Professional Documents
Culture Documents
Cho, H. (2017) - A Disposable Microfluidic Device With A Reusable Magnetophoretic Functional Substrate For Isolation of Circulating Tumor Cells
Cho, H. (2017) - A Disposable Microfluidic Device With A Reusable Magnetophoretic Functional Substrate For Isolation of Circulating Tumor Cells
View Journal
Lab on aChip
Miniaturisation for chemistry, physics, biology, materials science and bioengineering
Accepted Manuscript
This article can be cited before page numbers have been issued, to do this please use: H. Cho, J. Kim, C.
Jeon and K. Han, Lab Chip, 2017, DOI: 10.1039/C7LC00925A.
Volume 16 Number 1 7 January 2016 Pages 1–218 This is an Accepted Manuscript, which has been through the
Royal Society of Chemistry peer review process and has been
accepted for publication.
Lab on aChip
Miniaturisation for chemistry, physics, biology, materials science and bioengineering Accepted Manuscripts are published online shortly after
www.rsc.org/loc
rsc.li/loc
Page 1 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
1
Department of Nanoscience and Engineering
Center for Nano Manufacturing, Inje University
Gimhae 50834, Republic of Korea
2
Department of Surgery
Gospel Hospital, Kosin University
Busan 49267, Republic of Korea
*Corresponding Author
1
Lab on a Chip Page 2 of 31
View Article Online
DOI: 10.1039/C7LC00925A
Abstract
polymer thin film. It consists of a disposable polymeric superstrate and a reusable functional
substrate and they are assembled simply using vacuum pressure. The disposable polymeric
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
microchannel. The reusable functional substrate generates an intricate energy field that can
penetrate the micrometer-thick polymer film into the microchannel and control microfluids.
This is the first report to introduce a silicone-coated release polyethylene terephthalate (PET)
thin film as a bonding layer on a microstructured PDMS replica. The bonding strength was
~600 kPa, which is the strongest among bonding methods of PDMS and PET polymer.
Additionally, accelerated tests for bond stability and leakage demonstrated that the silicone-
coated release PET film can form a very robust bond with PDMS. To demonstrate the
device format and was evaluated for isolating circulating tumor cells (CTCs) from patients
2
Page 3 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
1. INTRODUCTION
Microfluidic devices usually include special functions for mixing, reaction, pumping, and
filtering, which are crucial for manipulating and analyzing fluids. The key functions are
processes and expensive materials. For example, as the key functional compartments, a
tiny magnet array for magnetic cell capture,4 and a thermocycler for micro-PCR.5
Additionally, a single-crystal piezoelectric wafer was used for generating a surface acoustic
materials allow microfluidic devices to be equipped with useful functionalities, they could
have been fabricated with polymeric materials8-10 and paper11 due to the advantages of a wide
range of mastering and prototyping methods with low-cost and high-volume manufacturing
methods.12, 13
Most polymeric microfluidic devices have been manufactured by limited
ablation,18 and soft lithography,19 which can only provide restricted structures and
functionalities. For this reason, a lot of efficiency and precision microfluidic devices are still
realized based on glass and semiconductor materials because of their compatibility with most
microfabrication processes.
