Professional Documents
Culture Documents
Affirm
Affirm
6 months
Geometry calibration – Verify that the geometry calibration was completed successfully.
Make sure that all labels are readable and properly affixed.
Verify the ID sensors correctly identify the Affirm system when installed.
QAS test for standard needle approach – Make sure the Affirm system is installed correctly and all
paddles are removed.
QAS test for standard needle approach – Make sure that ‘Affirm QAS’ appears in the Device field.
QAS test for standard needle approach – Verify that the target coordinates are within +/- 1 mm or X, Y,
and Z numbers as shown on the Diff values line or the Biopsy Control Module.
QAS test for standard needle approach – (for all unexposed views) Verify that the target coordinates are
within +/- 1 mm or X, Y, and Z numbers as shown on the Diff values line or the Biopsy Control Module.
QAS test for standard needle approach – (Tomosynthesis option: For the Tomo Biopsy QAS view, select
Manual exposure mode, 30 kV, 30 mAs, and Aluminum filter.) Verify that the target coordinates are
within +/- 1 mm or X, Y, and Z numbers as shown on the Diff values line or the Biopsy Control Module.
QAS test for lateral needle approach – Make sure the Affirm system is installed correctly and all paddles
are removed.
QAS test for lateral needle approach – Make sure that ‘Affirm Lateral QAS’ appears in the Device field.
QAS test for lateral needle approach – Verify that the target coordinates are within +/- 1 mm or X, Y, and
Z numbers as shown on the Diff values line of the Biopsy Control Module.
QAS test for lateral needle approach – (Move the Lateral Arm assembly to the right side of the Affirm
system.) Verify that the target coordinates are within +/- 1 mm or X, Y, and Z numbers as shown on the
Diff values line of the Biopsy Control Module.
1 year