Digital Portable X-Ray Systems: Manual Ver1.7
Digital Portable X-Ray Systems: Manual Ver1.7
Digital Portable X-Ray Systems: Manual Ver1.7
Digital portable
x-ray systems
Head office: 701, Hanshin IT Tower II, Gasan-Dong, Geumcheon-Gu, Seoul 153-023, Korea
Factory: 701-1, Hanshin IT Tower II, Gasan-Dong, Geumcheon-Gu, Seoul 153-023, Korea
TEL : +82-2-2025-3760 FAX: +82-2-2025-3763
URL : www.ecotron.co.kr
1
Manual ver1.7
CONTENTS
1. NOTICE OF THE SAFE OPERATION-------------------------------------------------------------3
2. RADIATION SAFETY--------------------------------------------------------------------------------4
3. WORKING ENVIRONMENT------------------------------------------------------------------------5
4. SYSTEM CONFIGURATION------------------------------------------------------------------------6
5. OPERATION MODE---------------------------------------------------------------------------------8
6. SPECIFICATIONS----------------------------------------------------------------------------------11
7. DIMENSION----------------------------------------------------------------------------------------13
8. TECHNIQUE CHART FOR SMALL ANIMALS--------------------------------------------------14
9. EQUINE CHART-----------------------------------------------------------------------------------15
TABLES
1. TABLE 1.2.3 SPECIFICATIONS------------------------------------------------------------------11
2. TABLE 4. CHART FOR SMALL ANIMALS------------------------------------------------------14
3. TABLE 5. EQUINE CHART-----------------------------------------------------------------------15
4. TABLE 6. EQUINE CHART-----------------------------------------------------------------------16
2
Manual ver1.7
This user’s guide is designed to ensure correct use and operation of EPX series. Please read all the
lines thoroughly before you use this equipment.
Incorrect use and operation exceeding described conditions in this manual may occur damage of the
machine and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.
This equipment should be used only by the legally qualified persons and practitioners.
EPX series is designed with the due consideration for users’ safety and product reliability. It, however, is
advisable to follow under mentioned rules to keep your additional safety and health.
1. This product should be operated only by or under the supervision of legally qualified persons.
2. EPX series is designed for the radiographic uses and not for fluoroscopy or other associated
applications.
3. EPX series should be used for the diagnosis, not for the therapy.
4. EPX series is specified as ClassⅠtype B equipment under the IEC 60601-1,2,3, IEC 60601-2-7,28
standard.
5. Do not modify the equipment at your discretion and in case any modification is required unavoidably,
ask the help of Ecotron or its authorized dealer for the service.
3
Manual ver1.7
2. RADIATION SAFETY
1. Users and operators should wear appropriate protecting devices and clothes.
2. Stay distance from the radiant sources and all the possible secondary radiation zones.
4. The distance from the focus to skin should be kept at least 8 inch (20cm).
5. For the experimental uses, apply the lowest possible values of kV, mAs and exposure time.
CAUTION
The ionizing radiation could be dangerous for patients and operators
unless following safety regulations are strictly observed.
4
Manual ver1.7
3. WORKING ENVIRONMENT
Avoid the following places for the normal operation and safe storage:
For safe operation of the equipment, it should be kept away from the place with strong vibration and
maintain proper environment and conditions.
Operation Environment
For safe storage and transportation, the following range of temperature and humidity must be kept.
5
Manual ver1.7
4. SYSTEM CONFIGURATION
EPX Series is a radiological device for veterinary uses. It should be applied for the radiographic
diagnosis and operated by qualified practitioners. Users have to comply with safety and health
regulations concerning the ionizing radiation protection and the electrical and mechanical safety of the
veterinary medical devices.
1. Operation Practices
b. Stay at least 6.5feet (2m) away from the unit or extend the cord of hand switch as long as possible
in order to secure enough space for your safety.
c. Use proper field size and technical values for each procedure to minimize x-ray exposure dose and
obtain the best radiographic result.
d. When you make exposure to the patient on the bed, ask visitors to leave the room first and keep
appropriate distance from the patient.
e. Pay attention to the maintenance schedule of the device and keep up with it.
f. Cumulative radiation dose usually does not exceed recommended maximum levels. However, if you
make radiographic exposures using high kV and mAs quite often, the evaluation of the specialist is
needed to check whether extra protective devices are necessary for the user or not.
b. Arrange the SID (Source to Image Receptor Distance) using a tape measure located on the side of
the collimator.
c. Turn on the collimator lamp and laser pointer by the pushing collimator/laser switch is on.
d. Adjust the size of the x-ray beam in accordance with the film size using the x-ray field
6
Manual ver1.7
adjustment knobs. The collimator and laser pointer will be turned off after 30 seconds by internal
timer.