3
Lab on a Chip Page 4 of 31
View Article Online
DOI: 10.1039/C7LC00925A
compartments, which work together with disposable microfluidic devices. The reusable
functional compartments generated external energy fields for manipulating fluid substances
in microchannel of the disposable device. The disposable microfluidic devices were usually
substrate was used as a thin layer for penetration of the external energy fields, such as
showed that the magnitude of the external energy field is dramatically decreased with the
distance from the energy source, thereby declining the force of the energy field transmitted
into the microchannel. Since the thickness of the used glass substrate was in a range of
microscaled space. Even though the force can be effectively transferred into the microchannel
by thinning the glass substrate, it could be easily broken with only a slight impact, making it
provides the main functional compartments, which may be structured in a complex manner
and fabricated using complicated fabrication processes with expensive materials. The
disposable superstrate and the reusable substrate can be assembled and disassembled simply
using vacuum pressure. The silicone-coated release PET film used here can also form strong
bonds with PDMS, and the energy fields, such as acoustic, electric, magnetic, and thermal,
can efficiently transmit through it. Silicone-coated release polymer films are widely used in a
range of shielding and packaging products, printing, protecting adhesive films, and electronic
4
Page 5 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
and medical applications.23 They are available with various polymer materials and
isolating circulating tumor cells (the ‘assembly-disposable CTC-µChip’) and evaluated its
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
5
Lab on a Chip Page 6 of 31
View Article Online
DOI: 10.1039/C7LC00925A
See the Supporting Information for detailed descriptions of the following methods: (a)
inlets and two outlets, and a vacuum hole leading to a vacuum trench surrounding the
microchannel. The reusable substrate includes a ferromagnetic wire array, which is inlaid into
the substrate and placed at an angle (θ = 5.7°) to the direction of flow. The disposable
superstrate and the reusable substrate can be assembled and disassembled simply using
vacuum contact, which strongly attaches them and removes the air gap between the PET film
After vacuum assembly, nucleated blood cells mixed with magnetic nanobeads and
phosphate-buffered saline (PBS) with 0.2% bovine serum albumin (BSA) were respectively
injected through the sample and buffer inlets, each 500 µm in width. To enhance the purity of
isolated CTCs, the widths of the waste and CTC outlets were designed to be 800 and 200 µm,
respectively. To generate a high-gradient magnetic field, the width and thickness of the
ferromagnetic wire were designed to be 70 and 40 µm, respectively. Detailed design of the
When a uniform external magnetic field is applied to the ferromagnetic wire, the
external magnetic field is deformed near the wire, generating a high-gradient magnetic field
over the whole area of the microchannel. Due to the vacuum contact between the superstrate
6
Page 7 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
and the substrate, the gradient magnetic field can effectively penetrate the micrometer-thick
PET film into the microchannel (Fig. 1(b)). Then, CTCs with bound magnetic nanobeads
passing over the wire experience magnetic force Fm and hydrodynamic drag force Fd, which
generates a lateral magnetic force, Fl, on CTCs (Fig. 1(c)). Consequently, with a uniform
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
external magnetic field, CTCs with bound magnetic nanobeads are forced laterally and flow
7
Superstrate
Inlets
Vacuum hole
Microchannel
Vacuum trench
Outlets
Vacuum Inlets Outlets
Substrate assembly
(a)
Vacuum
trench PDMS
Superstrate
CTC
Vacuum
SU-8 contact
Substrate
Glass
Drag force (Fd) Magnetic force (Fm) External magnetic field (H0)
(b)
polymeric superstrate and a reusable substrate. (b) A cross-sectional view of the gradient
magnetic field, generated from a ferromagnetic wire array inlaid into the reusable substrate and
penetrating the micrometer-thick polymer thin film into the microchannel. (c) Working
principle of the assembly-disposable CTC-µChip for isolating CTCs from nucleated blood
8
Page 9 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
cells.
silicone-coated release PET film and a microstructured PDMS replica. An SU-8 mold on a
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
glass master was first fabricated to make the PDMS replica, including the microchannel and
spun and patterned for creating a mold for the microchannel on the glass master, along with
an evaporated 1000 Å Cr layer to increase the adhesion between the SU-8 and the glass
master. Then, an acrylic square bar of 2 × 2 mm2 was bonded adhesively on the glass master
to define the vacuum trench (Fig. S1(a)). A polymer mold frame, fabricated by
stereolithography (ViperSI2; 3D Systems), was used to pour the liquid PDMS. The SU-8
mold was completed by assembling the glass master and the polymer mold frame. Liquid-
phase PDMS, prepared by mixing resin and curing agent in a 10 : 1 ratio (Sylgard 184; Dow
Corning), was poured into the SU-8 mold and cured for 1 h at 75°C in an oven (Fig. S1(b)).
After peeling the PDMS replica off the SU-8 mold, the inlet and outlet reservoirs and the
vacuum hole of the PDMS replica were generated using a punch, 1.5 mm in diameter. Then,
the PDMS replica and the 12-µm-thick silicone-coated release PET film (5 g release force,
Shanghai Guangtai Adhesive Products Co.) were bonded together after oxygen plasma
treatment for 60 s at 6.8 W RF power (PDC-32G-2; Harrick Plasma) (Figs. 2(a) and S1(c)).
