Download as pdf or txt
Download as pdf or txt
You are on page 1of 101

About This Manual

P/N: 4720.00180A01
Product Model: E5 Exp/E5/E5 Pro/E3 Exp/E3/E3 Pro/E2 Exp/E2/E2 Pro/E1 Exp/E1/E1 Pro
Release Date: February, 2018
Copyright © 2018 SonoScape Medical Corp. All rights reserved.

Statement
SonoScape Medical Corp. (hereinafter called SonoScape) owns the intellectual property rights to
this manual, and also maintains the contents of this manual as confidential information. This manual
is a reference for the operation, maintenance and cleaning of this product and does not convey any
license under the patent rights of SonoScape, nor the rights of others.
This manual contains the information protected by copyrights or patents, and should only be used
by the service personnel of SonoScape and the authorized agents. Reproduction, amendment or
translation of this manual in any manner whatsoever without the written permission of SonoScape is
strictly prohibited.
All information contained in this manual is believed to be correct. SonoScape shall not be liable for
errors contained herein or for incidental or consequential damages in connection with the furnishing,
performance or use of this manual. SonoScape does not assume any liability arising out of any
infringements of patents or other rights of third parties.
This manual is based on the maximum configuration and therefore some contents may not apply to
your product.
This manual is subject to change without prior notice and legal obligation.

Manufacturer’s Responsibility
SonoScape is responsible for the effects on safety, reliability and performance of this product, only if
all the following requirements are met.
●● All installation operations, expansions, changes, modifications and repairs of this product are
conducted by SonoScape authorized personnel.
●● The use or application of the product or the use of parts or accessories is approved by
SonoScape.
●● The electrical installation of the relevant room complies with the applicable national and local
requirements.

●● The product is used in accordance with the instructions for use.

Documentation
Understand the meanings of the following items clearly before reading this manual.
Item Meaning

Indicates a potentially hazardous situation which, if not avoided, could result in


! death or serious injury.

Indicates a potentially hazardous situation which, if not avoided, may result in


malfunction or damage of the system.

Indicates a potentially biological hazardous situation which, if not avoided, may


result in disease transmission.

NOTE Indicates precautions or recommendations that should be used in operating the


system.

Boldfaced Indicates controls on the control panel, or on-screen objects such as menu items
Word or keys.

Click Move the cursor to the controls on the display and press the confirm key on the
control panel.

> Select a menu item or a key following the path.

Contact Information
Manufacturer: SonoScape Medical Corp.
Address: 4/F, 5/F, 8/F, 9/F & 10/F, Yizhe Building, Yuquan Road, Nanshan, Shenzhen, 518051,
Guangdong, China
Tel: +86 755 2672 2890
Fax: +86 755 2672 2850
http://www.sonoscape.com
E-mail: sonoscape@sonoscape.net
Contents

1    Safety���������������������������������������������������������������������������������������������������������������������������������������� 1
1.1  Intended Use����������������������������������������������������������������������������������������������������������������������������2
1.2  Safety Precautions�������������������������������������������������������������������������������������������������������������������2
1.2.1  Electrical Safety������������������������������������������������������������������������������������������������������������2
1.2.2  Mechanical Safety���������������������������������������������������������������������������������������������������������4
1.2.3  Accessories Caring��������������������������������������������������������������������������������������������������������4
1.2.4  Biohazard Considerations���������������������������������������������������������������������������������������������5
1.3  Acoustic Power Principle��������������������������������������������������������������������������������������������������������5
1.3.1  Biological Safety����������������������������������������������������������������������������������������������������������5
1.3.2 ALARA�������������������������������������������������������������������������������������������������������������������������5
1.3.3  Mechanical and Thermal Indices����������������������������������������������������������������������������������6
1.3.4  Transducer Surface Temperature Limits�����������������������������������������������������������������������7
1.3.5  Imaging Functions that Change Acoustic Output���������������������������������������������������������8

2    Unpacking and System Checks���������������������������������������������������������������������������������������������� 9


2.1 Unpacking�����������������������������������������������������������������������������������������������������������������������������10
2.2  System Connection����������������������������������������������������������������������������������������������������������������13
2.3  Powering On/Off�������������������������������������������������������������������������������������������������������������������13
2.4  System Checks����������������������������������������������������������������������������������������������������������������������14
2.4.1  Unpacking Checks������������������������������������������������������������������������������������������������������14
2.4.2  General Checks�����������������������������������������������������������������������������������������������������������14
2.4.3  Functional Checks �����������������������������������������������������������������������������������������������������16
2.4.4  Image Tests�����������������������������������������������������������������������������������������������������������������17

3    Principle Descriptions����������������������������������������������������������������������������������������������������������� 19
3.1  System Block Diagram����������������������������������������������������������������������������������������������������������20
3.2  Hardware Description������������������������������������������������������������������������������������������������������������21
3.2.1  ABF Board������������������������������������������������������������������������������������������������������������������21
3.2.2 Keyboard���������������������������������������������������������������������������������������������������������������������23
3.2.3  Probe Board����������������������������������������������������������������������������������������������������������������23
3.2.4  Power Port Board��������������������������������������������������������������������������������������������������������24
3.2.5  Battery Port Board������������������������������������������������������������������������������������������������������24

I
Contents

3.2.6  ON/OFF Button Board�����������������������������������������������������������������������������������������������24


3.2.7  Port Extender Board����������������������������������������������������������������������������������������������������24
3.3  Software Description�������������������������������������������������������������������������������������������������������������24
3.3.1 Structure����������������������������������������������������������������������������������������������������������������������24
3.3.2 Workflow��������������������������������������������������������������������������������������������������������������������25

4    Parts Disassembly����������������������������������������������������������������������������������������������������������������� 27
4.1  Battery Disassembly��������������������������������������������������������������������������������������������������������������28
4.2  Bottom Components Disassembly����������������������������������������������������������������������������������������28
4.3  Back Cover Disassembly�������������������������������������������������������������������������������������������������������29
4.4  Control Panel Disassembly���������������������������������������������������������������������������������������������������30
4.5  LCD Monitor Disassembly���������������������������������������������������������������������������������������������������32
4.6  Fan Disassembly�������������������������������������������������������������������������������������������������������������������34
4.7  Grounding Cable and S-Video Disassembly�������������������������������������������������������������������������35
4.8  Battery Board Bracket Disassembly��������������������������������������������������������������������������������������35
4.9  ECG Components Disassembly��������������������������������������������������������������������������������������������36
4.10  Probe Board Disassembly���������������������������������������������������������������������������������������������������38
4.11  Hard Disk Disassembly�������������������������������������������������������������������������������������������������������39
4.12  SD Card Disassembly����������������������������������������������������������������������������������������������������������40
4.13  Speakers Disassembly���������������������������������������������������������������������������������������������������������40
4.14  ABF Board Disassembly�����������������������������������������������������������������������������������������������������41

5    Wiring Descriptions�������������������������������������������������������������������������������������������������������������� 43
5.1  Connections for Wiring���������������������������������������������������������������������������������������������������������44
5.2  Pin Connection����������������������������������������������������������������������������������������������������������������������45
5.2.1 Keyboard - TGC Board Signal Cable�������������������������������������������������������������������������45
5.2.2  ABF Board - Display EDP Cable��������������������������������������������������������������������������������45
5.2.3  ABF Board - Keyboard Signal Cable�������������������������������������������������������������������������46
5.2.4  ABF Board - ECG Board Signal Cable����������������������������������������������������������������������47
5.2.5  Adaptor Converter - Battery Board Cable������������������������������������������������������������������48
5.2.6  ABF Board - SATA Hard Disk Cable�������������������������������������������������������������������������48
5.2.7  Keyboard - Trackball Board Signal Cable������������������������������������������������������������������49
5.2.8  ABF Board - ON/OFF Button Board Signal Cable����������������������������������������������������49
5.2.9  ECG Board - BNC Socket Cable��������������������������������������������������������������������������������50
5.2.10  Keyboard - Indicator Board Cable����������������������������������������������������������������������������51
5.2.11  ABF Board - BNC Socket Cable������������������������������������������������������������������������������51

II Service Manual
Contents

5.2.12  ABF Board - BNC & Foot Switch����������������������������������������������������������������������������52

6    Software Installation and Update���������������������������������������������������������������������������������������� 53


6.1  Installing the Operating System��������������������������������������������������������������������������������������������54
6.1.1  Making the System USB Drive ���������������������������������������������������������������������������������54
6.1.2  Operating System Installation�������������������������������������������������������������������������������������55
6.2  Updating the Application�������������������������������������������������������������������������������������������������������58
6.2.1  Tool Preparation����������������������������������������������������������������������������������������������������������58
6.2.2 Operations�������������������������������������������������������������������������������������������������������������������59
6.2.3  Update Verification ����������������������������������������������������������������������������������������������������60
6.3  Importing the aconfig File ����������������������������������������������������������������������������������������������������60
6.3.1  Tool Preparation����������������������������������������������������������������������������������������������������������60
6.3.2 Operations ������������������������������������������������������������������������������������������������������������������61

7    Settings Import/Export��������������������������������������������������������������������������������������������������������� 63
7.1  Customizing a Preset�������������������������������������������������������������������������������������������������������������64
7.2  Exporting a Customized Preset���������������������������������������������������������������������������������������������64
7.3  Importing a Customized Preset���������������������������������������������������������������������������������������������65

8    Performance Test������������������������������������������������������������������������������������������������������������������� 67
8.1  Automatic Hardware Tests����������������������������������������������������������������������������������������������������68
8.1.1  Test Steps��������������������������������������������������������������������������������������������������������������������68
8.1.2  Underlying Hardware Tests����������������������������������������������������������������������������������������69
8.1.3  Peripheral Device Test������������������������������������������������������������������������������������������������71
8.2  Function Settings�������������������������������������������������������������������������������������������������������������������74
8.3  Firmware Update�������������������������������������������������������������������������������������������������������������������77
8.4  Burning Function�������������������������������������������������������������������������������������������������������������������78

9    System Troubleshooting�������������������������������������������������������������������������������������������������������� 79
9.1  Fault Diagnostics�������������������������������������������������������������������������������������������������������������������80
9.2  System Troubleshooting��������������������������������������������������������������������������������������������������������82
9.2.1  On-Site Inspection������������������������������������������������������������������������������������������������������82
9.2.2  Fault Identification������������������������������������������������������������������������������������������������������82
9.2.3  Fault Resolution����������������������������������������������������������������������������������������������������������83
9.2.4  Post Maintenance �������������������������������������������������������������������������������������������������������83

10    System Maintenance����������������������������������������������������������������������������������������������������������� 85
10.1  Cleaning the System������������������������������������������������������������������������������������������������������������86

Service Manual III


Contents

10.2  Probe Maintenance��������������������������������������������������������������������������������������������������������������87


10.2.1  Cleaning the Probe����������������������������������������������������������������������������������������������������87
10.2.2  Disinfecting or Sterilizing the Probe������������������������������������������������������������������������88
10.2.3  Sterilizing the Probe�������������������������������������������������������������������������������������������������91
10.2.4  Disinfecting and Sterilizing the Probe Cable������������������������������������������������������������91
10.2.5  Storage and Transportation���������������������������������������������������������������������������������������92
10.3  Biopsy Bracket Maintenance�����������������������������������������������������������������������������������������������93
10.3.1  Cleaning the Biopsy Bracket������������������������������������������������������������������������������������93
10.3.2  Sterilizing the Biopsy Bracket����������������������������������������������������������������������������������93
10.3.3 Storage����������������������������������������������������������������������������������������������������������������������94
10.4  Equipment Disposal������������������������������������������������������������������������������������������������������������94
10.5  Customer Service����������������������������������������������������������������������������������������������������������������94

Appendix  Renewal Parts���������������������������������������������������������������������������������������������������������� 95

IV Service Manual
1    Safety

This chapter describes the important information for operating this system. To ensure the safety of
both operator and patient, please read the relevant details in this chapter carefully before using this
system.

You should be thoroughly familiar with the precautions provided in this manual. Otherwise, the
manufacturer is not responsible for the effects on safety, reliability and performance of the system.

1
1 Safety

1.1  Intended Use


The system is intended for the following applications: Fetal, Abdominal, Pediatric, Small
Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal,
Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial),
Cardiac (pediatric and adult), OB/Gyn and Urology.

The system also provides the measurement and calculation packages used for clinical
diagnosis purposes. For details, please refer to the advanced user manual.

Contraindication: The system is not intended for ophthalmic use or any use causing the
acoustic beam to pass through the eye.

Precautions must be considered in the use of any application. Otherwise, it may result in
! system damage or serious injury.

1.2  Safety Precautions


Read and understand all precautions in this manual before attempting to use the system.
Keep this manual with the system at all times. Review the procedures for operation and
safety precautions periodically.

