Professional Documents
Culture Documents
Ivy 101 Monitor - Service Manual
Ivy 101 Monitor - Service Manual
Patient Monitor
R-wave Trigger
2409-00-20
Service Manual
Copyright © 1999 by Ivy Biomedical Systems, Inc., Branford, Connecticut. All rights reserved. No
part of this manual may be reproduced without the permission of Ivy Biomedical Systems, Inc.
101R/NR0999
2409-00-20 REV 00
Table of Contents
Table of Contents
WARRANTY .......................................................................................................................................................iv
INTRODUCTION ................................................................................................................................................1
SAFETY ................................................................................................................................................................2
Electrical................................................................................................................................................................2
Explosion ...............................................................................................................................................................2
Patient Connections..............................................................................................................................................2
Pacemakers ...........................................................................................................................................................2
Electrosurgery.......................................................................................................................................................3
Defibrillation Protection ......................................................................................................................................3
EMC.......................................................................................................................................................................3
Description of Labels............................................................................................................................................3
TRIGGER OUTPUT..........................................................................................................................................13
The Trigger Pulse ...............................................................................................................................................13
Trigger-Spot Display ..........................................................................................................................................13
ECG MONITORING.........................................................................................................................................14
Safety Considerations.........................................................................................................................................14
Patient Connections............................................................................................................................................15
ECG Waveform Amplitude (Size).....................................................................................................................15
Lead Selection .....................................................................................................................................................16
ECG Filter...........................................................................................................................................................16
Alarm Limits.......................................................................................................................................................17
CLEANING ........................................................................................................................................................24
Monitor................................................................................................................................................................24
Patient Cables .....................................................................................................................................................24
ACCESSORIES..................................................................................................................................................25
Disposal................................................................................................................................................................25
PURCHASING INFORMATION.....................................................................................................................49
Part List...............................................................................................................................................................49
Terms and Conditions........................................................................................................................................49
WARRANTY
All products warranted to be free from defects in material and workmanship and to operate within published
specifications, under normal use, for a period of two years from date of original shipment.
If an examination discloses such products or component parts to have been defective, then our obligation is limited
to repair or replacement (at our option). Fuses and batteries are not covered under this warranty.
During the first year of service, there is no charge for parts or labor. In the second year of service, the customer is
charged for labor only.
INTRODUCTION
This manual is to provide information on the correct use of the Model 101R/NR monitor. It is up to the user to
ensure that any applicable regulations respecting the installation and operation of the monitor are observed.
Manufacturer's Responsibility
The manufacturer of this equipment is responsible for the effects on safety, reliability, and performance of the
equipment only if:
• assembly operations, extensions, re-adjustments, or repairs are carried out by persons authorized by the
manufacturer
Incorrect operation or failure of the user to maintain the monitor in accordance with proper maintenance procedures
relieves the manufacturer or his agent from all responsibility for consequent non-compliance, damage, or injury.
For technical and service information, please refer to the Model 101R/NR Service Manual.
READ THE ENTIRE SAFETY INFORMATION SECTION BEFORE YOU OPERATE THE MONITOR.
SAFETY
Electrical
This product is intended to be operated from a mains power source of nominally 100, 120, 220 and 230 V ~. The
correct mains operating voltage is specified on the monitor's rear panel.
To prevent electrical hazards to all personnel, this monitor must be properly grounded. The power cable supplied
with this equipment provides for this protection. Do not attempt to defeat this protection by modifying the cable or
by using ungrounded adapters.
WARNING: Electric shock hazard! Do not remove covers or panels. Refer service to qualified service personnel.
WARNING: To avoid electrical shock, disconnect the monitor from its power source before changing fuses.
Replace only with same type and rating of fuse.
WARNING: This unit uses a common isolation path for ECG leads. Do not connect any non-isolated accessories to
the ECG input connected to a patient, as this may compromise the safety of the unit. When attached to other
devices, insure that the total chassis leakage currents of all units do not exceed 300µA.
Explosion
DANGER: Explosion hazard! Do not use this equipment in the presence of flammable anesthetics.
Patient Connections
Patient connections are electrically isolated Type CF ⎯⏐ ♥⏐⎯. For all connections use insulated probes. Don't let
patient connections contact other conductive parts, including earth. See instructions for patient connections in this
manual.
Carefully route patient cables to reduce the possibility of patient entanglement or strangulation.
Leakage current is limited internally by this monitor to less than 10 µA. However, always consider additional
leakage current that can be caused by other equipment used on the patient at the same time as this monitor.
To ensure that the leakage current protection remains within the specifications, use only the patient cables specified
in the patient connections instructions in this manual.
Pacemakers: Rate meters might continue to count the pacemaker rate during occurrences of cardiac arrest or some
arrhythmias. Do not rely on rate meter alarms. Keep pacemaker patients under close surveillance.
Electrosurgery
Avoid electrosurgery burns at monitoring sites by ensuring proper connection of the electrosurgery return circuit. If
improperly connected, some electrosurgery units might allow energy to return through the electrodes.
Defibrillation Protection
This equipment is protected against 360 J discharge and electrosurgery potentials. The monitor is internally
protected to limit current through the electrodes to prevent injury to the patient and damage to the equipment as long
as the defibrillator is used in conformance with the manufacturer's instructions.
EMC
This equipment has been certified to be protected to emissions and immunity according to IEC-601-1-2.
{ ⏐ { = OFF, ⏐= ON.
Output signal.
DANGER
HIGH
VOLTAGE
MONITOR DESCRIPTION
The Model 101R/NR Portable Patient Monitor is a dual trace monitor for ECG and heart rate, which produces an
output pulse corresponding in time to the R-wave. The monitor is intended primarily for use in timed imaging
studies, TMR/PMR and any application requiring precision R-wave synchronization.
Specifications
ECG
Isolation: Isolated from ground related circuits by >4 kV rms, 5.5 kV peak
Frequency Response
CRT: Filtered: 0.5 to 25 Hz
Unfiltered: 0.5 to 45 Hz
Frequency Response
X1000 output: Filtered: 0.5 to 25 Hz
Unfiltered: 0.2 to 100 Hz
Electrode Offset
Potential: ±0.5 V DC
Noise: <20 µV peak-to-peak, referred to the input with all leads connected
through 51 kΩ/47 nF to ground
Electrosurgical Interference
Protection: Standard
Cardiotach
Accuracy: ±1%
Resolution: 1 bpm
Alarms
Test Mode
Display
Trace: Dual, Fixed (moving bar), single trace normal operation. Dual trace in
“freeze” mode.
Mechanical
Trigger Output
Environmental
Operating Temperature
Range: 5°C to 45°C
Storage Temperature
Range: -25°C to 60°C
Power Requirements
Voltage Input: 100, 120, 220, 230, V ~, 50/60 Hz; +10% - 15%
(to change mains operating voltage see text on Monitor Setup).
Maximum ac Power
Consumption: 85 VA
Safety
Unit meets or exceeds the specifications for the AAMI Cardiac Monitor Standard. Units meet UL544,
CSA, and CE-MDD 93/42/ECC requirements. ISO-9001 Approval # 98-1044.
MAINS POWER
SWITCH: A switch to control ac power input to the monitor. The | position is on, O is off.
Puts monitor into setup mode and causes setup menu to be displayed on the right portion of the
screen. The monitor returns to normal mode after pressing MENU again or if no keys are pressed
for 20 seconds.
Disables the audible and visual alarms for a two-minute period to allow the operator perform
procedures that would otherwise set off the alarms. This avoids the problem of turning off the
alarms and forgetting to turn them back on. Press this key again to return the alarms to normal
before the two minutes have expired.
Generates a 1 mV pulse at 70 bpm that is displayed on the ECG trace and sent to the X1000
output on the rear panel while the switch is depressed. This is used to ensure that the monitor is
functioning correctly.
Turns the alarm audio on and off. When off, the ALARMS OFF message is displayed on the left
side of the screen.
Basic Keys
ECG Patient
Cable input
Power Switch
Programmable Keys
Programmable Keys
Each of the following keys, to the right of the CRT display, has a normal function, printed on the key, and two
menu functions, displayed on the screen in the setup modes. The additional functions are accessed by pressing the
MENU key once. Return to normal function by pressing MENU key again.
Selects the speed at which the ECG trace moves across the screen.
Stops the movement of the ECG waveform on the display for closer evaluation and displays the
waveform as a second trace below the actual trace. Press FREEZE again to resume movement.
Resets the audible and visual indicators for an alarm that has been activated.
Enables the Trigger-Spot feature, which highlights the point in the ECG waveform when the
trigger output pulse occurs.
MENU STRUCTURE
HIGH
LOW
TRIG. DELAY
(101R)
Display
HEART RATE: Displayed in beats per minute (bpm) on the upper left part of the screen.
SETUP: Selections made in the menu setup modes (alarm limits, lead selection, and filter on/off) are displayed in
small characters at the upper right corner.
ECG: Trace is displayed across the screen moving from right to left.
ALARMS: The following alarm indications are displayed in reverse video. Alarm indications appear on the upper
right portion of the screen and flash once per second except ALARMS OFF, which is displayed on the upper left
portion of the screen and is displayed as .
LEAD OFF: A lead has become disconnected. This alarm cannot be reset with the
ALERT RESET key.
HR HIGH: The high heart rate limit has been exceeded for four seconds.
HR LOW: The low heart rate limit has been exceeded for four seconds.
Rear Panel
When the monitor is connected to another piece of equipment through this rear panel connector, always make sure
that each piece of connected equipment has its own separate ground connection.
Do not attempt to connect cables to these connectors without contacting your Biomedical Engineering Department.