Figure 2(b) shows enlarged cross-sectional views of the microchannel made by the
The fabrication process for the reusable substrate used a 0.7-mm-thick glass slide
(Borofloat33 Pyrex; Schott). A Ti/Cu/Cr seed layer was first electron beam-evaporated on the
glass slide. A SU-8 3050 photoresist was spun and patterned to create 40-µm-thick
micromolds for ferromagnetic wires (Fig. S1(d)). Ferromagnetic permalloy (Ni0.8Fe0.2) wires
9
Lab on a Chip Page 10 of 31
View Article Online
DOI: 10.1039/C7LC00925A
were then electroplated on the glass slide (Fig. S1(e)). The ferromagnetic wire array, 40 µm
in thickness, was subsequently formed by a chemical and mechanical polishing technique and
was first placed on the center of two stacked neodymium-iron-boron (Nd-Fe-B) permanent
vacuum pressure of -50 kPa was applied through the vacuum hole to tightly attach the
superstrate and substrate (Fig. 2(d)). The vacuum contact also removed the air gap between
the PET thin film and the ferromagnetic wire array, so the high-gradient magnetic field
generated from the ferromagnetic wire array can efficiently penetrate the micrometer-thick
PET film into the microchannel and finely manipulate magnetic particles in fluids.
10
Page 11 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
(a)
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
(c)
(d)
Fig. 2 (a) A photograph of the fabricated disposable polymeric superstrate and (b) enlarged
cross-sectional views of the microchannel. (c) A photograph of the fabricated reusable
substrate, including the inlaid ferromagnetic wire array. (d) Vacuum-assembled assembly-
disposable CTC-µChip, placed on two stacked permanent magnets. (i) WBCs (red), spiked
with MCF7 breast cancer cell lines (green), mixed with magnetic nanobeads and PBS
solution are injected through sample and buffer inlets, respectively. (ii) With a uniform
external magnetic field, MCF7 cells bound with magnetic nanobeads are forced laterally and
(iii) flow into CTC outlet, while WBCs flow into the waste outlet.
11
Lab on a Chip Page 12 of 31
View Article Online
DOI: 10.1039/C7LC00925A
Healthy human peripheral blood samples were collected in Vacutainer tubes (G-Tube, Green
Cross) containing the anticoagulant EDTA and processed within 12 h. For the analytical
evaluation, MCF7 breast cancer cell lines were first stained with a membrane-permeable
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
green fluorescent nucleic acid dye (SYTO 13, Invitrogen). The fluorescently stained MCF7
Then, 10 µl of the concentrated solution was added to 5 ml of healthy whole blood in a 50-ml
conical tube, thereby spiking ~100 MCF7 cells into the blood. Patient blood samples were
drawn from patients (n = 20) with breast cancer using a protocol approved by the institutional
review board of Kosin University Gospel Hospital and informed consent was obtained from
the patients. The characteristics of the peripheral blood donors are summarized in Supporting
Information Table S2. All blood samples from cancer patients were first filtered using a 40-
µm grid strainer (Cell Strainer, SPL Life Sciences Co.) to eliminate matted substances and
debris.
Nucleated cells in prepared 5-ml blood samples were first extracted by a density
gradient centrifugation (700×g, 30 min) using 1.119 g ml-1 Ficoll solution (Histopaque-1119,
Sigma-Aldrich) and transferred into 10 ml of ice-cold PBS with 0.2% BSA in a 50-ml conical
tube. The extracted nucleated cells were washed by centrifugation (300×g, 10 min) and
resuspended in 1 ml of ice-cold PBS with 0.2% BSA in a 1.5-ml microcentrifuge tube. The
nucleated cells were washed once again and suspended in 200 µl of ice-cold PBS with 0.2%
BSA. The nucleated cells in 200 µl of medium were then mixed with anti-EpCAM antibodies
and magnetic nanobeads in sequence and incubated on ice for 60 and 90 min, respectively,
finally prepared by dilution with a 4-fold volume (800 µl) of ice-cold PBS with 0.2% BSA.
CTCs isolated from blood of patients was fixed with 100 µl of 4% paraformaldehyde
12
Page 13 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
for 5 min and subsequently permeabilized with 100 µl of 0.2% Triton X-100 (AMRESCO)
for 5 min. Then, CTCs were stained using a membrane-permeable nucleic acid fluorescent
dye (DAPI, Invitrogen) for nuclei, anti-cytokeratin-Alexa 488 (eBioscience) antibodies for
epithelial cells, and anti-CD45-Alexa 647 (Biolegend) antibodies for normal blood cells for
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
30 min. The fluorescently stained CTCs were identified and counted using a confocal
13
Lab on a Chip Page 14 of 31
View Article Online
DOI: 10.1039/C7LC00925A
The bond strength between the silicone-coated release PET film and PDMS was measured
from normal stress using a materials testing machine (LR10K Plus, Ametek) (Fig. 3(a)).