1.2.1  Electrical Safety

●● Only qualified physicians or sonographers can perform ultrasound scanning on human


! subjects for medical diagnostic reasons.
●● Any unauthorized personnel should not tamper with the main unit of the system.
●● Do not service or maintain the system while it is in use with a patient.
●● Do not position the system to make it difficult to disconnect the system from the
mains supply.
●● Do not use the system with flammable anaesthetics (category AP) or flammable
anaesthetics with oxidants (category APG).
●● Do not use the system with other equipment such as an electric knife, high-frequency
therapy equipment, or a defibrillator. Otherwise, electrical shock may occur.
●● Connect the protective earth conductor only before powering on the system.
Disconnect the ground wire only after powering off the system. Otherwise, electrical
shock may occur.
●● Connect the system to the other electrical equipment by using the potential-
equalization lead wire before connecting the power plug of the system to an electrical
outlet.
●● Do not place the multiple socket-outlet on the floor.

2 Service Manual
1 Safety

●● The video printer should be connected to the specific interface by using the cable
provided by the manufacturer. Otherwise, electrical shock may occur.
●● Within the environment that is 1.8 meters (6 feet) around a patient, connect peripherals
to the auxiliary power outlet which is capable of isolation protection; or, power the
peripherals by the auxiliary output cable or the isolation transformer complied with
EN/IEC 60601-1, or the power input of the same safety level.
●● Within the patient environment, when removing the covers or the connectors from the
non-medical electrical equipment without tools during maintenance, calibration or
other operations, do not touch the parts of the non-medical electrical equipment and
the patient simultaneously.
●● Use only the probes provided by the manufacturer. Otherwise, the system will not
perform, and an accident such as a fire may result in the worst case.
●● Only the peripherals and accessories provided or recommended by the manufacturer
can be used. Using other devices or accessories may degrade the system performance
and even cause electrical shock.
●● Do not pour any fluid onto the system surfaces, as fluid seepage into the electrical
circuitry may cause excessive electrical current leakage or system failure. If any
water is spilled onto the system carelessly, stop using the system and contact the local
distributor immediately.
●● The AC power plug for the system is a three-prong grounded plug and should never
be adapted to any two-prong outlet or by using an adapter. Connect the AC power
plug of the multiple socket-outlet to a hospital-grade power outlet.

●● If the system is transported to the operating environment with a great temperature


change, leave it for approximately 4 hours before powering it on. Ensure that the
temperature and humidity inside and around the system are equivalent before an
operation.
●● Do not use the system around a strong electric field, a strong electromagnetic field,
or the devices which generate radio waves, such as a radio, cellular telephones, or
transceivers. Using the system in an improper environment may result in malfunction
or damage.
●● Select the conformed multiple socket-outlet with protective grounding, and ensure
that its maximum output power exceeds the requirement of the system.
●● The multiple socket-outlet can only be used to provide power to the recommended
peripherals of the system.
●● Do not connect other devices to the multiple socket-outlet. Otherwise, the rated output
power of the multiple socket-outlet may be exceeded, and it may result in failure.

Service Manual 3
1 Safety

●● Accessory equipment connected to the analog and digital interfaces must be certified
according to the respective EN/IEC standards (for example, EN/IEC 60950 for data
processing equipment and EN/IEC 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the system standards EN/IEC 60601-1.
●● Do not use an endocavitary probe on the body surface of a patient. Otherwise, the
electromagnetic compatibility of the probe may be degraded.
●● If the non-medical electrical equipment used with the system is supplied by a multiple
socket-outlet with separating transformer, connect the AC power plug of the multiple
socket-outlet to a hospital-grade power outlet. Consult a professional to ensure that
the connection complies with the safety standards.

1.2.2  Mechanical Safety

●● To avoid possible personnel injury and system damage, move the system slowly and
! carefully.

●● Do not place other objects on top of the control panel. Do not sit on the control panel
or any other part of the system.
●● Disconnect the foot switch and the power cable before moving the system.
●● Do not knock or shake the system.
●● Always use the handle to move the system.

1.2.3  Accessories Caring

●● Do not use the foot switch in the operating room.


! ●● Use the probe carefully. If any part of the transducer surface is scratched, stop using the
probe immediately. Otherwise, electrical shock may occur.
●● After disinfecting the accessories, wash out chemicals or discharge gases thoroughly
from the accessories. Residual chemicals or gases could result in damage to the
accessories, and be harmful to human bodies.
●● Only trained physicians or sonographers under ultrasound guidance can handle the
biopsy needle guides. During the operation, the operator must observe proper needle
insertion sequencing with the needle guide to avoid undue discomforts, unnecessary
risks or injuries to the patient.
●● Use the legally marketed probe sheath. Refer to package labeling to determine latex
content. Natural rubber latex may cause anaphylactic reactions in some individuals.
Avoid contact with items containing natural rubber latex. Refer to the FDA Medical
Alert, March 29, 1991.

4 Service Manual
1 Safety

●● Use the legally marketed coupling gel in accordance with relevant local regulations.
Read and understand all precautions in the relevant manual of the coupling gel before
use.
●● Prepare, use, store and dispose the cleaner, disinfectant and sterilant according to the
instructions provided by manufacturers.

●● Do not disconnect the probe during the real-time scan. Otherwise, it may damage the
probe and the system.
●● Disconnect the probe from the system after freezing an image or powering off the
system. Otherwise, the system or the probe could be damaged.

1.2.4  Biohazard Considerations

●● To minimize the risk of cross-contamination or infectious diseases when performing a


biopsy, the operator should wear disposable gloves, protective clothing or protective
goggles if it is needed. Follow the working regulations strictly in case the skin
contacts the samples.
●● Some disinfectants or sterilants are acid or alkaline. Use them with caution, and
prevent hands or clothing from coming into direct contact with them. Wash hands or
eyes immediately in case of any contamination by disinfectants.
●● Dispose of cleaners, disinfectants or solutions in accordance with local standards or
regulations.

1.3  Acoustic Power Principle

●● Perform ultrasound procedures prudently under the guidance of the ALARA (as low
! as reasonably achievable) principle. Only expose the patient to the lowest practical
transmit power levels in the shortest possible period to achieve a satisfactory
diagnosis.
●● Freeze the image at any time if you are not operating the system for a long period of
time.
●● Do not scan the same part of an patient continuously or expose the patient to
prolonged scanning. Doing so may harm the patient.
●● Do not expose the fetus to prolonged scanning in the Doppler mode.
●● Although the output power is automatically controlled for the selected applications,
high TI values should be kept to a minimum or avoided in obstetric applications.
●● You should be familiar with the performances and operations of the system, observe
the ultrasound output parameters on the screen at all times.

Service Manual 5
1 Safety

1.3.1  Biological Safety


Diagnostic ultrasound is recognized as being safe, but the possibility of biological effects
exists when using it in high exposure levels and long exposure times. Thus ultrasound
should be used in a prudent manner to provide medical benefit to the patient.

1.3.2  ALARA
It is required to practice ALARA when using ultrasound energy. Practicing ALARA
ensures that the total energy level is controlled below a low enough level at which
bioeffects are not generated while diagnostic information is being accumulated. The total
energy is controlled by output intensity and total radiation time. The output intensity
necessary for examinations differs depending on the patient and the clinical case.

Not all examinations can be performed with an extremely low level of acoustic energy.
Controlling the acoustic level at an extremely low level leads to low-quality images or
insufficient Doppler signals, adversely affecting the reliability of the diagnosis. However,
increasing the acoustic power more than necessary does not always contribute to an
increase in quality of information required for diagnosis, rather increasing the risk of
generating bioeffects.

The operator must take responsibility for the safety of patients and utilize ultrasound
deliberately. Deliberate use of ultrasound means that output power of ultrasound must be
selected based on ALARA. Additional information regarding the concept of ALARA and
the possible bioeffects of Ultrasound is available in a document from the AIUM (American
Institute of Ultrasound Medicine) title “Medical Ultrasound Safety”.

1.3.3  Mechanical and Thermal Indices


The display of the system consists of two parts: Thermal Index (TI) and Mechanical Index
(MI).

■ MI/TI Explanation

In October 1987, the American Institute of Ultrasound in Medicine (AIUM) ratified a


report prepared by its Bioeffects Committee (Bioeffects Considerations for the Safety
of Diagnostic Ultrasound, J Ultrasound Med.,Sept. 1988: Vol. 7, No. 9 Supplement),
sometimes referred to as the StoweReport, which reviewed available data on possible
effects of ultrasound exposure. Another report “Bioeffects and Safety of Diagnostic
Ultrasound” dated January28, 1993, provides more current information.

6 Service Manual
1 Safety

●● Mechanical Index (MI)


Mechanical bioeffects are threshold phenomena that occur when a certain level of
output is exceeded. The threshold level varies, however, with the type of tissue. The
potential mechanical bioeffects varies with peak pressure and ultrasound frequency.
The MI accounts for these two factors. The higher the MI value, the greater the
likelihood of mechanical bioeffects occurring. There is no specific MI value that
means that a mechanical effect is actually occurring. The MI should be used as a
guide for implementing the ALARA principle.
●● Thermal Index (TI)
The TI value informs the operator about the conditions that might lead to an increase
in temperature at the surface of the body, within the body tissue, or at the point of
focus of the ultrasound beam on bone. That is, the TI value informs the operator about
the potential temperature rise in body tissue. It is an estimate of temperature increase
in body tissue with specific properties. The actual amount of any temperature rise is
influenced by factors such as tissue type, vascularity, mode of operation and others.
The TI value should be used as a guide for implementing the ALARA principle.
Depending on the examination and type of tissue involved, TI could be one of three
types:
−− Soft Tissue Thermal Index (TIS) is used when imaging soft tissue only, it provides
an estimate of potential temperature rise in soft tissue.
−− Bone Thermal Index (TIB) is used when bone is near the focus of the image as in
the third cropester OB examination, it provides an estimate of potential temperature
rise in the bone or adjacent soft tissue.
−− Cranial Bone Thermal Index (TIC) is used when bone is near the skin surface as in
transcranial examination, it provides an estimate of potential temperature rise in the
bone or adjacent soft tissue.

■ MI/TI Display

TI and MI values are displayed in real time on the screen. The operator should observe
these index values during examinations and ensure that exposure time and output values
are maintained at the minimum amounts needed for effective diagnosis.

The MI and TI precision is 0.1.

1.3.4  Transducer Surface Temperature Limits


For probes intended for internal applications, e.g. the endocavitary probe or
transesophageal probe, the surface temperature of the probe may change by adjusting
system parameters.

Service Manual 7
1 Safety

The maximum surface temperature of the endocavitary probes is 43°C. To protect the
patient against the harm of excessive temperature, the probe stops working automatically
when its temperature reaches the limit. The surface temperature of the probe is displayed
on the screen.

TIP <28°C Transducer Tip Temperature


PAT 37°C Patient Temperature

1.3.5  Imaging Functions that Change Acoustic Output


The qualified operator may use the system controls to limit the ultrasound output and to
adjust the quality of the images. The operator should observe the acoustic output display
for possible effects.

There are three categories of system controls relative to output. They are controls that
have direct effect on the output, controls that indirectly control output and controls that are
receiver controls.

8 Service Manual
2    Unpacking and System Checks

This chapter describes how to unpack, assemble, connect, power on/off this system, and system
checks are also detailed herein. The unpacking, general and functional checks are necessarily
performed after you assemble the system for the first time. The functional checks are necessarily
performed after you service the system.

9
2  Unpacking and System Checks

2.1  Unpacking
Before unpacking, ensure that the package is complete and take photos of the package. If
there is any problem about the package or the appearance of the system, send the photos to
the customer service department of the manufacturer.

One pair of scissors (or other specialized unpacking tools) and a knife are needed during
unpacking.

Perform the following steps.


1. Use the scissors to cut off all binding straps and use the knife to split wrapping films,
as shown in Figure 2-1.

Binding Straps

Film

Figure 2-1  Removing the Binding Straps

2. Use the knife to split adhesive tape, and open the carton, as shown in Figure 2-2.

Figure 2-2 Opening the Carton

10 Service Manual
2  Unpacking and System Checks

3. Take out the system.


The operation will be different depending on whether the system is equipped with
backpack or not. Refer to the following steps for unpacking with backpack.

■ With Backpack
1. Open the carton, remove the foams, and take out the user manuals and the accessories
in sequence.

1 2

Foam

3 4

Figure 2-3 Taking out the Accessories

2. Take out the probe box, left foam, right foam and backpack. Open the backpack,
remove the backpack belt and take out the system in sequence. As shown in
Figure 2-4.

