Download as pdf or txt
Download as pdf or txt
You are on page 1of 9

Designation: E319 − 85 (Reapproved 2014)

Standard Practice for the


Evaluation of Single-Pan Mechanical Balances1
This standard is issued under the fixed designation E319; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.

INTRODUCTION

The balance performs two basic functions: (1) it compares an unknown load with one or more
weights, and (2) it indicates the difference between the two loads for differences smaller than the
smallest weights normally used on the balance. The test procedure given herein measures the precision
with which the balance can compare the two loads, and the rates at which systematic errors may affect
the observed difference.

1. Scope 2.1.1 accuracy—the degree of agreement of the measure-


ments with the true value of the magnitude of the quantity
1.1 This practice covers testing procedures for evaluating
measured (2).
the performance of single-arm balances required by ASTM
standards. 2.1.2 correction for a weight—the correction for the error in
adjustment is:
1.2 This practice is intended for but not limited to sensitivity
ratios of 106 or better and on-scale ranges of 1000xd or more Cr.W 5 A 2 N (1)
where d = reability either directly or by estimation. where:
1.3 This practice can also be applied to other single-pan Cr.W = correction for the error in adjustment to nominal
balances with mechanical weight changing of different capaci- value,
ties or sensitivities with appropriate test loads and calibration A = actual value of the weight, and
weights. N = nominal value.
NOTE 2—In practice it is not possible to adjust weights exactly to their
NOTE 1—Mechanical balances of this type have largely been replaced nominal values.
by automatic electronic balances incorporating a variety of operational
principles. Nevertheless, some single-pan mechanical balances are still 2.1.3 correction for error in scale indication— the correc-
manufactured and many older balances will remain in service for years to tion for the scale indication, I, is:
come. One type of automatic electronic balance, the so-called “hybrid,”
Cr.I 5 A 2 I (2)
bears considerable similarity to single-pan mechanical balances of the null
type. (1)2 NOTE 3—The correction for the scale is taken with reference to the
measured value of a weight used as a test load during calibration of the
1.4 This standard does not purport to address all of the on-scale range.
safety problems, if any, associated with its use. It is the 2.1.4 index of precision—the standard deviation, computed
responsibility of the user of this standard to establish appro- in any acceptable manner, for a collection of measurements
priate safety and health practices and determine the applica- involving a given pair of mass standards (3).
bility of regulatory limitations prior to use.
NOTE 4—The standard deviation is computed from the data provided by
the instrument precision test (see Section 7) index of precision.
2. Terminology
2.1.5 null-type balance—a balance which requires, as the
2.1 Definitions(1):
final step in its operation, that the observer restore the angle of
the balance beam to its original (or null) position. The least
significant figures of the balance indication are obtained from
1
This practice is under the jurisdiction of ASTM Committee E41 on Laboratory this operation.
Apparatus and is the direct responsibility of Subcommittee E41.06 on Laboratory
Instruments and Equipment. 2.1.6 optical-type balance—in this type the least significant
Current edition approved Nov. 1, 2014. Published November 2014. Originally figures of the balance indication are related to the deflection
approved in 1968. Last previous edition approved in 2008 as E319 – 85 (2008).
angle of the beam from its original (or null) position. A scale
DOI: 10.1520/E0319-85R14.
2
The boldface numbers in parentheses refer to the list of references at the end of placed on the moving beam is optically projected onto the
this practice. (stationary) balance case to provide this indication.

Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States

1
E319 − 85 (2014)
2.1.7 precision—the repeatability of the balance indication test results are to indicate future performance, any disturbances
with the same load under essentially the same conditions. that occur exterior to the balance must be brought under control
(2).
NOTE 5—The more closely the measurements are grouped, the smaller
the index of precision will be. The precision must be measured under 4.2 Acceptance Tests—This practice may also be used as
environmental conditions that represent the conditions under which the acceptance tests for new balances. For this purpose, the tests
balance is normally used. should be conducted under favorable, but not necessarily ideal,
2.1.8 readability—the value of the smallest decimal subdi- conditions. Since systematic error in the course of the zero and
vision of a scale division in terms of mass units, that can be the course of the sensitivity may be caused by disturbances
read, when the balance is read in the intended manner. external to the balance, limits on these errors are not ordinarily
prescribed in acceptance requirements.
NOTE 6—The readability of a particular instrument is not a measure of
its performance as a weighing device. The relationship between the 5. Preparation of Apparatus
numerical value obtained by reading devices and the ability of the operator
to estimate the location of the reference device or index is important. It is 5.1 Balance (In all cases, the balance should be used in
possible to introduce a large number of readable subdivisions of the main accordance with the manufacturer’s instructions):
scale divisions that would increase the “readability” as defined but if the
5.1.1 The results obtained will depend on the environment.
reading device cannot be reset to the same numerical value when the beam
is in an immovable condition, or when the load on the beam is a constant Select an area which is free of excessive vibration and air
value, the readability becomes meaningless. Readability substantially less currents, where rapid changes in temperature and relative
than 1 standard deviation as determined by repeated measurement with a humidity will not be encountered, and where the floor is rigid
given test weight is usually superfluous. enough to be free of a tilting effect on the balance indication.
2.1.9 scale division—the smallest graduated interval subdi- Place the balance on a sturdy bench. If the balance has been
vided either by estimation or with the aid of a vernier. moved to a new location, permit it to come to thermal
Subdivisions which appear as divisions on the vernier are not equilibrium for at least 1 h before performing the test,
considered to be scale divisions, but rather parts of scale preferably several hours.
divisions. 5.1.2 Inspect and test the balance to make sure that it is in
proper mechanical order. Arrest and release the beam to make
2.1.10 sensitivity weight—a small weight used to measure
sure that readings are approximately repetitive. Observe the
the “on-scale” deflection of the balance indicator.
indication during arrest and release to ensure that there is no
NOTE 7—With single-pan balances the sensitivity weight should be “kick” that would indicate that arrestment points might be out
equal to the value of the smallest built-in weight represented by the first of adjustment. If necessary, have the balance adjusted by a
step on the dial for the smallest weights. competent balance technician.
2.1.11 test load—a load chosen to represent the sample load 5.1.3 Make a few trial measurements of the interval from
in the test procedure. zero to the full-scale indication.
2.1.12 value of the division—the change in load required to 5.2 Reading the Balance—The balance should be read in
change the balance indication by one scale division. The accordance with the instructions supplied by the manufacturer.
reciprocal of the sensitivity is its most useful function. Optical types should include the reading of verniers or mi-
crometres. Null types should include the indication of the
3. Summary of Practice device for restoring to null including verniers or micrometres.
3.1 The accuracy of the direct-reading scale, the smallest 6. Preliminary Testing of Single-Arm Balances
weight of the set of built-in weights, and uniformity of
6.1 Summary of Method—With single-pan balances the
sensitivity between the upper and lower halves of full-scale
smallest built-in weight, indicated by the first step on the dial,
deflections are verified by preliminary tests.
is compared with a calibrated weight. The direct-reading scale
3.2 Estimates of rate of change of the zero with time, rate of is tested for agreement with the smallest built-in weight and the
change in the value of the scale division with time, and a sensitivity is adjusted, if necessary, so that the indications of
quantitative measure of the variability or random error are the scale are precise in terms of the calibrated weight. A
provided by short tests for precision and bias. “fifty-fifty” test verifies the accuracy of the midpoint at
half-full scale. This test should be performed before proceed-
3.3 An overall test of the direct-reading capability is pro-
ing to other tests. After the accuracy of adjustment of the
vided by tests of the built-in weights.
smallest built-in weight is verified, this weight is used to test
full-scale deflections. Tests are also made for the uniformity of
4. Significance and Use
deflection over the lower and upper halves of the full-scale
4.1 Monitoring Weighing Performance—This practice pro- deflection. The preliminary tests show either that the balance is
vides results in the form of control charts which measure the operating properly, or that discrepancies indicate the presence
weighing capability at the time of the test. A series of tests at of sources of error. Uncertainties of perhaps one millionth of
appropriate intervals will monitor balance performance over a the balance capacity may be caused by dirt or foreign material
period of time. A marked change from expected performance in the bearings, or by unskilled handling, while larger discrep-
may result from a variety of causes including: maladjustment, ancies may be caused by worn or damaged knife-edges or other
damage, dirt, foreign material, and thermal disturbances. If the sources such as electrostatic effects. Any necessary cleaning or

