Professional Documents
Culture Documents
Clinical Chemistry Analyzers, Automated
Clinical Chemistry Analyzers, Automated
Purpose
Clinical chemistry analyzers determine the concen-
tration of certain metabolites, electrolytes, proteins,
and/or drugs in samples of serum, plasma, urine,
cerebrospinal fluid, and/or other body fluids. Auto-
mated chemistry analyzers provide the laboratory
with walkaway operation and a standardized method
of obtaining accurate and reproducible chemical con-
centration values.
By using separate reaction vessels for each test,
discrete clinical chemistry analyzers reduce the
2 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
C433UN3A
Each beam is transmitted through light guides to one
of eight photometer stations, where it is directed at the
moving cuvette track. Readings are taken at the ap-
propriate wavelengths at each station. The light that
passes through the cuvette is measured by photomul- Cuvette Area
tiplier tubes; the measurements are sent to a micro-
processor for analysis. If a sample’s analyte
concentration exceeds the procedure’s range of linear-
ity, the system notifies the operator, who can retrieve
the sample and rerun it in the dilute mode: the ana- Reagent Compartments Temporary Seals
lyzer withdraws one-fourth of the original sample vol- Permanent Seal
ume, adds more buffer, and repeats the test.
Figure 1. Reagent pack used with the Dade analyzers
These analyzers also have ISEs to measure concen-
trations of Na+, K+, Cl-, and lithium (Li+); four liquid
reagent channels are available for user-defined tests. form a 1 cm pathlength cuvette. Absorbance is meas-
ured, the concentration value for each test is calcu-
Another type of discrete clinical chemistry analyzer lated, and the results are printed; the used test packs
uses a special plastic test pack containing the reagents are automatically discarded into a waste container.
for one test and a double-walled reaction chamber that
This system uses ISEs to measure Na+, K+, and Li+
forms the cuvette for photometric analysis (see Fig. 1).
concentrations. For example, Li+ concentrations in a
On the top of each test pack is a rigid header, a plastic
sample are determined using the interaction of Li+
block used to move the pack through its processing
with an ionophore (a molecule that increases cell mem-
steps by means of a transport chain. The header may
brane permeability to specific ions) contained in a
incorporate a chromatographic column for removing
polyvinyl chloride membrane. The sample is aspirated
interfering substances or a gel-filtration matrix for
and delivered directly into the electrode measuring
retarding small molecules. Temporary seals surround
chamber, and the results are printed on a report form.
the compartments containing the tablet or liquid re-
agents; the reaction chamber itself is surrounded by a One analyzer uses fluorescence polarization (FP) for
permanent seal. the measurement of therapeutic drug levels and for
thyroid-function tests. With FP, the patient sample, the
The sample is introduced into the unit, followed by
FP sample dilution reagent, and the assay-specific an-
the reagent packs in the order in which the tests are
tibody are mixed together. Assay-specific tracer, a
to be performed. The sample is automatically aspi-
drug-fluorescein conjugate, is then added to the mix-
rated and delivered into each test pack, filtering
ture and binds to the assay-specific antibody. When
through the column to remove interfering substances
exposed to polarized light of the appropriate wave-
and proteins. The built-in computer reads an identifi-
length, a fluorescent molecule — a fluorophore — will
cation code on the pack header, which indicates the
absorb some of the light to emit it at a subsequently
test to be performed, its sample size, diluent type and
longer wavelength. Smaller molecules, such as free
volume, incubation time, and required photometric
analyte-tracer, can rotate rapidly, leading to a greater
wavelengths. The individual packets move on a trans-
depolarization of emitted light. Larger molecules, such
port chain through the temperature-controlled proc-
as the analyte-tracer bound to an antibody, rotate more
essing chamber, and applied pressure breaks the
slowly and emit light of lesser depolarization. Thus, as
temporary seals on the reagent compartments to add
the concentration of analyte increases, more unbound
the necessary reagents to the reaction vessel. The
analyte-tracer will be present, resulting in greater de-
packet contents are mixed by vibration and trans-
polarization at increasing analyte concentrations.
ported through delay stations, where the reaction
takes place. Once incubation is complete, the last re- In addition to performing FP analyses, this system
agent compartments are broken, the contents are re- also uses photometric and ISE methods to measure
mixed, and the packet is moved to the filter concentrations of enzymes, drugs of abuse, electro-
photometer, where it is mechanically compressed to lytes, and various other compounds.
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 3
Healthcare Product Comparison System
Reported problems
Although less of a problem than in continuous-flow
systems, carryover can occur on a sample probe or
reagent probes, in reusable cuvettes, or — in systems
using disposable cuvettes — as a result of residual
specimen in the pipetting system.
4 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
slide reagents and test packs should also determine • Requiring relatively little time and training to learn
appropriate waste disposal requirements. to operate, based on the degree of sophistication of
the equipment and the level of experience, compe-
Analyzers that require a water reagent hookup may
tence, and training of the laboratory personnel who
sometimes have water contamination problems — the
will operate it
presence of ions such as Na+ and Cl-, for instance —
that can affect test results. Special water treatment In addition to these basic requirements, other ana-
systems that feed directly into the analyzer may be lyzer features that should be considered are walkaway
necessary. capability, preferred report format, bar-code reading
capability, the ability to use different sample tube sizes
Some reagents/reagent slides must be stored at low
and run different types of samples, automated sample
temperatures (i.e., frozen or refrigerated). Prepara-
dilution, onboard reagent stability, and closed-tube
tion time must include thawing or warming before
sampling (which reduces risk of operator exposure to
use. Many analyzers have onboard refrigeration for
biohazardous materials). ECRI recommends that all
reagents.
instruments have alarms that can alert users to any
Problems occur with some analyzers’ software pack- system failure (e.g., insufficient reagents, out-of-range
ages; however, these are generally corrected by en- errors) that can produce false results.
hancements released by the manufacturer.
Buyers should be aware that an analyzer’s through-
Problems with mislabeled reagent containers, pack- put depends on the number and type of tests performed
age inserts, standards, and/or calibrators are fre- by the instrument during a particular time. Very high
quently reported. Labeling misinformation can cause throughput claims by a manufacturer may be based on
inaccurate or false chemistry results, which can lead tests that react rapidly to the addition of chromagenic
to misdiagnoses and inappropriate patient care. Users reagent (i.e., have short incubation times) and quickly
are advised to be alert for any recalls involving misla- show measurable results. Manufacturers should reveal
beled reagents. the test profile — the number and kinds of tests —
used to determine analyzer throughput claims.
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 5
Healthcare Product Comparison System
The high-complexity classification covers more spe- use a water reagent supply. Analyzer water costs can
cific testing, which requires extensive education and range from a relatively small amount to a significant
training in the specialty. Procedures that necessitate portion of an instrument’s overall operating costs,
a high degree of operator preparation, calibration, depending on water use and local water prices.
intervention, and analysis, such as clinical cytogenics
and histopathology applications, are placed in the An effective QC system is an important cost-con-
highly complex category. tainment consideration. An inappropriate system can
result in unnecessary test runs, delayed patient re-
Before purchasing new equipment or upgrading ex- sults, and failed proficiency tests, all of which waste
isting equipment, laboratories should thoroughly in- time and money. If the correct QC system is used, a
vestigate the CLIA regulations that apply to their large percentage of such waste can be eliminated. The
facility and to the devices being considered. In certain cost and expiration date of QC material and the fre-
situations, purchasing or upgrading a device may quency of test runs can significantly affect the overall
change the complexity category of the procedures. This reagent and analyzer costs.
could require additional staff training and certifica-
tion, as well as changes in QC, proficiency testing, QA State-of-the-art computer-driven instruments can
programs, and other laboratory procedures. become obsolete in a very short time if they cannot be
adapted for future advances in hardware development
With clinical chemistry analyzers, an important or accept software upgrades. Analyzers that have the
consideration is the system’s computer interface capa- flexibility to accept these kinds of improvements have
bilities. The effectiveness of the interface with the a long-term advantage over those that do not. By the
existing LIS or the hospital’s central computer system same token, analyzers that have open channels can be
is crucial to inputting test data, verifying testing accu- programmed to perform new analytical tests as they
racy, and maintaining QC, calibration, proficiency are developed by the instrument manufacturer.
testing, and patient files according to CLIA guidelines.