1 2 3
Probe box

Left Right
foam foam

4 5 6 7
Backpack

Backpack
belts

Figure 2-4 Taking out the System from the Backpack

Service Manual 11
2  Unpacking and System Checks

NOTE:
• Take out the system carefully.
• Do not scratch or hit the surface of the system.

■ Without Backpack
1. Open the carton, and take out the foam, user manuals, accessories and accessory box
in sequence , as shown in Figure 2-5.

1 2 3

Upper foam

4 5 6
Assessory box

Figure 2-5 Taking out Accessories

2. Take out the right foam, left foam in sequence, as shown in Figure 2-6.

1 2
Right foam

3 4

Left foam

Figure 2-6 Taking out the System

12 Service Manual
2  Unpacking and System Checks

2.2  System Connection

NOTE:
Ensure that the system works with the required voltage before you connect it to the
mains supply. The voltage and power supply of this system are 100-240V~, 50/60Hz,
1.5 - 0.75A.

Connect one end of the adaptor to the system firmly and connect the other end of the
adaptor to the mains supply firmly.

2.3  Powering On/Off

■ To Power On the System

Press the button on the right side of the screen to power on the system.

■ To Activate Screen Saver

If the system is idle for a set period of time, the system enables the screen saver
automatically. Press any key to exit the screen saver. For details, refer to System Settings
of the basic user manual.

■ To Enter the Standby Mode

Perform one of the following ways.

●● Press the button and click Standby on the pop-up menu to enter the standby
mode.
Press the button in standby mode to exit the standby mode.
●● Close the upper cover of the system, and the system automatically enters standby
mode.
Open the upper cover of the system, and the system automatically exits the standby
mode.
●● Set automatical standby time. If the system is idle for a set period of time, the system
will automatically enter the standby mode.
Press the button in standby mode, and the system automatically exits the standby
mode.

■ To Power Off the System

Press the button and select Shutdown on the pop-up menu to power off the system.

Service Manual 13
2  Unpacking and System Checks

2.4  System Checks


System checks consist of unpacking, general, functional, mechanical and image checks.

2.4.1  Unpacking Checks

■ To Check the System

You should check the LCD monitor and main unit covers, and ensure that there are not any
scratches or damages after unpacking.

■ To Check Accessories

You should check all accessories according to the packing list, and ensure that all
accessories are of the correct types and PN.

2.4.2  General Checks


General checks consist of overview, operating environment, system configuration, system
information and system running status checks.

■ Overview

You can check the following items or log with the user to make diagnostics.
●● Any exceptions occurred during system running
●● Accidental exception
●● Other requirements proposed by the user

■ Operating Environment

You should check the operating environment of the system. The measurement accuracy is
closely related to humidity, measurement application and circuit.

14 Service Manual
2  Unpacking and System Checks

The system insulation worsens with life-to-date depreciation or malfunction, and


fluctuation range of measurement errors may be larger with a greater humidity.

■ System Configuration

You should check the system configuration, including system settings, brightness and
contrast of the LCD monitor and date and time.

Item Inspection

System settings Press the Setup key to display the System Settings screen
and check the relevant settings by viewing this menu.

Brightness and contrast You can adjust the brightness of the LCD monitor on the
of the LCD monitor System Settings screen. LCD monitor’s brightness and
contrast is one of the most important factors for obtaining
an optimum image. If the settings is incorrect, you need to
frequently adjust the gain, TGC, dynamic range and output
power supply. Therefore, the lowest level of black should
just disappear into the background and the highest white
should be bright, but not saturated.

Time and Date Check whether the time and date displayed are correct. If
not, you need to adjust them.

■ System Information

Press the Setup key on the key panel to enter the System Settings screen and click About
to view the system information.

■ System Running Status


●● All menus or screens can be displayed normally.
●● Measurements can be performed in every exam mode.
●● All display mode can be switched normally.
●● The desired probe can be selected.

Service Manual 15
2  Unpacking and System Checks

2.4.3  Functional Checks

■ Workflow

Check Intital Operations

Check General Operations

Check Keyboard and Keys


检查常规操作

Check Trackball
检查常规操作

Check Peripheral Devices


检查常规操作

Check Fan
检查常规操作

Check Probe
检查常规操作

■ Checklist

Item Inspection

Initial Operations ●● Features can be applied in every exam mode.


●● Images can be scanned normally by using the corresponding
probe and can be printed out.
●● Peripheral devices can work normally.

General Operations ●● The Freeze key can respond.


●● The Gain knob can be rotated normally.
●● The CW/CFM/PW/M/B mode can be switched from one to
another.

Keyboard and Keys ●● Keys can respond to the up-down and left-right functional
tests.
●● The Print key can respond when the video printer is
connected.

16 Service Manual
2  Unpacking and System Checks

Item Inspection

Trackball 1. Press the Caliper or Calc key to start measurement.


2. The trackball can be moved well from left to right and up to
down.

Peripheral Devices ●● The probe can be disconnected normally.


●● The connected probe can work normally.

Fan ●● Fan works well when the system is powered on.


●● No noises are produced during fan working.

Probe 1. Connect the probe (such as L741) to the system and ensure
that the screen displays the correct icons for the probe.
2. Select a preferred probe and exam type, and the system
automatically enters the B mode.
3. Press the CFM key to enter the CFM mode, and ensure that
color ROI is displayed on the screen.
4. Adjust the color gain to the maximum to check whether color
noise is normal.
5. Press the Freeze key, replay the cine by using the trackball to
check whether data display correct.
6. Press the M key to enter the inactivated M mode, press the
Update key to ensure that the M trace is displayed, and TGC,
depth and frequency can be adjusted.
7. Press the PW key to enter the inactivated PW mode, press
the Update key to ensure that the spectral Doppler trace is
displayed and no nosies exist.
8. Press the THI key to ensure that the system enters the THI
mode.
9. Disconnect the probe from the system. Select another probe
to repeat the above steps.

2.4.4  Image Tests

■ B Image Test

Adjust the B gain to the maximum in accordance with depths or frequencies, and check the
following items in the B mode.

Service Manual 17
2  Unpacking and System Checks

●● The initial echo field of the image should be displayed in an appropriate brightness
without any aberrations.
●● No unexpected stripes caused by artifacts are displayed on the image.

■ THI Image Test

Adjust the B gain to the maximum in accordance with depths or frequencies, and check the
following items in the THI mode.
●● The initial echo field of the image should be displayed in an appropriate brightness
without any aberrations.
●● No unexpected stripes caused by artifacts are displayed on the image.

■ CFM/PDI Image Test

Adjust the B gain and ROI to the maximum with the default depth and frequency, adjust
the CFM/PDI gain until few color interferences occur. In this case, there are no unexpected
stripes on the image. However, it is normal if some color noises appears accidentally.

■ PW/CW Image Test

Adjust the PW/CW gain in the real time PW/CW mode, until a clear spectral Doppler
image displays. Adjust the pulse repetition frequency, and no noises appears on the image.

18 Service Manual
3    Principle Descriptions

This chapter provides hardware and software principle descriptions, you should read these
descriptions thoroughly before attempting to service the system.

19
3  Principle Descriptions

3.1  System Block Diagram


The system consists of the main unit, the probes and the adaptor.

Main unit consists of single boards, including the ABF board, probe board, ON/OFF button
board, battery port board, power port board, interface extension board, keyboard, TGC
board, trackball board and WiFi board, and relevant electronic components such as LCD
monitor, lithium battery, hard disk, fan, speaker and so on.

Connections between boards are shown in Figure 3-1.

Battery

SM BUS
PWR
WIFI&BT HDD/SSD
Fan1、2 Fan3、4
+12V Vehicle Battery Port Board
Adapter
PWR

SATA
SDIO &
UART

Power
PWR SM BUS

PCN
Port
PWR_AC PWR
Adapter Borad PWR

Port1
FSW FSW_IN
Extend ECG_USB
Board USB_OTG
ECG 3Lead
RS232
Port2
Video MPC
ON/OFF ONOFF Main Board
ON/OFF_IN Board

Matching Board Bottom

Matching Borad ToP or


Board

HDMI
S-Video

Left/Right
USB *3
Ethernet
Audio
Print_ctrl/busy Port3
Reset
Other Signals

USB
eDP
PWR

Speaker L Speaker R

LCD
Keyboard

TGC
Track Ball
Board

Figure 3-1 System Block Diagram

Functions of the single boards are shown in Table 3-1.

Table 3-1 Single Board Functions

Board Function

ABF board It is the main control board of the system that controls the
sending of the ultrasound pulse, the reception and display of
the echo signal and so on.
It consists of the power supply unit, ultrasound unit (including
FPGA and AFE), main control unit ( including CPU unit and
its peripheral circuits) and so on.

20 Service Manual
3  Principle Descriptions

Board Function

Keyboard, TGC board These three single boards constitute the key panel module,
and trackball board providing an input platform for the user.

WiFi board A single board that provides the WiFi connection function.

Probe board A board with three probe ports connects the ABF board and
the probes.

Power port board Connect the adaptor, the ABF board, the battery board and
the ABF board, switching power supply from the adaptor to
the battery, or from the battery to the adaptor.

Battery port board Connect the battery and the power port adaptor, switching
the battery status from charging to discharging, or from
discharging to charging.

ON/OFF button board Connect to the ABF board, powering the ultrasound system
on or off.

Interface extension Connect the ECG electrode connectors, the foot switch and
board the video cable.
USB_OTG port and RS232 port on the board can be used to
debug the mainframe.

3.2  Hardware Description

3.2.1  ABF Board


ABF board consists of the power supply unit, ultrasound unit and main control unit.

Power supply unit includes power supply charge and discharge management circuit,
voltage conversion circuit, power supply MCU control circuit and removable lithium
battery module.

Ultrasound unit includes FPGA module, AFE module, high-voltage switch module and so
on.

Main control unit, the core control unit of ABF board, is used to process the data, show
the processed data on the monitor and store the data. It consists of CPU and its peripheral
circuit.

Service Manual 21
3  Principle Descriptions

Charge and Power


AC Adaptor +19VDC discharge Voltage Voltage of
supply
management conversion circuits
module

Lithium
MCU
battery

Figure 3-2 Power Supply Unit

Ports of the ABF board includes the probe board port, the adaptor port, the peripheral
device ports and so on. The peripheral device ports consist of the USB port, the keyboard
port, the LCD monitor port, the audio output port, the network port, the interface extension
port and so on. All ports of the ABF board are listed in Table 3-2.

Table 3-2 ABF Board Port List

No. Name

J1 Back end CPU debugging serial port

J2 Ventilation fan port

J3/J23 Probe board port

J4 FPGA JTAG debugging port

J5 Ventilation fan port

J6 Keyboard port

J7 Left channel speaker port

J8 Power port

J9 Hardware power port

J10 Button cell port

J11 Right channel speaker port

J12 Hardware signal port

J13 Single chip on-line burning and debugging port

J14 Interface extension port

J15 Ventilation fan port

J16 WiFi module port

J17 Ventilation fan port

22 Service Manual
3  Principle Descriptions

No. Name

J18 EDP port

J21 Ventilation fan port

J22 Ventilation fan port

J24 FPGA debugging port

J26 Ventilation fan port

J27 ON/OFF button port

P1 USB port

P2 USB port

P4 High-definition display port

P5 microSD (TF) port

P6 RJ-45 network port

P7 S-VIDEO port

P8 Video print control socket

P9 USB port

P10 Audio port

P10 Microphone output port

3.2.2  Keyboard
Keyboard, TGC board and trackball board constitute an interactive control panel. This
control panel consists of ultrasound-specific keys, general alphabetic and numeric keys,
knobs, TGC sliders, trackball and so on. The keyboard consists of the USB port circuit,
MCU processing circuit, power supply module, LED indicator unit, TGC scan unit, key
processing module, knob encoder module and trackball module.

The keyboard and ABF board are connected by cables such as USB signal cable, I/O signal
control cables and power cable, and communicate with each other through the USB port.

3.2.3  Probe Board


Probe board connects the ABF board and probes. Probe board converts 64 channel signal
to 128 array signal by the 32PCS HV2701 high-voltage switch and the non-locking relay,
and switches array signals freely between three probe ports. Probe board and ABF board
are connected by board to board connector.

Service Manual 23
3  Principle Descriptions

3.2.4  Power Port Board


Power port board is equipped with adaptor port and connects with the ABF board by board
to board connector.

3.2.5  Battery Port Board


Battery port board is equipped with battery port and connects with the battery port by
cable.

3.2.6  ON/OFF Button Board


ON/OFF button board is equipped with tact switch and connects with the cable by ABF
board.

3.2.7  Port Extender Board


Port extender board contains ECG isolating circuit, OTG debugging port, VIDEO port,
foot switch port and serial debugging port.