This is to ensure the connection complies with leakage current requirements of one of the following applicable
standards: UL 544, CSA 22.2 No. 125 or IEC 601-1. The maximum non-destructive voltage that may be applied to
these connectors is 5V.
TRIGGER OUTPUT: A BNC type connector for the output of the trigger pulse indicating the timing of the peak
of the R-wave. The output can be delayed from the peak of the R-wave through user controls.
QRS VOLUME CONTROL: A switch to control the QRS beep audio volume. Alarm audio is not affected.
PEQ GROUND: Potential Equalization - A ground connection that can be used to ensure that no potential
differences can develop between this equipment and other electrical equipment.
FUSE: Replace only with the same type and rating of fuse as indicated on the fuse rating label.
REFERS TO
ECG (¼" stereo jack)
OPERATOR'S MANUAL ECG Output Tip=ECG X1000
QRS
X 1000 VOLUME QRS Volume
MAX Control
OFF
PHYS. TRIG.
PEQ Ground
MIN
OUTPUT
TRIGGER
Trigger Output/BNC
POLARITY Fuse rating
Selects Output
NEG POS
Trigger Polarity 100V ⏐ 120V 220V ⏐ 230V
T 1.6 Amp T 1.0 Amp
120 Vac Voltage Selection 250V 250V
wheel 5mm x 20mm 5 mm x 20 mm
Fuse
(Beneath Panel)
Power Input
85VA 50/60Hz
MONITOR SETUP
Check the monitor's rear panel to verify the voltage range required by the monitor.
2. Press the POWER switch at the left side of the front panel to turn power on.
3. Connect the patient cable to the ECG connector on the front panel.
4. Connect the Trigger Output from the monitor rear panel to the device being triggered.
4. Locate and remove voltage wheel and set for appropriate voltage.
6. Replace cover.
TRIGGER OUTPUT
Trigger-Spot Display
Press the TRIG MARK key on the front panel to highlight the portion of the ECG waveform corresponding to the
timing of the trigger pulse. Press TRIG MARK again to disable Trigger-Spot.
ECG MONITORING
When ECG monitoring, the display shows the ECG waveform across the center; the heart rate, heart rate limits, and
lead selection in the upper left corner; and alarm indications. Also, a heart symbol flashes each time a heartbeat is
detected.
Safety Considerations
Disposable products are intended for single-use only. Do not attempt to re-use these products.
Patient connections are electrically isolated Type CF ⎯⏐ ♥ ⏐⎯. For all connections use insulated probes. Don't let
patient connections contact other conductive parts, including earth. See instructions for patient connections in this
manual.
Leakage current is limited internally by this monitor to less than 10 µA. However, always consider additional
leakage current that can be caused by other equipment used on the patient at the same time as this monitor.
Avoid electrosurgery burns at monitoring sites by ensuring proper connection of the electrosurgery return circuit. If
improperly connected, some electrosurgery units might allow energy to return through the electrodes.
Rate meters might continue to count the pacemaker rate during occurrences of cardiac arrest or some arrhythmias.
Do not rely on rate meter alarms. Keep pacemaker patients under close surveillance.
Patient Connections
To ensure compliance with both safety and performance specifications, use the patient cables supplied by Ivy
Biomedical Systems (see Accessories).
2. Connect the patient cable to the monitor's front panel ECG input.
4. Use the procedures described in the following sections for alarm limit settings, lead selection, amplitude
adjustment, and enabling or disabling the filter.
ECG
HR
LIMITS
TRIGGER
DELAY
RECORDER
2. Press the first programmable key (SPEED 25/50 mm/s) once to select ECG.
×
ECG
SIZE
MM/MV
Ø
LEAD
II
FILT
3. Use the top two programmable keys to the right of the display to adjust the ECG waveform amplitude.
Lead Selection
Select [LEAD] to change the lead selection. The current lead selection is shown below the alarm limits in the upper
right portion of the display.
ECG Filter
Select [FILTER] to turn the filter on or off. The filter on/off indication is shown in the upper right portion of the
display. The filter sets the frequency response of the displayed waveform as follows:
Filtered: 0.5 to 25 Hz
Unfiltered: 0.5 to 45 Hz
Alarm Limits
1. With the first menu (heart rate alarm limits) displayed, select [HR LIMITS] to change the Heart Rate Alarm
Limits.
HR
LIMITS
×
HIGH
150
Ø
×
LOW
30
Ø
2. Use the four programmable keys to the right of the display to set the high and low heart rate limits.
Each time you press a key, the corresponding limit changes by 5 bpm. The current limits are always shown in
the upper left portion of the display.
RECORDER OPERATION
101R only
Changing Paper
Replace the roll of thermal paper as follows. (Recorder paper is Ivy P/N: 590035)
1. Press the paper eject button to open the door at the front of the recorder.
Press Here
If the door does not open completely, pull it toward you until it is completely open.
2. Reach in and remove the spent paper core by pulling it gently toward you.
3. Place a new paper roll between the two round tabs of the paper holder.
4. Pull some paper from the roll. Make sure the sensitive (shiny) side of the paper faces the print head. The
shiny side of the paper normally faces inside the roll.
Pinch
Roller
6. Hold the paper against the pinch roller and close the door.
7. The paper should advance one inch. If the paper doesn’t move, open the door and repeat starting at step 5.
Recorder Menus
The recorder user interface is through the menu structure of the 101R/NR.
NOTE: At the start of each printout, the printer produces a header consisting of the current readings for all
parameters in use at the time of the printout, whether the parameter is displayed or not.
Recorder Modes
Use the following procedure to select the printing mode to be used. Selections are DIRECT, TIMED and DELAY.
The print mode is indicated in the lower left corner of the display.
ECG
HR
LIMITS
TRIGGER
DELAY
RECORDER
RECORDER
SET CLOCK
Direct To print in direct, press the PRINT key. Press PRINT again to stop printing.
The plot is preceded by a header which contains all parameter readings and the time/date.
The speed of the plot and vertical resolution are the same as the display. The plot is labeled with the speed
of the plot in mm/s, the recorder mode, and the parameters.
Timed TIMED mode starts by pressing PRINT and prints for 30 seconds.
Delay Delay mode plots the 19 seconds before and 12 seconds after the occurrence of an alarm condition or if
print button is pushed.
3. Press the programmable key (TRIG. MARK) once to select SET CLOCK.
N
11:20 AM E
25 SEPT 98 X
T
ENTER
4. Press the programmable key (SPEED 25/50mm/s) to place the flashing cursor under the hour, minutes,
date, month or year.
5. Press the programmable key (FREEZE) to increase × what the cursor is under or press the
programmable key (ALARM RESET) to lower Ø what the cursor is under.
6. After changing the settings press the programmable key (TRIG MARK) to ENTER the settings.
ALARM MESSAGES
LEAD OFF A lead has been disconnected or the electrode offset potential has exceeded ±0.5 V.
This alarm cannot be reset with the ALERT RESET key.
HR HIGH The high heart rate alarm limit has been exceeded for four seconds.
HR LOW The low heart rate alarm limit has been exceeded for four seconds.
MONITOR TESTING
Press the TEST key to test the internal functions of the monitor. You should do this each time you begin monitoring
a patient.
The TEST function generates a 1 mV pulse at 70 bpm, causing a waveform and a 70 bpm indication on the display
and a signal at the rear panel connector. If these indications are not present, contact qualified service personnel.
Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments
should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance
with local or governmental regulations. In the event that internal adjustment or recalibration is necessary, refer to
the Service Manual for this equipment.
CLEANING
Monitor
WARNING: To avoid possible shock hazard, disconnect the monitor from the power source and detach the ECG
cable before cleaning.
Clean the exterior surfaces of the monitor with a cloth or swab dampened with a warm water and mild detergent
solution. Do not allow liquids to enter the interior of the instrument.
Patient Cables
Do not autoclave the patient cables.
Wipe the cables using soap and water or alcohol. Never submerge the cables in any liquid or allow liquids to enter
the electrical connections.
ACCESSORIES
Accessories
3 Lead Patient Cable 590170
Disposal
Disposal of devices or consumables must be done in accordance to natural environmental laws.
THEORY OF OPERATION
General
This section provides general descriptions of the circuitry in the R Wave Trigger patient monitor. Refer to the
schematic diagrams at the end of this manual to find the circuits described in this section. Also see the block
diagram in the Troubleshooting section for an overview of the signal flow in the monitor.
The High Voltage Power Supply generates every signal that goes to the crt except the yoke signals, which are
created by the Mother Board.
CRT Theory
Free electrons are created in the base of the crt by applying a voltage to the filament. A large voltage (6.5 kV in this
case) between the anode and cathode accelerate the free electrons from the filament to form a beam. This beam
strikes the phosphor at the face of the crt and forms a dot on the screen. The yokes, one horizontal (X yoke) and
one vertical (Y yoke), aim the beam by deflecting it with a magnetic field. This moves the dot across the screen, and
since phosphor takes some time to cool down, this forms a line across the screen when done quickly. The grids
(three in this case) can be biased to block the beam or to focus it. They are also used as intensity control.
Another way of changing the apparent intensity of the beam is to change the voltage at the cathode, or to quickly
switch the cathode on and off. Turning the crt beam on and off is often referred to as z axis control. This causes the
beam to appear and disappear. The phosphor holds some of the light from the beam while it is off, making it appear
that the beam is on but dim.
Two different strategies are used to display data on the crt: stroke writing for the waveform traces and vertical
raster writing for everything else. To simplify this explanation, we will break the screen down into three parts: a
vertical raster section on the left to show numbers and icons; as stroke written section in the center for writing
waveform data; and a vertical raster section on the right for writing scales, alarm messages, and menus.
