ADS1500 SHORT USER´S GUIDE

SENSOR CONNECTIONS

SENSOR TYPE CONNECTION INPUT MODE EXCITATION SCALE

VOLTAGE SOURCE DC Voltage -10V ... +10V


(DIFFERENTIAL) ---- to
(DC COUPLED)
(+/-) -3,33mV ... +3,33mV

VOLTAGE SOURCE
(DIFFERENTIAL) DC Voltage -10V ... +10V
---- to
(EXTERNALLY GROUNDED) (+/-) -3,33mV ... +3,33mV
(DC COUPLED)

VOLTAGE SOURCE
(NON-DIFFERENTIAL) DC Voltage -10V ... +10V
---- to
(GROUNDED) (+/Ground) -3,33mV ... +3,33mV
(DC COUPLED)

DC Voltage -10V ... +10V


POTENTIOMETRIC 2.5V, 5V or 10V to
(+/Ground) -1V ... +1V

FULL BRIDGE
(STRAIN GAGES, -33,3mV ... +33,3mV
PIEZORESISTIVE)
2.5V for 120Ω
Full Bridge to
10V for ≥ 350Ω -3,33mV ... +3,33mV
LOAD CELL

2.5V for 120Ω -33,3mV ... +33,3mV


HALF BRIDGE 1/2 Bridge to
(STRAIN GAGES) 5V for 350Ω -3,33mV ... +3,33mV

QUARTER BRIDGE 2.5V for 120Ω -33,3mV ... +33,3mV


3 WIRE CONNECTION 1/4 Bridge to
(STRAIN GAGE) 5V for 350Ω -3,33mV ... +3,33mV

QUARTER BRIDGE
Q
E
DC -33,3mV ... +33,3mV
2 WIRE CONNECTION 1/4 Bridge 2.5V for 120Ω to
+

-3,33mV ... +3,33mV


(STRAIN GAGE) G 5V for 350Ω
Select lowest
THERMOCOUPLE Thermocouple suitable
---- temperature
(DIFFERENTIAL) (+/-)
range
Select lowest
THERMOCOUPLE Thermocouple suitable
---- temperature
(GROUNDED) (+/Ground)
range
Select lowest
Pt100
suitable
Pt100 (Resistance) 5V temperature
(Temperature) range

H8707-M190009-0
EN

1
ADS1500 SHORT USER´S GUIDE
SENSOR CONNECTIONS

SENSOR TYPE CONNECTION INPUT MODE EXCITATION SCALE

AC VOLTAGE SOURCE AC Voltage -10V ... +10V


(GROUNDED) ---- to
(AC COUPLED)
(+/Ground) -3,33mV ... +3,33mV

IEPE ACCELEROMETER -10V ... +10V



(ICP , Deltatron ) IEPE ---- to
-100mV ... +100mV

IEPE MICROPHONE -10V ... +10V



(ICP )
IEPE ---- to
-100mV ... +100mV

ACTIVE CURRENT
SOURCE Current ---- -20mA ... +20mA
(DIFFERENTIAL)
(NOT GROUNDED)

ACTIVE CURRENT
SOURCE Current ---- -20mA ... +20mA
(NON-DIFFERENTIAL)
(GROUNDED)

PASSIVE CURRENT
SOURCE Current ---- -20mA ... +20mA
(CURRENT TRANSMITTER
WITH EXTERNAL POWER)

PASSIVE CURRENT
SOURCE Current 24V -20mA ... +20mA
(CURRENT TRANSMITTER
POWERED BY ADS1500)

SWITCH Switch ---- -1V ... +1V

OPEN COLLECTOR
(NPN OUTPUT)
Switch ---- -1V ... +1V

INDUCTIVE ROTATION DC Voltage -10V ... +10V


SENSOR ---- to
(+/Ground) -100mV ... +100mV
(MAGNETIC PICKUP)

H8707-M190009-0
EN

2
ADS1500 SHORT USER´S GUIDE
SENSOR CONNECTIONS

SENSOR TYPE CONNECTION INPUT MODE EXCITATION SCALE


DC Voltage or -10V ... +10V
MEMS
AC Voltage 5V to
ACCELEROMETER
(+/Ground) -1V ... +1V