3.3  Software Description

3.3.1  Structure
Software system consists of hardware layer, operating system layer, intermediate layer and
application layer.

Application
Ultrasound system applications + Associated applications
layer

Intermediate
layer File system Network protocol OpenGL, QT

Operating
LinuxOS
system layer

Hardware
layer Computer platform + Ultrasound module + Control panel

Figure 3-3 Software System Structure

System module can be divided by applications into imaging mode, system setting and
system function, as shown in Figure 3-4.

24 Service Manual
3  Principle Descriptions

System Function Modules

Imaging Modes System Settings System Functions

Patient
B M(B + M) General Annotation Power On/Off
Management

Probe
CFM PW(B+PW) Body Mark Peripheral Report
Identification

PDI / DPDI CW Measure Report Exam Review End Exam

PHI /THI Color +Doppler DICOM Load Default Annotation Body Mark

Image
TDI CFMM About Scanning Mode
Parameters
Measurements
Image Storage
/Calculation

System
Printer
Upgrade

Figure 3-4 System Module

3.3.2  Workflow
Software system is an system program on basis of Linux operating system. Its workflow is
shown in Figure 3-5.

PowerOn

Start uboot

Start kernel

Start file system

Start all kinds of servers

Start application

Figure 3-5 Workflow

Service Manual 25
3  Principle Descriptions

To start up the system,

1. Press the button on the control panel of the system to power on the system..
2. Start up uboot to check whether the hardware works normally.
3. If the hardware works normally, start up the kernel of Linux operating system.
4. Start up file system and various service processes.
5. Start up the software.
You can also control the system by operating and checking interfaces.

26 Service Manual
4    Parts Disassembly

This chapter contains information for disassembling the system. You can assemble the system
by performing disassembly procedures on the contrary. During the disassembly, cabling and
wiring should be paid great attention to, and disassembled parts should be taken care of. If you
are not familiar with the replacement procedures, you are recommended to take photos during the
disassembly and take them for assembly reference.

27
4  Parts Disassembly

NOTE:

•• Only authorized personnel can assemble or disassemble the system.


•• Only the parts provided or recommended by the manufacturer can be used as
renewal parts.
The system consists of the battery, hard disk, LCD monitor, keyboard, speaker, fan and
probe board assembly and disassembly.

4.1  Battery Disassembly


No tool is needed during battery disassembly.

Perform the following steps,

Switch the battery lock to the position, and remove the battery according to the arrow
direction as shown in Figure 4-1.

2
1

Figure 4-1 Removing the Battery

4.2  Bottom Components Disassembly


One screwdriver (for M3 screw) is needed during bottom component disassembly.

Perform the following steps,

Use the screwdriver to remove the screws marked with numbers, as shown in Figure 4-2.

1 2
3
9

4
8
5 6 7

Figure 4-2 Bottom Component Disassembly

28 Service Manual
4  Parts Disassembly

4.3  Back Cover Disassembly


One screwdriver (for M3 screw) and one pair of tweezers are needed during back cover
disassembly.

Perform the following steps,


1. Use the tweezer to remove the rubbers marked with numbers, as shown in Figure 4-3.

1 2 3

4
5

6 7

Figure 4-3 Removing the Rubbers from the Back Cover

2. Use the screwdriver to remove the screws marked with numbers, as shown in
Figure 4-4.

Service Manual 29
4  Parts Disassembly

1 2 3

4 5

6 7

Figure 4-4 Removing the Screws

3. Use the screwdriver to remove the screws marked with numbers and remove the back
cover, as shown in Figure 4-5.

1 3
2 4

Figure 4-5 Removing the Back Cover

4.4  Control Panel Disassembly


One screwdriver (for M2.5 screw) and tweezer are needed during control panel
components disassembly.

Perform the following steps,


1. Use the tweezer to remove the rubbers marked with numbers, as shown in Figure 4-6.

30 Service Manual
4  Parts Disassembly

1 2

3 4

5 6

Figure 4-6 Removing the Rubbers from the Control Panel

2. Use the screwdriver to remove the screws marked with numbers, as shown in
Figure 4-7.

1 2

3 4

5 6

Figure 4-7 Removing the Screws from the Control Panel

3. Slightly pull the cover marked with numbers with fingers and remove the upper cover
according to the arrow direction, as shown in Figure 4-8.

Service Manual 31
4  Parts Disassembly

1 2

Figure 4-8 Removing the Upper Cover from the Control Panel

4. Use the screwdriver to remove the screws for grounding cable, pull out the cable, and
remove the control panel.

Cable

Screws for earth wire

Figure 4-9 Removing the Components from the Control Panel

4.5  LCD Monitor Disassembly


One cross screwdriver (for M3 screw) is needed during LCD monitor disassembly.

Perform the following steps,


1. Use the screwdriver to remove the hinge screws marked with numbers, as shown in
Figure 4-10.

32 Service Manual
4  Parts Disassembly

Figure 4-10 Removing the Hinge Screws

2. Remove the acetate tape of the cable connected to the display and pull out the wires.

Figure 4-11 Removing the Cable Connected to the Display

Service Manual 33
4  Parts Disassembly

3. Use the screwdriver to remove the hinge screws marked with numbers and remove the
components from the display, as shown in Figure 4-12.

1 3
2 4

Figure 4-12 Removing the Screws from the Display Components

4.6  Fan Disassembly


One cross screwdriver (for M3 screw) is needed during fan disassembly.

Perform the following steps,


1. Use the screwdriver to remove the screws marked with numbers, pull out the cable
and remove the components, as shown in Figure 4-13.

1 2 7 8

3
4 5 6

Figure 4-13 Removing the Fan Components

34 Service Manual
4  Parts Disassembly

4.7  Grounding Cable and S-Video Disassembly


One cross screwdriver (for M3 screw) is needed during grounding cable and S-video
disassembly.

Perform the following steps,


1. Use the screwdriver to remove the screws marked with numbers, pull out the cable
and remove the grounding cable and S-video, as shown in Figure 4-14.

1 2

Figure 4-14 Removing Grounding Cable and S-video

4.8  Battery Board Bracket Disassembly


One cross screwdriver (for M3 screw) is needed during battery bracket disassembly.

Perform the following steps,


1. Pull out the cable. For details, refer to Section 4.7 Grounding Cable and S-Video
Disassembly.
2. Use the screwdriver to remove the screws marked with numbers, and remove the
battery board bracket, as shown in Figure 4-15.

Service Manual 35
4  Parts Disassembly

1 2

5
4

Figure 4-15 Removing the Battery Board Bracket

Figure 4-16 Removing the Battery Board

4.9  ECG Components Disassembly


One cross screwdriver (for M3 screw) is needed during ECG components disassembly.

Perform the following steps,


1. Use the screwdriver to remove the screws marked with numbers, pull out the ECG
bracket, as shown in Figure 4-17.

36 Service Manual
4  Parts Disassembly

2
1

3 4 5

Figure 4-17 Removing the Screws from the ECG Bracket

2. Pull out the wire.

Wire

Figure 4-18 Pulling out the Wire

3. Use the screwdriver to remove the screws marked with numbers, pull out the ECG
components, as shown in Figure 4-19.

Service Manual 37
4  Parts Disassembly

1 2

3 4

Figure 4-19 Removing the ECG Components

4.10  Probe Board Disassembly


One cross screwdriver (for M3 screw) is needed during probe board disassembly.

Perform the following steps,


1. Use the screwdriver to remove the screws marked with numbers and screws from
the probe isolated cover to bottom cover, and remove the isolated cover, as shown in
Figure 4-20.

Bottom cover

Probe isolated cover

1 7
2 8

3 9
4 10

5 11
6 12

Figure 4-20 Removing the Probe Isolated Cover

2. Use the screwdriver to remove the screws marked with numbers, and remove the
probe board, as shown in Figure 4-21

38 Service Manual
4  Parts Disassembly

1 3
2

5 6 7

Figure 4-21 Removing the Probe Board

4.11  Hard Disk Disassembly


One screwdriver (for M3 screw) is needed during hard disk disassembly.

Perform the following steps,


1. Use the screwdriver to remove the screws marked with numbers, remove wires at the
end of the ABF board, and remove the hard disk, as shown in Figure 4-22.

1 3

2 4

Figure 4-22 Removing the Hard Disk

Service Manual 39
4  Parts Disassembly

4.12  SD Card Disassembly


No tool is needed during SD card disassembly.

Perform the following steps,


1. Switch the SD card lock to right, and take out the SD card as shown in Figure 4-23.

SD card slot

Figure 4-23 Taking out the SD Card

2. Switch the SD card lock to left, and lock the SD card slot.

4.13  Speakers Disassembly


One screwdriver (for M3 screw) is needed during speakers disassembly.

Perform the following steps,


1. Use the screwdriver to remove the screws marked with numbers, as shown in
Figure 4-24.

Speaker bracket

1 2 4 6
3 5

Figure 4-24 Removing Speakers

40 Service Manual
4  Parts Disassembly

4.14  ABF Board Disassembly


One screwdriver and socket head wrench are needed during ABF board disassembly.

Perform the following steps,


1. Remove the plate and bottom cover.
a. Use the screwdriver to remove the screws marked with number 1, 2. 3, and remove
the plate, as shown in Figure 4-25.
b. Use the screwdriver to remove the screws marked with numbers (from 4 to 17), and
remove the bottom cover.

4 5
14 1
6
15
2
7
16

13 8

17 9
12 3
11 10

Figure 4-25 Removing the Plate and Bottom Cover

Service Manual 41
4  Parts Disassembly

2. Use the screwdriver to remove the screws and bolts, and remove the ABF board.

3 1 2

5 6 7 8

Figure 4-26 Removing the ABF Board

42 Service Manual
5    Wiring Descriptions

This chapter provides descriptions for pin connections between the boards inside the system. You
should get a general overview of pins before you service the system.

43
5  Wiring Descriptions

5.1  Connections for Wiring


The overall system connection is shown in Figure 5-1.

LCD monitor

ABF board
Hard disk

(ECG optional accessories)


(Foot switch optional accessories)

ECG optional
Keyboard

accessories
ECG board
ON/OFF board
ECG optional accessories

Battery board
Adaptor board

Pilot light board

Figure 5-1 Overall System Connection

For details about the connection cables, refer to Table 5-1 and the following sections.

Table 5-1 Codes and Names of Connection Cables

No. P/N Description Figure

1 2275.01073 Keyboard - TGC Board Signal Cable Figure 5-2

2 2275.01159 ABF Board - Display EDP Cable Figure 5-3

3 2275.01737 ABF Board - Key Board Signal Cable Figure 5-4

4 2275.01161 ABF Board - ECG Board Signal Cable Figure 5-5

5 2275.01162 Adaptor Board - Battery Board Power Cable Figure 5-6

6 2275.01168 ABF Board - SATA Hard Disk Cable Figure 5-7

7 2275.01175 Keyboard - Trackball Board Signal Cable Figure 5-8

8 2275.01179 ABF Board - ON/OFF Button Board Cable Figure 5-9

9 2275.01286 ECG Board - BNC Cable Figure 5-10

10 2275.01287 Keyboard - Indicator Cable Figure 5-11

11 2275.01430 ABF Board - BNC Cable Figure 5-12

12 2275.01465 ABF Board - BNC& Foot Switch Cable Figure 5-13

44 Service Manual
5  Wiring Descriptions

5.2  Pin Connection

5.2.1  Keyboard - TGC Board Signal Cable


P/N: 2275.01073

Cable Material:

Figure 5-2 Keyboard - TGC Board Signal Cable

Pins:

Table 5-2 Keyboard - TGC Board Signal Cable Pins

Color Red Red Black Black White Green Orange White Green Orange

CN1 20 19 18 17 16 15 14 13 12 11

CN2 1 2 3 4 5 6 7 8 9 10

Color White Green Blue Blue Blue Blue Blue Blue Blue Blue

CN1 10 9 8 7 6 5 4 3 2 1

CN2 11 12 13 14 15 16 17 18 19 20

Instructions:
●● Connect CN1 to the TGC board (J1).
●● Connect CN2 to the keyboard (J1).

5.2.2  ABF Board - Display EDP Cable


P/N: 2275.01159

Cable Material:

Service Manual 45
5  Wiring Descriptions

Figure 5-3 ABF Board - Display EDP Cable

Pins:

Table 5-3 ABF Board - Display EDP Cable Pins

Color Blue Blue Blue Blue Blue Blue Blue Blue Blue Blue Red Red Blue

CN1 1 5 6 4 2 3 7 8 9 10 11 12 13

CN2 2 3 4 5 6 7 8 9 10 11 12 13 14

Color Blue Blue Blue Blue Blue Blue Blue Blue Blue Red Red Red Red

CN1 14 15 16 17 18 19 20 21 22 25 26 27 28

CN2 15 16 17 18 19 20 21 22 23 26 27 28 29

Instructions:
●● Connect CN1 to the ABF board.
●● Connect CN2 to the display.

