Professional Documents
Culture Documents
GPE01006.004 ADVIA Chem Systems Operator Training Manual
GPE01006.004 ADVIA Chem Systems Operator Training Manual
post training.
GPE01006.004
GPE01006.004
CLASSROOM SAFETY REQUIREMENTS
In this classroom, you will be using materials and/or specimens that are potentially
hazardous. To that end, we at Bayer have requirements for Personal Protective
Equipment use that we strongly urge you to follow.
• Eating, drinking, storage of foods, the application of cosmetics or the handling of contact
lenses are not permitted
• When gloves are removed, they must be placed in the biohazard container provided.
Hands should be washed after gloves are removed
• After the use of any sharps, these items must be placed in the sharps disposal container
provided
• Keyboards are clearly marked as “Gloves Only” or “No Gloves” and should be used
accordingly
• Gloves and safety glasses are not required in the lecture area
• The lecture area may not be used to store or place reagents, calibrators, controls, waste
material or specimens derived from human blood or urine
• There is one sink in this classroom. It is not used for the disposal of any biohazard
material. It may be used for handwashing.
GPE01006.004
GPE01006.004
ADVIA ® Chemistry Operator
Training Objectives
At the conclusion of the day indicated below, the trainee will be able to:
DAY 1 DAY 2
___ Identify major components of the system ___ Perform New Start
___ Identify software screens used daily for routine ___ Practice Daily Maintenance
operations
___ Identify the ISE Hydraulic pathway
___ Identify system reagents and working solutions
___ Perform ISE Calibration
___ Practice Daily Maintenance with Wash 2
___ Perform Photometric Calibration
DAY 3 DAY 4
___ Locate calibration/ QC setup screens ___ Describe how a lamp is changed
___ Identify causes/corrections for ISE calibration ___ Perform LAMP ENERGY check procedure
failures
___ Perform & interpret CELL BLANK procedure
___ Process QC samples
___ Describe how to replace a probe
___ Use Daily Precision Control window to view
results ___ Perform a hard reboot
___ Perform sample run for troubleshooting ___ Perform Daily Operation: Day in the Lab
Trainee’s Signature:_______________________________________________________
GPE01006.004
GPE01006.004
ADVIA® Chemistry Operator
Training Course Agenda
Day1
12:00-1:00 Lunch
Continue exercise
• Review results
• Schedule a rerun
GPE01006.004
Day 2
ISE Module
• Hydraulics presentation
• Perform ISE Calibration Exercise
Photometric Calibration
• Discuss RBL/ photometric calibration
• Perform Calibration Operations Exercise
for single point and multipoint assays
12:00-1:00 Lunch
3:00-3:15 Break
ISE Maintenance
• Perform ISE Maintenance Exercise
• Review
GPE01006.004
Day 3
ISE Troubleshooting
• Perform ISE Troubleshooting Exercise
12:00-1:00 Lunch
Continue exercise
• Discuss “bugs”
3:00-3:15 Break
Day 4
11:30-12:30 Lunch
GPE01006.004
GPE01006.004
System Overview
Table of Contents
Operating Principle…………………………………………….. .3
System Components ……………………………………………4
Top View ………………………………………………………… 5
Sample Handling………………………………………………... 6
Sample Tray…………………………………………………….. 7
Sample Probe………………………………………………….. 9
Reaction Tray…………………………………………………....13
Reagent Handling…………………………………………….... 15
Reagent Trays………………………………………………….. 16
Reagent Probes………………………………………………... 18
Spectrophotometer……………………………………………. .20
Wash Mechanism (WUD)…………………………………….. .21
System Fluids (Front View)…………………………………….23
Hydraulics………………………………………………………..25
ISE Module………………………………………………...…….27
Display Panel…………………………………………………....28
Rear View………………………………………………………. 29
Computer…………………………………………………….…. 30
System Specifications & Acronyms..………………………….31
GPE 01006.004 1
GPE 01006.004 2
Operating principle
Illustration of the ADVIA® 1200 (top view)
This sequence summarizes a photometric analysis on the ADVIA 1200 Chemistry System:
1 The system prepares a reaction cuvette for analysis by passing it through the wash
station.
2 The probes are cleaned with probe wash 1 and 2.
3 The first reagent (R1) for a test is aspirated from reagent tray 1 (RTT1) and dispensed by
reagent probe1 (RPP1) into the RRV cuvettes.
4 Sample is aspirated from the sample tray and dispensed into the cuvette that contains
the R1 reagent. The sample and R1 reagent are mixed by reagent mixer 1 (MIX-1).
If a second reagent is required, the reagent probe (RPP2) dispenses the R2 reagent to
the same reaction cuvette and the solution is mixed again (MIX-2).
5 The reaction takes place for the amount of time designated in the assay. Concentration
data is obtained by the spectrophotometer every 13.5 seconds. For each measurement,
the RRV moves the cuvettes in front of the spectrophotometer.
Cell blank measurements are performed at each wavelength.
6 The results of the analysis are printed; they can also be viewed in the system.
7 The RRV cuvettes are washed when the measurement is complete.
GPE 01006.004 3
GPE 01006.004 4
GPE 01006.004 5
GPE 01006.004 6
GPE 01006.004 7
Sample tray operation
Initialization: The tray rotates clockwise until STT position 1 is in the aspiration position.
After you start a run, the sample tray can move in either direction to bring a sample to the
aspiration position.
If samples are analyzed by barcode, each sample detected by the barcode reader stops (in
turn) in the aspiration position. If samples are analyzed by position number, they move by
position number to the aspiration position. Regardless of the mode, a workorder for the
sample must be entered, or it will not be processed.
Monitor the progress of samples on the sample tray in the Test Result Monitor window.
If an assay is not running, you can operate the unit manually from the Manual Operation
window.
Container types
You place sample containers defined in the Order Entry window in the STT and CTT.
Usually you use 5 mL, 7 mL, or 10 mL collection tubes or 2 mL sample cups, which are
placed in plastic holders (the holders are then placed in tray positions).
The dead volume for the collection tubes is 200 µL, and the dead volume for the sample
cups is 50 µL.
Collection tubes often have barcode labels which can be read by the barcode reader, if
placed in the STT. However, you cannot use barcodes with sample cups
GPE 01006.004 8
Sampling mechanisms
The sample probe (SPP) aspirates sample from the sample tray (STT) and dispenses it into
the reaction tray (RRV) cuvettes for analysis, according to specified conditions. The
aspiration and dispensing functions are handled by the sampling pump (SP).
GPE 01006.004 9
Primary sampling - operation
Initialization: The SPP initializes by moving up and over to the RRV cuvette, then moves
back to stop at the wash port. The pumps stop when the SPP is positioned over the RRV
cuvettes.
Each sampling cycle consists of the following:
The SPP moves to the aspiration position of the STT and aspirates the sample. The SPP
moves back to the wash port, where deionized water washes the SPP’s outside surface.
The SPP moves to the RRV and dispenses the sample into a cuvette. The SPP returns to
the wash port, where the inside and outside of the SPP are washed.
After the sample is dispensed, the SPP moves back to the wash port, where its inside is
washed. The sample probe valve (SPEV1) opens, allowing the SRWP to send degassed
water through the SPP’s inside. The SPP is now ready for another cycle.
If an assay is not running, you can operate the unit manually from the Manual Operation
window.
GPE 01006.004 10
GPE 01006.004 11
Reaction tray mixer operation
When initialized, the mixers move to their mixer positions, then go to the down position. If a
mixer is already at the mixer position, it is raised, then lowered.
During each cycle of the RRV, the following steps occur:
NOTE
This sequence applies to each mixer; the mixers used depend on the analysis conditions.
1 The mixer moves up and over to the wash port, where it is washed with deionized water.
During this period, the reaction tray mixer wash valve (MWEV1 or MWEV2) is open,
depending on the mixer being washed, and briefly the mixer rod is on.
2 The mixer is raised and moved to the mixer position, where it is lowered into the RRV
cuvette. The rod is turned on to mix the sample and reagent.
3 The mixer is raised from the cuvette and moved back to the wash port. The next cycle
begins.
If an assay is not running, you can operate the unit manually from the Manual
Operation window.
GPE 01006.004 12
Reaction tray
For each sample assay, the reagent probes (RPP1 and RPP2) dispense reagent into
cuvettes in the reaction tray (RRV). Then the sample probe (SPP) dispenses sample into
the cuvettes. The mixture is then stirred by the reaction mixers (MIXR1 and MIXR2),
producing the desired reaction.
For analysis, the reaction tray rotates the cuvette in front of the spectrophotometer, where
the cuvette’s absorbance is measured. After analysis, the cuvettes are washed by the
reaction washer (WUD).
Reaction bath
The reaction bath contains nonreactive oil, which keeps the temperature of the liquid in
reaction tray (RRV) cuvettes at a constant 37° C ± 0.5° C. The temperature is controlled by
a heater and a thermostat.
The oil supply in the reaction tank is controlled by two reaction bath oil level sensors:
GPE 01006.004 13
Reaction tray operation
When initialized, the RRV tray turns clockwise to set the home position, pauses and then
turns until cuvette 1 is in position for the first reagent (RTT1).
After you start a run, the reaction tray begins the following cycle:
1 R1 is dispensed into the RRV cuvette.
2 The reaction tray turns clockwise past 42 cuvettes and stops. The reaction tray turns
clockwise past another 37 cuvettes and stops again. Finally, the reaction tray turns
counterclockwise past 3 cuvettes and stops for sample or reagent addition.
3 Each reagent addition cycle takes 4.5 seconds and advances the RRV by 76 cuvettes
(42+37-3)
4 The cuvette moves to the mixer 1 (MIX 1) position and mixes the reagent and sample.
5 As the analysis progresses, the spectrophotometer measures the absorbance of the
liquid in each of the cuvettes that pass it (across all 14 wavelengths).
6 R1 and sample are dispensed into subsequent cuvettes and mixed, until all sampling is
completed and data has been collected.
When required by assay conditions, the RRV stops for dispensation and mixing of R2.
You can view measurement data in the Reaction Monitor window.
7 After the required measurements are complete, cuvettes used for assay are washed and
the cell blank absorbance is checked.
During each reaction tray movement, unused cuvettes are prepared for use by the
system. Cuvettes are washed, a cell blank reading is taken, and the cuvettes are dried.
If an assay is not running, you can operate the unit manually from the Manual
Operation window.
GPE 01006.004 14
GPE 01006.004 15
Reagent trays
Reagent trays 1 and 2 (RTT1 and RTT2) contain reagents used for assays and the
detergents used for daily washing and contamination prevention. The reagent probes (RPP1
and RPP2) aspirate the required reagent and dispense it into the reaction tray (RRV)
cuvettes for analysis.
Each tray has 45 positions. RTT1 contains the first reagent (R1); RTT2 contains the second
reagent (R2).
Any reagent container can be used for more than one test item; a test item may require
more than one reagent container.
Each reagent tray has a barcode reader (RBC-1 and RBC-2).
GPE 01006.004 16
Reagent tray operation
Initialization: The trays rotate clockwise until reagent bottle 1 is in the aspiration position.
After you start a run, the reagent trays move clockwise to position 1, then they rotate
clockwise or counterclockwise (whichever results in a smaller rotation) to move reagents to
the aspiration position. The number of trays and reagents used depends on the specified
assay conditions.
To check the reagent Volume and # Tests in a container, use the Reagent Inventory
window.
NOTE: USE THE ROTATION BUTTONS TO VIEW CONTAINERS ON THE TRAYS WITHOUT REMOVING THE
COVERS.
Container types
The reagent container size is specified on the reagent barcode label. The trays can hold 20,
40 or 70-ml reagent containers.
Reagent mechanisms
The reagent probes (RPP1 and RPP2) aspirate reagent from the reagent trays (RTT1 and
RTT2) and dispense it into reaction tray (RRV) cuvettes for analysis, according to specified
conditions. The aspiration and dispensing functions are handled by the reagent pumps (RP1
and RP2).
Initialization: Each RPP moves (in the up position) to the RRV cuvette, then stops above the
wash ports. The pumps stop when the reaction probes are positioned over the RRV
cuvettes.
Each reagent cycle consists of these steps:
1 The reagent probes move to the aspiration position of the reagent trays and aspirate the
reagent.
2 The probes move back to the wash ports, where deionized water washes their outside
surfaces.
3 The probes move to the RRV and dispenses reagent into a cuvette.
4 The probes return to the wash ports, where deionized water washes the insides and
outsides of the probes.
5 The reagent probes are now ready for another cycle.
If an assay is not running, you can operate the unit manually from the Manual
Operation window
GPE 01006.004 17
GPE 01006.004 18
GPE 01006.004 19
Spectrophotometer
The photometer then measures the absorbance based on the lamp energy and the optical
density of the cuvettes. This process is repeated every 13.5 seconds on an individual
cuvette.
The output energy of the halogen lamp is monitored during the cell blank check and after
each assay. The operator is alerted if the lamp performance is abnormal.
Use the Lamp Energy Monitor window to ensure that the halogen lamp is functioning
normally.
GPE 01006.004 20
Reaction tray wash mechanisms
The reaction washer (WUD) washes reaction tray (RRV) cuvettes after sample analysis is
complete, so they can be reused for the next analysis.
The WUD has seven nozzles, each performing a stage of the wash. Each nozzle works on a
different cuvette, so the WUD washes seven cuvettes simultaneously (the cuvettes are
being washed in different stages at the same time).
After a cuvette is washed by one nozzle, it moves to the next until washing is complete.
While the RRV rotates, the WUD is in the up position.
Some wash liquids are heated to approximately 37° C, to maintain the temperature of the
RRV cuvettes
When initialized, the WUD moves to the up position. If it is already up, it is lowered, then
raised.
To advance the cuvettes to the next wash nozzle, the RRV rotates a full-turn.
If an assay is not running, you can operate the unit manually from the Manual Operation
window.
GPE 01006.004 21
During each rotation, the nozzles operate as follows:
GPE 01006.004 22
GPE 01006.004 23
Components, Controls, and Indicators: Analyzer (front view)
1 ISE location
2 Power panel
3 ISE buffer bottle
4 Reaction bath oil bottle
5 Cuvette detergent bottle
6 Cuvette conditioner bottle
7 Sampling pump (SP)
8 Sampling and reagent wash pump (SRWP)
9 Reagent dispensing pump 1 (RP1)
10 Reagent dispensing pump 2 (RP2)
11 Fluidics control panel location
GPE 01006.004 24
GPE 01006.004 25
GPE 01006.004 26
ISE (electrolyte analyzer)
The ISE measures the amount of sodium (Na), potassium (K), and chloride (Cl) in serum or
urine samples through voltage measurement by ion-selective electrodes.
Sample aspirated by the sample probe (SPP) is used for the electrolyte analysis. The
electrolyte analysis uses buffer as reagent.
In a two-stage process, the buffer voltage is measured, then the sample voltage is
measured. The difference between these voltages, the reference voltage, and the
temperatures of the liquids determines the concentration of Na, Cl, and K in the sample
ISE operation
To ensure data accuracy during electrolyte analysis with the ISE, the undiluted sample from
the sample tray is used, and electrolyte sampling is always done before photometric
sampling.
GPE 01006.004 27
Display panel and power panel
GPE 01006.004 28
Rear View
GPE 01006.004 29
PC (front view)
PC (rear view)
1 PC power connector
2 Mouse connection
3 Keyboard connection
4 Communication port 1 (LIS-TDC)
5 Printer connection
6 LCD monitor connection
7 Analyzer ethernet connection
8 Communication port 2
9 Sleep ITF board potentiometer
GPE 01006.004 30
ADVIA 1200 Specifications
Item Description
Method
Process
Throughput rate
Photometric Maximum 800 tests/hour. Photometric analysis is run on a 4.5 second cycle.
ISE Maximum 600 tests/hour. ISE analysis is run on an 18 second cycle.
Sample throughput rate Maximum 800 samples/hour
Tests per sample Maximum 103 tests per sample, typically 41 (photometric) + 3 (ISE).
Sample
Trays
STT Used for general samples and calibrators for multi-point calibration assays
Dilution
Sample dilution mechanism The primary sample is undiluted in a reaction tray (RRV) cuvette.
Primary sample volume 1 to 25 μL in 0.1μL steps
Dilution reagent volume 5 to 300 μL in 0.1μL steps
Reaction cuvette material Plastic
GPE 01006.004 31
Reagent
Reaction
Assay
ISE
GPE 01006.004 32
Method Na, K, Cl ion selective electrolyte assay
Dilution measurement method
Tests Three tests, Na, K, and Cl are measured simultaneously
Electrode Na: Crown ether membrane
K: Crown ether membrane
Cl: Super-layer solid molecule orientation membrane
ref: Silver/silver chloride electrode
Throughput rate Maximum of 600 tests/hour (200 samples/hour) for serum samples
Sample volume 22 μl
Dilution ratio 1:33 (approximate)
Reagent volume Buffer: 2.7 ml / sample
Maintenance
Automatic maintenance Automatic startup and automatic shutdown by timer for weekly
maintenance
Dimensions
GPE 01006.004 33
LIST OF ACRONYMS
Acronym Meaning
BP Oil heat bath pump
CDEV1 Reaction tray wash unit drain valve 1
CDEV2 Reaction tray wash unit drain valve 2
CDEV3 Reaction tray wash unit drain valve 3
CDP-1 Drain pump 1
CDP-2 Drain pump 2
CTT Calibrator/control Tray
CWEV Reaction tray wash unit drain valve
DCEV Cuvette conditioner valve
DCP Dilution probe wash pump
DIP Dilution probe aspiration pump
DMEV Dilution mixer wash valve
DMIX Dilution mixer
DMUD Dilution mixer (up and down)
DOP Dilution probe discharge pump
DPEV1 Dilution probe valve 1
DPEV2 Dilution wash cup valve 2
DPEV3 Dilution wash cup valve 3
DPPLR Sample-dilution probe (rotating)
DPPUD Sample-dilution probe (up and down)
DTEV1 Reaction tray detergent valve 1
DTEV2 Reaction tray detergent valve 2
DTP1 Reaction tray wash pump 1
DTP2 Reaction tray detergent pump 2
DTT Dilution tray
DWEV1 Dilution wash valve 1
DWEV2 Dilution wash valve 2
DWPl Dilution-cuvette wash pump 1
DWP2 Dilution-cuvette wash pump 2
GPE 01006.004 34
DWUD Dilution tray wash unit
LWP Water-supply pump
MIXR-1 Mixer 1
MIXR-2 Mixer 2
MLR-1 Mixer 1 (rotating)
MLR-2 Mixer 2 (rotating)
MUD-1 Mixer 1 (up and down)
MUD-2 Mixer 2 (up and down)
MWEVI Reaction tray mixer wash valve 1
MWEV2 Reaction tray mixer wash valve 2
RBC-1 Reagent bar-code reader 1
RBC-2 Reagent bar-code reader 2
RP1 Reagent dispensing pump 1
RP2 Reagent dispensing pump 2
RPEV1-1 Reagent probe 1 valve 1
RPEV2-1 Reagent wash cup 1 valve 2
RPEV1-2 Reagent probe 2 valve 1
RPEV2-2 Reagent wash cup 2 valve 2
RPPLR-1 Reagent probe 1 (rotation)
RPPLR-2 Reagent probe 2 (rotation)
RPPUD-1 Reagent probe 1 (up and down)
RPPUD-2 Reagent probe 2 (up and down)
RRV Reaction tray
RTT-1 Reagent tray 1
RTT-2 Reagent tray 2
RWPL Reagent-wash pump I
RWP2 Reagent-wash pump 2
SBC Sample bar-code reader
SCP Sample probe wash pump
SP Sample aspiration/dispense pump
SPEV1 Sample probe valve 1
GPE 01006.004 35
SPEV2 Sample wash cup valve 2
SPPLR Sample probe (rotating)
SPPUD Sample probe (up and down)
STT Sample tray
VDEV1 Drain valve 1
VDEV2 Drain valve 2
VIEV1 Drain valve 1
VIEV2 Drain valve 2
VIEV3 Drain valve 3
VOEV1 Vacuum valve 1
VOEV2 Vacuum valve 2
VP Vacuum pump
WCV Switching valve
WEV Water supply tank valve
WPI Reaction tray wash pump 1
WP2 Reaction tray wash pump 2
WP3 Reaction tray wash pump 3
WUD Reaction tray wash unit
GPE 01006.004 36
System Overview
Table of Contents:
1
GPE 01006.004
2
GPE 01006.004
Learning Objectives:
At the conclusion of this training module, the learner will be able to:
3
GPE 01006.004
Ion Selective Electrode (ISE) Components
The amount of sodium (Na), potassium (K), and chloride (Cl) in serum or urine samples is
determined through voltage measurement by ion-selective electrodes.
Sample aspirated by the sample-dilution probe (DPP) is used for the electrolyte analysis.
The electrolyte analysis uses buffer to dilute the sample.
In a two-stage process, the buffer voltage is measured, then the sample voltage is
measured. The difference between these voltages, the reference voltage, and the
temperatures of the liquids determines the concentration of Na, Cl, and K in the sample.
4
GPE 01006.004
Components, Controls, and Indicators:
Not articulated
on 1650 or 2400
Not on 1650 or
2400
5
GPE 01006.004
Sample Tray Illustration
The sample tray holds patient samples, controls, calibrators, and diluents for measurement.
The tray rotates to move the samples to the aspiration position.
STT (outer section): Used for general samples and reference samples for multipoint
calibrations. It has two rows, each containing 42 positions (total 84). You can place
serum or urine samples into each position. Room temperature.
A barcode reader identifies samples in the STT. It can interpret barcode formats Code
39, Interleaved 2 of 5, Codabar, and Code 128.
CTT (inner section): Used for calibrators, controls, and special purpose diluents. It has
two rows. The outer row contains 34 positions; the inner row contains 27 positions (total
61). The CTT is water-cooled.
6
GPE 01006.004
Sample Tray Operation:
Initialization: The tray rotates clockwise until STT position 1 is in the aspiration
position.
