SOMATOM

CT

Software
Installation of SOMARIS/5 VA47C

This document is valid for:


Balance / Emotion / Emotion Duo / Esprit / Esprit+
Emotion (2003) / Emotion Duo (2003)
© Siemens AG 2002
The reproduction, transmission or
Volume Access / Zoom use of this document or its contents
is not permitted without express
Sensation 4 written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.

English
Print No.: CT02-021.816.09.04.02 Doc. Gen. Date: 03.04
Replaces: CT02-021.816.09.03.02
0-2 Revision

Chapter Page Revision


0 to 5 all 01
0 to 5 all 02
0 to 5 all 03
0 to 5 all 04

Document revision level


The document corresponds to the version/revision level effective at the time of system delivery. Revi-
sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.

Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.

SOMATOM CT02-021.816.09 Page 2 of 6 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
Contents 0-3

Page
1 _______General _______________________________________________________ 1 - 1

Definitions and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1


Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Software not approved by Siemens . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Software kit VA47 for Navigator, IRS and Wizard . . . . . . . . . . . . . . . . . . . .1-3
Backup media . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3

2 _______SW Installation Navigator ________________________________________ 2 - 1

Conditions for starting the SW installation. . . . . . . . . . . . . . . . . . . . . . . .2-1


Required data carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Required data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Service key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
Host name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
System type and serial numbers
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Configuration of CD-R Write mode . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Enabled Autoreports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Backup of Static Routes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Starting Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Installation running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Restore Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Restore 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Restore 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Initial setup of configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 10
Completion of configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Import Reference Images . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
Optional Configuration of CD-R Single Session Mode . . . . . . . . . . . . . . 2 - 11
Set Auto Filming on the Wizard (optional) . . . . . . . . . . . . . . . . . . . . . 2 - 12
Restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Siemens Remote Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Define static route . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Remote Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 13
Configuration of customer-specific setup . . . . . . . . . . . . . . . . . . . . . . . 2 - 14
Image text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 14
Minified topogram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 14
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Disable Windows NT Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 16

3 _______SW Installation of IRS P10 _______________________________________ 3 - 1

Installation of IRS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Final steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
No software reinstallation on the Navigator . . . . . . . . . . . . . . . . . . . . . 3 - 2

Siemens AG Page 3 of 4 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
0-4 Contents

Software also reinstalled on the Navigator . . . . . . . . . . . . . . . . . . . . . 3 - 2

4 ______ SW Installation Wizard __________________________________________4 - 1

Conditions for starting the SW installation . . . . . . . . . . . . . . . . . . . . . . . 4-1


Required data carrier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Required data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Service key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Host name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Configuration of CD-R Write mode . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Enabled Autoreports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Backup of Static Routes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Starting Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Installation running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Restore Steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Restore 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Restore 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Completion of configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 10
Check/enable Navigator-Wizard connection . . . . . . . . . . . . . . . . . . . .4 - 10
Check Configuration - Database/Satellite Console . . . . . . . . . . . . . . . . .4 - 11
Regional Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 12
Optional Configuration of CD-R Single Session Mode . . . . . . . . . . . . . . .4 - 12
Restart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 12
Configuration of customer-specific setup. . . . . . . . . . . . . . . . . . . . . . . .4 - 13
Image text. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 13
Minified topogram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 13
Siemens Remote Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 14
Define static route . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 14
Remote Access . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 14
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 15
Disable Windows NT Access . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 16

5 ______ Changes to previous version _____________________________________5 - 1

SOMATOM CT02-021.816.09 Page 4 of 4 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
General 1 1-1

Definitions and Abbreviations 1

Abbreviation Meaning
ACR-NEMA American College of Radiology - National Electrical
Manufacturers Association
BIOS Basic Input Output System, lowest PC configuration settings
AET Application Entity Title - Name of DICOM service participating in the
communication either as user (SCU) or provider (SCP)
CD Compact Disc
CDR Compact Disc Recordable
CT Computed Tomography
DICOM Digital Image Communication in Medicine (ACR-NEMA standard)
DHCP Dynamic Host Configuration Protocol - procedure for assigning
numeric IP address to nodes in which blocks of addresses are leased
to clients on a first-come, first-serve basis. Typically, DHCP is com-
bined with WINS, meaning an IP address is both dynamically
assigned and automatically associated with a domain name. This can
be combined with DNS, resulting in a dynamically assigned numeric
IP address that is automatically associated with both NetBIOS and
Internet names.
DNS Domain Name Services - address resolution procedure that associ-
ates numeric IP addresses with Internet-style names
FTP File Transfer Protocol - a client-server application that transfers files
from one computer to another
Gateway A display that controls traffic between two networks using factors
above the network layer
ICS = Navigator Image Control System - control computer for the entire CT scanner,
with interface to the acquisition system
IES = Wizard Imaging Evaluation System
IMS Imaging System, consisting of ICS, IRS and console
IRS Image Reconstruction System - reconstruction computer
LAN Local Area Network (here, always Ethernet)
LMHOSTS Address resolution procedure which statically associates numeric IP
addresses with NetBIOS names. The LMHOSTS file is checked to
see if the name is being addressed in this file.
MOD Magneto-Optical Disk
NetBIOS Network Basic Input Output System - a collection of network services
and programming interfaces. NetBIOS services run on network proto-
cols like TCP/IP.
NIC Network Interface Card - an expansion card that provides a physical
connection to the network