5.2.3  ABF Board - Keyboard Signal Cable


P/N: 2275.01737

Cable Material:

Figure 5-4 ABF Board - Keyboard Signal Cable

Pins:

46 Service Manual
5  Wiring Descriptions

Table 5-4 ABF Board - Keyboard Signal Cable Pins

Color Red Red Red Black Black Black Yellow Yellow Yellow White Green White Green White Black

CN1 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29

CN2 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29

Color Shielded Orange/Blue Orange Orange Black Black White Green White Green White Green White Black

Double twisted

CN1 2 4 6 10 12 8 14 16 18 20 22 24 26 28 30

CN2 2 4 6 10 12 8 14 16 18 20 22 24 26 28 30

Instructions:
●● Connect CN1 to the ABF board (J6).
●● Connect CN2 to the keyboard (J1).

5.2.4  ABF Board - ECG Board Signal Cable


P/N: 2275.01161

Cable Material:

Figure 5-5 ABF Board - ECG Board Signal Cable

Pins:

Table 5-5 ABF Board - Keyboard Signal Cable Pins

Color Black Green Black Black Yellow Brown Black Yellow Yellow Black Yellow Yellow Black Brown Black

CN1 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29

CN2 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29

Color Red Black Green Brown Black Orange Brown Black Yellow Black Orange Brown Black Orange Black

CN1 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30

CN2 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30

Instructions:
●● Connect CN1 to the ABF board (J14).
●● Connect CN2 to the ECG board (J2).

Service Manual 47
5  Wiring Descriptions

5.2.5  Adaptor Converter - Battery Board Cable


P/N: 2275.01162

Cable Material:

Figure 5-6 Adaptor Converter - Battery Board Cable Pins

Pins:

Table 5-6 Adaptor Converter - Battery Board Cable Pins

Color White Green Black Black Black Black

CN1 1 3 5 7 9 11

CN2 1 3 5 7 9 11

Color Red Red Red Red Red Black

CN1 2 4 6 8 10 12

CN2 2 4 6 8 10 12

Instructions:
●● Connect CN1 to the adaptor converter (J1).
●● Connect CN2 to the battery board (J2).

5.2.6  ABF Board - SATA Hard Disk Cable


P/N: 2275.01168

Cable Material:
SERIAL
ATA

Figure 5-7 ABF Board - SATA Hard Disk Cable

Pins:

48 Service Manual
5  Wiring Descriptions

Table 5-7 ABF Board - SATA Hard Disk Cable

CN1 1 4 7 2 3 5 6

CN2 1 4 7 2 3 5 6

Remarks Three pins are conducted / / / /

Signal 3.3V GND 5V GND 12V

Color Blank Black Red Black Yellow

CN3 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15

CN4 NC 2 1 3 4

Instructions:
●● Connect CN1 to the ABF board (J9), (J12).
●● Connect CN2 to the hard disk.

5.2.7  Keyboard - Trackball Board Signal Cable


P/N: 2275.01175

Cable Material:

Figure 5-8 Keyboard - Trackball Board Signal Cable

Pins:

Table 5-8 Keyboard - Trackball Board Signal Cable Pins

Color Red Black White Green White Green White Green White

CN1 1 2 3 4 5 6 7 8 9

CN2 1 2 3 4 5 6 9 8 7

Instructions:
●● Connect CN1 to the trackball board (J1).
●● Connect CN2 to the keyboard (J6).

5.2.8  ABF Board - ON/OFF Button Board Signal Cable


P/N: 2275.01179

Service Manual 49
5  Wiring Descriptions

Cable Material:

Figure 5-9 ABF Board - ON/OFF Button Board Signal Cable

Pins:

Table 5-9 ABF Board - ON/OFF Button Board Signal Cable Pins

Color Red White Green Shielded

CN1 1 2 3 4

CN2 1 2 3 4

Instructions:
●● Connect CN1 to the trackball board (J1).
●● Connect CN2 to the ABF board (J27).

5.2.9  ECG Board - BNC Socket Cable


P/N: 2275.01286

Cable Material:
Black(Sheilded)

Red

Figure 5-10 ECG Board - BNC Socket Cable

Pins:

Table 5-10 ECG Board - BNC Socket Cable Pins

Color Shielded Red

CN1 1 2

CN2 Steel cover Core

Instructions:
●● Connect CN1 to the ECG board (J3).
●● Connect CN2 to the BNC socket.

50 Service Manual
5  Wiring Descriptions

5.2.10  Keyboard - Indicator Board Cable


P/N: 2275.01287

Cable Material:

Figure 5-11 Keyboard - Indicator Board Cable

Pins:

Table 5-11 Keyboard - Indicator Board Cable Pins

Color White Green White Black Red White Green Red

CN1 1 2 3 4 5 6 7 8

CN2 8 7 6 5 4 3 2 1

Instructions:
●● Connect CN1 to the indicator (J1).
●● Connect CN2 to the keyboard (J7).

5.2.11  ABF Board - BNC Socket Cable


P/N: 2275.01286

Cable Material:

Figure 5-12 ABF Board - BNC Socket Cable

Pins:

Table 5-12 ABF Board - BNC Socket Cable Pins

Color Shielded Red Black Black Black Black Black Black Black Black Black Black

CN1 1 3 2 4 13 14 15 16 27 28 29 30

CN2 Steel Core OPEN


cover

Instructions:

Service Manual 51
5  Wiring Descriptions

●● Connect CN1 to the ABF board (J14).


●● Connect CN2 to the BNC socket.

5.2.12  ABF Board - BNC & Foot Switch


P/N: 2275.01465

Cable Material:
Black(Sheilded)

Red

NOTE: The notch


should match
with the angle
of the mold.

Figure 5-13 ABF Board - BNC & Foot Switch

Pins:

Table 5-13 ABF Board - BNC & Foot Switch


Color Shielded Red Black Black Black Red Shielded

CN1 1 3 7, 11 short circuit 15 17 19

CN2 S t e e l Core / / / / /
cover

CN3 / / / / 3 2 1, 4

Instructions:
●● Connect CN1 to the ABF board (J14).
●● Connect CN2 to the BNC socket.

52 Service Manual
6    Software Installation and Update

The service department or local distributer can obtain the upgraded software package from the
manufacturer by the CD disk or Email. You can copy the package into the USB drive to update the
system with a reference to this chapter.

53
6  Software Installation and Update

6.1  Installing the Operating System

6.1.1  Making the System USB Drive

Table 6-1 List of System USB Drive Preparation Tools

Tool Specification Remark

USB drive Capacity not less than 4G FAT32 format, system USB drive

Windows XP operating
PC USB port available
system or above

OS_package_3_14_52-V_ 3_14_52-V_x_x_x corresponds to


System USB
x_x_x-int-release-kernel_ 3.14.52-Vx.x.x, the Kernel Version
drive package
xx_xx-yyyymmdd.zip of the system.

NOTE:

•• The format of the system USB drive should be changed to FAT32. Please backup
the data in the USB drive before formatting.
•• After completing the system USB drive preparation, please choose “USB Safely
Remove” and then unplug the USB drive.

■ Operations
1. Copy the system USB drive preparation package OS_package_3_14_52-V_x_x_x-int-
release-kernel_xx_xx-yyyymmdd.zip to a computer and extract the package. The
following files are contained in the package.

Figure 6-1 Files of the System USB Drive

2. Insert a USB drive to the computer and format it as FAT32.

54 Service Manual
6  Software Installation and Update

3. Copy update_file (shown in Figure 6-1) to the root directory of the USB drive.
4. Copy the package adata_5.*.*.*.tgz and asecure_5.*.*.*.tgz to file /update_file/
update/.
5. Safely remove the USB drive after completing copying.
The preparation of the system USB drive is complete.

6.1.2  Operating System Installation

Table 6-2 List of Operating System Installation Tool

Tool Specification Remark

Free capacity not less than


USB drive FAT32 format, system USB drive.
1G

Used for pressing the update


OTG converter /
switching button

Windows XP operating
PC USB port available
system or above

Two installation scenes are provided on the software update screen (USB drive reboot
screen). For Formatting Installation, install the operating system and applications and
format the hard disk synchronously. User data will be erased. For Save Data Installation,
only install the operating system and applications. User data will not be erased.

Ensure that the file in the USB of the system is complete before the installation. For
details, refer to Figure 6-2.

Service Manual 55
6  Software Installation and Update

Figure 6-2 File Tree of the System USB Drive

Only the steps of Formatting Installation will be detailed in this section. Save Data
Installation can also be performed in accordance with the following steps.
1. Connect the software package USB drive to the system when the system is powered
off.

56 Service Manual
6  Software Installation and Update

Software USB Drive

Update Switching Button Hole

Figure 6-3 Update Switching Button

2. Power on the system, wait approx. 1 second and use the aciculiform tool to press the
update button for approx. 3 seconds and remove the aciculiform tool. Wait until the
USB reboot screen (shown in Figure 6-4) appears.
You can view the versions of the operating system and applications of the current
system USB drive on the USB reboot screen.

Figure 6-4 Safety Mode Screen

3. Click AllInstall Format and a prompt pops up.

Service Manual 57
6  Software Installation and Update

NOTE:
Patient data will be erased after clicking AllInstall Format.

4. Click OK to start installing.


The installation process includes four stages and the progress bar for each stage will
be displayed on the screen.
a. Install the operating system (Step1: system upgrade).
b. Install the application (Step2: software upgrade).
c. Format the hard disk (Step3: disk format).
d. Complete the installation.
5. Click Exit to close the progress bar dialog box when All is success (the progress bar
displays 100%) appears.
6. Click PowerOff > OK on the pop-up dialog box to power off the system
7. Verify the installation.
Power on the system and View Kernel Version on the System Information screen

6.2  Updating the Application

6.2.1  Tool Preparation

Table 6-3 List of Application Update Tool

Tool Specification Remark

4 G c a p a c i t y , FAT 3 2 f o r m a t .
Any models with capacity
USB drive Application update package is saved in
not less than 4G
the USB drive.

Windows XP operating
PC USB port available
system or above

Apply for the update package from the


Application adata_*.*.*.*.tgz
specific website. *.*.*.* refers to the
update package asecure_*.*.*.*.tgz
software version.

NOTE:

•• Ensure that the prepared USB drive works normally on the desired system.
You can confirm the USB drive works normally by storing the system image
successfully into the USB drive.
•• Ensure that only one update file is saved in the USB drive.
•• Do not alter the format and file name of the update file in the USB drive.
Otherwise, the system may fail to update.

58 Service Manual
6  Software Installation and Update

•• Ensure that the system displays the current date. Otherwise, the system may fail
to update.
•• Ensure that the system can work normally.
•• Name or format of the update file provided by the manufacturer may be
inconsistent with that of the file user receives if the update file is sent through
E-mail. In such case, you can correct its name manually. For example, you can
correct asecure_*.*.*.*.tgz (1) to asecure_*.*.*.*.tgz.

6.2.2  Operations
1. Power on the system and inset the USB drive written with the update package to the
USB port of the system.
2. Click Software Update on the System Information screen.
3. Input the year password in the dialog box shown in Figure 6-5, and then press the
Enter key.

Figure 6-5 Inputting the Year Password

4. Click Software Update in the dialog box shown in Figure 6-6.

Figure 6-6 Software Update Screen

5. Click Exit when Files successfully copied. Please exit and reboot system to
complete update. pops up.
6. Unplug the USB drive and reboot the system.

NOTE:

• The password is not case sensitive. If the password is wrong, invalid password
is displayed to remind you for password retry.

Service Manual 59
6  Software Installation and Update

• You need to re-insert the USB drive or use another USB port if No USB driver
found pops up.
• Do not remove the USB drive during system reboot. Otherwise, the system
cannot be started up. The stored images or patient information will not be
damaged or lost if you update the software by the USB drive.

6.2.3  Update Verification


View Software Build On, Software Version, Application Version on the System
Information screen.

6.3  Importing the aconfig File

6.3.1  Tool Preparation

Table 6-4 List of aconfig File Import Tools

Tool Specification Remark

Any models with capacity FAT32 format. aconfig file is saved in the
USB drive
not less than 4G root directory of the USB drive.

Windows XP operating
PC USB port available
system or above

Ensure that the 7 digits in the file name is


aconfig file aconfigxxxxxxx consistent with the control number of the
system.

NOTE:

•• Ensure that the 7 digits in the file name is consistent with the control number of
the system.
•• Ensure that the aconfig file is saved in the root directory of the USB drive.
•• Ensure that the system display date is the current date. If it is not, modify the date
into the current date and save the modification. Or the aconfig file can not be
input.