After you start a run, the sample tray moves clockwise to position 1, then it rotates clockwise
to move samples successively to the aspiration position.
If samples are analyzed by barcode, each sample detected by the barcode reader stops (in
turn) in the aspiration position. If samples are analyzed by position number, they move by
position number to the aspiration position. Regardless of the mode, a workorder for the
sample must be entered, or it will not be processed.
Container types:
You place sample containers defined in the Order Entry window in the STT and CTT. Usually
you use 5 mL, 7 mL, or 10 mL collection tubes or 2 mL sample cups, which are placed in
plastic holders (the holders are then placed in tray positions).
The dead volume for the collection tubes is 200 µL, and the dead volume for the sample
cups is 50 µL.
Collection tubes often have barcode labels which can be read by the barcode reader, if
placed in the STT. However, you cannot use barcodes with sample cups.
7
GPE 01006.004
Dilution Tray
The sample-dilution probe (DPP) dispenses diluted samples into cuvettes in the dilution
tray (DTT). After the sample is in a DTT cuvette, it is stirred by the dilution mixer (DMIX),
then aspirated by the sampling probe (SPP). After analysis, the cuvettes are washed by
the dilution washer (DWUD).
The DTT contains 120 reusable cuvettes (6 sets of 20 cuvettes). The maximum sample
volume for each cuvette is 300 μL.
1 Dilution Tray
(DTT)
2 DTT Cuvette
The DWUD has three nozzles, each performing a stage of the wash. Each nozzle works
on a different cuvette, so the DWUD washes three cuvettes simultaneously. The cuvettes
are being washed in different stages at the same time.
After a cuvette is washed by one nozzle, it moves to the next until washing is complete.
While the DTT rotates, the DWUD is in the up position.
8
GPE 01006.004
Reaction Tray Illustration
For each sample assay, the reagent probes (RPP1 and RPP2) dispense reagent into
cuvettes in the reaction tray (RRV). Then the sample probe (SPP) dispenses diluted sample
into the cuvettes. The mixture is then stirred by the reaction mixers (MIXR1 and MIXR2),
producing the desired reaction.
For analysis, the reaction tray rotates the cuvette in front of the spectrophotometer, where
the cuvette’s absorbance is measured. After analysis, the cuvettes are washed by the
reaction washer (WUD).
The RRV contains 221 reusable cuvettes (13 sets of 17 cuvettes) on 1650 & 1800. 231 on
1200.
Each cuvette contains 80 to 300 μl of reaction liquid.
The RRV contains 340 reusable cuvettes (20 sets of 17 cuvettes) on 2400.
Each cuvette contains 60 to 180 μl of reaction liquid.
The RRV cuvettes are kept at a constant temperature of 37 °C by being immersed in the
reaction tank, which contains an oil-heat bath.
9
GPE 01006.004
Reaction Tray Operation:
Initialization: The tray rotates counterclockwise until RRV cuvette 1 is in the position
where the first reagent is dispensed.
After you start a run, the reaction tray moves RRV cuvette 1 counterclockwise to the reagent
1 dispense position. Then the following cycle begins:
10
GPE 01006.004
Reaction Tray Wash Mechanism
The reaction washer (WUD) washes reaction tray (RRV) cuvettes after sample analysis
is complete. This allows the reuse of cuvettes without the risk of contaminating the next
sample.
The WUD has seven nozzles. Each nozzle performs a stage of the wash. Each nozzle
works on a different cuvette, so the WUD washes seven cuvettes simultaneously (the
cuvettes are washed in different stages at the same time).
After a cuvette is washed by one nozzle, it moves to the next until washing is complete.
While the RRV rotates, the WUD is in the up position.
The wash liquids pass through a preheater before they reach the WUD.
11
GPE 01006.004
Reaction Tray Oil Bath Level Sensors
Two liquid level sensors control the oil supply in the reaction tank:
¾ Liquid depletion
¾ Start replenishment
¾ Stop replenishment
12
GPE 01006.004
Reagent Tray Illustration
Reagent trays 1 and 2 (RTT1 and RTT2) contain reagents used for assays and the
detergents used for daily washing and contamination prevention.
Position on Position on Solution Position on
1200 1800 1650 & 2400
42 53 PW1 47
43 54 PW2 48
44 55 10%CW 49
45 56 DI H2O 50
The reagent probes (RPP1 and RPP2) aspirate the required reagent and dispense it into
reaction tray (RRV) cuvettes for analysis.
Each tray on 1800 has 56 positions. 29 positions for 40 ml containers and 27 positions for 70
ml containers. 20 ml reagent adapters can be used in either the 40- or the 70-ml positions.
Each tray on 1650 & 2400 has 50 positions for 70 ml containers with adapters available for
20 & 40 ml conversion. 1200 has 45 positions: 12 for 70ml containers; the remaining for
20ml with adapters or 40ml containers.
RTT1 contains the first reagent (R1); RTT2 contains the second reagent (R2).
Any reagent container can be used for more than one test item; a test item may require more
than one reagent container.
13
GPE 01006.004
1200 Reagent Tray
Initialization: The trays rotate clockwise until reagent bottle 1 is in the aspiration
position.
After you start a run, the reagent trays move clockwise to position 1, then they rotate
clockwise or counterclockwise (whichever results in a smaller rotation) to move reagents to
the aspiration position. The number of trays and reagents used depends on the specified
assay conditions.
Reagent probes (RPP1 and RPP2)
Articulated RPP1 on
1800 only
14
GPE 01006.004
DTT Mixer
Mixers
15
GPE 01006.004
Spectrophotometer
1 Photometer
2 Halogen Lamp
3 Cooling Tank
16
GPE 01006.004
Operating Principle
1. The system performs a cell blank measurement before reagents are added.
2. The first reagent (R1) for the test is aspirated from reagent tray 1 and
a. dispensed by the reagent probe into a cuvette in the reaction tray.
3. Samples on the sample tray or rack handler are aspirated and diluted by the
a. dilution probe, then dispensed into cuvettes in the dilution tray.
b. The 1:5 dilution is made by mixing 30 μl of sample with 120 μl of saline.
4. The diluted sample is stirred by the dilution mixer.
5. The required amount of diluted sample is dispensed by the sample probe into
a. the RRV cuvettes (the reagent is already in the cuvettes).
b. The diluted sample remaining in the DTT cuvettes can be saved for rerun or
reflex testing.
6. The reaction mixer 1 mixes the first reagent and the sample.
7. The second reagent (R2) for a test is aspirated from reagent tray 2 and dispensed by
the reagent probe into the cuvette in the reaction tray.
8. The reaction mixer 2 mixes sample and reagent 1 and reagent 2.
9. The reaction takes place for the amount of time designated in the assay.
10. Concentration data is obtained by the spectrophotometer every six seconds. The
spectrophotometer takes readings as the RRV turns.
11. You can view and print the results of the analysis.;
12. The RRV cuvettes are washed when measurement is complete.
13. When the analysis is complete, the lamp energy is checked at each wavelength.
17
GPE 01006.004
All Systems: Display Panel and Power Panel * 1650 has additional
LED for ISE drawer warning here
18
GPE 01006.004
1800: Front View
19
GPE 01006.004
2400: Front View
20
GPE 01006.004
1650: Front View
Horizontal
Pumps
Vertical
Pumps
21
GPE 01006.004
1800 Lower right compartment…
Oil Filter
Water regulator
Red valve
Oil heater not visible
22
GPE 01006.004
1800: Rear View
Main
Power
LAS
Fuse Temperature Interfa
External Panel Regulator
Water
23
GPE 01006.004
2400: Rear View
24
GPE 01006.004
All Systems: PC Front View
1 USB Ports
2 PC power switch. Used to turn the power for the
personal computer ON or OFF. Normally, it is left
ON.
3 DVD/CD-RW Drive
4 DVD/CD-RW drive access lamp. Lights when
reading the CD.
5 Emergency eject CD
6 DVD/CD-ROM drive eject button. Press to
remove the CD.
7 PC power lamp. Lights when the power for the
personal computer is ON.
8 Hard-drive access lamp. Lights when reading or
writing to the hard disk.
9 Reset Switch
10 Access Door
PC Rear View
1 Modem Connectors
2 Sleep ITF Board Potentiometer
3 Analyzer Connector
4 CRT Connector
5 Printer Connector
6 Serial Connector (COM1) LIS
7 Keyboard Connector
8 Mouse Connector
9 PC Power Connector
10 Serial Connector (COM2) URH
25
GPE 01006.004
ADVIA Chemistry Operation Panel Window
ITEM PURPOSE
26
GPE 01006.004
ADVIA Chemistry Operation Panel Window (continued)
ITEM PURPOSE
System(s)
When selected, places the ADVIA 1800 Operation Panel window on top
Top display of all other windows. Area occupied by the ADVIA 1650 Operation
Panel window covers other window.
Provides the version information for the ADVIA 1800 Operation panel
Version info (A)
window
Reconnect Reconnects the workstation and analyzer if communication was lost.
1. Where the System Operating Mode is displayed (see table on following
pages)
2. Displays messages that tell you whether or not you can add or replace
samples, change sample trays, or add and remove reagent containers.
3.
Alarm message box: The system displays events and abnormal
conditions (alarms) in this box.
4.
Displays the time remaining (in mm:ss format) until tasks running under
the following operating modes are complete:
• WASH (1, 2, 3)
• PRIME
• WATER BLANK
• CELL BLANK
• DET.CHECK
• DET.RRV CHECK
For other operating modes, no time
displays.
27
GPE 01006.004
Operating mode Description Operating mode Description
28
GPE 01006.004
1800 System Specifications
Item Description
Method
Measurement method Open discrete
Single-line, simultaneous, multi-item measurement
Process
Throughput rate 1800 tests/hour
Biochemistry 1200 tests/hour
Electrolyte 600 tests/hour
Sample throughput rate 1200 samples/hour
Simultaneous Normally 52 (main system) + 3 (ISE); up to 100 definable assays
measurement item
Sample
Measurement sample Blood serum, plasma, urine (method dependent), and CSF
Containers 5-mL (13 x 75 mm), 7-mL (13 x 100 mm), 10-mL (16 x 100 mm)
collection tubes; dead volume for collection tubes is 200 µL
Cups 1-mL sample cup (STT only), Hitachi
2-mL sample cups, and Ez Nest 2-mL sample cups in 7-mL (16 x 75
mm) collection tubes (URH) or STT sample adapter; dead volume for
sample cups is 50 µL.
Dilution
Dilution tray (DTT) Turntable system
Number of cuvettes 120 (6 sets of 20 cuvettes)
Maximum sample volume 300 μL
Dilution cuvette dead 35 μL
volume
Dilution ratio From 0 to 1:5625
Reaction cuvette material Plastic
29
GPE 01006.004
1800 System Specifications (continued)
Item Description
Trays
STT Used for general samples and calibrators for multipoint calibration
assays
Two lines (outer and inner) of 42 samples each. Total positions in STT
tray: 84
Sample barcode (13 digits): Code 39, Codabar, and Interleaved 2 of 5,
Code 128 format A, B and special characters ( . - + / * $ : % ), NW7
CTT Used for calibrators, controls, and diluents
Two lines, 34 samples in outer line and 27 samples in inner line. Total
positions in CTT tray: 61
Liquid contents on CTT tray are cooled to between 6°C and 14°C
URH Universal rack handler 5 position rack
Original sample volume 1 to 30 μL (0.1 μl increments)
Assay sample volume 1 to 25 μL (0.1 μl increments)
(after dilution)
Reassay
Container Dilution tray (DTT) cuvette
Minimum sampling 1 μL (0.1 μL increments)
volume
Special dilution Diluted sample can be rediluted directly from tray
Reagent
Dispensing system 2-reagent capability, 2-probe system
Trays Two trays, each holding 56 containers
Multiple reagent pack loads - 5 reagents per method with automatic
rollover upon depletion
Reagents on each tray are cooled to between
6°C and 14°C
Container capacity 20, 40, or 70 mL
Refrigerator All method reagents
Reagent volume/item 15 to 150 μL (0.1 μL increments)
Dilution reagent volume 15 to 150 μL (0.1 μL increments)
Reagent inventory Computed by the liquid-level sensing
Barcode Interleaved 2 of 5
30
GPE 01006.004
1800 System Specifications (continued)
Item Description
Reaction
Reaction tray (RRV) Turntable system
Number of cuvettes 221 (13 sets of 17 cuvettes)
RRV cuvette material Plastic
Reaction liquid volume 80 to 300 μL; minimum read-volume 80 μL
Stirring system Rotation and reciprocation
Strong and weak stirring options are available.
Stirring immediately after additions of samples and reagents (S, R1, R2)
Mixer 1 is used for R1; Mixer 2 is used for R2.
Reaction times 3, 4, 5, 10 minutes
Extended reaction times: 15 & 21 minutes
Reaction temperature 37°C
Temperature regulation: ±0.1°C
Reaction tank Inert liquid circulation system
Assay
Measurement point 98 detection points/6 seconds in 10-minute reaction
Photometer Concave diffraction grating, rear spectroscopy system
Measurement wavelength 14 fixed wavelengths (340, 410, 451, 478, 505, 545, 571, 596, 658,
694, 751, 805, 845, and 884 nm), 1 or 2 wavelength calculation
Light source 12v, 50W halogen lamp, cooled by forced water circulation
Assay method Colorimetric assays
• Endpoint assays (EPA)
• Reaction rate assays (RRA)
• 2-point rate assays (2PA)
• 3-item simultaneous measurement (parameter independent)
• Prozone checking (antigen excess check)
• Substrate depletion check with point forwarding option
• Sample blank correction
Reassay mechanism Automatic or manual (selectable)
Automatic correction Blood serum blank, cell blank, measurement point change, sample
volume change in reassay
31
GPE 01006.004
1800 System Specifications (continued)
Item Description
ISE
Method Na, K, Cl ion selective electrolyte assay
Dilution measurement method
Analysis item Simultaneous 3 item measurement of Na, K, and Cl
Electrode Na: Crown ether membrane
K: Crown ether membrane
Cl: Super-layer solid molecule orientation membrane
ref: Sealed silver/silver chloride electrode
Throughput rate 600 tests/hour (200 samples/hour)
Sample volume 22 μL, plus 4 μL dead volume
Dilution ratio 1:33
Reagent volume Buffer: 2.8 mL/sample
Maintenance
Automatic maintenance Automatic startup and automatic shutdown by timer for weekly
maintenance
Dimensions
Analyzer dimensions 1133(h) x 1480(w) x 876(d) mm
(44.6 x 58.3 x 34.5 in)
Weight 600 kg (1323 lb)
Optional universal rack 95.25(h) x 72.5(w) x 103.0(d) cm
handler dimensions
(37.5 x 28.5 x 40.55 in)
Weight 80.7 kg (178 lb)
32
GPE 01006.004
1800 System Specifications (continued)
Item Description
Environment
33
GPE 01006.004
1800 System Specifications (continued)
Item Description
Cooling / ventilation • Ventilation sufficient to maintain +18°C to +30°C (+64° to +86°F
requirements operating temperature.
• Environment should be free of corrosive gas, significant vibration and
electrical disturbances, such as electromagnetic and electrostatic
induction.
• The maximum temperature change the system can accommodate is
2°C/hour.
• System is for indoor use at an altitude of up to 2000 meters with a
pollution degree of 2.
• Maximum relative humidity allowable with system operating is 40% to
70% with no condensation.
• Heat output:
50 Hz 60 Hz
Power off mode:
400 W 350 W
1365 Btu/hr 1024 Btu/hr
344 kCal/hr 300 kCal/hr
Ready mode:
1260 W 1030 W
4299 Btu/hr 3023 Btu/hr
1084 kCal/hr 886 kcal/hr
Auto mode:
1480 W 1540 W
5050 Btu/hr 4518 Btu/hr
1273 kCal/hr 1324 kCal/hr
Water requirements The system is connected directly to a pressurized water source using
Type 1 or ISO 3696 water.
Direct plumbing deionized water pressure: 1.4 - 14.2 psi (9.6 - 96 kPa)
Drain requirements Minimum of 40 liters (10.6 gallons) per hour
If the local laboratory practices and/or applicable environmental
regulations prohibit the inclusion of concentrated waste into the
laboratory's drain, an optional concentrated waste bottle must be
ordered.
34
GPE 01006.004
2400 Specifications
Item Description
Method
Measurement Open discrete
method
Single-line, simultaneous, multi-item measurement. Full random-
access.
Process
Throughput rate Maximum 2400 tests/hour
Biochemistry Maximum 1800 tests/hour
Electrolyte Maximum 600 tests/hour
Sample throughput Maximum 1800 samples/hour
rate
Simultaneous Maximum 103 items
measurement item
Normally 50 (main system) + 3 (ISE)
Sample
Measurement Blood serum, plasma, and urine (method dependent)
sample
Containers 5 mL, 7mL, 10 mL collection tubes; dead volume for collection
tubes is 200 µL
Cups 1.8 mL JEOL sample cup (STT only), Hitachi 2 mL sample cups,
and Immuno 2 mL sample cups in 7 mL (16 x 75 mm) collection
tubes (URH) or STT sample adapter; dead volume for sample
cups is 50 µL.
Trays
STT Used for general samples and calibrators for multipoint calibration
assays
Two lines (outer and inner) of 42 samples each. Total positions in
STT tray: 84
Sample barcode, 13 digits: Code 39, Codabar, and Interleaved 2
of 5, Code 128 format A, B and special characters (. - + / * $ %)
CTT Used for calibrators, controls, and diluents
Two lines, 34 samples in outer line and 27 samples in inner line.
Total positions in CTT tray: 61
Dilution
Dilution tray (DTT) Turntable system
Number of cuvettes 120 (6 sets of 20 cuvettes)
Maximum sample 300 μL
volume
Dilution cuvette 35 μL
dead volume
Dilution ratio From 1:1 to 1:5625 ( 1:75 x 1:75 )
Reaction cuvette Plastic
material
Reagent
Dispensing system 2 reagent capability, 2 probe system
Trays Two trays, each holding 50 containers
Reagents on each tray are cooled to between
6 °C and 14 °C
Container capacity 20 or 70 mL
Refrigerator Recirculated liquid
Reagent 10 to 100 μL (0.1 μl increments)
volume/item
Dilution reagent 10 to 90 μL (0.1 μl increments)
volume
Reagent inventory Computed by the liquid-level sensing
Barcode Interleaved 2 of 5. JSCLA standard.
36
GPE 01006.004
2400 Specifications
Reaction
Reaction tray (RRV) Turntable system
Number of cuvettes 340 (20 sets of 17 cuvettes)
RRV cuvette Plastic
material
Reaction liquid 60 to 180 mL
volume
Stirring system Rotation and reciprocation
Stirring immediately after addition of R1 and sample and R2.
Mixer 1 is used for R1; Mixer 2 is used for R2.
Reaction times 3, 4, 5, 10 minutes
Extended reaction times: 15 & 21 minutes
Reaction 37 °C
temperature
Temperature regulation: ±0.1 °C
Reaction tank Inert liquid circulation system
Assay
Measurement point 41 detection points/14 seconds in 10 minute reaction
Photometer Concavity diffraction grating, rear spectroscopy system
Measurement 14 fixed wavelengths (340, 410, 451, 478, 505, 545, 571, 596,
wavelength 658, 694, 751, 805, 845, and 884 nm), 1 or 2 wavelength
calculation
Light source 12V, 50W halogen lamp, cooled by forced water circulation
Assay method Colorimetric assays
• Endpoint assays (EPA)
• Reaction rate assays (RRA)
• 2-point rate assays (2PA)
• 3-item simultaneous measurement (parameter independent)
• Prozone checking (antigen excess check)
• Substrate depletion check with point forwarding option
• Sample blank correction
Reassay mechanism Automatic or manual (selectable)
Automatic correction Blood serum blank, cell blank, measurement point change,
sample volume change in reassay
37
GPE 01006.004
2400 Specifications
ISE
Method Na, K, Cl ion selective electrolyte assay
Dilution measurement method
Analysis item Simultaneous 3 item measurement of Na, K, and Cl
Electrode Na: Crown ether membrane
K: Crown ether membrane
Cl: Super-layer solid molecule orientation membrane
ref: Silver/silver chloride electrode
Throughput rate Maximum of 600 tests/hour (200 samples/hour) for serum
samples
Sample volume 22 μl
Dilution ratio 1:33 (approximate)
Reagent volume Buffer: 2.7 ml / sample
Maintenance
Automatic Automatic startup and automatic shutdown by timer for weekly
maintenance maintenance
Dimensions
Analyzer dimensions 1340(h) x 1711(w) x 934(d) mm
52.8(h) x 68.44(w) x 37.36(d) in.
Weight 630 kg (1389 lb.)
Vacuum pump 546(h) x 340(w) x 500(d) mm
dimensions
(21.5 x 13.4 x 19.7 in.)
Computer 366(h) x 221(w) x 421(d) mm
dimensions
(14.4 x 8.7 x 16.6 in.)
Monitor dimensions 422(h) x 417(w) x 447(d) mm
(16.6 x 16.4 x 17.6 in.)
Optional universal 1126(h) x 730(w) x 1022(d) mm
rack handler
(44.33 x 28.74 x 40.24 in.)
dimensions
38
GPE 01006.004
2400 Specifications
Environment
39
GPE 01006.004
1650 Specifications
Item Description
Method
Measurement Open discrete
method
Single-line, simultaneous, multi-item measurement.
Process
Throughput rate Maximum 1650 tests/hour
Biochemistry Maximum 1200 tests/hour
Electrolyte Maximum 450 tests/hour
Sample throughput Maximum 1200 samples/hour
rate
Simultaneous Maximum 103 items
measurement item Normally 50 (main system) + 3 (ISE)
Sample
Measurement Blood serum, plasma, and urine (method dependent)
sample
Containers 5 mL, 7mL, 10 mL collection tubes; dead volume for collection
tubes is 200 µL
Cups 1 mL JEOL sample cup (STT only), Hitachi 2 mL sample cups,
and Immuno 2 mL sample cups in 7 mL (16 x 75 mm) collection
tubes (URH) or STT sample adapter; dead volume for sample
cups is 50 µL.
