AXIOM Iconos R200 Spotfilm Device

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 42

lz

o
h
s
a
V

AXIOM Iconos R200


AX

Replacement of Parts
AXIOM Iconos R200

Spotfilm Device

Also for AXIOM Iconos R100


10021999
08602000
07094555
07030930
07030518
07030500

© Siemens AG 1999
The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.

Print No.: AXD3-340.841.02.05.02 English


Replaces: AXD3-340.841.02.04.02 Doc. Gen. Date: 04.02
2 Revision / Disclaimer
1Revision / Disclaimer

Document revision level


The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service pro-
cedures.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 2 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Table of Contents 3
0Table of Contents

1 _______ Requirements / Remarks__________________________________________ 5

General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Lubrication Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Adjustment of Bearings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Securing of Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remarks Regarding Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Product-specific Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Test Equipment / Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Maximum Torque Values (Table). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Replacement and Adjustment of Mechanical Parts . . . . . . . . . . . . . . . . . . . . . . . . . . 8

2 _______ Access to the Spotfilm Device _____________________________________ 9

Spotfilm Device Cover Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9


Removing the Hygienic Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Load Slot Cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Spotfilm Device Electronics Box Cover Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Middle Wall . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Centering Template . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

3 _______ General Remarks Regarding Cassette Transport _____________________ 16

Cassette Transport Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

4 _______ Tensioning the Cassette _________________________________________ 18

Replacing the Drive or Potentiometer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18


Adjusting the Tension Potentiometer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Adjusting the Tension Force . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Adjusting the Brake . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

5 _______ Transverse Cassette Drive _______________________________________ 22

Replacing the Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22


Replacing the Transport Belt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Replacing the Auxiliary Belt . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Adjusting Belt Tension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Replacing the Loading Slot Photo-sensitive Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Replacing the Park Photo-sensitive Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

6 _______ Diaphragm Plates ______________________________________________ 28

Replacing the Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28


Replacing the Belt or the Diaphragm Plates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Adjusting the Belt Tension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

Siemens AG AXD3-340.841.02.05.02 Page 3 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
4 Table of Contents

Adjusting the Diaphragm Symmetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

7 _______ Grid __________________________________________________________ 33

Replacing and Adjusting the Grid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33


Adjusting Belt Tension. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Adjusting the Potentiometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35

8 _______ IONTOMAT ____________________________________________________ 36

Iontomat Chamber Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36

9 _______ I.I. Cover Panel _________________________________________________ 39

Replacing the I.I. Cover Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

10 ______ Changes to Previous Version _____________________________________ 42

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 4 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Requirements / Remarks 5
1-
General Remarks
1Requirements / Remarks

NOTE All tools, test equipment and aids, with the exception of the stan-
dard installation tool kit” are listed and specified in ARTD (Part 3).

Remarks 0

Damaged or worn parts may only be replaced with original parts.


If there is a malfunction, first check the mechanism of the drives.
Special attention should be paid to easy movement of all movable parts.
Jamming movement will cause a switch-off by the safety software.
Because of this, clean and lubricate rails and guide parts, check the chain and toothed
belt tension for the particular drive, check the function of the brakes, if needed clean them
of any wear residue and adjust the air gap.

Safety Information 0

CAUTION Health Hazard!


¹ Checks and adjustments that must be performed with
radiation switched on are labeled with the radiation warn-
ing symbol x . While performing adjustments labeled
this way, radiation protective clothing must be worn.

NOTE When carrying out the work steps and checks, the product-spe-
cific safety information contained in the documents as well as the
general safety information contained in ARTD must be observed.

Lubrication Locations 0

Prior to beginning the particular adjustments, the condition of all accessible chain and
geared drives, guide and roller bearings must be checked, and if necessary, lubricated
using Longtime PD2 (unless otherwise specified).

Adjustment of Bearings 0

Support and roller bearings are adjusted so that the roller bearing not under load can just
still be turned by hand.

Siemens AG AXD3-340.841.02.05.02 Page 5 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
6 Requirements / Remarks

Securing of Screws 0

All screws removed for adjustment purposes must be resecured to their original condition.
If securing a screw with Loctite is necessary, an exact reference is made to this in the text.
When working with Loctite, it is absolutely necessary to make sure that this securing
material does not penetrate into any bearings.
A bearing that would jam as a result of this would only become apparent days later.
Loctite can be softened with heat (hair dryer, etc.).

