~ Calibration Certificate ~

Sensor Information: Calibration Data Transducer Specifications


Model Number SEN041F Sensitivity @ 100 Hz: 1.05 mV/m/ss Amp. Range: 500 g
Serial Number P264421 Test Level: 9.81 m/ss RMS Resolution: 0.008 g
Manufacturer Resonant Freq: 55000 Hz
o
ID Number Temp. Range: -54 a +121 C
Description ICP® Accelerometer Axis: X

Deviation Plot
1.0
0.8
0.6
Deviation (%)

0.4
0.2 Calibration Data:
0.0 Sensitivity #N/D mV/m/ss
-0.2 at
-0.4 Ref. Frequency 0.00 Hz
-0.6
-0.8
-1.0
10 100 1000 10000
Frequency (Hz)

Data Table
Deviation
Frequency (Hz) (%) Notes
20 -0.2 Results relate only to the items calibrated.
100 0.0 This certificate may not be reproduced except in full, without written permission.
300 -0.5 Traceable to NIST (Project Number 822/271196) and PTB (Project Number 5399).
500 0.2 This calibration was performed with Portable Vibration Calibrator: M/N: 9210D, S/N: 30242, Firmware version 5.1.1.
800 0.3 Back-to-Back Comparison per ISO 16063-21
1000 0.3

User Notes El presente certificado no debe ser reproducido ni modificado, sujeto a


sanciones legales.
Los resultados documentados en este certificado corresponden solo a los
items testeados o calibrados.

Customer: Gestión QHSEC S.A.C.


Av. Malachowsky Nro. 218 Urb. Torres de Limatambo Lima - Lima - San Borja

Unit Condition Lab Conditions


o
As Found: In Tolerance Temperature: 21 C
As Left: In Tolerance Humidity: 50 %

Approval Information
Technician: Mirian Choccña Calibration Date: 17-Jul-21
Approval: Calibration Time: 10:43
Due Date: 17-Jul-22