Siemens AG Page 1 of 4 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
1-2 General

Abbreviation Meaning
NT "New Technology" Operating System - trademark of Microsoft, Inc.
OI Operating Instructions
PB Patient Browser
PPP Point-to-Point Protocol - a data link and network protocol designed for
use over serial connections (such as dial-up modems)
RAS Remote Access Services - a Windows NT service that supports dial-
in and dial-out connections using modems and ISDN.
Router A multiport computer or piece of dedicated hardware that examines
network layer addresses of packets on a network and retransmits
them on a different network if necessary
SCU Service Class User
SCP Service Class Provider
TCP/IP Transport Control Protocol/Internet Protocol - the most widely used
standard for networking
UI User Interface
USC Uptime Service Center
WINS Windows Internet Name Service - address resolution procedure that
dynamically associates numeric IP addresses with NetBIOS names.
It is possible to combine DNS and WINS.
WIP Work in Progress

SOMATOM CT02-021.816.09 Page 2 of 4 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
General 1-3

Overview 1

ICS + IRS = Navigator


In standard configuration, the ICS is the main console for interchangeable scanning and
imaging functions. The IRS is equipped with four computers as well as switchable key-
board, mouse and monitor.
IES = Wizard
The IES console may be installed in parallel to the main operator ICS console via a
shared database link. The IES is used as a second console for postprocessing applica-
tions.

NOTICE The Wizard (IES) can run only when the ICS is switched on; if the
ICS is switched off, the shared database cannot be accessed.

Software not approved by Siemens 1

Warning Unauthorized or incorrect operation or changes to software can


cause malfunctions in your system. This can result in injuries or
system defects. For this reason, the use or installation of software
not approved by Siemens is strictly prohibited.

Software kit VA47 for Navigator, IRS and Wizard 1

The software installation kit for SOMATOM systems is included in the delivery volume.

Note The CD SOMARIS/5 VA47B (A) includes the software for installa-
tion of both the Navigator (ICS) and the Wizard (IES).

Backup media 1

The standard backup device for Navigator and for the Wizard is the CD-Writer (drive [-R-]
CD-R). Alternatively, the MOD may be used, if available.
• Make sure a separate backup media (e.g. "CD Recordable") is on hand for both the
Navigator and the Wizard (if available).

Siemens AG Page 3 of 4 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
1-4 General
This page intentionally left blank.

SOMATOM CT02-021.816.09 Page 4 of 4 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Navigator 2 2-1

Conditions for starting the SW installation 2

To ensure that all archiving, filming, copying and networking jobs are finished and that the
respective queues are empty:
• Open the patient browser.
• Click Transfer -> Select "Local Job Status" and, if necessary, delete all entries.
• Click Transfer -> Select "Network Job Status" and, if necessary, delete all entries.
• Click Patient -> Select "Film Task Status" and, if necessary, delete all entries.

Required data carrier 2

To perform the SW installation of the Navigator (ICS), the following data carriers are
required:

• CD SOMARIS/5 VA47B (A)


• Backup CD-R

Siemens AG Page 1 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
2-2 SW Installation Navigator

Required data 2

To perform the SW installation of the Navigator (ICS), the following data is required:

Regional Settings 2

• From the syngo user interface, select Options -> Configuration.


• In the configuration panel, double-click the icon Regional Settings.
• Write down the following data:

Regional Settings Language


Time Time style Time separator:
Date Short date style Date separator:
Input Locales Input locales

Service key 2

• From the syngo user interface, select Options -> Local Service.
• Write down the service key.

Service key (level=7)


Service key = 20 characters ---------------------- 14 characters ---------------- --- last 6 characters ---

Host name 2

• In the main menu of the service software, select Configuration.


• Click Next.
• Select Local Host -> TCP/IP LAN.
• Write down the host name.
.

Computer name ICS (host name)

NOTE Naming convention for ICS computer name: ct<serial number>.


Please note that the computer name may not contain a blank
space or an underscore.

SOMATOM CT02-021.816.09 Page 2 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Navigator 2-3
System type and serial numbers
2

Name Material No. Serial No.