Remarks Regarding Adjustment 0

NOTE After adjusting or replacing parts, the function test described in


the SSW absolutely must be performed.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 6 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Requirements / Remarks 7

Product-specific Remarks 1.1

Test Equipment / Aids 0

• SSW ICONOS
• Digital multimeter (e.g. Fluke 8060A) 97 02 101 RE999
• cm tape measure
• Slide calipers
• Spring balance, 100 N 44 29 122 MH029
• Longtime PD2 (20 gr.) 34 91 271
• Loctite 221 20 48 874
• P4 silicon paste 20 49 914
• Allen wrench attachment for flat-head screws 81 59 980 RE999

Maximum Torque Values (Table) 0

Nominal thread Screw material Steel per DIN 267 Brass


diameter CuZn
alloy
Hardness rating 4.6 4.8 5.8 8.8 10.9 12.9
M3 0.5 0.67 0.83 1.3 1.9 2.2 0.62
M 3,5 0.76 1.0 1.3 2.0 2.8 3.4 0.95
M4 1.1 1.5 1.9 3.0 4.2 5.1 1.4
M5 2.2 3.0 3.7 6.0 8.4 10.1 2.8
M6 3.8 5.1 6.4 10.2 14.4 17.3 4.8
M8 9.5 12.6 15.8 25.3 35.6 42.7 11.9
M 10 18.7 24.8 31.1 49.8 70.0 84.0 23.3
M 12 32.9 43.8 54.8 87.6 123.3 147.9 41.1

Check checking mountings, use the appropriate torque values for hardness rating 8.8.
A tolerance of ± 10% is permissible for torque values.

Siemens AG AXD3-340.841.02.05.02 Page 7 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
8 Requirements / Remarks

Replacement and Adjustment of Mechanical Parts 0

Access to the spotfilm device is described in the chapter, “Creating Access to the Spotfilm
Device”.
Not every detail is described in this document, down to the last screw (e.g. on the drive),
because these details are obvious while performing the work.
Concluding function tests, as well as movements of drives into certain positions in excep-
tional cases are described in the SSW. Only a reference to the corresponding screen is
made in this document.
All mechanical adjustments are described.

Corrective Measures / Adjustments


Reference is made several times to the SSW in this document for corrective measures
and for adjustments.
For this, select the adjustments, etc. in the XCS SSW program under “Compo-
nents/ICONOS”.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 8 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Access to the Spotfilm Device 9
2-
Spotfilm Device Cover Panels
2Access to the Spotfilm Device

4 5

2 1
7

6
8

10

Fig. 1: Access to spotfilm device


• Move the unit into the 0 degree position.
• Remove the middle paneling screws (1/Fig. 1 / p. 9) from the spotfilm device cover
panel (2/Fig. 1 / p. 9).
- If necessary, move the column to the head end or foot end to do this.
• Move the column into the 0 degree position.
• Remove the paneling screws (3/Fig. 1 / p. 9) and the safety switch strike plate
(4/Fig. 1 / p. 9).
• Remove the paneling screws (5/Fig. 1 / p. 9) and the spotfilm device cover panel.
• Reinstall the safety switch strike plate (4/Fig. 1 / p. 9).
• Insert a 24 x 30 cm or 10 x 12 inch cassette.
• Select the 4-part subdivision.
• Disconnect a motor cable to the collimator plate drive.
• Remove the cassette.
• Move the unit into the +90 degree position.

Siemens AG AXD3-340.841.02.05.02 Page 9 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
10 Access to the Spotfilm Device

Removing the Hygienic Cover 2.1

Required only when working on the grid, collimator plates.....

4
2

3
2

1
2
2

1
2

Fig. 2: Hygienic cover


• On the spotfilm device, place the film (3/Fig. 2 / p. 10) on the plastic cover of the
radio-transparent window (4/Fig. 2 / p. 10) and affix it to the spotfilm device. If needed,
cut off any excess length at the head and foot ends approx. 5 mm from the edge (arrows
(Fig. 2 / p. 10)) with a sharp knife.
• Use the back of the knife or a blunt object to move along the edges (arrows
(Fig. 2 / p. 10))to press down the film.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 10 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Access to the Spotfilm Device 11

Load Slot Cover 2.2

• Remove 2 paneling screws (6/Fig. 1 / p. 9) on both the left and right.


• Remove the load slot cover (8/Fig. 1 / p. 9), if needed, snip off the cable ties on the
back. Place the load slot cover (8/Fig. 1 / p. 9) down on the spotfilm device carriage.

