Professional Documents
Culture Documents
Gpa STD 3132-1984
Gpa STD 3132-1984
Natural Gasoline
Specifications and Test Methods
Reprinted 7 980
Revised I984
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
Members:
Single copies.. ...................... .4.W
1O or more copies, each.. ............3.00
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
TABLE OF CONTENTS
NOTE: ASTM test methods included in this publication have been adopted
by GPA, and are included through a right-to-reprint agreement with,
and by special permission of, the American Society for Testing and
Materials, 1916 Race Street, Philadelphia, Pa
Natural Gasoline
Specifications and Test Methods
Scope: These specifications state the required properties of Natural Gasoline. Natural
gasoline is a mixture of liquid hydrocarbons extracted from natural gas, com-
posed principally of pentanes and heavier hydrocarbons, although varying
amounts of butanes may be included, depending on the commercial grade.
In addition to the above general specifications, natural gasoline shall be divided into
twenty-four possible grades on a basis of vapor pressure and percentage evaporated
a t 140°F. Each grade shall have a range in vapor pressure of four pounds, and a range in
the percentage evaporated a t 140°F of i 5 percent. The maximum vapor pressure of the
various grades shall be 14, 18, 22, 26, 30, and 34‘pounds respectively. The minimum
percentage evaporated a t 140°F shall be 25, 40, 5 5 , and 70 respectively. Each grade
shall be designated by its maximum vapor pressure and its minimum percentage evap-
orated a t 140”F, as shown in the following table:
-4-
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
@ Designation: 1%/78
Deutsche Nomi DIN 51759
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
Standard Method for
DETECTION OF COPPER CORROSION' FROM
PETROLEUM PRODUCTS BY THE COPPER STRIP
TARNISH TEST'
This standard is issued under the fixed designation D 130; the number immediately following the designation indicates the
year of original adoption or, in the case of revision. the year of last revision. A number in parentheses indicates the year of last
reapproval. This is also a standard of the Institute of Petroleum issued under the fixed designation IP 154. The final number
indicates the year of last revision.
This method was adopted as a joint A S T M - I P standard in 1964.
This merhod has been adopted for use by government agencies to replace Method 5325 of Federal Test Method Standard.No.
791b.
refining. However, of the sulfur compounds for protecting corroded strip for close inspec-
remaining in the petroleum product, some can tion or storage (see Appendix),
have a corroding action on various metals and
this corrosivity is not necessarily related di- 6. Materials
rectly to the total sulfur content. The effect can 6.1 Wash Solvent-Any volatile, suifur-free
vary according to the chemical types of sulfur hydrocarbon solvent may be used provided that
compounds present. The copper strip corrosive it shows no tarnish at all when tested at 50°C
test is designed to assess the relative degree of (122°F). Knock tesi grade isooctane is a Suita-
corrosivity of a petroleum product. ble solvent and should be used in case of dis-
pute.
5. Apparatus Danger! Extremely Flammable-see Annex
5.1 Test Tubes, 25 by 150-mm. A2.1.
5.1.1 Provide a bath capable of being main- 6.2 Polishing Materials-Silicon carbide grit
tained at a constant temperature of 50 f 1°C paper of varying degrees of fineness including
(122 f 2°F) or 100 +- 1°C (212 f 2"F), or both, 65-pm (240-grit) paper or cloth; also a supply
and having suitable supports to hold the test of 105-pm (150-mesh) silicon carbide grain and
tubes in a vertical position and immersed to a pharmaceutical grade absorbent cotton (cotton
depth of about 100 mm (4 in.). Either water, wool).
oil, or aluminum block baths are suitable. 6.3 Copper Strips:
5.2 Copper Strip Corrosion Test Bomb, con- 6.3.1 Specipcation-Use strips 12.5 mm (%
structed of stainless steel according to the di- in.) wide, 1.5 to 3.0 mm (%6 to ?hin.) thick, cut
mensions as given in Fig. 1, and capable of 75 mm (3 in.) long from smooth-surfaced, hard-
withstanding a test pressure of 100 psi (689 temper, cold-fmished copper of 99.9 + percent
kPa). Alternative designs for the bomb cap and purity; electrical bus bar stock is generally suit-
synthetic rubber gasket may be used provided able (see Appendix). The strips may be used
that the internal dimensions of the bomb are repeatedly but should be discarded when the
the same as those shown in Fig. 1. Provide a 25 surfaces become deformed on handling.
by 150-mm test tube as a liner for holding the 6.3.2 Surface Preparation-Remove ali sur-
sample. face blemishes from all six sides of the strip
5.2.1 Provide liquid baths capable of being with silicon carbide paper of such degrees of
maintained at 40 A 1°C (104 f 2°F) or 100 f fineness as,are needed to accomplish the de-
1°C (212 f 2"F), or both, and having suitable sired results efficiently. Finish with 65-pn (240-
supports to hold the test bomb in a vertical grit) silicon carbide paper or cloth. removing
position. The bath must be deep enough so that all marks that may have been made by other
the entire bomb will be submerged during the grades of paper used previously. Immerse the
test. As the bath medium, use water or any strip in wash solvent from which it may be
other liquid which ca,n be satisfactorily con- withdrawn immediately for final polishing or
trolled at the specified test temperature. in which it may be stored for future use.
5.3 Thermometers, total immersion, for in- 6.3.2.1 As a practical manual procedure for
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
dicating the required test temperature, with surface preparation, place a sheet of the paper
smallest graduations of 1°C (2°F) or less. No on a flat surface, moisten it with kerosine or
more than 25 mm (1 in.) of the mercury thread wash solvent, and tub the strip against the
should extend above the surface of the bath at paper with a rotary motion, protecting the strip
the test temperature. The ASTM 12C (12F) or from contact with the fingers with an ashless
IP 64C (64F) thermometer is suitable. filter paper. Alternatively, the surface of the
5.4 Polishing Vise, for holding the copper strip may be prepared by use of motor-driven
strip firmly without marring the edges while machines using appropriate grades of dry paper
polishing. Any convenient type of holder (see or cloth.
