Download as pdf or txt
Download as pdf or txt
You are on page 1of 13

An American National Standard

Designation: D 1016 – 05e2

Standard Test Method for


Purity of Hydrocarbons from Freezing Points1
This standard is issued under the fixed designation D 1016; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

e1 NOTE—Eq 54 in 25.3 was corrected editorially in June 2006.


e2 NOTE—Eq 3 and Eq 39 were corrected editorially in September 2007.

1. Scope* the concentration is known for each most probable impurity in the given
substance.
1.1 This test method covers the sampling and determination
of purity of essentially pure compounds for which the freezing 2. Referenced Documents
points for zero impurity and cryoscopic constants are given.2 2.1 ASTM Standards: 3
The compounds to which the test method is applicable are:
D 1015 Test Method for Freezing Points of High-Purity
(Warning—Extremely flammable liquids and liquefied gases.)
Hydrocarbons
n-butane 1,3-butadiene
isobutane isoprene(2-methyl-1,3-butadiene) 3. Summary of Test Method
n-pentane benzene
isopentane toluene (methylbenzene) 3.1 After measurement of the freezing point of the actual
n-hexane ethylbenzene
n-heptane o-xylene (1,2-dimethylbenzene)
sample, purity can be calculated from the value of the
n-octane m-xylene (1,3-dimethylbenzene) determined freezing point and the values given for the freezing
2,2,4-trimethylpentane p-xylene (1,4-dimethylbenzene) point for zero impurity and for the applicable cryoscopic
methylcyclohexane styrene (ethenylbenzene)
isobutene
constant or constants.4
3.2 For the equilibrium between an infinitesimal amount of
1.2 The values stated in SI units are to be regarded as the the crystalline phase of the major component and a liquid phase
standard. The values in parentheses are for information only. of the major component and one or more other components, the
1.3 This standard does not purport to address all of the thermodynamic relation between the temperature of equilib-
safety concerns, if any, associated with its use. It is the rium and the composition of the liquid phase is expressed by
responsibility of the user of this standard to establish appro- the equation:5
priate safety and health practices and determine the applica-
21n N1 5 21n ~1 2 N2! 5 A~t f 0 2 tf!@1 1 B~t f 0 2 tf! 1 ...] (1)
bility of regulatory limitations prior to use. For specific hazard
statements, see Sections 1, 6, 8, and 10-26. where:
NOTE 1—This test method covers systems in which the impurities form N1 = mole fraction of the major component,
with the major component a substantially ideal or sufficiently dilute
solution, and also systems which deviate from the ideal laws, provided
that, in the latter case, the lowering of the freezing point as a function of
3
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
1
This test method is under the jurisdiction of ASTM Committee D02 on the ASTM website.
4
Petroleum Products and Lubricants and is the direct responsibility of Subcommittee For a more complete discussion of this test method, see Glasgow, A. R., Jr.,
D02.04.0D on Physical and Chemical Methods. Streiff, A. J., and Rossini, F. D., “Determination of the Purity of Hydrocarbons by
Current edition approved Nov. 1, 2005. Published November 2005. Originally Measurement of Freezing Points,” Journal of Research, JRNBA, National Institute
approved in 1949. Last previous edition approved in 2004 as D 1016–99(2004)e1. of Standards and Technology, Vol 35, No. 6, 1945, p. 355.
2 5
Numerical constants in this test method were taken from the most recently For details, see Taylor, W. J., and Rossini, F. D., “Theoretical Analysis of
published data appearing in “Tables of Physical and Thermodynamic Properties of Time-Temperature Freezing and Melting Curves as Applied to Hydrocarbons,”
Hydrocarbons and Related Compounds,” or ASTM DS 4A, Physical Constants of Journal of Research, JRNBA, Nat. Bureau Standards, Vol 32, No. 5, 1944, p. 197;
Hydrocarbons C 1 to C10, or both, prepared by the American Petroleum Institute, also Lewis, G. N., and Randall, M., “Thermodynamics and the Free Energy of
Research Project 44. Chemical Substances,” 1923, pp. 237, 238, McGraw-Hill Book Co., New York, NY.

*A Summary of Changes section appears at the end of this standard.


Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.

1
D 1016 – 05e2

N2 = (1 − N1) = sum of the mole fractions of all the other inhaled. High concentrations can cause unconsciousness or
components, death. Contact can cause skin irritation and dermatitis. Use
tf = freezing point, in degrees Celsius, of the given refrigerant bath only with adequate ventilation!)
substance (in which the mole fraction of the major 6.2 Liquid Nitrogen or Liquid Air—(Warning—Extremely
component is N1), defined as the temperature at cold. Liberates gas which can cause suffocation. Contact with
which an infinitesimal amount of crystals of the skin causes burns or freezing, or both. Vapors can react
major component is in thermodynamic equilibrium violently with hot magnesium or aluminum alloys.) For use as
with the liquid phase (see Note 3 of Test Method a refrigerant. If obtainable, liquid nitrogen is preferable be-
D 1015), cause of its safety.
tf0 = freezing point for zero impurity, in degrees Celsius, 6.2.1 Use liquid nitrogen refrigerant only with adequate
for the major component when pure, that is, when ventilation. If liquid air is used as a refrigerant, it is imperative
N1 = 1 or N2 = 0, that any glass vessel containing hydrocarbon or other combus-
A = first or main cryoscopic constant, in mole fraction tible compound and immersed in liquid air be protected with a
per degree, and suitable metal shield. The mixing of a hydrocarbon or other
B = secondary cryoscopic constant, in mole fraction per combustible compound with liquid air due to the breaking of a
degree. glass container would almost certainly result in a violent
Neglecting the higher terms not written in the brackets, Eq 1 explosion. If liquid nitrogen is used as a refrigerant, no
can be transformed to the equation: hydrocarbon sample should ever be permitted to cool below the
log10 P 5 2.00000 2 ~A / 2.3026!~tf 0 2 tf!@1 1 B~t f 0 2 tf!# (2) condensation temperature of oxygen (−183°C at atm). This
would not be likely to occur in normal operation, but might
where: occur if the apparatus were left unattended for some time.
P = purity of the given substance in terms of mole percent
of the major component. 7. Procedure
7.1 Measure the freezing point as described in Test Method
4. Significance and Use D 1015, using the modifications and constants given in Sec-
4.1 The experimental procedures and physical constants tions 8-26 of this test method for the specific compounds being
provided by this test method, when used in conjunction with examined.
Test Method D 1015, allow the determination of the purity of
NOTE 2—The estimated uncertainty in the calculated value of the purity
the material under test. A knowledge of the purity of these
as referred to in Sections 8-26 is not equivalent to the precision defined in
hydrocarbons is often needed to help control their manufacture RR: D02-1007.
and to determine their suitability for use as reagent chemicals
or for conversion to other chemical intermediates or finished 8. n-Butane6 (Warning—Extremely flammable liquefied
products. gas under pressure. Vapor reduces oxygen available for
breathing.)
5. Apparatus
8.1 Determine the freezing point from freezing curves, with
5.1 Sampling Apparatus, as shown in Fig. 1, for withdraw- the cage stirrer, with a cooling bath of liquid nitrogen (or liquid
ing liquefied gases (for example, 1,3-butadiene) from pressure air), with a cooling rate of 0.3 to 0.8°C/min for the liquid near
storage cylinders. the freezing point, and with crystallization induced immedi-
5.2 Distilling Apparatus, as shown in Fig. 2, for removing ately below the freezing point by means of a cold rod.
small amounts of polymer from low-boiling compounds (for 8.2 The method of obtaining the samples shall be as
example, 1,3-butadiene) by simple distillation at atmospheric follows: Assemble the apparatus for obtaining the sample as
pressure. shown in Fig. 1, but with no lubricant on the ground-glass
5.3 Distilling Apparatus, as shown in Fig. 3, for removing joints and with the valve at the bottom of the cylinder, so that
small amounts of polymer from compounds with boiling points sampling is from the liquid phase. Attach to C an absorption
near room temperature (for example, isoprene) by distillation tube containing anhydrous calcium sulfate or other suitable
at atmospheric pressure. desiccant (except magnesium perchlorate) so that water is not
5.4 Vacuum Distilling Apparatus and Transfer Trap, as introduced into the system (Note 3). Fill the flask F with the
shown in Fig. 4, for removing dissolved air and large amounts carbon dioxide refrigerant to within about 51 mm (2 in.) of the
of polymer from a compound (for example, 1,3-butadiene or top. After about 20 or 30 min, when the system will have
styrene), by repeated freezing and evacuation, followed by cooled sufficiently, remove the absorption tube and begin the
distillation of the compound in vacuum in a closed system. collection of liquid n-butane by opening the valve K and
adjusting the needle valve J so that the sample is collected at
6. Materials
a rate of 1 to 2 mL (liquid)/min in the condensing tube E.
6.1 Carbon Dioxide Refrigerant—Solid carbon dioxide in a
suitable liquid. (Warning—Extremely cold (−78.5°C). Liber- NOTE 3—However, if some water does condense with the hydrocarbon,
ates heavy gas which can cause suffocation. Contact with skin
causes burns or freezing, or both. Vapors can react violently 6
For further details, see Glasgow, A. R., Jr., et al. “Determination of Purity by
with hot magnesium or aluminum alloys.) Acetone is Measurement of Freezing Points of Compounds Involved in the Production of
recommended.(Warning—Extremely flammable. Harmful if Synthetic Rubber,” Analytical Chemistry, ANCHA, Vol 20, 1948, p. 410.

