Professional Documents
Culture Documents
1 - 4720.00180A01 E Series Service Manual
1 - 4720.00180A01 E Series Service Manual
P/N: 4720.00180A01
Product Model: E5 Exp/E5/E5 Pro/E3 Exp/E3/E3 Pro/E2 Exp/E2/E2 Pro/E1 Exp/E1/E1 Pro
Release Date: February, 2018
Copyright © 2018 SonoScape Medical Corp. All rights reserved.
Statement
SonoScape Medical Corp. (hereinafter called SonoScape) owns the intellectual property rights to
this manual, and also maintains the contents of this manual as confidential information. This manual
is a reference for the operation, maintenance and cleaning of this product and does not convey any
license under the patent rights of SonoScape, nor the rights of others.
This manual contains the information protected by copyrights or patents, and should only be used
by the service personnel of SonoScape and the authorized agents. Reproduction, amendment or
translation of this manual in any manner whatsoever without the written permission of SonoScape is
strictly prohibited.
All information contained in this manual is believed to be correct. SonoScape shall not be liable for
errors contained herein or for incidental or consequential damages in connection with the furnishing,
performance or use of this manual. SonoScape does not assume any liability arising out of any
infringements of patents or other rights of third parties.
This manual is based on the maximum configuration and therefore some contents may not apply to
your product.
This manual is subject to change without prior notice and legal obligation.
Manufacturer’s Responsibility
SonoScape is responsible for the effects on safety, reliability and performance of this product, only if
all the following requirements are met.
●● All installation operations, expansions, changes, modifications and repairs of this product are
conducted by SonoScape authorized personnel.
●● The use or application of the product or the use of parts or accessories is approved by
SonoScape.
●● The electrical installation of the relevant room complies with the applicable national and local
requirements.
Documentation
Understand the meanings of the following items clearly before reading this manual.
Item Meaning
Boldfaced Indicates controls on the control panel, or on-screen objects such as menu items
Word or keys.
Click Move the cursor to the controls on the display and press the confirm key on the
control panel.
Contact Information
Manufacturer: SonoScape Medical Corp.
Address: 4/F, 5/F, 8/F, 9/F & 10/F, Yizhe Building, Yuquan Road, Nanshan, Shenzhen, 518051,
Guangdong, China
Tel: +86 755 2672 2890
Fax: +86 755 2672 2850
http://www.sonoscape.com
E-mail: sonoscape@sonoscape.net
Contents
1 Safety���������������������������������������������������������������������������������������������������������������������������������������� 1
1.1 Intended Use����������������������������������������������������������������������������������������������������������������������������2
1.2 Safety Precautions�������������������������������������������������������������������������������������������������������������������2
1.2.1 Electrical Safety������������������������������������������������������������������������������������������������������������2
1.2.2 Mechanical Safety���������������������������������������������������������������������������������������������������������4
1.2.3 Accessories Caring��������������������������������������������������������������������������������������������������������4
1.2.4 Biohazard Considerations���������������������������������������������������������������������������������������������5
1.3 Acoustic Power Principle��������������������������������������������������������������������������������������������������������5
1.3.1 Biological Safety����������������������������������������������������������������������������������������������������������5
1.3.2 ALARA�������������������������������������������������������������������������������������������������������������������������5
1.3.3 Mechanical and Thermal Indices����������������������������������������������������������������������������������6
1.3.4 Transducer Surface Temperature Limits�����������������������������������������������������������������������7
1.3.5 Imaging Functions that Change Acoustic Output���������������������������������������������������������8
3 Principle Descriptions����������������������������������������������������������������������������������������������������������� 19
3.1 System Block Diagram����������������������������������������������������������������������������������������������������������20
3.2 Hardware Description������������������������������������������������������������������������������������������������������������21
3.2.1 ABF Board������������������������������������������������������������������������������������������������������������������21
3.2.2 Keyboard���������������������������������������������������������������������������������������������������������������������23
3.2.3 Probe Board����������������������������������������������������������������������������������������������������������������23
3.2.4 Power Port Board��������������������������������������������������������������������������������������������������������24
3.2.5 Battery Port Board������������������������������������������������������������������������������������������������������24
I
Contents
4 Parts Disassembly����������������������������������������������������������������������������������������������������������������� 27
4.1 Battery Disassembly��������������������������������������������������������������������������������������������������������������28
4.2 Bottom Components Disassembly����������������������������������������������������������������������������������������28
4.3 Back Cover Disassembly�������������������������������������������������������������������������������������������������������29
4.4 Control Panel Disassembly���������������������������������������������������������������������������������������������������30
4.5 LCD Monitor Disassembly���������������������������������������������������������������������������������������������������32
4.6 Fan Disassembly�������������������������������������������������������������������������������������������������������������������34
4.7 Grounding Cable and S-Video Disassembly�������������������������������������������������������������������������35
4.8 Battery Board Bracket Disassembly��������������������������������������������������������������������������������������35
4.9 ECG Components Disassembly��������������������������������������������������������������������������������������������36
4.10 Probe Board Disassembly���������������������������������������������������������������������������������������������������38
4.11 Hard Disk Disassembly�������������������������������������������������������������������������������������������������������39
4.12 SD Card Disassembly����������������������������������������������������������������������������������������������������������40
4.13 Speakers Disassembly���������������������������������������������������������������������������������������������������������40
4.14 ABF Board Disassembly�����������������������������������������������������������������������������������������������������41
5 Wiring Descriptions�������������������������������������������������������������������������������������������������������������� 43
5.1 Connections for Wiring���������������������������������������������������������������������������������������������������������44
5.2 Pin Connection����������������������������������������������������������������������������������������������������������������������45
5.2.1 Keyboard - TGC Board Signal Cable�������������������������������������������������������������������������45
5.2.2 ABF Board - Display EDP Cable��������������������������������������������������������������������������������45
5.2.3 ABF Board - Keyboard Signal Cable�������������������������������������������������������������������������46
5.2.4 ABF Board - ECG Board Signal Cable����������������������������������������������������������������������47
5.2.5 Adaptor Converter - Battery Board Cable������������������������������������������������������������������48
5.2.6 ABF Board - SATA Hard Disk Cable�������������������������������������������������������������������������48
5.2.7 Keyboard - Trackball Board Signal Cable������������������������������������������������������������������49
5.2.8 ABF Board - ON/OFF Button Board Signal Cable����������������������������������������������������49
5.2.9 ECG Board - BNC Socket Cable��������������������������������������������������������������������������������50
5.2.10 Keyboard - Indicator Board Cable����������������������������������������������������������������������������51
5.2.11 ABF Board - BNC Socket Cable������������������������������������������������������������������������������51
II Service Manual
Contents
7 Settings Import/Export��������������������������������������������������������������������������������������������������������� 63
7.1 Customizing a Preset�������������������������������������������������������������������������������������������������������������64
7.2 Exporting a Customized Preset���������������������������������������������������������������������������������������������64
7.3 Importing a Customized Preset���������������������������������������������������������������������������������������������65
8 Performance Test������������������������������������������������������������������������������������������������������������������� 67
8.1 Automatic Hardware Tests����������������������������������������������������������������������������������������������������68
8.1.1 Test Steps��������������������������������������������������������������������������������������������������������������������68
8.1.2 Underlying Hardware Tests����������������������������������������������������������������������������������������69
8.1.3 Peripheral Device Test������������������������������������������������������������������������������������������������71
8.2 Function Settings�������������������������������������������������������������������������������������������������������������������74
8.3 Firmware Update�������������������������������������������������������������������������������������������������������������������77
8.4 Burning Function�������������������������������������������������������������������������������������������������������������������78
9 System Troubleshooting�������������������������������������������������������������������������������������������������������� 79
9.1 Fault Diagnostics�������������������������������������������������������������������������������������������������������������������80
9.2 System Troubleshooting��������������������������������������������������������������������������������������������������������82
9.2.1 On-Site Inspection������������������������������������������������������������������������������������������������������82
9.2.2 Fault Identification������������������������������������������������������������������������������������������������������82
9.2.3 Fault Resolution����������������������������������������������������������������������������������������������������������83
9.2.4 Post Maintenance �������������������������������������������������������������������������������������������������������83
10 System Maintenance����������������������������������������������������������������������������������������������������������� 85
10.1 Cleaning the System������������������������������������������������������������������������������������������������������������86
IV Service Manual
1 Safety
This chapter describes the important information for operating this system. To ensure the safety of
both operator and patient, please read the relevant details in this chapter carefully before using this
system.