Although CLIA does not mandate computerized re- Because chemistry analyzers entail ongoing main-
porting systems in hospital laboratories, it does re- tenance and operational costs, the initial acquisition
quire facilities to have a system in place to ensure cost does not accurately reflect the total cost of owner-
compliance with CLIA performance standards for QC ship. Therefore, a purchase decision should be based
and QA of patient testing instruments and procedures. on issues such as life-cycle cost (LCC), local service
An effective LIS interface is a fast and efficient way to support, discount rates and non-price-related benefits
manage the large volume of test data that a laboratory offered by the supplier, and standardization with ex-
generates each day, as well as a convenient method of isting equipment in the department or hospital (i.e.,
organizing and storing data needed to comply with purchasing all analyzers from one supplier).
CLIA and other inspection agencies’ requirements.
(See the Product Comparison titled INFORMATION SYS- An LCC analysis can be used to compare high-cost
TEMS, LABORATORY for more information.) alternatives and/or to determine the positive or nega-
tive economic value of a single alternative. For exam-
Cost containment ple, hospitals can use LCC analysis techniques to
examine the cost-effectiveness of leasing or renting
Hospitals should perform a cost assessment of the
equipment versus purchasing the equipment outright.
instrument’s operating features, such as its testing
Because it examines the cash-flow impact of initial
reliability and accuracy, cost per reportable result, stat
acquisition costs and operating costs over a period of
time, average downtime, QC and calibration methods
time, LCC analysis is most useful for comparing alter-
and frequency, start-up procedures and time, labor,
natives with different cash flows and for revealing the
consumables, maintenance, and service costs. Also in-
total costs of equipment ownership. One LCC tech-
cluded in this assessment should be nonoperational
nique — present value (PV) analysis — is especially
but essential expenses such as those for instrument
useful because it accounts for inflation and for the time
installation (e.g., some analyzers require special
value of money (i.e., money received today is worth
plumbing and/or water filters), the disposal costs for
more than money received at a later date). Conducting
solid and liquid waste generated by the unit, and costs
a PV/LCC analysis often demonstrates that the cost of
for reagent preparation and storage.
ownership includes more than just the initial acquisi-
In addition to reagents and disposables such as tion cost and that a small increase in initial acquisition
cuvettes, reaction cups, and ISEs, the cost of consu- cost may produce significant savings in long-term op-
mables should also include an estimate of the amount erating costs. The PV is calculated using the annual
of water used over a specified time by analyzers that cash outflow, the dollar discount factor (the cost of
6 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
capital), and the lifetime of the equipment (in years) in Alternative 1 — Outright Purchase
a mathematical equation.
Capital Costs
The following analyses are based on approximately • System = $164,000
40,000 tests per month on one analyzer with one full-
time employee (FTE). Suppliers frequently offer vari- Total Capital Costs = $164,000
ous purchase/lease options on their equipment. Three Operating Costs
common options on chemistry analyzers are outright
purchase, monthly lease, and reagent contracts. Pur- • Service contract, years 2 through 5 = $16,400/year
chase usually requires initial capital for acquisition of • Salary and expenses for 1 FTE = $40,000/year
the equipment, ongoing costs for reagents and consu-
mables, and purchase of a service contract, which is • Reagents = $100,000/year
typically 8% to 12% of the list price for the system. A • Consumables = $15,000/year
lease can involve monthly payments that include a
service contract and additional ongoing costs for re- Total Operating Costs = $155,000 for year 1;
agents and consumables. A reagent contract usually $171,400/year for years 2 through 5
does not entail an initial acquisition fee or a separate
service contract fee but does require an agreement to PV = ($947,844)
purchase all reagents from the supplier on a cost-per-
Alternative 2 — Reagent Contract
test (CPT) or cost-per-reportable-test(CPRT) basis; the
cost generally depends on the number and type of tests Capital Costs
that will be performed per year. Equipment and serv- • System = No charge
ice costs are usually included in the CPT/CPRT calcu-
lation. Prices for the various arrangements may vary Total Capital Costs = $0
greatly depending on facility size, number of units Operating Costs
being purchased, and previous experience with the
supplier. Prices for reagents and consumables may be • Salary and expenses for 1 FTE = $40,000/year
fixed for a certain number of years (e.g., two years) and
• Reagents and consumables = $250,000/year for
then increase up to 4% annually. Individual suppliers
years 1 and 2
should be contacted to find out what options are avail-
able for a specific facility. • Reagents and consumables = $257,500 for year 3
Disposables are a significant part of the operating • Reagents and consumables = $265,225 for year 4
cost, and that cost can vary greatly depending on • Reagents and consumables = $273,182 for year 5
supplier discounts and incentives. ECRI’s
PriceGuide™ service benchmarks the price you pay for Total Operating Costs = $290,000/year for years 1
single-use medical products. PriceGuide compares and 2, $297,500 for year 3, $305,225 for year 4, and
your hospital’s current pricing data with the national $313,182 for year 5
or regional average and lowest recorded prices paid.
For more information, contact ECRI. PV = ($1,387,038)
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 7
Healthcare Product Comparison System
Other costs not included in the above analyses that Buyers should also negotiate for a nonobsolescence
should be considered for budgetary planning include clause stating that the supplier agrees not to introduce
the following: a replacement system within the next one or two years
and that if a replacement system is introduced during
• Software upgrades not included under warranty or
this time period, 100% of the purchase price can be
in service contract
applied to the purchase of the new system.
• Utilities (e.g., water, installation)
In addition, given the current highly competitive
As illustrated by the above sample PV/LCC analy- market for clinical chemistry equipment, hospitals
ses, the initial acquisition cost is only a fraction of the should negotiate for a significant discount — some
total cost of operation over five years. Therefore, rather suppliers may discount up to 20% or 30%. The actual
than making a purchase decision based solely on the discount received will depend on the hospital’s negoti-
acquisition cost of a clinical chemistry system, buyers ating skills, the system configuration and model to be
should consider operating costs over the lifetime of the purchased, the number of units being purchased, pre-
equipment. vious experience with the supplier, and the extent of
Also, it can be concluded that the best option for this concessions granted by the supplier, such as extended
sample situation is outright purchase. However, dif- warranties, fixed prices for annual service contracts,
ferent options may be more economical in other situ- and guaranteed on-site service response. Buyers
ations, depending on a hospital’s needs and budget. should make sure that applications training is in-
Each facility must have a specific analysis done based cluded in the purchase price of the system. Some
on its own situation to determine its most effective suppliers offer more extensive on-site or off-site train-
choice. ing programs for an additional cost.
ECRI recommends that buyers consider the number
For further information on PV/LCC analysis, cus-
and types of tests performed before deciding on a
tomized analyses, and purchase decision support,
specific system configuration. Also, if multiple analyz-
readers should contact ECRI’s SELECT™ Group.
ers are being purchased, hospitals should consider the
Purchasers should carefully evaluate the manufac- types of systems and capabilities that need to be pur-
turer’s service, which should include 24-hour hotline chased to avoid paying for unnecessary analysis pack-
access. They should also inquire about the terms of the ages. For instance, a hospital may want to purchase
warranty (e.g., some warranties may not include pre- two analyzers: one dedicated to routine general chem-
ventive maintenance visits), as well as what service istry profiles and one used for stat testing. In this case,
contracts are available, what they include, and what purchasing both analyzers from one supplier could
discounts are available if the contracts are purchased result in a significant discount. Standardization of
at the same time as the instrument. Hospitals can equipment can make staff training easier, simplify
purchase service contracts or service on a time-and- servicing and parts acquisition, and provide greater
materials basis from the supplier. Service may also be bargaining leverage when negotiating new equipment
available from a third-party organization. purchases and/or service contract costs.
Most suppliers provide routine software updates, Current analyzer users are also valuable sources of
which enhance the system’s performance, at no charge information on the quality, reliability, and overall ef-
to service contract customers. Furthermore, software ficiency of the instruments. When considering an in-
updates are often cumulative; that is, previous soft- strument, buyers should ask analyzer manufacturers
ware revisions may be required in order to install and to supply an unedited list of their customers.
operate a new performance feature. Also, many suppli-
ers do not extend system performance and uptime Stage of development
guarantees beyond the length of the warranty unless
New models of discrete clinical chemistry analyzers
the system is covered by a service contract.
are continually being introduced that offer more varie-
ECRI recommends that, to maximize bargaining ties of tests, greater stat testing capabilities, higher
leverage, hospitals negotiate pricing for service con- throughput, and step-saving features, such as direct
tracts before the system is purchased. Additional serv- sampling from the primary tube. Discrete clinical
ice contract discounts may be negotiable for chemistry analyzers continue to offer greater automat-
multiple-year agreements purchased up front at the ion (largely through software enhancements), require
time of acquisition or for service contracts that are less direct specimen handling, and involve fewer steps
bundled with contracts on other analyzers in the de- for operation and maintenance. Most analyzers fea-
partment or hospital. ture bar codes for sample and/or reagent identification.
