Professional Documents
Culture Documents
Ansi C39 4
Ansi C39 4
Ansi C39 4
UDC 621.3.083.4
Automatic Nu1I-Balancing
Elect r ica I Measur ing Instr uments
Sponsor
Electrical Standards Board
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
7007273 OCIOLO43 bTB W
American Standard
Registered United Stoles Patent Ofice
Published by
AMERICAN STANDARDS ASSOCIATION
INCORPORATED
10 East 40th Street, New York, N. Y. 10016
Copyright i Y h h by .American Standards Association, Incorporated
Universal Decimal Classification 621.3.083.4
Printed in USA
EN3M266/9
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
m %O09273 0001044 534 m
Foreword
(This Foreword is not a part of .American Standard Specifications for Autometic Null-lhlancing Electrical Measuring Instruments, C:)'). i-
1965.1
This standard was originally prepared by a subcommittee of Sectional Committee C39 on Electrical Measuring
Instruments. The work began in 1952 and after considerable time and effort was issued as an American Standard in
1956.
In 1964 the subcommittee undertook the revision of the standard to bring it up to date and to add requirements
for miniature recorders. The definitions have also been brought up to date and coordinated with those of the Scientific
Apparatus Makers Association.
Suggestions for improvement gained through the use of this standard will be welcomed. They should be sent to the
American Standards Association, Incorporated, 10 East 40th Street, New York, N.Y. 10016.
~
The organizations which participated in this work and the names of their representatives, as listed at the time this
standard was submitted to the Sectional Committee for approval, are as follows:
F. X. Lamb, Chairman bl. A. Pisciotta, Secretary
Organization Represented Name of Representntive
American Society for Testing and .................................................................................................. Gordon Thompson
Defense Electronics Supply Center ..................................................................................... ..........Ross V. Doughty
Electric Light and Power Group .............................................................................................................. J. B. Leinbach
A. T. Higgins (.4ft/
General Services 4dministratio1-1,Federal Supply Service ................................................................................. .....J ohn F. Kirby
George R . Tucker /,4/t/
Institute of Electrical and Electronics Engineers ..............................................................................
C. Dale Owens
Dr M. B. Stout
M. L. Lush (.41t/
Instrument Society of Amerika .................................................................................................................... 4. Yï. Jacobson
National Bureau.of Standards, partment of Commerce ........... .................................................... F. K. Harris
National Electrical Manufacturers Association ..................................... .................................................... J. F. Connelly
J. M.Cunnien
P. .A. DePaolo
L. J. Lunas
G. Stolar
Scientific Apparatus Makers Association ................................................................................................................ G . L. Broomell
Telephone Group .................................................................................................. ............................... R. H. Van Horn
J . P. Messana / 4 1 t l
U.S. Department of the Navy, Bureau of Ships .....................................................................................
Subcommittee C39-4 on Automatic Null-Balancing Electrical Measuring Instruments, which is directly responsible
for the development of this standard, has the following personnel:
G. L. Broomell, Chairman M. A. Pisciotta, Secretary
E. A. Blomqvist L. J. Lunas
U'. H. Brand D. R . Hyer
4. W.Jacobson Robert McCarron
J. B. Leinbach R . S. Smith
C. A. Vogelsang
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
Contents
SECTION PAGE
1. Introduction .................................................................................................................................................... 7
2. Scope .............................................................................................................................................................. 7
3. Classification .................,.......................,,,..., ......, ,....,.........,.....,,.,.,..,.., .................. ............ ...... 7
3.1 Exhibiting Means ..................................... ....... ...... .................................... ........ 7
3.2 Recording Means ................................................................................................................................ 8
3.3 Number of Measured Variables ....................................................... ... I............................ 8
3.4 Scale Length ............. ....................................................................................................... 8
3.5 Type of Measuring Cir ............. ....,<..........................................
.......,........a... I 8
3.6 Range or Span .......................................................................... ...................................................... 8
3.7 External Circuit Resistance Rating ,......... ...... ...... .... ...__. ...... ...... ...... .......... .. ............................................ 8
3.8 Impedance of Measured Variable ......................................................... ..............,................................ 8
3.9 Power Supply for Measuring Circuit ........................................................................................................ 8
3.10 Reference Operating Conditions .......................,......................,............................. .... . 8
3.1 1 Rated Operating Conditions ....... 8
3.1 2 Performance Ratings ................................................................., , .........,,,.......... ........ ................... 8
8
8
3.15 Design Limits .... ...... ..... ................................................................. 8
4. Definitions ............................................ ................. 8
.......................................................................... <.......................... 8
......................... 9
9
4.4 Design Limits ,.....I................................................................................... ......................... 9
4.5 Detecting Means ................................. 9
4.6 . -End Device ................ ........................................................... <.<............................... 9
9
_.._.......... 9
....................................... 9
4.10 Position Influence. .......................................................................................... 9
4.1 1 Interference .................................................. . 9
4.12 Intermediate Means ................................................................................................... 9
....................................................... 9
............... 9
........................................................ 9
4.16 Operating Conditions ... ................................................................ ............. 10
10
................................................................................... 10
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
... 12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
m 9009273 O001046 307 m
SECTION PAGE
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
m 9009273 0001047 243 m
2. Scope
3. Classification
This section lists the characteristics by which the
2.1 This standard applies to indirect-acting electrical instruments included in this American Standard may be
measuring instruments of the automatic null-balancing
classified.
