Proteus Room Layout - 22992000.000
Proteus Room Layout - 22992000.000
APPLIES TO
Proteus XR/a System, PSI Code XRA560.
PROBLEM
Some wall stands are installed in a wrong direction.
SOLUTION
A notice will be added into Service Manual when it will be updated. Before this is implemented,
Field Engineers should follow the instruction in this Service Note.
PROCEDURE
PROPER ROOM LAYOUT FOR PROTEUS XR/A
1 - Table
2 - Wall Stand
3 - System Console
4 - Generator Cabinet
5 - XT Suspension
The direction of wall stand must be the same as shown in illustration 5-1, Sevice
NOTICE Manual, Direction 2260326-100. Any other direction will result in conflict with
system functions and will not conform to HHS requiements for Automatic
Collimation Systems. Do NOT attempt to make any field modifications to support
non-standard placement of the wall stand.
If the customer must have the wall stand 90 degrees to the table a system with
MANUAL collimation should be ordered as this configuration is allowable with a
MANUAL Collimator (Not a Automatic Collimator used in Manual mode)