You should be thoroughly familiar with the precautions provided in this manual. Otherwise, the
manufacturer is not responsible for the effects on safety, reliability and performance of the system.
1
1 Safety
The system also provides the measurement and calculation packages used for clinical
diagnosis purposes. For details, please refer to the advanced user manual.
Contraindication: The system is not intended for ophthalmic use or any use causing the
acoustic beam to pass through the eye.
Precautions must be considered in the use of any application. Otherwise, it may result in
! system damage or serious injury.
2 Service Manual
1 Safety
●● The video printer should be connected to the specific interface by using the cable
provided by the manufacturer. Otherwise, electrical shock may occur.
●● Within the environment that is 1.8 meters (6 feet) around a patient, connect peripherals
to the auxiliary power outlet which is capable of isolation protection; or, power the
peripherals by the auxiliary output cable or the isolation transformer complied with
EN/IEC 60601-1, or the power input of the same safety level.
●● Within the patient environment, when removing the covers or the connectors from the
non-medical electrical equipment without tools during maintenance, calibration or
other operations, do not touch the parts of the non-medical electrical equipment and
the patient simultaneously.
●● Use only the probes provided by the manufacturer. Otherwise, the system will not
perform, and an accident such as a fire may result in the worst case.
●● Only the peripherals and accessories provided or recommended by the manufacturer
can be used. Using other devices or accessories may degrade the system performance
and even cause electrical shock.
●● Do not pour any fluid onto the system surfaces, as fluid seepage into the electrical
circuitry may cause excessive electrical current leakage or system failure. If any
water is spilled onto the system carelessly, stop using the system and contact the local
distributor immediately.
●● The AC power plug for the system is a three-prong grounded plug and should never
be adapted to any two-prong outlet or by using an adapter. Connect the AC power
plug of the multiple socket-outlet to a hospital-grade power outlet.
Service Manual 3
1 Safety
●● Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective EN/IEC standards (for example, EN/IEC 60950 for data
processing equipment and EN/IEC 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the system standards EN/IEC 60601-1.
●● Do not use an endocavitary probe on the body surface of a patient. Otherwise, the
electromagnetic compatibility of the probe may be degraded.
●● If the non-medical electrical equipment used with the system is supplied by a multiple
socket-outlet with separating transformer, connect the AC power plug of the multiple
socket-outlet to a hospital-grade power outlet. Consult a professional to ensure that
the connection complies with the safety standards.
●● To avoid possible personnel injury and system damage, move the system slowly and
! carefully.
●● Do not place other objects on top of the control panel. Do not sit on the control panel
or any other part of the system.
●● Disconnect the foot switch and the power cable before moving the system.
●● Do not knock or shake the system.
●● Always use the handle to move the system.
4 Service Manual
1 Safety
●● Use the legally marketed coupling gel in accordance with relevant local regulations.
Read and understand all precautions in the relevant manual of the coupling gel before
use.
●● Prepare, use, store and dispose the cleaner, disinfectant and sterilant according to the
instructions provided by manufacturers.
●● Do not disconnect the probe during the real-time scan. Otherwise, it may damage the
probe and the system.
●● Disconnect the probe from the system after freezing an image or powering off the
system. Otherwise, the system or the probe could be damaged.
●● Perform ultrasound procedures prudently under the guidance of the ALARA (as low
! as reasonably achievable) principle. Only expose the patient to the lowest practical
transmit power levels in the shortest possible period to achieve a satisfactory
diagnosis.
●● Freeze the image at any time if you are not operating the system for a long period of
time.
●● Do not scan the same part of an patient continuously or expose the patient to
prolonged scanning. Doing so may harm the patient.
●● Do not expose the fetus to prolonged scanning in the Doppler mode.
●● Although the output power is automatically controlled for the selected applications,
high TI values should be kept to a minimum or avoided in obstetric applications.
●● You should be familiar with the performances and operations of the system, observe
the ultrasound output parameters on the screen at all times.
Service Manual 5
1 Safety
1.3.2 ALARA
It is required to practice ALARA when using ultrasound energy. Practicing ALARA
ensures that the total energy level is controlled below a low enough level at which
bioeffects are not generated while diagnostic information is being accumulated. The total
energy is controlled by output intensity and total radiation time. The output intensity
necessary for examinations differs depending on the patient and the clinical case.
Not all examinations can be performed with an extremely low level of acoustic energy.
Controlling the acoustic level at an extremely low level leads to low-quality images or
insufficient Doppler signals, adversely affecting the reliability of the diagnosis. However,
increasing the acoustic power more than necessary does not always contribute to an
increase in quality of information required for diagnosis, rather increasing the risk of
generating bioeffects.
The operator must take responsibility for the safety of patients and utilize ultrasound
deliberately. Deliberate use of ultrasound means that output power of ultrasound must be
selected based on ALARA. Additional information regarding the concept of ALARA and
the possible bioeffects of Ultrasound is available in a document from the AIUM (American
Institute of Ultrasound Medicine) title “Medical Ultrasound Safety”.
■ MI/TI Explanation
6 Service Manual
1 Safety
■ MI/TI Display
TI and MI values are displayed in real time on the screen. The operator should observe
these index values during examinations and ensure that exposure time and output values
are maintained at the minimum amounts needed for effective diagnosis.
Service Manual 7
1 Safety
The maximum surface temperature of the endocavitary probes is 43°C. To protect the
patient against the harm of excessive temperature, the probe stops working automatically
when its temperature reaches the limit. The surface temperature of the probe is displayed
on the screen.
There are three categories of system controls relative to output. They are controls that
have direct effect on the output, controls that indirectly control output and controls that are
receiver controls.