60 Service Manual
6  Software Installation and Update

6.3.2  Operations

■ Performing the First Import

When you have updated the system or no aconfig file exists, the prompt “No valid license,
please import aconfig file.” pops up after powering on the system. You should copy the
aconfig file corresponding to the control number of the system to the root directory of the
USB drive, insert this USB drive to the USB port of the system, and press the Update key
to start importing. A prompt will be displayed after a successful import.

■ Performing an Active Import


1. Copy the aconfig file corresponding to the control number of the system to the root
directory of the USB drive.
2. Power on the system and insert the USB drive to the USB port of the system after
entering the application mode screen.
NOTE:
The digits in the aconfig file name is the control number of the system. For
example, in a file name of aconfig6589947, “6589947” corresponds to the control
number of the system.
3. Enter the About screen and click Software Update. A dialog box is displayed on the
screen, as shown in Figure 6-7.

Figure 6-7 Inputting the Year Password

4. Input the year password and press the Enter key. A dialog box is displayed on the
screen, as shown in Figure 6-8.
5. Click Software Update.

Figure 6-8 Loading aconfig File

Service Manual 61
6  Software Installation and Update

6. Click Exit when the prompt “Config updated” pops up.

NOTE:
If the aconfig file update fails, please click OK to perform the update again. If failure
still occurs, please check whether the digits in the aconfig file name is consistent with
the control number of the system, or whether the aconfig file is copied to the root
directory of the USB drive.

■ Verifying Update

View License Valid Days on the System Information screen.

■ Acquiring the Control Number


1. Power on the system and enter the application mode screen.
2. Press the Setup key to enter the System Settings screen.
3. Click About and press the confirm key to enter the System Information screen.
Control Number shows the control number of the system.

■ Acquiring the Serial Number

View the system serial number (in bar code format) on the rating label of the system. For
example, 0486589947.

62 Service Manual
7    Settings Import/Export

This chapter describes how to import/export the settings, which is closely related to system settings,
and customized application mode.

63
7  Settings Import/Export

7.1  Customizing a Preset


1. Select the desired probe and exam type.
2. Select an imaging mode and optimize the image.
NOTE:
Parameter optimizing is only available for the current imaging mode.
3. Press the S key on the key panel to display a dialog box.
4. Click Yes to enter the User Pre-Set Menu screen, as shown in Figure 7-1.

User Pre-Set Menu

Name

Type Abdomen

Part C-Vascular

Save Exit

Figure 7-1 User Pre-set Menu Screen

5. Input the name of the application preset in the Name textbox.


6. Select the department and exam type from the Type and Part lists.
7. Click Save to create the customized application preset.
If the name of the application preset exists, you can click Yes in the pop-up dialog box
to replace it.

7.2  Exporting a Customized Preset


This feature allows you to export all customized presets to the USB drive.

Perform the following steps.


1. Insert a USB Drive to the system.
2. Press the Setup key to enter the System Settings screen and click About > Software
Update. A dialog box is displayed on the screen, as shown in Figure 7-2.

64 Service Manual
7  Settings Import/Export

Figure 7-2 Software Update Dialog box

3. Input the month password and the software update content pops up, as shown in
Figure 7-3.

Figure 7-3 Software Update Content

4. Click Export Probetable, and probetable export success is displayed after a


successfully export.
5. View the exported file sonoexport in the USB drive.

7.3  Importing a Customized Preset


The default system icons will not be replaced or deleted when you import application
presets to the system.

NOTE:

•• Ensure that only one sonoimport file is contained in the USB drive. Otherwise, the
import may fail or the imported file cannot be recognized.
•• Ensure that all the probe presettings are deleted before restoring the application
presets. Otherwise, the system cannot restore the presets.
Perform the following steps.
1. Insert a USB drive to the system.
2. Press the Setup key to enter the System Settings screen and click About > Software
Update. A dialog box is displayed, as shown in Figure 7-4.

Service Manual 65
7  Settings Import/Export

Figure 7-4 Software Update Dialog box

3. Input the month password and the software update content is displayed, as shown in
Figure 7-5.

Figure 7-5 Software Update Content

4. Click Import Probetable and probetable import success is displayed after a


successful import.
5. Exit the System Settings screen, and press the Probe key to view the imported
customized preset.

66 Service Manual
8    Performance Test

You should perform the performance tests with a reference to this chapter after software update or
restore. In case of any test fails, you should solve it and perform the performance tests again until the
system works normally.

67
8  Performance Test

NOTE:

•• Only service engineers/representatives trained or qualified by the manufacturer


can perform the performance tests on the system.
•• To avoid damages to the system and personal injury, do not remove or replace
any circuit boards when the system is powered on.
•• Due to the potential hazardous voltage in the system, do not remove the insulation
unit of the system.

8.1  Automatic Hardware Tests


The system has a built-in testing software which can be utilized to test the performance
of the system hardware. After the system passed all the necessary performance tests, you
should restart the system.

Hardware tests are divided into two parts: underlying hardware tests and peripheral device
tests.

During hardware tests, the test results are displayed in the output box at the right side of
the TestItem list. The results in green indicate that the tested item works normally and
passes the test. While, the results in red indicate that the tested item results in malfunction,
and you should re-test it after servicing it.

NOTE:
During hardware tests, if the system is not connected with the probe, I/O docking
station, ECG, probe port extender board or 64-channel analogue generator, the test
results of the corresponding test items will be in red, indicating the tests fail.

8.1.1  Test Steps


1. Press the Setup key to enter the System Settings screen.
2. Click About, input the correct password and press the Enter key to enter the
EngineerMode screen, as shown in Figure 8-1.

68 Service Manual
8  Performance Test

Figure 8-1 Engineering Mode Screen

3. Click ABF Test or Peripheral Test to enter the corresponding screen.


4. Select a test item.
−− Move the cursor to the check box beside the desired test item, and press the confirm
key to select the test item.
−− Move the cursor to the check box beside Select All, and press the confirm key to
select all test items.
NOTE:
Move the cursor to the check box beside the selected test item, and press the
confirm key to deselect the test item.
5. Click Start to perform the selected tests.
You can press any key on the key panel to stop other tests but only complete the
current test during test.
6. Click Exit to exit the EngineerMode screen.

8.1.2  Underlying Hardware Tests

■ ABF Board Information Test

This test is used to acquire and display the ABF board information, including the
encryption status, part number, model and serial number of the ABF board. The
corresponding test item is ABFInfo. For details on test steps, please refer to Section 8.1.1
Test Steps.

Service Manual 69
8  Performance Test

■ FPGA Information Test

This test is used to acquire and display FPGA version information, including the platform,
version No. and release date. The corresponding test item is FPGAInfo. For details on test
steps, please refer to Section 8.1.1 Test Steps.

■ FPGA Clock Test

This test is used to check the FPFA clock on the ABF board. The corresponding test item is
FPGAClock. For details on test steps, please refer to Section 8.1.1 Test Steps.

■ PCIe Port Test

This test is used to test the connecting status and read-write rights of the PCIe port on the
ABF board. The corresponding test item is PCIeTest. For details on test steps, please refer
to Section 8.1.1 Test Steps.

■ FPGA DDR3 Register Test

This test is used to check the port status of FPGA DDR3 chip memory, including the
number of DDR, bit width, DDR clock frequency, the number of the supported ports,
controller address bit width and controller data bit width. The corresponding test item is
FPGA DDR3. For details on test steps, please refer to Section 8.1.1 Test Steps.

■ FPGA Kernel Voltage Test

This test is used to acquire and display the kernel voltage of FPGA on the ABF board.
The corresponding test item is FPGAKernelVol. For details on test steps, please refer to
Section 8.1.1 Test Steps.

■ FPGA PN Temperature Test

This test is used to acquire and display the PN temperature of FPGA on the ABF board.
The corresponding test item is Temperature. For details on test steps, please refer to
Section 8.1.1 Test Steps.

■ ABF Board Voltage Test

This test is used to acquire and display the voltage value of the ABF board. The
corresponding test item is ABFVol. For details on test steps, please refer to Section 8.1.1
Test Steps.

■ ABF Board Power Supply Status Test

This test is used to acquire and display the power supply status of circuits on the ABF
board. The corresponding test item is ABFPowerState. For details on test steps, please
refer to Section 8.1.1 Test Steps.

70 Service Manual
8  Performance Test

■ ABF Board High Voltage Test

This test is used to acquire and display the voltage value of the high-voltage power supply
on the ABF board. The corresponding test item is ABFHVPowerState. For details on test
steps, please refer to Section 8.1.1 Test Steps.

NOTE:
The result item of HVC_TYPE(CW) Voltage will be marked in red because the tested
voltage value is inconsistent with the actual value. This is normal and does not indicate
an ABF board malfunction, if voltage value tested by the system is correct.

■ Digital Circuit Status Test

This test is used to check the statuses of all the digital circuits. The corresponding test item
is ADCRampTest. For details on test steps, please refer to Section 8.1.1 Test Steps.

■ Analogue Circuit Status Test

This test is used to check the statuses of all the analogue circuits. The corresponding test
item is AnlogChTest. For details on test steps, please refer to Section 8.1.1 Test Steps.

8.1.3  Peripheral Device Test

■ Probe Test

This test is used to acquire and display the statuses and specifications (such as connection,
encryption, P/N, SN, ID, probe code and so on.) of the probe connected to the system or
the probes connected to the MPC board. The corresponding test item is Probe & MPC.
For details on test steps, please refer to Section 8.1.1 Test Steps.

■ I/O Docking Station Test

This test is used to check the communication and encryption statuses of the I/O docking
station and view the version of the I/O docking station. The corresponding test item is DK.
For details on test steps, please refer to Section 8.1.1 Test Steps.

■ Probe Port Extender Board Test

This test is used to check the communication and encryption statuses of the probe port
extender board and view the firmware version. The corresponding test item is Probe
expander. For details on test steps, please refer to Section 8.1.1 Test Steps.

Service Manual 71
8  Performance Test

■ Trackball Color Test

This test is used to check the color of the trackball. The corresponding test item is
Trackball Color. After selecting this test item, the trackball changes to seven different
preset colors in turn. The current color is displayed in the output box at the right side of the
TestItem list.

■ LCD Monitor Test

This test is used to check whether the monitor has color aberration and displays the images
normally. The corresponding test item is Screen Test.

Perform the following steps.


1. Select this item and click Start to test brightness, color temperature and image display
in turn.
−− During the brightness test, the brightness of the monitor changes automatically and
constantly. The current brightness value is displayed in the output box at the right
side of the TestItem list.
−− During the color temperature test, the monitor changes from cool color to warm
color or the other way around. The corresponding prompt is displayed in the output
box.
−− During image display test, the standard colorimetric images, including black,
white, blue, green and red are displayed in full screen format.
2. Check whether the monitor displays the brightness and color temperature correctly
and has color aberration or light leak.
The system exits the full screen display after completing the test.

■ Monitor Correction Test


1. This test is used to check whether the color temperature and brightness of the LCD
monitor is corrected. The corresponding test item is Screen Correct. Select this item
and click Start to enter the Display Adjust screen, as shown in Figure 8-2.

72 Service Manual
8  Performance Test

Figure 8-2 Display Adjust Screen

−− The green round displayed at the upper right corner of the screen indicates that the
monitor is successfully corrected.
−− The yellow round displayed at the upper right corner of the screen indicates that the
monitor fails to be corrected.
You should re-correct the monitor by using the color temperature tester because the
color temperature and brightness of the monitor deviates from the normal values.
And perform the test again after re-correction.
2. Click Exit to exit the Display Adjust screen.

■ Keyboard Test

This test is used to check whether the keyboard can respond correctly. The corresponding
test item is KeyboardTest.
1. Select this item and click Start to enter the keyboard test screen, as shown in
Figure 8-3.

Service Manual 73
8  Performance Test

Figure 8-3 Keyboard Test Screen

−− Click Off > Yellow > White, and ensure that the LED indicator lights up according
to the operation.
−− Press any key on the keyboard, and ensure that a hand icon is displayed on the
corresponding position on the screen.
−− Rotate all the knobs on the control panel, and ensure that a hand icon is displayed
on the corresponding position to indicate the rotation direction.
−− Slide the TCG sliders, and ensure that a hand icon is displayed on the corresponding
position on the screen.
2. Click Exit to exit the keyboard test screen.
If the test fails, you should re-test the keyboard after servicing it.

■ Speaker Test

This test is used to check whether the speaker works normally without noises. The audio
frequency of the speaker is between 1000Hz and 125Hz. You can test the left and right
channel of the speaker. The corresponding test item is Speaker. After selecting this test
item, click Start to test the current channel and check its frequency, and press any key to
test the other channel.