2
E319 − 85 (2014)
servicing should be done at this point. If discrepancies 6.4.2 Compute S1 in scale divisions to verify the full-scale
continue, other possible sources of uncertainty should be value on the direct-reading scale as follows:
studied. There is no point in proceeding with routine test S1 5 ~ c 2 b1d 2 e ! /2 (4)
procedures until acceptable results can be obtained with the
preliminary tests. where c, b, d, and e are taken from Table 1. Adjust the
balance sensitivity if necessary so that the full-scale reading
NOTE 8—With null-type balances (including the hybrid) it is possible to
equals D1.
use the flexure of a segment of metal, quartz, etc. as the main pivots
instead of knife edges. A flexure pivot is by its nature free of problems of 6.4.3 Compute average scale difference, A, for lower 50 %
dirt. Flexures are also generally more robust than knives. The chief of direct-reading scale as follows:
problem associated with flexures is that they act like springs and thus add
A 5 ~ g 2 f1j 2 k ! /2 (5)
a restoring force which may vary with time or temperature. This drawback
can be minimized by careful design and all but eliminated by the use of 6.4.4 Compute average scale difference, B, for upper 50 %
servo-control in electronic balances.
of direct-reading scale as follows:
6.2 Materials:
B 5 ~ h 2 g1i 2 j ! /2 (6)
6.2.1 A watch or clock which indicates seconds,
6.2.2 Pencils for recording data, A and B should agree within 3 standard deviations (see
6.2.3 Columnar data sheets (If balance performance will be 7.5.3). Any discrepancy smaller than 3 standard deviations may
monitored periodically, it may be useful to enter data directly be ascribed to uncertainty in the preliminary measurements and
into a personal computer which has been programmed for this does not necessarily indicate a real change in the value of the
task.), scale divisions.
6.2.4 A calibrated weight designated S1 which has the 6.4.5 Inspect the no-load readings, a, f, and k for agreement
nominal value equal to the smallest interval on the dial- or zero drift.
operated weights, and 6.4.6 See Table 2 and Fig. 1 for examples of calculations
6.2.5 Two weights of half of the nominal value of S1 and observation form.
designated (1⁄2)1 and (1⁄2)2. (These weights need not be
calibrated but they should bear distinguishing marks, prefer- 7. Instrument Precision (4)
ably one, and two dots.) 7.1 Summary of Method:
6.3 Procedure—Adjust the “no-load” readings to a point 7.1.1 A set of four readings is repeated four times, or more,
near the center of the vernier so that zero drift or other to obtain pairs of readings with identical loads:
deviation will not cause a negative scale reading. Perform the 7.1.1.1 A reading near zero,
preliminary tests, loading the pan and changing the dial 7.1.1.2 A reading near the upper end of the scale,
settings according to the schedule in Table 1. Before releasing 7.1.1.3 A reading near the upper end of the scale with a test
the beam, record the load on the pan and the dial setting so that load plus a small weight, and
the observation will be confined to the scale reading. Release 7.1.1.4 A reading near zero with the test load but with the
the balance and observe the scale reading. Record the indica- small weight removed.
tion and verify the stability of the scale reading, then arrest the 7.1.2 Readings are taken at a steady pace as rapidly as
balance promptly. practicable, consistent with good practice, and the time is
6.4 Calculations for Preliminary Tests: observed at the start of each set of observations and at the end
6.4.1 Compute D1, the value of the smallest built-in weight of the test.
as follows: 7.1.3 The balance indications are plotted on a graph to
provide a visual presentation of errors. The zero readings are
D1 5 @ ~ a 2 b1f 2 e ! /2 # 1S1 (3)
connected to show the course of the zero with time. The
where: a, b, f, and e are taken from Table 1, and S1 = cali- response of the balance to the small weight is plotted. The
brated value of test weight. course of the sensitivity with time is represented by a plot of

TABLE 1 Schedule for Preliminary Tests of Single-Arm Balances


Observation Time Pan Load Dial Setting Scale Reading
a Record the time zero 0 ...
A
b S1 1 ...
c S1A 0 ...
d Wait 30 s S1A 0 ...
e S1A 1 ...
f Record the time 0 0 ...
g (1⁄2)1B 0 ...
h Add (1⁄2)2B (1⁄2)1 + (1⁄2)2B 0 ...
i Wait 30 s (1⁄2)1 + (1⁄2)2B 0 ...
j Remove (1⁄2)1B (1⁄2)2B 0 ...
k Record the time 0 0 ...
A
S1 = calibrated weight of nominal value equal to the smallest dial-operated weight.
B 1
( ⁄2 ) 1and (1⁄2)2 = weights of nominal value equal to 1⁄2 S1 (not necessarily calibrated but marked for identification).

