04 September 2001 SERVICE NOTE Page 1 of 4

GIB# 1C Proteus XR/a Wall Stand SN 3957


Rev. 0 S2299200-100

THE PROPER LAYOUT OF PROTEUS XR/A SYSTEM

APPLIES TO
Proteus XR/a System, PSI Code XRA560.

PROBLEM
Some wall stands are installed in a wrong direction.

SOLUTION
A notice will be added into Service Manual when it will be updated. Before this is implemented,
Field Engineers should follow the instruction in this Service Note.

PROCEDURE
PROPER ROOM LAYOUT FOR PROTEUS XR/A
1 - Table
2 - Wall Stand
3 - System Console
4 - Generator Cabinet
5 - XT Suspension

NUMBER SIZE REVISION


2299200-100SNT Letter 0

Copyright 2001 by General Electric Co.
04 September 2001 SERVICE NOTE Page 2 of 4
GIB# 1C Proteus XR/a Wall Stand SN 3957
Rev. 0 S2299200-100

EXAMPLES FOR WRONG DIRECTIONS OF WALL STAND:



The direction of wall stand must be the same as shown in illustration 5-1, Sevice
NOTICE Manual, Direction 2260326-100. Any other direction will result in conflict with
system functions and will not conform to HHS requiements for Automatic
Collimation Systems. Do NOT attempt to make any field modifications to support
non-standard placement of the wall stand.
If the customer must have the wall stand 90 degrees to the table a system with
MANUAL collimation should be ordered as this configuration is allowable with a
MANUAL Collimator (Not a Automatic Collimator used in Manual mode)

NUMBER SIZE REVISION


2299200-100SNT Letter 0

Copyright 2001 by General Electric Co.
04 September 2001 SERVICE NOTE Page 3 of 4
GIB# 1C Proteus XR/a Wall Stand SN 3957
Rev. 0 S2299200-100

FOR MORE INFORMATION


The publications listed are referred to in this Service Note.

DOCUMENT TYPE DOCUMENT # REV # TITLE


Preinstallation Manual 2260326-100 3 Proteus XR/a PIM

On-Line Centers:
OLC–Americas OLC–Europe OLC–Asia OLC–ANZ
1–800–321–7937 (USA/CDN) +33 (0)1 30 83 13 00 81–426–48–2940 61–2–3163700
1 (262) 524–5300 (Elsewhere)

Author:
Guo Feng
Documentation Engineer / X-ray
GEMS-A
Headquarters Information:
Are parts in stock? Yes _ _ No _ _ _ _ N/A _√ _
(If No, Service Note is not to be distributed)

Revision History

REV DATE REASON FOR CHANGE

0 04 September 2001 Initial release.

NUMBER SIZE REVISION


2299200-100SNT Letter 0

Copyright 2001 by General Electric Co.
04 September 2001 SERVICE NOTE Page 4 of 4
GIB# 1C Proteus XR/a Wall Stand SN 3957
Rev. 0 S2299200-100

This page intentionally left blank.

NUMBER SIZE REVISION


2299200-100SNT Letter 0

Copyright 2001 by General Electric Co.