Siemens AG AXD3-340.841.02.05.02 Page 11 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
12 Access to the Spotfilm Device

Spotfilm Device Electronics Box Cover Panel 2.3

• Remove the cover over the spotfilm device electronics box.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 12 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Access to the Spotfilm Device 13

Middle Wall 2.4

• Open the cable ties on the Iontomat cable (9/Fig. 1 / p. 9) and open up the cable loop.
• Remove the middle wall (10/Fig. 1 / p. 9) on the operator’s side and lay the Iontomat
cable until the middle wall is flush with the collimator plate.

Fig. 3: Tension arms


When inserting the middle wall, it is absolutely necessary that the tension arms on
the side of the wall be lifted.
• If the spacer nipples have already fallen out on a spotfilm device (1/Fig. 3 / p. 13)
(scratches on the middle wall), reinstall them.
- Completely remove the middle wall.
- Insert the spacer nipples accordingly (1/Fig. 3 / p. 13).
- Lift the tension arms on the wall side and reinsert the middle wall.
• There are also 2 spacer nipples on the bottom of the tension arms (2/Fig. 3 / p. 13).
They can possibly be sheared off during operation or a service situation.
• After completing replacement or adjustment work, reinstall the cover panels in the
reverse of the instructions to remove them.

Siemens AG AXD3-340.841.02.05.02 Page 13 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
14 Access to the Spotfilm Device

Centering Template 2.5

Zentrierschablone (C)
Center matrix (C)

Zentrierschablone (B)
Center matrix (B)

Zentrierschablone (A)
Center matrix (A)

Zentrierschablone (D)
Center matrix (D)
D

Serviceabdeckung
Service cover

Fig. 4: Centering template for the test exposures

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 14 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Access to the Spotfilm Device 15

Zentrierschablone
Center matrix

305 mm
Film

Serviceabdeckung
Service cover
382 mm

Markierung auf dem Film


Marker on the Film

Fig. 5:
If the centering template for the spotfilm device is not included in the shipment or is no
longer available, a new template can be ordered. For the order number, see the Spare
Parts Catalogue.
- Up to this time, there were 4 different versions (A - D) of the centering template
(Fig. 4 / p. 14).
- Only Version (D) is currently being shipped, even in a replacement part situation
(Fig. 4 / p. 14).
• The centering template can also be used in systems without the hole. The dimensions
for placing on the template can be found in (Fig. 5 / p. 15).
• Drilling a hole in the spotfilm device should not be done. Drill shavings can not easily be
removed. If necessary, mark the position on the spotfilm device with a scoring tool.

Siemens AG AXD3-340.841.02.05.02 Page 15 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
16 General Remarks Regarding Cassette Transport
3-
Cassette Transport Problems
3General Remarks Regarding Cassette Transport

The following is required for reliable cassette transport:


• The tension of the “tension cassette” drive must be adjusted.
• The tension of the “cassette transport” belt must be adjusted.
• The plastic cover and the metal frame on the “I.I. cover” assembly must centered to
each other when installed.
• The middle wall may not be bowed; if needed, replace it with a new one.

NOTE The core of the new middle wall is made of aluminum and no
longer of wood. Its stability is now significantly better.

• No spacer nipples installed!

Fig. 6: Tension arms


- If the spacer nipples have already fallen out in a spotfilm device (1/Fig. 6 / p. 16)
(scratches on the middle wall), reinsert them.
- Completely remove the middle wall.
- Insert the spacer nipples accordingly (1/Fig. 6 / p. 16).
- Lift the tension arms on the wall side and reinsert the middle wall.
• There are also 2 spacer nipples on the bottom of the tension arms (2/Fig. 6 / p. 16).
These can be sheared off during operation or when there is a service situation.
• The transport belt and the belt that runs with it must be free of grease and may not be
damaged.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 16 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
General Remarks Regarding Cassette Transport 17

NOTE Hand creme on the cassettes have the same effect as greased
belts. With smooth cassettes, this can cause transport problems.

• The I.I. cover may not be bowed in either the table transverse or in the table longitudi-
nal direction.
• With large, sharp-cornered cassettes, the bow in the cassette may not be more than 1.5
mm, otherwise it can cause malfunctions.
• The cassette must be higher than the mechanism in the area of the tension arm mecha-
nism.

NOTE With the new I.I. cover panels, there are two raised points (seam
beads). These reduce the drag surface to the cassette (reduction in
friction).
In principle, the seam beads make the cassette 1-2 mm higher. If
this is not sufficient, the height of the cassette can be adjusted by
placing strips underneath it.

