01 DEC 2001 SERVICE NOTE Page 1 of 6

GIB# 1C Proteus XR/a table SN 3971


Rev. 0 2326496-100

SYSTEM CHECKS - GENERATOR RESISTOR AND JUMPERS

APPLIES TO
Proteus XR/a System.

PROBLEM
If the system has a wrong resistor in the CAN bus, or the JTR jumper on TOMO I/F board or
collimator I/F board is closed, the system will experience intermittent lock ups or errors (error code
may be 80-602). This system check will help to avoid such potential problems.

SOLUTION
Please follow the procedure to do the system checks and remove the wrong resistor or jumper,
especially for these Jedi Serial numbers listed below.

Jedi SN
58555WM4 58573WM7 58600WM8 58619WM8 58639WM6
58556WM2 58574WM5 58601MW6 58620WM6 58640WM4
58557WM0 58575WM2 58602WM4 58621WM4 58641WM2
58558WM8 58576WM0 58603WM2 58622WM2 58642WM0
58559WM6 58577WM8 58604WM0 58623WM0 58643WM8
58560WM4 58578WM6 58605WM7 58624WM8 58644WM6
58561WM2 58579WM4 58606WM5 58625WM5 58645WM3
58562WM0 58580WM2 58607WM3 58628WM9 58646WM1
58563WM8 58581WM0 58608WM1 58629WM7 62033WM6
58564WM6 58582WM8 58609WM9 58630WM5 62034WM4
58565WM3 58583WM6 58610WM7 58631WM3 62035WM1
58566WM1 58584WM4 58611WM5 58632WM1
58567WM9 58585WM1 58612WM3 58633WM9
58568WM7 58595WM0 58614WM9 58634WM7
58569WM5 58596WM8 58615WM6 58635WM4
58570WM3 58597WM6 58616WM4 58636WM2
58571WM1 58598WM4 58617WM2 58637WM0
58572WM9 58599WM2 58618WM0 58638WM8

Copyright 2001 by General Electric Co.


01 DEC 2001 SERVICE NOTE Page 2 of 6
GIB# 1C Proteus XR/a table SN 3971
Rev. 0 2326496-100

PROCEDURE

1. Power off the system.


2. Shut down the facility power breaker.
3. Dismount the front cover of the cabinet by loosing two screws underneath the cover.
Note: Pay attention to the ground wire connecting to the cover.
4. Find the JTR jumper on the collimator I/F board (221-2994). Confirm the JTR jumper is opened.
5. Measure the resistance between U6-6 and U6-7 on the collimator I/F board. The resistance should be 60±5
Ö.
Note: Ensure the power is off and all the cables are well connected while measuring the
resistance.

- If the measured resistance is out of range, then go to step 6 to check the system.
- If the measured resistance is in this range, then go to step 7.

U6

Collimator I/F
Board

6. System checks.

6.1.Check the JTR jumper on the TOMO I/F board (221-2993).

Note: This check is needed only when the system has the TOMO link.

- Open the cover of TOMO I/F board box.


- Check the JTR jumper on the TOMO I/F board (221-2993). Make sure the JTR jumper is opened.

Copyright 2001 by General Electric Co.


01 DEC 2001 SERVICE NOTE Page 3 of 6
GIB# 1C Proteus XR/a table SN 3971
Rev. 0 2326496-100

- Reinstall the box cover.


- Measure the resistance between U6-6 and U6-7 on the collimator I/F board again. If the resistance
is 60±5Ö , then go to step 7; Else, go to step 6.2.

TOMO I/F board box TOMO I/F board (221-2993)

6.2. Check the resistor on the Jedi I/F board.

The Jedi I/F Board is situated in the Interface Unit.

BEFORE ANY MANUAL INTERVENTION, ENSURE THE MAIN POWER IS OFF. APPLY
DANGER
LOCK OUT - TAG OUT PROCEDURE FOR YOUR OWN SAFETY WHEN MANIPULATING
INSIDE THE EQUIPMENT IS REQUIRED.

NOTICE
The removing of the I/F Board can be done without removing the Interface Unit from the
Cabinet.

Copyright 2001 by General Electric Co.


01 DEC 2001 SERVICE NOTE Page 4 of 6
GIB# 1C Proteus XR/a table SN 3971
Rev. 0 2326496-100

• Unscrew twelve 5.5-mm screws that retain the

cover panel and remove this one 1 .

Interface
Unit

• 2
Disconnect the flat ribbon cable and the two
3
cables , if existing, coming from the AEC
Board (option).

• Disconnect the flat wide ribbon cable J10


4
coming from the KV Control Board .

5
3

4
6

The backside of
J1 connector
• Remove the four screws of connectors J8 and J9
5
.

• 6
Remove the ten 5.5-mm screws securing the I/F Board; remove the board.

• Check the backside of the J1 connector, cut the 150Ö resistor if have had.

Copyright 2001 by General Electric Co.


01 DEC 2001 SERVICE NOTE Page 5 of 6
GIB# 1C Proteus XR/a table SN 3971
Rev. 0 2326496-100

• Mount the Jedi I/F board and connect all the cables.

• Mount the cover panel.

6.3. Measure the resistance between U6-6 and U6-7 on the collimator I/F board again. If the resistance is still

out of 60±5Ö (which should be very rare), please forward this issue to

Tanya Quiles-Moreno or Mi, Kai Yun

7. Mount the cabinet front cover.


On-Line Centers:
OLC–Americas OLC–Europe OLC–Asia OLC–ANZ
1–800–321–7937 (USA/CDN) +33 (0)1 30 83 13 00 81–426–48–2940 61–2–3163700
1 (262) 524–5300 (Elsewhere)

Author:
Doris Yu
Documentation Engineer
GEMSA
Headquarters Information:
Are parts in stock? Yes __ No _ _ Available On Demand _ _ N/A _√_

Revision History

REV DATE REASON FOR CHANGE

0 01 December 2001 Initial release.

Copyright 2001 by General Electric Co.


01 DEC 2001 SERVICE NOTE Page 6 of 6
GIB# 1C Proteus XR/a table SN 3971
Rev. 0 2326496-100

This page left blank.