SOMATOM Esprit 38 15 490 34_ _ _
SOMATOM Esprit+ 38 10 137 35_ _ _
SOMATOM Balance 38 10 152 38_ _ _
SOMATOM Emotion 38 10 160 40_ _ _
42_ _ _
SOMATOM Emotion Duo 38 12 950 36_ _ _
33_ _ _
SOMATOM Emotion (2003) 38 04 890 46_ _ _
SOMATOM Emotion Duo (2003) 38 13 933 365_ _
Volume Zoom 48 06 118 24_ _ _
Volume Access 71 14 023 26_ _ _
Sensation 4 70 05 502 28_ _ _

Configuration of CD-R Write mode 2

The setting for the optional single session mode for CD-R burning is not contained in the
backup. The single session mode has to be configured explicitly after each SW installa-
tion; otherwise, the multi-session mode is configured by default.
Check the entry in the Windows NT registry for the current settting.
• From the <Start> button menu, select Run and type in regedit.
• In the registry path select
HKEY_LOCAL_MACHINE\Software\Siemens\MedCom\Config\Modality\
archnet_common\CDR_MULTISESSION_FLAG\V1
Check the parameter V1; a value = 1 indicates a multi-session mode.
If the value V1 = 0, the single session mode is set.

CD-R Write mode Multi-session mode (standard)


Single session mode (optional)

Enabled Autoreports 2

- From the main menu of the service software, select AutoReport and check for active
reports
(these reports have to be enabled again after SW Installation).
predefined

1 Mon_HQ 4
specific
site-

2 Perf_Data 5
3 TripReport 6
CustomerComplaint

Siemens AG Page 3 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
2-4 SW Installation Navigator
Backup of Static Routes 2

Static routes are not saved with the backup. As a result, static routes have to be noted
before software installation.

• In the main menu of the service software, select Utilities.


• Under Source select Escape to OS.
• In Command select "NT Command Interpreter".
• Under Parameters enter route print -p and press Go.

SRS target server

Fig. 1 Route print

• Note Persistent Route to the SRS target server (Fig.1). If no entry is found, no static
route was necessary for SRS.
(Europe: 194.138.39.0 / America: 129.73.116.0 / Asia: 194.138.243.0).

Network Address Netmask Gateway Address

SRS target server Netmask target server Gateway (Router) to SRS

SOMATOM CT02-021.816.09 Page 4 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Navigator 2-5

Backup 2

Note This step is only necessary if no up-to-date backup is available.

Note Important: WindowsNT Explorer has to be closed during the


backup procedure!

Note In case you have to install the SW at the Wizard as well, you may
start the SW installation during the backup procedure (refer to
pages 4-1 to 4-3 in chapter 4).

• Start the service software and select the Backup & Restore function in the main menu.
• Insert the backup CD-R for the Navigator.
• From Command, select Backup.
• From Drives, select the corresponding drive.
• From Packages, select CT-Scanner and click Go.

Command Drives
Backup [-R-] CD-R

Packages
CT-Scanner-

➪ Wait until the backup files are stored successfully (up to 15 min.).

• From Packages, select SW-Settings02-PXXX-<host name> and click Go.

Command Drives
Backup [-R-] CD-R

Packages
SW-Settings02

➪ Wait until the backup files are stored successfully (up to 3 min.).

Note Check the backup files by copying the files from the backup
media (e.g. CD) to hard disk. If this works without error, the files
are saved correctly.

• At "Ready," remove the backup media.


• Click Home and terminate the service software by closing the browser window.

Siemens AG Page 5 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
2-6 SW Installation Navigator
• From the Wizard shut down the application and switch off system.
• From the Navigator, select System/End and Shut down the system.
• Insert the CD SOMARIS/5 VA47B (A).
• Switch off the system at the control box.
• Disconnect the network cable of the ICS from the Ethernet switch.

Starting Installation 2

• Switch on the system at the control box (system-on).


➪ The computer boots from the CD-ROM.

Starting Windows 98
......

After a while the following window appears:

Siemens Medical Solutions SOMARIS VA47 installation (Int.NT4.0)


SELECTION MENU INFORMATION WINDOW

Turn off the computer to interrupt

or press any key to continue Temporarily partitioning first hard


drive to 2049MB.
WindowsNT will extend the size up to a
CONTINUE INSTALLATION maximum of 8GB.
Confirm to DESTROY ALL DATA
on the first hard drive

HARDWARE DETECTION IN PROGRESS

Checking first hard drive

• Press <Enter> to continue installation (the ICS is beeping). This reboots the ICS.

SOMATOM CT02-021.816.09 Page 6 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Navigator 2-7

Siemens Medical Solutions SOMARIS VA47 installation (Int.NT4.0)


SELECTION MENU INFORMATION WINDOW

Install: < select System Type >

Distribution type: CUSTOMER Content changing according selection!

Customer Machine Name: <host name >

Keep data on other hard drive(s)

Use ARROW-KEYS to navigate


CONTINUE INSTALLATION Use ENTER to change selection
HARDWARE DETECTION RESULTS

Matrox Graphic adapter (G100 or later) found


Ensoniq AudioPCI-97 (ES1371) Soundcard found
Intel 8255X chipset Network adapter found

• Select the system to be installed (Sensation 16, Emotion, etc.).


Use ENTER to change selection.
• Select distribution type: CUSTOMER (default).
• Enter the machine name (host name) by replacing the default entry MedPC.
(e.g.: ct12345) and press Enter.

NOTE Do not use the default name MEDPC.


