Professional Documents
Culture Documents
Doppler Angle and Flow Velocity Estimations Using The Classic and Transverse Doppler Effects
Doppler Angle and Flow Velocity Estimations Using The Classic and Transverse Doppler Effects
Doppler Angle and Flow Velocity Estimations Using The Classic and Transverse Doppler Effects
4RY 1999
rres ce -
Doppler Angle and Flow Velocity Estimations the center of the conduit, the mean frequency equals t o the
Using the Classic and Transverse energy peak frequency theoretically The mean frequency, how-
Doppler Effects ever, strongly depends on the velocity gradient of the conduit,
and it is likely to be affected by the wall-thump filter or low
Bor-Ray Lee, Huihua Kenny Chiang", Member, IEEE, frequency disturbance for low-velocity flow
Cheng-Deng Kuo, Win-Li Lin, a n d San-KanLee The classic Doppler equation (1) can be used to estimate
a blood flow velocity, if its Doppler angle 0 is known In clin-
Abstract-Current clinical Doppler ultrasound systems ical applications, the Doppler angle can be determined from
could only measure the flow vector parallel to the ultra- the duplex scan image of ultrasound The scan image shows
sound beam axis, and the knowledge of the Doppler angle the Doppler ultrasound beam axis and the Doppler sample vol-
(beam-to-flow angle) is needed to calculate the real flow ve- ume on the B-scan tissue image, including the image of the
locity. Currently, the Doppler angle is determined visually blood vessel [3] However, many difficulties are involved in de-
by manually aligning a vessel axis marker along the blood termining the Doppler angle For example, it may be difficult to
vessel on the duplex scan image of the ultrasound. The ap- obtain a longitudinal cross-section scan of a blood vessel in the
plication of this procedure is often limited by practical con-
ultrasound scanning plane to determine the Doppler angle In
straints; therefore, measurements are not reliable. In order
to overcome this problem, we developed a simple Doppler addition, it is difficult to determine a three-dimensional (3-D)
angle and flow velocity estimation method using a combi- Doppler angle from a two-dimensional (2-D) ultrasound scan
nation of the classic and transverse Doppler effects. This image.
method uses only a single focused annular array transducer Recently, several Doppler ultrasound methods for 2-D and
to estimate the Doppler angle and the flow velocity. We 3-D flow velocity estimation have been developed These tech-
have verified experimentally that this method is successful niques used two transducers to measure the 2-D flow vector,
for measuring constant flow in a flow phantom between 4 5 O and three transducers for 3-D measurements [4]-[6] Newhouse
and 80' Doppler angle. The standard deviation of the esti- et al [a], [7]and Censor et al [SIused two transducers to esti-
mated Doppler angles is less than 4.5O. This method could
mate the 3-D flow vector by supplementing the classic Doppler
be implemented easily in medical Doppler ultrasound sys-
equation (1) with the transverse Doppler equation (2) The
tems to automatically estimate the Doppler angle and the
flow velocity. need of using at least two Doppler transducers limit its imple-
mentation in medical ultrasound systems Thus, an automatic
flow velocity estimation has not been implemented yet in cur-
rent medical Doppler ultrasound systems
1. INTRODUCTION
Newhouse et a1 [7],Censor et a1 [8], and Newhouse and Reid
[9] have exploited the characteristics of the Doppler ultrasound
EDICAL ULTRASOUND uses the Doppler effect, which as- system using a focused transducer to transmit a focused ultra-
sumes an infinitely wide, single-frequency ultrasound sound beam, and they have established the transverse Doppler
beam crossing a uniform flow to measure the flow velocity. theory describing the relationship between the flow velocity v
The ultrasound beam is scattered from particles carried by the and the bandwidth of the Doppler spectrum Bd
flow, and the ultrasound frequency is changed by the amount of
velocity-induced Doppler shift frequency. The classic Doppler
equation is given by:
Authorized licensed use limited to: Bursa Uludag Universitesi. Downloaded on September 03,2021 at 10:42:58 UTC from IEEE Xplore. Restrictions apply.
LEE et al.: FLOW VELOCITY AND TRANSVERSE DOPPLER EFFECTS 253
111. EXPERIMENTS
A. Doppler System
Authorized licensed use limited to: Bursa Uludag Universitesi. Downloaded on September 03,2021 at 10:42:58 UTC from IEEE Xplore. Restrictions apply.
IEEE TRANSACTIONS ON ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL, VOL. 46, NO. 1 ~ JANUARY 1999
500
n
2 450
v
+,400
0
5 350
2 300
L4
250
f 200
-0
150
2E 100
\
!