The vertical raster written portion of the screen is divided into a grid of pixels, each pixel appearing at a fixed point.
The space between pixels, although small, is always off. As the beam is traveling vertically down one row of
pixels, the beam is turned rapidly on and off. It is on where a pixel is to be brightened and off where the pixel is
darkened. When the crt beam reaches the bottom of one row of pixels (one raster) a retrace occurs, sending the
beam to the top of the next raster to be drawn.
The vertical raster portion of the display is accomplished by turning the crt beam rapidly on and off as the beam
travels vertically from top to bottom of the display at a frequency of about 16 kHz. At the same time that the beam
is rapidly traveling vertically, it is also traveling in the horizontal direction with a trace time of about 4 ms for the
left side of the crt. The size of the raster display on the right side of the crt differs and, depending on what is
displayed, can take from 1.2 ms to approximately 3 ms. Waveform traces are displayed using stroke writing. Stroke
writing is used because it offers a smooth, continuous, high resolution display. Stroke writing entails turning on the
crt beam and varying the y-axis (vertical) deflection of the beam while the beam travels horizontally across the x
axis. It takes about 2.8 ms to draw one complete waveform trace.
Timer Circuit
The Timer Circuit creates a waveform to be used by the high voltage module in the creation of the high voltages
needed by the crt. It starts with the VERT SYNC input.
The VERT SYNC input is a 16 kHz square wave with a duty cycle of 56% that is created on the Mother Board. The
main purpose of this signal is to synchronize the video circuits on the Mother Board while the crt displays rasters. It
is a steady square wave that can be used as a clock.
The Timer Circuit inverts the VERT SYNC signal with a transistor and runs it to a monostable multivibrator (one-
shot timer). The one-shot changes the duty cycle of the inverted waveform from 44% to between 12% and 60%
(varied by adjusting the EFF potentiometer). The ability to change the duration of this signal allows the operation
of the high voltage module to be fine tuned. A small duty cycle means less current to the high voltage module
(longer battery life), but it also means lower display intensity during battery operation.
From the output of the one-shot, the signal is buffered by a power MOSFET and sent to the high voltage module.
The High Voltage Module contains a circuit board that produces several relatively high voltages to be used in the
control of the crt. That circuit board is the Power Module assembly, Part Number 5001-00-XX. More information
about the Power Module is included in a section about that board. For now, be aware that when the High Voltage
Module is supplied with a 16 kHz square wave and the feedback amplifier is working, it produces three voltages: -
100 V, +275 V, and 6,500 V.
Feedback Amplifier
The Feedback Amplifier takes a feedback signal from the High Voltage Module, inverts it, compares it to a voltage
reference (the HV potentiometer), and sends it back to the High Voltage Module. It is used to keep the 6500 volt
output of the High Voltage Module within tolerances. When the 6500 volt output is within tolerance, the -100 and
+275 volt outputs should also be within tolerance.
Adjustment Circuit
The Adjustment Circuit takes the -100 V and +275 V outputs from the High Voltage Module, adjusts them with
potentiometers, and applies them to the grids of the crt.
The first grid, G1, adjusts the intensity of the crt. It is adjusted with the INT potentiometer and should be between -
100 V and -14 V.
The second grid, G2, also adjusts the intensity of the crt. It is adjusted with the CUTOFF potentiometer and should
be between -14 V and +275 V.
The third grid, G3, adjusts the focus of the crt. It is adjusted with the FOCUS potentiometer and should be between
-100 V and +275 V.
The main function of the cathode generation circuit is to produce the CATHODE signal for the crt. CATHODE is a
mixture of the TREND INT and COAX Z signals. It turns the beam of the crt on and off.
The COAX Z IN is a digital signal that originates on the Mother Board, where it is called Z. It tells the crt when the
beam should be off (between waveform strokes or during a retrace). The beam of the crt should be off when this
signal is high.
TREND INT is an analog signal used to make the waveform traces brighter or darker in relation to the raster-written
data on the crt.
The CATHODE Z IN signal enters the cathode generation circuit at the input of the Z buffer. The Z buffer contains
an inverting buffer and transistor. The inverting buffer allows the signal to drive the following transistor without
being severely loaded. That transistor also inverts the signal again and changes the level of the output. The signal
produced by the transistor is similar to the CATHODE signal but not identical.
TREND INT enters the circuit at the trend intensity buffer, which contains a transistor, a diode, and the TREND
INT potentiometer. The transistor buffers and inverts the signal. Next, the TREND INT potentiometer changes
how much TREND INT signal is allowed to enter the CATHODE signal. Finally, the diode prohibits current from
flowing the wrong way through the circuit. As a whole, the trend intensity buffer sinks current from the
CATHODE signal for the desired effects.
The Low Voltage Power Supply is on the Mother Board. Its inputs are 13.4 V~ and 6.9 V~ from the power
transformer.
Regulators U3, U4, and U5 generate +12 V and -12 V from the 13.4 V~. D23 uses the -12 V as an input to produce
-7 V. Regulator U10 and pass transistors Q13 and Q18 generate ±8 V.
From the 6.9 V~, +8 V unregulated is taken from rectifier diodes D5 and D6. Regulator U2 generates +5 V.
Isolated DC Voltages
Isolation transformer T1 passes power from the non-isolated portion of the Mother Board to the isolated portion.
The 16 MICROSECOND SYNC signal triggers U31 which turns Q36 on and off. This creates a varying
current through the windings of T1 to allow power to be passed to transformer's secondary winding on the isolated
portion of the circuit. The rectifiers D50 through D51 generate an unregulated ±6.5 V. Regulator U39 produces ±5
V through pass transistors Q38 and Q39.
ECG Input
Isolated Amplifiers
The ECG input signals (LL, LA, RA) are applied to overvoltage protection devices and to input amplifiers U54 A,
B & C. The signals are then routed to lead selector switch U45. The selected signal is amplified by U46 and driven
by optical couplers U42 and U43, which pass the signal to the non-isolated portion of the circuit.
If any of the leads are disconnected, a large offset on the input signal is detected by U47, which generates the LEAD
OFF signal. This signal is passed to the non-isolated circuits through coupler U38 and sent to the digital circuits to
generate the LEAD OFF alarm. This signal, through Q46, disables the ECG amplifiers.
Shield Drive
The common-mode error signal that serves as input to the shield drive circuit is derived from the signals summed
through R242 and R243. This signal drives the shield drive amplifier U47. The output of the shield drive amplifier
returns to the patient cable serving to prevent 50/60 Hz power interference.
Lead Selection
Lead select switch U45 selects proper signals according to the levels of the L0 and L1 signals from the digital
circuitry. These signals are set according to the lead selection made by the user.
One of the combinations of the L0 and L1 signals indicates TEST mode. In this mode, the lead select switch is off,
and a 70 bpm signal generated by U44 is sent to ECG amplifier U55D.
ECG Waveform
ECG Filter
The signal from the output of coupler U40 and then through a low-pass filter to amplifier U26B. The signal at
U26A Pin 1 is the filtered output. This signal is switched in or out by FET Q30 which is controlled by the NOTCH
signal from the digital circuitry. Potentiometer RV13 (60 Hz) and RV14 (50 Hz) is used to adjust the notch filter.
The processor senses the line freq from U18B and automatically switches in the correct notch filter.
X1000 Output
The output at U26 Pin 1 is the X1000 signal that is sent to the rear panel for use with a recorder or other equipment.
ECG Gain
The signal from U24 Pin 1 goes to ECG gain amplifier U24 Pin 7. The gain of this circuit is determined by resistors
switched into the circuit by U25. This switch is controlled by signals G1, G2, and G3 from the digital circuitry.
The gain depends on the settings made by the user. The output signal from U24 Pin 7 goes to the A/D converter in
the digital circuitry.
A/D Converter
The A/D converter is U21 along with amplifier U22A. The converter samples the analog ECG signal and generates
a series of 1,024 eight-bit numbers which can be stored in memory and then sent through the D/A converter for
ECG waveform display.
Cardiotach
Peak Detector
The peak detector generates a pulse at the peak of each R wave. The time between these pulses is used to compute
the heart rate. See the Peak Detector block diagram below.
4.5 SECOND
TIME-OUT
FROM CPU
PACEMAKER
RESET
U20
The ECG signal is picked off by amplifier U26C through a filter made up of R126 and C66 and the components in
the feeback of U24D. This removes the baseline and T waves from the ECG signal. The signal is rectified by
diodes D33, D34 to make sure that the R waves are always positive. The output of the filter and rectifier at U23 Pin
8 is a group of positive pulses for each QRS complex. The first large pulse is the R wave.
For a pulse to be detected, it must exceed a threshold, which is based on the amplitude of the last accepted pulse and
is fine adjusted by potentiometer RV12. Sample and hold circuit U19 with U23D holds the peak level of each
accepted pulse to form the threshold for the next pulse.
When a pulse exceeds the threshold, U23B Pin 7 goes low. When the peak occurs, U23B Pin 7 goes high. The
time between when the signal exceeds the threshold and when the peak occurs must be between 7 and 70 ms for the
pulse to be accepted. The 7 and 70 ms timing pulses are generated by one-shots U30. If the signal is accepted, the
CPU generates a 10 ms EVENT pulse. The leading edge of the EVENT pulse occurs at the peak of the R wave.
This signal goes to the digital circuitry for heart rate computation. Each filtered and rectified QRS complex usually
generates more than one acceptable pulse. Therefore, the EVENT signal usually has two 10 ms pulses for each
heartbeat. A 200 ms blanking in the digital circuitry eliminates all but the first pulse.
