages
Operator messages
Operator messages
Operator messages
Operator messages
Operator messages
Table 4 - Operator Messages Displayed on the Readout Panel at the foot of the Gantry
Operator messages
Operator messages
Operator messages
Operator messages
Operator messages
Specifications
Chapter 18 Specifications
1 Introduction
Specifications are given here only for functions and items specific to Senographe Essential Stereotaxy.
For specifications applicable to the Senographe Essential, refer to the Senographe Essential Acquisition
System Operator manual.
Specifications
3 Physical specification
3-1 Dimensions and weight
Component Dimensions in mm (inches) Weight
Depth Width Height kg (lbs)
Stereotactic Positioner 365 (14.4) 390 (15.4) 355 (14) 10.5 (23)
4 Trademark information
• GE and the GE Monogram are trademarks of the General Electric Company.
• Bard, Magnum and Vacora are trademarks and/or registered trademarks of C.R. Bard, Inc. or an
affiliate.
• EnCor is a registered trademark of SenoRx-Bard, Inc.
• ATEC and Suros are trademarks or registered trademarks of Suros Surgical Systems, Inc., a Hologic
Company.
• Mammotome is a registered trademark of Devicor, Inc.
• All other trademarks, service marks, logos, company names and product names are the property of
their respective owners.