Appendix) may be used provided that the strip 6.3.3 Final Polishing-Remove a strip from
is held tightly and that the surface of the strip the wash solvent. Holding it in the fingers
being polished is supported above the surface protected with ashless filter paper. polish first
of the holder. the ends and then the sides with the 105-pm
5.5 Viewing Test Tubes, flat, are convenient (150-mesh) silicon carbide grains picked up
-6-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
G P A STD*3132 84 W 3824699 0000303 9
-
COPPERCORROSION
FROM PETROLEUM (ASTMD 130 - IP 154)
PRODUCTS
from a clean glass plate with a gad of cotton side of each plaque.
(cotton wool) moistewd with a drop of wash 7.1.2 If the surface of the plastic cover shows
solvent. Wipe vigorously with fresh pads of excessive scratching it is suggested that the
cotton (cotton wool) and subsequently handle plaque be replaced.
only with stainless steel forceps; do not couch
with thejìngers. Clamp in a vise and polish the 8. Samples
main surfaces with silicon-carbide grains on 8.1 It is particularly important that all types
absorbent cotton. Rub in the direction of the of fuel samples, which should pass a low-tar-
long axis of the strip, carrying the stroke be- nish strip classification, be collected in clean,
yond the end of the strip before reversing the dark glass bottles, plastic bottles, or other suit-
direction. Clean all metal dust from the strip able containers that will not affect the corrosive
by rubbing vigorously with clean pads of ab- properties of the fuel. Avoid the use of tin plate
sorbent cotton until a fresh pad remains un- cmtainers for collection of samples, since ex-
soiled. When the strip is clean, immediately perience has shown that they may contribute
immerse it in the prepared sample. to the corrósiveness of the sample.
6.3.3.1 It is important to polish the whole 8.2 Fill the containers as completely as pos-
surface of the,strip uniformly to obtain a uni- sible and close them immediately after taking
formly stained strip. If the edges show wear the sample. Take care during sampling to pro-
(surface elliptical) they will likely show more tect the samples from exposure to direct sun-
corrosion than the center. The use of a vise (see light or even diffused daylight. Make the test
Appendix) will facilitate uniform polishing. as soon as possible after receipt in the labora-
tory and immediately after opening the con-
'I. Corrosion Standards tainer.
7.i ASTM Copper Strip Corrosion Standurds' 8.3 If suspended water (haze) is observed in
consist of reproductions in color of typical test the sample, dry by filtering a sufficient volume
strips representing increasing degrees of tarnish of sample through a medium rapid qualitative
and corrosion. the reproductions being encased filter, into the prescribed clean, dry test tube.
in plastic in the form of a plaque. Carry out this operation in a darkened room or
7.1. I Keep the plastic-encased printed under a light-protected shield.
ASTM Copper Strip Corrosion Standards pro- 8.3.1 Contact of the copper strip with water
tected from light to avoid the possibility of before, during, or after the completion of the
fading. Inspect for fading by comparing two test run, will cause staining, making it difficult
different plaques, one of which has been care- to evaluate the strips.
fully protected from light (new). Observe both
sets in diffused daylight (or equivalent) first 9. Procedure
from a point directly above and then from an 9. i Those product classes. to which given pro-
angle of 45". If any evidence of fading is ob- cedural variations are intended to be applied, are
served. particularly at the left-hand end of the listed below. Some product classes, being quite
plaque. it is suggested that the one that is the broad, may be tested by more than one set of
more faded with respect to the other be dis- conditions; in such cases the copper strip quality
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
-7-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
#
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
PROBLEM HARD COPY
G P A STD*3L32 84 m 3 8 2 4 b î î 0000304 0 W
FROM PETROLEUM
COPPERCORROSION (ASTM D 130 - IP 1.54)
PRODUCTS
specificationsor by agreement between parties. Copper Strap Corrosion Standards. Hold both
9.1. I For aviation gasoline, and aviation tur- the test strip aiid the standard strip plaque in
bine jrel-Place 30 mL of sample, completely such a manner that light reflected from them
clear and free of any suspended or entrained at an angle of approximately 45" wül be ob-
water (see 8.3) into a chemically clean, dry 25 by served.
150-mtn test tube, and within I min after com- 9.2.2 in handling the test strip during the
pleting the final polishing, slide the copper strip inspection and comparison;- the danger of
inta the sample tube. Carefully slide the sample marking or staining can he avoided if it is
tube into the test bomb (Fig. I ) and screw the lid inserted in a flat tube (see Annex AI) which
on tight, Completely immerse the bomb in a can be stoppered with absarbent cotton.
boiling water bath at 100 f 1°C (212 f 2°F). _.