3
E319 − 85 (2014)
TABLE 2 Example of Preliminary Test of Direct Reading Balance
Observer: John Doe Date: (May 5)
Observation Time Load Dial Scale Reading
a 1:30 p.m. 0 0 6.4
b S1 1 6.45
c S1 0 1006.5
d 1:34 1⁄2 S1 0 1006.5
e S1 1 6.5
f 0 0 6.4
g 1:39 (500 mg)1 0 506.5
h (500 mg)1 + (500 mg)2 1006.25
i (500 mg)1 + (500 mg)2 0 1006.3
j (500 mg)2 0 506.15
k 0 0 6.4

Calculations:
D1 = [(a − b + f − e) ⁄ 2] + S1 = [(6.4 − 6.45 + 6.4 − 6.5) ⁄ 2 ] + 1000 = 999.925
S1 ( = 1000.006 mg) = [(c − b + d − e) ⁄ 2 ] = [(1006.5 − 6.45 + 1006.5 − 6.5) ⁄ 2] = 1000.025 scale divisions
A = (g − f + j − k) = (506.5 − 6.4 + 506.15 − 6.4) ⁄ 2 = 499.925
B = (h − g + i − j) = (1006.25 − 506.5 + 1006.3 − 506.15) ⁄ 2 = 499.95

FIG. 1 Calculations and Observation Form for Test of Precision

the interval from zero to full-scale angular motion of the beam, 7.1.4 A control chart formed by plotting the range within
or with balances having indication to a null point by a plot of pairs provides a quick estimate of the standard deviation for a
an interval measured on the device for restoring the beam to a single reading. The precision of the balance is estimated by
null indication. computations based on the range within pairs of observations.

4
E319 − 85 (2014)
7.2 Materials—The following materials are required: 7.3 Preparation:
7.2.1 A watch or clock which indicates seconds, 7.3.1 Draw, in the upper left quarter of the sheet of graph
7.2.2 Pencils, paper, a horizontal base line for points near zero and a second
7.2.3 Cross-section paper, 20 lines/in. or 10 lines/cm (If horizontal base line for points near full scale (100 or 1000
balance performance will be monitored periodically, it may be divisions).
useful to enter data directly into a personal computer which has
7.3.2 Time is represented in the horizontal direction. Space
been programmed for this task.),
7.2.4 One 50-g test load or other appropriate load (the observations one square apart and record the time showing
correction to nominal value need not be determined), hour, minutes, and seconds at the start of each set of four
7.2.5 A sensitivity weight (usually 1 or 0.1 g depending readings and at the end of the test. The vertical axis represents
upon the type of balance) with the nominal value equal to one change in balance indication. One square represents the small-
interval on the dial for the smallest internal weight. (The est unit that is intended to be read or estimated. It is, however,
correction for this weight is not needed since the full-scale proper to record one half a vernier division as half a square on
indication is compared directly with the smallest weight in the the graph.
self-contained set of weights.) 7.3.3 Refer to Fig. 2 for an example.