8 Service Manual
2 Unpacking and System Checks
This chapter describes how to unpack, assemble, connect, power on/off this system, and system
checks are also detailed herein. The unpacking, general and functional checks are necessarily
performed after you assemble the system for the first time. The functional checks are necessarily
performed after you service the system.
9
2 Unpacking and System Checks
2.1 Unpacking
Before unpacking, ensure that the package is complete and take photos of the package. If
there is any problem about the package or the appearance of the system, send the photos to
the customer service department of the manufacturer.
One pair of scissors (or other specialized unpacking tools) and a knife are needed during
unpacking.
Binding Straps
Film
2. Use the knife to split adhesive tape, and open the carton, as shown in Figure 2-2.
10 Service Manual
2 Unpacking and System Checks
■ With Backpack
1. Open the carton, remove the foams, and take out the user manuals and the accessories
in sequence.
1 2
Foam
3 4
2. Take out the probe box, left foam, right foam and backpack. Open the backpack,
remove the backpack belt and take out the system in sequence. As shown in
Figure 2-4.
1 2 3
Probe box
Left Right
foam foam
4 5 6 7
Backpack
Backpack
belts
Service Manual 11
2 Unpacking and System Checks
NOTE:
• Take out the system carefully.
• Do not scratch or hit the surface of the system.
■ Without Backpack
1. Open the carton, and take out the foam, user manuals, accessories and accessory box
in sequence , as shown in Figure 2-5.
1 2 3
Upper foam
4 5 6
Assessory box
2. Take out the right foam, left foam in sequence, as shown in Figure 2-6.
1 2
Right foam
3 4
Left foam
12 Service Manual
2 Unpacking and System Checks
NOTE:
Ensure that the system works with the required voltage before you connect it to the
mains supply. The voltage and power supply of this system are 100-240V~, 50/60Hz,
1.5 - 0.75A.
Connect one end of the adaptor to the system firmly and connect the other end of the
adaptor to the mains supply firmly.
Press the button on the right side of the screen to power on the system.
If the system is idle for a set period of time, the system enables the screen saver
automatically. Press any key to exit the screen saver. For details, refer to System Settings
of the basic user manual.
●● Press the button and click Standby on the pop-up menu to enter the standby
mode.
Press the button in standby mode to exit the standby mode.
●● Close the upper cover of the system, and the system automatically enters standby
mode.
Open the upper cover of the system, and the system automatically exits the standby
mode.
●● Set automatical standby time. If the system is idle for a set period of time, the system
will automatically enter the standby mode.
Press the button in standby mode, and the system automatically exits the standby
mode.
Press the button and select Shutdown on the pop-up menu to power off the system.
Service Manual 13
2 Unpacking and System Checks
You should check the LCD monitor and main unit covers, and ensure that there are not any
scratches or damages after unpacking.
■ To Check Accessories
You should check all accessories according to the packing list, and ensure that all
accessories are of the correct types and PN.
■ Overview
You can check the following items or log with the user to make diagnostics.
●● Any exceptions occurred during system running
●● Accidental exception
●● Other requirements proposed by the user
■ Operating Environment
You should check the operating environment of the system. The measurement accuracy is
closely related to humidity, measurement application and circuit.
14 Service Manual
2 Unpacking and System Checks
■ System Configuration
You should check the system configuration, including system settings, brightness and
contrast of the LCD monitor and date and time.
Item Inspection
System settings Press the Setup key to display the System Settings screen
and check the relevant settings by viewing this menu.
Brightness and contrast You can adjust the brightness of the LCD monitor on the
of the LCD monitor System Settings screen. LCD monitor’s brightness and
contrast is one of the most important factors for obtaining
an optimum image. If the settings is incorrect, you need to
frequently adjust the gain, TGC, dynamic range and output
power supply. Therefore, the lowest level of black should
just disappear into the background and the highest white
should be bright, but not saturated.
Time and Date Check whether the time and date displayed are correct. If
not, you need to adjust them.
■ System Information
Press the Setup key on the key panel to enter the System Settings screen and click About
to view the system information.
Service Manual 15
2 Unpacking and System Checks
■ Workflow
Check Trackball
检查常规操作
Check Fan
检查常规操作
Check Probe
检查常规操作
■ Checklist
Item Inspection
Keyboard and Keys ●● Keys can respond to the up-down and left-right functional
tests.
●● The Print key can respond when the video printer is
connected.
16 Service Manual
2 Unpacking and System Checks
Item Inspection
Probe 1. Connect the probe (such as L741) to the system and ensure
that the screen displays the correct icons for the probe.
2. Select a preferred probe and exam type, and the system
automatically enters the B mode.
3. Press the CFM key to enter the CFM mode, and ensure that
color ROI is displayed on the screen.
4. Adjust the color gain to the maximum to check whether color
noise is normal.
5. Press the Freeze key, replay the cine by using the trackball to
check whether data display correct.
6. Press the M key to enter the inactivated M mode, press the
Update key to ensure that the M trace is displayed, and TGC,
depth and frequency can be adjusted.
7. Press the PW key to enter the inactivated PW mode, press
the Update key to ensure that the spectral Doppler trace is
displayed and no nosies exist.
8. Press the THI key to ensure that the system enters the THI
mode.
9. Disconnect the probe from the system. Select another probe
to repeat the above steps.
■ B Image Test
Adjust the B gain to the maximum in accordance with depths or frequencies, and check the
following items in the B mode.
Service Manual 17
2 Unpacking and System Checks
●● The initial echo field of the image should be displayed in an appropriate brightness
without any aberrations.
●● No unexpected stripes caused by artifacts are displayed on the image.
Adjust the B gain to the maximum in accordance with depths or frequencies, and check the
following items in the THI mode.
●● The initial echo field of the image should be displayed in an appropriate brightness
without any aberrations.
●● No unexpected stripes caused by artifacts are displayed on the image.
Adjust the B gain and ROI to the maximum with the default depth and frequency, adjust
the CFM/PDI gain until few color interferences occur. In this case, there are no unexpected
stripes on the image. However, it is normal if some color noises appears accidentally.
Adjust the PW/CW gain in the real time PW/CW mode, until a clear spectral Doppler
image displays. Adjust the pulse repetition frequency, and no noises appears on the image.
18 Service Manual
3 Principle Descriptions
This chapter provides hardware and software principle descriptions, you should read these
descriptions thoroughly before attempting to service the system.
19
3 Principle Descriptions
Main unit consists of single boards, including the ABF board, probe board, ON/OFF button
board, battery port board, power port board, interface extension board, keyboard, TGC
board, trackball board and WiFi board, and relevant electronic components such as LCD
monitor, lithium battery, hard disk, fan, speaker and so on.
Battery
SM BUS
PWR
WIFI&BT HDD/SSD
Fan1、2 Fan3、4
+12V Vehicle Battery Port Board
Adapter
PWR
SATA
SDIO &
UART
Power
PWR SM BUS
PCN
Port
PWR_AC PWR
Adapter Borad PWR
Port1
FSW FSW_IN
Extend ECG_USB
Board USB_OTG
ECG 3Lead
RS232
Port2
Video MPC
ON/OFF ONOFF Main Board
ON/OFF_IN Board
HDMI
S-Video
Left/Right
USB *3
Ethernet
Audio
Print_ctrl/busy Port3
Reset
Other Signals
USB
eDP
PWR
Speaker L Speaker R
LCD
Keyboard
TGC
Track Ball
Board
Board Function
ABF board It is the main control board of the system that controls the
sending of the ultrasound pulse, the reception and display of
the echo signal and so on.
