Throughput Elemental
Speciation Analysis
using GC-ICP-MS
Eckard Jantzen, Jasmin Mingo,
Jürgen Kuballa, Sabrina Sievers
R&D Department, GALAB Laboratories,
Hamburg, Germany
Figure 1. Calibration curve for tributyltin from 0 to 100 ng/L (ppt)
GALAB is an independent service contaminants. Actual or expected an Agilent 7900 ICP-MS using the
laboratory that analyzes and evaluates changes to the regulatory framework Agilent GC-ICP-MS interface
contaminant elements and compounds create a demand for more sensitive and (G3158D). The GC was fitted with an
in food, food packaging, sanitary robust analytical procedures. In this Agilent J&W DB-5ms Ultra Inert
products, industrial products, article, a sensitive, high-throughput column. The ICP-MS was fitted with
biopharmaceutical products, and technique is presented for the platinum-tipped interface cones.
environmental samples. We are well- detection of organometallic species
equipped with the latest analytical in different matrices. Table 1. GC-ICP-MS operating parameters
equipment, and use a broad range of
standard and tailored analytical Currently, most laboratories use Parameter Value
methods to provide our clients with GC/MSD, GC-FPD, or GC-AED for RF power (W) 1100
high-quality data. Our analytical the detection of organometallic
Injection port temperature (°C) 280
database comprises more than 5000 compounds. None of these techniques
single substances and more than offers the required sensitivity, stability, Helium carrier gas flow rate (mL/min) 19.6
1000 different analytical methods. or robustness to meet evolving
Despite this range, we are constantly environmental requirements. However, Injection volume (µL) 1
developing new methods on behalf ultrasensitive high throughput
of our clients to ensure the quality elemental speciation analysis Table 2. GC temperature program
and safety of their products. (UHTESA) is now possible using Start temp (°C) 50
GC-ICP-MS with a suitable sample
Speciation of organotin compounds preparation technique [5]. Heating rate (°C/min) 40
Organometallic species may occur Max temp (°C) 320
naturally in the environment or be Experimental
present as a result of human activity Reagents and samples Hold time (min) 3
[1]. Organometallic compounds are Analytical grade reagents were
often more bioavailable and toxic bought from Merck, Germany, and Appropriate sample preparation
than inorganic compounds of the were used without further purification. (extraction and derivatization)
same element. This is especially true Single compound and mixed organo- techniques are an important
for organotin compounds (OTCs), such tin standards were obtained from prerequisite to the accurate
as tributyltin (TBT) and triphenyltin Campro Scientific, Germany. Sodium determination of organometallic
(TPhT). Trisubstituted OTCs are tetraethylborate was from Synthese species by GC. The species must be
extensively used as biocides in Nord, Germany. transformed into their peralkylated
agrochemicals, antifouling paints, nonpolar derivatives. A detailed
wood preservatives and for material Instrumentation description of the sample preparation
protection. Mono-alkyl and dialkyl An Agilent 7890 GC was coupled to steps is given in reference [1].
tin compounds are widely used as
PVC stabilizers in packaging and
coating materials, foils, and various
types of piping [2]. Both TBT and
TPhT are well-known endocrine
disruptors that have contaminated
the environment for more than 50
years [3]. The biological effects of
OTCs on organisms depend on both
the nature and number of organic
groups bound to the Sn(IV) cation,
with the highest level of toxicity
being exhibited by the trisubstituted
compounds like TBT [4].
Conclusions
The GC-ICP-MS method is robust,
reproducible, and highly sensitive. It
meets the requirements of current
legislation, including the low
specifications for tributyltin specified
in Germany’s WRRL (2013/39/EW)
framework for surface water
surveillance monitoring of 0.2 ng/L.