2
D 1016 – 05e2

A—Three-way T stopcock, borosilicate glass (similar to Corning Pyrex No. 7420).


B—Connection to vacuum for purging and for evacuating system CDEGHI.
C—Capillary tube for venting, to which drying tube is also connected.
D—Joint, standard taper, 12/30, borosilicate glass.
E—Condensing tube, borosilicate glass.
F—Dewar flask, 1-qt size, borosilicate glass (similar to American Thermos Bottle Co. No. 8645).
G—Tubing, borosilicate glass, 10 mm in outside diameter, with spherical ground-glass joints, 18/7.
H—Tubing, silicate glass, 10 mm in outside diameter, with spherical ground-glass joints, 18/7.
I—Metal connection, brass spherical male joint at one end fitting to connection to needle valve at other end.
J—Needle valve, brass.
K—Valve on cylinder containing hydrocarbon material.
L—Standard cylinder containing hydrocarbon material.
M—Fitting to connect needle valve J to valve K on cylinder.
FIG. 1 Apparatus for Obtaining Sample

the freezing point will not be affected significantly because of the sample is now ready for introduction into the freezing tube (O
extremely low solubility of water in the hydrocarbon at the freezing point in Fig. 1 of Test Method D 1015).
of the latter.
8.3 Assemble the freezing point apparatus. Place the cool- NOTE 4—In case the original sample contained water, there will remain
at I some water that may be discarded after the hydrocarbon portion has
ing bath in position around the freezing tube (O in Fig. 1 of
been collected as outlined above.
Test Method D 1015), letting the temperature as read on the
platinum thermometer reach about −80°C when all the sample 8.5 When the temperature of the platinum thermometer is
has been collected. near −80°C, remove the condensing tube (E in Fig. 1) from the
8.4 When 50 mL of liquid (temperature about −80°C) has Dewar flask. Wrap a cloth around the upper portion of the
been collected in the condensing tube, close the valve K (Fig. condensing tube (for ease of handling and for preventing the
1) and allow the liquid which has collected at I to warm and refrigerating liquid from contaminating the sample on pour-
transfer to the condensing tube (Note 4). Replace the attaching ing), and after removing the caps on the condensing tube, raise
tubes G and D on the condensing tube by caps. The liquid the stopper holding the platinum thermometer, and pour the

3
D 1016 – 05e2

C—Dewar vessel, 1-qt capacity, borosilicate glass.


D—Clamp.
E—Distilling tube, borosilicate glass, 25 mm in outside diameter.
F—Standard-taper ground-glass joint, 24/40 borosilicate glass. A—Standard-taper, ground-glass joint, 24/40, borosilicate glass
G—Tubing, 10 mm in outside diameter, borosilicate glass. B—Distilling flask, round bottom, 200-mL capacity, borosilicate glass.
H, H8—Spherical ground-glass joints, 18/7, borosilicate glass. C—Tubing, 10 mm in outside diameter, borosilicate glass.
I—Tubing, 6 mm in outside diameter, borosilicate glass. D, D8—Spherical ground-glass joints, 18/7, borosilicate glass.
J—Receiver, 35 mm in outside diameter, 150 mm in length, borosilicate glass. E—Dewar flask, 1-qt capacity, borosilicate glass.
F—Receiver, same as J in Fig. 2.
FIG. 2 Simple Distilling Apparatus for Normally Gaseous
Substances FIG. 3 Simple Distilling Apparatus for Normally Liquid
Substances

sample through the tapered male outlet of the condensing tube Calculated Purity, Uncertainty, plus or
into the freezing tube (O in Fig. 1 of Test Method D 1015). mole % minus, mole %
Over 99.5 0.08
Quickly replace the stopper holding the platinum thermometer, 99.0 to 99.5 0.09
and start the stirrer, with dry air flowing into the upper portion 98 to 99 0.10
of the freezing tube through M (Fig. 1 of Test Method D 1015). 97 to 98 0.12
96 to 97 0.15
8.6 Because the material is normally gaseous at room 95 to 96 0.20
temperature, care should be taken in disposing of the sample
safely. 9. Isobutane6 (Warning—Extremely flammable gas under
8.7 For n-butane, the freezing point for zero impurity, in air pressure. Vapor reduces oxygen available for breathing.)
at 1 atm, is as follows: 9.1 Determine the freezing point from freezing curves with
tf 0 5 –138.362 6 0.025°C (3) the cage stirrer, with a cooling bath of liquid nitrogen (or liquid
and the cryoscopic constants are: air), with a cooling rate of 0.3 to 0.8°C/min for the liquid near
the freezing point, and with crystallization induced immedi-
A 5 0.03085 mole fraction/°C and (4)
ately below the freezing point by means of a cold rod.
B 5 0.0048 mole fraction/°C. (5) 9.2 Obtain the samples as follows: Assemble the apparatus
8.8 The cryoscopic constants given in 8.7 are applicable to for obtaining the sample as shown in Fig. 1, but with no
samples of n-butane having a purity of about 95 mole % or lubricant on the ground-glass joints and with the valve at the
better, with no one impurity present in an amount that exceeds bottom of the cylinder, so that sampling is from the liquid
its eutectic composition with the major component. phase. Attach to C an absorption tube containing anhydrous
8.9 The estimated uncertainty in the calculated value of the calcium sulfate or other suitable desiccant (except magnesium
purity is as follows, in mole %: perchlorate) so that water is not introduced into the system