Trays
STT Used for general samples and calibrators for multipoint calibration
assays
Two lines (outer and inner) of 42 samples each Total positions in
STT tray: 84
Sample barcode (13 digits): Code 39, Codabar, and Interleaved 2
of 5, Code 128
CTT Used for calibrators, controls, and diluents
Two lines, 34 samples in outer line and 27 samples in inner line.
Total positions in CTT tray: 61
Liquid contents on CTT tray are cooled to between 6 °C and 14
Rack handler °C
40
GPE 01006.004
1650 Specifications
Dilution
Dilution tray (DTT) Turntable system
Number of cuvettes 120 (6 sets of 20 cuvettes)
Maximum sample 300 μL
volume
Dilution cuvette dead 35 μL
volume
Dilution ratio From 1:1 to 1:5625
Reaction cuvette Plastic
material
Reagent
Dispensing system 2 reagent capability, 2 probe system
Trays Two trays, each holding 50 containers
Reagents on each tray are cooled to between
6 °C and 14 °C
Container capacity 7, 20, or 70 mL
Refrigerator All reagents
Reagent volume/item 15 to 150 μL (0.1 μl increments)
Dilution reagent 15 to 150 μL (0.1 μl increments)
volume
Reagent inventory Computed by the liquid-level sensing
Barcode Interleaved 2 of 5
41
GPE 01006.004
1650 Specifications
Reaction
Reaction tray (RRV) Turntable system
Number of cuvettes 221 (13 sets of 17 cuvettes)
RRV cuvette material Plastic
Reaction liquid 80 to 300 μL
volume
Stirring system Rotation and reciprocation
Strong and weak stirring options are available.
Stirring immediately after additions of samples and reagents (S,
R1, R2)
Mixer 1 is used for R1; Mixer 2 is used for R2.
Reaction times 3, 4, 5, 10 minutes
Extended reaction times: 15 & 21 minutes
Reaction 37 °C
temperature
Temperature regulation: ±0.1 °C
Reaction tank Inert liquid circulation system
Assay
Measurement point 98 detection points/6 seconds in 10 minute reaction
Photometer Concavity diffraction grating, rear spectroscopy system
Measurement 14 fixed wavelengths (340, 410, 451, 478, 505, 545, 571, 596,
wavelength 658, 694, 751, 805, 845, and 884 nm), 1 or 2 wavelength
calculation
Light source 12V, 50W halogen lamp, cooled by forced water circulation
Assay method Colorimetric assays
• Endpoint assays (EPA)
• Reaction rate assays (RRA)
• 2-point rate assays (2PA)
• 3-item simultaneous measurement (parameter independent)
• Prozone checking (antigen excess check)
• Substrate depletion check with point forwarding option
• Sample blank correction
Reassay mechanism Automatic or manual (selectable)
Automatic correction Blood serum blank, cell blank, measurement point change,
sample volume change in reassay
42
GPE 01006.004
1650 Specifications
ISE
Method Na, K, Cl ion selective electrolyte assay
Dilution measurement method
Analysis item Simultaneous 3 item measurement of Na, K, and Cl
Electrode Na: Crown ether membrane
K: Crown ether membrane
Cl: Super-layer solid molecule orientation membrane
ref: Silver/silver chloride electrode
Throughput rate Maximum of 450 tests/hour (150 samples/hour) for serum
samples
Sample volume 28 μl
Dilution ratio 1:33
Reagent volume Buffer: 2.7 ml/sample
Reference (ref) liquid: 0.3 ml/sample
Maintenance
Automatic Automatic startup and automatic shutdown by timer for weekly
maintenance maintenance
Dimensions
Analyzer dimensions 1129(h) x 1480(w) x 867(d) mm
(44.45 x 58.27 x 34.13 in.)
Weight 530 kg (1178 lb.)
Vacuum pump 546(h) x 340(w) x 500(d) mm
dimensions
(21.5 x 13.4 x 19.7 in.)
Computer 366(h) x 221(w) x 421(d) mm
dimensions
(14.4 x 8.7 x 16.6 in.)
Monitor dimensions 422(h) x 417(w) x 447(d) mm
(16.6 x 16.4 x 17.6 in.)
Optional rack 936(h) x 730(w) x 1022(d) mm
handler dimensions
(36.86 x 28.74 x 40.24 in.)
Optional universal 1126(h) x 730(w) x 1022(d) mm
rack handler
(44.33 x 28.74 x 40.24 in.)
dimensions
43
GPE 01006.004
1650 Specifications
Environment
Electrical • A 3-kVA power source, single-phase, 2-pole, 3-wire
requirements configuration with Class III grounding
• The following input voltages can be tapped: 100 V, 110 V, 115
V, 200 V, 220 V or 230 V.
• Maximum current draw at in-rush is 26 amps at 115 V or 14
amps at 220 V.
• For the optional rack handler, the requirements are 110 Vac
50/60 Hz, 0.6 A, 70 W.
• For the optional universal rack handler, the requirements are
110 Vac 50/60 Hz, 0.6 A.
Cooling / ventillation • Ventilation sufficient to maintain +18° C to +30° C operating
requirements temperature
• The maximum temperature change the system can
accommodate is 2° C/hour.
• Maximum relative humidity allowable with system operating is
40% to 70% with no condensation.
• Heat output:
Power off mode:
0.117 kW
0.01856 Btu/s
0.02055 kCal/s
Ready mode:
0.871 kW
0.060717 Btu/s
0.15303 kCal/s
Auto mode:
1.36 kW
0.94805 Btu/s
0.23895 kCal/s
Water requirements Deionized (or demineralized) water from a nonpressurized water
reservoir with a 30 liter/hour capability
Drain requirements Minimum of 6.8 gallons (25 liters) per hour
44
GPE 01006.004
LIST OF ACRONYMS
Acronym Meaning
BP Oil heat bath pump
CDEV1 Reaction tray wash unit drain valve 1
CDEV2 Reaction tray wash unit drain valve 2
CDEV3 Reaction tray wash unit drain valve 3
CDP-1 Drain pump 1
CDP-2 Drain pump 2
CTT Calibrator/control Tray
CWEV Reaction tray wash unit drain valve
DCEV Cuvette conditioner valve
DCP Dilution probe wash pump
DIP Dilution probe aspiration pump
DMEV Dilution mixer wash valve
DMIX Dilution mixer
DMUD Dilution mixer (up and down)
DOP Dilution probe discharge pump
DPEV1 Dilution probe valve 1
DPEV2 Dilution wash cup valve 2
DPEV3 Dilution wash cup valve 3
DPPLR Sample-dilution probe (rotating)
DPPUD Sample-dilution probe (up and down)
DTEV1 Reaction tray detergent valve 1
DTEV2 Reaction tray detergent valve 2
DTP1 Reaction tray wash pump 1
DTP2 Reaction tray detergent pump 2
DTT Dilution tray
DWEV1 Dilution wash valve 1
45
GPE 01006.004
DWEV2 Dilution wash valve 2
DWPl Dilution-cuvette wash pump 1
DWP2 Dilution-cuvette wash pump 2
DWUD Dilution tray wash unit
LWP Water-supply pump
MIXR-1 Mixer 1
MIXR-2 Mixer 2
MLR-1 Mixer 1 (rotating)
MLR-2 Mixer 2 (rotating)
MUD-1 Mixer 1 (up and down)
MUD-2 Mixer 2 (up and down)
MWEVI Reaction tray mixer wash valve 1
MWEV2 Reaction tray mixer wash valve 2
RBC-1 Reagent bar-code reader 1
RBC-2 Reagent bar-code reader 2
RP1 Reagent dispensing pump 1
RP2 Reagent dispensing pump 2
RPEV1-1 Reagent probe 1 valve 1
RPEV2-1 Reagent wash cup 1 valve 2
RPEV1-2 Reagent probe 2 valve 1
RPEV2-2 Reagent wash cup 2 valve 2
RPPLR-1 Reagent probe 1 (rotation)
RPPLR-2 Reagent probe 2 (rotation)
RPPUD-1 Reagent probe 1 (up and down)
RPPUD-2 Reagent probe 2 (up and down)
RRV Reaction tray
RTT-1 Reagent tray 1
RTT-2 Reagent tray 2
46
GPE 01006.004
RWPL Reagent-wash pump I
RWP2 Reagent-wash pump 2
SBC Sample bar-code reader
SCP Sample probe wash pump
SP Sample aspiration/dispense pump
SPEV1 Sample probe valve 1
SPEV2 Sample wash cup valve 2
SPPLR Sample probe (rotating)
SPPUD Sample probe (up and down)
STT Sample tray
VDEV1 Drain valve 1
VDEV2 Drain valve 2
VIEV1 Drain valve 1
VIEV2 Drain valve 2
VIEV3 Drain valve 3
VOEV1 Vacuum valve 1
VOEV2 Vacuum valve 2
VP Vacuum pump
WCV Switching valve
WEV Water supply tank valve
WPI Reaction tray wash pump 1
WP2 Reaction tray wash pump 2
WP3 Reaction tray wash pump 3
WUD Reaction tray wash unit
47
GPE 01006.004
48
GPE 01006.004
Wash Solutions & Reagents
This document is designed for use during training classes only. Reference should be
made to the ADVIA Chemistry System Operator’s Guide and related Customer Bulletins for
product labeling information.
GPE01006.004 1
GPE01006.004 2
Learning Objectives:
reagent preparation
Loading reagent:
• REAGENT PAUSE
• Non-barcoded reagent
GPE01006.004 3
System Solutions (Room Temp Storage)
Note: This document is for classroom use only. Reference should be made to the Safety Data
Sheets for more detailed and current information.
GPE01006.004 4
System Solution Preparation
GPE01006.004 5
Reaction washer nozzles
R2 Dispenses detergent.
NOTE
The fourth (4) and fifth (5) nozzles are separated by a width of nine cuvettes.
GPE01006.004 6
Reagent packaging and storage
R1 only (reagent 1)
R1 & R2 (reagent 1 & 2)
R1 = white cap
R2 = blue cap
Perform a Prime 2
GPE01006.004 7
Reagent Inventory window
The Reagent Inventory window displays the status of all reagents loaded on the reagent trays
(RTT1 and RTT2). The window continuously refreshes with new information, and it updates after
each barcode scan.
2) Select Reagent Inventory from the taskbar at the bottom of the computer screen to
‘maximize’ the window.
¾ The
Reagent Inventory window automatically opens at start up and remains opened. When using
other windows or when you select the X in the upper right corner, the window minimizes to the
taskbar at the bottom. This was done to allow the software to post ‘reagent empty’ and ‘pack
switch’ and other audible reagent alarms to the alarm log and generate a beep. These alarms
would not be audible if the Reagent Inventory window was fully closed.
NOTE
When the Reagent Inventory window opens, the RTT1 reagents display by default. To display
the RTT2 reagents, select RTT2.
GPE01006.004 8
Reagent Inventory window (continued)
• The Test Name displays the reagent name from the System Test List window.
• The Posi. # displays the position number of the reagent container on the RTT1 or RTT2
reagent tray. The system automatically updates the position number based on the
reagent pack position during a barcode scan. Position numbers for reagent
containers without barcodes are entered manually in the System Test List window.
• P/B identifies the reagent container currently in use (Primary), and identical reagent
containers available (Backup).
GPE01006.004 9
Reagent Inventory window (continued)
• The Expiration Date is scanned from the reagent container barcode. Nonbarcoded
containers display the date from the Reagent Container Settings window or N/A if
no container set expiration date is entered.
• The Cal Interval Days is the number of days remaining to the end of the calibration interval.
¾ C in this column indicates a new lot number of reagent for which a valid calibration has
not yet been obtained.
¾ After calibrating the new reagent, the number in the Cal Interval Days column resets to
the maximum calibration interval determined by the value entered at the Reagent
Information window (provided for each assay in the (Siemens Method Sheet). On each
new day, the number counts down. If the Cal Interval Days field is not active, or a C is
not displayed when you put a reagent with a new lot number on the system, check that
the calibration interval number in the Reagent Information window is defined.
¾ When the Cal Interval reaches 0, the method line is highlighted in yellow. This indicates
that you should recalibrate the method. If the method is not calibrated at 0 days, the
counter will continue into negative numbers. Whenever you recalibrate a method, the
Cal Interval Days resets to the maximum value for that method.
NOTE: when you install 2 containers of the same reagent on the same tray with
different lot numbers and perform a barcode scan, the reagent designated by the
system as primary displays a C in the cal interval days, indicating that you must
calibrate the reagent. The second, or backup reagent is not marked with a C even
though this container also requires calibration.
• Days Remaining indicates the countdown for that reagent’s open pack on-board stability.
• Cal Status is the status of the calibration. The calibration can pass or fail
GPE01006.004 10
Reagent Inventory window (continued)
Select Total Test Summary to view (or print) a list of all methods on-board and their total tests
counts. If there is only 1 R1 and 1 R2 reagent present, the total tests value is the lower of the tests
remaining in the R1 and R2 reagent container. If there are multiple R1 and R2 reagents present on
the system, the value displayed is the minimum of the sum of the combined R1 reagents and the
sum of the combined R2 reagents. For example:
GPE01006.004 11
Reagent Inventory window (continued)
Use the Cal Interval Review list to review daily which methods you should calibrate. Instead of
an alphabetical list, this window re-sorts the Reagent Inventory list methods in order of Cal
Interval Days remaining. Methods with the shortest days remaining are at the top.
¾ Barcode Scan
GPE01006.004 12
The information below is extracted from page 4 of this bulletin
Action
GPE01006.004 13
LOADING REAGENTS
2 1
1) The barcode readers on RTT1 and RTT2 scan all positions for each reagent tray and….
2) The RPP probes sense the liquid level in each container to determine the volume.
When the Barcode Scan is complete, all information at the Reagent Inventory window refreshes
automatically.
GPE01006.004 14
LOADING REAGENTS (continued)
When the system is in the READY mode, the barcode scan can be accomplished by clicking on
Barcode Scan in the Reagent Inventory window.
When the system is running, sample processing can be paused for reagent loading by using the
REAGENT PAUSE button on the Operation Panel.
¾ After pressing the RGT Pause button the system displays three messages:
R PAUSE SHIFT indicates the system is completing operations prior to allowing reagent
addition.
RGT LOAD NG indicates the reagent probes may still move and it is not safe to load
reagent.
0403 R PAUSE SHIFT alarm is generated.
¾ Sampling will stop. Reagent will be added to all tests already in process in the RRV before the ADD
Reagent message is displayed.
GPE01006.004 15
REAGENT PAUSE (continued)
¾ When the system is available for new reagent to be added, a new window will appear.
Press the SCAN button to begin reagent barcode scanning and level sensing. Once this
operation is complete the system will return to sampling.
GPE01006.004 16
REAGENT PAUSE (continued)
NOTES:
¾ The RGT Pause function has one additional feature. The communication link to an external
devise (LAS / URH) is maintained. Samples held at the LAS interface gate will remain at the
gate and be sampled when the reagent scan is complete.
¾ The system may go into ADD REAGENT mode within a few seconds or it may take several
minutes depending on the system’s work in progress.
¾ To stop the RGT Pause function and resume sampling simply press the Start button and
acknowledge the prompt in pop up window.
GPE01006.004 17
Here are three scenarios and what happens when RGT Pause is requested
3. If RGT addition is needed after all the specimens have been resulted and the
ADVIA 1800 is in the “Processing” state – (in other words it is washing), RGT Pause is
grayed out and the user can’t do anything – UNLESS ….
a. With the system in “Processing”, the START button is available, and the
minute the user selects START, the RGT Pause button becomes available.
b. From there the user can either initiate a START and then select RGT
Pause, in which case the system will allow almost immediate access to reagent
addition.
With the scenarios listed above, the longest wait was about 6 minutes and the shortest wait was
about 1 minute. This type of RGT pause functionality is wonderful and clinically useful.
I did test out some specimens to see if results were affected and I did use Mg and Iron as the tests
for these specimens since both of them have some type of extra washes programmed in the
system. I found that the wash time was the same whether or not reagent addition was a part of the
process. The results were the same in all scenarios as well.
Please don’t change this back to the original scenario of waiting until the wash sequences are
completed before reagent addition is allowed. The R1 and R2 probes are not moving during the
washes or following the addition of R2 to all sampled specimens, so this current set-up works well.
GPE01006.004 18
Information from JEOL
A1) The criterion for a reagent pause to happen is that ‘if all the scheduled reagent aspiration has
been completed’.
1) When the reagent pause button is pressed, the sampling (DPP) will be stopped.
To be more precise, it means ‘When the scheduled sampling, already established at the time
of the reagent pause button being pressed, is completed, the sampling by DPP will be
stopped.’
Sampling will be stopped after maximum 2 samples if the reagent pause button is pressed.
2) ‘If all the scheduled reagent aspiration has been completed’ means that ‘if all the aspiration
and mixing of the reagent (R1, R2 and R3) for the samples that were sampled by DPP have
been completed’.
(Q1-2) Does it wait for all aspirated tests to complete result processing so that it knows if it needs
to do a repeat?
(A1-2) No. When the reagent aspiration is completed, it is ready to move to ‘PAUSE’ mode, so it
does not wait to complete the data output.
(Q1-4) In some cases we see reagent pause of 3-6 min and others it can take 40 mins.
What is the explanation for this?
(A1-4)
Information from JEOL (cont’d)
GPE01006.004 19
1) Firstly, please note that aspiration time for a sample largely varies depending on the number
of the order entry.
For example, aspiration for a sample with 1 order is once, but aspirations for a sample with 20
orders will be 20 times.
2) Secondly, each order requires different number of the reagents to use (ex. A test that requires
R1 only, or a test that requires R1, R2 and R3 etc), so this determines the aspiration time for the
reagent(s).
For example, with ADVIA 1650, each reagent aspiration timing is as follows;
R1 1 cycle before aspirating the diluted sample on DTT to RRV by SPP (3 seconds before)
R2: 23 cycles after aspirating diluted sample on DTT to RRV by SPP (1 minute after)
R3: 96 cycles after aspirating diluted sample on DTT to RRV by SPP (5 minutes after)
Therefore, an order that uses only R1 will complete the reagent aspiration faster than an order that
also uses R3 as well as R1 and R2.
It is possible that a reagent pause can take 3 minutes or 40 minutes depending on the combination
of 1) and 2).
Q2) Why is the reagent pause button disabled during washing after processing of samples is
complete?
A2) When an analysis completed, the system goes into a mode to stop. This is called ‘END’ mode.
In the ‘END’ mode, the system performs various operations, such as stopping each unit according
to the procedure, or vacuuming remaining water in the cuvette. Therefore, it is not possible to re-
start the analysis during the ‘END’ mode. In this mode, not only the reagent pause button but also
all other buttons are disabled.
GPE01006.004 20
Deselecting a reagent must be done in the Ready State
¾ To deactivate a reagent container, select the Deselect checkbox next to the applicable Test
Name.
¾ When the Deselect box is checked, the system cannot use the reagent. The Test Name is
red. The P/B field clears and if more than one of the same reagent is on the reagent trays,
the deselected reagent is no longer part of the reagent ranking process.
NOTE If you deselect a reagent container while the system is running, the deselect does
not take effect until the system returns to the READY state.
GPE01006.004 21
ACTIVE TEST LIST must be done in the Ready State
The Active Test List window is only available from the supervisor logon.
The Active Test List displays all tests defined on the analyzer, but the deactivated or unchecked
tests are grayed out.
Use the Active Test List window to temporarily deactivate a method, so that method-related
alarm messages (for example, missing reagents) are not generated. You would use this function
for tests that you do not run every day.
After you made your selections, activating or deactivating a method, the system must go back to
READY for the changes to take effect.
NOTE At the active test list window, you must select all the tests you plan
to run. If a test is not selected, the system will skip that test. An error message does
not display in the alarm message box on the operation panel, but is recorded in the
error message log.
GPE01006.004 22
MANUAL REAGENT PLACEMENT
GPE01006.004 23
Manual Reagent Placement (continued)
7. Perform a barcode scan
8. Click on REAGENT, Reagent Container Set
9. Locate the test and manually enter the container size, lot number and expiration date for that
reagent.
NOTE: Be aware that 3rd party assays need to have the lot#’s for R1 & R2 differentiated. Software
errors may occur if the lot#’s are not unique.
Example: R1 = 123456a
R2 = 123456b
10. Click the ‘o’ open button, select the test in the new windowÎOK. This starts the clock for the on
board stability. This needs to be done for both RTT1 & RTT2 if needed.
11. Select Save
NOTE: This field will only be active for the reagent whose barcode did not read
12. Check Reagent Inventory to be sure it has updated
NOTE: Be sure to remove the position from System Test List if the next new wedge of this
particular reagent reads properly.
GPE01006.004 24
Let’s Set up Shutdown Wash 2 (1800)
refer to STA
Refresh DI H2O
RTT 1 & 2 #56
CTT #16, #50 & #51
Execute
Refresh DI H2O
RTT 1 & 2 #50
CTT #16 (2400 only)
CTT #50 & #51
Execute 49 50
GPE01006.004 25
Let’s Setup Shutdown Wash 2 (1200) Refer to Quick Reference Guide
CTT-50 DI H2O
CTT-51 DI H2O (if applicable)
RTT1 - 45 DI H2O
RTT2 - 45 DI H2O
GPE01006.004 26
Daily Operation
GPE01006.004 1
GPE01006.004 2
Table of Contents:
GPE01006.004 3
ADVIA Chemistry Startup window
The Startup window is used to:
Set or clear the system data prior to startup.
Shut down and restart Windows.
Back up and restore system files.
If you click CANCEL in the Startup Window, the startup window closes, leaving you in Windows
NT. To start the ADVIA Chemistry software again;
double click on the HR start icon at the desktop
or restart Windows NT.
GPE01006.004 4
ADVIA Chemistry Daily Startup Procedures
The ADVIA must be in the READY mode to proceed with a NEW START
1. On the ADVIA Menu window, click System(s) to open the System menu, then click
Exit(x).