Please check whether the voltage and frequency of the input power
are in accordance with the figures written in the system labels,
which attached on the body of the machine. The level of input
WARNING voltage fluctuation should be within ±10% of the nominal values.
For the first time installation of the (machine) or operation after long
period of non-use, it needs to be preheated to secure long service
life of the tube. Following method of pilot test will relieve the
damage on the x-ray tube. Operate the (machine) with low KV/mAs
values (50KV/5mAs) three times in row and execute with higher
CAUTION
values (70KV/5mAs) three times in row again. Then high KV/mAs
values (90kV/5mAs) will be applied without trouble.
7
Manual ver1.7
6. OPERATION MODE
6 Ready indicator Lighted push the hand switch one time for preparation
10 Collimator lamp on Switch for operation of the lamp inside the collimator
12 Display reversal switch Help to read LED value of kV, mAs reversed
13 mAs & sec select switch The switch select between mAs and sec
8
Manual ver1.7
1. Normal Mode
a. Connect power plug into the device.
e. Positioning the machine at the ready condition by pushing the hand one time
switch.
f. When the “Ready LED” lighted, press the hand switch again and make exposure. During the x-
ray exposure, “X-ray LED” will be lighted on.
g. After the exposure is ended, “wait LED” sign will be lighted on for the waiting time.
2. APR Mode
a. Connect power plug into the device.
f. Save the selected KV and mAs figures with pushing the “save” button.
g. When the “Ready LED” sign appears, press the hand switch again and make
exposure. During the x-ray exposure, the “X-ray LED” will be lighted on.
i. After the exposure is ended, [5] “wait LED” will be lighted on for the waiting time.
9
Manual ver1.7
10. SPECIFICATIONS
[TABLE 1. EPX-F2400]
Output Power 2.4 KW
Input Power Voltage 110 V-120V / 220V-240V
Phase & Frequency Single, 50/60 Hz
Tube voltage current mAs
40KV ~ 60KV 40mA 0.4 ~ 100
61KV ~ 70KV 30mA 0.4 ~ 64
61KV ~ 70KV 25mA 80 ~ 100
71KV ~ 80KV 30mA 0.4 ~ 32
Radiography KV Range in 1KV steps
71KV ~ 80KV 25mA 40 ~80
81KV ~ 90KV 25mA 0.4 ~ 40
81KV ~ 90KV 20mA 50 ~ 80
91KV ~ 100KV 20mA 0.4 ~ 50
91KV ~ 100KV 16mA 64 ~ 80
mAs Range 0.4mAs – 100mAs, 25steps
Max. KV accuracy ±3 %
Max. mAs Deviation ±5 %
Display KV / mAs: 7-segment LED
Model TOSHIBA D-125
Focal Spot Size 1.2mm x 1.2mm
X-ray Tube Target Angle 16 degree
Anode Heat Storage 50 kHU
Heat Dissipation 350 HU/sec
Total Filtration 2.5 mm Al eq. @ 100kV
Type Double slit type, manual operation
Min. X-ray Field Size 5cm x 5cm @1m SID
Collimator
Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser
Laser Pointer Class ⅢA, 5mW
Pointer
Timer Push button illuminator with 30 sec timer
Lamp 12V, 50W Halogen
Weight 11 Kg
10
Manual ver1.7
[TABLE 2. EPX-F2800]
Output Power 2.8 kW
Input Power Voltage 110 V-120V / 220V-240V
Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs
40KV ~ 60KV 40mA 0.4 ~ 100
61KV ~ 70KV 30mA 0.4 ~ 64
61KV ~ 70KV 25mA 80 ~ 100
71KV ~ 80KV 35mA 0.4 ~ 10