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
Because the thickness of the PET film is 12 µm, it was first attached on an acrylic round bar
mm in diameter and 100 mm in length). The non-silicone coated side of the PET film was
permanently bonded on a single end of the acrylic round bar using a superglue. Then, the
PDMS round bar and the silicone-coated release PET film on the acrylic round bar were
bonded using oxygen plasma treatment. The two round bar assembly was fixed in the
materials testing machine and pulled apart at a speed of 20 mm/min to measure the interfacial
The bond strength was defined as the force required to separate the two round bars.
After the bond strength test, PDMS residue remained on the PET film surface (Fig. 3(b)).
This result indicates that PDMS permanently bonded to the silicone-coated release PET film.
The bond strength between PDMS and the silicone-coated release PET film was measured at
617±106 kPa (Fig. 3(c)). Then, 10 sets of identical tests were conducted for statistical
evaluation of the bond strength. The average of the bond strength was higher than that
between PDMS and a PDMS linker-coated PET film (514.0±255 kPa),24 which has shown
the highest bond strength among bonding methods for PDMS and PET polymer. Even though,
a statistical analysis (Table S1) indicates that the difference between the bond strengths of the
silicone-coated PET film (617±106 kPa) and the PDMS linker-coated PET film (514.0±255
kPa) to the PDMS is not meaningful. Indeed, the bond strength of more than 600 kPa is a
very high value in comparison with the driving pressure for fluid flow of the assembly-
14
Page 15 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
(b)
800
700
600
Bond strength (kPa)
500
400
300
200
100
0
Silicone PDMS–linker
(c) coated PET coated PET
Fig. 3 Photographs of (a) instrument set-up for bond strength test and (b) PDMS residue
remaining on the silicone-coated release PET film. (c) Measured bond strengths of PDMS
with the silicone-coated release PET film and PDMS-linker coated PET film. The error bars
15
Lab on a Chip Page 16 of 31
View Article Online
DOI: 10.1039/C7LC00925A
To measure the maximum pressure the bond between PDMS and the silicone-coated release
PET thin film could withstand, air pressure from 0–700 kPa was applied to the microchannel
through a gas regulator (Fig. 4(a)). Then, a vacuum pressure of -50 kPa was used for vacuum
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
assembly of the disposable superstrate and the reusable substrate. To confirm that the air
(XGZP6847001MPG, CFSensor) were connected to monitor pressure at each inlet and outlet.
The burst test was conducted at 100-kPa increases for 30 min each, which is the operating
time of the CTC-µChip, and identical burst tests were repeated using three devices (Fig. 4(b)
and Supporting Video 1). The results demonstrated that the assembly-disposable CTC-µChip
can withstand at air pressure up to ~600 kPa. Thus, the burst pressure is much higher than the
10–20 kPa normally required to drive fluid flow in the assembly-disposable CTC-µChip. The
measured burst pressure was also similar to the value of 617 kPa obtained in the bond
strength test. Also similar to the result of the bond strength test, PDMS residue remained on
the PET film after the burst test, again indicating that PDMS permanently bonds on the
The hydrolytic stability of the bonding between the PDMS and the silicone-coated
release PET film was measured over 0–28 days, during which the disposable polymeric
superstrate was immersed in PBS solution. As shown in Fig. 4(b), bonding was maintained
until at least 28 days, although the typical intended operating time of the assembly-disposable
CTC-µChip is only ~30 min. The results also demonstrated that the proposed silicone-coated
release PET film can form strong and stable bonds with PDMS and is useful for realizing
disposable microfluidic devices. The high bond strength and hydrolytic stability will allow
disposable microfluidic devices based on the silicone-coated release polymer thin film to be
used in other applications, such as on-chip cell culture and PCR functionalities.