Note that the computer name must not contain a blank space or
an underscore.
Naming convention for ICS computer name: ct<serial number>
(e.g. ct123456).

• Select Keep data on other hard drive(s) or


select Clear data on second hard drive (deletes patient database).
Use ENTER to change selection.

NOTE Do not use Clear all hard drive(s). This option will format all hard
drives only (removes all software). No reinstallation of SW will be
started.

• With CONTINUE INSTALLATION confirm the start of installation (ICS is beeping if


"Clear data" is selected).
• Press any key.
➪ All necessary files are copied from CD SOMARIS/5 VA47B (A) to the hard drive.

After approximately 5 to 8 minutes the CD SOMARIS/5 VA47B (A) is ejected and


the ICS reboots.
➪ Remove the CD SOMARIS/5 VA47B (A).

Siemens AG Page 7 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
2-8 SW Installation Navigator

Installation running 2

From now on, the ICS installation runs automatically. No user action regarding ICS instal-
lation is required.

Note P10 only!


While the installation is running, you may start the installation of
IRS, if necessary (refer to chapter 3).

Note You may now continue with the software installation on the Wiz-
ard, if necessary (refer to page 4-4).

Completion of ICS/IRS installation (approx. 75 minutes) is indicated by a pop-up window


with the message Please enter license.

Note You can reduce the installation time to approx. 60 min. by cancel-
ing the memory check by pressing ESC during startup.

• Click ’Click here to start the Local Service Configuration’ to start the service software
(complete service license required).

SOMATOM CT02-021.816.09 Page 8 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Navigator 2-9

Restore Steps 2

1. Restore 1: restore all groups of SW settings except User Settings and Service/
Database/DBOffset and all groups of CT scanner settings except IRS.
2. Reconnect the network cable and restart the Navigator.
3. Restore 2: restore the IRS tables and the User Settings.

Notice After Restore 1, the ICS and IRS synchronizes and the IRS
"default tables" are automatically transferred from the ICS to the
IRS. With Restore 2, the "site-specific tune-up tables" are
restored to the IRS.
If you install the IRS software, wait until the IRS installation fin-
ishes and ensure that the IRS is ready before performing the sub-
sequent steps.

Restore 1 2

• In the main menu of the service software, select the Backup & Restore function.
• Insert the backup CD Navigator.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select SW-Settings02.

(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R UserSettings
Service\LocalHost\Users
Archive Service\TCPIP\LAN-Settings
SW-Settings02 date&time.ar Service\TCPIP\RAS-Settings
Service\Database/SateliteConsole
Service\Database\DBOffset
Service\Applications\kWorklistResults

• From Groups, select all except User settings and Service\Database\DBOffset and
click Go.
➪ Wait until "Ready".

• From Archive, select CT-Scanner.


(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R Cardio
Tube
Archive Service
CT-Scanner date&time.ar IRS
General
Firmware
Bolus

• From Groups, select all except IRS and click Go.


➪ Wait until "Ready".

Siemens AG Page 9 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
2 - 10 SW Installation Navigator
• Click Home.
• Switch on the gantry, if not on already.
• Click the Start button in the task bar.
• Perform Shutdown -->Restart.
• Connect the network cable of the ICS to the Ethernet switch.

Restore 2 2

• When started up, cancel the checkup platform and access the service software.
• In the main menu of the service software, select the Backup & Restore function.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select SW-Settings02.

(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R User Settings
Service\LocalHost\Users
Archive Service\TCPIP\LAN-Settings
SW-Settings02-PXXX date&time.ar Service\TCPIP\RAS-Settings
Service\Database/SateliteConsole
Service\Database\DBOffset
Service\Applications\kWorklistResults

• From Groups, select User settings and click Go.


➪ Wait until "Ready".

• From Archive, select CT-Scanner.


(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R Cardio
Tube
Archive Service
CT-Scanner-PXXX date&time.ar IRS
General
Firmware
Bolus

• From Groups, select IRS and click Go.


• At "Ready", remove the backup CD Navigator.
• Click Home but do not close the service platform.

Initial setup of configuration 2

• In the main menu of the service software, select the Configuration function.
➪ Click Next until you reach the end of all entries.
When you reach the last page, click Finish.

SOMATOM CT02-021.816.09 Page 10 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Navigator 2 - 11

Completion of configuration 2

Import Reference Images 2

Mandatory, if "Clear data on second hard drive" was selected in the installation mask.
• From the patient browser, select Transfer -> Import from Offline.
• From the drop-down list, select the path c:\Somaris\service\ReferenceImages.
• Select all images and click OK.
➪ After the import, the images are available in the patient browser.

Regional Settings 2

• From the syngo user interface, select Options -> Configuration.


• In the configuration panel, double-click the icon Regional Settings.
• Change regional settings as described on page 2-2.

Optional Configuration of CD-R Single Session Mode 2

The multi-session mode is the standard default mode for the CD-R writer.
• If the single session mode is required by the customer, double-click the file
c:\Somaris\utils\single_session.reg with the Windows NT Explorer.
This setting is not contained in the backup and, as a result, has to be repeated after each
SW installation.