4
D 1016 – 05e2

A, A8—Standard-taper ground-glass joints, 14/35 borosilicate glass. G and G8—Ceramic (or glass) fiber pad.
B—Tubing, 27 mm in outside diameter, borosilicate glass. H, H8, H9—Stopcock, ground for high vacuum, borosilicate glass.
C, C8—Clamp. I—Spherical ground-glass joint, 18/7, borosilicate glass.
D— Brass cylinder, 273 mm (103⁄4 in.) in length, 28.6 mm (11⁄8 in.) in inside J—Condensing tube, used as trap (see E in Fig. 1).
diameter; for precautions in use of liquid nitrogen and liquid air, see R in K—Connection to vacuum system.
legend to Fig. 1 of Test Method D 1015 and Notes 2 and 3 of Test Method L, L8—Stopcock, ground for high vacuum, borosilicate glass.
D 1015. M—Standard-taper ground-glass joint, 24/40 borosilicate glass.
D8— Brass cylinder, 254 mm (10 in.) in length, 47.6 mm (17⁄8 in.) in inside diameter, N—Receiver withdrawal, 36 mm in outside diameter, borosilicate glass.
(see D above). O—Dewar flask, 0.0005-m3(1-pt) capacity, borosilicate glass.
E—Original sample. P—Connection to vacuum.
E8—Distilled sample. Q—Funnel with extension, 4 mm in inside diameter, borosilicate glass.
F, F8—Dewar flask, 0.0009-m3(1-qt) capacity, borosilicate glass. R—Connection to drying tube, borosilicate glass.
FIG. 4 Apparatus for Simple Vacuum Distillation

(Note 3). Fill the flask F with the carbon dioxide refrigerant to sample through the tapered male outlet of the condensing tube
within about 51 mm (2 in.) of the top. After about 20 or 30 min, into the freezing tube (O in Fig. 1 of Test Method D 1015).
when the system will have cooled sufficiently, remove the Quickly replace the stopper holding the platinum thermometer
absorption tube and begin the collection of liquid isobutane by and start the stirrer, with dry air flowing into the upper portion
opening the valve K and adjusting the needle valve J so that the of the freezing tube through M (Fig. 1 of Test Method D 1015).
sample is collected at a rate of 1 to 2 mL (liquid)/min in the 9.6 Because of the fact that the material is normally gaseous
condensing tube E. at room temperature, care should be taken in disposing of the
9.3 Assemble the freezing point apparatus. Place the cool- sample safely.
ing bath in position around the freezing tube (O in Fig. 1 of 9.7 For isobutane, the freezing point for zero impurity, in air
Test Method D 1015), letting the temperature as read on the at 1 atm, is:
platinum thermometer reach about −80°C when all the sample tf 0 5 2159.605 6 0.025°C (6)
has been collected.
9.4 When 50 mL of liquid (temperature about −80°C) has and the cryoscopic constants are:
been collected in the condensing tube, close the valve K (Fig. A 5 0.04234 mole fraction/°C and (7)
1) and allow the liquid which had collected at I to warm and B 5 0.0057 mole fraction/°C. (8)
transfer to the condensing tube (Note 4). Replace the attaching
tubes, G and D, on the condensing tube by caps. The liquid 9.8 The cryoscopic constants given in 9.7 are applicable to
sample is now ready for introduction into the freezing tube (O samples of isobutane having a purity of about 95 mole % or
in Fig. 1 of Test Method D 1015). better, with no one impurity present in an amount that exceeds
9.5 When the temperature of the platinum thermometer is its eutectic composition with the major component.
near −80°C, remove the condensing tube (E in Fig. 1) from the 9.9 The estimated uncertainty in the calculated value of the
Dewar flask. Wrap a cloth around the upper portion of the purity is as follows, in mole %:
condensing tube (for ease of handling and for preventing the Calculated Purity, Uncertainty, plus or
mole % minus, mole %
refrigerating liquid from contaminating the sample on pour- Over 99.5 0.10
ing), and after removing the caps on the condensing tube, raise 99.0 to 99.5 0.11
the stopper holding the platinum thermometer, and pour the 98 to 99 0.12

