Professional Documents
Culture Documents
AXIOM Iconos R200 Spotfilm Device
AXIOM Iconos R200 Spotfilm Device
AXIOM Iconos R200 Spotfilm Device
o
h
s
a
V
Replacement of Parts
AXIOM Iconos R200
Spotfilm Device
© Siemens AG 1999
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Lubrication Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Adjustment of Bearings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Securing of Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remarks Regarding Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Product-specific Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Test Equipment / Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Maximum Torque Values (Table). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replacement and Adjustment of Mechanical Parts . . . . . . . . . . . . . . . . . . . . . . . . . . 8
NOTE All tools, test equipment and aids, with the exception of the stan-
dard installation tool kit” are listed and specified in ARTD (Part 3).
Remarks 0
Safety Information 0
NOTE When carrying out the work steps and checks, the product-spe-
cific safety information contained in the documents as well as the
general safety information contained in ARTD must be observed.
Lubrication Locations 0
Prior to beginning the particular adjustments, the condition of all accessible chain and
geared drives, guide and roller bearings must be checked, and if necessary, lubricated
using Longtime PD2 (unless otherwise specified).
Adjustment of Bearings 0
Support and roller bearings are adjusted so that the roller bearing not under load can just
still be turned by hand.
Securing of Screws 0
All screws removed for adjustment purposes must be resecured to their original condition.
If securing a screw with Loctite is necessary, an exact reference is made to this in the text.
When working with Loctite, it is absolutely necessary to make sure that this securing
material does not penetrate into any bearings.
A bearing that would jam as a result of this would only become apparent days later.
Loctite can be softened with heat (hair dryer, etc.).
• SSW ICONOS
• Digital multimeter (e.g. Fluke 8060A) 97 02 101 RE999
• cm tape measure
• Slide calipers
• Spring balance, 100 N 44 29 122 MH029
• Longtime PD2 (20 gr.) 34 91 271
• Loctite 221 20 48 874
• P4 silicon paste 20 49 914
• Allen wrench attachment for flat-head screws 81 59 980 RE999
Check checking mountings, use the appropriate torque values for hardness rating 8.8.
A tolerance of ± 10% is permissible for torque values.
Access to the spotfilm device is described in the chapter, “Creating Access to the Spotfilm
Device”.
Not every detail is described in this document, down to the last screw (e.g. on the drive),
because these details are obvious while performing the work.
Concluding function tests, as well as movements of drives into certain positions in excep-
tional cases are described in the SSW. Only a reference to the corresponding screen is
made in this document.
All mechanical adjustments are described.
4 5
2 1
7
6
8
10
4
2
3
2
1
2
2
1
2
• Open the cable ties on the Iontomat cable (9/Fig. 1 / p. 9) and open up the cable loop.
• Remove the middle wall (10/Fig. 1 / p. 9) on the operator’s side and lay the Iontomat
cable until the middle wall is flush with the collimator plate.
Zentrierschablone (C)
Center matrix (C)
Zentrierschablone (B)
Center matrix (B)
Zentrierschablone (A)
Center matrix (A)
Zentrierschablone (D)
Center matrix (D)
D
Serviceabdeckung
Service cover
Zentrierschablone
Center matrix
305 mm
Film
Serviceabdeckung
Service cover
382 mm
Fig. 5:
If the centering template for the spotfilm device is not included in the shipment or is no
longer available, a new template can be ordered. For the order number, see the Spare
Parts Catalogue.
- Up to this time, there were 4 different versions (A - D) of the centering template
(Fig. 4 / p. 14).
- Only Version (D) is currently being shipped, even in a replacement part situation
(Fig. 4 / p. 14).
• The centering template can also be used in systems without the hole. The dimensions
for placing on the template can be found in (Fig. 5 / p. 15).
• Drilling a hole in the spotfilm device should not be done. Drill shavings can not easily be
removed. If necessary, mark the position on the spotfilm device with a scoring tool.
NOTE The core of the new middle wall is made of aluminum and no
longer of wood. Its stability is now significantly better.
NOTE Hand creme on the cassettes have the same effect as greased
belts. With smooth cassettes, this can cause transport problems.
• The I.I. cover may not be bowed in either the table transverse or in the table longitudi-
nal direction.
• With large, sharp-cornered cassettes, the bow in the cassette may not be more than 1.5
mm, otherwise it can cause malfunctions.
• The cassette must be higher than the mechanism in the area of the tension arm mecha-
nism.
NOTE With the new I.I. cover panels, there are two raised points (seam
beads). These reduce the drag surface to the cassette (reduction in
friction).
In principle, the seam beads make the cassette 1-2 mm higher. If
this is not sufficient, the height of the cassette can be adjusted by
placing strips underneath it.
M3
R3
• Adjusting the force of the tension arms can only be done with the speed.
For this, use the SSW:
- Adjustment/ Calibration/ Spotfilm Device/ Cassette tensioning: Open speed
Adjustment/ Calibration/ Spotfilm Device/ Cassette tensioning: Close speed
for the adjustment.
2
3
1
4
2
5
• Remove the 3 countersunk Allen screws (2/Fig. 8 / p. 22) on the motor assembly
(3/Fig. 8 / p. 22) and loosen the adjustment screw (1/Fig. 8 / p. 22).
• Take the belt out from the wall side.
• Insert the new belt in the reverse order of the above instructions; to pull the belt through,
it may be necessary to use a pull wire.