After 2 h _+ 5 min in the bath, withdraw the 10, Interpretation
bomb and immerse for a few minutes in tap 10.1 Interpret the corrosiveness of the sam-
bomb, 'withdraw the test tube ple accordingly as the appearance of the test
and examine the strip as described in 9.2. strip agrees with one of thestrips of the ASTM
9.1.2 For nuriiralgasoline-Carry out the test Copper Strip Corrosion Standards.
exactly as described in 9-1.1 but at 40°C (104°F) 10. I . 1 When a strip is in the obvious transi-
and for 3 h f 5 min. tion state between that indicated by any two
9.1.3 For diesel /irel, fìiel oil, qtitomorive gas- adjacent standard strips, judge the sample by
oline-Place 30 mL of sample, completely dear the more tariiished Standard Strip. Should a.
and free of ahy suspended or entrained water (see strip appear to have a darker orange color.than
8.31, into a chemically clean, dry 25 by 150-mm Standard Strip lb, consider the observed strip
test tube and, within 1 min after completing the as still belonging in Classification 1: however,
fTnal polishing, slide the copper strip into the if any evidence of red CO is observed, the
sample tube. StoppeI; with a vented cork and observed strip belongs in Classification 2,
place in a'bath maintained at 50 i 1°C*(132f 10.1.2 A claret red strip in Classification 2
2°F) (see 5.'1. I).' Protect the contents of the test can be mistaken for a magenta overcast on
tube from strong light,during the test: After 3 h brassy strip in Classification 3 if the brassy
t 5 min in the bath, examine the strip as.de- underlay of the latter is completely masked by
scribed in 9.2. For tests on fuel oil and diesel a magenta overtone. To distinguish, immerse
fuel, to specifications other than' ASTM Specifi- the strip in wash solvept; the former will appear
cations D 396 and D 975, a temperature of 100°C as a dark orange strip while the latter will not
(2 12°F)for 3 h is often used as an alternative set change.
of conditions. 10.1.3 To distinguish multicolored strips in
' 9. I .4 For cleaners (Stoddurd) solvent and ker- Classifications 2 and 3, place a test strip in a 20
osine-Carry out the test exactly as described in by 150-mm test tube and bring to a temperature
9.1.3- but at 100 -L 1°C (212 2 2°F). of 315 to 370°C (600 to 700°F) in 4 to 6 min
9. I .5 For lubricuting oil-Tests may be car- with the tube 1,ying on a hot plate. Adjust to
ried out for varying times and at elevated tem- temperature by observing. a high distillation
peratures other than 100°C (212°F).For the sake thermometer inserted int0.a second test tube. If
of Liniformity, it is suggested that even incre- the strip belongs in Classification 2, it will
ments of 50°F beginning with 350°F (or Celsius assume the color of a silver and then a gold
equivalents ta the nearest whole degree) be used. strip: If in Classification 3 it will take on the
9.2 Strip Examination: ' appearance of a transparent black. etc., as de-
9.2.1 Empty the contents of the test tube into scribed in Classification 4. . .
a 150-mL tall-form beaker, letting the strip 10.1.4 Repeat the lest if blemishes due to
slide in gently so as to avoid breaking the f î g e r prints are observed, or due to spots from
beaker. Immediately withdraw the strip with any particles of water droplets that may have
stainless steel forceps and immerse in wash touched the test strip duting the digestion pe-
solvent. Withdraw the strip at once, dry with riod.
quantitative filter paper (by blotting and not 10.1.5 Repeat the test also if the sharp edges
by wiping), and inspect for evidences of tar- along the flat faces of the strip appear to be in
nishing or corrosion by comparison with the a classificatian higher thans the greater portion
-8-
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
COPPERCORROSION
PROM PETROLEUM (ASTM D 130 - IP 154)
PRODUCTS
of the strip; in this case it is likely that the edges State the duration of the test and the test tem-
were burnished during polishing. perature.
12. Precision
11. Report 12.1 In the-case of pass/fail data no gener-
11. I Report the corrosiveness in accordance aliy accepted method for determining precision
with one of the classifications listed in Table 1. is currently available.
I slight tarnish a. Light orange, almost the same as freshly polished strip
b. ûarkorange
4 corrosion a. Transparent black, dark gray or brown with peacock green barely showing
... b. Graphite or lusterless black
c. Glossy or jet black
~
The ASTM Copper Strip Corrosion Standard is a colored reproduction of strips characteristic of these descriptions.
A
"The freshly polished strip is included in the series only as an indication of the appearance of a properly polished strip before a
test run: it is not possiblc io duplicate this appearance after a test even with a completely noncorrosive sample.
\CHAMFER
FREE FROM FREE SULPHUR
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
MAX. TEST PRESSURE 100 LBF/INZ
(7.03Kgkmzl
C
z -1.6 WALL SEAMLESS TUBE
('46-1
o
(r
-9-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
r
G P A STD*313Z 84 3824699 O000306 4 M
COPPERCORROSION
FROM PETROLEWN PRODUCTS (ASTM D 130 - IP 154)
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
ANNEXES
(Mandatory Information)
MINIMUM DIMENSION
TO BE SUCH THAT A TUBE TO BE WITHOUT
COPPER STRIP 3 2 X 13 STRIAE OR SIMILAR DEFECTS
SECTION WILL ENTER
-10-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
G P A STD*3L32 84 3 8 2 4 6 9 9 0000307 b
ASTM D 180-80
120 (4%4")
d
i
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
SECTION 0 E
SECTION A A
- 1
5
COPPERCORROSION
FROM PETROLEUM (ASTM D 130 - IP 154)
PRODUCTS
The American Society f Ò r Testing and Materials takes no position respecting the validity of any patent righrs asserted in
connection with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validir).
ofany such patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed everyfive Tears
and if not revised, either reapproved or withdrawn. Your comments are invited eitherf o r revision of this standard or for additional
standards and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If youfeel that your comments have not received afair hearingyou should
make your views known to the A S T M Committee on Standards, 1916 Race St.. Philadelphia. Pa. 19103.
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
- 12 -
This method has been adopled for use by government agencies to replace Method 4246 of Federal Test Method Standard No.
791b, and Method 101 of Federal Test Method Standard No. 141A.
