Mythic 18 V3.3.7 Release Notes Descmythic18-V3.3.7: 1 Objet / Object
Mythic 18 V3.3.7 Release Notes Descmythic18-V3.3.7: 1 Objet / Object
1 OBJET / OBJECT
This document aims to list software modifications done by V3.3.7 version.
These modifications are in regard with V3.3.5 and software requirements.
6 REFERENCES / REFERENCES
SRS_Mythic18-32
M18-ReleaseNoteFluid-CycV203-008
8 CONTRAINTES / CONSTRAINTS
Downgrade to previous version: NO
This version has been validated as V3.3.7-014
10 FINALISATIONS / COMPLETED
New set of fluidics cycles V2.03-008 to improve cleaner consumption accuracy.
The reagent almost empty is raised when remaining reagent is not enough to run 30 analysis (was 10).
11 AMELIORATIONS / IMPROVEMENTS
[10708] If an error is ever raised during a Shutdown cycle, the instrument remains ON. The Power Off is only
initiated if the Shutdown cycle is performed without any error.
[11061] Add Emergency Stop if Cleaner Line is detected "unprimed". Message: " CYCLE: PRESSURE DEFAULT"
followed by "CYCLE: VALVE 11 FAILED", more details in the fluidic release notes: M18-ReleaseNoteFluid-CycV203-
008.
[11320] The park syringe position to match old and new piston design.
12 CORRECTIONS / CORRECTIONS
[10467] When downloading a printer driver, the combo-box printers list remains now correct including the lastly
download name drive. DPU-S445 is no more lost. Moreover, prefix -3 and -4 have been deleted and the printer
driver access button is accordingly labeled PRINTER DRIVER.
[10589] Onto GUI, print reports have been corrected and replaced with the expected Mythic texts.