Siemens AG AXD3-340.841.02.05.02 Page 17 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
18 Tensioning the Cassette
4-
Replacing the Drive or Potentiometer
4Tensioning the Cassette

M3

R3

Fig. 7: Removing the M3 drive


• Remove all cover panels as shown in (Access to the Spotfilm Device / p. 9).
• Remove the I.I. cover panel:
- To do this, remove the 4 countersunk screws (1/Fig. 7 / p. 18) and remove the cover
panel.
The M3 cassette tension drive (2/Fig. 7 / p. 18) and the R3 potentiometer will now be
accessible.
• Using the SSW:
- SSW: Diagnostic/ Tests/ Movements with service buttons/ Spotfilm device:
Movements with service buttons, close the tension arms to approx. 35 cm.
• Loosen the pinion rack mount on the tension arms:
- First loosen the bottom tension arm, pay attention to any plate washers and hold the
tension arm, otherwise it will slide.
- Loosen the top tension arm in the same manner.
• Remove the 2 Allen screws, toothed washer and plate washer from the potentiometer
bracket.
• Remove the 4 Allen screws and toothed washers for the cassette tension drive.
• Remove the 4 countersunk Allen screws for the drive support (aluminum).
• Take both cables out of the top ceramic terminal and insulate them.
• Take both cables out of the bottom ceramic terminal.
• Take out the drive by turning and moving the drive support.
• Install the new drive in the same manner.
• Install the potentiometer bracket, but do not mesh the potentiometer with the pinion
rack.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 18 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Tensioning the Cassette 19

Adjusting the Tension Potentiometer 0

• SSW: Adjustment/ Calibration/ Spotfilm Device/ Tensioning the cassette/ Potenti-


ometer adjustment

NOTE For remarks regarding the adjustment, as well as the function


test, see Help in the SSW.

Siemens AG AXD3-340.841.02.05.02 Page 19 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
20 Tensioning the Cassette

Adjusting the Tension Force 4.1

• Adjusting the force of the tension arms can only be done with the speed.
For this, use the SSW:
- Adjustment/ Calibration/ Spotfilm Device/ Cassette tensioning: Open speed
Adjustment/ Calibration/ Spotfilm Device/ Cassette tensioning: Close speed
for the adjustment.

NOTE For remarks regarding the adjustment, as well as the function


test, see Help in the SSW.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 20 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Tensioning the Cassette 21

Adjusting the Brake 4.2

• Adjust the air gap to 0.3 - 0.5 mm.

Siemens AG AXD3-340.841.02.05.02 Page 21 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
22 Transverse Cassette Drive
5-
Replacing the Drive
5Transverse Cassette Drive

2
3
1

4
2
5

Fig. 8: Replacing the transverse cassette drive


• Remove all cover panels as shown in (Access to the Spotfilm Device / p. 9).
• Place a 24 x 30 cm cassette on the lower tension arm and using the SSW
- SSW: Diagnostic/ Tests/ Movements with service buttons/ Spotfilm device:
Movements with service buttons
close the tension arms to 30 cm cassette.
• Remove the cassette.
• Remove the adjustment screw (1/Fig. 8 / p. 22) with the counternut.
• Remove the 3 countersunk Allen screws (2/Fig. 8 / p. 22) on the motor assembly
(3/Fig. 8 / p. 22).
- When doing this, be careful of the threaded plates behind the motor assembly.
• Install the new drive in the same way.
• Perform “Adjusting Belt Tension” (Adjusting Belt Tension / p. 25).

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 22 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Transverse Cassette Drive 23

Replacing the Transport Belt 5.1

• Remove the 3 countersunk Allen screws (2/Fig. 8 / p. 22) on the motor assembly
(3/Fig. 8 / p. 22) and loosen the adjustment screw (1/Fig. 8 / p. 22).
• Take the belt out from the wall side.
• Insert the new belt in the reverse order of the above instructions; to pull the belt through,
it may be necessary to use a pull wire.
• Perform “Adjusting Belt Tension” (Adjusting Belt Tension / p. 25).

Siemens AG AXD3-340.841.02.05.02 Page 23 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
24 Transverse Cassette Drive

Replacing the Auxiliary Belt 5.2

• Loosen the 2 Allen screws (4/Fig. 8 / p. 22) on the return pulley (5/Fig. 8 / p. 22).
• Remove the return pulley towards the operator’s side.
• Place the new belt on in the same way as the old one; to pull it through it may be neces-
sary to use a pull wire.
• Perform “Adjusting Belt Tension” (Adjusting Belt Tension / p. 25).

