Professional Documents
Culture Documents
Cu Medical Systems - Life Gain, Disfibrilador - Manual Tecnico
Cu Medical Systems - Life Gain, Disfibrilador - Manual Tecnico
Cu Medical Systems - Life Gain, Disfibrilador - Manual Tecnico
LiFEGAIN CU-HD1
Service Manual
CU Medical Systems, Inc. reserves the right to make changes to the device specifications contained in this manual
Edition 1
Software Version
All rights are reserved. Permission is granted to copy and distribute this document for educational use.
No part of this manual may be reproduced without the permission of CU Medical Systems, Inc.
Table of Contents
1. INTRODUCTION ....................................................................................................................... 4
1.1 Introduction of Product ................................................................................................. 4
1.2 Overview of Service Manual ......................................................................................... 5
1.3 Warning, Caution, and Notice ....................................................................................... 5
1.4 Abbreviations ................................................................................................................. 5
2. DEVICE DESCRIPTION ........................................................................................................... 6
2.1 Description of Appearance ........................................................................................... 6
2.2 Accessories .................................................................................................................... 9
2.3 Configuration of Product Screen ............................................................................... 13
2.4 Dependence and Independence of Internal Devices ................................................ 16
3. MAINTENANCE ...................................................................................................................... 17
3.1 Basic Maintenance ....................................................................................................... 17
3.1.2 Maintenance of Accessories .................................................................................... 17
3.1.2.1 Paper Replacement ........................................................................................ 17
3.1.2.2 Battery Maintenance ....................................................................................... 18
3.1.2.3 ECG Cable Maintenance ................................................................................ 18
3.1.2.4 Maintenance of SpO2 Sensors and Connection Adapters .......................... 18
3.1.2.5 Maintenance of External Defibrillation Paddles .......................................... 19
3.1.3 Cleaning .............................................................................................................. 19
3.2 Maintenance Checklist ................................................................................................ 20
4. PERFORMANCE VERIFICATION/INSPECTION ................................................................... 21
4.1 Target for Verification .................................................................................................. 21
4.2 Tools or Equipment ...................................................................................................... 21
4.3 Performance Verification (Visual Inspection)............................................................ 21
4.4 Performance Verification (Functional Check) ........................................................... 21
4.4.1 Self Test (Daily, Weekly, Monthly) .................................................................... 22
4.4.2 Manual Self Test ................................................................................................. 23
4.4.3 Self Test Result (Error Code Interpretation) .................................................... 24
4.4.4 Defibrillation Shock Test ................................................................................... 25
4.4.5 Disarm Function Test ........................................................................................ 26
4.4.6 ECG Measuring Function Test .......................................................................... 27
i
4.4.7 Pacer Test ........................................................................................................... 28
4.4.8 SpO2 Test ............................................................................................................ 29
4.4.9 Text, Voice Prompt Test ..................................................................................... 29
4.4.10 Printer Test ....................................................................................................... 33
4.4.11 Bluetooth & Data Transmission Test.............................................................. 35
4.4.12 Menu Operation Test ....................................................................................... 36
5. Troubleshooting .................................................................................................................... 42
5.1 Power-on Test ............................................................................................................... 42
5.2 Periodic Automatic Self Test ....................................................................................... 43
5.3 Manual Self Test ........................................................................................................... 44
5.4 Possible Common Problems during Operation of Product and Solutions ............ 45
5.5 Defibrillation Function Problems and Solutions ...................................................... 45
5.6 Problems with ECG Measurement Function and Solutions .................................... 47
5.7 SpO2 Problems and Solutions .................................................................................... 48
5.8 Pacer Problems and Solutions ................................................................................... 48
5.9 Printer Problems and Solutions ................................................................................. 48
5.10 Problems in Information Transmission Function and Solutions .......................... 49
5.11 Problems in Information Saving Function and Solutions ...................................... 49
5.12 Audio Function Problems and Solutions ................................................................ 49
5.13 LCD Function Problems and Solutions ................................................................... 50
5.1.4 Key and Button Problems and Solutions ............................................................... 50
6. Repair ..................................................................................................................................... 51
6.1 Repair Philosophy ........................................................................................................ 51
6.2 Precautions for Repair ................................................................................................. 51
6.2.1. Handling Problem ............................................................................................. 51
6.2.2 Disposal & Recycling ........................................................................................ 52
6.3 Safety Considerations ................................................................................................. 52
6.3 Product Configuration ................................................................................................. 53
6.4 Internal Assemblies Disassembly .............................................................................. 58
6.6 Product Assembly ........................................................................................................ 68
7. Operation................................................................................................................................ 70
7.1 AED ................................................................................................................................ 70
7.2 Manual Defibrillation .................................................................................................... 70
7.3 Pacer.............................................................................................................................. 71
7.4 Patient Monitoring........................................................................................................ 71
8. Specifications ........................................................................................................................ 72
ii
Body .................................................................................................................................... 72
Defibrillation Feature ......................................................................................................... 72
Sensitivity & Specificity..................................................................................................... 73
Defibrillation Energy Variation Corresponding to the Impedance ................................ 73
Manual Defibrillation .......................................................................................................... 74
AED ...................................................................................................................................... 74
Pacer.................................................................................................................................... 75
ECG Monitoring .................................................................................................................. 75
SpO2 ..................................................................................................................................... 75
Display: 7” TFT LCD .......................................................................................................... 76
Data Storage ....................................................................................................................... 76
Printer .................................................................................................................................. 76
Bluetooth............................................................................................................................. 77
Battery ................................................................................................................................. 77
AC Power Module, SMPS .................................................................................................. 78
Cigar Jack. .......................................................................................................................... 78
AC Adapter .......................................................................................................................... 78
Appendix 1 ................................................................................................................................. 79
Appendix 2 ................................................................................................................................. 82
Appendix 3 ................................................................................................................................. 84
iii
1. INTRODUCTION
4
1.2 Overview of Service Manual
This Service Manual provides information on the use and maintenance of the product and
information for device managers on the repair and performance maintenance of the product.