-13- --`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
same operator with the same apparatus under tween two shgle and independent test results
constant operating conditions on identical test obtained by different opkrators working in dif-
material would, in the long run,in the normal ferent laboratories on identical test material
and correct operation of the test method, exceed would, in the long run, in the normal and
the foliowing value only in one case in twenty: correct operation of the test method exceed the
1 color unit following value only in one case in twenty:
13.2.2 Reproducibility-The difference be- 2 color units
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
ANNEX
(MANDATORY INFORMATION)
A l . APPARATUS
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
- 16-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
AL3 Daylight Lamp and the luminous transmission (Tw),when calculated
A1.3.1 Lamp-Use a lamp bulb rated at 60 W from the spectral transmission data using the 1931
and conforming to the A m e n u n Association of ICI Standard Illuminant A,5 are as shown in Table
Lamp Manufacturers' specification 60A. It shall be A 1.2.
constructed of clear glass, inside frost finish, and shall Note A3-A spectrophotometric test of an ac-
be rated at approximately 13 lm/W and 2750 K'color ceptable filter should indicate a transmission of ra-
temperature. Attach to a standard socket reflector, diant energy not less than 60 percent at 410 nm, with
hemispherical in form, Fig. Al. I , the interior surface a smooth curve down to a transmission below 10
ofwhich is finished with a brilliant aluminum bronze percent at 700 nm. This curve furthermore, should
powder, free from mica and other adulterants, and not have the pronounced hump that is characteristic
applied with a heat-resistant bronzing liquid sprayed of excess cobalt. The typical cobalt curve has an
uniformly over the surface. This finish shall be such increased transmission at a wavelength of 570 nm
as to be free from selective absorption, and have an above a straight line drawn between the points 540
initial reflectivity above 65 %. nm and 590 nm, and also a transmission band in the
A1.3.2 Daylight Filter Glass (Note A3), concave- red for wavelengths of 600 nm and greater. This
convex in form, and dust-tight, shall fit closely over variation in an acceptable filter shall not, at 570 nm,
the opening in the hemispherical reflector, Fig. Al. 1. exceed 0.03 above the straight line drawn between
The glass filter shall be finished on its concave surface 540 nm and 590 nm, nor shall the transmission for
by sand-blasting and acid-smoothing or fortifying. wavelength 700 nrn exceed the transmission for any
An acceptable daylight filter shall possess such char- shorter wavelength, such as 600 nm, by more than
acteristics that the trilinear coordinates ( x , y, and z), 0.03.
TABLE Al.1 Characteristics of Color Standards TABLE A1.2 Characteristics of Daylight Filter
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
(D156)
PRODUCTS
TESTPOR COLOXOE PETROLEUM
,€q=
o- HA OLL
NOTL- HAWOLC IS f U T D l f D TO "Da
@-
- 18 -
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
I I
The American Socieiy for Tesiing and Maierials takes no position respeciing ihe validity of any paient rights asseried in
connection wiih any item meniioned in ihis standard. Users of ihis siandard are expressly advised thaf determination of ihe validity
of any such paient righis, and ihe risk of infringemeni of such righfs, are eniirely their own responsibility.
This standard is subject io revision ai any rime by the responsible iechnical commiiiee and musi be reviewed every five years
and if nof revised, eiiher reapproved or wiihdrawn. Your cornmenis are invited eiiherfor re-visionof ihis standard orf o r additional
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
standards and should be addressed io A S T M Headquarters. Your commenis will receive careful consideration at a meeting of ihe
responsible iechnical commiiiee, which you may aiiend. q y o ufeeI ihai your comments have not received afair hearing you should
make your views known io fhe A S T M Commiiiee on Standards, 1916 Race St., Philadelphia, Pa. 19103.
-19-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST PROBLEM HARO C O P Y
G P A STD*3l132 84 m 3 8 2 4 b î î 0 0 0 0 3 1 b 7 m
THIS STANDARD HAS BEEN REAPPROVED
WITHOUT CHANGE IN 1984
4CTb C G 2
AMERICAN NATIONAL ANSI/ASTM D 216 - 77 Method 1015-Federal Test Method
Standard No. 791 b
British Standard 4717
Designation: 191/ 65
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
Standard Method for
DISTILLATION OF NATURAL GASOLINE'
This Standard is issued under the fixed designation D 216; the number immediately following the designation indicates the
year of original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of
last reapproval. This is rilso a standard of the Institute of Petroleum issued under the fixed designation IP 191. The final
number indicates the year of last revision.
Thb method was issued as a joint ASTM - IP standard in 1965.
-21
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
_.
-
.I
OF NATUBAL
TESTFOR DISTILLATION GASOLINE(ASTM D 216-IP Ni)
the series ordinarily employed by the labora- report the figure as the volume evaporated, in
tory making the test. the second, as the volume recovered.
8.2 When the liquid residue in the distilla-
tion flask is approximately 5 ml, the heat may
be increased because of the presence of heavy 9. Correctioii trf Temperature Poinús Po S t a ~
ends which have relatively high boiling points. dard Atmospheric Pressure
However, do not apply any further increase of 9.1 Ascertain and record the actual baro-
heat after this adjustment. The 4 t a 5-ml rate metric pressure, but do not make any correc-
cannot always be maintained from this point tion except in case of dispute. In such cases
to the end of the 'distillation, but in no case correct the temperature points to 760 mm
should the period between the point when (29.92 in.) HE, by the use of the Sydney
approximateiy 5 ml of the liquid remains in Young equation, as follows:
the flask and the end point be more than 5 For Celsius readings:
min.
8.3 Continue the heating until the mercury C, = 0.00012 (760 - P)(273 + fe)
reaches a maximum and starts to fqll consist- For Fahrenheit readings:
ently after the bottom of the flask has become Cf= 0.00012 (760 - P)(460 + 1,)
dry. Record the highest temperature observed where:
on the distillation thermometer as the maxi- Ce and CI = the correction to be added to the
mum temperature or end point.
observed temperature te or lf, respec-
8.4 Observe the total volume of the distil- tively, aiici
late collected in the receiving cylinder at the
P = the actual barometric pressure, mm Hg.
conclusion of the distillation, and thereafter at
In case of dispute, however, no test shall be
not less than 2-min intervals, until successive
regarded as official when conducted under an
observations agree. Then record this volume
absolute pressure lower than 735 mm (28.94
as the recovery
in.) Hg.