8 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
Additional enhancements to components such as sam- Griffiths J. Automation and other recent develop-
ple trays include separate quadrant pieces that allow ments in clinical chemistry. Am J Clin Pathol 1992
continuous addition of samples, as well as adapters Oct;98(4 Suppl 1):S31-4.
that conform to cups, sample tubes, and dilution tubes.
Liu-Allison LY, Hui RA, Schwenzer KS. Analysis of
Data management systems play an important role lidocaine on the Cobas™ Integra. Ann Clin Lab Sci
in chemistry analyzers and many other kinds of auto- 1997;27(2):105-15.
mated analytical devices in the clinical laboratory.
These systems provide a bidirectional interface be- Ng RH. More on CK-MB [tips on technology]. MLO
tween the instrument and the LIS or hospital informa- Med Lab Obs 1991 Nov;23(11):11-2.
tion system and often link two or more kinds of
Palmer SM, Kaufman RA, Salamone SJ, et al. Cobas™
equipment to these computer systems, helping to coor-
Integra: clinical laboratory instrument with con-
dinate the ordering of tests to various instruments in
tinuous and random-access capabilities. Clin Chem
the laboratory and the reporting of test results to
1995;41(12):1751-60.
several locations in the facility.
A significant new development is complete clinical Plaut D, Schaefer L. A timesaving QC planning and
laboratory automation using robotics technology. A design tool that uses QAP data. Am Clin Lab 1996
wide range of systems is available; some automate the Jan-Feb:6-8.
entire laboratory, while others merely automate cer-
Solsky RL. Ion-selective electrodes. Anal Chem 1990
tain procedures. Systems can be configured to meet the
Jun 15;62(12):21R-33R.
needs and testing volumes of individual laboratories.
Labs may purchase devices for just front-end sample U.S. Department of Health and Human Services. Clini-
handling or conveyor-belt systems with mobile robots cal Laboratory Improvement Amendments of 1988
to completely automate almost all the laboratory pro- (CLIA). Fed Regist 1992 Feb 28;57(40):7002-288.
cedures. Many analyzer manufacturers provide adap-
tations on their units to accommodate laboratory Standards and guidelines
robotics.
Note: Although every effort is made to ensure that the
Automation can be used for a variety of reasons; following list is comprehensive, please note that other
some labs may want to increase test volume and applicable standards may exist.
throughput without adding staff. Others may use ro-
botics to reduce staffing costs. Each facility’s situation American Association for Clinical Chemistry, Inc.
will determine the type of system (if any) that is most Guidelines for providing quality stat laboratory
appropriate. Automation can be very expensive in- services. Laboratory Quality Assurance Committee.
itially but may contribute to cost savings in the long 1987.
run. There is little information currently available on American Board of Clinical Chemistry. Standards,
the cost versus savings of total laboratory automation. requirements, and procedures for certification.
More information is continually being gathered and 1996.
published; facilities should carefully investigate the American National Standards Institute. Preparation
benefits and drawbacks of the available systems and of manuals for installation, operation, and repair of
configurations before making a decision. laboratory instruments [standard]. ANSI/NCCLS
ASI-1-1981. 1981.
Bibliography
Specification for low-level protocol to transfer mes-
Bertholf RL, Savory MG, Winborne KH, et al. Lithium sages between clinical laboratory instruments and
determined in serum with an ion-selective elec- computer systems [standard]. ANSI/ASTM E1381.
trode. Clin Chem 1988 Jul;34(7):1500-2. 1996.
Burtis CA, Ashwood ER. Tietz fundamentals of clinical American National Standards Institute/Association
chemistry. 5th ed. Philadelphia: WB Saunders; 2001. for the Advancement of Medical Instrumentation.
Safe current limits for electromedical apparatus
Felder RA. Laboratory automation: strategies and pos- [standard]. 3rd ed. ANSI/AAMI ES1-1993. 1985 (re-
sibilities. Clin Lab News 1996 Mar;22(3):10-1. vised 1993).
Geary TD. Understanding analytical systems technol- American National Standards Institute/Underwriters
ogy for selective analysis. J Auto Chem 1988 Oct- Laboratories. Standard for safety for laboratory
Dec;10(4):163-87. equipment. ANSI/UL 1262-1989. 1990.
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 9
Healthcare Product Comparison System
American Society for Testing and Materials. Specifica- NCCLS — The Clinical Laboratory Standards Organi-
tion for transferring information between clinical zation. Assessment of the clinical accuracy of labo-
instruments and computer systems [standard]. ratory tests using receiver operating characteristic
ASTM Subcommittee E31.1400 on Clinical Labora- (ROC) plots [guideline]. GP10-A. 1995.
tory Instrument Interface. E1394-97. 1991.
Cholesterol: approved summary of methods and ma-
Canadian Society for Medical Laboratory Science. terials credentialed by the NRSCL Council [guide-
Laboratory safety [guideline]. 4th ed. 1996. line]. RS3-A. 1988.
Canadian Standards Association. Laboratory equip-
ment [standard]. C22.2 No. 151-M1986 (R1992). Clinical laboratory safety [guideline]. GP17-A. 1996.
1986 (reaffirmed 1992).
Clinical laboratory waste management [guideline].
Safety requirements for electrical equipment for GP5-A. 1993.
measurement, control, and laboratory use, part 1:
general requirements [standard]. CAN/CSA-C22.2 Internal quality control testing — principles and
No. 1010.1-92. 1992. definitions [guideline]. C24-A. 1991.
International Electrotechnical Commission. Medical Laboratory instruments and data management sys-
electrical equipment — part 1: general requirements tems: design of software user interfaces and end-
for safety [standard]. IEC 60601-1 (1988-12). 1988. user software systems validation, operation, and
Medical electrical equipment — part 1: general re- monitoring [guideline]. GP19-A. 1995.
quirements for safety. Amendment 1 [standard].
IEC 60601-1-am1 (1991-11). 1991. Preliminary evaluation of quantitative clinical labo-
ratory methods [guideline]. EP10-A. 1998.
Medical electrical equipment — part 1: general re-
quirements for safety. Amendment 2 [standard]. Preparation and testing of reagent water in the
IEC 60601-1-am2 (1995-03). 1995. clinical laboratory [guideline]. 3rd ed. C3-A2. 1997.
Medical electrical equipment — part 1: general re-
quirements for safety. Section 1. Collateral standard: Procedures for the handling and processing of blood
safety requirements for medical electrical systems. specimens [guideline]. 2nd ed. H18-A. 1990 (revised
IEC 60601-1-1 (1992-06). 1992. 1999).
Medical electrical equipment — part 1: general re- Protection of laboratory workers from instrument
quirements for safety. Section 1. Collateral standard: biohazards and infectious diseases transmitted by
safety requirements for medical electrical systems. blood, body fluids, and tissue [guideline]. M29-A.
Amendment 1 [standard]. IEC 60601-1-1-am1 (1995- 1997.
11). 1995.
Medical electrical equipment — part 1: general re- U.S. Department of Health and Human Services. Food
quirements for safety. Section 2. Collateral standard: and Drug Administration. Clinical chemistry and
electromagnetic compatibility — requirements and clinical toxicology devices. 21 CFR 862. 2000.
tests. IEC 60601-1-2 (2001-09). 2001.
Safety requirements for electrical equipment for U.S. Department of Health and Human Services.
measurement, control, and laboratory use — part 1: Health Care Financing Administration. Laboratory
general requirements [standard]. IEC 61010-1 requirements. 42 CFR 493. 2000.
(1990-09). 1990.
Standards and certification: laboratory require-
Safety requirements for electrical equipment for ments. 42 CFR 493. 1998.
measurement, control, and laboratory use — part 1:
general requirements. Amendment 1 [standard].
IEC 61010-1-am1 (1992-09). 1992.
Safety requirements for electrical equipment for Citations from other ECRI publications
measurement, control, and laboratory use — part 1:
general requirements. Amendment 2 [standard]. Health Devices
IEC 61010-1-am2 (1995-07). 1995.
National Registry of Certified Chemists. Standards for What’s happening to clinical laboratory testing? [clini-
certification. 1995. cal perspective]. 1995 May-Jun;24(5-6):168.