The LiFEGAIN CU-HD1 may not include all the options described in this manual, depending on
the purchase options. The purchase options therefore need to be checked through the purchase
staff, or the device management documents need to be checked before repairing the product.
For information on the operation of the LiFEGAIN CU-HD1 which is not described in this Service
Manual, refer to the User’s Manual.
1.4 Abbreviations
AED: Automated External Defibrillator
VF: Ventricular Fibrillation
VT: Ventricular Tachycardia
ECG: Electrocardiography
SpO2 : Saturation of oxygen in arterial blood flow
NSR: Normal Sinus Rhythm
HR: Heart Rate
MST: Manual Self Test
5
2. DEVICE DESCRIPTION
Front
6
Right Side
Left Side
7
Back
8
2.2 Accessories
Accessories are shown in the table below and each accessory may be replaced when old or
worn. When disposing of an accessory, follow the waste disposal guidelines, depending on the
kind of waste.
Accessory List
Part Number Accessory Name
External Paddles
9
Pads Connection Adapter
10
Extension Cable of SpO2 Measurement Sensor
AC Power Module
Battery Module
AC Power Adapter
Bed Rack
11
SD Card
Printer paper
Conductive Gel
Carrier Bag
12
2.3 Configuration of Product Screen
In the product, a 7-inch TFT LCD is used, and the ECG of the patient and various kinds of
information are differently displayed on the LCD depending on the mode. The screen displayed
in each mode is shown below.
13
- Configuration of Screen in Pacer Mode
14
ECG & SpO2 Measurement Mode
15
2.4 Dependence and Independence of Internal Devices
The information on the dependence and independence of the internal and external components
of the LiFEGAIN CU-HD1 is as described below. When repairing the product, dependent parts
should be replaced together.
The classification of dependence and independence of boards, modules, and devices
constituting the product is shown in the figure. The components in the solid lines do not
influence other modules or boards; however, the main board and the therapy board should be
replaced together because they require calibration before shipment.
Since each instance that necessitates the replacement of the component is not additionally
explained in this Manual, the information described above should be remembered when
repairing the product, in order to solve any problems.
16
3. MAINTENANCE
3. Put the printer paper into the printer in the direction shown in the following figures, and
pull out some of the paper.
4. Push the printer cover back to its default setting, until it makes a "click" sound.
5. If the Rotary switch is set to the monitor mode, the green printer power lamp will be on.
You can use the “Feed” button to take as much printer paper as you want.
17
3.1.2.2 Battery Maintenance
Rechargeable batteries should always be maintained at a certain charge level, or above, in
case of a situation arising where AC power is unavailable. The charge level may be checked
using the Battery Capacity Indication Button at the back of the battery module.
CAUTION Avoid any short circuits caused by the insertion of metallic materials into
18
3.1.2.5 Maintenance of External Defibrillation Paddles
External Defibrillation Paddles are used in defibrillation therapy and require continuous
maintenance. When the External Defibrillation Paddles are used, their conductive gel should be
wiped off of the metallic electrodes. The External Defibrillation Paddles should also be wiped
clean before use, in the same manner. If the metallic plates dry with a coating of foreign
substances or conductive gel, the patient may get burnt when defibrillation therapy is performed,
and the contact between the External Defibrillation Paddles and the patient may not be
recognized.
3.1.3 Cleaning
For any device problems resulting from negligence of the below cleaning instructions, the free
repair service may not be applicable, even during the warranty period.
Keep the main body and cables away from dust and pollution, and under usual circumstances,
clean them with a soft cloth. There is a risk of device damage from forcible pressure or shock.
19
NOTICE When cleaning the product with a common cleaner, avoid possible corrosion of
the terminals and electrodes and melting of the resin of the main body through an
Maintenance Checklist
LiFEGAIN CU-HD1
Date
Schedule
Exterior Condition
Accessory Maintenance
Status
Error Correction or
Measures
Inspected by Signature
of Operator
20
4. PERFORMANCE VERIFICATION/INSPECTION
WARNING Use a Simulator/Analyzer that has been inspected and calibrated by authorized
personnel.
21
4.4.1 Self Test (Daily, Weekly, Monthly)
The self test is automatically performed by the LiFEGAIN CU-HD1, depending on the situation.
A description of the items in the self test is given below.
- This test is the same as the battery capacity test done during the "power on" self test.
- Checks the Charge button status of the device to prevent unintended recharging.
- Checks the Shock button status of the device.
22
CAUTION For a proper self test of the LiFEGAIN CU-HD1, the AC power module or the
battery module should be installed.
product.)
B. A window prompting for verification will appear. Please check it again.
The product cannot be used during the manual self test session. The following
section will give you the step-by-step testing guidelines.