Fig. 2 Single Session mode for CD-R burning

• To re-configure from single session to the standard multi-session mode, double-click the
file c:\Somaris\utils\multi_session.reg with Windows NT Explorer.

Siemens AG Page 11 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
2 - 12 SW Installation Navigator
Set Auto Filming on the Wizard (optional) 2

Note Auto filming has to be enabled on the Navigator, if the customer


wants auto filming on the first or second Wizard.
Please ask the customer!
Switch on the IES and wait for the complete start-up of the
Wizard.

• From the Navigator press the <CTRL> and <ESC> buttons (Windows NT Access has to
be enabled).
• Click the <Start> button in the task bar and select Programs / CommandPrompt.
• Type in perl c:\Somaris\utils\determinewizard.pl.
➪ Choose the corresponding Wizard according to the host name.
Type in 1 or 2, then press Return.

• Press Return and close the command prompt window.

Restart 2

• Perform System/End -> Restart the System


(Restart is necessary to enable options and make the option tab cards available in the
customer UI (user interface)).

SOMATOM CT02-021.816.09 Page 12 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Navigator 2 - 13

Siemens Remote Services 2

Notice System Management and SOMATOM Life have to be installed/con-


figured.
For detailed information on how to install and configure the
"Siemens Remote Services," refer to the document
Installation of Siemens Remote Services (SRS),
print no. CT00-000.816.01.xx.02.

Define static route 2

In case a static route was used, configure it as described in chapter 2.


• Select Utilities from Service Software Home menu.
• Select Escape to OS under Source in the navigation frame.
• Select NT Command Interpreter under Command.
• Type in the following line using the parameters (in < >) in your notes:

route add <SRS target server> mask <Netmask of target server> <Gateway to SRS> -p

Remote Access 2

• Set Remote Access as it was prior to the installation.

Siemens AG Page 13 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
2 - 14 SW Installation Navigator

Configuration of customer-specific setup 2

Image text 2

• Open the Windows NT Explorer.


• Double click the file C:\SOMARIS\Service\Config\Config_ImageTexr.bat
➪ in the displayed command window, select the appropriate configuration and type
in the correct number.

Minified topogram 2

Notice It is necessary only to instruct the customer to switch the minified


topogram on or off acording to the user manual.

SOMATOM CT02-021.816.09 Page 14 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Navigator 2 - 15

Backup 2

• Start the service software and select the Backup & Restore function in the main menu.
• Insert the backup CD-R for the Navigator.
• From Command, select Backup.
• From Drives, select the CD drive, e.g. [-R-] CD-R
• From Packages, select CT-Scanner and click Go.

Command Drives
Backup [-R-] CD-R

Packages
CT-Scanner

➪ Wait until the backup files are stored successfully.


• From Packages, select SW-Settings02 and click Go.

Command Drives
Backup [-R-] CD-R

Packages
SW-Settings02

➪ Wait until the backup files are stored successfully.


• At "Ready," remove the backup media.
• Click Home and terminate the service software by closing the browser window.

Siemens AG Page 15 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
2 - 16 SW Installation Navigator

Disable Windows NT Access 2

This is to disable/enable Windows NT access for the customer. Administrator privileges


are required; for this reason, use the command shell of the service software.
• From the Service Software Home Menu, select Utilities.
• Under Source, select Escape to OS.

Fig. 3 Disable Windows NT Access

• Under Parameters type in regedit c:\Somaris\Tools\bin\disable_windows_key.ini.


• Activate with Return or GO.
• Shift the Service Software UI to confirm the following notice window.

Fig. 4 Confirmation for Disable Windows NT Access

• Click OK.
• Reboot is necessary to make the change effective.

(To enable Windows NT, type in


regedit c:\Somaris\Tools\bin\enable_windows_key.ini)

SOMATOM CT02-021.816.09 Page 16 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation of IRS P10 3 3-1

Installation of IRS 3

It is not possible to keep raw data files. Make sure that raw data files are archived or
no longer necessary.

Notice Please use the correct IRS SW Installation CD:

If the IRS computer is a Primergy C150 (3814568), use the installation


CD "CD_Som5_IRS-2_VA47A."
If the IRS computer is a Celsius R610 (7733418), use the installation
CD "Som5 IRS VA47C_SP1"
For all other IRS computers, use the installation CD "CD_Som5_IRS-
1_VA47A"

• Insert the Som5 IRS CD (delivered with the computer)


Leave the CD tray open to prevent blocking automatic CD ejection.
• Reboot the IRS.
• When ejected, remove the Som5 IRS CD from the CD drive.
• Power Off the IRS and Power On again.
➪ The IRS installation continues and will be finished within approx. 20 min.

Siemens AG Page 1 of 2 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
3-2 SW Installation of IRS P10

Final steps 3

No software reinstallation on the Navigator 3

Notice During the IRS software installation, the IRS "default tables" are
deleted. The IRS "default tables" have to be transferred manually
to the IRS. As a result, the batch file IrsTablesInstall.bat has to be
executed. Afterwards, the "site-specific tune-up tables" have to
be restored.