71KV ~ 80KV 30mA 13 ~ 32
71KV ~ 80KV 25mA 40 ~ 80
81KV ~ 90KV 25mA 0.4 ~ 10
Radiography kV Range in 1kV steps 81KV ~ 90KV 25mA 13 ~ 40
81KV ~ 90KV 20mA 50 ~ 80
91KV ~ 100KV 20mA 0.4 ~ 13
91KV ~ 100KV 20mA 16 ~ 50
91KV ~ 100KV 16mA 64 ~ 80
101KV ~ 110KV 18mA 0.4 ~ 32
101KV ~ 110KV 16mA 40 ~ 64
111KV ~ 120KV 16mA 0.4 ~ 20
111KV ~ 120KV 14mA 25 ~ 50
MAs Range 0.4mAs – 100mAs, 25steps
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name D-125 TOSHIBA
Focal Spot 1.2mmx1.2mm
X-ray Tube Target Angle 16 degree
Anode Heat Storage 50 kHU
Heat Dissipation 350 HU/sec
Total Filtration 2.5 mm Al eq. @ 120kV
Min. X-ray Field Size ≦5cm x 5cm @1m SID
Collimator Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser Laser Pointer Class : ⅢA 5mW
Pointer Timer Push button illuminator with 30 sec timer
Lamp 12V 50W Halogen
Weight 11.5 Kg
11
Manual ver1.7
[TABLE 3. EPX-F3200]
Output Power 3.2 KW
Input Power Voltage 220V-240V
Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs
40KV ~ 50KV 60mA 0.4 ~ 100
51KV ~ 60KV 50mA 0.4 ~ 80
61KV ~ 70KV 40mA 0.4 ~ 100
71KV ~ 80KV 40mA 0.4 ~ 32
Radiography kV Range in 1KV steps
71KV ~ 80KV 35mA 40 ~80
81KV ~ 90KV 30mA 0.4 ~ 40
81KV ~ 90KV 25mA 50 ~ 80
91KV ~ 100KV 30mA 0.4 ~ 50
91KV ~ 100KV 25mA 64 ~ 80
MAs Range 0.4mAs – 100mAs, 25steps
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name D-205S TOSHIBA
Focal Spot 2.0mmx2.0 mm
X-ray Tube Target Angle 16 degree
Anode Heat Storage 40 kHU
Heat Dissipation 378 HU/sec
Total Filtration 2.5 mm Al eq. @ 100KV
Type Double slit type, Manually operation
Min. X-ray Field Size ≦5cm x 5cm @1m SID
Collimator
Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser
Laser Pointer Class : ⅢA 5mW
Pointer
Timer Push button illuminator with 30 sec timer
Lamp 12V 50W Halogen
Weight 12 Kg (26.4 lb.)
12
Manual ver1.7
[EPX-F4000]
Output Power 4 KW
Input Voltage 200V-250V
Power Phase & Frequency Single / 50/60 Hz
Tube voltage current mAs Sec
50KV 80mA 0.4~16mAs 0.2
50KV 70mA 18~32mAs 0.45
50KV 60mA 40~100mAs 1.666
51KV ~ 60KV 60mA 0.4~32mAs 0.533
51KV ~ 60KV 50mA 40~100mAs 2
61KV ~ 70KV 50mA 0.4~32mAs 0.64
61KV ~ 70KV 40mA 40~100mAs 2.5
Radiography kV Range in 1KV steps
71KV ~ 80KV 50mA 0.4~32mAs 0.64
71KV ~ 80KV 40mA 40~80mAs 2
81KV ~ 90KV 40mA 0.4~32mAs 0.8
81KV ~ 90KV 30mA 40~80mAs 2.666
91KV ~ 100KV 35mA 0.4~32mAs 0.914
91KV ~ 100KV 25mA 40~80mAs 3.2
101kV ~ 110KV 30mA 0.4~32mAs 1.066
101kV ~ 110KV 20mA 40~80mAs 4
MAs Range 0.4mAs – 100mAs
Max. kV Deviation ±3 %
Max. mAs Deviation ±5 %
Display KV /mAs: 7-segment LED
Model Name CEI OX/105-2
Focal Spot 1.8mmx1.8mm
X-ray Tube
Target Angle 19 degree
Anode Heat Storage 30000J
Total Filtration 2.5 mm Al eq. @ 100KV
Type Double slit type, Manually operation
Min. X-ray Field Size ≦5cm x 5cm @1m SID
Collimator
Max X-ray Field Size 40cm x 40cm @ 75cm SID
with Laser
Laser Pointer Class : ⅢA 5mW
Pointer
Timer Push button illuminator with 30 sec timer
Lamp 12V 50W Halogen
13
Manual ver1.7
Side view
14
Manual ver1.7
Front view
* NOTE
The contents in the table are recommended for starting radiographic techniques.