2. Click Yes when prompted, then click yes when prompted again.
3. The ADVIA Startup Window will appear.
a. be sure today’s date displays in the system date box before you continue.
Type today’s date in the box, in yyyymmdd format, if needed
b. Click New Start.
c. In the User name, and password boxes, type in the default password of
advia. Click OK. While the system starting message is on your screen,
startup is taking place.
d. Once startup has completed, the ADVIA 1800 menu panel and ADVIA 1800
Operation Panel appear. Reagent Inventory will also automatically open.
e. In the ADVIA 1800 Operation system status window, when the system
status WAIT appears in the upper right corner of the Operation Panel, the
START indicator is off, and the READY indicator is off…..click Initialize to
reinitialize the system.
WARNING to avoid possible injury and damage to the analyzer, make sure that all probes
and mixers are free to move without obstruction and that all analyzer covers are
in place.
GPE01006.004 5
ADVIA Chemistry Startup Procedures (continued)
Check the volumes of the containers in positions 53 to 56 on RTT1 and RTT2. Replenish
if necessary. To avoid contamination, we recommend replacing these with fresh
solutions in new wedges monthly for PW1 and PW2, and weekly for 10% cuvette wash
and water.
Inspect probes and mixing rods, clean if required with lint-free tissue
Inspect WUD and DWUD, probe wash stations, wipe down spills.
GPE01006.004 6
ADVIA Startup Procedures (continued)
4. From the Menu Panel, click on REAGENT, then click on Reagent Inventory
to check number of tests left in the reagent containers of RTT1 and RTT2.
Replace if necessary. NOTE: A barcode scan must be performed if
reagents are replaced.
Calibration
5. System Wash
1650 15 NA 49 50 49 50 49 50
2400 15 16 49 50 49 50 49 50
1200 15 NA 49 50 44 45 44 45
Material ISE DI H2O 10% cuvette DI water 10% DI water 10% DI water
detergent wash cuvette cuvette
wash wash
Type of J-cup / 10 ml tube 10 ml tube 10 ml 70 ml 70 ml 70 ml 70 ml
container adapter tube wedge wedge wedge wedge
GPE01006.004 7
START CONDITIONS Window
GPE01006.004 8
Starting a Sample Run
From STT
a. load samples
GPE01006.004 9
SAMPLE CONFIRMATION window
baby-smith
When running in the barcode mode, the Sample Confirmation window will appear after
all labels are read. The barcodes display next to their position numbers.
Names can also be entered in this window when a barcode is not available. This will
allow you to run a non barcoded sample in barcode mode rather than tray/cup mode.
This window will be displayed for only 20 seconds. Click timer off to keep this window
displayed while you verify the information.
If the box to the left of the barcode is checked, a priority analysis (STAT) was requested
for that sample. You can manually reprioritize by selecting or de-selecting these check
marks.
Once you have verified the correct information, click OK to begin the run.
To read the barcodes again before you begin dispensing sample, click Retry.
GPE01006.004 10
Confirmation Window Colors
A rerun is pending for this sample, but has not yet been
aspirated.
GPE01006.004 11
TEST RESULT MONITOR
GPE01006.004 12
The Test Result Monitor Window
To monitor an analysis while it is running:
3. When the run is completed, the operating mode display becomes END, then
returns to READY if samples were run from the STT; the system returns to the WAIT
mode when samples were run from a rack handler or Laboratory Automation system.
The left side of the Test Result Monitor window displays three panels.
The System Status panel shows the current operating mode of the system.
The Sample information panel shows the sample number and sample
position of the currently selected position on the STT/CTT graphic. If this is a
barcode analysis, the barcode number displays, and the position number is 0-
00.
The code panel shows the color codes used in the STT/CTT graphic to
represent sample status.
GPE01006.004 13
Processing status
The center of the window resembles the sample tray. The outer ring is the STT tray; the
inner ring is the CTT tray. Each tray position is a circle.
You can double click on each sample circle to display “Time to completion”
As the run continues, the circles containing samples change color. The colors indicate
the current status of each sample. The color codes are displayed at the lower left of the
window.
In the middle of the sample tray display, the tray (TT) number for the current run displays
in the TT No. list box. If this is a barcode analysis, the TT number is 0.
To view the status of prior tray samples that are still in process, click the down arrow of
the TT No. list box, then select the number of the tray you want to view.
The button bar at the top of the window displays the Sample Search and Rack or LAS.
Sample Info. buttons.
Press the Sample Search button to display a dialog box where you can
search for sample information by sample number by STT position (tray
number and sample position). The option to search by Rack/LAS position is
no longer valid.
The search returns a Sample Information window showing the sample
number, position number, sample status, and the time remaining to complete
the analysis of the sample.
If a rack handler or laboratory automation system (LAS) is in use, click the
Rack or LAS. Sample Info. button to display the following sample
information in the Rack or LAS. Sample Information window:
Sample number
Sample status (see the status color codes listed on the Test Result
Monitor window).
The time remaining to complete the processing of the sample.
GPE01006.004 14
REAL TIME MONITOR
The system monitors new calibration, control, and patient sample results in real time.
(Only patient and control sample results are transmitted to a host computer.) The
system reports results as each sample completes analysis. Results are displayed for
each sample after the sample analysis is complete. You can print sample results from
this window.
NOTE
You can view and print data only after all results for that sample display.
At the RealTime Monitor window, the latest tests display in the window on the left.
3. Select a Sample # from the list to display the specific data for a selected test.
a. Each test display consists of concentration, mark, and absorbance
information.
b. Selecting a sample while holding down the Ctrl key allows more than
one sample to be selected.
GPE01006.004 15
Calibration ID
ISESTDSL011
• ISE = ELECTROLYTE ASSAY
• STD = STANDARD
• S = SERUM
• L = LOW ( IF HIGH “H” )
• 01 = FIRST REQUEST OF DAY
• 1 = FIRST ASPIRATION
ISESTDUL011
• ISE = ELECTROLYTE ASSAY
• STD = STANDARD
• U = URINE
• L = LOW ( IF HIGH “H” )
• 01 = FIRST REQUEST OF DAY
• 1 = FIRST ASPIRATION
Chemistry Calibration ID
C0101
• C = CALIBRATOR/BLANK
• 01 = POSITION ON CTT
• 01 = # OF REQUEST FOR BLANK FOR CURRENT DAY
Control Sample ID
PA01
• P = PILOT
• A = CONTROL “A - Z”
• 01 = FIRST ASPIRATION OF CONTROL “A”
GPE01006.004 16
REAL TIME MONITOR
If the DETAIL format is selected, results will appear with the following measurement data:
ABS-RB ABS data minus reagent baseline; for reagent baseline this is the difference between the
current blank and the previous blank
E1 RRA- main wavelength absorbance value at check DP1 (point set in Analytical parameters (Chemistry)
window), typically defined as a point either before or after the addition of the second reagent (substrate
depletion check)
E2 Main wavelength absorbance data collected as the medial value of read points 6, 7 and 8. This value is
used for sample absorbance correction and to determine the absorbance of R1 + sample.
N Number of data points used to calculate the final concentration. For RRA this number must
be greater than or equal to 6
S A precision measurement of the data points within the read window. The difference of 2 absorbance
readings expressed as a % of the mean of the 2 readings; if S > 10 an * flag is reported (Variance = 10.0)
RRVNo Number of the reaction tray cuvettes used for this analysis
DTTNo Number of the dilution tray cuvettes used for this analysis
GPE01006.004 17
REAL TIME MONITOR (continued)
If the STANDARD format is selected, results will appear with recorded audit trail information, ie:
reagent lot numbers
GPE01006.004 18
User Code Set window
Use this window to enter a user code, user name, and user password for up to 50
operators who are authorized to use the ADVIA® Chemistry System.
2. In the please enter password box, type the user password. (not case sensitive)
GPE01006.004 19
Alarm (result) flags
Displayed on
d Photometric tests – N N N
Abnormal reaction
absorbance (lower)
ISE tests - dilution error
GPE01006.004 20
Alarm (result) flags
n Abnormal Insufficient valid data points Y Y N
number of
effective
points
s Photometric test - Y Y N
insufficient sample
ISE test - insufficient sample N N N
t Insufficient diluent N N N
GPE01006.004 21
REVIEW AND EDIT window
Click Request , then Review and Edit
The following information appears on the left side of the Review / Edit window.
Column Heading Column Contents
color coded circles showing sample test status
Order # (See Sample Test Status Color Codes for explanation of color code)
workorder sequence number
GPE01006.004 22
Review and Edit Window Color Coding:
GPE01006.004 23
To review/edit the sample results
1. On the ADVIA Menu window, click Request then click Review / Edit.
2. Select the date of data you wish to review (dropdown list)
3. In the Sample List, click on the order number of the sample you want to
review.
NOTE
On that day (current day in date dropdown list) sample results can be edited and rerun.
Sample results from previous days are locked and cannot be changed.
TIP: Use the standard format of the real-time print feature on the System Monitor
window to obtain additional information about the sample analysis.
GPE01006.004 24
Review/edit the sample results (cont’d)
In the Samp.type area, select the sample type you want represented in
the displayed list.
To search for a specific sample, use one or more of the following means
of identification in the Search condition area:
a. In the Samp.# area, select the check box, then type the sample
number.
b. In the LAS # Rack # area, select the check box, then type the
number of the rack containing the samples you want to review.
c. In the Comment 1 area, select the check box, then type the
applicable comment text.
d. In the Comment 2 area, select the check box, then type the
applicable comment text.
e. Execute.
GPE01006.004 25
Review/edit the sample results (cont’d)
CAUTION
To avoid incorrect positioning of results on the print report, items on the Print Report
form must be in the same order as they appear in the System Test List window. You
must not modify the order of the test items on the System Test List window after
initial setup. If you reposition any test items on the System Test List window, results for
samples already run could be associate with the wrong test name.
GPE01006.004 26
Exercise #1
1. With J-cups inserted in the barcoded tubes, ½ fill cups with pooled serum for testing.
Remember to wear your personal protective equipment!
2. Click on START from the Operation Panel
At the Ordinary Sample field Select Analyze
For Analyze Mode Select Barcode
Select START at the bottom of the window
Click OK when prompted
3. Click Timer Off on Sample Confirmation Window to review. Then click OK to
continue.
4. Open the Test Result Monitor Window
from the Upper Left Click Request
Select Test Result Monitor
5. Observe the ADVIA 1800 in progress and the changes to the Test Result
Monitor screen.
6. When results are complete (dots on Test Result Monitor Window are blue), click on
Request, then select Realtime Monitor to watch results post as they are completed.
Then look at Request Æ Review & Edit
7. Manually request a rerun on any assay on one of your samples. Verify that the
sample is still on the STT, and re-Start the analyzer so that the rerun is sampled and
processed.
8. From Review and Edit, select Print Report. Verify the following settings:
Sample type = Normal Sample
File data = today
Specification Range = Input Range
Number entry Format = you can choose:
a. Order No
b. Sample No.
c. Position No.
These are all referenced in the list on the left of this screen.
Last no. entry format = Number of sample
Report form file = c:\A002\ETC\chart.frm (use OPEN to find this file if it is
not already selected)
Enter whatever Start No., and Last No. you’d like from the list displayed at the
left.
GPE01006.004 27
On the ADVIA Menu window, click Request, then click Sample Log.
GPE01006.004 28
Delete a specific sample log entry.
a. Click the sample log entry you want to delete.
b. Click Clear, then click YES to confirm.
GPE01006.004 29
Functions in the ORDER ENTRY window:
1. 2. 3. 4. 5. 6. 7. 8. 9.
1. Entry Setup = You can select the auto increment mode; determine how the cursor moves;
select the color for each sample status.
2. Batch Entry = create multiple workorders.
3. Batch Func. = changes can be made to Batch orders
4. Delete tests = delete all tests on the displayed workorder
5. Delete workorder = deletes displayed workorder
6. Create profile = you can customize the profiles buttons on lower left of window
7. Create list = a load list will contain only workorders that have position numbers.
To include workorders currently without position numbers, use Create List.
8. View worklist = creates a custamizable list of pending and completed workorders
9. Host request = a way to manually download workorders from the host computer.
GPE01006.004 30
Creating Workorders
1. On the ADVIA Menu window, click Request then click Order Entry.
2. Click Routine or Interr. (STAT)
3. In the Posi.no. boxes, type the sample position number (tray ID# and position #).
(not necessary if using barcodes)
4. In the Samp.no. box, type the sample identification number (barcode).
5. Verify that the System Dilution mode, Container, Samp. Type, Dil.factor,
Comment (name), Sex and Collection date entries are correct..
6. Order tests.
In the Test table, click each test or ratio you want to run.
In the Test tbl.no. box, type the number of each test you want, then press
the period (.) key.
In the Profiles area, click each profile you want to run.
7. You can
Click Delete Tests to deselect all tests. Click Yes to confirm.
Click Delete Workorder to erase all entries. Click Yes to confirm.
8. Click Enter.
The Nbr.o.regist box may incremented. Automatic incrementing is assigned
using the Entry Setup window. If autoincrement is on, a new workorder appears
with the next sample number and position number incremented.
9. You can create another workorder or click Exit to leave.
If necessary, click New to clear the window for entry of the next workorder.
GPE01006.004 31
Order Entry Exercise #1:
Scheduling by Barcode
GPE01006.004 32
Order Entry Exercise #2:
Scheduling by Tray & Cup Position
1. Select Request from the upper left
Select Order Entry
2. Click on NEW
3. Select Routine as sample type
4. Place the cursor in the Position No. field
Enter 1 in the Tray field
Enter 1 in the Cup field
GPE01006.004 33
ORDER ENTRY – Batch Orders
GPE01006.004 34
Order Entry Exercise #3:
Scheduling by Tray and Cup position using the BATCH option
GPE01006.004 35
GPE01006.004 36
GPE01006.004 37
GPE01006.004 38
GPE01006.004 39
GPE01006.004 40
ISE’s on 1200, 1650, 1800 & 2400
white
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
1 GPE01006.004
LEARNING OBJECTIVES:
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
2 GPE01006.004
Ion Selective Electrodes (ISE’s)
In Serum or Urine
On 1650
1:33 dilution of the sample is used…..
28 μl of sample
+ 900 μl of buffer
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
3 GPE01006.004
ISE Measurement Sequence
3 stages:
Equilibration Chamber
Waste Nozzle
Dilution Bowl
Probe Guide
Stirrer motor
Ground Thermistor
Buffer in
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
5 GPE01006.004
1200 & 1800 ISE HYDRAULICS
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
6 GPE01006.004
(Older vs. )2400 ISE MODULE
Electrode
Cell Pot (Mixing chamber) Retainer
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
7 GPE01006.004
2400 ISE HYDRAULICS
BUFFER PUMP
WASTE
BLOCK
DRAIN PUMP
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
8 GPE01006.004
1650 ISE MODULE
1 DILUTION BOWL
2 SPACER
3 O-RING
4 PACKING GASKET
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
9 GPE01006.004
1650 ISE HYDRAULICS
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
10 GPE01006.004
1200/1650/1800/2400 Electrode Part Numbers
O RINGS
PN 073-0071-01
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
11 GPE01006.004
Calibration Schedule for ISE’s
ISE calibration is
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
12 GPE01006.004
Calibration of ISE’s from ISE Operation window
2. Click Maint.
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
13 GPE01006.004
Calibration of ISE’s (continued)
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
14 GPE01006.004
An Observation During Bufferprime
System Status indicates READY, but the ISE Operation banner tells you
the bufferprime is running, and gives you a countdown.
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
15 GPE01006.004
Calibration of ISE’s (continued)
7. Calibration is completed
when analyzer returns to
Ready with no error
messages.
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
16 GPE01006.004
Exercise:
Let’s calibrate ISE’s from ISE Operation window
Perform ISE Calibration Exercise
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
17 GPE01006.004
ISE CALIBRATION REAL TIME REPORT
First replicate is
a dummy
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
18 GPE01006.004
ISE Calibration Criteria
For the calibration to PASS, the data must
first be accurate & precise……
Accuracy (Range)
• Sample-Buffer readings need to fall within the
expected ranges
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
19 GPE01006.004
ISE Calibration Criteria (continued)
Precision (Clearance)
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
20 GPE01006.004
ISE Calibration Summary
If the summary
prints, you know
that the
ACCURACY
and PRECISION
checks have
passed……
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
21 GPE01006.004
Other criteria for successful calibrations
Slope (H, h, L, l)
After accuracy and precision, these criteria
Dilute factor (D) are also checked before a calibration attains
a PASS/FAIL status.
Thermistor variation (T)
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
22 GPE01006.004
ISE Calibration Criteria (Flag Limits)
Dilution factor= 25 - 60
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
23 GPE01006.004
Ways to View ISE Calibration Data
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
24 GPE01006.004
ISE Calibration Data
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
25 GPE01006.004
Calibration /RBL History
“latest” = all
calibrations will
appear in
chronological order
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
26 GPE01006.004
Data Entry Options…..
RealTime Monitor
In addition to
calibration data, it
displays and stores
electrode information,
the standards’ lot
numbers, and
expiration dates for
complete audit trail
capture…..
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
27 GPE01006.004
Data Entry Options…..
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
28 GPE01006.004
Data Entry Options…..
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
29 GPE01006.004
ISE Troubleshooting
Objectives:
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
30 GPE01006.004
ISE Calibration Criteria
Check Flag Explanation
Also called a range check, because it verifies that the Sample-Buffer readings fall within the expected
ranges:
Calibrator Na range K range Cl range
Serum low 270 to 430 (12001800 2400) 160 to 330 -160 to -40
270 to 400 (1650)
Accuracy C Serum high 380 to 510 290 to 460 -270 to -130 (12001800 2400)
-270 to -150 (1650)
Urine low -15 to 385 75 to 825 -180 to-55
Urine high 285 to 685 625 to 1375 -375 to -130
An alarm message will post:
“ISE Calibration range error”
Also called a clearance check. The difference between two consecutive aspirations of calibrator,
for Sample-Buffer, must be:
≤ 1.0 for serum
Precision * ≤ 2.0 for urine
An alarm message will post:
“Calibration Low STD error” or
“Calibration High STD error”
Accuracy and precision are assessed, and * or C flags will post, as the calibration progresses.
If accuracy and Precision passes, then a summary will be posted and the following criteria assessed.
Dilute D Must be 25 - 60
The following checks will not cause a calibration failure, but notify the operator of the need for part replacement.
Chloride Bias NG If the Chloride bias fall outside of +2 to -4 replace the electrode
Reference electrode d Reference electrode voltage below 500. If this happens, replace the electrode
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
31 GPE01006.004
Accuracy (Range) errors
C Flag
Common Causes:
- High and low calibrators are reversed
- Urine/Serum Calibrators are switched
- Hardware Failure (electrode or wire connector)
Troubleshooting Suggestions:
- Repour calibrators and repeat calibration
- Determine how many electrolytes are flagged……
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
33 GPE01006.004
Precision (clearance) errors
If the summary
prints, you know
that the
ACCURACY
and PRECISION
checks have
passed……
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
35 GPE01006.004
Slope
ELASTDSH011 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
Na 473.4 -156.5 -539.9 -106.6 -106.6 -124.0 -123.7
h, l, H, L flags K
Cl
361.5
-243.9
160.3
870.3
-101.2
1014.2
ELASTDSH012 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
Na 473.9 -168.7 -542.5 -106.2 -106.0 -123.8 -123.3
K 364.4 160.8 -103.6
Cl -243.4 870.8 1014.2
ELASTDSH013 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
Na 474.5 -169.5 -544.0 -106.2 -106.0 -123.8 -123.3
High slope: K
Cl
363.7
-243.5
160.1
870.3
-103.6
1013.8
ELASTDSL011 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
repeat calibration Na
K
318.6
259.0
-125.8
257.9
-544.4
-101.1
-106.2 -106.7 -123.8 -123.3
Calibration: Serum Meas. Date: 2004/11/10 15:17 ISE Serial Number 96BE16
Item Name Mark H-STD Buffer L-STD Buffer Slope Dilute
Na H 473.7 543.2 318.4 -544.4 65.2 48.4
K 364.2 101.7 258.6 -101.8 56.8 37.8
Cl -243.7 1012.6 154.0 1012.8 -55.9 38.6
TH1 106.4 106.5 123.8
TH2 124.3 123.6 123.7
Cl bias 0.9
Ref. Electrode 678.3
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
36 GPE01006.004
Slope (continued)
ELASTDSH011 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
Na 473.4 -156.5 -539.9 -106.6 -106.6 -124.0 -123.7
h, l, H, L flags K
Cl
361.5
-243.9
160.3
870.3
-101.2
1014.2
ELASTDSH012 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
Na 473.9 -168.7 -542.5 -106.2 -106.0 -123.8 -123.3
K 364.4 160.8 -103.6
Cl -243.4 870.8 1014.2
ELASTDSH013 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
Na 474.5 -169.5 -544.0 -106.2 -106.0 -123.8 -123.3
K 363.7 160.1 -103.6
Low slope: Cl
ELASTDSL011
-243.5
S-B Mark
870.3
Sample
1013.8
Buffer TH1 Sample Buffer TH2 Sample Buffer
Na 315.6 -126.7 -543.3 -106.6 -106.6 -124.0 -123.7
- old electrode K
Cl
257.4
-153.7
253.8
839.5
-103.5
1013.3
ELASTDSL012 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
must be replaced Na 318.4 -126.4 -544.8 -106.2 -106.0 -123.8 -123.3
K 258.9 257.0 -101.9
Cl -153.7 839.2 1012.9
ELASTDSL013 S-B Mark Sample Buffer TH1 Sample Buffer TH2 Sample Buffer
Na 318.6 -125.8 -544.4 -106.2 -106.7 -123.8 -123.3
K 259.0 257.9 -101.1
Cl -154.4 838.9 1013.3
replacement Na
K
473.7
364.2
543.2
101.7
318.4
258.6
-544.4
-101.8
65.2
56.8
48.4
37.8
Cl L -243.7 1012.6 154.0 1012.8 -37.9 38.6
TH1 106.4 106.5 123.8
TH2 124.3 123.6 123.7
Cl bias 0.9
Ref. Electrode 678.3
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
37 GPE01006.004
Case Study
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
38 GPE01006.004
Successful calibration No flags (mark), ISE Summary printed
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
39 GPE01006.004
Let’s perform the ISE Troubleshooting Exercise!