type. In these instruments the position of the end device 3.1 Exhibiting Means
is a function Of the measured when NOTE: Throuehout this standard. references to charts or indi-
is achieved. These instruments include intermediate cating scales ap$y only where such 'devices are provided. Instru-
means to control the power applied to the end device in ments classified under 3.1(3)and ( 4 ) have no charts, nor is it the
intent of the standard to imply that indicating scales are neces-
with the difference between the magnitude Of sarily provided in any instruments except those classified uiider
the measured signal and an opposing electrical quantity 3.1(31.
whose magnitude is a function of the position of the end ( 1 ) Recording-strip chart
'rvice. ( 2 ) Recording-round chart
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
7
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
9009273 O O O L 0 4 8 L 8 T W
c39.4
8 AMERICAN STANDARD SPECIFICATIONS FOR
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
~ ~~
~
~ ~~~
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
of the dead band, they denote the minimum change in measured
signal required to initiate response. Since this condition of balance interference which appears between any measuring
is not normally achieved and is not readily recognizable, these circuit terminal and ground.
terms are not used in this standard and the various requirements
are stated in terms of dead band. 4.11.2 Differential Mode Interference. (SeeNormal
Sensitivity is frequently used to denote the same quantity. How- Mode Interference.)
ever, its usage in this sense is deprecated since it is riot in accord 4.11.3 Longitudinal Interference. (See Common
with accepted standard definitions of the term.
Mode Interference.)
4.3.1 Dead Band Rating. The limit which the dead
4.11.4 Magnetic Field Interference. A form of in-
band will not exceed when the instrument is used under
terference induced in the circuits of a device due to the
rated operating conditions.
presence of a magnetic field.
4.4 Design Limits. Design aspects of the instrument in NOTE: It may appear as common mode or normal mode inter-
terms of certain limiting conditions to which the instru- ference in the measuring circuit.
ment may be subjected without permanent physical dam- 4.11.5 Normal Mode Interference. A form of inter-
age or iinpairment of operating characteristics. ference which appears between measuring circuit term-
4.5 Detecting Means. The first system element or group inals.
of elements that responds quantitatively to the measured 4.11.6 Transuerse Interference. (See Normal Mode
variable and performs the initial measurement operation. i nte rfe re nce .)
The detecting means performs the initial conversion or 4.12 Intermediate Means. Includes all system elements
control of measurement energy. or devices that are used to perform necessary and dis-
4.6 End Device. The final instrument element that tinct operations in the measurement sequence between
responds quantitatively to the measured signal and per- the primary detector and the end device. The inter-
forms the final measurement operation. An end device mediate means, where necessary, adapts the operational
performs the final conversion of measurement energy to results of the primary detector to the input requirements
an indication, record, or the initiation of control. of the end device.
4.7 Error. The difference between the indication and 4.13 Measured Signal. The electrical quantity applied
the true value of the measured signal. to the measuring circuit terminals of the instrument.
NOTE: .4 positive error denotes that the indication of the instru- NOTE: It is the electrical analog of the measured variable.
ment is greater than the true value.
Error = Indication - True 4.14 Measured Variable (Measurand).The physical
See Correction. quantity, property, or condition which is to be measured.
NOTE: Common measured variables are temperature, pressure,
4.8 Influence thickness, speed, etc.
4.8.1 Operating Influence. The change in a desig- 4.15 Null Balance. As applied to instruments within the
nated performance characteristic caused solely by a scope of this American Standard, null balance is the
prescribed change in a specified operating variable from condition which exists in the circuits of an instrument
its reference operating condition to its extreme operating when the difference between an opposing electrical quan-
condition, all other operating variables being held within tity within the instrument and the measured signal does
the limits of reference operating conditions. not exceed the dead band. The value of the opposing
NOTE 1: It is usually expressed as a percentage of span.
NOTE 2: If the magnitude of the influence is affected by direc- electrical quantity produced within the instrument is
tion, polarity, or phase, the greater value shall be taken. related to the position of the end device.
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
m 9009273 0001050 838 m
c39.4
10 AMERICAN STANDARD SPECIFICATIONS FOR
4.16 Operating Conditions. Conditions (such as am- etc, is less than the lower range-value. Zero does not
bient temperature, ambient pressure, vibration, etc) to appear on the scale.
which a device is subjected, but not including the vari- NOTE: For example: 20 to 100.
able measured by the device. 4.18.6 Suppression Ratio (of a Suppressed-Zen,
4.16.1 Extreme Operating Conditions. The range Range). The ratio of the lower range-value to the span.
of operating conditions within which a device is designed NOTE: For example: Range 20 to 100
to operate and under which operating influences are Suppression Ratio =-
20
80
= 0.25
usually stated.
4.16.2 Rated Operating Conditions. The limits of 4.19 Repeatability. The closeness of agreement among
specified variables or conditions within which the per- a number of consecutive measurements by the instrument
formance ratings apply. for the same value of the measured signal under the same
' 4.16.3 Reference Operating Conditions. The con- operating conditions, approaching from the same direc-
ditions under which reference performance is stated and t ion.
the base from which the values of operating influences NOTE: It is usually expressed as a maximum nonrepeatability
are determined. in percent of span.
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
9009273 O O O L 0 5 L 774
c39.4
AUTOMATIC NULLBALANCING ELECTRICAL MEASURING INSTRUMENTS 11
4.21.3 Ramp Response. A criterion of the dynamic 4.24 Time per Point ( Multiple-Point Recorders).
response of an instrument when subjected to a measured The time interval between successive points on printed
signal which varies at a constant rate. records.
4.21.4 Ramp Response Time. The time lag, ex- NOTE: For some instruments this intirval is variable and de-
pends on the magnitude of change in measured signal. For such
pressed in seconds, between the measured signal and the instruments, time per point is specified as the minimum and maxi-
equivalent positioning of the end device when the meas- mum time intervals.
ured signal is varying at constant rate.