3
200
I
400 600 800
-6dB
1000 1200 1400
50
0
7
45
25.2 cm/s
50 55 60 65 70 75 80 85
Frequency (Hz) Measured Angle
.- 100
E.Flow Measurement
2 50
Two different flows were generated in the UHDC flow phan- 0- 1 I
tom. The measured maximum Doppler frequency fmax and the 45 50 55 60 65 70 75 80 85
energy peak frequency f d of these two flows in 12 measure- Measured Angle
ments are shown in Figs. 3(a) and (b), respectively. The solid
lines represent the mean value, and the error bars represent
the standard deviations. We observed that fmax and f d de-
crease with the increase of the Doppler angle. After finax and Fig. 3. The measured (a) maximum frequency fmax, and (b) energy
fd were measured, the Doppler angles and flow velocities were peak frequency fd of the 18.9 and 25.2 cm/s flows.
estimated using (4) and ( 5 ) , respectively. The calculated re-
sults are shown in Figs. 4 and 5 . The standard deviation of the
estimated Doppler angles is less than 4.5", and the standard is estimated visually by using a vessel axis marker on the ul-
deviation of the estimated flow velocities is less than 4.7 cm/s. trasound duplex scan image. Therefore, the current methods
are limited because the vessel orientation needs to be manually
set by the operator, and the vessel must lie in the ultrasound
IV. DISCUSSION scanning plane.
Measurements shown in Figs. 4 and 5 demonstrate that the
We have experimentally demonstrated a Doppler angle application of this method for Doppler angle and flow velocity
and flow velocity estimation method by combining the classic estimation is valid for a wide range of angles between 45" and
Doppler equation (I) and transverse Doppler equation (3). This 80". For Doppler angles larger than 80",the lower frequency
method has two major advantages. It is very simple, only one portion of the Doppler shift frequency spectrum might be fil-
focused transducer is needed. It can be implemented directly tered out by the wall-thump filter of the ultrasound receiver
in current medical Doppler ultrasound system with a clinical Thus, the energy peak frequency f d of the filtered Doppler spec-
annular array transducer. The Doppler angle and flow velocity trum may be overestimated; and the Doppler angle may be
estimation method is very important for obtaining quantita- underestimated, especially for flows with a lower velocity such
tive flow velocity measurements. Currently, the Doppler angle as the 18.9 cm/s flow measured at 85", in Fig. 5. However, a
Authorized licensed use limited to: Bursa Uludag Universitesi. Downloaded on September 03,2021 at 10:42:58 UTC from IEEE Xplore. Restrictions apply.
LEE et al.: FLOW VELOCITY AND TRANSVERSE DOPPLER EFFECTS 255
90 1.0
85
--
80
75 Wall-thump filter
W
?70
$65
I-
(d
.g 60
x
55
-25.2 cm/s
50
45 - 1 - L I .
. ..I
0- I 200 400 600 800 1000 1200 1400
45 50 55 60 65 70 75 80 85 90 Frequency (Hz)
Measured Angle
Fig. 6. The energy peak frequency f d of the Doppler spectrum of the
Fig. 4. The actual (X-axis) and estimated (Y-axis) Doppler angles of 25.2 cm/s flow measured at 60' Doppler angle before and after the
two flows with different flow velocity. wall thump filter.
0.5
n
B 0.45
0.4
*sr
-$
*g 0.35
0.3
0.25
Ag 0.2
E: 0.15
;
t i
Fig. 5 . The calculated maximum flow velocities of two different flows Fig. 7. Three Doppler spectra of the 25.2 cm/s flow measured at 50°,
measured at different Doppler angles (maximum flow velocity # 70°, and 8 5 O Doppler angle. A represents fmax, and B represents fd.
mean flow velocity).
Authorized licensed use limited to: Bursa Uludag Universitesi. Downloaded on September 03,2021 at 10:42:58 UTC from IEEE Xplore. Restrictions apply.
256 IEEE TRANSACTIONS O N ULTRASONICS, FERROELECTRICS, AND FREQUENCY CONTROL, VOL. 46, NO. 1,JANUARY 1999
the Doppler spectrum is affected by the position of sample vol- M. D. Fox and M. Gardiner, “Three-dimensional Doppler ve-
ume. If the sample volume is set at the center of a conduit, the locimetry of flow jets,” IEEE Trans. Biomed. Eng., vol. 35, pp.
effect of changing the size of the sample volume on the vari- 834-841, Oct. 1988.
