Instruction Manual
2
1 Contents
1 Contents
1 Contents ........................................................................................ 3
2 Safety and the environment ........................................................ 5
2.1. About this document ........................................................................5
2.2. Safety ..............................................................................................5
2.3. Protecting the environment ..............................................................6
3 Specifications ............................................................................... 6
3.1. Use ..................................................................................................6
3.2. Technical data .................................................................................6
4 Product description.................................................................... 11
5 Using the product ....................................................................... 15
5.1. Commissioning ..............................................................................15
5.2. Switching on / off ...........................................................................15
5.3. Setting the date / time....................................................................16
5.4. Showing / hiding the date / time.....................................................16
5.5. Switching the display illumination on / off ......................................16
5.6. Changing the frequency weighting ................................................16
5.7. Changing the time weighting .........................................................17
5.8. Measuring ......................................................................................17
5.9. Using the min / max hold function..................................................18
5.10. Using individual value storage .......................................................19
5.11. Using measurement series storage ...............................................19
5.12. Using AC / DC signal output ..........................................................20
5.13. Using PC interface.........................................................................21
6 Maintaining the product ............................................................. 21
6.1. Cleaning the instrument.................................................................21
6.2. Changing batteries ........................................................................21
6.3. Calibrating / adjusting the instrument ............................................21
3
1 Contents
4
2 Safety and the environment
Warnings
Always pay attention to any information denoted by the following
warnings, indicated with warning pictograms. Implement the
precautionary measures specified!
Display Explanation
2.2. Safety
> Only use the product properly, for the purpose it is intended for
and within the parameters specified in the technical data. Do
not apply any force.
> Do not store the product together with solvents. Do not use any
desiccants.
> Only perform maintenance and repair work on this instrument
that is described in the documentation. Follow the prescribed
steps exactly. Use only original spare parts from Testo.
> Protect the instrument from rain and humidity. Make sure that
no fluids get into the microphone.
5
3 Specifications
3 Specifications
3.1. Use
The testo 816-1 is a sound level meter with a measuring range of
30 to 130 dBA and 35 to 130 dBC, two time weightings, two
frequency weightings, minimum / maximum value function,
individual value storage and measurement series storage.
Measurement configurations and results are shown on the LCD
display. Measurement data can be stored in the instrument or
transferred to a Windows® PC via a PC interface.
Using the calibrator (accessory item), the meter can be adjusted
with the enclosed adjustment screwdriver.
The instrument fulfils the requirements of IEC 61672-1 Class 2.
6
3 Specifications
Feature Values
DC output 10 mV / dB
Voltage supply 4 x IEC LR6P (AA) batteries
Battery life approx. 30 h (alkaline batteries)
Power consumption approx. 0.3 W
Mains connection 9 V DC (8-10 V DC max)
Operating 0 to 40 °C
temperature
Operating humidity 10 to 90 % RH
Operating / storage max. 2000m above sea level
height
Storage temperature -10 to 60 °C
Storage humidity 10 to 75 % RH
Dimensions 272 × 83 × 42 mm
(L x W x H)
Weight 390 g (including batteries)
Laws, guidelines, IEC 61672-1 Class 2, ANSI S 1.4 Type 2
standards ISO 9001:2008
7
3 Specifications
8
3 Specifications
Frequency weighting
Frequency [Hz] dBA(dB) dBC(dB) Error limits (dB)
20 -50.5 -6.2 ±3.5
31.5 -39.4 -3.0 ±3.5
63 -26.2 -0.8 ±2,5
125 -16.1 -0.2 ±2,0
250 -8.6 0.0 ±1,9
500 -3.2 0.0 ±1,9
1000 0.0 0.0 ±1,4
2000 1.2 -0.2 ±2,6
4000 1.0 -0.8 ±3,6
8000 -1.1 -3.0 ±5,6
9
3 Specifications
Temperature dependency
Values relating to:
• Ambient humidity: 65 %RH
• Sound pressure reference value: 124 dB
• Temperature range with discrepancy < 0.5 dB: 10 to 40 °C
Temperature [°C] Offset value [dB]
-10 -0.7
0 -0.7
5 -0.6
50 +1
10
4 Product description
4 Product description
11
4 Product description
12
4 Product description
Display
13
4 Product description
14
5 Using the product
Switching off
> Hold the key down until P-OFF is displayed.
15
5 Using the product
Automatic switch-off
The display illumination is switched off automatically after 30 s.
16
5 Using the product
5.8. Measuring
Information and recommendations
• Sound waves can be reflected off walls, ceilings and other
objects. If not handled correctly, the measuring instrument
housing and the person conducting the measurement are also
disruptive factors within the sound field and can result in
incorrect measurement results.
• Not only can the instrument housing and the person operating
the instrument impede the sound which comes from a particular
direction, they may even cause reflections, resulting in
significant measurement errors. Experiments have shown, for
example, that at frequencies around 400 Hz the body can
cause errors of up to 6 dB if a measurement is carried out less
than one meter away from the person. This error is smaller at
other frequencies, but a minimum distance should still be
observed. Generally, it is recommended that the measuring
instrument is held at least 30 cm - even better 50 cm - away
from the body.
• We recommend fitting the instrument to a tripod for exact
measurements
• Absolute pressure dependency: The measuring instrument is
calibrated at a height of 0 m above sea level in the factory.
Measurements at other heights give rise to measurement errors
which can be corrected using a table (see technical data).
Please subtract the corresponding offset value from the
measured value (e.g. - 0.1 dB for measurements at a height of
500 m above sea level). Prevent this measuring error by
adjusting the measuring instrument at the corresponding height
before (and after) each measurement. Follow the instruction
manual for the calibrator.
17
5 Using the product
18
5 Using the product
19
5 Using the product
DC Signal
AC Signal Ground
20
6 Maintaining the product
21
7 testo 816-1 software
22
7 testo 816-1 software
Real Time
Function Description
Run Start real-time measurement.
Stop End real-time measurement.
Data Logger
Function Description
Load Data Load measurement series from the
instrument memory into the software.
23
7 testo 816-1 software
Function Description
Erase Memory Delete measurement series stored in the
instrument memory.
RECALLl
Function Description
RECALL Display individual readings from the
instrument memory.
View
Function Description
Control Panel Display the meter control window.
Real-Time Display window for real-time display of the
Graph current readings.
Window
Function Description
Tile Arrange windows one above another.
Cascade Freely arrange windows.
Help
Function Description
Contents Open help file.
Info Display program information.
24
7 testo 816-1 software
Making settings
Function Description
Real-Time data Input field: enter the number of readings that
no. to be are to be recorded.
recorded
Real-Time Input field: enter the recording rate.
sampling rate
Recording Information field: calculated recording period
Period is displayed.
Start Start real-time display
Real-Time Graph
25
7 testo 816-1 software
Function bar
Function Description
Show / hide bar for displaying the recording
information.
Show / hide bar for displaying the evaluation
information.
Standard mouse cursor.
Zoom
1. Activate standard mouse cursor.
2. Click in the graph and draw a rectangle by holding down and
dragging your mouse.
- The selected graph area is displayed.
3. To return to the standard view, press the button.
26
7 testo 816-1 software
27
8 Tips and assistance
28
8 Tips and assistance
For a complete list of all accessories and spare parts, please refer
to the product catalogues and brochures or visit our website
www.testo.com
29
Testo SE & Co. KGaA
Celsiusstr. 2
79822 Titisee-Neustadt
Germany
Tel.: +49 7653 681-0
E-Mail: info@testo.de
www.testo.com
0970 8161 en 04