It consists of the power supply unit, ultrasound unit (including
FPGA and AFE), main control unit ( including CPU unit and
its peripheral circuits) and so on.
20 Service Manual
3 Principle Descriptions
Board Function
Keyboard, TGC board These three single boards constitute the key panel module,
and trackball board providing an input platform for the user.
WiFi board A single board that provides the WiFi connection function.
Probe board A board with three probe ports connects the ABF board and
the probes.
Power port board Connect the adaptor, the ABF board, the battery board and
the ABF board, switching power supply from the adaptor to
the battery, or from the battery to the adaptor.
Battery port board Connect the battery and the power port adaptor, switching
the battery status from charging to discharging, or from
discharging to charging.
ON/OFF button board Connect to the ABF board, powering the ultrasound system
on or off.
Interface extension Connect the ECG electrode connectors, the foot switch and
board the video cable.
USB_OTG port and RS232 port on the board can be used to
debug the mainframe.
Power supply unit includes power supply charge and discharge management circuit,
voltage conversion circuit, power supply MCU control circuit and removable lithium
battery module.
Ultrasound unit includes FPGA module, AFE module, high-voltage switch module and so
on.
Main control unit, the core control unit of ABF board, is used to process the data, show
the processed data on the monitor and store the data. It consists of CPU and its peripheral
circuit.
Service Manual 21
3 Principle Descriptions
Lithium
MCU
battery
Ports of the ABF board includes the probe board port, the adaptor port, the peripheral
device ports and so on. The peripheral device ports consist of the USB port, the keyboard
port, the LCD monitor port, the audio output port, the network port, the interface extension
port and so on. All ports of the ABF board are listed in Table 3-2.
No. Name
J6 Keyboard port
J8 Power port
22 Service Manual
3 Principle Descriptions
No. Name
P1 USB port
P2 USB port
P7 S-VIDEO port
P9 USB port
3.2.2 Keyboard
Keyboard, TGC board and trackball board constitute an interactive control panel. This
control panel consists of ultrasound-specific keys, general alphabetic and numeric keys,
knobs, TGC sliders, trackball and so on. The keyboard consists of the USB port circuit,
MCU processing circuit, power supply module, LED indicator unit, TGC scan unit, key
processing module, knob encoder module and trackball module.
The keyboard and ABF board are connected by cables such as USB signal cable, I/O signal
control cables and power cable, and communicate with each other through the USB port.
Service Manual 23
3 Principle Descriptions
3.3.1 Structure
Software system consists of hardware layer, operating system layer, intermediate layer and
application layer.
Application
Ultrasound system applications + Associated applications
layer
Intermediate
layer File system Network protocol OpenGL, QT
Operating
LinuxOS
system layer
Hardware
layer Computer platform + Ultrasound module + Control panel
System module can be divided by applications into imaging mode, system setting and
system function, as shown in Figure 3-4.
24 Service Manual
3 Principle Descriptions
Patient
B M(B + M) General Annotation Power On/Off
Management
Probe
CFM PW(B+PW) Body Mark Peripheral Report
Identification
PHI /THI Color +Doppler DICOM Load Default Annotation Body Mark
Image
TDI CFMM About Scanning Mode
Parameters
Measurements
Image Storage
/Calculation
System
Printer
Upgrade
Figure 3-4 System Module
3.3.2 Workflow
Software system is an system program on basis of Linux operating system. Its workflow is
shown in Figure 3-5.
PowerOn
Start uboot
Start kernel
Start application
Figure 3-5 Workflow
Service Manual 25
3 Principle Descriptions
1. Press the button on the control panel of the system to power on the system..
2. Start up uboot to check whether the hardware works normally.
3. If the hardware works normally, start up the kernel of Linux operating system.
4. Start up file system and various service processes.
5. Start up the software.
You can also control the system by operating and checking interfaces.
26 Service Manual
4 Parts Disassembly
This chapter contains information for disassembling the system. You can assemble the system
by performing disassembly procedures on the contrary. During the disassembly, cabling and
wiring should be paid great attention to, and disassembled parts should be taken care of. If you
are not familiar with the replacement procedures, you are recommended to take photos during the
disassembly and take them for assembly reference.
27
4 Parts Disassembly
NOTE:
Switch the battery lock to the position, and remove the battery according to the arrow
direction as shown in Figure 4-1.
2
1
Use the screwdriver to remove the screws marked with numbers, as shown in Figure 4-2.
1 2
3
9
4
8
5 6 7
28 Service Manual
4 Parts Disassembly
1 2 3
4
5
6 7
2. Use the screwdriver to remove the screws marked with numbers, as shown in
Figure 4-4.
Service Manual 29
4 Parts Disassembly
1 2 3
4 5
6 7
3. Use the screwdriver to remove the screws marked with numbers and remove the back
cover, as shown in Figure 4-5.
1 3
2 4
30 Service Manual
4 Parts Disassembly
1 2
3 4
5 6
2. Use the screwdriver to remove the screws marked with numbers, as shown in
Figure 4-7.
1 2
3 4
5 6
3. Slightly pull the cover marked with numbers with fingers and remove the upper cover
according to the arrow direction, as shown in Figure 4-8.
Service Manual 31
4 Parts Disassembly
1 2
4. Use the screwdriver to remove the screws for grounding cable, pull out the cable, and
remove the control panel.
Cable
32 Service Manual
4 Parts Disassembly
2. Remove the acetate tape of the cable connected to the display and pull out the wires.
Service Manual 33
4 Parts Disassembly
3. Use the screwdriver to remove the hinge screws marked with numbers and remove the
components from the display, as shown in Figure 4-12.
1 3
2 4
1 2 7 8
3
4 5 6
34 Service Manual
4 Parts Disassembly
1 2
Service Manual 35
4 Parts Disassembly
1 2
5
4
36 Service Manual
4 Parts Disassembly
2
1
3 4 5
Wire
3. Use the screwdriver to remove the screws marked with numbers, pull out the ECG
components, as shown in Figure 4-19.
Service Manual 37
4 Parts Disassembly
1 2
3 4
Bottom cover
1 7
2 8
3 9
4 10
5 11
6 12
2. Use the screwdriver to remove the screws marked with numbers, and remove the
probe board, as shown in Figure 4-21
38 Service Manual
4 Parts Disassembly
1 3
2
5 6 7
1 3
2 4
Service Manual 39
4 Parts Disassembly
SD card slot
2. Switch the SD card lock to left, and lock the SD card slot.
Speaker bracket
1 2 4 6
3 5
Figure 4-24 Removing Speakers
40 Service Manual
4 Parts Disassembly
4 5
14 1
6
15
2
7
16
13 8
17 9
12 3
11 10
Service Manual 41
4 Parts Disassembly
2. Use the screwdriver to remove the screws and bolts, and remove the ABF board.
3 1 2
5 6 7 8
42 Service Manual
5 Wiring Descriptions
This chapter provides descriptions for pin connections between the boards inside the system. You
should get a general overview of pins before you service the system.