A 4.5 second time-out is produced by the CPU. If no acceptable pulses occur for 4.5 seconds, this one-shot times
out and generates a signal that turns on U19. This causes the threshold circuit to acquire a new level.
Whenever a pulse is detected that is less than 7 ms wide, it is assumed to be a pacemaker spike. In this case, U20 is
triggered, generating a 30 ms pulse, which turns on Q22 and Q23 to discharge the filter capacitors and turns on Q31
to inject the last threshold level into the peak detector comparator input.
The Trigger Output at the rear panel connector is derived from the EVENT signal. This output is one 1 ms pulse for
each QRS complex, with the leading edge corresponding to the peak of the R wave. The signal level is -12 V to
+12 V. Other output pulses (+5 TTL) may be generated by removing U12 and changing C56 and R111. See
component matrix on schematic.
Rate Computation
The EVENT signal goes to PLD U36 in the digital section where the 200 ms blanking removes all but one pulse for
each heartbeat. The ouptut of the flip-flop goes to the interrupt input of the microprocessor U48. The
microprocessor measures the time between interrupt pulses and computes the heart rate.
Alarm Generation
The microprocessor compares the current heart rate to the limits set by the user. If the limits are exceeded, the
appropriate alarm signal is generated.
CRT Signals
Refer to the Key Signals portion of the Testing and Troubleshooting section of this manual for diagrams of each of
the crt signals.
Vertical Deflection
The display generation is broken up into two timing periods, one for the raster display (digital information and
alarms) and the other for the stroke display of the ECG waveform. Each time period is 9.2 ms.
The raster display also contains two time periods. The first is for the left display and the second for the right
display. The left display contains the heart rate display, the setup information, and the ALARMS OFF indicator.
The right display contains the alarm indicators.
During the stroke display period, the vertical deflection is the ECG waveform signal. This signal is selected
through FET Q15 to the vertical deflection amplifiers U6 and U9 and transistors Q1, Q2, Q5 & Q6. During the
raster display period, the vertical deflection is the vertical ramp generated by U8B, with its timing set by the
V SYNC signal. The vertical ramp is passed through Q15 to the vertical deflection amplifiers.
Horizontal Deflection
The horizontal ramps are generated by U16 and U17 with timing controlled by the H SYNC signal. Four ramps are
generated in each timing cycle. The two small ramps are for raster display, and the large ramps are for stroke
display. An offset for the right side raster display is added to the circuit through U11D.
Horizontal Correction
Because the face of the crt is flat and the beam originates at a single point, the speed of the beam as it strikes the
face is greater near the edges than at the center. Therefore, correction circuitry is required to produce a waveform
that is linear over the entire screen. The linearity correction is provided by U17 and associated components.
The correction circuitry produces two signals that are added to the horizontal ramp to correct the linearity. First,
U16A integrates the horizontal ramp to produce a parabolic waveform. This waveform is the mathematical
equivalent of the speed of the beam across the crt face. This signal is inverted by U16B. The original signal and its
inverse are fed into the SYM potentiometer RV6, which selects the portion of the positive or negative waveform
required to produce a symmetrical trace between the left and right sides of the crt.
The second signal is an S-shaped waveform generated by U17A, which integrates the parabolic signal from U17B.
This S-shaped signal, when added to the horizontal ramp has the effect of slowing the beam near the edges of the
screen. The amount of this signal added to the ramp is adjusted by the S potentiometer RV4. This allows
adjustment of linearity between the edges and the center of the screen.
Using the adjustments described above, it is possible to achieve a linear signal across the entire crt screen for any
crt. These adjustments should be done whenever a crt or Mother Board is replaced. The complete horizontal
deflection signal including correction is summed at U7 Pin 2 and goes through amplifier U7 and transistors Q3, Q4,
Q7 & Q8.
Unblanking
The unblanking (Z) signal indicates when to brighten the crt. This signal is created in the digital section. For the
raster display, short pulses cause the brightening of each dot needed to generate the characters to be displayed. For
the stroke portion, the signal is high during the entire horizontal sweep displaying the ECG waveform.
Audio Circuits
Two types of audio is produced by the monitor: a QRS beep and an alarm tone. The QRS beep is a short tone that is
generated each time a QRS complex is detected by the monitor. The alarm tone is produced when any of the alarms
are triggered. The alarm tone can be disabled by pressing the front panel
ON/OFF button.
ECG Audio
The audio signal originates in the digital circuitry. The ECG AUD signal is enabled at U51 by an output from the
processor's I/O port U51 Pin 11. This signal allows the audio frequency from U51 to be passed to the audio
amplifier. Two levels of ECG AUD, through resistor R90, go to the rear panel QRS VOLUME switch as AUDIO
HI and AUDIO LO, where the user can select a high, low, or off setting. The selected signal then goes to the audio
amplifier U13 and transistors Q19 and Q20, which drive the speaker.
Alert Audio
The ALRT AUDIO is enabled by the output from U36 Pin 18 in the digital circuit The ALERT AUD signal passes
through volume adjust potentiometer RV11 to the audio amplifier U13.
Digital Circuitry
The digital circuitry consists of the microprocessor and its associated components. The 8 MHz clock is generated by
the F1100.. The microprocessor is U48. U49 is the read only memory (ROM), which contains the program to
operate the system. U32 is an input/output (I/O) port for passing information to and from the microprocessor
circuits. U52 is random access memory (RAM). This memory is for temporary storage of data that the
microprocessor must store. U36 generate timing signals for the system.
For the stroke display, the microprocessor sends the digital ECG waveform through FIFOs U27 and U28 to D/A
converter U29 to generate the D/A signal that gets switched into the vertical deflection signal during ECG
waveform display.
For raster display, the microprocessor sends the data through U36 and U15 to generate the Z signal. This signal
brightens the crt at the proper times to generate the characters.
ECG Test
Equipment Required:
1. Connect the patient cable and simulator to the monitor and set the simulator as follows:
• Amplitude of 1.0 mV
• The heart next to the heart rate indicator flashes with every QRS complex displayed.
4. Change the heart rate on the simulator to 40 bpm. Notice the following:
12. Set the simulator heart rate to 120 bpm. The HR HIGH message is displayed, and an audible alarm
sounds.
Troubleshooting
The monitor is made up of the following parts:
This manual provides information to help determine which of these parts might be the cause of a problem. A simple
block diagram of the monitor is included in this section to help. See the Disassembly procedure in this manual for
removing the top cover.
General Checks
The first step is to determine what doesn't work. If nothing works, check those things that you would check on any
instrument.
3. Check the fuses on the low voltage side inside the monitor.
CRT Display
The display is generated from signals developed on the Mother Board. The crt gets its high voltage and cathode
signal from the High Voltage Power Supply board.
If there is no picture at all, the problem is most likely the crt itself or the HVPS board. See the Key Signals section
to check these components.
If there is a display but either the horizontal or vertical deflections are wrong, the problem is most likely the Mother
Board.
Mother Board
The Mother Board has all circuitry for processing the ECG, computing heart rate, and generating the alarms and the
audio. Therefore, any problems with these functions are caused by problems on the Mother board.
Once you have determined the major component causing the problem, refer to the Disassembly section for
instructions on replacement.
Troubleshooting Table
Block Diagram
AUDIO
CIRCUITS SPEAKER
U13,Q19,Q20
FRONT
PANEL
RAMP HORIZONTAL
GENERATOR DEFLECTION
U16A U7
HORIZONTAL
CORRECTION
U17
HIGH VOLTAGE
POWER SUPPLY
AND VIDEO CRT
AMPLIFIER
RASTER
GENERATOR
U8B
A/D
CONVERTOR
U21
TRIGGER
PEAK CIRCUITRY TRIGGER
DETECTOR U18A, U12 OUTPUT
U23B
ISOLATED
TEST
CIRCUIT
U44
ECG
INPUT
SHIELD
DRIVE
ECG X1000
BUFFER
MOTHER BOARD
Voltage Measurements
The following procedures include voltage checks, key waveforms, and some adjustments. These procedures are
only for qualified technicians with appropriate equipment, such as an oscilloscope and digital voltmeter. In most
cases, the procedures in the Troubleshooting section of this manual are adequate for determining which major
component of the monitor is the cause of a problem.
Schematic diagrams of the monitor are included at the end of this manual. See the Disassembly procedure in this
manual for removing the top cover.
Measure the following voltages with a DVM using the ground connection lug on the rear panel as the reference.
See Mother Board layout for test point (TP) locations.
Unregulated Voltages
Regulated Voltages
Isolated Voltages
Measure the following in the isolated section of the Mother Board using ISOL GND TP as the reference. See
Mother Board layout for test point (TP) locations.
Key Signals
The following are a few signals to help verify the proper operation of the monitor. Use an oscilloscope and refer to
Mother Board layout for test point locations.
ECG Waveform
1. Press and hold the TEST button on the monitor's front panel.
2. Put the scope probe at the rear panel X1000 output. The signal should appear as 1 V peak pulses. The
same waveform should appear on the monitor's display. The rate indication on the display should be 70
bpm.
This verifies the operation of the ECG circuitry from very near the front of the monitor's ECG section. To
verify the remaining circuit, use a simulator at the ECG input to generate a 1 mV ECG signal.
CRT Display
The following are signals generated by the Mother Board for creating the display. If these signals are correct but
the display isn't, the problem is not on the Mother Board. See Mother Board layout for test point locations.
For each of the following the signals, trigger the oscilloscope (negative) on Raster Enable (RE test point).
Raster Enable
RE Test Point
Vi Test Point
Hi Test Point
To verify operation of the High Voltage Power Supply board, check the dc voltages as indicated in the Voltage
Measurements section (High Voltage Power Supply).
Also check the following dc filament voltages at the crt socket harness.