The final radiography of proper image density and resolution depends on many factors. Please follow the
next adjustment and common rules, if necessary.
1. Add or subtract 2 KV for 1 cm increase or decrease in thickness of body section.
2. If films are too dark, decrease mAs value to the next lower station.
3. If films are too light, increase mAs value to the next higher station
15
Manual ver1.7
9. EQUINE CHART
16
Manual ver1.7
Mid-Contrast
ANATOMY VIEW DISTANCE
KV mAs
LAT 18 inch distance 78 1.2
STIFLE
PA 18 inch distance 78 2.5
AP 26 inch distance 68 2.5
NAVICULAR
LAT 26 inch distance 68 2.0
FOAL 33 inch distance 90 0.8
CHEST 300LBS 33 inch distance 100 1.2
500LBS 33 inch distance 100 4.0
SINUS LAT 33 inch distance 88 1.2
LAT 30 inch distance 86 3.2
SHOULDER
OBI 30 inch distance 86 4.0
17
Manual ver1.7
Chapter 1. Introduction
This manual is designed to ensure correct use and operation of EPX model portable x-ray units. Please
read all the lines thoroughly before you use the equipment.
Incorrect use and operation exceeding described conditions in this manual may cause damage of the
equipment and shorten its life. Particular attention must be paid to all the warnings, cautions and notes
incorporated herein.
This equipment should be used only by the legally qualified persons and practitioners.
Regarding the AC cable and software, only the ones developed and supplied by Ecotron Inc. should be
used.
No responsibility is taken by Ecotron for any infringement of patients or other rights of the third parties
which many results from the use of this manual.
NOTE: If the output calibration is needed after periodical inspections or repairs, please observe following
rules and methods described in this manual.
CAUTION
The ionizing radiation is dangerous for the operator if the following safety measures
are not strictly observed.
18
Manual ver1.7
Chapter 2. Specification
2-1. System Block Diagram
19
KV Up/Dn Switch
mAs Up/Dn Switch
APR(1~6) Switch
20
Save Switch
2-2.1 Control Panel Layout
LAYOUT
Reverse Switch
Manual ver1.7
2-2
Manual ver1.7
21
Manual ver1.7
22
Manual ver1.7
23
Manual ver1.7
24
Manual ver1.7
25
Manual ver1.7
h. Push APR S/W ① and then the display should indicate as follows,
where, P01 is Filament preheat calibration code (Preheat calibration code of 30mA)
and 1.22 is Filament preheat calibration data (Preheat calibration data of 30mA).
l. Adjust data by using the mAs Up-Down switch on the Control Panel.
m. After it is adjusted, push the Store switch (see operation sheet).
n. DIP SWITCH #4 on the PCB OP BOARD is set to OFF position then exposure X-ray.
26
Manual ver1.7
o. If overshooting or undershooting occurs in the waveform, set the calibration data again.
3-2.4 DIP Switch setting
3-3.1 Introduction
The Exposure Hand Switch consists of a two step hand switch.
The first (1) stage of the hand switch is called the “Preparation (Prep or READY)” stage.
The second (2) stage of the hand switch is called the “Exposure (Exp or X-Ray Out)“ stage.
3-3.2 Hand switch Layout
SWITCH 1
SWITCH 2
27
Manual ver1.7
There are two ways to use the hand switch to get the exposure.
a. Press and hold the Switch #1 for longer than 1 second, which will get the “Prep or
Ready” status on, then press Switch #2 to get the “Exposure”.
b. Press and hold the Switch #1 and Switch #2 simultaneously for longer than 1
second. This action will result in “Prep or Ready” status and “Exposure” at the same
time.