5
D 1016 – 05e2
97 to 98 0.14 and the cryoscopic constants are:
96 to 97 0.16
95 to 96 0.20
A 5 0.04829 mole fraction/°C and (13)
B 5 0.0058 mole fraction/°C. (14)
10. n-Pentane (Warning—Extremely flammable liquid. 11.4 The cryoscopic constants given in 11.3 are applicable
Harmful if inhaled. Vapors can cause flash fire.) to samples of isopentane having a purity of about 95 mole % or
10.1 Determine the freezing point from freezing curves with better, with the usual impurities and with no one impurity
the cage stirrer, with a cooling bath of liquid nitrogen (or liquid present in an amount which exceeds the composition of the
air), with a cooling rate of 0.3 to 0.8°C/min for the liquid near eutectic with the major component.
the freezing point, and with crystallization induced immedi- 11.5 The estimated uncertainty in the calculated value of the
ately below the freezing point by means of a cold rod. purity is as follows, in mole %:
10.2 To obtain the sample, cool the container and n-pentane Calculated Purity, Uncertainty, plus or
to near 0°C and transfer a sample of about 60 mL (liquid at the mole % minus, mole %
Over 99.5 0.07
given temperature) to a graduated cylinder which has been kept 99.0 to 99.5 0.08
refrigerated slightly below 0°C. The sample is now ready for 98 to 99 0.09
introduction into the freezing tube, which should be precooled 97 to 98 0.10
96 to 97 0.12
to near −80°C. 95 to 96 0.14
10.3 For n-pentane, the freezing point for zero impurity, in
air at 1 atm, is as follows: 12. n-Hexane (Warning—Extremely flammable. Harmful if
inhaled. Can produce nerve cell damage. Vapors can
tf 0 5 2129.730 6 0.015°C (9)
cause flash fire.)
and the cryoscopic constants are: 12.1 Determine the freezing point from freezing curves with
A 5 0.04906 mole fraction/°C and (10) the cage stirrer, with a cooling bath of liquid nitrogen (or liquid
B 5 0.0042 mole fraction/°C. (11) air), with a cooling rate of 0.3 to 0.8°C/min for the liquid near
the freezing point and with crystallization induced immediately
10.4 The cryoscopic constants given in 10.3 are applicable
below the freezing point by means of a cold rod.
to samples of n-pentane having a purity of about 95 mole % or
12.2 Obtain a sample of 50 mL (measured at room tempera-
better, with the usual impurities and with no one impurity
ture) directly from its original container by means of a pipet or
present in an amount that exceeds the composition of its
by pouring into a graduated cylinder.
eutectic with the major component.
12.3 For n-hexane, the freezing point for zero impurity, in
10.5 The estimated uncertainty in the calculated value of the
air at 1 atm, is as follows:
purity is as follows, in mole %:
Calculated Purity, Uncertainty, plus or tf 0 5 295.322 6 0.010°C (15)
mole % minus, mole % and the cryoscopic constants are:
Over 99.5 0.07
99.0 to 99.5 0.08 A 5 0.04956 mole fraction/°C and (16)
98 to 99 0.09
97 to 98 0.10 B 5 0.0039 mole fraction/°C. (17)
96 to 97 0.12 12.4 The cryoscopic constants given in 12.3 are applicable
95 to 96 0.14
to samples of n-hexane having a purity of about 95 mole % or
11. Isopentane (Warning—Extremely flammable liquid. better, with the usual impurities and with no one impurity
Harmful if inhaled. Vapors can cause flash fire.) present in an amount that exceeds the composition of its
eutectic with the major component.
11.1 Determine the freezing point from melting curves with 12.5 The estimated uncertainty in the calculated value of the
the double helix stirrer, with a cooling bath of liquid nitrogen purity is as follows, in mole %:
(or liquid air) to obtain the slurry of crystals and liquid, and a
Calculated Purity, Uncertainty, plus or
warming bath of carbon dioxide refrigerant, with a cooling rate mole % minus, mole %
of 0.3 to 0.8°C/min for the liquid near the freezing point and Over 99.5 0.05
99.0 to 99.5 0.06
with crystallization induced immediately below the freezing 98 to 99 0.07
point, by seeding with crystals. (Crystallization may also be 97 to 98 0.08
induced with a cold rod, but the recovery from undercooling 96 to 97 0.10
95 to 96 0.12
will not be as rapid.)
11.2 To obtain a sample, cool the container and isopentane 13. n-Heptane (Warning—Flammable. Harmful if inhaled.)
to near 0°C and transfer a sample of about 65 mL (liquid at the 13.1 Determining the freezing point from freezing curves
given temperature) to a graduated cylinder which has been kept with the cage stirrer, with a cooling bath of liquid nitrogen (or
refrigerated slightly below 0°C. The sample is now ready for liquid air), with a cooling rate of 0.3 to 0.8°C/min for the liquid
introduction into the freezing tube which should be precooled near the freezing point, and with crystallization induced
to near −80°C. immediately below the freezing point by means of a cold rod.
11.3 For isopentane, the freezing point for zero impurity, in 13.2 Obtain a sample of 50 mL (measured at room tempera-
air at 1 atm, is as follows: ture) directly from its original container by means of a pipet or
tf 0 5 2159.905 6 0.015° C (12) by pouring into a graduated cylinder.

6
D 1016 – 05e2
13.3 For n-heptane, the freezing point for zero impurity, in with a cooling rate of 0.3 to 0.8°C/min for the liquid near the
air at 1 atm, is: freezing point and with crystallization induced immediately
tf 0 5 290.581 6 0.010°C (18) below the freezing point by means of a cold rod.
15.2 For samples having a purity less than about 99.5
and the cryoscopic constants are:
mole %, determine the freezing point from freezing curves
A 5 0.05065 mole fraction/°C and (19) with the cage stirrer, with a cooling bath of liquid nitrogen (or
B 5 0.0033 mole fraction/°C. (20) liquid air), with a cooling rate of 0.3 to 0.8°C/min for the liquid
13.4 The cryoscopic constants given in 13.3 are applicable near the freezing point, and with crystallization induced
to samples of n-heptane having a purity of about 95 mole % or immediately below the freezing point by means of a cold rod.
better, with the usual impurities and with no one impurity 15.3 Obtain a sample of 50 mL (measured at room tempera-
present in an amount that exceeds the composition of its ture) directly from its original container by means of pipet or
eutectic with the major component. by pouring into a graduated cylinder.
13.5 The estimated uncertainty in the calculated value of the 15.4 For 2,2,4-trimethylpentane, the freezing point for zero
purity is as follows, in mole %: impurity, in air at 1 atm, is as follows:
Calculated Purity, Uncertainty, plus or minus, tf 0 5 2107.373 6 0.010°C (24)
mole % mole %
Over 99.5 0.05 and the cryoscopic constants are:
99.0 to 99.5 0.06
98 to 99 0.07
A 5 0.04032 mole fraction/°C and (25)
97 to 98 0.08 B 5 0.0043 mole fraction/°C. (26)
96 to 97 0.10
95 to 96 0.12 15.5 The cryoscopic constants given in 15.4 are applicable
to samples of 2,2,4-trimethylpentane having a purity of about
14. n-Octane (Warning—Flammable. Harmful if inhaled.)
95 mole % or better, with the usual impurities and with no one
14.1 Determine the freezing point from freezing curves with impurity present in an amount that exceeds the composition of
the cage stirrer, with a cooling bath of carbon dioxide refrig- its eutectic with the major component.
erant at a cooling rate of 0.3 to 0.8°C/min for the liquid near the 15.6 The estimated uncertainty in the calculated value of the
freezing point, and with crystallization induced immediately purity is as follows, in mole %:
below the freezing point by means of a cold rod. Calculated Purity, Uncertainty, plus or
14.2 Obtain a sample of 50 mL (measured at room tempera- mole % minus, mole %
ture) directly from its original container by means of a pipet or Over 99.5 0.05
99.0 to 99.5 0.06
by pouring into a graduated cylinder. 98 to 99 0.07
14.3 For n-octane the freezing point for zero impurity, in air 97 to 98 0.08
at 1 atm, is as follows: 96 to 97 0.10
95 to 96 0.12
tf 0 5 256.764 6 0.010°C (21)
and the cryoscopic constants are: 16. Methylcyclohexane (Warning—Flammable. Harmful if
A 5 0.05329 mole fraction/°C and (22) inhaled.)
B 5 0.0031 mole fraction/°C. (23) 16.1 Determine the freezing point from melting curves with
the double helix stirrer, with a cooling bath of liquid nitrogen
14.4 The cryoscopic constants given in 14.3 are applicable
(or liquid air) to obtain the slurry of crystals and liquid, and a
to samples of n-octane having a purity of about 95 mole % or
warming bath of carbon dioxide refrigerant, with a cooling rate
better, with the usual impurities and with no impurity present
of 0.3 to 0.8°C/min for the liquid near the freezing point and
in an amount that exceeds the composition of its eutectic with
with crystallization induced immediately below the freezing
the major component.
point by seeding with crystals. (Crystallization may also be
14.5 The estimated uncertainty in the calculated value of the
induced with a cold rod, but the recovery from undercooling
purity is as follows, in mole %:
will not be as rapid.)
Calculated Purity, Uncertainty, plus or minus,
mole % mole %
16.2 Obtain a sample of 60 mL (measured at room tempera-
Over 99.5 0.05 ture) directly from the original container by pouring into a
99.0 to 99.5 0.06 graduated cylinder.
98 to 99 0.07
97 to 98 0.08 16.3 For methylcyclohexane, the freezing point for zero
96 to 97 0.10 impurity, in air at 1 atm, is as follows:
95 to 96 0.12
tf 0 5 2126.596 6 0.015°C (27)
15. 2,2,4-Trimethylpentane (Warning—Extremely and the cryoscopic constants are:
flammable. Harmful if inhaled. Vapors can cause flash
fire.) A 5 0.03779 mole fraction/°C and (28)
A 5 0.0032 mole fraction/°C. (29)
15.1 For samples having a purity greater than about 99.5
mole %, determine the freezing point from melting curves with 16.4 The cryoscopic constants given in 16.3 are applicable
the cage stirrer, with a cooling bath of liquid nitrogen (or liquid to samples of methylcyclohexane having a purity of about 95
air), and a warming bath of solid carbon dioxide refrigerant, mole % or better, with the usual impurities and with no one