C. If the mode is switched to the "Self Test" mode, product information (model name,
serial number), software version information, and the date when the self test was last
done, will be displayed on the screen.
D. System Test: A common system test is performed.
E. Mode Rotary Key Test: The rotary switch is checked. If the instruction window appears,
move the rotary switch to 150J of the manual defibrillator mode.
F. Charge Button Test: The operational status of the Charge button is checked. Press
the Charge button according to the instructions.
G. Shock Button Test: The operational status of the Shock button is checked. Press the
Shock button according to the instructions.
H. Audio Test: Checks the speaker functions. After the voice test, there is a process to
check whether to operate the voice function by pressing the menu selection switch.
I. Defibrillation Test: Checks the defibrillation functionality, including a test for the automatically
23
charging 200J and the delivery of defibrillation energy. To test the 200J defibrillation
operation, you must press the shock button, and use the checkup resistance to implement
the test. If there is no resistance or simulator connection to deliver the defibrillation energy,
the device discharges the internally charged energy, leading to failure.
J. Pacer Test: Pacer function is checked.
K. ECG Lead Test: The ECG measuring part (12 Lead ECG) is checked.
L. Battery Part A Test: Power terminal « A » is inspected.
M. Battery Part B Test (Battery Part B Test): Power terminal « B » is checked.
N. SpO2 Test: Inspects the functionality of the SpO2 measurement module.
O. Printer Test: Inspects the functionality of the printer.
If the above manual test function has been completed, a message window will appear, stating
that the device must be closed.
If the test results are indicative of a problem during the self test, solve the problem by referring
to the Manual Self Test in “Chapter 5 Troubleshooting.”
24
Error Item Code Cause of Error
Pacing Current No proper pacing current control communication with the
80
Control Error high-voltage board.
Charge Button When pressing the charge button, the line condition is
100
Error abnormal.
When pressing the shock button, the line condition is
Shock Button Error 200
abnormal.
The mode rotary is not operating properly.
Mode Rotary Error 400
(Is found only in the MST Mode Rotary Test.)
Audio Current is not normal.
Audio Current Error 800
(Is found only in the MST Audio Current Test)
ECG Board is not operating properly.
ECG Control Error 1000
(Is found only in the MST ECG Board Test)
SpO2 Module Error 2000 No proper communication with the SpO2 Module.
Impedance Error 4000 Stored Impedance is abnormal.
25
B. Mount the Battery Module.
C. Connect the defibrillation simulator and the LiFEGAIN CU-HD1 using pad connectors.
D. Set the defibrillation simulator to meet the energy calculation.
E. Turn the rotary switch to 200J so as to charge the defibrillation energy to 200J.
F. Push the “Charge” button to charge the defibrillation energy.
G. When charging is completed, push the “Shock” button to deliver the defibrillation
energy.
H. Check the delivered defibrillation energy.
The energy should be in the range of 200J±30J.
26
1. Automatic Disarm Test
A. Connect the pads with the defibrillation simulator.
B. Turn the rotary switch to 200J.
C. Push the “Charge” button to charge the defibrillation energy.
D. When the “Shock” button is turned on, indicate that charging with defibrillation
energy is finished for automatic discharge, and wait for 15 seconds.
E. Internal discharge may be identified by turning-off of the “Shock” button.
F. Check that defibrillation energy is not delivered to the defibrillation simulator.
Check if the value is zero (0) or that there is no change in the display of the
simulator.
A. Set the ECG waveform of the simulator as NSR (Normal Sinus Rhythm) and set
the electric potential as 1mV.
B. Connect the simulator and the main body using a 3-Lead, 5-Lead, or 12-Lead ECG
measuring cable depending on the kind of ECG to be measured.
C. Turn the rotary switch of the LiFEGAIN CU-HD1 to the ‘Monitor’ position.
D. Check the ECG displayed on the LCD screen.
Check whether the ECG is clearly displayed.
Check whether a normal waveform is displayed for each lead with changing
leads by pushing the lead selection buttons.
27
Check whether the HR (Heart Rate) is 30bpm.
E. Change the alarm setting to a minimum of 30bpm and check whether the alarm
sounds.
F. Re-check items D and E in terms of the ECG Waveform NSR at 60bpm of HR and
1mV of electric potential.
G. Re-check items D and E in terms of the ECG Waveform NSR at 200bpm of HR
and 1mV of electric potential. (Set the alarm’s maximum alarm generation HR to
no higher than 200bpm.)
H. Remove each lead from the simulator and check whether they recognize the
connection status of the ECG cable.
Check whether a removed lead is recognized, using the lead selection buttons.
I. Remove the ECG cable from the terminal of the product and check whether
removal of the ECG cable is recognized by the product.
Check whether a message, « Connect ECG Cable, » is displayed.
Check, when the ECG cable is connected to the upper part of the LCD, whether
the icon disappears.
J. Set the ECG Waveform of the simulator as Aystole.
K. Check whether the paper is loaded into the built-in printer of the LiFEGAIN CU-
HD1.
L. Print the ECG by pushing the “Print” button and check the printed result.
Check whether the baseline is clearly printed.
Check whether the items shown below are printed on the paper.
Details of print: Patient Information, Product Information, Information of Printed
Waveform, HR, Attributes of Printer, and Date & Time.
A. Remove the connection of the AC Power Module and connect the main body with
the battery only. (Pacer function may be checked when at least two columns’
capacity is secured on the battery when checking the battery.)