If the speaker works abnormally or has noises, you should re-test the speaker after
servicing it.

8.2  Function Settings


Click the Function tab on the EngineerMode screen to enter the function test screen, as
shown in Figure 8-4.

74 Service Manual
8  Performance Test

Figure 8-4 Function Configurations Screen

■ Setting Virtual Probe

NOTE:
The virtual probe is only a demo, and do not use it when performing an exam.

You can perform the relevant operations, for example, testing the system function without
probes, after selecting a virtual probe.

Perform the following steps.


Move the cursor to the check box beside VProbe by using the trackball, and press the
confirm key.
Press the left or right arrow key to select the desired virtual probe.
The probe code and model of the virtual probe is displayed beside VProbe.
Click Exit, and the desired virtual probe model and icon are displayed on the screen.
You can stop using the virtual probe by entering and exiting the EngineerMode
screen again.

■ Deleting Patient Data

NOTE:
This function is used to delete all patient data in the hard disk. Ensure that you back
up all patient data before performing this function.

Perform the following steps.

Service Manual 75
8  Performance Test

Move the cursor to the check box beside Delete patient data by using the trackball, and
press the confirm key.
Input del or DEL according to the pop-up prompt and then click OK.
The system starts deleting patient data. Waiting time varies according to the size of
patient data. After all patient data is deleted, the pop-up prompt disappear, and a green
prompt is displayed in the output box at the right side of the Item list.
Reboot the system.

■ Showing the Probe SN

The system displays the information of the connected probe.

Perform the following steps.


Use the trackball to move the cursor to the check box beside Show Probe SN, and click
Confirm.
Click Exit to exit function settings screen. Information of the connected probe will be
displayed on the left corner of the Application Mode, including probe ID, name and
serial number.
NOTE:
Normally, use the probe port extenders to connect three probes. When connected
probes are less than three, the information display of the unconnected probes
is blank. Enter the Engineer Mode again to click Exit to disable the showing the
probe SN feature.

■ Real Time Monitoring

The system displays the temperature of hard disk in operation on Main screen.

Perform the following steps.


Use the trackball to move the cursor to the check box beside the probe SN, and click
Confirm. Click Exit to disable function settings feature.
On Scanning Mode, press Ctrl+Alt+M key on the control panel, and Real Time
Monitoring menu is displayed on the left corner.
NOTE:
Monitoring data displayed in red font indicates that the test value is out of the
normal range.
On Scanning Mode, press Ctrl+Alt+M key or re-enter the Engineer Mode and
click Exit to disable the real time monitoring feature.

■ Adjusting Depth Range

The feature is used to indicate the maximum depth range of the system.

Perform the following steps.

76 Service Manual
8  Performance Test

Use the trackball to move the cursor to the check box beside adjusting depth range, and
click Confirm. Click Exit to exit the function settings screen.
Click Depth on the Scanning Mode. Current depth range can be viewed on the Scanning
Mode.
NOTE:
Do not enable this function when performing patient exam to avoid reducing
accuracy of the data. Re-enter the Engineer Mode and click Exit to exit the
adjusting depth range feather.
The depth the probe can reach in engineering mode may be different with the
depth the probe can reach when performing patient exam.

■ Optional Functions

Functions of the current system can be viewed in Optional Function menu.

Use the trackball to move the cursor to the check box beside adjusting depth range, and
functions of the current system are displayed on the right display box.

8.3  Firmware Update


This function allows you to update the FPGA firmware on the ABF board, ABF power
supply board, keyboard, ECG firmware, U-boot firmware. The firmware update screen is
shown in Figure 8-5.

Figure 8-5 Firmware Update Screen

NOTE:
Only professional engineers can perform this operation, as the operation may change
the settings of the fundamental devices and hardware.

Service Manual 77
8  Performance Test

Perform the following steps.


1. Copy the update file to the root directory of the USB drive. Ensure that the firmware
version is correct.
2. Insert the USB drive to the system.
3. Enter the firmware update screen on the engineer mode, and click the desired update
item to copy the corresponding update file.
After the copying is complete and the firmware update succeeds, a green prompt is
displayed in the output box at the right side of the UpdateItem list.
4. Reboot the system.

NOTE:

•• For keyboard update, you should reboot the system to start blue screen update,
after copying the update file.
•• For power supply firmware update, the system is automatically powered off to
complete the update, after copying the update file.

8.4  Burning Function


This function allows you to burn the Gamma color temperature of the LCD monitor,
encryption, EDP of the LCD monitor and MAC address.

NOTE:
Only professional engineers can perform this operation, as the operation may change
the settings of the fundamental devices and hardware.

Encryption burning should be performed for the system if the following tests fail in the
hardware tests.
●● Encryption test of the probe
●● Encryption test of the MPC
●● Encryption test of the I/O docking station
●● Encryption test of the ABFinfo

78 Service Manual
9    System Troubleshooting

This chapter contains important information for system troubleshooting. You can make fault
diagnostics and resolve it with a reference to this chapter.

79
9  System Troubleshooting

●● Potentially hazardous voltage exists inside the system. Remove personal jewelry that
! contains metallic substances when you operate the system.
●● Do not disassembly or remove any circuit boards when the system is powered on.
●● Identify whether the system malfunction is caused by incorrect operations or system
hardware/software failures to make a fault diagnostics.
●● Propose all possible factors, including system running time and external interferences,
that affect system functions.
●● Identify whether the system malfunction is a common fault. Fault caused by power
supply socket, (extension) power cord or fuse can be attributed to a common fault.
●● If the system malfunction is identified as a unusual fault, you can make the fault
recurrence in a similar test environment.
●● You can identify the system malfunction and solve it with your knowledge, experience
or a reference to 3 Principle Descriptions, Section 8.1 Automatic Hardware Tests and
9.1 Fault Diagnostics.
●● Do not replace a component if you are unsure if it is the cause.
●● If certain component is identified to be the fault cause, you should replace it only,
but not replace the whole circuit board that holds it. For example, a faulty capacitor
is identified to be fault cause, you should replace it only, but not the PCB board that
holds it.
●● Only the fault cause is identified, you should solve it and perform necessary tests
described in this chapter to ensure the system performance.

9.1  Fault Diagnostics


The common faults are listed as follows, you can take them as reference.

NOTE:
Fault diagnostics shown in Table 9-1 and Table 9-2 are for reference only, you should
make a on-site fault diagnostics with care.

80 Service Manual
9  System Troubleshooting

●● Startup Faults

Table 9-1 Startup Fault Diagnostics

Fault Description Solution

Unable to start The system cannot be Check whether the power supply is
up the system started up by pressing the correctly connected.
Power on/off button and the Check whether the Power on/off
back light of the keyboard button is pressed.
does not light up.
Check whether the requirement of the
voltage is met.

No Display No display on the monitor. Check whether the connecting cable


of the monitor is securely connected.

Blue Screen The system stops at a blue Press and hold the Power on/off
screen after startup. button to power off the system, and
reboot it.
If the fault cannot be solved, re-install
the system software.

●● Operating Faults

Table 9-2 Operating Faults Diagnostics

Fault Description Solution

Flashing Probe icons appear after Replace the probes.


Probe Icons the system enters the Ensure that the probe conversion board
Exam screen, but they are and ABF boards are connected well or
constantly flashing. there are no dry joints.

No initial No initial echo displayed on Ensure that voltages on the ABF board
echo on the the image. are normal.
image Ensure that there are high voltage
output.

Trackball The confirm key cannot Reboot the system.


Malfunction respond. Clean the trackball.
The Update key cannot If the fault cannot be solved, replace the
respond. trackball.
The cursor cannot be moved
by using the trackball.

Service Manual 81
9  System Troubleshooting

Fault Description Solution

Auto The system automatically Identify the operation causing the auto
Shutdown shuts down when you shutdown, and inform the malfunction
perform general system frequency to the manufacturer.
operations, such as Update the system software.
performing measurements,
connecting the USB drive or
freezing an image.

Unable to The system crashes during Press and hold the Power on/off button
shut down operation. to shut down the system.
the system If the fault cannot be solved, re-install
the system software.

9.2  System Troubleshooting


Ensure that you are familiar with the internal system structure, software installation, parts
assembly, corresponding principles and you can make a correct fault diagnostics before
you attempt to service the system.

9.2.1  On-Site Inspection


Get detailed information from the user before the on-site inspection, including system
working environment, fault description, fault frequency and frequency of system being
moved.
1. Check power supply and operating environment of the system, also get a knowledge
of local weather.
2. Make a detailed record of the system malfunction, including malfunction duration
and fault cause proposed by the user, and then make an initial solution plan with a
reference to 3 Principle Descriptions or your experiences.
Fault Diagnostics in Table 9-1 and Table 9-2 can also be used as a reference.

9.2.2  Fault Identification


1. If there is burn smell from the system, identify the burned board by checking the ABF
board and keyboard successively.
2. Ensure that all connection cables are complete and connected well before servicing
the system.

82 Service Manual
9  System Troubleshooting

9.2.3  Fault Resolution

■ Disassembling Parts

Observe and record the location and connection direction of parts when you replace them.
The parts in incorrect location or connection direction may result in system startup failure,
system malfunction, board burning, or even serious damages.

■ Replacing Parts

Replace failure parts one by one and test it after each replacement until the fault is solved.
You are not recommended to replace failure parts before system test. Meanwhile, make a
detailed record of the replacement sequence.

■ Assembling Parts

Install all replaced parts in the system by fastening all necessary screws. Meanwhile,
ensure that there are no screws left.

■ Taking a Simple Test


●● Take a simple test with a reference to Chapter 8 after you replace all failure parts.
●● If the system is connected to a server, ensure that all images can be backed up to the
server.
●● Ensure that all images can be displayed normal and available in all working modes,
such as B/PW/CW.
●● Adjust contrast and brightness of the monitor or other imaging parameters, such as
dynamic range, grayscale or gain, based on the user’s needs.

9.2.4  Post Maintenance

■ Inspection after Critical Parts Replacement

The system mainly consists of keyboard, ABF board, LCD monitor and probe conversion
board. It is very important to ensure the full connections among these parts. Therefore, you
should check the connections both before and after servicing the equipment.
●● Keyboard
Ensure that all keys, knobs, switches and trackball work normally after re-installation.
●● Probe Board
Ensure that the probe board and ABF board are connected well after re-installation.
●● Trackball
Ensure that the cursor can be moved smoothly by using the trackball after re-
installation.

Service Manual 83
9  System Troubleshooting

●● ABF Board
Ensure that the ABF board and keyboard are connected securely after re-installation.

■ Maintenance Items
●● Start up or shut down the system for several times to ensure no malfunction occurs.
●● Clean the system if it has been used for a long period. For details, refer to 10 System
Maintenance.
●● Record or solve problems proposed by the user.
●● Make a detailed service record for future maintenance or service. The fault cause,
replaced parts and their corresponding part numbers, repairing procedures should
be detailed in this record. Moreover, a signature from the hospital on this record is
required.

84 Service Manual
10    System Maintenance

To maintain the safety and functionality of the system, you should periodically perform the
maintenance for the system and accessories.

85
10  System Maintenance

●● To avoid electrical shock and damage to the system, power off and unplug the system
! from the AC power outlet before cleaning.
●● To maintain the safety and functionality of the system, maintenance must be
performed at least every 1 year. Electrical safety tests must also be performed at
regular intervals as specified by local safety regulations.

10.1  Cleaning the System

■ To Clean the System Surface


1. Power off and unplug the system from the AC power outlet.
2. Dampen a lint-free soft cloth with 75% ethyl alcohol.
3. Use this lint-free soft cloth to clean the surfaces of the monitor screen and the control
panel.
4. Use a cotton swab to remove solids from between keys and controls.
5. In case of blood or any infectious fluids or materials, use a lint-free soft cloth
dampened with 70% isopropyl alcohol.

■ To Clean the Trackball


1. Power off and unplug the system from the AC power outlet.
2. Press the bulges in the ring with both hands.
3. Turn the ring anticlockwise until the ring lifts.

4. Take out the ring and trackball (be careful not to drop the trackball) from the control
panel.
5. Use a lint-free soft cloth dampened with 75% ethyl alcohol to clean the trackball.
6. Use a cotton swab dampened with 75% ethyl alcohol to remove stains from the ring.
7. Place the trackball and the ring back in the control panel, press the bulges in the ring
with both hands and turn the ring clockwise to install the trackball.

86 Service Manual
10  System Maintenance

10.2  Probe Maintenance

10.2.1  Cleaning the Probe

●● To avoid electrical shock, disconnect the probe from the system before cleaning.
! ●● To avoid potential disease transmission, you should wear sterile medical gloves and
protective goggles while cleaning.