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 24 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Transverse Cassette Drive 25

Adjusting Belt Tension 5.3

25 - 30 mm

F = 20 - 22 N

F = 20 - 22 N

20 - 25 mm

Fig. 9: Cassette transport belt


• Transport belt
- Hook on a spring balance as shown i (Fig. 9 / p. 25)
- Adjustment dimension: at 20-25 mm deflection: 20-22 N
• Auxiliary belt
- Hook on a spring balance as shown in (Fig. 9 / p. 25)
- Adjustment dimension: at 25-30 mm deflection: 20-22 N
• Install the loading slot cover.
Perform a movement test of the cassette drive using the SSW:
- SSW: Diagnostic/ Tests/ Movement tests/ Movement test/ Transverse cassette
drive in park position direction
- SSW: Diagnostic/ Tests/ Movement tests/ Movement test/ Transverse cassette
drive in loading position direction

NOTE For remarks regarding the adjustment, as well as the function


test, see Help in the SSW.

Siemens AG AXD3-340.841.02.05.02 Page 25 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
26 Transverse Cassette Drive

Replacing the Loading Slot Photo-sensitive Cell 5.4

• Remove the cover over the loading slot (see “Access to the Spotfilm Device”)
• Replace the photo-sensitive cells in the loading slot cover; a mechanical adjustment is
not possible.
After completing the replacement, using the SSW:
• SSW: Diagnostic / tests/ Movement tests/ Transverse cassette drive in loading
position direction and
• SSW: Diagnostic / tests/ Movement tests/ Transverse cassette drive in park posi-
tion direction
the position of the park photo-sensitive cell should be checked; this is important for
positioning of the exposure position and of cassette detection.

NOTE For remarks regarding the adjustment, as well as the function


test, see Help in the SSW.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 26 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Transverse Cassette Drive 27

Replacing the Park Photo-sensitive Cell 5.5

2 2
1

Fig. 10: Replacing the park photo-sensitivecell


• Remove the spotfilm device cover panel (Access to the Spotfilm Device / p. 9).
• Loosen the mount (1/Fig. 10 / p. 27) for the photo-sensitive cell bracket
(2/Fig. 10 / p. 27).
• Remove the connector from the lower board (3/Fig. 10 / p. 27).
• Install a new photo-sensitive cell on the bracket.
• Plug the cable on the lower board back in (3/Fig. 10 / p. 27).
• Secure the photo-sensitive cell bracket (2/Fig. 10 / p. 27) approx. in the middle of the
long holes (1/Fig. 10 / p. 27).
Using the SSW:
• SSW: Diagnostic / tests/ Movement tests/ Transverse cassette drive in loading
position direction and
• SSW: Diagnostic / tests/ Movement tests/ Transverse cassette drive in park posi-
tion direction
check the position of the park photo-sensitive cell and, if necessary, adjust it.

NOTE For remarks regarding the adjustment, as well as the function


test, see Help in the SSW.

Siemens AG AXD3-340.841.02.05.02 Page 27 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
2 8 Diaphragm Plates
6-
Replacing the Drive
6Diaphragm Plates

4 3
2

6
2

5 1

10

10
9
8

10

10

Fig. 11: Replacing the diaphragm plates


• Remove all cover panels as shown in (Access to the Spotfilm Device / p. 9).
• Loosen the nut (1/Fig. 11 / p. 28) on the motor assembly (2/Fig. 11 / p. 28); when this is
done, the tension on the belt is relaxed.
• Take the belt off the drive wheel (3/Fig. 11 / p. 28).
• Disconnect the cable to the motor.
• Remove the 3 countersunk Allen screws (4/Fig. 11 / p. 28) on the back of the spotfilm
device.
• Remove the encoder cable (5/Fig. 11 / p. 28) below the drive.
• The drive can now be moved towards the middle of the spotfilm device and removed, if
needed, remove a cable clamp (6/Fig. 11 / p. 28) that does not need to be replaced.
• Install the new drive in the reverse order of the instructions above.
• Perform the adjustment of the belt.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 28 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Diaphragm Plates 29

Replacing the Belt or the Diaphragm Plates 6.1

• Loosen the nut (1/Fig. 11 / p. 28) of the motor assembly (2/Fig. 11 / p. 28); when this is
done, the belt tension is relaxed.
• Remove the hygienic film.
• Remove the countersunk Allen-head screws (7/Fig. 11 / p. 28) on the operator’s side
return pulley (8/Fig. 11 / p. 28), remove the return pulley.
• If necessary, move the tabletop transversely to the operator’s side.
• Take the belt off the drive wheel (3/Fig. 11 / p. 28).
• The two diaphragm plates (9/Fig. 11 / p. 28) can be removed carefully from the opera-
tor’s side.
• The diaphragm plates (9/Fig. 11 / p. 28) must be placed down flat.