10 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 11
Healthcare Product Comparison System
Biotecnica Hycel
Biotecnica Instruments SpA [175052] Hycel Diagnostics [175047]
via Licenza 18 4 rue Galvani
I-00156 Roma F-91745 Massy Cedex
Italy France
Phone: 39 (06) 4112316 Phone: 33 (1) 64531515
Fax: 39 (06) 4103079 Fax: 33 (1) 64531520
E-mail: bt@biotecnica.it E-mail: hycel.com@wanadoo.fr
Internet: http://www.biotecnica.it Internet: http://www.hycel-diagnostics.com
CGA
Instrumentation
CGA Strumenti Scientifici SpA [336997]
via Luca Giordano 7/B Instrumentation Laboratory Co [101922]
I-50132 Firenze FL 101 Hartwell Ave
Italy Lexington MA 02421-3125
Phone: 39 (055) 571476 Phone: (781) 861-0710, (800) 955-9525
Fax: 39 (055) 5000889 Fax: (781) 861-1908
E-mail: support@ilww.com
Dade Behring Internet: http://www.ilus.com
Dade Behring Asia Ptd Ltd [329466] Instrumentation Laboratory GmbH [283451]
15 McCallum Street Klausnerring 4
#04-01 Natwest Centre D-85551 Kichheim bei Muenchen
Singapore 069045 Germany
Republic of Singapore Phone: 49 (89) 907070
Phone: 65 2967787 Fax: 49 (89) 90907116
Fax: 65 2966478 E-mail: support@ilww.com
Internet: http://www.dadebehring.com Internet: http://www.ilww.com
12 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 13
Healthcare Product Comparison System
14 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 15
Healthcare Product Comparison System
16 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
When reading the charts, keep in mind that, unless comparison among models. However, keep in mind
otherwise noted, the list price does not reflect supplier that exchange rates change often.
discounts. And although we try to indicate which
features and characteristics are standard and which Need to know more?
are not, some may be optional, at additional cost.
For further information about the contents of this
For those models whose prices were supplied to us Product Comparison, contact the HPCS Hotline at +1
in currencies other than U.S. dollars, we have also (610) 825-6000, ext. 5265; +1 (610) 834-1275 (fax); or
listed the conversion to U.S. dollars to facilitate hpcs@ecri.org (e-mail).
About ECRI . . .
ECRI is a nonprofit health services research agency and a Collaborating Center of the World Health
Organization, providing information and technical assistance to the healthcare community to support
safe and cost-effective patient care for more than 25 years. The results of ECRI’s research and
experience are available through its publications, information systems, databases, technical assis-
tance program, laboratory services, seminars, and fellowships.
Our full-time staff includes a wide range of specialists in healthcare technology, hospital admini-
stration, financial analysis, risk management, and information and computer science, as well as
hospital planners, attorneys, physicists; biomedical, electrical, electronic, chemical, mechanical, and
registered engineers; physicians; basic medical scientists; epidemiologists and biostatisticians; and
writers, editors, and communications specialists.
Underlying ECRI’s knowledge base in healthcare technology are its integrity and objectivity. ECRI
accepts no financial support from medical product manufacturers, and no employee may own stock
in or consult for a medical equipment or pharmaceutical company.
The scope of ECRI’s resources extends far beyond technology. ECRI keeps healthcare professionals,
manufacturers, legal professionals, information specialists, and others aware of the changing trends
in healthcare, healthcare standards and regulations, and the best ways to handle environmental and
occupational health and safety issues. ECRI also advises on management issues related to healthcare
cost containment, accreditation, risk management, human resources, quality of care, and other
complex topics.
ECRI has more than 35 publications, databases, software, and services to fulfill the growing need
for healthcare information and decision support. They focus on three primary areas: healthcare
technology, healthcare risk and quality management, and healthcare environmental management.
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 17
Healthcare Product Comparison System
SPECIAL/TIME, min
Acid phosphatase Yes No No Yes
Ethanol No Yes Yes Yes
Aldolase No No No No
Amylase Yes Yes Yes Yes
C3 Yes No No No
C4 Yes No No No
Cholinesterase Yes No No Yes
CK-MB Yes Yes Yes Yes
CSF protein Yes User defined User defined Yes
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes Yes Yes
18 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF Yes Yes Yes Yes
SAMPLE TRAY CAPACITY 231 150 150 78
THROUGHPUT,
20 test profiles/hr Up to 100 Profile dependent Profile dependent 280
REAGENT TYPE Liquid, ready to use Liquid Liquid Liquid
REAGENT DELIVERY Automatic Automatic with level Automatic with level Automatic
sensing sensing
REAGENT SUBSTITUTION Yes No No Yes
REAGENT PREPARATION On 2 assays No No No
AUTO DILUTION Yes Yes Yes Yes
AUTO QUALITY CONTROL Yes Yes Yes Yes
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 19
Healthcare Product Comparison System
AUTO VERIFICATION Not specified Not specified Not specified Not specified
AUTO CALIBRATION Yes Yes Yes Yes
CALIBRATION FREQ Varies Test dependent, up Test dependent, up 2 weeks
to 30 days to 30 days
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Closed tube Closed tube Yes
WATER REQUIREMENTS,
L/min Up to 0.75 None None 0.03
Onboard supply Yes NA NA Yes
DISPLAY CRT, printer Color CRT, 8.5 x 11" CRT, 8.5 x 11" CRT, printer
report printer report printer
LIGHT SOURCE Tungsten-halogen Xenon flash tube Xenon flash tube Tungsten-halogen
PREVENTIVE MAINT
FREQUENCY 3/year 1/year 1/year 4/year
H x W x D, cm (in) 111 x 189 x 112 40 x 69 x 57 61 x 69 x 65.5 42 x 65 x 100
(43.7 x 74.4 x 44.1) (15.8 x 27.2 x 22.5) (24 x 27.2 x 25.8) (16.5 x 25.6 x 39.4)
20 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
ASSAY METHODS End point, kinetic End point, ISE, End point, kinetic, End point, rate,
kinetic rate indirect ISE
BASIC/TIME, min
Albumin Yes Yes Yes No
ALP Yes Yes Yes No
ALT Yes Yes Yes No
Amylase Yes Yes Yes No
AST Yes Yes Yes No
BUN Yes Yes Yes Yes
Ca++ Yes Yes Yes ISE
Cholesterol Yes Yes Yes No
CK Yes Yes Yes No
Cl- Yes Yes Yes Yes
CO2 Yes Yes Yes Yes
Creatinine Yes Yes Yes Yes
Direct and total
bilirubin Yes Yes Yes No
Glucose Yes Yes Yes Yes
Inorgan phosphorus Yes Yes Yes No
Iron Yes Yes Yes No
K+ Yes Yes Yes Yes
LDH Yes Yes Yes No
Mg Yes Yes Yes No