4.25 Transient Overshoot. An excursion beyond the
4.21.5 Ramp Response-Time Rating. The maxi-
final steady-state value of output as the result of a step-
mum ramp response time for all rates of change of meas-
input change..
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
ured signal not exceeding the average velocity corres- NOTE: It is usually referred to as the first excursion; expressed
ponding to the span step-response-time rating of the as a percent of the steady-state output step.
instrument when the instrument is used under rated
4.26 Transportation and Storage Conditions. The
operating conditions. Example: If the span step-response-
conditions to which a device may be subjected between
time rating is 4 seconds, the ramp response-time rating
the time of construction and the time of installation.
shall apply to any rate of change of measured signal not
Also included are the conditions that may exist during
exceeding 25 percent of span per second.
shutdown.
4.21.6 Step Response. A criterion. of the dynamic NOTE: No permanent physical damage or impairment of operat-
response of an instrument when subjected to an instan- ing characteristics shall take place under these conditions, but
taneous change in measured quantity from one value to minor adjustments may be needed to restore performance to normal.
another. 4.27 Travel
4.21.7 Step-Response Time. The time required for 4.27.1 hdiC8tOr Travel. The length of the path
the end device to come to rest in its new position after an described by the indicating means or the tip of the pointer
abrupt change to a new constant value has occurred in in moving from one end of the scale to the other.
the measured signal. NOTE 1: The path may be an arc or a straight line.
4.21.8 Span Step-Response-Time Rating. The lim- NOTE 2: In the case of knife-edge pbinters and others extending
beyond the scale division marks, the pointer shall be considered
it which the step-response time will not exceed for a as ending at the outer end of the shortest scale division marks.
change in measured signal essentially equivalent to full 4.27.2 Pen Travel. The length of the path described
span when the instrument is used under rated operating by the pen in moving from one end of the chart scale to
conditions. The actual span step-response time shall not the other. The path may be an arc or a straight line.
be less .than 2/3 of the span step-response-time rating.
(For example, for an instrument of 3-second span step-
responsc-time rating, the span step-response time, under
rated operating conditions, will be between 3 and 2
5. General Requirements
seconds.)
The requirements in this section apply to the design
I t must be recognized that the step-response time for
and construction of all instruments w'ithin the scope of
smaller steps is not proportional t o the step-response
this American Standard. Additional requirements or
time for fuZ1 span.
NOTE: The end device shall be considered to be at rest when it exceptions for design and performance are covered by
remains within a band of plus and minus the accuracy rating from the Detailed Requirement Sheets.
its final position.
5.1 Construction. The construction of instruments shall
4.22 Source Resistance. The resistance presented to be mechanically sound, suitable to their purpose, and
the input of a device by the source. shall be such as to give assurance of satisfactory per-
4.22.1 Source Resistance Rating. The value of formance. All materials shall be suitable for the purpose
source resistance which, when injected in an external for which they are used. Workmanship shall be such as
circuit having essentially zero resistance, will either to provide a neat appearance commensurate with the
( 1 ) double the dead band, or ( 2 ) shift the dead band by service for which intended.
one-half its width.
5.2 Safety Requirements. The applicable provision
4.23 Span. The algebraic difference between the upper of American Standard Electrical Safety Requirements for
and lower range-values. Electrical Measuring and Controlling Instrumentation,
NOTE 1: For example: C39.5-1964, shall be applied to the instruments covered
( a ) Range O to 150, Span 150
( b ) Range -20 to 200, Span 220 by this specification.
( c ) Range 20 to 150, Span 130
( d ) Range -100 to -20, Span 80
5.3 Enclosure. Instruments shall be designed and
NOTE 2: The following compound terms are used with suitable constructed with close-fitting joints in the enclosure to
modifications in the units: Measured Variable Span, Measured minimize the entrance of dust and moisture that might
Signal Span, etc.
NOTE 3: For multirange devices, this definition applies to the impair operation over an extended period of time or when
particular range that the devicr is set to measure. subjected to any of the tests specified herein.
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
W 7009273 0001052 600
c39.4
12 AMERICAN STANDARD SPECIFICATIONS FOR
5.4 Protective Finishes. Exposure to the atmosphere 5.9.2 General Date. The following information shall
and conditions which can be reasonably expected in be legibly marked on the instrument and accessible from
applications for which the instrument is intended shall the front, but need not be visible with the cover closed:
not cause rust or corrosion which affect the performance. ( 1 ) Manufacturer's name, trademark, or symbol
Protective finishes shall be used where necessary to ( 2 ) Manufacturer's type or model number or both
assure this requirement. (3) Manufacturer's serial number
( 4 ) Range of measured variable
5.5 SC*W Threads. All threaded parts normally em-
( 5 ) Range of measured signal, if feasible
ployed in the application or normal use of instruments
( 6 ) Instruction book number
shall comply with American Standard:
(7) Nominal power supply, including voltage, fre-
Unified Screw Threads, Bl.1-1960
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
quency, and' maximum load in volt-amperes
Screw Thread Cages and Gaging, B1.2-1951
( 8 Primary detector and external intermediate device,
Square and Hexagon Bolts and Nuts, B18.2-1960
if required, where same is not evident from other stated
Slotted and Recessed Head Wood Screws, B18.6.1-
data
1961 ( 9 ) Response-time classification: Span step response-
Hexagon Head Cap Screws, Slotted Head Cap Screws,
time rating for single-point instruments or time per point
Square Head Set Screws, and Slotted Headless Set Screws,
for multiple-point instruments
B18.6.2-1956
(10) Chart number
Slotted and Recessed Head Machine Screws and Ma-
( 1 1 ) Chart speed
chine Screw Nuts, B18.6.3-1962
(12) Identification and, if not self-evident, the location
Slotted and Recessed Head Tapping Screws and Metal-
of parts subject to replacement in normal maintenance
lic Drive Screws, B18.6.4-1958
Where practicable, all other screw threads shall also 5.10 Chart Drive. The chart shall be driven by a
comply with these standards. synchronous motor.