43
5 Wiring Descriptions
LCD monitor
ABF board
Hard disk
ECG optional
Keyboard
accessories
ECG board
ON/OFF board
ECG optional accessories
Battery board
Adaptor board
For details about the connection cables, refer to Table 5-1 and the following sections.
44 Service Manual
5 Wiring Descriptions
Cable Material:
Pins:
Color Red Red Black Black White Green Orange White Green Orange
CN1 20 19 18 17 16 15 14 13 12 11
CN2 1 2 3 4 5 6 7 8 9 10
Color White Green Blue Blue Blue Blue Blue Blue Blue Blue
CN1 10 9 8 7 6 5 4 3 2 1
CN2 11 12 13 14 15 16 17 18 19 20
Instructions:
●● Connect CN1 to the TGC board (J1).
●● Connect CN2 to the keyboard (J1).
Cable Material:
Service Manual 45
5 Wiring Descriptions
Pins:
Color Blue Blue Blue Blue Blue Blue Blue Blue Blue Blue Red Red Blue
CN1 1 5 6 4 2 3 7 8 9 10 11 12 13
CN2 2 3 4 5 6 7 8 9 10 11 12 13 14
Color Blue Blue Blue Blue Blue Blue Blue Blue Blue Red Red Red Red
CN1 14 15 16 17 18 19 20 21 22 25 26 27 28
CN2 15 16 17 18 19 20 21 22 23 26 27 28 29
Instructions:
●● Connect CN1 to the ABF board.
●● Connect CN2 to the display.
Cable Material:
Pins:
46 Service Manual
5 Wiring Descriptions
Color Red Red Red Black Black Black Yellow Yellow Yellow White Green White Green White Black
CN1 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29
CN2 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29
Color Shielded Orange/Blue Orange Orange Black Black White Green White Green White Green White Black
Double twisted
CN1 2 4 6 10 12 8 14 16 18 20 22 24 26 28 30
CN2 2 4 6 10 12 8 14 16 18 20 22 24 26 28 30
Instructions:
●● Connect CN1 to the ABF board (J6).
●● Connect CN2 to the keyboard (J1).
Cable Material:
Pins:
Color Black Green Black Black Yellow Brown Black Yellow Yellow Black Yellow Yellow Black Brown Black
CN1 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29
CN2 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29
Color Red Black Green Brown Black Orange Brown Black Yellow Black Orange Brown Black Orange Black
CN1 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
CN2 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30
Instructions:
●● Connect CN1 to the ABF board (J14).
●● Connect CN2 to the ECG board (J2).
Service Manual 47
5 Wiring Descriptions
Cable Material:
Pins:
CN1 1 3 5 7 9 11
CN2 1 3 5 7 9 11
CN1 2 4 6 8 10 12
CN2 2 4 6 8 10 12
Instructions:
●● Connect CN1 to the adaptor converter (J1).
●● Connect CN2 to the battery board (J2).
Cable Material:
SERIAL
ATA
Pins:
48 Service Manual
5 Wiring Descriptions
CN1 1 4 7 2 3 5 6
CN2 1 4 7 2 3 5 6
CN3 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
CN4 NC 2 1 3 4
Instructions:
●● Connect CN1 to the ABF board (J9), (J12).
●● Connect CN2 to the hard disk.
Cable Material:
Pins:
Color Red Black White Green White Green White Green White
CN1 1 2 3 4 5 6 7 8 9
CN2 1 2 3 4 5 6 9 8 7
Instructions:
●● Connect CN1 to the trackball board (J1).
●● Connect CN2 to the keyboard (J6).
Service Manual 49
5 Wiring Descriptions
Cable Material:
Pins:
CN1 1 2 3 4
CN2 1 2 3 4
Instructions:
●● Connect CN1 to the trackball board (J1).
●● Connect CN2 to the ABF board (J27).
Cable Material:
Black(Sheilded)
Red
Pins:
CN1 1 2
Instructions:
●● Connect CN1 to the ECG board (J3).
●● Connect CN2 to the BNC socket.
50 Service Manual
5 Wiring Descriptions
Cable Material:
Pins:
CN1 1 2 3 4 5 6 7 8
CN2 8 7 6 5 4 3 2 1
Instructions:
●● Connect CN1 to the indicator (J1).
●● Connect CN2 to the keyboard (J7).
Cable Material:
Pins:
Color Shielded Red Black Black Black Black Black Black Black Black Black Black
CN1 1 3 2 4 13 14 15 16 27 28 29 30
Instructions:
Service Manual 51
5 Wiring Descriptions
Cable Material:
Black(Sheilded)
Red
Pins:
CN2 S t e e l Core / / / / /
cover
CN3 / / / / 3 2 1, 4
Instructions:
●● Connect CN1 to the ABF board (J14).
●● Connect CN2 to the BNC socket.
52 Service Manual
6 Software Installation and Update
The service department or local distributer can obtain the upgraded software package from the
manufacturer by the CD disk or Email. You can copy the package into the USB drive to update the
system with a reference to this chapter.
53
6 Software Installation and Update
USB drive Capacity not less than 4G FAT32 format, system USB drive
Windows XP operating
PC USB port available
system or above
NOTE:
•• The format of the system USB drive should be changed to FAT32. Please backup
the data in the USB drive before formatting.
•• After completing the system USB drive preparation, please choose “USB Safely
Remove” and then unplug the USB drive.
■ Operations
1. Copy the system USB drive preparation package OS_package_3_14_52-V_x_x_x-int-
release-kernel_xx_xx-yyyymmdd.zip to a computer and extract the package. The
following files are contained in the package.
54 Service Manual
6 Software Installation and Update
3. Copy update_file (shown in Figure 6-1) to the root directory of the USB drive.
4. Copy the package adata_5.*.*.*.tgz and asecure_5.*.*.*.tgz to file /update_file/
update/.
5. Safely remove the USB drive after completing copying.
The preparation of the system USB drive is complete.
Windows XP operating
PC USB port available
system or above
Two installation scenes are provided on the software update screen (USB drive reboot
screen). For Formatting Installation, install the operating system and applications and
format the hard disk synchronously. User data will be erased. For Save Data Installation,
only install the operating system and applications. User data will not be erased.
Ensure that the file in the USB of the system is complete before the installation. For
details, refer to Figure 6-2.
Service Manual 55
6 Software Installation and Update
Only the steps of Formatting Installation will be detailed in this section. Save Data
Installation can also be performed in accordance with the following steps.
1. Connect the software package USB drive to the system when the system is powered
off.
56 Service Manual
6 Software Installation and Update
2. Power on the system, wait approx. 1 second and use the aciculiform tool to press the
update button for approx. 3 seconds and remove the aciculiform tool. Wait until the
USB reboot screen (shown in Figure 6-4) appears.
You can view the versions of the operating system and applications of the current
system USB drive on the USB reboot screen.
Service Manual 57
6 Software Installation and Update
NOTE:
Patient data will be erased after clicking AllInstall Format.