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
40 GPE01006.004
Photometric Calibration
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
1 GPE01006.004
Objectives
Perform
RBL
Single point calibration
Multipoint calibration
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
2 GPE01006.004
Definitions
Full Calibration
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
3 GPE01006.004
Required Calibration Materials
MultiPoint Assays
Specific multi-level calibrators (6)
Std 1 is used to calculate the
blank
Assigned to STT positions on
Virtual tray 98 or 99
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
4 GPE01006.004
Calibrator Preparation
Chem Cal
Follow package insert to reconstitute
Stability
Opened calibrator = 48 hrs @ 2-8°C
Exception (Tbil & Dbil): calibrator needs to be <8hrs old
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
5 GPE01006.004
Calibration Frequency
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
6 GPE01006.004
Scheduling from Start Conditions window
3. Ordinary or
2.
special
5. Always schedule QC
with your cal
6. Start the
cal & QC
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
7 GPE01006.004
Temporary Item Select Tests to calibrate
1650,1800 & 2400
;One-pt. smp.
~Ordinary calib.
Temp. Item Select
A default will be set up for daily
RBL’s and calibrations
Use Clear button to clear defaults
and select your tests
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
8 GPE01006.004
Cal/Temporary Item Select Tests to calibrate
X
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
9 GPE01006.004
Temporary Sample Select
Samples to put on the CTT
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
10 GPE01006.004
Schedule Multipoint Calibration
1650, 1800 & 2400
; Multipnt.smp. Analyze
~ Select tray
~ Ordinary calib.
NOTE: there is no difference between ordinary & special
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
11 GPE01006.004
Schedule Multipoint Calibration
1200
; Multipnt.smp. Analyze
~ Select tray ~98
~ Special calib.
NOTE: there is no difference between ordinary & special
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
14 GPE01006.004
What next…
Evaluate calibrations
On Reagent Inventory:
Pass or Fail
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
15 GPE01006.004
Test Result Monitor
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
16 GPE01006.004
Check the Calibration Status
Reagent Inventory
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
17 GPE01006.004
ID’s for Calibrators and QC
Chemistry Calibration ID
C0101
C = CALIBRATOR/BLANK
01 = POSITION ON CTT
01 = # OF REQUEST FOR BLANK FOR CURRENT DAY
Control Sample ID
PA01
P = PILOT
A = CONTROL “A - Z”
01 = FIRST ASPIRATION OF CONTROL “A”
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
18 GPE01006.004
Review & Print the Cal Report
Real Time Monitor
Review QC
Daily Precision Control
ADVIA QC
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
19 GPE01006.004
Additional Calibration Information
Calibration/RBL History
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
20 GPE01006.004
Print Calibration Reports
View Cal Curve
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
21 GPE01006.004
Calibration Troubleshooting
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
22 GPE01006.004
Troubleshooting Failed Calibrations
View Cal Curve
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
23 GPE01006.004
Troubleshooting Failed Calibrations
View Cal CurveÎRBL Check
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
24 GPE01006.004
Troubleshooting Failed Calibrations
View Cal CurveÎCal Check (One Point)
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
25 GPE01006.004
Troubleshooting Failed Calibrations
View Cal CurveÎCal Check (MultiPoint)
c Absorbance Change:
Compares the Low std to the High std to evaluate
the shape of the curve and reaction rates.
c
d Monotonicity:
Verifies the cal curve is increasing or decreasing at
d
each calibrator level. This is method dependent.
e
e RMS:
Comparison of a calculated root mean square
f
(RMS) to a predetermined maximum RMS.
f Calibration Fit:
Evaluates the calculated FV from the new curve to
the original FV for the assay.
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
26 GPE01006.004
Interpreting //// on Sample Results
1.0 main
SAMPLE ID 070012345 tray/cup position LAST NAME FIRST NAME sex age S dilution collection date
Test Conc User Mark ABS-RB ABS R1 R2 ABS1 ABS2 N S P RRV DTT
Alb 4.8 bay h 0.4698 0.6347 0.7093 0.6583 0.6347 2 0 59 53
ALP 205 bay h 0.0142 0.0241 0.3344 0.0508 0.0341 30 0.16 168 53
ALT 164 bay h -0.0329 -0.0333 1.6018 0.5592 -0.0323 24 0.23 56 53
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
27 GPE01006.004
Determine the STD-1 and STD-5 Absorbance values
-0-0955
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
28 GPE01006.004
Calibration Setup Windows
GPE01006.004
Objectives
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
30 GPE01006.004
Calibration Setup; single point assay
Single point
assays Single point assays
Enter cup position for
BLK posi.
STD posi.
Enter Coeff or FV
Enzyme FV’s come in
software
Occasionally updated via
Customer Bulletin
Others FV’s on product insert
Click Ctrl/Cal Setup
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
31 GPE01006.004
Calibration SetupÎCtrl/Cal Setup
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
32 GPE01006.004
Calibration Setup; Multi - point assay
Multi-point assays
Click Settings
TT No. ~98
Enter
Lot No.
Multi-point assays Lot Name (+ lot#)*
Exp. Date (yyyymmdd)
Cup position on STT
FV values from product insert
* By including the Lot# with the
name, it shows up in the Start
window
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
33 GPE01006.004
CalibrationÎTest Select
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
34 GPE01006.004
Auto Calibration Setup
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
35 GPE01006.004
Auto Calibration Setup cont’d
6. CalÎTest Select
7. ~4 Select test(s)
8. ~6 Select test(s)
9. Save
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
36 GPE01006.004
Priority
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
37 GPE01006.004
QUALITY CONTROL
Objectives
At the end of this module you will be able to:
Run and evaluate QC
Recognize the use of :
• Daily Precision Control window
File QC to the Cumulative File
GPE01006.004 1
GPE01006.004 2
Table of Contents
When do we run QC .................................................................................................5
How do we schedule QC? .......................................................................................5
Select the test(s) to be QC’d................................................................................6
Select the QC sample to be used ........................................................................7
Position QC on the CTT .......................................................................................8
Start the run ..........................................................................................................8
Daily Precision Control............................................................................................9
View option: X-Charts ..........................................................................................9
Format the chart display ....................................................................................10
Set the display order ..........................................................................................11
View Option: QC List.............................................................................................12
View Option: control data.....................................................................................13
Documenting Corrective Actions/Omitting outliers ............................................14
Daily Precision Control..........................................................................................15
Saving Daily QC to cumulative ..........................................................................17
QC Cumulative .......................................................................................................19
Set the presentation parameters .......................................................................19
Set the display order ..........................................................................................19
View the daily mean and range (x-r) charts ......................................................20
Specify the time interval you want to review ...................................................21
View the details and omit outliers .....................................................................22
View the cumulative QC statistics list (summary) ...........................................22
Establish the QC Cumulative mean and 1 SD values ......................................23
Print the cumulative charts................................................................................23
Real Time QC Window ...........................................................................................24
Setting presentation options for the X-Charts .................................................24
Select the controls and tests to be monitored .................................................25
Quality control overview........................................................................................26
QC & Calibration Setup………………………………………………………………….25
ADVIAQC……………………………………………………………………………………27
GPE01006.004 3
GPE01006.004 4
When do we run QC?
GPE01006.004 5
1. Select the test(s) to be QC’d
GPE01006.004 6
2. Select the QC sample to be used
GPE01006.004 7
3. Position QC on the CTT
GPE01006.004 8
QC Æ Daily Precision Control
Recalulated
Mean
If a control sample value exceeds the ±2 SD limit, the chart color changes from green
to red.
Control sample values omitted by the operator are colored red and are not included in
the ±2 SD evaluation.
GPE01006.004 9
To Format the chart display
Recalulated
Mean
GPE01006.004 10
To set the display order for the X-Charts
1. Click Display order test.
2. Each test you want to review must be in the Select Test(S) list.
a. To add another test, select the test you want to review in the Meas. Test(I) list, then
click >>.
b. To remove a test from the Select Test(S) list, click it then click <<.
3. To set the display order for a test, in the Select Test(S) list, select the test you want to
move, then click Up or Down as applicable.
4. Click OK when done.
Recalulated
Mean
GPE01006.004 11
View Option: QC List
Recalulated
Mean
GPE01006.004 12
View Option: control data
Recalulated
Mean
GPE01006.004 13
Documenting Corrective Actions/Omitting outliers:
Do not edit or erase a questionable control value, but omit it as follows:
Recalulated
Mean
1. In the Comment box for the control value you want to omit, type the reason for omitting
the result.
2. Click O button located beside the Comment box.
a. The control result information is now colored red and your user code appears in the
User box.
b. The control data points are colored red on the daily control charts.
c. The daily statistics are recalculated without the omitted control value.
3. Click Save when done. Do not click the Save control data button.
GPE01006.004 14
Daily Precision Control
Recalulated
Mean
The first control you a viewing (in this case A) will be plotted with a
black line and it’s corresponding mean and SD marks.
The overlay line will be yellow, and plotted to display it’s recovery
relative to it’s own mean and SD settings – though those numbers
are not displayed
Recalulated
Mean
GPE01006.004 15
Daily Precision Control
Saves any changes
(comments/omissions)
Recalulate
dMean
Recalulated
Mean
GPE01006.004 16
Saving Daily QC to cumulative:
Prior to each New Start, Daily QC statistics should be saved to the QC Cumulative file.
This can be done either of 2 ways. Either way you choose, the daily mean and fluctuation
range (R) for every test in each selected control are saved in the QC cumulative window
as a data point for the current day.
1. When exiting the software, one popup box will ask if you want
to save the daily QC to cumulative.
a. If you have already updated the QC Cumulative window or don't want to at this time,
click No in the first message box, then click Yes in the second one to close the
window.
2. To update the QC Cumulative window:
a. a. Read the displayed message box. Click Yes if the daily QC results are valid and
ready to be included in the cumulative QC data.
b. In the Control (C) list of the QC Cumulative Calculation dialog box, select the
controls you want to update. By default all controls are selected.
c. Click OK.
GPE01006.004 17
Saving Daily QC to cumulative (cont’d):
3. Use the QC Cumulative Calculation button…
Recalulated
Mean
a. After reviewing QC, and documenting corrective action and omitting if needed…
b. Click QC Cumulative Calculation.
c. Read the displayed message box. Click Yes if the daily QC results are valid and ready
to be included in the cumulative QC data.
d. In the Control (C) list of the QC Cumulative Calculation dialog box, select the
controls you want to update. By default all controls are selected.
e. Click OK.
IMPORTANT TIPS To avoid losing QC statistics, please observe the following:
• Update the QC Cumulative window before more than 20 daily control samples are run.
• The QC Daily Precision Control window can manage a maximum of 20 results for each
control product (level). Only the most current QC data are saved when more than twenty
controls are run in a day. (If an additional control sample is run, the oldest control sample
is deleted to make room for the new one. Similarly, if multiple repetitions (up to 5) of a
control are requested when 20 results have already been stored, the system aspirates the
sample multiple times and saves the results as the last ones in the sequence, discarding
the same number at the beginning of the sequence.)
• To avoid losing control results, you must perform a QC cumulative save on the QC Daily
Precision Control window at the end of each day, and perform a New Start when the
system is turned on the next day.
• Do not update the QC Cumulative window more than once each day.
• If you return to the QC Daily Precision Control window after performing a "QC
Cumulative save," you must not perform another update.
• Dot not perform an update when there is no daily control data available. The cumulative
data point will be deleted for today and no additional control data can be saved for the
day.
GPE01006.004 18
QC Cumulative Window
Recalulated
Mean
5. In the Drawing area, select how you want the chart presented.
6. When done, click OK.
GPE01006.004 19
QC Cumulative Window (continued)
To view the daily mean and range (x-r) charts
1. In the Display list, click X-Chart.
2. In the Control list, select the control you want to review.
Each control product is identified by a letter (A to Z)
3. The cumulative control charts for the selected control appear.
There are two charts for each test.
• The daily means are plotted on the top chart
• the daily ranges are plotted on the bottom chart. This is the range between the
highest and lowest value for that assay on that control for each day.
The heading identifies the test and provides the mean and SD for the daily means.
For the daily mean chart, the expected daily mean, the ±2 SD limits, and the ±3 SD limits
are plotted on the vertical axis for reference.
If a daily mean or range value exceeds the ±3 SD limit, the chart color changes from blue-
green to red. Corrective action for unsatisfactory control results
Daily mean and range values omitted by the operator are colored red and are not
included in the ±3 SD evaluation.
GPE01006.004 20
To specify the time interval you want to review
1. From the File(F) menu, click Specification or Value at date(T).
2. In the Beginning Day box, enter the date for the first day you want to review.
3. In the End Day box, enter the date for the last day you want to review.
• The QC Cumulative window can display data for up to 180 days.
4. Click OK.
• The new date range appears in the title bar for the QC Cumulative window.
• Each time a new date range is entered, the display order must be re-established.
Recalulated
Mean
GPE01006.004 21
To view the details and omit outliers
1. In the Display [view (V)] list, click Control Data.
2. In the Control list, select the control you want to review. Each control product is identified
by a letter (A to Z) that is assigned in the QC Sample Definition window.
3. In the Test list in the lower panel, select the test you want to review.
4. Review the following QC cumulative information for the test.
• Levey-Jennings (x-r) charts: top chart is for daily means and bottom chart is for daily
ranges
Optional: You can add data from other control products. In the Overlap (D) area, select
the desired control(s).
5. Do not edit or erase a questionable daily data point, but omit it as follows:
a. In the Comment box for the daily data point you want to omit, type the reason for
omitting the result.
b. Click O button located beside the Comment box.
• The daily data point information is now colored red and your user code appears in
the User box.
• The daily mean and range points are colored red on the cumulative charts.
• The cumulative statistics are recalculated without the omitted control value.
6. Click Save when done.
GPE01006.004 22
To use cumulative statistics to establish the QC Cumulative mean
and 1 SD values
Optional: You can use the cumulative statistics to establish the QC Cumulative mean and 1
SD values for a control on the Control Data Registration window.
NOTE
For commercial control products, the average and SD values are typically obtained
from the package insert for commercial control products. You can use this feature to
generate average and SD values for pooled samples used for control purposes.
1. In the Control list, select the desired control.
2. Click Save Control Data.
3. In the Test list of the Save Control Data dialog box, select the desired tests.
4. Click OK. The selected cumulative control information is sent to the Control Data
Registration window.
Optional: You can obtain a print preview of the cumulative charts as follows:
1. In the Control list, select the desired control.
2. In the File(F) menu, click Print Preview(V).
3. In the Test list of the Print-Test Select dialog box, select the desired tests.
4. Click OK. The cumulative mean and range charts for the selected tests appear.
GPE01006.004 23
Real Time QC Window
Use the Real Time QC window to review the performance for two controls simultaneously
using a Levey-Jennings chart (x-chart) or a twin chart.
NOTE: This window is intended to be used when two controls are run for each
test. However, you can still use this window if only one control is being run.
6. Click Automatic calculation to have the x-axis scaling varied for optimum presentation
depending on the number of data points (control results) available. For example, if 6
control results are available, the x-axis is scaled from 1 to 6.
7. Click Number of data for fixed x-axis scaling. That is, the axis is always scaled from 1 to
20 regardless of the number of control results.
8. In the Drawing area, select how you want the chart presented (data points only, line
graph only, or both).
9. When done, click OK.
GPE01006.004 24
To set up the Real-Time QC window (cont’d)
GPE01006.004 25
Quality Control Overview
When to run control samples
Refer to the ADVIA Chemistry system methods documentation for the quality control
recommendations.
If control results fail to meet the laboratory’s established criteria for acceptability, all
patient test results obtained in the unacceptable test run must be evaluated to
determine if patient test results were adversely affected.
The laboratory should take and document appropriate corrective actions, which may
include recalibration, before reporting patient results
Verify that the controls and reagents were prepared properly and have not expired.
GPE01006.004 26
Advia QC is…
An optional application for the collection, evaluation,
and management of QC data from the analyzer.
Features
Passive application – QC automatically captured from analyzer
Monitors 2 SD and 3SD Quality Control errors (Westgard Rules)
User Options:
Omit/Reinstate outliers
Document corrective action with user ID
Accept new data by
• individual assays or
• all data across assays at once
GPE01006.004 27
When launched from ADVIA QC, the first window in view is the QC Panel.
• Real-time updates
(dynamic screen)
• Alert indicators
• Panel Button= Control Group
• Max capacity = 300 Groups
Use Setup for customization
GPE01006.004 28
QC PANEL BUTTONS
Color is determined by the highest level alert of any
control file within a control group.
GPE01006.004 29
Select a QC Panel and click ‘OK’ to open the PRIMARY WINDOW
PRIMARY WINDOW
The Primary Window provides data and gives access to other functions.
Minimize program
Task bar
User ID
GPE01006.004 30
PRIMARY WINDOW VIEWS
GPE01006.004 31
PRIMARY WINDOW VIEWS (continued)
GPE01006.004 32
CONTROL TREE: Function
• The color border of primary branch indicates highest alert level for any data
within a file in that group.
• The color border of secondary branch indicates highest alert level for data in
that file.
-
Yellow: Unaccepted result; warning
GPE01006.004 33
GRAPH PANE
From the Control Tree, files are selected with checkmarks. Then the data is available as...
1 3 5 7 9
•Nine colors are used in order
as files are selected 2 4 6 8
Level 3
.….. Not in use closed control file
General Control File
GPE01006.004 34
DATA FUNCTION TOOLBAR
Tools used with the GRAPH PANE
Comment: Select to review or add
comments to a selected data point
•An omitted data point is encompassed in a circle & is not connected to any other data points
• Omit/reinstate is a toggle
Error Report
¾ From the Control Tree, select QC with rule violations. Error button becomes available.
¾ Click on Error.
¾ Graph and List displays details relating to the data points which violated selected rules
GPE01006.004 35
PRIMARY WINDOW: Summary Pane
• Target Data that was entered for the selected control definition
• Subset Summary of the statistical data for results currently in view in graph pane
Ns displays number of data points in the subset
• All Data Summary of all the statistical data for all selected results in the control file
Ns displays total number of data points gathered for that file
• Each line represents one data point-result, omitted result, or non-numeric result
GPE01006.004 36
PRIMARY WINDOW: List Pane
Provides control results and associated data for selected files.
These columns
can be resized
Ra Range Flag
S System flag
Us User
SD Standard deviation index
Target Target value from definition
These columns
can be resized
GPE01006.004 37
DATA FUNCTION TOOLBAR
Tools used with the Summary and List Panes
View control definition plus summary statistics for a single control file
Accept: Change status of all new data points in a selected control file to “accepted”
GPE01006.004 38
Task Bar - Setup Button
Customize
application of
Westguard Rules
of QC Evaluation
and
set flagging levels
GPE01006.004 39
Total used out of 500
Severity symbols
! Warning
X Error
Scroll buttons
To top
By page
By message
GPE01006.004 40
Task Bar - Reports
How are reports formatted?
• Preview, if desired
• Select Print
GPE01006.004 41
REPORTS: Select Type
QC Errors & Omissions lists each data point that has a QC error
or is an omitted data point for each file
GPE01006.004 42
Task Bar - Tools Button
Manages control files at the file level :
GPE01006.004 43
Help: Access ADVIA QC online help
Reduce: Reduce the size of the primary window to display only QC Panel button alert
indicator.
DSC
GPE01006.004 44
ISE Maintenance
NOTE: This document has been created using v1.02 of the ADVIA 1200
Electronic Operators Guide.
GPE01006.004 1
GPE01006.004 2
ISE Maintenance
Table of Contents
Daily
3 Days
Wash electrode lines (only if running 500 or more dialysis samples a month) 9
Weekly
Wash electrode lines (only if running 100 or more dialysis samples a month) 9
Monthly
Wash electrode lines (only if running less than100 dialysis samples a month 9
or 330 routine samples a day)
3 Months
As Required
GPE01006.004 3
GPE01006.004 4
ISE maintenance schedule
This schedule is based on your laboratory running 1000 ISE tests or less per day.
Customer replaceable parts
CAUTION
To avoid reporting faulty data, deterioration of reproducibility, or damage to the ISE
module and its parts, follow the precautions listed below:
• Immediately clean up any spills or leaks. Repair the source of leaks.
• Do not leave the chloride electrode wet pack open. The electrode will dry and become
inactive.
• Always condition electrodes before first use.
After you have finished the ISE maintenance tasks, do the following:
On the ISE Operation window, click Execute to the right of the word Initialize and click Yes. ISE Pri.
Wash will automatically return to ON.
GPE01006.004 5
Once a day If your laboratory runs dialysis samples
As required
If you notice contamination in the lines,
wash all lines.
Check and clean the cell tray (dilution
bowl and nozzle).
If the slope or selectivity is incorrect,
replace the bad electrode.
To prepare replacement electrodes for
operation, condition the electrodes.
GPE01006.004 6
ISE Maintenance – Once a day at Shutdown
Time: 5 minutes
Analyzer mode: Manual operation
NOTE
Executing Wash2 will automatically perform the electrode wash by default. This default
can be modified on the ISE parameters of Setting window.
1 On the ADVIA 1200 Menu panel, click Maint., then click ISE Operation.
2 Enter the CTT position number of the ISE detergent container in the Detergent
position area. The recommended CTT position number is 15.
3 In the Container field, select the type of container for the wash solution. The
recommended type of container is a Jcup in adapter.
4 Put the ISE detergent solution in the container and place it on the CTT position you
entered.
5 Click the Execute button to the right of the word Wash Electrode.
9 Click Exit.
GPE01006.004 7
ISE maintenance – Daily
GPE01006.004 8
ISE Maintenance – As required (refer to the schedule)
1 Remove the Na, K, Cl, and Ref electrodes and install the dummy electrode in their
place.