Na+ Yes Yes Yes Yes
Total protein Yes Yes Yes Yes
Triglycerides Yes Yes Yes No
Urea Yes Yes Yes No
Uric acid Yes Yes Yes No
Others Ammonia, TIBC, LDL Ammonia, TIBC, LDL See footnote ** None specified
SPECIAL/TIME, min
Acid phosphatase Yes Yes Yes No
Ethanol Yes Yes Yes No
Aldolase Yes Yes Yes No
Amylase Yes Yes Yes No
C3 Yes Yes Yes No
C4 Yes Yes Yes No
Cholinesterase Yes Yes Yes No
CK-MB Yes Yes No No
CSF protein Yes Yes No Yes
GGT Yes Yes Yes No
HDL cholesterol Yes Yes Yes No
IgA Yes Yes Yes No
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 21
Healthcare Product Comparison System
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF Yes Yes Yes Yes
SAMPLE TRAY CAPACITY 55 55 84 42 (uses centrifug-
able sectors)
22 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
AUTO VERIFICATION Not specified Not specified Not specified Yes, with DataLink
AUTO CALIBRATION Yes Yes Yes No
CALIBRATION FREQ 2-4 weeks 2-4 weeks Test dependent 24 hr, 7 days
(total protein)
POWER FAILURE
CALIBRATION MEMORY Yes Yes Calibration, patient Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes Yes
WATER REQUIREMENTS,
L/min 0.05 0.05 25 L/hr Not specified
Onboard supply Yes Yes Yes No
DISPLAY PC PC CRT, Epson printer CRT, printer
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 23
Healthcare Product Comparison System
BASIC/TIME, min
Albumin Yes Yes Yes Yes
ALP Yes Yes Yes Yes
ALT Yes Yes Yes Yes
Amylase Yes Yes Yes Yes
AST Yes Yes Yes Yes
BUN Yes Yes Yes Yes
Ca++ Yes Yes ISE Yes
Cholesterol Yes Yes Yes Yes
CK Yes Yes Yes Yes
Cl- No : Yes Yes Yes Yes
CO2 No : Yes Yes Yes No
Creatinine Yes Yes Yes Yes
Direct and total
bilirubin Yes Yes Yes Yes
Glucose Yes Yes Yes Yes
Inorgan phosphorus Yes Yes Yes Yes
Iron Yes Yes Yes Yes
K+ No : Yes Yes Yes No
LDH Yes Yes Yes Yes
Mg Yes Yes Yes Yes
Na+ No : Yes Yes Yes No
Total protein Yes Yes Yes Yes
Triglycerides Yes Yes Yes Yes
Urea Yes Yes Yes Yes
Uric acid Yes Yes Yes Yes
Others IBCT, direct HDL, IBCT, direct HDL, IBCT, direct HDL, Copper, lipase
lipase lipase lipase
SPECIAL/TIME, min
Acid phosphatase No No No Yes
Ethanol Yes Yes Yes Yes
Aldolase No No No No
Amylase Yes Yes Yes Yes
C3 No No No Yes
C4 No No No Yes
Cholinesterase Yes Yes Yes Yes
CK-MB Yes Yes Yes Yes
CSF protein Yes Yes Yes Yes
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes Yes Yes
24 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
THROUGHPUT,
20 test profiles/hr Not specified Not specified Not specified 75
REAGENT TYPE Liquid Liquid Liquid Liquid
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 25
Healthcare Product Comparison System
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes Yes
WATER REQUIREMENTS,
L/min Not specified Not specified 0.6 0.01
Onboard supply Not specified Not specified Not specified Yes
DISPLAY Flat-panel LCD Flat-panel LCD Flat-panel LCD CRT, printer
monitor, printer monitor, printer monitor, printer
LIGHT SOURCE Pulsed xenon lamp Pulsed xenon lamp Pulsed xenon lamp Tungsten-halogen
26 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
SPECIAL/TIME, min
Acid phosphatase Yes Yes Yes Yes
Ethanol Yes Yes Yes Yes
Aldolase No Yes Yes Yes
Amylase Yes Yes Yes Yes
C3 Yes Yes Yes Yes
C4 Yes Yes Yes Yes
Cholinesterase Yes Yes Yes Yes
CK-MB Yes Yes Yes Yes
CSF protein No Yes Yes Yes
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes Yes Yes
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 27
Healthcare Product Comparison System
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF No No No No
SAMPLE TRAY CAPACITY 45 70 70 90
THROUGHPUT,
20 test profiles/hr 70 250 480 with ISE 500 with ISE
REAGENT TYPE Liquid Liquid Liquid Liquid
28 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes Yes
WATER REQUIREMENTS,
L/min 0.01 0.05 0.05 0.05
Onboard supply Yes Yes Yes Yes
DISPLAY CRT, forms printer CRT, forms printer Color CRT, forms LCD color STF,
printer 12.1" VGA
LIGHT SOURCE Tungsten-halogen Dichroic lamp Dichroic lamp Dichroic lamp
COMPUTER INTERFACE Bidirectional RS232C Bidirectional RS232C Bidirectional RS232C Bidirectional RS232C
PREVENTIVE MAINT
FREQUENCY 2/year 3/year 3/year 2/year
H x W x D, cm (in) 47 x 58 x 53 45 x 100 x 50 45 x 100 x 50 68 x 100 x 57
(18.5 x 22.8 x 20.9) (17.7 x 39.4 x 19.7) (17.7 x 39.4 x 19.7) (26.8 x 39.4 x 22.4)
LINE POWER, VAC 200-240, 110 200-240, 110 200-240, 110 200-240, 110
POWER CONSUMPTION 200 W 350 W 450 W 450 W
PURCHASE INFORMATION
List price,
std configuration $22,000 $45,000 $65,000 $75,000
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 29
Healthcare Product Comparison System
PROGRAMMED TESTS 50 95 57 62
USER-DEFINABLE TESTS Yes No Yes Yes
OPTICAL SYSTEM Photometric Photometric, Photometric, Photometric,
bichromatic, bi/polychromatic, bi/polychromatic,
turbidimetric turbidimetric turbidimetric
ASSAY METHODS End point, kinetic End point, rate End point, ISE, End point, ISE,
kinetic, rate kinetic, rate, het-
erogeneous immuno
BASIC/TIME, min
Albumin Yes Yes Yes Yes
ALP Yes Yes Yes Yes
ALT Yes Yes Yes Yes
Amylase Yes Yes Yes Yes
AST Yes Yes Yes Yes
BUN Yes Yes Yes Yes
Ca++ Yes Yes Yes Yes
Cholesterol Yes Yes Yes Yes
CK Yes Yes Yes Yes
Cl- Yes Yes Yes Yes
CO2 Yes Yes Yes Yes
Creatinine Yes Yes Yes Yes
Direct and total
bilirubin Yes Yes Yes Yes
Glucose Yes Yes Yes Yes
Inorgan phosphorus Yes Yes Yes Yes
Iron Yes Yes Yes Yes
K+ Yes Yes Yes Yes
LDH Yes Yes Yes Yes
Mg Yes Yes Yes Yes
Na+ Yes Yes Yes Yes
Total protein Yes Yes Yes Yes
Triglycerides Yes Yes Yes Yes
Urea Yes No No Yes
Uric acid Yes Yes Yes Yes
Others Ammonia, TIBC, LDL, TIBC, ammonia TIBC BUN, TIBC, NPT-HDL
turbidimetric meth-
ods
SPECIAL/TIME, min
Acid phosphatase Yes Yes Yes Yes
Ethanol Yes Yes Yes Yes
Aldolase Yes No No No
Amylase Yes Yes Yes Yes
C3 Yes No No No
C4 Yes No No No
Cholinesterase Yes No No No
CK-MB Yes Yes Yes Yes
CSF protein Yes Yes Yes Yes
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes No No
30 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF Yes Yes Yes Yes
SAMPLE TRAY CAPACITY 44 14 60 60
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 31
Healthcare Product Comparison System
POWER FAILURE
CALIBRATION MEMORY Yes Yes UPS UPS
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING No Yes Yes Yes
WATER REQUIREMENTS,
L/min Not specified Not specified 0.03-0.07 0.03-0.07
Onboard supply Yes Yes Yes Yes
DISPLAY Printer Thermal printer, Thermal printer, Thermal printer,
color CRT ** color CRT ** color CRT
LIGHT SOURCE Tungsten-halogen Tungsten-halogen Tungsten-halogen Tungsten-halogen