8.5 Pour the cooled residue from the flask
9.2 The following table is a convenient a p
into a small cylinder graduated in 0.1 ml,
proximation of the corrections as calculated
measure when cool, and record the volume as
by the above equation:
residue at 32 to 40°F (O to 4.5"C).
8.6 Calculate the difference between 100 Correctiona/lO-mm
ml and the sum of the recovery and the resi- Temperature Range Difference in Pressure
due and record as distillation loss. "C "F "C "F
NOTE 3 -Abnormal distillation losses, when l o t o 30 50 to 86 0.35 0.63
caused by low barometric pressure, may. be cor- 30 to $0 86 to 122 0.38 0.68
rected by the equation given 'below to an approxi- 50 to 70 122 to 158 0.40 0.72
mation of the loss to be expected under normal 70 to 90 158 to 194 0.42 0.76
pressure. The corrected losses shall not be used in 90 to 110 194 to 230 0.45 0.81
the calculation of percentages evaporated. 110 to 130 230 to 266 0.47 0.85
Corrected loss = A L + B 130 to 150 266 to 302 0.50 0.89
150 to 170 302 to 338 0.52 0.94
where: 170 to 190 338 to 374 0.54 0.98
L = observed distillation loss, 190 to 210 374 to 410 0.57 1.o2
A and B = empirical constants, as given in Table 210 to 230 410 to 446 0.59 1.O6
1, for varioüs values of the observed baromet- 230 to 250 446 to 482 0.62 1.11
ric pressure. 250 to 270 482 to 518 0.64 1.15
270 to 290 518 tq 554 0.66 1.19
8.7 The sum of the volume collected in the 290 to 310 554 to 590 0.69 1.24
cylinder at any specified temperature, includ- 310 to 330 590 to 626 0.71 1.28
ing the initiai boiling point, and the distillation 330 to 350 626 to 662 0.74 1.32
350 to 370 662 to 698 0.76 1.37
loss may be recorded as the percentage evapo- 370 to 390 698 to 734 0.78 1.41
rated at the temperature in question. Make 390 to 410 734 to 770 0.81 1.45
clear distinction between the inclusion or ex-
clusion of the distillation loss figure in the
To be added in case barometric pressure is below 760
report of the volume of disiillate obtained at mm Hg (101.3 kPa); to be subtracted in case barometric
any prescribed temperature. in. the first case pressure is above 760 mm Hg (101.3 kPa).
-22-
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
OP NATURAL
TESTSOB DISTILLATION GASOLINE(ASTM D 216-IP 191)
The American Society for Testing and Materials takes no position respecting the ialidity of any patent rights asserted
in connection with any item mentioned in this standard. Users of this standard are expressly advised that determination
of the validity of any such patent rights, and the risk of infringement of such rights, is entirely their own responsibility.
-23-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
G P A STD*3132 84 W 3824699 0000320 9 W
-24-
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
D 323
However, samples taken in containers of other sizes (O to 1°C) for at least 10 min.
as prescribed in Method D 4057 may be used if it is 6.4 Preparation of Air Chamber-After
recognized that the precision might be affected. In
the case of referee testing, the I-qt (1-L) sample purging and rinsing the air chamber and pres-
container shall be mandatory. sure gage in accordance with 7.5, connect the
gage to the air chamber. Immerse the air cham-
5.4 Precautions:
ber to at least 1 in. (25 mm) above its top in the
5.4.1 The Reid vapor pressure determina-
water bath mainfained at 100 f 0.2"F (37.8 f
tion shall be the first test run on a sample, nor
0.1"C) for not less than 10 min just prior to
may more than one sample be withdrawn from
coupling it to the gasoline chamber. Do not
the sample container for this test.
remove the air chamber from the bath until the
5.4.2 Samples shall be protected from exces-
gasoline chamber has been filled with sample
sive heat prior to testing.
as described in 7.1.
5.4.3 Samples in leaky containers shall not
be tested. They should be 'discarded and new 7. Procedure
samples obtained. 7.1 Sample Transfer-With everything in
5.5 Sample Handling Temperature-In all readiness, remove the chilled sample container
cases, the sample container and contents shall from the bath, uncap it, and insert the chilled
be cooled to 32 to 34°F (O to 1°C) before the transfer apparatus (see Fig. 1). Quickly empty
container is opened. Sufficient time to reach the water from the chilled gasoline chamber
this temperature shail be assured by direct mea- and place it, in an inverted position, over the
surement of the temperature of a similar liquid sample delivery tube of the transfer apparatus.
in a like container placed in the cooling bath at Invert the p t i r e system rapidly so that the
the same time as the sample. gasoline chamber is upright with the end of the
delivery tube touching the bottom of the gaso-
6. Preparation for Test line chamber. Fill the gasoline chamber to
6.1 Verification of Sample Container Fill- overflowing. Withdraw the delivery tube from
ing-With the sample at a temperature of 32 to the gasoline chamber while allowing the sample
34'F (O to l"C), take the container from the to continue flowing up to the moment of com-
cooling bath, unseal it, and examine its ullage. plete withdrawal.
The sample content, as determined by use of a 7.1.1 Caution-Provision should be made
suitable gage, shall equal 70 to 80% of the for suitable restraint and disposal of the over-
container capacity. flowing gasoline to avoid fue hazard.