5.6 Interchangeability of Parts and Subassemblies. 5.11 Terminals
Parts and subassemblies required for general maintenance 5.11.1 Mounting of Terminals. The terminals,
shall be interchangeable in instruments of the same studs, or binding-post bodies shall be firmly anchored
type and model. to the instrument in such a manner as to prevent turning
or movement sufficient to cause damage to internal con-
5.7 Scale and Chart Visibility. The instrument shall
struct ion.
be so designed that the scale and the chart may be easily
5.11.2 Marking of Terminals. The terminals shall
read while operating normally with cover closed. On
be legibly marked and information shall be provided to
strip-chart instruments, a portion of the most recent
identify the proper connections.
record shall be visible with the cover closed.
5.12 Potentiometer Current Standardization
5.8 Indicating Scales. The legibility of indicating
5.12.1 Regulated Supply. For potentiometer circuits
scales shall be consistent with the accuracy rating. The
where the source of potentiometer current is a regulated
style and size of the scale numerals and the maximum
voltage supply, means shall be provided to permit check-
number of digits shall be such as to provide legibility at
ing the supply with an external standard.
a distance consistent with intended applications.
5.12.2 Periodic Adjustment. For potentiometer
5.8.1 Scale and Chart Divisions. The value of each
circuits where the source of potentiometer current is such
scale and chart division, wherever practical, shall be
that periodic adjustment by comparison with a standard
either 1, 2, or 5 units or an integral power of 10 times
is required in order to meet performance ratings, means
these values. The spacing between divisions on the chart
for accomplishing such adjustment shall be automatic.
preferably shall be between 0.04 and 0.1'25 inch.
When the adjustment operation occurs periodically,
means shall also be pro;ided to adjust manually the
5.9 Data Marking. Instruments shall be legibly marked, current at any time. If the period is one day or more,
with the applicable details specified below. If made on nonautomatic adjustment is acceptable for round-chart
the indicating scale plate, these markings shall not be instruments.
so conspicuous as to distract attention from the scale 5.12.3 CircuitSupply-Condition Indicator. When
markings specified in 5.8.1. the bridge or potentiometer is energized by a source
5.9.1 Scale Data. The following infoimation shall be subject to deterioration, means shall be provided to
visible from the front of the instrument without opening indicate when the source is no longer adequate.
the cover and on or associated with the indicating scale
if one is provided: 5.13 Thermocouple Reference-Junction Compensa-
( i ) Designation of the unit of measurement such as tion. When the purpose of the instrument is to measure
millivolts, degrees C, pH, ohms, etc temperature by means of a thermocouple, compensation
( 2 ) Multiplying factor, if required for the reference-junction temperature shall be automatic.
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
~
tubes and semiconductors, when used, shall be com- indicated on the indicating scale where provided.
mercially available types.
5.15 Gain Adjustment. In instruments where the per- 7. Reference Standards
formance is affected by the amount of amplification pro-
vided in the intermediate means, provision shall be made 7.1 Accuracy. Reference standards in which the com-
for adjusting the gain of the intermediate means. bined uncertainties are less than the accuracy ratings
of the instrum,ent under test, by a factor of at least 3,
5.16 Time per Point. With a change in measured sig-
shall be used in testing the accuracy of instruments.
nal between successive points equivalent to the span, the
instrument shall be capable of balancing and printing 7.2 Verification. Standards shall be tested periodically
within its performance ratings. and certified in terms of national standards. These tests
can be made in any laboratory capable of substantiating
5.17 Standard Accessories. The following accessories
measurements of the accuracies required.
shall be supplied with the instrument where applicable:
.~
Box of 100 charts or one chart roll
Mounting hardware 8. Test Procedures
Ink and pen filler, or ink pads
Lubricating oil The procedures in this section shall be used to de-
Pen cleaner termine compliance with the detailed performance rat-
Battery ings, operating influences, and design limits. The instru-
Instructions for installation and operation ment shall be set up and adjusted in accordance with
Special tools required for adjustment the manufacturer's standard instructions. Unless other-
Available Accessories. The following material wise stated in this section, test conditions shall be main-
shall be available from the manufacturer: tained in accordance with Section 6.
( 1) Internal wiring diagram and circuit-component
Values for rated and extreme operating conditions,
values performance ratings, operating influences, and design
( 2 ) Maintenance manual and replacement parts list limits shall be determined by reference to the applicable
Detailed Requirement Sheets.
8.1 Tests for Reference Performance. Reference per-
6. Test Conditions formance shall be determined under reference operating
conditions as listed in Table 1.
When conducting a test to determine compliance with 8.1.1 Connecting Means. Where instrument per-
a detailed requirement, all conditions shall be main- formance and accuracy are affected by the characteristics
tained as stated in Table 1, unless otherwise specified by of the connecting means, the connecting means recom-
the applicable test procedure of Section 8. If additional mended by the manufazturer shall be used.
conditions are specified in the manufacturer's standard 8.1.2 Test Procedures. The reference standards
instructions, they shall also be maintained. shall be so located that they are not subjected to tem-
6.1 Chart and Ink. The chart and ink supplied by the perature gradients resulting from rapidly varying temper-
manufacturer shall be used. atures, drafts, or localized sources of heat.