Adjustment Procedure
The following adjustments are not meant as field procedures. However, if these procedures are attempted, they
must be done correctly and according to the procedures listed here. See the Disassembly procedure in this manual
for removing the top cover.
Equipment Required
ECG Signal
1. Press and hold the TEST button on the monitor's front panel.
2. With an oscilloscope, look at the signal at the X1000 output on the rear panel.
1. With the monitor's front panel controls, set the filter to ON and the lead selection to Lead I.
2. Insert a 10 mV p-p 50 or 60 Hz sinusoidal signal into the monitor's front panel ECG input. See the
diagram below for a suggested input circuit.
3. With an oscilloscope, look at the signal at the NOTCH test point on the Mother Board.
4. Adjust RV13 for 60 Hz and RV14 for 50 Hz for minimum amplitude signal at the NOTCH test point. This
signal should be ≤600 mV p-p. To turn on the correct filter set TP21 high (+5) for 60 Hz and ground for 50
Hz. A clip lead tied to +5 or ground will accomplish this.
5. From the front panel, set the filter to OFF. The signal at the NOTCH test point should be approximately 10
V p-p.
6. To test the auto notch function, plug unit into a variable freq. power generator and slowly change the line
freq. from 50 to 60 Hz.
Peak Detector
2. Connect the ECG simulator and adjust the ECG level so that only one pulse is present at U23 Pin 7.
5. Set the simulator for an ECG input signal of 300 µV. Put the oscilloscope probe at U23 Pin 7.
6. Adj RV12 (T'HLD) until the signal at U23 Pin 7 changes to two pulses.
CRT Display
• Beam Centering
1. Reduce the intensity using the intensity pot at the top of the High Voltage Power Supply board.
3. Adjust the rings at the rear of the crt to place the beam in the center of the screen.
• Yoke Straightening
1. With no ECG input signal, loosen the screw holding the yoke clamp at the rear of the crt.
2. Rotate the yoke so that the trend line on the display is horizontal.
• Horizontal Adjustments
1. Put the oscilloscope on U16 Pin 1. There should be two large ramps and two small ramps.
3. Adjust RV5 so that the trend line (ECG waveform line) is 132 mm long.
4. Adjust RV16 (SS ADJ) so that the ALARMS OFF display area is 33 mm across.
5. Press the TEST button on the monitor's front panel. When pulses appear across the entire screen, press
FREEZE to stop the motion.
6. Adjust RV4 (S) and RV6 (SYM) to make the pulses evenly spread across the screen. The S adjustment
controls the linearity between the middle and ends, and the SYM adjustment controls the linearity from end
to end.
7. Press FREEZE and release TEST so that there are no pulses. Press the MENU button.
8. Adjust RV9 (RSTP) so that the right end of the trend line just touches the left of side the menu box.
• Vertical Adjustments
2. Adjust RV1 (RASTER) so that the negative peak of the signal at U8 Pin 7 is at -2.5 V.
5. Adjust RV3 (V BAL) to center the menu box between the top and bottom of the screen.
7. Adjust RV8 (DA0) so that the trend line is at the same level as the center horizontal line in the second
menu.
1. Adjust the Focus and Intensity adjustments near the top of the High Voltage Power Supply board for the
desired display.
2. Adjust the Trend Intensity adjustment on the HVPS board to set the trend line intensity separately.
Use a leakage tester or use the ac voltage range on a digital voltmeter to measure across a 1 kΩ resistor in parallel
with a 0.15 µF capacitor and 10 Ω resistor.
1. Measure between a known ground and the monitor's ground post on the rear panel. Wth the line cord
properly grounded, with the ground removed, and with the line cord reversed, the readings should be less
than 20 µA.
2. With the line cord grounded, measure between shorted patient leads and the hot side of the ac line. The
reading should be less than 10 µA.
This section describes procedures for disassembly of the monitor for replacement of major components. For
reassembly, use the same procedures but work in reverse. Always remember to save all hardware and to re-install
any connectors removed during disassembly.
2. Remove the two screws in the rear that attach the top cover to the rear panel.
4. To remove the bottom cover, turn the monitor upside-down and use the same procedure.
The High Voltage Power Supply board is mounted vertically near the front of the monitor near the crt.
4. Remove the crt anode cap from the front of the crt by pulling up on the rubber from the edges.
5. Pull the High Voltage Power Supply board up and out of the card guides.
4. Loosen the screw holding the yoke clamp on the crt and slide the yoke assembly off.
3. Remove the two screws holding the crt bracket on the center side of the crt.
4. Tilt the crt toward the side of the monitor and guide it out of the unit.
5. After replacing the crt, do the CRT Display adjustments described in the Testing and Troubleshooting
section of this manual.
2. Loosen the two captive Philips-head screws inside the recorder housing.
3. Turn the monitor over and remove the screw at the front of the Mother Board that attaches the board to the
front panel.
4. Remove the four screws across the bottom of the rear panel.
5. Squeeze the tabs on each of the plastic standoffs and gently pull on the board to remove it.
3. Use a knife to work a corner of the Touch Panel loose and pull the panel away from the front bezel.
4. Feed the ribbon cable through the slot in the front plate.
5. Make sure the clear plexiglas window is in place in the bezel. Then feed the ribbon cable on the new
Touch Panel through the slot in the bezel.
6. Peel the protective backing from the Touch Panel and carefully press the panel onto the front bezel.
PURCHASING INFORMATION
Parts List
The following are the part numbers for the major components in the monitor.
CRT 570000
LED 1062-00-01
• Minimum order is $100 on all orders requiring an invoice prior to payment. When payment accompanies
purchase order, no minimum applies.
• A 15% restocking charge will be made on an instrument returned for credit or exchange, unless a waiver is
issued, in writing, by an employee of Ivy Biomedical Systems, Inc.
• One instruction manual and service instructions are supplied with each instrument. Additional copies may be
ordered at $25 each.
SCHEMATICS
GND
OPTION U1 C1 +12V POWER ON 03 REFERENCE REJECTED ECR 803. GP 7-27-98
TP9 LED 04 REVISED PER ECN 805. GP 7-27-98
H.S. .1UF CER J207 05 REVISED PER ECN 817. GP 8-5-98
+12V LT1085CT-12 R81 06 REVISED PER ECN 868. GP 1-25-99
1
J201-16 2
07 REVISED PER ECN 934. GP 4-12-99
A 470 OHM 08 REVISED PER ECN 970. GP 8-6-99
1/2W 10% A 09 REVISED PER ECN 1052. GP 11-20-00
+10/13.5VDC R97 10 REVISED PER ECN 1104. GP 7-17-01
J201-32 11 REVISED PER ECN 1105. GP 7-17-01
100 OHM 12 REVISED PER ECN 1216. GP 12-12-02
2W 5% +12X 13 REVISED PER ECN 1217. GP 12-12-02
U4 TP5 14 REVISED PER ECN 1218. GP 12-12-02
F3 D8 LM78L12ACZ 15 REVISED PER ECN 1251. GP 2-4-03
13.4_VAC
VI VO
J201-25 FC FC +8V
GND
3ASB 2AG 1N5624 +12X Q13 TP18
D
C4 2N3906 D
2
+12V 49.9K
MC34002P
D11 D9 + C13 VI VO
+ 5
GND
1N5624 1N5624 10000UF R43 R47 7
FILTER_GND 35V U10B 6 + C23
J201-21 + C3 499 OHM 1K -
U3 2200UF 1000UF 16V
LT1085CT-12 35V R44
FILTER_GND H.S. 9.09K
J201-22 R46
D7 20K
1N5624 + C30
A_GND A GND D15
J201-3 ICL8069CCZR 6.8UF TANT
TP24
A_GND
J201-33
A A C31
A_GND A
J201-34 + C12 R1 + C6 +12V
10000UF 1K 6.8UF R48
A .1UF
8
35V TANT CER 10K .1%
REC_GND D12 2 10PPM
R2 R51 R49 1
TO SHT. 4 1N5624 U10A 3
121 OHM 499 OHM 1K
MC34002P
+ C5 C32 R50 + C34
4
2200UF 10K .1% 1000UF 16V
35V B 10PPM
2
U5 A .1UF
ADJ
F4 D10 LT1033 CER -12V
C 13.4VAC H.S. R52 E1 3C C
J201-29 VI VO -8V
FC FC 10 OHM
3ASB 2AG 1N5624 Q18
-12V D23 2N3903
-12V TP10 TP11
*
J201-11 1N5338B
-12V -8V
-7V
TRIGGER PULSE OUTPUT
TO SHT. 4
R272 TP20
+5V TRGOUT 1ms 100ms/150ms 200ms REAR PANEL
1K -7V J201
+5V
TP17 TO SHT. 4 194000 TO SHT. 4
ICM7556 TRIGGER C56 .01UF 1UF .22UF
A_GND Z_GND 1 2 Z
14
R113 U18A 4 3 4 AUDIO_HI
10K 5 6
TO SHT. 4 C59
* R111 60.4K 60.4K 806K A D
R
1000UF U2
* C45 C46 D ECG_VOLUME
25 26
D
VO
27 28
+
16V LM2940T-5.0 D 13.4VAC
ECG_AUDIO 29 30 ALERT_OUT +10/13.5VDC
H.S. D 31 32
D 100UF 25V 100UF 25V A_GND A_GND
GND FIL_GND 33 34 FIL_GND
D D 35 36
14
A A
1
VI
+5V
+V
-V
2 1A 1Y 3 1
D28
1N4148 2
U18B OUTPUT
10
4 2A 2Y 6 3
ICM7556 5
F1 D5 D21 TO SHT. 4 2B
OPTION 2 D 4
R
B B
6.9_VAC R83 8 9 9 8
J201-13 TR Q LNFREQ 10 3A 3Y
+5V J206
FC FC 100K 3B 174000
1.5ASB 1N5624 1N4148 DIS 13
12 11 +12V
GND
2AG R84 11 12 13 4A 4Y
10K CV THR 4B R42 +5V
F2 D6 + C14 D U12 1K
6.9VAC 15000UF MC1488P R40
J201-17 FC FC 16V SH D C47 R271
7
402 22.1 OHM
1.5ASB 1N5624 D
2AG OHM
3
FIL_GND
J201-35
.01UF
MP
D
OPTION 1 2N3903
Q12
2 R41
10K
R39
10 OHM
2 Q11
2N3903
1
FIL_GND D
J201-36 D GND R38
TP15 475 OHM
D_GND +12V
J201-4
D
R75
OPTION 3 D
D_GND
J201-5 47.5 OHM TO SHT. 4
Q20 R96
T150- +5V
3
+12V 2N3903 C
D_GND C50 B2 SPEAKER 10K NOTES:
J201-9 TP14 AUDIO_HI
E J201-6 1. ALL RESISTORS ARE 1/4W 1% U.O.S.