• At the Navigator (SOMARIS has been started up) open the Windows NT Explorer:
Click the Start button -> select Programs -> Windows NT Explorer.
• Select C:\Somaris\utils via the Explorer.
• Double-click the file IrsTablesInstall.bat.
➪ The "default tables" are transferred to the IRS.

• In the Service Software Main Menu, select the function Backup & Restore.
• Insert the backup CD Navigator.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select CT-Scanner.
(This is only an example!)

Command Drives Groups


Restore [-R-] CD-R Cardio
Tube
Archive Service
CT-Scanner date&time.ar IRS
General
Firmware
Bolus

• From Groups, select IRS and click Go.


➪ The "site-specific tune-up tables" are transferred to the IRS.

• At "Ready," remove the backup CD Navigator and restart the complete system.

Software also reinstalled on the Navigator 3

Notice After Restore 1 (refer to chapter 2), the ICS and IRS synchronizes
and the IRS "default tables" are transferred automatically from
the ICS to the IRS. With Restore 2 (refer to chapter 2) the "site-
specific tune-up tables" are restored to the IRS. No further steps
are necessary.

SOMATOM CT02-021.816.09 Page 2 of 2 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Wizard 4 4-1

Note These instructions are valid for both the first and second Wizards.

Conditions for starting the SW installation 4

To ensure that all archiving, filming, copying and networking jobs are finished and that the
respective queues are empty:
• Open the patient browser.
• Click Transfer -> Select "Local Job Status" and, if necessary, delete all entries.
• Click Transfer -> Select "Network Job Status" and, if necessary, delete all entries.
• Click Patient -> Select "Film Task Status" and, if necessary, delete all entries.

Required data carrier 4

To perform the SW installation of the Wizard, the following data carriers are required:

• CD SOMARIS/5 VA47B (A)


Backup CD-R

Siemens AG Page 1 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
4-2 SW Installation Wizard

Required data 4

To perform the SW installation of the Wizard, the following data is required:

Regional Settings 4

• From the syngo user interface, select -> Options -> Configuration.
• In the configuration panel, double-click the icon Regional Settings.
• Write down the following data:

Regional Settings Language


Time Time style Time separator:
Date Short date style Date separator:
Input Locales Input locales

Service key 4

• From the syngo user interface, select -> Options -> Local Service
• Write down the service key

Service key (level=7)


Service key = 20 characters -------------------- 14 characters ------------------ -- last 6 characters --

Host name 4

• In the service software main menu, select Configuration.


• Click Next.
• Select Local Host -> TCP/IP LAN.
• Write down the host name.

Computer name IES (host name)

NOTE Naming convention for IES computer name: ctw<serial number>


Please note that the computer name may not contain blanks or
underscores.

SOMATOM CT02-021.816.09 Page 2 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Wizard 4-3
Configuration of CD-R Write mode 4

The setting for the optional single session mode for CD-R burning is not contained in the
backup. The single session mode has to be configured explicitly after each SW installa-
tion; otherwise, the multi-session mode is configured by default.
Check the entry in the Windows NT registry for the current setting.
• From the <Start> button menu, select Run and type in regedit.
• In the registry path, select
HKEY_LOCAL_MACHINE\Software\Siemens\MedCom\Config\Modality\
archnet_common\CDR_MULTISESSION_FLAG\V1
Check the parameter V1; the value =1 indicates multi-session mode. If the V1 value = 0,
the single session mode is set.

CD-R Write mode Multi-session mode (standard)


Single session mode (optional)

Enabled Autoreports 4

- from the service software main menu select AutoReport and check for active reports
(these reports have to be enabled again after SW Installation)
predefined

1 "Mon_HQ" 4
specific
site-

2 " 5
3 6

Siemens AG Page 3 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
4-4 SW Installation Wizard
Backup of Static Routes 4

Static routes are not saved with the backup. As a result, static routes have to be noted
before software installation.

• In the main menu of the service software, select Utilities.


• Under Source select Escape to OS.
• In Command select "NT Command Interpreter".
• Under Parameters enter route print -p and press Go.

SRS target server

Fig. 1 Route print

• Note Persistent Route to the SRS target server (Fig.1). If no entry is found, no static
route was necessary for SRS.
(Europe: 194.138.39.0 / America: 129.73.116.0 / Asia: 194.138.243.0).

Network Address Netmask Gateway Address

SRS target server Netmask target server Gateway (Router) to SRS

SOMATOM CT02-021.816.09 Page 4 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Wizard 4-5

Backup 4

Note This step is only necessary if no up-to-date backup is available.

Note Important: the WindowsNT Explorer has to be closed during the


backup procedure!

• Start the service software and select the Backup & Restore function.
• Insert the backup media (e.g. CD) for the Wizard.
• From Command, select Backup.
• From Drives, select the corresponding drive.
• From Packages, select CT-Settings and click Go.

Command Drives
Backup [-R-] CD-R

Packages
CT-Settings

➪ Wait until the backup files are stored successfully (up to 5 min.).