4 G c a p a c i t y , FAT 3 2 f o r m a t .
Any models with capacity
USB drive Application update package is saved in
not less than 4G
the USB drive.
Windows XP operating
PC USB port available
system or above
NOTE:
•• Ensure that the prepared USB drive works normally on the desired system.
You can confirm the USB drive works normally by storing the system image
successfully into the USB drive.
•• Ensure that only one update file is saved in the USB drive.
•• Do not alter the format and file name of the update file in the USB drive.
Otherwise, the system may fail to update.
58 Service Manual
6 Software Installation and Update
•• Ensure that the system displays the current date. Otherwise, the system may fail
to update.
•• Ensure that the system can work normally.
•• Name or format of the update file provided by the manufacturer may be
inconsistent with that of the file user receives if the update file is sent through
E-mail. In such case, you can correct its name manually. For example, you can
correct asecure_*.*.*.*.tgz (1) to asecure_*.*.*.*.tgz.
6.2.2 Operations
1. Power on the system and inset the USB drive written with the update package to the
USB port of the system.
2. Click Software Update on the System Information screen.
3. Input the year password in the dialog box shown in Figure 6-5, and then press the
Enter key.
5. Click Exit when Files successfully copied. Please exit and reboot system to
complete update. pops up.
6. Unplug the USB drive and reboot the system.
NOTE:
• The password is not case sensitive. If the password is wrong, invalid password
is displayed to remind you for password retry.
Service Manual 59
6 Software Installation and Update
• You need to re-insert the USB drive or use another USB port if No USB driver
found pops up.
• Do not remove the USB drive during system reboot. Otherwise, the system
cannot be started up. The stored images or patient information will not be
damaged or lost if you update the software by the USB drive.
Any models with capacity FAT32 format. aconfig file is saved in the
USB drive
not less than 4G root directory of the USB drive.
Windows XP operating
PC USB port available
system or above
NOTE:
•• Ensure that the 7 digits in the file name is consistent with the control number of
the system.
•• Ensure that the aconfig file is saved in the root directory of the USB drive.
•• Ensure that the system display date is the current date. If it is not, modify the date
into the current date and save the modification. Or the aconfig file can not be
input.
60 Service Manual
6 Software Installation and Update
6.3.2 Operations
When you have updated the system or no aconfig file exists, the prompt “No valid license,
please import aconfig file.” pops up after powering on the system. You should copy the
aconfig file corresponding to the control number of the system to the root directory of the
USB drive, insert this USB drive to the USB port of the system, and press the Update key
to start importing. A prompt will be displayed after a successful import.
4. Input the year password and press the Enter key. A dialog box is displayed on the
screen, as shown in Figure 6-8.
5. Click Software Update.
Service Manual 61
6 Software Installation and Update
NOTE:
If the aconfig file update fails, please click OK to perform the update again. If failure
still occurs, please check whether the digits in the aconfig file name is consistent with
the control number of the system, or whether the aconfig file is copied to the root
directory of the USB drive.
■ Verifying Update
View the system serial number (in bar code format) on the rating label of the system. For
example, 0486589947.
62 Service Manual
7 Settings Import/Export
This chapter describes how to import/export the settings, which is closely related to system settings,
and customized application mode.
63
7 Settings Import/Export
Name
Type Abdomen
Part C-Vascular
Save Exit
64 Service Manual
7 Settings Import/Export
3. Input the month password and the software update content pops up, as shown in
Figure 7-3.
NOTE:
•• Ensure that only one sonoimport file is contained in the USB drive. Otherwise, the
import may fail or the imported file cannot be recognized.
•• Ensure that all the probe presettings are deleted before restoring the application
presets. Otherwise, the system cannot restore the presets.
Perform the following steps.
1. Insert a USB drive to the system.
2. Press the Setup key to enter the System Settings screen and click About > Software
Update. A dialog box is displayed, as shown in Figure 7-4.
Service Manual 65
7 Settings Import/Export
3. Input the month password and the software update content is displayed, as shown in
Figure 7-5.
66 Service Manual
8 Performance Test
You should perform the performance tests with a reference to this chapter after software update or
restore. In case of any test fails, you should solve it and perform the performance tests again until the
system works normally.
67
8 Performance Test
NOTE:
Hardware tests are divided into two parts: underlying hardware tests and peripheral device
tests.
During hardware tests, the test results are displayed in the output box at the right side of
the TestItem list. The results in green indicate that the tested item works normally and
passes the test. While, the results in red indicate that the tested item results in malfunction,
and you should re-test it after servicing it.
NOTE:
During hardware tests, if the system is not connected with the probe, I/O docking
station, ECG, probe port extender board or 64-channel analogue generator, the test
results of the corresponding test items will be in red, indicating the tests fail.
68 Service Manual
8 Performance Test
This test is used to acquire and display the ABF board information, including the
encryption status, part number, model and serial number of the ABF board. The
corresponding test item is ABFInfo. For details on test steps, please refer to Section 8.1.1
Test Steps.
Service Manual 69
8 Performance Test
This test is used to acquire and display FPGA version information, including the platform,
version No. and release date. The corresponding test item is FPGAInfo. For details on test
steps, please refer to Section 8.1.1 Test Steps.
This test is used to check the FPFA clock on the ABF board. The corresponding test item is
FPGAClock. For details on test steps, please refer to Section 8.1.1 Test Steps.
This test is used to test the connecting status and read-write rights of the PCIe port on the
ABF board. The corresponding test item is PCIeTest. For details on test steps, please refer
to Section 8.1.1 Test Steps.
This test is used to check the port status of FPGA DDR3 chip memory, including the
number of DDR, bit width, DDR clock frequency, the number of the supported ports,
controller address bit width and controller data bit width. The corresponding test item is
FPGA DDR3. For details on test steps, please refer to Section 8.1.1 Test Steps.
This test is used to acquire and display the kernel voltage of FPGA on the ABF board.
The corresponding test item is FPGAKernelVol. For details on test steps, please refer to
Section 8.1.1 Test Steps.
This test is used to acquire and display the PN temperature of FPGA on the ABF board.
The corresponding test item is Temperature. For details on test steps, please refer to
Section 8.1.1 Test Steps.
This test is used to acquire and display the voltage value of the ABF board. The
corresponding test item is ABFVol. For details on test steps, please refer to Section 8.1.1
Test Steps.
This test is used to acquire and display the power supply status of circuits on the ABF
board. The corresponding test item is ABFPowerState. For details on test steps, please
refer to Section 8.1.1 Test Steps.
70 Service Manual
8 Performance Test
This test is used to acquire and display the voltage value of the high-voltage power supply
on the ABF board. The corresponding test item is ABFHVPowerState. For details on test
steps, please refer to Section 8.1.1 Test Steps.
NOTE:
The result item of HVC_TYPE(CW) Voltage will be marked in red because the tested
voltage value is inconsistent with the actual value. This is normal and does not indicate
an ABF board malfunction, if voltage value tested by the system is correct.
This test is used to check the statuses of all the digital circuits. The corresponding test item
is ADCRampTest. For details on test steps, please refer to Section 8.1.1 Test Steps.