B. Connect the pads with the simulator equipped with an analysis function.
C. Connect the pads with the main body of the LiFEGAIN CU-HD1.
D. Turn the rotary switch to ‘Pacer Mode’.
E. Set pacer as shown below.
Pacer Setting: Fixed Mode, Rate of 70ppm, and Current of 30mA
F. Push “Start/Stop” soft button.
G. Check the result analyzed by the simulator.
Pacer Speed: 70ppm±1.5%
Current: 30±5mA
28
H. Set Pacer as shown below.
Pacer Setting: Fixed Mode, Speed of 180ppm, and Current of 200mA
I. Check the result analyzed by the simulator.
Pacer Speed: 180ppm±1.5%
Current: 200±5mA
A. Turn the rotary switch of the LiFEGAIN CU-HD1 to the ‘Monitor’ position.
B. Connect the main body and the SpO2 sensor with an SpO2 Cable.
C. Connect the sensor to a finger, or to the simulator, and perform the measurement. .
(Set the SpO2 as 98 %.)
Check whether the measurement curve of the SpO2 is clearly displayed.)
If the range of the SpO2 exceeds a 95%~100% range when measuring with a
finger, re-check the position of the finger by referring to the User’s Manual.)
Check whether the pulse and SpO2 are displayed on the LCD.
29
Soft Key Symbols & Description
AED Mode /
The charged electric shock energy is internally
Manual
discharged.
Defibrillation Mode
Monitor Mode
Starts the 12-lead mode.
(12-Lead ECG)
Monitor Mode
Stops the 12-lead mode.
(12-Lead ECG)
Monitor Mode
Transfers the 12-lead data.
(12-Lead ECG)
Transfers the information of the patient being examined
Monitor Mode to the computer in real time.
To see this, Bluetooth must be connected.
The symbols on the screen are common and show the accessories and specific status.
Display Symbols
30
Display Symbols
Indicates it is printing.
Indicates that the voice and ECG storage functions are being executed.
In AED Mode, the processes of AED and CPR are guided with the use of voice messages and
texts on the screen.
The voice and text messages to guide the AED process are shown below.
Indicates that the pads connector or the paddles adapter must be connected to the main body.
Attach pads.
Denotes that the disposable defibrillation electrode must be adhesively attached to the skin area where
the clavicle ison the right side of the chest, and the last rib bone on the left side, and the opposite
connector of the electrode must be accurately connected to the defibrillation electrode connector hole.
31
Indicates that contact with the patient must be avoided for an accurate ECG analysis.
Indicates that the patient's ECG signal is being analyzed. Do not touch the patient. Also, when
analyzing the ECG signal, if a patient makes any movements, the analysis may contain errors.
Shock advised.
Stand clear.
It means that a patient must avoid all contact with other people.
Charging
Indicates that the user must press the shock button for electric shock treatment.
Shock delivered.
No shock advised.
Indicates that a prompt to press the shock button was given, but that the shock button was not
(The device will discharge through the internal circuit after 15 seconds.)
Check pulse.
32
If there is no pulse, CPR must be started.
Breathe.
Indicates that there is no pulse in manual mode, and that the analyze button must be pressed.
A. Turn the rotary switch to the “OFF” position and stop the operation.
B. When the “Power” LED of the printer is turned off, keep the “FEED” button pushed
in.
C. When the “FEED” button of the printer is pushed in, turn the rotary switch of the
LiFEGAIN CU-HD1 to ‘Monitor’.
D. Check printed result.
When checking, refer to the test print result (Figure) shown below.
33
34
4.4.11 Bluetooth & Data Transmission Test
The information transmission function using Bluetooth may be checked by connecting Bluetooth
communication and actually transmitting information.
The process for sending information using Bluetooth is described below.
35
4.4.12 Menu Operation Test
The menu has a branch structure as shown below, and each menu is described in the table
below. Check whether the menus described below correspond with the product.
성
이름
ID
환자 정보 환자분류
나이
성(sex)
페이서 유/무
심박수
맥박
SpO2
알람
VT/VF
무수축
알람 일시정지 시간
자동분석
심폐소생술
데이터관리
음성녹음
볼륨
날짜와 시간
장비관리 필터
자가 진단 테스트
블루투스
심전도 크기
36
Menu Category Menu Screen
Main Menu
Patient
Information
Patient
Information
Patient
Information
Alarm
Main Menu
37
Menu Category Menu Screen
Alarm
Alarm
Alarm
Alarm
Alarm
38
Menu Category Menu Screen
Alarm
Printer
Automatic
Analysis
CPR
Device
Management
Main Menu
39
Menu Category Menu Screen
Device
Management
Device
Management
Device
Management
Device
Management
Device
Management
Device
Management
(Bluetooth
Main Menu)
40
Menu Category Menu Screen
Bluetooth
Device
Management
41
5. Troubleshooting
In this chapter, the symptoms and solutions in case of instances where errors are generated in
the modules configuring the LiFEGAIN CU-HD1 and each function are suggested.
Among the problems found during use, minor errors are not described in this chapter. For the
solutions of the problems that may occurr during operation of the product due to inexperienced
method of use, refer to the problem solving items in the User’s Manual.
The LiFEGAIN CU-HD1 checks the status of the product by itself and lets the user identify the
occurrence of a problem.