5 3

X Z

40 40
1 A B Y 2

Fig. 12: Diaphragm plates belt winding


Explanation about (Fig. 12 / p. 29):
• Pulley on the operator’s side (1/Fig. 12 / p. 29)
• Drive wheel (2/Fig. 12 / p. 29)
• Operator’s side diaphragm plate (A/Fig. 12 / p. 29)
• Wall side diaphragm plate (B/Fig. 12 / p. 29)
Total length of belt = 1853 mm ± 1 tooth
• Belt length spacing of X to Y = 698 mm to be set (or next tooth of the toothed belt).
Insert the belt accordingly in the belt clamp (3/Fig. 12 / p. 29).
• The belt must extend at both ends by 2 teeth (3/Fig. 12 / p. 29) out of the belt clamp
(4/Fig. 12 / p. 29).
• The dimension of X to Z results automatically from the belt length.
• Lightly lubricate the glide pieces (10/Fig. 11 / p. 28) with silicon paste.
• Insert the collimator plate, with the drag plate it configured.
• Take note of the belt winding (Fig. 12 / p. 29).

Siemens AG AXD3-340.841.02.05.02 Page 29 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
30 Diaphragm Plates

• Place the belt on the drive pulley (3/Fig. 11 / p. 28).


• Take note of the position of the return pulley (8/Fig. 11 / p. 28), place the belt on it and
install the return pulley (8/Fig. 11 / p. 28).

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 30 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Diaphragm Plates 31

Adjusting the Belt Tension 6.2

10 - 12 mm

F = 10 - 12 N

Fig. 13: Diaphragm plates belt tension


• Insert a 24 x 30 cm or 10 x 12 inch cassette.
• Select the 4-part subdivision.
• Disconnect a motor cable to the collimator plate drive.
• Remove the cassette.
• Adjust the belt tension by shifting the position of the motor assembly as shown in
(2/Fig. 11 / p. 28).
• Perform a function test of the diaphragm plates using the SSW:
- SSW: Adjustment/ Calibration/ Spotfilm Device/ Diaphragm plates/ Basic set-
ting.

NOTE For remarks regarding the adjustment, as well as the function


test, see Help in the SSW.

Siemens AG AXD3-340.841.02.05.02 Page 31 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
32 Diaphragm Plates

Adjusting the Diaphragm Symmetry 6.3

Prerequisite
The symmetry can only be checked using an exposure made under specific conditions.
Expose a film as described in the SSW.
• Adjustment/ Calibration/ Spotfilm Device/ Diaphragm plates: Basic setting
Explanation:
1. The diaphragm plate opening results from the distance of the diaphragm plate (wall
side) to the photo-sensitive cell and an offset that can be influenced by the SSW.
2. The symmetry depends only on the position of the belt clamp (5/Fig. 12 / p. 29) on the
operator’s side diaphragm plate (A/Fig. 12 / p. 29).
Procedure:
• Remove the diaphragm plates.
• Mark the position of the belt clamp (5/Fig. 12 / p. 29) on the operator’s side diaphragm
plate (A/Fig. 12 / p. 29), move the belt clamp accordingly and secure it in place.
Perform the function check again:
• SSW: Adjustment/ Calibration/ Spotfilm Device/ Diaphragm plates: Basic setting

NOTE For remarks regarding the adjustment as well as the function test,
see Help in the SSW.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 32 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Grid 33
7-
Replacing and Adjusting the Grid
7Grid

6 6 4
7 3

1
2

Fig. 14: Replacing the grid


• Remove all cover panels as shown in (Access to the Spotfilm Device / p. 9).
• Remove the Allen-head countersunk screws (1/Fig. 14 / p. 33) on the return pulley on
the operator’s side (2/Fig. 14 / p. 33).
• It may be necessary to move the tabletop transversely to the operator’s side.
• Remove the 3 Allen-head countersunk screws (3/Fig. 14 / p. 33).
• Remove the drive with the belt.
• Remove the belt from the drive pulley (4/Fig. 14 / p. 33).
• Remove the bottom return pulley (5/Fig. 14 / p. 33) on the motor assembly
(6/Fig. 14 / p. 33).
• Measure belt tension of the diaphragm plates and make a note of the value.
• Remove the hygienic film.
• Remove the diaphragm plate return pulley (7/Fig. 14 / p. 33).
• The grid can be removed from the operator’s side.
• The grid must be placed down flat.
• Replace the belt on the grid.
• Install the drive.
• Insert the grid.
• Be careful of how the belt is installed; first thread the belt in through the removed motor
assembly.
• Install the return pulley (2/Fig. 14 / p. 33).