AUD. VOL.
D
1 1
.1UF
7
2
TO SHT. 4 C58 LF351 499 OHM C 5.11K 3. NOT USED: JP2,JP3,U19,U37,U50.
G3 Q21 NPO 3G
2N5460
3
1
A 4.75K MP 1.47M 1N4148 R79 |LINK A
30.1K ECG_VOLUME 3.01K |C:\ORCAD\SHEET\SH5119\5119-002.SCH
R92 J201-27 D22 R80 |C:\ORCAD\SHEET\SH5119\5119-003.SCH
D C52 49.9K .1UF D -12V 1N4148 1.47M
D -12V CER |C:\ORCAD\SHEET\SH5119\5119-004.SCH
10000PF |C:\ORCAD\SHEET\SH5119\5119-005.SCH
D NPO D R95
C53 AUDIO_LO
AUDIO_OFF J201-15 IVY BIOMEDICAL SYSTEMS INC.
J201-10 R87 R77 1K D
.1UF MP 11 BUSINESS PARK DRIVE
3.01K 301K AUDIO_SHIELD BRANFORD, CT. 06405
SPEAKER_GND C48 J201-23
.022UF MP R86 C49 (203) 481-4183
J201-18 30.1K .0047UF MP D
D D Title
MOTHER BOARD 1 ASSY, WITH RECORDER OPTION
SPEAKER
J201-14 Size Document Number Rev
C 5119-00-01 (-03 SCHEMATIC) 15
+
09 REVISED PER ECN 1052. GP 11-20-00
6.8UF A 10 REVISED PER ECN 1104. GP 7-17-01
R64 11 REVISED PER ECN 1105. GP 7-17-01
1.5K TANT 12 REVISED PER ECN 1216. GP 12-12-02
PN4393 +12V 13 REVISED PER ECN 1217. GP 12-12-02
14 REVISED PER ECN 1218. GP 12-12-02
Q14
S 15 REVISED PER ECN 1251. GP 2-4-03
1
D17
RV8 3G -12V
D/A 0 R8 + C9
D
1N4148 100UF D
20K TOP D C17 499 OHM E 25V
2
+8V -8V R56
3
+12V 2M A
C28 R7 B1 Q2
R57 A H.S. 2N6490
D/A R36 .1UF 4.99K C
R65 1K CER
3
TP8 A 1M C
2
49.9K .1UF B2 Q6
CER
7
TO SHT. 4 Q16 C35 2N3903
R66 1 2 2
+ 6.8UF TANT
D/A D S 6 E VERT.
1
6.04K PN4393 3 U9 D2 TP1
1N4148
7
D18 LF351 A 1
G A 3 + 8
3
1N4148 6 R19 R5 R6
4
1
5
RV2
V. GAIN 2 - U6 5
R67 C29 100 OHM 100 OHM 27 OHM
1K TOP LM318 1/2W 5% 1/2W 5%
200K C15
A
4
100PF
D26 .1UF A NPO C8
1N4148 -12V CER A .47UF
TO SHT. 4 Q15 +8V MP
Z_INT PN4393 D1 A
MC34004P 2 1 1N6263
S D E
1
10 + RV3 R26 C18 R9
8 D16 V. BAL. B2 Q5 470 OHM
9 -U16C G 1N4148 20K TOP 80.6K 2N3906 C 1/2W 10%
2
.1UF A C
+5V A B1 Q1
3
CER R3 2N6487
R55 A 4.99K H.S.
3.01K R4 E
3
R62 R60 499 OHM
10K C7
Vi YOKE
200K TP3 J202
+
A
100UF A VERT. YOKE 1 5
R63 D24 -12V 25V 4
R23 VERT. YOKE 2
10K 1N4148 VRAMP 3
4.99K 2
C
______ TP7 C16 +5V 1 C
A RASTER R22 R17
12 + ______ 1 OHM ____
TO SHT. 4 14 ENABLE 806 OHM MOVE
470PF NPO 5W 5% R58
R_ENABLE- 13 -U16D TP23 10K TP12
A A
HOR. YOKE 1
HOR. YOKE 2
D25 TO SHT. 4
MC34004P R59
______ +5V MOVE-
H SYNC 10K
TP13 R61 1N6263
+5V 16 R68
TO SHT. 4 10K D19 14 15 49.9K
-8V
H_SYNC-
U11D RV9
C36 1N6263 DG201ACJ D20 "RT. SIDE" TEXT
13 +12V 1N4148 50K TOP
A .1UF U11A
CER DG201ACJ A
2 3 U11C
DG201ACJ
5 1 9
+12V
A 10 11
12
U11B
4
DG201ACJ
-12V C37 8 C44 + C11
.1UF -8V R15 100UF
CER 6 7 C21 499 OHM 25V
A -8V .01UF MP E A
3
C43 R110 R14
C38 +12V 162K .1UF A 4.99K B1 Q4 HORIZ.
R106 C42 CER H.S. 2N6490 TP2
3
+12V C
C40 .1UF 52.3K R109 2 Q8 C
2
MP .1UF MP B 2N3903
.1UF A
4
RAMP R105 R74 13 - 10K
14 R73 2 - CER E Hi
A
1
RV15 .1UF
4
7
B B
1 R104 6 - A MC34004P 10K 5 +U17B 1
80.6K MC34004P
3 +U16A 7 A 2 8 R13 R12
10K 5 +U16B
11
MC34004P A 6 R20
S MC34004P 3 U7 5 100 OHM 470 OHM
A C41
2
100 OHM
11
4
.1UF 100PF
1N4148 .1UF A -12V CER NPO 470 OHM
D -12V A 1/2W 10%
1
R72 CER
200K RV4 D3
"S" 1N6263 A
5K TOP E
1
A R135 D31 C22
______ 110K 1N4148 B2 Q7
V SYNC R27 2N3906 C
A
2
TP6 9.09K A C
.1UF
+5V A B1 Q3
3
RV6 R29 CER R10
C24 SYM. 4.99K H.S. 2N6487
R33 RV16
50K TOP 5K TOP 30.1K R11 E
3
1K SLOW SWEEP ADJ. 499 OHM
220PF
NPO E R70
1
A G 15K 15K
CER
3
A 2 RV5 R18
Q10 H. GAIN R30 2.4 OHM NOTES:
-12V 1 80.6K
S2 1D 3 U8A 20K TOP 5W 5%
PN4393 1. ALL RESISTORS ARE 1/4W 1% U.O.S.