• From Packages, select SW-Settings02-IES-<host name> and click Go.

Command Drives
Backup [-R-] CD-R

Packages
SW-Settings02

➪ Wait until backup files are stored successfully (up to 3min.).

➪ .

Note Check the backup files by copying the files from the backup
media (e.g. CD) to hard disk. If this works without error, the files
are saved correctly.

• At “Ready”, remove the backup media.

Siemens AG Page 5 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
4-6 SW Installation Wizard

Starting Installation 4

• Click Home and terminate the service software by closing the browser window.
• Select System/End and Shut down the system.
• Insert the CD SOMARIS/5 VA47B (A).
• Power off IES.
• Disconnect network cable of IES from the Ethernet switch.
• Power on IES.

➪ The computer boots from the CD-ROM.

Starting Windows 98
......

After a while the following window appears:

Siemens Medical Solutions SOMARIS VA47 installation (Int.NT4.0)


SELECTION MENU INFORMATION WINDOW

Turn off the computer to interrupt

or press any key to continue Temporarily partitioning first hard


drive to 2049MB.
WindowsNT will extend the size up to a
maximum of 8GB.
CONTINUE INSTALLATION
Confirm to DESTROY ALL DATA
on the first hard drive

HARDWARE DETECTION IN PROGRESS

Checking first hard drive

• Press <Enter> to continue installation (the IES is beeping). This restarts the IES.

SOMATOM CT02-021.816.09 Page 6 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Wizard 4-7

Siemens Medical Solutions SOMARIS VA47 installation (Int.NT4.0)


SELECTION MENU INFORMATION WINDOW

Install: Wizard

Distribution type: CUSTOMER Content changing according selection!

Customer Machine Name: <host name >

Keep data on other hard drive(s)

Use ARROW-KEYS to navigate


CONTINUE INSTALLATION Use ENTER to change selection
HARDWARE DETECTION RESULTS

Matrox Graphic adapter (G100 or later) found


Ensoniq AudioPCI-97 (ES1371) Soundcard found
Intel 8255X chipset Network adapter found

• Select Install: Wizard.


Use ENTER to change selection.
• Select distribution type CUSTOMER (default).
• Enter the machine name (host name) by replacing the default entry MedPC.
(e.g.: ctw12345) and press Enter.

NOTE Do not use the default name MEDPC.


Please note that the computer name may not contain blank or
underscores.
Naming convention for IES computer name: ctw<serial number>
(e.g. ctw12345, wizard connected to ct12345).

• Keep data on other hard drive(s) is default and may not be modified (only one
hard drive available)

NOTE Do not use Clear all hard drive(s). This option will format all hard
drives only (removes all software). No reinstallation of SW will be
started.

• Via CONTINUE INSTALLATION, confirm the start of installation.


• Press any key.
➪ All necessary files are copied from CD SOMARIS/5 VA47B (A) to the hard drive.

After approximately 15 minutes, the CD SOMARIS/5 VA47B (A) is ejected and


the IES reboots.
➪ Remove the CD SOMARIS/5 VA47B (A).

Siemens AG Page 7 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
4-8 SW Installation Wizard

Installation running 4

From now on, the IES installation runs automatically. No user action regarding IES instal-
lation is required.

Completion of IES installation (approx. 55 minutes) is indicated by a pop-up window with


the message Please enter license.

Note You can reduce the installation time to approx. 45 minutes by


pressing ESC during startup to cancel the memory check.

Note Continue only when the Navigator is fully installed and started up.

• Click ’Click here to start the Local Service Configuration’ to start the service software
(complete service license required).

SOMATOM CT02-021.816.09 Page 8 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Wizard 4-9

Restore Steps 4

Note The subsequent steps require the Navigator to be already


installed and running.
• Switch on the Navigator.

1. Restore 1: restore all groups of SW-settings except User Settings and all groups
of CT settings.
2. Reconnect the network cable and restart the system.
3. Restore 2: restore the User Settings.

Restore 1 4

• In the service software main menu, select the Backup & Restore function.
• Insert the backup media for the Wizard.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select SW-Settings02.
• From Groups, select all except User Settings and click Go.
➪ Wait until "Ready" (about 5 min.).
• From Archive, select CT-Settings.
• From Groups, select all and click Go.
➪ Wait until "Ready" (about 2 min.)
• Click Home.

Restore 2 4

• Click the Start button in the task bar.


• Perform Shutdown -->System restart.

Note The reboot takes a while because the database is being created.

• Connect network cable of IES to the Ethernet switch.


• When started up, access the service software.
• In the service software main menu, select the Backup & Restore function.
• From Command, select Restore.
• From Drives, select the corresponding drive.
• From Archive, select SW-Settings02-IES-<host name>.
• From Groups, select User Settings and click Go.
➪ Wait until "Ready" (about 5 min.).

Siemens AG Page 9 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
4 - 10 SW Installation Wizard
• Click Home and terminate the service software by closing the browser window.
• Perform System/End -> Restart the System.