WARNING
ISE detergent is a sodium hypochlorite solution. When handling bleach wear
protective clothing, gloves, and safety glasses. It is harmful if swallowed and may
cause eye or skin irritation. In case of skin or eye contact, flush with large amounts of
water.
2 Remove the cap of the dummy electrode and put about 5 ml of ISE detergent solution
into the dummy electrode.
CAUTION
Be sure to tighten the cap on the dummy electrode. If the cap is loose or defective,
the ISE detergent solution may leak into the module and cause damage.
GPE01006.004 9
The wash starts and buffer prime is performed 10 times. The message "Line Wash 2
Running" and the approximate amount of time remaining appears in the display area.
7 Verify that there are no leaks or bubbles and that the buffer is going to the waste
during the priming cycle.
8 On the ISE Operation window, next to the Initialize button, click Execute.
9 On the ADVIA 1200 Operation panel, click ISE-Wash-ON.
10 Perform calibration and run controls.
GPE01006.004 10
ISE Maintenance – As required
1 On the ADVIA 1200 Menu panel, click Maint., then click ISE Operation.
3 Select Set
1 Pull out the system-solution shelf. Replace the buffer solution with another buffer bottle
containing 500 mL of deionized water.
GPE01006.004 12
ISE Maintenance – As required
Cleaning the dilution bowl (cell pot) and the waste-drain nozzle
1 On the ADVIA 1200 Menu panel, click Maint., then click ISE Operation.
3 Select Set
CAUTION
Be careful not to scratch or bend the nozzle. Damaging the nozzle may cause faulty
results.
5 Remove the thumb screw (1) that secures the liquid-supply nozzle and guide. Lift out
the nozzle with its guide.
6 Using deionized water, moisten a lint-free gauze and wipe off any crystals remaining
on the nozzle.
7 Using deionized water, moisten a lint-free gauze and wipe up water and dirt particles
remaining in the dilution bowl.
8 Replace the nozzle with its guide and secure it with the thumb screw.
GPE01006.004 13
9 On the ISE Operation window, enter 5 in the Bufferprime Times box, then click
Execute.
10 When prompted, click Yes to execute buffer prime.
11 Check that buffer is draining and not overflowing the Cell Pot.
CAUTION
Be careful not to bend the nozzle. Damaging the nozzle may cause faulty results.
1 Remove the thumb screw (1) to the waste-drain nozzle. Lift out the nozzle.
2 Using deionized water, moisten a lint-free gauze and clean the nozzle. If any crystals
remain, using a pointed toothpick, carefully scrape the crystals that are attached.
3 Replace the nozzle and secure it with the thumb screw.
4 On the ISE Operation window, enter 4 or 5 in the Bufferprime Times box, then click
Execute.
IMPORTANT
Verify that no buffer collects in the wash block. If there is water, crystals or other
matter may be clogging the drain.
GPE01006.004 14
After the dilution bowl and waste-drain nozzle are clean
GPE01006.004 15
3 Remove the sponge from the bottom
of the electrode case and place the
electrode to be conditioned back into
the case.
4 Using a dropper or pipette, pour 0.5
mL of pool serum into the flow path of
the electrode. Be sure to apply the
serum thoroughly.
5 Add buffer solution to the case,
covering the whole electrode with
solution. Let the electrode condition
overnight.
6 After aging is complete, take out the
electrode, wash it with deionized
water, and thoroughly wipe it off.
WARNING
To prevent infection from contacting
serum directly, wear suitable
protective gloves when you remove
the electrode from the solution.
NOTE
High-concentrated salt water is used as a preservation solution to maintain electrode
performance. When the electrode package is opened, wash the electrode with
sufficient water and wipe well before use. Small amounts of salt on the electrode may
cause rust on the electrode connector.
NOTE
To store the reference electrode, remove it from the ISE module, rinse it with water
then place it into an appropriate container. Cover the reference electrode with ADVIA
1200 ISE Buffer solution. Cover the container and store at 0 to 40 °F (-18 to 4.5 °C).
Rinse with water prior to next use.
7 Replace the electrodes on the instrument with the newly conditioned ones.
8 Calibration is performed as part of the electrode replacement. If the calibration fails,
repeat calibration again. If data continues to be unstable after electrode conditioning,
perform an electrode wash.
GPE01006.004 16
ISE Maintenance – As required
Replacing electrodes
l = 38.0 - 44.9
L < 38.0
1 On the ADVIA 1200 Menu panel, click Maint., then click ISE Operation.
3 Select Set
GPE01006.004 17
To install electrodes
NOTE
Make sure that the K and Na electrodes have been conditioned. When Cl and Ref
electrodes are taken out of their packaging, they are wet. Wipe the Cl electrode
thoroughly, and wash the Ref electrode using water.
CAUTION
If there is space between the electrode connections, the plate retaining the
electrodes cannot be closed. If you cannot close it, move each electrode left and
right little by little. Do not force the electrode. Fasten the thumbscrew tightly. If the
retaining plate loosens during measurement, liquid could leak, causing a problem
with the instrument.
4 On the ISE Operation window, click Execute to the right of the word Initialize. Click
Yes when prompted to execute arrangement.
5 In the Bufferprime area, enter 3 into the Times field. Click Execute, then click Yes
when prompted to execute buffer prime.
Verify that the liquid is discharged smoothly from the dilution bowl during priming. If
the liquid is increasing without being discharged, there is a leak, an incorrectly
positioned electrode, or a clog in the drain system. Immediately stop the instrument.
IMPORTANT
If clogging occurs, most probably the flow path is clogged inside the electrode.
Remove the Na and K electrodes, and check them by transmitted light to see
whether the flow path is clogged or not. You cannot do this for the Cl electrode due
to its construction. When in doubt, even if you cannot find a problem by the above
check, try replacing the electrode.
GPE01006.004 18
9 On the ADVIA 1200 Menu panel, click Maint., then click ISE Operation.
10 On the ISE Operation window, click Execute to the right of the word Calibration,
then click Yes when prompted to execute calibration.
NOTE
The electrodes may have to stabilize on the system before a successful calibration is
achieved.
GPE01006.004 19
System Maintenance
This document is designed for use during training classes only. Reference should be made to the
ADVIA Chemistry System Operator’s Guide and related Customer Bulletins for product labeling
information.
GPE01006.004 1
GPE01006.004 2
Table of Contents
GPE01006.004 3
AS REQUIRED MAINTENANCE
RESPONDING DURING A POWER FAILURE WHILE ELECTRIC POWER IS STILL OFF .............................................. 58
BACKING UP SYSTEM FILES TO A CD ............................................................................................................ 59
RESTORING SYSTEM FILES DONE BY TAS OR FSE ........................................................................................ 61
MAINTENANCE: SYSTEM MAINTENANCE MONITOR WINDOW ........................................................................... 62
ENTERING MAINTENANCE SCHEDULE INFORMATION ....................................................................................... 62
GPE01006.004 4
Maintenance schedule
Perform maintenance procedures at the recommended frequency to maintain the operating efficiency of your
analyzer. Procedures marked with an * may require more frequent performance (described in each procedure).
Print the maintenance procedures. When you select to print maintenance procedures, an Adobe file opens. At the
menu bar, select the Print icon.
Clean the reaction and dilution cuvette washers. Every 3 months (print 3-month maintenance)
Inspect the cuvette splash covers. Every 4 months (print 4-month maintenance)
Perform shutdown wash or modified shutdown Replace the reaction and dilution cuvettes.
wash.*
As required (print As required maintenance)
Perform additional ISE wash.*
Back up system files.
Record ISE slopes .
Replace probes
Weekly (print Weekly maintenance)
Replenish the RRV-bath oil bottle
Perform the weekly wash.*
Recover from a power failure.
Check the lamp coolant level.
Perform preventive cleaning of the wash station lines.
Perform lamp energy check.
Wash all ISE lines (if contaminated).
Perform cuvette blank (cell blank) measurement
Condition the ISE electrodes.
Clean the exterior module panels.
Replace ISE electrodes.
Monthly (print Monthly maintenance)
Clean the ISE dilution bowl and waste nozzle.
Clean the turntable interiors
GPE01006.004 5
Daily maintenance
1650-1800-2400
1200 1 Dilution Probe (DPP)
1 Sample Probe (SPP) 2 Sample Probe (SPP)
2 Reagent Probe 2 (RPP2) 3 Reagent Probe 1 (RPP1)
3 Reagent Probe 1 (RPP1)
• Replace clogged probes. See Replacing the Probes in the ‘As Required Maintenance’
section of this module.
TIP TO PREVENT THE PROBES FROM CLOGGING, PERFORM THE SHUTDOWN WASH
AND WEEKLY WASH AS SCHEDULED.
To position a probe allowing access for cleaning, use one of the one of the
Cleaning the probes using automatic advance probe motion (preferred method)
2. At the Manual Operation window, double-select the code for the probe you want
to move as follows:
Probe Code 1200 Code
Dilution 3.DPPLR NA
probe
3. Select Toggle the number of times necessary to move the probe to the
accessible location, then select Exit to close the probe window.
GPE01006.004 7
Daily maintenance
Power panel to
CAUTION
If you are performing this procedure with the
power off, manually support (lift) the probe to
avoid damaging the probe tip.
Be careful not to strike the probe against other
components on the analyzer.
2. Lift and manually rotate the probe arm to an accessible location. The movement
may feel a bit awkward and tight.
GPE01006.004 8
Daily maintenance
1. To access the DPP probe, use a Phillips screwdriver to remove the screws (1) that secure
the DPP shield to the analyzer panel.
2. Push the DPP shield to the right and slowly lift the DPP shield until it reaches approximately
a 90° angle, then gently lift the tab of the DPP shield and remove.
3. Unscrew all thumb screws, then remove the splash guard protective covers from the wash cups.
NOTE: If the system includes an anti-rotation bracket, avoid hitting it while removing the splash
guard protective cover.
4. Using alcohol prep pads or lint-free towels and alcohol (antiseptic ethanol), wipe the probe.
CAUTION
Do not use excessive force while cleaning, to avoid bending the probes.
NOTE verify the probe tips do not contain any imperfections, which could cause
contamination. Replace probes as necessary.
5. Visually inspect each probe wash cup for cleanliness, and clean them if necessary:
a. Pour deionized water into the wash cups and overflow sensor unit.
b. Using a lint-free towel, clean and dry these areas.
GPE01006.004 9
Daily maintenance
NOTE If an overflow error message displays, there may be water on the sensor. dry the
sensor.
9. Verify the analyzer mode is READY before performing any further actions.
10. Verify that the analyzer mode is READY before performing any further actions.
Inspecting and cleaning the mixing rods and mixer wash cups
The analyzer should be in the READY mode
1200 1650-1800-2400
1. Reagent Mixer 1 1. Dilution Mixer
2. Reagent Mixer 2 2. Reagent Mixer 1
3. Reagent Mixer 2
GPE01006.004 10
Inspecting and cleaning the mixing rods and mixer wash cups
1. Visually inspect each mixing rod and mixer wash cup for cleanliness.
2. Clean any dirty mixing rods or wash cups to avoid contamination of the mixers which results
in carryover:
a. With the instrument in the READY state, visually
verify that each mixing rod is at its upper limit.
b. Using a lint-free towel moistened with deionized
water, wipe the mixer rod.
CAUTION
Do not use excessive force while cleaning, to avoid
bending the mixing rods.
3. Inspect the mixing wash tank for cleanliness, then clean if dirty:
a. Pour deionized water into the mixer wash cup.
b. With a lint-free towel and cotton-tipped applicators, clean the mixer wash cup
CAUTION
Do not apply excessive force while cleaning, to avoid damaging the sensor.
NOTE
If an overflow error message appears, there is probably water on the sensor. Dry the
sensor.
GPE01006.004 11
Daily maintenance
NOTE
Cuvette covers are installed around the probes to prevent water and reagent from entering the
dilution and reaction cuvettes.
1200 1650-1800-2400
2. Using lint-free towels moistened with deionized water, wipe down the covers.
3. If splattering is extensive or enters the cuvettes, call your local technical support provider
or distributor.
GPE01006.004 12
Daily maintenance
Inspecting and cleaning the reaction (WUD) and dilution (DWUD) cuvette washers
(DWUD not applicable to 1200)
The analyzer should be in the READY mode
1. Inspect the exterior of the reaction cuvette washer (WUD) and dilution cuvette washer
(DWUD) tubing for cleanliness.
TIP
To keep the WUD or DWUD from clogging, perform this inspection in addition to the
startup, shutdown, and weekly washes. In the event of a clog, call your local technical
support provider or distributor for assistance.
3. Remove the wash head of each wash station (DWUD and WUD):
a. Cover nearby cuvettes with lint-free toweling to protect them from dust.
b. Loosen the wash station retaining screw (1) with a 4-mm hex wrench (see figure).
c. Lift the wash head off of the wash station assembly.
GPE01006.004 13
Daily maintenance
Inspecting and cleaning the reaction (WUD) and dilution (DWUD) cuvette washers
(continued)
4. Look for signs of wear or damage to the drying nozzle (2) on the WUD.
5. If there is wear or damage, call your local technical support provider or distributor.
6. Using alcohol prep pads or a lint-free towel moistened with alcohol or antiseptic ethanol,
wipe each wash head nozzle.
8. Verify the wash head nozzles are correctly centered in the cuvettes:
a. At the Menu Panel, select Maint, then select Manual operation.
b. At the Manual Operation window, double-select 14.DWUD or 23.WUD.
c. Select Move to slightly lower the washer nozzles and verify that they are correctly positioned.
If not, call your local technical support provider or distributor.
d. Confirm that the nozzles are correctly centered.
e. At the DWUD or WUD window, select Init., then select Exit, to raise the washer
nozzles.
9. At the Operation Panel, select Initialize, then verify the DWUD and the WUD
are in the up position and the instrument is in the READY state.
GPE01006.004 14
Daily Maintenance
Checking Reagents
6. Visually check the wash solutions on the CTT and both RTT trays.
7. Visually check the lamp coolant level. The fluid level should be between the green and
the red lines indicated on the reservoir. If low, fill with a 5% of lamp coolant additive.
GPE01006.004 15
Daily Maintenance
Checking Reagents
CAUTION
You must perform a barcode scan following any changes you make to the reagent
containers on the RTT1 and RTT2 trays. Failure to scan after moving, adding, or
deleting reagents can cause erroneous results.
If you accidentally switch barcoded reagents (R1 reagent on RTT-2 and R2 reagent on
RTT1), and perform a reagent barcode scan, an error message displays.
NOTE
If you replace only 1 of the reagents of a 2-reagent method and the replaced reagent
has a new lot number, after the barcode scan, the system alerts you that a reagent pair
does not exist. You must calibrate the new set of reagents before you continue to run
samples.
d. After replacing the reagents, do the following:
(1) At the Menu Panel, select Reagent, then select Reagent Inventory.
(2) Execute a Barcode Scan at the Reagent Inventory window.
e. Evaluate calibration status.
Recommendations after replacing reagents
You must calibrate before using a new lot of reagent.
Running controls is recommended before using a new container of the
same reagent lot.
GPE01006.004 16
Daily maintenance
Inspecting pumps
The analyzer should be in the READY mode
The pumps are located behind the system solution tray. Pull the solution tray forward to
facilitate inspection of the pumps.
A decrease in liquid flow or air bubbles in the lines may be due to a leaking pump. Inspect the
pumps for leaks daily to identify potential problems.
1800
1 SP
2 DIP
3 DOP
4 SCP
5 SRWP
6 RP-1
7 RP-2
1200
1 Sample pump (SP)
2 Sample and reagent wash pump (SRWP)
3 Reagent dispensing pump 1 (RP1)
4 Reagent dispensing pump 2 (RP2)
GPE01006.004 17
1. Inspect the SP and DIP pumps:
NOTE
Liquid leaking from the seal on the sample aspiration pump (SP) or the dilution aspiration
pump (DIP) flows to the drive lever unit. If the drive lever unit is wet, the pump seal must be
replaced.
GPE01006.004 18
Daily maintenance
The daily shutdown wash uses a detergent to clean the probe lines, reaction and dilution
cuvettes, and ISE components.
NOTE
The daily shutdown wash does not need to be performed on the day you perform the
weekly wash.
2. Ensure that the 10-mL tube at CTT (1) position #49 contains a 10% solution of Cuvette
Wash Solution, the cup at CTT (1) position #15 contains ISE Detergent and that the cup
at CTT (1) position 16 contains DI water.
3. Ensure that the bottle at RTT1 (2) and RTT2 (3) position #55 contains a 10% solution of
cuvette wash solution.
4. Ensure that the 10 ml tube at CTT (1) position #50 contains DI water.
5. Ensure that the bottle at RTT1 (2) and RTT2 (3) position #56 contains DI water.
NOTE
At your laboratory’s discretion, you may use other positions for 49 and 50 on the CTT,
position 55 and 56,RTT1, and RTT2. Change the entries for the alternate positions in the
appropriate fields on the WASH Set window.
6. At the WASH Set window: (For additional information concerning the Wash Set window,
refer to Using the Wash Set window in the On-line Operator’s Guide)
a. Select WASH 2.
b. Select 2 for Cycles.
c. Enter 49 in the CTT cup position 1st time field and 50 in the CTT cup position 2nd
time field.
d. Enter 55 in the RTT1 and RTT2 cup positions 1st time fields and 56 in the RTT1
and RTT2 cup positions 2nd time fields.
7. Select Execute.
GPE01006.004 19
Positions of wash solutions
RTT2 -44 RTT2 - 55 RTT2 - 49 RTT2 - 49 10% Cuvette Wash Solution (Daily)
5% Reagent Probe Wash 3 (Weekly)
Example of 1800
GPE01006.004 20
Daily maintenance
ISE Detergent Solution is automatically run through the ISE module as part of the shutdown
wash procedure (WASH2).
Manually perform additional ISE washes (described in the following procedure) once
per shift under either of the following conditions:
¾ Do NOT perform the ISE electrode wash more than 3 times per day (once as part of the
shutdown wash and twice on a per shift basis).
¾ Pour fresh ISE Detergent into a cup, not a tube, before each wash, from the CTT.
GPE01006.004 21
Record ISE slopes daily
Once a day, record the slopes from a successful ISE
calibration on the Maintenance Log. The slopes are
provided on the:
GPE01006.004 22
Weekly maintenance
The weekly wash is the same as the daily shutdown wash, except that a 5% solution of
reagent probe wash 3 is substituted for 10% cuvette wash solution.
NOTE
The daily shutdown wash is not required on the day you perform the weekly wash.
2. Ensure that the 10-mL tube at CTT position #49 contains a 5% solution of reagent
probe wash 3, and that the cup at CTT position #15 contains ISE detergent.
3. Ensure that the bottles at RTT1 and RTT2 contain a 5% solution of reagent probe
wash 3. See chart p.20 for positions
4. Ensure that the 10-mL tube at CTT position #50 contains DI water.
5. Ensure that the bottles at RTT1 and RTT2 contains DI water. See chart p.20 for
positions
NOTE
At your laboratory’s discretion, you may use positions other than 49 and 50 on the CTT,
RTT1, and RTT2. Change the entries for the alternate positions in the appropriate fields on
the WASH Set window.
6. At the WASH Set window (For additional information concerning the Wash Set window,
refer to Using the Wash Set window in the On-line Operator’s Guide)
a. Select WASH2.
b. Select 2 for Cycles (the default setting).
c. Enter 49 in the CTT cup position 1st time field and 50 in the CTT cup position
2nd time field.
d. Enter the RTT1 and RTT2 cup positions 1st time fields in the RTT1 and RTT2 cup
positions 2nd time field. See chart p.19 for positions
7. Select Execute.
8. After the wash, check the lamp energy and run the cell blank measurement test.
GPE01006.004 23
Weekly maintenance
The lamp is cooled by circulating liquid coolant. As the volume of coolant decreases, the heat
of the lamp increases.
NOTE
Check the lamp coolant level weekly and whenever the system generates a lamp coolant
warning and turns off the lamp.
1. Lift off the lamp access cover (1) to gain access to the lamp coolant reservoir.
GPE01006.004 24
Weekly maintenance
1. At the
Menu
panel,
select
Maint, then
select
Lamp
Energy
Monitor.
1200 45
1800 56
3. At the Lamp Energy Monitor window, in the Luminous Energy Check area, verify the settings:
a. Type 1000 in the Meas. times field.
b. Type 25 in the Meas. cycle field.
c. Select AD.
d. Select Auto.
4. Select Check Energy. The Lamp Energy Monitor dialog box displays.
5. Type # in the RTT1 bottle posi. field, then select 4: 70 mL brown for the Container field.
GPE01006.004 25
Weekly maintenance
Checking lamp energy (continued)
7. Select Meas. Energy. The message, "Execute the lamp energy check?" displays.
8. Select OK.
9. On the Lamp Energy Monitor window, select Collect Data.
10. Calculate the scatter plot:
a. Note the value of the 340-nm AD count field.
b. Add 50 to the 340-nm AD count and type the sum in the top field to the left of the
graph, then press Enter.
c. Subtract 50 from the 340-nm AD count (noted in step 10 a) and type the difference in
the bottom field, to the left of the graph, then press Enter.
d. The lamp energy displays as a scatter plot:
11. Replace the lamp if any of the following is true:
• The AD points are not within ±40 of the center line
• The voltage reading (Volts column) for any of the 14 wavelengths is outside the range
5.0 to 9.0 volts
• The attenuation [ATTENU(%) column] for any of the 14 wavelengths falls below 80%
NOTE: Only if you replaced the lamp, select Regist Data, then select OK in the
Registration window.
12. If not, proceed to step 13.
13. At the Menu Panel, select System(s), then select Screen Print to print the window
contents.
14. Save the printout with your laboratory maintenance records.
15. Exit the Lamp Energy Monitor window, then select Initialize to switch the system from the
WAIT state to the READY state.
GPE01006.004 26
Weekly maintenance
Reaction cuvettes cause changes in absorption with use. After the weekly wash, perform the
cuvette blank measurement to determine the change. The cuvette blank is only run weekly,
even if your lab runs the Weekly Wash as a daily procedure.