LINE POWER, VAC 110/220 115/230, 50/60 Hz 115/230, 50/60 Hz 115/230, 50/60 Hz
POWER CONSUMPTION 500 W Not specified Not specified Not specified
PURCHASE INFORMATION
List price,
std configuration $10,500 Not specified Not specified Not specified
32 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
ATAC 6000 ATAC 8000 Lisa 300 Plus Lisa 500 Plus
ASSAY METHODS EIA, end point, EIA, end point, ISE, End point, kinetic, End point, kinetic
kinetic, rate, fixed kinetic, rate, fixed ISE optional direct potentio-
time, sample BLK time, sample BLK metric, ISE
BASIC/TIME, min
Albumin Yes Yes 3 3
ALP Yes Yes 5 5
ALT Yes Yes 5 5
Amylase Yes Yes 7 7
AST Yes Yes 5 5
BUN Yes Yes 3 3
Ca++ Yes Yes 5 5
Cholesterol Yes Yes 12 12
CK Yes Yes 7 7
Cl- Yes Yes 1 1
CO2 Yes Yes 5 5
Creatinine Yes Yes 5 5
Direct and total
bilirubin Yes Yes 7 7
Glucose Yes Yes 10 10
Inorgan phosphorus Yes Yes 5 5
Iron Yes Yes 7 7
K+ Yes Yes 1 1
LDH Yes Yes 5 5
Mg Yes Yes 5 5
Na+ Yes Yes 1 1
Total protein Yes Yes 3 3
Triglycerides Yes Yes 10 10
Urea Yes Yes 3 3
Uric acid Yes Yes 5 5
Others TIBC TIBC, microalbumin, TIBC TIBC
albumin, micropro-
tein
SPECIAL/TIME, min
Acid phosphatase Yes Yes 7 7
Ethanol No No 7 7
Aldolase No No 10 10
Amylase Yes Yes 10 10
C3 No No 5 5
C4 No No 5 5
Cholinesterase No No 10 10
CK-MB Yes Yes 5 5
CSF protein No No 7 7
GGT Yes Yes 7 7
HDL cholesterol Yes Yes 10 10
IgA No No 7 7
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 33
Healthcare Product Comparison System
ATAC 6000 ATAC 8000 Lisa 300 Plus Lisa 500 Plus
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF No Yes Yes Yes
SAMPLE TRAY CAPACITY 45 70 96 96
THROUGHPUT,
20 test profiles/hr Not specified 18 8.3 10
REAGENT TYPE Liquid Liquid Liquid Liquid
34 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
ATAC 6000 ATAC 8000 Lisa 300 Plus Lisa 500 Plus
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING No Yes Yes Yes
WATER REQUIREMENTS,
L/min Not specified 0.02 0.006 L/min 0.007 L/min
Onboard supply Yes Yes ~21-day supply 20-day supply
DISPLAY CRT, printer CRT, printer Color screen, Color screen,
printer printer
LIGHT SOURCE Halogen Halogen Halogen Halogen
COMPUTER INTERFACE Bidirectional RS232 Bidirectional RS232 Bidirectional RS232C Bidirectional RS232C
PREVENTIVE MAINT
FREQUENCY 4/year Cycle dependent 1/year 1/year
H x W x D, cm (in) 47 x 58 x 53 47 x 99.1 x 52.1 190 x 307 x 183 190 x 307 x 183
(18.5 x 23 x 21) (18.5 x 39 x 20.5) (74.8 x 120.8 x 72) (74.8 x 120.8 x 72)
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 35
Healthcare Product Comparison System
ASSAY METHODS End point, ISE, End point, End point, ISE, End point, reaction
kinetic kinetic, rate kinetic, rate rate
BASIC/TIME, min
Albumin Yes Yes Yes Yes
ALP Yes Yes Yes Yes
ALT Yes Yes Yes Yes
Amylase Yes Yes Yes Yes
AST Yes Yes Yes Yes
BUN Yes Yes Yes Yes
Ca++ Yes Yes Yes Yes
Cholesterol Yes Yes Yes Yes
CK Yes Yes Yes Yes
Cl- Yes Yes Yes Yes
CO2 Yes Yes Yes No
Creatinine Yes Yes Yes Yes
Direct and total
bilirubin Yes Yes Yes Yes
Glucose Yes Yes Yes Yes
Inorgan phosphorus Yes Yes Yes Yes
Iron Yes Yes Yes Yes
K+ Yes Yes Yes Yes
LDH Yes Yes Yes Yes
Mg Yes Yes Yes Yes
Na+ Yes Yes Yes Yes
Total protein Yes Yes Yes Yes
Triglycerides Yes Yes Yes Yes
Urea Urea nitrogen Urea nitrogen Urea nitrogen Yes
Uric acid Yes Yes Yes Yes
Others Acid phosphatase, Acid phosphatase, Acid phosphatase, Ammonia, UIBC, LDL
CK-MB, lipase, TIBC CK-MB, lipase, TIBC CK-MB, lipase, TIBC
SPECIAL/TIME, min
Acid phosphatase Yes Yes Yes Yes
Ethanol Yes Yes Yes No
Aldolase No No No Yes
Amylase Yes Yes Yes Yes
C3 Yes Yes Yes Yes
C4 Yes Yes Yes Yes
Cholinesterase Yes Yes Yes Yes
CK-MB Yes Yes Yes Yes
CSF protein Yes Yes Yes No
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes Yes Yes
36 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF Yes Yes Yes Not specified
SAMPLE TRAY CAPACITY 75 150 150 84
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 37
Healthcare Product Comparison System
AUTO VERIFICATION Not specified Not specified Not specified Not specified
AUTO CALIBRATION Yes Yes Yes Yes
CALIBRATION FREQ Test dependent Test dependent Test dependent Not specified
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes Yes
WATER REQUIREMENTS,
L/min 0.3 0.3 0.5 0.41
Onboard supply Yes Yes Yes Yes
DISPLAY Color CRT, printer Color CRT, printer Color CRT, printer CRT, printer
PREVENTIVE MAINT
FREQUENCY 2/year 4/year 4/year 2/year
H x W x D, cm (in) 118 x 98 x 76 118 x 152 x 101.5 118 x 223 x 101.5 116 x 148 x 93
(46.5 x 38.6 x 29.9) (46.5 x 59.8 x 40) (46.5 x 87.8 x 40) (45.7 x 58.3 x 36.6)
WEIGHT, kg (lb) 300 (661.5) 550 (1,212.5) 850 (1,873.9) 600 (1,322.8)
38 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
SPECIAL/TIME, min
Acid phosphatase Yes Yes No Yes
Ethanol No No No No
Aldolase Yes Yes No No
Amylase Yes Yes Yes Yes
C3 Yes Yes No No
C4 Yes Yes No No
Cholinesterase Yes Yes No No
CK-MB Yes Yes No CK, yes; MB, no
CSF protein No No No No
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Direct method Yes
IgA Yes Yes No No
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 39
Healthcare Product Comparison System
SAMPLE TYPE
Serum and plasma Yes Yes Yes Serum
Urine Yes Yes No No
CSF Not specified Not specified No No
SAMPLE TRAY CAPACITY 84 84 1 48
THROUGHPUT,
20 test profiles/hr 1,650 2,250 63 12-13
REAGENT TYPE Liquid Liquid Dry Liquid, powder
40 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes No
WATER REQUIREMENTS,
L/min 0.41 0.5 None 0.0007
Onboard supply Yes Yes NA Yes
DISPLAY CRT, printer CRT, printer LCD PC monitor, printer
PREVENTIVE MAINT
FREQUENCY 2/year 2/year 1/year Not specified
H x W x D, cm (in) 116 x 148 x 93 116 x 170 x 93 16 x 20 x 33 61 x 61 x 35.6
(45.7 x 58.3 x 36.6) (45.7 x 66.9 x 36.6) (6.3 x 7.9 x 13) (24 x 24 x 14)
WEIGHT, kg (lb) 600 (1,322.8) 650 (1,433) <5 (<11) 29.5 (65)
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 41
Healthcare Product Comparison System
ASSAY METHODS End point, kinetic, End point, kinetic, End point, kinetic, End point, kinetic,
immunoassay, coagu- ISE ISE ISE
lation time, ISE *
BASIC/TIME, min
Albumin Yes Yes Yes Yes
ALP Yes Yes Yes Yes
ALT Yes Yes Yes Yes
Amylase Yes Urine Urine Urine
AST Yes Yes Yes Yes
BUN Yes Yes Yes Yes
Ca++ Yes Urine Urine Urine
Cholesterol Yes Yes Yes Yes
CK Yes Yes Yes Yes
Cl- Yes Urine Urine Urine
CO2 Yes Yes Yes Yes
Creatinine Yes Urine Urine Urine
Direct and total
bilirubin Yes Yes Yes Yes
Glucose Yes Urine Urine Urine
Inorgan phosphorus Yes Urine Urine Urine
Iron Yes Yes Yes Yes