- Identification when an error is found in the Power-on Test
- Identification when an error is found in the Daily/Weekly/Monthly Self Tests
- Identification with use of the result of the manual self test - Pass or Fail
- Identification when an error is found during use of the product
If a problem is generated, a service lamp is turned on after all kinds of tests; if an error is found
in the self test, when the rotary switch is turned off, the mode changes to manual self test mode
and inspection of the major functions of the product is performed.
If the product fails to pass the manual self test, the product should be repaired.
Symptom at Occurrence of a
Inspection Item Troubleshooting
Problem
Battery Capacity - Occurrence of Low Battery - Charge the battery using AC power module or
Test: Error replace the battery with a charged battery.
- Use the product with AC power.
Button Functionality - Charge Button Error - Check whether the button is pushed through
Test: - Shock Button Error external pressing.
- Exterior of the button may be damaged; if not
damaged, replace the keyboard.
42
5.2 Periodic Automatic Self Test
The Periodic Automatic Self Test is performed daily, weekly, and monthly at the same time. The
Periodic Automatic Self Test is performed under conditions where power is supplied and the
rotary switch is set to the ‘Off’ position at 02:00 am. If the product is in use at the time that the
Periodic Automatic Self Test performs, the test is omitted.
- Daily Automatic Self Test: The test is performed on the same inspection items as those of the
Power-on Test.
SpO2 Test - SpO2 Module Error - Check whether SpO2 Module Cable is
43
5.3 Manual Self Test
MST may be performed to check the functions of LiFEGAIN and may also be performed to
check all the functions in the process of turning on and off the product when a problem is found
in the Periodic Automatic Self Test.
Symptom at Occurrence of a
Inspection Item Troubleshooting
Problem
System Test - System Stop / Abnormal - Send the product to the company to be
Operation repaired.
Charge Button Test - Failure in Charge Button Test - Check whether the button is pushed by
external pressing.
- Exterior of the Button may be damaged; if not
damaged, replace Keyboard.
Shock Button Test - Failure in Shock Button Test - Check whether the Button is pushed by
external pressing.
- Exterior of the button may be damaged; if not
damaged, replace the keyboard.
Defibrillation Test - Charge Error - Check the status and connection of the therapy
- Discharge Error board cable.
- If cable has no problem, replace therapy
board.
Pacer Test - Pacer Error - If therapy board cable has no problem, replace
therapy board.
ECG Lead Test - ECG Control Error - Check whether ECG board cable is damaged;
if damaged, replace it.
- Replace ECG board.
44
Symptom at Occurrence of a
Inspection Item Troubleshooting
Problem
SpO2 Test - SpO2 Error - Check the status and connection of SpO2
module cable.
- If cable has no problem replace SpO2 module.
Printer Test - Printer Failure - Check whether printer paper is loaded and
perform printer function test.
- If no problem is found, check the status and
connection of printer cable.
- If function test and cable connection shows no
problem, replace the printer.
Product
Button Key does not Incorrect Contact - Perform MST to check any problem with key.
self test
Charge/Shock button - Pushed status - Check whether charge button is pushed in.
does not work.
- Mechanical Damage - Check keyboard connection and cable status..
- Replace keyboard.
External Paddle is -Incorrect Fastening - Check the connection between external paddles
not recognized. and main body and, if fastened wrong, fasten
correctly.
45
Symptom Expected Cause Troubleshooting
External Paddle - Incorrect Fastening - Check the connection between external paddles
Charge/Shock button and main body and, if fastened wrong, fasten
does not work. correctly.
Internal Discharge is - Incorrect Attachment - Check the attachment status of pads and external
spontaneously done paddles.
in the status that If they are not attached within the permitted range,
Paddles and Pads internal discharge is done.
are attached.
- External Paddle - Clean the metallic electrode parts of external
Contamination paddles.
Defibrillation Energy - Charge Button Failure - Perform MST to check button status.
Charging is not (Manual Defibrillator - If needed, check the status of keyboard and
done. Mode) keyboard cable.
Low Charge Speed - Battery Capacity - Check battery level and, if insufficient, charge it.
46
Symptom Expected Cause Troubleshooting
Defibrillation Energy - Wrong Connection of - Replace the connector of the electric shock cable.
is not delivered. Electric Shock Cable
ECG Measurement - Incorrect Attachment of - Check the attachment status of ECG electrode on
is wrong. ECG Electrode. the skin.
- Incorrect Skin Status - If the skin to be measured is wet, dry the skin.
Lead Fault Message - Incorrect Attachment of - Check the attachment status of ECG electrode on
is repeatedly ECG Electrode. the skin.
generated.
- Incorrect Skin Status - If the skin to be measured is wet, dry the skin.
- ECG Cable Defect - As ECG cable may be cut, replace ECG cable.
47
5.7 SpO2 Problems and Solutions
Symptom Expected Cause Troubleshooting
Defect
Terminal Failure
SpO2 module.
Printer does not - Paper Jam - Check whether paper is jammed in the printer.
work.
- Cable Defect - Check printer cable connection.
Printing is not done. - Paper Placing Direction - If printer is set so that the paper moves in the
Error reverse direction, printing is not done; setting
should be checked.
- Printer Head Defect - Clean printer head part with a cotton bud as if
sweeping the part.
48
5.10 Problems in Information Transmission Function and Solutions
Symptom Expected Cause Troubleshooting
initialized. Manual.
- Bluetooth Setting Error - Contact A/S Center and ask Bluetooth connection
setting method.