This test is used to check the statuses of all the analogue circuits. The corresponding test
item is AnlogChTest. For details on test steps, please refer to Section 8.1.1 Test Steps.
■ Probe Test
This test is used to acquire and display the statuses and specifications (such as connection,
encryption, P/N, SN, ID, probe code and so on.) of the probe connected to the system or
the probes connected to the MPC board. The corresponding test item is Probe & MPC.
For details on test steps, please refer to Section 8.1.1 Test Steps.
This test is used to check the communication and encryption statuses of the I/O docking
station and view the version of the I/O docking station. The corresponding test item is DK.
For details on test steps, please refer to Section 8.1.1 Test Steps.
This test is used to check the communication and encryption statuses of the probe port
extender board and view the firmware version. The corresponding test item is Probe
expander. For details on test steps, please refer to Section 8.1.1 Test Steps.
Service Manual 71
8 Performance Test
This test is used to check the color of the trackball. The corresponding test item is
Trackball Color. After selecting this test item, the trackball changes to seven different
preset colors in turn. The current color is displayed in the output box at the right side of the
TestItem list.
This test is used to check whether the monitor has color aberration and displays the images
normally. The corresponding test item is Screen Test.
72 Service Manual
8 Performance Test
−− The green round displayed at the upper right corner of the screen indicates that the
monitor is successfully corrected.
−− The yellow round displayed at the upper right corner of the screen indicates that the
monitor fails to be corrected.
You should re-correct the monitor by using the color temperature tester because the
color temperature and brightness of the monitor deviates from the normal values.
And perform the test again after re-correction.
2. Click Exit to exit the Display Adjust screen.
■ Keyboard Test
This test is used to check whether the keyboard can respond correctly. The corresponding
test item is KeyboardTest.
1. Select this item and click Start to enter the keyboard test screen, as shown in
Figure 8-3.
Service Manual 73
8 Performance Test
−− Click Off > Yellow > White, and ensure that the LED indicator lights up according
to the operation.
−− Press any key on the keyboard, and ensure that a hand icon is displayed on the
corresponding position on the screen.
−− Rotate all the knobs on the control panel, and ensure that a hand icon is displayed
on the corresponding position to indicate the rotation direction.
−− Slide the TCG sliders, and ensure that a hand icon is displayed on the corresponding
position on the screen.
2. Click Exit to exit the keyboard test screen.
If the test fails, you should re-test the keyboard after servicing it.
■ Speaker Test
This test is used to check whether the speaker works normally without noises. The audio
frequency of the speaker is between 1000Hz and 125Hz. You can test the left and right
channel of the speaker. The corresponding test item is Speaker. After selecting this test
item, click Start to test the current channel and check its frequency, and press any key to
test the other channel.
If the speaker works abnormally or has noises, you should re-test the speaker after
servicing it.
74 Service Manual
8 Performance Test
NOTE:
The virtual probe is only a demo, and do not use it when performing an exam.
You can perform the relevant operations, for example, testing the system function without
probes, after selecting a virtual probe.
NOTE:
This function is used to delete all patient data in the hard disk. Ensure that you back
up all patient data before performing this function.
Service Manual 75
8 Performance Test
Move the cursor to the check box beside Delete patient data by using the trackball, and
press the confirm key.
Input del or DEL according to the pop-up prompt and then click OK.
The system starts deleting patient data. Waiting time varies according to the size of
patient data. After all patient data is deleted, the pop-up prompt disappear, and a green
prompt is displayed in the output box at the right side of the Item list.
Reboot the system.
The system displays the temperature of hard disk in operation on Main screen.
The feature is used to indicate the maximum depth range of the system.
76 Service Manual
8 Performance Test
Use the trackball to move the cursor to the check box beside adjusting depth range, and
click Confirm. Click Exit to exit the function settings screen.
Click Depth on the Scanning Mode. Current depth range can be viewed on the Scanning
Mode.
NOTE:
Do not enable this function when performing patient exam to avoid reducing
accuracy of the data. Re-enter the Engineer Mode and click Exit to exit the
adjusting depth range feather.
The depth the probe can reach in engineering mode may be different with the
depth the probe can reach when performing patient exam.
■ Optional Functions
Use the trackball to move the cursor to the check box beside adjusting depth range, and
functions of the current system are displayed on the right display box.
NOTE:
Only professional engineers can perform this operation, as the operation may change
the settings of the fundamental devices and hardware.
Service Manual 77
8 Performance Test
NOTE:
•• For keyboard update, you should reboot the system to start blue screen update,
after copying the update file.
•• For power supply firmware update, the system is automatically powered off to
complete the update, after copying the update file.
NOTE:
Only professional engineers can perform this operation, as the operation may change
the settings of the fundamental devices and hardware.
Encryption burning should be performed for the system if the following tests fail in the
hardware tests.
●● Encryption test of the probe
●● Encryption test of the MPC
●● Encryption test of the I/O docking station
●● Encryption test of the ABFinfo
78 Service Manual
9 System Troubleshooting
This chapter contains important information for system troubleshooting. You can make fault
diagnostics and resolve it with a reference to this chapter.
79
9 System Troubleshooting
●● Potentially hazardous voltage exists inside the system. Remove personal jewelry that
! contains metallic substances when you operate the system.
●● Do not disassembly or remove any circuit boards when the system is powered on.
●● Identify whether the system malfunction is caused by incorrect operations or system
hardware/software failures to make a fault diagnostics.
●● Propose all possible factors, including system running time and external interferences,
that affect system functions.
●● Identify whether the system malfunction is a common fault. Fault caused by power
supply socket, (extension) power cord or fuse can be attributed to a common fault.
●● If the system malfunction is identified as a unusual fault, you can make the fault
recurrence in a similar test environment.
●● You can identify the system malfunction and solve it with your knowledge, experience
or a reference to 3 Principle Descriptions, Section 8.1 Automatic Hardware Tests and
9.1 Fault Diagnostics.
●● Do not replace a component if you are unsure if it is the cause.
●● If certain component is identified to be the fault cause, you should replace it only,
but not replace the whole circuit board that holds it. For example, a faulty capacitor
is identified to be fault cause, you should replace it only, but not the PCB board that
holds it.
●● Only the fault cause is identified, you should solve it and perform necessary tests
described in this chapter to ensure the system performance.
NOTE:
Fault diagnostics shown in Table 9-1 and Table 9-2 are for reference only, you should
make a on-site fault diagnostics with care.
80 Service Manual
9 System Troubleshooting
●● Startup Faults
Unable to start The system cannot be Check whether the power supply is
up the system started up by pressing the correctly connected.
Power on/off button and the Check whether the Power on/off
back light of the keyboard button is pressed.
does not light up.
Check whether the requirement of the
voltage is met.
Blue Screen The system stops at a blue Press and hold the Power on/off
screen after startup. button to power off the system, and
reboot it.
If the fault cannot be solved, re-install
the system software.
●● Operating Faults
No initial No initial echo displayed on Ensure that voltages on the ABF board
echo on the the image. are normal.
image Ensure that there are high voltage
output.
