1 Phasics creates the

revolution in infrared optics


quality control with the Kaleo
Bench-IR series. These inno-
vative benches offer both MTF
and wavefront quality (aber-
rations, WFE) in one single
fast acquisition. They cover
the full IR range from 1.2 µm
to 14 µm.
Their unique patented techno-

This document is not contractual.


logy of unrivalled high resolu-
tion and sensitivity ensures an
accurate characterization
while simplifying set-up and
alignment.

"MTF & WAVEFRONT ERROR WITH ONE BENCH"


6 GET THE MTF... λ = 3.39 µm λ = 10.6 µm
• Along any direction − 2.650
3.201 − 3.201
− 3.000 − 3.000
• For any pupil size − 2.250
− 2.000
− 2.500 − 2.500

• On and off-axis − 2.000


− 1.500
− 2.000

− 1.500 − 1.500
• Up to cut-off frequency − 1.000
− 1.000 − 1.000

• With various focusing methods − 0.500


− 0.500 − 0.500
RMS=0.545 µm RMS=0.541 µm
− 0.000 − 0.000

... AND MORE SA = 0.513 µm RMS (λ/6.61) SA = 0.498 µm RMS (λ/21.28)


100 100

• EFL, F#, NA 90
Horizontal
90
Horizontal
80 Vertical 80 Vertical

• Aberrations: Zernike, Seidel 70


Diffraction Limit
70
Diffraction Limit

60 60

• Real time filtering of phase map


Y (%)

Y (%)

50 50
40 40
(Zernike, Kernel…) 30 30
20 20

• Chromatic aberrations 10 10
0 0

• TF MTF for any pupil size 0 20 40 60


X (cc/mm)
80 100 120 0 10 20 30
X (cc/mm)
40 50

• Comparison to design
Chromatic behaviour measured with one unique sensor on a ZnSe
lens of 25.4 mm diameter. The focal spot shifts 560 µm due to
refractive index dispersion.
BENCH FOR INFRARED OPTICS TESTING

Collimated source Optical component PHASICS


quality of λ/20 PV under test wavefront sensor

Simple direct set up with Phasics unique sensor

Applies to lens, objective, zoom, strongly aberrated subassembly.

4 UNIQUE TECHNOLOGY 1 FULLY INTEGRATED BENCH


• High resolution & high • Powerful software with optional modules such as Design
sensitivity for robust calcu- Pro for rigorous comparison to design files and RetroPro for
lations and small defects characterization in the exit pupil
detection
• Closed structure & motorization for safety compliance
• High dynamics for testing
Flexible Chromatic high
& matching strongly aberrated characterization
High resolution
sensitivite
characterization
optics subassemblies at 3.39
at 10.6 µm
characterization
& 10.6 µm in MWIR region
• True achromaticity through LWIR
DWIR IR-MCT
the full IR range for chromatic
HeNe laser Black body
response testing Source and/or CO2 CO2 laser with spectral
laser filters
• Direct measurement up
to F/1 Entrance pupil Up to 100 mm
diameter
... for a simple set-up
Maximum
- No relay lens aperture F/1
- Easy to align
Detector
- Cost effective technology
Micro bolometer Micro bolometer Cooled MCT

... for an easy interpretation


Wavelength 3 - 5 µm & 8 - 14 µm 1.2 - 5.5 µm
- Measurement in working range 8 - 14 µm
conditions
Phase spatial
- Rigorous calculation resolution
140 µm 100 µm 60 µm

- Results in any plane Phase


96 x 72 160 x 120 160 x 128
(best focus, paraxial) sampling

Sensibility 25 nm RMS 25 nm RMS 3 nm RMS