Failure
Transmission is not - Connection Release - Retry Bluetooth connection according to User’s
done. Manual.
Failure
Prohibited.”
Connection
49
5.13 LCD Function Problems and Solutions
Symptom Expected Cause Troubleshooting
50
6. Repair
A/S Department
CU Medical Systems, Inc.
1647-1 Donghwa, Munmak,
Wonju, Gangwon, Republic of Korea
(Sales Department)
Tel +82-31-421-9700
Fax +82-31-421-9911
51
The tools and equipment required for repairing LiFEGAIN are as listed below:
- (+) Driver
- High-voltage energy discharge tool
(Is not supplied. If required, make a request. )
- A sharp tool to remove the labels and rubber from the product.
WARNING Before opening the product to commence repairs, all the power supplies, including the
AC Power Module and the Battery Module should be removed from the main body.
NOTICE 1. To repair the product, the person should have good knowledge and
understanding of the Service Manual and the product repair process.
Therefore, the equipment manager should keep a Service Manual and a
User’s Manual close at hand to be referred to during repair.
2. The problems that may occur during use of the product are described in the
User’s Manual. The solutions to the problems that may occur during repair
of the product are described in this Manual.
3. The components of the product may be different depending upon the
purchaser’s option. When evaluating the functions of the product, check the
functions installed in the product.
52
6.3 Product Configuration
The product is configured with devices consisting of labels, metal, and plastic injection-molded
materials, cables, connecting boards and accessories, parts to support fastening of devices,
and PCB boards and various accessories containing IC elements.
The connections of the various parts are shown in the figure below.
<Product Configuration>
53
The connections and names of parts are described below in detail. The front part containing the Paddle Cradle and the Main Board, and the rear part containing the
Therapy Board are shown.
54
<Front Part containing Therapy Board of
Main Body>
55
<Rear Part containing Therapy Board>
56
Parts List Table
57
6.4 Internal Assemblies Disassembly
The product disassembly process is described with respect to disassembling the entire modules,
including devices other than the disassembly of a specific module. Therefore, to disassemble a
specific module, the product disassembly process should be referred to.
2. Loosen the screws on the paddle cradle and separate the paddle cradle as shown in
the photos below.
58
3. Disassemble the screws from the back of the main body as shown in the photos. Under
the damage-preventive sticker, one more screw is inserted. Disassemble the product
as shown in the photos below.
Screw separation from power Input section Screw separation from upper section
59
5. Separate the front part and the rear part by pulling forward and backwards, respectively.
Do not pull with excessive power. If the screws are not completely removed, the front
part and the rear part are not separated.
6. Detach the AC adapter and the car power jack connection terminal from the power
board.
60
7. Separation of the front part and the rear part of the product.
8. Remove the screws from the power board and the ECG board.
61
9. Detach the power board cable.
62
11. Remove the screws from the left/right of the aluminum panel.
12. Detach the cable connecting the main board and the therapy board.
13. Photo showing the detached main board and therapy board
(Do not touch the therapy board with hands.)
63
14. Detach the LCD module flat cable and the LCD module power.
15. Remove the screws connecting the SpO2 module and the main board.
16. Loosen the screws of the main board and separate the main board.
17. Remove the screws fixing the aluminum plate and remove the aluminum plate.
64
18. Detach the screws fixing the LED onto the bracket.
65
20. Loosen the screws of the keyboard and detach the keyboard.
21. Discharge the high-voltage condenser using a high-voltage discharge device. The
arrow is shown below the high-condenser connection terminal and is used to ground
the discharge device.
66
22. Loosen the screws on the aluminum plate covering the high-voltage condenser and
remove the aluminum plate.
23. The speaker, the defibrillation paddle and pad connection terminal, the ECG cable
terminal, and the SpO2 probe connection terminal are shown in the photos below.
67
Separation of Built-In Printer
3. When pulling the built-in printer forward, it is detached from the main body. To detach
the built-in printer, the printer cable should be detached.
68
The kinds of connector cables connected with the PCB board and the connector terminals
constituting the product and their connection positions on the PCB are shown in the table below.
69
7. Operation
For a detailed operational method, refer to the Operator’s Manual. In this chapter, the description
of the operational method is summarized.
7.1 AED
AED Mode is applied to patients who are exhibiting the symptoms of SCA (Sudden Cardiac
Arrest) including ventricular fibrillation and ventricular tachycardia, in order to bring them back to
life by applying electric shock and restoring the normal ECG rhythm.
SCA patient means the person is showing no response, no movement, no normal breathing, and
no pulse.
When setting the product to AED Mode, it must be used by medical professionals or paramedics
with Class 2 or a higher level.
In the AED mode, the defibrillation pads must be used. The defibrillation pads work to obtain the
ECG signal of patients and deliver electric shock according to the patient’s condition.
The AED Mode’s function is to treat patients with acute cardiac arrhythmia, analyzing the ECG of
the patient to determine whether the patient is in the cardiac arrhythmia state, including ventricular
tachycardia and ventricular fibrillation. In this mode, the product provides you with voice and text
prompts.
70
level of electric shock energy from a range of 1-200 joules for the use of pads or paddles.
On the other hand, when used for synchronous cardio version, it is used on patients with the
symptoms of atrial fibrillation. It is designed to analyze the R-wave of the ECG QRS and apply the
R-wave synchronized electric shock.