16
Page 17 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
the physical bonding stability of PDMS and the silicone-coated release PET film. Red and
blue colored solutions were injected into sample and buffer inlets at several flow rates, 20, 60,
100, 200, and 300 ml h-1, for 10 min (see Supporting Video 2) so 10, 30, 50, 100, and 150
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
times higher, respectively, than the ordinary injection flow rate of 2 ml h-1. The assembly-
injection syringe pumps (Legato 100, KD Scientific) were stopped due to high repulsive
forces. This result confirmed again that a strong bond was formed between PDMS and the
17
Lab on a Chip Page 18 of 31
View Article Online
DOI: 10.1039/C7LC00925A
Vacuum (−
−50 kPa) (mm)
2
2
Air pressure
Pressure 4 (0~700 kPa) 4 Pressure
sensor 1 0.5 0.8 sensor 3
5 1 5
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
Pressure 5 22 Pressure
sensor 2 0.5 0.2 sensor 4
800
700
Burst pressure (kPa)
600
500
400
300
200
100
0
0 7 14 21 28
(b) Time (days)
Fig. 4 (a) Illustration of instrument set-up for burst test of the assembly-disposable CTC-
µChip. (b) Measured burst pressure of the silicone-coated release PET film and PDMS of the
disposable polymeric superstrate, immersed in PBS solution over 0–28 days after bonding the
18
Page 19 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
3.3. Analytical evaluation using cancer cells spiked into whole blood
thicknesses, from 12–30 µm, were evaluated using nucleated blood cells, prepared by Ficoll
density gradient centrifugation of 5 ml whole blood spiked with ~100 MCF7 breast cancer
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
cell lines. For the analytical evaluation, the CTC-µChip was placed on a stack of two
horizontal direction to the microchannel. During injection of the nucleated blood cells, MCF7
cells separated into the CTC outlet were counted using a fluorescence microscope and the
recovery rate of spiked MCF7 cells was evaluated (Fig. 5(a)). The analytical evaluation
demonstrated that the assembly-disposable CTC-µChip with 12-µm-thick PET film could
recover > 90% of the MCF7 cells from spiked blood samples at a sample flow rate of 1–4 ml
h-1. The result demonstrated that the recovery rate decreased as the thickness of the PET film
increased. It also corresponds with the theoretical analysis (Fig. S2(a)) that the lateral
decreased as the thickness of the PET film increased. As the flow rate increased, more MCF7
cells easily passed over the ferromagnetic wires, as shown in Fig. S2(b), and thus the
recovery rate decreased. Interestingly, in the condition of 12-µm-thick PET film, the recovery
rate at the flow rate of 1 ml h-1 is slightly lower than that at the flow rate of 2 ml h-1. It might
be because some MCF7 cells adhere to the ferromagnetic wires due to the strong magnetic
force in this condition. The majority of MCF7 cells were laterally forced and separated by the
first and second ferromagnetic wires. However, a few MCF7 cells were not strongly
influenced by the magnetic forces of the forward wires due to their initial levitation height in
the microchannel.
The number of contaminating WBCs ranged from 153–741 cells per 5-ml blood
sample (Fig. 5(b)). The average contaminating number of WBCs was 386.6 cells and thus the
19
Lab on a Chip Page 20 of 31
View Article Online
DOI: 10.1039/C7LC00925A
average WBC depletion rate could be calculated as 64,666-fold (4.81 log), on the assumption
there are 5×106 WBCs per milliliter of blood. Figure 5(b) also shows that the contaminating
WBCs were reduced as the flow rate increased. From the measured results, the purity of
MCF7 cells isolated by the assembly-disposable CTC-µChip could be calculated as the ratio
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
of the number of isolated MCF7 cells to the total number of isolated nucleated cells, which
bonding and the thickness of PDMS spun-layer right on the ferromagnetic wire substrate was
~10 µm. As shown in Figs. S3(a) and S3(b), the x-directional and z-directional magnetic
forces generated from the assembly-disposable CTC-µChip with a 12-µm-thick PET film
were similar to those generated with the non-disposable CTC-µChip with a 10-µm-thick
PDMS spun-layer. Similar to the experimental results with the non-disposable CTC-µChip,
the assembly-disposable CTC-µChip with a 12-µm-thick PET film yielded a ~90% recovery
rate. This finding suggests that both PET thin film and PDMS spun-layer are favorable for the