8.1.2.1 Potentiometer Instruments for Measurement
6.2 Potentiometer Current Adjustment. For poten-
of emf (Electromotive Force) -
tiometer circuits, the potentiometer current shall be ( 1 ) Use copper wires for connecting the reference
adjusted when required, in accordance with the manu- potentiometer to the instrument.
facturer's instructions, immediately prior to performing ( 2 ) Adjust the current of the reference potentiometer
tests. to the standard value.
( 3 ) Short-circuit or clamp the coil of the galvanom-
6.3 Gain Adjustment. The gain of the intermediate
eter in the reference potentiometer so that motion of the
device shall be adjusted for minimum dead band in ac-
galvanometer will not introduce spurious voltages.
cordance with manufacturer's instructions.
( 4 ) Adjust the potentiometer circuit current of the
6.4 Enclosure. The cover of the case shall be closed instrument where required in accordance with manu-
and other openings in the case shall be closed to the facturer's instructi,ons.
degree indicated by 5.3. ( 5 ) Set the reference potentiometer to the emf value
at which it is desired to check the instrument accuracy.
6.5 Warm-up Period. The instrument shall be ener-
Approach the balance point from each direction. Use
gized at least 30 minutes before starting test.
indicating scale, where provided, for observing instru-
6.6 Use of Indicating Scale. Observation for deter- ment reading.
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
c39.4
14 AMERICAN STANDARD SPECIFICATIONS FOR
I I I I I I I I I I I I 2
E
.d
O
P-
O
O In Y
Y
Q\ I I I I I I I I I 0 I Y
m
-
O
o
7 a
Y
O
m o
Y
O
QI
O
QI
O 0 0
o
m d
o
2.
m
o
~
O
3
O
m
In
I -m
U
m
I ü
."m
-u
O O U x
d rn YI
.
>
O m
O
Y $12 0 O
I 2M
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
e4
m
U 2 QI
m ci
O
-
+I
m
In
O
CiI
x
+I O 0 0
5: er) O
I
m
c4 0
U
m 2 5 2 ._
2 m
o
m
.w
Y
S
W
o 8 %%% Y o> E
3
U
Li
op
mLi
8 U
5 bL
E
L
-X
E
-
e
m
EO
.-x
U
S
CI
c
o X
m
O
5O
-n 1 -E
c)
u
2
3 E
Y
2u
c
o a- o ."mOY
&
.-m 5 m u u O m
I .-
o o71 E E
-
Li
a m YI
. -Y
>* o E E Z I
-
Y
I m O u
M CI E S E S
E o 0 - " .-S
.I
E
E CI u
Z E E f 2 .-uO
.I
c
2o o
1-
n
E
o
.I
ci
* E E L
" O 0 0
o
2
E
.-
O
.-
Y -m
5 uuz
oa d
E E E
U
I
e
n a
c
O
m
> I
1
E
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
’7009273 QOOL055 3LT
c39.1
AUTOMATIC NULLBALANCING ELECTRICAL MEASURING INSTRUMENTS 15
8.1.2.2 Potentiometer Instruments for Measurement instrument where required in accordance with the man-
of Temperature with Thermocouples ufacturer’s instructions.
( i ) Read the temperature indicated by the mer-
( i ) Method 1 (see Fig. 1 ) cury thermometer and obtain the emf equivalent of this,
( a ) Use a melting crushed-ice bath for maintain- temperature from a table of emf-temperature equivalents’
ing the temperature of the test reference junction. for the type of thermocouple used.
(b) Use thermocouple extension wires of appro- ( j ) Obtain the emf equivalent of the temperature
priate type for connecting the instrument to the test at which it is desired to check the instrument accuracy.
reference junction. The thermoelectrically positive and ( k ) Set .the reference potentiometer to the value
negative wires must be connected to the respective posi- which is the algebraic difference between the emf equiv-
tive and negative terminals of the instrument. alents of the checking temperature and the.test reference-
( c ) Join a copper wire to each of the extension junction temperature. Approach the balance point from
wires at the test reference junction. Insulate the junction each direction. Use indicating scale, where provided, for
with a suitable material (e.g., paraffin) and place in good observing instrument reading.
thermal contact with the ice bath but electrically in- Example: It is desired to check an instrument having
sulated therefrom. Allow sufficient time for junctions to a range of O-1,000 F with an iron constantan thermo-
reach temperature of ice bath. couple at 100 F, 500 F, and 900 F. The temperature of
( d ) Connect the copper wires to the reference the test reference junction is 90 F. The emf equivalents
potentiometer, observing proper polarity. (reference temperature 32 F) of these temperatures are:
( e ) Adjust the current of the reference poten-
tiometer to the standard value. Temperature, emf,
Degrees F Millivolts
( f ) Short-circuit or clamp the coil of the gal-
vanometer of the reference potentiometer. 90 1.65
( g ) Adjust the Potentiometer circuit current of 1O0 1.94
the instrument where required in accordance with the 500 14.12
manufacturer’s instructions. 900 26.40
( h ) Set the reference potentiometer to the emf
equivalent of the temperature at which it is desired to Set the reference potentiometer as follows:
check the instrument accuracy. [Use table of emf-temper- Reference
ature with reference junction of O C (32 F ) for the type Check Junction Reference
of thermocouple used.] Approach the balance point from Temperature, Temperature, Potentiometer
each direction. Use indicating scale, where provided, for Temperature, Equivalent Equivalent Setting,
observiig instrument reading. Degrees F Millivolts Millivolts Millivolts
( 2 ) Method 2 (see Fig. 2) 1O0 1.94 1.65 0.29
( a ) Use a metal block with three holes drilled in 500 14.12 1.65 12.47
proximity to each other and near the center of the block 900 26.40 1.65 24.75
to provide a uniform temperature for the test reference
junction. 8.1.2.3 Resistance Bridges
( b ) Use thermocouple extension wires of appro- ( i ) Use copper wires for connecting the reference
priate type for connecting the instrument to the test resistance box to the instrument in accordance with
reference junction. The thermoelectrically positive and the manufacturer’s wiring instructions.
negative wires must be connected to the respective posi- ( 2 ) Obtain the resistance equivalent of the desired
tive and negative terminals of the instrument. check point from the specified table.