In the manual mode, synchronous cardio version treatment can be applied to patients with rapid
atrial fibrillation, ventricular tachycardia, and cardiac ischemia.
7.3 Pacer
Pacing is a method applied to patients who have lost the natural cardiac movement functions, and
is mostly used on patients with bradycardia. The LiFEGAIN CU-HD1 supports noninvasive pacing,
a way of helping to maintain the patient’s pulse by attaching an electrode to the patient’s skin and
delivering artificial electric stimulation to the heart.
Pacing mode is divided into the fixed mode and the demand mode.
The currents that can be delivered from the pacing mode ranged from 5 to 200mA (±5mA); its
pacing rate is 30~180 PPM (±1.5%), the pacing time is 20ms, and the pacing wave appears in a
monophasic rectangular form.
71
8. Specifications
The specifications of the LiFEGAIN CU-HD1 are as shown below.
Body
Defibrillation Feature
▪ Semi-automatic
Operation Mode
▪ Manual: Sync, Asynchronous Defibrillation
Output Waveform
(Manual / Automatic)
72
Sensitivity & Specificity
ECG Test 90% One Sided
Minimum test Performance Shock No Shock Observed
Rhythm Rhythms
sample size goal
sample
Decision Decision Performance
Lower Confidence
Class size Limit
97.26%
Coarse >90%
SHOCKABL
81.02%
>75%
Fast VT 50 137 111 26 (111/137) 76%
sensitivity
sensitivity
Normal 100 100%
> 99%
Sinus minimum 100 0 100 (100/100) 97%
specificity
Rhythm (arbitrary) specificity
NON SHOCKABLE
AF,SB,S
VT,
heart 99.54%
30 > 95%
block, 219 1 218 (218/219) 98%
(arbitrary) specificity
idioven- specificity
tricular
PVC’s
96.21%
> 95%
Asystole 100 132 5 127 (127/132) 93%
specificity
specificity
1 1 1 1 1 1 1 1 ±1 J
2 2 2 2 2 2 2 2 ±1 J
3 3 3 3 3 3 3 3 ±1 J
4 4 4 4 4 4 4 4 ±1 J
5 5 5 5 5 5 5 5 ±2 J
6 6 6 6 6 6 6 6 ±2 J
7 7 7 7 7 7 7 7 ±2 J
8 8 8 8 8 8 8 8 ±2 J
9 9 9 9 9 9 9 9 ±2 J
10 10 10 10 10 10 10 10 ±2 J
15 15 15 15 15 15 15 15 ±3 J
20 20 20 20 20 20 20 20 ±3 J
30 30 30 30 30 30 30 30 ±15 %
50 50 50 50 50 50 50 50 ±15 %
70 70 70 70 70 70 70 70 ±15 %
100 100 100 100 100 100 100 100 ±15 %
120 120 120 120 120 120 120 120 ±15 %
150 150 150 150 150 150 150 150 ±15 %
170 170 170 170 170 170 170 170 ±15 %
200 200 200 200 200 200 200 200 ±15 %
73
Manual Defibrillation
▪ Within 7 seconds: When rechargeable battery is completely charged
▪ Within 6 seconds: When an AC power module is used (only when power is
more than 90%).
Charging Time
(200 Joules)
▪ Within 7 seconds: When the battery has been discharged more than 15
times after full charging.
▪ Within 7 seconds: When the battery has been discharged more than 15
times after replacement of battery module with a new one.
Shock
1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 100, 120, 150, 170, 200 Joules
Energy Selection
Indicator LCD for ECG display, Power/Error display indicators
▪ Text prompts for charge energy
Charging Indicator ▪ Beep when charging
▪ Shock button flashes orange
Patient Impedance
▪ 25~175 Ohms
Recognition Range
▪ Use SYNC button for synchronous cardioversion.
SYNC ▪ Analyze the patient's ECG signals and synchronize R-wave of QRS in
ECG with shock delivery within 60ms.
AED
▪ Within 7 seconds: When Battery Module is completely charged
▪ Within 7 seconds: When a new Battery Module is used
▪ Within 6 seconds: When an AC power module is used (only when power is
Charging Time more than 90%).
(200 Joules) ▪ Within 7 seconds: When the battery module has been discharged more than
15 times after full charging.
▪ Within 7 seconds: When the battery has been discharged more than 15 times
after replacement of battery module with a new one.
AED Energy 200 Joules, fixed
Text and Gives step-by-step guidelines to the user on how to take proper
Voice Prompts measures for an emergency situation.
AED Operation Key Analyze button, Stop Analysis button, Shock button, CPR Type 30:2 / 15:2,
and Button Start/Stop CPR
▪ Progress bar of the amount of energy charged and text prompts
Charging Indicator ▪ Beep when charging is finished
▪ Shock button flashes orange
▪ Analyzes the patient's ECG to determine whether defibrillation is needed or
Patient Analysis
not. (When the automatic patient analysis feature is enabled)
Defibrillation-needed
Ventricular Fibrillation or Fast Ventricular Tachycardia, 150bpm or above
rhythm
Algorithm Sensitivity
and Specification That AHA 2005 guidelines are met.