FIG. 2 Computation Form for Preliminary Test of Direct-Reading Scale

5
E319 − 85 (2014)
7.4 Precision Test (Short Test Procedures)—See Table 3 for sample statistics. A more reliable estimate of the standard
summary of test. In the following procedure a 50-g test load is deviation may be secured by using more sets of observations.
used as an example. Any other test load can be used with If desired, additional data can be taken. The total number of
appropriate dial settings. sets of four readings must be an even number so that two
7.4.1 The 50-g weight and sensitivity weight equal to one adjacent sets can be paired to provide the difference between
step on the dial for the smallest weights shall have been in the pairs of readings with identical loads. Fig. 2 shows the data and
balance case for at least 1 h. Load the pan with the sensitivity computations for six sets of four readings arranged so that all
weight (1 or 0.1 g for most balances) and leave it on the pan for data and computations are on a single sheet of graph paper.
the entire series of observations. Close the door. Set the dial for Successive points are to be plotted one square apart, from left
the smallest weights to “1”. Release the beam and adjust the to right.
zero to a reading in the range from +0.5 to 1 mg so that change 7.4.7 After completing the desired even number of sets of
in the zero will not take the indication off scale. Arrest the four readings, finish with a zero reading. This last reading will
beam. After the door has been closed for at least 30 s, proceed be used to complete the course of the zero with time. Record
with the following test cycle. the time.
7.4.2 Record the time on the graph. Release the beam at the 7.4.8 See 7.5 for calculations, and Fig. 2 for an example.
time recorded. When the beam comes to rest, observe the
7.5 Calculations—See Fig. 1 for an example.
indication and record as a zero reading on the graph. Enclose
the point within a small circle. Look again at the balance 7.5.1 The Course of the Zero—Connect the points obtained
indication to verify the reading. Without observing the time, by Step 1 of the test cycle to show the course of the zero with
but working at a steady pace, continue with the sequence. time.
7.4.3 Arrest the beam, change the dial from “1” to “0”, 7.5.2 The Course of the Sensitivity—Subtract the indications
release the beam, read as soon as practicable, record one square for the first observation from the second, and the fourth from
to the right of the first point as a point near full scale (near an the third, and plot to obtain the values of scale difference
indication of “100” or “1000” divisions as the case may be), corresponding to the first step on the dial for the smallest
encircle this point, verify the reading, and arrest the beam. weights. Connect these points to show the course of the
7.4.4 Open the door and place the 50-g test load on the pan sensitivity with time as illustrated in Fig. 1.
using weight forceps long enough so that your hand is not put 7.5.3 Quick Estimate of the Standard Deviation(s)—(4,5) :
inside the weighing chamber. Shut the door promptly, change Average difference between pairs = (2/π)s = 0.798s
the dial to the appropriate setting, that is “50” or “50.0” leaving It then follows that s = average difference/0.798
the dial for the smallest weight at “0”. Release the beam, read The estimate computed in this manner may not agree with
and record as a point near an indication of “100” or “1000” that computed by more sophisticated procedures if the collec-
scale divisions, as the case may be. Enclose the point in a small tion of data is small. This, however, does not impair its
square to indicate the 50-g load. Arrest and change the dial for usefulness since either value may differ from the long range
the smallest weight from “0” to “1”. Release, read and record estimate of the standard deviation computed on sufficiently
as a point near zero enclosed in a small square, verify, and large collections of data.
arrest the beam. 7.5.4 Report on Balance Test—Identifying Data—The bal-
7.4.5 Open the door and remove the 50-g test load from the ance should be identified by model and serial number. The
pan and place it on the floor of the balance case next to the pan observer, date, and location should be stated. Distinctive
where it will be most convenient to be picked up again. Close features of the balance should be listed.
the door. Change the appropriate dial, that is the 10-g incre- 7.5.5 Systematic Error—If there was a trend in the data,
ment dial from “5” to“ 0”. Observe the time estimating a few state an estimate of the rate of change with time in mass units
seconds ahead to the time the balance will be released. per minute, or describe the change in the course of the zero. For
7.4.6 Repeat the observations as listed from 7.4.4 and 7.4.5 direct reading balances, compare the full scale deflection with
at least three times more to obtain four or more sets of four the first dialed interval.
readings. Four sets of readings will provide the minimum 7.5.6 Random Error—State the estimate of the standard
number of observations required to treat the data by small deviation.

TABLE 3 Summary of Short Test of Precision—Single-Pan Constant-Load Balance


Record starting time on graph.
Step Operation Weight on Pan Dial Setting Observation Record
1 Record time and release beam, read, record 1 g or 0.1 g 1 or 0.1 point near “0” |b. at “0” time near “0” line
2 Arrest beam, change dial, release beam, read, 1 g or 0.1 g 0 or 0.0 point near 1000 |b. at time “1” near “1000” or “100”
record, arrest beam or 100 line
3 Open door, place test load on pan, shut door, set 51 g or 50.1 g 50 or 50.0 point near 1000 at time “2” near “1000” or “100”
dial, release beam, read, record or 100 line
4 Arrest, change dial, release, read, record 51 g or 50.1 g 51 or 50.1 point near 0 at time “3” near “0” line
5 Open door, remove 50-g weight, close door, 1 g or 0.1 g 1 or 0.1
change dial, close door, observe time

Repeat steps 1 to 5 four times for a total of at least 16 points recorded on the graph (see Fig. 2).
Finish with a zero reading (Step 1). Record the time.