●● Do not use the cleaner that contains sodium sulfate and/or sodium hypochlorite to
clean the probe. Otherwise, the probe may be damaged.
●● To avoid probe damages, do not bump the probe on hard surfaces.
●● Use only the cleaner recommended by the manufacturer to clean the probe. Otherwise,
the probe may be damaged.

You should clean the probe before and after each use. Probe cleaning is an essential step of
the disinfection and sterilization.

Perform the following steps to clean the probe.


1. Remove the coupling gel from the probe surface after exam.
2. Disconnect the probe from the system and remove the probe sheath and biopsy
bracket from the probe.
3. Use a lint-free soft cloth dampened with mild soapy water or the recommended
cleaner (see the basic user manual) to wipe the probe.
If there are dried stains on the surface or in the grooves of the probe, remove them by
a brush dampened with an enzymatic clearer.
4. Rinse the surfaces of the probe with the fresh running water to completely remove any
residual cleaner.
5. Dry the probe with a lint-free soft dry cloth.

Service Manual 87
10  System Maintenance

10.2.2  Disinfecting or Sterilizing the Probe

●● There is a cleared list of sterilants and high-level disinfectants for use in processing
! reusable medical and dental devices on the FDA website. That list can be consulted
to find agents that may be useful for probe disinfection. Please refer to the following
URL for FDA-Cleared Sterilants and High Level Disinfectants: http://www.fda.gov/
MedicalDevices/DeviceRegulationandGuidance/ReprocessingofSingle-UseDevices/
ucm133514.htm.
●● Use the liquid disinfectant/sterilant that meets local laws and regulations or the
recommended disinfectant/sterilant.
●● The probe sheath may be damaged during exam and contamination risk level caused
by the damaged sheath is difficult to estimate. Therefore, probe sheath cannot change
the level of disinfection for a probe. Please clean and sterilize (or high level disinfect)
the endocavitary probe and clean and sterilize the probe for biopsy and coming into
contact blood after exam.
●● Do not use an expired disinfectant/sterilant.
●● Store the probe in a sterile environment, and review its expiration date before use.

●● If you use the recommended disinfectant/sterilant, ensure that the requirements for
the disinfection level and probe soaking period are met. Otherwise, the probe could
be damaged and your warranty might be void. If you have further use or needs for the
probe, ensure that the level of the make-up disinfectant/sterilant and the probe soaking
period are suitable for the intended clinical application.
●● Do not use the following methods to sterilize the probe.

−− Autoclave
−− Ultraviolet sterilizer
−− Gamma ray sterilizer
−− Ethylene oxide gas
−− Ethylene oxide
−− Thermal disinfection/sterilization. Temperature higher than 66°C (150°F) will
damage the probe.
●● Do not allow any disinfectant/sterilant to be air-dried on the probe.

■ Disinfection Levels

To choose an appropriate disinfectant, you must first determine the required level of
disinfection, based on the probe classification.

88 Service Manual
10  System Maintenance

Table 10-1 Disinfection Classifications

Classification Definition Level of Disinfection Application

Critical A device that enters Sterilization Intraoperative,


normally sterile tissue biopsy use or
or the vascular system. blood contacting
probe

Semi-critical A device that comes High-level Endocavitary


into contact with intact disinfection probe
mucous membranes
and does not ordinarily
penetrate sterile tissue.

Noncritical Devices that do no Medium or low-level Body surface


ordinarily touch the disinfection probe
patient or touch only
intact skin.

■ Soaking Requirements

Do not immerse the probe beyond its junction line.

■ To Disinfect the Body Surface Probe

You should disinfect the body surface probe after each use.

Perform the following steps.


1. Clean the probe.
For details, refer to Section 10.2.1 Cleaning the Probe.
2. Perform the medium-level disinfection to the probe as shown in Table 10-2.
Or select a recommended disinfectant (see the user manual) to disinfect the probe by
following the instruction provided by disinfectant’s manufacturer.
3. Rinse the probe with fresh running water to completely remove the residual
disinfectant.
4. Dry the probe with a lint-free soft dry cloth.

Service Manual 89
10  System Maintenance

Table 10-2 Recommended Method for Medium-level Disinfection

Disinfectant Manufacturer Active Level for Active Contact Contact


Ingredient Ingredient Type Period

70% Isopropyl ALL 70% Isopropyl 70% Spray/ <10


alcohol alcohol Wipe minutes

T-spray II Pharm. Inc. Quat. / Spray/ <10


Ammonia Wipe minutes

T-spray Pharm. Inc. Quat. / Spray/ <10


Ammonia Wipe minutes

■ To Disinfect the Endocavitary Probe

You should disinfect the endocavitary probe before and after each use.

Perform the following steps.


1. Clean the probe.
For details, refer to Section 10.2.1 Cleaning the Probe.
2. Perform the high-level disinfection to the probe as shown in Table 10-3.
Or select the recommended disinfectant (see the basic user manual) to disinfect the
probe by following the instruction provided by disinfectant’s manufacturer.
3. Rinse the probe with running sterile water to completely remove the residual
disinfectant.
4. Dry the probe with a lint-free soft dry cloth.

Table 10-3 Recommended Method for High-level Disinfection

Disinfectant Manufacturer Active Level for Active Contact Contact


Ingredient Ingredient Type Period

Cidex™ J&J Glutaraldehyde 2.4% Soak 45-50


Activated minutes
Dialdehyde
Solution

Resert XL HLD STERIS 2.0% H2O2 2.0% Soak 8 minutes

Tristel Duo for Tristel ClO2 / Wipe 0.5-2


Ultrasound minutes

Tristel Trio Tristel ClO2 0.180-0.220g/L Wipe 0.5-2


Wipes System minutes

90 Service Manual
10  System Maintenance

10.2.3  Sterilizing the Probe


You should sterilize the intraoperative, biopsy use probe or any probe that comes into
contact with blood after each use.

Perform the following steps.


1. Clean the probe.
For details, refer to Section 10.2.1 Cleaning the Probe.
2. Sterilize the intraoperative probe as shown in Table 10-4.
Or select the recommended disinfectant (see the basic user manual) to sterilize the
probe by following the instruction provided by disinfectant’s manufacturer.
3. Rinse the probe with running sterile water to completely remove the residual sterilant.
4. Dry the probe with a lint-free soft dry cloth.

Table 10-4 Recommended Method for Sterilization

Disinfectant Manufacturer Active Level of Active Contact Contact


Ingredient Ingredient Type Period

Cidex™ Activated J&J Glutaraldehyde 2.4% Soak 10 hours


Dialdehyde
Solution

10.2.4  Disinfecting and Sterilizing the Probe Cable

To avoid potential disease transmission, you should wear sterile medical gloves and
! protective goggles during disinfection or serialization.

Do not immerse the probe cable in alcohol or isopropyl alcohol at any concentration for a
long period of time. Avoid any contact with iodine or phenols. Otherwise, it will accelerate
the aging of the cable and decrease the service time of the probe, or even result in cable
damages.

■ Disinfection

Perform the following steps to disinfect the probe cable.


1. Use a lint-free soft dry cloth dampened with mild soapy water to wipe the probe cable.
2. Rinse the cable with fresh running water to completely remove the residual cleaner.
3. Perform the medium-level disinfection to the probe cable as shown in Table 10-5.
4. Rinse the surfaces of the probe cable with fresh running water to completely remove
the residual disinfectant.

Service Manual 91
10  System Maintenance

5. Dry the probe cable with a lint-free soft dry cloth.

Table 10-5 Recommended Method for Medium-level Disinfection

Disinfectant Manufacturer Active Level of Active Contact Contact


Ingredients Ingredients Type Period

T-spray II Pharm. Inc. Quat. / Spray/ <10 minutes


Ammonia Wipe

T-spray Pharm. Inc. Quat. / Spray/ <10 minutes


Ammonia Wipe

■ Sterilization

If the probe cable comes into contact with any blood or body fluid, you should sterilize it.
For details, refer to the information about probe sterilization.

10.2.5  Storage and Transportation

■ To Store the Probe

Follow the following items to store the probe.


●● Store the probe in a probe holder or specific package when not in use.
●● Ensure that the probe holder is clean before storing it.
●● To avoid contaminating the package, disinfect or sterilize the probe before storage or
store it in a disposable probe package.
●● Do not expose it to direct sunlight or sudden environmental temperature changes.
●● Store the probe alone to avoid any impact on the probe head. 
●● Carefully wind the probe cable.
●● Dry the probe before storing it.

■ To Transport the Probe

Perform the following steps to transport the probe.


1. Clean and disinfect the probe before transportation.
2. Place the probe fully into the carrying case and carefully wind the cable.
3. Close the cover of the carrying case.
4. Pack the carrying case with sponge and place it in a carton.

92 Service Manual
10  System Maintenance

10.3  Biopsy Bracket Maintenance

10.3.1  Cleaning the Biopsy Bracket

●● Clean the biopsy bracket after each use.


! ●● Wear sterile medical gloves and protective goggles during cleaning.

Perform the following steps.


1. Disconnect the probe from the system, and remove the biopsy bracket and biopsy
guide tube from the probe.
2. Brush and rinse the bracket with enzymatic cleaner to remove any residue.
3. Dry the bracket with a lint-free soft dry cloth.

10.3.2  Sterilizing the Biopsy Bracket

●● Disinfect the biopsy bracket after each use.


! ●● Use the liquid sterilant that meets local laws and regulations.
●● If you use the recommended sterilant, ensure that the requirements for the sterilant
level and biopsy bracket soaking period are met. Otherwise, the biopsy bracket could
be damaged and your warranty may be void. If you have further use or needs for the
probe, ensure that the level of the make-up sterilant and the biopsy bracket soaking
period are suitable for the intended clinical application.
●● Do not use the expired sterilant.
●● If you use other contact type to sterilize the biopsy bracket, ensure that it is guaranteed
by the manufacturer or the local distributer.
●● Do not sterilize the biopsy bracket with ethylene oxide.
●● Do not allow any sterilant to be air-dried on the biopsy bracket.

Perform the following steps.


1. Clean the biopsy bracket.
For details, refer to Section 10.3.1 Cleaning the Biopsy Bracket.
2. Sterilize the biopsy bracket as shown in Table 10-6 for the entire recommended
soaking period.
3. Rinse the bracket with running sterile water to completely remove the residual
sterilant.
4. Dry the bracket with a lint-free soft dry cloth.

Service Manual 93
10  System Maintenance

Table 10-6 Recommended Method for Sterilization

Sterilant Manufacturer Active Level for Active Contact Contact


Ingredients Ingredient Type Period

Cidex™ Activated J&J Glutaraldehyde 2.4% Soak 10 hours


Dialdehyde
Solution

10.3.3  Storage
Ensure that the biopsy bracket is cleaned, sterilized and completely dried before storage.
Store the biopsy bracket in a sterile environment and keep it away from the environment of
sudden temperature changes or direct sunlight.

10.4  Equipment Disposal


Date of manufacture is located on the product label. Under normal conditions, the service
life of the probe is about twenty-five thousand hours. Once the time expires, you should
dispose of it in accordance with local laws or regulations.

You should dispose of the main unit, the probe, the biopsy bracket and other accessories in
accordance with local laws and regulations.

For the detailed disposal information, consult the manufacturer or the local distributor.
The manufacturer is not responsible for any system content or accessories that have been
discarded improperly.

10.5  Customer Service


Only the service personnel of or authorized by the manufacturer can service the system.
Any feedback or inquires concerning our product or service should be directed to the
following address.

Address: 4/F, 5/F, 8/F, 9/F & 10/F, Yizhe Building, Yuquan Road, Nanshan, Shenzhen,
518051, Guangdong, China

Tel: +86-755-26722890

Fax: +86-755-26722850

Email: service@sonoscape.net

94 Service Manual
Appendix  Renewal Parts

Parts P/N Quantity

64 channel ABF board 7500.01271 1

32 channel ABF board 7500.01304 1

64 channel No 1 probe board 7500.01392 1

64 channel No 2 probe board 7500.01301 1

64 channel No 3 probe board 7500.00823 1

32 channel No 1 probe board 7500.01393 1

32 channel No 2 probe board 7500.01303 1

SD card 2100.00960 1

Port converter (including ECG, foot switch and CVBS 7500.00985 1


port)

Keyboard 3600.02199 1

LCD monitor 3600.02198 1

Button cell 2100.00673 1

ON/OFF button board 7500.00986 1

Battery board 7500.00969 1

Adaptor converter 2100.00968 1

Lithium battery 2100.00883 1

Hard disk 2100.00110 1

Speaker 2100.00921 2

Fan 2100.00973 4

WiFi board 3600.02467 1

95

You might also like