( c ) Join a copper wire to each of the extension (3) Set the resistance box to the resistance equivalent
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
wires at the test reference junction. Electrically insulate of the check point. Approach the balance point from
each junction and insert it into a hole in the metal block. each direction. Use indicating scale, where provided, for
( d ) Connect the copper wires to the reference obtaining instrument reading.
potentiometer, observing proper polarity. 8.1.2.4 Other Circuits. Use manufacturer’s recom-
( e ) Insert a calibrated mercury thermometer in mendation for testing for accuracy.
the third hole of the metal block and allow sufficient 8.1.3 Dead Band
time for the thermometer to reach the temperature of the 8.1.3.1 Method 1
block. NOTE: In employing this method it is essential that a reference
standard be used which can be reproducibly set to any fixed value
( f ) Adjust the current of the reference poten- with an error significantly less than the dead band rating of the
tiometer to the standard value. instrument under test.
( g ) Short-circuit or clamp the coil of the gal- ( 1 I Connect the reference equipment to the instrument
vanometer of the reference potentiometer. and set it at a value corresponding approximately to
( h ) Adjust the potentiometer circuit current of the midspan of the instrument.
LI. -.
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
~~
~~
- ~ ~
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
Fig. 1
Potentiometer Instrument fur Measurement of Temperature with Theriiiocouples
Method 1 - Crushed-Ice Bath
( 2 ) Set the gain adjustment of the instrument to pro- 8.1.3.1(4)and 8.1.3.1(6),expressed in percent of span,
vide minimum dead band in accordance with manu- is the dead band.
facturer’s instructions.
( 3 ) Change the setting of the reference equipment 8.1.3.2 Method 2
NOTE: This method is applicable to instruments in which the
until the instrument indication is greater than the orig- balancing means can be readily displaced manually and without
inal value by an amount equal to approximately twice damage.
the accuracy rating. ( 1 ) Connect the reference equipment to the instrument
( 4 ) Slowly reset the reference equipment to the orig- and set it at a value corresponding approximately to mid-
inal value [8.1.3.1(1 I]. Observe instrument reading. span of the instrument.
( 5 ) Change the setting of the reference equipment ( 2 ) Set the gain adjustment of the instrument to pro-
until the instrument indication is less than the original vide minimum dead band in accordance with manufac-
value by an amount equal to approximately twice the turer’s instructions.
accuracy rating. (3) Manually displace the balancing motor shaft or
( 6 ) Slowly reset the reference equipment to the orig- associated gearing until the instrument indication is
inal value [8.1.3.1(1 I].Observe instrument reading. greater than the original value by an amount equal to
17) T h e difference in t h e readings obtained in approximately twice the accuracy rating/”
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
W 9009273 O O O L 0 5 7 L92 W
c39.4
AUTOMATIC NULLBALANCING ELECTRICAL MEASURING INSTRUMENTS 17
INSTRUMENT
THERMOCOUPLE
EXTENSION WIRES
THERMOMETER
COPPER WIRES
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
REFERENCE
POTENTIOMETE
Fig. 2
Potentiometer Instrument for Measurement of Temperature with Thermocouples
Method 2 - Metal Block
(4) Allow the system to return to balance slowly so change in measured signal equivalent to 99 percent of
that the edge of the dead band may be determined. span is applied and the instrument allowed to balance
(5) Manually displace the balancing motor shaft or as defined in Section 4. Test is repeated for opposite
associated gearing until the instrument indication is direction of balancing action.
less than the original value by an amount equal to ap- Method 2-The abrupt change in measured signal
proximately twice the accuracy rating. is applied for a precisely timed interval. Length of in-
( 6 ) Allow the system to return to balance slowly so terval is varied in small steps, and length of interval
that the edge of the dead band may be determined. during which balance, as defined in Section 4 is achieved,
( 7 ) T h e difference in t h e readings obtained in represents response time for 99 percent of span.
8.1.3.2(4)and 8.1.3.2(6),expressed in percent of span, 8.1.4.2 Span Frequency Response. Apply a sinus-
is the dead band. oidally varying measured signal having a double ampli-
8.1.4 Dynamic Response or Time per Point tude (peak-to-peak value) equivalent to approximately
8.1.4.1 Span Step-Response Time 99 percent of span at a frequency which will allow the
Method 1 -Chart speed is increased sufficiently to instrument to come to rest momentarily at the peak
allow an accurate time analysis of record. An abrupt values. The instrument is considered to follow chan
/-
--
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
9009273 0001058 O29
c39.4
18 AMERICAN STANDARD SPECIFICATIONSFOR
at higher frequencies if the indicated or recorded double Requirement Sheets. The instrument shall be maintained
amplitude does not differ from the initially determined at that temperature for a period of 24 hours and shall
double amplitude by an amount exceeding 10 times the be energized from the operating power supply during
accuracy rating. that time. At the conclusion of the 24-hour period, and
8-1-43Ramp Response Time. To determine the while still at the low temperature, the following períorm-
ramp response time of an instrument, a technique for ance characteristics shall be measured and compared
the particular instrument must be devised. Because of with the data obtained prior to the test: accuracy, dead
the wide variation in instrument designs, no standard band, step-response time, or time per point.
