Professional Documents
Culture Documents
Uritest-50 Training Manual (2020!08!20 02-50-44 UTC)
Uritest-50 Training Manual (2020!08!20 02-50-44 UTC)
Uritest
Training Manual
1
1 Introduction
1.1 Appearance
1
4
2
2 Installation
2.1 Unpacking and Inspection
Check the contents of the accessories according to the packing list.
Standard Component List
3
2.4 Strip Holder Installation
Gently push the holder in the entrance until the end of the holder parallels to the instrument
edge.
NOTE
OTE::
1. The bar code scanner option is only available for the single test mode
mode.
2. The bar code can only be inputed in the testing interface which displays after clicking the start
key.
4
2.6 Power Connection
3 Operation Instruction
3.1 Start Up
After finishing the Self-checking, the analyzer get into the standby interface as the following
picture shows:
On the Ready screen, you may see the icon or displayed on the upper of the screen:
5
Sensitivity has been adjusted. Refer to System, Sensitivity Setting.
NOTE:
There should be no urine strip or other stuff on the strip holder, when turning on the power
switch. Otherwise the system will warn that the strip holder is not clean.
3.2 Set up
Setup
Press the MENU key to get into the MAIN MENU screen:
3. 2.1 System
3.2.1
Press System key to get into the System setting menu.
In Manual menu,you can adjust the sensitivity for each testing item.
For Positive
Positive, "+" denotes increasing reflectance ratio and decreasing cutoff value, "-"denotes
decreasing reflectance ratio and increasing cutoff value.
For Linear
Linear, "+" denotes increasing the reflectance ratio of high concentration, "-"denotes
decreasing the reflectance ratio of high concentration.
6
NOTE:
The adjustable range is from “+10 +10”” to “-10
-10””. This figure is just a relative value of adjustable
rang, not a percentage. It's recommended that operator collect enough QC test data before
percentage.It's
using the function of sensitivity adjustment, and check the conformance with clinical test after
adjustment.
Press following keys to adjust the sensitivity:
[+]: Increases the reflectance rate
[-]: Reduces the reflectance rate
[← →]: ]: Switches between Positive and Linear
Displays the next item to be adjusted. Keeping pressing it for 3 seconds will save the
adjustment and exit.
NOTE:
1. The function is not recommended for use.
3. If the icon displays on the Ready screen, it means the sensitivity is not in default. The
system will prompt user to resume the default each time turning on the instrument. Press [YES] to
reset the sensitivity to default; press [NO] to preserve the current setting.
3.2.2 Initialize
Pressing [OK] will clear all test results stored in memory and reset all features to default;
pressing [DATA] will only clear test results stored in memory.
NOTE:
All records will be cleared up when initializing the instrument. Please select it carefully.
3.2.3 RS232
BAR CODE:
The bar code ID will also be shown on the report of internal printer, RS 232 exporting and
LPT exporting, bur not be stored in the instrument memory, i.e., there won't be bar code ID in the
Databank screen.
7
3. 2.4 SN(Serial Number)
3.2.4
Press the SN key form the Ready screen to display the SN screen to set the serial number. The
serial number will always be reset to 000001 each time turning on the instrument.
3.2.5 Strip
Press the STRIP key from the Ready screen to display the Strip Setting screen to set the strip
mode. There three strip modes namely,
NOTE:
All pads should be within the scale range marked on the holder as shows in the following
picture:
Scale
8
3.4 Calibration
You should use the Uritest check strip to perform calibration for the analyzer frequently. There
are two check strips that come with the analyzer. One is for daily use, and the other for standby use.
Compare the check strip results with the reference value. If the results are consistent, it means the
analyzer is in order and can be used for routine testing. If not, use the other check strip for
calibration again. If the results are not consistent yet, check whether malfunction occurs and solve it.
It is recommended that control solution testing be run under the following situations:
1. When using a new bottle of reagent strips.
2. When replacing the operator.
3. When you question about the test results.
Please contact the manufacturer to get the control solution.
4 Maintenance
4.1 General Cleaning
Keep the instrument clean and dust-free. If cleaning is needed, use wet fabric or soft detergent.
Any oil, ester, silica gel and lubricant is forbidden for use. Use only soft fabric to clean the screen.
Water, ketene and aromatic solvent are forbidden for use.
4.2 Daily Cleaning
Clean the instrument at least one time everyday. If uncleanness affects test results, clean the
instrument at any time. If there is too much urine dirt on the strip holder s groove or flank, the test
result might not be exact. In this case, remove out the holder and clean as below:
1. Switch off the power supply. Gently pull out the holder toward operator.
2. Clean the holder with water and dry it with soft, lint-free cloth.
3. After cleaning, gently push the holder in the entrance until the end of the holder parallels to the
instrument edge.
4. Switch on the power supply. The instrument will automatically adjust the holder to its proper
position.
NOTE:
DO NOT wipe across the white calibration block on the top end of the holder. If dust exists,
use a cotton swab, wetted with plain water, to clean the block.
9
5 System Introduction
5.1 Mainboard Structure
10
6 Troubleshooting
6.1 Power on with No Response
1. Check the voltage of TP6 (DC 12± 0.4V).
2. Check whether the power adapter, power input plug, power switch, power input connector and
power input cable are normal, and replace the defective parts.
11
Picture 6-2 Display Trouble
6.3 Reagent Strip Detector Is Abnormal
1. Check the voltage of 4 pin on the detector output connector
2. If reagent strip is on the holder, the voltage is more than 3.5V
3. If reagent strip is not on the holder, the voltage is less than 1.5V
12
Picture 6-4 Time Trouble
6.5 Light Intensity Trouble
Trouble -1
1. Reference light is too strong(Trouble Trouble-4
-1) or reference light is too weak(Trouble-4
Trouble-4):
a. Adjust the LED 1 potentiometer to make sure the voltage of Tp3 is within the normal range
(DC3V~4.7V)
b. If the voltage is normal, while the same error persists, contact the distributor.
Trouble-2
2. Red light is too strong(Trouble-2 Trouble-5
Trouble-2) or red light is too weak(Trouble-5
Trouble-5):
a. Adjust the LED 2 potentiometer to make sure the voltage if Tp4 is within the normal range (DC
4.6V 0.1V)
b. If the voltage is normal, while the same error persists, contact the distributor.
Trouble-3
3. Green light is too strong(Trouble-3 Trouble-6
Trouble-3) or green light is too weak(Trouble-6
Trouble-6):
a. Adjust the LED 3 potentiometer to make sure the voltage if Tp5 is within the normal range
(DC3.8V 0.1V)
b. If the voltage is normal, while the same error persists, contact the distributor.
13
3. If the same error persists, contact the distributor.
14
Appendix A Test Principle
The optical-electronic sensor system is composed of a light source and a light receptor . The
light from the light source falls on the reagent pads on the strip. The absorbance and reflectance
vary according to the color of the reagent pad, i.e. if the reagent pad is darker, more light is
absorbed, and less light is reflected, vice versa. The degree of color development is proportional to
the concentration of analyte in urine.
The reflected light is transmitted into the optical-electronic sensor system, which transforms
the optical signal into electrical signal. The strength of the electrical signal correlates with
reflectance. After that, the electric signals are processed by CPU following transformed by I/V
converter, and the test results are printed out.
15
Appendix B (Symbols and Dat a)
Data)
16
Appendix C Troubleshooting Tables
17
Table-2 Troubles and Remedy
18