K+ Yes Urine Urine Urine
LDH Yes Yes Yes Yes
Mg Yes Urine Urine Urine
Na+ No Urine Urine Urine
Total protein Yes Urine Urine Urine
Triglycerides Yes Yes Yes Yes
Urea Yes Urine Urine Urine
Uric acid Yes Urine Urine Yes
Others None specified Lipase, UIBC, Lipase, UIBC, Lipase, UIBC,
ammonia, lactate, ammonia, lactate, ammonia, lactate,
direct HDL, direct direct HDL, direct direct HDL, direct
LDL LDL LDL
SPECIAL/TIME, min
Acid phosphatase Yes Yes Yes Yes
Ethanol Yes Yes Yes Yes
Aldolase Yes No No No
Amylase Yes Yes Yes Yes
C3 Yes Yes Yes Yes
C4 Yes Yes Yes Yes
Cholinesterase Yes No No No
CK-MB Yes Yes Yes Yes
CSF protein Yes Yes Yes Yes
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes Yes Yes
42 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF Yes Yes Yes Yes
SAMPLE TRAY CAPACITY 48 80 150 300
THROUGHPUT,
20 test profiles/hr Not specified 20 samples 45 samples 89 samples
REAGENT TYPE Liquid Liquid Liquid Liquid
REAGENT DELIVERY Automatic, level Automatic pipetting Automatic pipetting Automatic pipetting
detection
REAGENT SUBSTITUTION Yes Yes Yes Yes
REAGENT PREPARATION No No No No
AUTO DILUTION Yes Yes Yes Yes
AUTO QUALITY CONTROL Yes Yes Yes Yes
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 43
Healthcare Product Comparison System
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes Yes
WATER REQUIREMENTS,
L/min 0.002 : 1.5/500 test 0.43 0.67 1.1
Onboard supply Yes Yes Yes Yes
DISPLAY Video, SVGA, color CRT, printer CRT, printer CRT, printer
printer
LIGHT SOURCE Tungsten-halogen Tungsten-halogen Tungsten-halogen Tungsten-halogen
PREVENTIVE MAINT
FREQUENCY 2/year 2/year 2/year 2/year
H x W x D, cm (in) 40 x 81 x 54 121.9 x 144.8 x 76.2 127 x 188 x 81 126 x 200 x 114
(15.7 x 31.9 x 21.3) (48 x 57 x 30) (50 x 74 x 31.9) (49.6 x 78.7 x 44.9)
44 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 45
Healthcare Product Comparison System
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF Yes Yes Yes Yes
SAMPLE TRAY CAPACITY 300 300 40 40
THROUGHPUT,
20 test profiles/hr 178 samples 267 samples Not specified Not specified
REAGENT TYPE Liquid Liquid MicroSlide Dry chemistry slides
46 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
AUTO VERIFICATION Not specified Not specified Not specified Not specified
AUTO CALIBRATION Yes Yes No No
CALIBRATION FREQ Test dependent Test dependent 6 months 6 months
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes Yes
WATER REQUIREMENTS,
L/min 2.0 3.0 None None
Onboard supply Yes Yes NA NA
DISPLAY CRT, printer CRT, printer CRT, forms printer CRT, forms printer
PREVENTIVE MAINT
FREQUENCY 2/year 2/year 1/year 2/year
H x W x D, cm (in) 126 x 377 x 114 126 x 507 x 114 119 x 114 x 71 140 x 173 x 96.5
(49.6 x 148 x 44.9) (49.6 x 200 x 44.9) (47 x 44.9 x 28) (55 x 68 x 38)
WEIGHT, kg (lb) 1,370 (3,021) 1,780 (3,916) 272 (600) 681 (1,500)
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 47
Healthcare Product Comparison System
ASSAY METHODS End point, ISE, End point, kinetic, End point, kinetic, End point, kinetic,
kinetic rate FxT ratio, FxT ratio, double kinetic,
differential, MSTD differential, MSTD turbidimetric, ISE
BASIC/TIME, min
Albumin Yes Yes Yes Yes
ALP Yes Yes Yes Yes
ALT Yes Yes Yes Yes
Amylase Yes Yes Yes Yes
AST Yes Yes Yes Yes
BUN Yes Yes Yes Yes
Ca++ Yes Yes Yes Yes
Cholesterol Yes Yes Yes Yes
CK Yes Yes Yes Yes
Cl- Yes Yes Yes Yes
CO2 Yes Yes Yes Yes
Creatinine Yes Yes Yes Yes
Direct and total
bilirubin Yes Yes Yes Yes
Glucose Yes Yes Yes Yes
Inorgan phosphorus Yes Yes Yes Yes
Iron Yes Yes Yes Yes
K+ Yes Yes Yes Yes
LDH Yes Yes Yes Yes
Mg Yes Yes Yes Yes
Na+ Yes Yes Yes Yes
Total protein Yes Yes Yes Yes
Triglycerides Yes Yes Yes Yes
Urea Yes Yes Yes Yes
Uric acid Yes Yes Yes Yes
Others Ammonia, TIBC, LDL Yes Yes Direct HDL, direct
LDL, lactate,
lipase, UIBC
SPECIAL/TIME, min
Acid phosphatase Yes Yes Yes No
Ethanol Yes Yes Yes Yes
Aldolase Yes Yes Yes No
Amylase Yes Yes Yes Yes
C3 Yes Yes Yes Yes
C4 Yes Yes Yes Yes
Cholinesterase Yes Yes Yes Yes
CK-MB Yes Yes Yes Yes
CSF protein Yes Yes Yes Yes
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes Yes Yes
48 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 49
Healthcare Product Comparison System
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes No Yes Yes
WATER REQUIREMENTS,
L/min 0.005 No No 0.0083
Onboard supply No NA NA No
DISPLAY CRT, printer LCD 4 x 40 dots External CRT, printer
50 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
ASSAY METHODS End point, FP, End point, FP, End point, FP, End point, kinetic,
kinetic, turbidi- kinetic, turbidi- kinetic, turbidi- turbidimetric, ISE
metric, ISE metric, ISE metric, ISE
BASIC/TIME, min
Albumin Yes Yes Yes Yes
ALP Yes Yes Yes Yes
ALT Yes Yes Yes Yes
Amylase Yes Yes Yes Yes
AST Yes Yes Yes Yes
BUN Yes Yes Yes Yes
Ca++ Yes Yes Yes Yes
Cholesterol Yes Yes Yes Yes
CK Yes Yes Yes Yes
Cl- Yes Yes Yes Yes
CO2 Yes Yes Yes Yes
Creatinine Yes Yes Yes Yes
Direct and total
bilirubin Yes Yes Yes Yes
Glucose Yes Yes Yes Yes
Inorgan phosphorus Yes Yes Yes Yes
Iron Yes Yes Yes Yes
K+ Yes Yes Yes Yes
LDH Yes Yes Yes Yes
Mg Yes Yes Yes Yes
Na+ Yes Yes Yes Yes
Total protein Yes Yes Yes Yes
Triglycerides Yes Yes Yes Yes
Urea Yes Yes Yes Yes
Uric acid Yes Yes Yes Yes
Others Direct HDL, direct Direct HDL, direct UIBC, direct HDL, Direct HDL, direct
LDL, lactate, LDL, lactate, direct LDL, lipase, LDL
lipase, micro- lipase, micro- lactate
albumin, UIBC albumin, UIBC
SPECIAL/TIME, min
Acid phosphatase Yes Yes Yes Yes
Ethanol Yes Yes Yes Yes
Aldolase No No No No
Amylase Yes Yes Yes Yes
C3 Yes Yes Yes Yes
C4 Yes Yes Yes Yes
Cholinesterase Yes Yes Yes No
CK-MB Yes Yes Yes No
CSF protein Yes Yes Yes No
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Direct
IgA Yes Yes Yes Yes
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 51
Healthcare Product Comparison System
THROUGHPUT,
20 test profiles/hr Not specified 35 Not specified Not specified
REAGENT TYPE Liquid Liquid Liquid Liquid/dry
reconstitute
REAGENT DELIVERY Automatic Automatic Automatic Automatic
52 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
AUTO VERIFICATION Not specified Not specified Not specified Not specified
AUTO CALIBRATION Yes Yes Yes Yes
CALIBRATION FREQ 4/year, test 4/year, test 4/year, test Test dependent;
dependent dependent dependent 30 days, minimum
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Not specified
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes Yes
WATER REQUIREMENTS,
L/min 0.03 0.07 0.07 0.004