3-4.1 Introduction
All ORANGE-F series models can be equipped with an optional remote control
system using either Wireless Bluetooth Interface or RS-232 communication cable
between PC and the x-ray unit. This option will allow the full control of the x-ray unit
from PC.
28
Manual ver1.7
handshaking type. In this cable, only the data lines and signal ground are cross connected as
2 3 Rx Tx
3 2 Tx Rx
5 5 Signal ground
a. Connect the HPS-120 and HPU-100 modules to x-ray unit and PC, respectively. Both
x-ray unit and PC powers should be turned on before the connection.
b. The HPS-120 module’s red LED (OPR) light should be on, and the HPU-100
module’s red LED (OPR) should blink periodically once it is successfully installed.
29
Manual ver1.7
30
Manual ver1.7
3-5 Colimator
3-5.1 Specification
a. Halogen Lamp * 1
b. Laser pointer * 1
c. Horizontal and vertical lighting range control
3-5.3 Layout
a. External Layout
31
Manual ver1.7
b. Internal Layout
c. Knob Layout
32
Manual ver1.7
3-6 DR-InterfaceConnection
a. Normally, X-ray unit goes through READY, EXPOSURE process to get the x-ray output by using a
hand switch or remote communication signals. However, once the DR interface is used, it controls the
exposure exclusively by the EXP_OK_INPUT signal.
b. As shown in the timing chart above, activating the Ready SW and Exposure SW will result in the
outputs of Ready_SW_OUPUT and EXP_SW_OUTPUT, respectively, and waiting for the (external,
user provided) EXP_OK_INPUT signal to produce x-ray output.
c. Therefore, only when the EXP_OK_INPUT signal is available, the x-ray unit generates x-ray output.
When the x-ray generation is stopped by the preset time which is set by the control panel,
EXP_END_SIGNAL_OUTPUT is generated for the duration of 0.1 second.
33
+5V +5V
RA2 2
+3.3V 1 3 U2 ATMEGA128
D1 BZ1 4.7K 4
SY S LED R1 5
15 1/4W 1 HS_PRE 12 36 LED_LOAD
HS_EXP 13 P1.0/TACLK P4.0/TB0 37 LED_DATA
2 1 8 14 P1.1/TA0 P4.1/TB1 38 LED_CLK
2 7 CS_ERROR 15 P1.2/TA1 P4.2/TB2 39 WDT_OUT
QFP-03A
D2 3 6 16 P1.3/TA2 P4.3/TB3 40 READY _SW_OUTPUT
1 2 4 5 17 P1.4/SMCLK P4.4/TB4 41 EXP_SW_OUTPUT
SW1 18 P1.5/ TA0 P4.5/TB5 42 EXP_OK_INPUT
P1.6/ TA1 P4.6/TB6 +3.3V
1
SW DIP-4 19 43 EXP_END_SIGNAL_OUTPUT
C1 1N4148 P1.7/ TA2 P4.7/TBCLK
R2 1uF/25V TANTAL/SMD BUZZER 20 44 SW_SA
P2.0/ACLK P5.0/STE1
2
1K SY SLED 21 45 SW_SB
P2.1/TAINCLK P5.1/SIMO1
1
/LAMP_ON LAMP_ON SW_SC
D3 1N4148
D4 1N4148
D5 1N4148
D6 1N4148
22 46
/HV_ON HV_ON 23 P2.2/ CAOUT/ TA0 P5.2/SOMI1 47 SW_SD
/PREP PREP 24 P2.3/ CA0/ TA1 P5.3/UCLK1 48 SW_R1