Completion of configuration 4

Check/enable Navigator-Wizard connection 4

• From the service software main menu, select Configuration.


• Activate the checkbox "Main Console with Satellites or Satellite Console" in the list
of options, if not activated already.

Fig. 2 Configuration - list of system options

• Click Next.

SOMATOM CT02-021.816.09 Page 10 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Wizard 4 - 11
Check Configuration - Database/Satellite Console 4

Fig. 3 Satellite Console 1

• Verify the presence of the Navigator (e.g. CT24600) and its IP.
• Select the console type as Satellite Console with <Navigator> remote Main Console
Host.

Fig. 4 Satellite Console 2

• Check/correct database offset of local DB’s.

Note - 1st Wizard offset = 500


- 2nd Wizard (optional) offset = 1000

• Click Save, Finish, and then Home.


• The system now reboots.

Siemens AG Page 11 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
4 - 12 SW Installation Wizard
Regional Settings 4

• From the syngo user interface, select -> Options -> Configuration.
• In the configuration panel, double-click the icon Regional Settings.
• Change regional settings as described on page 4-2.

Optional Configuration of CD-R Single Session Mode 4

The multi-session mode is the standard default mode for the CD-R writer.
• If the single session mode is required by the customer, double-click the file
c:\Somaris\utils\single_session.reg with the Windows NT Explorer.
This setting is not contained in the backup and has, as a result, to be repeated after each
SW installation.

Fig. 5 Single Session mode for CD-R burning

• To re-configure from single session to the standard multi-session mode, double-click the
file c:\Somaris\utils\multi_session.reg with the Windows NT Explorer.

Restart 4

• Perform System/End -> Restart the System


(Restart is necessary to enable options and make the option tab cards available in the
customer UI (user interface)).

SOMATOM CT02-021.816.09 Page 12 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Wizard 4 - 13

Configuration of customer-specific setup 4

Image text 4

• Open the Windows NT Explorer.


• Double click the file C:\SOMARIS\Service\Config\Config_ImageTexr.bat
➪ in the displayed command window, select the appropriate configuration and type
in the correct number.

Minified topogram 4

Notice It is necessary only to instruct the customer to switch the minified


topogramon or off according to the user manual.

Siemens AG Page 13 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
4 - 14 SW Installation Wizard

Siemens Remote Services 4

Notice System Management and SOMATOM Life have to be installed/con-


figured.
For detailed information on how to install and configure the
"Siemens Remote Services," refer to the document
Installation of Siemens Remote Services (SRS),
print no. CT00-000.816.01.xx.02.

Define static route 4

In case a static route was used, configure it as described in chapter 2 (2.2.1.4).


• Select Utilities from Service Software Home menu.
• Select Escape to OS in the navigation frame under Source.
• Select NT Command Interpreter under Command.
• Type in the following line using the parameters (in < >) in your notes:

route add <SRS target server> mask <Netmask of target server> <Gateway to SRS> -p

Remote Access 4

• Set Remote Access as it was prior to installation.

SOMATOM CT02-021.816.09 Page 14 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
SW Installation Wizard 4 - 15
Backup 4

• Start the service software and select the Backup & Restore function in the main menu.
• Insert the backup CD-R for the Wizard.
• From Command, select Backup.
• From Drives, select the corresponding drive.
• From Packages, select CT-Settings and click Go.

Command Drives
Backup [-R-] CD-R

Packages
CT-Settings

➪ Wait until the backup files are stored successfully.

• From Packages, select SW-Settings02 and click Go.

Command Drives
Backup [-R-] CD-R

Packages
SW-Settings02

➪ Wait until backup files are stored successfully.

• At “Ready”, remove the backup media.


• Click Home and terminate the service software by closing the browser window.

Siemens AG Page 15 of 16 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
4 - 16 SW Installation Wizard
Disable Windows NT Access 4

This is to disable/enable Windows NT access for the customer. Administrator privileges


are required; for this reason, use the command shell of the service software.
• From the Service Software Home Menu, select Utilities.
• Under Source, select Escape to OS.

Fig. 6 Disable Windows NT Access

• Under Parameters type in regedit c:\Somaris\Tools\bin\disable_windows_key.ini


• Activate with Return or GO.
• Shift the Service Software UI to confirm the following notice window.

Fig. 7 Confirmation for Disable Windows NT Access

• Click OK.
• Reboot is necessary to make the change effective.

(To enable Windows NT, type in


regedit c:\Somaris\Tools\bin\enable_windows_key.ini)

SOMATOM CT02-021.816.09 Page 16 of 16 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions
Changes to previous version 5 5-1

Chapter Change Reason


all name of IRS CD changed 128324

Siemens AG Page 1 of 2 CT02-021.816.09 SOMATOM


Medical Solutions CSPS22 Rev. 04 03.04
5-2 Changes to previous version
This page intentionally left blank.

SOMATOM CT02-021.816.09 Page 2 of 2 Siemens AG


Rev. 04 03.04 CSPS22 Medical Solutions