6
E319 − 85 (2014)
8. Test for Built-In Weights (Single-Pan Constant-Load ate schedules for the various balance models. Observe and
Balance) record the scale readings according to the schedule of test loads
8.1 Summary of Method—In 6.4.2 it was specified that the and dial settings, without readjusting the zero.
sensitivity of the balance is to be adjusted, if necessary, so that 8.3.3 Load the pan with the sensitivity weight equivalent to
the indications of the direct reading scale are accurate. After the full-scale deflection (100 mg in the sample Observation
this adjustment has been performed, a test load, from a set of Schedule, Table 3). Close the door. Set the dial for the smallest
weights whose values are known (5), is compared with two weights to “1”. Record the time, the load on the pan, and dial
combinations of the built-in weights in the smallest decade of setting in the proper columns on the observation sheet. (See
the dial. One dial setting will have the same nominal value as Table 7.) Release the beam at the time recorded. When the
the test load, the other will be smaller by one step on the dial
for the smallest decade. In this manner, suitable test loads are
selected to test all the built-in weights in the smallest decade.
TABLE 6 Example of Report of Test on Build-In Weights
Each decade is tested in a similar manner. (See Table 7)A
8.2 Materials Required—The materials required for this test
are: Balance:.......................... Date:.............................
Serial No:........................ Observer:.......................
8.2.1 Watch or clock which indicates seconds, Capacity: 160 g, digital readout to 0.1 mg, one step on dial 1 g, optical readout
8.2.2 Pencils, 1000 mg.
8.2.3 Columnar data sheets (If balance performance will be A
Various combinations of built-in weights were compared with standards. The
monitored periodically, it may be useful to enter data directly largest indicated deviation of any combination from the nominal value was 0.1 mg.
The corrections for standards were negligible so that they were not taken into
into a personal computer which has been programmed for this account.
task.), and
8.2.4 A set of weights to provide known test loads. (It is
convenient to use weights which have been adjusted close TABLE 7 Test of Built-In Weights
enough to the nominal values so that the corrections are NOTE 1—(Corrections to standard are negligible)
negligible by comparison with the readability of the balance. It
Time Load Dial Scale
may be necessary to account for the errors in adjustment of the
1:42 1g 1 0.6.4
weights). Dial 0 to 1 → 1000.0
div.
NOTE 9—Caution: All direct reading balance indications are in terms of 1g 0 1006.4
apparent mass on some defined or implied basis. The built-in weights of 0 0 6.4
many older single-pan mechanical balances are adjusted on the basis of S2g 2 1006.5
8.4 g/cm3 (that is, the weights, which are usually stainless steel, are j
S1g 3 6.4
adjusted to balance brass standards under normal laboratory conditions). 1000.1
More recently, built-in weights have been adjusted on an 8.0 g/cm3 basis, 0 0 6.4
1:47 S5 4 1006.4
which better approximates the actual density of modern grade S weights j
S5 5 6.4
(6). Using error-free weights adjusted in one basis to check the calibration 1000.0
of weights adjusted in the second basis will require a mass correction of 0 0 6.4
about 0.0007 % (0.7 mg for each 100 g). These corrections are necessary S 20 g 19 1006.4
j
because of the buoyant effect of laboratory air (7,8). S 20 g 20 6.4
1000.0
8.3 Procedure: 0 0 6.4
S 50 g
8.3.1 The set of weights should be placed inside the pan
S 20 g 79 1006.3
compartment of the balance for at least 1 h before starting the S 10 g
j
80 6.3
j
test of the built-in weights. Weights should be in thermal 1000.0
0 0 6.4
equilibrium with the balance. 1g 1 6.4
8.3.2 Adjust the no-load reading to some point near the 1g 0 1006.4
center of the vernier so that variability will not produce a Dial 0 to 1 → 1000.0
div.
negative reading. Table 5 shows a typical test load and
dial-setting schedule. The manufacturer can furnish appropri-

TABLE 4 Example of a Report on a Balance Test


Balance:.......................... Date:............................ beam comes to rest, observe the indication and record on the
Serial No:........................ Observer:..................... observation sheet, on the same line as the load and the weight,
Capacity: 160 g, digital readout to 0.1 mg, one step on the dial 1 g, optical
readout 1000 mg.
but in the column headed “Reading.” Verify the reading.
Systematic Error: The zero readings increased 0.15 mg in 15 min for an Without observing time, but working at a steady pace, continue
estimated rate of 10 µg/min. The sensitivity in scale divisions per gram the sequence.
varied between 1000.2 and 999.95 scale divisions per gram during 18 min.
Random Error: The estimated standard deviation for the difference between a 8.3.4 Arrest the beam, change the dial from “1” to “0”,
pair of observations is 0.073 mg. release the beam, and record the load and dial settings. Read
Comment: On the basis of 12 pairs of duplicate readings taken during an the balance indication when the beam comes to rest and record
18-min period, it appears that the systematic error in the balance indication
is relatively unimportant by comparison with the random error. the indication on the observation sheet, verify the reading, and
arrest the beam.