Requires Defibrillation
74
Pacer
Pacing Type Non-invasive Pacing
Energy Waveform Monophasic Rectangular
Pacing Mode Demand mode, Fixed mode
Energy Magnitude 5 ~200 mA (±5mA)
Pulse Width 20ms (±10%)
Pacing Rate 30 ~ 180 ppm (±1.5%)
Impedance Range
In Which Pacing is 25 ~ 175 Ohms
Possible
Operation Key and Mode button, Rate button, Print button, MENU key, HOME button, and
Button LEAD Selection button.
LCD for ECG display etc., text prompt, QRS detection display, patient monitoring
Indicator
information display, pacing signal delivery display.
Demand Mode Analyzes the patient ECG to determine whether to deliver pacing energy or not in
Analysis the Demand mode.
ECG Monitoring
▪ Kinds of ECG measurement: 3-Lead, 5-Lead, and 12-Lead
ECG Input
▪ Able to see ECG results using LCD or an external printer.
Detects when the ECG cable(s) is detached (if the ECG cable is
Lead Fault
disconnected from the patient or the device)
Heart rate display 30 ~ 300 bpm (Accuracy: ±3 bpm)
▪ Heart rate alarm setting range
▪ Lower Limit: 30~300 bpm (it should be set to a value lower than the
Heart rate Alarm Setting maximum, however)
▪ Upper Limit: 30~300 bpm (it should be set to a value higher than the
minimum, however)
▪ 5 mm/mV, 10 mm/mV, 20 mm/mV
ECG Size
▪ Auto: 0.3 ~ 5.5 mV, Display inputted ECG signals as 10mm on the screen.
▪ Emergency: 1 ~ 30 Hz (-3 dB)
▪ Monitor: 0.5 ~ 40 Hz (-3 dB)
Frequency Range
▪ Diagnosis: 0.05 ~ 150 Hz (-3 dB)
▪ 50Hz, 60Hz Notch Filter
Patient Isolation
CF Type
(Defibrillation Check)
SpO2
Pulse Rate 20 ~ 250 bpm (± 3 bpm)20 ~ 250 bpm (± 3 bpm)
SpO2 값SpO2 value 70 ~ 100 % (± 3 digit) 70 ~ 100 % (±3 digit)
Perfusion 0.2%
▪ Lower Limit: 70% ~ 100% (it should be set to a value lower than the
maximum, however)
SpO2 Alarm Setting
▪ Upper Limit: 70% ~ 100% (it should be set to a value higher
than the minimum, however)
Display Update Interval 6 secs.
75
Resolution 1%
SpO2 Measuring Sensor: Nellcor Sensor (DS100A Sensor)
660 nm (Red), 890 nm (Infrared) 660 nm (Red), 890 nm
Sensor Light
(Infrared)
Power Consumption 15mW or below
Backlight LED Life Time 20,000 hours (time when brightness is reduced to 50%)
Data Storage
External Storage
Store more than 192 hours of events and ECG data.
SD Card
Or, store more than 8 hours of events, ECG data, and voice data
(if 1GB)
Output ECG directly from LiFEGAIN CU-HD1 or output ECG data by
CG Data Print
transferring it through Bluetooth communication.
Printer
Thermal line
Print Method
printing
2
Resolution 203dpi X 406dpi (dpi: dot per inch )
Print Width 48mm
Print Rate 25mm/sec
Feed Rate About 62.5mm/second
7.2 V DC
Input Power Power consumption in a standby state: 70mA (maximum power
consumption: 2.4A)
5 °C ~ 40 °C
Operation Temperature
Humidity: 30%~85%, Non-condensing
-10 °C ~ 50 °C
Storage Temperature
Humidity: 30%~90%, Non-condensing (without printer papers)
Printer paper
Type Roll Type
Width: 58mm
Size
Role Size: Minimum diameter 40mm
76
Bluetooth
Communication
Within 30m (Based on open space)
distance
Operation
10°C ~ 55°C (Humidity: 90%, Non-condensing)
Temperature
Transmission Method:
FHSS (Frequency Hopping Spread Spectrum)
Miscellaneous Modulation Method:
GFSK (Gaussian-filtered Frequency Shift Keying)
Battery
Battery Type Lithium Polymer
14.8 VDC
Output
3100 mAh
-20°C ~ 45°C
Storage Temperature
Humidity: 90% or below, Non-condensing
77
AC Power Module, SMPS
18 VDC, 5 A
Output
12 VDC, 0.5 A
-20°C ~ 40°C
Operation Temperature
Humidity: 90% or below, Non-condensing
-20°C ~ 60°C
Storage Temperature
Humidity: 90% or below, Non-condensing
Cigar Jack
-20°C ~ 40°C
Operation Temperature
Humidity: 90% or below, Non-condensing
-20°C ~ 60°C
Storage Temperature
Humidity: 90% or below, Non-condensing
AC Adapter
Output 12V/3.6A
-20°C ~ 40°C
Operation Temperature
Humidity: 90% or below, Non-condensing
-20°C ~ 60°C
Storage Temperature
Humidity: 90% or below, Non-condensing
78
Appendix 1
Defibrillator Shock Waveform
79
Defibrillating waveform at 200J to 75-ohm load
80
Defibrillating waveform at 200J to 150-ohm load
81
Appendix 2
Pacing Waveform
82
Pacing waveform at 200mA to 125-ohm Pacing waveform at 200mA to 150-ohm
load load
83
Appendix 3
Additional Notice
Manufacturer
CU Medical Systems, Inc.
1647-1 Donghwa, Munmak,
Wonju, Gangwon, Republic of Korea
+82-33-747-7657
84