7
D 1016 – 05e2
impurity present in an amount that exceeds the composition of 17.6 Because of the fact that the material is normally
the eutectic with the major component. gaseous at room temperature, care should be taken in disposing
16.5 The estimated uncertainty in the calculated value of the of the sample safely.
purity is as follows, in mole %: 17.7 For isobutene, the freezing point for zero impurity, in
Calculated Purity, Uncertainty, plus or air at 1 atm, is as follows:
mole % minus, mole % tf 0 5 2140.337 6 0.020° C (30)
Over 99.5 0.05
99.0 to 99.5 0.06 and the cryoscopic constants are:
98 to 99 0.07
97 to 98 0.08 A 5 0.04044 mole fraction/°C and (31)
96 to 97 0.10 B 5 0.005 mole fraction/°C. (32)
95 to 96 0.12
17.8 The cryoscopic constants given in 17.7 are applicable
17. Isobutene6 (Warning—Extremely flammable liquefied to samples of isobutene having a purity of about 95 mole % or
gas under pressure. Vapor reduces oxygen available for better, with no one impurity present in an amount that exceeds
breathing.) its eutectic composition with the major component.
17.9 The estimated uncertainty in the calculated value of the
17.1 Determine the freezing point from freezing curves,
purity is as follows, in mole %:
with the cage stirrer, with a cooling bath of liquid nitrogen (or
Calculated Purity, Uncertainty, plus or
liquid air), with a cooling rate of 0.3 to 0.8°C/min for the liquid mole % minus, mole %
near the freezing point, and with crystallization induced Over 99.5 0.08
immediately below the freezing point by means of a cold rod. 99.0 to 99.5 0.09
98 to 99 0.10
17.2 Obtain the samples as follows: Assemble the apparatus 97 to 98 0.12
for obtaining the sample as shown in Fig. 1, but with no 96 to 97 0.15
95 to 96 0.20
lubricant on the ground-glass joints and with the valve at the
bottom of the cylinder, so that sampling is from the liquid
18. 1,3-Butadiene6 (Warning—Extremely flammable
phase. Attach to C an absorption tube containing anhydrous
liquefied gas under pressure. May form explosive
calcium sulfate or other suitable desiccant (except magnesium peroxides upon exposure to air. Harmful if inhaled.
perchlorate) so that water is not introduced into the system Irritating to eyes, skin, and mucous membranes.)
(Note 3). Fill the flask F with the carbon dioxide refrigerant to
within about 51 mm (2 in.) of the top. After about 20 or 30 min, 18.1 Determine the freezing point from freezing curves with
when the system will have cooled sufficiently, remove the the cage stirrer, with a cooling bath of liquid nitrogen (or liquid
absorption tube and begin the collection of liquid isobutene by air), with a cooling rate of 0.3 to 0.8°C/min for the liquid near
opening the valve K and adjusting the needle valve J so that the the freezing point and with crystallization induced immediately
sample is collected at a rate of 1 to 2 mL (liquid)/min in the below the freezing point by means of a cold rod.
condensing tube E. 18.2 Obtain the samples as follows: Assemble the apparatus
for obtaining the sample as shown in Fig. 1, but with no
17.3 Assemble the freezing point apparatus. Place the cool-
lubricant on the ground-glass joints and with the valve at the
ing bath in position around the freezing tube (O in Fig. 1 of
bottom of the cylinder, so that sampling is from the liquid
Test Method D 1015), letting the temperature as read on the
phase. Attach to C an absorption tube containing anhydrous
platinum thermometer reach about −80°C when all the sample calcium sulfate or other suitable desiccant (except magnesium
has been collected. perchlorate) so that water is not introduced into the system
17.4 When 50 mL of liquid (temperature about −80°C) has (Note 3). Fill the flask F with the carbon dioxide refrigerant to
been collected in the condensing tube, close the valve K (Fig. within about 51 mm (2 in.) of the top. After about 20 or 30 min,
1) and allow the liquid which has collected at I to warm and when the system will have cooled sufficiently, remove the
transfer to the condensing tube (Note 4). Replace the attaching absorption tube and begin the collection of liquid 1,3-butadiene
tubes G and D on the condensing tube by caps. The liquid by opening the valve K and adjusting the needle valve J so that
sample is now ready for introduction into the freezing tube (O the sample is collected at a rate of 1 to 2 mL (liquid)/min in the
in Fig. 1 of Test Method D 1015). condensing tube E.
17.5 When the temperature of the platinum thermometer is 18.3 Assemble the freezing point apparatus. Place the cool-
near −80°C, remove the condensing tube (E in Fig. 1) from the ing bath in position around the freezing tube (O in Fig. 1 of
Dewar flask. Wrap a cloth around the upper portion of the Test Method D 1015), letting the temperature as read on the
condensing tube (for ease of handling and for preventing the platinum thermometer reach about −80°C when all the sample
refrigerating liquid from contaminating the sample on pour- has been collected.
ing), and after removing the caps on the condensing tube, raise 18.4 When 50 mL of liquid (temperature about −80°C) has
the stopper holding the platinum thermometer, and pour the been collected in the condensing tube, close the valve K (Fig.
sample through the tapered male outlet of the condensing tube 1) and allow the liquid, which has collected at I, to warm and
into the freezing tube (O in Fig. 1 of Test Method D 1015). transfer to the condensing tube (Note 4). Replace the attaching
Quickly replace the stopper holding the platinum thermometer tubes G and D on the condensing tube by caps. The liquid
and start the stirrer, with dry air flowing into the upper portion sample is now ready for introduction into the freezing tube (O
of the freezing tube through M (Fig. 1 of Test Method D 1015). in Fig. 1 of Test Method D 1015).