FIL_IP_CHANGE1 FIL_IP_CHANGE 25 P2.4/ CA1/ TA2 P5.4/MCLK 49 SW_R2
/LASER LASER 26 P2.5/ Rosc P5.5/ SMCLK 50 SW_R3
J1
P2.6/ ADC12CLK/DMAE0 P5.6/ ACLK
2
/ZC /100V_SEL 27 51 SW_R4
1 P2.7/ TA0 P5.7/ TBOUTH/SVSOUT R4 100
2 +5V /TRIAC_ON 28 59 THEMAL_MON
THERMAL_ERROR 29 P3.0/ STE0 P6.0/ A0 60 R5 100 EP_FB_ADC
CON2
30 P3.1/ SIMO0/SDA P6.1/ A1 61 R6 100 IP_FB_ADC
J2 TP7
P_P_ERR 31 P3.2/ SOMI0 P6.2/ A2 2 FIL_FB_ADC
1 /TRIAC_ON TXD232C_CPU P3.3/ UCLK0/SCL P6.3/A3 TP12
32 3 1 R7 100
2 RXD232C_CPU 33 P3.4/ UTXD0 P6.4/A4 4 1
P3.5/URXD0 P6.5/A5
1
+5V TXD422_CPU 34 5 EP_REF
D7 1N4148
D8 1N4148
D9 1N4148
D10 1N4148
CON2
P3.6/UTXD1 P6.6/A6/DAC0
2
/ZC RXD422_CPU 35 6 IP_REF CON14AP
C3 100pF
C4 100pF
C5 100pF
C6 100pF
J14
P3.7/URXD1 P6.7/A7/DAC1/SVSIN 13 14
1 /LASER 58 /RST 11 12
2 RST/ NMI
1
9 10
2
CON2 +7V 8 57 TCK TCK 7 8
9 XIN TCK 56 TMS TMS 5 6
XOUT TMS 55 TDI TDI 3 4 +3.3V
TDI/TCLK
2
53 54 TDO TDO 1 2
52 XT2IN TDO/ TDI
XT2OUT
2
R8 R9 +3.3V
270 270 C9 64 J3 +5V
+3.3V AVcc 7
20p/0805 V REF+ R82
1
10 JP1
X2 1 Ve REF+
HS_PRE 63 DVcc 11 1
CRYSTAL DVss V REF-/Ve REF- 2
1
2
62 10/0805
AVss
2
1
1 20p/0805
2
2
3 HS_EXP TP3 +5V
4
1
34
CON4 ISO2
TLP121 C43 C44 RS232C DISPLAY
104 100u/25V RXD232C_CPU SW_R1 LED_DATA
RXD232C_CPU SW_R1 LED_DATA
2
3
2
1
1
PULSE BY PULSE
3 2 PULSE BY PULSE EP_PWM
GND
J6 IN OUTB 4
PW_EC1 OUTA PW_EC2
1 EP_FB + +
DR_INTF
1
2 PW_C1 LM1117-3.3/SOT223 PW_C2 IP_PWM
3 THEMAL_MON 104/DIP 22uF/16V 47uF/16V 104/DIP IP_REF IP_FB_ADC
4 FIL_IP_CHANGE1 IP_REF IP_FB_ADC FIL_FB_ADC
5 IP_FB(A) IP_FB(A) FIL_IP_CHANGE FIL_FB_ADC
/PREP IP_FB(A) FIL_A
FIL_FB /PREP FIL_A FIL_B
CON5 FIL_FB FIL_B
3-7.1 MAIN circuit
IP_PWM
3-7 Schematics Title
Manual ver1.7
Size
MAIN
Document Number Rev
Custom<Doc>
Date: Monday , October 16, 2006 Sheet 1 of 6
TP8
IP_FB 1
-12V
+12V
1
D35 LM358
4
1N4148 2 -
R34 4.7k 1
2
3 TP10
IP_FB(A)
+
1
1
7.5K 2K 15K
1N4148 1 2 1 3 1 2 (1J-5)
FIL_FB_ADC
2
2
+12V
2
R39
-12V
-12V 10K 1%
1
U10 +12V
4
8
OP07 R42
R41 2 - 5.6K U12
4.7K 6 1 2 R46 1 12
1 2 3 1 2 +IN1 VCC
FIL_FB
+
10K 2 8
R44 2 1 1 2 -IN1 C1 9
E1
7
1
2
1
5K 1% 4
DT R52 R53
+12VC25 MN104 7 1K 1K
2 1 IP_REF 1 2 15 GND 14
-IN2 VR
1
16 13
+IN2 OC
2
R55 U17B R54 R56 C26
8
10K 1% 2 1 1 2
1 2 5 TL494
35
IP_REF +
7 IP_CONT
6 - R59
1
LM358
1
C27 IP_FB 2 1
4
10K 1% 2 1
5.6K
2
1
-12V MN104 C29
+5V
U4B
14
2
14066_9 +5V
11 10 U4D
14
14066_9
8 9
12
6
/PREP
3-7.2 IP PWM circuit
Title
IP_PWM
Manual ver1.7
D37
8
1N4148
1
EP_FB 2 -
LM358
2
R66
4
2
D38 R R67
C35 MN104 5.6K
1N4148 2 1 1 2
1
2
1
MN104
+12V
EP_REF U15
1
1 12
1 2 +IN1 VCC
R68 2 8
5.6K -IN1 C1 9
E1 C1
2
1
VR3 2K C37 3
1 3 F.B.