• Perform “Adjusting Belt Tension” (Adjusting Belt Tension / p. 25).
• Loosen the 2 Allen screws (4/Fig. 8 / p. 22) on the return pulley (5/Fig. 8 / p. 22).
• Remove the return pulley towards the operator’s side.
• Place the new belt on in the same way as the old one; to pull it through it may be neces-
sary to use a pull wire.
• Perform “Adjusting Belt Tension” (Adjusting Belt Tension / p. 25).
25 - 30 mm
F = 20 - 22 N
F = 20 - 22 N
20 - 25 mm
• Remove the cover over the loading slot (see “Access to the Spotfilm Device”)
• Replace the photo-sensitive cells in the loading slot cover; a mechanical adjustment is
not possible.
After completing the replacement, using the SSW:
• SSW: Diagnostic / tests/ Movement tests/ Transverse cassette drive in loading
position direction and
• SSW: Diagnostic / tests/ Movement tests/ Transverse cassette drive in park posi-
tion direction
the position of the park photo-sensitive cell should be checked; this is important for
positioning of the exposure position and of cassette detection.
2 2
1
4 3
2
6
2
5 1
10
10
9
8
10
10
• Loosen the nut (1/Fig. 11 / p. 28) of the motor assembly (2/Fig. 11 / p. 28); when this is
done, the belt tension is relaxed.
• Remove the hygienic film.
• Remove the countersunk Allen-head screws (7/Fig. 11 / p. 28) on the operator’s side
return pulley (8/Fig. 11 / p. 28), remove the return pulley.
• If necessary, move the tabletop transversely to the operator’s side.
• Take the belt off the drive wheel (3/Fig. 11 / p. 28).
• The two diaphragm plates (9/Fig. 11 / p. 28) can be removed carefully from the opera-
tor’s side.
• The diaphragm plates (9/Fig. 11 / p. 28) must be placed down flat.
5 3
X Z
40 40
1 A B Y 2
10 - 12 mm
F = 10 - 12 N
Prerequisite
The symmetry can only be checked using an exposure made under specific conditions.
Expose a film as described in the SSW.
• Adjustment/ Calibration/ Spotfilm Device/ Diaphragm plates: Basic setting
Explanation:
1. The diaphragm plate opening results from the distance of the diaphragm plate (wall
side) to the photo-sensitive cell and an offset that can be influenced by the SSW.
2. The symmetry depends only on the position of the belt clamp (5/Fig. 12 / p. 29) on the
operator’s side diaphragm plate (A/Fig. 12 / p. 29).
Procedure:
• Remove the diaphragm plates.
• Mark the position of the belt clamp (5/Fig. 12 / p. 29) on the operator’s side diaphragm
plate (A/Fig. 12 / p. 29), move the belt clamp accordingly and secure it in place.
Perform the function check again:
• SSW: Adjustment/ Calibration/ Spotfilm Device/ Diaphragm plates: Basic setting
NOTE For remarks regarding the adjustment as well as the function test,
see Help in the SSW.
6 6 4
7 3
1
2
F= 10 - 12 mm
NOTE For remarks regarding the adjustment, as well as for the function
test, see Help in the SSW.
• After each exposures, check the mAs post display; approximately the same mAs
should be displayed for all exposures.
NOTE With the new I.I. cover panels, there are 2 raised locations (strips).
These minimize the contact surface to the cassette (lessening of
friction). In addition, the strips in principle raise the cassette 1-2
mm higher. If this is not sufficient, the height of the cassette can
be adapted by placing strips (Y/Fig. 18 / p. 39) underneath it.
• Remover all cover panels as shown (Access to the Spotfilm Device / p. 9).
Fig. 18:
• Remove the I.I. cover panel:
- To do this, remove the 4 countersunk screws (A/Fig. 18 / p. 39).
- The I.I. component group cover panel can now be removed towards the front.
Fig. 19:
The two parts, the metal frame and the plastic panel, of the I.I. cover panel must be cen-
tered when installed.
To assure this, perform the following work steps.
• Remove the 4 Allen screws (1/Fig. 19 / p. 40) on both sides.
• Loosen the 3 Allen screws (2/Fig. 19 / p. 40).
• Loosen the nuts (3/Fig. 19 / p. 40) on both sides.
• Carefully insert the cover panel.
- When doing this, lift the plastic cover on the wall side, otherwise the cover will jam
against the I.I. or the tension mechanism.
• Insert the 4 Allen screws (A/Fig. 18 / p. 39).
• Carefully move the metal frame to the stop on the wall side.
• Tighten the middle Allen screws (2/Fig. 19 / p. 40).
• Remove the 4 Allen screws (A/Fig. 18 / p. 39).
• Carefully pull out the I.I. cover panel approx. 10 cm.
- When this is done, the plastic panel and the metal frame may not move towards one
another.
• Carefully lift the plastic panel and remove the protective film (4/Fig. 19 / p. 40) on the
adhesive strips.
• Press the plastic panel lightly onto the adhesive strips.
• Tighten the 6 Allen screws (1, 2/Fig. 19 / p. 40).
• Carefully pull out the I.I. cover panel completely.
• Tighten the nuts (3/Fig. 19 / p. 40) on both sides.
• Carefully lift the plastic cover panel and remove the protective film (5/Fig. 19 / p. 40) on
the adhesive strips.