test procedure for ramp response time is included in this 8.3.2 Operating-Powersupply Influence
standard. Refer to the manufacturer for the method to 8.3.2.1 Operating-Power-Supply Voltage Influence.
be employed for particular instruments. The operating-power-supply voltage shall be varied
8.1.4.4 Time per Point (Multiple-Point Instru- slowly over the range stated for extreme operating con-
ments). To determine conformance with 5.16, values of ditions in the Detailed Requirement Sheets. Throughout
measured signal equivalent as nearly as practicable to the test the performance characteristics shall be meas-
the upper and lower range-values are applied alternately ured and compared with the data obtained prior to the
to successive measurement points. The instrument shall test.
balance and indicate or record each value of measured 8.3.2.2 Operating-Power-Supply Transient-Voltage
signal to within the accuracy rating. Influence. The operating-power-supply voltage shall be
8.1.5 Maximum Transient Overshoot. The dead changed abruptly in either direction by the amount
band of the instrument is adjusted to within the dead stated for extreme operating conditions in the Detailed
band rating and the instrument is caused to balance Requirement Sheets and the transient shift in instrument
near one end of the span. An abrupt change in measured indication shall be measured.
signal equivalent to approximately 90 percent of span is 8.3.2.3 Opera t ing-Po we r-S u p p l y F r e q u e n c y In-
applied to the instrument. The amount of overshoot fluence. The operating-power-supply frequency shall be
beyond the point of final balance expressed in percent varied over the range stated for extreme operating con-
of span is the transient overshoot. Test is repeated for ditions in the Detailed Requirement Sheets. Throughout
opposite direction of balancing action. the test, the performance characteristics shall be meas-
ured and compared with the data obtained prior to the
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
o#mnNc
POWER SuPP'V{ I I
Fig. 3
Test for Normal Mode Interference Influence for Instruments with dc Measui --ig Circuits
the phase angle shall be varied through 360 degrees. meter shall preferably be of the battesr-operated elec-
Throughout the test the error (accuracy) and dead band tronic type.
shall be measured and compared with the data obtained The rms magnitude of the common mode interference
prior to the introduction of the normal mode interference. shall be adjusted to the value stated for extreme operat-
8.3.3.2 Common Mode interference Influence. ing conditions in the Detailed Requirement Sheets, and
IRefer to Fig. 3 . ) With the instrument balanced at ap- the phase angle shall be varied through 360 degrees.
proximately midspan, common mode interference shall Throughout the test the error (accuracy) and dead band
be artificially introduced by applying a voltage of operat- shall be measured and compared with the data obtained
ing-power-supply frequency between each measuring prior to the introduction of the common mode inter-
circuit terminal of the instrument and the instrument ference.
case. This voltage shall be adjustable in magnitude and 8.3.3.3 Magnetic-Field Influence. With the instru-
means shall be provided for continuously shifting its ment balanced at midspan, it shall be subjected to an
phase angle. The rms magnitude of this voltage shall be external magnetic field derived from a Helmholtz coil
measured by a voltmeter connected between instrument set energized at operating-power-supply frequency. The
case and the measuring circuit terminal, and this volt- coils shall be 40 inches in diameter, not over 5 inch$$,.
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
c39.4
20 AMERICAN STANDARD SPECIFICATIONS FOR
MEASURING
-
A A (
O
STABLE
Dc
SOURCE
AC VACUUM-TUBL
VOLTMETER
- --``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
Fig. 4
Test for Common Mode Interference Influence for I n s t r u m e n t s with dc Measuring Circuits
long, and spaced 20 inches apart, and shall carry suf- major axes. Throughout the test the error (accuracy)
ficient current to produce the required field. (56.5 ampere and dead band shall be measured and compared with the
turns per coil will produce a field intensity of approxi- data obtained prior to the introduction of the external
mately 1 oersted.) The instrument under test shall be magnetic field.
placed in the center of the coil system. 8.3.4 Source Resistance. With the source resistance
The magnetic-field strength shall be adjusted to the not exceeding one-tenth the source resistance rating,
value stated for extreme operating conditions in the adjust the dead band of the instrument in accordance
Detailed Requirement Sheets and the phase angle shall with the manufacturer's instructions. Then insert re-
be varied through 360 degrees. The test shall be re- sistance in the external circuit as specified in the tabu-
peated with the coil oriented along each of the three lated Detailed Requirement Sheets under extreme oper-
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
~- ~- ~ ~ ~ ~
9009273 O O O L O b L b L 3
(39.4
4CTOM4TIC YULLB4L4YCIYG ELECTRICAL ME4S1lRING INSTRUMENIS 21
ating conditions. The values of error, dead band, span by means of a voltmeter coil placed in the testing trans-
step-response time, and transient overshoot p d e r these former or through an auxiliary potential transformer.