I. A. Hein, “3-D flow velocity vector estimation with a triple-
ance of fd is less than 8%. Law et al. [14] concluded that fmax beam lens transducer experimental results,” IEEE Trans. Ul-
and f d are affected by the position of the sample volume but trason., Ferroelect., Freq. Contr., vol. 44, pp. 85-89,1997.
to a lesser degree by its size, and the quantification of pulsed V. L. Newhouse, D. Censor, T. Vontz, J. A. Cisneros, and B.B.
Doppler spectrum is feasible if the sample volume is positioned Goldberg, “Ultrasound Doppler probing of flows transverse with
respect to beam axis,” IEEE Trans. Bzomed. Eng., vol. BME-34,
within the central 25% of the tube. pp. 779-789, Oct. 1987.
D. Censor, V. L. Newhouse, and T. Vontz, “Theory of ultra-
sound Doppler-spectra velocimetry for arbitrary beam and flow
V. CONCLUSIONS configurations,” IEEE Duns. Biomed. Eng., vol. BME-35, pp.
740-451, Sep. 1988.
V. L. Newhouse and J. Reid, “Invariance of the Doppler band-
We have demonstrated a Doppler angle and flow velocity width with flow displacement in the illuminating field,” J .
estimation method for constant flow measurement using only Acoust. Soc. Amer., vol. 90, no. 5, pp. 2595-2601, 1991.
one focused annular array transducer. The Doppler angle can Y.Chabria and V. L. Newhouse, “Estimation of axial blood
velocity using the Doppler equation corrected for broadening,”
be computed from the combination of the classic Doppler ef- Ultrasound Med. Biol., vol. 23, pp. 967-968, 1997.
fect (I) and the transverse Doppler effect (3). For pulsatile flow G. Guidi, V. L. Newhouse, and P. Tortoli, “Doppler spectrum
measurements, the Doppler angle and flow velocity need to be shape analysis based on the summation of flow-line spectra,”
estimated in a very short time to obtain accurate results. This IEEE Trans. Ultrason., Ferroelect., Freq. Contr., vol. 42, no. 5,
pp. 907-915, 1995.
problem may be resolved by using fast digital signal process- W. L. McCabe, J. C. Smith, and P. Harriot, Unit Operations
ing techniques, which can be implemented by a customer-built of Chemical Engineering, 5th ed. New York: McGraw-Hill, pp.
circuit to speed up the computation. Further work with more 96-98, 1992.
sophisticated algorithms is needed to apply the Doppler angle J. W. Andrew and W. Junru, “Correction of intrinsic spectral
and flow velocity measurement method for pulsatile flow mea- broadening errors in Doppler peak velocity measurement made
with phased sector and linear array transducer,” Ultrasound
surements. Med. Biol., vol. 21, pp. 1928-1035, 1995.
Y. F. Law, P. A. J. Bascom, E(. W. Johnston, P. Vaitkus, and
R. S.C. Cobbold, “Experimental study of the effects of pulsed
Doppler sample volume size and position on the Doppler spec-
REFERENCES trum,” Ultrasonics, vol. 29, pp. 404-410, Sep. 1991.
G. P. Tortoli, G. Guidi, and P. Pignoli, “Transverse Doppler
P. Tortoli, G. Guidi, and V. L. Newhouse, “Improved blood ve- spectral analysis for a correct interpretation of flow sonograms,”
locity estimation using the maximum Doppler frequency,” Ul- Ultrasound Med. Biol., vol. 19, no. 2, pp. 115-121, 1993.
trasound Med. B i d , vol. 21, pp. 527-532, 1995. S.A. Jones and D. P. Gidedens, “A simulation of transit time
V. L. Newhouse, K. S. Dickerson, D. Cathignol, and J. Y. effects in Doppler ultrasound signals,” Ultrasound Med. Biol.,
Chapelon, “Three-dimensional vector flow estimaiton using two vol. 16, pp. 607-619, 1990.
transducers and spectral width,” IEEE Trans. Ultrason., Ferro- P. Tortoli, G. Guidi, and P. Pignoli, “Transverse Doppler spec-
elect., Freq. Contr., vol. 41, pp. 90-95. tral analysis for a correct interpretation of flow sonograms,” Ul-
P. Fish, Physics And Instrumentation of Diagnostzc Medical Ul- trasound Med. Biol., vol. 19, pp. 115-121, 1993.
trasound. Chichester: Wiley, 1990.
J. R. Overbeck, K. W. Beach, and D. E. Strandness, ‘Vector
Doppler: Accurate measurement of blood velocity in two dimen-
sions,’’ Ultrasound Med. Biol., vol. 18, no. 1, pp. 19-31, 1992.
Authorized licensed use limited to: Bursa Uludag Universitesi. Downloaded on September 03,2021 at 10:42:58 UTC from IEEE Xplore. Restrictions apply.