Professional Documents
Culture Documents
Laparoscopic Insufflators: Biomedical Guide
Laparoscopic Insufflators: Biomedical Guide
Laparoscopic Insufflators
Procedure Number 466-20210319
Scope
Applies to pneumatically and electronically controlled insufflators intended for introduction of CO 2 gas into the peritoneal
space.
Interval
12 Months
Time Required
0.5 Hour
Questions?
Please contact BiomedicalGuide@ecri.org
Adapt Procedure For
Insufflators, Laparoscopic [16-849]
Scope
Applies to pneumatically and electronically controlled insufflators intended for introduction of CO 2 gas into the peritoneal
space.
Interval
12 Months
Time Required
0.5 Hour
Overview
Insufflators are used to establish and maintain the pneumoperitoneum during laparoscopic procedures. Gas, typically CO 2, is
used to inflate the peritoneal cavity to provide a viewing and working space within the abdomen.
The primary function of the insufflator is to act as a pressure-controlled gas flow regulator. The insufflator uses compressed
gas from a supply cylinder (700 to 850 psi) or wall outlet (50 to 100 psi) and delivers it into the patient’s peritoneal cavity,
typically at 10 to 15 mm Hg (0.2 to 0.3 psi). In pneumatic insufflators, abdominal pressure control is accomplished by limiting
the delivery pressure of the gas to the desired abdominal pressure. Electronic insufflators typically deliver gas at a pressure
higher than that desired in the pneumoperitoneum; these units limit abdominal pressure by slowing and then suspending flow
when the abdominal pressure reaches the user-selected pressure.
Pneumatic and electronic insufflators have controls for setting the pressure and maximum flow rate. For example, they feature
displays, gauges, or other indicators for set and detected abdominal pressure and flow, volume of gas consumed, and external
gas cylinder pressure or volume remaining. With pneumatic insufflators, flow is typically specified as high or low. With
electronic insufflators, flow rates are specified either as a time-averaged flow or as an instantaneous flow. The maximum flow
possible from a given insufflator varies depending on flow resistance introduced by in-line tubing filters and by the stopcock
connection through which the insufflator is connected to the patient.
Special Precautions
If there is evidence of blood or body fluid contamination, submit the device for cleaning and decontamination before
inspecting it. (See IPM Safety in Biomedical Guide HTM Resources, Technical Documents for additional precautions and
guidelines.)
When testing insufflators that employ nitrous oxide or helium, any exhausted gas should be scavenged to protect personnel.
Laparoscopic Insufflators | 2
Procedure Number: 466-20210319
Procedure
Be sure that you understand how to operate the equipment and the purpose of each control, indicator, and alarm. Before
beginning an inspection, carefully read this procedure and the inspection and preventive maintenance procedure
recommended by the manufacturer (typically included in the service manual). Use Biomedical Guide’s Support Assessment
Form to document a scheduled maintenance interval and procedure that reflects past experience with this model and type of
equipment and the environment where it is used. Also consider BiomedicalBenchmark Maintenance Data for this model and
type of equipment.
The following procedure contains tasks for Acceptance and Scheduled inspections; tasks not labeled [ Acceptance] or
[Scheduled] apply to both types of inspections. The form provided for the General Devices IPM procedure can be used to
develop a form for documenting performance of this procedure.
Qualitative Tasks
Chassis/Housing [Acceptance]
Check for shipping damage; report any damage to the manufacturer, shipper, or service organization, and arrange for repair
or replacement.
Check that the insufflator is suitably constructed to withstand normal hospital use and abuse. For instance, a unit with venting
on the top of the housing or poorly protected or sealed controls and indicators may be prone to fluid entry. If not correctable,
warn users or take other preventive measures.)
Examine the exterior of the insufflator for cleanliness and general physical condition. Ensure that plastic housings are intact,
that all assembly hardware (e.g., screws, fasteners) is present and tight.
Chassis/Housing [Scheduled]
Examine the exterior of the insufflator for cleanliness and general physical condition. Ensure that plastic housings are intact,
that all hardware is present and tight, and that there are no signs of spilled liquids or other serious abuse. Inspect the gas
outlet port for evidence of fluid entry, which can indicate contamination by body fluids.
Mount/Fasteners [Acceptance]
If the insufflator is mounted on a stand or cart, ensure that the assembly and weight distribution is stable and that the
insufflator will not tip over during transport. If it is attached to a wall, check the security of this attachment. If the insufflator
is designed to rest on a shelf, ensure that it has nonslip legs or supports.
Mount/Fasteners [Scheduled]
If the insufflator is mounted on a stand or cart, examine the condition of the mount. If it is attached to a wall, check the
security of this attachment.
AC Plug [Acceptance]
A solidly constructed, good quality plug with adequate strain relief is acceptable, but the use of a Hospital Grade plug
(identifiable by a green dot and/or labeling) will eliminate guesswork and ensure a plug of acceptable construction quality.
Right-angle plugs are unacceptable for devices that are moved frequently. A good quality two-prong plug is acceptable for
double-insulated devices. Replace the plug or have the supplier replace it if it is not Hospital Grade or otherwise suitable.