6.1.1 Discard the sample if its volume is less 7.2 Assembly of Apparatus-Immediatelyre-
than 70 % of the container capacity. move the air chamber from the water bath and,
6.1.2 If the container is more than 80 % full, as quickly as possible, attach the air chamber
pour out enough sample to bring the container to the gasoline chamber. Not more than 40 s
contents within the 70 to 80 % range. Under no shall be consumed in coupling the two cham-
circumstance may any sample poured out be bers.
returned to the container. NOTE5-When the air chamber is removed from
the bath, aliow it to drain oniy briefly and, without
6.2 Air Saturation of Sample in Sample Con- shaking, connect it to the gasoline chamber without
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
-25 -
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
G P A STD*3132 84 m 3824699 0000322 2 m
D 323
the gasoline and air chambers is below the least five times. A4fterthoroughly removing the
water level and may be carefully examined for previous sample from the gasoline chamber,
leaks. If no leaks are observed, further immerse immerse the chamber in the ice bath for the
the apparatus to at least 1 in. (25 mm) above ‘nexttest. After disconnecting the pressure gage
the top of the air chamber. Observe the appa- from its manifold connection with the manom-
ratus for leakage throughout the test. Discard eter, remove trapped flvid in the Bourdon tube
the test at any time a leak is detected. òf the gage by repeated centrifugal thrusts. This
NOTE6-Liquid leaks are more difficult to detect may be accomplished in the following manner:
than vapor leaks; and because the coupling between hold the gage between the palms of the hands
the chambers is normally in the liquid section of the with the right hand on the face side and the
apparatus, give it particular attention. threaded connection of the gage forward. Ex-
7.4 Measurement of Vapor Pressure-After tend the arms foward and upward at an angle
the assembled vapor pressure apparatus has of 45” with the coupling of the gage pointing
been immersed in the bath for at least 5 min, in the same direction. Swing the arms down-
tap the pressure gage lightly and observe the ward through an arc of about 135” so that the
reading. Withdraw the apparatus from the bath centrifugal force aids gravity in removing the
and repeat 7.3. At intervals of not less than 2 trapped liquid. Repeat this operation three
min, perform 7.3 until a total of not less than times to expel all liquid. Purge the pressure
five shaking and gage readings have been made gage by directing a small jet of air into its
and continuing thereafter if necessary until the Bourdon tube for at least 5 min (Note 9).
last two consecutive gage readings are constant, NOTE8-If the purging of the äir chamber is done
indicating equilibrium attainment. These op- in a bath, be sure to avoid small and unnoticeable
erations normally require 20 to 30 min. Read films of floating sample by keeping the bottom and
the final gage pressure to the nearest 0.05 psi top openings of the chamber closed as they pass
through the water surface.
(0.25 kPa) for gages with intermediate gradua- NOTE9-In the case of crude oil, the Bourdon
tions of 0.1 psi (0.5 kPa) or less and to the tube must be rinsed with a volatile solvent after each
nearest.0.1 psi €ar gages with graduations of 0.2 test.
to 0.5 psi (1.O to 2.5 kPa), and record this value
as the “uncorrected vapor pressure’’ of the Sam- 8. Precautioiis
ple. Without undue delay remove the pressure 8.1 Gross errors can be obtained in vapor
gage (Note 7) and, without attempting to re- pressure measurements if the prescribed pro-
move any liquid which may be trapped in the cedure is not followed carefully. The following
gage, check its reading against that of the ma- list emphasizes the importance of strict adher-
nometer while both are subjected to a common ence to the precautions given in the procedure:
steady pressure which is no more than 0.2 psi 8.1.1 Checking the Pressure Gage-Check all
(1.O kPa) different from the recorded gages against a manometer after each test in
“uncorrected vapor pressure”. If a difference is order to ensure higher precision of results (7.4).
observed between the gage and manometer Read all gages while the gage is in a vertical
readings, the difference should be added to or position and after tapping it lightly.
subtracted from the “uncorrected vapor pres- 8.1.2 Shake the container vigorously to en-
sure” recorded for the sample being tested and sure equilibrium of the sample with the air in
the resulting value recorded as the Reid vapor the container (6.2).
pressure of the sample. 8.1.3 Checkingfor Leaks-Check all appa-
NOTE7-Cooling the assembly prior to discon- ratus before and during each test for both liquid
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
necting the gage will facilitate disassembly and re- and vapor leaks (Al. 1.6 and Note 6).
duce the amount of hydrocarbonvapors released into 8.1.4 Sumpzing-Because initial sampling
the room. and the handling of samples will greatly affect
7.5 Preparation of Apparatusfor Next Test- the final results, employ the utmost precaution
Disconnect the air and gasoline chambers and and the most meticulous care to avoid losses
discard contained sample. Thoroughly purge through evaporation and even slight changes in
the air chamber of residual sample by filling it composition (Sections 5 and 7.1). In no case
with warm water above 90°F (32°C) and allow- shall any part of the Reid apparatus itself be
ing it to drain, (Note 8). Repeat this purging at used as the sample container previous to ac-
-26-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
GPA STD*3L32 84 3824699 0000323 4 m
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
D 323
D 323
-28 -
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
G P A STD*3132 84 3 8 2 4 b 9 9 0000325 8 M
may be calculated by means of the foilowing equation: AFor other temperatures anä pressures, the corrections
c = I@ - pd(r - 100)/460 + 11 - - pi) may be calculated by means of the foilowing equation:
where:
c = correction, psi,
c = [@ - pt)(t - 37.8)/273 + 4 - b 3 7 . ü - p i )
f = air chamber temperature at beginning of test, OF, where:
p = barometric pressure, psi, at time of test (if a barom- c = correction, kPa,
eter is not available. the normal barometric pressure p = barometric pressure, kPa, at the time of the test,
may be used), pi = vapor pressure of water, kPa, at t OC,
pl = vapor pressure of water, psia, at P F , and t = initial air chamber temperature, OC, at the begin-
pia>= vapor pressure of water, psia, at 100OP = 0.95. ning of the test, and
Calculated corrections are to be rounded to the nearest p37.8 = vapor pressure of water, kPa, at 37.8OC, namely
0.05 psi. 6.56 kPa.