conditions of test shall not exceed the values for operat- To avoid voltage surges, the applied voltage shall be
ing influences shown in the tabulated Detailed Require- raised to its full value gradually and, upon completion
ment Sheets. of the 60-second test period, shall then gradually be
8.3.5 Reproducibility. The instrument shall be reduced.
operated for a period of 24 hours, during which all The rms values of test voltage which the instrument
operating conditions shall be held constant. As frequently shall withstand without damage or flashover shall be
as desired during this period, the error (accuracy1 shall determined by reference to the design limits in the De-
be measured and compared with data obtained prior to tailed Requirement Sheets. The test voltage shall be
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
~
9009273 0 0 0 1 0 b 2 S S T
c39.4
22 AMERICAN STANDARD SPECIFICATIONS FOR
means shall be shut off and the instrument shall be de- The case of the instrument shali be subjected to shock
energized and allowed to cool for 48 hours in the test along each of its major axes ( a total of 6 directions) hav-
chamber. Inasmuch as the relative humidity rises as the ing the acceleration stated in the Detailed Requirement
temperature decreases, precautions should be taken to Sheets as the design limit for which subsequent operation
prevent dripping on the instrument. At the end of the of the instrument will not be impaired. The duration of
cooling period, the instrument shall be removed from the the shock shall not exceed 50 milliseconds.
chamber and maintained de-energized under the test
conditions of Section 6 for 24 hours. Its performance
characteristics shall then be measured and compared
with the data obtained prior to the test.
8.4.4 Storage Temperature Test. Standard cells, 9. Revision of American Standards
batteries, and ink shall be removed from the instrument Referred to in This Document
for this test. The instrument shall be subjected for 48
' hours to each extreme value of the storage temperature When the following American Standards referred to
range stated in the Detailed Requirement Sheets as the in this document are superseded by a revision approved
design limits for which subsequent operation of the in- by the American Standards Association, Incorporated,
strument will not be impaired. Following exposure at the the revision shall. apply:
temperature extremes, the instrument shall be main- American Standard Electrical Safety Requirements for
tained de-energized under the test conditions of Section 6 Electrical Measuring and Controlling Instrumentation,
for 24 hours. Its performance characteristics shall then C39.5-1964
be measured and compared with the data obtained prior American Standard Definitions of Electrical Terms,
to the test. Instruments, Meters and Meter Testing (Croup 30),
8.4.5 Vibration Test. The instrument shall be vi- C42.30-1957
brated in the direction of each of its three major axes. American Standard Terminology for Automatic Con-
Components which are normally removed during ship- trol, C85.1-1963
ment shall be removed for the test, and shipping clamps American Standard Unified Screw Threads, Bl.1-1960
or other securing methods normally used during ship- American Standard Screw Thread Gages and Caging,
ment shall be employed. B1.2-1951
The instrument shall be vibrated over the frequency American Standard Square and Hexagon Bolts and
tange and at the acceleration stated in the Detailed Nuts, B18.2-1960
Requirement Sheets as the design limits at which dam- American Standard Slotted and Recessed Head Wood
$ging resonance shall not occur. Audible or visible reso- Screws, B18.6.1-1961
nance shall be considered to be potentially damaging American Standard Hexagon Head Cap Screws, Slotted
unless proved otherwise by vibration testing at these Head Cap Screws, Square Head Set Screws, and Slotted
frequencies for one hour. Headless Set Screws, B18.6.2-1956
8.4.6 Shock Test. Components which are normally American Standard Slotted and Recessed Head Ma-
removed during shipment shall be removed for this test, chine Screws and Machine Screw Nuts, B18.6.3-1962
and shipping. clamps or other securing methods normally American Standard Slotted and Recessed Head Tap-
used during shipment shall be employed. ping Screws and Metallic Drive Screws, B18.6.4-1958
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
m 9009273 0001063 496 m
c39.4
AUTOMATIC NULLBALANCING ELECTRICAL MEASURING INSTRUMENTS 23
DETAILED REQUIREMENTS
General Characteristics
Direct-Current Potentiometer
Direct-Current Bridge
Strip Chart -3 inches to less than 6 inches wide
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
~ ~
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
m 9009273 00010b5 2b9 m
c39.4
AUTOMATIC NULL-BALANCING ELECTRICAL MEASURING INSTRUMENTS 25
DETAILED REQUIREMENTS
General Characteristics
Direct-Current Potentiometer
Direct-Current Bridge
-
Strip Chart 6 inch minimum width and round chart
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
-~
~~ ~
9009273 OOOLObb L T 5
(39.4
26
.-s
I
E
L
cr;
.-C
L
3
-
L
I
L
a
U
dt
-z8
.a
2w
I
S
E
I
2.
Y.
.-C .
ua5
fi
I
E)
--``,`,````,,``,```,,,,,,,,,-`-`,,`,,`,`,,`---
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
12/30/2003 08:08:08 MST Questions or comments about this message: please call
the Document Policy Group at 1-800-451-1584.
~- ___
AMERICAN STANDARDS
The standard in this booklet is one of over 2,600 standards approved to date
by the American Standards Association, incorporated.
The ASA provides the machinery for creating voluntary standards. It serves
to eliminate duplication of standards activities and to weld conflicting standards
into single, nationally accepted standards under the designation "American
Standards."
Each standard represents general agreement among maker, seller, and user
groups as to the best current practice with regard to some specific problem. Thus
the completed standards cut across the whole fabric of production, distribution,
and consumption of goods and services. American Standards, by reason of
ASA procedures, reflect a national consensus of manufacturers, consumers, and
scientific, technical, and professional organizations, and governmental agencies.
The completed standards are used widely by industry and commerce and often
by municipal, state, and federal governments.
The ASA, under whose auspices this work i s being done, i s the American
clearinghouse and coordinating body for standards activity on the national
level, Founded in 1918, it i s a federation of 135 trade associations, technical
societies, professional groups, and consumer organizations. Some 2,000 com-
panies are affiliated with the ASA as company members.
COPYRIGHT 2003; American National Standards Institute Document provided by IHS Licensee=IHS Employees/1111111001, User=EEPPM03,
4 MST Questions or comments about this message: please call
12/30/2003 08:08:08
the Document Policy Group at 1-800-451-1584.