8
D 1016 – 05e2
18.5 In some cases, it will be desirable to remove the dimer, 18.10 The estimated uncertainty in the calculated value of
other C8 hydrocarbons, and higher polymer from the sample of the purity is as follows, in mole %:
1,3-butadiene before determining the purity. For this removal, Calculated Purity, Uncertainty, plus or
the procedure is as follows: Assemble the apparatus shown in mole % minus, mole %
Over 99.5 0.05
Fig. 2 with a small amount (10 to 100 ppm) of tertiary butyl 99.0 to 99.5 0.06
catechol or other suitable inhibitor placed in the bottom of the 98 to 99 0.07
distilling tube E, with no lubricant on the ground-glass joints. 97 to 98 0.08
96 to 97 0.09
It is also desirable to place at the bottom of the flask a piece of 95 to 96 0.10
carborundum or other suitable material to prevent bumping.
Make a connection to the atmosphere through an absorption 19. Isoprene (2-Methyl-1,3-Butadiene)6 (Warning—
tube (as previously described in this section) at H8 so that Extremely flammable liquefied gas under pressure. Vapor
entering air is freed of carbon dioxide and water. Place a bath reduces oxygen available for breathing.)
containing carbon dioxide refrigerant around the distilling tube
19.1 For samples having a purity greater than about 98
E, and also around the receiver J so that the small entrance and
mole %, determine the freezing point from melting curves,
exit tubes of J are covered with at least 5 cm of the bath. After
with the double helical stirrer, with a cooling bath of liquid
about 20 to 30 min, when the system will have precooled
nitrogen (or liquid air), and a warming bath of carbon dioxide
sufficiently, disconnect the connection to the atmosphere at H8,
refrigerant, with the jacket of the freezing tube open to the high
remove the cap F, and introduce the liquid butadiene (tempera-
vacuum system during the entire melting part of the curve, with
ture near −80°C) by pouring through a precooled funnel (such
a cooling rate of 0.3 to 0.8°C/min for the liquid near the
as Q in Fig. 4 which may be cooled without contamination by
freezing point and with crystallization induced immediately
liquid air or liquid nitrogen) into the distilling tube. Grease the
below the freezing point by means of a cold rod.
cap F and replace immediately after the introduction of the
sample. Then distill the material by removing the bath from the 19.2 For samples having a purity less than about 98 mole %,
distilling tube and allowing it to warm in contact with the air determine the freezing point from freezing curves, with either
of the room. Distillation is complete when the distilling tube the aluminum cage stirrer or the double helical stirrer, with a
has warmed to room temperature. Disconnect the receiver with cooling bath of liquid nitrogen (or liquid air), with a cooling
the bath around it, cap it at H and H8, and transfer 50 mL rate of 0.3 to 0.8°C/min for the liquid near the freezing point
(liquid at about − 80°C) of the liquid butadiene to the freezing and with crystallization induced immediately below the freez-
tube by pouring through I in a manner similar to that described ing point by means of a cold rod.
for a sample collected in the condensing tube. 19.3 The method of obtaining the sample is as follows:
18.6 When the temperature of the platinum thermometer is When the material is in a cylinder, assemble the apparatus
near −80°C, remove the condensing tube (E in Fig. 1) or the shown in Fig. 1, with a suitable lubricant on the ground-glass
receiver ( J in Fig. 2) from the Dewar. Wrap a cloth around the joints, and with the valve below the body of the cylinder so that
upper portion of the condensing tube or receiver (for ease of the sample is obtained from the liquid phase. Evacuate the
handling and for preventing the refrigerating liquid from system by connecting, through heavy-walled tubing, the open-
contaminating the sample on pouring), and after removing the ing B to a vacuum line. After evacuation, close the stopcock A
caps on the condensing tube or receiver, raise the stopper to the outlets B and C, and collect the sample of isoprene (55
holding the platinum thermometer, and pour the sample mL, liquid, at about −80°C) in the refrigerated condensing tube
through the tapered male outlet of the condensing tube or the E, in which was previously placed a small amount (about 10 to
exit tube I of the receiver into the freezing tube (O in Fig. 1 of 100 ppm) of tertiary butyl catechol or other suitable inhibitor.
Test Method D 1015). Quickly replace the stopper holding the The sample as thus collected will contain the bulk of any dimer
platinum thermometer and start the stirrer, with dry air flowing present in the original material. The sample, including substan-
into the upper portion of the freezing tube through M (Fig. 1 of tially all of the dimer, is now ready for introduction into the
Test Method D 1015). freezing tube, which should be precooled to near −100°C.
When the isoprene is contained in capped bottles or sealed
18.7 Because of the fact that the material is normally
ampoules, cool the container and isoprene to near 0°C and
gaseous at room temperature, care should be taken in disposing
transfer a sample of about 65 mL (liquid at the given
of the sample safely.
temperature) to a graduated cylinder which has been kept
18.8 For 1,3-butadiene, the freezing point for zero impurity,
refrigerated slightly below 0°C. The sample, including such
in air at 1 atm, is
amount of dimer and higher polymer as was originally present,
tf 0 5 2108.902 6 0.010° C (33) is now ready for introduction into the freezing tube, which
and the cryoscopic constants are: should be precooled to near −100°C.
A 5 0.03560 mole fraction/°C and (34) 19.4 In most cases, it will be desirable to remove the dimer
and higher polymer from the sample of isoprene before
B 5 0.0053 mole fraction/°C. (35)
determining the purity. For this removal, the procedure is as
18.9 The cryoscopic constants given in 18.8 are applicable follows: Assemble the apparatus shown in Fig. 3 with no
to samples of 1,3-butadiene having a purity of about 95 mole % lubricant on the ground-glass joints D and D8. Place a small
or better, with no one impurity present in an amount that amount of tertiary butyl catechol or other suitable inhibitor
exceeds its eutectic composition with the major component. (about 10 to 100 ppm) in the receiver F and a larger amount

9
D 1016 – 05e2
(about 100 to 1000 ppm) in the distilling flask B. It is also Surround the receiver N8 by carbon dioxide refrigerant. Re-
desirable to place at the bottom of the flask B a piece of move the material by inserting the inlet tube at L into the
carborundum or other suitable material to prevent bumping. receiver and then opening the stopcock L. This procedure
Place a cooling bath of water-ice around the distilling flask B avoids loss by evaporation. Then introduce the material into
and a bath containing carbon dioxide refrigerant around the the freezing tube, previously precooled to near −100°C, by
receiver F. Make a connection to the atmosphere at D8 through pouring through the tapered joint at M. For the procedure for
which the air is first freed of carbon dioxide and water, using introducing the sample into the tube, see 18.6 on 1,3-butadiene.
a tube containing Ascarite and anhydrous calcium sulfate or 19.6 For isoprene (2-methyl-1,3-butadiene), the freezing
other suitable desiccant. Introduce the sample (at 0°C) into the point for zero impurity, in air at 1 atm, is:
flask B, place the cap A in position with a suitable lubricant tf 0 5 2145.964 6 0.020°C (36)
between the grindings, and remove the connection to the
atmosphere at D8. Place a water bath (at 40 to 50°C) around the and the cryoscopic constants are:
flask B and distill the material into F. Stop the distillation when A 5 0.0330 mole fraction/°C and (37)
a small residue remains in B with the water bath at 50°C. B 5 0.0030 mole fraction/°C. (38)
Detach the receiver F at D and cap at D and D8 with the bath
containing carbon dioxide refrigerant still surrounding it. 19.7 The cryoscopic constants given in 19.6 are applicable
to samples of isoprene having a purity of about 95 mole % or
Remove the sample, with the upper portion of the container
better, with no one impurity present in an amount that exceeds
wrapped with a cloth (for ease of handling and for preventing
the eutectic composition with the major component.
the refrigerating liquid from contaminating the sample on
pouring), from the flask E, remove the caps and introduce the 19.8 The estimated uncertainty in the calculated value of the
sample into the freezing tube, previously precooled to near purity is as follows, in mole %:
−100°C, by pouring through D8. For the procedure for intro- Calculated Purity, Uncertainty, plus or
mole % minus, mole %
ducing the sample into the tube, see 18.6 on 1,3-butadiene. Over 99.5 0.08
19.5 If the sample contains a very large amount of dimer 99.0 to 99.5 0.10
98 to 99 0.12
and polymer, then the simple preceding procedure outlined will 97 to 98 0.15
not suffice because the required distilling temperature will be 96 to 97 0.20
too high, and a more complicated procedure is used, as follows: 95 to 96 0.25