Esample-The pressure gage gives an “uncorrectedvapor Example-The pressure gage gives an “uncorrectedvapor
pressure” reading of I 1.6 psi. When the gage is compared to pressure” reading of 68.9 kPa. When the gage is compared
a mercury column, a “manometer reading” of 11.5 psi is with a mercury manometer, a manometer reading of 68.2
obtained. For an “initial air temperature” of 80°F and kPa is obtained when the pressuregage reads 68.9 kPa. If the
atmospheric pressure of 700 m m Hg, the correctionshown in initial air chamber temperature is 30°C, the vapor pressure
Table 1 is 0.90 psi. Because the “initial air temperature” is of water (pi) would be 4.2 kPa. If the observed barometric
below 100°F. this correction of 0.90 psi is suhtracted from pressure is 90.0 kPa, the correction c is then calculated to be
the “manometer reading” of I I .5 psi giving, a “Reid vapor -4.6 kPa. The corrected vapor pressure is then: 68.2 - 4.6
pressure” of 10.60 psi. = 63.6 H a .
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
-29-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
b
D323
ANNEX
(MANDATORY INFORMATION)
A l . l Reid Vapor Pressure Bomb, consisting of two the air chamber. The inner surface of the end con-
chambers-an air chamber (upper section) and a taining the coupling member shall be sloped to pro-
gasoline chamber (lower section)-shall conform to vide complete drainage when inverted. The other end
the following requirements: of the gasoline chamber shall be completely closed.
NOTEAl.1: Caution-Tomaintain thecorrect vol- NOTE Al.2--The ratio for units to be used for
ume ratio between the air chamber and the gasoline aviation gasoline testing shall be between 3.95 to
chamber, the units shall not be interchanged without 4.05.
recalibratingto ascertain that the volume ratio is within Al. 1.3 Gasóline Chamber (Two-Opening)-For
satisfactory limits. sampling from closed vessels, the lower section or
Al.l.l Air Chamber-The upper section or air gasoline chamber, as shown in Fig. Al.l shall be
chamber, as shown in Fig. A 1.1. shall be a cyiindrical essentially the same as the gasoliie chamber de-
vessel 2 & % in. (51 f 3 mm) in diameter and 10 f scribed in A1.1.2, except that a %-in. valve shall be
% in. (254 & 3 mm) in length, inside dimensions, with attached near the bottom of the gasoline chamber
the inner surfaces of the ends slightly sloped to and a %-in. straight-through, full-opening valve shall
provide complete drainage from either end when be introduced in the coupling between the chambers.
held in a vertical position. On one end of the air The volume of the gasoiine chamber, including only
chamber, a suitable gage coupling with an internal the capacity enclosed by the valves, shall fulfill the
diameter not less than %6 in. shall be provided to volume ratio requirements as set forth in Al. 1.2.
receive the %-in.gage connection. In the other end of NOTE Al .3-Iai determining capacities for the
the air chamber an opening approximately % in. in two-opening gasoline chamber (Fig. Al. l), the ca-
diameter shall be provided for coupling with the pacity of the gasoline chamber shall be considered as
gasoline chamber. Care shall be taken that the con- that below the %-in.valve closure. The volume above
nections to the end openings do not prevent the the %-in.valve closure including the portion of the
chamber from draining completely. coupling permanently attached to the gasoline cham-
A 1.1.2 Gasoline Chamber (One-Opening)-The ber shall be considered as a part of the air chamber
lower section or gasoline chamber, as shown in Fig.
A 1.1, shall be a cylindrical vessel of the same inside capacity.
diameter as the air chamber and of such volume that A1.1.4 Method of Coupling Air and Gasoline
the ratio of the volume of the air chamber to the Chambers-Any method of coupling the a u and
volume of the gasoline chamber shall be between the gasoline chambers may be employed, provided that
limits of 3.8 and 4.2 (see Note A1.2). In one end of no gasoline is lost during the coupling operation, that
the gasoline chamber an opening approximately ?4 no compressioneffect is caused by the act of coupling,
in. in diameter shall be provided for coupling with and that the assembly is free of leaks under the
-30-
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST
G P A STD*3132 84 3824699 O000327 I =
D 323
conditions of the tests. To avoid displacement of calibration correction shall not be greater than O. 15
gasoline during assembly, it is desirable that the male psi (0.3 kPa) for a O to 15-psi (O tc 30-kPa) gage or
fitting of a suitable coupling be on the gasoline 0.3 psi (0.9 kPa) for a O to 30-psi (9 to 90-kPa) gage.
chamber. To avoid compression of air during the
assembly of a suitable screw coupling, à vent hole NOTEAIS-Gages 3% in. (90 mm) in diameter
may be used to ensure atmospheric pressure in the may be used in the O to 5-psi (O to 35-kPa) range.6
air chamber at the instant of sealing. A1.3 Water Cooling Bath-A water cooling bath
NOTEA 1.4: Caution-Some commercially avail- shall be provided of such dimensions that the sample
able equipment does not make adequate provision containers and gasoline chambers may be completely.
tòr avoiding air compression effects. Before employ- immersed. Means for maintaining the bath at a tem-
ing any apparatus, it shall be established that the act perature of 32 to 40°F (O to 4.5"C) shall be provided.
of coupling does not compress the air in the air NOTEAl.6-Solid carbon dioxide shall not be
chamber. This may be accomplished by tightly stop- used to cool samples in storage or in the preparation
pering the gasoline chamber opening and assembling of the air saturation step. Carbon dioxide is appreci-
the apparatus in the normal manner, utilizing the O ably soluble in gasoline, and its use has been found
to 5-psi (O to 35-kPa) gage. Any observable pressure to be the cause of erroneous vapor pressure data.