A
C25 MC34002P A A
C27 2. LAST USED: C144,D73,F4,FB4,J207,JP5,
4
5
RA SHIELD 20K 10K 10K X500 10 REVISED PER ECN 1104. GP 7-17-01
1 A
LA DRIVE U38 REMOVE R273 11 REVISED PER ECN 1105. GP 7-17-01
2 LL TP43 D64 R239 CNY17-3 FOR FILTER 12 REVISED PER ECN 1216. GP 12-12-02
1N4148 I .1UF 30.1K
8
3
4 CER OUT ONLY 13 REVISED PER ECN 1217. GP 12-12-02
+ 3 14 REVISED PER ECN 1218. GP 12-12-02
4
5
SHIELD_DRIVE FB4 R267 1 15 REVISED PER ECN 1251. GP 2-4-03
6 U47A 2 R238 I I
150013 10K TL062 - R204 RV17 TO SHT. 4
10K ECG BAL +12V
C144 1K +5VI C94
.01UF D65 C133 + C131 20K TOP LEADS_OFF
4
CD 1N4148 6.8UF TANT -8V +8V +8V
A FILTER
I .1UF A X1000
R228
4
I I .1UF C I -6.5V IN TP29 TO SHT. 1
80.6K CER
3
-6.5V CER -6.5V R185 3 + J201-24
D B2 Q42 C119 R186 JP4 1 R181 D
402K 162K ECG_X1000
2N3903 R218 R184 2 -U26A
10K U40 75K A FILTER
OUT MC34004P 100 OHM
E 150PF NPO
1
Q40
11
IL300G C93 C91
I 2N3906 1 8 .033UF
-12V
1
6 - E C95 RV18 INSTALL R183 MP
7 R217 2 B .01UF 100K TOP FOR FILTER .1UF A A
R235 5 +U39B R219 2 7 MP ECG OUT ONLY
499K 499 OHM C +6.5V GAIN CER
3 6 6 - R182
I MC34004P 100 7 R183
3
-6.5V OHM 5 +U26B 100K C88
4 5 100K
SHIELD +5VI A MC34004P C92 +12V
.1UF A
RA +5VI I A
J204-1 R147 CER
+6.5V .0047UF MP
4
C139 NOTCH
100K TP30 2 - MC34004P
R234 R146 13 - 1 R169 5 + TO SHT. 4
FB1 D66 R246 10K C85
150013 FDH333 50M 5% .1UF I 2N3906 14 3 +U24A 7
1
GUARD 100K 12 +U26D MC34004P 100K 6 -U24B ECG
CER
4
Q41
2 1UF MP
R244 R245 R171 D45
11
3 + D55
10K 1W 100K 2 -U54A
1 LT1058ACN
ISOLATED R187
90.9K
A MC34004P
R167 R170 C87
121K 1N4148 C89
3
10% R233 INSTALL R166 60.4K 301K C90 C86
C137 Q43 1N4148 150K FOR 16.9K .022UF .022UF
LD1 C71 .1UF A .01UF
NE2HC2A
220PF D1 2S 11 D53 D60 D61 NEONATAL A -12V CER MP MP D46 MP
NPO S
2
PN4393 D
I
NON-ISOLATED .033UF 5% MP
1
I PN4393 3G R144 A
G 1N4148 1N4148 1N4148 R148 R168 R176 G3 1N4148
C140 Q32
3
1
D35 C65 S 110K
.1UF I C72
2
1N4148 .1UF MP
CER 1N4148 9 - R173
-5VI R247 8 R145 A
R220 10 +U26C 10K
1M 100K R137 R136 .033UF 100K
+6.5V 30.1K 30.1K 5% MP +5V R175
A MC34004P Q28 C67 -8V
R242 D36 C3 1E
I 100K R259 1N4148 7.5K
INSTALL
C122 C129 4.75M FOR
2N3906
.1UF +5V A B NEONATAL
CER +6.5VI C142 RV13 RV14 R143 R134 R177
2
LA +5VI R243 60 HZ ADJ. 50 HZ ADJ. +12V 10K
J204-2 100K .1UF 200K TOP 200K TOP 10K 221 OHM
CER X500 R101
16
I LT1101CN8 I 470PF 10K R142
5
2 I NPO TP42 -12V 33.2K
8
C
FB2 12 13 6 7 C125 LT1058ACN D 60/50 C
VD
R250 TL062
1
D68 14 X0 X + 8 R256 12 + TP21
150013
GUARD FDH333 50M 5% 15 X1
3 C126 3 U46 1 14 R229 3G R98 1 + 3 R178
LT1058ACN X2 Y .01UF 402K PN4393
11 X3 MP
- 1UF 13 -U54D U14A 2
-
R100
60/50
R248 R249 5 + I 49.9K Q27 1M 210K
MP S 10K TO SHT. 4 R179
7 1
2
10K 1W 100K Y0
6 -U54B 5 C130 D C141
4
1
10% Y1 80.6K
2 Y2 -6.5VI R257 A
C138 Q44 4 U45 G3 R126 R180
LD2 220PF D1 2S 6 Y3
CD4052BE 402K D -12V
.1UF I 1000PF NPO
1
NE2HC2A NPO 10 INH
Q46 649K (ADULT) 33.2K
PN4393 9 A CER PN4393 S PN4393 3G R99 7 + 5 130K (NEONATAL) R138
VS
VE
2
B R255 U14B 6
I 1M - 12.7K
G D72 I Q26
3
2
TL062 R102 FOR NEONATAL
FDH333 -6.5V 1M R254 10K 6.04K
1N4148 1K A R140
C124 -8V C66 D33
-5VI R251 R252 +6.5V +5V A
.1UF CER 2.21K
1
1M I I 22.1 .047UF MP R141
OHM R253 U42 R131 1N4148 R132
21.5K 1M 10K 1K
CNY17-3 +5V A
C64 Q23 R127 R128 U25
3
PN4393
G
10
11
12
DG508A
7
6
5
4
R223 +5V C68 100K 100K C70 -12V
R124 D1 2S
A .1UF 6.8UF
S8
S7
S6
S5
S4
S3
S2
S1
LL R224 1K 1.47M +
14
+5VI 10K CER TANT GND
J204-3 +5VI R123 A C69 .0047UF MP 13 +12V
4
8
RATE V+
RV19 49.9K A 1UF MP A
V+
R
200K TOP R241 R133
AOUT
FB3 R258 R232 -6.5V 2 3 A 3 -12V
3
150013 D71 TR Q G V-
EN
A2
A1
A0
GUARD FDH333 50M 5% 499K 499K +5VI +6.5V +5V U20 7 10K
LT1058ACN R163 ICM7555 DIS D1
D59 +5VI +5VI 100K 2S 13 -
1
R261 R260 R227 R122 R130
15
16
10 + 5 6 14 9 -
1
2
8
8 U43 CV THR +5V 12 +U24D 8
R230 Q22
V-
10K 1W 100K 9 -U54C 1M C63 301K 10K 10 +U24C
14
10
1N4148 10K CNY17-3 A PN4393
4
10% .01UF A MC34004P +5V
C143 Q45 D63 CD C62
C123 A MC34004P
1
R
V+
R
LD3 220PF D1 2S D56 1N4148 6 5 8 9 .1UF MP G1
NE2HC2A NPO TR Q TR Q
R225 A R129 D34
PN4393 I I DIS 1 .01UF DIS 13 10K A 1N4148 G2
I 1N4148 MP R226 1K
G +6.5V 3 2 11 12
3
1M C L1
2N3903 2B R264 C114 R222
100K 1UF MP L0
Q47E R263
I 1K
1
1M
-6.5V RV12
C112 T'HOLD
-6.5V 20K TOP
+8V -8V
.01UF MP R118
49.9K C84
+6.5V R156 R154 .22UF MP
C118 C61 200K 2M
R158 R164
.1UF I 1UF MP 60.4K 100K
CER 10 + +12V
4
8 R120 D3 1S 13 - C78
R214 2 - D49 9 -U23C 14 R155 D44
1 R213 49.9K Q48 12 +U23D
I 90.9K 3 +U39A R197 2N5116 100K A
TL034ACN .1UF
4
C113 MC34004P 20K D73
1N4148 G A TL034ACN 1N4148
2
R215 1N4148 CER
2
.01UF 10K A 2 - B
402K C83
2
4
11
MP C117 .01UF MP 1
U41 A 3 +U23A E1 3C
CNY17-3 TL034ACN
.1UF I R165 A Q31
+5V
11
-6.5V C77 -8V
1
R208 +12V 6 - T2
374 OHM 499 OHM 25UH R159 TO SHT. 4
1N5819 7 TP31
R161 5 +U23B D41
MC34004P R206 -8V 10K 1N6263 ADULT = /7MS NEONATAL = /4MS
C121 + 12 + 1K D48 R202 3.24M C79 +5V T2
100UF 14 1N5819 100 OHM C81
R216 C82
25V 13 -U39D 20K
C127 + R200 R198 .15UF MP
470UF .001UF
25V 10K 1K R157 D40 T1 70MS
C104 R189 A 680PF NPO MP 100K 1N6263
R207 + 100UF 6.04K +5V TP35 +5V
121 25V
5
A
OHM C115 + ISOL GND 730MA A A C98 A
100UF D52 TP37 U31 +5V R160 NOTES:
25V ICL8069CCZR R192
ICM7555 150K
8
4
16
C101
V+
VD
R209 499 7 C99 RC Q RC
T1,TP43,U54,Y1.
10K OHM DIS 150PF .22UF MP .47UF MP
9 - C 1 15
7
NPO
3
CX CX
8 C128 + 6 5 C U30A 3. NOT USED: JP2,JP3,U19,U37,U50.
R194
3
THR CV
C120 + 10 +U39C 470UF D 2N3903 2 TO SHT. 4 4 CD4538B 12
V-
10K
2
100UF R196 +T +T
R211 25V B 2N3903 2 5 11 +5V
25V Q37 Q36E B 16uSEC_SYNC D
-T
7 9 -T
D Q34 IVY BIOMEDICAL SYSTEMS INC.