DC Voltage -10V ... +10V


INCLINOMETER 5V to
(+/Ground) -1V ... +1V

RESISTANCE
Resistance ---- 10kΩ to 10MΩ
THERMISTOR

H8707-M190009-0
EN

3
ADS1500 SHORT USER´S GUIDE
SENSOR CONNECTIONS

COUNTING, FREQUENCY AND PERIOD


SENSOR TYPE CONNECTION PULSE OPERATION MODE

A/B: Quadrature Counting


ENCODER
(x1, x2, x4)

SIMPLE PULSE Frequency


GENERATOR Period

PULSE A: Counting B: Enable R:


GENERATOR WITH Reset
ENABLE AND (Enable active high; Reset active
RESET low)

GROUNDING AND SHIELDING ADVICE


CONFIGURATION SENSOR SIDE ADS1500 SIDE

- CABLE SHIELD NOT CONNECTED


TO SENSOR

- SENSOR CONNECTED OR NOT TO


EARTH

- CABLE SHIELD CONNECTED TO


SENSOR

- SENSOR NOT CONNECTED TO EARTH

- CABLE SHIELD CONNECTED TO


SENSOR

- SENSOR CONNECTED TO EARTH

H8707-M190009-0
EN

4
ADS1500 SHORT USER´S GUIDE
PANEL LEDS

SYSTEM LED MODE STATUS


OFF OFF

FAST BLINK PROGRAM LOADER RUNNING

SLOW BLINK STARTUP / SELF-TEST

ON READY

FAST BLINK PRODUCT IDENTIFICATION

SLOW BLINK RECORDING / WAITING FOR TRIGGER

ON BUSY (SENDING DATA)

FAST BLINK HARDWARE / SOFTWARE FATAL ERROR

SLOW BLINK PROGRAM LOADER FAILURE

ON RECOVERABLE ERROR

SYNC LED MODE STATUS


OFF OFF / SELF-TEST (SEE SYSTEM LED)

FAST BLINK PROGRAM LOADER RUNNING

SLOW BLINK SYNCHRONIZED

ON SYNCHRONIZED
1. SYNCHRONIZING (PHASE 1)
2. PRODUCT IDENTIFICATION
FAST BLINK
3. IF PROGRAM LOADER IS RUNNING (SYSTEM LED FAST
BLINKING GREEN), NO VALID PROGRAM WAS FOUND
SLOW BLINK SYNCHRONIZING (PHASE 2)

ON SYNCHRONIZING (PHASE 3)

FAST BLINK SOFTWARE FATAL ERROR

SLOW BLINK --------

ON EXTERNAL SYNC, INTERNAL CLOCK OR GPS FAILURE

ACT FRONTAL LED MODE STATUS


OFF POWER OFF

SLOW BLINK STARTUP / SELF-TEST

ON NORMAL OPERATION

FAST BLINK PRODUCT IDENTIFICATION

DOUBLE BLINK HARDWARE / SOFTWARE ERROR

H8707-M190009-0
EN

5
ADS1500 SHORT USER´S GUIDE
POWER-ON OPTIONS

SWITCH SW LED SIGNALING EFFECT


PRESSED AFTER POWER-ON
FORCE FACTORY DEFAULT IP ADDRESS (192.168.1.11x)
AND HOLD FOR 10 sec.

HOLD FOR 10 sec. MORE FORCE MODEL DEFAULT IP ADDRESS (192.168.1.115)

FORCE DEFAULT CONFIGURATION DATA


HOLD FOR 10 sec. MORE FORCE MODEL DEFAULT IP ADDRESS (192.168.1.115)
ATTENTION: PERMANENT CHANGES

SWITCH SW LED SIGNALING EFFECT


PRESSED BEFORE POWER-ON ENTER PROGRAM LOADER MODE
AND HOLD TO EXIT, TURN OFF EQUIPMENT

FORCE MODEL DEFAULT IP ADDRESS (192.168.1.115) AND


HOLD FOR 10 sec. MORE AFTER
ENTER PROGRAM LOADER MODE
POWER-ON
TO EXIT, TURN OFF EQUIPMENT

DIGITAL INPUT/OUTPUT CONECTOR