7
E319 − 85 (2014)
TABLE 5 Sample Observation Schedules for Test of Built-in Weights
Series of Denominations 80, 40, 20, 10, 5, 2, 1, 1 g Series of Denominations 5, 2,1, 1 g, for
All Decades
Smallest Interval = 1 g Smallesr interval = 0.1 g
Capacity = 160 g Capacity = 100 g
A
Pan Load Dial Setting Scale Reading Pan LoadA Dial Setting Scale Reading
1g 1 ... 100 mg 0.1 ...
1g 0 ... 100 mg 0.0 ...
Zero 0 ... Zero 0.0 ...
S3g 2 ... S 0.3 g 0.2 ...
S3g 3 ... S 0.3 g 0.3 ...
Zero 0 ... Zero 0.0 ...
S5g 4 ... S 0.5 g 0.4 ...
S5g 5 ... S 0.5 g 0.5 ...
Zero 0 ... Zero 0.0 ...
S 20 g 19 ... S3g 2.9 ...
S 20 g 20 ... S3g 3.0 ...
Zero 0 ... Zero 0.0 ...
S 80 g 79 ... S5g 4.9 ...
S 80 g 80 ... S5g 5.0 ...
Zero 0 ... Zero 0.0 ...
1g 1 ... S 30 g 29.9 ...
1g 0 ... S 30 g 30.0 ...
Zero 0.0 ...
S 50 g 49.9 ...
S 50 g 50.0 ...
Zero 0.0 ...
100 mg 0.1 ...
100 mg 0.0 ...
A
Standard weights are designated by S before the denomination.

8.3.5 Open the door, remove the weight from the pan, close 8.4 Calculations—See Table 7 for example in which correc-
the door, release the beam, record the time, load (in this tions to the standard weights are neglected, and Table 8 for an
instance no load), and dial settings. Read as soon as example in which corrections are taken into account.
practicable, record reading, verify reading, and arrest beam. 8.4.1 The value of the weight or combination of weights
8.3.6 Open the door and place the first weight listed in the removed from the balance suspension when the dial is at the
Observation Schedule (in this case, 0.3 g) on the pan. Close the indicated setting may be computed from the following expres-
door, change the dial or dials to the appropriate setting sion:
indicated in the Observation Schedule (in this instance, the dial Value of dialed weight = value of standard (s)
for the smallest weights is set at “2”). Release the beam, record
+ nearest zero reading (a) − load reading (b)
the load and dial settings, read as soon as practicable, record
reading, verify reading, and arrest beam. Example from Table 8:
8.3.7 Change the dials to the next setting indicated in the Dial s a b
Observation Schedule (in this instance the dial for the smallest 0.2 = 300.00 + 0.53 − 100.51 = 200.02 mg
weights from “2” to “3”). Release the beam, record the load 0.3 = 300.00 + 0.54 − 0.53 = 300.01 mg
and dial setting and read as soon as practicable, record reading,
8.4.2 The values of the weights and combinations of
verify reading, and arrest the beam.
weights represented by the various dial settings may be shown
8.3.8 Repeat the observations as listed above from 8.3.5 –
on the observation sheet.
8.3.7 inclusive, for all of the test loads given in the Observation
Schedule. Use appropriate weights and dial settings. After the
last test-load observation, make another zero or no-load obser- 9. Keywords
vation as in 8.3.5 then repeat observations 8.3.3 and 8.3.4. 9.1 balances; scales; weighing instruments