BOX102
C2
2
5 11
CT C2 10
6 E2
RT
4
DT
+12V 7
GND R76
15 14
16 -IN2 VR 13
+IN2 OC +5V
R78 U8B
8
100K 1%
1 2 5 + TL494 200
EP_REF
36
7
1
1
6 -
+5V
2
1
C42 LM358 C40 + R74
1
14
1
2
100K 1%
2
2
4 3
-12V
2
1
R75
5
14066_9 2K
2
/HV_ON
3-7.3 EP PWM circuit
Title
Size
EP_PWM
Document Number Rev
Custom<Doc> <Rev Code>
Date: Friday , April 28, 2006 Sheet 5 of 6
Manual ver1.7
+5V
R14 C12 C13
15k 100u/16V 104
18
19
ISET
VCC
14 A
A 16 B
B 20 C
C 23 D
D 21 E
E 15 F
F 17 G
12 G 22 DP
LOAD DP
1
DATA
13 2
10
10
10
10
10
10
CLK DIG0
7
5
11
DIG1 6
A
DP
DP
DP
DP
DP
DP
C
F
DIG2 7
MAX7219 DIG3 3
DIG4 10
DIG5 5
DIG6 8
GND_A
GND_B
DIG7
CC
CC
CC
CC
CC
CC
CC
CC
CC
CC
CC
CC
3
8
3
8
3
8
3
8
3
8
3
8
U5A SEG1 uD_SR105K SEG2 uD_SR105K SEG3 uD_SR105K SEG4 uD_SR105K SEG5 uD_SR105K SEG6 uD_SR105K
4
U5
Name = I10294726
LED
READY 2 LD1 1 A
U6B U6A U6C
U6D
4
8
74LS04_0 X-RAY 2 LD2 1 B
+5V +5V +5V
14 7 14 7 14 7 +5V LED
37
14 7 WAIT 2 LD3 1 C
3
9
LED
M1 2 LD4 1 D
LED
LED_LOAD LED_DATA LED_CLK M2 2 LD5 1 E
D11 SW2 SKHJAB D12 SW3 SKHJAB D13 SW4 SKHJAB D14 SW5 SKHJAB
SW_R1 2 1 2 1 2 1 2 1 LED
M3 2 LD6 1 F
1N4148 KVDN 1N4148 KVUP 1N4148 MASDN 1N4148 MASUP
R15
2 1 LED
CONNECTING 2 LD11 1 G
33K
LED
D15 SW6 SKHJAB D16 SW7 SKHJAB D17 SW8 SKHJAB D18 SW9 SKHJAB ERROR 2 LD12 1 DP
SW_R2 2 1 2 1 2 1 2 1
1N4148 MEM1 1N4148 MEM2 1N4148 MEM3 1N4148 MEM4 M4 LED
R16 2 LD9 1 A
2 1
LED
M5 2 LD10 1 B
33K
SW12 SW13 LED
3-7.4 Display circuit
VCC
6 2
V- V+
MXC4
15
16
MXC3 MXC5
1uF/BOX 1uF/BOX
1uF
+3.3V
+3.3V
R80
U19 200/0805
SN75179
1
J11
38
RXD422 2 7
RXD422_CPU RX- RX-
VCC
R B 8 1
A RX+ RX+ 2
TXD422 TX+ 3
TXD422_CPU 3 6
GND
D Z TX- TX- 4
5
Y TX+
CON4
4
3-7.5 Communication circuit
200/0805
R84
Title
Manual ver1.7
Size
COM
Document Number Rev
A4 <Doc> <Rev Code>
Date: Tuesday , March 21, 2006 Sheet 6 of 6
+3.3V
1
RA3
C
390
7
6
5
4
3
2
PC1
1 4
1
2 3
LD39 LD40 LD41
TLP121
LED LED LED
TD62083 PC2
2
9 10 1 4
GND COM CN1
8 11 DR+5~12V_INPUT
7 I8 O8 12 READY_SW_OUTPUT 1
6 I7 O7 13 2 3 EXP_SW_OUTPUT 2
READY _SW_OUTPUT I6 O6 3
5 14 EXP_END_SIGNAL_OUTPUT
EXP_SW_OUTPUT I5 O5 4
4 15 TLP121
EXP_OK_INPUT
EXP_END_SIGNAL_OUTPUT I4 O4 5
3 16 DR_GND