magnetic field to penetrate into the microchannel, because PET and PDMS are non-magnetic
materials. In particular, the non-magnetic properties of the PET film are compatible with the
high-gradient magnetic separation (HGMS) method used here, allowing the realization of a
performance was compared with that of EasySep Human EpCAM Positive Selection Kit
cells, the number of contaminating WBCs and isolation purity of MCF7 cells were
respectively measured as 60.7%, 18,756 WBCs and 0.32%, as shown in Fig. S4. Although
the CTC-isolation procedure from the prepared nucleated blood cells takes only 15 min,
20
Page 21 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
which is faster than 30 min of the assembly-disposable CTC-µChip, the results showed that
conventional MACS method due to the precise magnetophoretic isolation of target cells in
100 times that (0.32%) by the conventional MACS method. Since high purity of isolated
21
Lab on a Chip Page 22 of 31
View Article Online
DOI: 10.1039/C7LC00925A
-1
1 ml h
100 -1
2 ml h
-1
4 ml h
80
Recovery rate (%)
60
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
20
0
12 µm 19 µm 25 µm 30 µm
-1
1 ml h
-1 5
800 2 ml h 2.0x10
The number of contaminated WBCs
-1
4 ml h
700 WBC depletion rate
5
600 1.5x10
5
400 1.0x10
300
4
200 5.0x10
100
0 0.0
12 µm 19 µm 25 µm 30 µm
22
Page 23 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
60
-1
50 1 ml h
-1
2 ml h
-1
4 ml h
40
Purity (%)
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
30
10
0
12 µm 19 µm 25 µm 30 µm
Fig. 5 (a) Recovery rates of MCF7 cells, (b) the number of contaminating WBCs, and (c) the
purities of MCF7 cells, isolated with the assembly-disposable CTC-µChips with various
thicknesses of the silicone-coated release PET film and flow rates. Breast cancer cell lines
(~100 MCF7 cells) were spiked into 5 ml of human peripheral blood and tagged by anti-
EpCAM magnetic nanobeads and isolated at sample and buffer flow rates from 1–4 ml h-1
with an external magnetic flux of 0.2 T. The error bars represent one standard deviation of
23
Lab on a Chip Page 24 of 31
View Article Online
DOI: 10.1039/C7LC00925A
To isolate CTCs from patients with breast cancer, 5-ml samples of peripheral blood from
patients (n = 20) were collected in vacuum tubes. The thinnest film, a silicone-coated release
12-µm-thick PET film, was used to obtain the highest performance. The sample and the
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
buffer flow rates were 2 ml h-1 each, with the separation process taking < 30 min. The number
(33.6±29.4; mean ± SD) CTCs per 5 ml of blood from patients with breast cancer (n = 20), as
shown in Fig. 6(b) and Table S2. Neither of the two healthy donors had any identifiable CTCs.
The number (33.6±29.4) of CTCs isolated from 5-ml peripheral blood samples of patients can
be compared with previous experimental results3 of the number (4±2) of CTCs isolated from
200-µl peripheral blood samples of patients with breast cancer. Because the quantity of blood
used in this study (5 ml) was 25 times more than the 200 µl used previously, the previously
isolated 4 CTCs corresponds to 100 CTCs here. Thus, the average number of isolated 33.6
CTCs in this study is less than that reported in the previous study. This might have been
caused by use of the 40-µm grid strainer, RBC removal, and washing steps during blood
sample preparation. Moreover, because the present and previous results were obtained from
only 20 and 3 patients, respectively, larger clinical trials are needed to obtain more reliable
results. The number of CTCs would seem to be correlated with the stage of the cancer (Fig.
6(b)). Many studies25-28 have examined the possibility of such a relationship. Interestingly,
the numbers of CTCs isolated from two patients (nos. 19 and 20) under treatment were
markedly reduced.
The average purity of CTCs isolated from patients, defined as the ratio of
was 9.5%. This means that the WBC depletion rate of the assembly-disposable CTC-µChip
was ~71,429-fold (4.85 log) in the clinical experiments. The WBC depletion rate obtained
24
Page 25 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
using clinical samples was similar, at 64,666-fold (4.81 log), measured from the biomimetic
blood samples spiked with MCF7 breast cancer cell lines. Because the WBC depletion rate of
the assembly-disposable CTC-µChip, based on a positive separation method using only anti-
EpCAM, is almost constant, the purity of CTCs tends to increase with the stage of the cancer.