Hospital Grade molded plugs are acceptable.
Shake the plug and listen for rattles that could indicate loose screws. If any damage is suspected, open the plug and inspect
it.
AC Plug [Scheduled]
Examine the AC power plug for damage. If any damage is suspected, open the plug and inspect it.
Line Cord [Acceptance]
Ensure that the line cord is long enough for the incubator's intended application; an extension cord should not be required. (A
length of 10 ft [3 m] is suitable for most applications.)
Laparoscopic Insufflators | 3
Procedure Number: 466-20210319
Equipment having a detachable power cord should also have adequate capture devices, cleats, or channels to hold the cord in
place. If these are absent, request that the supplier provide suitable means of securing the cord. Verify that the incubator has
adequate alarms or indicators for line-power loss and battery depletion and an adequate battery-charging indicator.
Tubes/Hoses
Check the condition of gas-supply hoses. They should be clean and without cracks, particularly at the hose fittings.
Filters [Acceptance]
Check for the presence and condition of internal gas filters. Record the manufacturer's recommended interval for service of
internal filters (typically 2 years).
Filters [Scheduled]
Check the condition of internal gas filters. Clean or replace as appropriate. Follow the manufacturer's recommended interval
for service of internal filters (typically 2 years) and instructions for replacement.
Controls/Switches [Acceptance]
Verify that software setup parameters accessible through hidden or service menus are correctly set for the appropriate
application and are consistent for all insufflators. Instruction and service manuals may contain instructions regarding such
modes. If they do not, contact the manufacturer. Discuss appropriate settings with the department head and users. If alarm
capabilities are included, see the Acceptance test for Alarms.
Examine all controls and switches for physical condition, secure mounting, and correct motion. If a control has fixed-limit
stops, check for proper alignment, as well as positive stopping.
For the model configured with a touch screen, check every available key on the screen, ensure the screen is working properly
and execute right operation. Ensure the screen is normal with no missing characters or any bright spots and dark shadows on
the screen.
Laparoscopic Insufflators | 4
Procedure Number: 466-20210319
Controls/Switches [Scheduled]
Examine all controls and switches for physical condition, secure mounting, and correct motion. Check that control knobs have
not slipped on their shafts. Where a control should operate against fixed-limit stops, check for proper alignment, as well as
positive stopping. During the course of the inspection, check that each control and switch performs its proper function.
Indicators/Displays
Confirm the operation of all lamps, indicators, meters, gauges, and visual displays on the insufflator. Observe that all
segments of a digital display function.
User Calibration
Verify that any pre-use calibration function operates.
Alarms
Induce alarm conditions to activate audible and visual alarms. Check that all associated interlocks or features function. If the
insufflator has an alarm-silence feature, check the method of reset (i.e., manual or automatic) against the manufacturer's
specifications.
Note the bag pressures at which intermittent or continuous audible alarms and visual indicators are activated during the
Pressure Setting Accuracy and Pressure Relief Mechanism quantitative inspection tasks. Confirm appropriate volume, as well
as the operation of a volume control if so equipped.
Labeling
Check that all necessary placards, labels, conversion charts, and instruction cards are present and legible.
Manuals [Acceptance]
An operators and service manual (electronic or hard copy), including schematics, should be provided with the monitor.
Software
Ensure all the software are up to date.
Accessories
Verify that all necessary accessories have been supplied with the unit.
Quantitative Tasks
Grounding Resistance [Acceptance] 0.5
Measure the resistance between the grounding pin of the power cord (if so equipped) and exposed (unpainted and not
anodized) metal on the chassis. Grounding resistance should not exceed 0.5 . Grounding resistance measurement is not
applicable to double insulated devices.
Depending upon circumstances, including prior history of grounding failures or power cord damage, frequency of use, and
environmental factors, this task should be considered for inclusion in scheduled inspections.
Laparoscopic Insufflators | 5
Procedure Number: 466-20210319
Inspect AC adapters used to power (or recharge) certain devices for CE mark or UL (or other testing laboratory) listing and to
verify that it is labeled to identify the device with which it is to be used. ECRI recommends testing of adapters, particularly
those that are not listed, by measuring the leakage current from each secondary (low voltage) connection to ground. The
leakage current should not exceed the limit for the device touch current.
Laparoscopic Insufflators | 6
Procedure Number: 466-20210319
With the insufflator set to its maximum flow setting, measure the time required to fill an empty 3 L solution bag to a typical
pressure setting (e.g., 15 mm Hg) through the tubing/filter set intended for use with the insufflator, and record it. Calculate
the flow using the following formula:
Clean
Clean the exterior (interior, if required). Wipe the surface of the device with a moistened soft cloth with the surface
disinfectant and make sure no water gets into the device. Do not attempt to sterilize the device.
Lubricate
Lubricate per the manufacturer's instructions.
Calibrate
Calibrate pressure settings, if required.
Replace
Replace filters, if required.
Laparoscopic Insufflators | 7
Procedure Number: 466-20210319