P_P_ERR I3 O3 6
2 17 PC3
CHARG_ERR I2 O2
1 18 35312-06
THERMAL_ERROR I1 O1 1 4
U7
TP DR PLATE
2 3
2
TLP121
R22
PC4
1K/0805
4 1
EXP_OK_INPUT
1
R23
220
39
3 2
TLP121
JH2 JH3 JH5 JH6 JH7 JH8
1
1
1
1
+3.3V
JH1 JH4 JH13 JH9 JH10 JH11 JH12
R25
1
1
PC5
R26 1.5K/0805
1
3
Q1 2 470/0805 4 1
2N2222/SOT23
3-7.6 DR-Interface circuit
1
R31
10K/0805 3 2
TLP121
Title
Size
DR INTERFACE
Document Number Rev
Manual ver1.7
40
Manual ver1.7
C05 C21 71~80(3) 40 80 171.4 25.0 25.0 4.29
C06 C22 81~90(1) 0.4 10 171.4 30.0 30.0 5.14
C07 C23 81~90(2) 13 40 171.4 25.0 25.0 4.29
C08 C24 81~90(3) 50 80 171.4 20.0 20.0 3.43
C09 C25 91~100(1) 0.4 13 171.4 25.0 25.0 4.29
C10 C26 91~100(2) 16 50 171.4 20.0 20.0 3.43
C11 C27 91~100(3) 64 80 171.4 16.0 16.0 2.74
C12 C28 101~110(1) 0.4 32 171.4 18.0 18.0 3.09
C13 C29 101~110(2) 40 64 171.4 16.0 16.0 2.74
C14 C30 111~120(1) 0.4 20 171.4 16.0 16.0 2.74
C15 C31 111~120(2) 25 50 171.4 14.0 14.0 2.40
41
Manual ver1.7
42
Manual ver1.7
43
Manual ver1.7
40~50 40 0.4 100 11.4 2857.1 1600 2000
51~60 40 0.4 100 11.4 2857.1 2040 2400
61~70(1) 30 0.4 64 13.3 2133.3 1800 2100
61~70(2) 25 80 100 3200.0 4000.0 1500 1750
71~80(1) 30 0.4 32 13.3 1066.7 2100 2400
2.4kW
71~80(2) 25 40 80 1600.0 3200.0 1750 2000
81~90(1) 25 0.4 40 16.0 1600.0 2000 2250
81~90(2) 20 50 80 2500.0 4000.0 1600 1800
91~100(1) 20 0.4 50 20.0 2500.0 1800 2000
91~100(2) 16 64 80 4000.0 5000.0 1440 1600
40~50 60 0.4 100 6.7 1666.7 2400 3000
51~60 50 0.4 80 8.0 1600.0 2550 3000
61~70 40 0.4 100 10.0 2500.0 2440 2800
71~80(1) 40 0.4 32 10.0 800.0 2840 3200
3.2kW
71~80(2) 35 40 80 1142.9 2285.7 2485 2800
81~90(1) 30 0.4 80 13.3 2666.7 2430 2700
91~100(1) 30 0.4 50 13.3 1666.7 2730 3000
91~100(2) 25 64 80 2560.0 3200.0 2275 2500
44
Manual ver1.7
ERROR
No. Code Description Indication
2 010 standby FIL FB ERROR Excess from 200mA (10% of max. current, 2A)
3 011 standby EP FB ERROR Excess from 12KV (10% of max. voltage, 120KV)
4 012 standby IP FB ERROR Excess from 3.5mA (10% of MAX current, 35mA)
8 015 run EP FB or AC POWER LOW ERR Out of range from reference-10% or AC power line poor
45
Manual ver1.7
46
Portable manual