MC34004P 1K IRF522 1G Q35 3.32K Q Q
VS
1
R210
R
D51 C102 C103 + E 11 BUSINESS PARK DRIVE
1
2
R212 S R188 .001UF 22UF R193 B 2N3903 BRANFORD, CT. 06405
3
13
330 OHM MP TANT (203) 481-4183
8
3
2
24
+5V MK48T02-12
U51
26
1
TO SHT. 3 71055C-10 A0 8 A0 17 D7
VCC
D0 A1 DQ7 D6
4 PA0 34 7 A1 16
VCC
L0 3 PA1 D0
33 D1 A2 6 A2 DQ6
15 D5
RN2 L1 2 PA2 D1
32 D2 A3 5 A3 DQ5
14 D4
1K G1 1 D2
31 D3 A4 4 A4 DQ4
13 D3
D FRONT G2 40 PA3 D3
30 D4 A5 3 DQ3
11 D2 D
PANEL CHASSIS_GND G4 39 PA4 D4
29 D5 A6 2 A5 DQ2
10 D1 +5V
NOTCH- 38 PA5 D5
1 A6 DQ1
10
J203 28 D6 C136 A7 9 D0
2
3
4
5
6
7
8
9
14 37 PA6
PA7
D6
D7 27 D7
+5V
A8 23 A7 DQ0
13 D A9 22 A8
12
PB1 18 PB0 RD 5 RD- .1UF D A10 19 A9
11
PB2 19 PB1 WR 36 WRT-
CER 21 A10 C105 C106 C135 C111 C57 C110 C109 C108
10
PB3 20 PB2 A0 9 A0 20 W-
G-
.1UF .1UF .1UF .1UF .1UF .1UF .1UF .1UF
PB4 21 8 A1 18 E- CER CER CER CER CER CER CER CER
GND
9 PB3 A1
8
ALARM_ON/OFF 22 PB4 RESET 35
2_MIN_PRCDR 23 6 CS55
7 PB5 CS
TEST-
28
6 24 PB6
12
5
MENU 25 PB7 D
GRAPH ____ D0 11 10 A0
VCC
4 MOVE D1 D0 A0 A1
3 14 PC0 12 D1 A1 9 D
15 D2 13 8 A2
2 PC1 D3 D2 A2 A3
1 16 PC2 15 D3 A3 7
TO SHT. 3 17 D4 16 6 A4 CST02
PC3 D5 D4 A4 A5
182001 60/50 13 PC4 17 D5 A5 5
BEEP 12 D6 18 4 A6
HONK PC5 D7 D6 A6 A7
11 19 3
GND
VRTD PC6 D7 A7 A8
10 PC7 A8 25
24 A9
R269 A9 A10
3.32K A10 21
23 A11
7
A11 A12
A12 2
D 26 A13
A13 A14
A14 1
TO SHT. 2
TO SHT. 3 27 WRT-
WE
22 RD- MOVE-
+5V LEADS_OFF OE
20 CSRAM
GND
CS1
TO SHT. 2
RN1
1
10K R_ENABLE-
14
U52
HM62256LP-12
D
ECG_AUDIO
10
+12VREC
2
3
4
5
6
7
8
9
3
ALERT_AUDIO
+12VREC TO SHT. 1
2 Q33
TO SHT. 1 RDBK0 2N3903 HONK
RECSYNC T500
/RDRDY
1
C100 + /WRTRDY D BEEP
T150-
C
10000UF 16V C
R94 R93
AGND 10K 10K
TO SHT. 1 D[0..7]
D
C107
1 26 D0
2 1A 1B
27 D1 +5V TO SHT. 1
3 2A 2B
28 D2 R190 TO SHT. 2
3A 3B 49.9K .1UF D TRGOUT
D3 U35
13
+5V 4 4A 4B 29 CER H_SYNC-
5 30 D4 Z84C1008
5A 5B D5 D0 A6
6 31 39 A0 7
16US
VCC
6A 6B D6 D1 D0 A7
CSROM
CSRAM
CST02
CS54
CS55
RECRD
RECWR
BEEP
HONK
VRTD
RDBK0
/WRTRDY
/RDRDY
RECSYNC
7 7A 7B 32 38 D1 A1 5 TP32
8 33 D7 D2 37 A8 TO SHT. 3
8A 8B D3 D2 A2 4 A9
D 9 9A 9B 34 36 D3 A3 3 +5V 16uSEC_SYNC
10 35 RECWR D4 35 A0
10A 10B /WRTRDY D5 D4 A4 2 A1
11 11A 11B 36 33 D5 A5 1
12 37 D6 32 A2
12A 12B RECRES D7 D6 A6 44 A3 C134
13 13A 13B 38 31 D7 A7 43 .1UF CER
14 39 RECRD A4
14A 14B /RDRDY BREQ- A8 26 A5
28
31
11
40
41
39
37
36
33
30
23
14
24
76
80
25
27
84
35
15 40 17 A9 25 D D
5
15A 15B RECSYNC BACK- BUSREQ
16 16A 16B 41 16 BAI A10 24
EVENT
17 42 RDBK0 15 A11 D0 A0 TP28
A11 23 11 10 7
CSROM
CSRAM
VCC
VRTD
/RDY
RECRD
RECWR
/TMOT
BEEP
HONK
/WRTRDY
/RDRDY
RECSYNC
CST02
CS54
CS55
RDBK0
RDBK1
PIN001
PIN002
17A 17B A0 BAO A12 D1 O0 A0 A1 GND
18 18A 18B 43 A12 22 12 O1 A1 9 GND 19 TO SHT. 3
19 44 A1 30 A13 D2 A2
19A 19B A2 M1 A13 21 A14 D3
13 O2 A2 8
A3 GND 32 EVENT
20 20A 20B 45 11 IORQ A14 20 15 O3 A3 7 GND 42
21 46 A3 MREQ- 14 A15 D4 A4
21A 21B A4 RD- MREQ A15 19 D5
16 O4 A4 6
A5
8 LDOFF GND 47
22 22A 22B 47 10 RD 17 O5 A5 5 61 /PRINT GND 59 T2
23 48 WRT- 9 27 RDY D6 18 4 A6 A15 2 72
23A 23B WR RDY D7 O6 A6 A7 A14 A15 GND
24 24A 24B 49 CE/WAIT 18 19 O7 A7 3 69 A14 GND 82 70MS
25 50 41 25 A8 A13 79
25A 50B INT/PULSE A8 A9 A12 A13
42 IEI A9 24 74 A12 /4MS 45 SAMPLE TO SHT. 3 TO SHT. 1
40 21 A10 A11 75 56 TP26 J201-7
J205 12 IEO A10
23 A11 A2 63 A11 70MS
22
GND
RECORDER INTERFACE CLK 8 /RES
U49 A11
2 A12 A1 6 A2 SAMPLE SAMPLE TO SHT. 2
202002 CLK A12 A13 A0 A1
(330043) A13 26 73 A0 V_SYNC-
D1 50 D1
34
2366-01-04 20 CSROM D0 44 Z
SOCKET R221 A10 CE
22 RD- D0
TP16
INSTALL WITH 44 PIN FOR 101R OE
27 29
U36
GRG_GUU J201-2
GND
10K PGM /WRT
2367-01-04 1 1 18
RECORDER OPTION D
FOR 101NR
VPP +5V 70
10
/RD
/MREQ
2452-00-04
(330093)
HONKOT
BEEPOT 20 Z
/BACK
14
SOCKET FLD21 46
____ 55 TO SHT. 2
28 PIN INTR 48 8MHZ
71
D R270 22 /HSYNC Z_INT
TP41 VCC 3
VCC 13 +5V +12V
B 150 OHM 49 26 B
INTRPT VCC
21 NMI VCC 38 R71
28 43 JP1 TO SHT. 1
4.99K
7
EVENT VCC C96
53 1 STD J201-20
REINTEN
VCC .1UF
/UNLFIF
/VSYNC
4 66 2 8
DTAD8
DTAD7
/RDFIF
/MOVE
TREND_INT
/ZOUT
60HZ VCC
CER
EVNT
57 /10MS VCC 78 D30 6
FD7
FD6
FD5
FD4
FD3
FD2
FD1
FD0
34 D 3 U15 5
25
ADCS TM
A 1N4148 LM318
58
83
12
81
62
54
77
15
16
17
64
68
65
52
60
51
67
TO SHT. 1
4
SOCKET
3
84 PIN J201-1
R103 2 Q24
Z_GND
F7
F6
F5
F4
F3
F2
F1
F0
2N3903 -12V
+5V 4.99K D
1
R149 A
U48 C73 100 OHM
11
M1 A1
20
RESET A1 31 A2 R150 U29 MC34002P
13
TP34 19 MREQ A2 32
20 A3 D0 475 OHM MC1408-8N TO SHT. 2
A3 33 18 6 6
VCC
WRT- IORQ A4 D1 DB0 VI+ F7
22 A4 34 17 5 7
VCC
1
5
7
RD- 21 WR A5 D2 DB1 F6 A1
5 U22B D/A
RD A5 35 A6 D3
16 DB2 F5
6 A2 IO 4
U33 A6 36 15 7 C75 D38 7
VO
PFO
RS
A12 2 1
VB
GND
D GND
VEE
NMI- INT A14 RD VREF RC
17 NMI A14 4 3 WR R174 +5V +5V 16 CMP A
A15
A15 5 5 8
4
6
8
16
15
14
10
11
12
13
16
15
14
10
11
12
13
D 14 C97
3
2
D0 U21
10
BREQ- 25 15 D1
BACK- BUSRQ D1 D2 ADC0803
10
23 12 A
SO
SO
VCC
DOR
VCC
DOR
Q3
Q2
Q1
Q0
Q3
Q2
Q1
Q0
BUSAK D2 D3
D3 8 .01UF
12 9 CLK 6 7 D4 A -12V
PR
CLK PG D6 D7
13 74HCT40105
RST
RST
D7
DIR
DIR
SIN
SIN
8
OE
OE
CL
D3
D2
D1
D0
D3
D2
D1
D0
Q
29
14
U34B
13
9
1
2
3
7
6
5
4
9
1
2
3
7
6
5
4
74HCT74 8 R268 D
+5VDC
FO OUT
C74
150 OHM
D7
D6
D5
D4
D3
D2
D1
D0
A DELTRG .1UF 28 9 D0 D D A
GND
D5 D6
18 CLK1 D6 3 L1,LD3,Q48,R274,RN2,RV19,
2N3903 2 R115 D 17 2 D7 T1,TP43,U54,Y1.
G1 D7
16 OUT1
TO SHT. 3
Q29 20K +5V 26 +8V +12V 3. NOT USED: JP2,JP3,U19,U37,U50.
21 RD
27 ECG
1
CLK2 WR A0
D 19 G2 A0 22
20 23 A1 IVY BIOMEDICAL SYSTEMS INC.
R152
1
OUT2 A1
6 14 24 CS54 9.09K 2 11 BUSINESS PARK DRIVE
CL
Q GND CS
1 BRANFORD, CT. 06405
7 PG CLK 3 D SOCKET 3 U22A (203) 481-4183
U32
5 2 N82C54-2 28 PIN MC34002P Title
PR
15K 6.8UF
U34A TANT Size Document Number Rev
4