Siemens AG AXD3-340.841.02.05.02 Page 33 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
34 Grid

Adjusting Belt Tension 7.1

F= 10 - 12 mm

Fig. 15: Diaphragm plate - grid belt


• Manually slide the grid into the exposure position.
• Adjust the belt tension by adjusting the position of the lower return pulley on motor
assembly. Hook on a spring balance as shown in (Fig. 15 / p. 34).
Adjustment value: with 4-5 mm deflection, 10-12 N
• Install the diaphragm plate return pulley (7/Fig. 14 / p. 33).
• Adjust the diaphragm plate belt tension to the value that was noted.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 34 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
Grid 35

Adjusting the Potentiometer 7.2

• Using the SSW, perform:


- SSW: Adjustment/ Calibration/ Spotfilm Device/ Grid/ Potentiometer adjust-
ment
- SSW: Adjustment/ Calibration/ Spotfilm Device/ Grid/ Exposure position
Adjust the potentiometer and the exposure position.

NOTE For remarks regarding the adjustment, as well as the function


test, see Help in the SSW.

• It is also absolutely necessary to perform a function test using the SSW:


- SSW: Adjustment/ Calibration/ Spotfilm Device/ Diaphragm plates: Basic set-
ting

NOTE For remarks regarding the adjustment, as well as for the function
test, see Help in the SSW.

Siemens AG AXD3-340.841.02.05.02 Page 35 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
36 IONTOMAT
8-
Iontomat Chamber Replacement
8IONTOMAT

Fig. 16: Iontomat chamber replacement


• Remove all cover panels as shown in (Access to the Spotfilm Device / p. 9).
• Completely remove the middle wall (1/Fig. 16 / p. 36) on the operator’s side.
- When doing this, guide the Iontomat cable through!
- The Iontomat chamber is pressed tightly against the walls of the middle wall; because
of this, insert a ruler or a similar object between the plastic wall and the Iontomat
chamber at the top and bottom.
• The chamber can be slid out of the middle wall by inserting an approx. 25 cm long
screwdriver into an opening in the middle wall (2/Fig. 16 / p. 36) from the head end,
approx. in the middle of the Iontomat chamber.
• Install the new chamber in the reverse of the steps above. Be especially careful of the
connectors on the chamber.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 36 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
IONTOMAT 37

Fig. 17: Tension arms


When inserting the middle wall, it is absolutely necessary to lift the tension arms on
the wall side.
• If the spacer nipples have already fallen out in a spotfilm device (1/Fig. 17 / p. 37)
(scratches on the middle wall), they must be reinstalled.
- Completely remove the middle wall.
- Insert the spacer nipples accordingly (1/Fig. 17 / p. 37).
- Lift the tension arms on the wall side and reinsert the middle wall.
• There are also 2 spacer nipples on the bottom of the tension arms (2/Fig. 17 / p. 37).
These can be sheared off during operation or during a service situation.
• Insert the middle wall.
- To insert the middle wall, a pull wire should be attached to the Iontomat cable.
- When inserting it, the middle wall hits against a grabber block for the drag plate.
- The grabber block (visible from the operator’s side) must be lifted slightly; this can be
done using a ruler or a small piece of wood.
• Take note of how the cables are laid.
• Plug in the Iontomat connector and secure the cable loop with the cable tie
(3/Fig. 16 / p. 36).
Function Test of Chamber Selection
- Connector properly plugged in?
• Insert the 2.1 mm copper filter into the collimator guide rails.
• Trigger an Iontomat exposure for each chamber.

Siemens AG AXD3-340.841.02.05.02 Page 37 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
38 IONTOMAT

• After each exposures, check the mAs post display; approximately the same mAs
should be displayed for all exposures.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 38 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
I.I. Cover Panel 39
9-
Replacing the I.I. Cover Panel
9I.I. Cover Panel

The following is required for reliable cassette transport:


• The tension force of the “cassette tension” drive must be adjusted.
• The “cassette transport” belt tension must be adjusted.
• The plastic cover and the metal frame of the “I.I. cover” component must be centered to
one another when installed.
• The I.I. cover may not be buckled in either the table transverse or the table longitudinal
direction.
• With large, sharp-cornered cassettes, the cassette buckling may not be more than 1.5
mm, otherwise it can lead to malfunctions.
• In the area of the tension arm mechanism, the cassette must be higher than the mecha-
nism.