Assemble the apparatus shown in Fig. 4, with inhibitor placed


in the distilling tube and receiver (plus some carborundum in 20. Benzene (Warning—Poison. Carcinogen. Harmful or
the distilling tube to prevent bumping) as previously described fatal if swallowed. Extremely flammable. Vapors can
in 19.4, and with all the ground joints except that at A cause flash fire. Vapor harmful, can be absorbed through
lubricated. Place a cooling bath of carbon dioxide refrigerant skin.)
around the distilling tube E. Permit air, freed of carbon dioxide 20.1 Determine the freezing point from freezing curves with
and water, to enter the system through RH9 H8H in order to the cage stirrer, with a cooling bath of carbon dioxide refrig-
compensate for the change in volume. When the sample is erant, with a cooling rate of 0.3 to 0.8°C/min for the liquid near
cooled, remove the cap A and introduce the sample through the the freezing point and with crystallization induced immediately
funnel Q, which has been precooled with liquid air or liquid below the freezing point by means of a cold rod.
nitrogen. Then lubricate the cap A and close the stopcocks H, 20.2 Obtain a sample of 50 mL (measured at room tempera-
H8, and H9. Place liquid air or liquid nitrogen around the ture) directly from the original container by means of a pipet or
condensing tube E (Fig. 1), which serves as a trap, and also by pouring into a graduated cylinder. Then filter the sample
replace the carbon dioxide refrigerant around the distilling tube directly into the freezing point tube (O in Fig. 1 of Test Method
E (Fig. 4) with liquid nitrogen or liquid air. After the isoprene D 1015), through silica gel to remove water. See 9.3, and Fig.
has solidified, evacuate the system by opening H and H8 to the 5 of Test Method D 1015.
vacuum system. Close the stopcocks H and H8 and remove the 20.3 For benzene, the freezing point for zero impurity, in air
bath from E to allow the material to melt and release dissolved at 1 atm, is:
air. Crystallize the material again and evacuate the system as
tf 0 5 5.531 6 0.010°C (39)
before. Repeat the process again, if necessary, to remove
substantially all the air. (If any hydrocarbon has been caught in and the cryoscopic constants are:
the trap J, it should be distilled back into the tube E, with the A 5 0.01523 mole fraction/°C and (40)
stopcock H open and H8 closed.) Distill the material into E8 by B 5 0.0032 mole fraction/°C. (41)
placing carbon dioxide refrigerant around the receiver and a
water-ice bath around E (after the latter has warmed to near 20.4 The cryoscopic constants given in 20.3 are applicable
0°C). Halt the distillation when the transfer of material into the to samples of benzene having a purity of about 95 mole % or
receiver has substantially halted, by admitting air (freed of better, with the usual impurities and with no one impurity
water and carbon dioxide) into the system through RH9 H8H. present in an amount that exceeds the composition of its
Remove the sample from the receiving tube E with the eutectic with the major component.
withdrawal receiver N. Evacuate the system LMNL, with L8 20.5 The estimated uncertainty in the calculated value of the
open and L closed, through P and then close the stopcock L8. purity is as follows, in mole %:

10
D 1016 – 05e2
Calculated Purity, Uncertainty, plus or 22.4 The cryoscopic constants given in 22.3 are applicable
mole % minus, mole % to samples of ethylbenzene having a purity of about 95 mole %
Over 99.5 0.02
99.0 to 99.5 0.03
or better, with the usual impurities and with no one impurity
98 to 99 0.04 present in an amount that exceeds the composition of its
97 to 98 0.05 eutectic with the major component.
96 to 97 0.06
95 to 96 0.08 22.5 The estimated uncertainty in the calculated value of the
purity is as follows, in mole %:
21. Toluene (Warning—Flammable. Vapor harmful.) Calculated Purity, Uncertainty, plus or
mole % minus, mole %
21.1 Determine the freezing point from freezing curves with Over 99.5 0.05
the cage stirrer, with a cooling bath of liquid nitrogen (or liquid 99.0 to 99.5 0.06
98 to 99 0.08
air), with a cooling rate of 0.3 to 0.8°C/min for the liquid near 97 to 98 0.10
the freezing point, and with crystallization induced immedi- 96 to 97 0.12
ately below the freezing point by means of a cold rod. 95 to 96 0.14
21.2 A sample of 50 mL (measured at room temperature) is
obtained directly from its original container by means of a 23. o-Xylene (Warning—Flammable. Vapor harmful.)
pipet or by pouring into a graduated cylinder. 23.1 The freezing point is determined from freezing curves
21.3 For toluene, the freezing point for zero impurity, in air with the cage stirrer, with a cooling bath of carbon dioxide
at 1 atm, is: refrigerant, with a cooling rate of 0.3 to 0.8°C/min for the
tf 0 5 294.965 6 0.012°C (42) liquid near the freezing point, and with crystallization induced
immediately below the freezing point by seeding with crystals.
and the cryoscopic constants are: (Crystallization may also be induced with a cold rod, but the
A 5 0.02508 mole fraction/°C and (43) recovery from undercooling will not be as rapid.)
B 5 0.0019 mole fraction/°C. (44) 23.2 A sample of 50 mL (measured at room temperature) is
21.4 The cryoscopic constants given in 21.3 are applicable obtained directly from its original container by means of a
to samples of toluene having a purity of about 95 mole % or pipet or by pouring into a graduated cylinder.
better, with the usual impurities and with no one impurity 23.3 For o-xylene, the freezing point for zero impurity, in
present in an amount that exceeds the composition of its air at 1 atm, is:
eutectic with the major component. tf 0 5 225.167 6 0.005°C (48)
21.5 The estimated uncertainty in the calculated value of the and the cryoscopic constants are:
purity is as follows, in mole %:
A 5 0.02659 mole fraction/°C and (49)
Calculated Purity, Uncertainty, plus or
mole % minus, mole % B 5 0.0030 mole fraction/°C. (50)
Over 99.5 0.03
99.0 to 99.5 0.04 23.4 The cryoscopic constants given in 23.3 are applicable
98 to 99 0.05 to samples of o-xylene having a purity of about 95 mole % or
97 to 98 0.06 better, with the usual impurities and with no one impurity
96 to 97 0.08
95 to 96 0.10 present in an amount that exceeds the composition of its
eutectic with the major component.
22. Ethylbenzene (Warning—Flammable. Vapor harmful.) 23.5 The estimated uncertainty in the calculated value of the
22.1 Determine the freezing point from melting curves with purity is as follows, in mole %:
the cage stirrer, with a cooling bath of liquid nitrogen (or liquid Calculated Purity, Uncertainty, plus or
mole % minus, mole %
air) to obtain the slurry of crystals and liquid, and a warming Over 99.5 0.02
bath of carbon dioxide refrigerant, with a cooling rate of 0.3 to 99.0 to 99.5 0.03
98 to 99 0.04
0.8°C/min for the liquid near the freezing point, and with
97 to 98 0.05
crystallization induced immediately below the freezing point 96 to 97 0.06
by seeding with crystals. (Crystallization may also be induced 95 to 96 0.08
with a cold rod, but the recovery from undercooling will not be
as rapid.) 24. m-Xylene (Warning—Flammable. Vapor harmful.)
22.2 Obtain a sample of 50 mL (measured at room tempera- 24.1 Determine the freezing point from freezing curves with
ture) directly from its original container by means of a pipet or the cage stirrer, with a cooling bath of solid carbon dioxide
by pouring into a graduated cylinder. refrigerant, with a cooling rate of 0.3 to 0.8°C/min for the
22.3 For ethylbenzene, the freezing point for zero impurity, liquid near the freezing point, and with crystallization induced
in air at 1 atm, is: immediately below the freezing point by seeding with crystals.
tf 0 5 294.949 6 0.015°C (45) (Crystallization may also be induced with a cold rod, but the
recovery from undercooling will not be as rapid.)
and the cryoscopic constants are:
24.2 Obtain a sample of 50 mL (measured at room tempera-
A 5 0.03471 mole fraction/°C and (46) ture) directly from its original container by means of a pipet or
B 5 0.0029 mole fraction/°C. (47) by pouring into a graduated cylinder.

