Professional Documents
Culture Documents
Ultrasonic Flowmeter Design: Chapter Fifteen
Ultrasonic Flowmeter Design: Chapter Fifteen
Ultrasonic Flowmeter Design: Chapter Fifteen
15.1 General
The measurement system should be designed, operated, and main-
tained to provide measurements within *0.50% or &1.00%overall
uncertainty. (Note: To achieve an uncertainty of *0.50%, the multi-
path ultrasonic flowmeter must be linearized as a function of MF
versus qvg, or MF versus qav.) The multipath ultrasonic flowmeter
consists of three or more chordal paths using invasive transducers
design, an high-performance flow conditioner, and associated
upstream and downstream pipe spools (flowmeter run). The ultra-
sonic flowmeter must comply with A.G.A. Report No. 9, latest
361
362 Natural Gas Measurement Handbook
edition, the Federal Gas Pipeline Safety Regulations, and all other
applicable codes and standards. Failure or removal of one pair of
the transducers should not cause the flowmeter to lose all measure-
ment functions.
The flowmeter and direct mount electronics should be designed
for an environmental temperature range of -13°F to 131°F in the
presence of condensing humidity. The ultrasonic flowmeter(s)
should be leak-tested by the manufacturer after final assembly and
prior to shipment to the dynamic calibration facility. The leakage
test uses nitrogen gas at 200 psig for a minimum of 15 minutes with
no leaks detectable using a noncorrosive liquid solution (soap test).
Ultrasonic flowmeters are calibrated in two steps: (1) a static
(or zero flow) calibration using pure nitrogen as the test medium at
the manufacturer’s facility and (2) a dynamic calibration of the
flowmeter assembly(s) using natural gas as the test medium at an
approved flow laboratory as close as practical to the normal operat-
ing pressure of the facility. The flowmeter assembly should be
recalibrated dynamically at an approved flow laboratory at a mutu-
ally agreed-on interval (i.e., every five years).
Where the delivery is required to be continuous and it is im-
practical to shut down in the event of a flowmeter failure, a spare
flowmeter@)is installed, complete with associated accessories, and
symmetrical with the initial flowmeter assembly.
Where it is likely for parallel flowmetering to occur, special
consideration must be given to balanced flow and pressure control.
For parallel flowmetering installations, the design must be symmet-
rical with and of similar size to the other flowmeter assemblies.
Automated valve sequencing is installed to provide for accurate
measurement, turndown, and flowmeter failure. For parallel
MUSM flowmeters, flow control valves should be installed in a sta-
tion control valve manifold to allow for control valve maintenance.
The pressure regulation or flow control valves should be installed
downstream of the flowmeter assemblies.
Bidirectional flowmeter facilities are designed with dedicated
totalization for each direction of flow, piping spools, HPFCs, com-
plex instruments, and directional meter factors for the MUSM.
Ultrasonic Flowmeter Design 363
Flowmeter Alarms
The following alarms should be provided in the form of digital
communications or fail-safe, dry, relay contacts, or voltage-free
solid-state switches isolated from the ground:
Output invalid (when indicated flow rate at line conditions is
invalid).
Partial failure (when one or more of the path results is not
usable).
Ultrasonic Flowmeter Design 369