NOTE With the new I.I. cover panels, there are 2 raised locations (strips).
These minimize the contact surface to the cassette (lessening of
friction). In addition, the strips in principle raise the cassette 1-2
mm higher. If this is not sufficient, the height of the cassette can
be adapted by placing strips (Y/Fig. 18 / p. 39) underneath it.

• Remover all cover panels as shown (Access to the Spotfilm Device / p. 9).

Fig. 18:
• Remove the I.I. cover panel:
- To do this, remove the 4 countersunk screws (A/Fig. 18 / p. 39).
- The I.I. component group cover panel can now be removed towards the front.

Siemens AG AXD3-340.841.02.05.02 Page 39 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
40 I.I. Cover Panel

Fig. 19:
The two parts, the metal frame and the plastic panel, of the I.I. cover panel must be cen-
tered when installed.
To assure this, perform the following work steps.
• Remove the 4 Allen screws (1/Fig. 19 / p. 40) on both sides.
• Loosen the 3 Allen screws (2/Fig. 19 / p. 40).
• Loosen the nuts (3/Fig. 19 / p. 40) on both sides.
• Carefully insert the cover panel.
- When doing this, lift the plastic cover on the wall side, otherwise the cover will jam
against the I.I. or the tension mechanism.
• Insert the 4 Allen screws (A/Fig. 18 / p. 39).
• Carefully move the metal frame to the stop on the wall side.
• Tighten the middle Allen screws (2/Fig. 19 / p. 40).
• Remove the 4 Allen screws (A/Fig. 18 / p. 39).
• Carefully pull out the I.I. cover panel approx. 10 cm.
- When this is done, the plastic panel and the metal frame may not move towards one
another.
• Carefully lift the plastic panel and remove the protective film (4/Fig. 19 / p. 40) on the
adhesive strips.
• Press the plastic panel lightly onto the adhesive strips.
• Tighten the 6 Allen screws (1, 2/Fig. 19 / p. 40).
• Carefully pull out the I.I. cover panel completely.
• Tighten the nuts (3/Fig. 19 / p. 40) on both sides.
• Carefully lift the plastic cover panel and remove the protective film (5/Fig. 19 / p. 40) on
the adhesive strips.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 40 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions
I.I. Cover Panel 41

• Press the plastic panel lightly onto the adhesive strips.


• Under no circumstances may the I.I. cover panel have a belly in it in the area of the 3
drawn lines (Fig. 19 / p. 40). If necessary, lightly bend up the plastic panel and the
metal frame to remove the belly.
• Carefully insert the I.I. cover panel.
• Insert the 4 Allen screws (A/Fig. 18 / p. 39) and tighten them.
• Manually insert the largest available cassette and slide it into the park position.
• The lightly press down on the cassette and slide it in the direction of the load position.
• If the cassette hits against the tension mechanism in the area of X (X/Fig. 18 / p. 39)
place an aluminum strip underneath at Y (Y/Fig. 18 / p. 39).
- To do this, remove the 4 Allen screws (A/Fig. 18 / p. 39).
• Remove the protective film from the adhesive strip on the aluminum strip.
• Stick the aluminum strip on the plastic panel (there should be approx. 5 mm spacing
between edge X (X/Fig. 18 / p. 39) of the tension mechanism and the aluminum strip).
• Insert the 4 Allen screws (A/Fig. 18 / p. 39) and tighten them.
• Perform the test again, if necessary, place a 2nd aluminum strip underneath.
• After completing the work steps, reinstall the cover panels in the reverse order of the
instructions above.

Siemens AG AXD3-340.841.02.05.02 Page 41 of 42 AXIOM Iconos R200


Medical Solutions 04.02 TD SD 21
42 Changes to Previous Version
10Changes to Previous Version 10-

Establishing access Removing the hygienic film Added.


to the spotfilm device Middlewall Illustration added and text
Centering template expanded.
Added.
General Remarks Cassette transport prob- Illustration added and text
Regarding Cassette lems expanded.
Transport
IONTOMAT Iontomat chamber replace- Illustration added and text
ment expanded.

AXIOM Iconos R200 AXD3-340.841.02.05.02 Page 42 of 42 Siemens AG


04.02 TD SD 21 Medical Solutions

You might also like