The measured cuvette blank values for all cuvettes and a list of abnormal cells are printed
in approximately 15 minutes.
GPE01006.004 27
Weekly maintenance
Abnormal cuvettes
A list containing abnormal cuvettes is printed as part of the cell blank. The list contains
HmarksH indicating abnormality. Cuvettes appearing on the list are not used for analysis.
If the absorbance of a cuvette differs by more than 0.4 (typical setting) from the mean
absorbance reading from all cuvettes, that cuvette is marked with an "H" or "L."
In a cell blank measurement, 2 measurements are taken and their mean value is the actual
registered value. If the 2 measurements differ by more than the cell breakup limit value
(usually set at 0.1), the measurement is marked "N", and the cuvette is considered
abnormal.
In a cell blank measurement, if the cell’s absorbance differs from its registered absorbance
by more than the “skip absorbance value” (usually set at 0.04), the cell is considered
abnormal (marked E), and therefore not used for analysis.
Cuvettes exceeding the lamp energy voltage limits are skipped (marked U or D), and
therefore not used for analysis.
Reference value
The reference value (the average value of the measurements of all cuvettes) remains the
same until the next measurement.
GPE01006.004 28
Weekly maintenance
1. Save all data and close all open windows on the workstation.
WARNING
Turn off the main power switch at the rear of the analyzer, to avoid catching the toweling in
the cooling fans.
WARNING
GPE01006.004 29
Weekly maintenance
5. Dampen lint-free towels with the solution and wipe the following exterior surfaces:
• top cover
• side panels
• front panel
• rear panel
8. Return the analyzer to Operating Mode and the rack handler to the ON mode (if
applicable).
The Power indicator (2) is on, and the System Stop (3), Start (4), and Ready (5)
indicators are flashing.
2. At the Startup window, start the workstation by selecting New Start or Re-start.
WARNING
Make sure all probes and mixers are free to move without obstruction and all
analyzer covers are in place, to avoid possible injury and damage to the analyzer.
4. Verify the Operating mode field displays READY before performing any further
actions.
GPE01006.004 30
Monthly maintenance
NOTE
Use the two procedures that follow to clean inside the STT/CTT housing and RTT refrigerated
housing to remove accumulated sample, reagent, dust, and other materials.
GPE01006.004 31
Monthly maintenance
3. Remove the CTT (1) and STT (2) trays (see figure below):.
a. Pull up on the 2 Nylatch fasteners (3) securing the CTT tray in place.
b. Lift out the CTT tray by the center handle (4)
c. Pull up the 2 Nylatch fasteners (5) securing the STT tray in place.
d. Lift out the STT tray by the two metal handles (6).
4. Using lint-free towels, wipe the interior of the STT and CTT housings.
5. Replace the CTT and STT trays and covers (see figure above):
a. Orient each tray to the locator screw (7).
b. Ensure the trays are securely in position, then push down the 4 Nylatch fasteners (3 and
5) to lock the trays in place.
c. Replace the CTT cover and the STT evaporation cover.
d. Position the DPP splash tray, then secure it in place with the 3 Phillips screws.
e. Replace the DPP shield and secure it in place with the 2 Phillips screws.
1 CTT Tray
2 STT Tray
3 CTT Nylatch Fasteners - 2 places
4 CTT Handle
5 STT Nylatch Fasteners - 2 places
6 STT Handles - 2 places
7 Locator Screw
GPE01006.004 32
Monthly maintenance
2. Using lint-free towels, wipe the interior of the refrigerated housing and clean the glass
window of the reagent bar code reader.
3. Replace the reagent tray, aligning the 3 holes in the center of the tray with the 3 posts on
the hub.
4. Replace the cover, aligning the rectangular hole in the cover with the tab on the rim of
the housing.
GPE01006.004 33
To clean inside the reagent tray refrigerated housing on 1200-1650 & 2400
1 Using the hand grip opening on the top cover, remove the cover.
2 Remove the reagent tray (RTT1 and RTT2) cover.
a Move the cover latch toward the front of the analyzer.
b Using the handle, lift and remove the cover.
8 Grasp the hand grip on the top protective cover and lower the cover securely in place
GPE01006.004 34
Monthly maintenance
Cleaning or replacing the wash solution reagent containers
1. Remove the wash solution reagent containers from RTT1 and RTT2 See Table below
2. Replace the containers with new ones or clean the old containers with DI water.
3. Refill the containers with fresh solutions as specified in the table below.
Probe Wash 1 42 53 47
Probe Wash 2 43 54 48
10% Cuvette Wash 44 55 49
DI Water 45 56 50
GPE01006.004 35
Monthly maintenance
cuvette wash
cuvette conditioner
Materials required BIOHAZARD
Time:10 minutes Wear personal protective equipment.
Use universal precautions.
Analyzer mode: READY
NOTE
Cleaning each bottle can be performed at time of bottle refill, but must be performed at least
once a month.
CAUTION
Make a note of the bottle position on the shelf, to avoid mixing up the fluid bottles.
2. Empty the remaining contents of the bottle.
3. Rinse the bottle with deionized water and drain well.
4. Refill the bottle with the solution matching the label on the container.
5. Replace the bottle in the same position on the shelf in the cabinet.
GPE01006.004 36
Monthly maintenance
connect the level sensor connector, then push the connector in and
turn clockwise.
Be certain that all straws and filters are at the bottom of the solution bottle.
GPE01006.004 37
Monthly maintenance
NOTE
The chiller filter is located on the right inside bottom shelf of the analyzer cabinet. It must be
accessed through the panel door on the right side of the analyzer.
1. On the right side of the analyzer, push and release the panel door to gain access
to the chiller unit.
5. Slide the filter back in place and close the panel on the right side of the cabinet.
GPE01006.004 38
Monthly maintenance
Cleaning the large water pump (LWP) filter (1200 & 1800 only)
NOTE
This filter removes dust and foreign matter from the water circulation system. This water dilutes
samples and is used to clean the probes, mixers, and cuvettes.
CAUTION
TURN THE ANALYZER OFF WHILE CLEANING THE LWP FILTER.
IMPORTANT
Be sure to label the right and left sides of the LWP filter before removing it.
Large water pump (LWP) filter
4. Gently pull the filter toward you to gain access to the attached, blue inlet and outlet
tubes.
GPE01006.004 39
Monthly maintenance
b. Repeat for the other tube to completely remove the filter from the system.
c. Label the tubing, if necessary, to maintain proper
orientation.
6. Unscrew the filter and set aside the black o-ring
1. Prepare 1.5 liters of 5% Probe Wash 3 solution diluted with deionized water.
3. Install clean cuvette segments on the dilution tray (DTT) and fasten the
thumbscrews by hand.
NOTE Verify that the Operating mode field displays READY before performing
any further actions.
5. Perform the daily shutdown WASH 2 routine; then verify the operation.
GPE01006.004 41
Every 3 months maintenance
2. Lift and remove the access panel in front of the Rotating Reaction Tray to expose the lamp
housing.
WARNING
The lamp housing is hot. To avoid burns, allow it to cool down (approximately 10 minutes)
before touching any components.
GPE01006.004 42
Replacing the lamp (continued)
3. Loosen the lead wire connectors (1 and 2) and remove the wires.
4. Unfasten the lamp screws (3) on the plate (4) and remove the halogen lamp from the
housing.
5. When installing the new lamp, align the hole (1) to the locating pin (2).
Do not touch the glass portion of the lamp. Touching the lamp could damaging it. If the
lamp is dirty, clean it using lint-free toweling moistened with ethanol.
6. Install the lamp screws.
7. Install the lead wires and fasten the knobs.
8. Replace the access panel.
9. Return the system to Operating Mode.
10. Wait 40 minutes for the lamp to stabilize.
11. Then, check the lamp energy.
12. Perform the cell blank measurement test.
NOTE
Bayer HealthCare recommends that the assays on the system be calibrated after the lamp
is replaced.
13. Run QC controls to verify that all assays are within the laboratory’s established
control ranges.
GPE01006.004 43
Every 4 months maintenance
1 Thumbscrew
2 Tab
3 Detector Unit
WARNING
Turn off the power before removing or replacing cuvettes in order to allow the RRV to move
freely.
CAUTION
Be careful not to get RRV bath oil inside the cuvette. If you do, allow the cuvette to dry
overnight.
CAUTION
Be careful not to drop the cuvette set screws into other components of the instrument.
Do not remove the cuvette if it is in front of the detector (3).
b. Hold the cuvette set by the tab (2) and lift it from the tray.
c. To remove the cuvette sets located by the detector unit or under the cuvette wash
station (WUD), rotate the reaction tray by hand until the cuvettes are in an accessible
location.
CAUTION
d. Work away from the area of the lamp housing. You want to avoid
splashing oil on any of the lamp or spectrophotometer components.
GPE01006.004 44
Every 4 months maintenance
3. While the RRV cuvette segments are removed from the system, make a visual
inspection of the oil. Remove any foreign substances, if present.
4. Install the new cuvette sets on the RRV and fasten the set screws.
CAUTION
To avoid damaging the cuvettes, do not to touch or scratch the cuvette surfaces or wipe
the cuvette interior.
4. Install the new cuvette sets on the DTT and fasten the set screws by hand.
6. Perform the daily shutdown WASH 2 routine and verify the operation.
7. Perform the cell blank measurement, and if the cell blank run was completed successfully,
save the results.
GPE01006.004 45
Every 4 months maintenance
Cleaning the on-board pure-water bottle filter and the reagent bottle filters
There is a hose in each of the Pure Water, Saline, Cuvette Detergent, and Cuvette Conditioner
bottles. The filter in the end of each of these hoses should be checked and cleaned every 4
months. Refer to the On-line Operator’s Guide for specific instructions.
Clogged filters create an insufficient flow rate and produce air bubbles.
NOTE
A set of filters is included in the supplies kit. To avoid system down-time, replace the filters with
those in the kit, resume operation, then clean and store the removed filters for the next
scheduled maintenance.
If any of the filters are ripped or damaged, replace them with new filters.
Filters can be cleaned in a freshly made 10% solution of water and bleach. Rinse thoroughly
with distilled water before replacing the filter on the system.
GPE01006.004 46
As required maintenance
NOTE
Use this procedure to replace SPP and RPP probes not equipped with crash detection. For
dilution probes (DPP) equipped with crash detection, go to: Replacing DPP probes - with crash
detection
Caution
Manually support the probe and be careful not to strike it against anything on the analyzer, to avoid
damaging the probe tip when the power is off.
2. Cover the cuvettes, wash cups, and other analyzer surfaces with lint-free towels to catch
any screws that might fall.
4. Loosen but do not remove the setscrews (1) on each side of the probe cover. Lift the
cover off the probe.
GPE01006.004 47
As required maintenance
1. Terminal 2
2. Joint Connector
3. Joint Holder
4. Phillips Screws (4 places)
5. Probe
6. Terminal 1
7. Flange
5. Using pliers, loosen the joint connector (2) counterclockwise, then unfasten
and remove it by hand.
8. Slowly insert the new probe through the guide hole (5) until the flange (4) is seated
against terminal 1 (6).
9. Verify that the probe is correctly positioned in terminal 2 (1) and the joint holder (3).
10. Tighten the 4 Phillips screws (4) while maintaining the probe position in terminal 2 and
the joint holder.
Caution
Do not cross thread or force the joint connector in too far, to avoid damaging the threads or
introducing leaks or air bubbles.
12. Replace the probe-arm cover and tighten the two probe cover screws.
GPE01006.004 48
As required maintenance
13. Lift up the probe arm to the end of its travel, then manually rotate the probe over the
probe wash cup but not within the wash port.
15. At the Menu Panel, select Maint, then select Manual Operation to move the probes.
16. At the Operation Panel, select Initialize to return the probes back to home
(over the wash cups).
17. At the Operation Panel, select PRIME, PRIME 2, and then Execute to ensure
proper water flow through the probe.
NOTE
Make sure that no water is leaking from the joint connector.
GPE01006.004 49
Chemistry Probe Cleaning Tool Chemistry Probe Cleaning Tool
Purpose: Description:
To provide the Chemistry system Operator The Chemistry Probe Cleaning Kit consists of
with the means to flush obstructions from a a 10 mL syringe coupled to a Push-In fitting
pipette probe. which can be screwed into the ADVIA®
Chemistry pipette probe.
Kit contents:
1) 10 mL Luer-Lock Syringe
2) Barbed Female Luer Fitting
3) Short Piece of Polyurethane Tubing
4) Mini Push-In Fitting
5) Instructions
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Assembled Tool:
Instructions:
1 Ensure the system is in the Standby Push-In Fitting
mode.
2 Remove the pipette probe by following
the instructions (Replacing the Probes)
in the Analyzer As Required section Tubing
under Maintenance of the Operator
Guide.
3 While holding the joint holder end of the
pipette probe secure, screw in the Push- Luer Fitting
In fitting of the tool into the probe clock
wise.
Syringe
(over)
06.14.2006
GPE01006.004 50
Instructions (continued):
Instructions (continued):
4 Unscrew the syringe at the luer fitting.
10 If bleach was used to clean the probe
5 Fill the syringe with DI-H2O and reattach then follow the bleach with a syringe of
it to the luer fitting. DI-H2O to remove any traces of bleach.
6 Depress the syringe plunger while 11 Remove the tool by holding the joint
holding the pipette probe over a sink or holder end of the screw and turning the
suitable container. Push-In fitting counter clockwise.
7 Observe that any obstruction is cleared
and a steady uniform stream exits the
probe tip.
8 If the obstruction does not clear then
immerse the probe tip in a beaker of DI-
H2O and move the syringe plunger back
and forth to dislodge the obstruction.
9 After the obstruction is cleared,
optionally a 10% solution of bleached
may be used to clean the probe further.
WARNING
06.14.2006
GPE01006.004 51
As required maintenance
2. Using a Phillips screwdriver, remove the screws that secure the DPP shield to the
analyzer panel.
3. Push the DPP shield to the right and slowly lift it until it reaches approximately a 90°
angle, then gently lift the tab of the DPP shield and remove.
4. Unscrew all thumb screws and remove the splash guard protective covers from the wash
cups.
NOTE
If the system includes an anti-rotation bracket, avoid hitting it while removing the splash guard
protective cover.
CAUTION
Manually support the probe to avoid damaging the probe tip when the power is off. Be
careful not to strike it against any other parts on the analyzer.
6. To catch any screws that might fall, cover the cuvettes, wash cups, and other analyzer
surfaces with lint-free towels.
7. Lift and manually rotate the probe over the sample tray.
8. Loosen but do not remove the screws on each side of the probe cover (1).
9. Lift the cover from the probe arm and set it aside.
GPE01006.004 52
As required maintenance
1 Probe Tubing
2 Joint Connector
3 Joint Holder
4 Probe Wire Screw
5 Black Wire
6 Spring Clips
7 Probe Guide
8 Probe
9 Wire Lock Screw
1. Gently loosen the probe joint connector (2) (use pliers, if necessary), then slide it back on
the tubing (1) approximately 1 cm.
2. To remove it from the end of the probe body, gently flex and pull back on the tubing (1).
CAUTION Be careful not to damage the flared end or kink the tube.
3. Loosen but do not remove the probe wire screw (9), then remove the orange probe wire
from the post.
4. Hold the probe arm securely and open the two spring clips (6) by grasping each at the
side closest to the black wire (5), then gently raising each to an open, locked position.
CAUTION There is some spring resistance when attempting to open the clips. Do
not allow the probe arm to swing side to side when opening the clips.
7. Gently snap the gold joint holder (3) from its post.
8. Gently lift the probe (8) up through the probe guide (7), then carefully remove it from the
probe arm.
GPE01006.004 53
As required maintenance
1. Carefully insert the new probe into the probe guide (7), then route the other end through
the joint holder (3).
CAUTION
Do not allow the clips to snap on the probe shaft. This may damage to the probe.
3. Reconnect the black wire (5) under the wire lock screw (4) and tighten the screw.
CAUTION
Do not flex the wire more than necessary. Over-flexing may damage the wire.
CAUTION
If the screw does not fully tighten, or the standoff spins, tighten the screw on the probe
arm base until the standoff no longer spins; otherwise the liquid-level-sensing capability
may be adversely affected.
4. Slip the orange wire onto the probe wire screw (9), then tighten the screw.
5. Carefully flex the tubing (1) and slip the flared end into the joint connector (2).
6. Slide the knurled nut of the joint connector (2) into the joint holder (3) and carefully tighten
until snug.
CAUTION
Do not cross thread or force the joint connector in too far, to avoid damaging the threads
or introducing leaks or air bubbles.
8. Replace the probe arm cover and tighten the two probe cover screws .
9. Replace the splash guards and secure them back in place with the thumb screws.
10. Manually lift and rotate the probe over the probe wash cup but not within the wash port.
GPE01006.004 54
As required maintenance
4. In the Probe posi.adjust area, select Position adjust start to move all the probes (DPP,
SPP, RPP1, and RPP2) over cuvettes.
5. Ensure that the probe is perpendicular to the arm and centered over the cuvette.
If it is not, call your local technical support provider or distributor.
6. At the Operation Panel, select Initialize to return the probes back to home (over the
wash cups).
7. At the Operation Panel, select PRIME, then PRIME 2, then Execute to ensure proper
water flow through the probe.
NOTE
Make sure that no water is leaking from the joint connector (2).
GPE01006.004 55
As required maintenance
If you experience a problem with clogs in the wash station aspiration nozzles and lines, use
this procedure to clean the WUD and DWUD wash station aspiration nozzles and lines. Refer
to the On-line Operator’s Guide to follow specific instructions for this procedure.
The Wash Head is removed (1) and the nozzles placed into a tray (dryer paddle (1) remains
outside the tray)
You will activate a sequence which allows you to flush these lines with distilled water, then
a cleaning solution, and then water to rinse.
GPE01006.004 56
As required maintenance
Shutting down and Powering up the ADVIA Chemistry System (Recovering from a
Power Failure)
1. Turn off the workstation and analyzer power by performing the normal shutdown operation:.
1. From the System (s) menu, on the Menu Panel, select Exit.
4. Select Shutdown.
5. After a brief delay, the message box displays to allow you to turn off the computer.
2. If you expect the power supply to be cut off for a long period of time, refrigerate the
reagents.
3. When the power supply returns, perform the normal startup procedure
1. After you apply power and the Windows operating system loads, the ADVIA
Chemistry system Startup window displays.
3. At the system power panel, set the Operate/Standby switch (1) to Operate.
5. When the Start and Ready indicators are off, and the Initialize button at the
Operation Panel activates (turns black), select Initialize.
GPE01006.004 57
As required maintenance
2. When the Startup window opens, turn the Operate/Standby switch to Operate.
3. Select the system reset button on the analyzer unit power supply panel.
6. If possible, repeat the task that you were performing prior to the power failure and
verify that the data were stored.
7. If reagent was dispensed, you must perform a Weekly WASH2 before resuming
operation..
1. If the Startup window is open, select Shutdown and perform the normal
shutdown operation.
2. Turn off the workstation power and turn the Operate/Standby switch on the
analyzer to Standby.
4. Perform the normal startup operation and open the Startup window.
7. If possible, repeat the task prior to the power failure and verify that the data were
stored.
8. If reagent was dispensed, you must perform a Weekly Wash before resuming
operation.
GPE01006.004 58
As required maintenance
Materials required: • 1 Blank CD-R (for systems equipped with a read/write CD drive)
GPE01006.004 59
4. At the ADVIA Backup window, select Make a Backup Copy, then select the Target Files
to be backed up from the following options:
• System Files - Approximately 30 MB of disk space is required.
• Data Files - Disk space required is dependent on the amount of data stored on the C:/
drive. A new CD holds approximately 650 MB.
5. Select Execute.
NOTE
The system names the backup automatically, which consists of a yyyymmdd format.
Accept the destination folder default for the DVD disk drive letter (usually D:) or select
Browse to choose a different destination. If a recordable disk is not available, then the
backup can be stored on the partitioned storage drive (E:).
7. When the back up completes, at the Backup complete window, select OK.
GPE01006.004 60
As required maintenance
2. At the ADVIA Backup window, select Restore a Backup Copy, then browse to the
source folder that contains the backup files to restore and select Execute.
4. If the CD contains all the backed up files to restore, then at the Restore window, select
Exit Restore.
5. If the backed up files are on more than one CD, select Continue.
GPE01006.004 61
Maintenance: System Maintenance Monitor window
Use this window to enter schedules for maintenance tasks and to monitor the maintenance
status of the system.
At the Menu Panel, select Maint., then select System Maintenance Monitor.
Entering maintenance schedule information
Located at the right of the System Maintenance Monitor window, the monitor
displays colored bars representing the amount of time (as a percentage) that
has elapsed between the last maintenance and the next scheduled
maintenance (for each task).
The monitor only displays bars for tasks not marked No setting.
Color Meaning
Blue Less than 50% of the time between maintenance dates has elapsed.
Green 50%-75% of the time between maintenance dates has elapsed.
Yellow 75%-100% of the time between maintenance dates has elapsed.
Red The next scheduled maintenance date has passed.
GPE01006.004 62
ISE maintenance – as required
Condition the electrodes the day before you need to use it. A new electrode may take a long
time to stabilize.
WARNING
To prevent infection from contacting serum directly, wear
suitable protective gloves when you remove the electrode
from the solution.
7. Replace the electrodes on the instrument with the newly conditioned ones.
GPE01006.004 63
ISE Maintenance - As required
Replacing electrodes
The acceptable ISE slope is between 45.0 and Mark ISE Slope Range
63.0. Slopes beyond this range are flagged as
shown in the following table. A flagged slope fails H > 65.0
the calibration. The slope limits are defined at the h 63.1 to 65.0
ISE Parameter Settings window.
l 38.0 - 44.9
L < 38.0
GPE01006.004 64
ISE Maintenance - As required
1. Assemble the new electrodes in the correct order (marked on the baseplate on the
analyzer ISE module)
2. Set the electrodes in place, paying careful attention not to leave a space between them.
Make sure that there is an 0-ring between each electrode and make sure that the ridges
on the side of each electrode fits into the depressions on the side of the electrode next to it.