Onboard supply 1L Not specified Not specified 2.5 L
DISPLAY Color LCD, laser Color CRT, laser Color CRT, laser CRT, thermal
printer printer printer dot-matrix printer
LIGHT SOURCE Tungsten-halogen Tungsten-halogen LED Xenon flash tube
COMPUTER INTERFACE Bidirectional RS232C Bidirectional RS232C Bidirectional RS232C Bidirectional RS232
PREVENTIVE MAINT
FREQUENCY 2/year 4/year 4/year 2/year
H x W x D, cm (in) 75 x 135 x 66 150 x 120 x 85 120 x 190 x 90 66 x 74 x 58
(29.5 x 53.1 x 26) (58 x 47.2 x 33.5) (47.3 x 74.8 x 35.4) (26 x 29 x 23)
WEIGHT, kg (lb) 230 (507) 420 (925) 512 (1,126) 100 (221)
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 53
Healthcare Product Comparison System
ASSAY METHODS End point, kinetic, End point, kinetic, End point, kinetic, End point, kinetic,
double kinetic, 2-point, direct ISE 2-point, direct ISE 2-point, direct ISE
turbidimetric, ISE
BASIC/TIME, min
Albumin Yes Yes Yes Yes
ALP Yes Yes Yes Yes
ALT Yes Yes Yes Yes
Amylase Yes Yes Yes Yes
AST Yes Yes Yes Yes
BUN Yes Yes Yes Yes
Ca++ Yes No Yes Yes
Cholesterol Yes Yes Yes Yes
CK Yes Yes Yes Yes
Cl- Yes Yes Yes Yes
CO2 Yes Yes Yes Yes
Creatinine Yes Yes Yes Yes
Direct and total
bilirubin Yes Yes Yes Yes
Glucose Yes Yes Yes Yes
Inorgan phosphorus Yes Yes Yes Yes
Iron Yes Yes Yes Yes
K+ Yes Yes Yes Yes
LDH Yes Yes Yes Yes
Mg Yes Yes Yes Yes
Na+ Yes Yes Yes Yes
Total protein Yes Yes Yes Yes
Triglycerides Yes Yes Yes Yes
Urea Yes Yes Yes Yes
Uric acid Yes Yes Yes Yes
Others Direct HDL, direct Li, HBDH Li, HBDH Li, HBDH
LDL, LD1, UIBC
SPECIAL/TIME, min
Acid phosphatase No Yes Yes Yes
Ethanol Yes Yes Yes Yes
Aldolase No No No No
Amylase Yes Yes Yes Yes
C3 Yes Yes Yes Yes
C4 Yes Yes Yes Yes
Cholinesterase Yes Yes Yes Yes
CK-MB Yes Yes Yes Yes
CSF protein Yes Yes Yes Yes
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes Yes Yes
54 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
THROUGHPUT,
20 test profiles/hr 520 200 300 600
REAGENT TYPE Primarily liquid Liquid Liquid Liquid
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 55
Healthcare Product Comparison System
COMPUTER INTERFACE Bidirectional RS232 Bidirectional RS232C Bidirectional RS232C Bidirectional RS232C
PREVENTIVE MAINT
FREQUENCY 4/year 2/year 2/year 2/year
H x W x D, cm (in) Varies 130 x 85 x 79 115 x 120 x 79 115 x 150 x 79
(51.2 x 33.5 x 31.1) (45.3 x 47.2 x 31.1) (45.3 x 59 x 31.1)
56 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
PROGRAMMED TESTS 48 90 90 31
USER-DEFINABLE TESTS Yes Yes Yes Yes
OPTICAL SYSTEM Filter photometric, Diffraction grating Diffraction grating Photometric,
bichromatic of 12 wavelengths of 12 wavelengths bichromatic
(340-750 mm) (340-750 mm)
ASSAY METHODS End point, kinetic Rate A, rate B, 1- Rate A, rate B, 1- End point, initial
point, 2-point, point, 2-point, rate, kinetic
multipoint multipoint differential
BASIC/TIME, min
Albumin Yes Yes Yes Yes
ALP Yes Yes Yes Yes
ALT Yes Yes Yes Yes
Amylase Yes Yes Yes Yes
AST Yes Yes Yes Yes
BUN Yes Yes Yes Yes
Ca++ Yes Yes Yes Yes
Cholesterol Yes Yes Yes Yes
CK Yes Yes Yes Yes
Cl- Yes Yes Yes Yes
CO2 Yes Yes Yes No
Creatinine Yes Yes Yes Yes
Direct and total
bilirubin Yes Yes Yes Yes
Glucose Yes Yes Yes Yes
Inorgan phosphorus Yes Yes Yes Yes
Iron Yes Yes Yes Yes
K+ Yes Yes Yes Yes
LDH Yes Yes Yes Yes
Mg Yes Yes Yes Yes
Na+ Yes Yes Yes Yes
Total protein Yes Yes Yes Yes
Triglycerides Yes Yes Yes Yes
Urea Yes Yes Yes Yes
Uric acid Yes Yes Yes Yes
Others Hemoglobin, UIBC, ammonia, LDL UIBC, ammonia, LDL None specified
ammonia, TIBC, LDL microalbumin, CRP microalbumin, CRP
SPECIAL/TIME, min
Acid phosphatase Yes Yes Yes Yes
Ethanol Yes Yes Yes No
Aldolase Yes Yes Yes No
Amylase Yes Yes Yes Yes
C3 Yes Yes Yes No
C4 Yes Yes Yes No
Cholinesterase Yes Yes Yes Yes
CK-MB Yes Yes Yes Yes
CSF protein Yes Yes Yes Yes
GGT Yes Yes Yes Yes
HDL cholesterol Yes Yes Yes Yes
IgA Yes Yes Yes Yes
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 57
Healthcare Product Comparison System
SAMPLE TYPE
Serum and plasma Yes Yes Yes Yes
Urine Yes Yes Yes Yes
CSF Yes Yes Yes Yes
SAMPLE TRAY CAPACITY 45 50 x 10 trays, 25 50 x 10 trays, 25 40
stat, 8 controls stat, 8 controls
58 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
POWER FAILURE
CALIBRATION MEMORY Yes Yes Yes Yes
ABNORMAL VALUES FLAG Yes Yes Yes Yes
DIRECT SAMPLING Yes Yes Yes Yes
WATER REQUIREMENTS,
L/min 0.03 0.1 0.1 0.001
Onboard supply No Yes Yes Yes
DISPLAY CRT, printer CRT CRT with optional CRT, printer
touchscreen
LIGHT SOURCE Tungsten-halogen Tungsten-halogen Tungsten-halogen Tungsten-halogen
WEIGHT, kg (lb) 34 (75) 150 (331) max, main 150 (331) max, main 60 (132.3)
unit with monitor unit with monitor
LINE POWER, VAC 110/220 110/220 110/220 110/220 ±10%
POWER CONSUMPTION 100 W 1 kVA 1 kVA 400 W
PURCHASE INFORMATION
List price,
std configuration $12,500 $45,000 $52,000 $20,000
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 59
Healthcare Product Comparison System
BASIC/TIME, min
Albumin Yes Yes
ALP Yes Yes
ALT Yes Yes
Amylase Yes Yes
AST Yes Yes
BUN Yes Yes
Ca++ Yes Yes
Cholesterol Yes Yes
CK Yes Yes
Cl- Yes Yes
CO2 Yes Yes
Creatinine Yes Yes
Direct and total
bilirubin Yes Yes
Glucose Yes Yes
Inorgan phosphorus Yes Yes
Iron Yes Yes
K+ Yes Yes
LDH Yes Yes
Mg Yes Yes
Na+ Yes Yes
Total protein Yes Yes
Triglycerides Yes Yes
Urea Yes Yes
Uric acid Yes Yes
Others Bicarbonate, CK-NAC, Bicarbonate, CK-NAC,
lactate, LAP, lactate, LAP,
LDH(Tris), lipase, LDH(Tris), lipase,
LDL cholesterol LDL cholesterol
SPECIAL/TIME, min
Acid phosphatase Yes Yes
Ethanol Yes Yes
Aldolase Yes Yes
Amylase Yes Yes
C3 Yes Yes
C4 Yes Yes
Cholinesterase Yes Yes
CK-MB Yes Yes
CSF protein Yes Yes
GGT Yes Yes
HDL cholesterol Yes Yes
IgA Yes Yes
60 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.
Clinical Chemistry Analyzers, Automated, Discrete
SAMPLE TYPE
Serum and plasma Yes Yes
Urine Yes Yes
CSF Yes Yes
SAMPLE TRAY CAPACITY 51 80
THROUGHPUT,
20 test profiles/hr 180 + ISE 460 + ISE, 360
REAGENT TYPE Liquid Liquid
©2002 ECRI. Duplication of this page by any means for any purpose is prohibited. 61
Healthcare Product Comparison System
POWER FAILURE
CALIBRATION MEMORY Yes Yes
ABNORMAL VALUES FLAG Yes Yes
DIRECT SAMPLING Yes Yes
WATER REQUIREMENTS,
L/min Variable Variable
Onboard supply Yes Yes
DISPLAY CRT, printer CRT, printer
PREVENTIVE MAINT
FREQUENCY 2/year 2/year
H x W x D, cm (in) 115 x 49 x 58 117 x 115 x 77
(45.3 x 19.3 x 22) (46.1 x 45.3 x 30.3)
62 ©2002 ECRI. Duplication of this page by any means for any purpose is prohibited.