increase on the gage is an indication that the appa: A 1.4 Water Bath-The water bath shall be of such
ratus does not adequately meet the specifications of
the method. If this problem is encountered, the man- dimensions that the vapor pressure apparatus may be
immersed to at least 1 in. (25 mm) above the top of
ufacturer should be consulted for remedy. the air chamber. Means for maintaining the bath at
A1.1.5 Volumetric Capacity of Air and Gasoline
Chambers-In order to ascertain if the volume ratio a constant temperature of 100 & 0.2'F (37.8 f O. 1OC)
of the chambers is between the specified limits of 3.8 shall be provided. In order to check this temperature,
to 4.2 (see Note A1.2), measure a quantity of water the bath thermometer shall be immersed to the 98°F
greater than wiil be needed to fiil the gasoline and (37°C) mark throughout the vapor pressure deter-
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
-31 -
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST PROBLEM H A R D C O P Y
G P A STD*3132 84 m 3824699 0000328 3 =
I
A
1 I"g.
1
Gorolinr Chombrr
(Two Oponingr]
Coupling
G
vrnt
Gasoline Chombrr
(Onr Op«ring)'
Air Chomkr
The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted ih
connection with any item mentioned in this standard. Users of this standard are expressly advised that determination of the valiaYty
of any such patent rights, and the risk of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed everyfive years
and qnot revised, either reapproved or withdrawn. Your comments are invited eitherfor revision of this standord orfor &ional
standards and should be addressed to A STM Heudquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. v y o ufeel that your comments have not received afair hearingyou sñould
make your views known to the ASTM Committee on Standards, 1916 Race SI.,Philadelphia, Pa. 19103.
--`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---
- 32 -
Copyright Gas Processors Association
Provided by IHS under license with GPA Licensee=Agencia Nacional De Hidrocarburos/5993362001, User=Tapias, Erik
No reproduction or networking permitted without license from IHS Not for Resale, 12/28/2016 09:21:56 MST P R O B L E M HARD C O P Y
~~~
GPA
STANDARD DOCTOR TEST
for
NATURAL GASOLINE
DOCTOR TEST and the interfacial area between the sample and the plumbite
solution. Both of these factors determine the concentration
APPARATUS of the dark precipitate which discolors the sulfur film and is
1. The test shall be carried out in standard cylindrical used as the criterion of judgment as to whether a sample iS
four-ounce sample bottles, the inside diameter of which shall doctor sour or not.
be’approximately 33 mm. The bottles shall be fitted with Stopper tightly with a clean cork and shake vigorously for
clean cork stoppers. 15 seconds. Add a quantity of pure, dry sulfur just sufficient
to cover the interface between the sample and the plumbite
REAGENTS REQUIRED solution, stopper the bottle, shake vigorously for 15 seconds
.
2. (a) Sodium Plumbite (Doctor Solution). A plumbite
solution shall be prepared containing 125 grams of NaûH per
and allow to settle.
Nom.-It is important to avoid the addition of more sulfur
liter and 25 grams of PbO per liter. Where available sublimed than will just cover the interface, otherwise the dark precipi-
Litharge should be used since it is more easily dissolved. tate will be diluted by the excess sulfur and the interpretation
(b) Sulíur. Pure, dry flowers of Sulfur, ground and screened of the test thereby affected. With finely ground sulphur, as
to 100-200mesh, shall be used, and shall be kept in a closed specified, 25 to 30 milligrams is the proper quantity. This
container to protect from dust. amount can be accurately estimated with a little practice.
NoTE.-The reagents used shall be chemically pure. Ex- INTERPRETATION OF RESULTS
treme care must be taken in storing all chemicals and the
sodium plumbite solution in strictly clean, tightly closed 4. If, after agitation of the plumbite-treated sample with
receptacles. sulfur, the film immediately becomes discolored or flecked
This solution is most conveniently prepared by dissolving with brown or black, the test shall be reported positive and
125 grams of NaOH in one liter of distilled water adding 25 the sample considered “sour.”
grams of PbO and agitating the mixture for a period sufficient NoTE.-In many samples which have been examined, it has
to dissolve the reagents a t a temperature of 150-175°F.After been noticed that if the sample is over slightly sour, a colored
.cooling to room temperature, the solution may be brought “lace like” emulsion tends to form between the sample and
back to its original volume by addition of distilled water. the plumbite solution.
Undissolved litharge should be allowed to settle out ar.d clear When agitation of the plumbite-treated solution with sulfur
liquid should be siphoned off into a container protected from results in no discoloration of the yellow sulfur film, the test
’the air. Filteration through an asbestos mat may be employed shall be reported negative and the sample considered “sweet.”
if the solution does not settle clear.
NoTE.-The use of these exact amounts of PBO and NcOH NmE.-The appearance of an immediate black precipitate
will avoid undersaturated or oversaturated solutions. on the addition of the sample to the plumbite solution is in-
dicative of the presence of hydrogen sulfide. If no immediate
PROCEDURE precipitate is formed but the sample becomes opalescent and
much darker in color after shaking with the plumbite solu-
3. To 5 mi. of the plumbite solution in a standard test tion, the presence of dissolved elemental sulfur and mer-
bottle, add 10 ml. of the sample to be tested. captans is indicated. In the presence of mercaptans and
NorE.-Given a standard plumbite solution, the two most absence of hydrogen sulfide and dissolved elemental sulfur, a
important variables are the volume of sample to be tested clear yellow solution of varying intensity usually results.
-3 3 - --`,```,``,,`,,`,,,``,,,,,,`,,`,-`-`,,`,,`,`,,`---