11
D 1016 – 05e2
24.3 For m-xylene, the freezing point for zero impurity, in 97 to 98 0.06
air at 1 atm, is: 96 to 97 0.08
95 to 96 0.10
tf 0 5 247.844 6 0.020°C (51)
and the cryoscopic constants are: 26. Styrene (Ethenylbenzene)6 (Warning—Flammable.
Vapor harmful.)
A 5 0.02741 mole fraction/°C and (52)
26.1 Determine the freezing point from freezing curves with
B 5 0.0027 mole fraction/°C. (53)
the cage stirrer, with cooling bath of carbon dioxide refrigerant,
24.4 The cryoscopic constants given in 24.3 are applicable with a cooling rate of 0.3 to 0.8°C/min for the liquid near the
to samples of m-xylene having a purity of about 95 mole % or freezing point, and with crystallization induced immediately
better, with the usual impurities and with no one impurity below the freezing point by means of a cold rod.
present in an amount that exceeds the composition of its 26.2 Obtain a sample of 50 mL (measured at room tempera-
eutectic with the major component. ture) directly from the original container by means of a pipet or
24.5 The estimated uncertainty in the calculated value of the by pouring into a graduated cylinder.
purity is as follows, in mole %: 26.3 If the previous treatment or storage condition of
Calculated Purity, Uncertainty, plus or material was such that dimerization or polymerization may
mole % minus, mole %
Over 99.5 0.05
have occurred, the dimer or polymer should be removed by a
99.0 to 99.5 0.06 simple vacuum distillation (Fig. 4), using the same procedure
98 to 99 0.07 as described under isoprene (see 19.5) except that carbon
97 to 98 0.08
96 to 97 0.10
dioxide refrigerant is used to refrigerate the receiver and the
95 to 96 0.12 sample is distilled at room temperature.
26.4 For styrene, the freezing point for zero impurity, in air
25. p-Xylene (Warning—Flammable. Vapor harmful.) at 1 atm, is:
25.1 Determine the freezing point from freezing curves with tf 0 5 230.610 6 0.008°C (57)
the cage stirrer, with a cooling bath of carbon dioxide refrig-
erant, with a cooling rate of 0.3 to 0.8°C/min for the liquid near and the cryoscopic constants are:
the freezing point, and with crystallization induced immedi- A 5 0.02365 mole fraction/°C and (58)
ately below the freezing point, by seeding with crystals. B 5 0.0044 mole fraction/°C. (59)
(Crystallization may also be induced with a cold rod, but the 26.5 The cryoscopic constants given in 26.4 are applicable
recovery from undercooling will not be as rapid.) to samples of styrene having a purity of not less than about 95
25.2 Obtain a sample of 50 mL (measured at room tempera- mole %, with the usual impurities and with no one impurity
ture) directly from the original container by means of a pipet or present in an amount that exceeds the composition of its
by pouring into a graduated cylinder. The sample is then eutectic with the main component.
filtered directly into the freezing point tube, (O in Fig. 1 of Test 26.6 The estimated uncertainty in the calculated value of the
Method D 1015), through silica gel to remove water. See 9.3 purity is as follows, in mole %:
and Fig. 5 of Test Method D 1015.
Calculated Purity, Uncertainty, plus or
25.3 For p-xylene, the freezing point for zero impurity, in mole % minus, mole %
air at 1 atm, is: Over 99.5 0.04
99.0 to 99.5 0.05
tf 0 5 13.258 6 0.012°C (54) 98 to 99 0.06
97 to 98 0.07
and the cryoscopic constants are: 96 to 97 0.08
A 5 0.02509 mole fraction/°C and (55) 95 to 96 0.09

B 5 0.0028 mole fraction/°C. (56) 27. Precision and Bias


25.4 The cryoscopic constants given in 25.3 are applicable 27.1 Precision—The precision for this test method is gov-
to samples of p-xylene having a purity of about 95 mole % or erned by the precision of Test Method D 1015. Test Method
better, with the usual impurities and with no one impurity D 1015 must be used for the freezing point determinations in
present in an amount that exceeds the composition of its this test method.
eutectic with the major component. 27.2 Bias—The bias for this test method is governed by the
25.5 The estimated uncertainty in the calculated value of the bias of Test Method D 1015. Test Method D 1015 must be used
purity is as follows, in mole %: for the freezing point determinations in this test method.
Calculated Purity, Uncertainty, plus or
mole % minus, mole % 28. Keywords
Over 99.5 0.03
99.0 to 99.5 0.04 28.1 crystallization; freeze point; LPG; pure hydrocarbons;
98 to 99 0.05 purity

12
D 1016 – 05e2
SUMMARY OF CHANGES

Subcommittee D02.04 has identified the location of selected changes to this standard since the last issue
(D 1016–99(2004)e1) that may impact the use of this standard.

(1) Replaced “asbestos” with “ceramic (or glass) fiber” in


Fig. 4.

ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).

13

You might also like