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
For the sample injection flow rate of 2 ml h-1, the Reynolds number in the
important for high purity because it allows for a multi-stream fluid in which CTCs are
selectively transported across the sample and buffer stream boundary into the CTC outlet.
depletion rate of the assembly-disposable CTC-µChip to be 4.85 log, which is much higher
than that of other approaches.30-32 Indeed, the high purity of 9.5% of the assembly-disposable
CTC-µChip will make it an ideal tool for high-precision genetic analyses, such as next-
To determine whether CTCs isolated from patients were suitable for subsequent
genetic analysis, RT-PCR analyses were conducted for detecting EpCAM and keratin 19
KRT19 transcripts amplified from CTCs isolated from patients (nos. 5, 6, 12) and a healthy
donor. EpCAM transcript was detected in 12 (63.2%) of 19 patients with breast cancer (Table
S2). However, among the 20 patients, the KRT19 transcript could be detected in 5 (25.0%),
meaning that the KRT19 transcript is not always expressed in CTCs. This result may support
the explanation that the phenotype of CTCs could be changed due to the epithelial-
25
Lab on a Chip Page 26 of 31
View Article Online
DOI: 10.1039/C7LC00925A
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
100
The Number of CTCs per 5 ml Blood
80
60
40
20
0
P1 P2 P3 P4 P5 P6 P7 P8 P9 P10 P11 P12 P13 P14 P15 P16 P17 P18 P19 P20
Fig. 6 (a) Magnified images of fluorescently stained CTCs isolated from peripheral blood of
patients with breast cancer. (b) Frequency of CTCs per 5 ml of peripheral blood from patients
26
Page 27 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
(287 bp) transcripts using CTCs isolated from peripheral blood of patients (nos. 5, 6, 12) with
breast cancer. Lanes 1 and 14, 100-bp DNA size markers; lanes from 2–10, β-actin, EpCAM,
and KRT19 amplified using CTCs isolated from peripheral blood of patients (nos. 5, 6, 12)
with breast cancer; lanes from 11–13, controls using CTCs isolated from peripheral blood of
a healthy donor.
27
Lab on a Chip Page 28 of 31
View Article Online
DOI: 10.1039/C7LC00925A
4. CONCLUSIONS
release polymer thin film. The disposable format is an important feature for medical and
biological microfluidic devices. In this study, as the test vehicle, a lateral magnetophoretic
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
microseparator for isolating CTCs (CTC-µChip) was developed in the proposed assembly-
and a reusable substrate. The disposable superstrate was fabricated simply by oxygen plasma
bonding of a microstructured PDMS replica and a silicone-coated release PET film and does
not include any complex structures in the microchannel. Thus, no clogging problems
occurred while nucleated blood cells flowed through the microchannel for isolating CTCs.
The analytical and clinical evaluations demonstrated that the assembly-disposable CTC-
µChip is a practical device to isolate CTCs from cancer patients with high performance.
Typically, polymers have superior mechanical and electrical properties, making them
readily compatible with microfluidic devices. Because various polymeric materials can be
used for silicone-coated release films, there are many choices depending on the application of
used for various functionalities based on energy fields, such as acoustic, electric, magnetic,
and thermal fields, which can transmit through the polymer thin film to manipulate target
device can be realized simply with multiple functions, it could be further advanced as a
28
Page 29 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
ACKNOWLEDGMENTS
REFERENCES
29
Lab on a Chip Page 30 of 31
View Article Online
DOI: 10.1039/C7LC00925A
14, 3790-3799.
17. R. Schirhagl, E. W. Hall, I. Fuereder and R. N. Zare, Analyst, 2012, 137, 1495-1499.
18. S. M. Grist, N. Oyunerdene, J. Flueckiger, J. Kim, P. C. Wong, L. Chrostowski and K.
C. Cheung, Analyst, 2014, 139, 5718-5727.
19. G. M. Whitesides, E. Ostuni, S. Takayama, X. Jiang and D. E. Ingber, Annu Rev
Published on 27 October 2017. Downloaded by Kings College London on 31/10/2017 10:44:44.
30
Page 31 of 31 Lab on a Chip
View Article Online
DOI: 10.1039/C7LC00925A
Parce, Annual Review of Biophysics and Biomolecular Structure, 2000, 29, 155-181.
31