8
E319 − 85 (2014)
TABLE 8 Example Test of Built-in WeightsA
Balance M-1
Date: Observer:
Observations Computations
Mass of Standard Mass of Dial Weight,
Load Dial Scale Zero Load
Time Weight, mg Nominal + mg Nominal +
on Pan Setting, g Reading, mg ReadingB Reading, mgB
Correction Correction
9:00 100 mg 0.1 0.55 = a 100 mg − 0.004 +0.53 (c) −0.55 (a) = 100 mg − 0.02
100 mg 0.0 100.50 = b
9:02 0 0.0 0.53 = c 300 mg − 0.002 +0.53 (e) −100.51 (d) = 200 mg + 0.02
300 mg 0.2 100.51 = d 300 mg − 0.002 +0.54 (f) −0.53 (e) = 300 mg + 0.01
300 mg 0.3 0.53 = e
9:05 0 0.0 0.54 = f 500 mg + 0.001 +0.54 (f) −100.54 (g) = 400 mg + 0.00
500 mg 0.4 100.54 = g 500 mg + 0.001 +0.50 (i) −0.54 (h) = 500 mg − 0.04
500 mg 0.5 0.54 = h
9:08 0 0.0 0.50 = i 3 g + 0.00 +0.50 (i) −100.51 (j) = 2.9 g − 0.01
3g 2.9 100.51 = j 3 g + 0.00 +0.52 (l) −0.47 (k) = 3.0 g + 0.05
3g 3.0 0.47 = k
9:11 0 0.0 0.52 = l 5 g + 0.02 +0.52 (l) −100.54 (m) = 4.9 g + 0.00
5g 4.9 100.54 = m 5 g + 0.02 +0.48 (o) −0.50 (n) = 5.0 g + 0.00
5g 5.0 0.50 = n
9:14 0 0.0 0.48 = o 30 g − 0.05 +0.48 (o) −100.41 (p) = 29.9 g + 0.02
30 g 29.9 100.41 = p 30 g − 0.05 +0.44 (r) −0.45 (q) = 30.0 g − 0.06
30 g 30.0 0.45 = q
9:17 0 0.0 0.44 = r 50 g + 0.01 +0.44 (r) −100.41 (s) = 49.9 g + 0.04
50 g 49.9 100.41 = s 50 g + 0.01 +0.43 (u) −0.45 (t) = 50.0 g + 0.01
50 g 50.0 0.45 = t
9:20 0 0.0 0.43 = u 100 mg − 0.004 +0.43 (u) −0.42 (v) = 100 mg + 0.01
100 g 0.1 0.42 = v
100 g 0.0 100.36 = w
A
An example of the test of the built-in weights of a semi-micro balance with a readability of 0.01 mg in which the corrections for the weights are taken into account. The
corrections are from NBS 2.6 Test No. 105, Set E. The computations are: mass of the standard weight plus the zero reading minus the load reading equals the mass of
the weights indicated by the dial setting.
B
The letters in parentheses indicate the observation indicated in the column, Scale Reading.

REFERENCES

(1) Schoonover, R. M., “A Look at the Electronic Analytical Balance,” (6) ASTM E617, Specification for Laboratory Weights and Precision
Analytical Chemistry, Vol 53, 1982, pp. 973A–980A. Mass Standards, Annual Book of ASTM Standards , Vol 14.02.
(2) Macurdy, L. B., “Response of Highly Precise Balances to Thermal (7) Jaeger, K. B., and Davis, R. S., A Primer for Mass Metrology,
Gradients,” Journal of Research of the National Bureau of National Bureau of Standards Special Publication 700-1, August 1984
Standards—C. Engineering and Instrumentation , Vol 68C, No. 3, .
July–September 1964. (8) Schoonover, Randall M., and Jones, Frank, “Air Buoyancy Correc-
(3) Eisenhart, Churchill, “Realistic Evaluation of the Precision and tions in High-Accuracy Weighing on Analytical Balances,” Analytical
Accuracy of Instrument Calibration Systems,” Journal of Research of Chemistry, Vol 53, 1981, pp. 900–902.
National Bureau of Standards—C. Engineering and Instrumentation, (9) Macurdy, L. B., “Tests of the Performance of Laboratory Balances,”
Vol 67C, No. 2, April–June 1963.
Instruments and Control Systems, Vol 38, No. 9, pp. 127–133,
(4) Natrella, M. G., “Experimental Statistics,” National Bureau of Stan-
September 1965.
dards Handbook 91 , October 1966.
(5) Tippett, L. H. C., “On the Extreme Individuals and the Range of
Samples taken from a Normal Population,” Biometrika, Vol 17, 1925,
p. 364.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/

You might also like