3. Fasten the thumbscrew (1) while holding down each electrode with the retaining plate.
CAUTION
If a space is between the electrode
connections, the plate retaining the electrodes cannot close. If you cannot close it, move each
electrode left and right little by little. Do not force the electrode. Fasten the thumbscrew tightly. If the
retaining plate loosens during measurement, liquid could leak, causing a problem with the
instrument.
5. At the ISE Operation window, select Execute to the right of the word Initialize.
8. Select Execute, then select Yes when prompted to execute buffer prime.
GPE01006.004 65
ISE Maintenance - As required
Installing electrodes
9. Verify that the liquid is discharged smoothly from the dilution bowl during priming.
If the liquid is increasing without being discharged, a leak exists, an electrode is incorrectly
positioned, or a clog is in the drain system. If the liquid increases, immediately stop the
instrument.
IMPORTANT
If clogging occurs, the most probable cause is that the flow path is clogged inside the
electrode. Remove the Na and K electrodes, and check them by transmitted light to see
whether the flow path is clogged or not. You cannot do this for the Cl electrode because of
its construction. When in doubt, even if you cannot find a problem, try replacing the
electrode.
10. Mount the stainless steel cover of the top of the ISE unit by sliding it inside and fasten the
screw retaining the cover.
NOTE
When sliding it, be careful not to scratch the tubes and dilution bowl. When fastening the screw,
verify that the cover is not caught in the groove and is not loose.
12. At the ISE Operation window, select Execute to the right of the word Initialize, then select
Yes.
13. At the Menu Panel, select Maint., then select ISE Operation.
14. At the ISE Operation window, select Execute to the right of the word Calibration.
16. If the calibration fails, repeat calibration again and if data continues to be unstable, perform
an electrode wash.
NOTE
The electrodes may have to stabilize on the system before a successful calibration is
achieved.
17. At the ISE Operation window, select the Electrode Info button and enter the new
electrode information.
GPE01006.004 66
ISE Maintenance - As required
2. Unscrew the acrylic cover at the position where the electrolyte sample is dispensed at
the left front of the analyzer.
3. Loosen the screw retaining the stainless steel cover at the top of the ISE unit, and
remove that cover by sliding it toward you.
4. At the ISE Operation window, next to Final operation, type 16 in the field next to Pure
water position.
6. Fill a 10-mL tube with deionized water and place it on the CTT tray in position 16.
9. To dissolve the crystals attached to the liquid-supply nozzle, let it stand for about five
minutes.
10. At the ISE Operation window, next to Dil Bowl drain, select Execute.
The water in the dilution bowl drains.
11. Wipe the water or dirty parts around the liquid-supply nozzle (1) using a wet
cotton stick or a similar material.
12. At the ISE Operation window, enter 5 in the Bufferprime Times box, then select Execute.
GPE01006.004 67
ISE Maintenance - As required
CAUTION
Be careful not to scratch the nozzle. Damaging the nozzle may cause faulty results.
1. Using a pointed toothpick, carefully scrape the crystals that are attached to the waste-
drain nozzle (2).
2. At the ISE Operation window, enter 4 or 5 in the Bufferprime Times box, then select
Execute.
IMPORTANT
Verify that no buffer collects in the wash block. Buffer remains in the wash block may clog
the drain.
Maintaining the ISE Unit after the dilution bowl and waste-drain nozzle are clean
1. Replace the stainless steel cover of the ISE unit by sliding it into place, then secure the
retaining screw.
CAUTION
When sliding the cover, be careful not to scratch the tubes and dilution bowl. Also, when fastening
the screw, verify that the cover is not caught in the groove and is not loose.
3. At the ISE Operation window, select Execute to the right of the word Initialize.
GPE01006.004 68
ISE Maintenance - As required
This procedure would be run when ISE line contamination is suspected. Please refer to the
On-line Operator’s Guide for step-by-step instructions for performing this procedure.
Similar to the Washing Electrode Lines procedure, you will remove the electrodes from the ISE
module, and place a dummy electrode on the system.
Additionally, you will be replacing the ISE buffer bottle with a bottle filled with water, then a
bottle filled with 5% Probe Wash 3.
¾ reinstall the buffer bottle and the electrodes, and perform a buffer prime.
GPE01006.004 69
Table of Contents
GPE01006.004 1
GPE01006.004 2
1650 SPECIFIC Maintenance
Monthly Maintenance
Occasionally, cuvette waste water collects in the vacuum tank. Over time the vacuum tank can
accumulate an unacceptable level of waste water, preventing the vacuum pump from completely
removing waste water from the cuvettes. Monthly draining of the vacuum tank should preclude
this problem.
Note
The vacuum tank is located within the analyzer. Access to the tank is not required to drain the
tank.
GPE01006.004 3
Monthly Maintenance
Drain the Vacuum tank liquid (continued)
1. Open the center door on the analyzer to gain access to the vacuum tank drain
hose on the lower right shelf of the cabinet (1).
2. Squeezing the cap, remove the cap from the waste water discharge tube and
place the tube in a container.
GPE01006.004 4
4 Month Maintenance
Cleaning the On-board Pure-Water bottle filters
Materials required
ten (10) 10R filters, PN 073-0033-01
one (1) 18R filter, PN 073-0034-01
hex wrench
Time: 15 minutes
Analyzer mode: Ready
Clogged filters create an insufficient flow rate and produce air bubbles.
Note: A set of filters is included in the supplies kit. To avoid system down-time, you
may want to replace the filters with the ones in the kit, resume operation, then
clean and store the removed filters for the next scheduled maintenance.
GPE01006.004 5
4 Month Maintenance
Cleaning the On-board Pure-Water bottle filters (cont’d)
1. Using a hex wrench if necessary, unfasten the filter holder from the end of the
return hose and remove the filter.
Note: If any of the filters are ripped or damaged, replace it with a new filter.
GPE01006.004 6
Cleaning the On-board Pure-Water bottle filters (cont’d)
4. To clean the 18R filter, place in a beaker filled with a freshly made 10% solution of
water household bleach. After 30 minutes, remove the filter, rinse it in deionized water, and
replace into its holder.
5. Remove the ten small 10R Teflon filter hoses from the top front of the water bottle.
6. First unfasten the filter holder from the end of each tube, and then remove the
filter.
GPE01006.004 7
Cleaning the On-board Pure-Water bottle filters (cont’d)
1. To clean the 10R filters, place them in a beaker filled with a freshly made
10% solution of water and household bleach. (can use 5% Probe Wash 3)
After 30 minutes, remove the filters, rinse them in deionized water, and replace
onto their holders.
CAUTION: To avoid having the filter shift, ensure that it is properly
positioned within the filter holder.
2. Using a gauze pad soaked in alcohol, clean the outside surfaces of the filter
holders and hoses.
9. Insert the ten Teflon filter hoses in the tank, 3 or 4 at a time.
10. Insert the silicon filter hose in the water bottle.
11. Prime the lines.
a. On the Operation panel, click the PRIME button.
b. In the PRIME Set dialog box, select PRIME 2 and enter 10 or more for the
number of times in all fields.
c. Click Execute.
GPE01006.004 8
1650 Operator’s Guide to
ISE (Ion Selective Electrodes) maintenance
NOTE: This Maintenance Document was referenced from Operator’s Guide v.8.00.00.
In the event a new Operator’s Guide is released, please disregard this document and
refer to the new version. Also Referenced: Customer Bulletins and Best Practice
Observations.
After you have finished the ISE maintenance tasks, do the following:
On the ISE Operation window, click Execute to the right of the word Initialize and click Yes.
GPE01006.004 9
GPE01006.004 10
ISE Operation Window
GPE01006.004 11
Ise Maintenance
Washing the electrodes is normally done during the daily Shutdown wash (WASH2)
If required, manually request through ISE Operation Window as follows:
Materials Required:
ISE Detergent Solution: B01-4174-01
Time: 5 minutes Wear personal protective equipment.
Analyzer mode: Manual operation Use universal precautions.
To wash the electrodes
1. On the ADVIA 1650 Menu panel, click Maint.,
then click ISE Operation.
2. Enter the position number of the ISE detergent container.
3. In the Container field, select the type of container for the wash
solution.
4. Put the ISE detergent solution in the container and place it on the
CTT position you entered. Recommended container is a J-cup (to
ensure fresh detergent is loaded)
5. Click the Execute button to the right of the word Wash Electrode.
6. Click Exit.
7. Click Yes when prompted to execute electrode wash.
GPE01006.004 12
ISE Maintenance –Weekly
1. Dilution Bowl
2. Bowl Base Cap
3. Dilution Bowl Base
GPE01006.004 13
To clean the dilution bowl
5. Place the dilution bowl into a 10% solution of household bleach and water,
and allow it to soak for 30 minutes. Rinse the dilution bowl thoroughly in
running water, then in deionized water.
6. If there is any dirt left on the dilution bowl, use a piece of lint-free gauze and
wipe it clean. Rinse the dilution bowl thoroughly in deionized water.
7. Reinstall the o-ring and spacer, then reinstall dilution bowl and the bowl base
cap.
8. Verify that the bowl is pushed all the way down into the packing gasket.
Important: You should feel some resistance when your are pushing the
bowl into the base. If this is not the case, replace the packing gasket
(PN 073-0076-01).
GPE01006.004 14
To clean the dilution bowl (cont’d)
GPE01006.004 15
ISE Maintenance –Monthly
Materials Required
Lint-free gauze
1. On the ISE Operation window, click Execute to the right of the word
Initialize.
2. Click Yes when prompted to execute arrangement.
3. On the ISE Operation window, click Exit.
4. Click Yes when prompted whether ISE is over.
GPE01006.004 16
ISE Maintenance –Monthly
1. Peristaltic pump
GPE01006.004 17
To replace the peristaltic-pump tube
GPE01006.004 18
6. Push up the pump platen (4) until it clicks into place, then verify that it is
securely in place by shaking it lightly.
7. On the ISE Operation window, double-click 5.PP to test pump operation.
NOTE
When the pump rotates for the first time after a new tube is in place, you
may hear a scratching sound. This is normal and after the first few turns
the sound should go away.
8. Double-click 5.PP again to turn off the testing operation.
1. On the ISE Operation window, click Execute to the right of the word
Initialize.
2. Click Yes when prompted to execute arrangement.
3. On the ISE Operation window, click Exit.
4. Click Yes when prompted whether ISE is over
GPE01006.004 19
ISE Maintenance –Monthly
1. BUFFER LINE
2. PUMP
3. PLASTIC COVER
GPE01006.004 20
REPLACING THE PUMP SEAL AND PACKING (CONT’D)
GPE01006.004 21
REPLACING THE PUMP SEAL AND PACKING (CONT’D)
13. On the ISE Operation window, click Execute to the right of the word
Initialize.
14. Click Yes, when prompted to execute arrangement.
15. On the ISE Operation window, in the Bufferprime area, enter 20 into the
Times field. Click Execute. Click Yes, when prompted to execute buffer
prime.
16. Verify that the syringe fills with the buffer solution and does not leak.
GPE01006.004 22
ISE Maintenance –Monthly
BIOHAZARD
All products or objects that come in contact with human or animal body fluids should be handled,
before and after cleaning, as if capable of transmitting infectious diseases. Wear facial
protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of infectious agents
in health-care settings as recommended for potentially infectious specimens in Protection of
Laboratory Workers from Infectious Disease Transmitted by Blood, Body Fluids, and Tissue, 2d
edition; Approved Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user protection and it can
be used as reference material for instructions on laboratory safety.
GPE01006.004 23
ISE Maintenance – Once a Month
Clean the ISE Waste Block (cont’d)
NOTES:
• Use caution when removing the back screw of the waste block. Do not
damage the Teflon line next to the block.
• The thumbscrews from an old halogen lamp may be substituted for the
existing screws to facilitate this procedure.
5. Slightly tilt the wash block towards you to view the drainpipe.
6. Use a cotton tipped applicator and deionized water to clean the salt off the stainless
steel drainpipe.
GPE01006.004 24
ISE Maintenance – Once every Three Months*
*NOTE: This procedure should be done more frequently if your lab meets the following criteria:
Time: 10 minutes
Analyzer Mode: Manual Operation
GPE01006.004 25
WASHING THE ELECTRODE LINES (CONT’D)
GPE01006.004 26
WASHING THE ELECTRODE LINES (CONT’D)
GPE01006.004 27
ISE Maintenance – as required (if contamination is suspected)
WASHING ALL THE LINES
Materials Required
• Deionized water
• Dummy electrode, PN 073-0342-01
• ISE detergent solution, B01-4174-01
• Reagent Probe Wash 3, B01-4183-01
Time: 15 minutes
GPE01006.004 28
To wash all lines
1. Open the lower left analyzer door and replace the buffer-solution bottle with a
bottle of deionized water.
2. Pull out the ISE drawer and replace the electrodes with a dummy electrode.
Suggestion: Immerse the electrodes in ISE
buffer to keep the electrodes moist while you perform maintenance.
3. Close the drawer and on the ISE Operation window, click Execute to the
right of the word Initialize. Click Yes when prompted to execute
arrangement.
4. Replace the deionized-water bottle with a bottle of 5% solution of Reagent
Probe Wash 3.
5. On the ISE Operation window, in the Bufferprime area, enter 50 into the
Times field. Click Execute. Click Yes when prompted to execute buffer
prime.
6. Replace the Reagent Probe Wash 3 bottle with a bottle of deionized water.
7. Enter 50 again in the Times field of the Bufferprime area, then click
Execute.
8. Reinstall the electrodes and click Execute to the right of the word Initialize.
9. Reinstall the buffer-solution bottle.
10. On the ISE Operation window, in the Bufferprime area, enter 10 into the
Times field. Click Execute, then click Yes to prime the line with buffer.
11. When the priming is finished, pull out the ISE drawer and verify that the
electrodes are not leaking.
12. Close the drawer. On the ISE Operation window, click Exit, then click Yes
when prompted whether ISE is over.
13. Run ten pooled serum samples, or do an ISE CV check, to verify ISE
performance and data stability.
GPE01006.004 29
ISE Maintenance – as required
CONDITIONING THE Na and K ELECTRODES
Materials Required
• 10 mL of serum pool
• 30 mL of ISE Buffer, B01-4171-51
• 2 mL or 3 mL plastic, disposable pipette
A new electrode may take a long time to stabilize, so carry out aging of the electrode on
the day prior to start of use.
GPE01006.004 30
CONDITIONING THE Na AND K ELECTRODES (CONT’D)
warning
To prevent infection from contacting serum directly, wear
suitable protective gloves when you remove the electrode
from the solution.
NOTE
High-concentrated salt water is used as a preservation solution to maintain electrode
performance. When the electrode package is opened, wash the electrode with sufficient water
and wipe well before use. Small amounts of salt on the electrode may cause rust on the electrode
connector.
7. Replace the electrodes on the instrument with the newly conditioned ones.
8. Calibration is performed as part of the electrode replacement. If the calibration
fails, repeat calibration again. If data continues to be unstable after electrode
conditioning, perform an electrode wash.
GPE01006.004 31
ISE Maintenance – as required
REPLACING THE ELECTRODES
Replace electrodes if slope or selectivity is incorrect. The acceptable ISE slope must be between
45.0 and 63.0. Slopes beyond this range are flagged as shown in the following table. A flagged
slope fails the calibration. The slope limits are defined in the ISE Parameter Of Setting window.
Materials required:
• Electrodes
• Cl, PN 073-0049-01
• K, PN 073-0050-01
• Na, PN 073-0051-01
• O-rings, 3, PN 073-0071-01
Time: 5 minutes
Analyzer mode: Manual operation
GPE01006.004 32
4. REPLACING THE ELECTRODES (CONT’D)
1 ELECTRODE CONNECTORS
2 BOWL BASE
3 THUMBSCREW
4 ELECTRODES
GPE01006.004 33
REPLACING THE ELECTRODES (CONT’D)
11. On the ISE Operation window, click Execute to the right of the word
Initialize area.
Click Yes when prompted to execute arrangement.
12. In the Bufferprime area, enter 15 into the Times field. Click Execute, then
click Yes when prompted to execute buffer prime.
13. Make sure that the electrodes are not leaking.
14. Reinstall the electrode cover.
15. Close the ISE drawer. On the ISE Operation window, click Execute to the
right of the word Initialize and click Yes. After initialization is complete, click
Exit and select Yes.
16. On the ADVIA 1650 Operation panel, click Initialize.
17. On the ADVIA 1650 Menu panel, click Maint., then click ISE Operation.
18. On the ISE Operation window, click Execute to the right of the word Calibration, then click
Yes when prompted to execute calibration.
NOTE
The electrodes may have to stabilize on the system before a successful calibration is achieved.
GPE01006.004 34
ISE Maintenance – as required
Materials required
• 10 mL of serum pool
• 30 mL of ISE Buffer, B01-4171-51
• 2 mL or 3 mL plastic, disposable pipette
biohazard warning
Time: 45 minutes Wear personal protective equipment.
Analyzer mode: Manual operation Use universal precautions.
GPE01006.004 35
To condition the electrodes on the system
GPE01006.004 36
ISE Maintenance – as required
GPE01006.004 37
GPE01006.004 38
Best Practices &
Common Alarm Messages
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info.
What is the difference between a Best Practice and
a Recommendation?
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
2
Best Practices & Recommendations
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
3
Daily
NOTE: remember to put 10% cuvette wash back into these positions
during normal operation for assay contamination sets
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
4
Saline
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
5
Water
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
6
Monthly
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
7
Replace RRV Segments every 4 months
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
8
Troubleshooting Error Codes
If multiple error codes occur, the earliest error code is the most
significant
Errors may cascade
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
9
Common Reagent Installation Error Codes
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
10
Common Reagent Installation Error Codes
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
12
4097 ISE bowl high
Hint: was any ISE maintenance just performed? - go back and recheck it
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
13
4097 ISE bowl high
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
14
6178 System Error
Liquid level status was abnormal when checking the liquid-level sensor
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
15
6202 RPP1(2) Full Stroke Error
6139 = DPP Full Stroke Error (1650,1800, 2400)
6139 = SPP Full Stroke Error (1200)
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
16
6207 DWUD Nozzle Sensor Error
DWUD nozzle crashed into DTT (not applicable to 1200)
Initialize system
Check the dilution tray for
obstructions
Check that the DTT cuvettes are
installed properly
Observe DWUD that the green LED
light is on
Call TS, phone support
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
17
6208 WUD Nozzle Sensor Error
WUD nozzle crashed into DTT
Initialize system
Check the RRV area for
obstructions
Verify that the RRV wedges have
been installed properly
Observe WUD that the green
LED light is on.
Call TS, phone support
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
18
6222 Probe Crash Detected DPP (SPP on 1200)
This picture displays the DPP for 1650, 1800 & 2400
The 1200 SPP crash detection differs slightly, but the principles above
& steps below apply
Initialize System
Visually assess probe for any damage and for proper alignment
Remove Obstruction
Change probe if necessary
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
19
Probe Liquid Level Sensing Errors
General suggestions for troubleshooting: applies to “Liquid Sensing” or
“surface verification error”)
Verify that
the probe is clean
Confirm screws Confirm LLS Amp
LLS wire connections on probe are
are tight is firmly mounted
tight
Be certain the DPP probe is below
the Crash Detect clamps
SPP on 1200
the tubing connection at the back of
the probe is tight and not leaking
Wipe condensation from the
CTT/RTT trays 1650, 1800, 2400 PROBE
Verify that water pressure gauge is
reading 76 kPa during operation
Replace probe
If problem continues, LLS sensor
may need to be replaced.
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
20
Clot Alarms
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
21
8722 Lamp Energy Low
D flags with Cell Blank Check
Lamp output (voltage) was less than the lower limit of the acceptable
A/D range
If all cuvettes are affected
Verify that cuvette conditioner is being dispensed properly from the WUD
station 4
Replace the lamp
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
22
9004 LWP pressure low
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
23
9005 Vacuum Pressure Alarm
9039 Standby Check
Liquid in the vacuum overflow tank
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
24
Wash Station Overflow
Overflow sensors are wet
Possible causes:
Filter needs to be replaced
Pump failure
Call TS to report the alarm
TS may guide you through
adjusting flow rate
Slight turns of the red valve
Use System Monitor to determine
flow rate adjustment
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
26
Leak Sensors differ & not all are accessible to the
customer. Call Tech Support.
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
27
Leak Sensor Errors on 1650, 1800 & 2400
9030 Leak: DTTÎ Leak was detected inside or under the DTT.
9031 Leak: J4 manifold or water bottle Î Leak was detected under the J4 manifold located at the right
side of the analyzer or under the water bottle located in the lower center compartment at the front of the
analyzer.
9032 Leak: solenoid valve panel Î Leak was detected under the solenoid valve panel in the upper
compartment at the rear of the analyzer.
9033 Leak: ISE compartment Î Leak was detected under the ISE compartment at the left side of the
analyzer.
9064 Standby Check (DTT Leak) Î Leak was detected inside or under the DTT.
9065 Standby Check (Leak under J4 manifold or water bottle) Î Leak was detected under the J4
manifold located at the right side of the analyzer or under the water bottle located in the lower center
compartment at the front of the analyzer.
9066 Standby Check (solenoid valve panel leak) Î Leak was detected under the solenoid valve panel in
the upper compartment at the rear of the analyzer.
9067 Standby Check (leak under ISE compartment) Î Leak was detected under the ISE compartment
at the left side of the analyzer
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
28
Leak Sensor Errors on 1200
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
29
Leak Sensor Errors on 1800
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
30
Leak Sensor Errors cont’d
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
31
Leak Sensor errors on 1800
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
32
Turning off the ISE module
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
33
9152 ISE position alarm (1650 only)
Do not lean on the system top cover. This may cause the ISE pull out
drawer to twist and set off the ISE position alarm.
SUPPLEMENTAL INFO FOR TRAINING PURPOSES ONLY Refer to Product Labeling for updated info
GPE01006.004
34