Service Manual 81
9 System Troubleshooting
Auto The system automatically Identify the operation causing the auto
Shutdown shuts down when you shutdown, and inform the malfunction
perform general system frequency to the manufacturer.
operations, such as Update the system software.
performing measurements,
connecting the USB drive or
freezing an image.
Unable to The system crashes during Press and hold the Power on/off button
shut down operation. to shut down the system.
the system If the fault cannot be solved, re-install
the system software.
82 Service Manual
9 System Troubleshooting
■ Disassembling Parts
Observe and record the location and connection direction of parts when you replace them.
The parts in incorrect location or connection direction may result in system startup failure,
system malfunction, board burning, or even serious damages.
■ Replacing Parts
Replace failure parts one by one and test it after each replacement until the fault is solved.
You are not recommended to replace failure parts before system test. Meanwhile, make a
detailed record of the replacement sequence.
■ Assembling Parts
Install all replaced parts in the system by fastening all necessary screws. Meanwhile,
ensure that there are no screws left.
The system mainly consists of keyboard, ABF board, LCD monitor and probe conversion
board. It is very important to ensure the full connections among these parts. Therefore, you
should check the connections both before and after servicing the equipment.
●● Keyboard
Ensure that all keys, knobs, switches and trackball work normally after re-installation.
●● Probe Board
Ensure that the probe board and ABF board are connected well after re-installation.
●● Trackball
Ensure that the cursor can be moved smoothly by using the trackball after re-
installation.
Service Manual 83
9 System Troubleshooting
●● ABF Board
Ensure that the ABF board and keyboard are connected securely after re-installation.
■ Maintenance Items
●● Start up or shut down the system for several times to ensure no malfunction occurs.
●● Clean the system if it has been used for a long period. For details, refer to 10 System
Maintenance.
●● Record or solve problems proposed by the user.
●● Make a detailed service record for future maintenance or service. The fault cause,
replaced parts and their corresponding part numbers, repairing procedures should
be detailed in this record. Moreover, a signature from the hospital on this record is
required.
84 Service Manual
10 System Maintenance
To maintain the safety and functionality of the system, you should periodically perform the
maintenance for the system and accessories.
85
10 System Maintenance
●● To avoid electrical shock and damage to the system, power off and unplug the system
! from the AC power outlet before cleaning.
●● To maintain the safety and functionality of the system, maintenance must be
performed at least every 1 year. Electrical safety tests must also be performed at
regular intervals as specified by local safety regulations.
4. Take out the ring and trackball (be careful not to drop the trackball) from the control
panel.
5. Use a lint-free soft cloth dampened with 75% ethyl alcohol to clean the trackball.
6. Use a cotton swab dampened with 75% ethyl alcohol to remove stains from the ring.
7. Place the trackball and the ring back in the control panel, press the bulges in the ring
with both hands and turn the ring clockwise to install the trackball.
86 Service Manual
10 System Maintenance
●● To avoid electrical shock, disconnect the probe from the system before cleaning.
! ●● To avoid potential disease transmission, you should wear sterile medical gloves and
protective goggles while cleaning.
●● Do not use the cleaner that contains sodium sulfate and/or sodium hypochlorite to
clean the probe. Otherwise, the probe may be damaged.
●● To avoid probe damages, do not bump the probe on hard surfaces.
●● Use only the cleaner recommended by the manufacturer to clean the probe. Otherwise,
the probe may be damaged.
You should clean the probe before and after each use. Probe cleaning is an essential step of
the disinfection and sterilization.
Service Manual 87
10 System Maintenance
●● There is a cleared list of sterilants and high-level disinfectants for use in processing
! reusable medical and dental devices on the FDA website. That list can be consulted
to find agents that may be useful for probe disinfection. Please refer to the following
URL for FDA-Cleared Sterilants and High Level Disinfectants: http://www.fda.gov/
MedicalDevices/DeviceRegulationandGuidance/ReprocessingofSingle-UseDevices/
ucm133514.htm.
●● Use the liquid disinfectant/sterilant that meets local laws and regulations or the
recommended disinfectant/sterilant.
●● The probe sheath may be damaged during exam and contamination risk level caused
by the damaged sheath is difficult to estimate. Therefore, probe sheath cannot change
the level of disinfection for a probe. Please clean and sterilize (or high level disinfect)
the endocavitary probe and clean and sterilize the probe for biopsy and coming into
contact blood after exam.
●● Do not use an expired disinfectant/sterilant.
●● Store the probe in a sterile environment, and review its expiration date before use.
●● If you use the recommended disinfectant/sterilant, ensure that the requirements for
the disinfection level and probe soaking period are met. Otherwise, the probe could
be damaged and your warranty might be void. If you have further use or needs for the
probe, ensure that the level of the make-up disinfectant/sterilant and the probe soaking
period are suitable for the intended clinical application.
●● Do not use the following methods to sterilize the probe.
−− Autoclave
−− Ultraviolet sterilizer
−− Gamma ray sterilizer
−− Ethylene oxide gas
−− Ethylene oxide
−− Thermal disinfection/sterilization. Temperature higher than 66°C (150°F) will
damage the probe.
●● Do not allow any disinfectant/sterilant to be air-dried on the probe.
■ Disinfection Levels
To choose an appropriate disinfectant, you must first determine the required level of
disinfection, based on the probe classification.
88 Service Manual
10 System Maintenance
Table 10-1 Disinfection Classifications
■ Soaking Requirements
You should disinfect the body surface probe after each use.
Service Manual 89
10 System Maintenance
You should disinfect the endocavitary probe before and after each use.
90 Service Manual
10 System Maintenance
To avoid potential disease transmission, you should wear sterile medical gloves and
! protective goggles during disinfection or serialization.
Do not immerse the probe cable in alcohol or isopropyl alcohol at any concentration for a
long period of time. Avoid any contact with iodine or phenols. Otherwise, it will accelerate
the aging of the cable and decrease the service time of the probe, or even result in cable
damages.
■ Disinfection
Service Manual 91
10 System Maintenance
■ Sterilization
If the probe cable comes into contact with any blood or body fluid, you should sterilize it.
For details, refer to the information about probe sterilization.
92 Service Manual
10 System Maintenance
Service Manual 93
10 System Maintenance
10.3.3 Storage
Ensure that the biopsy bracket is cleaned, sterilized and completely dried before storage.
Store the biopsy bracket in a sterile environment and keep it away from the environment of
sudden temperature changes or direct sunlight.
You should dispose of the main unit, the probe, the biopsy bracket and other accessories in
accordance with local laws and regulations.
For the detailed disposal information, consult the manufacturer or the local distributor.
The manufacturer is not responsible for any system content or accessories that have been
discarded improperly.
Address: 4/F, 5/F, 8/F, 9/F & 10/F, Yizhe Building, Yuquan Road, Nanshan, Shenzhen,
518051, Guangdong, China
Tel: +86-755-26722890
Fax: +86-755-26722850
Email: service@sonoscape.net
94 Service Manual
Appendix Renewal Parts
SD card 2100.00960 1
Keyboard 3600.02199 1
Speaker 2100.00921 2
Fan 2100.00973 4
95